Agilent Technologies Water Dispenser 8591E Users Manual 5963 6909e
8596E to the manual e33a60eb-27b3-4398-bbb2-90f9c208e3a1
2015-02-02
: Agilent-Technologies Agilent-Technologies-Agilent-Technologies-Water-Dispenser-8591E-Users-Manual-413557 agilent-technologies-agilent-technologies-water-dispenser-8591e-users-manual-413557 agilent-technologies pdf
Open the PDF directly: View PDF .
Page Count: 7
These specifications apply to the Agilent
Technologies 8591E, 8593E, 8594E, 8595E, and
8596E spectrum analyzers.
Specifications
All specifications apply over O°C to +55°C. The
analyzer will meet its specifications after 2 hours
of storage at a constant temperature, within the
operating temperature range, 30 minutes after the
analyzer is tumed on, and after CAL FREQ and
CAL AMPTD (and for the 8593E, 8595E, and
8596E CAL YTF) have been run.
Frequency Specifications
Frequency Range
8591E
50 Ω9 kHz to 1.8 GHz
75 Ω1 MHz to 1.8 GHz
8593E 9 kHz to 22 GHz
Option 026/027 9 kHz to 26.5 GHz
Band LO harmonic = N
0 1 9 kHz to 2.9 GHz
1 1 2.75 GHz to 6.5 GHz
2 2 6.0 GHz to 12.8 GHz
3 3 12.4 GHz to 19.4 GHz
4 4 19.1 GHz to 22.0 GHz
4 4 (Opt. 026/027) 19.1 GHz to 26.5 GHz
8594E
dc coupled 9 kHz to 2.9 GHz
ac coupled 100 kHz to 2.9 GHz
8595E
dc coupled 9 kHz to 6.5 GHz
ac coupled 100 kHz to 6.5 GHz
8596E
dc coupled 9 kHz to 12.8 GHz
ac coupled 100 kHz to 12.8 GHz
Band LO harmonic = N
0 1 9 kHz to 2.9 GHz (dc coupled)
0 1 100 kHz to 2.9 GHz (ac coupled)
1 1 2.75 GHz to 6.5 GHz
2 2 6.0 GHz to 12.8 GHz
Frequency Reference (Opt. 004)
Aging ±2 x 10-6/year ±1 x 10-7/year
Temperature Stability ±5 x 10-6 ±1 x 10-8
Initial Achievable Accuracy ±0.5 x 10-6 ±2.2 x 10-8
Agilent 8590 E-Series
Portable Spectrum Analyzers
Data Sheet
2
Frequency Readout
Accuracy
(Start, Stop, Center, Marker) ±(frequency readout x frequency
reference error1+span accuracy +1%
of span +20% of RBW+100 Hz x N*)
Marker Count Accuracy
Frequency Span
≤10 MHz x N* ±(marker frequency x frequency
reference error1+ counter resolution
+100 Hz x N*)
Frequency Span ±(marker frequency x frequency
>10 MHz x N* reference error1+ counter
resolution +1 kHz x N*)
Counter Resolution
Frequency Span
≤10 MHz x N* Selectable from 10 Hz to 100 kHz
Frequency Span
>10 MHz x N* Selectable from 100 Hz to 100 kHz
Frequency Span
Range 0 Hz (zero span), and
Opt. 130 Std.
Min. Min. Max
(KHz) (KHz) (GHz)
8591E 1 10 1.8
8593E 1 x N* 10 x N* 19.25
8594E 1 10 2.9
8595E 1 10 6.5
8596E 1 x N* 10 x N* 12.8
Resolution Four digits or 20 Hz x N*
whichever is greater
Accuracy
Span ≤10 MHz x N* ±2% of span
Span >10 MHz x N* ±3% of span
Frequency Sweep Time
Range
Span = 0 Hz, >1 kHz 20 ms to 100 s
Span = 0 Hz (Opt. 101) 20 µs to 100 s
Accuracy
20 ms to 100 s ±3%
20 µs to <20 ms (Opt. 101) ±2%
Sweep Trigger Free run, single, line, video,
external
Resolution Bandwidth 1 kHz to 3 MHz (3 dB) in 1-3-10
sequence.
9 kHz and 120 kHz (6 dB) EMI
bandwidths.
Option 130 Adds 30, 100, and 300 Hz (3 dB)
bandwidths and 200 Hz (6 dB) EMI
bandwidth.
Accuracy ±20%
Selectivity (Characteristic)
–60 dB/–3 dB
3 kHz to 10 kHz 15:1
100 kHz to 3 MHz 15.1
1 kHz, 30 kHz 16:1
–40 dB/–3 dB
30 Hz to 300 Hz 10:1
Video Bandwidth Range 30 Hz to 1 MHz in 1,3 sequence
1 Hz to 1 MHz (Opt 130)
Stability
Noise Sidebands (1 kHz RBW, 30 Hz VBW and sample detector)
>10 kHz offset from CW signal ≤–90 dBc/Hz + 20 Log N*
>20 kHz offset from CW signal ≤–100 dBc/Hz + 20 Log N*
>30 kHz offset from CW signal <–105 dBc/Hz + 20 Log N*
Residual FM
8591E
1 kHz RBW, 1 kHz VBW ≤250 Hz pk-pk in 100 ms
30 Hz RBW, 30 Hz VBW ≤30 Hz pk-pk in 300 ms
8593E, 94E, 95E, 96E
1 kHz RBW, 1 kHz VBW ≤(250 x N*) Hz pk-pk in
100 ms
30 Hz RBW, 30 Hz VBW ≤(30 x N*) Hz pk-pk in 300 ms
System-Related Sidebands
>30 kHz offset from CW signal ≤–65 dBc + 20 Log N*
Comb Generator Frequency
8593E, 96E I00 MHz fundamental
frequency
Accuracy ±0.007%
* N = LO harmonic. N = 1 for 91E, 94E, 95E
1. Frequency reference error = (aging rate x period of time since
adjustment + initial achievable accuracy + temperature
stability).
3
Amplitude Specifications
Amplitude specifications do not apply for Analog+
mode and negative peak detector mode except as
noted in “Amplitude Characteristics.”
Amplitude Range
Displayed average noise level
to +30 dBm
8591E (Opt. 001) Displayed average noise level to
+72 dBmV
Maximum Safe Input Level (input attenuator ≥10 dB)
Average Continuous Power +30 dBm (1 W)
8591E (Opt. 001) +72 dBmV (0.2 W)
Peak Pulse Power
8591E +30 dBm (1 W)
8591E (Opt. 001) +72 dBmV (0.2 W)
8593E, 94E, 95E, 96E +50 dBm (100 W) for < 10 µs pulse
width and <1 % duty cycle, input
attenuation ≥30 dB.
dc
8591E 25 Vdc
8591E (Opt. 001) 100 Vdc
8593E 0 Vdc
8594E, 95E, 96E 0 V (dc coupled)
50 V (ac coupled)
Gain Compression
>10 MHz ≤0.5 dB (total power at input mixer2 = –10 dBm)
Displayed Average Noise Level
(Input terminated, 0 dB attenuation, 1 Hz/30 Hz VBW, sample-
detector)
8591E 30 Hz RBW 1 kHz RBW
400 kHz to 1 MHz ≤–130 dBm ≤–115 dBm
1 MHz to 1.5 GHz ≤–130 dBm ≤–115 dBm
1.5 GHz to 1.8 GHz ≤–128 dBm ≤–113 dBm
8591E (Opt. 001)
1 MHz to 1.5 GHz ≤–78 dBmV ≤–63 dBmV
1.5 GHz to 1.8 GHz ≤–76 dBmV ≤–61 dBmV
8593E
400 kHz to 2.9 GHz ≤–127 dBm ≤–112 dBm
2.75 GHz to 6.5 GHz ≤–129 dBm ≤–114 dBm
6.0 GHz to 12.8 GHz ≤–117 dBm ≤–102 dBm
12.4 GHz to 19.4 GHz ≤–113 dBm ≤–98 dBm
19.1 GHz to 22 GHz ≤–107 dBm ≤–92 dBm
8593E (Opt. 026/027)
19.1 GHz to 26.5 GHz ≤–102 dBm ≤–87 dBm
8594E
400 kHz to<5 MHz ≤–122 dBm ≤–107 dBm
5 MHz to 2.9 GHz ≤–127 dBm ≤–112 dBm
8595E
400 kHz to 2.9 GHz ≤–125 dBm ≤–110 dBm
2.75 GHz to 6.5 GHz ≤–127 dBm ≤–112 dBm
8596E
400 kHz to 2.9 GHz ≤–125 dBm ≤–110 dBm
2.75 GHz to 6.5 GHz ≤–127 dBm ≤–112 dBm
6.0 GHz to 12.8 GHz ≤–115 dBm ≤–100 dBm
Sensitivity 30 Hz RBW
Third Order Intermod
Second Order Distortion
30
40
50
60
70
80
90
100
–70 –60 –50 –40 –30 –20 –10 0 +10
19.1 TO 22 GHz
6.0 TO 12.8 GHz
400 kHz TO 6.5 GHz
12.4 TO 19.4 GHz
>2.75 GHz
Nominal Dynamic Range
Agilent 8593E (Option 130)
Mixer Level (dBm)
Dynamic Range (dB)
(30 Hz Res BW)
10 MHz to 2.9 GHz
2. Mixer Power Level (dBm) = Input Power (dBm) Input Atten. (dB)
4
Spurious Responses
Second Harmonic Distortion
5 MHz to 1.8 GHz (91E) <–70 dBc for –45 dBm tone at
input mixer.2
10 MHz to 2.9 GHz (93E) <–70 dBc for –40 dBm tone at
>10 MHz (94E, 95E, 96E) input mixer.2
>2.75 GHz (93E, 95E,96E) <–100 dBc for –10 dBm tone at
input mixer2(or below displayed
average noise level).
Third Order Intermodulation
Distortion
5 MHz to 1.8 GHz (91E) <–70 dBc for two –30 dBm tones
>10 MHz (93E, 94E, at input mixer2and >50 kHz
95E, 96E) separation.
Other Input Related Spurious
≤1.8 GHz (91E) <–65 dBc at ≥30 kHz offset, for
≤2.9 GHz (94E) –20 dBm tone at input mixer2
≤6.5 GHz (95E)
≤12.8 GHz (96E)
≤18 GHz (93E)
≤22 GHz (93E) <–60 dBc at ≥30 kHz offset, for
–20 dBm tone at input mixer2
Residual Responses (input terminated and 0 dB attenuation)
1 MHz to 1.8 GHz
(91E Opt. 001) <–38 dBmV
150 kHz to 1.8 GHz (91E) <–90 dBm
150 kHz to 2.9 GHz (94E) <–90 dBm
150 kHz to 6.5 GHz <–90 dBm
(93E, 95E, 96E)
Display Range
Log Scale 0 to –70 dB from reference level is
calibrated; 0.1, 0.2, 0.5 dB/division
and 1 to 20 dB/division in 1 dB steps;
eight divisions displayed.
Linear Scale Eight divisions
Scale units dBm, dBmV, dBuV, V, and W
Marker Readout Resolution 0.05 dB for log scale
0.05% of reference level for linear
scale
Fast Sweep Times for Zero Span (Opt. 101 or 301)
20 µs to 20 ms
≤1 GHz 0.7% of reference level for linear
scale
>1 GHz 1.0% of reference level for linear
scale
Reference Level
Range same as amplitude range
Resolution 0.1 dB for log scale, 0.12% of
reference level for linear scale
Accuracy ±0.3 dB @ –20 dBm
0 dBm to –59.9 dBm ±(0.3 dB +.01 x dB from –20 dBm)
Frequency Response (10 dB input attenuation)
8591E Absolute3Relative Flatness4
9 kHz to 1.8 GHz ±1.5 dB ±1.0 dB
8593E Preselector peaked in band > 0
Absolute3Relative Flatness4
9 kHz to 2.9 GHz ±1.5 dB ±1.0 dB
2.75 GHz to 6.5 GHz ±2.0 dB ±1.5 dB
6.0 GHz to 12.8 GHz ±2.5 dB ±2.0 dB
12.4 GHz to 19.4 GHz ±3.0 dB ±2.0 dB
19.1 GHz to 22 GHz ±3.0 dB ±2.0 dB
19.1 GHz to 26.5 GHz ±5.0 dB ±2.0 dB
8594E, 95E, 96E (dc coupled preselector peaked)
Absolute3Relative Flatness4
9 kHz to 2.9 GHz ±1.5 dB ±1.0 dB
2.75 GHz to 6.5 GHz ±2.0 dB ±1.5 dB
6.0 GHz to 12.8 GHz ±2.5 dB ±2.0 dB
Calibrator Output
Amplitude –20 dBm ±0.4 dB
8591E Opt.001 +28.75 dBmV ±0.4 dB
Resolution Bandwidth
Switching Uncertainty
(Referenced to 3 kHz RBW, at ref. level)
3 kHz to 3 MHz RBW ±0.4 dB
1 kHz RBW ±0.5 dB
30 Hz to 300 Hz RBW ±0.6 dB
Linear to Log Switching ±0.25 dB at reference level
Display Scale Fidelity
Log Maximum Cumulative
0 to –70 dB from reference level
3 kHz to 3 MHz RBW ± (0.3 + 0.01 x dB from reference
level)
30 Hz to 1kHz RBW ± (0.4 + 0.01 x dB from reference
level)
Log Incremental Accuracy ±0.4 dB/4 dB
0 to –60 dB from reference level
Linear Accuracy ±3% of reference level
3. Referenced to 300 MHz CAL OUT.
4. Ref. to midpoint between highest and lowest freq.
response deviations.
5
Option Specifications
Option 010 and 011 Tracking Generator
Frequency Range
8591E 100 kHz to 1.8 GHz
(Opt. 011, 75 Ω)1 MHz to 1.8 GHz
8593E, 94E, 95E, 96E 9 kHz to 2.9 GHz
Output Level
Range
8591E 0 to –70 dBm
8591E (Opt. 011) +42.8 to –27.2 dBmV
8593E, 94E, 95E, 96E –1 to –66 dBm
Resolution 0.1 dB
Absolute Accuracy
(@ 300 MHz, –20 dBm, +28.8 dBmV)
8591E ±1.0 dB
8593E, 94E, 95E, 96E ±0.75 dB
Vernier
Range
8591E 10 dB
8593E, 94E, 95E, 96E 9 dB
Accuracy
8591E ±0.75 dB
8593E, 94E, 95E, 96E ±0.5 dB
Output Attenuator
Range
8591E 0 to 60 dB, 10 dB steps
8593E, 94E, 95E, 96E 0 to 56 dB, 8 dB steps
Output Flatness
8591E ±1.75 dB
8593E, 94E, 95E, 96E
(>10 MHz) ±2.0 dB
Effective Source Match (Characteristic)
8591E 1.6:1 (10 dB attenuation)
8593E, 94E, 95E, 96E 1.5:1 (8 dB attenuation)
Spurious Output
Harmonic Spurs
8591E
(0 dBm, +42.8 dBmV output) <–25 dBc
8593E, 94E, 95E, 96E
(–1 dBm Output)
Nonharmonic Spurs
8591E <–30 dBc
8593E, 94E, 95E, 96E
300 kHz to 2.0 GHz ≤–27 dBc
2.0 GHz to 2.9 GHz ≤–23 dBc
Dynamic Range (Characteristic)
Dynamic Range5TG Feedthrough
8591E 106 dB ≤–106 dBm
8591E (Opt. 011) 100 dB ≤–57.24 dBmV
8593E (> 400 kHz) 111 dB ≤–112 dBm
8594E (> 400 kHz) 106 dB ≤–107 dBm
(> 5 MHz) 111 dB ≤–112 dBm
8595E (>400 kHz) 109 dB ≤–110 dBm
8596E (> 400 kHz) 109 dB ≤–110 dBm
Power Sweep
Range
8591E (–15 dBm to 0 dBm) –(source
attenuator setting)
8591E (Opt 011) (+27.8 to 42.8 dBmV)–(source
attenuator setting)
8593E, 94E, 95E, 96E (–10 dBm to –1 dBm)–(source
attenuator setting)
Resolution 0.1 dB
Option 103 Quasi-Peak Detector
Amplitude response conforms with Publication 16 of Comité
International Spécial des Perturbations Radioélectriques (CISPR)
Section 1, Clause 2.
Option 105 Time Gated Spectrum Analysis
Gate Delay
Range 1 µs to 65.535 ms
Resolution 1 µs
Accuracy ±(1 µs + 0.01% x Gate Delay Readout)6
(From Gate Trigger Input to positive edge of Gate Output)
Gate Length
Range 1 µs to 65.535 ms
Resolution 1 µs
Accuracy ±(0.2 µs + (0.01% x Gate Length
Readout))
(From positive edge to negative edge of Gate Output)
Additional Gate Amplitude Error7
Log Scale <2 µs ±0.8 dB
≥2 µs ±0.5 dB
General Specifications
Temperature Range
Operating 0°C to +55°C
Storage –40°C to +75°C
EMI Compatibility Conducted and radiated
interference CISPR Pub. 11
and Messempfaenger
Postverfuegung 526/527/79.
Audible Noise <37.5 dBa pressure and <5.0 Bels
power (ISODP7779)
5. Maximum output level minus TG feedthrough.
6. Up to 1 Vs jitter due to 1 µs resolution of gate delay clock.
7. With GATE ON enabled and triggered, CW Signal, Peak Detector Mode.
6
Power Requirements
ON (Line 1) 90 to 132 V rms, 47 to 440 Hz
195 to 250 V rms, 47 to 66 Hz
Power consumption <500 VA;
<180W
Standby (Line 0) Power consumption <7 W
User Program Memory 238 Kbytes non-volatile RAM
Data Storage (nominal)
Internal 24 traces or 32 states
External 50 traces, 8 states
Memory card (85700A) 32 Kbytes
lnputs/Outputs
Front Panel Connectors
Input 50 ΩType N
(Opt 001) 75 ΩBNC female
(Opt 026) APC 3.5 mm male
(Opt 027) 50 ΩType N female
Cal Output 50 ΩBNC, –20 dBm, 300 MHz
100 MHz Comb Out 100 MHz ±0.007%, SMA
Probe Power +15 Vdc, –12.6 Vdc, and Gnd
(150mA max each)
Rear Panel Connectors
Earphone (Opt 102 and 103) 1/8 inch monaural jack
LO Output (Opt 009) 50 ΩSMA Female, 3.0 to 6.8214 GHz
TV Trigger Output BNC, TTL levels, negative edge
(Opt 101 and 102) trigger after sync pulse
Gate Trigger Input (Opt 105) 50 ΩBNC, Pulsewidth >30 ns (TTL)
Gate Output (Opt 105) 50 ΩBNC (TTL)
SWEEP + Tune Output
(Opt 009) 2 k ΩBNC, 0 to +10V,
0.36V/GHz of CF
Ext. ALC Input 1 MW, –66 dBV to +6 dBV
Sweep Output BNC, 5 k W, 0 to +10 V ramp
High Sweep In/Out BNC, high TTL = sweep,
low TTL = Retrace
Aux Video Out 50 ΩBNC, 0-1 V Uncalibrated
Aux IF Output 50 ΩBNC, –10 to –60 dBm,
21.4 MHz
Keyboard (Opt. 041 or 043) 5 Pin mini-DIN, compatible with
HP C1405B and most IBM/AT key
boards
Ext. Trigger Input BNC, TTL levels, positive edge
trigger
GPIB and Parallel (Opt 041) SH1, AH1, T6, L4, ST1, RL1, PPO,
DC1, C1 C2, C3, & C28 and 25 Pin
subminature D-shell female for
parallel
RS-232 and Parallel (Opt 043) 9 Pin subminature D-shell female
and 25 Pin subminature D-shell
female for parallel
Ext Ref In 50 ΩBNC, 10 MHz, –2 to +10 dBm
10 MHz Ref Output 50 ΩBNC, 10 MHz, 0 dBm
Aux Interface 9 pin “D” subminiature
Pin 1-4, TTL Output
Pin 5 TTL Input
Pin 6 Gnd
Pin 7 –15 vdc ±5%; 150 mA max
Pin 8 +5 vdc ±5%; 150 mA max
Pin 9 +15 vdc ±5%; 150 mA max
Monitor Out 50 ΩBNC,
Selectable Format NTSC, 15.75 kHz, 60 Hz
PAL, 15.625 kHz, 50 Hz
Dimensions (Nominal)
(Without handle, feet, 163 mm (H) x 325 mm (W)
or cover) x 427 mm (D)
(Overall) 184 mm (H) x 373 mm (W)
X 461 mm (D)
Weight (Nominal)
8591E 15.4 kg (34 lb)
8593E 16.4 kg (36 lb)
8594E 16.4 kg (36 lb)
8595E 16.4 kg (36 lb)
8596E 16.4 kg (36 lb)
Related Literature Pub. Number
General Purpose Information
Agilent 8590L and 8592L
Product Overview 5962-7275E
Agilent 8590C/E/L and EM Series
Configuration Guide 5963-6858E
Agilent 8590 E-Series Brochure 5963-6908E
Product Feature Briefs
Analog + Display 5091-4054E
Transmitter Power Measurements
(ACP, OBW) 5091-4055E
Zoom Window 5091-4051E
Measuring AM with FFT 5091-4049E
Time Gated Spectrum Analysis 5091-4053E
Editing Keyboard 5091-4048E
Marker and Peaks Table 5091-4050E
Third Order Intermodulation,
N & B Bandwidth, and Percent AM 5091-4052E
Product Notes
Time-Gated Spectrum Analysis
(Agilent 8590-2) 5952-3685
Analog + Display 5091-2364E
Maximizing Accuracy in Noise Figure
Measurements (Agilent 85791-1) 5091-4801E
Application Notes
Spectrum Analysis Basics (150) 5952-0292
Amplitude and Frequency
Modulation (150)-1) 5954-9130
Agilent Technologies’ Test and Measurement
Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive,
while minimizing your risk and problems. We strive to ensure
that you get the test and measurement capabilities you paid
for and obtain the support you need. Our extensive support
resources and services can help you choose the right Agilent
products for your applications and apply them successfully.
Every instrument and system we sell has a global warranty.
Support is available for at least five years beyond the produc-
tion life of the product. Two concepts underlie Agilent’s
overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
“Our Promise” means your Agilent test and measurement equip-
ment will meet its advertised performance and functionality.
When you are choosing new equipment, we will help you with
product information, including realistic performance specifica-
tions and practical recommendations from experienced test
engineers. When you use Agilent equipment, we can verify that
it works properly, help with product operation, and provide
basic measurement assistance for the use of specified capabili-
ties, at no extra cost upon request. Many self-help tools are
available.
Your Advantage
“Your Advantage” means that Agilent offers a wide range of
additional expert test and measurement services, which you
can purchase according to your unique technical and business
needs. Solve problems efficiently and gain a competitive edge
by contracting with us for calibration, extra-cost upgrades, out-
of-warranty repairs, and on-site education and training, as well
as design, system integration, project management, and other
professional services. Experienced Agilent engineers and tech-
nicians worldwide can help you maximize your productivity,
optimize the return on investment of your Agilent instruments
and systems, and obtain dependable measurement accuracy
for the life of those products.
Get assistance with all your test and measurement needs at:
www.agilent.com/find/assist
Product specifications and descriptions in this
document subject to change without notice.
Copyright © 1996, 2000 Agilent Technologies
Printed in U.S.A. 5/00
5963-6909E