ADInstruments NZ HU11 Short range telemetry transmitter User Manual

Millar Instruments Ltd Short range telemetry transmitter

User Manual

 User Manual for Millar Mouse Telemetry MKT0006/A
 User Manual for Millar Mouse Telemetry MKT0006/A Contents   Important Information ............................................................................................................... 1Disclaimer ......................................................................................................................... 1Warranty ........................................................................................................................... 1FCC Compliance .............................................................................................................. 2Introduction .............................................................................................................................. 4Telemetry System Components ............................................................................................... 5Quick Start Guide ..................................................................................................................... 6Configurator System Setup ...................................................................................................... 7Configurator Software Installation ..................................................................................... 7Connecting the Configurator ........................................................................................... 10Configurator Software - Overview ................................................................................... 11tBase - Overview .................................................................................................................... 13tBase Laboratory Setup .................................................................................................. 14Status lights .................................................................................................................... 14Analog Outputs and Calibration Values .......................................................................... 15The Mouse telemeter ............................................................................................................. 16Channel and serial numbers ........................................................................................... 16System setup and configuration ............................................................................................. 17Changing telemeter Channel .......................................................................................... 18Changing tBase Channel ................................................................................................ 20Positioning the telemeter during surgery ................................................................................ 22Telemeter Care and Handling ................................................................................................ 24Diagnostics ............................................................................................................................. 25tBase Technical Specifications .............................................................................................. 26Telemeter Technical Specifications ....................................................................................... 27Frequently Asked Questions........ .......................................................................................... 28Contact Information ................................................................................................................ 28  To all users; We pride ourselves in partnering with you to achieve excellent results.  If you have any questions related to the operation of your system, please contact us at support@millar.com We are here to help.  Contact Information  US Head Office  New Zealand Office Houston, Texas Toll free (USA only) 1 800 669 2343 TEL: +1 832 667 7000 FAX: +1 713 714 8497 Auckland, New Zealand TEL: +64 9 367 7126 FAX: +64 9 367 7157   EMAIL support@millar.com or sales@millar.com WEB www.millar.com
 User Manual for Millar Mouse Telemetry MKT0006/A
 User Manual for Millar Mouse Telemetry Systems MKT0006/A  Page 1 Important Information Disclaimer Millar believes that information in this user manual is correct at the time of publishing.  In the event errors may exist, Millar reserves the right to make changes to future editions without notice to holders of this edition.  The reader should consult Millar if errors are suspected.  In no event shall Millar be held liable for damages arising out of the use of this document or information contained within. Except as specified below, Millar makes no warranties, express or implied, and specifically disclaims any warranty of the products fitness for a particular purpose. The customer’s right to recover damages by fault or negligence of Millar are limited to the amount paid by the customer.  Any action against Millar must be brought within one year after the cause of action accrues.  Millar will not be held liable for any loss of data, research funding or profits and neither will it be liable for incidental or consequential damages even if advised of the possibility thereof. Millar expresses that its products are for research purposes only. They are not designed with components and testing  intended  for  use  in medical treatment  or diagnosis of humans  or  animals  and  doing  so  may result  in serious harm. Warranty Telemeter Warranty Millar warrants that at the time of sale and shipment to the original Purchaser, Millar mouse telemeters shall be free from defects in material and workmanship.  This warranty voids on first use of the telemeter.  If there is such a defect, Millar will, at no charge, replace the equipment as appropriate.   Telemetry Hardware Warranty Millar warrants that at the time of sale to the original Purchaser, the telemetry hardware (excluding telemeters – see above) shall be free from defects in materials and workmanship for a period of one (1) year from its date of shipment to the original purchaser. If there is such a defect, Millar will, at no charge and at its option, either repair or replace the equipment as appropriate.   Should the hardware become damaged after expiration of the one year warranty, Millar will accept the product for a no charge evaluation with a signed RMA (return material authorization).  If repairs are required, Millar will provide a detailed quote and lead time for the repair process.  Repair work will not commence until approval is received from the customer.   As  part  of  Millar’s  continued  commitment  to  customer  support,  free  online, phone  and  email  support  in  the  use  of  all telemetry equipment is available beyond all warranty periods. The warranties above do not cover damage caused by failure to follow instructions for use, owner’s abuse, misuse or negligence, user tampering or modifications, power failures or surges and events or accidents considered force majeure such as flooding, fire, etc. outside the reasonable control of Millar. In no case shall liability exceed the purchase price of the original product. Millar provides free product evaluations during the  warranty period and free  return  to customer shipment. All  returned products should be packed safely, preferably in original packaging, cleaned/sterilized and shipped to Millar with the RMA number visible on the packaging and with all RMA documentation. In order to receive the highest level of support, the customer  is  responsible  for  notifying  Millar  immediately  of  a  problem  within  the  warranty  period.  Following  product evaluation, Millar will replace or repair (excluding telemeters which are non-repairable) any product found to be defective and covered within the warranty period, while operated in accordance with instructions for use and specifications.    Copyright © Millar, November 2016 This  publication  may not  be  reproduced  or  transmitted in  any  form, electronic or  mechanical,  including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of Millar.    For complete product instructions, general insights and surgical recommendations, visit millar.com and access Millar’s free online Knowledge Center. http://millar.com/knowledge-center Millar is committed to delivering the collaborative support you expect in facilitating scientific insight and advancing medical innovation.
 Page 2  User Manual for Millar Mouse Telemetry MKT0006/A FCC Compliance Telemeter (MT10B) This  device  complies  with  part  15  of  the  FCC  Rules.  Operation  is  subject  to  the  following  two conditions: 1. This device may not cause harmful interference, and 2.  This  device  must  accept  any  interference  received,  including  interference  that  may  cause undesired operation. Note:  This  equipment  has  been  tested  and  found  to  comply  with  the  limits  for  a  Class  B  digital device,  pursuant  to  part  15  of  the  FCC  Rules.  These  limits  are  designed  to  provide  reasonable protection against harmful interference in a residential installation. This equipment generates, uses and  can  radiate  radio  frequency  energy  and,  if  not  installed  and  used  in  accordance  with  the instructions,  may  cause  harmful  interference  to  radio  communications.  However,  there  is  no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful  interference  to  radio  or  television  reception,  which  can  be  determined  by  turning  the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:   Reorient or relocate the receiving antenna.   Increase the separation between the equipment and receiver.   Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.   Consult the dealer or an experienced radio/TV technician for help. Changes or modifications not expressly approved by Millar Inc. could void the user's authority to operate the equipment. TR190 Configurator This  device  complies  with  part  15  of  the  FCC  Rules.  Operation  is  subject  to  the  following  two conditions: 1. This device may not cause harmful interference, and 2.  This  device  must  accept  any  interference  received,  including  interference  that  may  cause undesired operation. Note:  This  equipment  has  been  tested  and  found  to  comply  with  the  limits  for  a  Class  B  digital device,  pursuant  to  part  15  of  the  FCC  Rules.  These  limits  are  designed  to  provide  reasonable protection against harmful interference in a residential installation. This equipment generates, uses and  can  radiate  radio  frequency  energy  and,  if  not  installed  and  used  in  accordance  with  the instructions,  may  cause  harmful  interference  to  radio  communications.  However,  there  is  no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful  interference  to  radio  or  television  reception,  which  can  be  determined  by  turning  the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:   Reorient or relocate the receiving antenna.   Increase the separation between the equipment and receiver.   Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.   Consult the dealer or an experienced radio/TV technician for help. Changes or modifications not expressly approved by Millar Inc. could void the user's authority to operate the equipment.
 User Manual for Millar Mouse Telemetry Systems MKT0006/A  Page 3 MT110 tBase The Wireless Power Transfer System is FCC Part 18 Compliant Note: This equipment has been tested and found to comply with part 18 of the FCC limits for non-consumer equipment. These limits are designed to provide reasonable protection against harmful interference  in  an  industrial  installation.  This  equipment  generates  uses  and  can  radiate  radio frequency  energy  and,  if  not  installed  and  used  in  accordance  with  the  instructions,  may  cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television  reception,  which  can  be  determined  by  turning  the  equipment  off  and  on,  the  user  is encouraged to try to correct the interference by one or more of the following measures:   Reorient or relocate the receiving antenna.   Increase the separation between the equipment and receiver.   Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.   Consult the dealer or an experienced radio/TV technician for help. Note: There is no required maintenance of this device from a FCC compliance perspective. The 2.4GHz Data Transmitter is FCC Part 15 Compliant This  device  complies  with  part  15  of  the  FCC  Rules.  Operation  is  subject  to  the  following  two conditions: 1. This device may not cause harmful interference, and 2.  This  device  must  accept  any  interference  received,  including  interference  that  may  cause undesired operation. Note:  This  equipment  has  been  tested  and  found  to  comply  with  the  limits  for  a  Class  B  digital device,  pursuant  to  part  15  of  the  FCC  Rules.  These  limits  are  designed  to  provide  reasonable protection against harmful interference in a residential installation. This equipment generates, uses and  can  radiate  radio  frequency  energy  and,  if  not  installed  and  used  in  accordance  with  the instructions,  may  cause  harmful  interference  to  radio  communications.  However,  there  is  no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful  interference  to  radio  or  television  reception,  which  can  be  determined  by  turning  the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:   Reorient or relocate the receiving antenna.   Increase the separation between the equipment and receiver.   Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.   Consult the dealer or an experienced radio/TV technician for help. Changes or modifications not expressly approved by Millar Inc. could void the user's authority to operate the equipment.
 Page 4  User Manual for Millar Mouse Telemetry MKT0006/A Introduction  This User Manual contains information to help with the correct installation and basic use of the Millar Mouse Telemetry System.  It is strongly recommended that you also visit our online Knowledge Center for more information and to view instructional videos showing telemeter implantation and recommendations for general use and care of your system. www.millar.com/knowledge-center  The Millar Mouse Telemetry system uses a unique and efficient patented inductive power transfer system  for  making  long-term  recordings  in  conscious  mice  (>22  g).    No  battery  is  required  to wirelessly record high quality biological signals 24/7. Telemetered mice can be housed in a single room without the need for shielding as each mouse telemeter communicates with a tBase on one of 40 independent transmission frequencies (Channels).  The  system  comprises  of  a  telemeter,  tBase  and  Configurator  System.  The  mouse  biopotential telemeter measures the electrical potential (ECG, EMG, EEG) between the positive and negative leads  and  transmits  the  data  wirelessly  to  the  tBase.  The tBase  provides  wireless  power  to  the telemeter, acts as the receiver for the digital data (sampled at 2 kHz) transmitted by the telemeter and converts it to an analog signal.  The Millar Mouse Telemetry System is therefore compatible with all acquisition systems with analog inputs.  The Configurator System allocates telemeter and tBase data transmission frequencies and provides diagnostic information on the telemeter or tBase.    Due to continual product improvement, your system may look slightly different to the images and graphics contained in this manual. If you  have  any  questions or need help  please  contact  us  at support@millar.com  Key Features  Accuracy: For long term, high frequency physiological recordings.  Digital wireless recording of data with the highest quality and signal integrity (sampling 2 kHz).  24/7: Data can be collected 24 hours a day while the animal is on the tBase.  Versatility: Telemeters send digital data wirelessly from within the animal to the tBase where it is converted to an analog signal compatible with existing data acquisition systems.  Low cost of ownership: Millar mouse telemeters are single use items which eliminates the time and cost associated with telemeter explant and refurbishment.  Efficient housing: Multiple animals can be housed in one location for simultaneous recordings.  No specialized cages or  shielding  is  required  as  each  telemeter  operates  independently  on  one  of  40  transmission frequencies.  Service & support: Millar  prides  itself  on  being  responsive  to  our  customers’  current  and  future  needs.    We  have experienced engineers and scientists on staff to provide professional, expert and timely support and advice.
 User Manual for Millar Mouse Telemetry Systems MKT0006/A  Page 5 Telemetry System Components  The Mouse telemeter All telemeters are shipped individually packaged, sterile and ready for use.  The telemeters will need to be configured to a data collection Channel prior to implantation.  PLEASE NOTE the telemeter does NOT need to be removed from its packaging for Configuration.  When handling the telemeter, care must be taken not to damage the telemeter body or lead wires as telemeters are non-repairable.  The tBase The  tBase  is  able  to  detect  the  location  of  the  telemeter  within  the  operating  area  to  efficiently provide power to the telemeter.  For safe and normal operation of the tBase it needs to be placed on a flat non-metallic surface.  There must not be any metal or another tBase within 10 cm of the tBase and we recommend at least a minimum of 40 cm separating distance above or below another tBase.   The Configurator System The  Configurator  System  (Configurator  hardware  and  ConfigSoft  software)  wirelessly  controls mouse telemeters and tBases to:   Change/configure mouse telemeter and tBase transmission Channels (this must be done before implantation).   Run and provide diagnostic information on mouse telemeters and tBases.  ONLY a single Configurator System is required to control all Millar (rat or mouse) telemetry systems in a laboratory.  The Configurator System is not required for data acquisition. Note: Existing Millar rat  telemetry  customers  with  a  Configurator  System  may  only  require  a  firmware  and  software update  to  their  Configurator  for  use  with  the  mouse  telemetry  system.    Please  contact support@millar.com for assistance.   ** IMPORTANT: The ConfigSoft software supplied on the USB drive must be installed on a computer before connecting the Configurator hardware to the computer.
 User Manual for Millar Mouse Telemetry MKT0006/A  Page 6 Quick Start Guide
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 7 Configurator System Setup Introduction The  Configurator  System  (TR190  Configurator  hardware  and  ConfigSoft  software),  wirelessly communicates with the mouse telemeters and tBases. The system allows the user to: 1)  Change/configure the transmission Channel of the tBase  2)  Change/configure the transmission Channel of the telemeter. 3)  Run Diagnostics on telemeters and tBases.  Only a  single Configurator  System is  required  for  each  laboratory  and  it is  NOT required during data acquisition.  Existing Configurator hardware in use with the Millar Rat Telemetry Systems  can  be  used  with  the  Millar  Mouse  Telemetry  System  however  a  firmware  and ConfigSoft software upgrade may be required.  Contact support@millar.com for assistance.  Configurator Software Installation NOTE: ConfigSoft must be installed on the computer before connecting the Configurator hardware.  Existing versions of ConfigSoft must also be uninstalled prior to installation. System requirements: Windows 7, 8 or 10.  There are two separate installers 32 bit and 64 bit, please install the version compatible with your Windows operating system.  Internet connection is necessary to access and install any drivers not already installed.  1)  Connect the supplied USB drive containing ConfigSoft to your computer. 2)  Open the ConfigSoft folder and go to folder “ConfigSoft 32 bit” or “ConfigSoft 64 bit” according to your Windows Operating System. 3)  Double click on “setup” to start the installation.    4)  Once the welcome screen appears, follow the instructions for installation.
 Page 8  User Manual for Millar Mouse Telemetry MKT0006/A 5)  Previous  versions  of  the  ConfigSoft  software  must  be  removed  before  continuing  with  the installation.  If you need to remove the program, follow the instructions in this window before clicking Next.      6)  Confirm installation.  Click Next   7)  After the installation has been completed.  Click Close.
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 9 8)  The following window will automatically appear to install the driver for the TR190 Configurator if it is not already located on the computer.  Click Next.    9)  After successful installation of the TR190 Driver, Click Finish.   10)  Installation of ConfigSoft and the TR190 Configurator drivers has been completed.
 Page 10  User Manual for Millar Mouse Telemetry MKT0006/A Connecting the Configurator  Front  Rear     1)  Attach the antenna to the rear of the Configurator.  2)  After installation  of  ConfigSoft,  connect  the Configurator to the computer using  the  supplied USB cable.  The status light on the front of the Configurator should glow red or green indicating that it is receiving power from the computer.  The status light on the front panel of the Configurator indicates the following information:  Off    Not connected, USB cable unplugged, Configurator is off Red   Configurator is receiving power but not communicating with computer Green  Configurator is communicating with the computer  3)  The computer will install any additional drivers and the status light should glow green.  If there are any problems please contact support@millar.com.  4)  The Configurator System is now ready for use with all your mouse telemeters and tBases.   Status light  Antenna connection  USB port
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 11 Configurator Software - Overview ConfigSoft can be started from either the   Desktop icon   OR   Start>Programs>Millar Telemetry>ConfigSoft
 Page 12  User Manual for Millar Mouse Telemetry MKT0006/A From the main menu the user has access to the: 1.  Rat Telemeter (not required for the mouse telemetry system) 2.  SmartPad (not required for the mouse telemetry system) 3.  Mouse Telemeter 4.  tBase 5.  SmartPad Status (not required for the mouse telemetry system)  Mouse Telemeter Tab    tBase Tab
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 13 tBase - Overview The tBase is both a wireless power supply and receiver for the mouse telemeter.  The top of the tBase has an indent (335mm (L) x 190mm (W) x 10mm (D)) which a mouse cage must be positioned in for the best access to the inductive power field.  Only use cages that fit into the indent area with the cage base in contact with the surface of the indent. The tBase generates an inductive power field which extends approximately 7cm above the surface of  the  indent.    When  the  mouse  telemeter  is  within  the  tBase  power  field,  the  telemeter  will  be powered  and  data  recorded.    Only  telemeters paired  on the  same  transmission  Channel as  the tBase will communicate, be powered and transmit data efficiently and accurately. No data will be recorded when the telemeter is not in the power field.  For optimal powering, it is recommended that the cage is centered in the tBase indent. NOTE: The use of cages with base dimensions that exceed those of the tBase indent is NOT recommended and may compromise data collection.  tBase Front      tBase Rear      When initially setting up a laboratory with multiple tBases, it is important that only ONE tBase is turned on at a time.  The tBases and telemeters are all shipped set to Channel “Setup”.  Before using the telemetry system each tBase and mouse telemeter needs to be configured and paired to one of the 40 transmission Channels.  It is important that only one tBase and one telemeter are used on any particular channel to prevent signal interference. The recommended arrangement of multiple tBases in a laboratory is detailed below. In particular, note the distances from other tBases and metal objects.  Power socket  Analog Outputs Indent for cage positioning Status light  Air vent inlet Air vent outlet Serial number
 Page 14  User Manual for Millar Mouse Telemetry MKT0006/A tBase Laboratory Setup The tBase provides  power  to  the telemeter using an  electromagnetic  field  generated above  and below the tBase.  For this reason, it is important that a tBase is NOT placed on a metal surface or above or below other tBases (unless 40 cm apart).  We recommend the use of plastic shelving for your tBases.  Multiple  tBases  should  be  arranged  so  that  there  is more  than  10  cm  horizontal  and  40  cm  vertical separation  between  them.    If  using  shelving  that  has metal  supports  or  framing  then  the  tBases  must  be >10cm away from these as well (don’t forgot to check the underside of the shelves for metal supports).  There should also be more than 10 cm separation between the top of a metal cage lid and any tBase on a shelf above.   If metal shelves are used, the tBase needs to be raised at least 10 cm from the shelf surface but still maintaining a vertical separation of at least 40 cm between tBases.   If the tBase shows a solid red light, then the tBases need to be further separated (vertically and horizontally) and/or  move  them  further  away  from  metal  brackets, supports, cables or power packs.  Reset the tBase by disconnecting  and  reconnecting  the  power.  If  the problem continues, please contact support@millar.com.  IMPORTANT:   Connect ONLY the supplied and labelled “TR181 or MT110” Power Supply to the power socket at the rear of the MT110 tBase.  Please note: Using any other Power Supply will invalidate the warranty and may damage the tBase.   DO NOT obstruct the air vents at the front and rear of the MT110. Ensure that any power cables and other items are kept clear of the air vents.   Status lights The status light at the front of the tBase provides information on the communication status between the tBase and the telemeter.  The  status  lights  on  the  tBase  may  flicker  between  solid  green  and  solid  orange  during  normal operation as the telemeter enters and leaves the power field with normal activity of the mouse i.e. mouse climbing on and off large objects in the cage. *Troubleshooting:  When the status light is solid red, the power field of the tBase has been disabled and the telemeter will not be powered.  This means that there will be no data recorded from the telemeter.  Cause: may be due to the tBase being placed too close to a metal surface or to another tBase.  Move the tBase away from any metal and then reset by turning the power off and then on Status light  MT110 tBase Solid Green  The tBase is communicating with a telemeter on the same Channel Solid Orange  The tBase is turned on but is not communicating with a telemeter Solid Red* The tBase has detected an issue with the telemetry setup and has disabled the powering.  Please see troubleshooting below for more information.   Flashing Red  The tBase is in diagnostic mode.  No telemeter data will be output Example tBase arrangement on plastic shelves with metal framing.
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 15 again. If the problem persists, record a diagnostic file in ConfigSoft while the tBase is showing a red status light and contact support@millar.com. Analog Outputs and Calibration Values After pairing the tBase/telemeter Channels, the analog outputs of the tBase will correspond to the following signal outputs:  Telemeter model  Output 1 Output 2 Output 3 MT10B Activity Biopotential Data Received Calibration values  0V = 0 RAU 3.84V = 10 RAU 2.048V = 0mV 4.096V = 2.5mV  ~0V = no data received ~3.3V = data received   Activity: As the mouse moves in the power field the strength and orientation of the electromagnetic field it receives changes.  These changes are detected by the tBase and collated every ~1 second to calculate an  index  of  animal activity.  The  output  is  updated  approximately  every 1  second  as ranges from 0V – 3.84V in 13 levels (see picture below). We recommend converting the voltage output to Relative Activity Units (RAU). The activity index can be used to determine relative changes in activity between two time periods (e.g. night vs day).   Biopotential:  This is your ECG, EEG or EMG signal.  The potential difference between the two electrode leads is measured by the telemeter (sampling rate 2 kHz) and transmitted to the tBase. The tBase low-pass filters the signal at 1 kHz and reconstructs the analog signal.  Data Received:  This provides an easy visual representation of when data is collected or when it has been missed.  When the tBase is receiving data from a telemeter, the Data Received analog output will be ~3.3V. If no data is received from a telemeter for 12 milliseconds this output will drop to ~0V until data is received again (see picture below). The  most  likely  cause  of  data  not  being received is that the mouse and telemeter have  moved outside  the  inductive  power  field  e.g.  the mouse is hanging off the bars of the cage lid and is more than 7cm above the tBase surface. The Data Received output can be useful during data analysis to differentiate between real periods of low signal amplitude and missing data e.g. skipped beats in ECG recordings.  Example recording showing the Activity Index, Biopotential and Data Received outputs.
 Page 16  User Manual for Millar Mouse Telemetry MKT0006/A The Mouse telemeter  Millar mouse telemeters are designed for single use and for 24/7 recordings from mice living in their home cages. The telemeters are designed, with a unique contour, for use as subcutaneous implants in mice >22 grams.  Telemeters are shipped sterile and ready for implantation.  ALL telemeters are shipped  on  Channel  “Setup”  and  each  telemeter  MUST  be configured to one of 40 different data recording Channels before implantation.  DO NOT change the Channel of the telemeter while implanted in an animal.  There is no need to unpack the telemeter from the sterile packaging (see right) to change the channel.    The telemeters are precision electronic devices. Care must be taken not to damage the telemeter body or electrodes as they cannot be repaired.  Please read the section about Care and Handling of your telemeters.      The MT10B telemeter (above) can measure a single biopotential signal between the positive and negative  leads  for  applications  including  ECG,  EMG  and  EEG.  For  ease  of  identification,  the negative lead is trimmed shorter than the positive lead so the wire coil is flush with the tubing. Before implanting a telemeter make sure to write down the serial number and Channel setting.  Channel and serial numbers All telemeters are factory set to Channel “Setup” and need to be configured to different channels before use (see page 18). The telemeter model and serial number can be found on the telemeter packaging.  In the  example below, the model is MT10B, the serial number is 00000.  The serial number can also be found on the end of the telemeter.     Negative lead
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 17 System setup and configuration Before  implanting  a  telemeter  into  an  animal,  the  telemeter  and  the  tBase  to  be  used  with  the telemeter must have their Channels paired.  Please ensure that only one tBase and one telemeter are  being  configured and paired  at  any one  time.  All tBases and telemeters are shipped on Channel “Setup” and should be changed to one of the 40 available data collection Channels before animal implantation.  The tBase and telemeter need to be within 1 meter of the Configurator during configuration.    1)  Turn on the tBase (Status light should be orange).  If this is the first time the tBase has been turned on, it should be on the Channel “Setup”.  If not, use the Configurator System to configure the tBase Channel to “Setup” (see page 20 for details).         2)  Place a telemeter (in its sterile packaging) in the center of the indent of the tBase.  The tBase will  automatically  start  powering  the  telemeter  as  both  the  tBase  and  telemeter  will  be  on Channel “Setup”.  The tBase status should turn green.   3)  Using the Configurator system, first change the Channel of the mouse telemeter to one of the 40 data collection Channels (detailed instructions in the next two sections).  The tBase status light should now be orange.  Remove the mouse telemeter from the tBase.  4)  Configure  the  tBase  to  the  same  data  collection  Channel  as  the  mouse  telemeter  using ConfigSoft (detailed instructions in the next two sections).  Place the telemeter on the tBase, the status light should be green.  Your system is now paired and ready for use.     Insert power plug here Status light
 Page 18  User Manual for Millar Mouse Telemetry MKT0006/A Changing telemeter Channel  1)  Place  the  mouse  telemeter  in  the  center of  the  indent  of  the  tBase  configured  to  the  same Channel (first use Channel = “Setup”).  Ensure the telemeter and tBase are within 1 meter of the Configurator. The tBase status light should be green to show the telemeter is powered and communicating with the tBase. 2)  Open ConfigSoft 3)  Click on the “Mouse Telemeter” tab  4)  Enter  the  serial  number  of  the  telemeter.    The  number  is  on  the  outside  of  the  telemeter packaging or on the label at the end of the telemeter.  5)  Select or enter the current Channel number of the telemeter (first use: Channel = “Setup”). 6)  To change the Channel number, select or enter the new Channel number in the “Configure your telemeter” section. 7)  Click Go to activate the new Channel. 8)  A dialog will pop up with “Only change the telemeter transmission Channel before implantation”. Click OK to confirm the choice.
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 19 9)  A message confirming the new telemeter Channel should appear.  To make it easier for the future, record the telemeter Serial number and Channel number for all telemeter configurations.    Troubleshooting: Configurator – ConfigSoft - telemeter If the Configurator cannot communicate with the telemeter an error message may appear.    Solution:  Do the following and try again:   Check that you have entered the correct telemeter Serial number   Check that you have entered the correct telemeter Channel number   Make sure that the telemeter is on a tBase set to the same Channel   Move the telemeter and paired tBase closer to the Configurator   Check you only have one tBase/telemeter set to the same Channel   Move the telemeter to a different position on the tBase
 Page 20  User Manual for Millar Mouse Telemetry MKT0006/A Changing tBase Channel 1)  To  change  the  tBase  Channel  and  pair  this  with  a  telemeter,  click  on  the  “tBase”  tab  in ConfigSoft.    2)  Enter the serial number of the tBase (found on the front and rear of the tBase). 3)  Enter current Channel number of the tBase (first use: Channel = “Setup”). 4)  Under “Configure your tBase”, enter or select a new Channel number for the tBase (this should be same as telemeter being used). 5)  Click Go to activate the new Channel. 6)  Click OK to confirm your choice.   6.  The tBase status light will flash red for a few seconds before a message confirming that the tBase channel has been changed appears.  Record the tBase serial number and new Channel number for future reference.
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 21 7)  If a telemeter on the same Channel is placed on the tBase, the Status light should change to green.    The  telemeter  and  tBase  are  now  on  the  same  Channel,  paired,  and  able  to communicate for data collection and transmission.  Troubleshooting: Configurator – ConfigSoft - tBase If the Configurator cannot communicate with the tBase the following error message may appear.    Click OK and perform the following then retry:   Check that the light on the front of the tBase is orange.   Check that you entered the correct Serial and/or Channel numbers.   Move tBase closer to Configurator.   Check you only have one tBase set to the same Channel     Repeat the above process for each of the tBases and telemeters to configure each pair to a different Channel.  Your Millar Mouse Telemetry System is now configured and ready for use.
 Page 22  User Manual for Millar Mouse Telemetry MKT0006/A Positioning the telemeter during surgery   The telemeter has a unique contoured shape with a curved convex front and a concave underside for better positioning in the mouse.  The telemeter body is designed to fit subcutaneously, with the concave underside sitting against the flank of the mouse and the convex front side next to the skin.  Biopotential lead placement MT10B Mouse biopotential telemeters are supplied with 8cm long stainless steel bipolar electrodes suitable for measurement of biological electric potential signals including ECG, EEG and EMG.  To reduce noise in the recorded signal it is recommended that the electrode leads be trimmed to length at the time of surgery rather than coiling any excess wire. It is important however to allow some extra  length  in  the  electrode leads  to allow  for  animal  movement.  For  ease  of  identification,  the negative lead is trimmed shorter than the positive lead so the wire coil is flush with the tubing.  After trimming the electrode leads to the desired length, it will be necessary to expose the coiled stainless steel electrode within the lead tubing. Care must be taken to avoid damaging the wire as this can lead to wire breakages and poor signal quality during recordings. The recommended way to expose the electrode wire is to use a surgical microscope and carefully trim the tubing from around the wire. Do not trim, stretch or disturb the wire coil within 2.5 cm of the telemeter body.  The  stainless  steel  electrodes  are  coiled  for  maximum strength  and  flexibility.  The  coil  can  slip  out  of  a  suture however so it may be useful to carefully stretch the coils at the  tip  of  the  electrode  to  allow  a  suture  thread  to  pass between the coils.   It is not recommended that sutures be placed around the leads other than around the point of contact with  the  tissue  as  this  may  provide  a  stress  point  on the  leads and  cause them  to  break.  More information and surgical videos are available online at http://millar.com/knowledge-center Negative lead
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 23                     Telemeter Body Placement for Powering and Signal Maximization The  tBase  provides  an  electromagnetic  field  to power  the  mouse  telemeter.    This  field  provides effective powering of the telemeter within up to 7cm vertical distance from the bottom of the indent on the tBase.  However, when the telemeter body is more than 7cm above the tBase, the telemeter will not be powered and data collection will not be available.  This can occur when the mouse hangs from the wire cage lid or is physically sitting more than 7cm above the tBase. The recommended site for telemeter placement is on the flank/side of the animal between the front and the rear limbs with the concave surface of the telemeter against the body.  Placement on the right hand side of the mouse means the electrode leads will exit from the top corner of the telemeter body and can be tunneled over the shoulder for ECG placement or to the skull for measurement of EEG.  Powering may not be as effective if the telemeter is placed on the back of the animal or up near the shoulders as that creates additional distance from the tBase, especially when the animal is up on their hind legs. More information and surgical videos are available online at the Millar Knowledge Center (http://millar.com/knowledge-center).   It is also strongly recommended that bedding within the cage is kept to a minimum and tall items, which the mouse can climb on top of, are removed.  For example, large enrichment tubes, multi-level mouse houses or other toys with elevated flat surfaces should be avoided.       Example placement of electrode leads for recording of high quality ECG
 Page 24  User Manual for Millar Mouse Telemetry MKT0006/A Telemeter Care and Handling All  telemeters  are  shipped  pre-sterilized  using  ethylene  oxide  gas  sterilization  and  can be implanted without the need for further sterilization.  Telemeters will remain sterile as long as the  pouch  is  intact  and  not  compromised.  It  is  recommended  that  they  remain  in  their  original packaging until surgical implantation.   However,  if  the  telemeters  have  been  accidently  removed  from  the  sterile  packaging  prior  to implantation,  or  if  they  come  in  contact  with  non-sterile  surfaces  during  surgery,  they  can  be sterilized using one of the methods below: Chemical Sterilants Glutaraldehyde may be used to sterilize the telemeters.  Glutaraldehyde (Fisher Scientific) must be diluted to 2% before use.  Check your local chemical supply company for availability.  Do not use gluteraldehyde  solutions  containing  surfactants  (i.e.  do  not  use  Cidex  7,  Cidex  Plus  28  Day  or Metrocide 28).  Suitable products are in the table below:    Rinse with Sterile Saline Telemeters sterilized with a chemical sterilant should always be washed thoroughly in sterile saline before implantation to remove all traces of the sterilant.    Ethylene Oxide Gas Sterilization Telemeters  are  suitable  for  Ethylene  Oxide  sterilization  as  long  as  they  are  not  subjected  to temperatures above 60°C.  *Important note:  Under no circumstances should the telemeter be autoclaved or subjected to temperatures over 60°C as this will damage the telemeter. Do not sterilize by radiation (gamma or e-beam), plasma, alcohol, peroxide or formaldehyde vapor solutions. This will damage the telemeter. Customers are liable for product replacement if  recommended  products and  instructions are not used or followed.  If there are any questions about a chemical or procedure please contact Millar (support@millar.com). Warning: Some examples of chemicals that will cause damage to telemeters include, but are not limited to: alcohols, phenols, iodophors, and hypochlorite.  Please confirm with Millar before using any product other than the approved products list above. ** Products with similar names are available but may not be suitable.  Please use only products specified or contact Millar for clarification (support@millar.com).  Removal The  mouse  telemeters  are  designed  as  a  single  use  product.  Depending  on  your  biohazard  or medical waste disposal services, it is possible to discard the biopotential leads and/or the telemeter body while still implanted in the mouse at the end of the experimental protocol.  Trade Name  Manufacturer  Active Ingredient  Soak Time / Temperature Cidex Activated Dialdehyde Solution Advanced Sterilization Products (J&J) Gluteraldehyde  1-2 hours / 25°C (77°F)  Cidex OPA  Advanced Sterilization Products (J&J) Orthopthalaldehyde  16-30 mins / 20°C (68°F) MetriCide Metrex  Gluteraldehyde  1-2 hours / 25°C (77°F)
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 25 Diagnostics Using the Configurator System the user can view and change the Channel of the telemeter.  The “Search for telemeter” function can be used to find the serial number of an implanted telemeter if this was not recorded prior to implantation.   The “Search for tBase” function will find the Channel of a specific tBase. The Configurator System can also be used to investigate the status of the temperature and power consumption of the tBase.  Diagnostics files When  sending  emails  to  support@millar.com  for  troubleshooting,  you  may  be  asked  to  send  a diagnostics file. To run diagnostics for the mouse telemeter or tBase, enter the Serial number and the Channel number in the “mouse telemeter” or “tBase” dialog, press the “Start Diagnostics” button and choose to save the file. ConfigSoft will ask you to confirm the name and location for the log file. The  default  filename  includes  the  telemeter  or  tBase  serial  number  and  the  date  and  time  the diagnostics was started (see example below). The log file and display are both updated every 1 s. A  count  of  the  updates  is  shown  in  the  bottom  left  hand  corner  of  the  window.  Diagnostics  will continue to run until you press “Stop Diagnostics”. The log file is then automatically saved. If you choose to save a log file you will not be able to stop the diagnostics for at least 2 minutes.   Please note that you cannot record data from your telemeter using a tBase that is in Diagnostics mode (the status light flashes red and a saw tooth pattern will output from the tBase analog outputs).  Mouse telemeter diagnostics  tBase diagnostics MTx10289_log_2015_6_9_16_11.txt Tb10253_log_2015_6_9_16_48.txt  Temperature normal  = the  temperature  of the tBase  is at acceptable levels.  =  the  temperature  is  above  acceptable  limits.  Please contact Millar (support@millar.com).  Power consumption normal  = tBase power consumption is normal.  = Excess power consumption has been detected.  The Status light on the front of tBase will be red.  The tBase may be too close to a metal surface or another tBase.
 Page 26  User Manual for Millar Mouse Telemetry MKT0006/A tBase Technical Specifications  tBase MT110 tBase functions Provides power only to mouse telemeters using inductive wireless charging, outputs standard analog voltage compatible with any data acquisition system (BNC), has a built-in fan for optimal temperature operation.  Weight 3.0 Kg Dimensions 400(w) x 450(d) x 65(h) mm / 15.7(w) x 17.7(d) x 2.6(h) in Cage platform dimensions 190(w) x 335(d) mm / 7.5(w) x 13.2(d) in Power input 100-240 V, 50-60 Hz Max Power draw 90W Temperature Operating Range 10 to 40 degrees Celsius Output Connectors BNC x 3 Output Voltage Range 0 to 4V Low Pass filtering on signal outputs Cut off frequency = 1000 Hz
 User Manual for Millar Mouse Telemetry Systems MKT006/A  Page 27 Telemeter Technical Specifications The technical specifications in the table below are for the mouse telemeters.    Biopotential Input range  +2.5 mV Biopotential Resolution   12 bit A/D Biopotential High pass characteristics  AC coupled, single pole, -3dB point at 2 Hz Biopotential Low pass characteristics  AC coupled, single pole, -3dB point at 440 Hz Biopotential Electrode leads  8cm length, coiled stainless steel Biopotential Electrode diameter  0.42 mm Temperature operating range   34 to 41 degrees Celsius Sampling frequency  2000 Hz Low Pass filtering by tBase  Cut off frequency = 1000 Hz Transmitted signal  Fully digital at 2.4 GHz Channels  40 transmission frequencies are available, user set Minimum animal weight  22 grams Outer material  Epoxy (telemeter body) and polyurethane (leadset) Volume  1.6 cc On-off mechanism  Powered when placed on the tBase and deactivated when removed from the tBase.  Calibration  No user intervention required (calibration values stored within each telemeter). Analog output calibration values Activity:   0V output = 0 RAU     3.84V output = 10 RAU Biopotential:   2.048V output = 0mV input,      4.096V output = 2.5 mV input Data received:   ~0V output = 0 (no data received),      ~3.3V = 1 (data received)
 Page 28  User Manual for Millar Mouse Telemetry MKT0006/A Frequently Asked Questions........  Would the animal be harmed by staying in the tBase powering field? There are no known biological effects of the inductive field.  Studies have shown that field strengths much higher than generated by the Millar tBase have no effect.  The level of magnetic field is quite low and remains useable only ~7 cm above the tBase.  How many animals can be monitored at one time?  Each telemeter sends its signal on one of 40 unique transmission Channels.  Therefore up to 40 telemeters and animals can be monitored at one time (on separate tBases).  Can I use the Millar Mouse Telemetry System alongside the Millar Rat Telemetry System?  Both the Millar Rat Telemetry System and Millar Mouse Telemetry System can be operated on any of  the  40  independent  data  collection  Channels  with  minimal  interference  between  animals.  Therefore, both systems can be operated side-by-side to monitor up to a total of 40 rats or mice, or a combination of both rats and mice, at any one time. Your existing rat Configurator system may require a firmware or software update before use with mouse telemeters (please contact Millar).  Is there any interference between mouse telemeters? No.  As long as each telemeter is set to use a different Channel they use separate transmission frequencies.  No shielding is required between cages.  Close proximity is possible.  Do I need to use special mouse cages with the tBase? Special caging is not required for use with this system.  The only limitation is that the cage base must sit on the surface of the tBase indent.  The indent of the tBase for the mouse cage is 190mm x 335mm and is compatible with any non-metallic cage bottom which fits these dimensions.     Contact Information  US Head Office Houston, Texas Toll free (in USA) 1 800 669 2343 TEL: +1 832 667 7000 FAX: +1 713 714 8497 New Zealand Office Auckland, New Zealand TEL: +64 9 367 7126 FAX: +64 9 367 7157  EMAIL  support@millar.com or sales@millar.com WEB  www.millar.com  Further questions? Millar is proud to offer unlimited technical support and advice to all its customers.  Register on our free online Knowledge Center (http://millar.com/knowledge-center) for surgical recommendations and hardware troubleshooting advice.  If you have any further questions, our team of engineers and  physiologists  can  advise  on  specific  applications  and  equipment  configurations.  Please contact us.

Navigation menu