Well Life Healthcare WR2605AOTCRX Wireless Electrical Stimulator User Manual

Well-Life Healthcare Limited. Wireless Electrical Stimulator

User manual

OPERATION MANUAL
WELL LIFE Wireless TENS/EMS Stimulator
Model no. WR-2605A (OTC)
WELL LIFE HEALTHCARE LTD.
1Fl, No.16, Lane 454, Jungjeng Rd., Yunghe District,
New Taipei City 23455, Taiwan
Operation Manual
Read Before Using
IM-WR2605A-USA
Date: 2016-4-19
0120
TABLE OF CONTENTS
INDICATIONS AND CONTRAINDICATIONS..………………………………… 1
WARNINGS AND PRECAUTIONS……………………………………………… 2
GENERAL INFORMATION ………………………....……………………………. 4
ABOUT RECHARGEABLE BATTERY ………………. ……..…………………. 6
Charging the Remote Using the AC Adaptor……………..……………………
Charging the Receiver…………………………………………………………………
Charging the System via USB Port……………………………
STEP BY STEP OPERATION GUIDE FOR TRAINING ……..………………… 8
Preparing Skin for Running a Session…………………………………………………….
Placement of Electrodes …………………………………………………………………..
Turning ON/OFF Remote Control…………………………………………………………..
Turning ON/OFF Wireless Unit…………………………….………………………………
Communication between Remote and Receiver(s) …………………………………….
Selecting the Program.……………………………………………………………………..
Selecting the Therapy Intensity Level… …………………………………………………
Selecting the Therapy Time……………………………………………………………….
CARE AND MAINTENANCE…………………………………………………........ 16
TROUBLESHOOTING.………..……………………………………………….… ... 17
TECHNICAL SPECIFICATIONS…………………………………………………… 18
FCC STATEMENT ..…………………………………………................................ 23
WARRANTY…………………………………………............................................. 24
1
INDICATIONS AND CONTRAINDICATIONS
Read the operation manual before using.
Read instruction manual before operation. Be sure to comply with all “CAUTIONS” and “WARNINGS” in the manual. Failure
to follow instructions can cause harm to user or device.
Observe your physician’s precise instructions and let them show you where to apply the electrodes. For successful therapy,
the correct application of the electrodes is an important factor.
Indications for use
The WR-2605A (OTC) is a dual channel digital electrical stimulator for active treatment application as the following intended
use:
Transcutaneous Electrical Nerve Stimulator (TENS) can be used for the following applications:
For temporary relief of pain associated with sore and aching muscles in yours shoulder, upper and lower back due to
strain from exercise, normal household chores and work related activities.
Electrical Muscle Stimulator (EMS ) stimulation can be used for the following applications:
To stimulate healthy muscles in order to facilitate muscle performance.
Relax muscle spasms
Prevent or retard disuse atrophy
Re-educate muscles
Maintain or increase the range of motion
WARNINGS AND PRECAUTIONS
Warnings
If you are under the care of a Physician, consult with your Physician before using this System.
The long-term effects of this System are not known.
Do not place the pads on or close to your heart.
Do not place the pads around or close to your neck. Do not apply stimulation over the neck. Severe spasm of the
muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing.
Stimulation over the neck could also have adverse effect on hearing or blood pressure.
Do not apply stimulation across the chest because the introduction of electrical current into the chest may cause
rhythm disturbances to the heart.
Do not place the pads on or around your head. The effects of stimulation of the brain are unknown.
Do not use the pads over or close to sores.
Do not place the pads on the front or sides of the neck across or through the heart (one pad on the front of the chest
and one on the back), in the genital region, or on the head, because of the risk of stimulating inappropriate muscles
and organs.
Do not place the pads over any recent scars, broken or inflamed areas of infection or susceptibility to acne, thrombosis
or other vascular problems (e.g. varicose veins), or any part of the body where feeling is limited.
Do not place the pads over areas of injury or restricted movement (e.g. fractures or sprains).
2
Do not use this System while sleeping.
Do not use if you feel numbness.
Do not use this System in or close to water.
Do not apply stimulation across the chest because the introduction of electrical current into the chest may cause
rhythm disturbances to the heart, which could be lethal.
Do not use the pads over or close to cancerous lesions.
Use the pads only on normal, healthy, clean and dry skin. Do not use the pads on open wounds or rashes, or over
swollen, red, infected or inflamed skin.
If you have ever had back surgery, consult your Physician before using this System.
Electronic monitoring equipment (such as ECG and ECG alarms) may not operate properly when electrical stimulation
is in use.
You must position the pads and operate the unit ONLY as indicated in this manual.
Avoid areas in injury or restricted movement (e.g. fractures or sprains)
Avoid placing the pads over metal implants.
Do not use in the bath or shower, or in an environment of elevated humidity (e.g. Sauna, hydrotherapy, etc).
Operation in close proximity, such as 3 feet (1 meter), to shortwave or microwave therapy equipment may produce
instability in the device output and may shut the device off.
Do not use the device in an environment where flammable or explosive fumes may exist.
Patient should never operate potentially dangerous machinery such as power saws, automobiles, etc. during electrical
stimulation. It might result in burs at the site of the stimulator electrodes and possible damage to the stimulator if you
connect to high frequency equipment.
Application of electrodes near the thorax may increase the risk of cardiac fibrillation.
The system is internally powered (power form battery pack), IP 22, continuous operation.
Wait before using the unit :
At least 6 weeks after the birth of your baby (you must consult your doctor before use).
One month after an IUD contraceptive device (e.g. coil) has been fitted (you must consult your doctor before use).
At least 3 months after having a caesarean section (you must consult your doctor before use).
The heavy days of your period have finished, because vigorous abdominal exercise is not recommended at this time.
Precautions
Read User Manual before using this System for the first time.
Keep this manual available whenever you use your System.
The System is intended for personal use on healthy adult muscle only.
The safety of using the System during pregnancy or birth has not been established.
The effectiveness of the System depends greatly on a person’s individual physical condition. It may not always be
effective for every user.
The safety of neuromuscular stimulation during pregnancy has not been established.
Use caution when/if:
- Sensory nerve damage is present by a loss of normal skin sensation.
- Use caution prior to using this device on patients suspected of having heart disease.
- Use caution for patients with suspected or diagnosed epilepsy when using this device.
- Use caution following recent surgical procedures when muscle contraction may disrupt the healing process.
- Use caution when there is a tendency to hemorrhage, such as following acute trauma or fracture.
- A menstruating or pregnant uterus.
- Patient experiences skin irritation due to electrical stimulation or the electrical conductive medium used, remove
the electrodes, discontinue stimulation, and consult the clinician. Irritation may be reduced by an alternative
conductive medium or an alternative electrode placement. Isolated cases of skin irritation may occur at the site of
electrode placement following long term application.
Place electrodes in accordance with illustrations in the User Manual.
This unit should not be used while driving, operating machinery or during any activity in which involuntary muscle
contractions may place the user at undue risk of injury.
Some users may experience skin irritation or hypersensitivity due to the electrical stimulation or the conductive
medium.
Keep this device out of the reach of children. If the patient is a child, make sure he/she is properly supervised during
electrical stimulation. Strangulation resulting from baby or child entanglement in monitoring cables.
3
Application of moderate heat (thermal wrap) to muscles as well as moistening skin prior to treatment improves
treatment efficacy; use of cold packs on treated muscles after treatment is likewise recommended.
This unit should only be used with the leads, electrodes and accessories provided by the manufacturer.
The device is not intended for medical use, for the treatment of any medical condition or for any permanent physical
changes.
Contact Well-Life, or an authorized dealer, if your unit is not working correctly. Do not use in the meantime.
An effective session should not cause discomfort.
For first time users, muscle stimulation can be an unusual sensation. We recommend that you begin in a seated
position with low stimulation intensity settings to familiarize yourself with the sensation before progressing to higher
intensity settings.
The leads and pads must not be connected to other objects.
Do not over exert yourself while using muscle stimulation. Any workout should be at a comfortable level for you.
Do not place pads over jewelry or body pierecings.
Use Caution and consult Physician before using System if any of the following conditions
apply to you:
You have any serious illness or injury not mentioned in this guide.
You have recently undergone a surgical procedure.
You take insulin for diabetes.
You use the unit as part of a rehabilitation program.
If you have suspected or diagnosed heart problem.
If you have suspected or diagnosed epilepsy.
If you have a tendency to bleed internally following an injury.
If you recently had surgery, or have ever had surgery on your back.
If areas of skin lack normal sensations, such as skin that tingles or is numb.
During menstruation or during pregnancy.
Some people may feel skin irritation or experience a very sensitive feeling in the skin due to electrical stimulation. If
this occurs, stop using your System and consult your Physician.
If skin under one of more pads feels irritated after using the stimulator for a long period of time, use the stimulator for a
shorter period of time.
Minor redness at stimulation placement is a normal skin reaction. It is not considered as skin irritation, and it will
normally disappear within 30 minutes after the electrodes are removed. If the redness does not disappear after 30
minutes from the removal of electrodes, do not use the stimulator again until after the excessive redness has
disappeared.
Turn off the stimulator if the stimulation feels unpleasant or does not provide pain relief.
Keep your System out of the reach of children.
Use your stimulator only with the pads, snap cables and accessories recommended by the manufacture.
Do not use this System when driving, operating machinery or when swimming.
Before removing the pads, be sure to power off device to avoid unpleasant stimulation.
After strenuous exercises or exertion:
Always use lower intensity to avoid muscle fatigue.
Important :
Effectiveness is highly dependent upon patient selection by a clinician qualified in the management of pain or
rehabilitation.
Do not use your unit at the same time as any other device, which transfers an electrical current into the body (e.g.
another muscle stimulator).
Cease using your unit if you are feeling light headed or faint. Consult doctor if this happens.
Do not touch the pads or metal studs while the unit is switched on.
Do not use unit if you are wearing a belly button ring. Remove ring before session.
Use the device with only the leads and electrodes provided for use by the Well-Life with your device. Any others many
not be compatible with your unit and could degrade the minimum safety levels. Use only the electrode placements and
stimulation settings prescribed by your practitioner.
This device is for external use only.
Choking resulting from a child swallowing a small part that has become detached from the device.
4
Note: If you are in any doubt about using device for any reason, please consult your doctor .
Pad/Electrode Precautions
To reposition the pads during a session, always pause the program currently running, reposition the pads and then
restart the program.
The pads are for single person use only.
Do not plunge the pads into water.
Do not apply solvents of any kind to the pads.
Always ensure the unit is OFF before removing the pads.
Apply the whole surface of the pads firmly to the skin. Do not use pads, which do not adhere properly to the skin.
If your skin is red under the pad after a session, do not start another session in the same area until your redness has
completed disappeared.
Adverse Reactions
You may experience skin irritation due to prolonged exposure to the stimulation electrodes applied to your skin.
You may experience potential allergic reactions to accessible materials used in the electrodes.
You may experience headache and other painful sensations during or following the application of electrical stimulation
near your eyes and to your head and face.
You should stop using the device and should consult with your physician if you experience adverse reactions from the
device.
Conditions that may affect your Well Life Wireless TENS/EMS System
Since the stimulator is a battery-operated electronic System, its output performance and safety may be affected greatly in
extreme humidity. Therefore, it is very important to keep the system device(s) dry to ensure the safety and performance of
the stimulator.
The patient may also be an intended operator, at least 15 years old and 8 years intensive reading experience( school),
no maximum.
This system is for home use, indoor, not intended for professinal use
This system is for daily use, no operation time limit on device but it is recommended not exceeding 60 minutes per day.
If there is any other problem, please cosult or return the device to your distributor. Do not try to repair a defective device.
For plug-in equipment , the power outlet socket must be located near the equipment and must be easily accessible
WARNING: No modification of this system is allowed.
WARNING: Other cables and accessories may negatively affect EMC performance .
WARNING: Do not stack and store theis system closed to other equipment.
5
GENERAL INFORMATION
Package Content
1. Remote x1
2. Receivers (Ch1 & Ch2) x2
3. Self Adhesive Electrodes
Size: 150x100mm, 2 pcs/bag
4. USB Charging Cables x 2
(Length 60 cm )
5. AC Adaptor x1
About Remore Control
1. Power ON key
2. CH key
3. SET key
4. Power OFF + key
5. ON + key : increase
6. OFF key : decrease
7. USB charger socket
About Recivers
8. CH1 Receiver : Power On/Off key
9. CH2 Receiver : Power On/Off key
10. Snaps for Electrode
11. USB charger Socket
6
About Rechargeable Batteries
Please fully charge the batteries in both the remote unit and the receivers before using this device.
Charging the Remote Using the AC Adaptor
1. Connect the small end of the USB cable to the remote and the larger end to the AC adaptor..
2. Plug the adaptor into any standard wall outlet.
When the remote has finished charging the battery an icon will indicate “full”.
Charging the Receiver
1. Connect the small end of the USB cable to the receiver and the other end into the AC adaptor.
2. Plug the adaptor into any standard wall outlet.
A Orange light indicates the receiver is now in charge mode, once fully charged the Orange light will change to
Green indicating that the receiver is fully charged.
7
Note : for important precautions for use with AC adaptor , please be informed :
To prevent the risk of electric shock, make sure power cord is unplugged from wall socket. To fully disengage the power
to the unit, please disconnect the power cord from the AC outlet. Do not remove cover (or back). No user serviceable
parts inside. Refer servicing to qualified service personnel. The AC outlet shall be readily available and accessible.
Note : for important precautions regarding the batteries please be informed:
It is recommended to use a manufacturer supplied or approved battery charger in order to maximize battery life.
Do not connect the battery to metal objects placed in your pocket or backpack or other containers.
Do not short the metal (+)(-) terminals. Wishing to put batteries or electrical devices do not end when (+)(-) device error.
Do not dismantle or modify the battery.
Please do not take a hammer or other items to hammer blow, or throw batteries and trampling, falls, and fall causing a
strong impact, heavy blow, etc.
Do not use sharp utensils or metal objects to scratch or puncture the battery.
Do not place the battery into a microwave, oven, or dryer. Do not place the battery into high-pressure or high
temperature environments.
Do not mix this product with other brands or other types of batteries
Before charging or discharging the battery, read the manual.
Stay away from conductive objects during battery charging and discharging.
Keep batteries away from children.
When the device(s) system is not in use, store in a low humidity, low temperature environment.
In the event the battery fluid comes into contact with your eyes, do not rub your eyes. Rinse with water and immediately
seek treatment.
If you detect abnormal heat, odors, or flames coming from the battery, discontinue using the battery.
If the battery fluid leaks and comes into contact with the skin, wash with clean water.
Do not throw into a fire.
When discarding or recycling batteries, make sure not to short-circuit (+)(-) the terminals.
Please recycle. Do not dispose of old batteries with your household waste; dispose of them safely at your recycling
centre or business where the batteries were purchased.
8
STEP BY STEP OPERATION GUIDE FOR TREATMENT
Preparing the Skin for Running a Session
Proper preparation of the skin covered by the electrodes allows more stimulation to reach targeted tissues, prolongs
electrode life, and reduces the risk of skin irritation. After connecting the wireless units to the electrode pads, use the
following steps to prepare your skin at the electrode placement sites:
1. Determine the placement sites for the electrodes.
2. Wash the area with mild soap and water (do not use alcohol). Rinse and dry thoroughly.
3. Trim excess body hair from the area with scissors (do not shave).
4. Optionally, apply skin prep to the area to form a protective barrier on your skin. Apply, let dry, and apply electrode as
directed. This will both reduce the chance of skin irritation and extend the life of your electrodes.
5. When removing electrodes, always remove by pulling in the direction of hair growth.
6. It may be helpful to apply skin lotion on electrode placement area when not wearing electrodes.
9
Placement of Electrode
1.Place the electrode with the contact plugs up onto a flat
surface. Place the wireless unit with the contact jacks down
onto the Electrode pad so that plugs and jacks are aligned.
Then press down until completely snapped in.
2. Do not remove the clear plastic shield from Electrode until
the unit is ready to be placed on the body.
10
Turning ON/OFF Remote
1. To turn on the Remote , press and hold the ON/+ button for one (1)
second to turn on the device.
All the treatment status displays clearly on the big LCD screen,
with backlight for 10 seconds, with each entry of button.
The most recently selected program will display on the screen.
2. To turn off the Remote, press and hold the OFF/- button for three (3)
seconds to turn off the device. Or it will be automatically off when
no button is pressed for 60 seconds.
Note: Do not turn the unit on until the receiver(s) are properly
attached.
Note : When stimulating the muscles of the arms or legs bear
in mind that the muscle contraction may cause involuntary limb
movement which could cause injury to you or others. Always
ensure the limb is secured to prevent movement.
Turning On/Off Receiver
1. Turn on the receiver, press the Power button.
2. Green backlight will be lighted up, and lasting on 60 seconds, then
it will light off if no other entry.
3. Green backlight of the receiving unit is then flashing whist the
identical wireless unit receives the signal from the remote .
4. To Turn off the receivers , press and hold the OFF/- button for three
(3) seconds.
Note: In an emergency you may take off the receiver(s) and
electrodes directly from treatment site(s).
Note : To prevent unpleasant electric shocks, never remove off the
device while it is still turned on.
Communication between Remote and Receiver(s)
11
Radio Frequency Wireless technology 2.4G is used in this system, for good quality of service between the remote and
wireless unit(s) requirement requires 3 meters (9 feet).
1. LCD symbol on the remote control and LED indicator guide on wireless unit(s):
Indicator Description
LCD symbol on Remote
1. It shows Mode, Timer, Intensity Level, and wireless unit by Channel
number on the LCD display when Power ON, LED backlight for 10
seconds.
2. If the paired wireless is connected, it will display :
or
3. If low battery symbol shows , battery power is low.
When charging the remote , the device will shut off automatically.
Receivers ( Ch1 / Ch2) :
1. When Power ON, Green LED light on.
2. LED in Green blinking on the Channel in use, when it receives signal
transmitting from the Remote Control.
3. LED in Red Flashing, battery power is low.
Receiver unit will turn on when charging the battery,
Orange light lasting until fully charged, then it turns off.
2. There is one (1) Remote with two (2) receivers (Ch1 or Ch2 printed on the front) supplied for use with this system.
Additional two (2) receivers (Ch3,Ch4) might be available for replacement purpose, please contact authorized
distributors.
3. Whenever the receivers turns on, the remote control unit shall display the identical number of receiver(s) .
Make sure the identical receiver is turned on and properly placed before setting up Program /Timer / Intensity Level.
4. The Remote will display treatment status on the LCD sequentially in turn to the Channel of receiver(s). Or,
you could turn off the remote when all set.
Selecting the Program
12
The Device offers 8 different preset treatment programs respectively for TENS and 6 programs for EMS modes; the
programs differ with respect to varying pulse widths and frequencies.
Choice of the appropriate mode and treatment program
The mode you choose determines the kind of work that is imposed upon the stimulated muscles. Choose the mode that is
appropriate to your needs or gives you the greatest pleasure.
How to Select TENS/EMS Mode
1. Press the “SET” button , the preset (default) therapy
mode TENS/EMS will flash on the display.
Use ON + or the button OFF –, if you would like to
change the therapy mode.
2. Press the SET button again, the numeric number of
program is then flashing.
3. Press the button ON + (to increase) or the button
OFF – (to decrease) for choice of program of the
selected modality.
Note : Make sure the identical receiver is turned on and
properly placed.
For TENS programs:
When using any of the 8 programs for pain relief always start with the lowest intensity and gradually increase the
level of intensity until you feel a “tingling” sensation. All programs are different and therefore feel differently. You may
try all 8 programs in the beginning and choose one that feels pleasant. Never increase the intensity to a level so that it hurts,
always stay under the point of discomfort. Start with short sessions of 5 or 10 minutes until your body gets used to the
stimulation.
Program Max. Phase duration Rate Function mode
P1 80mA 260uS 15Hz Constant
P2 80mA 260uS 60Hz Modulated
P3 80mA 260uS 60Hz Constant
P4 80mA 260~150uS 2 ~ 60Hz Modulated
P5 80mA 260~150uS 60Hz Modulated
P6 80mA 260uS 7 <->60Hz Modulated
P7 80mA 260~156 uS 60Hz Modulated
P8 80mA 210 uS 2.45~245 Hz Cycle
All electrical specification ±20%
Program./mode Benefits You should feel
13
P1
Continuous comfortable tingling. The underlying pain
should decrease gradually after treatment.
P2
Comfortable pulsing sensation.
The underlying pain should decrease almost immediately.
P3
Comfortable pulsing sensation. The underlying pain
should decrease almost immediately.
P4
Variable comfortable tingling and pulsing sensation
(sensation should appear to come in waves).
Pain should ease and there should be relief after
treatment.
P5
Variable comfortable mild tingling sensation (sensation will
appear to come in waves).
P6
Variable comfortable pulsing and pumping action (action
will appear to come in waves).
P7
-for temporary relief of pain associated with
sore and aching muscles in the lower back
due to strain from exercise or normal
household and work activities.
- for temporary relief of pain associated with
sore and aching muscles in the upper and
lower extremities ( arm and/or leg) due to
strain from exercise or normal household
and work activities.
Variable comfortable tingling and pumping action (action
should appear to come in waves).
P8
-for symptomatic relief and management of
chronic, intractable pain and relief of pain
associated with arthritis.
Variable comfortable tingling and pulsing sensation
(sensation should appear to come in waves).
Pain should ease and there should be relief after
treatment.
For EMS programs:
When using the device for muscle stimulation (EMS) any of the 6 programs may be used. The intent is to cause a muscle to
contract, and then release. All 6 programs will achieve contraction and vary mainly by the rate and duration of the
contractions. As with any exercise regiment, start out slowly with low intensity levels for a warm-up (5~10min). You may
increase intensity level and treatment time as you progress with your muscle performance. Use the device regularly over a
longer period of time as to maintain the benefit you may have gained during “exercise”.
Pulse Width
(uS)
Ramp up
(sec)
Hold on
(sec)
Ramp down
(sec)
Off Time
(sec)
Pulse rate
(Hz)
P1 300 - 2 - 1 40~99
P2 200 - - - - 4
P3 300 - - - - 5
P4 200 - 2 - 1 99
P5 200 2 6 2 1 4~20
P6 300 2 5 3 10 50
All electrical specification ±20%
Mode /
Exercise You should Feel & Benefits Suggestion
14
Program
P1
Exercise
Preparation
This program gently warms up the muscles
prior to exercise; it feels like a rhythmic
massage.
Increase the intensity until you get a
strong but comfortable muscle movement,
10 min/duration.
P2
Active
Recovery
This program produces muscle twitches at
very low frequency and it feels like a tapping
massage, for muscle recovery from fatigue
and becoming more relaxed with reduced
stiffness.
Use it after intense exercise to promote
recovery and relaxation, 30 min/duration.
P3
Active
Recovery
This program is similar to P2, except that the
muscle twitch rate slows down during the
session. It feels like a tapping massage, but
softer than P2.
Use it after intense exercise to promote
recovery and relaxation, 20 min/duration.
P4
Active
Recovery
This program activates the muscle in a short
contraction/relaxation cycle.
It feels like a kneading massage, smother
than P2/P3.
Use it after intense exercise to promote
recovery and relaxation, 20 min/duration.
P5
Build
Endurance
This program uses a low frequency pulse
train which favours slow twitch fibers, for
developing aerobic capacity and capillary
supply. It improves fatigue resistance during
long duration moderate intensity exercise.
The exercise comprises an alternating
sequence of work and rest phases lasting
several seconds. Increase the intensity
until you get a strong and deep muscle
contraction. Do not exceed your comfort
level, 20min/duration.
P6
Muscle
Strengthening
This program uses a pulse frequency
appropriate to fast twitch muscle fibers.
It improves their anaerobic capacity and is
used for improving maximum muscle
strength.
The exercise comprises a sequence of
work phases separated longer relaxation
phases. Increase the stimulation intensity
until you get a strong and deep
contraction. Do not exceed your comfort
level, 20 min./Duration.
Selecting the Therapy Time
1. Press the “SET” button again, the preset (default)
treatment time will flash on the display.
2. To increase or decrease the treatment time, press
the button ON + (to increase) or the button OFF –
(to decrease) repeatedly until the desired duration
appears on the display.
3. The treatment time you selected will appear on the
display the next time you turn the device on.
Note : If you change programs during the course of a
therapy session, the treatment time will not set
unless you manually reset it by performing the
steps described above.
Selecting the Therapy Intensity Level
15
The Device offers 25 intensity levels!
Intensity is adjustable according to the receiver
selected. Make sure the identical wireless unit is
turned on and properly placed.
1. Press the “SET” button, the numeric number of
program is then flashing.
2. To increase or decrease the intensity, press the
button On+ (to increase) or the button OFF– (to
decrease) repeatedly until the desired intensity level
flashes on the display.
Note: You will feel the intensity increase or
decrease as you select the intensity level.
You can use this as a guide to select a level that is
comfortable for you.
Note : If you change therapy mode/program during
the course of a therapy session, the intensity
level will reset to “0” for safety reason.
CARE AND MAINTENANCE
The Well-Life Wireless TENS/EMS System is recommended for single person use at home.
Remote and Receivers
The system device(s) may be wiped clean with a small amount of soapy water on a clean cloth. Do not
submerge the stimulator in liquids or expose it to large amounts of water.
For pet and pests consideration within the user area, you should clear the system device(s) each
time before use and keep them safe away in a drawer or somewhere pets and children won’t be
able to reach.
Never use aggressive cleaning products of stiff brushes to clean the device.
Do not clean the system devices when charging. Always un-plug the charger first before cleaning
the device(s).
Do no use the device again until it is completely dry.
Do not expose the device to direct sunlight and protect it from dirt and moisture.
Store the system in a clean, dry place.
Do not dispose of the device(s) in a fire. The battery could explode causing injury or death.
16
Electrode
The electrode pads are disposable and use an adhesive that will dry after prolonged usage or storage.
Pads should be replaced when they lose their adhesive quality, or you sense a change in stimulation
sensation.
If you’re in doubt about the integrity of the pads, order fresh pads please order online at
www.well-life.com.tw, or contact authorized distributor(s).
Rechargeable Battery
The rechargeable Battery is built-in the device(s). Please do not attempt to disassemble or force
open the built-in battery.
Batteries have a finite lifetime for 12 months. To maximize battery life, please charge the system every
3 months when not in use.
If you find the battery can not be charged effectively, or the fully charged battery can only be used for
very short time, please order replacement units online at www.well-life.com.tw, or contact authorized
distributor(s).
Do not dispose of the device(s) in a fire. The battery could explode causing injury or death.
How to Store Your System
1. Store your System at room temperature in a dry place, out of the reach of children.
2. Please store in a low humidity, low temperature environment.
17
TROUBLE SHOOTING
Always check the unit and accessories before use to prevent damage and defects; these are some of
the simple checks:
1. Make sure the Remote and Receiver(s) have a sufficient charged.
2. Make sure the Remote and Receiver(s) turn on and in 3 meters (9 feet) away from each other, for
good quality of receiving distance …
If this happens … Try this solution…
The Remote is not turning on or
I cannot see anything on the screen
Ensure that the remote has been fully charged.
The power light on the Receiver(s)
flashes continuously.
Charge the Receiver(s) or it will turn off automatically.
The Remote and/or Receiver(s) do not
seem to be charging.
Ensure the USB Cable is correctly into the Remote/
Receiver(s) and firmly plugged into the AC adaptor and
into the electrical socket on the other side.
Receiver(s) does not pair with remote
The remote does not control Receiver(s)
well
Make sure the receiver is fully charged, and turned
on, away from Remote in 3 meter distance.
Try to restart the remote and receiver(s).
Stimulation causes discomfort Electrodes lose their adhesive capacity and no
longer provide adequate contact with skin.
Electrodes are worn and must be replaced.
Change the positing of the electrodes slightly.
Choose Program that is appropriate to your needs
or gives you the greatest pleasure.
Stimulation does not produce the usual
sensation
Check that all settings are right and check
electrodes are properly positioned.
Change the positing of the electrodes slightly.
18
TECHNICAL SPECIFICATIONS
Channel: Dual , 2 wireless units independent and individually adjustable channels, electrically insulated from one another.
Pulse Rate : As pre-programmed, in operation mode.
Pulse Width : As pre-programmed, in operation mode.
Timer : 5~60 min. selectable.
Wave Form : Symmetrical Bi-Phasic square pulse.
Max Charge per Pulse : 20.8 micro-coulombs maximum.
** All electrical specifications are ±10% at 500 load.
Input rating: 5Vdc, 1A
Power Source : Remote Control Battery 3.7V /260 mAh Lithium polymer (LiPo) battery.
Wireless unit 3.7V /180 mAh Lithium polymer (LiPo) battery.
AC Adaptor manufacturer/model: FranMar International Inc. / FRM06-S05-z, 28.1x64.1x42 mm( LxWxH)
AC Adaptor I/P: 100-240Vac, 50-60Hz, 0.3-0.15A ; O/P: 5Vdc, 1.2A
Weight & Dimension: Remote Control 60 grams (battery included), 156.98x47x15.50 mm
Wireless Unit 30 grams (battery included) , 60x60x15.65 mm
Operating Conditions: + 50°F (10) to +104° (40), 40-90% max. Relative humidity
Transport and Storage Conditions: +14°F (-10) to +140° (60), 30-95% max. Relative humidity
Operation altitude: 3000m.
Operating Atmospheric pressure range: 700~1013 hPa
Transport and Storage Atmospheric pressure range: 500~1060 hPa
RF Module :
Radiations Frequency2.4GHz
Frequency Band2410~~2450MHz
GFSK typicalGFSK
(i) There are a number of technical symbols on your system, explained as follows:
Remote Label Receiver Label
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received,
(2) including interference that may cause undesired operation.
19
This symbol means “Serial number “on the back of remote control.
This symbol means “Lot number “on the back of wireless unit.
This symbols means “ Attention, consult the accompanying documents”
Follow instructions for use
This symbol means “ Manufacturer “
This symbol means type BF equipment; this device offers protection against
electrical shock by standard compliance to leakage currents of electrode pad.
This device shall be disposed in accordance with national laws after their
useful lives
IP22
Interference may occur in the vicinity of
equip
m
en
t
marked with the following
sym
bol
:
Drip proof. Do not immerse power unit in any liquid or spray any liquids directly
on power unit.
Stand-by.
(ii) There is a label on the package explained as:
This symbol means “used before”, represent as “YYYY-MM”
(for year and month).
ELECTROMAGNETIC COMPATIBILITY
The device complies with current specifications with regard to electromagnetic compatibility and
is suitable for use in all premises, including those designated for private residential purposes.
The radio frequency emissions of the device are extremely low and in all probability do not cause
any interference with other devices in the proximity.
It is recommended that you do not place the device on top of or close to other electronic devices.
Should you notice any interference with other electrical devices, move the device or connect it to
a different socket.
Radio equipment may affect the operation of this device.
20
Recommendedseparation distancesbetween
portableandmobileRFcommunications equipment and the ME equipment
TheWirelessElectricalStimulatorisintendedforuseinanelectromagneticenvironmentinwhichradiatedRF
disturbancesarecontrolled.ThecustomerortheuseroftheWirelessElectricalStimulatorcanhelpprevent
electromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommunications
equipment(transmitters)andtheWirelessElectricalStimulatorasrecommendedbelow,accordingtothemaximum
outputpowerofthecommunicationsequipment.
Separationdistanceaccordingtofrequencyoftransmitter
m
Ratedmaximumoutput
poweroftransmitterW150kHzto80MHz
80MHzto800MHz
800MHzto2.5GHz
0.010.120.120.23
0.10.370.370.74
11.171.172.33
103.73.77.37
10011.6711.6723.33
Declaration – electromagnetic emissions and immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING and are specified for use only in a
shielded location
TheWirelessElectrical Stimulatordeclaration electromagnetic immunity
TheWirelessElectricalStimulatorsystemisintendedforuseintheelectromagneticenvironmentspecifiedbelow.
Thecustomerortheusero
f
theWirelessElectricalStimulato
r
system should assure thatitisused in suchanenvironment.
ImmunitytestIEC60601testlevelCompliancelevelElectromagneticenvironment‐guidance
ConductedRF
IEC6100046
3Vrms
150kHzto80MHz
3V
RadiatedRF
IEC6100043
3V/m
80MHzto2.5GHz
3V/m
PortableandmobileRFcommunicationsequipmentshouldbe
usednoclosertoanypartoftheEQUIPMENTorSYSTEM
includingcables,thantherecommendedseparationdistance
calculatedfromtheequationapplicabletothefrequencyofthe
transmitte
Interferencemayoccurinthevicinityofequipmentmarked
withthefollowingsymbol.
21
Declaration electromagnetic immunity
TheWirelessElectricalStimulatorsystemisintendedforuseintheelectromagneticenvironmentspecifiedbelow.
Thecustomerortheusero
f
theWirelessElectricalStimulato
r
system should assure thatitisused in suchanenvironment.
ImmunitytestIEC60601testlevelCompliancelevelElectromagneticenvironment‐guidance
Electrostatic
discharge(ESD)
IEC6100042
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteorceramic
tile.Iffloorsarecoveredwithsynthetic
material,therelativehumidityshouldbeat
least30
%
.
Electricalfast
transient/burst
IEC6100044
±2kVforpowersupplylines
±1kVforinput/outputlines
±2kVforpowersupplylines
Mainspowerqualityshouldbethatofa
typicalcommercialorhospitalenvironment.
Surge
IEC6100045
±1kVdifferentialmode
±2kVcommonmode±1kVdifferentialmodeMainspowerqualityshouldbethatofa
typicalcommercialorhospitalenvironment.
Voltagedips,
short
interruptionsand
voltage
variationson
powersupply
inputlines
IEC61000411
5%UT
(95%dipinUT)for0.5cycle
40%UT
(60%dipinUT)for5cycles
70%UT
(30%dipinUT)for25cycles
5%UT
(95%dipinUT)for5sec
5%UT
(95%dipinUT)for0.5cycle
40%UT
(60%dipinUT)for5cycles
70%UT
(30%dipinUT)for25cycles
5%UT
(95%dipinUT)for5sec
Mainspowerqualityshouldbethatofa
typicalcommercialorhospitalenvironment.
IftheuseroftheEQUIPMENTorSYSTEM
requirescontinuedoperationduringpower
mainsinterruptions,itisrecommendedthat
theEQUIPMENTorSYSTEMbepoweredfrom
anuninterruptiblepowersupplyora
battery.
Powerfrequency
(50/60Hz)
magneticfield
IEC6100048
3A/m
3A/m
Powerfrequencymagneticfieldsshouldbeat
levelscharacteristicofatypicallocationina
typicalcommercialorhospitalenvironment.
Declarationelectromagneticemissions
TheWirelessElectricalStimulatorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerorthe
useroftheWirelessElectricalStimulatorshouldassurethatitisusedinsuchanenvironment.
EmissionstestComplianceElectromagneticenvironment‐guidance
CEemissions
CISPR11
Group1
TheWirelessElectricalStimulatorusesRFenergyonlyforitsinternalfunction.Therefore,
itsRFemissionsareverylowandarenotlikelytocauseanyinterferenceinnearby
electronice
q
ui
p
ment.
REemissions
CISPR11
ClassB
Harmonicemissions
IEC6100032
ClassA
Voltagefluctuations/
Flickeremissions
IEC6100033
Complies
TheWirelessElectricalStimulatorissuitableforuseinallestablishments,including
domesticestablishmentsandthosedirectlyconnectedtothepubliclowvoltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
22
FCC Information
The Federal Communication Commission Radio Frequency Interference Statement includes the
following paragraph:
The equipment has been tested and found to comply with the limits for a Class B Digital Device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instruction, may
cause harmful interference to radio communication. However, there is no grantee that interference will
not occur in a particular installation. If this equipment dose cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on , the user is
encouraged to try to correct the interference by one or more of the following measures:
--Reorient or relocate the receiving antenna.
--Increase the separation between the equipment and receiver.
--Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
--Consult the dealer or an experienced radio/TV technician for help.
The user should not modify or change this equipment without written approval from
Well-Life Healthcare Limited . Modification could void authority to use this equipment.
Notice : The changes or modifications not expressly approved by the party responsible for compliance could
void the users authority to operate the equipment.
IMPORTANT NOTE:
To comply with the FCC RF exposure compliance requirements, no change to the antenna or the device is
permitted. Any change to the antenna or the device could result in the device exceeding the RF exposure
requirements and void user’s authority to operate the device.
23
Warranty
This Well-Life Stimulator carries a one-year warranty from the date of purchase.
Product Service Life: Three (3) years.
The warranty applies to the main device and necessary parts and labor relating thereto.
Consumable items like : batteries, electrodes, and other accessories are warranted to be free from
defects in workmanship and materials at the time of delivery.
The warranty does not apply to damage resulting from failure to follow the operating instructions,
accidents, abuse, alterations or disassembly by unauthorized individuals.
The distributors reserve the right to replace or repair the unit at their discretion.
Well-Life Healthcare Ltd.
1Fl. No.16 , Lane 454. Jungjeng Road,
Yunghe District, New Taipei City 23455,
Taiwan
www.well-life.com.tw
SANICO Srl
Via Turati,7, 20121 Milano

Navigation menu