eNeura HL7618RD sTMS mini User Manual

eNeura Inc. sTMS mini

User manual

DRAFT 2/12/2019 SPW 1
sTMS mini
Instructions for Use
sTMS mini Patient Manual
Caution: Federal law (U.S.) restricts this device
to sale by or on the order of a physician.
DRAFT 2/12/2019 SPW 2
Table of Contents
Before You Begin ......................................................................................................................................................... 3
Intended Use ................................................................................................................................................. 4
Warnings and Precautions ............................................................................................................................ 4
Contraindications .......................................................................................................................................... 6
Clinical Trial and Adverse Reactions ............................................................................................................. 8
Your sTMS mini ............................................................................................................................................. 9
Getting to Know Your Device .................................................................................................................................... 10
Using Your Device ..................................................................................................................................................... 12
Setting Up the Device ................................................................................................................................. 12
Recommended Treatment .......................................................................................................................... 12
Preparing for Treatment ............................................................................................................................. 13
Positioning the Device ................................................................................................................................ 15
Delivering the Treatment ............................................................................................................................ 16
Renewing Your Prescription by Remote Activation .................................................................................... 17
Renewing Your Prescription by optional Micro SIM Rx card ...................................................................... 17
Replacing the Micro SIM Rx card ................................................................................................................ 17
Additional Information .............................................................................................................................................. 18
Caring For Your sTMS mini ........................................................................................................................................ 20
Understanding System Display Messages ................................................................................................................. 22
Troubleshooting ........................................................................................................................................................ 24
Service ......................................................................................................................................................... 25
Return Goods Policy .................................................................................................................................... 25
Technical Specifications ............................................................................................................................................ 25
Operating Environment .............................................................................................................................. 26
Storage Environment .................................................................................................................................. 26
Industry Standard Classification ................................................................................................................. 26
REACH and Warning Statement .................................................................................................................. 27
EMC Compliance and Warning Statement.................................................................................................. 27
EMC Guidance ............................................................................................................................................. 28
Glossary of Abbreviations ......................................................................................................................................... 33
Medical Device Reporting ......................................................................................................................................... 34
Warranty and Limitation of Liability.......................................................................................................................... 34
Customer Care Contact Information ......................................................................................................................... 34
DRAFT 2/12/2019 SPW 3
Before You Begin
Thank you for choosing the eNeura sTMS mini, the portable single-pulse Transcranial Magnetic Stimulation (sTMS) device. This
manual provides the Information you need to use the sTMS mini. You will see cautions, warnings, and helpful information placed
near the related steps. Call Customer Care if you don’t understand something in this manual.
eNeura and your doctor are committed to helping you manage your migraine pain. The sTMS mini is a new, safe, non-drug therapy
to treat migraine at the first sign of pain.
Please read this entire manual before using the sTMS mini. Learn the contraindications, cautions, warnings and notes about the use
of the device. As the manufacturer, eNeura cannot and does not intend to give medical advice. Contact your doctor for all medical
advice. This manual should be kept near the sTMS mini and be available at all times.
eNeura is committed to the service and support of our customers. If there are any questions about the use of the eNeura sTMS
mini, please contact Customer Care or your local representative at the following:
Manufactured by: eNeura Inc.
715 North Pastoria Avenue
Sunnyvale, CA 94085
eNeura Inc.
715 North Pastoria Avenue
Sunnyvale, CA 94085
Tel: +1 408.245.6400
Toll free (USA only): +1 855.366.8355
Fax: +1 877.874.9584
www.eNeura.com
info@eNeura.com
customercare@eNeura.com
eNeura (UK) Ltd.
6th Floor
One London Wall London EC2Y 5EB
United Kingdom
www.eNeura.co.uk
Tel. +44 (0) 20.3695.4063
Fax: +44 (0) 20.7785.8152
customercare@eneura.co.uk
Authorized Representative
Emergo Europe BV
Prinsessegracht 20
2514 AP The Hague
The Netherlands
Tel: +31 (0) 70.345.8570
Fax: +31 (0) 70.346.7299
www.emergoeurope.eu
Australia Sponsor
Emergo Australia
Level 20, Tower II, Darling Park
201 Sussex Street
Sydney, NSW 2000
Australia
www.emergogroup.com
DRAFT 2/12/2019 SPW 4
Intended Use
The sTMS mini (The System) is indicated for the acute and prophylactic treatment of migraine headache.
The System is designed for self-treatment and delivers a non-invasive, brief, single pulse of magnetic energy to the back of the head.
This creates a brief electrical current in the brain intended to stop or reduce the effects of migraine headaches. This type of
stimulation is called single-pulse Transcranial Magnetic Stimulation or sTMS.
The System is a drug-free treatment option that can be used in the home or away from home based on your doctor’s instructions.
After treatment, there are no restrictions. You can resume your normal activities.
WARNING: This device should be used under the supervision of a physician.
Keep the sTMS mini out of the reach of children.
Safety and effectiveness have not been established in pregnant women, children under the age of 18
and adults over the age of 65.
The long-term effects of single-pulse transcranial magnetic stimulation are unknown.
Warnings and Precautions
The words WARNING, Precaution and NOTE have special meanings in this manual. Read them throughout the manual to ensure
the safe and effective use of your sTMS mini.
WARNING: A WARNING tells you that the personal safety of the patient may be involved. Ignoring a
WARNING could result in injury to the patient. WARNINGS in the manual are shown in an orange box.
Precaution: A Precaution means that exact steps must be followed to prevent damage to the product. Precautions
in the manual are shown in a purple box.
NOTE: A NOTE gives special information to ease product use or to explain important information.
NOTES in the manual are shown in a dashed box.
DRAFT 2/12/2019 SPW 5
WARNING: The sTMS mini should be used under the continued supervision of a physician. The System
has been prescribed by your physician only for you.
Inspect the System for any signs of damage before use. Do not use it if it is cracked or wet. If you
suspect damage to the device, call eNeura at: +1 855.366.8355 option 1 for assistance.
Do not operate the System in or near an area where explosive gases are being used or have been used.
Do not operate near gasoline or natural gas.
Do not operate the System in or near the presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR
or WITH OXYGEN or NITROUS OXIDE.
Risk of electrical shock. Do not open the System. There are no parts that can be serviced or replaced
by the user. High voltage may be present.
Risk of electrical shock. Do not allow the System or power cords to get wet. Quickly wipe up spills on
or near the sTMS mini. Do not use the System in or near water. For example, do not use while in the
bathtub or shower, in the rain, or while standing in water or on a wet surface.
Do not use the System if the cause of your headache is illness, underlying pathology, trauma or overuse
of medication. See your physician if you are uncertain.
Do not use the System if you have suspected or diagnosed epilepsy or a personal or family history of
seizures. Consult your physician before using the System if a family member has epilepsy or seizures
or if you have had seizures, a head trauma or head injury or take any medication such as tricyclic
antidepressants, neuroleptic agents, or other drugs that lower the seizure threshold.
Do not use the System if you have a history of stroke.
The System has not been demonstrated as safe or effective when treating cluster headache.
Do not use the System if you use a wearable cardioverter defibrillator (WCD).
The long-term effects of chronic magnetic stimulation are unknown.
Transcranial magnetic stimulation should only be applied to the back of the head as described in the
“Using Your Device” section of this manual.
Do not stimulate over the front of the neck or mouth. Severe spasm of the laryngeal and pharyngeal
muscles may occur and the contractions may be strong enough to close the airway or cause difficulty
in breathing.
Do not stimulate over the upper side of the neck. Stimulation of the carotid sinus nerves, particularly
in patients with a known sensitivity to the carotid sinus reflex, could result in a sudden drop in blood
pressure, slowing of the heart or loss of consciousness.
Do not stimulate the chest or back. The induction of electrical current into the heart may cause cardiac
arrhythmias.
DRAFT 2/12/2019 SPW 6
Contraindications
WARNING: Failure to follow the restrictions listed below could result in serious injury or death.
The sTMS mini creates a very strong single-pulse magnetic field. The System has been prescribed by your doctor for your use only.
The System may not be used in patients who have metals, conductive materials, or metal-containing implants in their head, neck or
upper body. Metals and conductive materials can be affected by a magnetic field. You should discuss this with your doctor before
use.
Do not use the System if you have a cardiac pacemaker, vagus stimulator (VNS) or other implanted neurostimulator, implanted
cardioverter defibrillator (ICD) or any implanted medical device that stimulates the body or uses any signal from the body.
Talk to your doctor before using the System if you receive an implant. Patients with implants affected by a magnetic field should not
use the System. Examples of such implants include:
Aneurysm clips or coils
Cochlear implants
Cerebral spinal fluid shunts
Bullets or pellets lodged in the head or upper
body
Metal plates, screws, staples or sutures in
skull, neck, shoulders, arms or hands
Radioactive seeds
Magnetically programmable shunt valves
Stents
Filters
Metallic artificial heart valves
Facial tattoos with metallic ink
Electrodes
Dental implants, fillings or other dental appliances are okay and are not affected by the device.
NOTE: If you have an implanted medical device but are not sure if it has metals or conductive
materials, please talk to your doctor before using the sTMS mini.
DRAFT 2/12/2019 SPW 7
WARNING: The System should not be used while driving, operating machinery or during any activity in
which involuntary muscle contractions may put the user at risk of injury.
Stay at least 2 feet (0.6 meter) from others when using the System. The System could be harmful to
anyone with an electronic implant such as a pacemaker. Anyone with a hearing aid or cochlear implant
may hear an audible click.
The device could be disrupted by RF-emitting equipment including: wireless home network devices,
mobile phones, cordless telephones and their base stations and walkie-talkies. See “EMC Compliance
and Warning Statement” section for additional information on preventing unwanted interference.
Precaution: Keep the System away from other electronic devices that depend-on (receive) or radiate
(transmit) radio frequency energy, when it is powered on.
The operation of the System may be impaired when operated near home devices such as wireless
network routers, mobile phones, cordless telephones and their base stations and walkie-talkies. Keep
the sTMS mini device at least 2 feet (0.6 meter) from these devices when it is powered on and in use.
Keep credit cards, audio/video tapes, computers, computer disks, flash memory sticks, cell phones,
personal digital assistants (PDAs), MP3 players, headphones, digital cameras, portable glucose meters
and other electronic devices or electronic storage media more than 2 feet (0.6 meter) away from the
System when it is in use.
Keep any loose metal objects such as eyeglasses, keys, coins, jewelry, watches and hair clips more than
2 feet (0.6 meter) away from the System when it is in use.
Keep wearable medical devices such as insulin pumps, medicinal pumps, monitors, bone grow
stimulators and Transcutaneous Electrical Nerve Stimulator (TENS) devices more than 2 feet (0.6 meter)
away from the System when it is in use.
Safety and effectiveness have not been established in pregnant women, children under the age of 18
and adults over the age of 65.
Caution should be used for patients with suspected or diagnosed heart problems.
The System is only intended to be serviced or maintained by the manufacturer. Do not attempt to open
the device. The warranty may be invalidated. If the device is opened, contact eNeura at: +1
855.366.8355 option 1.
Keep the System out of the reach of children.
Side effects can include minor dizziness, nausea, vomiting, application site tenderness, muscle spasm,
headache and migraine.
Special precautions regarding Electromagnetic Compatibility (EMC) are required when installing and
using the System. Portable and mobile communications devices can affect proper operation of the
System. See “EMC Compliance and Warning Statement” section of the Instructions for Use for more
information.
DRAFT 2/12/2019 SPW 8
Clinical Trial and Adverse Reactions
eNeura completed a clinical study in Headache Centers in the United States designed to support safety and effectiveness for use of
the Spring TMS device for the acute treatment and prophylaxis of migraine headache. Baseline medication and symptoms were
recorded for 28 days via patient diary. Subjects were instructed to use the SpringTMS device daily (morning and evening) and for
the acute treatment of attacks for three months with no change in preventive medication. Study results showed statistically
significant reduction in migraine headache days of 2.8 days from a baseline mean of 9.1 days. Forty-six percent of the patients saw
at least a 50% reduction in headache days (responder rate). Reduction in acute medication was 2.9 days. A statistically significant
improvement in quality of life was reported. Six patients reported using prophylactic medications at baseline (2.8%), of these, 5 were
using topiramate and propranolol.
The efficacy of the eNeura SpringTMS for prophylactic treatment is based on the result of an open label study. In open label studies,
bias may affect the result. Additionally, open-label studies may introduce placebo rates of 10-25%. This is consistent with placebo
rates reported in Randomized Controlled studies for migraine prevention.*
For further information about the clinical trial and any adverse reaction, please contact eNeura at 1-855-366-8355, option 1 or
consult your prescribing physician
*Macedo A, Banos JE, Farre M. Placebo response in the prophylasix of migraine: a meta-analysis. Eur J Pain 2008; 12;68-75
DRAFT 2/12/2019 SPW 9
Your sTMS mini
The complete sTMS mini includes this manual and the following items:
The battery-powered, rechargeable device
Prescription card with SIM (Subscriber Identity
Module) Rx card (SIM Rx card is optional - your prescription
may be installed via remote activation)
NOTE: Your first prescription is preinstalled at the factory.
Battery Charger 12V DC 1.5A 18 watts
(reorder no. DWG-0505)
DRAFT 2/12/2019 SPW 10
Getting to Know Your Device
A. Power Button In the center of the panel on the top of the System. Press the power button to turn the device on and
off.
B. Power Indicator LED light on the top of the System inside the Power button. Static white illumination shows the device is
on and ready. Pulsing blue illumination indicates device is connecting to the cellular network to transfer data.
C. Treatment Progress Indicator LED light around the power button on the top of the System, shows status as it prepares
for treatment.
D. Prescription Status Indicator Located on left side on the top of the System. Confirms a valid prescription is available and
shows the status of the prescription: White illumination indicates Rx is active; static orange indicates Rx will expire soon; static
orange with orange Power Button indicates Rx has expired
E. Contact eNeura Customer Care Action required. Call Customer Care.
F. Temperature warning The device temperature is not in range for safe use.
G. Lock Indicator On the right of the Power button. Indicator is visible when device is turned on if the security lock switch is
enabled.
H. AC Adapter On the right of the Power button. Indicator is lit when the AC adapter is connected to the device.
I. Battery Capacity On right side on the top of the System. Indicates whether or not battery power is enough to allow
treatment. White illumination indicates battery power is available; orange illumination indicates battery power will soon be or has
been exhausted.
J. Treatment Buttons On the right and left edge of the panel on the top of the device. Press one or both buttons to deliver a
treatment.
Read Understanding System Display Messages” for more information.
A
B
C
F
G
H
I
D
DRAFT 2/12/2019 SPW 11
Battery Charger (Power Supply) and Micro SIM Rx card Ports
K. Lock ON/OFF Switch Located behind accessory port door, on the left. Slide the switch to the left or right to unlock or lock the
device
L. Volume ON/OFF Switch Located behind accessory port door, on the right. Slide the left or right to turn the sound on or off
M. Battery Charger Port Located behind accessory port door, in the center. Recharge the batteries by plugging the AC battery
charger, DWG-0505, into this port
AC battery charger 12V DC
1.5A 18 watts
(reorder no. DWG-0505)
Plug that connects to
the System
N. Micro SIM Rx card port Located on the back of the System beneath the Accessory Door. Insert the micro SIM RX card into this
port.
Precaution: Use only eNeura-supplied accessories with your System: Battery charger (reorder no.
DWG-0505) and Micro SIM Rx card.
M
N
L
K
Door side of device, with door open
DRAFT 2/12/2019 SPW 12
Using Your Device
Precaution: Prior to use of the System, this manual must be completely read and understood.
Improper use may cause personal injury and damage to the System and may void the
warranty.
You, as the user of the System, must become familiar with the functions and features of the
device prior to use.
Setting Up the Device
Be prepared to treat your migraine when needed. Set up your sTMS mini as soon as it arrives and keep your device charged
and ready to treat.
NOTE: Use only as directed. The device should only be used per the instructions of your
physician who prescribed the System.
Remove the System and the battery charger from the box.
NOTE: Do not discard box or packing materials. They will be required if you need to return this
product.
Your device should arrive ready to use. If it is not ready, plug the battery charger into a standard wall outlet. Connect
the other end of the battery charger to the port on the device. Charge the battery for about 4 to 6 hours.
Treatment delivery is not available during battery charging.
If the device is on while charging, the AC adapter symbol and the battery capacity symbol will be lit.
Unplug the battery charger when the device is fully charged. A fully charged battery should deliver at least 12
treatments.
Keep the accessory door closed when not in use.
Store the device and its battery charger in a cool, dry place, away from excessive dust and direct sunlight.
Precaution: Only use the battery charger (DWG-0505) included with the sTMS mini. Contact
eNeura if you have any questions.
Recommended Treatment
The treatment with the sTMS mini should be performed per the instructions of your prescribing physician.
Prevention
Treat with 4 pulses each morning and evening
2 consecutive pulses
Wait 15 minutes
Repeat the 2 consecutive pulses
Acute
3 sequential pulses (early) at the onset of migraine pain
Wait 15 minutes
If needed treat with additional 3 pulses
Wait 15 minutes
If needed treat with additional 3 pulses
DRAFT 2/12/2019 SPW 13
Read the instructions below prior to operating the device. Then, follow the entire procedure as described below. Consult
your physician for additional treatment information.
NOTE: Become familiar with the System and how to use it before you experience a migraine.
Preparing for Treatment
Consult your doctor about your specific condition and the best timing of your sTMS treatment.
Precaution: Keep the System away from other electronic devices that depend-on (receive) or
radiate (transmit) radio frequency energy, when it is powered on.
The operation of the System may be impaired when operated near home devices such as
wireless network routers, mobile phones, cordless telephones and their base stations and
walkie-talkies. Keep the sTMS mini device at least 2 feet (0.6 meter) from these devices when
it is powered on and in use.
The System emits a strong magnetic pulse that may interfere with the operation of common
home electronic devices such as radios, televisions, wireless network routers, mobile phones,
cordless telephones and their base stations and walkie- talkies, if they are not installed and
used in accordance with the manufacturer’s instructions. There is no guarantee that
interference will not occur in a particular installation. If the System causes interference with
other devices, try to correct the matter by:
o Reorienting or relocating the device receiving the interference.
o Increase the separation distance between the System and the device.
o Connect the System to an AC outlet on a circuit different from that to which the other
device is connected.
Use the System in a comfortable, seated position. Remove the System from its storage location.
WARNING: Inspect the System for any signs of damage before use. Do not use if it is cracked
or wet. If you suspect damage to the device, call eNeura at +1 855.366.8355 option 1 for
assistance.
The System needs to be in direct contact with the back of your head to work properly. Before treatment, remove any hat,
head covering, or hairpiece that covers the back of your head and undo any braids, ponytails or buns.
DRAFT 2/12/2019 SPW 14
WARNING: Remove any personal medical devices such as insulin pumps or other medical
pumps and bone growth stimulators.
Press the power button
Watch the LED track light progress around the power button:
It takes 30-60 seconds to prepare the System for treatment.
As the device prepares for treatment, each segment will light up to indicate progress.
When the system is ready, the device lights up all 6 segments and sounds an audio tone:
Once the System is ready, you have 45 seconds to position the device and deliver treatment.
WARNING: Stay at least 2 feet (0.6 meter) from others when using the sTMS mini. This device
could be harmful to anyone with an electronic implant such as a pacemaker.
Remove any personal medical devices such as insulin pumps, other medicinal pumps, bone
growth stimulators, Transcutaneous Electrical Nerve Stimulator (TENS) devices and hearing
aids.
Do not use the System if you use a wearable cardioverter defibrillator (WCD).
The System should not be used while driving, operating machinery or during any activity in
which involuntary muscle contractions may put you at risk of injury.
Risk of electrical shock: Do not allow the System or power cords to get wet. Quickly wipe up
spills on or near the device. Do not use the device in or near water. For example, do not use
while in the bathtub or shower, in the rain or while standing in water or on a wet surface.
Do not operate the System in or near an area where explosive gases are being used or have
been used. Do not operate near diesel fuel, gasoline or natural gas.
Do not operate the System in or near the presence of a FLAMMABLE ANESTHETIC MIXTURE
WITH AIR or with OXYGEN or NITROUS OXIDE.
Do not use the System if the cause of your headache is illness, trauma or excess medication.
Consult your doctor if you are unsure.
Do not use the System if you have suspected or diagnosed epilepsy. Consult your doctor
before using the System if a family member has epilepsy. Consult your doctor if you have had
a head trauma, head injury or take any medication such as tricyclic antidepressants,
neuroleptic agents or other drugs that lower the seizure threshold.
Do not use the System if you have a history of stroke.
Caution should be used for patients with suspected or diagnosed heart problems.
DRAFT 2/12/2019 SPW 15
Precaution: Keep the sTMS mini away from metal or conductive objects, medical devices and
magnetic media when it is powered on.
When activated, the System emits a strong magnetic pulse and may interfere with other metal,
electronic or magnetic products within 2 feet (0.6 meter). A person with a hearing aid or
cochlear implant may hear a click when the System is activated.
The operation of the sTMS mini System may be impaired when operated near home devices
such as wireless network routers, mobile phones, cordless tele- phones and their base stations,
and walkie-talkies. Keep the sTMS mini device at least 2 feet (0.6 meter) from these devices
when it is powered on and being used.
o Move credit cards, audio/video tapes, computers, computer disks, flash memory sticks,
cell phones, personal digital assistants (PDAs), MP3 players, headphones, digital
cameras, portable glucose meters and other electronic devices or electronic storage
media more than 2 feet (0.6 meter) away from the System.
Remove all metal objects from your head, face, neck, arms and hands. Remove eyeglasses,
hearing aids, removable metallic orthodontic appliances, hair clips and earphones and move
them more than 2 feet (0.6 meter) away from the System.
Remove any loose metal objects from your shirt pockets such as keys, coins, clips, pens and
mechanical pencils and move them more than 2 feet (0.6 meter) away from the System.
Positioning the Device
With the sTMS mini in front of you: The treatment buttons should be facing up and the accessory door should be
facing away from you. Place your hands on the sides of the device with your thumbs on the treatment buttons.
Lift the System over your head. The device should be placed so that it comfortably and naturally cradles the base of
your skull. Hold it firmly against your head.
The curved surface positions the device to fit comfortably against the head and optimize pulse delivery.
DRAFT 2/12/2019 SPW 16
WARNING: Transcranial magnetic stimulation should only be applied to the back of the head
as described in the “Positioning the Device” section of this manual.
Do not stimulate over the front of the neck or mouth. Severe spasm of the laryngeal and
pharyngeal muscles may occur. Contractions may be strong enough to close the airway or
cause difficulty breathing.
Do not stimulate over the side of the neck. Stimulation of the carotid sinus nerves, particularly
in patients with a known sensitivity to the carotid sinus reflex, could result in a sudden drop
in blood pressure, slowing of the heart or loss of consciousness.
Do not stimulate the chest or back. Electrical current into the heart may cause cardiac
arrhythmias.
NOTE: If the device detects an error when it is on, a symbol is displayed. See the
“Understanding System Display Messages” and “Troubleshooting” sections of this manual.
NOTE: The System is a handheld device that comes in direct contact with skin. The housing of
the device is made of polycarbonate, a plastic common in consumer medical products, and
poses no handling risk.
Delivering the Treatment
NOTE: Two hands are needed to stabilize the device, but only one treatment button needs to
be activated to deliver treatment.
NOTE: The device shuts itself off if treatment is not delivered within 45 seconds. Press the
power button to turn the device back on and restart the treatment procedure.
Once the device is in place and your thumbs are on the treatment switches, press one or both switches and hold for at
least 2 seconds.
The device emits a soft, audible click as the treatment is delivered.
After you have successfully delivered the first pulse, remove the device from your head.
The LED light on the POWER button softly pulses to show that treatment is completed. The device is now ready for
another pulse. You may select to treat again by pressing the power button.
Repeat the subsequent procedures from page 15, Positioning the Device.
After delivering the final pulse and 30 seconds of inactivity, the device will attempt to connect to a cellular network,
transfer data and will then power off.
Delivery of several pulses in a row may cause the device’s surface temperature to exceed 105.8 F (41 C). The surface
may get as warm as 118.4 F (48 C). Let the device cool for 15 minutes. Try again.
DRAFT 2/12/2019 SPW 17
Renewing Your Prescription by Remote Activation
NOTE: The System is intended for use under the care of a doctor. For more information on renewing
your prescription, please contact eNeura Customer Care. An active prescription is required.
NOTE: The System only delivers sTMS treatments if the programmed prescription has not
expired.
The device will display an ORANGE warning light when prescription is about to expire: Renew your prescription
promptly.
Contact Customer Care to request a refill.
Upon receipt of payment, Customer Care will initiate reactivation of your device via the cellular network.
Press the Power Button on your device. Pulsing blue illumination on the power button indicates the device is
attempting to connect to a cellular network.
After the device has successfully connected to a network, the new Rx data will transfer and the ORANGE warning light
will become white again. You may resume use.
Renewing Your Prescription by optional Micro SIM Rx card
The Micro SIM Rx card from your prescription card may be installed to activate the device to deliver sTMS treatments.
The device will display an ORANGE warning light when your Micro SIM Rx card is about to expire: Renew your
prescription promptly.
After renewing your prescription, you will receive a prescription card
containing a new Micro SIM Rx card within 2-3 business days.
Replacing the Micro SIM Rx card
Open the accessory door on the back of the device.
Push on the edge of the Micro SIM Rx card. It pops out about 3 mm (1/8 inch).
DRAFT 2/12/2019 SPW 18
Pull the Micro SIM Rx card out with your fingers. This expired Micro SIM Rx card is of no value.
Dispose as eWaste or recycle by appropriate methods.
WARNING: The Micro SIM Rx card may be swallowed and present a choking hazard for infants
or small children if it is ejected from the SIM card port and left loose and unattended.
Remove your new Micro SIM Rx card (the cut-out portion) from the prescription card by pressing down firmly on the
Micro SIM Rx card. Check that it is the same shape as shown below. Remove any excess plastic.
Insert the Micro SIM Rx card with the gold side up and the edge with the blunt corner pointing into the Micro SIM Rx
card port
Push the Micro SIM Rx card until it is flush with the opening and snaps into place. Close the accessory door.
Additional Information
Battery Capacity Indicator
Your sTMS mini has rechargeable batteries. The battery capacity indicator on your device shows how much charge is
remaining in the battery. eNeura recommends that you keep your device’s battery charged so that it is always ready when
you need it. Typically, a fully charged battery can deliver 18 treatments. Your battery pack life is roughly a minimum of 100
charge cycles.
Recharging the Battery
To recharge your battery, push the round plug on the battery charger into the battery charger port on the back of the
device. Plug the battery charger into a standard wall outlet.
The AC adapter and battery capacity indicators are lit while the device is on and the battery is charging. Treatment
while recharging is disabled for safety. To use the device for treatment, unplug the battery charger and proceed with
Micro SIM Rx card
DRAFT 2/12/2019 SPW 19
your treatment.
WARNING: The battery charger cable may present a strangulation or trip hazard if the System
is left unattended around infants and small children. Recharge the System’s battery in a safe
location, away from foot traffic and children. Store the battery charger in a safe place when
not in use.
DRAFT 2/12/2019 SPW 20
Locking the System
The System lock switch is located beneath the Accessory door.
The System lock may be controlled by sliding the switch to the desired position (ON or OFF).
When the System is locked, the device will prevent others from using it by disabling
treatment functionality.
Activating the System lock switch places the device in Airplane Mode. Place the System in
Airplane Mode whenever you are in a location where use of cellular devices is discouraged or
forbidden.
Adjusting the Audible Tone
The System sounds an audible tone when the device is ready to deliver a treatment.
The switch to control the audible tone is located beneath the Accessory door.
Slide the switch to the desired position (ON or OFF).
Travel
Your sTMS mini may travel with you. Place the device in Airplane Mode and pack it securely to avoid damage during
travel. Before using the unit after traveling, check it for any damage to ensure it appears to be working correctly
before you use it.
For international travel, use either the correct international plug adapter or contact eNeura to purchase the correct
plug for your battery charger. Only use the AC adapter that was provided with the System.
Caring For Your sTMS mini
Inspect your device regularly
Before using your sTMS mini always check to make sure your device is in good working order. If you notice damage to
the exterior of the device or rattling when you shake the device DO NOT USE the device and contact your eNeura
representative for a replacement.
Rechargeable Battery
The rechargeable battery in the sTMS mini is in a sealed compartment and may only be replaced by a trained service
technician. If the battery fails to recharge, please contact Customer Care to arrange for a device exchange.
Cleaning Procedure
eNeura recommends cleaning the sTMS mini every three months, similar to the way you would clean a telephone
handset. It may be necessary to clean the System more often if it becomes dirty or sticky. It may be cleaned as many
times as needed.
Only the external surfaces of the System may be cleaned. Wipe dust and lint away with a dry cloth. Take care not to
drip or spray liquid directly onto your device or get liquid into any of the ports, the buttons or the display. Dampen a
soft cloth in water premixed with a mild detergent or an alcohol- based, hospital-grade, hand-disinfecting solution.
Wring out any excess liquid, and then wipe the device. Immediately dry the device with a soft dry cloth to make sure
that no liquid remains on the surface.
Do not use any spray cleaners because fluid should not enter openings. Do not use detergents or other cleaning
solutions to clean the display.
DRAFT 2/12/2019 SPW 21
Precaution: Disconnect the sTMS mini from the battery charger prior to cleaning. NEVER
immerse the System. Do NOT drip or splash any liquid on the device. Contact eNeura if the
device gets wet. Do not attempt to sterilize the System.
DRAFT 2/12/2019 SPW 22
Understanding System Display Messages
If your device displays one of the following images, please take the steps listed to clear the message.
Battery charge is full
Battery charge is low
The rechargeable battery charge is low. There is only enough power to deliver approximately 10
treatments so charge the battery at your earliest convenience.
Charge the battery for at least 15 minutes. To fully charge the battery, allow the device to charge for
about 4 hours.
If the problem persists, check that only the eNeura battery charger (DWG-0505) is in use. Check that the
battery icon on the display is lit to show it is charging.
If the problem persists, contact eNeura or your local Agent for help.
Prescription expires in less than 15 days
If the Rx symbol is orange, please contact eNeura or your local Representative to order a new prescription
card. An active prescription is required.
Valid treatment card installed
If the Rx symbol is white, the installed prescription is valid and active.
Invalid treatment card installed
If the Rx symbol is orange and pulsing, the Micro SIM Rx card installed in the device cannot be read by
your device and may be invalid.
Turn off the device, remove and reinstall the Micro SIM Rx card.
If the Rx symbol is still orange and pulsing, contact eNeura or your local Representative for assistance.
Prescription expired
The prescription installed in the device has expired and can no longer be used. Each prescription is valid
for a specific date range based on the prescription issued by your doctor. If you have received a new
Micro SIM Rx card, install it in your device.
If you need a new prescription, contact eNeura or your local Representative for assistance. An active
prescription is required. eNeura can contact your physician if a renewal prescription is required.
Security Lock Enabled
The security lock is enabled and the device is non-operable. To unlock the device and make the device
operable, slide the lock switch to the “off” position.
When enabled, the security lock places the System into Airplane Mode and will prevent wireless
transmissions from the device.
DRAFT 2/12/2019 SPW 23
DRAFT 2/12/2019 SPW 24
Device temperature not in operating range
The device does not operate when its temperature is below 16°C (60°F) or above 40°C (104°F). Move the
device to a more temperate environment, such as inside your home or office, and wait 15 minutes. Try
again.
Precaution: Do NOT actively heat or cool your sTMS mini in any way.
Contact eNeura Customer Care
If this symbol appears with the Prescription low or expired symbol, contact Customer Care (+1-408-245-
6400, option 1) to renew or refill your prescription.
Otherwise, this symbol means service is needed and the device will not deliver treatment. Contact
Customer Care for assistance.
Treatment not allowed (Orange power button)
If this symbol appears with the Prescription Expired, Device Locked, Battery Low, Charger Connected or
other symbols, then the device will prohibit treatment until the other system messages are cleared. Clear
these symbols. Restart the device.
Contact Customer Care for help if other symbols do not appear with the Orange power button.
If treatment is not allowed due to a device failure, an expired or invalid prescription, or temperature out
of range condition, then, after 30 seconds of inactivity, the device will attempt to connect to a cellular
network, transfer data and power off. During data transfer, the power button indicator will pulse blue..
Troubleshooting
None of the device indicator lights are on
The device may not be powered on. Press the power button for at least one second.
The battery may be completely discharged. Plug in the battery charger to charge the battery for at least
15 minutes. Be sure that only the eNeura battery charger (DWG-0505) is being used. When you plug in
the battery charger, the display should light. Check that the battery icon on the display is lit to indicate
that the battery is charging.
If the problem persists, contact eNeura or your local Representative for assistance.
WARNING: There are no user serviceable parts inside. Do not attempt to open the device, as
it could expose user to hazardous voltages. Do not try to modify the unit in any way.
The device is charging correctly, but I can’t make it deliver a treatment.
Check the power button indicator to confirm the device is ready for treatment.
Check the lock switch to make certain it is not locked.
Disconnect the battery charger if it is connected to the device.
The Micro SIM Rx card may not be installed or may have expired. Check the indicator and install a
current Micro SIM Rx card if required.
The device may be too warm or too cold. Check the indicator. The device does not operate when its
temperature is below 15°C (60°F) or above 40°C (104°F). Move the device to a normal indoor
DRAFT 2/12/2019 SPW 25
environment and wait 15 minutes before trying again.
Precaution: Do NOT actively heat or cool your sTMS mini in any way.
Service
For questions about your sTMS mini, please call or email Customer Care at eNeura.
Please have your device available when calling eNeura Customer Care.
When the System is no longer usable, return it to eNeura or your local Representative. Do not dispose of the device.
For questions, contact eNeura or your local Representative for assistance.
Precaution: Only authorized service personnel should service or repair components of the
sTMS mini. There are no user serviceable parts. Improper service, repair or modifications
performed by unauthorized personnel may pose a hazard and void the warranty. Do not heat,
incinerate or dispose of the device.
Return Goods Policy
Contact eNeura or your local Representative for information on product returns and return authorization. Contact
information is available at the beginning of this manual and on the last page.
NOTE: Do not discard the box or packing materials. They will be required if you need to return
this product.
Technical Specifications
Weight:
3.2 lb (1.4 kg)
Size:
8.8 in long x 5.1 in wide x 2.7 in deep
(22.4 cm long x 13 cm wide x 6.9 cm deep)
Electrical Power
Internally powered
(Rechargeable battery)
Battery charger:
Input: 100-240V AC, 400 mA (max), 50/60 Hz
Output: 12V DC, 1.5A
Plug Type:
Class II (ungrounded 2-prong plug)
Rechargeable Battery:
7.2V, 2150 mAh, 15.48 Wh lithium ion battery
Typical battery life is 100 charge cycles
(minimum)
Magnetic Pulse Output
0.9 Tesla (nominal) at 1 cm from the center
of curved surface of the sTMS mini per
treatment
FCC ID: 2AR4G-HL7618RD
The sTMS Mini device complies with FCC
rules.
DRAFT 2/12/2019 SPW 26
Operating Environment
The sTMS mini is a battery-operated device.
The sTMS mini requires a regular home or office electrical outlet for proper battery recharging.
The sTMS mini operates best in typical household and office environments where the temperature is 15°C to 40°C (60°F
to 104°F); relative humidity range of 10% to 90%, non-condensing; an atmospheric pressure range of 700 hPa to 1013
hPA; operating altitude of less than or equal to 3000m.
The sTMS mini is not affected by exposure to dust or lint normally found in household and office environments.
The sTMS mini and its accessories may be exposed to sunlight for a few days duration, but prolonged or repeated
exposure to direct sunlight is not recommended.
It is recommended that a cold device (below 60°F or 15°C) be allowed to warm to room temperature for at least 15
minutes prior to use. Similarly, it is recommended that a hot device (above 104°F or 40°C) be allowed to cool for at
least 15 minutes prior to use.
The sTMS mini is suitable for home use, and complies with IEC 60601-1-11:2015.
Precaution: Do NOT actively heat or cool your sTMS mini in any way.
Storage Environment
The sTMS mini may be stored at -40°C to 60°C (-40°F to 140°F) with a relative humidity of up to 90% non-condensing
and an atmospheric pressure range of 700 hPa to 1060 hPa and storage altitude of less than or equal to 3000m.
Keep your sTMS mini dry during operation, storage and transport. When not in use, store the sTMS to protect it from
household dust and lint, accidental spills or exposure to the elements (rain, dust, insects, vermin, prolonged direct
sunlight, etc).
During extended storage periods, it is recommended that the battery be periodically recharged every 6 months, so that
the device will be available for use when needed.
Industry Standard Classification
(IEC 60601-1)
Classification
Type of Electric Shock Protection
(sTMS mini)
(Battery Charger, external)
Internally powered
Class II
Degree of Electric Shock Protection
Type B applied part
Degree of Protection Against
Harmful Ingress of Liquids and
Particulates
IP22
Safety in the Presence of
Flammable Atmosphere
Not suitable for use in
flammable environment
Mode of Operation
Short-time operation
individual 1ms pulses
EMC
Class B
Service Life (sTMS mini device, AC
Adapter and Rx SIM Card)
1800 treatments
DRAFT 2/12/2019 SPW 27
REACH and Warning Statement
Thresholds and Substances of Very High Concern (SVHCs) are defined in Article 7 and 57 of Regulation (EC) No
1907/2006 (REACH Regulation). This article contains a substance on the Candidate List in a concentration above 0.1%
(w/w), namely: 4-Nonylphenol, branched and linear -substances with a linear and/or branched alkyl chain
with a carbon number of 9 covalently bound in position 4 to phenol, covering also UVCB- and well-defined substances
which include any of the individual isomers or a combination thereof. This substance is completely enclosed in the
device, and does not pose a safety concern during use.
EMC Compliance and Warning Statement
IEC 60601-1-2: 2014
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of
the FCC Rules and with the limits of the standard for medical devices, IEC 60601-1-2, suitable for use in all
environments including domestic. The unit also complies with the requirements of EN 60601-1-2, providing the
presumption of compliance to the European Union’s Medical Device Directive2007/42/EC. The limits are designed to
provide reasonable protection against harmful interference in a typical medical installation. This equipment generates,
uses, and can radiate radio-frequency energy, (such as wireless home networks, mobile phones, cordless telephones
and their base stations, and walkie-talkies), and, if not installed and used in accordance with the manufacturer’s
instructions, may cause harmful interference to other devices in the vicinity. Use of this equipment adjacent to or
stacked with other equipment should be avoided because it could result in improper operation. Portable RF
communications equipment (including peripherals such as antenna cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the sTMS mini, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result. However, there is no guarantee that interference will
not occur in a particular installation. If this equipment causes interference with other devices, which may be
determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
Reorient or relocate the device receiving the interference.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other
device(s) are connected.
Consult the manufacturer or field service technician for help.
DRAFT 2/12/2019 SPW 28
EMC Guidance
In accordance with Clause 4.3 of IEC 60601-1, the Essential Performance of the sTMS Mini is to not deliver a pulse
above 0.95T.
EMC interference from other equipment may disrupt normal operation of the sTMS Mini and may impede the delivery
of the 0.9T magnetic pulse to the patient’s head. If other equipment interferes with normal sTMS Mini operation,
which may be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
Reorient or relocate the sTMS Mini device away from the equipment.
Increase the separation between the sTMS Mini and the equipment.
Connect the equipment into an outlet on a circuit different from that to which the sTMS Mini is
connected.
Consult eNeura Customer Service for assistance.
Guidance and Manufacturer’s Declaration - Emissions
The DWG-0600 is intended for use in the electromagnetic environment specified below. The customer or user of the
DWG-0600 should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF Emissions
CISPR 11
Group 1
The DWG-0600 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class B
The DWG-0600 is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonics
EN 61000-3-2
Class A
Voltage
Fluctuations/Flicker
emissions
Complies
DRAFT 2/12/2019 SPW 29
Guidance and Manufacturer’s Declaration – Immunity
The DWG-0600 is intended for use in the electromagnetic environment specified below. The customer or user of the DWG-0600 should
ensure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output
lines
±2 kV for power
supply lines
±1 kV for
input/output
Lines*1
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Voltage Dips 30%
reduction, 25/30
periods
At 0°
Voltage Dips 30%
reduction, 25/30
periods
At 0°
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
EQUIPMENT requires
continued operation during power
mains interruptions, it is recommended
that the EQUIPMENT be powered from an uninterruptible
power
supply or a battery.
Voltage Dips >
95% reduction, 0.5
period
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
Voltage Dips >
95% reduction, 0.5
period
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
Voltage Dips >
95% reduction, 1
period At 0°
Voltage Dips >
95% reduction, 1
period At 0°
Voltage
Interruptions > 95%
reduction, 250/300
periods
Voltage
Interruptions >
95% reduction,
250/300 periods
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
1 The EUT does not contain signal input/output lines longer than 3 m in length.
DRAFT 2/12/2019 SPW 30
EMC Guidance (continued)
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The DWG-0600 is intended for use in the electromagnetic environment specified below. The customer or user of the
DWG-0600 should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
(6 Vrms in ISM and
amateur radio Bands
within 150kHz 80MHz)
10 V/m
80 MHz to 2.7 GHz
3 Vrms
10 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the
EQUIPMENT, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2P
d = 1.2P 80 MHz to 800 MHz
d = 2.3P 800 MHz to 2.7 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site surveya,
should be less than the compliance level in each
frequency range.b
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the EQUIPMENT is used exceeds the applicable RF
compliance level above, the EQUIPMENT should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the EQUIPMENT.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
DRAFT 2/12/2019 SPW 31
Immunity to RF Wireless Communications Equipment
Test
Frequency
(MHz)
Band a)
(MHz)
Service a)
Modulation b)
Maximum
Power
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385
380 - 390
TETRA 400
Pulse modulation b)
18 Hz
1.8
0.3
27
450
430 - 470
GMRS 460,
FRS 460
FM c)
± 5 kHz deviation
1 kHz sine
2
0.3
28
710
745
780
704 - 787
LTE Band 13,
17
Pulse modulation b)
217 Hz
0.2
0.3
9
810
870
930
800 - 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation b)
18 Hz
2
0.3
28
1720
1845
1970
1700 - 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse modulation b)
217 Hz
2
0.3
28
2450
2400 - 2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation b)
217 Hz
2
0.3
28
5240
5500
5785
5100 - 5800
WLAN 802.11a/n
Pulse modulation b)
217 Hz
0.2
0.3
9
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.
DRAFT 2/12/2019 SPW 32
The device and the method of use that comprise the eNeura sTMS mini are protected by U.S. Patents 6,402,678, 7,294,101, 7,494,458
and 7,601,116; by European Patents 1 307 260 B1 and 1 977 787 B1; and by Canadian Patent 2423840. Additional patents are pending.
Recommended separation distances between
portable and mobile RF communications equipment and the sTMS Mini
The sTMS Mini is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the sTMS Mini can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the sTMS Mini as recommended below, according to the maximum output
power of the communications equipment.
Rated
maximum
output power
of transmitter
(W)
Separation distance according to frequency of transmitted
(m)
150 kHz to 80 MHz
d = 1.2P
80 MHz to 800 MHz
d = 1.2P
800 MHz to 2.7 GHz
d = 2.3P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
DRAFT 2/12/2019 SPW 33
IP22
Glossary of Abbreviations
CE Mark
CE mark certifies that a product has met EU consumer safety, health or environmental
requirements.
Intertek
ETL
Intertek Mark certification marks indicate that this product has been tested to a U.S. standard and
it meets the requirements of an applicable ETL standard or another recognized document used as
a basis for certification. Conforms to AAMI STD ES60601-1, IEC STDS 60601-1-6, 60601-1, 60601-1-
11, 62304 and 62366
EMC
Electromagnetic Compatibility
IEC
International Electro-technical Commission
SN
Serial Number
TMS
Transcranial Magnetic Stimulation, which is a method of using a brief pulse of magnetic energy to
stimulate nerves in the brain
Key to Symbols
CE Mark
Attention: See Instructions
for Use
Intertek Mark
5003910
Risk of strong magnetic
field
Protection against falling liquids
Warning, electricity
Type BF Applied part
Keep Dry
Caution: Federal law (US) restricts
this device to sale by or on the
order of a physician
Rx Only
Federal Communications
Commission
Class B certified for home use
Warning
Authorized representative
in European Community
Temperature limitation
EMC non-ionizing
radiation
Humidity limitation
Smart Card status
Contact eNeura Customer Care
Power button On/Off
Treatment not allowed
Preparing for Treatment
Ready for Treatment
Lock switch enabled
0344
DRAFT 2/12/2019 SPW 34
AC Adapter connected
Rechargeable Battery capacity level
Medical Device Reporting
Any potential adverse incident involving eNeura products should be reported immediately by calling your local Representative
or Customer Care at eNeura.
Users and physicians may also report adverse events to FDA through MedWatch at 1-800-332-1088 or at
https://www.fda.gov/safety/medwatch.
Warranty and Limitation of Liability
eNeura Inc. warrants that the product when delivered is free from defect in materials and workmanship and conforms to the
manufacturer’s product specifications. eNeura shall not be liable for any incidental or consequential loss, damage, or expense,
directly or indirectly arising from the use of, or inability to use, its product.
THE FOREGOING WARRANTY IS IN LIEU OF AND EXCLUDES ALL OTHER WARRANTIES, WHETHER EXPRESSED OR IMPLIED,
ARISING BY OPERATION OF LAW OR OTHERWISE, INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS
FOR A PARTICULAR PURPOSE.
eNeura does not assume, nor authorizes any person to assume for it, any other additional liability or responsibility with respect
to this product other than as set forth in writing herein.
Customer Care Contact Information
Manufactured by: eNeura Inc.
715 North Pastoria Avenue
Sunnyvale, CA 94085 USA
eNeura Inc.
715 North Pastoria
Sunnyvale, CA 94085
Tel: +1 855.366.8355
Fax: +1 877.874.9584
www.eNeura.com
info@eNeura.com
customercare@eNeura.com
eNeura (UK) Ltd.
6th Floor
One London Wall London EC2Y 5EB
United Kingdom
Tel. +44 (0) 20.3695.4063
Fax: +44 (0) 20.7785.8152
www.eNeura.co.uk
customercare@eneura.co.uk
Authorized Representative
Emergo Europe BV
Prinsessegracht 20
2514 AP The Hague
The Netherlands
Tel: +31 (0) 70.345.8570
Fax: +31 (0) 70.346.7299
www.emergoeurope.eu
Australia Sponsor
Emergo Australia
Level 20, Tower II, Darling Park
201 Sussex Street
Sydney, NSW 2000
Australia
www.emergogroup.com
DRAFT 2/12/2019 SPW 35
sTMS mini
© 2018 eNeura Inc.
eNeura® and sTMS mini are trademarks of eNeura Inc.
The device and the method of use that comprise the eNeura sTMS mini are protected by U.S. Patents 6,402,678, 7,294,101,
7,494,458 and 7,601,116; by European Patents 1 307 260 B1 and 1 977 787 B1; and by Canadian Patent 2423840.
Additional patents are pending.
LBL-0103-US Rev F 10/2018