Manually set the COM Port number. Set COM port. Step3: Start to Download. 3-1. To download the patient data from Apex's CPAP device, click on "Download" ...

Easy Compliance Software | Instructions for Use

* * To ensure proper use of the software, please read all instructions prior to ... Following the instructions will guide you to complete the installation of the Easy Compliance...

Instructions for Use

instructions will guide you to complete the installation of the Easy Compliance Software. 3 In the “Destination folder” box, you can type in a directory path or press the “Browse” button to establish a folder and path on your hard disc that will contain the software. The default path is

PDF easy-compliance-lite-software-help-gb
Multilingual version 3.4
Lite version is intended for patient use only; Some physician features are disabled.
Instructions for Use
Table of Contents
1 Introduction ....................................................................................... 1 2 Installing and Starting Software............................................................ 1 3 Using the Software ............................................................................. 5 3.1 New ................................................................................................. 5 3.2 Close ................................................................................................ 6 3.3 Save ................................................................................................. 7 3.4 Save As ............................................................................................. 8 3.5 Save Reports as PDF ........................................................................... 8 3.6 Open ................................................................................................ 9 3.7 Print ............................................................................................... 10 3.8 Print Setup ...................................................................................... 14 3.9 Exit................................................................................................. 14 3.10 Edit Patient Profile ........................................................................... 15 3.11 Edit Clinical Details........................................................................... 15 3.12 Edit Patient Notes ............................................................................ 16 3.13 Edit Phone-in Data ........................................................................... 16 3.14 Using a miniSD or SD Card ................................................................ 18 3.15 Download via USB cable ................................................................... 19 3.16 Help ............................................................................................... 23 3.17 Patient Profile Tab & Clinical Details Tab............................................. 24 3.18 Patient Notes Tab............................................................................. 25 3.19 Total Time Tab ................................................................................. 25 3.20 Compliance Tab ............................................................................... 26 3.21 Long Term Data ............................................................................... 28

1 Introduction
The Easy Compliance Lite Software is compatible with Apex's CPAP System which is intended for the treatment of patients with Obstructive Sleep Apnea (OSA). The Easy Compliance Lite Software is used with Windows® based PCs and allows you to download usage data collected by the Apex's CPAP device after treatment. The memory capacity in the Apex's CPAP system can hold up to 730 sessions of summary data. In order to meet some FDA regulations, some features (mark with *) have been disabled in this version
*
*
To ensure proper use of the software, please read all instructions prior to downloading data from the Apex's CPAP.
2 Installing and Starting Software
 Preparations 1. Make sure your PC is running on Windows® based system. It will not run on Mac systems. 2. There are two steps to the software installation: i. Download the Easy Compliance software from website. ii. Install the USB drivers for the Apex's CPAP.
 System requirements Your computer must meet the following requirements: y Available Windows® supported USB port (USB 2.0) y Pentium-class processor; Pentium III 800-megaherz (MHz) or higher is recommended. y Microsoft Windows® XP SP2, Windows® Vista SP1, or Windows® 7 (32 bit). y 128 MB RAM. y CD-ROM drive compatible with multimedia desktop computer specifications.
1

y 1024 x 768 or higher resolution VGA monitor. y Minimum 100MB hard disk space for software and data. y Internet Explorer 7 or greater  Steps to Install Easy Compliance software 1. Close all applications before installation. 2. Find the installation program and double-click it.
¾ Install USB-to-Serial: PL2303 Driver Note: Once this step is executed, PC will display the PL-2303 USB-to-Serial InstallShield WiZARD. It will guide you to complete the installation of the PL-2303 driver. If this driver has already been installed in your PC, just press "Cancel" button on the InstallShield WiZARD to skip this step.
¾ Install PDF creation: CutePDF Writer ¾ Install Easy Compliance Software
Note: Once this step is executed, PC will display the Setup WiZARD. Following the instructions will guide you to complete the installation of the Easy Compliance Software.
2

In the "Destination folder" box, you can type in a directory path or press the "Browse" button to establish a folder and path on your hard disc that will contain the software. The default path is "C:\Program Files\Apex Medical\Easy Compliance 3.4 Lite\". Press the "Next" button to complete the installation. Once the installation process is complete, click "Close" button to exit the program.

 Install Software

The software icon the desktop.

will appear in the folder where you installed the software and also on

 Installing the USB drivers for Apex's CPAP
The USB driver should be installed automatically after the setup program is completed. For the first time to connect the device to PC with the USB cable, the message box below may appear. Found New Hardware WiZARD starts when Windows finish searching for the new hardware. Windows will ask you if it can connect to Windows Update to search for software. Select "Yes, this time only" then click "Next".

When the window below appears, select "Install the software automatically {Recommended}" then click "Next".
3

The Found New Hardware WiZARD will start to search the hardware and show window below.
During the search process, there may appear the warning message window below, click "Continue Anyway" to continue installation. The driver installation starts.
Click "Finish" to complete the installation. Now you can download the data via USB Cable.
Note: Some systems may need to restart the computer to complete the installation of the driver.
4

 Starting the software
To use the software, double-click on the software icon software functions can be accessed.

. A new file will be displayed, and all

3 Using the Software
3.1 New
If you would like to create a file for a new patient, click the "New" Toolbar button or click "File" on the Menu Bar and then select the "New" item from the drop-down menu. Note that when you first start the software, the screen will already have a blank patient profile available (no need to press the "New" button).

File - > New
Note: Before downloading new user data, it is necessary to have a patient file open first (create a new file, open an existing file or have a blank profile on the screen). Otherwise, the download function will fail.

New File 5

When you open a new file, the main window with empty patient record will pop up. The functions of the various parts of the main window are described below: Patient Record There are seven tabs in patient record to present the data of patient profile, clinical details, patient notes, total usage time, compliance usage time, detail and long term report. Menus From the menus user can perform several functions to load/save file, edit patient record, download data from CPAP device, and see help file. Toolbar Users also can quickly perform the functions of menu by clicking 11 icons.
Icon meaning
3.2 Close
If you would like to close the current patient file, click "File" on the Menu Bar and select the "Close" item.
File - > Close
Once you execute this function, the current window will be closed and you will be asked to confirm whether you would like to save the current data or not. It is recommended that you save data before closing the patient file.
6

Confirmation during the Close function
After closing the patient file, the main window will be displayed.
Main Window
3.3 Save
If you would like to save the current data to the hard disk, click "File" from the Menu Bar and then select the "Save" item. You can also press the "Save" Toolbar button.
File - > Save 7

3.4 Save As
Click "File" from the Menu Bar and then click "Save As" to the save data to another file.
File - > Save As
To save the current data to a different filename, type a new name into the "File name" box and click the "Save" button.
"Save As" window
3.5 Save Reports as PDF
If you would like to save usage data reports as an electronic file, click on "File" from the Menu Bar and then "Save Reports as PDF". All the usage data during the selected period will be converted into a PDF file.
File - > Save Reports as PDF 8

3.6 Open
If you would like to open an existing file, click "File" from the Menu Bar and then select the "Open" item. You can also press the "Open" Toolbar button.
File - > Open
1. When you select the "Open" item, the Open dialog box will appear. 2. Search for your file (with .paf extension) in the dialogue box and select it by double clicking
on it. 3. By default, the Main Window will display the Patient Profile screen for the selected patient
data.
9

3.7 Print
If you would like to print the current window, click on "File" from the Menu Bar and then "Print". You can also press the "Print" Toolbar button.
File - > Print
When you select the "Print" function, a dialog box appears allowing you to select the reports to be printed. You can choose to preview the content before printing or to convert the report to a PDF file.
Report Type
10

Summary Report 11

Compliance Time Report, with limited 90 days of data 12

Long Term Trend Report, with limited 90 days of data shown in the pressure graph 13

3.8 Print Setup
If you would like to adjust the printing options, go to "File" (Menu Bar) -> "Print Setup". You can adjust paper size, printer, landscape etc.
File - > Print Setup
3.9 Exit
If you would like to exit the Easy Compliance software, go to "File" (on Menu Bar) -> Exit.
File - > Exit
When you exit the software, you will be asked to confirm whether you would like to save the current data or not. It is recommended you save all data before exiting.
Confirmation window while exiting the software
14

3.10 Edit Patient Profile
If you would like to edit the patient profile data, click on "Edit" from the Toolbar buttons and then select the "Patient Profile" tab towards the top of the "Patient Information" window. A patient profile edit window will then display. Simply fill in or update the appropriate information and click the "OK" button.
Edit - > Patient Profile
Edit Patient Profile
3.11 Edit Clinical Details
If you would like to edit the clinical data for the patient, click on "Edit" from the Toolbar buttons and then select the "Clinical Data" tab towards the top of the window. Simply fill in or update the appropriate information and click the "OK" button.
Edit -> Clinical Details
Edit Clinical Details 15

3.12 Edit Patient Notes
You can also add or edit the Patients Notes. When the Edit Patient Notes screen appears, click on any of the available 20 lines that you want to enter data for. The maximum number of Patient Notes is 20.
Edit Patient Notes
3.13 Edit Phone-in Data
If you would like to edit the phone-in data, click on "Edit" from the Toolbar buttons and then select the "Phone-in Data" tab towards the top of the window. A phone-in data edit window will then display. Simply fill in or update the appropriate information and click the "Print" button.
Edit -> Phone-in Data
Edit Phone-in Data 16

Summary Report of Phone-in Data 17

3.14 Using a miniSD or SD Card
The patient can use the miniSD or SD card to backup the most recent compliance data in the Apex's CPAP system (for some models). Once the data is backed up, the patient will send the miniSD or SD card to you and the compliance data can be retrieved and analyzed using the Easy Compliance software.
Insert the miniSD or SD card to the card reader connected to your PC. Click the "Card" icon to open the files saved in the miniSD or SD card. The files of type is "CPAP SDCard Files (*.cpe) or APAP SDCard Files (*.ape).
18

3.15 Download via USB cable
This section describes how to download data from the Apex's CPAP to your PC. To download data, there are three steps to follow ­
Step1. Connect Apex's CPAP device to your computer with USB cable. Step2. Start the software and establish an internal connection (between the CPAP
system and PC). Step3. Download data. Each step is described in detail below.

Step1: Connect Apex's CPAP device to your computer with USB cable.
1-1. Plug the power cord into the power socket of the Apex's CPAP device. Plug the other end of the cord into a wall outlet. Keep the CPAP system in Standby mode (no airflow). Download will fail if the device is providing airflow.
1-2. Connect the Apex's CPAP device to your computer using the USB cable supplied in package. The Mini-USB end (the smaller end) is for the CPAP system and the standard USB end (the larger end) is for the computer.

1-3. Start the software on the PC by double-clicking on the icon directory.

in the software

Step2: Start the software and establish an internal connection (between CPAP system and PC).
Aside from the physical cable connections, it is necessary to establish another internal connection between the two devices, which is done through the PC's Serial Ports (not to be mistaken for the physical Serial connectors).
In most cases, you can have the software automatically establish these connections as follows: 2-1. Click on the "Download" item in the Menu Bar and select the "COM port Auto
Connect" item.
2-2. The Software will automatically detect an appropriate connection for you.
2-3. If the connection was established successfully, the message "Auto Connect Successful" will appear. Otherwise, the message "Auto Connect Fail" will appear.

Download / COM Port Auto Connect 19

Auto Connect Successfully
Auto Connect Failed
-- PC fails to connect automatically ­ If the internal connection to the CPAP system fails, check the following: ¾ Check if the power cord is securely attached to the Apex's CPAP device, and the CPAP
device is in standby mode ¾ Check if the connecting cable is securely attached to the CPAP system and the computer. ¾ Check the USB driver on the PC. Go to the Device Manager in your system to determine
the appropriate driver installation steps. Open the Device Manager (located in "Control Panel\System" then select the "Hardware" tab and click "Device Manger..."), and select "View > Devices by Type" to make sure the device you install is correctly displayed. The device appears as an additional COM port with the label "USB Serial Port". If you don't find an additional "USB Serial Port", try to reinstall the USB driver.
Manually set Connect setting You can also manually set the connection setting but you must know the correct COM port number which Apex's CPAP device uses. Check Device Manager in your system to find the correct COM port number. Select "View > Devices by Type," and under "Ports (COM & LPT)," look for a USB Device with "Prolific USB Serial Port (COMn)." Note that n is the number of the COM port assigned to the Apex's CPAP device.
2-1. Go into "Control Panel">System and click the Device Manager button on the Hardware tab
2-2. Select the menu "COM & LPT," and find the COM port of "Prolific USB Serial Port (COMn)". Right mouse click (or double click) for the "Properties"
2-3. Select the Port Settings Tab. Click the Advanced button. 2-4. Manually set the COM Port number. 2-5. Go back to the menu bar of Easy Compliance and select correct COM port, and click
"OK" button to confirm.
20

Click "Device Manager"
Select "COM & LPT" and find "Prolific USB Serial Port". Right mouse click (or double click) for the "Properties".
Select the Port Settings Tab. Click the Advanced button 21

Manually set the COM Port number
Set COM port
Step3: Start to Download 3-1. To download the patient data from Apex's CPAP device, click on "Download" from the
Menu Bar and select the "Extract Data" function; the user's therapy data will be copied to your computer's memory (you can also click the "Download" Toolbar button). Note that after a download, the data on the Apex's CPAP system will not be removed. Data in Apex's CPAP system can only be erased through the control panel on the device. While downloading data, a progress indicator displays the percentage of data retrieved.
Download progress indicator
Download complete indicator 22

Once the data retrieval process has been completed, click "OK" and the Patient Profile screen will appear with the new information. Software fails to download the data If the connection to the CPAP Device fails, you will be informed by an error message. Please check the following: ¾ Check if the Apex's CPAP device is powered on and in standby mode. ¾ Check if the USB cable is secured at both ends. ¾ Check if the connection between CPAP device and the computer is established by
following Step 2 described above. To stop the download, simply press the Close button in the top right corner of the progress window. Important: Every time you download data to an existing patient file, all existing usage data for that patient will be erased. So, if it is important to retain existing data, remember to save the existing data to another file before downloading new data. The other option is to ALWAYS save the data to a new file after downloading.
3.16 Help
To obtain help for the software, click on the "Help" Tool Bar button. The Help documentation will appear in PDF format.
23

3.17 Patient Profile Tab & Clinical Details Tab
General patient information can be viewed from two of the Information Tabs at the bottom of the main menu. This includes the "Patient Profile" and "Clinical Details" tab. Both tabs have an "Edit" Button in the lower right corner that you can press to access an "Edit" screen where you can modify/add information.
Patient Profile View
Clinical Details View
24

3.18 Patient Notes Tab
The Patients Notes tab is a simple way to attach your own notes to the current patient file. To view and edit the patient notes, go to "Edit" (Menu Bar) -> "Patient Notes". When the "Patient Notes" window appears, you can click on any of the line items where you can add or edit your notes. The maximum number of notes is 20.
Patient Notes
3.19 Total Time Tab
Total Time Tab displays a general overview of the device operating records for each day. The Total Time data has two sections. The left panel is the cumulated operating time for the device which is defined as the time airflow is flowing through the tube (whether or not the patient has donned the mask). The right panel shows the times during the day that the device was operating.
Total Time Tab
25

View of Usage Profile Range (right panel)
The Usage Profile Range (right panel) displays the time of day during which the device is running. A 24-hour period is shown as a horizontal bar, and starts at 12pm (noon) in order to capture one night on a continuous single line (since the devices is typically used at night).
View of the cumulated daily device operating time
3.20 Compliance Tab
Compliance Tab displays an overview of the compliance records for each day. The Compliance tab is composed of two sections; the left panel is the cumulated daily compliance time recorded for each day; the right panel is the times during the day that the patient is receiving treatment.
Compliance time is the time that the device is delivering airflow AND the patient is wearing the mask.
Compliance Tab 26

View of compliance usage profile range
The Compliance Usage Profile Range displays the times during the day where the patient is receiving treatment. A 24-hour period is shown as a horizontal bar and begins at 12pm in order to capture one night on a continuous single line.
Note: In order to leave the patient enough time to settle into a relaxed state, the first 15 minutes of each session is not recorded by the Apex's CPAP device.
Note: The compliance time and pressure data will be captured every 15-minute in the Apex's CPAP system and user may turn off the device by pressing "START/STANDBY" button on the device. If a usage session is ended up by low pressure auto-off function instead of normally turning off manually, the compliance record shorter than 15-minute session will NOT be shown on the compliance view even though it is stored in the device memory. Therefore, there might be a few minutes difference in the meter record between software and the device LCD.
There might be a slight discrepancy between the information on the Compliance tab and the data on the Apex's CPAP device. If, at the end of a usage session, there was a drop in pressure in the mask (mask slips off), the last few minutes of a 15-minute session may NOT show on the Compliance tab (though it is recorded on the CPAP device).
View of cumulated daily compliance time
27

3.21 Long Term Data
The Long Term data shows various statistical information regarding patient data. This tab is separated to two sections; the left panel is the daily pressure graph and right panel is the analysis data. The analysis data is composed of three pages which are Pressure, Compliance and Events.
Time Edit To change the parameters for the Long Term data, you can click "Time Edit" button at the top-right of Long Term tab. And then an "Analysis Time Edit" window will display. Here, you can configure the select period and minimal compliance. ¾ Selected period: The time period will be covered in the long term data. ¾ Minimal Compliance: This is used to calculate the percentage of days that the user
reached the "Minimum" hours of treatment
Time Edit.
Edit Analysis settings
Pressure This page provides the information about average pressure & leakage. ¾ Used Pressure y Average Pressure: The average pressure over the time period. y Average 90th Percentile Pressure: The pressure that user is stayed equal or lower of it for
90% of the time. y Average Leak: The average leak for the time period. y Lowest Leak: The average lowest leak for the time period. y Highest Leak: The average highest leak for the time period. y Average Large Leak Time: The average time of large leak for any one day. ¾ Pressure Graph: Pressure graph is a histogram that displays the average pressure levels
delivered on any particular a day.
28

Long Term Data ­ Pressure
Compliance This page provides the information about compliance. ¾ Used Days y Number of Days: The total number of days the user used (Utilization) and did not use
(No Utilization) the device. y Average of Days/Week: The average of days per week the user used (Utilization) and did
not use (No Utilization) the device. y % Average/Period: The percentage of days with device usage and without device usage. ¾ Util Time (Utilization Time) y Average Compliance: The average daily compliance time per day. y Min. Compliance: The least amount compliance time for any one day. y Max. Compliance: The most amount of compliance time for any one day. ¾ Analysis
This section provides the information about the days that reach a minimum number of compliance hours. The minimum compliance hour is defined in Long Term Æ Time Edit Æ Minimal Compliance y Selected Period >= x h/d: The number of days that reached a minimum number of compliance hours. y % of Utilization Days days with usage >= x h/d: The percentage of days that reached a minimum number of compliance hours.
Long Term Data ­ Compliance
29

Events This page provides the information about total number or index (average events/hours) of detected events.
y AHI: The total number or index of apnea plus hypopnea. y Apnea: The total number or index of apnea. y Hypopnea: The total number or index of Hypopnea. y Vibratory Snoring: The total number or index of snoring.
Long Term Data ­ Events
30

