Hydraulic service kit Model CPG-KITH - Instrumentation2000 ...

WIKA -Cal software. The collected data from the data logger of the digital pressure gauge can be transmitted wirelessly to the. WIKA-Cal calibration software ...

mensor-cpg-kith-data-sheet
Hydraulic service kit Model CPG-KITH

Calibration technology
WIKA data sheet CT 93.02

Applications
 Oil and gas industry  Maintenance and service facilities  Calibration service companies and service industry  Simple calibrations directly on site
Special features
 Testing and adjustment of pressure gauges, pressure sensors, pressure switches or safety valves
 Reference instrument model CPG1500 with 0.05 % FS  Hydraulic version up to 700 bar (10,000 psi)  Clear and simple operation

Hydraulic service kit, model CPG-KITH

Description
Application range The CPG-KITH hydraulic service kit is the ideal tool for the hydraulic testing, adjustment and calibration of mechanical and electronic pressure measuring instruments through comparative measurements or for the setting of pressure switches. These and other tests can take place quickly and reliably in the laboratory or workshop, or on site at the measuring point. With the service kit, service and maintenance are more complete, safer and more flexible. The kit is high-quality finish, handy and robust and is also suitable for service use.
Components The service kit consists of a hydraulic hand test pump for the generation of a defined test pressure and also a highly accurate model CPG1500 precision digital pressure gauge as a reference instrument. Air is used as the pressure medium. Various connection adapters are also available in the test case. All components are carefully stowed in the case and protected for in-house transport.

Simple function If one connects the test item and the model CPG1500 precision digital pressure gauge to the test pump, on actuating the pump, the same pressure will act on both instruments. Using the fine adjustment valve, the test pressure is set exactly. By comparison of the two measured values at any given pressure value, a check of the accuracy and/or adjustment of the pressure measuring instrument under test can be carried out.
WIKA-Cal software The collected data from the data logger of the digital pressure gauge can be transmitted wirelessly to the WIKA-Cal calibration software for further evaluation. WIKA-Cal also offers, over and above PC-supported calibration, the management of the calibration and instrument data in an SQL database. A WIKA-Wireless interface is available for the data transfer.

WIKA data sheet CT 93.02  08/2017
Data sheets showing similar products: Precision digital pressure gauge; model CPG1500; see data sheet CT 10.51 Hydraulic hand test pump; models CPP700-H and CPP1000-H; see data sheet CT 91.07 Pneumatic service kit; model CPG-KITP; see data sheet CT 93.01

Page 1 of 6

Specifications
Specifications
CPG1500 Pressure range Pressure connection Wetted parts
CPP700-H Pressure range Pressure transmission medium Pressure connection
Fine adjustment Fluid reservoir Material Sealings Dimensions (L x W x H) Weight Standard accessories
1) Other pressure transmission media on request. 2) Liquids are not included in the scope of delivery.

Metric version CPG-KITH-EUxx-Z

Imperial version CPG-KITH-USxx-Z

See ordering information G ½ male thread Stainless steel

¼ NPT male thread

0 ... 700 bar

0 ... 10,000 psi

Hydraulic fluid based on mineral oil or single distilled water 1) 2)

G ½ female thread for reference instrument, freely rotating, hose with G ¼ female thread for test item

¼ NPT female thread via adapter for reference instrument, hose with ¼ NPT via adapter for test item

Fine adjustment valve/volume adjustment

200 cm³ (12.2 in³)

anodised aluminium, brass, stainless steel, ABS

FKM and NBR (standard)

280 x 170 x 120 mm (11.02 x 6.69 x 4.72 in)

1.9 kg (4.19 lbs.)

Minimess® test item connection hose; length 1 m (3.3 ft)

Certificates

Certificate Calibration
Recommended recalibration interval

Standard: 3.1 calibration certificate per EN 10204 Option: DKD/DAkkS calibration certificate
1 year (dependent on conditions of use)

Approvals and certificates, see website

Ordering information
CPG-KITH-EU52-5-ZZ CPG-KITH-EU53-5-ZZ CPG-KITH-EU54-5-ZZ CPG-KITH-EU55-5-ZZ CPG-KITH-EU51-5-ZZ CPG-KITH-EU50-5-ZZ CPG-KITH-EU52-1-ZZ CPG-KITH-EU53-1-ZZ CPG-KITH-EU54-1-ZZ CPG-KITH-EU55-1-ZZ CPG-KITH-EU51-1-ZZ CPG-KITH-EU50-1-ZZ
3) FS = Full Span

Metric version Measuring range 0 ... 60 bar 0 ... 100 bar 0 ... 160 bar 0 ... 250 bar 0 ... 400 bar 0 ... 700 bar 0 ... 60 bar 0 ... 100 bar 0 ... 160 bar 0 ... 250 bar 0 ... 400 bar 0 ... 700 bar

Accuracy 0,05 % FS 3) 0,05 % FS 3) 0,05 % FS 3) 0,05 % FS 3) 0,05 % FS 3) 0,05 % FS 3) 0.1 % FS 3) 0.1 % FS 3) 0.1 % FS 3) 0.1 % FS 3) 0.1 % FS 3) 0.1 % FS 3)

Page 2 of 6

WIKA data sheet CT 93.02  08/2017

Ordering information
CPG-KITH-US52-5-ZZ CPG-KITH-US53-5-ZZ CPG-KITH-US54-5-ZZ CPG-KITH-US55-5-ZZ CPG-KITH-US51-5-ZZ CPG-KITH-US50-5-ZZ CPG-KITH-US52-1-ZZ CPG-KITH-US53-1-ZZ CPG-KITH-US54-1-ZZ CPG-KITH-US55-1-ZZ CPG-KITH-US51-1-ZZ CPG-KITH-US50-1-ZZ
3) FS = Full Span
Ordering information CPG-A-15-ZZ2-Z WIKA-Cal CPG-A-15-ZZB-Z CPH-A-15-ZZF-Z CPG-A-15-ZZD-Z CPG-A-15-ZZC-Z

Imperial version Measuring range 0 ... 700 psi 0 ... 1,500 psi 0 ... 2,000 psi 0 ... 3,000 psi 0 ... 5,000 psi 0 ... 10,000 psi 0 ... 700 psi 0 ... 1,500 psi 0 ... 2,000 psi 0 ... 3,000 psi 0 ... 5,000 psi 0 ... 10,000 psi

Accuracy 0,05 % FS 3) 0,05 % FS 3) 0,05 % FS 3) 0,05 % FS 3) 0,05 % FS 3) 0,05 % FS 3) 0.1 % FS 3) 0.1 % FS 3) 0.1 % FS 3) 0.1 % FS 3) 0.1 % FS 3) 0.1 % FS 3)

WIKA-Wireless USB stick for wireless data transfer WIKA-Cal BSP adapter set NPT adapter set Sealing set, consisting of 4 x G ½ USIT seals and 2 x G ¼ USIT seals 3 x AA alkaline batteries

Adapter set for test item connection

Pressure range

BSP

Adapter

G ¼ male to: G  female G  female G ½ female

NPT
G ¼ male to:  NPT female ¼ NPT female  NPT female ½ NPT female

Dimensions in mm (in)
Various test connections

32.5 (1.28) 25.5 (1.00)
31.5 (1.24) 25.5 (1.00)

G 1/2

G 1/4

WIKA data sheet CT 93.02  08/2017

NPT 1/2

NPT 1/4
Page 3 of 6

Dimensions in mm (in)
CPG1500 without protective rubber cap

Hydraulic hand test pump model CPP700-H

Front view

1

2

3

4

Rear view 4

8

1

9 7
7
6 6
5 5

1 Test item connection hose; Minimess® system with a G ¼ female pressure gauge bulkhead connector
2 Pressure relief valve 3 Plug screw for fluid reservoir 4 Fine adjustment valve (volume adjustment) 5 Pump handles

6 Toggle control, system and high-pressure generation
7 Fluid reservoir
8 Connection G ½ female, freely rotating for reference instruments
9 Blind plug, do not open!

Page 4 of 6

WIKA data sheet CT 93.02  08/2017

WIKA-Cal calibration software
Easy and fast creation of a high-quality calibration certificate
The WIKA-Cal calibration software is used for generating calibration certificates or logger protocols for pressure measuring instruments and is available as a demo version for a cost-free download. A template helps the user and guides him through the creation process of a document. In order to switch from the demo version to a full version of the respective template, a USB stick with the template has to be purchased. The pre-installed demo version automatically changes to the selected full version when the USB stick is inserted and is available as long as the USB stick is connected to the computer.
 Creation of calibration certificates for mechanical and electronic pressure measuring instruments
 A calibration assistant guides you through the calibration  Automatic generation of the calibration steps  Generation of 3.1 certificates per DIN EN 10204  Creation of logger protocols  User-friendly interface  Languages: German, English, Italian and more due with
software updates
For further information see data sheet CT 95.10

Calibration certificates can be created with the Cal-Template and logger protocols can be created with the Log-Template.

Cal Demo Generation of calibration certificates limited to 2 measuring points, with automatic initiation of pressures via a pressure controller.
Cal Light Generation of calibration certificates with no limitations on measuring points, without automatic initiation of pressures via a pressure controller.

Calibration certificate Kalibrierzeugnis

General Allgemein
Calibration certificate Kalibrierschein
Customer Kunde
Customer order No. Kundenbestellnummer Order date Bestelldatum Date of calibration Datum der Kalibrierung Operator Bearbeiter

Demo customer Demo street DE-12345 Demo city
01/02/2017 Demo User

Environmental conditions Umgebungsbedingungen

Temperature Temperatur
Atmospheric pressure Atmosphärischer Luftdruck
Rel. humidity Rel. Luftfeuchte

(20,0 ± 0,1) °C (1013,0 ± 0,1) hPa (40 ± 1) %

Measurement conditions Messbedingungen
Working standard (WS) Pressure controller Gebrauchsnormal (GN) Druckcontroller
Model Typ Calibration mark Kalibrierzeichen Scale range Anzeigebereich Accuracy Genauigkeit

Demo CPC Demo 12345 (0 ... 10) bar g 0.02 % FS

Certificate No. Zeugnis-Nr.

00000122

Device under test (DUT) Kalibriergegenstand (KG)

Electrical gauge Elektrisches Druckmessgerät
Model Typ Serial number Seriennummer Scale range Anzeigebereich Accuracy Genauigkeit Resolution (scale division) Auflösung (Skalenwert) Measuring point No. Messstellen-Nr

D-10 demo1234 (0 ... 10) bar g 0.05 % FS 0.0001 bar demo tag

Place of calibration Ort der Kalibrierung
Demo place Musterort

g = (9.80946 ± 0.01) m/s²

Demo laboratory Demo street 123 12345 Demo city Demo country

Tel.: 01234/56789 Fax: 01234/56789 E-Mail: demo@demo.com www.demo.com

Page Seite

1 / 2

Calibration certificate Kalibrierzeugnis

Certificate No. Zeugnis-Nr.

Measurement results Messergebnisse

Reference value WS Referenzwert GN bar
0.00000 10.00000

Reading DUT Ablesung KG bar

 0.0000 10.0010

 0.0010 10.0020

Statement of compliance Konformitätsaussage
Object keeps the specifications! Der Kalibriergegenstand hält die Herstellerspezifikationen ein!

00000122

Demo laboratory Demo street 123 12345 Demo city Demo country

Tel.: 01234/56789 Fax: 01234/56789 E-Mail: demo@demo.com www.demo.com

Page Seite

2 / 2

Log Demo Creation of data logger test reports, limited to 5 measured values.

Log Creation of data logger test reports without limiting the measured values.

WIKA data sheet CT 93.02  08/2017

Page 5 of 6

Scope of delivery
 Precision digital pressure gauge model CPG1500  Hydraulic hand test pump model CPP700-H  Adapter set  Protective rubber cap for model CPG1500  Case with matching foam insert for CPG1500, CPP700-H
and adapter set  Operating instructions CPG1500 and CPP700-H

Options
 DKD/DAkkS calibration

Accessories
 WIKA-Cal calibration software  WIKA-Wireless USB stick

© 10/2016 WIKA Alexander Wiegand SE & Co. KG, all rights reserved. The specifications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modifications to the specifications and materials.
Page 6 of 6

WIKA data sheet CT 93.02  08/2017

08/2017 EN


sejda.com (4.1.2) SAMBox 2.2.3