WaveMark HF3000 RFID Cabinet User Manual Title here

WaveMark, Inc. RFID Cabinet Title here

User manual_07-0002 HF3000-Series Install Manual.v.5.pdf

Download: WaveMark HF3000 RFID Cabinet User Manual Title here
Mirror Download [FCC.gov]WaveMark HF3000 RFID Cabinet User Manual Title here
Document ID4180751
Application IDiNkCNYg53n/nFya0vYzUBw==
Document DescriptionUser manual_07-0002 HF3000-Series Install Manual.v.5.pdf
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize88.04kB (1100474 bits)
Date Submitted2019-02-26 00:00:00
Date Available2019-02-26 00:00:00
Creation Date2019-02-22 16:29:10
Producing SoftwareMicrosoft® Word 2013
Document Lastmod2019-02-22 16:29:10
Document TitleTitle here
Document CreatorMicrosoft® Word 2013
Document Author: Administrator

Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 1 of 14
HF3000-Series Installation Manual
Contents
1 Purpose .............................................................................................................................................................................. 2
2 Scope .................................................................................................................................................................................. 2
3 Reference Documents ........................................................................................................................................................ 2
4 Product Specification: ........................................................................................................................................................ 2
5 Responsibility ..................................................................................................................................................................... 3
6 Work Safety ........................................................................................................................................................................ 3
7 Unpacking and Transport ................................................................................................................................................... 4
7.1 Unpacking the HF3000 .................................................................................................................................................. 4
7.2 Cabinet and Shelf Transportation ................................................................................................................................. 5
8 Cabinet and Shelf Installation Process ............................................................................................................................... 7
8.1 Cabinet Installation ....................................................................................................................................................... 7
8.2 Shelf installation............................................................................................................................................................ 8
8.3 Shelf Removal................................................................................................................................................................ 9
9 Cabinet Wall Securing Process ......................................................................................................................................... 10
9.1 Directions for installing with Screw Inserts................................................................................................................. 10
9.2 Directions for installing with Expanding Inserts. ........................................................................................................ 10
10 Cabinet Servicing Process ................................................................................................................................................ 12
11 Cabinet Cleaning Process ................................................................................................................................................. 12
11.1 Cleaning with a detergent ......................................................................................................................................... 12
11.2 Cleaning with wipes .................................................................................................................................................. 12
12 Appendix .......................................................................................................................................................................... 13
Appendix 1: Notice of FCC Compliance .......................................................................................................................... 13
Appendix 2: Notice of Industrie Canada Compliance ..................................................................................................... 13
13 Revision History................................................................................................................................................................ 14
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 2 of 14
1 PURPOSE
This procedure describes the specifications, installation steps, and regulatory information for
the HF3000-series SmartCabinet.
2 SCOPE
This procedure applies to installation of HF3000-series SmartCabinets
3 REFERENCE DOCUMENTS
None
4 PRODUCT SPECIFICATION:
Power Brick
100-240 VAC, 1.5A Max, 50-60 Hz, power supply is rated to IEC/EN/UL-62368-1
and IEC/EN/UL-60601-1 (medical grade).
Cabinet Power
12.5V DC, 7.2A
AC Cable
The cabinet is supplied with a power cable.
Fuse
None
Environment
Operating environment between 50 and 90 degrees F (10C – 32C), noncondensing; must be installed in an environment free of corrosive agents and
excessive dust
Size
78 inches high by 36 inches wide by 27 inches deep
Weight
Weight in lbs.
Base Cabinet with 2 shelves
Cabinet with 3 shelves
Cabinet with 4 shelves
Cabinet with 5 shelves
Cabinet with 6 shelves
Cabinet with 7 shelves
Cabinet with 8 shelves
Cabinet with 9 shelves
Cabinet with 10 shelves
Cabinet with 11 shelves
Without Doors
160
175
190
205
220
235
250
265
280
295
Weight Capacity
The weight capacity of each shelf if 25 lbs.
Network
Auto-select 10/100Mbps Ethernet
With Doors
220
235
250
265
280
295
310
325
340
355
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 3 of 14
5 RESPONSIBILITY
Activity
Shipping
Receiving
Unpacking
Moving Cabinet and Shelves from Receiving to Department
Training on shelf installation and shelf movement
Shelf Installation
Placing Inventory on shelves
Connecting cabinet to power and Ethernet
Installing wall bracket
Responsible Party
WaveMark
Customer
Customer
Customer
WaveMark
Customer
Customer
Customer
Customer
6 WORK SAFETY
It is our highest priority that all field installations comply with the company’s installation and safety
procedures and any safety procedures in place at the location of installation.
In order to assure that equipment is installed in a safe manner and in compliance
with installation guidelines only trained personnel are authorized to activate or
service cabinets. This product contains no user-serviceable components.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 4 of 14
7 UNPACKING AND TRANSPORT
NOTE: Two people are required to move the cabinet from the shipping container to its final location.
Cabinet contains sensitive electronic components. Take care in moving the cabinet from the
shipping container.
7.1 Unpacking the HF3000
Step 1
Prior to unpacking the cabinets, inspect all packing for shipping damage. Take pictures of
any observed damage and immediately report to WaveMark Customer Support toll free at
877.492.8362 or email support@wavemark.com. Include all relevant pictures, a brief
description of damage, as well as cabinet serial number.
Note: Side marked “Ramp” must be facing out with 10ft (3 meters) of clearance to
accommodate ramp and cabinet unloading.
Step 2
Cut and remove strapping. Remove outer layers of packaging and discard. Be careful not
to damage the cabinet when cutting the packaging.
Step 3
Lower ramp:
1. Remove packing foam from ramp.
2. Lift the wooden ramp out of the groove and position ramp by inserting pin on the
ramp into the hole on the pallet. The edge of the ramp should sit flat against the
crate edge.
Step 4
With one person holding each side of the cabinet, guide cabinet down the ramp to the floor.
Step 5
Inspect the cabinet for damage. Take pictures of observed damage and immediately
report to WaveMark Customer Support toll free at 877.492.8362 or email
support@wavemark.com. Include all relevant pictures, a brief description of damage, as
well as cabinet serial number.
Don’t proceed with the installation of any damaged cabinet.
Step 6
If required, clean the cabinet exterior following procedure in Section 11.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 5 of 14
7.2 Cabinet and Shelf Transportation
Step 1
Identify cabinet location and locate nearest electrical outlet and network jack.
Confirm that the cabinet location is free from airborne corrosive agents, excessive dust or
dripping water. The floor should be level and dry
Step 2
Before moving a cabinet to its location, walk the route to identify obstructions such as signs,
door frames, or flooring issues such as elevator thresholds that might be a problem for the
cabinet wheels.
The cabinet is designed to be rolled over smooth flooring; the cabinet should never
be tilted, placed on its side, or rolled across parking lots.
Step 3
Except for the top and bottom shelves, cabinet shelves will be shipped separately from the
cabinet. Remove shelves from their shipping containers and carefully place on the bottom
shelf for transport.
Shelves should first be placed vertically and then tilted gently to create points of contact with
both side walls. This reduces the risk of the shelves being damaged during transport inside
the hospital facility.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 6 of 14
While not necessary, packing foam and bags can be used for added safety during
transport.
To avoid injury and/or damage to equipment during cabinet transport,
follow these safety precautions:
 Do not place more than 10 shelves inside a cabinet.
Step 4

Do not use the cabinet to transport any other materials (e.g. RFID tags, point of use
stations, box readers, etc.) as it may damage the cabinet or materials and will void
warranty.

Do not transport shelves in the following orientations. This will increase the risk of injury
or damage to the shelves.
Move the cabinet with shelves to its final location by pushing/pulling only the lower half of
the cabinet. Move the cabinet safely to avoid objects and people who may be traveling in the
same corridors.
The cabinet is designed to be very stable but moving too quickly and pushing against the top
of the cabinet may cause it to tip.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 7 of 14
When entering and exiting elevators, the cabinet’s wheels may lodge in the
elevator door gap. If this happens, gently move the cabinet from side to side and
pull/push to dislodge it. DO NOT USE EXCESSIVE FORCE!
CABINET AND SHELF INSTALLATION PROCESS
8.1 Cabinet Installation
Step 1
Locate power and network cables shipped with the cabinet and remove all packaging
and safety covers.
Step 2
The cabinet network and power connections are located in the base of the cabinet back
panel near the floor.
Step 3
Connect the DC power cable (attached to the power brick) to the cabinet and twist the
connector until secure. This requires several full turn twists until completely tightened.
Connect the electrical AC cable to the power brick.
Step 4
Connect network cable to the back of the cabinet next to the power connector.
Step 5
Tuck power brick into the cavity as shown below:
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 8 of 14
Step 6
Connect network cable to network port on wall, and connect AC power cord to an
electrical outlet on wall. The wall outlet must be easily accessible.
NOTE: If the network cable is not long enough, obtain a longer straight-thru network
cable (a crossover cable will not work). In the final installation the network cable should
not be under physical stress and should not be resting against electrical equipment such
as motors or heating units.
When successfully connected, the cabinet shelf illumination lights, shelf indicator LEDs,
and eBox LEDs will turn on in about one minute. The cabinet is now ready for
configuration.
If this does not occur and power/network connections and sources are confirmed to be
correct and live, contact WaveMark Customer Support via phone at 877.492.8362 or via
email at support@wavemark.com
Step 7
Place cabinet into its final position.
Step 8
Secure cabinet to the wall with the included mounting hardware. See Section 9 for more
information.
8.2 Shelf installation
NOTE: Only trained technicians should install or adjust shelf location.
The bottom and top shelves must remain in place and should not be moved.
Step 1
Stabilize cabinet against a wall or equivalent.
Note: If this is the initial installation, remove shelves transported in cabinet and set aside.
Step 2
Use the allen wrench provided with the cabinet to remove the end corner caps by removing
the four screws from each corner cap. Set corner caps aside in a manner that will prevent
them from falling over.
Step 3
Insert Shelf:
1. Orient shelf so the red receptacle is facing out towards front of cabinet and on the right
to be closest to the red network cables.
2. Place shelf in the desired location ensuring that it is aligned in the same numbered
track on each side of the cabinet. Shelf center-to-center spacing should not be less
than 6.5”.
3. Push shelf toward the back of the cabinet until secure.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 9 of 14
Step 4
Insert nearest red cable into the red receptacle on the front-right of each shelf. Repeat
steps 3 and 4 for additional shelves.
Step 5
Secure all unused network connectors and excess cables are secured in the cable
chase. Be sure that cables are not twisted or crossed.
Step 6
Replace end covers and secure with the screws. The cabinet is now ready for final
configuration. If cabinet was previously configured the cabinet will resume normal
inventory cycle counting.
8.3 Shelf Removal
Step 1
Stabilize cabinet against a wall or equivalent.
Step 2
Use the allen wrench provided with the cabinet to remove the end corner caps by removing
the four screws from each corner cap. Set corner caps aside in a manner that will prevent
them from falling over.
Step 3
Unplug red cable from shelf to be removed.
Step 3
Slide shelf out and set aside.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 10 of 14
8 CABINET WALL SECURING PROCESS
Safety Warning: The HF3000-series must be secured to a wall to provide
stability and prevent personal injury from tipping. Hospital personnel must
secure cabinet using appropriate hardware.
9.1 Directions for installing with Screw Inserts.
Step 1
Place a mark on the wall 79” from the floor at the point that will be the center of the cabinet.
Step 2
Place bottom edge of bracket on the mark.
Step 3
Place a mark in the center of the 2 holes.
Step 4
Drill approximately 1 ½” deep using a ¼” drill where you placed the marks indicating the center
of the holes.
Prior to drilling the hole, ensure that the wall can accommodate the bracket. Confirm
with the appropriate authorities that the target section of the wall may be drilled,
penetrated, and can support hardware without adverse impact on safety, health or
operating environment.
Step 5
Remove screws from plastic inserts and tap the plastic inserts into drilled hole until the flanges
are flush with the wall.
Step 6
Place the bracket in place and insert the screws. Tighten the screws only tight enough to hold
the bracket in place until the cabinet is in place.
Step 7
Position cabinet so it interfaces with bracket and tighten screws.
9.2 Directions for installing with Expanding Inserts.
Step 1
Place a mark on the wall 79” from the floor at the point that will be the center of the cabinet.
Step 2
Place bottom edge of bracket on the mark.
Step 3
Place a mark in the center of the 2 holes.
Step 4
Drill approximately 1 ½” deep using a ¼” drill where you placed the marks indicating the
center of the holes.
Prior to drilling the hole, ensure that the wall can accommodate the bracket.
Confirm with the appropriate authorities that the target section of the wall may be
drilled, penetrated, and can support hardware without adverse impact on safety,
health or operating environment.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 11 of 14
Step 5
Remove screws from plastic inserts and tap the plastic inserts into drilled hole until the
flanges are flush with the wall.
Step 6
Place the expanding inserts in the holes. Tighten the screws until the inserts are snug.
Remove screws from the inserts.
Step 7
Hold the bracket in position and insert screws back into the inserts. Tighten the screws
enough to hold the bracket in place until the cabinet is in position.
Step 8
Position cabinet so it interfaces with bracket and tighten screws.
Illustrations for mounting bracket and securing cabinet:
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 12 of 14
9 CABINET SERVICING PROCESS
The cabinet must be serviced only by trained service personnel.
10 CABINET CLEANING PROCESS
11.1 Cleaning with a detergent
Mix a mild detergent solution such as 1:100 dilution of Alconox, Liquinox, or other mild
detergent. Isopropyl Alcohol is an acceptable alternative. Quaternary Ammonium (Quat) is
also an acceptable alternative. Do not use Sodium Hypochlorite or other bleach solutions as
they could oxidize stainless steel and other electronic parts.
11.2 Cleaning with wipes
Clean all surfaces by spraying liberally with the cleaning solution and wiping down with
general purpose wipes. Wipe dry with general purpose wipes. Dispose of wipes in
accordance with local and hospital regulations.
To clean under the cabinet, carefully move the cabinet (disconnect wall-mount
hardware, unplug the cabinet, then roll cabinet away from its location).
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 13 of 14
11 APPENDIX
Appendix 1: Notice of FCC Compliance
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates, uses,
and can radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the user will be required to correct the interference
at his/her own expense.
Changes or modifications not expressly approved by Cardinal Health could void the user’s authority to operate
the equipment.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1)
this device may not cause harmful interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
This product must be professionally activated by Cardinal Health/ WaveMark trained personnel and must only
be serviced by Cardinal Health/ WaveMark trained service personnel.
Appendix 2: Notice of Industrie Canada Compliance
Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this
device must accept any interference, including interference that may cause undesired operation of the device.
This device has been designed to operate with the antennas listed below and having a maximum gain of 0
dB. Antennas not included in this list are strictly prohibited for use with this device. Acceptable antenna is
Cardinal Health part number 04-0175-R.
To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that
the equivalent isotropically radiated power (e.i.r.p.) is not more than that permitted for successful
communication.
Son fonctionnement est soumis aux deux conditions suivantes: (1) cet appareil ne doit pas causer
d’interférences et (2) cet appareil doit accepter toute 13ause1313rence, y compris les 13ause1313rences
pouvant provoquer un fonctionnement indésirable de l’appareil..
Cet appareil a été conçu pour fonctionner avec les antennes énumérées ci-dessous et ayant un gain
maximal de 0 dB. Les antennes ne figurant pas dans cette liste sont strictement interdites d’utilisation avec
cet appareil. L’antenne acceptable est la pièce numéro 04-0175-R de Cardinal Health.
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.
Doc. No. Revision
HF3000-Series Installation Manual
07-0002 Rev.3
Page 14 of 14
Pour réduire le 14ause14 de brouillage radioélectrique 14ause aux autres utilisateurs, le type d’antenne et
son gain doivent être choisis de manière à ce que la puissance rayonnée isotrope équivalente (par
exemple) ne soit pas supérieure à celle autorisée pour une communication réussie.
12 REVISION HISTORY
Rev.
Date
12.12.2018
1.30.2019
2.20.2019
Reviewer
Description of the Revision
Lars
New Document
Rohrberg
Rich
Updates, change-bars activated
Leitermann
Bob Quinn Updates
Change No.
NA
NA
N/A
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and ESSENTIAL TO CARE are trademarks or registered
trademarks of Cardinal Health. All other marks are the property of their respective owners.

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 14
Language                        : en-US
Tagged PDF                      : Yes
Title                           : Title here
Author                          : Administrator
Creator                         : Microsoft® Word 2013
Create Date                     : 2019:02:22 16:29:10-05:00
Modify Date                     : 2019:02:22 16:29:10-05:00
Producer                        : Microsoft® Word 2013
EXIF Metadata provided by EXIF.tools
FCC ID Filing: UQYHF3000

Navigation menu