IRay Technology 02112031 Wireless Digital Flat Panel Detector User Manual

iRay Technology (Shanghai) Ltd. Wireless Digital Flat Panel Detector

User manual

Download: IRay Technology 02112031 Wireless Digital Flat Panel Detector User Manual
Mirror Download [FCC.gov]IRay Technology 02112031 Wireless Digital Flat Panel Detector User Manual
Document ID2430284
Application IDGGUKCirWgZapWBfX7VKK9g==
Document DescriptionUser manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize140.62kB (1757758 bits)
Date Submitted2014-10-29 00:00:00
Date Available2014-10-29 00:00:00
Creation Date2014-10-10 15:42:45
Producing SoftwareAcrobat Distiller 9.5.5 (Windows)
Document Lastmod2014-10-13 14:01:34
Document TitleMicrosoft Word - 032-202-01 User Manual-Mars1417V User Manual A2.docx
Document CreatorPScript5.dll Version 5.2.2
Document Author: wei.pan

To customers
Mars1417V serial
Wireless Digital Flat Panel Detector
User Manual
iRay Technology (Shanghai) Ltd.
ADDRESS:2F, Building 7, No. 590, Ruiqing Road,
Zhangjiang East, Pudong New Area,
Shanghai 201201, China
TEL:+86‐21‐50720560
FAX:+86‐21‐50720561
DATE: 2013‐08‐09
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
To customers
To customers
Thank you for purchasing iRay’s Mars1417V serial Medical X-ray
Radiography Flat Panel Detector. This User’s Manual explains the general
usage of the detector and peripheral equipment. Before using this product,
please read this manual thoroughly. Our company will not be responsible for
any abnormalities, malfunction and body harms caused by violations of the
user’s manual.
Important information on usage and management of
equipment
1. Only a physician or a legally certified operator is allowed to use this product.
2. The equipment should be maintained in a safe and operable condition by
maintenance personnel.
3. Use only computers and image display monitors complying with IEC 60601-1 or
IEC 60950-1. For details, consult our sales representative or local iRay dealer.
4. Use only the dedicated cables. Do not use any cables other than those supplied with
this product.
5. Opening the detector by the unauthorized personnel of our company is forbidden.
6. Prevent the liquid or electrical conductive substance entering into the detector.
7. The detector should not be used in circumstance with flammable gas or corrosive
gas.
8. The detector’s radiation protection complies with IEC 60601-1-3.
9. Prevent the detectors from sharp/hard things, or the cover will be broken. If the
painting or the cover is broken, do not contact with surface directly.
10. To pull out the plug is the only way to cut down the main power supply.
11. For product available in the USA/Canada market, only channel 1~11 can be
operated. Selection of other channels is not possible.
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
To customers
Disclaimer
1. In no event shall iRay be liable for any damage or loss arising from fire, earthquake,
any action or accident by a third party, any intentional or negligent action by user,
any trial usage, or other usage under abnormal conditions.
2. Roentgenography, image processing, image reading, and image data storage must
be performed in accordance with the laws of the country or region in which the
product is being used. The user is responsible for maintaining the privacy of image
data.
3. In no event shall iRay be liable for personal physical harm or property damage that
is sustained when the instructions are not followed or the product is misused.
4. It is the responsibility of the attending physicians to provide medical care services.
iRay will not be liable for faulty diagnoses.
5. In no event shall iRay be liable for direct or indirect consequential damages arising
from the use or unavailability of this product. iRay shall not be liable for loss of
image for any reason.
6. In no event shall iRay be liable for any damage arising from moving, alteration,
inspection or repair by a person other than authorized service engineers.
7. Specifications, composition, and appearance of this product may change without
prior notice.
About CE
MARS1417V SERIAL is compliant with the standard of
IEC 60601-1,General requirements for basic safety and essential performance
and IEC 60601-1-2, Electromagnetic Compatibility –Medical Electrical
Instrument.
The detector can connect to public power supply as defined in CISPR 11 (the class
defines GROUP1 Class B).
Please refer to chapter V: Regulatory information for CE testing standard.
For more information, please contact with iRay.
About FCC
FCC Compliance
This device complies with Part 15 of the FCC Rules. Operation is subject to the
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
To customers
following tow conditions:
1. This device may not cause harmful interference.
2. This device must accept any interference received, including interference that may
cause undesired operation.
Caution: Any changes or modifications not expressly approved by the party
responsible for compliance could void the user's authority to operate this equipment.
NOTE: This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses, and can radiate radio frequency energy
and, if not installed and used in accordance with the instruction, may cause harmful
interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If this equipment dose cause
harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encourage to try to correct the
interference by one or more of the following measures;
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
- FCC requires this product to be used indoors for the frequency range 5.15 to 5.25
GHz to reduce the potential for harmful interference to co-channel Mobile Satellite
systems.
Federal Communication Commission (FCC) Radiation Exposure Statement.
This EUT is compliance with SAR for general population/uncontrolled exposure limits
in ANSI/IEEE C95.1-1999 and had been tested in accordance with the measurement
methods and procedures specified in OET Bulletin 65 Supplement C. This equipment
should be installed and operated contact with the radiator & your body.
Note on installation
Please request your sales representative or local iRay dealer to install this product.
Note on disposal of this product
Disposal of this product in an unlawful manner may have a negative impact on
health and environment. Be absolutely sure to follow the procedure which is in
conformity with the laws and regulations applicable in your area when disposing
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
To customers
of this product.
European Union (and EEA) only.
This symbol indicates that this product can not be disposed of with
your household waste, according to the WEEE Directive (2002/96/EC)
and your national law. This product should be handed over to a
designated collection point, e.g., on an authorized one-for-one basis
when you buy a new similar product or an authorized collection site
for recycling waste electrical and electronic equipment (EEE).
Improper handling of this type of waste could have a possible
negative impact on the environment and human health due to
potentially hazardous substances that are generally associated with
EEE. At the same time, your cooperation in the correct disposal of
this product will contribute to fully utilizing natural resources. For
more information about where you can drop off your waste equipment
for recycling, please contact your local city office, waste authority,
approved WEEE scheme or your household waste disposal service.
For more information about return and recycling of WEEE products,
please contact iRay.
About Service Information

Please refer to iRay or iRay’s authorized service personnel only.

iRay or iRay’s authorized agents provide telephone service and site services if it is
determined necessary.

Expense of service and component may be charged to the customer according to
the warranty.

Any unauthorized attempt to repair instrument under warranty voids that warranty.

Client shall provide product’s serial number to obtain service.

Replaced component becomes the manufacturer’s property.

Customers shall be responsible for the instruments damage because of
transportation from customer to iRay or authorized agency.
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
To customers
Symbols
The list of the symbols used in this instrument is as follows:
Symbol for “caution, consult accompanying documents”
For an instrument which has passed the CE test, the CE mark shall be placed on
the instrument’s label followed by the CE number
a. This symbol shall be accompanied by the manufacturer’s serial number. The
serial number shall be after, below or adjacent to the symbol.
b. The relative size of the symbol and the serial number are not specified.
c. So far our product serial number is made of thirteen digits:
A1A2A3A4
C1C2 M DD Y XXX
Numerical Order
Year
Date
Month
Version
Product Code
a. The symbol shall be accompanied by the name and address of the
manufacturer,adjacent to the symbol. The address is not required with the
symbol on an immediate container as specified in EN375 AND EN376,
except the immediate container is also the outer container.
b. The relative size of the symbol ,the name and address is not specified.
a. The symbol shall be accompanied by the name and address of the
authorized representative in the European Community, adjacent to the
symbol. The address is not required with the symbol on an immediate
container as specified in EN375 AND EN376, except the immediate
container is also the outer container.
b. The relative size of the symbol, the name and address is not specified.
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
To customers
Symbol for “consult instructions for use”, “consult operation instructions”.
Safety Signs: Symbol for “refer to instruction manual”
Safety Signs:Symbol for “Dangerous Voltage”
“ON” ( power )
“OFF” ( power )
Protective Earth Conductor(ground)
This Mark indicates that this equipment must be handled with care
Do not jolt or apply excessive load to the equipment
The equipment shall be operated during the upper and lower limitsof
temperature
Fragile, the package should be handledwith care, package symbol
Keep awayfrom sunlight, package symbol
Keep dry, package symbol
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
To customers
The humidity limitationshall be indicated adjacentto the upper horizontal
lines, package symbol
The storage temperatureshall be indicated adjacentto the upper horizontal
lines, package symbol
Indicates correctupright position ofthe transportpackage, package symbol
Transport packageshall not be rolled, package symbol
Stacking limit by number,package symbol
Trademarks

The iRay name and iRay logo are registered trademarks of Shanghai iRay
Technology Ltd.
Copyright

All rights reserved.

Under copyright laws, this manual may not be reproduced, in whole or in part,
without the written permission of iRay.
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Contents
Contents
CONTENTS .......................................................................................................................................9
SAFETY NOTICES ............................................................................................................................ 11
1.
CHAPTER I SAFETY INFORMATION ........................................................................................... 12
1.1 SAFETY PRECAUTIONS ......................................................................................................................... 12
1.2 NOTES FOR USING THE EQUIPMENT ....................................................................................................... 16
2.
CHAPTER II GENERAL DESCRIPTION ......................................................................................... 20
2.1SCOPE .............................................................................................................................................. 20
2.2MODEL ............................................................................................................................................ 21
2.3CHARACTERISTIC ................................................................................................................................ 21
2.4INTENDED USE ................................................................................................................................... 22
2.5STANDARD CONFIGURATION.................................................................................................................. 22
2.6 STANDARD PACKAGE LIST .................................................................................................................... 23
2.7 MAIN SPECIFICATIONS ........................................................................................................................ 25
2.8 DESCRIPTION OF INDICATORS ............................................................................................................... 28
2.9 IMAGE DIRECTION ............................................................................................................................. 30
3. CHAPTER III INSTALLATION AND OPERATION .............................................................................. 31
3.1QUICK START ..................................................................................................................................... 31
3.2FIRST IMAGE ACQUISITION ................................................................................................................... 32
3.3 OPERATION NOTES ............................................................................................................................ 32
4. CHAPTER IV CONTROL BOX INTERFACE ....................................................................................... 35
4.1AC INPUT ......................................................................................................................................... 36
4.2 SERIAL PORT ..................................................................................................................................... 36
4.3GIGABIT ETHERNET PORT ..................................................................................................................... 36
4.4HVG INTERFACE................................................................................................................................. 37
4.5 INNERTRIGGERMODULE ..................................................................................................................... 38
4.6 X‐RAY IMAGE ACQUISITION.................................................................................................................. 38
5. CHAPTER V REGULATORY INFORMATION.................................................................................... 45
5.1 MEDICAL EQUIPMENT SAFETY STANDARDS .............................................................................................. 45
5.2 GUIDANCE AND MANUFACTURE’S DECLARATION FOR EMC ........................................................................ 46
6. CHAPTER VI IMAGINGSOFTWARESDK ......................................................................................... 51
7. CHAPTER VII SERVICE INFORMATION ......................................................................................... 52
PRODUCT LIFETIME ............................................................................................................... 52
REGULAR INSPECTION AND MAINTENANCE..................................................................... 52
REPAIR ....................................................................................................................................... 52
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Contents
REPLACEMENT PARTS SUPPORT .......................................................................................... 52
APPENDIX A INSTALLATION OF SDK ................................................................................................ 54
APPENDIX B DESCRIPTION OF DIFFERENT CABLE SERIES ................................................................. 57
APPENDIX C DIFFERENT VERSION OF CONTROL BOX ...................................................................... 58
APPENDIX DRECOMMENDED COMPUTER PLATFORM .................................................................... 60
APPENDIX EFIRMWARE UPDATE .................................................................................................... 62
APPENDIX F DUMMY LINES ........................................................................................................... 68
APPENDIX GDEFINITION OF DEFECTS ............................................................................................. 70
APPENDIX H INFORMATION OF SERVICE OFFICE ............................................................................. 71
APPENDIX I INFORMATION OF MANUFACTURES ............................................................................ 72
10
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Safety notices
Safety notices
The following safety notices are used to emphasize certain safety instructions. This
manual uses the caution symbol along with a caution message.
This notice is used to identify conditions under which improper
use of the product may cause death or serious personal injury.
This notice is used to identify conditions under which improper
use of the product may cause minor personal injury.
This notice is used to identify conditions under which improper
use of the product may cause property damage.
This is used to indicate a prohibited operation.
This is used to indicate an action that must be performed.
This is used to indicate important operations and restrictions. Be
sure to read this notice to prevent property damage or malfunction.
This is used to indicate operations for reference and
complementary information. Users are recommended to read this
notice.
11
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter I Safety Information
1.
Chapter I Safety Information
1.1 Safety precautions
Follow these safeguards and properly use the equipment to prevent injury and damage to any
equipment/data.
WARNING
Installation and environment of use

Do not use or store the equipment near flammable chemicals such as alcohol, thinner, benzene, etc.
If chemicals are spilled or evaporate, it may result in fire or electric shock through contact with electric
parts inside the equipment. Also, some disinfectants are flammable. Be sure to take care when using them.

Do not connect the equipment with anything other than specified.
Doing so may result in fire or electric shock.
Power supply

Do not operate the equipment using any type of power supply other than the one indicated on the
rating label.
Otherwise, it may result in fire or electric shock.

Do not handle the equipment with wet hands.
You may experience electric shock that could result in death or serious injury.

Do not place heavy object such as medical equipment on cables and cords. Do not pull, bend, bundle,
or step on them to prevent their sheath from being damaged, and do not alter them neither.
Doing so may damage the cords which could result in fire or electric shock.

Do not supply power to more than one piece of equipment using the same AC outlet.
Doing so may result in fire or electric shock.

Do not turn ON the system power when condensation has formed on the equipment.
Doing so may result in fire or electric shock.

Do not connect a multiple portable socket-outlet or extension cord to the system.
Doing so may result in fire or electric shock.

To avoid the risk of electric shock, this equipment must only be connected to power supply with
protective earth.
Not doing so may result in fire or electric shock.
12
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter I Safety Information

Securely plug the power cord into the AC outlet.
If contact failure occurs, or if metal objects come into contact with the exposed metal prongs of the plug,
fire or electric shock may result.

Be sure to turn OFF the power to each piece of equipment before connecting or disconnecting the
cords.
Otherwise, you may get an electric shock that could result in death or serious injury.

Be sure to hold the plug or connector to disconnect the cord.
If you pull the cord, the core wire may be damaged, resulting in fire or electric shock.
If the detector is intended to assemble mobile DR system, the sensor cable and Detector cable must be
ordered no longer than4m, and the cable must be collected in the fixed position, not doing this may cause
the DR system be toppled.
13
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter I Safety Information
WARNING
Handling

Never disassemble or modify the equipment. No modification of this equipment is allowed
Doing so may result in fire or electric shock. Also, since the equipment incorporates parts that may cause
electric shock as well as other hazardous parts, touching them may cause death or serious injury.

Do not place anything on top of the equipment.
The object may fall and cause an injury. Also, if metal objects such as needles or clips fall into the
equipment, or if liquid is spilled, it may result in fire or electric shock.

Do not hit or drop the equipment.
The equipment may be damaged if it receives a strong jolt, which may result in fire or electric shock if
the equipment is used without being repaired.

Do not put the equipment and pointed objects together.
The equipment may be damaged. If so, the equipment should be used in bucky.

Have the patient take a fixed posture and do not let the patient touch parts unnecessarily.
If the patient touches connectors or switches, it may result in electric shock or malfunction of the
equipment.
When a problem occurs

Should any of the following occurs, immediately turn OFF the power to each piece of equipment,
unplug the power cord from the AC outlet, and contact your sales representative or local iRay
dealer:

When there is smoke, an odd smell or abnormal sound

When liquid has been spilled into the equipment or a metal object has entered through an opening

When the equipment has been dropped and damaged
Maintenance and inspection

Please turn OFF the power of the equipment and unplug the power cord from the AC outlet before
cleaning.

NEVER use alcohol, ether and other flammable cleaning agent for safety. NEVER use methanol,
benzene, acid and base because they will erode the equipment.

DON’T dip the equipment into the liquid.

Please make sure that the equipment’s surface & plugs are dry before turning ON.
Otherwise, it may result in fire or electric shock.

Clean the plug of the power cord periodically by unplugging it from the AC outlet and removing
dust or dirt from the plug, its periphery and AC outlet with a dry cloth.

If the cord is kept plugged in for a long time in a dusty, humid or sooty place, dust around the plug
will attract moisture; this could cause insulation failure that may result in a fire.
For safety reasons, be sure to turn OFF the power to each piece of equipment when the performing
inspections indicated in this manual.
Otherwise, electric shocks may occur.
14
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter I Safety Information
CAUTION
Installation and environment of use

Do not install the equipment in any of the locations listed below. Doing so may result in failure,
malfunction, equipment falling, fire or injury.



Close to facilities where water is used

Where it will be exposed to direct sunlight

Close to the air outlet of an air-conditioner or ventilation equipment

Close to heat source such as a heater

Where the power supply is unstable

In a dusty environment

In a saline or sulfurous environment

Where temperature or humidity is high

Where there is freezing or condensation

In areas prone to vibration

On an incline or in an unstable area
Because the equipment cable is long, take care that cables do not become tangled during use. Also,
be careful not to get your feet caught in the cable.
Otherwise, it may cause a malfunction of the equipment or the injury of the user due to tripping over
the cable.
Non-medical equipment such as the battery charger, access point and IR data communication unit
cannot be used in patient’s vicinity.
Power supply

Always connect the three-core power cord plug to a grounded AC power outlet.

To make it easy to disconnect the plug at any time, avoid putting any obstacles near the outlet.
Otherwise, it may not be possible to disconnect the plug in an emergency.

Be sure to ground the equipment to an indoor grounded connector. Also, be sure to connect all the
grounds for the system to a common ground.

Do not use any power source other than the one provided with this equipment.
Otherwise, fire or electric shock may be caused due to leakage.
Handling

Do not spill liquid or chemicals onto the equipment. In case the patient is injured, it is not allowed
to contact with blood or other body fluids.

Doing so may result in fire or electric shock.
In such a situation, protect the equipment with a disposable cover as necessary.
Turn OFF the power and pull out the plug to each piece of equipment for safety when not used.
15
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter I Safety Information
CAUTION
Handling

Handle the equipment carefully.

Do not submerge the equipment in water.

The internal image sensor may be damaged if something hits against it, or if it is dropped. If the
equipment is dropped, the drop sensor will turn red and the equipment will not be warranted by
iRay.



Do not place excessive weight on the detector.
Otherwise, the internal image sensor may be damaged and image may be incorrect.
Be sure to use the detector on a flat surface so it will not bend. Otherwise, the internal image sensor
may be damaged. Be sure to securely hold the detector while using it in upright positions.
Otherwise, the detector may fall over, resulting in injury to the user or patient, or may flip over,
resulting in damage to the inner device.
The detector should be used in the bucky or holder to keep the same load (same pressure) on the
detector when acquiring the image. Or the image will be incorrect.
1.2 Notes for using the equipment
When using the equipment, take the following precautions. Otherwise, problems may occur and the
equipment may not function correctly.
Before exposure

Be sure to check the equipment daily and confirm that it works properly.

Check the cable MPN No. & control box version.

Sudden heating of the room in cold areas will cause condensation to form on the equipment.
In this case, wait until the condensation evaporates before performing an exposure. If the
16
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter I Safety Information
equipment is used while condensation is formed on it, problems may occur in the quality of
captured images. When an air-conditioner is used, be sure to raise/lower the temperature
gradually so that a difference of temperature in the room and equipment does not occur, to
prevent condensation.

The detector should warm up for 25 minutes before exposure or updating the gain map or
defect map.
During exposure

Do not move the power or Ethernet Cables during exposure, or it may cause image noise or
artifacts, even incorrect images.

Do not use the devices near the equipment generating a strong magnetic field. Otherwise, it
may cause image noise, artifacts or even incorrect images.
Disinfection and Cleaning

After every examination, wipe the patient contact surfaces of the detector using disinfectants
such as ethanol, to prevent the risk of infection. For details on how to sterilize, consult a
specialist.

Do not spray the detector directly with disinfectants or detergents.

Wipe it with a cloth slightly damped with a neutral detergent. Do not use solvents such as
alcohol, thinner, benzene, acid and base. Doing so may damage the surface of the equipment.

It’s recommended to use a waterproof non-woven cover as the isolated layer between detector
and the blooding patient.
17
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter I Safety Information
Replace Cables


Turn OFF the power of the equipment and unplug the power cord from the AC outlet before
operation. Unplug the detector cable from the float outlet, or it may result in fire or electric
shock.
Eliminate the static before replacing cable, including operating platform, tools and operator,
or ESD may damage the detector.
Operating/storage environment
Be sure to use and store this equipment under the conditions described below:
Temperature
Operating
5~30℃
Storage
‐20~60℃

Temperature
change
<1k/min
Humidity
Atmospheric
Pressure
10~70% RH
700~1000hPa1
0~70% RH
Pressure Change
<10kp/min
(1kp=1.0197E‐5Pa)
≤93% RH
Do not expose this equipment to high temperatures and/or high humidity. Malfunctions may
occur.

Keep air flow around the equipment, especially when the environment temperature is above
30℃.

When the environment temperature is above 33℃ or running for a long time, the equipment
should be used in a holder or a bucky. The surface temperature may be above 41℃.

When not in use, keep the detector and grid in a designated and safe location.

The labeling for Packaging marked with special handling instructions for transport and
storage:

The product should be storage or transport under the temperature from -20~60℃, the
humidity under 93% RH.
18
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter I Safety Information
19
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
2.
Chapter II General Description
The Mars1417V-PSI is the cassette-size (digital radiographic) X-ray flat panel detector based on
amorphous silicon thin-film transistor technologies. It is designed to provide the highest quality of
radiographic image, which contains an active matrix of2304×2800 pixels with 150μm pitch.Users can
select the type of scintillator such as Standard GOS(Gadolinium Sulfoxylate), DRZ Plus, or CsI
(Cesium Iodide).The data interface is Gigabit Ethernet which can afford high data transfer rates. In
addition, Mars1417V-PSI is equipped with iRay’s unique sync-shot function to make the hardware
connection between the equipment and HVG (High Voltage Generator)unnecessary.
Along with the equipment, iRay also provides a Microsoft Visual Studio (2008) library
®
IrayAcquireLib.lib, and a demo program iDemo with source code to assist users’ integration of the
equipment into their existing DR system. The library IrayAcquireLib.lib contains functions for image
®
acquisition and equipment configuration. iDemo demonstrates the usage of IrayAcquireLib.lib and
provides a simple user interface to acquire and preview X-ray images.
2.1Scope
This manual contains information about the iRay Mars1417V. All operators must read and understand
this manual before using the equipment. All information in this manual, including the illustrations, is
based on equipment prototype. If configuration of your equipment does not have any of these items,
information about these items in the manual does not apply to your equipment.
20
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
2.2Model
Mars
□
□ - □
Scintillator type:PSI, Gadolinium Sulfoxylate
product use:V,cassette and portable detector
product dimension:1417,14inch×17inch
product model:wireless digital flat panel detector
Specification of Mars1417V serial Flat Panel Detector:
No.
Model
Brand Name
Type of Use
Mars1417V-PSI
Mars1417V-PSI
Cassette and
portable
Active
area(cm2)
34.3×41.6
Pixel matrix
Pixel Pitch(μm)
2304×2800
150
2.3Characteristic

Fixed static Flat Panel Detector used for general radiography.

Sync-Shot exposure trigger

Gigabit Ethernet

DRZ Plus or CsI scintillation screen.

Easy to change the cable and upgrade firmware.
21
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
2.4Intended use
This equipment provides digital X-ray imaging for diagnosis of disease, injury, or any applicable health
problem. The image is obtained as the result of X-ray passing through the human body and detected
by the equipment. This device is intended to be used in the holder or bucky which is well insulated to
the detector. Or the holder or the bucky is well grounded. This device is not intended for directly
touching the person.
iRay will provide equipment and software support for integration of system. The total length of
Detector cable and Extension cable cannot exceed 9 m, or the impedance of protective earth
connections may exceed the safety threshold.
This device is not intended for mammography or dental applications,and prohibited for pregnant
women and children
2.5Standard configuration
The Mars1417Vcomes with a power supply which connects to 100-240 AC outlets and generates DC
voltages needed by the equipment.
Mars1417V sub types ending with suffix -I are equipped with internal X-ray sensors which
automatically detect the X-ray and synchronize image acquisition with the X-ray exposure, therefore,
they do not need to be connected to high voltage generator of the user’s DR system. Mars1417V
subtypes ending with the suffix -M are not equipped with any internal X-ray sensor and need to be
connected to the user’s high voltage generator through a supplied cable for synchronization between
image acquisition and X-ray exposure.
The equipment shall be connected to a computer through the Ethernet cable for data transfer. There is
also a serial cable connecting the detector to the computer for configuration and image acquisition
commands.
22
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
2.6 Standard Package List
Item
Qty
Detector:
Mars1417V
(with detector cable)
Medical Adaptor:
(Control box is specified as a part of ME
Equipment)
Battery:
Wireless AP
LAN Cable
Ethernet Cable
Power Cable
23
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
Power convert cable
24
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
2.7 Main Specifications
2.7.1
Useful entrance field size
All the iRay’s detectors should meet the following requirements; the useful entrance field size should
be more than 95% nominal entrance field size.
Table 2.7.1 Entrance field size
Model
Nominal entrance field size(cm)
Useful entrance field size(cm)
Mars1417V serial
35 × 43
34.3×41.6
2.7.2
Minimum detectable dose should be less than100nGy.
2.7.3
Maximum linearity dose should be more than 60μGy,and the correlation coefficient(r²)
should be not less than 0.99.
2.7.4
Limiting spatial resolution should be not less than2.8Lp/mm(Standard GOS)/ 3.1Lp/mm(DRZ
Plus or CsI).
2.7.5
Low‐contrast resolution should be more than 0.0055(with the air kerma should be 50μGy
25
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
in the test process).
2.7.6
Defects
Defects
Full Class 0
Full Class 1
Full Class 2
Full Class 3
Full Class 4
Full Class 5
Full Class 6
Full Class 7
Full Class 8
Meta 7+8 Class 1
Meta 7+8 Class 2
Meta 7+8 Class 3
Meta 7+8 Class 4
Meta 7+8 Class 5
Meta 7+8 Class 6
Meta 7+8 Class 7
Meta 7+8 Class 8
Single Horizontal Defect Lines
Single Vertical Defect Lines
Double Horizontal Defect Lines
Double Vertical Defect Lines
Double Vertical Defect Lines in Central Area
Triple Vertical or Horizontal Lines
Min Gap between Vertical Defect Lines/Line Pairs
Min Gap between Horizontal Defect Lines/Line Pairs
2.7.7
Flat uniformity should be not exceed 2.2%
2.7.8
Modulation transfer function[MTF(u,v)]
Max Number Allowed
13500
1500
1050
300
150
120
10
10
The MTF should meet the following table
Standard radiation
quality No.RQA/μGy
Spatial frequency (lp/mm)
MTF (DRZ‐Plus)
MTF(CSI)
≥0.65
≥0.77
≥0.35
≥0.58
5/2.5
0.5
1.0
1.5
2.0
2.5
3.3
≥0.20
≥0.39
≥0.10
≥0.25
≥0.05
≥0.15
≥0.01
≥0.10
26
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
2.7.9
Detective quantum efficiency(DQE)
The DQE should meet the following table
Standard radiation
quality No.RQA/μGy
5/2.5
Spatial frequency(lp/mm)
DQE (DRZ‐Plus)
DQE (CSI)
≥0.20
≥0.58
0.5
≥0.15
≥0.50
1.0
≥0.10
≥0.40
2.0
≥0.02
≥0.21
2.5
≥0.01
≥0.14
2.7.10 Lag effect should be less than 3%.
2.7.11 adaptor (Control Box is specified as a part of ME equipment)
Power: Input:
110-240V~ 50-60Hz
Output: 12V
1A
Rated Input: 110VA
27
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
2.8 Description of Indicators
Fig. 2.8.1 Detector Indicators
Table 2.8.1 Description of Detector Indicators
Items
Power
Status
Color
Descriptions
OFF
Power OFF
Green(steady)
Power supply is normal
OFF
Power OFF or Idle
Yellow(steady)
Power up, load firmware or there exists error
Green(blinking)
Data transmission between equipment and workstation
28
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
Link
OFF
Power OFF
Green(steady)
Gigabit Ethernet link is normal
29
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter II General Description
2.9 Image Direction
Please notice that the image direction is defined as follows:
30
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter III Installation and Operation
3. Chapter III Installation and Operation

The following is a general installation guide. The interface & operation may be different
between different firmware and software editions.
Please contact iRay service office for the
details.

Before installation, please check the MPN (Manufactory Product Number) of the equipment
cable and the version of the control box. Please see Appendix B and Appendix C.

Do not place the detector/control box/cable near the AC-DC adapter, the electromagnet or the
RF equipment. Please find details from chapter 5.2.

iDemo 1.0.6.1(SDK 1.0.6.1) or above supports Software Auto-Clear Mode, when the mode is
enabled, please note the status message. Commands should be sent only when it is “Exposure
Enabled” or “Ready”, if it is “Exposure Prohibit (Auto-Clearing), the commands will be
ignored if they are sent. Please find details from SDK Manual and iRayDR User Guide.

The operator should have the experience of computer operation and X-ray diagnostic system
operation. Before using the product, the operation should be trained by iRay’s FAE or iRay’s
authorized engineers.
The equipment should be used with the specified power supply.
1. Turn on the computer.
2. Install a Gigabit Ethernet card in the computer if not installed.
3. Install WinPcap 4.1.2 for the Gigabit Ethernet interface.
4. Connect the power supply to a 110-240V AC outlet.
5. Connect the extension cable to the equipment.
6. Connect control box to PC and high voltage generator with interface cable(serial cable and
Ethernet cable).
7. Turn on power supply and the green indicator of the detector shall be lit up.
8. Run one image acquisition software such as iDemo.
3.1Quick Start
Before using the equipment, the user shall read the appendix A for details of SDK installation.
31
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter III Installation and Operation
®
Then the user can use the Quick Start software iDemo for image acquisition.
3.2First Image Acquisition
iRay provides a Microsoft Visual Studio (2008) library IrayAcquireLib.lib and a demo program
®
iDemo with source code to assist the users’ integration of the equipment into their existing DR system.
The library IrayAcquireLib.lib contains functions of image acquisition and detector configuration.
iDemo demonstrates the usage of IrayAcquireLib.lib and provides a simple user interface to acquire
and preview X-ray images.
The user can start the iDemo by double clicking on the executable file iDemo.exe included in the
CD ROM. iDemo loads the acquisition driver and sends commands through Gigabit Ethernet port. The
image data acquired by Mars1417V serial will be sent through the Gigabit Ethernet cable to the
computer. iDemo can perform offset, gain and defect correction on the image and then display the
image on screen.
After setting up the FPD and running iDemo, Please find the details from the SDK Manual or
contact with iRay Service Office.
3.3 Operation Notes
The following notes should be noticed before using the detector.

Basic Configuration Setting
There are some basic parameters on the “Config” Dialog when click the “Config” button on the
main window of the iDemo. The parameters should be set and checked as following, or the
image will be incorrect or no images will be gotten.
Parameters
Set Delay
Description
Comment
The time between the end of Clear
Range from 0 to 9500
process and the beginning of the Acquire
and should be less than
process of the “PREP Acquire”
Exp Window-200
Command.
32
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
Acquire Delay The time between the end of the trigger
Range from 0 to 9500
signal and the beginning of the Acquire
and should be less than
process of the “Acquire” command.
Exp Window-200
It’s recommended to keep the
default value.
Exp Window
The exposure window.
Range from 100 to 10000
Trigger
To set the mode of the trigger
Should be checked before
installation
Please keep the other parameters as the default. Or contact with the service office of iRay.

Acquisition
and
post-offset calibration
Post-offset calibration needs the interval from the end of “Clear” and the beginning of the
“Acquire”. So with post-offset calibration, be sure to send “Clear” command before “Acquire” or
it’ll take long time to get the image or no images will be gotten.
33
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface

Gigabit Ethernet Connection
The detector can set stable connection with the Ethernet Card from Intel or Broadcom.
The
computer should meet the minimum requirements of the hardware and software. Please find the
details from Appendix D If there is a connection problem, please check the hardware settings,
including connector, cable or the computer.
Re-plugging the Ethernet cable may be helpful to
set the connection.

To acquire a normal clinical diagnostic image
The clinical diagnostic image does not contain the defect pixel, defect data line and any other
image defect. Human tissue should be fit for the requirements of clinical diagnosis, the image
should be clearly and identifiable.
34
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
4. Chapter IV Control Box Interface
Control box is the relay device between computer and equipment. It is specified as a part of ME
equipment. As shown in the figure, there are 1 connection ports & 4 LEDs at the front panel and 6
connection ports at the back panel of the control box. Some of these interfaces may not provide the
function in some versions, please contact iRay service office. All the plug and connector should be
placed in a portable position. To avoid the connection error, all the plug and connector are designed to
insert error hardly.

AC Input: Three-phase (electric) outlet for AC110-240V power input.

LAN: Gigabit Ethernet port connected to PC.

RS232: Reserved.

HVG: Control interface connected to high voltage generator.

FS : Foot Switch interface

DETECTOR: Integrated interface with power and data connected to Equipment.

MS: Hand Switch (Manual Switch) interface.

Status indicators:

POWER: indicate power status

READY: indicate equipment is ready for X-ray exposure

EXPOSURE: indicate there is X-ray shot

ERROR: indicate power up or there exists error
Please find the details from §2.8 Description of Indicators
35
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
4.1AC Input
AC input shall be connected to AC 100-240V three-phase power supply. Be careful that the AC socket
must have right earth connection terminals. The control box shall conform to IEC 60601-1.
4.2 Serial Port
The serial port is reserved for firmware updating.
4.3Gigabit Ethernet Port
The GBE port shall be connected with an Ethernet cable to a Gigabit Ethernet
Port of the user’s computer. The GBE port transmits image data through Gigabit
Ethernet protocol. For NIC (Network Interface Card), Intel Pro EXP9301CT
PRO Gigabit Network Adapter with PCIe interface is recommended. If the user
prefers any other Gigabit Network Adapter, only those with PCIe interface
should be considered.
36
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
4.4HVG Interface
The HVG interface with 16 pins is to synchronize HVG with the equipment.
The descriptions of HVG interface & HVG Cable
Pin
No.
Shield
NC
White
Syn_OUT_N
Red/white
Xray_On_N/
Syn_IN_N
Reserved for dynamic mode, the negative
part of trigger signal provided by the HVG.
Wire Color
Name
I/O
Signal Description
Not Connected
Reserved for dynamic mode, the negative
part of trigger signal provided by the
equipment.
Black
SYN_OUT_P
Reserved for dynamic mode, the positive
part of trigger signal provided by the
equipment.
Green
Reserved
Please left float
Gray
Expose_OK_N
Negative part of ready signal for high
voltage generator. When the signal is
active, it indicates that the equipment is
ready for exposure.
Brown
Reserved
Please left float
Brown/
white
Expose_Ok_P
Blue
Frame_Request_P
10
EXT_VOUT
Frame_Request_N
12
Yellow
Black/
white
Pink
13
Red
EXT_GND
Xray_ON_P
/Syn_IN_P
14
Cyan
G_SW_XON
15
Purple
G_SW_EN
16
Orange
G_SW_COM
11
37
Positive part of ready signal for high
voltage generator. When the signal is
active, it indicates that the equipment is
ready for exposure.
Positive part of request provided by the
HVG. Ask for preparation of the
equipment.
+5V power supply
Negative part of request provided by the
HVG. Ask for preparation of the detector.
Ground of EXT_VOUT
Reserved for dynamic mode, positive part
of trigger signal provided by the HVG.
Reserved for hand switch or foot switch,
relay pin for X-ray exposure
Reserved for hand switch or foots witch,
relay pin for X-ray preparation
Reserved for hand switch or foot switch,
relay common pin
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
Table 4.4.1 DC characteristic of HVG interface
Input(P‐N)
Output(P‐N)
Power Output
Ityp(mA)
Imax(mA)
Umin(V)
Utyp(V)
Umax(V)
‐16
‐1
‐25
‐10
1000
4.5
6.5
6.5
5.5
Urev_max(V)
4.5 InnerTriggerModule
The detector has anembedded well-proved Inner Trigger module consisting of X-ray sensor and
inner trigger board.
It can simplify the connection with the HVG.The energy range is from 50kV to
150kV.
The voltage and current of tube should be set properly to get sufficient dose rate.
Otherwise, no images will be gotten after the exposure and another “Acquire” command should be
sent to get the image back.
4.6 X‐ray Image Acquisition
Mars1417V serial provides carefully designed functions and commands for the terminal users to
acquire X-ray images under various environments. The users need to select one of three acquisition
trigger modes: Prep-trigger mode, Inner-trigger mode and Soft-trigger mode.
The trigger mode must be set correct before image acquisition; otherwise the FPDs
may not work properly.
38
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
Fig 4.6.1 Basic Connection
4.6.1 Prep-trigger mode
Before connecting to the HVG (High Voltage Generator), please check the
control box version. The HVG connection is different between two versions, but both of them
can support Prep-Trigger mode .Please see Appendix C or find the details from iRay service
office. The Following is for V1.
In the Prep-trigger mode, HVG (High Voltage Generator) needs to be connected to the control box.
The whole acquisition process (including prepare, exposure, acquisition and transmission) is trigged
by the request from HVG. If the process is started, the images will be acquired and transmitted to the
Ethernet port, no matter X-ray is fired or not.

System installation
Control box provides 2pairs of signals and 1 pair of power (Fig. 4.6.2).
39
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
For Prep-trigger mode, only Frame_Request,Expose_OK are needed. To work properly some
pull-up/pull-down circuits may be necessary for different HVG, please find details from your HVG
manual. The control box also has output power supply for pull-up/pull down circuits.
Fig. 4.6.2 HVG Port of Control Box
The Frame_request is low level valid.

X-ray image acquisition timing
In Prep-trigger mode, the acquisition process is triggered by the Frame_Request from the HVG
when there is no “Exposure Prohibited (auto clear)” shown in software status box. And
equipment will start preparing for exposure and sending the message “Exposure Prohibited” to the
workstation to indicate that the X-ray should not be fired before the equipment is ready.
After the equipment is ready, it sends the message “Exposure Enabled” to the workstation to
indicate that the detector is ready for exposure,and the signal “Exposure_OK” is given to HVG
at the same time.
40
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
When the detector is ready, it starts the “prep trigger window”. X-ray should be fired in the “prep
trigger window” to get the proper images. After “preptrigger window”, an image will be acquired
no matter the X-ray is fired or not.
If the “post-offset” is enabled, DO NOT trigger another exposure or
acquisition before the previous image is on screen.
Fig. 4.6.3 Prep‐Trigger Timing
If the “post-offset” is enabled, the dark image acquisition will start automatically after the light
image acquisition. The dark image provides the offset calibration information to workstation.
The calibrated image will be shown when workstation receives light image and dark image, or
there may be no image shown.
Table 4.6.2
Typical Value of Mars1417V serial in prep‐trigger
mode (unit:sec)
T1
Prepare time of equipment.
0.5
T2
Interval between end of prep‐trigger window
and beginning of acquisition process.
T4
Data acquire and transmission time of
detector
0.7
T5
Interval between the end of light image
acquisition and the beginning of dark image
acquisition.
0.3
T7
Prep trigger window
Set Delay
Time
TA2
Cycle time of the image acquisition when
“post‐offset” is enabled
41
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
4.6.2 Inner-trigger mode,

System installation
In Inner-trigger mode, the acquisition is triggered by sync-shot module in the equipment. The
connection between HVG and control box is unnecessary.

X-ray image acquisition timing
The acquisition is triggered by the “Prep” (“Prepare”) command provided by the workstation
when there is no “Exposure Prohibited (auto clear)” shown in software status box. Then the
equipment will start preparing and feedbacking the message “Exposure Prohibited” to the
workstation to indicate that X-ray should not be fired before the equipment is ready.
When the equipment is ready, it starts “inner trigger window” and feedbacks the message
“Exposure Enabled” to the workstation to indicate that the equipment is ready for exposure.
To get proper image, X-ray should be fired in the “inner triggerwindow” (see fig.4.6.4),or
equipment gives the message“Expired & Exposure Prohibited” to indicate that no image is
acquired and the equipment is not ready for X-ray.(see fig. 4.6.5).
If the “post-offset” is enabled, DO NOT trigger another exposure or
acquisition before the previous image is on screen.
If the “post-offset” is enabled, the dark image acquisition will start automatically after the light
image received. The dark image provides the offset calibration information to workstation.
The
calibrated image will be shown when workstation receives light image and dark image, or there
may be no image shown.
Fig. 4.6.4 Timing of Inner Trigger Mode: Normal Operation
Fig.4.6.5 Timing of Inner Trigger Mode: No Exposure
42
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
Table 4.6.3
Typical Value of Mars1417V serial in inner‐trigger
mode (unit:sec)
T1
Prepare time of equipment.
0.5
T2
Interval between end of prep‐trigger window
and beginning of acquisition process.
T3
Inner‐trigger window
<10
T4
Data acquire and transmission time of
detector
0.7
T5
Interval between of the end of light image
acquisition and the beginning of dark image
acquisition.
0.3
TA1
Cycle time of the image acquisition when
“post‐offset” is enabled
2.7+2T3
4.6.3 Soft-trigger mode

System installation
In soft-trigger mode, the acquisition is triggered by the “PREP Acquire” command provided by
workstation when there is no “Exposure Prohibited (auto clear)” shown in software status
box. The connection between HVG and equipment is unnecessary.

X-ray image acquisition timing
In soft-trigger mode, equipment starts preparation when receiving “PREP Acquire” command
provided by workstation. And the message “Exposure Prohibited” will be sent to the workstation
to indicate that X-ray should not be fired before the equipment is ready.
After the equipment is ready, it sends the message “Exposure Enabled” to indicate that equipment
is ready for exposure. To acquire X-ray image (light image), X-ray should be fired(see
fig.4.6.6).Please note that the image acquisition is just triggered by the “PREP Acquire” command,
so an image will be acquired if the detector gets the “PREP Acquire” command no matter X-ray
is fired or not.
If “post-offset” is enabled, the dark image acquisition starts automatically after the light image
acquisition is successful. The dark image provides the offset calibration information to
workstation. The calibrated image will be on screen when workstation receives light image and
dark image, otherwise there may be no image shown.
43
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter IV Control Box Interface
If the “post-offset” is enabled, DO NOT triggers another exposure or
acquisition before the previous image is shown.
Fig. 4.6.6 Timing of Software Trigger Mode: Normal Operation
Table 4.6.4
Typical Value of Mars1417V serial in soft‐trigger mode
(unit:sec)
T1
Prepare time of equipment.
0.5
T2
Interval between receiving image acquisition
command and beginning of acquisiton
process.
T4
Data acquire and transmission time of
equipment
0.7
T5
Interval between the end of light image
acquisition and the beginning of dark image
acquisition.
0.3
T9
Interval between the equipment is ready and
the beginning of the image acquisition
95% dip in
<5% UT (>95% dip in
UT) for 0.5 cycle.
UT) for 0.5 cycle.
40% UT (60% dip in
40% UT (60% dip in
voltage variations on
UT) for 5 cycle.
UT) for 5 cycle.
power
70% UT (30% dip in
70% UT (30% dip in
lines
UT) for 25 cycle.
UT) for 25 cycle.
IEC 61000-4-11
<5% UT (>95% dip in
<5% UT (>95% dip in
UT) for 5 sec.
UT) for 5 sec.
Voltage
dips,
short
Mains power quality should be that of a typical
interruptions
and
commercial or hospital environment. If the user of
that requires continued operation during power
supply
input
mains interruptions, it is recommended that (?) be
powered from an uninterruptible power supply.
Power
frequency
Power frequency magnetic fields should be at
(50/60Hz)
magnetic
3 A/m
3 A/m
levels characteristic of a typical location in a
field
typical commercial or hospital environment.
IEC 61000-4-8
UT is the a.c. mains voltage prior to application of the test level.
 Guidance and manufacturer’s declaration----electromagnetic immunity
Mars1417V serial is intended for use in the electromagnetic environment specified below.
Thecustomer or the user of Mars1417V serial should assure that it is used in such an environment.
IEC
60601
Test
Immunity Test
Compliance Level
Electromagnetic Environment - Guidance
3 Vrms
Portable
Level
Conducted RF
3 Vrms
IEC 61000-4-6
150kHz to 80MHz
and
mobile
RF
communications
equipmentshould be used no closer to any part of the
ModelMars1417V serial, including cables, than the
recommendedseparation distance calculated from the
equationapplicable to the frequency of the transmitter.
Recommended separation distance
Radiated
RF
IEC 61000-4-3
3 V/m
d = 1,2
d = 1,2
p 80 MHz to 800 MHz
d = 2,3
p 800 MHz to 2,5 GHz
3 V/m
80MHz to 2.5GHz
Where P is the maximum output power rating of
thetransmitter
in
watts
(W)
according
to
the
transmittermanufacturer and d is the recommended
separationdistance in meters (m).
47
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter V Regulatory Information
Field
strengths
from
fixed
RF
transmitters,
as
determinedby an electromagnetic site survey, shouldbe
less than the compliance level in each frequency range.
Interference
may
occur
in
the
vicinity
of
equipmentmarked with the following symbol:
NOTE: UT is the a.c. mains voltage prior to application of the test level.
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which Mars1417V serialis used exceeds the applicable RF
compliance level above, Mars1417V serialshould be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating Mars1417V serial.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
 Recommended separation distances between portable and mobile RF communications
equipment and Mars1417V serial
Mars1417V serial is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of Mars1417V serial can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and Mars1417V serial as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter /W
Separation distance according to frequency of transmitter /m
150kHz~80 MHz
d = 1.2
80 MHz~800 MHz
d = 1.2
800 MHz ~2.5GHz
d = 2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
48
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter V Regulatory Information
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d inmeters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
 Below cables information are provided for EMC reference.
Cable
Recommended
cable length
Shielded or Unshielded
Number
Cable classification
Adaptor Power Cable
3m
Unshielded
1 Set
AC Power
LAN port Cable
>3m
Shielded
1 pcs
Signal
Data Cable
>3m
Shielded
1 pcs
Signal
 Important information regarding Electro Magnetic Compatibility (EMC)
Mars1417V serialneeds special precautions regarding EMC and needs to be installed only by
iRay or authorization engineer and put into service according to the EMC information provided in
the user manual; Mars1417V serial in use may be susceptible to electromagnetic interference from
portable and mobile RF communications such as mobile (cellular) telephones. Electromagnetic
interference may result in incorrect operation of the system and create a potentially unsafe situation.
Mars1417V serial conforms to this EN60601-1-2:2007 standard for both immunity and
emissions.
Nevertheless, special precautions need to be observed:
The use of accessories, transmitters and cables other than those specified by this User Manual,
with the exception of accessories and cables sold by iRay of Mars1417V serial as replacement parts
for internal components, may result in increased EMISSIONS or decreased IMMUNITY of
Mars1417V serial.
Mars1417V serial should not be used adjacent to or stacked with other equipment. In case
adjacent or stacked use is necessary, Mars1417V serial should be observed to verify normal
49
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter V Regulatory Information
operation in configuration.
50
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter VI Imaging Software SDK
6. Chapter VI ImagingSoftwareSDK
Please find the details from iRaySDKManualortheupdated versions. Please contact iRay service office.
51
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter VII Service Information
7. Chapter VII Service Information
Product lifetime
The estimated product lifetime is up to 8 years under appropriate regular inspection and maintenance.
Regular inspection and maintenance
In order to ensure the safety of patients, operator and third parties, and maintain the performance and
reliability of the equipment, be sure to perform regular inspection at least once a year. If necessary,
clean up the equipment, make adjustments, or replace consumables. There may be cases where
overhaul is recommended depending on conditions. Contact your sales representative or local iRay
dealer for regular inspection or maintenance.
Repair
If a problem cannot be solved even taking the measures indicated in troubleshooting, contact your
sales representative or local iRay dealer for repairs. Please refer to the name label and provide the
following information:
Product Name: VenuSerial Number:
Description of Problem: In details as clear as possible.
Replacement parts support
Performance parts (parts required to maintain the functioning of the product) of this product will be
stocked for 5 years after discontinuance of production, to allow for repair.
Trouble Shooting
When you encounter problems or error message in using MARS1417V SERIAL detector, search the
table below for the problem or error message and try the solutions. If the problem persists, turn off the
detector and consult your sales representative or local dealer. Please refer to the detail of the problem
or error message.
52
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Chapter VII Service Information
symptom
Causer / Error messages Remedy
in control system.
Failed to turn on
Damage of control box
Check the power supply connector and
power plug, plug the power cord
again.
Check the fuses holder.
Consult iRay’s or iRay’s authorized
Filed Application Engineer.
Link lamp does not light The
up
communication Check the LAN cable whether loose,
circuit is not secured.
plug the LAN cable again.
Consult iRay’s or iRay’s authorized
Filed Application Engineer.
Power lamp does not The power circuit is not Check the power supply connector and
light up
secured.
power plug, plug the power cord
again.
Check the fuses holder.
Consult iRay’s or iRay’s authorized
Filed Application Engineer.
Detector’s
parameter The control interface is Consult iRay’s or iRay’s authorized
does not read out.
not secured.
Filed Application Engineer.
When send acquisition The control interface is Consult iRay’s or iRay’s authorized
command to detector, not secured.
Filed Application Engineer.
no image can acquire.
Service Office
Tel: +86 21 50720560 ‐ 8059
Fax: +86 21 50720561
E‐mail: service@iraychina.com
Location: 2F, Building 7, No.590, Ruiqing Rd, Pudong, Shanghai, China PC: 201201
53
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix A Installation of SDK
Appendix A Installation of SDK
The equipment provides two interfaces for image acquisition. For using, user must
install the related software correctly.
A.1Installation of Gigabit Ethernet Driver
In order to use the Gigabit Ethernet interface for data transfer, the uses needs to
install the driver WinPcap4.1.2, which has been included in the CD-ROM.
Click “next” button to start installation.
Click the “I Agree” button to confirm with the license agreement.
54
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix A Installation of SDK
Do not select the “Automatically start the WinPcap driver at boot time” item. The
function will be loaded bytheimage acquisition software.
Then installation completed
When using a new computer, the user shall configure the settings
of network card.
As shown below, click “Configure” button of Local Area Connection Properties, set
the value of Large Send Offload as Disabled, and the value of Speed & Duplex as 1.0
Gbps Full Duplex. Error settings may consult in abnormal network connection.
55
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix A Installation of SDK
56
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix BDescription of Different Cable Series
Appendix B Description of Different Cable Series
A. Check the MPN of the Extension cables and the Detector cables of the equipment.
MPN Label of the Detector Cable
There is a printed label near the float plug/socket on the cable. The first line is the MPN. IF there is no label or
different label on the cable, please find the details from iRay service office.
B. Check the color label of the cable
There are color labels on the Detector cable and Extension cable. If the label color is same that means the
definition of the Detector cable can match the Extension cable. For Mars1417V serial, ONLY series B is available
whose color label is BLUE.
The same color means the pin definitions of the cables are same or matching.
But before installation, please check the cable length. The total length of the
Detector cable and Extension cable should be less than 9m.
C. Compatibility of different cable series
Cable Series
Detector Cable
Extension Cable
02040002
02040003 (long)
02040023 (short)
02030144
02040013 (long)
02040019 (short)
Cables can ONLY be used with the same series. Mixed use of cables in series
A and B may result in damage of the equipment or Control Box circuit.
57
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix CDifferent Version of Control Box
Appendix C Different Version of Control Box
There are two versions of control box. They can be told by the label on the bottom of the control
box. The V0 used the label only in English or Chinese (Fig. C1). The V1’s label has both English
and Chinese descriptions (Fig. C2)
Both versions share the same circuit architecture and have the same performance. But the definition
of the HVG port is different.
The descriptions of HVG interface & HVG Cable ofV0
Pin
No.
Wire Color
Name
I/
Signal Description
Shield
NC
No Connection
White
Reserved
No Connection
The synchronizing signal to the X-ray given by the
Red/white
Xray_On
HVG. Detector will start data acquire after the
signal shift from the low level to the high level
Black
Reserved
Green
Frame_Request
Gray
Expose_OK
Brown
Reserved
No Connection
Brown/white
EXT_GND
GND
Blue
Reserved
No Connection
10
Yellow
EXT_VOUT
+12V(Imax=450mA) power supply
11
Black/white
Reserved
No Connection
12
Pink
Reserved
No Connection
13
Red
Reserved
No Connection
14
Cyan
Reserved
No Connection
15
Purple
Reserved
No Connection
16
Orange
Reserved
No Connection
No Connection
Request given by the HVG, it’s a prepare signal of
the detector.
When the signal is low , it indicates that the
detector is ready for the exposure
The V0 control box also supports the prep-trigger mode.
There are pull-up/pull-down circuits
inside the control box, so the external pull-up/pull-down circuits are unnecessary. (Fig.C3)
58
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix CDifferent Version of Control Box
59
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix D Recommended Computer Platform
Appendix DRecommended Computer Platform
Recommended
Recommended
Desktop
Laptop
Minimum
CPU
Intel Core i3 2.8G
Intel Core i73.6G
Intel Core i73.6G
Memory
2G DDR3
4G DDR3
4G DDR3
Hard Disk
160 G
640 G
320 G
Monitor
1280×1024
1680 x 1050
1600×900
Memory of
Display
Adapter
512Mbits
1Gbits
1Gbits
Intel Pro EXP9301CT
Gigabit Network
PRO
Intel(R) 82579LM
Gigabit Network
Gigabit Network
Adapter with PCIe
Connection
Adapter with PCIe
Ethernet Card
interface of Intel or
Broadcom
interface
Others
USB2.0
OS
Windows XP 32bit
Please set the computer to “high performance” and set “power
saving” disabled, when the SDK or detector is running. It’s also
recommended to set “screen saver” disabled.
60
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix D Recommended Computer Platform
Function List of Different Hardware/Firmware
Versions
Detector
Control
Box
V0
V0
V0
V1
V1
V1
Cable
Series A
Or
Series B
SeriesB
Firmware
Function
Detector: 2.2.x
Control Box: 1.2.x
Detector: 2.2.x
Control Box: 1.2.x
Detector: 3.3.x
Control Box:1.3.x
61
1. RS232 command interface
2. Static image acquisition
3. Default mode:inner-triggermode
1. Static image acquisition
2. Gigabit Ethernet command
interface
3. Default mode: inner-trigger
mode, but can be configured by
customers
4. Temperature feedback
5. Detector SN. Feedback
6. Replaceable sensor cable
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix E Firmware Update
Appendix EFirmware Update
Mars1417V serial allows user upgrade firmware without removing cover of the equipment.
Please do as follows:
a)Turn off the power
b)Open the cover of the JTAG port on it. You will see the port as FigE1.
Fig.E1 JTAG Port
c)Insert the FPD JTAGBoard into the port.
Fig.E2JTAGBoard
d)Connect the Xilinx JTAG Adapter to the JTAGBoard and Computer. Then turn on the power.
The LED on the Adapter will turn on, Fig A/E3.
FigE3. Xilinx JTAG Adapter
a) Open the Xilinx iMPACT® software;
62
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix E Firmware Update
Select “No”
Select “No”
63
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix E Firmware Update
Select“Create a new project”
Select“Configure devices using Boundary-Scan(JTAG)”, then“OK”
Select“NO”
64
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix E Firmware Update
Select“Cancel”
f)Program the FPGA
Select the first FPGA,right click and select“Add SPI/BPI Flash”
65
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix E Firmware Update
Select“slave.mcs”
Select Flash type“N25Q128”,then“OK”
Right click and select“Program”
66
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix E Firmware Update
Ittakes several minutes to finish updating
b) As the same steps, upgrade the master FPGA’s firmware
c) Reboot the Flat Panel Detector;
67
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix F Dummy Lines
Appendix F Dummy Lines
There are 8 dummy lines on the each long edge of the image.
Fig D1
68
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix F Dummy Lines
In details
Fig D2
69
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix G Definition of Defects
Appendix GDefinition of Defects
Defect Pixel:
A defect pixel is defined as a pixel whose value is deviated from the median of its neighboring
32x32 pixels by more than 6 times of the standard deviation of this regionafter gain-offset
correctionunder any X-ray condition. In addition, if there is defect which is not included in the defect
map the supplier defined, then new defect should be added to the defect map. If the final statistics is
above the quality control limit, then the detector should be returned.
Defect Line:
The horizontal or verticallinesof which more than 5% of the pixels are determined as defect pixels.
Full Class 0-8:
"Class N" defect is a defective pixel with N adjacent defect(s). "Full" means that all defective
pixels (even included in a defective line or column) are individually counted.
Meta 7+8 Class1-8:
"Meta 7+8" means that on the original defect map, isolate the defective pixels corresponding to
Class 7 and Class 8: this gives a new defect map where each Class 7 or Class 8 cluster is replaced by
one single defective pixel. On this new defect map, re-calculate the distribution of defect Class N, this
gives the "Meta 7+8 Class N" defect.
Central Area:
The area in which both the x-coordinates and y-coordinates of the pixels are between 1/6 to 5/6 width
and height of the panel respectively. For example, for a 2304x2800 panel, the central area is defined
as the area where 384≤x≤1920 and 467≤y≤2333.
70
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix H Information of Service Office
Appendix H Information of Service Office
Tel:
+86 21 50720560 - 8059
Fax:
+86 21 50720561
E-mail: service@iraychina.com
Address: 2F, Building 7, No.590, Ruiqing Rd, Pudong, Shanghai, China PC: 20120
71
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com
Appendix I Information of Manufactures
Appendix I Information of Manufactures
COMPANY: iRayTechnology(Shanghai) Ltd.
ADDRESS: Rm202, Building 7, No. 590, Ruiqing Rd, Zhangjiang East, Pudong New
Area, Shanghai, China
ZIPCODE: 201201
TELEPHONE: +86-21-50720560
FAX: +86-21-50720561
HOMEPAGE: WWW.IRAY-TECH.COM
SERVICE: SERVICE DEPARTMENT of iRay
SERVICE TEL: +86-21-50720560
MARKET TEL: +86-21-50720560
72
Service Office:
Tel: +86 21 50720560
E‐mail:service@iraychina.com

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.7
Linearized                      : Yes
Author                          : wei.pan
Create Date                     : 2014:10:10 15:42:45+08:00
Modify Date                     : 2014:10:13 14:01:34+08:00
XMP Toolkit                     : Adobe XMP Core 4.2.1-c041 52.342996, 2008/05/07-20:48:00
Creator Tool                    : PScript5.dll Version 5.2.2
Metadata Date                   : 2014:10:13 14:01:34+08:00
Producer                        : Acrobat Distiller 9.5.5 (Windows)
Format                          : application/pdf
Title                           : Microsoft Word - 032-202-01 User Manual-Mars1417V User Manual A2.docx
Creator                         : wei.pan
Document ID                     : uuid:5b420b42-f29f-4d44-b9d2-069af94e2dc3
Instance ID                     : uuid:4f6fcf6e-9871-46f5-b2f6-12c998a3b307
Page Count                      : 72
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2ACHK-02112031

Navigation menu