Genexus Integrated Sequencer User Guide (Pub. No. MAN0017910 D.0)

Genexus Integrated Sequencer

Thermo Fisher Scientific (25 November 2020)

Genexus Integrated Sequencer - Thermo Fisher Scientific

For Research Use Only. Not for use in diagnostic procedures. Genexus™ Integrated Sequencer USER GUIDE Catalog Number A45727 Publication Number MAN0017910 Revision D.0

Genexus™ Integrated Sequencer - USER GUIDE - Thermo ...

Life Technologies Holdings Pte Ltd | ... Life Technologies Corporation | ... This user guide provides detailed instructions for operating the Genexus™ ...

PDF MAN0017910 GenexusIntegratedSequencer UG
GenexusTM Integrated Sequencer
USER GUIDE
Catalog Number A45727
Publication Number MAN0017910 Revision D.0
For Research Use Only. Not for use in diagnostic procedures.

Manufacturer: Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256
Manufacturer: Life Technologies Corporation | 200 Oyster Point Blvd | South San Francisco, CA 94080 | USA
Manufacturer: Life Technologies Corporation | 7335 Executive Way | Frederick, MD 21704 | USA

Product: GenexusTM Integrated Sequencer

Product: GenexusTM Software

Products: GX5TM Chip and GenexusTM Coupler GenexusTM Library Strips 1 and 2AS GenexusTM Library Strips 1 and 2HD GenexusTM Templating Strips 3GX5TM and 4 GenexusTM Barcodes 1­96 AS GenexusTM Barcodes 1­32 HD GenexusTM Primer Pool Tubes GenexusTM Pipette Tips

GenexusTM Sequencing Kit GenexusTM Controls GenexusTM Conical Bottles GenexusTM Filter GenexusTM GX5TM Starter PackAS GenexusTM GX5TM Starter PackHD OncomineTM GX assays

The information in this guide is subject to change without notice.

DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.

Revision history: Pub. No. MAN0017910

Revision

Date

Description

D.0

25 November Corrected an error in GenexusTM Software version number in "Software compatibility and requirements" on

2020

page 11.

C.0

29 October · Updated the contents of the GenexusTM Installation and Training Kit. See "GenexusTM Integrated Sequencer".

2020

· Corrected guidance on reuse of GenexusTM Barcodes plates. See "Guidelines for GenexusTM Integrated

Sequencer operation" on page 26.

· Added the topic "Guidelines for expired reagents and chips" on page 27.

· Changed the recommended concentration for manually prepared libraries from 125 pM to 200 pM.

· Clarified use of "Do Force Clean" checkbox. See "Load the sequencer and start a run" on page 83.

· Added screenshots for chip verification and leak checking in "Load the sequencer and start a run" on page 83.

· Updated "Options for an expired sequencer initialization" and moved to Chapter 7.

· Updated recommended actions for instrument error and warning messages in "GenexusTM Integrated Sequencer error and warning messages" on page 140.

· Performance qualification results table updated with removal of MAPD metric to align with a new version of the Performance Qualification Assay (v1.4.0). See "Performance Qualification results" on page 135.

B.0

7 July 2020 · Updated for GenexusTM Software 6.2.0.

· Updated for library and template strip part numbers.

· Updated the components of the GenexusTM Installation and Training Kit.

· Added guidance for using inline controls for troubleshooting.

· Added the troubleshooting section "GenexusTM Integrated Sequencer error and warning messages".

· Added the topic "Replace the GenexusTM Conical Bottles" on page 165.

· Added the section "Library QC Archive: recover library preparations from the GenexusTM Integrated Sequencer for reuse" on page 172.

A.0

13 November New user guide for the GenexusTM Integrated Sequencer

2019

Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these products, you accept the terms and conditions of all applicable Limited Use Label Licenses.

TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. Microsoft and Excel are trademarks of Microsoft Corp. Google and Chrome are trademarks of Google Inc. Ubuntu is a trademark of Canonical Limited. TaqMan is a registered trademark of Roche Molecular Systems, Inc., used under permission and license. Eppendorf and LoBind are trademarks of Eppendorf AG. Agencourt and AMPure are trademarks of Beckman Coulter, Inc.
©2020 Thermo Fisher Scientific Inc. All rights reserved.

Contents
 CHAPTER 1 Product information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 GenexusTM Integrated Sequencer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Software compatibility and requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Reagents and supplies--Ion AmpliSeqTM library chemistry . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
GenexusTM Library Strips 1 and 2AS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 GenexusTM Barcodes AS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 GenexusTM GX5TM Starter PackAS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Reagents and supplies--Ion AmpliSeqTM HD library chemistry . . . . . . . . . . . . . . . . . . . . . . . . 13 GenexusTM Library Strips 1 and 2HD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 GenexusTM Barcodes 1­32 HD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 GenexusTM GX5TM Starter PackHD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Shared reagents and supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 GenexusTM Templating Strips 3GX5TM and 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 GenexusTM Primer Pool Tubes and Pipette Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 GX5TM Chip and GenexusTM Coupler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 GenexusTM Sequencing Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 GenexusTM Conical Bottles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 GenexusTM Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 GenexusTM Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 OncomineTM GX assays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Required materials not supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Recommended materials for nucleic acid isolation and quantification . . . . . . . . . . . . . . . . . 19 GenexusTM Integrated Sequencer components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 GenexusTM Integrated Sequencer deck stations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 GenexusTM Integrated Sequencer input and output connections . . . . . . . . . . . . . . . . . . . . . . 23 Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

4

GenexusTM Integrated Sequencer User Guide

Contents
 CHAPTER 2 Before you begin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Avoid nucleic acid contamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Avoid chip damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Avoid strong electromagnetic radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Protection by equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Guidelines for GenexusTM Integrated Sequencer operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Guidelines for expired reagents and chips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Power the GenexusTM Integrated Sequencer on or off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Power on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Power off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Get started with GenexusTM Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 About the GenexusTM Software user interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 User-access levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 System tracking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Request and sign in to a new account . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Sign in . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
 CHAPTER 3 Create and manage assays (manager/administrator) . . . . . . . . . . . . 32
About assays in GenexusTM Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 System-installed assays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Manage assays (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Create a new assay (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Copy an assay (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Import an assay (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
 CHAPTER 4 Enter samples and libraries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Create a new sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 System-installed sample attributes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Create a custom sample attribute (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . 48
Import samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Download an example samples file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Manage samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Sort, search, and filter samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Export samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 View notes or add a note to a sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Edit a sample (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Edit a sample and amend a report after a run (manager/administrator) . . . . . . . . . . . . 53

GenexusTM Integrated Sequencer User Guide

5

Contents

Review sample edit history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Delete samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Prepare or import a library batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Prepare a library batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Import a library batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
 CHAPTER 5 Plan and manage runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Plan a sample run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 Plan a library run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
 CHAPTER 6 Dilute the samples and load the sample plate . . . . . . . . . . . . . . . . . . . 75
Guidelines for nucleic acid isolation and quantification--sample runs . . . . . . . . . . . . . . . . . 75 Dilute or concentrate the samples (if needed) and load the sample plate--sample run . . 76 Guidelines for library quantification--library runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Dilute and pool libraries, and load the sample plate--library run . . . . . . . . . . . . . . . . . . . . . . 78
 CHAPTER 7 Load the sequencer and start a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Before you begin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 Fill GenexusTM Primer Pool Tubes (custom assays only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 Load the sequencer and start a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Clear the instrument deck and perform a UV Clean . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Options for an expired sequencer initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
 CHAPTER 8 Monitor the run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
View run progress on the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
 CHAPTER 9 Review data and results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Review sample results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Review run results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Assign PCR Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Lab Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Download the Lab Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 View sequencing results and assay metrics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Summary of the Sample Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102 Assay metrics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 QC results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 View SNV/INDEL results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112 View Fusion results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114 View CNV results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 View annotation sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121

6

GenexusTM Integrated Sequencer User Guide

Contents

Create and assign variant classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 Filter results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 View OncomineTM TCR BetaLR Assay GX run results . . . . . . . . . . . . . . . . . . . . . . . . . . 124 Reanalyze a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 Sign off on the run results (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 Generate customized reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128 Results files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130 Review coverageAnalysis plugin results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 Upload sample results files to Ion ReporterTM Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 Verification runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133 View and sign off on the verification run results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
 APPENDIX A Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Troubleshoot GenexusTM Integrated Sequencer performance with CF-1 and inline controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
GenexusTM Integrated Sequencer--general and QC troubleshooting . . . . . . . . . . . . . . . . . 138 GenexusTM Integrated Sequencer error and warning messages . . . . . . . . . . . . . . . . . . . . . . 140 GenexusTM Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
 APPENDIX B Touchscreen reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Touchscreen icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Network Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149 Perform a Clean instrument procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152 System Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154 Data Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158 Instrument settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
 APPENDIX C Supplemental information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Maintain the sequencer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163 Materials required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163 Clean or decontaminate the sequencer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164 Replace the GenexusTM Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164 Replace the GenexusTM Conical Bottles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165 GenexusTM Integrated Sequencer power off and power on before and after a long-term shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Quantify FFPE DNA with the QubitTM Fluorometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171 Library QC Archive: recover library preparations from the GenexusTM Integrated
Sequencer for reuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172 Required materials and equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172 Recover libraries from the sequencer and purify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172 Quantify the purified libraries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175

GenexusTM Integrated Sequencer User Guide

7

Contents
Combine libraries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175 Store libraries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 175 Guidelines for using custom assays with the GenexusTM Integrated Sequencer . . . . . . . . 176 Planning sequencing runs for efficient use of consumables . . . . . . . . . . . . . . . . . . . . . . . . . 178 Configure an Ion ReporterTM Server account (administrator) . . . . . . . . . . . . . . . . . . . . . . . . . 179 Tag an Ion ReporterTM Software analysis workflow for use with the
IonReporterUploader plugin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181 Configure Thermo Fisher Accounts in GenexusTM Software (administrator) . . . . . . . . . . . . 182
 APPENDIX D coverageAnalysis plugin in GenexusTM Software . . . . . . . . . . . . . . 184
Reads statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184 Example Coverage Analysis Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187 Example charts generated by the coverageAnalysis plugin . . . . . . . . . . . . . . . . . . . . . . . . . 189 Output files generated by the coverageAnalysis plugin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
 APPENDIX E Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193
Symbols on this instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193 Standard safety symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193 Additional safety symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 194 Location of safety labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 195 Control and connection symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196 Conformity symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
Instrument safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197 Physical injury . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197 Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198 Cleaning and decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199 Instrument component and accessory disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
Safety and electromagnetic compatibility (EMC) standards . . . . . . . . . . . . . . . . . . . . . . . . . 200 Safety standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 EMC standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 Environmental design standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201 Radio compliance standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201
Chemical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203 Biological hazard safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
 APPENDIX F Documentation and support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
Related documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206 Customer and technical support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206 Limited product warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207

8

GenexusTM Integrated Sequencer User Guide

1

Product information

 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9  GenexusTM Integrated Sequencer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10  Software compatibility and requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11  Reagents and supplies--Ion AmpliSeqTM library chemistry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11  Reagents and supplies--Ion AmpliSeqTM HD library chemistry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13  Shared reagents and supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15  OncomineTM GX assays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18  Required materials not supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19  Recommended materials for nucleic acid isolation and quantification . . . . . . . . . . . . . . . . . . . . . . . 19  GenexusTM Integrated Sequencer components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21  GenexusTM Integrated Sequencer deck stations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22  GenexusTM Integrated Sequencer input and output connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23  Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
IMPORTANT! Before using this product, read and understand the information in the "Safety" appendix in this document.

Product description
The Ion TorrentTM GenexusTM Integrated Sequencer is a next-generation (NGS) sequencing system that integrates library preparation, template preparation, and sequencing into a single-day, single-instrument automated run. The instrument supports both sample-to-results and library-to-results sequencing runs (up to 32 DNA or RNA samples/run). Ion TorrentTM GenexusTM Software streamlines the NGS workflow by integrating the setup-to-report workflow within a single software system. Key features include:
· Go from nucleic acid to report in a single day · Flexible and cost-effective run planning making use of a multi-lane, multi-run sequencing chip: the
Ion TorrentTM GX5TM Chip · Automated library preparation, including cDNA synthesis, for up to 400 base-read libraries using
either standard Ion AmpliSeqTM or Ion AmpliSeqTM HD library chemistry · Automated Ion SphereTM Particle (ISP)-loading and template preparation with on-chip amplification · Support for up to four compatible assays on a GX5TM Chip in a single run, with an output of 12­
15 million reads from each lane. · As little as five minutes total hands-on time required per run · Real-time consumables tracking by the instrument to guide consumable loading during run setup

GenexusTM Integrated Sequencer User Guide

9

1

Chapter 1 Product information GenexusTM Integrated Sequencer

· Consumables that are usable for up to 14 days after loading · On-instrument sequencing data analysis requiring no external server

GenexusTM Integrated Sequencer

The GenexusTM Integrated Sequencer includes the following components.

Components

Cat. No.

GenexusTM Integrated Sequencer

A45727

GenexusTM Installation and Training Kit
[1] Not available for separate purchase.

A40278[1]

The Ion TorrentTM GenexusTM Installation and Training Kit (Cat. No. A40278) is available to first-time owners of a GenexusTM Integrated Sequencer and is shipped with the instrument. The kit contains the following reagents, supplies, and controls that are used during the installation, training, and operation of the instrument.

GenexusTM Installation and Training Kit

Contents GenexusTM Controls GenexusTM Strip 1 GenexusTM Strip 2AS GenexusTM Barcodes 1­32 AS GenexusTM Strip 3GX5TM GenexusTM Strip 4 GenexusTM Cartridge GenexusTM Bottle 2 GenexusTM Bottles 1 and 3 GenexusTM Pipette Tips GenexusTM Conical Bottles GenexusTM Filter GX5TM Chip and GenexusTM Coupler Adhesive PCR Plate Foils

Part No. A40267 A46812 A46813 A40258 A46815 A46816 A40272 A40273 A40274 A40266 A40275 A40302 A40269 AB0626

Quantity 1 kit
8 strips 8 strips 1 plate 8 strips 8 strips 2 cartridges 4 bottles 2 bottles each 12 racks 2 sets of 5 bottles 1 filter 2 each 1 box (100 foils)

Storage -30°C to -10°C
2°C to 8°C -30°C to -10°C
15°C to 30°C 2°C to 8°C
-30°C to -10°C -30°C to -10°C
15°C to 30°C

10

GenexusTM Integrated Sequencer User Guide

Chapter 1 Product information Software compatibility and requirements

1

Software compatibility and requirements
The procedures in this guide are designed for use with GenexusTM Software 6.2.0 or later. Versionspecific information is provided in the software release notes for your version of the software. An administrator-level user can view the software version in the (Settings) / Software Updates screen.
GenexusTM Software is supported on GoogleTM ChromeTM browser version 64 and later and is best viewed with 1440 × 900 screen resolution. GoogleTM ChromeTM browser is recommended for use with the software.
The operating system of the sequencer is UbuntuTM 18.041.
For more information on using the software, see the online GenexusTM Software 6.2 Help.

Reagents and supplies--Ion AmpliSeqTM library chemistry
GenexusTM Integrated Sequencer reagents and supplies can be ordered in convenient combo kits and starter packs, but most consumables can also be ordered individually as your needs require. The following tables provide information on the various ordering options that are available for Ion AmpliSeqTM library chemistry.
Note:
· Consumables that have catalog numbers are orderable. Components that have part numbers cannot
be ordered individually.
· Reagents that are specific to Ion AmpliSeqTM library chemistry have an AS suffix.

GenexusTM Library Strips 1 and 2AS
Ion TorrentTM GenexusTM Library Strips 1 and 2AS (Cat. No. A40252) for standard Ion AmpliSeqTM librarybased chemistry are ordered as kits with eight pairs of strips/kit.
Note: GenexusTM Library Strips 1 and 2AS with part numbers listed in the following table are compatible only with GenexusTM Software 6.2.0 and later.

Component GenexusTM Strip 1

Carrier color Light red

Part No. A46812

Quantity per kit 8 strips

Storage 2°C to 8°C

GenexusTM Strip 2AS

Light blue

A46813

8 strips

-30°C to -10°C

GenexusTM Integrated Sequencer User Guide

11

1

Chapter 1 Product information Reagents and supplies--Ion AmpliSeqTM library chemistry

GenexusTM Barcodes AS

Ion TorrentTM GenexusTM Barcodes AS are supplied in plates containing 32 dual barcodes per plate. The barcodes can be ordered as a set of three plates (Cat. No. A40257), or ordered individually.

Item GenexusTM Barcodes 1­96 AS GenexusTM Barcodes 1­32 AS GenexusTM Barcodes 33­64 AS GenexusTM Barcodes 65­96 AS

Label color Blue Blue Blue Blue

Cat. No. A40257 A40258 A40259 A40260

Quantity 3 plates 1 plate 1 plate 1 plate

Storage 15°C to 30°C

GenexusTM GX5TM Starter PackAS
Ion TorrentTM GenexusTM GX5TM Starter PackAS (Cat. No. A40279) supplies the following components for Ion AmpliSeqTM library preparation and sequencing using a GenexusTM-ready assay. Note: For custom assays, GenexusTM Primer Pool Tubes (Cat. No. A40262) must be ordered separately.

Component GenexusTM Strip 1 GenexusTM Strip 2AS GenexusTM Strip 3GX5TM GenexusTM Strip 4 GenexusTM Barcodes 1­32 AS GenexusTM Pipette Tips GenexusTM Cartridge GenexusTM Bottle 2 GenexusTM Bottles 1 and 3

Part or Cat. No. A46812 A46813 A46815 A46816 A40258 A40266 A40272 A40273 A40274

Quantity 8 strips 8 strips 8 strips 8 strips 1 plate 12 racks 2 cartridges 4 bottles 2 bottles each

Storage 2°C to 8°C -30°C to -10°C 2°C to 8°C -30°C to -10°C 15°C to 30°C
-30°C to -10°C 15°C to 30°C

12

GenexusTM Integrated Sequencer User Guide

Chapter 1 Product information Reagents and supplies--Ion AmpliSeqTM HD library chemistry

1

Reagents and supplies--Ion AmpliSeqTM HD library chemistry
GenexusTM Integrated Sequencer reagents and supplies can be ordered in convenient combo kits and starter packs, but most consumables can also be ordered individually as your needs require. The following tables provide information on the various ordering options that are available for Ion AmpliSeqTM HD library chemistry.
Note:
· Consumables that have catalog numbers are orderable. Components that have part numbers cannot
be ordered individually.
· Reagents that are specific to Ion AmpliSeqTM HD library chemistry have an HD suffix.

GenexusTM Library Strips 1 and 2HD
Ion TorrentTM GenexusTM Library Strips 1 and 2HD (Cat. No. A40255) for Ion AmpliSeqTM HD library-based chemistry are ordered as kits with eight pairs of strips/kit.
Note: GenexusTM Library Strips 1 and 2HD with part numbers listed in the following table are compatible only with GenexusTM Software 6.2.0 and later.

Component GenexusTM Strip 1

Carrier color Light red

Part No. A46812

Quantity per kit 8 strips

Storage 2°C to 8°C

GenexusTM Strip 2HD

Violet

A46814

8 strips

-30°C to -10°C

GenexusTM Integrated Sequencer User Guide

13

1

Chapter 1 Product information Reagents and supplies--Ion AmpliSeqTM HD library chemistry

GenexusTM Barcodes 1­32 HD

Ion TorrentTM GenexusTM Barcodes 1­32 HD are supplied in a plate containing 32 dual barcodes.

Item GenexusTM Barcodes 1­32 HD

Label color Purple

Cat. No. A40261

Quantity 1 plate

Storage 15°C to 30°C

GenexusTM GX5TM Starter PackHD
Ion TorrentTM GenexusTM GX5TM Starter PackHD (Cat. No. A40280) supplies the following components for Ion AmpliSeqTM HD library preparation and sequencing using a GenexusTM-ready assay. Note: For custom assays, GenexusTM Primer Pool Tubes (Cat. No. A40262) must be ordered separately.

Component GenexusTM Strip 1 GenexusTM Strip 2HD GenexusTM Strip 3GX5TM GenexusTM Strip 4 GenexusTM Barcodes 1­32 HD GenexusTM Pipette Tips GenexusTM Cartridge GenexusTM Bottle 2 GenexusTM Bottles 1 and 3

Part or Cat. No. A46812 A46814 A46815 A46816 A40261 A40266 A40272 A40273 A40274

Quantity 8 strips 8 strips 8 strips 8 strips 1 plate 12 racks 2 cartridges 4 bottles 2 bottles each

Storage 2°C to 8°C -30°C to -10°C 2°C to 8°C -30°C to -10°C 15°C to 30°C
-30°C to -10°C 15°C to 30°C

14

GenexusTM Integrated Sequencer User Guide

Chapter 1 Product information Shared reagents and supplies

1

Shared reagents and supplies
The following reagents and supplies are used in both Ion AmpliSeqTM library chemistry and Ion AmpliSeqTM HD library chemistry runs.
Note: Consumables that have catalog numbers are orderable. Components that have part numbers cannot be ordered individually.

GenexusTM Templating Strips 3GX5TM and 4
Ion TorrentTM GenexusTM Templating Strips 3GX5TM and 4 (Cat. No. A40263) are ordered as kits with eight pairs of strips/kit.
Note: GenexusTM Templating Strips 3GX5TM and 4 with part numbers listed in the following table are compatible only with GenexusTM Software 6.2.0 and later.

Component GenexusTM Strip 3GX5TM

Carrier color Brown

Part No. A46815

Quantity per kit 8 strips

Storage 2°C to 8°C

GenexusTM Strip 4

Yellow

A46816

8 strips

-30°C to -10°C

GenexusTM Primer Pool Tubes and Pipette Tips

GenexusTM Primer Pool Tubes and Pipette Tips can be ordered individually. GenexusTM Primer Pool Tubes are required for custom assays.

Item GenexusTM Primer Pool Tubes

Cat. No. A40262

Quantity
50 assemblies (2 tubes/assembly)
Bag of 100 caps

Storage 15°C to 30°C

GenexusTM Pipette Tips

A40266

12 racks

GenexusTM Integrated Sequencer User Guide

15

1

Chapter 1 Product information Shared reagents and supplies

GX5TM Chip and GenexusTM Coupler

The GX5TM Chip and GenexusTM Coupler (Cat. No. A40269) are ordered as a set that contains two chips and two couplers, sufficient for up to eight sequencing runs.

GX5TM Chip

Component

Part No. 100081364

Quantity 2 chips

Storage 15°C to 30°C

GenexusTM Coupler

100081252

2 couplers

GenexusTM Sequencing Kit

The Ion TorrentTM GenexusTM Sequencing Kit is ordered as a combo (Cat. No. A40271). Each combo kit is sufficient to sequence up to 2 full chips. The components of the combo are also orderable as listed.

Component GenexusTM Cartridge

Cat. No. A40272

Quantity 2 cartridges

Storage -30°C to -10°C

GenexusTM Bottle 2 GenexusTM Bottles 1 and 3
16

A40273 A40274

4 bottles
2 bottles each (4 bottles total)

15°C to 30°C

GenexusTM Integrated Sequencer User Guide

Chapter 1 Product information Shared reagents and supplies

1

GenexusTM Conical Bottles

GenexusTM Conical Bottles (Cat. No. A40275) are installed in the sequencing reagents bay and serve as reservoirs for nucleotide reagent dilutions. For information on when and how to replace the bottles, see "Replace the GenexusTM Conical Bottles" on page 165.

Component GenexusTM Conical Bottles

Quantity 5 bottles

Storage 15°C to 30°C

GenexusTM Filter

The GenexusTM Filter (Cat. No. A40302) is installed in the liquid waste disposal port on the instrument deck to prevent liquid waste line blockage. For information on installation, see "Replace the GenexusTM Filter" on page 164.

Component GenexusTM Filter

Quantity 2 filters

Storage 15°C to 30°C

GenexusTM Controls
The Ion TorrentTM GenexusTM Controls kit (Cat. No. A40267) provides sufficient GenexusTM Control LibraryAS to perform four library runs. The kit also provides sufficient GenexusTM Control PanelAS and GenexusTM DNA Control to perform eight sample runs.

IMPORTANT! GenexusTM Strip 2AS is required for sequencing GenexusTM Controls. For ordering information, see "GenexusTM Library Strips 1 and 2AS" on page 11.

Note: The GenexusTM Control LibraryAS is barcoded with IonCodeTM 0101.

Component GenexusTM Control LibraryAS GenexusTM Control PanelAS GenexusTM DNA Control

Quantity 1 tube
8 carriers (white) 2 tubes

Storage -30°C to -10°C

GenexusTM Integrated Sequencer User Guide

17

1

Chapter 1 Product information OncomineTM GX assays

OncomineTM GX assays

Ion TorrentTM OncomineTM GX assays are GenexusTM-ready assays sufficient for 32 reactions, and are supplied in pre-measured readytoload GenexusTM Primer Pool Tubes. Assays are provided with GenexusTM Library Strips 1 and 2AS (Cat. No. A40252) or GenexusTM Library Strips 1 and 2HD (Cat. No. A40255) in the amount listed.

Contents

Carrier color

OncomineTM Comprehensive Assay v3 GX (Cat. No. A46296)

OncomineTM Comprehensive Assay v3 DNA GX

Magenta

Pale green

OncomineTM Comprehensive Assay v3 RNA GX Pale orange

Blue

GenexusTM Strip 1

Light red

GenexusTM Strip 2AS

Light blue

Pool
DNA Pool 1 DNA Pool 2 RNA Pool 1 RNA Pool 2
-- --

Carriers per kit
4 4 4 4 2 × 8 2 × 8

OncomineTM Precision Assay GX (Cat. No. A46291)

OncomineTM Precision Assay GX (panel only)

Magenta OPA Pool 1

8

(FWD and

REV primers)

GenexusTM Strip 1

Light red

--

8

GenexusTM Strip 2HD

Violet

--

8

OncomineTM TCR BetaLR Assay GX (Cat. No. A46297)

OncomineTM TCR BetaLR Assay GX (panel only) Magenta RNA Pool 1

8

GenexusTM Strip 1 GenexusTM Strip 2AS

Light red

--

8

Light blue

--

8

Part No.
A40281 A44351 A46812 A46813
A44350
A46812 A46814
A40282 A46812 A46813

Storage
-30°C to -10°C
2°C to 8°C -30°C to
-10°C
-30°C to -10°C
2°C to 8°C -30°C to
-10°C
-30°C to -10°C
2°C to 8°C -30°C to
-10°C

Note:
· Assays using Ion AmpliSeqTM chemistry have primer pools that are loaded in capped tubes in
position 1 of the primer pool tube carriers. Empty uncapped tubes are loaded in position 2.
· Assays using Ion AmpliSeqTM HD chemistry have FWD and REV primer pools that are loaded in
capped tubes in both positions of the primer pool tube carrier.

18

GenexusTM Integrated Sequencer User Guide

Chapter 1 Product information Required materials not supplied

1

Required materials not supplied

Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that the material is available from fisherscientific.com or another major laboratory supplier. Catalog numbers that appear as links open the web pages for those products.

Item

Source

MicroAmpTM EnduraPlateTM Optical 96-Well Clear Reaction Plates with Barcode

4483352, 4483354

Adhesive PCR Plate Foils

AB0626

20, 200, and 1,000µL pipettors and appropriate filtered tips

MLS

Microcentrifuge tubes, 1.5-mL or 1.7-mL (low retention for nucleic acids)

MLS

Vortex mixer with a rubber platform

MLS

Gloves, powder-free nitrile

MLS

Ice buckets and ice

--

Nuclease-free water, molecular biology grade

AM9932

Isopropyl alcohol, 70% solution

MLS

Wipes, disposable lint-free

MLS

(Optional) Uninterruptible Power Supply (UPS)[1]

MLS

[1] For laboratories that experience frequent power outages or line voltage fluctuations, we recommend that you use an uninterruptible power supply that is compatible with 2500 W output or higher.

Recommended materials for nucleic acid isolation and quantification

Unless otherwise indicated, all materials are available through thermofisher.com. Catalog numbers that appear as links open the web pages for those products.

Item Nucleic acid isolation
Ion AmpliSeqTM Direct FFPE DNA Kit RecoverAllTM Total Nucleic Acid Isolation Kit for FFPE RecoverAllTM Multi-Sample RNA/DNA Workflow MagMAXTM FFPE DNA/RNA Ultra Kit PureLinkTM Genomic DNA Mini Kit MagMAXTM CellFree DNA Isolation Kit

Source
A31133, A31136 AM1975 A26069 A31881 K1820-00 A29319

GenexusTM Integrated Sequencer User Guide

19

1

Chapter 1 Product information Recommended materials for nucleic acid isolation and quantification

(continued) Item
MagMAXTM Cell-Free Total Nucleic Acid Isolation Kit RNaseZapTM RNase Decontamination Solution
Nucleic acid quantification TaqManTM RNase P Detection Reagents Kit (Recommended for DNA only) QubitTM 4 Fluorometer[1] One or more of the following kits for use with the QubitTM 4 Fluorometer: · QubitTM dsDNA HS Assay Kit (High-sensitivity DNA) · QubitTM dsDNA BR Assay Kit (Broad range DNA) · QubitTM RNA HS Assay Kit (High-sensitivity RNA) · QubitTM RNA BR Assay Kit (Broad range RNA)
Library quantification (library runs only) Ion Library TaqManTM Quantitation Kit
[1] QubitTM 2.0 Fluorometer and later are supported.

Source A36716 AM9780
4316831 Q33238
Q32851, Q32854 Q32850, Q32853 Q32852, Q32855 Q10210, Q10211
4468802

20

GenexusTM Integrated Sequencer User Guide

Chapter 1 Product information GenexusTM Integrated Sequencer components

1

GenexusTM Integrated Sequencer components

1
2 10

3

9

4

8

11

11

5

6

7

Major features and components of the exterior and sequencing reagents bay of the GenexusTM Integrated Sequencer

1 Door to deck chamber. The door is locked in the closed position during an instrument run.
2 Power button 3 GenexusTM Bottle 1 (Chemical Waste) 4 GenexusTM Conical Bottles (Reusable conical bottles for
GenexusTM Cartridge reagent dilution) 5 GenexusTM Bottle 2 (Sequencing Solution)

6 GenexusTM Bottle 3 (Cleaning Solution) 7 Waste Carboy 8 Waste pipette tip bin 9 GenexusTM Cartridge 10 Touchscreen 11 Sequencing reagents bay door. Doors are locked in the
closed position during an instrument run.

GenexusTM Integrated Sequencer User Guide

21

1

Chapter 1 Product information GenexusTM Integrated Sequencer deck stations

GenexusTM Integrated Sequencer deck stations

1

16

2

3

4

15

7

8

5

14

9

10

13

6

11

12

Interior GenexusTM Integrated Sequencer deck components and stations

1 PCR amplification station
2 Microcentrifuge 3 GenexusTM Barcodes station 4 GenexusTM Primer Pool Tube station 5 GenexusTM Coupler station
6 Chip install station 7 Zone 1 station (GenexusTM Strip 1) 8 Zone 2 station (GenexusTM Strip 2AS or GenexusTM
Strip 2HD, depending on your assay,)

9 Zone 3 station (GenexusTM Strip 3GX5TM) 10 Zone 4 station (GenexusTM Strip 4) 11 Enrichment plate station 12 Liquid waste disposal port 13 Waste pipette tip disposal port 14 Sample plate station 15 GenexusTM Pipette Tips station 16 Robotic pipettor

22

GenexusTM Integrated Sequencer User Guide

Chapter 1 Product information GenexusTM Integrated Sequencer input and output connections

1

GenexusTM Integrated Sequencer input and output connections
The connection panel, power port, and an on/off switch are located on the right side of the rear panel of the instrument.
2

3

1

4

5

1 Power port--100­240VAC port that provides power to the instrument.
2 On/off switch--Power switch, where the states are on ( | ) or off ( O ).
3 USB ports--Connects a USB device to the instrument.
4 Ethernet port--An RJ45 port that provides Ethernet (Gigabit) communication between the sequencer and a local area network.
5 Ethernet port--An RJ45 port that provides Ethernet (Gigabit) communication between the sequencer and an accessory instrument.

GenexusTM Integrated Sequencer User Guide

23

1 Chapter 1 Product information Workflow
Workflow
Sample-to-results sequencing run using the GenexusTM Integrated Sequencer
Create an assay (page 32) System installed assays that are specifically configured for each sample type are available in GenexusTM Software. You can use the system-installed assays in your run without change. If you want to modify any assay settings, copy the system-installed assay that best represents your experiment, then edit the assay settings as needed.
Enter samples (page 46) Enter samples in GenexusTM Software to assign sample names and provide other information such as sample collection date, gender, type, and disease category.
Plan a sample run (page 62) Runs planned in GenexusTM Software contain all of the settings that are used in library preparation, templating, sequencing, and analysis, including sample information and plate location, assays, and barcodes.
Dilute the samples and load the sample plate (page 75) Quantify and dilute your nucleic acid samples, then load the sample plate.
Load the sequencer and start a run (page 79) Follow the step-by-step instructions on the sequencer touch screen to load the sample plate and consumables in the GenexusTM Integrated Sequencer.
Monitor the run (page 94) Monitor the run in GenexusTM Software in real time.
Review data and results in GenexusTM Software (page 95) Review data and results in GenexusTM Software, or analyze data in Ion ReporterTM Software using an Ion ReporterTM analysis workflow.

24

GenexusTM Integrated Sequencer User Guide

2

Before you begin

 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25  Guidelines for GenexusTM Integrated Sequencer operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26  Guidelines for expired reagents and chips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27  Power the GenexusTM Integrated Sequencer on or off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28  Get started with GenexusTM Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Precautions
Avoid nucleic acid contamination
IMPORTANT! A primary source of contamination is spurious DNA fragments from previous sample processing steps. Do not introduce amplified DNA into the work area where the instrument is located.
Avoid chip damage
IMPORTANT! To avoid possible damage due to electrostatic discharge, ground yourself before picking up a chip or placing a chip on a surface such as a lab bench. For example, touch the deck surface or one of the metal deck stations.
Avoid strong electromagnetic radiation
WARNING! Do not use the instrument in close proximity to sources of strong electromagnetic radiation (for example, unshielded intentional RF sources), as these sources can interfere with proper operation.
Protection by equipment
WARNING! The protection that is provided by the equipment can be impaired if the instrument is operated outside the environment and use specifications, the user provides inadequate maintenance, or the equipment is used in a manner that is not specified by the manufacturer (Thermo Fisher Scientific).

GenexusTM Integrated Sequencer User Guide

25

2

Chapter 2 Before you begin Guidelines for GenexusTM Integrated Sequencer operation

Guidelines for GenexusTM Integrated Sequencer operation

· Follow guidance that is provided by GenexusTM Software when you plan a run to determine which consumables must be loaded and which consumables can be reused from a previous run.
· Follow guidance that is provided by the software when you plan a run to determine how many samples can be run with a given assay or assays in an instrument run. The number of samples that can be included in a sequencing run depends on multiple factors.

Limiting factor
The number of available barcodes in the barcode plate

Description
The maximum number of available barcodes per run is 32. IMPORTANT! When libraries are prepared on the GenexusTM Integrated Sequencer, each target amplification reaction for a sample requires a unique barcode.

Maximum number of target amplification reactions per run

One library strip pair has the reagents necessary for 4 target amplification reactions, or 4 barcodes. With a maximum of 8 library strip pairs loaded, a maximum of 32 samples can be run using an assay with one primer pool.

The number of primer pools per assay

Given the limits of 32 target amplification reactions, and 32 available barcodes, the number of samples in a run multiplied by the total number of primer pools in the assays that are used in a run cannot exceed 32.
For one singlepool assay, a maximum of 32 samples can be run on a single chip. If you are using 2 assays with two primer pools each, you can sequence a maximum of 8 samples in a run. Similarly, for one assay with 4 primer pools, you can sequence a maximum of 8 samples in a run, if the minimum read count per sample allows it.

The number of unused lanes on an installed chip

A maximum of 4 lanes are available on a single GX5TM Chip.

The minimum read count per sample for an assay

The minimum read count per sample parameter is set during assay creation.

· Two assays cannot share a chip lane, so a maximum of 4 assays can be run per chip.
· The assays that are used in a single run must use the same chemistry (Ion AmpliSeqTM or Ion AmpliSeqTM HD), and have compatible cycling parameters to allow amplification in the instrument thermal cycler. The thermal cycler has two independently controlled heating zones. After you select an assay, GenexusTM Software restricts the list of available assays to use in the run to those that are compatible with the selected assay or assays.
· One library strip pair is required for each primer tube position 1­8 that is filled in a run.
· One template strip pair is required for every chip lane that is used in a run.
· Consumables are configured to support sample batch sizes in multiples of four samples. The most efficient use of consumables occurs when samples are run in multiples of four.
· If a chip installed in a sequencer has unused lanes, do not remove it unless you are sure that you want to replace it with a new chip. After a partially used chip has been removed from the sequencer, it cannot be reinserted and reused. The sequencer cannot track lane usage after chip removal.

26

GenexusTM Integrated Sequencer User Guide

Chapter 2 Before you begin Guidelines for expired reagents and chips

2

· You can remove a chip in one of the following situations.
­ After all the lanes of a chip are used in a run, the chip shuttles to the install position and you are asked to remove the used chip.
­ When you select a run plan that requires more lanes than are available on the installed chip, you are asked to remove the partially used chip, and the sequencer performs a post-chip clean. In addition, you must clear consumables from the lower sequencing reagents bay, even if only a single lane of the chip was used.
· The GenexusTM Integrated Sequencer can track used and unused barcodes in barcode plates in GenexusTM Software 6.2.0 and later, enabling you to swap plates between runs as needed, and reload a partially used barcode plate for a run if a sufficient number of barcodes are available on the plate.
· After loading in the sequencer, reusable consumables, such as barcodes, chips, and sequencing reagents bay components, must be used within 14 days for optimal results.
· An assay that is selected in a library run cannot include library batches that share one or more samples. However, two different assays in a run can include shared samples, because assays are run in separate lanes of a chip.

Guidelines for expired reagents and chips
Follow these guidelines for using reagents and sequencing chips that are at or near expiration. We do not recommend using components past the expiration date, but under certain circumstances, a sequencer warning for expired reagents can be overridden to allow a sequencing run to proceed.
· For all reagents except for GenexusTM barcode plates, the instrument bypasses an expired reagent warning after you tap the Help button in the sequencer screen.
· If a barcode plate has expired, a run cannot proceed, even if you tap Help.
· If a GX5TM Chip is expiring in a given month, ensure that you use up all lanes of the chip within that calendar month. If you start a run in the next calendar month with the same chip installed, the sequencer does not allow the run to proceed. The sequencer prompts you to perform a Clean instrument procedure and install a new chip and sequencing consumables, even if the initialization has not expired. For further information, see "Perform a Clean instrument procedure" on page 152.

GenexusTM Integrated Sequencer User Guide

27

2

Chapter 2 Before you begin Power the GenexusTM Integrated Sequencer on or off

Power the GenexusTM Integrated Sequencer on or off
Note: If the GenexusTM Integrated Sequencer is powered on, and the touchscreen is blank, touch the screen to "wake" the touchscreen.
Power on
If the touchscreen is unresponsive, check the power switch on the back of the instrument to ensure that the switch is in the on ( | ) position. If the power switch is in the off ( O ) position, proceed with step 1. If the power switch is already in the on position, proceed to step 2.
1. Turn the power switch on the back of the instrument to the on ( | ) position.
2. Press the power button on the front of the instrument. The button illuminates.
3. In the Sign In screen, enter the username and password created by the field service engineer when the instrument was set up. When the instrument home screen appears, the instrument is ready for use.
Power off
It is not necessary to power off the instrument overnight or over the weekend. If the instrument or GenexusTM Software will not be used for more than 3 days, power off the instrument as follows:
1. In the home screen, tap Settings4System Tools4Shut down.
2. Select either Shutdown or Reboot.
3. If you select Shutdown, a confirmation message appears. Select Yes to power off the instrument.
Note: If you power off the instrument with a partially used chip installed, the chip and consumables status is saved. When you power back on, the saved chip and consumable information enables you to use the chip for up to 14 days after the chip was installed.
IMPORTANT! Do not press the power button during a run. Interrupting power to the instrument during a run can result in sequencing run failure and loss of sample.

28

GenexusTM Integrated Sequencer User Guide

Chapter 2 Before you begin Get started with GenexusTM Software

2

Get started with GenexusTM Software

About the GenexusTM Software user interface

The GenexusTM Software user interface provides menus to help you add, select, and manage samples, libraries, runs, and assays. You can also view and manage your sequencing results, monitor GenexusTM Integrated Sequencer runs in progress, and manage software settings.

1

2

3

4

5

67

1 Samples: add new samples, import samples, prepare library batches, import library batches and manage attributes. 2 Runs: plan a run starting from a sample (Sample Run), or library (Library Run). View, edit, and manage runs. 3 Monitor: view a sequencer run in progress. 4 Results: view sample results, run results, and verification results. 5 Assays: manage, create, and import assays. Manage assay preset parameters and panels. 6 Settings: access audit records and run logs, configure network settings, and manage data archiving, disk space, and users. 7 Profile: access Help, manage and edit user profile settings and SSH key, sign out.

GenexusTM Integrated Sequencer User Guide

29

2

Chapter 2 Before you begin Get started with GenexusTM Software

User-access levels

Users at this level... Operator
Manager
Administrator

· Add and import samples · Prepare library batches · Plan and save runs · Monitor runs · View results and reports

Can...

Operator functions plus: · Create and edit sample attributes · Delete runs · Create and import assays · Manage reference sequences and panel, hotspot, and other sequence files · Access services information

Operator and manager functions plus: · View, export, and print audit records · Configure network settings · View and manage software updates · Configure data archive and storage settings · Manage instrument and software log files · Add and manage users

System tracking
The system tracks and checks user, sample, workflow, reagents and QC metrics for auditable records. If the software detects an error at any step--for example, a scanned barcode is inconsistent with the information given for the run--the software alerts the user and does not proceed with the run.
Request and sign in to a new account
Only administrator-level users can create user accounts. After account creation, the GenexusTM Integrated Sequencer automatically sends an email to the new user with the username and password information.
· To request a new account, contact your local administrator.
· To sign in to a new account for the first time: a. Open the GenexusTM Software, then enter your username and password.
b. Press Enter, or click Sign In.
c. Click Accept to accept the End User Software License Agreement.

30

GenexusTM Integrated Sequencer User Guide

Chapter 2 Before you begin Get started with GenexusTM Software

2

d. In the Change Password screen, enter your temporary password in the Current Password field. Type a new password in the New Password field, then confirm the password. ­ Passwords must be between 6 and 10 alphanumeric characters (0­9, Aa­Zz) with no spaces or special characters. ­ Passwords must contain at least one alphabetic character (Aa­Zz). ­ Passwords must contain at least one numeric character (0­9). ­ Passwords are case-sensitive.
e. Click Change.
Sign in
1. Open the software home page.
2. Select your preferred language from the dropdown list in the upper right corner of the page.
3. Enter your username and password, then press Enter or click Sign In. IMPORTANT! Your username and password must be unique and not shared with other users.
The software opens to the Manage Samples screen.

GenexusTM Integrated Sequencer User Guide

31

3

Create and manage assays (manager/administrator)

 About assays in GenexusTM Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32  System-installed assays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33  Manage assays (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33  Create a new assay (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35  Copy an assay (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43  Import an assay (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Manager- and administrator-level users can create and manage assays in GenexusTM Software. This chapter describes how to create, import, and manage assays, and how to copy and edit an assay, including a system-installed assay. If you are using a system-installed assay without change, proceed to Chapter 4, "Enter samples and libraries".
About assays in GenexusTM Software
Assays contain the settings and parameters for library preparation, templating, controlling the sequencing run, and analyzing the results. Assays also define the panels, kits, and chips that are used in a run, and specify the reference files and threshold values for quality control and variant detection. Assays can be created from system-installed templates or assays, or from custom assays that are copied and edited.
An assay is a reusable experimental design that contains predefined settings appropriate for use with common types of research applications. An assay can be used to plan many runs and plays an important role in enabling rapid throughput across the sequencing instrument. Assays help reduce the chance of errors, because information is stored and then applied to runs instead of entered manually for each run.
An assay can be copied and edited for a different use. Before you can create a custom assay, you must add a panel file, and hotspot and copy number baseline files (if needed for your assay), to the software. Custom assays are for advanced users. For assistance, contact your local Field Service Engineer.

32

GenexusTM Integrated Sequencer User Guide

3 Chapter 3 Create and manage assays (manager/administrator) System-installed assays

Assays can contain system-installed annotation sets, filter chains, copy number baselines and report templates. You can also create custom versions of each of these presets.
The software provides to tools to:
· Create, import, and manage assays. · Create and manage annotation sets, report templates, filter chains, and copy number baselines
(Manage Presets). · Add and manage panels (Manage Panels).

System-installed assays
GenexusTM Software includes system-installed assays that are preconfigured for use with OncomineTM GX assays. System-installed assays are available for download at Software Updates in the (Settings) menu. System-installed assays are locked and cannot be changed, but the assays can be copied, then edited.

Manage assays (manager/administrator)

Manager- and administrator-level users create and manage assays for use in GenexusTM Software.
In the menu bar, click Assays4Manage Assays to open the Manage Assays screen. The following tools are available in this screen.

To...
Review and export the assay audit trail

Do the following...
1. In the row of an assay, in the Actions column, click Audit. 2. In the Audit Trail dialog box, in the Record column, click (Audit Trail
Details) in the row of an action to view the details of that action. 3. In the Audit Record Details dialog box, click Export to export a PDF
file of the record.

Lock a draft assay Edit a draft assay

When a manager- or administrator-level user first creates an assay, in the Assay column, the assay is listed as assay name (Draft). To use the assay in a run, you must lock it. In the row of a draft assay, in the Actions column, for an assay with a status of Draft, click Lock. Locked assays cannot be edited or deleted.
When a manager- and administrator-level user first creates an assay, the assay name is followed by (Draft) in the Assay column. While the assay is in draft status, it can be edited. A locked assay cannot be edited.
1. In the row of a draft assay, in the Actions column, click Edit. The Create Assay workflow reopens.
2. Edit the options on each assay step as desired, then click Save.

GenexusTM Integrated Sequencer User Guide

33

3

Chapter 3 Create and manage assays (manager/administrator) Manage assays (manager/administrator)

(continued)

To...

Do the following...

Copy an assay to create a new assay Only locked assays can be copied.

1. In the row of an assay, in the Actions column, click Copy.
The Create Assay workflow reopens.
2. Edit the options on each assay step as desired, enter a new name for the assay, then click Save.

Delete an assay Remove a locked assay

Only draft assays can be deleted. When an assay is locked, it can be removed from use in the software by designating it obsolete. In the row of a draft assay, in the Actions column, click Delete, then confirm the deletion.
A manager-or administrator-level user can remove a locked assay from use in the software by designating it obsolete. The assay is not deleted and a record of it is maintained in the audit trail. The results for any runs already performed with the assay remain on the sequencer.
1. In the row of a locked assay, in the Actions column, click Obsolete.
2. Click Yes to confirm the operation.

Export an assay Download parameters

An assay can be exported, for example if you want to use that assay in another GenexusTM Integrated Sequencer in your lab. Only locked assays can be exported.
In the row of a locked assay, in the Actions column, click Export. The assay parameter files are downloaded to your local drive as a ZIP folder and are available for import to another sequencer.
Panel reference files are not included in the exported folder.
In the row of an assay, in the Actions column, click (More Options). Assay parameter files are downloaded to your local drive as a ZIP folder containing assay parameter JSON files.

34

GenexusTM Integrated Sequencer User Guide

Chapter 3 Create and manage assays (manager/administrator) Create a new assay (manager/administrator)

3

Create a new assay (manager/administrator)
Manager- and administrator-level users can create a new assay. Assays can be copied from an existing system-installed assay or other assay, then modified as needed. For more information, see "Copy an assay (manager/administrator)" on page 43. To create a new assay from an assay template, follow these steps.
1. In the menu bar, click Assays4Create Assay .

The Create Assay screen opens. Four oncology and one generic sequencing assay templates are available. The templates have assay-specific configuration steps that are prepopulated with default settings and parameters.

· Generic Sequencing Application

· DNA Somatic

· DNA Germline

· Fusions

· DNA and Fusions

Note: Clicking + Create Assay takes you to the same screen.

GenexusTM Integrated Sequencer User Guide

35

3

Chapter 3 Create and manage assays (manager/administrator) Create a new assay (manager/administrator)

2. Select the assay template that you want to use, then click anywhere inside its box.

Note:
· Although each assay template has a specific set of steps in the setup wizard, the setup
procedures for all are similar. The following is an example of assay creation after selecting the DNA and Fusions template.
· If you are using a custom assay, see "Guidelines for using custom assays with the GenexusTM
Integrated Sequencer" on page 176 for guidelines for setting parameters for Minimum Read Count Per Sample and target amplification in the Panel and Parameters steps.

36

GenexusTM Integrated Sequencer User Guide

Chapter 3 Create and manage assays (manager/administrator) Create a new assay (manager/administrator)

3

3. In the Panel step, make the following selections or entries, then click Next. a. The panel for the assay from the Panel list.
Note: To add a new panel, click Assays4Manage Panels, then click + Add New.

b. The hotspot file for the panel.
Note: To add a hotspot file, click Assay4Manage Panels4Hotspots, then click + Add New.
c. The CNV baseline for CNV calling, if used in the assay. Select Do Not Run CNV Algorithm if you are not performing CNV (copy number variant) analysis. To add a CNV baseline, click Assay4Manage Presets4Copy Number Baselines, then click + Create New or Import Copy Number Baseline.
d. (Optional) If you selected a CNV baseline, select a value to copy number based on tumor cellularity, if you want the CNV analysis to adjust for heterogenous tumor content in your samples. Samples with % cellularity below this value when created are rejected. If you do not want to make this adjustment, select Do Not Adjust CN and Do Not Reject Samples.
e. The extraction method type.
f. The minimum read count per sample coverage setting (required).

GenexusTM Integrated Sequencer User Guide

37

3

Chapter 3 Create and manage assays (manager/administrator) Create a new assay (manager/administrator)

g. The annotation set to use in variant reporting.
To add an annotation set, click Assays4Manage Presets4Annotation Sets, then click + Add New.
4. In the Reagent step, make the following selections or entries, then click Next. confirm or select names of library, barcode, templating, and sequencing kits, select the sequencing chip, and inline control check boxes, then select or enter values for template size and sequencing flows, if different from the prepopulated values. Click Next. a. Confirm or select names of library, barcode, templating, and sequencing kits.

Note: If you select Other AmpliSeq from the Library Kit dropdown list, the Barcode Kit dropdown list includes barcodes sets, such as IonCodeTM and Ion XpressTM barcodes, that are compatible with manually or Ion ChefTM-prepared libraries.
b. Select the sequencing chip.
c. Select one or both of the DNA and RNA inline control check boxes if you want to include inline controls in the quality control analysis.
d. Select or enter values for template size and sequencing flows, if different from the prepopulated values.

38

GenexusTM Integrated Sequencer User Guide

Chapter 3 Create and manage assays (manager/administrator) Create a new assay (manager/administrator)

3

5. In the QC step, enter parameters in the CF-1 Control and Run QC sections, and in the NTC (no template control) QC and Sample QC fields that appear appropriate to the sample types specified in the assay: NTC QC - DNA, NTC QC - RNA, Sample QC - DNA, and Sample QC - RNA. Leave parameters that are not applicable to your assay as Not Set. Click Next when finished.

Parameters for QC pass/fail thresholds

Parameter CF-1 Control Average Reads Per Lane Base Call Accuracy Mean AQ20 Read Length Run QC Key Signal

Description
The average number of CF-1 reads per chip lane. The percentage of accuracy of the CF-1 calls aligned to the CF-1 reference. The mean length of CF-1 reads having 99% accuracy at each position.
The average signal after software processing for all ISPs that identically match the library key (TCAG). A measure of the efficiency of template amplification.

GenexusTM Integrated Sequencer User Guide

39

3

Chapter 3 Create and manage assays (manager/administrator) Create a new assay (manager/administrator)

Parameters for QC pass/fail thresholds (continued)

Parameter

Description

Percent Loading

The percentage of addressable wells on a chip lane that are loaded with an ISP.

Raw Read Accuracy

The percentage of raw reads mapping to the reference sequence.

Sample QC - DNA

Deamination score

Deamination is reported as the estimated SNP proportion consistent with deamination (low allele frequency C:G>T:A SNVs). The deamination score can be used to determine the quality of an FFPE sample.

MAPD

The Median of the Absolute values of all Pairwise Differences; a quality metric that estimates coverage variability between adjacent amplicons in copy number variant (CNV) analyses. A MAPD value 0.5 generally indicates an acceptable level of coverage variability in the DNA Library or DNA Control.

Mapped Reads

The total number of bases mapped to target amplicons.

Mean AQ20 Read Length The mean length of sample reads aligned to a reference sequence that have 99%

(bp)

accuracy at each position.

Mean Read Length (bp) The mean length of all sample reads.

Uniformity of Amplicon Coverage

The percentage of reads showing a depth of coverage 20% of the mean base coverage.

Sample QC - RNA

Mapped Reads

The total number of reads mapping to a fusion reference sequence.

Mean AQ20 Read Length The mean length of called fusion reads that have 99% accuracy at each position. (bp)

Mean Read Length (bp) The mean length of all sample reads.

NTC QC- DNA

Average Base Coverage Depth

Average base coverage depth of reads aligned to a reference sequence in the no template control.

Mean Read Length (bp) The mean length of reads in the no template control.

NTC QC - RNA

Mapped Reads

The total number of reads in the no template control mapping to a fusion reference sequence.

Mean Read Length (bp) The mean length of reads in the no template control.

40

GenexusTM Integrated Sequencer User Guide

Chapter 3 Create and manage assays (manager/administrator) Create a new assay (manager/administrator)

3

6. In the Parameters step, review the prepopulated analysis settings, then modify if needed. Click Next when finished.
Note:
· Not all parameters are adjustable. To modify primary analysis parameters, select the Customize
Parameters checkbox.
· Select Yes under UDG Treat DNA in the Library Prep & Templating Parameters section to
include uracil DNA glycosylase (UDG) treatment of DNA during library preparation. Removal of uracil residues can increase sequencing quality for FFPE samples that have undergone significant cytosine deamination. We recommend using multiple samples in runs that include UDG treatment. Single sample runs can result in low read number.

Scrolling from the top, the settings are grouped in the following categories:

GenexusTM Integrated Sequencer User Guide

41

3

Chapter 3 Create and manage assays (manager/administrator) Create a new assay (manager/administrator)

· Library Prep & Templating Parameters

· CNV Finding

If needed, change the default settings for the cycling and input parameters used in library and template preparation.
· Primary Analysis Parameters

· Fusions · Read Mapping · Variant Finding

· Annotation

A parameter category can be quickly brought to the top of the screen by selecting it in the Filter Parameters by... list at the upper left.

Parameters can also be set by uploading an Advanced Parameter Configuration file, which overrides default settings. Click Upload , then click Select files... to navigate to this file on your drive and upload. Click Download to download parameters settings as a JSON file to your hard drive.
7. In the Plugins step, select plugins that you want to include in the sequencing data analysis from the plugin list, then click Next.

42

GenexusTM Integrated Sequencer User Guide

Chapter 3 Create and manage assays (manager/administrator) Copy an assay (manager/administrator)

3

8. In the Save step, enter a name and a short name for the assay, an optional description, then click Save.

The assay appears in the Manage Assays screen with the name you entered.
9. In the Manage Assays screen, click Lock in the Actions column of the assay to prevent changes to the assay. The assay must be locked to be available for use when planning a run.
Copy an assay (manager/administrator)
Manager- and administrator-level users can create a new assay by copying an existing system-installed assay or other custom assay and modifying it if needed. Only locked assays can be copied.
1. In the menu bar, click Assays4Manage Assays.
2. In the Manage Assays screen, in the Actions column for the assay that you want to copy, click Copy.

The Copy Assay screen opens to the Panel step. The assay settings can be modified for the new assay.
3. Proceed through the workflow steps, and modify assay settings if needed.

GenexusTM Integrated Sequencer User Guide

43

3

Chapter 3 Create and manage assays (manager/administrator) Import an assay (manager/administrator)

4. When finished, enter a new name and short name for the copied assay in the Assay Name field, then click Save. The newly created assay is added to the list of assays in the Assays / Manage Assays screen. The assay name is followed by (Draft) in the Assay column. The assay remains in draft status until it is locked.
5. In the Actions column, in the row of the assay, click Lock to enable its use in a run.

Import an assay (manager/administrator)
A manager- or administrator-level user can import an assay from another GenexusTM Integrated Sequencer if the assay has been first exported to a local drive.
1. In the menu bar, click Assays4Manage Assays.
2. In the Manage Assays screen, click Import Assay.

3. In the Import Assay dialog box, click Select Assay file, navigate to the exported assay ZIP folder on your drive, then select it.

44

GenexusTM Integrated Sequencer User Guide

Chapter 3 Create and manage assays (manager/administrator) Import an assay (manager/administrator)

3

4. Navigate to the exported assay ZIP folder on your drive, select it, then click Upload.

The assay appears in the list of assays in the Assays4Manage Assays screen.

GenexusTM Integrated Sequencer User Guide

45

4

Enter samples and libraries

 Create a new sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46  Import samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49  Manage samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51  Prepare or import a library batch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

Before planning a run in GenexusTM Software for either a sample run or a library run, you must first enter sample information in the software to assign sample names and provide other information. From the Samples menu, you can add samples in two ways. You can manually enter sample information or import sample information from a file.
Create a new sample
You can add a new sample in the Samples / Manage Samples screen. The new sample will be available to use in your run.
1. In the menu bar, click Samples4Manage Samples.
2. In the Manage Samples screen, click Create Sample.

3. In the Create Sample dialog box, complete the required fields.
Attributes identified with a red asterisk (*) in the Create Sample dialog box are required when adding a new sample. If attribute information is not available when adding a new sample, substitute mock information to complete the required fields.

46

GenexusTM Integrated Sequencer User Guide

4 Chapter 4 Enter samples and libraries Create a new sample

For more information, see "System-installed sample attributes" on page 47.
4. Click Save. The new sample is listed in the Manage Samples screen and will be available to use in your run.

System-installed sample attributes

The following table lists and describes system-installed sample attributes. System-installed sample attributes cannot be edited. Custom sample attributes are not listed in this table.

Sample attribute Sample Name[1]

Description
A unique identifier representing the sample.
The sample name can contain only alphanumeric characters (0­9, Aa­Zz), full stops/periods (.), underscores (_), or hyphens (-), cannot contain spaces, and is limited to a maximum of 20 characters.

IMPORTANT! To prevent erroneous sample selection during run planning, make sure that you assign a unique and distinguishable sample name for each sample.

Collection Date[1] Gender[1]

Note:
· Samples that have been used in a run cannot be deleted. · To prevent duplication, the software checks all sample names and returns an error
message if a non-unique sample name is detected.
The date that the sample was collected. Click Calendar to select the date in the correct format.
The biological sex of the sample: Female, Male, or Unknown.

IMPORTANT! Male or Female must be selected for proper measurement of AR CNV.

Sample Type[1]

A term that describes the sample, for example, FFPE, DNA, DNA & RNA. You can also select Other, then enter a custom sample type.

Disease Category[1] The disease type of the sample.
Note: If you select Cancer in this list, the Cancer Stage, Cancer Type, % Cellularity, and % Necrosis attributes listed below become available in the Add New Sample dialog box.

Cancer Type[1] Cancer Stage[1]

The type of cancer that is represented by the sample. Select the type of solid or hematologic cancer. If cancer type is unknown, select Unknown Primary Origin.
The stage of the cancer from which the sample was collected. Select Stage 0­IV, or Primary, Unknown, or Other.

GenexusTM Integrated Sequencer User Guide

47

4 Chapter 4 Enter samples and libraries Create a new sample

(continued)

Sample attribute

Description

% Cellularity

The percentage of tumor cells over normal cells in the sample. This is a whole number between 1 and 100. The % Cellularity attribute is applicable to FFPE samples only.

IMPORTANT!
· If not set, % Cellularity is assumed to be 100% in calculations that use this attribute. · % Cellularity is a required attribute for CNV analyses. Do not leave the field blank. · (FFPE samples only) If % Cellularity value is set to <100, then the magnitude of copy
number gain or loss can be decreased. For more information, see "CNVs table" on page 119.

% Necrosis
Notes
[1] Required attribute

The percentage of cellular necrosis in the sample. This is a whole number between 1 and 100.
An open-entry field for any additional sample information.

Create a custom sample attribute (manager/ administrator)

A manager- or administrator-level user can create custom sample attributes. Sample attributes, such as the date of collection and type of cancer, characterize and define a nucleic acid sample. A sample attribute can be made mandatory, in which case you are required to enter the attribute information for each new sample. Custom attributes can be used to further characterize samples.
Custom samples attributes cannot be edited or deleted after the samples that use them have been added to a library. To remove a custom sample attribute from use, see "Remove a sample attribute (administrator)" on page 49.
1. In the menu bar, click Samples4Manage Attributes.

2. In the Manage Attributes screen, click Add New Attribute.

3. Complete the fields in the Add New Attribute dialog box.
a. In Attribute Name, enter the name of the attribute.
Attribute names are limited to 20 alphanumeric characters (0­9 and Aa­Zz), full stops/periods (.), underscores (_), or hyphens (-).

b. In Data Type, specify whether the attribute is text or a number.

c. To require users to select the new attribute when adding or importing samples, click the toggle button to designate the attribute as Required.

d. Click Submit.

The new sample attribute is listed in the Attribute Name column and is available when you add a new sample.

48

GenexusTM Integrated Sequencer User Guide

4 Chapter 4 Enter samples and libraries Import samples
The new sample attribute is available in the Create Sample dialog box and in the Edit Sample dialog box, even for samples that were created before the new sample attribute was created. If the new sample attribute is a required attribute, it must be specified in the Edit Sample dialog box in order to save the changes. For more information, see "Edit a sample (manager/ administrator)" on page 52.
Remove a sample attribute (administrator)
An administrator-level user can remove user-created sample attributes from use in the software. Custom sample attributes that have been made obsolete can be reactivated. A record of how the attribute is used, removed, or changed, and by whom, is maintained in the audit trail of samples that are created using that attribute. You cannot designate as obsolete system-installed sample attributes.
1. In the menu bar, click Samples4Manage Attributes.
2. In the Manage Attributes screen, select a custom sample attribute to remove.
3. Click Obsolete in the Actions column, then confirm the action. Reactivate replaces Obsolete | Edit in the Actions column.
4. (Optional) To reactivate an attribute, click Reactivate in the Actions column. Active sample attributes are listed in the Add New Sample dialog box.
Import samples
Sample data files can be used to capture, manage, and edit sample data. You can import sample data files in the following formats: TXT, XLS, XLSX, or CSV. For a list of the sample attributes that are included in the import file, see "System-installed sample attributes" on page 47. For ease of use, you can download a MicrosoftTM ExcelTM example file to create an import file. For more information, see "Download an example samples file" on page 50.
1. In the menu bar, click Samples4Manage Samples, then click Import Samples.

2. In the Import Samples dialog box, click Select samples file.
3. Navigate to the file, then click Open.
4. Click Upload. A progress bar followed by an import report appears. If the import process fails, an error message indicates the reason for failure (for example, an invalid character was used). For troubleshooting, see "Batch sample import fails" on page 144.

GenexusTM Integrated Sequencer User Guide

49

4 Chapter 4 Enter samples and libraries Import samples Successfully imported samples are listed in the Samples / Manage Samples screen.
Download an example samples file
Download a MicrosoftTM ExcelTM example file to create a sample import file. The example file contains two tabs. The Instruction tab in the spreadsheet lists and indicates mandatory and optional attributes, which are the column headings in the Specimen Format tab. Use the Specimen Format tab to enter samples.
1. In the menu bar, click Samples4Manage Samples, then click Import Samples.
2. In the Import Samples dialog box, click Click here to download the MicrosoftTM ExcelTM example file.

The example file contains default sample attributes as columns. If custom sample attributes have been configured in the software, add these attributes as columns to the example file.
3. Save the file. You can now import the file in the Import Samples dialog box. For more information, see "Import samples" on page 49.

50

GenexusTM Integrated Sequencer User Guide

4 Chapter 4 Enter samples and libraries Manage samples
Manage samples
You can find tools for creating, searching, sorting, editing, deleting, and exporting samples, and for viewing the sample history in the Samples / Manage Samples screen.

Sort, search, and filter samples

Use the Samples / Manage Samples screen to sort, search, and filter the list of samples. The samples are listed by name with the most recently created sample at the top of the list.

· Use the Filter Samples by list and the Sample Name search box to search, sort, and filter the list of samples.

To
List only the samples that have not been extracted.
List only the samples that have not been prepared as a library.
List all samples.
Sort the samples list.

Do this In Filter Samples by, select To Be Extracted.
In Filter Samples by, select To Be Prepared.
In Filter Samples by, select All. 1. Click a column heading to sort the list by the entries in that column. 2. Click the column heading again to reverse the order.

Search for samples of interest.

1. In the Sample Name search box, enter the full or partial sample name.
2. Click (Search) or press Enter.
The sample or samples that match the sample name search entry are listed.
3. Click (Remove) to return to the complete list of samples.

GenexusTM Integrated Sequencer User Guide

51

4 Chapter 4 Enter samples and libraries Manage samples

Export samples

The Export function generates an XLS file that contains details about the selected sample.
1. In the menu bar, click Samples4Manage Samples.
2. In the Manage Samples screen, select the checkbox in the row of each sample that you want to export. To select all samples, select the checkbox in the column heading row.
3. Click Export. An XLS file is created that contains the details of the selected samples. Depending on your browser settings, the software automatically downloads the file or prompts you to open or save the file.
4. Open the XLS file in an appropriate viewer to review or print.

View notes or add a note to a sample

You can add notes to a specific sample or view existing notes. Use Notes to capture time-stamped written comments and observations for a sample. A sample can contain multiple notes from different users.
1. In the menu bar, click Samples4Manage Samples.

2. In the row of the sample of interest, in the Actions column, click Notes, then review existing notes or add a new note.

Action

Description

View existing notes for a sample.

In the Notes dialog box, notes that are associated with the sample are listed by user, date, and time.

Add a new note to a sample.

1. In the Notes dialog box, click Add Notes. 2. In Add Notes, enter the note, then click Save.

Edit a sample (manager/ administrator)
A manager- or administrator-level user can edit samples. In the Samples / Manage Samples screen, samples that can be edited are identified by the presence of the Edit link in the Actions column.
Samples can be edited at any point before a sample run that uses that sample is complete using the Edit link. After the sample is used in a run, the Edit link is no longer available. To edit a sample after a run is complete, use the Edit Sample & Amend Report link instead. For more information, see "Edit a sample and amend a report after a run (manager/ administrator)" on page 53.

52

GenexusTM Integrated Sequencer User Guide

4 Chapter 4 Enter samples and libraries Manage samples
To add a custom sample attribute to a sample, see "Create a custom sample attribute (manager/ administrator)" on page 48.
1. In the menu bar, click Samples4Manage Samples.
2. In the Manage Samples screen, in the Actions column, click Edit in the row of the sample of interest.
3. In the Edit Sample dialog box, edit the sample attributes. See "System-installed sample attributes" on page 47 for more information about the systeminstalled sample attributes.
4. Click Save.
Edit a sample and amend a report after a run (manager/ administrator)
After a sequencing run and its analysis have completed, a manager- or administrator-level user can edit a sample and amend the Lab Report for up to 30 days. Editing a sample after a run triggers an automatic update of the report and other files associated with the sample. In the Samples / Manage Samples screen, editable samples are identified by the presence of the Edit link in the Actions column. After 30 days, the Edit link is unavailable, only the Audit link is available.
1. In the menu bar, click Samples4Manage Samples.
2. In the Manage Samples screen, in the Actions column, click Edit in the row of the sample of interest.
3. In the Edit Sample & Amend Report dialog box, edit sample attributes as necessary. See "System-installed sample attributes" on page 47 for more information about system-installed sample attributes.
4. Click Save. The Edit Sample and Amend Report confirmation dialog box opens.
5. Click Yes to confirm the changes and continue. The sample information is edited, and all runs and reports associated with that sample are updated. Any test report, lab report, tab files and info.csv file that is associated with the sample will be updated with the changes.

GenexusTM Integrated Sequencer User Guide

53

4 Chapter 4 Enter samples and libraries Manage samples
Review sample edit history
The entire history of defining a sample is available to review and export using the audit feature. The history shows the original sample values and any new values or activities. The following actions generate an audit record:
· Create a new sample · Edit a sample · Obsolete a sample · Delete a sample This feature allows you to meet Title 21 CFR Part 11 of Federal Regulations that establishes the United States Food and Drug Administration regulations on electronic records and signatures, password policies, and user activity auditing.
1. In the Samples / Manage Samples screen, in the Actions column, click Audit in the row of the sample of interest. The Audit Trail Details window opens, listing each action that was performed for the sample, such as sample creation or edits.
2. (Optional) Click Download to export the audit record to a PDF file.
3. Click Back to close the Audit Trail Details window and return to the list of samples.
Delete samples
You can delete samples that have not been assigned to a run. Samples that are assigned to a run are locked and cannot be edited or deleted. Locked samples display Audit in the Actions column.
1. In the menu bar, click Samples4Manage Samples.
2. In the Manage Samples screen, select the checkbox in the row of each sample that you want to delete. To select all samples, select the checkbox in the column heading row.
3. Click Delete. The Delete Sample dialog box opens with the message Are you sure you want to delete selected sample(s)?
4. Click Yes to delete the selected samples.
The sample is removed from the Manage Samples screen.

54

GenexusTM Integrated Sequencer User Guide

Chapter 4 Enter samples and libraries Prepare or import a library batch

4

Prepare or import a library batch
A library batch is a group of prepared libraries that are sequenced in the same library run. If you are planning a run starting from libraries that you have already prepared manually, you must first create a library batch in GenexusTM Software from samples. You can enter samples in the software or import a sample file. For more information see:
· "Create a new sample" on page 46 · "Import samples" on page 49
Select the library batch when you plan the run. If you are planning a run starting from nucleic acid samples, skip this step and proceed to "Plan a sample run" on page 62.
Note:
· Each library in a library batch must have a unique library name. · When combining libraries in the same run, each library must have a unique barcode. · Fields identified with a red asterisk (*) are required.

Prepare a library batch
A library batch is a group of prepared libraries that are sequenced in the same library run. If you are planning a run starting from libraries that you have already prepared manually, you must first create a library batch in GenexusTM Software from samples that you have added. If you are planning a run starting from nucleic acid samples, skip this step and proceed to "Plan a sample run" on page 62.
1. In the menu bar, click Samples4Manage Samples.
2. In the Manage Samples screen, in the Filter Samples by... dropdown menu, apply the To Be Prepared filter to limit the displayed samples to those samples that have not been placed in a library batch. See "Sort, search, and filter samples" on page 51 for more information on the Filter samples feature.

GenexusTM Integrated Sequencer User Guide

55

4

Chapter 4 Enter samples and libraries Prepare or import a library batch

3. Select samples in the list by clicking the checkbox to the left of each sample, then click Prepare Library Batch.

4. In the Create Library batch screen, in Select Assay, select the assay that you want to run. The assay determines specific parameters of the run, including any required controls and post-run data analysis settings. Only locked assays appear in the Select Assay dropdown menu. For instructions to lock an assay, see "Manage assays (manager/administrator)" on page 33.
5. In the expanded screen, in Library Batch ID, enter a unique identifier for the library batch. Note: Library Prep Type: automatically fills for the nucleic acid type specified by the assay you selected: DNA, RNA, DNA+RNA, or TNA.

56

GenexusTM Integrated Sequencer User Guide

Chapter 4 Enter samples and libraries Prepare or import a library batch

4

Library Batch IDs can contain only alphanumeric characters (0­9, Aa­Zz), full stop/period (.), underscore (_), and hyphen (-). Required fields are indicated with a red asterisk (*).

6. Select the barcodes from the kit boxes into the appropriate fields.
7. Select the Include NTC checkbox to add no template control sample processing and reporting to the library batch.
8. Type a unique library name for each DNA and/or RNA library in the appropriate field. · Library names can contain only alphanumeric characters (0­9, Aa­Zz), full stop/period (.), underscore (_), and hyphen (-). · If your assay requires specific controls, they are automatically listed in the dialog box. These controls each require a unique barcode ID within the library batch, but do not require library names.
9. Select the barcode ID of the adapter used to prepare each library. If appropriate, swap the default barcodes in the dialog box between DNA, RNA, and Fusions by clicking the Swap Barcodes swap image. For example, click the DNA and Fusions swap button.

GenexusTM Integrated Sequencer User Guide

57

4

Chapter 4 Enter samples and libraries Prepare or import a library batch

Each library in a library batch must have a different barcode ID. When preparing the physical libraries, best practice is to swap barcodes between DNA and RNA libraries in consecutive sequencing runs to prevent carryover contamination. The barcodes that are listed in the DNA Barcode or RNA Barcode dropdown list belong to the barcode set that was selected when the assay was created.
IMPORTANT! Ensure that the actual barcodes that you used to create the libraries match the barcodes that you enter in the Create Library batch screen.
10. Enter the Input Quantity for each library.
11. Click Submit to save and submit your selections. The Manage Libraries screen opens, listing the library batch that you created. Libraries that are prepared in the same batch have the same Library Batch ID.
Import a library batch
You can import library batch information in the form of an .XLS or .XLSX file. The import file must include all required library and kit information.
1. In the menu bar, click Samples4Manage Libraries.

2. In the Manage Libraries screen, click Import Library Batch.
3. In the Import Library Batch dialog box, click Click here to select an assay for which the libraries are prepared, and to download an example file for import.

58

GenexusTM Integrated Sequencer User Guide

Chapter 4 Enter samples and libraries Prepare or import a library batch

4

4. Select an assay from the list, then click Download.
The assay name is auto-populated in the MicrosoftTM ExcelTM template file that downloads to your drive.

5. In the template file, enter or confirm the library batch information.

Template item Reagents tab Assay Name Library Batch ID
Library kit barcode
Panel kit barcode
Libraries tab Sample Name Library Name Barcode
Nucleic Acid Type Input Quantity No Template Control

Description
Auto-populated when assay is selected in step 4 (required) Must be alphanumeric (0­9, Aa-Zz), full stop/period (.), underscore (_), and hyphen (-) (required) For example, GenexusTM Library Strips 1 and 2AS barcode (optional) For example, OncomineTM Comprehensive Assay v3 DNA GX barcode (optional)
Same character requirement as Library Batch ID (required) Same character requirement as Library Batch ID (required) Barcodes used for each sample and control library preparation (required) DNA, RNA, or TNA (required) Library input quantity (optional) To Include a no template control, add a row with Sample Name as NTC, Library Name as NA, Barcode and Nucleic Acid Type similar to sample rows (optional)

IMPORTANT! For DNA+Fusions assays, the DNA library and RNA libraries must be listed in sequential order per pool for each sample. For example, for a 1pool DNA+Fusions assay, order should be DNA, RNA for sample 1, DNA, RNA for sample 2. For a 2pool DNA+Fusions assay, library order should be DNA, RNA (pool 1), DNA, RNA (pool 2) for sample1, then DNA, RNA (pool 1), DNA, RNA (pool 2) for sample 2.
6. Save the file.
7. Click Browse, navigate to the saved file, then select it.

GenexusTM Integrated Sequencer User Guide

59

4

Chapter 4 Enter samples and libraries Prepare or import a library batch

8. Click Upload.

A progress bar followed by an import report displays. If the import process fails, an error message indicates the reason for failure (for example, an invalid character was used). For additional troubleshooting support, see "Library batch import fails" on page 145.

60

GenexusTM Integrated Sequencer User Guide

5

Plan and manage runs

 Plan a sample run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62  Plan a library run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

Runs created in GenexusTM Software contain all of the settings that are used in library preparation, templating, sequencing, and analysis, including sample information and plate location, assays, and barcodes. Runs are used to track samples, consumables, and chips throughout the library preparation, templating, sequencing, and data analysis workflow.
You can plan runs for sequencing runs that use either nucleic acid samples (sample run) or libraries that you have previously prepared manually (library run) as input. GenexusTM Software guides you step-by-step to set up a run that tells you what consumables are needed, and provides a printed run setup guide to help you load the GenexusTM Integrated Sequencer with the required consumables.

GenexusTM Integrated Sequencer User Guide

61

5 Chapter 5 Plan and manage runs Plan a sample run
Plan a sample run
You can plan a run starting from isolated nucleic acid samples. Planning a sample run is organized into steps: Setup, Assays, Samples, Sample Plate, and Review. Progress through the steps is tracked in the upper left corner of the Runs / Plan Sample Run screen.
Before planning a run for a sample run, you must enter sample information into GenexusTM Software. For more information, see "Create a new sample" on page 46, or "Import samples" on page 49.
1. In the menu bar, click Runs4Plan Sample Run.
Note: You can also click Plan Sample Run in the Runs / Manage Runs screen.

62

GenexusTM Integrated Sequencer User Guide

5 Chapter 5 Plan and manage runs Plan a sample run
2. In the Setup step, enter or make the following selections. a. In the Plan section, enter a unique name.

b. (Optional) In the Reporting (Optional) section, select one or more options if needed. You can select both options, or leave both options deselected.

Reporting option Generate Report
Upload BAM files to Ion ReporterTM Software

Description
Select this option to generate a Lab Report using a report template that you specify in the list. To create a report template, click Assays4Manage Presets, then in the Report Templates tab, click Add New.
Select this option to automatically upload data for further analysis with Ion ReporterTM Software. You can also upload BAM files after a run if you leave this option unselected. Select your Ion ReporterTM Software account and release version. To configure an Ion ReporterTM Server account, see "Configure an Ion ReporterTM Server account (administrator)" on page 179. To configure an Ion ReporterTM Software on Connect account, see "Configure Thermo Fisher Accounts in GenexusTM Software (administrator)" on page 182.

c. Click Next.
If a chip is installed in the sequencer, the Chip View graphic in the lower left corner indicates the lanes that are available for sequencing.

GenexusTM Integrated Sequencer User Guide

63

5 Chapter 5 Plan and manage runs Plan a sample run
3. In the Assays step, select the assay or assays that you want to use in the run. Use the Filter Assays By list and the Assay Name search box to search, sort, and filter the list of assays.

Note:
· After selecting an assay, the list is filtered to show compatible assays that can be selected and
run at the same time.
· To create a new assay, see Chapter 3, "Create and manage assays (manager/administrator)".
IMPORTANT! Ensure that you select the assay that corresponds with the sample type that you will use in the run. If you select the wrong assay when you plan a run, the instrument will use incorrect settings during the run, resulting in invalid sequencing results. Available assays are listed in the Assays / Manage Assays screen.

64

GenexusTM Integrated Sequencer User Guide

5 Chapter 5 Plan and manage runs Plan a sample run
If you selected the Upload BAM files to Ion ReporterTM Software reporting option in substep 2b, make the following selections from the dropdown list in the Ion ReporterTM Workflow column in the row of each assay that you selected.
· Select Upload Only to upload sample data to Ion ReporterTM Software automatically upon run completion.
· Select the desired Ion ReporterTM analysis workflow to upload sample data and launch an analysis in Ion ReporterTM Software automatically upon run completion.
Note: In order for the Ion ReporterTM analysis workflow to appear in the list, you must tag the analysis workflow for use with the IonReporterUploader plugin. For more information, see "Tag an Ion ReporterTM Software analysis workflow for use with the IonReporterUploader plugin" on page 181.
4. In the Include NTC column for each assay that you select, click the Include NTC checkbox to include a no template control for the assay.
5. After you select an assay (or assays) and make the appropriate Ion ReporterTM Software selections (if applicable), click Next.
6. In the Samples step, select the samples from the list that you want to run with the assay, then click Assign.

The Chip View updates to show the lanes to be used in the run. Lane usage is calculated based on the number of samples (including a no template control, if selected), assay type, primer pools used, and minimum reads per sample entered at assay setup. Green denotes a chip lane that will be used in the run containing assigned samples within lane capacity. If the minimum reads per sample × the number of samples exceeds the chip or lane well capacity, a dialog box appears

GenexusTM Integrated Sequencer User Guide

65

5 Chapter 5 Plan and manage runs Plan a sample run
after you click Assign asking you to confirm that you want to continue. After confirmation, the Chip View updates and shows the lane color as red instead of green. The run is allowed, but you may not achieve the required reads per sample to pass QC metrics.

Green lane color denotes lane usage and sample assignment within lane capacity.

Red lane color denotes sample assignment that exceeds lane capacity.

7. If you selected more than one assay, repeat step 6 for each additional assay.
8. If needed, edit samples in one of the following ways, then click Next. · Click View & Remove, make your selections, then click Update. · Click Remove All, make your selections, then click Assign.

9. In the Sample Plate step, review sample positions in the sample plate. Drag and drop samples and no template controls to edit the location of samples and controls, if desired.

66

GenexusTM Integrated Sequencer User Guide

10. Modify the concentration of samples, if needed.

5 Chapter 5 Plan and manage runs Plan a sample run

Click Bulk Edit to modify sample concentration of all samples at one time, then click Submit to return to the Sample Plate screen.

If a sample concentration is 1,024X of the target concentration for the assay, which is displayed as a default value for each sample in the Sample Plate screen, the sequencer automatically dilutes the sample to the target concentration during the run. If a sample concentration is greater than this value, you must manually dilute the sample to the target concentration, or to a value within

GenexusTM Integrated Sequencer User Guide

67

5 Chapter 5 Plan and manage runs Plan a sample run
range for automated dilution before loading on the sample plate. For more information, see "Dilute or concentrate the samples (if needed) and load the sample plate--sample run" on page 76.
Note: The sample volume that is required for library preparation is not adjustable. The volume depends on the number of primer pools in the assay, sample type, and library chemistry. For more information, see "Dilute or concentrate the samples (if needed) and load the sample plate--sample run" on page 76 for specific sample volumes to load on the sample plate.
11. If sample plate information is correct, click Next.
12. In the Review step, review the run plan summary, then click Save & Print to print the run setup guide, if desired. Click Save to save the run without printing. The run plan summary lists the following details: · the consumables that are required for this run · how much sample volume to load · where to load samples and primer pool tubes · the sample concentration.

68

GenexusTM Integrated Sequencer User Guide

5 Chapter 5 Plan and manage runs Plan a sample run

1 2

3

4

5

8

6 7

1 Run information 2 List of consumables required for the run 3 List of consumables that are installed on the sequencer and
available for the run
4 Positions to load primer pool tubes 5 Chip view showing the lanes to be used in the run

6 Positions in the sample plate to load the samples
7 Table listing the sample plate position, sample type, volume to load, concentration, dilution factor, and assay for each sample
8 Description of each primer pool and its position

Note: If you are using an assay with Ion AmpliSeqTM HD library chemistry, the primer pool positions show that HD primer pools occupy both rows:

After saving, the run appears in the Manage Runs screen in the run list with the name you specified.
After selecting the run and loading the sequencer, the run is started on the sequencer screen.

GenexusTM Integrated Sequencer User Guide

69

5

Chapter 5 Plan and manage runs Plan a library run

Plan a library run
Before planning a library run, you must enter sample information and prepare a library batch in GenexusTM Software. The library batch selects the assay to be used in the run, and the assay in turn specifies the barcode set that was used to prepare the sample libraries. If your sample libraries were prepared using a barcode set not specified in an assay you want to use in the run, you must
1. create a new assay, or copy an existing assay, and specify the new barcode set in assay setup. 2. prepare a library batch that selects the new assay.
For more information, see "Create a new assay (manager/administrator)" on page 35, "Copy an assay (manager/administrator)" on page 43, and "Prepare a library batch" on page 55.
GenexusTM Software guides you through the four steps of planning a library run: Setup, Assays, Library Batches, and Review. Progress through the steps is tracked in the upper left corner of the Runs / Plan Library Run screen.

1. In the menu bar, click Runs4Plan Library Run. Note: You can also click Plan Library Run in the Runs / Manage Runs screen.

70

GenexusTM Integrated Sequencer User Guide

5 Chapter 5 Plan and manage runs Plan a library run
2. In the Setup step, enter a name for the run, then configure the reporting options. a. In Run Name, enter a unique name.

b. (Optional) In the Reporting (Optional) section, select one or more options if needed. You can select both options, or leave both options deselected.

Reporting option Generate Report
Upload BAM files to Ion ReporterTM Software

Description
Select this option to generate a Lab Report using a report template that you specify in the list. To create a report template, click Assays4Manage Presets, then click Add New.
Select this option to upload data for further analysis with Ion ReporterTM Software. You can also upload BAM files after a run if you leave this option unselected. Select your Ion ReporterTM Software account and release version. To configure an Ion ReporterTM Server account, see "Configure an Ion ReporterTM Server account (administrator)" on page 179. To configure an Ion ReporterTM Software on Connect account, see "Configure Thermo Fisher Accounts in GenexusTM Software (administrator)" on page 182.

c. Click Next.
If a chip is installed in the sequencer, the Chip View graphic in the lower left corner indicates the lanes that are available for sequencing.

GenexusTM Integrated Sequencer User Guide

71

5 Chapter 5 Plan and manage runs Plan a library run
3. In the Assays step, select the assay or assays that you want to use in the run, then click Next. Use the Filter Assays By list and the Assay Name search box to search, sort, and filter the list of assays.

Note: For the assay to be selectable at this step, you must have prepared a library batch that assigns the assay to the batch. For more information on preparing a library batch, see "Prepare a library batch" on page 55. The assay specifies the barcode set that was used to prepare the sample libraries. To create a new assay, or copy an existing assay, see "Create a new assay (manager/administrator)" on page 35, or "Copy an assay (manager/administrator)" on page 43.
IMPORTANT! Ensure that you select the assay that corresponds with the sample type that you will use in the run. If you select a wrong assay when you plan a run, the instrument will use incorrect settings during the run, resulting in invalid sequencing results. Available assays are listed in the Assays / Manage Assays screen.
If you selected the Upload BAM files to Ion ReporterTM Software reporting option in substep 2b, make the following selections from the dropdown list in the Ion ReporterTM Workflow column in the row of each assay that you selected.
· Select Upload Only to upload sample data to Ion ReporterTM Software automatically upon run completion.
· Select the desired Ion ReporterTM analysis workflow to upload sample data and launch an analysis in Ion ReporterTM Software automatically upon run completion.
Note: In order for the Ion ReporterTM analysis workflow to appear in the list, you must tag the analysis workflow for use with the IonReporterUploader plugin. For more information, see "Tag an Ion ReporterTM Software analysis workflow for use with the IonReporterUploader plugin" on page 181.

72

GenexusTM Integrated Sequencer User Guide

Chapter 5 Plan and manage runs Plan a library run

5

4. In the Library Batches step, select the library batch or batches that you want to use in the run.
Note: Only one library batch can be selected per assay. However, you can plan a multi-assay library run if you select multiple, different assays in the Assays step.

The Chip View updates to show the lanes to be used in the run as green. Lane usage is calculated based on the number of samples, assay type, primer pools used, and minimum reads per sample entered at assay setup.

Green lane color denotes lane usage and sample assignment within lane capacity.

Red lane color denotes sample assignment that exceeds lane capacity.

If the minimum reads per sample × the number of samples exceeds the chip or lane well capacity, a dialog box appears after you click Next asking you to confirm that you want to continue. After clicking Yes, the Chip View updates and shows the lane color as red instead of green. In the example shown at right, seven samples were included in a library batch instead of six. The run is allowed, but you may not achieve the required reads per sample to pass QC metrics.

GenexusTM Integrated Sequencer User Guide

73

5

Chapter 5 Plan and manage runs Plan a library run

5. After you select a library batch (or batches), click Next.
6. In the Review step, review the run plan summary, then click Save and Print to print the run setup guide, if desired. Click Save to save the run without printing. The run plan summary lists the consumables that are required for the run, where to load the library batch on the sample plate, and how much library volume to load.

1 2
3 4
7

6

1 Run information

5 Chip view showing the lanes to be used in the run

2 List of consumables required for the run

6 Table listing assay and run information

3 List of consumables installed on the sequencer and available for the run

7 Table listing the well position, library batch ID, volume to load, and assay for each library batch

4 Position(s) in the sample plate to load the library batch

After saving, the run appears in the run list on the Manage Runs screen with the name you

specified.

The run is started on the sequencer screen after selecting a run and loading the sequencer.

74

GenexusTM Integrated Sequencer User Guide

6

Dilute the samples and load the sample plate

 Guidelines for nucleic acid isolation and quantification--sample runs . . . . . . . . . . . . . . . . . . . . . . . 75  Dilute or concentrate the samples (if needed) and load the sample plate--sample run . . . . . . . . . 76  Guidelines for library quantification--library runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77  Dilute and pool libraries, and load the sample plate--library run . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Before starting a run on the instrument, you must quantify and dilute the samples or sample libraries, then load the sample plate.
Guidelines for nucleic acid isolation and quantification-- sample runs
These are general guidelines for isolating and quantifying DNA and RNA. For assay-specific guidelines, see the assay user guide.
· See "Recommended materials for nucleic acid isolation and quantification" on page 19 for recommended kits for isolating DNA and RNA.
· We recommend the TaqManTM RNase P Detection Reagents Kit (Cat. No. 4316831) for quantifying amplifiable human genomic DNA (see Demonstrated Protocol: Sample Quantification for Ion AmpliSeqTM Library Preparation Using the TaqManTM RNAse P Detection Reagents Kit (Pub. No. MAN0007732) available at thermofisher.com).
· The QubitTM dsDNA HS Assay Kit (Cat. No. Q32851 or Q32854) can also be used for quantification, particularly for FFPE DNA, and highly degraded DNA samples. See "Quantify FFPE DNA with the QubitTM Fluorometer" on page 171 for a detailed procedure for quantifying FFPE DNA.
· We recommend the QubitTM RNA HS Assay Kit (Cat. No. Q32852 or Q32855) for quantifying RNA. · Quantification methods such as densitometry (for example, using a NanoDropTM spectrophotometer)
are not recommended, because they are not specific for DNA or RNA. Use of these methods can lead to gross overestimation of the concentration of sample nucleic acid, under-seeding of the target amplification reaction, and low library yields. · The Ion AmpliSeqTM Direct FFPE DNA Kit bypasses nucleic acid isolation when preparing libraries from FFPE sections on slides. See the Ion AmpliSeqTM Direct FFPE DNA Kit User Guide (Pub. No. MAN0014881) for a protocol for using this kit to prepare gDNA from FFPE tissue. · The Direct FFPE DNA preparation can be stored for up to 6 months at ­30°C to -10°C before library preparation.

GenexusTM Integrated Sequencer User Guide

75

6

Chapter 6 Dilute the samples and load the sample plate Dilute or concentrate the samples (if needed) and load the sample plate--sample run

Dilute or concentrate the samples (if needed) and load the sample plate--sample run

Isolate DNA and RNA samples using one of the procedures and kits that are recommended in "Recommended materials for nucleic acid isolation and quantification" on page 19.
Samples with concentrations up to 1,024X of the target concentration for an assay (displayed as default values in the Sample Plate step screen in run planning) are in range for automated dilution and require no manual dilution. Enter the concentrations during sample run planning at the Sample Plate step (see step 9 on page 66).

1. For samples with concentrations that are out of range for automated dilution, manually dilute the sample with nuclease-free water, or concentrate the sample to a concentration 1,024X of the target concentration. For samples that are in range, go to step 2.

If the sample concentration is...

Then...

<0.11 ng/µL

Concentrate the sample to greater than or equal to the target concentration.

0.11 ng/µL, but less than the target concentration

Run is allowed but sample concentration may not be optimal for library preparation. Concentrate the sample to greater than or equal to the target concentration.

1,024X of the target concentration

No manual dilution is necessary. The sequencer dilutes the sample to the target concentration automatically during the run.

>1,024X of the target concentration

Manually dilute to the target concentration based on assay type, or to a concentration in range for automated dilution by the sequencer.

Note:
· If you enter a concentration <0.11 ng/µL or >10,000 ng/µL target concentration, a warning that
the concentration is out of range appears, and you are not allowed to proceed to the next step.
· If the concentration is 10,000 ng/µL, but >1,024X of the target concentration, you can proceed,
but because the instrument cannot dilute samples more than 1,024fold, the diluted sample concentration will be greater than the target concentration.

2. Add samples to the sample plate at the volume and positions that are specified in the run setup guide.
The sample volume is not adjustable and depends on sample type, the number of primer pools in the assay, and library chemistry. The following table also provides loading volume.

Sample type
DNA DNA RNA RNA

Number of primer pools Ion AmpliSeqTM chemistry
1 2 1 2

Volume
15 µL 25 µL 15 µL 25 µL

76

GenexusTM Integrated Sequencer User Guide

Chapter 6 Dilute the samples and load the sample plate Guidelines for library quantification--library runs

6

(continued) Sample type
DNA RNA TNA

Number of primer pools Ion AmpliSeqTM HD chemistry
1 1 1

Volume
20 µL 20 µL 20 µL

3. Seal the plate with a sheet of Adhesive PCR Plate Foils (Thermo Fisher Scientific Cat. No. AB0626). Note: The use of other plate seals may affect performance.

4. Keep the plate on ice until ready to load it in the sequencer.

Guidelines for library quantification--library runs
· We recommend that you use libraries that are freshly quantified and diluted before pooling in a library batch.
· Pre-prepared libraries can be quantified by one of the following three methods: ­ Quantification using the AgilentTM 2100 BioanalyzerTM instrument ­ Quantification using the QubitTM Fluorometer ­ Quantification by qPCR using the Ion Library TaqManTM Quantitation Kit
See one of the following guides for specific procedures. ­ Ion AmpliSeqTM Library Kit 2.0 User Guide (Pub. No. MAN0006735) ­ Ion AmpliSeqTM Library Kit Plus User Guide (Pub. No. MAN0017003) ­ Ion AmpliSeqTM HD Library Kit User Guide (Pub. No. MAN0017392)

GenexusTM Integrated Sequencer User Guide

77

6

Chapter 6 Dilute the samples and load the sample plate Dilute and pool libraries, and load the sample plate--library run

Dilute and pool libraries, and load the sample plate--library run
1. Dilute each manually prepared and quantified sample library to 200 pM with nuclease-free water.
Note: Each library must be barcoded with a unique barcode or barcode pair. Use this concentration as a starting point, then titrate up or down based on sequencing results, if needed.
2. Add equal volumes of each library to a new 1.5mL low DNA retention tube so that the total volume is greater than the volume specified in the run setup guide provided by the software. Note: For information on combining DNA and RNA libraries recovered from sample runs using assays that include DNA and fusions, see "Combine libraries" on page 175.
3. Mix well by pipetting up and down five times, then transfer the specified volume of each library batch to the sample plate position specified in the run setup guide.
4. Seal the plate with a sheet of Adhesive PCR Plate Foils (Thermo Fisher Scientific Cat. No. AB0626).
Note: The use of other plate seals may affect performance.
5. Keep the plate on ice until you are ready to load it in the sequencer.

78

GenexusTM Integrated Sequencer User Guide

7

Load the sequencer and start a run

 Before you begin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79  Fill GenexusTM Primer Pool Tubes (custom assays only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81  Load the sequencer and start a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83  Clear the instrument deck and perform a UV Clean . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90  Options for an expired sequencer initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
After you have planned a run in GenexusTM Software, use the run setup guide provided by the software to load samples in the sample plate, and to determine which consumables to load in the sequencer. Follow the stepbystep instructions in the sequencer touchscreen during run setup. The vision system of the sequencer tracks the addition of consumables in real-time and alerts you if a component is loaded in an incorrect position, or if an incorrect quantity is loaded.
Before you begin
1. Remove the library and templating strips from their boxes in the refrigerator or freezer, and ready them for loading in the sequencer. · GenexusTM Strip 1 and GenexusTM Strip 3GX5TM: equilibrate to room temperature for 30 minutes. · GenexusTM Strip 2AS or GenexusTM Strip 2HD, depending on your assay, and GenexusTM Strip 4: thaw at room temperature for 30 minutes. If you are delayed in loading, keep the thawed strips on ice or at 4 until you load them in the sequencer.
IMPORTANT! Confirm that the strip contents are completely thawed before installing in the sequencer.
2. Visually check tube 3 of the GenexusTM Strip 2HD for precipitation. If needed, flick the tube or gently vortex the strip to dissolve the precipitate.
3. Remove primer pool tubes in tube carriers that are needed for the run from the freezer, then thaw for at least 30 minutes on ice or at 4. After thawing, gently tap the primer pool tube or tubes on a bench surface to ensure that contents are collected at the bottom of the tubes. Keep the tubes and carriers on ice or at 4 until you load them in the sequencer.
4. If you are installing a new GenexusTM Cartridge, thaw the cartridge at room temperature for 30 minutes before installing in the sequencer.

GenexusTM Integrated Sequencer User Guide

79

7 Chapter 7 Load the sequencer and start a run Before you begin
5. GenexusTM Strip 1 and GenexusTM Strip 3GX5TM contain magnetic beads in one or two positions, yellow or brown in color, that sometimes get trapped in the upper "keyhole" of the tube. Dislodge these beads from the keyhole before installing the strip in the sequencer. Use the following procedure for each strip. a. Invert the strip 3­4 times to dislodge beads that are trapped in the keyholes.
b. To remove any remaining beads and liquid from the keyholes, grasp the strip at one end with the strip seal facing up, then swing the strip with a rapid, downward centrifugal arm motion, ending with a sharp wrist-flick.
c. Grasp the strip at the other end, then repeat the centrifugal motion.
d. Check tube positions for significant amounts of beads that are still trapped in keyholes (see the following figure), then repeat the centrifugal motion, if needed. It is acceptable if a few beads remain in the keyhole or on the tube wall, but most should be either in suspension or in a pellet at the bottom of the tube.
1

2

Example GenexusTM Strip 3GX5TM before (upper) and after (lower) inversion. The carrier has been removed to show tube contents more easily.

1 Magnetic beads trapped in keyholes

2 Magnetic beads dislodged from keyholes

80

GenexusTM Integrated Sequencer User Guide

Chapter 7 Load the sequencer and start a run Fill GenexusTM Primer Pool Tubes (custom assays only)

7

Note:
· It is not necessary to resuspend the magnetic beads completely--it is only necessary to
dislodge most of the beads that may be trapped in the keyhole. The instrument resuspends the beads during the run when needed.
· Fine bubbles can form above the liquid in some tubes after inversion. These bubbles do not
affect the run.
6. Inspect all strips for large bubbles lodged under the surface of the liquid or at the bottom of each tube or well. Gently tap the strips on a benchtop to dislodge any bubbles without splashing the contents onto the upper tube walls. If tapping fails to dislodge a bubble, use the technique that is described in substep 5b until large bubbles are dislodged.
Fill GenexusTM Primer Pool Tubes (custom assays only)
If you are using a custom assay, GenexusTM Primer Pool Tubes must be manually filled with the custom Ion AmpliSeqTM or Ion AmpliSeqTM HD panels at the appropriate volume and in the correct primer pool tube positions. For Ion AmpliSeqTM library panels, use one carrier per DNA or RNA assay primer pool. The two positions in the primer pool tube carrier are designated as shown in the following figure:

1

2

1 Position 1 tube: Contains Ion AmpliSeqTM DNA, Ion AmpliSeqTM RNA, or Ion AmpliSeqTM HD FWD primer pool 2 Position 2 tube: Contains Ion AmpliSeqTM HD REV primer pool
Note: When you order assays from Ion AmpliSeqTM Designer (AmpliSeq.com), be sure to order a sufficient amount of panel for your needs, and request the tube format, not the plate format. Library preparation on the GenexusTM Integrated Sequencer requires greater panel volume per sample than manual library preparation, or library preparation on the Ion ChefTM System.

GenexusTM Integrated Sequencer User Guide

81

7

Chapter 7 Load the sequencer and start a run Fill GenexusTM Primer Pool Tubes (custom assays only)

1. Add primer pool at the indicated volume, appropriate to your assay type, to the GenexusTM Primer Pool Tubes using the following tables as a guide. Fill the number of tubes specified by the run plan summary.

Ion AmpliSeqTM DNA assays

Number of primer pairs per pool

Concentration

Volume in position 1

Volume in position 2

12­96

2X (400 nM)

140 µL

--

97­3,072

2X (100 nM)

140 µL

--

>3,072

2X ([3,072 / Number of primer

140 µL

--

pairs per pool] × 100 nM)[1]

[1] For example, if a panel pool has 3,500 primer pairs, the 2X concentration is (3,072 / 3,500) × 100 nM = 87.8 nM.

Ion AmpliSeqTM RNA assays

Number of primer pairs per pool

Concentration

Volume in position 1

Volume in position 2

12-1,228

5X (250 nM)

75 µL

--

>1,228

5X ([1,228 / Number of primer

75 µL

--

pairs per pool] × 250 nM)[1]

[1] For example, if a panel pool has 1,500 primer pairs, the 5X concentration is (1,228 / 1,500) × 250 nM = 205 nM.

Ion AmpliSeqTM HD assays

Primer pool type
Ion AmpliSeqTM HD FWD Ion AmpliSeqTM HD REV

Concentration
10X 10X

Volume in position 1
50 µL
--

Volume in position 2
--
50 µL

IMPORTANT!
· If you are using Ion AmpliSeqTM library chemistry, leave the tube in position 2 empty and
uncapped, but do not remove the tube from the carrier before loading in the sequencer. Do not add a second Ion AmpliSeqTM primer pool to the position 2 tube.
· If you are using Ion AmpliSeqTM HD library chemistry, add the FWD and REV primer pools to the
appropriate tubes in the same carrier.
· Ensure that no bubbles are introduced at the bottom of the tube when adding the primer pool.

2. If you do not install the primer pool tube carriers in the sequencer immediately, cap the tubes that contain primer pools, then store the tube carriers on ice. Remember to uncap all tubes before installing.

82

GenexusTM Integrated Sequencer User Guide

Chapter 7 Load the sequencer and start a run Load the sequencer and start a run

7

Load the sequencer and start a run
1. Tap Run on the sequencer home screen to start the loading procedure.

2. In the Run Selection screen, select the run that you want to use from the list.

Note: If you select a run that requires more lanes than are available on a currently installed chip, a dialog appears giving you the option to install a new chip, or cancel. If you proceed with a new chip, a postChipClean is performed, then the sequencer prompts you to perform the Clear Deck, UV Clean, Load Deck, Clear Sequencing Reagents, and Load Sequencing Reagents steps.

GenexusTM Integrated Sequencer User Guide

83

7

Chapter 7 Load the sequencer and start a run Load the sequencer and start a run

3. In the Review Run screen, confirm the run and assay selections, then tap Next.

The deck door opens automatically.
Note:
· If the instrument vision system detects consumables loaded on the deck, the sequencer
prompts you to remove the consumables, then starts a UV Clean.
· Select the Do Force Clean checkbox if there will be an unused lane or lanes on the installed
chip after the run, but you want to start your next run on a new chip after the current run. A force clean automatically cleans the instrument after the run, eliminating the need for an operator to execute the cleaning procedure between the completion of the current run and the next run. Selecting Do Force Clean renders all lanes of the installed chip unusable after the run.
4. In the Load Deck screen, the sequencer instructs you step by step to load each required consumable in a highlighted position on the deck. The sequencer detects the loading of each consumable in real time and advances to the next component automatically.

84

GenexusTM Integrated Sequencer User Guide

Chapter 7 Load the sequencer and start a run Load the sequencer and start a run

7

IMPORTANT!
· Ensure that you remove the primer pool tube cap or caps before installing the tube carrier on the
deck.
· Ensure that you load the correct type of barcode plate and library strip 2 for the type of run
you are setting up. The sequencer displays a warning if you have installed consumables that are incompatible with the run you have selected, for example, a GenexusTM Barcodes AS plate or GenexusTM Strip 2AS in an HD run.

Note:
· A primer pool tube carrier can only be installed with
the position 1 tube in the back row of the Primer Pool Tube Station. Follow the guidance in the run setup guide for loading the primer pool tube carrier or carriers in the correct position and order in the station.
· If the sequencer cannot read the correct loading of
an unexpired consumable, tap Help in the lower left corner of the screen to override the block. After using this override, the name of the consumable will not appear in the run summary consumables list.

1

2

1 Position 1 2 Position 2

5. If prompted, insert a new GX5TM Chip and GenexusTM Coupler. Insert the chip into the chip install slot with the chip notch oriented down and toward the front of the instrument.

1

1 Notched corner of chip
Note: A chip shuttle under the deck moves the installed chip to loading and sequencing positions during the run.
IMPORTANT! Insert the GenexusTM Coupler so that it is level to ensure it will properly align with the GX5TM Chip. A coupler that is installed at an angle or is not level will not align properly to the chip and can result in a failed run.

GenexusTM Integrated Sequencer User Guide

85

7

Chapter 7 Load the sequencer and start a run Load the sequencer and start a run

6. When the deck consumables have been loaded, lock the library and templating strips in place by sliding the latches toward the rear of the deck.

If a chip is detected and the strip latches are closed, the Close Deck Door screen appears. 7. Close the deck door, then tap Next.

· If you installed a new chip in the sequencer, the sequencer prompts you to open the sequencing reagents bay doors to empty the waste and remove used sequencing reagents bay consumables. Proceed to step 8.
· If you are using a chip that was previously installed and has sufficient lane capacity for the run, the sequencer prompts you to start the run.
IMPORTANT! The cartridge and bottles in the sequencing reagents bay must be replaced every time that a new chip is installed, regardless of how many lanes were used in the previous chip.

86

GenexusTM Integrated Sequencer User Guide

Chapter 7 Load the sequencer and start a run Load the sequencer and start a run

7

8. Follow on-screen instructions to empty the waste in the Waste Carboy, remove waste pipette tips, remove the used GenexusTM Bottle 1, GenexusTM Bottle 2, GenexusTM Bottle 3, and GenexusTM Cartridge, then tap Next.

IMPORTANT!
· Ensure that you empty and replace the Waste Carboy and the waste pipette tip bin. · After replacing the emptied Waste Carboy, ensure that you reinsert the waste tube into the
carboy.
· Follow all applicable local, state/provincial, and/or national regulations when recycling or
disposing of consumables and liquid waste.
9. Install a new GenexusTM Bottle 1, GenexusTM Bottle 2 (two required), GenexusTM Bottle 3, and GenexusTM Cartridge.

Note: The installed reagents can be used for up to 14 days on the sequencer with full performance. After 14 days, you may observe reduced performance.
After reagents have been installed, the Close Sequencing Reagent Bay Door screen appears.

GenexusTM Integrated Sequencer User Guide

87

7

Chapter 7 Load the sequencer and start a run Load the sequencer and start a run

10. Close the sequencing reagents bay doors.

After the doors are closed, the sequencer automatically starts the run. IMPORTANT! Do not tap Start Run. Tapping Start Run can cancel the run.
At the beginning of the run, the instrument verifies the chip, checks for leaks, then calculates run time.

88

GenexusTM Integrated Sequencer User Guide

Chapter 7 Load the sequencer and start a run Load the sequencer and start a run

7

A sequencing run encompasses the following stages:

· Starting

· Pre-sequencing

· Initializing

· Sequencing

· Library Prep

· Cleaning

· Templating

At each stage, the instrument shows the time remaining on the touchscreen.

Note: The time remaining shown on the screen does not include run analysis time.

When the run finishes, the sequencer displays the Run Complete screen.
Note: If all the lanes of a chip are used, the chip shuttles to the install position. You are asked to remove the chip and coupler, and clear the sequencing reagents.

GenexusTM Integrated Sequencer User Guide

89

7

Chapter 7 Load the sequencer and start a run Clear the instrument deck and perform a UV Clean

Clear the instrument deck and perform a UV Clean
After a run completes, remove used consumables from the deck and perform a UV Clean to ready the instrument for the next run.
1. In the Run Complete screen, tap Next to start removal of used consumables.

The deck door opens.
2. In the Clear Deck screen, the sequencer provides stepbystep instructions by highlighting the components to be removed. Unlock the library and templating strips by sliding the latches toward the front of the deck, then remove the used strips. Remove the remaining deck components specified by the sequencer.

90

GenexusTM Integrated Sequencer User Guide

Chapter 7 Load the sequencer and start a run Clear the instrument deck and perform a UV Clean

7

3. Inspect the GenexusTM Filter in the liquid waste disposal port and verify that no standing liquid is present. If standing liquid is present, manually remove the liquid with a pipette, then pull out the filter. Test the filter with water to determine if a clog is present.
· If the GenexusTM Filter is clogged, replace it with a new filter. For more information, see "Replace the GenexusTM Filter" on page 164.
· If the GenexusTM Filter does not appear to be clogged, a line clog downstream of the filter is implicated. Contact Technical Support and report a possible deck liquid waste line clog.
4. When finished, close the deck door, then tap Next.

A two-minute UV Clean starts.

GenexusTM Integrated Sequencer User Guide

91

7

Chapter 7 Load the sequencer and start a run Clear the instrument deck and perform a UV Clean

5. After UV cleaning, if all the chip lanes were used, the sequencing reagents bay doors unlock. Open the doors, remove used components from the bay and empty the Waste Carboy, then tap Next.

IMPORTANT! Do not discard or remove the conical bottles, unless alerted by the sequencer to replace the bottles after a conical bottle flow rate test. For more information, see "Replace the GenexusTM Conical Bottles" on page 165.
IMPORTANT! Follow all applicable local, state/provincial, and/or national regulations when recycling or disposing of GenexusTM Integrated Sequencer consumables and liquid waste.
CAUTION! The GenexusTM Bottle 1 (small waste bottle) contains small amounts of formamide. Dispose of this waste appropriately.
6. After removal of used components, close the sequencing reagents bay doors, then tap Next. The sequencer returns to the home screen.

92

GenexusTM Integrated Sequencer User Guide

Chapter 7 Load the sequencer and start a run Options for an expired sequencer initialization

7

Options for an expired sequencer initialization
A sequencer initialization is defined by the installation of a new chip, chip coupler, sequencing bottles, and reagent cartridge on the sequencer before a run. Reagents are stable on the sequencer for 14 days, after which you may experience reduced performance. An initialization expires after 28 days, after which the following warning appears. Three options are available: Cancel, Ignore, and Proceed.

The outcome of each option is described in the following table.

Option Cancel
Ignore
Proceed

Outcome
The sequencer returns to the home screen and the run is canceled. The expired initialization warning appears again if a run is started on the same initialization. After tapping Cancel, select Settings4Clean instrument to clean the instrument before starting a run with a new chip. Lane assignment for a new run starts with lane 1.
The instrument moves forward with the run and ignores the expired initialization warning. The warning appears for each run that uses the expired initialization until the instrument is cleaned and ready for a new initialization. Troubleshooting efforts are not supported on runs that start 28 days after an initialization or later.
The instrument starts a clean before run setup. After cleaning, you are asked to remove the used chip and coupler, and to load the deck for a new run. You are also asked to remove the used sequencing bottles and cartridge and install new consumables. A new run starts.

IMPORTANT! We strongly recommend that you not tap Proceed, and instead use Cancel or Ignore. After tapping Proceed, the lane assignment for the new run after the clean remains the originally assigned chip lane before the clean (not necessarily lane 1). Subsequent runs will assign the next available lane. For example, if the originally assigned chip lane was lane 3, lane 3 will be assigned in the newly installed chip and you will lose the use of lanes 1 and 2.

GenexusTM Integrated Sequencer User Guide

93

8

Monitor the run

In the Monitor menu, you can view the status of the sequencer in an idle condition, or the status of a run in progress.

View run progress on the instrument
1. In the menu bar, click Monitor4Instrument View.
2. In the Instrument View screen, view the status of a run in progress, or the status of an instrument. The following information is provided: · Sequencer, operator, and run names · PQ status of the instrument · Instrument status at various stages of the run and post-run · Status of the results analysis · Run start and completion times
3. Click Refresh to update the information.

94

GenexusTM Integrated Sequencer User Guide

9

Review data and results

 Review sample results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96  Review run results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97  Lab Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99  View sequencing results and assay metrics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100  Reanalyze a run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126  Sign off on the run results (manager/administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127  Generate customized reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128  Results files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130  Review coverageAnalysis plugin results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132  Upload sample results files to Ion ReporterTM Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132  Verification runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133

Use the Results menu to review results and data analysis and perform data management tasks. You can view results sorted by sample or by run.

Selection Click Results4Sample Results Click Results4Run Results Click Results4Verification

Description
Select this option to review completed sample results and reports.
Select this option to review completed run results and reports by assay.
Select this option to review data from completed verification runs that were performed during sequencer installation or performance qualification.

GenexusTM Integrated Sequencer User Guide

95

9 Chapter 9 Review data and results Review sample results
Review sample results
In the Results / Sample Results screen, samples that have been sequenced are listed by sample name. You can search the list of results by sample name. Enter a search term, then click (Search).

The following information appears in the Sample Results screen.

Column Sample Name
Sample Name followed by (Signed Off)
Assay Short Name Run Name Sample Status
QC Status

Description
The unique identifier created when the sample was entered into the software. Click the Sample Name to open the Sample Details screen for the sample. Use the tabs above the Sample Details to view the run summary, assay metrics, quality control, detailed variant results, and results for plugins that are associated with the selected assay, if any.
Manager- and administrator-level users can provide their electronic signature on sample results for completed runs. A sample name followed by (Signed Off) indicates that a manager- or administrator-level user has approved the sample results. The signature information appears in the Lab Report PDF file or a user-created report, if selected. For more information, see "Sign off on the run results (manager/ administrator)" on page 127.
A shortened version of the assay name you imported or created.
The unique name of the run given when it was created in the software.
The status of the run or sample (for example: Completed, Running, Failed, Terminated, Pending, Stalled).
The QC status of a completed run.
Note:
· (Passed) indicates the sample passed all QC metrics. · (Failed) indicates the sample failed a QC metric. · (Not Calculated) indicates a sample did not undergo QC analysis.

Started On Last Updated On

The date and time when the run analysis was started. The date and time when the last action was completed on the run.

96

GenexusTM Integrated Sequencer User Guide

9 Chapter 9 Review data and results Review run results

(continued) Column
Actions

Description
Click the appropriate link. To see more actions, click (More Options). · Lab Report--Download the Lab Report (available only for samples with a sample status of completed. · Audit--View the audit trail for the run. · Notes--View or add notes to a run. · CSA--Download customer support archive (CSA) log files for the run to help with troubleshooting.

Click the Compare button in the upper right corner of the Results / Sample Results screen to compare variant results between multiple samples, or from a single sample source over time. For detailed information, see GenexusTM Software 6.2 Help, or the GenexusTM Software 6.2 User Guide (Pub. No. MAN0018955).

Review run results

In the Results / Run Results screen, runs that are pending, running, or completed are listed.
You can search the list of results by run name or PCR plate number. Enter a search term, then click (Search).
The following run information appears in the Results / Run Results screen.

Column Run Name
Assay Short Name
Total Samples Run Status
PCR Plate Number
Started On Updated On

Description
The unique name of the run given when it was created in the software. Click a run name to open the Run Summary.
The shortened unique identifier of an Assay name. You can view the complete list of Assays and Assay Short Names in the Assays4Manage Assays screen.
The total number of samples in a run.
The status of the run (for example: Not Started, Pending, Analysis Running, Executing Plugin, Completed, Terminated, Archival: In Progress).
A unique identifier for the 96-well plate used for library preparation and templating. For more information, see "Assign PCR Plate" on page 98.
The date and time when the run analysis was started.
The date and time when the last action was completed on the run.

GenexusTM Integrated Sequencer User Guide

97

9 Chapter 9 Review data and results Review run results

(continued) Column
Actions

Description
Click the appropriate link. To see more actions, click (More Options).
· Audit--View the audit trail for the run.
· CSA--Download customer support archive (CSA) log files for the run to help with troubleshooting.
· Reanalyze--Reanalyze the results of a completed run, starting from alignment or basecalling. Click Reanalyze, then select a new reanalysis assay that you have created with modified settings. For more information, see "Reanalyze a run" on page 126.
· Upload Samples to IR--Upload sample results files to Ion ReporterTM Software for secondary analysis. For more information, see "Upload sample results files to Ion ReporterTM Software" on page 132.
· Assign PCR Plate--Enter a unique identifier for the 96well plate used for library preparation and templating. For more information, see "Assign PCR Plate" on page 98.

Assign PCR Plate

GenexusTM Software allows you to track and associate a run with the PCR plate used in the run. The PCR plate is the 96well plate that is used for library preparation and templating. You can assign a unique identifier (PCR Plate Number) to completed runs and runs in progress. The PCR plate number that you enter is generated on the Lab Report and if needed, can help you track libraries and troubleshoot.
1. In the menu bar, click Results4Run Results.

2. In the Run Results screen, in the Actions column, click number in the row of the run of interest.

(More Options)4Assign PCR Plate

3. In the Assign PCR Plate dialog box, confirm, edit, or enter the PCR Plate Number.
The PCR plate number must be between 1 and 10 characters. Only alphanumeric characters (numbers 0 to 9 and letters A to Z), periods, underscores, and hyphens are allowed. Spaces are not permitted.

4. Click Submit to associate the PCR plate with the run.

98

GenexusTM Integrated Sequencer User Guide

9 Chapter 9 Review data and results Lab Report

Lab Report

The Lab Report is a PDF report of the results for each sample in a sequencing run. The assay used in the run determines the data that is included in the report.
To automatically generate a Lab Report for each sample during data analysis of a run, select the Generate Report checkbox in the Setup step when you plan the run (for more information, see Chapter 5, "Plan and manage runs"). To generate a Lab Report for each sample after a run is complete, see "Generate customized reports" on page 128.
When a Lab Report has been generated for a sample, a link is available in the Results / Sample Results screen in the Actions column for that sample. Click the link to download the PDF.
Lab reports can be electronically signed by manager- and administrator-level users. Electronically signed reports have (Signed Off) after the sample name in the Sample Results screen. The electronic signature is included in the footer of the report. For more information, see "Sign off on the run results (manager/ administrator)" on page 127.
A Lab Report typically contains the following sections and information, depending on the assay used.

Section

Description

Sample Details

The sample information that is entered into the software. You can customize the format of the Sample Details section when you create a new report template. To create a new report template, click Assays4Manage Presets, then in the Report Templates tab, click
Add New.

Sequence Variations: Variants and fusions detected in the sample, based on the targets defined by the assay.

Detected

Allele frequencies are also reported.

Test Description

A description of the report or assay that was entered in the report template.

Sequence Variations: Variants and fusions not detected in the sample, based on the targets defined by the assay. Not Detected

Comments

Laboratory comments entered in the report template.

Sequencing Run Details

Contains the following subsections: · Assay--the assay name and panel used · Analysis--the run date and name of the user who sent the run to the instrument · Run Details--the consumables used in the run · Control QC Evaluation Metrics--a summary of the CF-1 quality control metrics · Run QC Evaluation Metrics--a summary of the run quality control metrics · Sample QC Evaluation Metrics--a summary of the sample quality control metrics

GenexusTM Integrated Sequencer User Guide

99

9

Chapter 9 Review data and results View sequencing results and assay metrics

Download the Lab Report
You can download the Lab Report for a sample of interest from the Results / Sample Results screen.
Note: The Lab Report is also available for download as part of the results files in the Results screen for a specific sample. For more information, see "Results files" on page 130.
1. In the menu bar, click Results4Sample Results.
2. In the Sample Results screen, in the Actions column, click Lab Report in the row of the sample of interest.

A ZIP file that contains the PDF report downloads automatically. 3. Extract the downloaded files, then open the PDF file in an appropriate viewer.

View sequencing results and assay metrics

For every run, you can view assay-specific results and sample-specific results. Assay-specific results include assay metrics, such as final read data, and assay-level plugin information, such as execution of the customer support archive. For more information, see "Assay metrics" on page 103 and "Review coverageAnalysis plugin results" on page 132.
The following sample-specific result information is available.

Tab

Description

Summary
QC Variants Plugins

An overview of the results for the sample, including Sample Details, Key Metrics, and the Variant Summary. For more information, see "Summary of the Sample Results" on page 102.
The quality metrics for the sample sequenced in the run. For more information, see "QC results" on page 107.
Detailed variant results for SNVs/Indels, Fusions, and CNVs. For more information, see "View SNV/INDEL results" on page 112.
Results generated from the plugins associated with the assay used to analyze the sequenced sample. For more information, see "Review coverageAnalysis plugin results" on page 132.

100

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

1. To view sequencing results for a particular sample, including variant and fusion calls, click Results4Sample Results.
2. In the Sample Name column, click a sample name.
3. In the Results screen, click the QC, Variants, Plugins, and Summary tabs to view the different types of sample-specific results and data. In the Variants tab, you can toggle between variant types. You can also export data in tabular format.

4. Toggle between different assay-specific and sample-specific results for the run with the dropdown menus.

1 Run Result: The run name is listed.
2 Select Assay: Click the assay name of interest to view the assay metrics for the run.
3 Select Sample: Click a sample name to view the sequencing results for the sample. Click a different sample name to view other sample results for the run and selected assay.

Options

Description

Run Results dropdown The run name is listed. Multiple runs are listed only if the run has been

menu

reanalyzed. For more information, see "Reanalyze a run" on page 126.

Select Assay dropdown menu

Toggle between different assays used in the run. If only one assay is associated with the run, only one assay name is listed.
To remove the assay selection, select Select Assay from the dropdown menu. Removing the assay selection opens the Run Summary tab for the run selected in the Run Results dropdown menu.

Select Sample dropdown menu

Toggle between different sample results for the selected run and assay.
To remove the sample selection, select Select Sample from the dropdown menu. Removing the sample selection opens the Assay Metrics tab for the assay selected in the Select Assay dropdown menu.

All of the samples that were run with the same assay share the same assay metrics. All other results are sample-specific.

GenexusTM Integrated Sequencer User Guide

101

9

Chapter 9 Review data and results View sequencing results and assay metrics

5. To view more options, click (More Options).

Options

Description

Reanalyze

Reanalyze a run with a new assay. For more information, see "Reanalyze a run" on page 126.

Run Plugin

Run plugins on your sequencing data after a sequencing run is complete.

Run Report

The run report includes assay metrics and the record of reagents that were used in a run.
For more information, see "View or download a Run Report" on page 107.

Download Files Download all or selected variant, results, audit and logs, and lab report files. For more information, see "Results files" on page 130.

Generate Report Generate a Lab Report each sample in a sequencing run. For more information, see "Lab Report" on page 99.

Sign Off

Manager- and administrator-level users can provide their electronic signature on sample results for completed runs.
For more information, see "Sign off on the run results (manager/ administrator)" on page 127.

CSA Upload to IR

Customer support archive (CSA) log files for the run to help with troubleshooting.
Upload results to Ion ReporterTM Software for further analysis. For more information, see "Upload sample results files to Ion ReporterTM Software" on page 132.

Summary of the Sample Results

The Results screen displays a summary of the variant and fusion calls for each sample in a run, as well as details about the sample and a summary of the metrics for the run.
You can view the Results screen for a sample starting from sample or run results, but navigating from sample results requires fewer steps.
· In the menu bar, click Results4Sample Results, then click a sample name to open the Results screen.
· In the menu bar, click Results4Run Results, then click a run name to open the Results / Run Results screen. In the Run Name column, click a run name to open the Results screen. Select a sample from the Select Sample dropdown list to open the Results screen.
The information that is displayed depends on the assay that was used in the run. You can toggle between different assays used in a run with the Assays dropdown list at the top of the screen.

Section Sample Details Sample Name

Description
A unique identifier representing the sample. Click the Sample Name to open the Sample Details screen for the sample. Use the tabs above the Sample Details to view the run summary, assay metrics, quality control, detailed variant results, and results for plugins that are associated with the selected assay, if any.

102

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

(continued) Section
Collection Date Gender Sample Type Disease Category Cancer Type Cancer Stage % Cellularity
Key Metrics Average Base Coverage Depth
Uniformity Of Base Coverage
% Base Reads On Target
Variant Summary SNVs/Indels
Fusions CNVs

Description
The date that the sample was collected.
The biological sex of the sample: Female, Male, or Unknown.
A term that describes the sample, for example, FFPE, DNA, DNA & RNA
The disease type of the sample.
The type of cancer that is represented by the sample.
The stage of the cancer from which the sample was collected.
The percentage of tumor cellularity in the sample. This is a whole number between 1 and 100. The % Cellularity attribute is applicable to FFPE samples only.
The average number of reads of all targeted reference bases. This is the total number of base reads on target divided by the number of targeted bases, and therefore includes any bases that had no coverage.
The percentage of bases in all targeted regions (or whole genome) that are covered by at least 20% of the average base coverage depth reads. Cumulative coverage is linearly interpolated between the nearest integer base read depths.
The percentage of filtered reads that are mapped to any targeted region relative to all reads mapped to the reference. A read is considered on target if at least one aligned base overlaps at least one target region. If no target regions (file) was specified, this value will be the percentage of reads passing uniquely mapped and/or non-duplicate filters, or 100% if no filters were specified.
Lists and describes the SNV, MNV, and INDEL variants that are detected in the sample.
Lists and describes the fusions that are detected in the sample.
Lists and describes the copy number variants (CNVs) that are detected in the sample.

Assay metrics
Assay metrics include various chip metrics for the run, such as well and Ion SphereTM Particles (ISPs) statistics. For runs with multiple assays, metrics are provided for each assay in the run. Summary metrics are displayed at the top of the screen, followed by sample-specific metrics in the Run Samples table. Barcode-specific metrics are listed in the Barcodes With Reads Reported table. All barcodes with reads detected, even if unassigned or not planned in the run, are listed to allow you to readily identify the source of any barcode cross-contamination.

GenexusTM Integrated Sequencer User Guide

103

9

Chapter 9 Review data and results View sequencing results and assay metrics

To view the assay metrics for a run, in the Results / Run Results screen, in the Run Name column, click the run name of interest. In the Run Summary screen, select an assay from the Select Assay dropdown menu. To view the metrics for another assay in the run, select a different assay from the dropdown menu.
Assay metrics are assay-specific and cannot be viewed within the sample results screens. To view assay metrics, ensure that Select Sample is selected in the Select Sample dropdown menu.

Assay Metrics Metric
Loading Enrichment Library Final Reads
Total Bases

Description
The number and percentage of total addressable wells on the chip that contain an ISP.
The number and percentage of wells ISPs that contain live ISPs.
The number and percentage of wells with live ISPs that contain Library ISPs.
Library reads passing all filters that are recorded in the output BAM files. This value can be different from the total number of reads due to technicalities associated with read trimming beyond a minimal requirement.
The number of filtered and trimmed base pairs that are reported in the output BAM file.

104

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

Assay Metrics (continued)

Metric

Description

Mean Raw Accuracy 1x

The mean raw accuracy across each individual base position in a read calculated as, (1-[total errors in the sequenced reads]/[total bases sequenced]) × 100. Raw read accuracy is measured at each base across the length of the read and is based on 1x sequencing coverage; raw read accuracy is not based on consensus accuracy across multiple reads for the same base position.

Wells with ISPs

The number of wells that contain an ISP.

Total Addressable Wells Wells on the chip that can be physically reached by a library.

Empty Wells

The percentage of total addressable wells on the chip that do not contain an ISP.

Wells with Live ISPs

Loaded wells with ISPs with a signal of sufficient strength and composition to be associated with the library or control fragment key.

Wells with Library ISPs

Loaded wells with live ISPs with a key signal that is identical to the library key signal.

Control ISPs

Loaded wells with live ISPs with a key signal that is identical to the control fragment key signal.

Polyclonal

Wells with a live ISP that carries clones from two or more templates.
To view polyclonal metrics, mouse over the first low quality portion (gray) of the Final Reads bar plot.

Low Quality

Loaded wells with a low or unrecognizable signal.
To view polyclonal metrics, mouse over the second low quality portion (gray) of the Final Reads bar plot.

Filtered Adapter Dimer

The total percentage of filtered reads, or the sum of the percentages of polyclonal, low quality, and adapter dimer reads.
Loaded wells with a library template of an insert size less than 8 bases.

GenexusTM Integrated Sequencer User Guide

105

9

Chapter 9 Review data and results View sequencing results and assay metrics

Assay Metrics (continued)

Metric

Description

Loading Density

A visual representation of chip loading. Red color indicates areas of higher density of loading. Blue color indicates areas of lower density of loading. The following example illustrates a sequencing experiment where two lanes on the chip are uniformly loaded with ISPs.

Run Samples
The Run Samples table lists read data for each individual sample in the assay.

Column

Description

Sample Name

The unique identifier created when the sample was entered in the software.

Nucleic Acid Type

The sample nucleic acid type, such as DNA, RNA, or TNA.

Barcode

The unique identifiers of the dual barcode pair assigned to the DNA and/or RNA library for a sample.

Total Reads

The total number of filtered and trimmed reads with the listed dual barcodes assigned to the sample. The reads are independent of length reported in the output BAM file.

Mean Read Length

The average length, in base pairs, of usable library reads for each sample.

Q20 Bases

The total number of called bases that have 99% accuracy (or less than 1% error rate) aligned to the reference for the sample.

Uniformity

The percentage of bases in all targeted regions (or whole genome) with a depth of coverage 20% of the mean read coverage.

Read Length Histogram

A histogram that presents all filtered and trimmed library reads that are reported in the output BAM file (Yaxis) and the mean read length in base pairs (Xaxis). The shape of the histogram should closely resemble the library size distribution trace without the adapter sequences.

106

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

View or download a Run Report
You can view a run report or download a run report summary in PDF format. The run report includes assay metrics and the record of reagents that were used in a run.

1. In the menu bar, click Results4Run Results.

2. In the Run Results screen, in the Run Name column, click the run name of interest.

3. In the Run Summary screen, click A Run Report window opens.

(More Options)4Run Report.

4. To download a run report summary in PDF format, click Download as PDF.

QC results

The QC screen displays quality metrics for each sample that was sequenced in a run. This information is also accessible through the Monitor menu within 72 hours of starting the run on the sequencer.

To view the QC screen, in the Results / Sample Results screen, click a sample name in the Sample Name column. In the Results screen, click the QC tab. The QC status for each metric is indicated beneath each QC test (Run QC, Templating Control QC­CF­1, Sample QC­DNA, and Sample QC­ RNA).

If a sample fails a single test metric, the sample fails that QC test. A sample must meet all QC parameter thresholds of a particular QC test in order to pass. The QC status is broken down into the following categories.

·

(Passed) indicates the sample passed all QC metrics.

·

(Failed) indicates the sample failed a QC metric.

·

(Not Calculated) indicates a sample did not undergo QC analysis.

Note: If a sample fails to meet one or more QC parameters, you can reanalyze a run (see "Reanalyze a run" on page 126).

The data displayed in the screen depend on the assay that was used in the run.

GenexusTM Integrated Sequencer User Guide

107

9

Chapter 9 Review data and results View sequencing results and assay metrics

Metric Run QC Key Signal
Percent Loading
Raw Read Accuracy
Templating QC--CF-1 Control Average Reads Per Lane

Description
General run quality control information.
The average signal after software processing for library ISPs that identically match the library key (TCAG).
The number of wells with ISPs divided by the number of the total addressable wells in a run.
The average raw accuracy across each individual base position in a read, where raw read accuracy is calculated as 100 * (1 - (sum(per base error)/sum(per base depth))).
Sequencing quality metrics of the control fragment. These metrics indicate templating success.
The number of CF-1 reads divided by the number of chip lanes used in the run.

108

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

(continued) Metric
Base Call Accuracy
Mean AQ20 Read Length (bp) Sample QC--DNA MAPD
Mapped Reads Mean AQ20 Read Length (bp) Mean Read Length (bp) Uniformity of Amplicon Coverage
Median Mol Cov
Uniformity of Base Coverage
Read Length Histogram

Description
The probability that a given base is called correctly. 1 ­ (total number of errors for all positions in CF-1) / (total number of CF-1 base reads).
Average length, in base pairs, at which the accuracy rate is 99% for CF-1 reads.
Sequencing quality metrics of the sample DNA library.
MAPD (Median of the Absolute values of all Pairwise Differences) is a quality metric that estimates coverage variability between adjacent amplicons in CNV analyses. A MAPD value of 0.5 indicates an acceptable level of coverage variability. High MAPD value typically translates to a lower coverage uniformity. Lower coverage uniformity can result in missed or erroneous CNV calls. If the MAPD QC threshold is not met, CNVs do not get called. The MAPD metric does not affect SNVs/INDEL calls.
The number of reads that are mapped to the reference file.
The average length, in base pairs, at which the accuracy rate is 99% for all aligned reads of a library.
The average length, in base pairs, of final library reads for the sample.
The percentage of amplicons that had at least 20% of the average number of reads per amplicon. Cumulative coverage is linearly interpolated between nearest integer read depth counts.
The median number of functional molecule reads per amplicon calculated over all amplicons in the assay. This metric is applicable to Ion AmpliSeqTM HD library chemistry only.
The percentage of reads with a depth of coverage 20% of the mean read coverage at each position. This metric is applicable to Ion AmpliSeqTM HD library chemistry only.
The histogram presents all filtered and trimmed DNA library reads that are reported in the output BAM file (Yaxis) and the mean read length in base pairs (Xaxis). The shape of the histogram should closely resemble the library size distribution trace without the adapter sequences.

GenexusTM Integrated Sequencer User Guide

109

9

Chapter 9 Review data and results View sequencing results and assay metrics

(continued)
Metric
Read Ratio for Inline Controls (Endogenous to Spike-ins)

Description The ratio of genomic sample reads to control reads for each inline control amplicon.

Sample QC ­ RNA Mapped Reads Mean Read Length (bp) RNA Expression Ctrls Detected
Mean AQ20 Read Length
Read Length Histogram

Note: The Read Ratio for each inline control amplicon is expected to be 3 with 10 ng DNA input for both Ion AmpliSeqTM and Ion AmpliSeqTM HD chemistries.
Sequencing quality metrics of the sample RNA library.
The total number of final library reads that are aligned to the reference file for the sample.
The average length, in base pairs, of the final library reads for the sample.
The number of expression control genes detected for the sample. This metric measures the RNA input integrity, input amount, and the fidelity of the reverse transcriptase that was used in library preparation.
Fusion panels include primer pairs that cover 5­7 control housekeeping genes.
The average length, in base pairs, of the final library reads with no more than 1% error rate for the sample.
The average length, in base pairs, at which the accuracy rate is 99% for all aligned reads of a library.
The histogram presents all filtered and trimmed library reads that are reported in the output BAM file (Yaxis) and the mean read length in base pairs (Xaxis). The shape of the histogram should closely resemble the library size distribution trace without the adapter sequences.

110

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

(continued)
Metric
Read Ratio for Inline Controls (Endogenous to Spike-ins)

Description
The ratio of genomic sample reads to control reads for each inline control amplicon. You can use this metric to measure the level of contaminating genomic DNA in RNA libraries when you select the RNA checkbox for Include Inline Controls in assay set up. Using inline controls in RNA assays will reduce the total number of panel reads.

NTC QC ­ DNA Average Base Coverage Depth Mean Read Length (bp)
Read Length Histogram
Read Ratio for Inline Controls (Endogenous to Spike-ins) NTC QC ­ RNA Mapped Reads
Mean Read Length (bp)
RNA Expression Ctrls Detected Read Ratio for Inline Controls (Endogenous to Spike-ins)

Note: The Read Ratio for each inline control amplicon is expected to be 3 with 10 ng DNA input for both Ion AmpliSeqTM and Ion AmpliSeqTM HD chemistries.
DNA Sequencing quality metrics of the no template control. The average number of DNA reads of all targeted reference bases.
The average length, in base pairs, of final DNA library reads for the no template control. The histogram presents all filtered and trimmed reads for the no template control that are reported in the output BAM file (Yaxis) and the mean read length in base pairs (Xaxis). The ratio of no template control sample reads to control reads for each DNA inline control amplicon.
RNA Sequencing quality metrics of the no template control. The total number of final library reads that are aligned to the reference file for the NTC. The average length, in base pairs, of final RNA library reads for the no template control. The number of expression control genes detected for the no template control.
The ratio of no template control sample reads to control reads for each RNA inline control amplicon.

GenexusTM Integrated Sequencer User Guide

111

9

Chapter 9 Review data and results View sequencing results and assay metrics

View SNV/INDEL results
The SNVs/Indels table lists the calls and other information for the SNV and INDEL variants that are analyzed in each sample in a run.
To view the SNVs/Indels table for a sample, click Results4Sample Results in the menu bar, then in the Sample name column, click the name of the sample of interest. In the Variants tab, click SNVs/Indels to display the data. To export the data in tabular format, click Export in the upper right corner of the screen.

SNVs/Indels table
The data displayed in the table depend on the assay that was used in the run. Results in the table can be filtered using the filtering tools. For more information, see "Filter results" on page 122.

Column

Description

User Classifications User-defined classification to selected from the list. For more information, see "Create and assign variant classifications" on page 121.

Locus

The chromosome and position of the detected variant.

Oncomine Variant Class

The type of SNV or INDEL at the locus based on OncomineTM annotations.
This information is available if the Apply Oncomine Variant Annotations checkbox is selected in the assay used in the run.

Oncomine Gene Class

The change in molecular function of the altered gene product due to the mutation, based on OncomineTM annotations:
· Gain-of-function -- the altered gene product has a new molecular function or pattern of gene expression compared to the wild-type gene
· Loss-of-function -- the altered gene product lacks the molecular function of the wildtype gene
This information is available if the Apply Oncomine Variant Annotations checkbox is selected in the assay used in the run.

Gene

The gene name. Click the link to open the View Annotation Sources window to view additional information. For more information, see "View annotation sources" on page 121.

AA Change

Identification of the amino acid change using Human Genome Variation Society (HGVS) nomenclature.

Ref

The reference base or bases at that locus.

Alt

The alternate base or bases at that locus.

Type

The type of variant that is detected. · snp (single nucleotide polymorphism) · mnp (multi-nucleotide polymorphism) · ins (insertion) · del (deletion) · complex

112

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

(continued) Column
Call

Description
Indicates the presence or absence of an SNV/Indel variant. When the default filter chain is applied, only the variant calls that are designated with PRESENT (HOMOZYGOUS) or PRESENT (HETEROZYGOUS) are displayed in the results table. To view all calls, including calls that do not pass the required filter thresholds, apply the No Filter option or download the Variants (.vcf) file (see "Results files" on page 130).
· PRESENT (HOMOZYGOUS) or PRESENT (HETEROZYGOUS) ­ indicates a high confidence call that passes all filter thresholds at a given variant position.
Note:
· When the default filter chain is applied, PRESENT (HOMOZYGOUS) or PRESENT
(HETEROZYGOUS) indicates the presence of the ALT (alternative) allele.
· When the No Filter option is applied or when viewing the Variants (.vcf) file, Present
does not imply the presence of the ALT (alternative) allele. To infer the presence of the ALT allele, refer to the Alt column.
· NO CALL ­ while some evidence for the presence of a variant exists, the call does not pass one or more filters that are required for a high confidence variant call.
· ABSENT ­ indicates the presence of a variant that differs from the reference allele at a given position, however, this nucleotide is not an ALT allele that is targeted by this assay.

No Call Reason

The reason why a variant is reported as No Call.

Phred QUAL Score

The relative probability of either the "reference" hypothesis interval [0,cutoff], or the "variant" hypothesis interval [cutoff,1], Phred-scaled (10*log10). A higher score means more evidence for the variant call.

Raw Read Depth

Total read coverage across amplicon containing SNV/INDEL hotspot locations. Count of chip-level reads aligned at this locus that participate in variant calling.

Effective Read Depth The number of reads covering the position.

Alt Allele Read Counts

The number of reads containing the alternate allele.

Variant ID

The name of the hotspot as defined in the Browser Extensible Data (BED) file.

Nuc Change

The position and identity of the nucleic acid change.

Allele Fraction

The number of variant read counts divided by the total number of read counts for the sample.

Mol Depth[1]

Reports number of interrogated DNA molecules containing target. It defines limit of detection at hotspot position in a particular run and sample. For instance, if molecular depth is 1,500, you can have high confidence that no variant is present at 0.2% LOD. If molecular depth is 2,500, you can have high confidence that no variant is present down to 0.1% LOD. For reference calls, Molecular Depth provides measurable metric that serves as confirmation for variant absence among a large number of interrogated molecules.

WT Mol Counts[1]

Number of detected molecules containing the wildtype allele.

GenexusTM Integrated Sequencer User Guide

113

9

Chapter 9 Review data and results View sequencing results and assay metrics

(continued)

Column

Description

Alt Allele Mol Counts[1]

Number of detected molecules containing the alternate allele.

Mol Freq %[1]

Molecular frequency percentage. The percentage of mutant reads over total reads at the locus.

% LOD[1]

Limit of detection (LOD) of a variant allele expressed as a percentage of the WT allele. LOD is the lowest possible variant frequency in the sample that can be detected by the system with a true positive rate greater than 98% for FFPE samples or 95% for cfTNA samples. LOD is dependent on the molecular read depth at the locus. %LOD is reported when there are no variant calls for the gene.

[1] Column appears in analyses of Ion AmpliSeqTM HD sequencing data only.

View Fusion results

The Fusions table lists the calls and other information for the fusions analyzed in each sample in a run.
To view the Fusions table for a sample, click Results4Sample Results in the menu bar, then in the Sample name column, click the name of the sample of interest. In the Variants tab, click Fusions to display the data. To export the data in tabular format, click Export in the upper right corner of the screen.

Fusions table
The data displayed in the table depend on the assay that was used in the run. Filter the results list in the table using the filtering tools. For more information, see "Filter results" on page 122.

Column

Description

User Classifications A user-defined classification selected from the list. For more information, see "Create and assign variant classifications" on page 121.

Locus

The chromosome positions in the reference genome that define the fusion junction.

Oncomine Variant Class

Oncomine variant class annotation that indicates fusion type based on OncomineTM annotations. This information is available if the Apply Oncomine Variant Annotations checkbox is selected in the assay that is used in the run.

Oncomine Gene Class

The change in molecular function of the altered gene product due to the mutation, based on OncomineTM annotations:
· Gain-of-function -- the altered gene product has a new molecular function or pattern of gene expression compared to the wild-type gene
· Loss-of-function -- the altered gene product lacks the molecular function of the wildtype gene
This information is available if the Apply Oncomine Variant Annotations checkbox is selected in the assay used in the run.

Genes (Exon)

The name of fusion target and representative acceptor and donor exons.

114

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

(continued) Column
Read Counts Type
Call

Description
The frequency that the fusion was detected in the sample.
Assay type (for example, Fusion, RNA exon variant (exon skipping), RNAExon Tile, Proc Control).
Indicates the presence or absence of a fusion or RNA exon variant. When the default filter chain is applied, only the fusion/RNA exon variant calls that are designated with PRESENT are displayed in the results table. To view all calls, including calls that do not pass the required filter thresholds, apply the No Filter option or download the Variants (.vcf) file (see "Results files" on page 130).
· PRESENT ­ indicates a high confidence call that passes all filter thresholds at a given variant position.
· ABSENT ­ indicates the absence of a fusion due to the variant call failing internal quality control.
· NO CALL ­ while some evidence for the presence of a fusion exists, the call does not pass one or more filters that are required for a high confidence fusion call.

No Call Reason
Failed Reason
Variant ID
Read Counts Per Million
Oncomine Driver Gene Mol Cov. Mutant[1] Imbalance Score[2]

The reason for reporting a fusion as NOCALL.
The reason for reporting a fusion as Absent.
The name of the fusion target as defined in the BED file.
The number of fusion read counts detected per number of total reads (in millions).
The gene believed to be associated with increased oncogenic properties. The gene is inappropriately activated by the fusion.
The median molecular coverage across a fusion amplicon.
Each fusion gene exhibits a characteristic Imbalance Score threshold. Scores that exceed this threshold value indicate a high likelihood of the presence of the fusion in the test sample.
· Observed = (sum of read count coverage of amplicons downstream (3') of a predicted breakpoint in a target gene) / (sum of read count coverage of all amplicons of the gene) [Read counts from test sample]
· Expected = (sum of baseline value for amplicons downstream of the breakpoint/sum of baseline values) / (sum of baseline values of all amplicons of the gene) [Baseline values computed from normal samples]

Imbalance PValue[2] The statistical significance of measure of imbalance relative to a control gene.

Predicted Breakpoint Range[2]

The exonic range for predicted fusion break point in exon tiling assays.

GenexusTM Integrated Sequencer User Guide

115

9

Chapter 9 Review data and results View sequencing results and assay metrics

(continued)

Column

Description

Ratio To Wild Type[1] The molecular ratio for exon skipping assays relative to wild type control amplicons.

Norm Count Within (Lung panel only) Exon skipping assay coverage normalized to molecular coverage of wild

Gene[1]

type (WT) MET control amplicons.

[1] Column appears in analyses of Ion AmpliSeqTM HD sequencing data only. [2] Column appears in analyses that use the exon tiling fusion detection method.

View RNA Exon Variants
The RNA Exon Variant data view displays a bar graph summary of intragenic exon rearrangements or fusions. The displayed RNA exon variants are defined in the BED file that is associated with an assay. The RNA Exon Variant data view is available for all RNA and Fusion assays.
Note: To review RNA Exon Tile Fusion Imbalance analysis plots, see the user guide for your assay.

1. In the menu bar, click Results4Sample Results.

2. In the Sample name column, click the name of the sample of interest.

3. Click the Variants tab, then click Fusions. The Fusions table opens to display fusions results.

4. In the top right corner of the screen, click Visualization4RNA Exon Variant, then review the RNA Exon Variants plot.

116

GenexusTM Integrated Sequencer User Guide

Representative RNA Exon Variant plots
1

Chapter 9 Review data and results View sequencing results and assay metrics

9

2

The Xaxis represents specific exon variants, where each variant is labeled with a gene ID followed by a sequence of adjacent exons. The Yaxis measures the read counts for each variant, normalized to the wild type.
1 Example analysis where only the wild type EFGR (EFGR.E6E7) was detected. 2 Example analysis where RNA exon 2­7 deletion occurred in the EFGR gene. The deletion of exons 2­7 resulted in an increase
of normalized read counts for the EFGR variant that contains the intragenic fusion of exon 1 and exon 8 (EFGR.E1E8.Del1) and a decrease of normalized read counts for the wild type EFGR (EFGR.E6E7).
To return to the table view of fusions, click (Remove) next to the Visualization dropdown menu.
View RNA Exon Tile Fusion Imbalance
The RNA Exon Tile Fusion Imbalance data view provides a visual representation of the RNA fusion imbalance analyses.
1. In the menu bar, click Results4Sample Results.
2. In the Sample name column, click the name of the sample of interest.
3. Click the Variants tab, then click Fusions. The Fusions table opens to display fusions results.
4. In the top right corner of the screen, click Visualization4RNA Exon Tile Fusion Imbalance, then review the RNA Exon Tile Fusion Imbalance plots.

GenexusTM Integrated Sequencer User Guide

117

9

Chapter 9 Review data and results View sequencing results and assay metrics

Representative RNA Exon Tile Fusion Imbalance plots

1

2

3 4 5 6
7 8
118

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

1 The Fusion Imbalance Call: Imbalance Score & P-value plot shows the imbalance scores and pvalues for all the genes in the selected sample. The dashed gray lines mark the threshold for an imbalance call, which is applied to all genes across all samples. Points that fall within the blue shaded area of the plot represent fusionpositive genes ( ). All other points that are outside of the blue shaded area represent fusionnegative genes ( ). Control genes are marked with .
2 The Normalized Read Depth by Gene plot shows the mean read counts of each gene that is captured on the chip for the selected sample. For each gene, the read counts are normalized to the number of amplicons.
3 The Sample Coverage for Imbalance Analysis plots show the expression profile for each exonexon tiling amplicon for each gene. The Yaxis represents the corrected molecular counts. The Xaxis represents individual exonexon junctions, which are listed from 5 to 3. The imbalance score and pvalue are listed in the panel of each gene that was called positive for fusion.
4 Baseline (a cluster of gray lines), generated from a fusionnegative sample. 5 Test sample corrected read coverage (blue line), normalized to the baseline. Each point on the line represents a unique
exonexon junction that was covered by the assay and normalized to the baseline. 6 Predicted range for the fusion break point for a fusionpositive gene (dashed red line). 7 Sample coverage profile for the control gene (orange line). 8 If the collected data are insufficient to determine an imbalance score, the INVALID SAMPLE message appears in the panel for
that gene.
To return to the table view of fusions, click (Remove) next to the Visualization dropdown menu.
View CNV results
The CNVs table lists the calls and other information for the copy number variants (CNVs) analyzed for each sample in a run.
To view the CNVs table for a sample, click Results4Sample Results in the menu bar, click the name of the sample of interest, then click Variants4CNVs. Click Export in the upper right corner of the screen to export the data in tabular format.
CNVs table
The data displayed in the table depend on the assay that was used in the run. Results in the table can be filtered using the filtering tools. For more information, see "Filter results" on page 122.
IMPORTANT! (FFPE samples only) If % Cellularity value for a sample is set to <100, then the magnitude of copy number gain or loss can be decreased. For more information, see "System-installed sample attributes" on page 47.

Column

Description

User Classifications A user-defined classification selected from the list. For more information, see "Create and assign variant classifications" on page 121.

Locus

The starting position of the first amplicon covering the CNV gene.

Oncomine Variant Class

Annotation that indicates whether CNV is an amplification or deletion.
This information is available if the Apply Oncomine Variant Annotations checkbox is selected in the assay used in the run.

GenexusTM Integrated Sequencer User Guide

119

9

Chapter 9 Review data and results View sequencing results and assay metrics

(continued) Column
Oncomine Gene Class
Gene
Copy Number Call

Description
The change in molecular function of the altered gene product due to the mutation, based on OncomineTM annotations:
· Gain-of-function -- the altered gene product has a new molecular function or pattern of gene expression, compared to the wild-type gene
· Loss-of-function -- the altered gene product lacks the molecular function of the wildtype gene
This information is available if the Apply Oncomine Variant Annotations checkbox is selected in the assay used in the run.
The gene name, which provides a link to the View Annotation Sources dialog box with additional information about the HotSpot ID.
For more information, see "View annotation sources" on page 121.
The copy number of a CNV gene locus per genome. This column is available when a positive call is made.
Indicates the presence or absence of a CNV. When the default filter chain is applied, only the CNV-positive calls that are designated with PRESENT are displayed in the results table. To view all calls, including calls that do not pass the required filter thresholds, apply the No Filter option or download the Variants (.vcf) file (see "Results files" on page 130).
· PRESENT ­ indicates a high confidence call that passes all filter thresholds. · PRESENT (GAIN) ­ a CNV-positive call that indicates gene amplification; a high
confidence variant call that passes all filter thresholds. · PRESENT (LOSS) ­ a CNV-positive call that indicates gene deletion; a high confidence
variant call that passes all filter thresholds. · ABSENT ­ the absence of a variant; result is below detection threshold for a CNV call. · NO CALL­ while some evidence for the presence of a variant exists, the call does not
pass one or more filters that are required for a high confidence variant call.

P-Value

The statistical significance of the CNV ratio measurement.

No Call Reason

The reason for reporting a CNV as No Call.

CNV Confidence

The CNV confidence interval associated with the call. The 5% lower confidence bound value is the ploidy estimate, where there is a 5% chance that the true ploidy is below that value. The 95% upper confidence bound is the ploidy estimate, where it is 95% certain that the true ploidy is below that value.

Variant ID

The name of the hotspot as defined in the BED file.

CNV Ratio

The ratio of measured CNV gene locus coverage relative to coverage of non-CNV loci.

Med Read Cov Gene The median read coverage of targeted CNV gene.

Med Read Cov Ref The median read coverage of non-CNV reference loci.

Valid CNV Amplicons The number of amplicons spanning the CNV call.

120

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

View annotation sources
You can view more information for each HotSpot ID in the sample Results screen in the SNVs/Indels, Fusions, and CNVs tables.
In the menu bar, click Results4Sample Results, then click the sample name in the Sample Name column in the row of a sample of interest. In the Variants tab, click SNVs/Indels, Fusions, or CNVs.
1. In the SNVs/Indels, Fusions, or CNVs table, in the Gene column, click the gene symbol.

[Sample name]

[Run name] [Assay name] [Sample name]

The View Annotation Sources window opens, which displays information for the particular variant.
2. Review annotation information in the View Annotation Sources window. a. Click View IGV to download a JNPL file that can be opened in the Integrative Genomics Viewer (IGV) from the Broad Institute (see software.broadinstitute.org/software/igv).
b. If available, click a link in the row of an annotation source to navigate to the online annotation information.
Create and assign variant classifications
You can create and assign user-defined variant classifications in the SNVs/Indels, Fusions, and CNVs tables in the Variants tab of the Results screen.
1. In the menu bar, click Results4Sample Results.
2. Click a sample name in the Sample Name column in the row of a sample of interest to open the Results screen for the sample.
3. Click the Variants tab.
4. In the Variants table, click SNVs/Indels, Fusions, or CNVs to display the list of variants for the selected variant type.

GenexusTM Integrated Sequencer User Guide

121

9

Chapter 9 Review data and results View sequencing results and assay metrics

5. In the Variants table, in the User Classification column, perform any of the following actions. · To assign an existing classification to a variant, select it from the list.

· To create a new classification, enter a name for the classification in the text box, select a color for the new classification, then click Add.

· To remove a classification from a variant, click (Remove). · To delete a classification from the list, click Delete next to the classification name. The
classification will be removed from all variants in all results.
Filter results
You can filter results in the SNVs/Indels, Fusions, and CNV tables in the Variants tab of the Results screen in two ways. You can apply filters to columns of information that appear in the screen. The filters, available at the top of each column, immediately narrow the list of information in any columns to which filters are applied.
You can also apply a filter chain, a set of filters that GenexusTM Software uses to narrow the list of variants that are included in results. A manager-or administrator-level user creates filter chains from system-installed filters.

122

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

Filter the view of results in the Results screen
You can filter results to immediately narrow the list of results that is shown in columns in Results screen. Filters are available for each column in the list of results.
1. In the menu bar, click Results4Sample Results.
2. Click a sample name in the Sample Name column.
3. In the Results screen, click the Variants tab.
4. Select the variant class to display the results: SNVs/Indels, Fusions, or CNVs.
5. In the results table, click (Filter) in the column heading of interest, enter filter options, then click Filter. The options that are available depend on the column and variant class. For example, you can filter data in the Type column to show one specific variant type.
The column or columns to which you applied a filter change to reflect the filter and selected options.

Filter results using a filter chain
You can filter the results that are listed in the Results screen with a system-installed or custom filter chain. A filter chain is a set of filters that GenexusTM Software uses to narrow the list of variants that are included in results.
Select a filter chain to change the list of variants that are included in the results. You can apply the filter chain temporarily, then review the results before you decide whether to save the updated results, or discard the changes.
If you save the filter chain to a result, the variants that are included reflect the filtered results when the results are later opened.

GenexusTM Integrated Sequencer User Guide

123

9

Chapter 9 Review data and results View sequencing results and assay metrics

For information about system-installed filter chains, and how manager- and administrator-level users can create custom filter chains, see GenexusTM Software 6.2 Help, or the GenexusTM Software 6.2 User Guide (Pub. No. MAN0018955).
1. In the menu bar, click Results4Sample Results.
2. In the Sample Results screen, in the Sample Name column, select a sample of interest.
3. In the Results screen, click the Variants tab.
4. Select the variant type to display the results: SNVs/Indels, Fusions, or CNVs.
5. Above the variant table in the Results screen, in the Filter dropdown menu, select a filter chain. The list of results changes to reflect the selected filter chain.
6. (Optional) Click Save next to the filter chain name if you want the filter chain to be applied to the results when the results are later opened. The filter chain is selected in, and applied to the results when the results are reopened.
View OncomineTM TCR BetaLR Assay GX run results

· To view the analysis of OncomineTM TCR BetaLR Assay GX expression data, in the Results / Sample Results screen, click the sample name in the Sample Name column, then click the Results tab. Results are viewed in two tabs: Sample Results and Sample QC.

· In the Sample Results tab, select from the following options in the Views dropdown menu.

­ Spectratyping Plots

­ Clones detected per variable gene

­ V-gene usage

­ CDR3 histogram

­ VJ-gene usage

­ Clone Summary

124

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results View sequencing results and assay metrics

9

· In the Sample QC tab, select from the following options in the Views dropdown menu. ­ Read classification ­ Proportion of full length, quality-trimmed, and reads lacking P1key, by read classification ­ Base composition of overcalled and undercalled homopolymers ­ Downsampling analysis ­ QC metrics
For detailed information on viewing and analyzing OncomineTM TCR BetaLR Assay GX results, see the OncomineTM TCR BetaLR Assay GX User Guide (Pub. No. MAN0018513).

GenexusTM Integrated Sequencer User Guide

125

9 Chapter 9 Review data and results Reanalyze a run
Reanalyze a run
If a sequencing run fails to meet one or more QC parameters defined by the assay, you can relax the assay parameters and reanalyze a run. Runs can be reanalyzed starting from Alignment or Basecalling. When you reanalyze a run, the reanalysis is applied to all samples in the assay. For more information, see "View sequencing results and assay metrics" on page 100.
Note:
· Manager- and administrator-level users can reanalyze a sequencing run only if the run completed
without any critical alarms or errors. If the run aborted or produced major alarms or errors, the run cannot be reanalyzed.
· QC parameters at the limits of stringency cannot be relaxed further. The sample library must be
resequenced or a new library prepared.
· An assay can be used only once.
Before you begin, create a new assay or copy the original assay that was used in a run and modify assay parameters as needed. For more information, see Chapter 3, "Create and manage assays (manager/administrator)".
1. In the menu bar, click Results4Run Results.
2. In the Results / Run Results screen, in the Run Name column, click the run name of interest.
3. In the upper right corner of the screen, click (More Options)4Reanalyze.
4. In the Reanalyze dialog box, enter or select the following information. a. In Reanalysis Run Name field, enter a reanalysis run name.
b. In Start Reanalysis from dropdown list, select Alignment or Basecalling.

126

GenexusTM Integrated Sequencer User Guide

Chapter 9 Review data and results Sign off on the run results (manager/ administrator)

9

c. In the Current Assay in Run column, select the checkbox in the row of each assay that you want to reanalyze, then in the Choose Assay to Reanalyze column, select an assay that you want to use for each reanalysis from the dropdown list.

5. Click Reanalyze.
Follow the progress of the reanalysis in the Results / Run Results screen in the Run Status column and in the Results / Sample Results screen in the Sample Status screen. When reanalysis is complete, the new results can be viewed by clicking the run name corresponding to the reanalysis assay in the Results / Run Results screen. Runs that have been reanalyzed are appended with (Reanalysis) after the run name.

Sign off on the run results (manager/ administrator)
Manager- and administrator-level users can provide their electronic signature on sample results for completed runs. In the Results / Sample Results screen, a sample name followed by (Signed Off) indicates that a manager- or administrator-level user has approved the sample results. The signature information appears in the Lab Report PDF file or a user-created report, if selected. For more information, see "Lab Report" on page 99. For information on how to create a report template, see GenexusTM Software 6.2 Help, or the GenexusTM Software 6.2 User Guide (Pub. No. MAN0018955).
A manager- or administrator-level user can update report template selections during sign off.

GenexusTM Integrated Sequencer User Guide

127

9

Chapter 9 Review data and results Generate customized reports

This feature allows you to meet Title 21 CFR Part 11 of Federal Regulations that establishes the United States Food and Drug Administration regulations on electronic records and signatures, password policies, and user activity auditing.
1. In the menu bar, click Results4Sample Results.
2. In the Sample Results screen, click the sample of interest in the Sample Name column.
3. In the upper right-hand corner of the screen, click (More Options)4Sign Off.
4. In the Electronic Signature dialog box, enter your user name, password, and comments. Items identified with a red asterisk (*) are required fields.
5. In the Meaning of Signature dropdown list, select Approval.
6. In the Report Template dropdown list, select the report template that you want to use.
7. In the Report Customizations section, in the Lab Report pane, select the types of variant calls that you want to include in each report. For assays that use Reporting Gene Lists, you can customize the variants by reporting category.
8. In Footer Field, enter any text that you want to appear in the footer of the PDF report pages. If you entered footer information in the Footer Field when you created a report template, the same footer information appears in the Electronic Signature dialog box. You can enter new footer information to override the report template.
9. Click Sign Off to confirm your electronic signature.
10. In the menu bar, click Results4Sample Results to return to the Results / Sample Results screen.
11. In the row of the sample of interest, in the Actions column, click Lab Report to download the report.
Generate customized reports
When generating a customized report, you can update any report template selections. For more information on results files, see "Results files" on page 130.
1. In the menu bar, click Results4Sample Results.
2. In the Results / Sample Results screen, click the sample of interest in the Sample Name column.

128

GenexusTM Integrated Sequencer User Guide

3. In the Results screen, click Generate Report.

(More Options)4

4. In the Generate Report dialog box, select the report template and variants to include in the report.

Chapter 9 Review data and results Generate customized reports

9

5. Click Generate. A ZIP file that contains all the selected reports and other files is downloaded.

GenexusTM Integrated Sequencer User Guide

129

9 Chapter 9 Review data and results Results files

Results files

The following files can be downloaded from the View Results screen for each sample. The files that are available for download vary depending on the assay used. For a list and descriptions of plugin output files, see "Output files generated by the coverageAnalysis plugin" on page 190.
To download the files, click Results4Sample Results, then click a sample name in the Sample Name column. In the Results screen, click (More Options)4Download Files. In the Download Files dialog box, select the files to download, then click Download. The selected results files are downloaded in one ZIP folder.

Selection

File name

Description

Variants

Filtered Variants (.vcf)

<Filter name> _filtered.vcf

Summary of filtered variant results in variant call format (VCF).

All Variants (.vcf)

Oncomine_<LibPrepID>_ Summary of variant results in variant call format

<analysisID>.vcf

(VCF).

Variant Summary (.tsv)

Summary.tsv

File that lists SNV/INDEL, copy number, and fusion results in tab-separated value format (TSV).

Snvindel (.tsv)

Snvindel.tsv

File that lists SNV/INDEL variant results in tabseparated value format (TSV).

Fusion (.tsv)

Fusion.tsv

File that lists fusion results in tab-separated value format (TSV).

CNV (.tsv)

Cnv.tsv

File that lists copy number variant results in tabseparated value format (TSV).

Sequencing results

DNA Unmapped Bam File (.bam)

<barcode>

Unmapped DNA barcode BAM file; output after

_rawlib.basecaller.bam mapping reads to reference.

DNA mapped bam file (.bam) merged.bam

Mapped BAM file of combined barcode reads.

DNA Mapped Bam Index File merged.bam.bai (.bai)

Mapped BAM Index file.

DNA Basecaller FASTAQ File <barcode>_rawlib.

(.fastq)

basecaller.fastq

FASTQ file of the DNA barcodes used.

DNA Processed Bam File

merged.bam.ptrim.bam Mapped BAM file of combined barcode reads.

DNA Processed Bam Index merged.bam.ptrim.bam .bai

Mapped BAM index file.

RNA Unmapped Bam File (.bam)

<barcode>_rawlib. basecaller.bam

Unmapped RNA BAM file; output of base calling, contains unmapped reads.

130

GenexusTM Integrated Sequencer User Guide

9 Chapter 9 Review data and results Results files

(continued)

Selection

File name

Description

RNA Mapped Bam File (.bam)

<barcode>_rawlib. basecaller_alignments .bam

Mapped BAM file of combined barcode reads.

RNA Mapped Bam Index File <barcode>_rawlib.

(.bai)

basecaller_alignments

.bam.bai

Mapped BAM index file.

RNA Basecaller FASTAQ File <barcode>_rawlib.

(.fastq)

basecaller.fastq

FASTQ file generated from the unmapped BAM file of the RNA barcodes used.

Test Fragment Basecaller FASTAQ File (.fastq)

rawtf.basecaller.fastq FASTQ file for the test fragment.

Audit and Log

Analysis Log File

analysis.log

Analysis log file.

Run Summary

Info.csv

Contains information about the run and analysis, including software version details, sample details, library details, run details, assay details, reagent and consumable information, run and sample QC metrics, and instrument summary.

Run Audit

PlannedRun-AuditTrail Contains all audit records pertaining to the run. .pdf

Reports

Lab Report

<language>_<sample name>_AD_Lab_Report_ <assay name>_<date> .pdf

A PDF report that contains samplespecific results. For more information, see "Lab Report" on page 99.

GenexusTM Integrated Sequencer User Guide

131

9

Chapter 9 Review data and results Review coverageAnalysis plugin results

Review coverageAnalysis plugin results
The coverageAnalysis plugin generates a Coverage Analysis Report. This report includes read statistics and several charts. The statistics and charts that are presented depend on the library type for the analysis. The report summary lists the barcode, the sample, the number of mapped reads, the percentage of on target reads, mean base coverage depth, and base coverage uniformity. MicrosoftTM ExcelTM-compatible reports are also generated, including differential expression tables. Additional details regarding read coverage are also provided on a per-barcode basis, along with a list of gene annotations for each sequenced region. You can download coverageAnalysis plugin output files from the Results screen for a sample. For more information, see "Results files" on page 130 and "Output files generated by the coverageAnalysis plugin" on page 190. For more information on how to review and interpret read statistics and output files, see Appendix D, "coverageAnalysis plugin in GenexusTM Software".
1. In the menu bar, click Results4Sample Results.
2. In the Sample Results screen, in the Sample Name column, click the sample of interest.
3. Click the Plugins tab. A summary table of the coverage analysis, by barcode, is included in the coverageAnalysis summary pane.
4. (Optional) From the Executed At dropdown list, select an alternate timestamp, if available, to view additional reports.
5. (Optional) Click View Log to view the coverageAnalysis log.
6. (Optional) Click Delete to delete the coverageAnalysis plugin output for the selected timestamp.
IMPORTANT! If you click Delete, the report is deleted without the appearance of confirmation dialog window. Ensure that you intend to delete the report before clicking Delete.
7. In the coverageAnalysis summary pane, in the Barcode Name column, click the link in the row of the barcode of interest. The detailed Coverage Analysis Report for the barcode opens in a separate window.

Upload sample results files to Ion ReporterTM Software
When a run completes successfully in GenexusTM Software, and the results were not automatically uploaded to Ion ReporterTM Software, you can upload the sample results (BAM files) to Ion ReporterTM Software for further analysis.
Note: To automatically upload BAM files at the completion of a run, select the Upload BAM files to Ion ReporterTM Software checkbox in the Setup step of planning a run.

132

GenexusTM Integrated Sequencer User Guide

9 Chapter 9 Review data and results Verification runs

Before you can upload sample results files to Ion ReporterTM Software, you must configure your Ion ReporterTM Software account.
· To configure an Ion ReporterTM Server account, see "Configure an Ion ReporterTM Server account (administrator)" on page 179.
· To configure an Ion ReporterTM Software on Connect account, see "Configure Thermo Fisher Accounts in GenexusTM Software (administrator)" on page 182.
1. In the menu bar, click Results4Run Results.
2. In the Run Results screen, in the Actions column, click Upload to IR in the row of a run of interest.
3. In the Upload Samples to Ion Reporter Software dialog box, make the following selections. a. From the Ion ReporterTM Software Account list, select your Ion ReporterTM Server account.
b. In the Assay column, select the appropriate assay from the list.
c. In the Ion Reporter Workflow column, do one of the following: · Select Upload Only to upload the sample results (BAM files) Ion ReporterTM Software. You will need to sign in to Ion ReporterTM Software to launch an analysis. · Select an appropriate Ion ReporterTM Software analysis workflow to upload the sample results (BAM files) and automatically launch the selected analysis workflow in Ion ReporterTM Software. This results in an analysis in Ion ReporterTM Software.
d. Click Upload.
The sample results are uploaded to Ion ReporterTM Software.

Verification runs

The following information is available in the Results / Verification Results screen.

Column Run Name
Assay Name Field Engineer Name Instrument Name Run Status QC Status
Last Updated on

Description
The name of the run, created when the run was planned. Click the name to open the View Sample Run dialog box.
The name of the assay used in the run.
The name of the support specialist who performed the run.
The name of the sequencer that was verified.
The current status of the full sequencing run, including analysis.
Indicates whether a sequencing run passed or failed, based on the sequencing QC metrics selected for the assay.
The date and time that the sequencing run was last updated.

GenexusTM Integrated Sequencer User Guide

133

9 Chapter 9 Review data and results Verification runs

(continued) Column
Actions

Description
· Audit--Click this link to display the list of users who created/edited the run. You can export and print information from the list from the Audit Trail dialog box.
· CSA--Customer Support Archive. Click this link to download all of the sequencer log files. Log files contained within the CSA may be useful when troubleshooting issues with the sequencer. For more information, see the GenexusTM Software 6.2 User Guide (Pub. No. MAN0018955).

View and sign off on the verification run results
Manager- and administrator-level users can view verification run results, but we recommend that a qualified Thermo Fisher Scientific support specialist signs off on verification results.
1. In the menu bar, click Results4Verification Results.
2. In the Verification Results screen, click a run name in the Run Name column to view verification results for a verification run.
3. Click Sign Off in the upper right corner of the screen to open the Electronic Signature dialog box, then enter your user name, password, and comments. Use the Meaning of Signature dropdown list to indicate approval.
4. In Footer Field, enter any text that you want to appear in the footer of the PDF report pages.
5. Click Sign Off to confirm your electronic signature.

System Install Qualification results
For a System Install Qualification run, basic information about the verification run is provided, and the pass/fail QC status for the following metrics are reported for the controls that are used in the run:

QC metric Run QC AQ20 Mean Read Length
Average Uniformity of base coverage Raw Read Accuracy Run Balance Factor
Total Reads

Description

Pass threshold

Average read length, in base pairs, at which the error rate is 1% for all aligned reads.
The percentage of bases having a depth of coverage 20% of the mean coverage at each position.
The percentage of raw reads mapped to the reference genome.
The minimum Sample Balance Factor in a set of four tested samples A­D.
The total number of filtered and trimmed reads with the listed barcodes assigned to the samples.

90 90 98.5 0.5 10,000,000

134

GenexusTM Integrated Sequencer User Guide

9 Chapter 9 Review data and results Verification runs

(continued)

QC metric

Description

Templating Control QC

Average Reads Per Lane

Average number of CF-1 reads per GX5TM Chip lane.

Base Call Accuracy

1 ­ (total number of errors for all positions in the control / total number of aligned bases).

Mean AQ20 Read Length (bp) Average length, in base pairs, at which the error rate is 1% for all aligned reads of the CF-1 control fragment.

Library QC

AQ20 Mean Read Length

Average read length, in base pairs, at which the error rate is 1% for all aligned reads.

Percent Reads On Target

The percentage of reads mapped on target to total reads.

Sample Balance Factor

For a given sample in a system install qualification run, Sample Balance Factor = Number of mapped reads / Mean number of mapped reads for four tested samples A­D.

Pass threshold
0 0 0
90 90 0.5

Performance Qualification results
For a Performance Qualification (PQ) run, basic information about the verification run is provided, and the reference range and pass/fail QC status for the following metrics are reported for the samples and controls that are used in the run.
Manager- and administrator-level users can sign off on the PQ Report, at which point it becomes a locked PQ Report. However, we recommend that only qualified support specialists sign off on a PQ Report.

QC metric

Description

Pass threshold

Run QC

AQ20 Mean Read Length (bp) Average read length, in base pairs, at which the error rate is 1% for all aligned reads.

Average Uniformity of base coverage

The percentage of bases having a depth of coverage 20% of the mean coverage at each position.

Raw Read Accuracy

The percentage of raw reads mapped to the reference genome.

Run Balance Factor

The minimum Sample Balance Factor in a set of four tested samples A­D.

Total Reads

The total number of filtered and trimmed reads with the listed barcodes assigned to the samples.

90 90 98.5 0.5 10,000,000

GenexusTM Integrated Sequencer User Guide

135

9 Chapter 9 Review data and results Verification runs

(continued)

QC metric

Description

Templating Control QC

Average Reads Per Lane

Average number of CF-1 reads per GX5TM Chip lane.

Base Call Accuracy

1 ­ (total number of errors for all positions in the control / total number of aligned bases).

Mean AQ20 read length (bp)

Average read length, in base pairs, at which the error rate is 1% for all aligned reads of the CF-1 control fragment.

Control QC

COSM12558 COSM28747

Detection of variant yields call of PRESENT and QC Status = Pass

COSM476

COSM521

COSM6223

COSM6224

COSM760

Library QC

Mapped Reads

The total number of reads mapped to the reference sequence.

Mean AQ20 Read Length (bp) Average length, in base pairs, at which the error rate is 1% for all aligned reads of the sample DNA library.

Percent Reads On Target

The percentage of reads mapped on target over total reads.

Sample Balance Factor

For a given sample in a PQ run, Sample Balance Factor = Number of mapped reads / Mean number of mapped reads for four tested samples A­D.

Pass threshold
0 0 0
Present
1,100,000 90 90 0.5

136

GenexusTM Integrated Sequencer User Guide

A

Troubleshooting

Troubleshoot GenexusTM Integrated Sequencer performance with CF-1 and inline controls

You can use quality control results to troubleshoot GenexusTM Integrated Sequencer runs to help identify the cause of performance problems. If you select the Include Inline Controls checkboxes for DNA and RNA in the Reagent step of the Create Assay workflow (see "Create a new assay (manager/ administrator)" on page 35), you include the inline control analysis in the post-run results analysis. Addition of a set of six control amplicons (covering a range of amplicon length) and spike-in nucleic acid into sample library preparation reactions helps you identify whether poor performance is due to insufficient sample input and/or poor sample quality, or is unrelated to sample input and quality. With 10 ng sample input, the read ratio of endogenous sample reads to inline control reads is expected to be ~3. Using more than 10 ng sample input results in a proportionally higher read ratio. For example, if you load 20 ng of sample, the read ratio should be ~6.
The CF-1 templating control serves as a check on templating and sequencing performance that is independent of library preparation.
Use the following table as a guide to help identify the source of performance problems. For recommended actions, see the troubleshooting topics under "GenexusTM Integrated Sequencer-- general and QC troubleshooting" on page 138.

QC category
Sample QC (endogenous sample reads)
Read ratio for inline controls (endogenous to spike-in reads)
Templating Control QC - CF-1

Successful run Passed
Normal Read ratio ~3
Passed

Run diagnostic

Sample input and/or quality problem

Library preparation problem unrelated to
sample

Templating or Sequencing problem

Failed

Failed

Failed

Low Read ratio <<3
Passed

Normal or variable Passed

-- Failed

GenexusTM Integrated Sequencer User Guide

137

A

Appendix A Troubleshooting GenexusTM Integrated Sequencer--general and QC troubleshooting

GenexusTM Integrated Sequencer--general and QC troubleshooting

Observation Consumable is not recognized by instrument after loading on the deck
Run Status = Failed Details: In the GenexusTM Software Run Result screen, the Run Status for a completed run is listed as "Failed". In the Sample Results screen, the Sample Status is listed as "BaseCallingActor FAILED". Used lane not crossed out in the sequencer screen Details: After completion of a run, the lane used in the run was not crossed out, so that the next run could reuse the lane.

Possible cause The consumable (for example, a strip, barcode plate, pipette tip box) is correctly placed but is not completely inserted into its position, causing it to be misaligned with its expected position. The barcode of the consumable is not readable by the instrument.
Consumable version does not match the GenexusTM Software version. For example, a consumable compatible with GenexusTM Software 6.0 is installed in a sequencer updated for GenexusTM Software 6.2.0. Chip calibration failed due to a chip problem, or an instrument problem.
A chip problem caused a datacollect failure to write efuse.

Recommended action Ensure that the consumable is pressed completely into place. Apply firm pressure on the item so that it fits snugly into its deck position.
Tap Help in the lower left corner of the Load Instrument screen and follow on-screen instructions to override the block manually. Note that the name of the consumable does not appear in the list of consumables in the run summary. If the behavior continues in subsequent runs, contact Technical Support. Ensure that you are using consumables compatible with the software installed on the sequencer.
Repeat the run with a new chip. If the problem persists, contact Technical Support.
In the sequencer screen, tap Settings4Clean instrument to perform a clean instrument. For details, see "Perform a Clean instrument procedure" on page 152. After cleaning, start a new run.

138

GenexusTM Integrated Sequencer User Guide

Appendix A Troubleshooting GenexusTM Integrated Sequencer--general and QC troubleshooting

A

Observation
The number of sample reads is low, CF-1 metrics pass QC, but read ratio of inline controls is low.
Details: If CF-1 reads per lane, accuracy, and mean AQ20 read length are good, and read ratio of inline controls (endogenous vs. spike-in) is low (<< 3), a problem with sample input is indicated. For more information, see "Troubleshoot GenexusTM Integrated Sequencer performance with CF-1 and inline controls" on page 137.
The number of sample reads is low, but CF-1 metrics pass QC, and read ratio of inline controls is normal
Details: If CF-1 metrics passed QC, and read ratio of inline controls is normal (~ 3), a problem in library preparation unrelated to sample input or quality may be indicated. For more information, see "Troubleshoot GenexusTM Integrated Sequencer performance with CF-1 and inline controls" on page 137.
The number of sample reads is low, and CF-1 metrics fail QC
Details: If CF-1 metrics failed QC, a problem in templating or sequencing is indicated. For more information, see "Troubleshoot GenexusTM Integrated Sequencer performance with CF-1 and inline controls" on page 137.

Possible cause Nucleic acid input may have been insufficient, and/or the nucleic acid was degraded.
One or more of the GenexusTM Strip 1 strips used in the run had magnetic beads trapped in the tube 5 keyhole. An incorrect assay was selected for the run, or library amplification parameters were not optimal. Library strips were inadequately equilibrated to room temperature (GenexusTM Strip 1), or incompletely thawed (GenexusTM Strip 2AS or GenexusTM Strip 2HD) before loading in the sequencer. One or more of the GenexusTM Strip 3GX5TM strips used in the run may have had an excessive amount of magnetic beads trapped in the tube 6 or 7 keyhole. Template strips were inadequately equilibrated to room temperature (GenexusTM Strip 3GX5TM), or incompletely thawed (GenexusTM Strip 4) before loading in the sequencer. The sequencing chip or coupler was faulty or leaky.

Recommended action For a sample run, re-quantify nucleic acid samples and/or perform sample QC to ensure that the expected nucleic acid input and size was loaded. If needed, re-isolate and purify nucleic acid samples.
Repeat the run with strips that you have verified have no trapped beads. For more information, see "Before you begin" on page 79.
Ensure that you have selected the correct assay and reviewed assay parameters.
Ensure that GenexusTM Strip 1 strips are fully equilibrated to room temperature, and GenexusTM Strip 2AS strips are completely thawed before loading in the sequencer.
Repeat the run with strips that you have verified have no trapped beads. For more information, see "Before you begin" on page 79.
Ensure that GenexusTM Strip 3GX5TM strips are fully equilibrated to room temperature, and GenexusTM Strip 4 strips are completely thawed before loading in the sequencer.
Repeat the run with new chip and coupler. If low performance continues, contact Technical Support.

GenexusTM Integrated Sequencer User Guide

139

A

Appendix A Troubleshooting GenexusTM Integrated Sequencer error and warning messages

Observation
The number of sample reads is low, and CF-1 metrics fail QC
Details: If CF-1 metrics failed QC, a problem in templating or sequencing is indicated. For more information, see "Troubleshoot GenexusTM Integrated Sequencer performance with CF-1 and inline controls" on page 137.
(continued)

Possible cause
The run was started >14 days after the last initialization was performed, or on an expired initialization.

Recommended action
Perform a Clean instrument procedure (Settings4Clean instrument), install new a chip, and new sequencing reagent bottles and cartridge in the sequencing reagents bay, then repeat the run.
Note: Reagents are stable on the sequencer for 14 days, after which you may experience reduced performance. For further information, see "Options for an expired sequencer initialization" on page 93.

GenexusTM Integrated Sequencer error and warning messages

Observation Error message: Library Prep Failed

Possible cause

Recommended action

· The sequencer failed xy Confirm that no analyses are

homing, or

in progress, then reboot the

· The sequencing chip

instrument. Repeat the run. If

failed to engage at pipette the failure continues, contact

position.

Technical Support.

Error message: Templating Failed

· The sequencer failed xy homing, or
· The GenexusTM Coupler was misaligned.

Confirm that no analyses are in progress, then reboot the instrument. Repeat the run, and ensure that the coupler is installed with proper alignment. If the failure continues, contact Technical Support.

140

GenexusTM Integrated Sequencer User Guide

Appendix A Troubleshooting GenexusTM Integrated Sequencer error and warning messages

A

Observation Error message: Init Failed: XXX did not reach fill
Error message: XXX clog check failed

Possible cause

Recommended action

GenexusTM Bottle 2 flow was restricted, or a flow sensor malfunctioned.
Note: XXX is the name of the conical tube that did not reach fill.

Allow the run to complete, then select Settings4Clean instrument to do a clean. After the clean, reboot the instrument, then start a new run. If the error message appears again, call Technical Support for service.

Clog check following Clean Instrument failed. Incomplete fluid flush or debris has caused a clog.

Note: After the run completes, you can recover libraries from the amplification plate, if desired. For more information, see "Library QC Archive: recover library preparations from the GenexusTM Integrated Sequencer for reuse" on page 172.
Contact Technical Support for recommended action.

Note: XXX refers to the failed waste lines, chip, or main 1-4.

Error message: Flow rate from W3 failed ­ below 50 L/s

During the post-chip clean the GenexusTM Bottle 3 was empty, or GenexusTM Bottle 3 flow was
restricted.

Select Settings4Clean instrument to do a clean. Ensure that the GenexusTM Bottle 3 is installed correctly, and the bottle contains sufficient volume. Replace as needed. For details, see "Perform a Clean instrument procedure" on page 152. If the failure persists, contact Technical Support.

GenexusTM Integrated Sequencer User Guide

141

A

Appendix A Troubleshooting GenexusTM Integrated Sequencer error and warning messages

Observation Error message: PostChipClean failed
Warning message: Pressure Leak Detected in vac pressure drop leakrate. Press Ok to open doors and fix the leak.
Warning message: Pressure Leak Detected in vac time to pressure. Press Ok to open doors and fix the leak.
Warning message: Pressure Leak Detected in vac pressure baseline. Press Ok to open doors and fix the leak.

Possible cause PostChipClean aborted due to a chip error.
GenexusTM Bottle 1 leak was detected.
A GenexusTM Bottle 1 leak or over-fill was detected.
GenexusTM Bottle 1 leak was detected.

Recommended action
In the sequencer screen, tap Settings4Clean instrument to re-do a Clean instrument procedure. For details, see "Perform a Clean instrument procedure" on page 152. You may or may not need to replace the chip during deck set up.
Remove the GenexusTM Bottle 1 and ensure that the cap is tight, then replace the bottle securely on the instrument. Continue with the run. If the pressure leak message appears again, replace the GenexusTM Bottle 1 with a new bottle, then continue with the run. If the failure persists, contact Technical Support.
Ensure that a new GenexusTM Bottle 1 is installed, and the cap is tight and the bottle is secure on the instrument. Continue with the run. If the pressure leak message appears again, replace the GenexusTM Bottle 1 with a new bottle, then continue with the run. If the failure persists, contact Technical Support.
Remove the GenexusTM Bottle 1 and ensure that the cap is tight, then replace the bottle securely on the instrument. Continue with the run. If the pressure leak message appears again, replace GenexusTM Bottle 1 with a new bottle, then continue with the run. If the failure persists, contact Technical Support.

142

GenexusTM Integrated Sequencer User Guide

Appendix A Troubleshooting GenexusTM Integrated Sequencer error and warning messages

A

Observation Warning message: Pressure Leak Detected in W2P. Press Ok to open doors and fix the leak.
Warning message: Pressure Leak Detected in W22P. Press Ok to open doors and fix the leak.
Warning message: Pressure Leak Detected in W3P. Press Ok to open doors and fix the leak.
Warning message: Pressure Leak Detected in Conicals & Reagent Cartridge. Press Ok to open doors and fix the leak.

Possible cause A leak was detected in the GenexusTM Bottle 2 at the left position.
A leak was detected in the GenexusTM Bottle 2 installed at the right position.
A leak was detected in the GenexusTM Bottle 3.
A leak was detected in one or more of the GenexusTM Conical Bottles or the GenexusTM Cartridge.

Recommended action
Remove GenexusTM Bottle 2 installed at the left position, ensure that the cap is tight, then reinstall the bottle securely. Continue with the run. If the pressure leak message is seen again, replace the left GenexusTM Bottle 2 with a new bottle, then continue with the run. If the failure persists, contact Technical Support.
Remove the GenexusTM Bottle 2 installed at the right position, ensure that the cap is tight, then reinstall the bottle securely. Continue with the run. If the pressure leak message appears again, replace the right GenexusTM Bottle 2 with a new bottle, then continue with the run. If the failure persists, contact Technical Support.
Remove the GenexusTM Bottle 3, ensure that the cap is tight, then reinstall the bottle securely. Continue with the run, or repeat the clean if this message appears in a postChipClean. If the pressure leak message appears again, replace the GenexusTM Bottle 3 with a new bottle. Continue with the run, or repeat the clean if this message appears in a postChipClean. If the failure persists, contact Technical Support.
Tighten all GenexusTM Conical Bottles one bottle at a time to prevent accidental position changes. Inspect the GenexusTM Cartridge for defects. Continue with the run. If the failure persists, contact Technical Support.

GenexusTM Integrated Sequencer User Guide

143

A Appendix A Troubleshooting GenexusTM Software

Observation
Warning message: Liquid detected in conicals. Press Ok to open doors and fix the issue.

Possible cause
The previous postChipClean did not remove all of the liquid in the GenexusTM Conical Bottles, causing failure of the Conicals empty test and Conical volume test.

Recommended action
Replace the GenexusTM Conical Bottles with new bottles, then tap Settings4Clean instrument to clean the instrument. After cleaning, start a new run.

Note: If a leak check fails in a run starting with lanes 2­4, tap Help in the lower left corner of the screen to allow use of the remaining sequencing reagents in the next run.

GenexusTM Software

Observation Cannot sign in to GenexusTM Software
Batch sample import fails

Possible cause You either entered an incorrect password or you are signed out due to several failed login attempts. One or more entries in the sample-import spreadsheet contains special characters, lines breaks, unexpected spaces, incorrect entry length, incorrect date formatting, or other formatting errors. Blank rows were copied into the sample-import template file from a different source.
The sample import spreadsheet contains a nonunique sample name.
The headings in the sample import spreadsheet do not match the sample attributes in the software.

Recommended action Contact the GenexusTM Software system administrator.
Check each entry for correct formatting, correct any errors, and repeat the import.
Rows that appear empty can contain hidden formatting that conflicts with the import function. Start with a clean sample-import template file, and be careful to copy only those rows that contain actual data. Every sample name in the software must be unique. Ensure that the spreadsheet does not contain any duplicate sample names, and repeat the import. Note that the system check is not case-sensitive, so a sample name of ABC1 conflicts with abc1. The headings must match the sample attributes in the software exactly. Check the headings for spelling or other errors.

144

GenexusTM Integrated Sequencer User Guide

A Appendix A Troubleshooting GenexusTM Software

Observation Library batch import fails
The assay I created does not appear in the menu when I plan a run Cannot upload my panel or hotspots Allele coverage does not match hotspot coverage

Possible cause
One or more entries in the library batch import spreadsheet contains special characters, lines breaks, unexpected spaces, incorrect entry length, incorrect date formatting, or other formatting errors. Blank rows were copied into the library batch import template file from a different source.
The library batch import spreadsheet contains a nonunique Library Batch ID.
A sample name entered in the library batch import spreadsheet does not match a sample name listed in the Manage Samples screen. The Barcode ID name format does not exactly match the format that is used in the Prepare Library Batch dialog box. An invalid library, control, or panel kit barcode has been entered in the spreadsheet. Forgot to lock your assay.

Recommended action Check each entry for correct formatting, correct any errors, and repeat the import.
Rows that appear empty can contain hidden formatting that conflicts with the import function. Start with a clean library batch import template file, and be careful to copy only those rows that contain actual data. Every Library Batch ID in the software must be unique. Ensure that the spreadsheet does not contain any duplicate IDs, and repeat the import. Note that the system check is not case-sensitive, so a Library Batch ID of ABC1 conflicts with abc1. Ensure that the sample names entered into the spreadsheet are correct and match an existing sample name added to the software.
Use the name format following the Barcode ID name format found in the Barcode Set reference lists (Settings4References4Barcode Set), for example: IonDual_0101 through IonDual_0196, or IonHDdual_0101 to IonHDdual_0132. Ensure that you have correctly entered a valid kit barcode in the appropriate cell of the spreadsheet. Go to Assay tab4Manage Assays and make sure that the assay is locked.

Issues with BED file format or files do not end in (.bed).
The coverage value reported under the Variants tab and Allele Coverage tab can be different.

Ensure your file is in the correct BED format and has a (.bed) extension.
No action required. The coverage value reported under the Variants tab is the coverage after down-sampling, while the Allele Coverage tab reports the raw coverage without downsampling. Down-sampling can speed up variant calling for some over-sampled positions.

GenexusTM Integrated Sequencer User Guide

145

A Appendix A Troubleshooting GenexusTM Software

Observation Variants tab is missing hotspot entries. Details: The remaining entries are present.
The results of the run are not showing up in the Completed Runs & Reports screen Cannot download run result files

Possible cause

Recommended action

Hotspot BED file contains entries that are incorrectly formatted.

Check that BED file entry is correctly formatted. See the following examples:
SNP entry: chr1 2337276 2337277 SVA_322 0 + REF=C;OBS=T;ANCHOR=G AMPL

Deletion entry: chr1 201341175 201341180 SVA_497 0 + REF=AGAAG;OBS=;ANCHOR=C AMPL

The instrument disk space is full.

Insertion entry: chr1 236978992 236978992 SVA_621 0 + REF=;OBS=TCTG;ANCHOR=T AMPL
Confirm the REF values match the actual reference coordinate of hg19.
Clear disk space on the sequencer. For more information, see "Manually delete run data" on page 159.

The run failed. Links to run files are not available for runs that fail QC.

Create a reanalysis assay with reduced QC parameters, then reanalyze your sample.
Repeat the run.

146

GenexusTM Integrated Sequencer User Guide

B
Touchscreen icons

Touchscreen reference

123

4

Number 1

2

3

Icon

5

6

Description Network connectivity ­ connected Network connectivity ­ not connected Instrument idle Sequencing in progress Instrument ready Error Chip status ­ Absent

GenexusTM Integrated Sequencer User Guide

147

B Appendix B Touchscreen reference Touchscreen icons

(continued)

Number

Icon

Description

3

Chip status ­ Standby

Chip status ­ Connecting

Chip status ­ Ready

Chip status ­ Imaging

Chip status ­ Error

4

1234

Chip lane status ­ 4 lanes available

234

Chip lane status ­ 1 lane in use or used, 3 lanes available

34

Chip lane status ­ 2 lanes in use or used, 2 lanes available

4

Chip lane status ­ 3 lanes in use or used,1 lane available

Chip lane status ­ 4 lanes in use or used, 0 lanes available

5

Instrument File System Space, the percent of file space used is

indicated.[1] The instrument checks for sufficient disk space before

each run and notifies the user if there is not enough.

6

Instrument Server File System Space, the percent of file space used

is indicated.

Note: If the indicator turns red, archive data from the server to free up disk space. Refer to the GenexusTM Software Help for more information on archiving data.

[1] Indicator turns yellow when disk space is 67% full, indicator turns red when 90% full.

148

GenexusTM Integrated Sequencer User Guide

B Appendix B Touchscreen reference Settings
Settings
In the Settings menu, users can view and/or change instrument settings, manage data and network configurations, and clean the instrument.
Note: The System tools option is for use by trained service personnel only.
Network Settings
The Network Settings menu allows you to configure IP address, Ion TorrentTM Server, and FTP settings.

GenexusTM Integrated Sequencer User Guide

149

B Appendix B Touchscreen reference Settings

Item
IP Address Configuration

Description
Allows users to set or change the IP configuration (DHCP or Static IP).

When/How to use
1. Select either the DHCP or Static IP radio button.
2. Touch the screen in the field you want to edit to activate the field. A virtual keyboard appears.
3. Enter the new information, then press Done.

Ion TorrentTM Server Configuration

Allows users to change the Ion TorrentTM Server IP address and user information.

When a change to the Ion TorrentTM Server IP address or user information is required.
Note: Do not change address unless directed to do so by a Thermo Fisher Scientific field service engineer.

Enter the following IP address in the Torrent Server URL field for your instrument: 127.0.0.1

150

GenexusTM Integrated Sequencer User Guide

B Appendix B Touchscreen reference Settings

(continued)
Item
FTP Server Configuration

Description
Allows users to change the FTP Server IP address and user information.

When/How to use
When a change to the FTP Server IP address or user information is required.
Note: Do not change address unless directed to do so by a Thermo Fisher Scientific field service engineer.

Enter the following IP address in the FTP server field for your instrument: 127.0.0.1

GenexusTM Integrated Sequencer User Guide

151

B Appendix B Touchscreen reference Settings
Perform a Clean instrument procedure
Cleaning is normally performed automatically at the completion of the previous sequencing run. Perform a Clean instrument procedure if the sequencing run:
· was aborted or had a power failure during a run, or · the normal post-sequencing run cleaning was not completed.
The Clean instrument command initiates the following cleans: · UV clean--irradiation of the deck surface with UV light (see "Clear the instrument deck and perform a UV Clean" on page 90) · Vacuum clean--cleaning of the four chip vacuum lines and robotic waste line. · postChipClean--cleaning of sequencer fluidic lines with conical flow rate test.
IMPORTANT! The Clean instrument procedure renders remaining sequencing reagents and unused lanes on the installed chip unusable after the cleaning. Tap Help in the lower left corner of the screen to allow use of remaining sequencing reagents in the clean procedure.
1. In the Settings menu, tap Clean instrument.
2. Follow the on-screen instructions, then tap Next to go to the following sequence of screens: Clear Deck if there are items to be cleared, UV Clean, Load Deck, Clear Sequencing Reagents, and Load Sequencing Reagents. For more information, see "Clear the instrument deck and perform a UV Clean" on page 90.
3. After cleaning is complete, tap Next. Follow the on-screen instructions in the Clear Deck and Clear Sequencing Reagents screens. After sequencing reagents are cleared, tap Next.
The user interface returns to the home screen.

152

GenexusTM Integrated Sequencer User Guide

B Appendix B Touchscreen reference Settings
Replace GenexusTM Conical Bottles during a Clean instrument
If the sequencer detects restricted flow from the GenexusTM Conical Bottles during a Clean instrument procedure, the following alert is displayed.

If you see this alert, follow these steps to replace the five GenexusTM Conical Bottles. 1. Tap OK. The screen returns to the run screen.
2. Tap Settings4Clean instrument again.
3. Follow the screen instructions to clear the deck, then load the deck.
4. When the sequencing reagent bay doors open, do the following. a. Remove, but do not discard, the GenexusTM Bottle 1, the two GenexusTM Bottle 2 bottles, and the GenexusTM Bottle 3 to gain easier access to the GenexusTM Conical Bottles.
b. Remove the five used GenexusTM Conical Bottles, then install five new conical bottles.
c. Replace the four reagent bottles that you removed in their original positions.
d. Tap Help in the lower left corner of the screen to ignore the reagent bottles and GenexusTM Cartridge as used so they can be reused in the next clean.
5. Close the sequencing reagent bay doors. The Clean instrument procedure begins automatically.
6. After the cleaning finishes, follow the screen instructions to clear the deck and remove the sequencing reagents. For further information, see "Replace the GenexusTM Conical Bottles" on page 165.

GenexusTM Integrated Sequencer User Guide

153

B Appendix B Touchscreen reference Settings
System Tools
The System Tools menu enables you to upload instrument diagnostics, manage data, and shut down or reboot the instrument.

Item RFID

Description

When/How to use

Lists the product, product expiration date, and remaining To determine if reagents that were left

uses of the product.

on an instrument during a period of

instrument shutdown have expired.

1. Tap the screen within the row of the product to view an RFID Tag Detail page.
2. Tap Done to return.

Note: The RFID screen can also be used to track remaining days of reagent use before a sequencer initialization expires. We recommend that you use reagents within 14 days of initialization. After 28 days, the sequencer displays a warning at run start with options to proceed to a clean and new initialization, ignore to continue without clean, or cancel the run. For further information, see "Options for an expired sequencer initialization" on page 93.

154

GenexusTM Integrated Sequencer User Guide

(continued) Item
Noise Test

Description
Provides real-time measurement of electrical noise readings on the chip.

B Appendix B Touchscreen reference Settings
When/How to use For troubleshooting if directed to do so by Technical Support.

Chip Calibration Checks the status of a chip.
IMPORTANT! Do not perform a chip calibration unless instructed to do so by a field service engineer.

1. Insert a chip into the chip install slot.
2. Tap Chip Calibration to start the chip calibration.
3. In the dropdown lists, select the Image to display and the type of Calibration to perform.
4. After a chip calibration, user should perform a Clean instrument before removing the chip.
Note: Tap Cancel to quit the chip calibration.

GenexusTM Integrated Sequencer User Guide

155

B Appendix B Touchscreen reference Settings

(continued)
Item
Filesystems and disks

Description
Provides real-time drive activity status and disk space used.

When/How to use
Use if directed to do so as part of a troubleshooting procedure:
1. In the System Tools menu, press Filesystems and disks. The Filesystems screen opens.
2. Tap the screen within the row of the Filesystem you want to view, then press Details. The Disk Drives in <Filesystem> screen opens.
3. Tap the screen within the row of the Drive to view details of the disk drive. The Details for <Drive> screen opens.
4. Tap Done to return to the previous screen.

156

GenexusTM Integrated Sequencer User Guide

B Appendix B Touchscreen reference Settings

(continued)
Item
Upload Diagnostics

Description
Allows upload of instrument diagnostics files for review by support personnel.

When/How to use
Upload an instrument diagnostics file after an aborted run, or if otherwise directed to do so by Technical Support.
Note: The file is pushed to GenexusTM Software. To retrieve the file using GenexusTM Software, navigate to (Settings)4About, click Instrument Diagnostics, then follow the on-screen instructions.

Shut Down

Access to "Shut Down" and "Reboot" commands.

If directed to do so by Technical Support as part of a troubleshooting procedure, or if the instrument will not be used for more than three days.
Note: It is not necessary/recommended to shut down the instrument overnight or over the weekend. To power on the instrument after a shut down, see "Power the GenexusTM Integrated Sequencer on or off" on page 28.

GenexusTM Integrated Sequencer User Guide

157

B Appendix B Touchscreen reference Settings

Data Management

The Data Management function allows users to delete run data manually, or transfer data in the event of failure of automatic transfer. Under normal conditions, run data are automatically transferred to the analysis server, then deleted from the instrument hard drive.

Item Delete data

Description Manually delete run data from the instrument.

Transfer data

When/How to use
If the instrument hard drive becomes full, see "Manually delete run data" on page 159.
Transfer run data from the instrument hard drive to the Ion TorrentTM Server.
1. Tap the screen in the Experiment Name row of the individual experiment to be transferred.
2. Tap Transfer. 3. Select the radio button of the
action to be performed. 4. Select the radio button of the files
to be transferred. 5. Tap OK.

Analysis Controls

Allows the user to stop analysis the data of individual runs.

1. Tap the screen in the Experiment Name row of the individual experiment to be analyzed, then tap Analysis.
2. Select the Stop Analysis radio button, then tap OK.
Note: Restart analysis is currently not supported.

158

GenexusTM Integrated Sequencer User Guide

B Appendix B Touchscreen reference Settings
Manually delete run data
To troubleshoot data management problems the Data Management function allows users to delete run data manually or transfer the data to an external server.
1. In the Settings menu, tap Data Management to access the Data Management screen, then tap Manage.

2. Tap Select All to select all the available experiments, or tap the screen in the Experiment Name row of the individual experiment to be managed.

3. Tap Delete Selected.

GenexusTM Integrated Sequencer User Guide

159

B Appendix B Touchscreen reference Settings
Instrument settings
The Instrument Settings menu provides information about the instrument and allows you to set the instrument name and calibrate the touchscreen.

Item About

Description Provides instrument details.

When/How to use
To view instrument reference information or access regulatory information. Click Regulatory info to view the following screen.

160

GenexusTM Integrated Sequencer User Guide

(continued)

Item

Description

Regulatory Info Lists instrument-specific regulatory information.

B Appendix B Touchscreen reference Settings
When/How to use To view instrument RFID regulatory information. See "Radio compliance standards" on page 201 for compliance information.

Instrument Name

To change the instrument name.

GenexusTM Integrated Sequencer User Guide

161

B Appendix B Touchscreen reference Settings

(continued)
Item
Screen Calibration

Description

When/How to use
For troubleshooting if directed to do so by Technical Support.
Touch the red cross with your finger or a stylus each time it appears. In total, you touch the screen 4 times, one time in each corner.

Smart Monitor -- Enabled

Select the checkbox to enable remote monitoring of instrument runs by Thermo Fisher Scientific for troubleshooting.

162

GenexusTM Integrated Sequencer User Guide

C

Supplemental information

 Maintain the sequencer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163  Quantify FFPE DNA with the QubitTM Fluorometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171  Library QC Archive: recover library preparations from the GenexusTM Integrated
Sequencer for reuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
 Guidelines for using custom assays with the GenexusTM Integrated Sequencer . . . . . . . . . . . . . . 176  Planning sequencing runs for efficient use of consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178  Configure an Ion ReporterTM Server account (administrator) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179  Configure Thermo Fisher Accounts in GenexusTM Software (administrator) . . . . . . . . . . . . . . . . . . 182
Maintain the sequencer
Materials required
Cleaning · Lint-free wipes · 70% isopropanol · (Optional) 10% bleach solution
Liquid waste filter replacement · GenexusTM Filter (ordered separately [Cat. No. A40302])
Conical bottle replacement · GenexusTM Conical Bottles (from the GenexusTM Installation and Training Kit, or ordered separately [Cat. No. A40275])

GenexusTM Integrated Sequencer User Guide

163

C Appendix C Supplemental information Maintain the sequencer
Clean or decontaminate the sequencer
If a spill or leak occurs inside the instrument, perform the following steps to clean or decontaminate the sequencer.
Note: Dispose of all waste in appropriate liquid or solid waste containers.
1. Remove each GenexusTM Bottle 2, then remove and empty the GenexusTM Bottle 1 and Waste Carboy.
2. Remove the GenexusTM Cartridge.
3. Inspect the floor of the sequencing reagents bay and GenexusTM Cartridge port for liquid.
4. Using absorbent paper, soak up as much liquid as possible, then wash the affected area with 10% bleach solution.
5. Wipe the affected surfaces with 70% isopropanol, then allow to air-dry.
Replace the GenexusTM Filter
The GenexusTM Filter captures particulate matter in the liquid waste to prevent blockage of the waste line over time. The filter needs to be replaced with a new filter after one year of regular instrument use.
1. Remove the old filter from the liquid waste disposal port on the instrument deck by grasping the filter firmly and pulling up. Dispose of the filter as regular waste.
Note: If an interim filter is installed in the port, use a straight needle probe to pierce the filter, then lift out the filter using the probe.
2. Insert a new GenexusTM Filter into the liquid waste disposal port, then press firmly to seat the filter O-ring securely.

1

2

3

4

1 GenexusTM Filter 2 Liquid waste disposal port 3 Used pipette tip disposal port

164

5 4 Sample plate station 5 Installed filter
GenexusTM Integrated Sequencer User Guide

C Appendix C Supplemental information Maintain the sequencer
Replace the GenexusTM Conical Bottles
GenexusTM Conical Bottles need to be replaced when filters in the bottles become partially clogged, resulting in reduced flow rate. The GenexusTM Integrated Sequencer measures flow rate in a Conical Flow Rate test that is performed during every postChipClean. If the conical flow rate test fails, the sequencer automatically redirects to the conical bottle replacement procedure. During a normal run, a postChipClean is performed in three situations:
· At the end of the run if all four lanes have been used. · At the end of the run if you selected Do Force Clean at run setup. · At run setup when a run is selected but insufficient unused lanes in the installed chip are available
for the run. If you continue with the run, a postChipClean is needed before the run can proceed. Follow these steps to replace the GenexusTM Conical Bottles.
1. After a run, In the Run Complete screen, tap Next.
If the Conical Flow Rate test failed, a screen appears directing you to open the sequencing reagents bay doors.

GenexusTM Integrated Sequencer User Guide

165

C Appendix C Supplemental information Maintain the sequencer 2. Open the sequencing reagents bay doors, then tap Next.
The Replace Conicals screen appears.

166

GenexusTM Integrated Sequencer User Guide

C Appendix C Supplemental information Maintain the sequencer

3. Remove, but do not discard, the GenexusTM Bottle 1, the two GenexusTM Bottle 2 bottles, and the GenexusTM Bottle 3 to gain easier access to the GenexusTM Conical Bottles.

2

2

1

3

4

1 GenexusTM Bottle 1 2 GenexusTM Bottle 2

3 GenexusTM Bottle 3 4 GenexusTM Conical Bottles

4. Remove the five used GenexusTM Conical Bottles, then install five new conical bottles.

Note: Discard the used conical bottles appropriately.

5. Replace the four reagent bottles that you removed in step 3 in their original positions.
6. After replacing the bottles, close the sequencing reagents bay doors, then tap Next in the Replace Conicals screen. After you tap Next, the sequencer starts a Leak Test.

GenexusTM Integrated Sequencer User Guide

167

C Appendix C Supplemental information Maintain the sequencer

· If the sequencer passes the Leak Test, no action is needed. The sequencer proceeds to a cleaning cycle.

· If the sequencer fails the Leak Test, a notification appears. Tap OK to return to step 2 to fix the leak. Open the sequencing reagents bay doors, check the tightness of each conical bottle that you installed, then ensure that the four reagent bottles are installed correctly. Close the sequencing reagents bay doors, then tap Next in the Replace Conicals screen to repeat the Leak Test.
After completion of a successful Leak Test and cleaning, the sequencer moves to a normal workflow.
· If the sequencer is at the end of the run, the next screen is a Clear Deck screen, followed by a Clear Sequencing Reagents screen.
· If the sequencer is at the start of the run, the next screen is a Clear Deck screen, followed by a Load Deck screen.

168

GenexusTM Integrated Sequencer User Guide

C Appendix C Supplemental information Maintain the sequencer
GenexusTM Integrated Sequencer power off and power on before and after a long-term shutdown
Follow these procedures to power off the sequencer if it will not be used for more than 28 days, and power on after a long-term shutdown.
Power off before a long-term shutdown
· If all four lanes of a chip have been used, and a postchip clean has completed successfully, follow these steps to power off the sequencer, as described in "Power off" on page 28. a. In the home screen, tap Settings4System Tools4Shut down.
b. Select Shutdown. A confirmation message appears. Select Yes to power off the instrument.
c. Turn the power switch located at the back of the instrument to the off ( O ) position.
· If a partially used chip is installed on the instrument, follow these steps to perform a Clean instrument procedure as described in page 152, then power off the instrument. a. In the home screen, tap Settings4Clean Instrument.
b. Follow the on-screen instructions to start the cleaning.
Note: If an alarm appears, follow the on-screen prompts, or contact technical support for further help. Do not continue until the problem is resolved.
c. After the cleaning completes, power off the system in the home screen by tapping Settings4System Tools4Shut down.
d. Select Shutdown. A confirmation message appears. Select Yes to power off the instrument.
e. Turn the power switch located at the back of the instrument to the off ( O ) position.
Power on after a long-term shutdown
After a long-term shutdown (>28 days), follow these steps to power on the sequencer, as described in "Power on" on page 28.
1. Turn the power switch on the back of the sequencer to the on ( | ) position.
2. Press the power button on the front of the instrument. The button illuminates.
3. Sign in with your user name and password. After the home screen appears, the sequencer is ready for use. Check for instrument alarms, if any. If a postchip clean was performed before the system was powered off, the instrument is ready to use.

GenexusTM Integrated Sequencer User Guide

169

C Appendix C Supplemental information Maintain the sequencer
4. If a postchip clean was not performed before the sequencer was powered off, perform a Clean instrument procedure as described in "Perform a Clean instrument procedure" on page 152. a. In the home screen, tap Settings 4Clean Instrument.
b. Follow the on-screen instructions to start the cleaning.
Note: If an alarm appears, follow the on-screen prompts, or contact technical support for further help. Do not continue until the problem is resolved.
c. After the cleaning is complete, the instrument is ready to use.
5. If the instrument was in an initialized state before the system was powered off, and shows the following alarm for an expired initialization, perform the following steps to clean the instrument as described in "Options for an expired sequencer initialization" on page 93.

a. Select Cancel to return to the home screen, then tap Settings 4Clean Instrument. b. Follow the on-screen instructions to start the cleaning.
Note: If an alarm appears, follow the on-screen prompts, or contact technical support for further help. Do not continue until the problem is resolved.
c. After the cleaning is complete, the instrument is ready to use.

170

GenexusTM Integrated Sequencer User Guide

C Appendix C Supplemental information
Quantify FFPE DNA with the QubitTM Fluorometer
Quantify FFPE DNA with the QubitTM Fluorometer
When using the Ion AmpliSeqTM Direct FFPE DNA Kit, the DNA concentration can be estimated using a QubitTM Fluorometer and the QubitTM dsDNA HS Assay Kit (Cat. No. Q32851). See the QubitTM dsDNA HS Assay Kits User Guide (Pub. No. MAN0002326) for more information.
1. Set up the required number of 0.5mL QubitTM Assay tubes for standards and samples. The QubitTM dsDNA HS Assay requires 2 standards.
2. Prepare sufficient QubitTM working solution for all samples and standards by diluting QubitTM dsDNA HS Reagent 1:200 in QubitTM dsDNA HS Buffer.
3. Combine 2 L of the FFPE DNA sample with 198 L (200-µL final volume) of working solution, mix well, then incubate for at least 2 minutes.
4. Prepare each QubitTM standard as directed in the user guide.
5. Measure the concentration of each sample and standard on the QubitTM Fluorometer.
6. (QubitTM 2.0 Fluorometer only.) Calculate the concentration of the undiluted sample by multiplying by the dilution factor. Alternatively, use the Calculate Stock Conc. feature on your instrument.
Proceed to "Dilute or concentrate the samples (if needed) and load the sample plate--sample run" on page 76.

GenexusTM Integrated Sequencer User Guide

171

C

Appendix C Supplemental information Library QC Archive: recover library preparations from the GenexusTM Integrated Sequencer for reuse

Library QC Archive: recover library preparations from the GenexusTM Integrated Sequencer for reuse

After a sample run completes on the GenexusTM Integrated Sequencer, a variable volume of library prepared during the run is left over, and can be manually recovered from Zone 2 wells (columns 5­8) of the amplification plate. The leftover volume depends on the number of samples used per library strip. If one sample is used per strip, 7 L is present; if four samples are used per strip, 55 L is present. The recovered pre-pooled libraries can be purified, quantified, pooled, and reused in a library run in the GenexusTM Integrated Sequencer, or in a sequencing run in another Ion TorrentTM sequencer after template preparation. The recovered and purified libraries can also be archived for later use.

Required materials and equipment

Unless otherwise indicated, all materials are available through thermofisher.com.

Item AgencourtTM AMPureTM XP Reagent
DynaMagTM­96 Side Magnet, or equivalent Ethanol, Absolute, Molecular Biology Grade
Nuclease-free Water Low TE buffer (10 mM Tris, 0.1 mM EDTA, pH 8.0) EppendorfTM DNA LoBindTM Microcentrifuge Tubes, 1.5mL
MicroAmpTM Optical 96Well Reaction Plate with Barcode MicroAmpTM Clear Adhesive Film Ion Library TaqManTM Quantitation Kit Applied BiosystemsTM 7500 Fast RealTime PCR Instrument, or equivalent

Source
NC9959336, NC9933872 (fisherscientific.com)
12331D
BP2818500 (fisherscientific.com)
AM9932
--
13-698-791 (fisherscientific.com)
4306737
4306311
4468802
4351106

Recover libraries from the sequencer and purify
After a sequencing run, leftover volume from each library prepared from a sample and primer pool (7­55 L, depending on the number of samples used per library strip) is present in the plate loaded in the PCR amplification station. The library order in the plate is provided in the Run Summary tab (Results 4Run Results4Run name). The following figure shows the library positions for two samples in a run using the OncomineTM Comprehensive Assay v3 GX with two primer pools each for DNA and RNA.

172

GenexusTM Integrated Sequencer User Guide

Appendix C Supplemental information Library QC Archive: recover library preparations from the GenexusTM Integrated Sequencer for reuse

C

1
1 Library position in PCR plate
The aqueous library layer is overlayed with approximately 25 L of mineral oil. Use the following procedure to recover and purify the libraries. Before you start the procedure, do the following.
· Prepare sufficient 70% ethanol to have 400 L for each library to be recovered and purified. · Warm the AgencourtTM AMPureTM XP Reagent to room temperature and vortex thoroughly to
disperse the beads before use. Pipet the suspension slowly.
IMPORTANT! Do not substitute a DynabeadsTM-based purification reagent for the AgencourtTM AMPureTM XP Reagent.
1. Remove the PCR amplification plate from the sequencer deck, then transfer 30 L of the lower aqueous layer to a new PCR plate
Note:
· If a well contains less than 30 L, transfer the entire aqueous volume. · It is not critical if you transfer some of the oil overlay. The oil does not interfere in the purification
procedure.
2. Add 30 L of Nuclease-free Water to each well with library.

GenexusTM Integrated Sequencer User Guide

173

C

Appendix C Supplemental information Library QC Archive: recover library preparations from the GenexusTM Integrated Sequencer for reuse

3. Add 60 L (1X sample volume) of AgencourtTM AMPureTM XP Reagent to each library, then pipet up and down 5 times (with the pipettor set to 60 L) to mix the bead suspension with the DNA thoroughly. Visually inspect each well to ensure that the mixture is homogeneous. Use a new pipette tip for each library.
4. Incubate for 5 minutes at room temperature.
5. Place the plate in a magnetic rack such as the DynaMagTM-96 Side Magnet, then incubate for 2 minutes, or until the suspension clears.
6. Carefully remove the supernatant without disturbing the pellet, then discard the supernatant.
7. Add 150 L of freshly prepared 70% ethanol, then move the plate side-to-side in the two positions of the magnet for 30 seconds to wash the beads.
Note: If your magnet does not have two positions for shifting the beads, remove the plate from the magnet and gently pipet up and down 5 times (with the pipettor set at 100 L), then return the plate to the magnet and incubate for 2 minutes or until the solution clears.
8. Carefully remove the supernatant without disturbing the pellet, then discard.
9. Repeat step 7 and step 8 for a total of two 70% ethanol washes.
10. Ensure that you remove all ethanol droplets from the wells. Keeping the plate in the magnet, air-dry the beads for 5 minutes at room temperature. IMPORTANT! Residual ethanol can inhibit later reactions. If needed, centrifuge the plate and remove remaining ethanol before air-drying the beads. Under conditions of low relative humidity the beads air-dry rapidly. Do not overdry.
11. Remove the plate from the magnet, then add 50 L of Low TE to each pellet to disperse the beads.
12. Seal the plate with MicroAmpTM Clear Adhesive Film, vortex thoroughly, then briefly centrifuge to collect droplets. Alternatively, mix by pipetting at least half the total volume up and down at least 5 times before sealing the plate.
13. Incubate the plate for at least 2 minutes at room temperature.
14. Place the plate on the magnet for at least 2 minutes.
15. Remove the supernatant, which contains your purified library, and transfer it to a new labeled Eppendorf LoBindTM tube.
Proceed to "Quantify the purified libraries".

174

GenexusTM Integrated Sequencer User Guide

Appendix C Supplemental information Library QC Archive: recover library preparations from the GenexusTM Integrated Sequencer for reuse

C

Quantify the purified libraries
1. Dilute 2 L of each purified library with 198 L Nuclease-free Water for a 100-fold dilution.
2. Use the dilution to quantify your libraries with the Ion Library TaqManTM Quantitation Kit. For detailed procedures, see the Ion Library TaqManTM Quantitation Kit User Guide (Pub. No. MAN0015802), available at thermofisher.com.
Quantified libraries can be combined with sample libraries of similar panel and barcode type (Ion TorrentTM Dual Barcode) and used in library runs on the GenexusTM Integrated Sequencer, or in templating reactions for sequencing on other Ion TorrentTM platforms. For details, see "Combine libraries".
Combine libraries
After quantification, combine libraries that were prepared with different barcodes according to the assay used.

IMPORTANT! Be careful not to combine libraries barcoded with the same barcode adapter.

· For single primer pool assays, and for multiple primer pool DNA or fusion assays, adjust sample library concentration to 200 pM, then combine an equal volume of each library so that the total volume is 125 µL, or the volume specified in the setup guide for the library run that you plan.

· For DNA and fusion assays, adjust sample library concentration to 200 pM, then follow the recommendations in the table below to combine the DNA and RNA libraries prepared for a given sample in the appropriate ratio. Prepare sufficient volume so that the total volume of combined libraries is 125 µL.

Assay

DNA:RNA library ratio[1]

FFPE sample

High-molecular weight sample

OncomineTM Comprehensive v3 - GX5 - DNA and Fusions

70:30

80:20

[1] DNA:RNA library pooling ratio in system-installed assays can be found in the Library Pooling Percent DNA and Library Pooling Percent RNA assay parameters.

Note: For DNA or RNA panels with two primer pools, the fraction of each pool should be equal.

Proceed to "Plan a library run" on page 70 to use the purified and combined libraries in a library run on the GenexusTM Integrated Sequencer. To achieve the required number of reads per sample specified in the assay that is used, follow guidance during library run planning on how many samples you can combine in a single run with the sequencing chip that is loaded.
Store libraries
You can store libraries at 4­8°C for up to 1 month. For longer lengths of time, store at ­30°C to ­10°C.

GenexusTM Integrated Sequencer User Guide

175

C

Appendix C Supplemental information Guidelines for using custom assays with the GenexusTM Integrated Sequencer

Guidelines for using custom assays with the GenexusTM Integrated Sequencer

Follow these guidelines for setting up assays on the GenexusTM Integrated Sequencer if you are using a custom Ion AmpliSeqTM or Ion AmpliSeqTM HD assay.
Ion AmpliSeqTM library chemistry:
· For germline Ion AmpliSeqTM assays, start with a coverage depth of 150 reads per amplicon for calculating the Minimum Read Count Per Sample that you enter in the Panel step of assay setup. Example: Your panel has 500 amplicons in each of two primer pools. Minimum Read Counts Per Sample = (500 amplicons × 2 pools) × 150 reads/amplicon/sample = 150,000
· For somatic Ion AmpliSeqTM assays, start with a coverage depth of 2,500 reads per amplicon for calculating the Minimum Read Count Per Sample when you set up your assay. Example: Your panel has 500 amplicons in each of two primer pools. Minimum Read Counts Per Sample = (500 amplicons × 2 pools) × 2,500 reads/amplicon/sample = 2,500,000
· Refer to the information in the following table for entering the number of target amplification cycles and anneal/extend time parameters for a custom Ion AmpliSeqTM panel.

Primer pairs per pool
RNA fusion panels 12­24 25­48 49­96 97­192
193­384 385­768 769­1,536 1,537­3,072 3,073­6,144 6,145­24,576

Recommended number of amplification cycles
(10 ng high-quality DNA/RNA)[1] 28 22 21 20 19 18 17 16 15 14 13

Anneal/Extend time[2]
4 minutes 4 minutes 4 minutes 4 minutes 4 minutes 4 minutes 4 minutes 8 minutes 8 minutes 16 minutes 16 minutes

[1] Add 3 cycles for low quality (FFPE) samples. [2] For Ion AmpliSeqTM panels using a 375-bp amplicon design, add 4 minutes to the anneal/extend time recommended in the
table.

176

GenexusTM Integrated Sequencer User Guide

Appendix C Supplemental information Guidelines for using custom assays with the GenexusTM Integrated Sequencer

C

Ion AmpliSeqTM HD library chemistry:

· To calculate the Minimum Read Count Per Sample parameter that you enter during assay setup, estimate the coverage depth that you require to achieve the limit of detection (LOD) needed for your samples. The limit of detection depends on coverage depth and amount of input material.
For more information, see the Ion AmpliSeqTM HD Library Kit User Guide (Pub. No. MAN0017392).

LOD
5% 1% 0.5% 0.1%

Minimum amplicon coverage 1,400 7,000 14,000 50,000

Examples:

­ Your panel has 200 amplicons in each of two primer pools, and you require 5% LOD and 1,400 reads/amplicon for a solid tumor sample.

Minimum Read Counts Per Sample =

(200 amplicons × 2 pools) × 1,400 reads/amplicon/sample = 560,000

­ Your panel has 100 amplicons in each of two primer pools, and you require 0.1% LOD and 50,000 reads/amplicon for a cfDNA sample.

Minimum Read Counts Per Sample =

(100 amplicons × 2 pools) × 50,000 reads/amplicon/sample = 10,000,000

Note: These calculations are based on 100% uniformity and 100% on target reads. Depending on the actual uniformity of the panel, the quality of input sample, and panel on-target percentage, minimum read counts per sample may need adjustment.

· Refer to the information in the following table for entering the number of library amplification cycles for a custom Ion AmpliSeqTM HD panel.

Primer pairs per pool 12­500
501­1,000 1,001­2,000 2,001­5,000

Recommended number of cycles 20 19 18 17

GenexusTM Integrated Sequencer User Guide

177

C Appendix C Supplemental information Planning sequencing runs for efficient use of consumables
Planning sequencing runs for efficient use of consumables
GenexusTM Integrated Sequencer consumables are designed to be used most efficiently when samples are grouped in multiples of four. For example, running one sample uses the same number of primer pool tubes, library and template strips, and chip lanes as four samples. Running five or six samples requires the same consumables as eight samples. For the most economical use of reagents, therefore, you should try to group samples in your runs in multiples of four, if possible.
The following figures illustrate the relationship between sample number, the number of pools in an assay, the required number of reads per sample, and the amount of deck consumables required for a sequencing run.

One-pool assay, requiring 500K reads per sample

Two-pool assay, requiring 1M reads per sample
178

GenexusTM Integrated Sequencer User Guide

Appendix C Supplemental information Configure an Ion ReporterTM Server account (administrator)

C

Four-pool assay, requiring 8M reads per sample
Note: Eight samples at 8M reads per sample exceed the capacity of the GX5TM Chip. You can proceed with the run using eight pairs of library strips and four pairs of template strips, but you may not achieve the minimum reads per sample set for the assay.
Configure an Ion ReporterTM Server account (administrator)
An administrator-level user can configure a link to an Ion ReporterTM Server account in GenexusTM Software. When the server is configured, you can upload sample results files from GenexusTM Software directly to the Ion ReporterTM Server and use Ion ReporterTM Software for further data analysis, annotation and reporting. The uploads contain unmapped BAM files.
1. Click (Settings)4Thermo Fisher Account.
2. In the Thermo Fisher Account Settings screen, click Create Account.

GenexusTM Integrated Sequencer User Guide

179

C

Appendix C Supplemental information Configure an Ion ReporterTM Server account (administrator)

3. In the Create Thermo Fisher Account dialog box, complete the information required to configure the account.

Option

Description

Account Type

Select IonReporter Local

Name User Name Password Server

A name to identify the account in the list of Thermo Fisher Account Settings
This must a username for a valid Ion ReporterTM Server account.
Enter the password for the Ion ReporterTM Server account.
Enter the name of the Ion ReporterTM Server that you will use to upload sample results files.

Version Port

Click Get Ion Reporter Software Versions, then select the software version.
Enter the port of the Ion ReporterTM Server.

(Optional) Set as Default Account

Enable the toggle switch if you want to make this account a default account to upload sample results files.

4. Click Submit.
The configured Ion ReporterTM Server account is added to the Thermo Fisher Account Settings screen. The account can be selected in the Setup step of a run plan to automatically upload sample results files and launch analyses in Ion ReporterTM Software. Completed sample results files can also be uploaded to Ion ReporterTM Software from the Actions column in the Results / Run Results screen.
The configured account is listed in the (Settings) / Thermo Fisher Account screen. A successfully authenticated account has Active in the Status column. The software version for the Ion ReporterTM Software is listed in the Version column.

180

GenexusTM Integrated Sequencer User Guide

Appendix C Supplemental information Configure an Ion ReporterTM Server account (administrator)

C

Tag an Ion ReporterTM Software analysis workflow for use with the IonReporterUploader plugin
If you configure GenexusTM Software for use with Ion ReporterTM Software, you can select an Ion ReporterTM Software analysis workflow to transfer sample results files to and launch an analysis automatically in Ion ReporterTM Software. To limit the number of analysis workflows that are listed as available in the GenexusTM Software, you can tag workflows for use with the IonReporterUploader plugin.

IMPORTANT! If you use the Tag for IRU option to tag analysis workflows for use with the IonReporterUploader plugin, you must tag all analysis workflows that you want to be available in GenexusTM Software. After you use the tag for the first time, only analysis workflows that include the tag are visible.

1. Sign in to Ion ReporterTM Software. 2. In the Workflows tab, click the row of the workflow of interest. 3. In the Details pane, click (Actions)4Tag for IRU.
Note: To remove a previously tagged workflow from the list of workflows that are available for selection in the GenexusTM Software, click Actions4Untag from IRU.

The analysis workflows are available for selection when you perform one of the following actions.

Plan a run

Action

Description
1. In the Setup step of run planning, select Upload BAM files to Ion ReporterTM Software, then select an Ion ReporterTM Software account and software version.
2. In the Assays step, in the row of each assay, the workflow that was tagged for use with the IonReporterUploader plugin is available for selection from the dropdown list in the Ion ReporterTM Workflow column.

Upload sample results files to Ion ReporterTM Software after the sequencing run is complete

1. In the Results / Run Results screen, in the Actions column, click Upload to IR in the row of a run.
2. In the Upload Samples to Ion ReporterTM Software window, the workflow that was tagged for use with the IonReporterUploader plugin is available for selection from the dropdown list in the Ion ReporterTM Workflow column.

GenexusTM Integrated Sequencer User Guide

181

C

Appendix C Supplemental information Configure Thermo Fisher Accounts in GenexusTM Software (administrator)

Configure Thermo Fisher Accounts in GenexusTM Software (administrator)

Administrator-level users can configure a link in GenexusTM Software to one or more Connect user accounts. This account type is called a Thermo Fisher Connect account in GenexusTM Software.
When a Thermo Fisher Connect account is configured and active, administrator-level users can perform the following tasks in the GenexusTM Software:
· Download the latest software updates · Download additional plugin software · Download assay configuration packages · Download software configuration packages
Before you configure a Thermo Fisher Connect account you must have a valid Connect account on the thermofisher.com website.
1. Click (Settings)4Thermo Fisher Account.

2. If you do not have a Connect account, create a new Connect account. If you already have a Connect account, proceed to step 3.
a. Navigate to thermofisher.com/connect. The internet browser opens to the Connect home page.

b. Click Sign up now to navigate to the registration page, then enter the information that is required to create a new account.

c. Click Create account at the bottom of the screen to complete the registration.

If you are configuring your Ion ReporterTM Software on Connect account, you can also create an account by clicking Ion Reporter Software on Connect. This action opens your internet browser to the Connect sign-in page where you can click Create Account to navigate to the registration page.

3. In the Thermo Fisher Account Settings screen, click Create Account.

4. In the Create Thermo Fisher Account dialog box, enter the information that is required to create the account.

Item Account Type Name
User Name Password

Description
Select Thermo Fisher Connect.
Enter a name to identify the account in the Thermo Fisher Account Settings screen in GenexusTM Software. The name can contain only alphanumeric characters (09, AaZz), periods (.), underscores (_), or hyphens (). For example, enter Lab_Admin.
The username is the email that you used to register for the Connect account.
Enter the password for the Connect account.

182

GenexusTM Integrated Sequencer User Guide

Appendix C Supplemental information Configure Thermo Fisher Accounts in GenexusTM Software (administrator)

C

The configured account is listed in the (Settings) / Thermo Fisher Account screen. A successfully authenticated account has Active listed in the Status column.
When the Thermo Fisher Account is active, software updates from the Connect App Store are automatically uploaded to the GenexusTM Software and administrator-level users can download the updates. For more information, see Empty.
When the Thermo Fisher Account is active, software updates from the Connect App Store are automatically uploaded to the GenexusTM Software and administrator-level users can download the updates.

GenexusTM Integrated Sequencer User Guide

183

D

coverageAnalysis plugin in GenexusTM Software

 Reads statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184  Example Coverage Analysis Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187  Example charts generated by the coverageAnalysis plugin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189  Output files generated by the coverageAnalysis plugin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190
Use the coverageAnalysis plugin to view statistics and graphs that describe the level of sequence coverage produced for targeted genomic regions. The results for a run analyzed with the plugin vary based on the library type that you select when you configure the plugin. You can export some charts as graphics, such as the Amplicon Coverage Chart and the Reference Coverage Chart.

Reads statistics

The library type determines which statistics are presented. The following tables describe the statistics that are generated by the coverageAnalysis plugin. The statistics that are displayed in your report depend on the type of library that is used in your sequencing experiment. Definitions are in tooltips. Almost every statistic, plot, link, and functional widget in the report provides tooltips with definitions. Hover over a heading or description in the report to view the tooltip.
General statistics

Statistic Number of mapped reads Percent reads on target

Description
The total number of reads mapped to the reference genome.
The percentage of filtered reads mapped to any targeted region relative to all reads mapped to the reference. If no target regions file is specified, this value will be the percentage of reads passing uniquely mapped and/or nonduplicate filters, or 100% if no filters were specified. A read is considered on target if at least one aligned base overlaps at least one target region. A read that overlaps a targeted region but where only flanking sequence is aligned, for example, due to poor matching of 5' bases of the read, is not counted.

Amplicon read coverage statistics
The following statistics describe the reads that are assigned to specific amplicons. Each sequence read is assigned to exactly one of the amplicons specified by the targets file. If a read spans multiple amplicon targets, the target region that the reads covers most is assigned. In the event of a tie, the target that is the closest to the 3' end is assigned.

184

GenexusTM Integrated Sequencer User Guide

D Appendix D coverageAnalysis plugin in GenexusTM Software Reads statistics

Statistic

Description

Number of amplicons

The number of amplicons that is specified in the target regions file.

Percent assigned amplicon reads

The percentage of reads that were assigned to individual amplicons relative to all reads mapped to the reference. A read is assigned to a particular (inner) amplicon region if any aligned bases overlap that region. If a read might be associated with multiple amplicons, it is assigned to the amplicon region that has the greatest overlap of aligned sequence.

Average reads per amplicon The average number of reads assigned to amplicons.

Uniformity of amplicon coverage

The percentage of amplicons that had at least 20% of the average number of reads per amplicon. Cumulative coverage is linearly interpolated between nearest integer read depth counts.

Amplicons with at least N reads

The percentage of all amplicons that had at least N reads.

Amplicons with no strand bias The percentage of all amplicons that did not show a bias towards forward or reverse strand read alignments. An individual amplicon has read bias if it has 10 reads and the percentage of forward or reverse reads to total reads is greater than 70%. Amplicons with <10 reads are considered to have no strand bias.

Amplicons reading end-toend

The percentage of all amplicons that were considered to have a sufficient proportion of assigned reads (70%) that covered the whole amplicon target from 'endtoend'. To allow for error, the effective ends of the amplicon region for read alignment are within 2 bases of the actual ends of the region.

Amplicon base composition bias

A number that represents the proportion of amplicons showing low representation (<0.2x mean reads) in the lower and/or upper quartiles of amplicons ordered by increasing G/C base pair content of their insert sequences. The value is relative to that in the center 50th percentile of amplicons and weighted by the standard deviation of representation over all amplicons. An RMS (root mean square) value is used so that a bias greater in either upper or lower quartiles produces a larger value than a mean bias seen more equally in both outer quartiles. The value is 0 if the uniformity of amplicon coverage metric is 100%, however, the value is not necessarily high at lower amplicon uniformity.

Target base coverage statistics
The following statistics describe the targeted base reads of the reference. A base covered by multiple target regions is counted only once per sequencing read.

Statistic

Description

Bases in target regions

The total number of bases in all specified target regions of the reference.

Percent base reads on target

The percent of all bases covered by reads aligned to the reference that covered bases in target regions. Clipped bases, deletions, and insertions (relative to the reference) are not included in this percentage.
If no specific target regions was specified, the targeted regions is the whole genome.

GenexusTM Integrated Sequencer User Guide

185

D Appendix D coverageAnalysis plugin in GenexusTM Software Reads statistics

Target base coverage statistics (continued)

Statistic

Description

Average base coverage depth The average number of reads of all targeted reference bases. This is the total number of base reads on target divided by the number of targeted bases, and therefore includes any bases that had no coverage.

Uniformity of base coverage

The percentage of bases in all targeted regions (or whole genome) that is covered by at least 20% of the average base coverage depth reads. Cumulative coverage is linearly interpolated between nearest integer base read depths.

Target base coverage at Nx The percentage of target bases covered by at least N reads.

Target bases with no strand bias

The percentage of all target bases that did not show a bias toward forward or reverse strand read alignments. An individual target base is considered to have read bias if it has 10 reads and the percentage of forward or reverse reads to total reads is greater than 70%. Target bases with <10 reads are considered to have no strand bias.

Percent endtoend reads

The percentage of ontarget reads that fully cover their assigned amplicon (insert) from 'endtoend'. To allow for error, the effective ends of the amplicon region for read alignment are within 2 bases of the actual ends of the region.

186

GenexusTM Integrated Sequencer User Guide

D Appendix D coverageAnalysis plugin in GenexusTM Software Example Coverage Analysis Report
Example Coverage Analysis Report
The following is an example of a Coverage Analysis Report generated by the coverageAnalysis plugin for an Ion AmpliSeqTM DNA and fusions run. For a DNA and fusions run, the DNA and fusion coverage results are viewable by clicking the respective links in the Barcode Name column of the coverageAnalysis results table.

GenexusTM Integrated Sequencer User Guide

187

D Appendix D coverageAnalysis plugin in GenexusTM Software Example Coverage Analysis Report

188

GenexusTM Integrated Sequencer User Guide

Appendix D coverageAnalysis plugin in GenexusTM Software Example charts generated by the coverageAnalysis plugin

D

Example charts generated by the coverageAnalysis plugin
The charts that are generated by the coverageAnalysis plugin include Plot, Overlay, or Display menus that allow you to customize the data that is displayed in each chart.
Click (Search) (in the top right corner of a chart) to open the chart Viewing Options panel, where you can further customize a chart. Click (Help) to open a description of the chart.

Figure 1 Representative Depth of Coverage Chart
The Depth of Coverage Chart shows the distribution of targeted base coverage. The Xaxis represents the base read depth. The left Y-axis represents the number of reads at a given base read depth or a range (bin) of base read depths, as a percentage of the total number of base reads. The right Y-axis represents the cumulative count of the number of reads at a given read depth or greater, as a percentage of the total number of reads. The individual orange bars represent the percentage of reads in the specific range of base read depths. The blue curve measures the cumulative reads at a given base read depth or greater. If your analysis includes a regions of interest file, this chart reflects only target regions (reads that fall within a region of interest). Use the Plot dropdown list to switch between Maximum Read Depth, 99.9% of All Reads, and Normalized Coverage plots.

Figure 2 Representation Plots
Use the Display dropdown list to switch between Pass/Fail vs. Target G/C Content, Pass/Fail vs. Target Length, Representation vs. Target G/C Content, and Representation vs. Target Length plots. This figure shows an example Pass/Fail vs. Target G/C Content plot.

GenexusTM Integrated Sequencer User Guide

189

D

Appendix D coverageAnalysis plugin in GenexusTM Software Output files generated by the coverageAnalysis plugin

Output files generated by the coverageAnalysis plugin

You can download coverageAnalysis plugin results files from links that are contained in the File Links section.
Sometimes the file name can be too long to open in applications such as MicrosoftTM ExcelTM. To resolve this problem, rightclick the file and click Save As to rename the downloaded file.
Click (Help) next to the file to open a description of the file.
The following is an example of the content of a results file that is generated by the coverageAnalysis plugin.
The list of files depends on the application type selected.

File
Coverage statistics summary
Base depth of coverage

Description
A summary of the statistics presented in the tables at the top of the plugin report. The first line is the title. Each subsequent line is either blank or a particular statistic title followed by a colon (:) and its value.
Coverage summary data used to create the Depth of Coverage Chart. This file contains the following fields:
· read_depth: the depth at which a (targeted) reference base has been read. · base_cov: the number of times any base was read (covered) at this depth. · base_cum_cov: the cumulative number of reads (coverage) at this read depth or greater. · norm_read_depth: the normalized read depth (depth divided by average base read
depth). · pc_base_cum_cov: same as base_cum_cov but represented as a percentage of the
total base reads.

190

GenexusTM Integrated Sequencer User Guide

Appendix D coverageAnalysis plugin in GenexusTM Software Output files generated by the coverageAnalysis plugin

D

(continued) File
Amplicon coverage summary
Chromosome base coverage summary

Description
Coverage summary data used to create the Amplicon Coverage Chart. This file contains these fields:
· contig_id: the name of the chromosome or contig of the reference for this amplicon. · contig_srt: the start location of the amplicon target region.
This coordinate is 1-based, unlike the corresponding 0-based coordinate in the original targets BED file. · contig_end: the last base coordinate of this amplicon target region.
Note: The length of the amplicon target is given as tlen = (contig_end - contig_srt + 1). · region_id: the ID for this amplicon as given as the 4th column of the targets BED file. · gene_id: the gene symbol as given as the last field of the targets BED file. · gc_count: the number of G and C bases in the target region. %GC = 100% * gc / tlen. · overlaps: the number of times this target was overlapped by any read by at least one
base. Individual reads might overlap multiple amplicons where the amplicon regions themselves overlap. · fwd_e2e: the number of assigned forward strand reads that read from one end of the amplicon region to the other end. · rev_e2e: the number of assigned reverse strand reads that read from one end of the amplicon region to the other end. · total_reads: the total number of reads assigned to this amplicon. This value is the sum of fwd_reads and rev_reads and is the field that rows of this file are ordered by (then by contig id, srt and end). · fwd_reads: the number of forward strand reads assigned to this amplicon. · rev_reads: the number of reverse strand reads assigned to this amplicon. · cov20x: the number of bases of the amplicon target that had at least 20 reads. · cov100x: the number of bases of the amplicon target that had at least 100 reads. · cov500x: the number of bases of the amplicon target that had at least 500 reads.
Base reads per chromosome summary data used to create the default view of the Reference Coverage Chart. This file contains these fields:
· chrom: the name of the chromosome or contig of the reference. · start: the coordinate of the first base in this chromosome. This is always 1. · end: the coordinate of the last base of this chromosome. Also its length in bases. · fwd_reads: the total number of forward strand base reads for the chromosome. · rev_reads: the total number reverse strand base reads for the chromosome. · fwd_ontrg (if present): the total number of forward strand base reads that were in at least
one target region. · seq_reads: the total sequencing (whole) reads that are mapped to individual contigs.

GenexusTM Integrated Sequencer User Guide

191

D

Appendix D coverageAnalysis plugin in GenexusTM Software Output files generated by the coverageAnalysis plugin

(continued)

File

Description

Aligned reads BAM file

Contains all aligned reads that are used to generate this report, in BAM format. This is the same file that can be downloaded from the main report (for the specific barcode). See the current SAM tools documentation for more file format information.

Aligned reads BAI Binary BAM index file as required by some analysis tools and alignment viewers such as IGV.

file

This is the same file that can be downloaded from the main report (for the specific barcode).

192

GenexusTM Integrated Sequencer User Guide

E

Safety

WARNING! GENERAL SAFETY. Using this product in a manner not specified in the user documentation may result in personal injury or damage to the instrument or device. Ensure that anyone using this product has received instructions in general safety practices for laboratories and the safety information provided in this document.
· Before using an instrument or device, read and understand the safety information provided in the
user documentation provided by the manufacturer of the instrument or device.
· Before handling chemicals, read and understand all applicable Safety Data Sheets (SDSs) and use
appropriate personal protective equipment (gloves, gowns, eye protection, and so on). To obtain SDSs, see the "Documentation and Support" section in this document.
Symbols on this instrument
Symbols may be found on the instrument to warn against potential hazards or convey important safety information. In this document, the hazard symbol is used along with one of the following user attention words.
· CAUTION!--Indicates a potentially hazardous situation that, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.
· WARNING!--Indicates a potentially hazardous situation that, if not avoided, could result in death or serious injury.
· DANGER!--Indicates an imminently hazardous situation that, if not avoided, will result in death or serious injury.
Standard safety symbols
Symbol and description
CAUTION! Risk of danger. Consult the manual for further safety information.
CAUTION! Risk of electrical shock.
CAUTION! Hot surface.

GenexusTM Integrated Sequencer User Guide

193

E

Appendix E Safety Symbols on this instrument

(continued)

Symbol and description

CAUTION! Potential biohazard.

CAUTION! Ultraviolet light.
Symbole et description
MISE EN GARDE ! Risque de danger. Consulter le manuel pour d'autres renseignements de sécurité.

MISE EN GARDE ! Risque de choc électrique.

MISE EN GARDE ! Surface chaude.

MISE EN GARDE ! Danger biologique potentiel.

MISE EN GARDE ! Rayonnement ultraviolet.

Additional safety symbols
CAUTION! Moving parts.

Symbol and description

CAUTION! Piercing hazard.

Symbole et description MISE EN GARDE ! Parties mobiles.
MISE EN GARDE ! Danger de perforation.

194

GenexusTM Integrated Sequencer User Guide

Location of safety labels
Label and location

Appendix E Safety Symbols on this instrument

E

6191

A44096

6191

A44096

Rear panel

GenexusTM Integrated Sequencer User Guide

195

E

Appendix E Safety Symbols on this instrument

Control and connection symbols

Symbols

On (Power)

Descriptions

Off (Power)

Protective conductor terminal (main ground)

Alternating current

Conformity symbols

Conformity mark

Description Indicates conformity with safety requirements for Canada and U.S.A.

Indicates conformity with China RoHS requirements.

Indicates conformity with European Union requirements.
Indicates conformity with Australian standards for electromagnetic compatibility.
Indicates conformity with the WEEE Directive 2012/19/EU.
CAUTION! To minimize negative environmental impact from disposal of electronic waste, do not dispose of electronic waste in unsorted municipal waste. Follow local municipal waste ordinances for proper disposal provision and contact customer service for information about responsible disposal options.

196

GenexusTM Integrated Sequencer User Guide

Appendix E Safety Instrument safety

E

Instrument safety
General
CAUTION! Do not remove instrument protective covers. If you remove the protective instrument panels or disable interlock devices, you may be exposed to serious hazards including, but not limited to, severe electrical shock, laser exposure, crushing, or chemical exposure.
Physical injury
CAUTION! Moving and Lifting Injury. The instrument is to be moved and positioned only by the personnel or vendor specified in the applicable site preparation guide. Improper lifting can cause painful and permanent back injury. Things to consider before moving the instrument:
· Depending on the weight, moving may require two or more persons. · If you decide to move the instrument after it has been installed, do not attempt to do so without the
assistance of others, the use of appropriate moving equipment, and proper lifting techniques.
· Make sure that the path from where the object is to where it is being moved is clear of
obstructions.

CAUTION! Moving Parts. Moving parts can crush, pinch and cut. Keep hands clear of moving parts while operating the instrument. Disconnect power before servicing.

GenexusTM Integrated Sequencer User Guide

197

E Appendix E Safety Instrument safety
Electrical safety
WARNING! Ensure appropriate electrical supply. For safe operation of the instrument:
· Plug the system into a properly grounded receptacle with adequate current capacity. · Ensure the electrical supply is of suitable voltage. · Never operate the instrument with the ground disconnected. Grounding continuity is required for
safe operation of the instrument.
AVERTISSEMENT ! Veiller à utiliser une alimentation électrique appropriée. Pour garantir le fonctionnement de l'instrument en toute sécurité :
· Brancher le système sur une prise électrique correctement mise à la terre et de puissance
adéquate.
· S'assurer que la tension électrique est convenable. · Ne jamais utiliser l'instrument alors que le dispositif de mise à la terre est déconnecté. La
continuité de la mise à la terre est impérative pour le fonctionnement de l'instrument en toute sécurité.
WARNING! Power Supply Line Cords. Use properly configured and approved line cords for the power supply in your facility.
AVERTISSEMENT ! Cordons d'alimentation électrique. Utiliser des cordons d'alimentation adaptés et approuvés pour raccorder l'instrument au circuit électrique du site.
WARNING! Disconnecting Power. To fully disconnect power either detach or unplug the power cord, positioning the instrument such that the power cord is accessible.
AVERTISSEMENT ! Déconnecter l'alimentation. Pour déconnecter entièrement l'alimentation, détacher ou débrancher le cordon d'alimentation. Placer l'instrument de manière à ce que le cordon d'alimentation soit accessible.

198

GenexusTM Integrated Sequencer User Guide

E Appendix E Safety Instrument safety
Cleaning and decontamination
CAUTION! Cleaning and Decontamination. Use only the cleaning and decontamination methods that are specified in the manufacturer user documentation. It is the responsibility of the operator (or other responsible person) to ensure that the following requirements are met:
· No decontamination or cleaning agents are used that can react with parts of the equipment or with
material that is contained in the equipment. Use of such agents could cause a HAZARD condition.
· The instrument is properly decontaminated a) if hazardous material is spilled onto or into the
equipment, and/or b) before the instrument is serviced at your facility or is sent for repair, maintenance, trade-in, disposal, or termination of a loan. Request decontamination forms from customer service.
· Before using any cleaning or decontamination methods (except methods that are recommended by
the manufacturer), confirm with the manufacturer that the proposed method will not damage the equipment.
MISE EN GARDE ! Nettoyage et décontamination. Utiliser uniquement les méthodes de nettoyage et de décontamination indiquées dans la documentation du fabricant destinée aux utilisateurs. L'opérateur (ou toute autre personne responsable) est tenu d'assurer le respect des exigences suivantes:
· Ne pas utiliser d'agents de nettoyage ou de décontamination susceptibles de réagir avec certaines
parties de l'appareil ou avec les matières qu'il contient et de constituer, de ce fait, un DANGER.
· L'instrument doit être correctement décontaminé a) si des substances dangereuses sont
renversées sur ou à l'intérieur de l'équipement, et/ou b) avant de le faire réviser sur site ou de l'envoyer à des fins de réparation, de maintenance, de revente, d'élimination ou à l'expiration d'une période de prêt (des informations sur les formes de décontamination peuvent être demandées auprès du Service clientèle).
· Avant d'utiliser une méthode de nettoyage ou de décontamination (autre que celles
recommandées par le fabricant), les utilisateurs doivent vérifier auprès de celui-ci qu'elle ne risque pas d'endommager l'appareil.
Instrument component and accessory disposal
To minimize negative environmental impact from disposal of electronic waste, do not dispose of electronic waste in unsorted municipal waste. Follow local municipal waste ordinances for proper disposal provision and contact customer service for information about responsible disposal options.

GenexusTM Integrated Sequencer User Guide

199

E

Appendix E Safety Safety and electromagnetic compatibility (EMC) standards

Safety and electromagnetic compatibility (EMC) standards

The instrument design and manufacture complies with the following standards and requirements for safety and electromagnetic compatibility.
Safety standards

Reference
EU Directive 2014/35/EU IEC 61010-1 EN 61010-1 UL 61010-1 CAN/CSA C22.2 No. 61010-1 IEC 61010-2-010 EN 61010-2-010
IEC 61010-2-020 EN 61010-2-020 IEC 61010-2-081 EN 61010-2-081
IEC 61010-2-101 EN 61010-2-101

Description European Union "Low Voltage Directive" Safety requirements for electrical equipment for measurement, control, and laboratory use ­ Part 1: General requirements
Safety requirements for electrical equipment for measurement, control and laboratory use ­ Part 2-010: Particular requirements for laboratory equipment for the heating of materials Safety requirements for electrical equipment for measurement, control and laboratory use ­ Part 2-020: Particular requirements for laboratory centrifuges
Safety requirements for electrical equipment for measurement, control and laboratory use ­ Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment

EMC standards

Reference EU Directive 2014/30/EU EN 61326-1 IEC 61326-1 EN 61326-2-6 IEC 61326-2-6 FCC Part 18 (47 CFR) AS/NZS CISPR 11

Description
European Union "EMC Directive"
Electrical Equipment for Measurement, Control and Laboratory Use ­ EMC Requirements ­ Part 1: General Requirements
Electrical equipment for measurement, control and laboratory use. EMC requirements. Particular requirements. In vitro diagnostic (IVD) medical equipment
U.S. Standard "Industrial, Scientific, and Medical Equipment"
Limits and Methods of Measurement of Electromagnetic Disturbance Characteristics of Industrial, Scientific, and Medical (ISM) Radiofrequency Equipment

200

GenexusTM Integrated Sequencer User Guide

Appendix E Safety Safety and electromagnetic compatibility (EMC) standards

E

(continued)

Reference

Description

ICES-001, Issue 4

Industrial, Scientific and Medical (ISM) Radio Frequency Generators

FCC Part 15 Subpart B (47 CFR)

U.S. Standard Radio Frequency Devices
This equipment has been designed and tested to CISPR 11 Class A. In a domestic environment it may cause radio interference, in which case, you may need to take measures to mitigate the interference.

Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g. unshielded intentional RF sources), as these can interfere with proper operation.

Environmental design standards

Reference Directive 2012/19/EU Directive 2011/65/EU
SJ/T 11364-2014

Description
European Union "WEEE Directive"--Waste electrical and electronic equipment
European Union "RoHS Directive"--Restriction of hazardous substances in electrical and electronic equipment
"China RoHS" Standard--Marking for the Restricted Use of Hazardous Substances in Electronic and Electrical Products For instrument specific certificates, visit our customer resource page at www.thermofisher.com/us/en/home/technical-resources/rohs-certificates.html.

Radio compliance standards

Reference Directive 2014/53/EU RFID

Description
European Union "RE Directive"--Radio equipment
FCC Notice (for U.S. Customers): This device complies with Part 15 of the FCC Rules: Operation is subject to the following conditions: 1. This device may not cause harmful interference, and 2. This device must accept any interference received, Including interference that may cause undesired operation. Changes and modifications not expressly approved by Thermo Fisher Scientific can void your authority to operate this equipment under Federal Communications Commissions rules.

GenexusTM Integrated Sequencer User Guide

201

E

Appendix E Safety Safety and electromagnetic compatibility (EMC) standards

(continued) Reference
RFID
RFID

Description
Canada (English):
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions:
(1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Canada (Français):
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes :
(1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage adioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.

202

GenexusTM Integrated Sequencer User Guide

Chemical safety

Appendix E Safety Chemical safety

E

WARNING! GENERAL CHEMICAL HANDLING. To minimize hazards, ensure laboratory personnel read and practice the general safety guidelines for chemical usage, storage, and waste provided below. Consult the relevant SDS for specific precautions and instructions:
· Read and understand the Safety Data Sheets (SDSs) provided by the chemical manufacturer
before you store, handle, or work with any chemicals or hazardous materials. To obtain SDSs, see the "Documentation and Support" section in this document.
· Minimize contact with chemicals. Wear appropriate personal protective equipment when handling
chemicals (for example, safety glasses, gloves, or protective clothing).
· Minimize the inhalation of chemicals. Do not leave chemical containers open. Use only with
sufficient ventilation (for example, fume hood).
· Check regularly for chemical leaks or spills. If a leak or spill occurs, follow the manufacturer
cleanup procedures as recommended in the SDS.
· Handle chemical wastes in a fume hood. · Ensure use of primary and secondary waste containers. (A primary waste container holds the
immediate waste. A secondary container contains spills or leaks from the primary container. Both containers must be compatible with the waste material and meet federal, state, and local requirements for container storage.)
· After emptying a waste container, seal it with the cap provided. · Characterize (by analysis if needed) the waste generated by the particular applications, reagents,
and substrates used in your laboratory.
· Ensure that the waste is stored, transferred, transported, and disposed of according to all local,
state/provincial, and/or national regulations.
· IMPORTANT! Radioactive or biohazardous materials may require special handling, and disposal
limitations may apply.

AVERTISSEMENT ! PRÉCAUTIONS GÉNÉRALES EN CAS DE MANIPULATION DE PRODUITS CHIMIQUES. Pour minimiser les risques, veiller à ce que le personnel du laboratoire lise attentivement et mette en oeuvre les consignes de sécurité générales relatives à l'utilisation et au stockage des produits chimiques et à la gestion des déchets qui en découlent, décrites ci-dessous. Consulter également la FDS appropriée pour connaître les précautions et instructions particulières à respecter :
· Lire et comprendre les fiches de données de sécurité (FDS) fournies par le fabricant avant de
stocker, de manipuler ou d'utiliser les matériaux dangereux ou les produits chimiques. Pour obtenir les FDS, se reporter à la section « Documentation et support » du présent document.
· Limiter les contacts avec les produits chimiques. Porter des équipements de protection appropriés
lors de la manipulation des produits chimiques (par exemple : lunettes de sûreté, gants ou vêtements de protection).
· Limiter l'inhalation des produits chimiques. Ne pas laisser les récipients de produits chimiques
ouverts. Ils ne doivent être utilisés qu'avec une ventilation adéquate (par exemple, sorbonne).
· Vérifier régulièrement l'absence de fuite ou d'écoulement des produits chimiques. En cas de fuite
ou d'écoulement d'un produit, respecter les directives de nettoyage du fabricant recommandées dans la FDS.
· Manipuler les déchets chimiques dans une sorbonne.

GenexusTM Integrated Sequencer User Guide

203

E

Appendix E Safety Chemical safety

· Veiller à utiliser des récipients à déchets primaire et secondaire. (Le récipient primaire contient
les déchets immédiats, le récipient secondaire contient les fuites et les écoulements du récipient primaire. Les deux récipients doivent être compatibles avec les matériaux mis au rebut et conformes aux exigences locales, nationales et communautaires en matière de confinement des récipients.)
· Une fois le récipient à déchets vidé, il doit être refermé hermétiquement avec le couvercle fourni.
· Caractériser (par une analyse si nécessaire) les déchets générés par les applications, les réactifs et
les substrats particuliers utilisés dans le laboratoire.
· Vérifier que les déchets sont convenablement stockés, transférés, transportés et éliminés en
respectant toutes les réglementations locales, nationales et/ou communautaires en vigueur.
· IMPORTANT ! Les matériaux représentant un danger biologique ou radioactif exigent parfois une
manipulation spéciale, et des limitations peuvent s'appliquer à leur élimination.

WARNING! HAZARDOUS WASTE (from instruments). Waste produced by the instrument is potentially hazardous. Follow the guidelines noted in the preceding General Chemical Handling warning.

WARNING! 4L Reagent and Waste Bottle Safety. Four-liter reagent and waste bottles can crack and leak. Each 4-liter bottle should be secured in a low-density polyethylene safety container with the cover fastened and the handles locked in the upright position.

204

GenexusTM Integrated Sequencer User Guide

Biological hazard safety

Appendix E Safety Biological hazard safety

E

WARNING! Potential Biohazard. Depending on the samples used on this instrument, the surface may be considered a biohazard. Use appropriate decontamination methods when working with biohazards.

WARNING! BIOHAZARD. Biological samples such as tissues, body fluids, infectious agents, and blood of humans and other animals have the potential to transmit infectious diseases. Conduct all work in properly equipped facilities with the appropriate safety equipment (for example, physical containment devices). Safety equipment can also include items for personal protection, such as gloves, coats, gowns, shoe covers, boots, respirators, face shields, safety glasses, or goggles. Individuals should be trained according to applicable regulatory and company/ institution requirements before working with potentially biohazardous materials. Follow all applicable local, state/provincial, and/or national regulations. The following references provide general guidelines when handling biological samples in laboratory environment.
· U.S. Department of Health and Human Services, Biosafety in Microbiological and Biomedical
Laboratories (BMBL), 5th Edition, HHS Publication No. (CDC) 21-1112, Revised December 2009; found at:
https://www.cdc.gov/labs/pdf/CDC-BiosafetymicrobiologicalBiomedicalLaboratories-2009P.pdf
· World Health Organization, Laboratory Biosafety Manual, 3rd Edition,
WHO/CDS/CSR/LYO/2004.11; found at:
www.who.int/csr/resources/publications/biosafety/Biosafety7.pdf

GenexusTM Integrated Sequencer User Guide

205

F

Documentation and support

Related documentation
Document GenexusTM Software 6.2 User Guide GenexusTM Integrated Sequencer Quick Reference OncomineTM Precision Assay GX User Guide OncomineTM Comprehensive Assay v3 GX User Guide OncomineTM TCR BetaLR Assay GX User Guide GenexusTM Integrated Sequencer Site Preparation Guide GenexusTM Integrated Sequencer IT Checklist

Publication number MAN0018955 MAN0017912 MAN0018508 MAN0018512 MAN0018513 MAN0017918 MAN0018466

Customer and technical support
Visit thermofisher.com/support for the latest service and support information. · Worldwide contact telephone numbers · Product support information ­ Product FAQs ­ Software, patches, and updates ­ Training for many applications and instruments · Order and web support · Product documentation ­ User guides, manuals, and protocols ­ Certificates of Analysis ­ Safety Data Sheets (SDSs; also known as MSDSs)
Note: For SDSs for reagents and chemicals from other manufacturers, contact the manufacturer.

206

GenexusTM Integrated Sequencer User Guide

Appendix F Documentation and support Limited product warranty

F

Limited product warranty
Life Technologies Corporation and/or its affiliate(s) warrant their products as set forth in the Life Technologies' General Terms and Conditions of Sale at www.thermofisher.com/us/en/home/ global/terms-and-conditions.html. If you have any questions, please contact Life Technologies at www.thermofisher.com/support.

GenexusTM Integrated Sequencer User Guide

207

Genexus Integrated Sequencer_UG_MAN0017910-v14-GUID-462E12B2-26FB-474CBC86-15CF9B0ED2F7-2020/11/20 21:26:03 en 17:28:07.905Z thermofisher.com/support | thermofisher.com/askaquestion thermofisher.com
25 November 2020

