Welch Allyn 3021-100 Propaq G Series Wireless Option User Manual cs ec ud bk

Welch Allyn, Inc. Propaq G Series Wireless Option cs ec ud bk

Contents

User Manual Addendum for Radio

EXHIBIT L – User Manual Addendum for Radio  FCC ID#  PGU3021-100
UPDATEPropaq CS® Directions For UseUpdate for Software Version 3.4XKeep this update in the back pocket of your manual.
 Copyright © 2001 by Welch Allyn Protocol, Inc.  Welch Allyn ®  is a registered trademark of Welch Allyn, Inc. Protocol ® , Propaq ® , Propaq ®  CS, Acuity ® , and Flexible Monitoring ®  are registered trademarks and FlexNet ™  is a trademark of Welch Allyn Protocol, Inc. Welch Allyn Protocol, Inc. is protected under various patents and patents pending. Copyright Notice:  Software in this Product is Copyright© 2001 by Welch Allyn Protocol, Inc., or its vendors.  All rights are reserved.  The  software is protected by United States of America copyright laws and international treaty provisions applicable all over the world.  Under such laws, you are licensed to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded, but the software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form.  This is not a sale of the software or any copy of the software; all right, title and ownership of the software remains with Welch Allyn Protocol or its vendors.  Welch Allyn Protocol will make available specifications necessary for interoperability of this software on request; however, users should be aware that use of Welch Allyn Protocol hardware and software with devices or software not sold by Welch Allyn Protocol or its authorized dealers and affiliates may lead to erroneous results and consequent danger in patient care, and may also void Welch Allyn Protocol's warranty.This device complies with Part 15 of the FCC rules and the Canadian ICES-003. Operation is subject to the following conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including any interference that may cause undesired operation. Disclaimers:   Welch Allyn Protocol, Inc. cautions the reader of this manual:• This manual may be wholly or partially subject to change without notice.• All rights are reserved. No one is permitted to reproduce or duplicate, in any form, the whole or part of this manual without permission from Welch Allyn Protocol, Inc.• Welch Allyn Protocol, Inc. will not be responsible for any injury to the user or other person(s) that may result from accidents during operation of the Propaq monitor.• Welch Allyn Protocol, Inc. assumes no responsibility for usage not in accordance with this manual that results in illegal or improper use of the Propaq monitor.• No implied license:  Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device. For information concerning this document or any Welch Allyn product, contact:Welch Allyn Protocol, Inc. Customer Service8500 SW Creekside PlaceBeaverton, Oregon  97008-7107  USA  Within USA, toll free:  Phone: (800) 289-2500Phone Technical Services: (800) 289-2501 WorldWide:  Phone:  (503) 526-8500Fax: (503) 526-4200Fax Technical Services:  (503) 526-4910 Internet:   http://www.protocol.comEmail Technical Services:  solutions@protocol.comEmail Marketing Dept.:  marketing@protocol.com Protocol Medical Systems, Ltd.: NW Europe Derby Service Centre, St. Georges House, Vernon GateDerby DE1 1UQ, United KingdomPhone: 44 1332 206208 Fax: 44 1332 206209Email: uk@protocol.com Welch Allyn U.K. Ltd. Cublington Road, Aston AbbottsBuckinhamshire HP22 4ND, EnglandPhone: 44-1296-682140 Fax: 44-1296-682104 Welch Allyn Italia Srl Via Napo Torriani, 28, 20124 Milano, ItalyPhone: 39-02-6699-291 Fax: 39-02-6671-3599 Welch Allyn GmbH Germany Postfach 31, Zollerstrasse 2-472471 Jungingen, GermanyPhone: 49-7477-92-710 Fax: 49-7477-92-7190 Welch Allyn France 814 Rue Charles de Gaulle77100 Mareuil les Meaux, FrancePhone: 01-6009-3366 Fax: 01-6009-6797 Welch Allyn – South Asia and Pacific P.O. Box 39-293 Howick, Auckland, New ZealandPhone: 64-9-532-9524 Fax: 64-9-532-9526 Welch Allyn Hong Kong Room 601, Kornhill Metro Tower1 Kornhill Road, Quarry Bay, H.K.Phone: 852-2886-8980 Fax: 852-2886-8360 Welch Allyn Protocol, Inc. – Latin America MD Intl., 7324 SW 48th Street, Miami, FL 33155  USAPhone: (305) 669-9003 Fax: (305) 669-8951 Update Part No: 810-xxxx-00Rev. 0.12 09/01Printed in USA0123
 Update     Propaq CS DFU 3.4X 3 About this Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 General Warnings and Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Understanding the Propaq CS, Acuity, and the FlexNet Network  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 The Acuity Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8 Hardwired Connection to Acuity Connect the Monitor to the Acuity System . . . . . . . . . . . . . . . . . . . . . . . . . . .9Disconnect the Hardwired Monitor from the Acuity System . . . . . . . . . . . . .10Switch from Hardwired to Wireless Monitoring . . . . . . . . . . . . . . . . . . . . . . .11 Wireless Connection to Acuity  . . . . . . . . . . . . . . . . . . . .12 Connect a New Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12Monitor a Patient Out of Range of Acuity . . . . . . . . . . . . . . . . . . . . . . . . . . .14Stop Monitoring a Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15Reconnect a Recently Monitored Patient . . . . . . . . . . . . . . . . . . . . . . . . . . .16Reassign a Monitored Patient to a New Room in the Same Unit . . . . . . . . .17Transfer a Monitored Patient to a New Room in a Different Unit . . . . . . . . .18Reassign the Wireless Propaq CS to a New Patient  . . . . . . . . . . . . . . . . . .19Switch from Wireless to Hardwired Monitoring . . . . . . . . . . . . . . . . . . . . . . .20 Equipment Alert Messages  . . . . . . . . . . . . . . . . . . . . . . .21Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22 Contents
 4 Welch Allyn  About this Update This Update describes the operation of the Propaq ®  CS monitor (software version 3.40) with the Acuity ®  FlexNet™ network. It describes both the Acuity hardwired and wireless communication options for the Propaq CS monitor.Use the Acuity System in compliance with the instructions in this Update, the  Propaq CS Directions for Use , the  Acuity System Reference Guide , and accepted hospital and clinical protocols. General Warnings and CautionsWarning  When connecting the Acuity hardwire connection, connect the Propaq CS monitor to an Acuity system only. Connecting to other networks could damage the monitor or injure the patient. If in doubt about the network jacks or devices, consult your facility’s Biomedical Engineering Department.Make sure the Acuity network cable is not damaged. If the wireless connection is not an option or not available, the Acuity network cable is the sole link between the Propaq CS monitor and the Acuity Central Station.When considering a treatment protocol that involves wireless communication of patient data, be sure to recognize some limitations inherent in wireless communications. When the Propaq CS monitor is not connected to the network by wireless (or hardwired) connection:- There are no patient alarms or alerts at the Acuity Central Station.- Acuity does not perform arrhythmia and ST analysis on the patient data and does not generate related alarms.If wireless communication is no longer available or is not functioning properly, consider using the hardwire connection to Acuity.If you don’t set alarm limits, the Acuity system uses preset settings (for arrhythmia test limits), and the powerup default settings for the Propaq CS monitor.The radio in this monitor has been authorized by the FCC for mobile use only. Mobile use as defined by the FCC is for operation 20 cm or more away from a person’s head or torso. The distance does not apply to transient exposure due to incidental passage closer than the maximum permissible exposure (MPE) limit.
 Update     Propaq CS DFU 3.4X 5Caution  Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC standards (e.g., EN 60950 for data processing equipment and EN 60601-1 for medical equipment). Furthermore, all configurations shall comply with the system standard IEC 601-1-1. Anyone connecting additional equipment to the signal input or output connectors is configuring a medical system, and is therefore responsible that the system complies with the requirements of the system standard IEC 601-1-1. If in doubt, consult your Biomedical Engineering Department.Note   WHEN USED IN CANADA:To prevent radio interference to the licensed service, this device is intended to be operated indoors and away from windows to provide maximum shielding.  Equipment (or its transmit antenna) that is installed outdoors is subject to licensing.Pour empêcher que cet appareil cause du brouillage au service faisant l’objet d’une licence, il doit être utilisé à l’intérieur et devrait être placé loin des fenêtres afin de fournir un écran de blindage maximal. Si le matériel (ou son antenne d’émission) est installé à l’extérieur, il doit faire l’objet d’une licence.
 6 Welch Allyn  Understanding the Propaq CS, Acuity, and the FlexNet ™  Network The Propaq CS monitor can communicate with an Acuity ®  Central Station as part of Welch Allyn’s FlexNet network. FlexNet allows multiple devices to communicate through hardwired Ethernet networks and Wireless Local Area Networks (WLANs). The Acuity Central Station provides central patient monitoring for monitoring devices connected to the network.As shown below, the Propaq CS can communicate through a hardwire Acuity connection. In addition, the Propaq CS can also be equipped with the Wireless option which allows two-way wireless communication with an Acuity Central Station through an access point in the FlexNet network. The access point is a digital radio transceiver that connects to the FlexNet network.To Other Access PointsAlternative hardwire connectionFlexNet NetworkPropaq CSwith hardwire Acuity connectionAccess PointAcuity Central StationTo Other Acuity SystemsOther wireless devicesPrimary hardwire connectionWireless Propaq CS(hardwire connection available as alternative)
 Update     Propaq CS DFU 3.4X 7 While connected to the network, the Propaq CS sends patient data to Acuity. Acuity continuously analyzes the data and provides appropriate alarm or alert messages at the Central Station and other network devices such as a hallway message panel or the Propaq CS itself. Acuity also stores the patient data for viewing or report printing.If the Wireless Propaq CS is moved out of range or loses communication with the FlexNet network and Acuity, it continues to monitor the patient and display patient data. While not communicating with Acuity, the Propaq CS continues to generate local patient alarms or alert messages. Acuity does not perform waveform analysis or generate arrhythmia messages while the Propaq CS is not communicating with Acuity. When the Wireless Propaq CS is returned to within range of the FlexNet network, it automatically reconnects to Acuity.A Propaq CS with the Acuity hardwire connection can be identified by the Acuity connector on the right side panel.A Wireless Propaq CS can be identified by the blue antenna cap on the top right corner of the monitor in front of the carrying handle. Every Wireless Propaq CS also has the Acuity hardwire connection.If a Wireless Propaq CS monitor is communicating with the network via the wireless interface and then the Acuity hardwire cable is used to connect the bedside Acuity network jack and the monitor, the monitor disconnects from the wireless connection and communicates through the hardwire connection. There may be a short interruption in the display of patient waveforms at Acuity while this change occurs.DEFIB SYNCHROMONITOREKG x 1000!3A2AG!!!12-28V, 3ARight Side PanelAcuity hardwire connectorWireless Propaq CSYellow antenna cap indicates Wireless Propaq CS
 8 Welch Allyn  The Acuity Menu The Acuity Menu on the Propaq CS allows you to control some of the interaction with the Acuity network. The Acuity Menu is only accessible while the monitor is connected to the Acuity network.The Acuity Menu for the Wireless Propaq CS provides additional functions that allow you to manage the patient assignment and patient location from the monitor. These functions are described later in this document.ACUITY MENU (HARDWIRED CONNECTION)NETOFF SNAPSHOT PREVIOUSMENUPress to disconnect from the Acuity network.Press to print a waveform displayed on the Propaq CS monitor screen at the Acuity printer.NEWROOM TRANSFER NEWPATIENT PATIENTINFO PREVIOUSMENUCHANGE PATIENT MENUACUITY MENU (WIRELESS CONNECTION)ENDTELE SNAPSHOT PREVIOUSMENUPress to disconnect from the Acuity network.Press to print a waveform displayed on the Propaq CS monitor screen at the Acuity printer.CHANGEPATIENTPress to reassign a patient to a new room in the same unit.Press to transfer a patient to a new room in a new unit.Press to assign the monitor to a new patient.Press to display patient information such as ID, name, unit and room.Press to access the Change Patient Menu with additional functions.
 Update     Propaq CS DFU 3.4X 9 Hardwired Connection to Acuity Connect the Monitor to the Acuity System 1. If the Propaq CS monitor has already been connected to the patient, save the patient’s Trends and Alarm Limit settings by keeping the monitor turned on. The monitor transmits up to five hours of trend information when you connect it to the Acuity network. If the monitor has not been connected to the patient, clear any prior patient’s trends and alarm limit settings by turning off the monitor, then turning it on after a few seconds. 2. If the monitor is not already connected to the patient, attach leads and sensors to the patient as described in the  Propaq CS Directions For Use . 3. Plug in the Acuity network cable to the Acuity network jack on the monitor side panel as shown. Plug in the other end of the cable to the bedside Acuity network jack. 4. Connect the ac power adapter to the monitor and the wall outlet to charge the battery. Check to see that the green battery charging light on the monitor’s right side panel is on. 5. When the monitor completes the connection to Acuity, it displays the message  ON NETWORK  (or alternating messages  ON NETWORK  and  SERIAL  if the monitor has wireless capability). Confirm the patient identification at the bedside or enter the patient information at the Acuity Central Station using the Patient ID Setup Window. 6. If alarm limits have not been set, do so at the monitor or at the Acuity Central Station using the Alarms Setup Window.DEFIB SYNCHROMONITOREKG x 1000!3A2AG!!!12-28V, 3AAcuity ConnectorBattery charging lightPower input connector
 10 Welch Allyn Note   When a Propaq CS monitor in Adult or Pediatric Mode is connected to an Acuity System, the audible alarms at the bedside Propaq CS monitor can be delayed up to 4 minutes and 15 seconds. The delay time is selected in Acuity software at the time of Acuity installation. Visual alarm indications are not delayed. Disconnect the Hardwired Monitor from the Acuity System 1. To permanently disconnect the Propaq CS monitor from the Acuity network, press the  NET OFF  key on the upper left corner of the monitor display. (From the Main Menu you can also press  SETUP,   ACUITY ,  NET OFF .) 2. Within 15 seconds, disconnect the Acuity network cable from the Propaq CS monitor side panel and the bedside jack. If the patient will no longer be monitored with this monitor, turn off the monitor to erase trend information. If you want to temporarily disconnect the Propaq CS monitor from the Acuity network and reconnect the same patient to the Acuity network later, see the  Acuity System Reference Guide .Note   If the Propaq CS has the wireless option, be aware that after you disconnect the hardwire connection and disconnect the Acuity network cable, the Wireless Propaq CS will attempt to establish a wireless connection to Acuity as long as the power is on. Switch from Hardwired to Wireless MonitoringADULTMON17:05:10 PRINT FAULT II P(P2NETOFF ON NETWORKmVcm1HBPALLEN KIMPress to disconnect from Acuity
 Update     Propaq CS DFU 3.4X 11 Follow these steps to switch a Wireless Propaq CS monitor from a hardwired Acuity connection to a wireless Acuity connection. 1. Press the  NET OFF  key on the upper left corner of the monitor display to disconnect from Acuity. (From the Main Menu you can also press  SETUP,   ACUITY ,  END TELE .) 2. Within 15 seconds unplug the Acuity network cable from the Wireless Propaq CS Acuity connector and the Acuity bedside network jack.The Wireless Propaq CS automatically seeks to establish a wireless connection to the Acuity network. 3. After the network connection is established, select the Acuity Unit, patient, and room from the Wireless Propaq CS as described on page 16.
 12 Welch Allyn  Wireless Connection to Acuity Connect a New Patient 1. Make sure the Acuity network cable is not plugged into the Acuity connector and an Acuity bedside network jack. Turn on power to the Wireless Propaq CS. The monitor displays the powerup screen for about 10 seconds, then displays the Main Menu. The monitor is in the powerup patient mode with the associated settings. 2. Confirm that the Wireless Propaq CS displays the flashing message  SEARCHING  in the upper left corner of the display. Within a short time the monitor displays the flashing message  CONNECTING . This indicates the monitor has associated with an access point but is not fully connected to the Acuity network. 3. Confirm that within one minute the monitor displays these alternating messages: ON NETWORK  and  WIRELESS This indicates the Wireless Propaq CS is connected to the Acuity network. 4. After the network connection is established, the monitor may prompt you to select an Acuity Unit (if your facility has more than one Acuity unit). Scroll up or down to highlight the desired Acuity Unit, then press  Select . 5. The Wireless Propaq CS displays a list of possible patients. If your patient has been pre-admitted to the selected Acuity unit, they will be included in the list. Monitor is searching for a network connection.SEARCHING(Need artwork -- Propaq CS display . . .)
 Update     Propaq CS DFU 3.4X 13 • If your patient is NOT in the list, highlight  Select Patient at Acuity  and press  Select . The patient name will need to be entered later at the Acuity Central Station.Warning  If you do not select the patient name at the Wireless Propaq CS at this time, do not adjust any alarm limits until  after  the patient name and ID are confirmed at Acuity. When the patient name and ID are confirmed at Acuity, Acuity downloads the default settings and patient alarm limits for that Acuity unit to the Wireless Propaq CS, thereby overriding any previous settings and alarm limits. • If your patient is in the list, scroll to highlight your patient’s name, then press  Select . 6. Within a few seconds the Wireless Propaq CS displays a list of unassigned rooms.• If you want to assign the patient to a room, highlight the room and press  Select .• If you do not want to assign a room at this time, highlight  Select Room at Acuity  and press  Select . The patient room will need to be entered later at the Wireless Propaq CS (see page 17) or at Acuity (see the  Acuity System Reference Guide  or  Acuity Directions For Use ). 7. Customize alarm limits for your patient if needed. Refer to the  Propaq CS Directions For Use .
 14 Welch Allyn  Monitor a Patient Out of Range of Acuity While out of range of Acuity, the Wireless Propaq CS continues to monitor the patient and provide local alarms or alerts at the Propaq CS as needed.When the patient with the Wireless Propaq CS goes out of range of Acuity, do the following: 1. A DROPOUT equipment alert occurs at the Acuity Central Station.Acknowledge the alert at Acuity. 2. An equipment alert occurs at the Wireless Propaq CS with this message: NETWORK FAULTCHECK ACUITY/NETWORK CONNECTIONPress a key on the Propaq CS to acknowledge the alert and silence the alert tones (if tones are enabled).When the patient with the Wireless Propaq CS returns within range of Acuity, the Propaq CS automatically reconnects to Acuity. No clinician intervention is required.
Update     Propaq CS DFU 3.4X 15Stop Monitoring a PatientIf you want to discontinue monitoring the patient, follow these steps.1. Press the NET OFF key on the upper left corner of the monitor display. (From the Main Menu you can also press SETUP, ACUITY, END TELE.)2. Disconnect the leads and sensors from the patient and turn off the power to the monitor.If you do not turn off power to the monitor within about 30 seconds, the Wireless Propaq CS will automatically try to reconnect to the network.If you do not use NET OFF or END TELE to disconnect from the network as described above, the Acuity Central Station generates a DROPOUT equipment alert at Acuity.If you want to monitor this same patient at a later time, you will need to re-select the patient name from Wireless Propaq CS or confirm the patient ID at Acuity.
16 Welch Allyn Reconnect a Recently Monitored Patient1. Turn on power to the Wireless Propaq CS and confirm that the monitor displays the powerup screen.2. The Wireless Propaq CS will then present a series of menus and messages requesting you to provide information about the connection and patient. The actual screens presented depend on how long the patient has been disconnected.Provide the information as requested. This may include:• Select an Acuity unit.• Select a patient from the patient list.• Select a patient room from the room list.If you do not select the patient name or room while connecting the patient, you will need to do that later at the Acuity Central Station. See the Acuity System Reference Guide or Acuity Directions For Use for more information.
17 Welch Allyn Reassign a Monitored Patient to a New Room in the Same UnitIf a patient is being monitored and you want to assign them to a new room in the same unit, follow these steps.1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, NEW ROOM.Within a few seconds the Wireless Propaq CS displays a list of all available rooms, including the patient’s current room.• If you decide not to change the patient’s current room assignment, press Select (the patient’s current room is the default selection in the list).• To assign the patient to a new room, scroll up or down to highlight the room and press Select.• If you want to cancel the patient’s current room assignment, but do not want to assign a new room at this time, you can highlight Select Room at Acuity and press Select. You can then assign the room later from the Acuity Central Station, or you can repeat this procedure and assign a new room from the Wireless Propaq CS.
18 Welch Allyn Transfer a Monitored Patient to a New Room in a Different UnitIf a patient is being monitored and you want to assign them to a new room in a different unit, follow these steps.1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, TRANSFER. Within a few seconds the Wireless Propaq CS displays a list of units.2. Scroll up or down to highlight the new unit, then press Select.The patient is not monitored at Acuity during the short time required by Acuity to process the transfer to the new unit (typically less than one minute). However, the patient continues to be monitored by the Wireless Propaq CS.(If the selected unit is currently not available, the Wireless Propaq CS displays an appropriate message; press a key to acknowledge the message and cancel the transfer.)3. After the patient is assigned to the new unit, the Wireless Propaq CS displays a list of unassigned rooms. (The patient’s previous unit and room assignment is cancelled.)• To assign the patient to a new room, highlight the room and press Select.• If you decide not to assign the patient to a new room at this time, you can highlight Select Room at Acuity and press Select. You can then assign the room later from the Acuity Central Station, or you can assign a new room from the Wireless Propaq CS later using the procedure on page 17.
19 Welch Allyn Reassign the Wireless Propaq CS to a New PatientIf you want to discontinue monitoring a patient and connect the Wireless Propaq CS to a new patient, follow these steps.1. From the Main Menu, press SETUP, ACUITY, CHANGE PATIENT, NEW PATIENT.The monitor then presents a series of menus and messages requesting you to provide information about the connection and patient. The actual screens presented depend on how the Acuity System is configured.Provide the information as requested. This may include:• Select an Acuity unit.• Select a patient from the patient list. (After you select a new patient, all monitor operating settings are reset to the Acuity System default powerup settings.)• Select a patient room from the room list.If you do not select the patient name or room while connecting the patient, you will need to do that later at the Acuity Central Station. See the Acuity System Reference Guide or Acuity Directions For Use for more information.Switch from Wireless to Hardwired Monitoring1. Plug the Acuity network cable into the Wireless Propaq CS Acuity connector and an Acuity bedside network jack.The Wireless Propaq CS will automatically switch to a hardwired Acuity network connection. No equipment alerts are generated.
20 Welch Allyn Equipment Alert MessagesThe following equipment alerts can occur when operating the Propaq CS monitor with the Acuity network.CHECK ACUITY/SERIAL CONNECTION. The Propaq CS detects a problem in the hardwired (serial) communication with Acuity. Check the Acuity network cable to be sure it is plugged into the side panel and the bedside Acuity jack. If the cable is damaged, replace the cable. If the cable appears undamaged and the Acuity system is operating normally, contact biomedical engineering for assistance.CHECK ACUITY/NETWORK CONNECTION. The Wireless Propaq CS detects a problem with the wireless communication with Acuity. The monitor might be out of range of the network, or there might be a problem involving the monitor radio card, the access point, or the Acuity system. If the problem persists, contact biomedical engineering for assistance.
21 Welch Allyn SpecificationsWireless Propaq CS Radio SpecificationsCharacteristic SpecificationFlexNet Network 2.4 GHz Wireless Local Area Network (WLAN) and 10/100 Base-T Ethernet networkFrequency11. When used within certain countries, authorization for use is restricted as follows:• France:  The equipment is internally restricted to the 2.448-2.482 GHz frequency range.• Spain:  The equipment is internally restricted to the 2.447-2.473 GHz frequency range.• Japan:  The equipment is internally restricted to the 2.473-2.495 GHz frequency range.• Italy:  Operation requires a user license.Note: The frequency ranges specified above are subject to geographic-specific regulatory authorities.2.402 to 2.480 GHzModulation GFSK, Frequency Hopping Spread Spectrum (FHSS)Output Power 112 mW (maximum)IEEE 802.11 compliant YesWireless Propaq CS monitors per Access Point 7 (max.)
22 Welch Allyn