Alcon Research NGPFSW1 Centurion Vision System Footswitch User Manual

Alcon Research Ltd. Centurion Vision System Footswitch

User Manual

Operator's ManualManufacturer:Alcon Laboratories, Inc.  Alcon Laboratories (UK) Ltd.6201 South Freeway  Frimley Business ParkFort Worth, Texas 76134-2099  Frimley, CamberleyU.S.A.  Surrey, GU16 7SR, United Kingdom Produced By:Alcon Research, Ltd.15800 Alton ParkwayIrvine, California 92618-3818U.S.A.Telephone: 949/753-1393 800/832-7827FAX: 949/753-66148065751772, CATALOG NUMBER905-2150-001 P4, TEXT ONLY © 2012 NovartisDirective 93/42/EECEC  REP
ii  8065751772Centurion® Vision System Operator's Manual8065751772MANUAL REVISION RECORD  DATE  REVISION  ECN NUMBER AND DESCRIPTIONMarch 2011  P2  2011XXXX - Preliminary release of Centurion® Vision System Operator's Manual with catalog number 8065751772, and 905-2150-001 text (applies to Centurion® Vision System consoles with software version X.00).June, 2012  P3  2012XXXX - On page 1.23 changed Maximum Input Current in Table 1-3 from 6A to 10A. On page 1.25 changed amperage on label from 12A to 10A. Text Cover Sheet and pages i & ii also updated.June, 2012  P4  2012XXXX - On page 1.25 added IPX1 to remote control label. Text Cover Sheet and pages i & ii also updated.END USER LICENSE AGREEMENT:This product contains software licensed from Microsoft Corporation.* Registered in the U.S. Patent & Trademark Ofce.** Mackool is a trademark of Richard J. Mackool, M.D.Cycoloy and Lexan are registered trademarks of Sabic Innovative Plastics IP
vi  8065751772PREFACEThis operator's manual is your written guide to the Centurion® Vision System and considers all options available to the customer; therefore, when reading this manual, ignore the options which do not apply to your specic unit.Please read the entire manual carefully before operating the instrument. Recommended settings are given only as guidelines, and are not meant to restrict the surgeon; however, before trying other settings, the surgeon and support personnel should be experienced with the system and familiar with the new settings.NOTE: If an inconsistency exists between the instructions in the operator's manual and the Directions For Use (DFU) supplied with a consumable pack or accessory, follow the DFU.Equipment improvement is an on-going process and, as such, changes may be made to the equipment after this manual is printed.Pay close attention to Warnings, Cautions, and Notes in this manual. A WARNING! statement is written to protect individuals from bodily harm. A Caution statement, with the CAUTION heading centered above the text, is written to protect the instrument from damage. A NOTE: is written to bring attention to highlighted information. If you have questions, or want additional information, please contact your local Alcon representative or the Alcon Technical Services Department at:Alcon Research, Ltd.15800 Alton ParkwayIrvine, California 92618(949) 753-1393FAX (949) 753-6614CAUTION: U.S. Federal Law restricts this device to sale by or on the order of a physician.
8065751772 1.3GENERAL INFORMATION The Centurion® Vision System is indicated for emulsication, separation, irrigation, and aspiration of cataracts, residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intra-ocular lens injection. The AutoSert® IOL Injector Handpiece is intended to deliver qualied AcrySof® intraocular lenses into the eye following cataract removal. The AutoSert® IOL Injector Handpiece achieves the functionality of  injection of intraocular lenses. The AutoSert® IOL Injector Handpiece is indicated for use with AcrySof® lenses SN60WF, SN6AD1, SN6AT3 through SN6AT9, as well as approved AcrySof® lenses that are specically indicated for use with this inserter, as indicated in the approved labeling of those lenses. The Centurion® Vision System, including accessories approved by Alcon, constitutes a complete surgical system and is intended exclusively for use by licensed ophthalmic surgeons and their surgical teams. These surgical teams are experienced at conducting phacoemulsication procedures in a properly maintained surgical environment (qualied personnel, availability of backup equipment) and are familiar with the operation of the equipment used as outlined in operator's manuals and directions for use (setup/checkout procedures to be completed before the surgical procedure; processing of reusable devices; maintenance; etc.).  Patient selection for use with the Centurion® Vision System (such as age, ophthalmic pathology, and other factors) is determined by the surgeon. The general patient age can range from newborn to geriatric, although there have been studies that have identied the mean age of patients that underwent cataract surgery was 72.32 yrs  - men and 74.89 yrs  - women.1 Intended Use Environments The Centurion® Vision System is intended for use in hospitals and ambulatory surgery centers. Phaco Handpiece Note  Throughout the rest of this manual the CENTURION® OZil® handpiece and the INFINITI® OZil® handpiece will be referred to as phaco handpieces, unless one or the other must be referred to exclusively.  Trademark Note  A button, mode, or step labeled OZil®, AutoSert®, or UltraChop refers to a display screen control used with a phaco handpiece, INTREPID® AutoSert® IOL injector, or ALCON® UltraChopper® tip, respectively.  Abbreviation Descriptions  Many of the abbreviations used in this manual and on the Centurion® Vision System are described in Table 1-x. Icons are identied in Figure 1-x.  Accessory Equipment  Accessory equipment connected to or used with this equipment must be certied according to the respective IEC Standard (e.g., IEC 60950-1 for data processing equipment, and IEC 60601-1 for medical equipment). Furthermore, all congurations shall comply with clause 16 of IEC 60601-1:2005 (as amended). Anyone connecting additional equipment or otherwise causing a different system conguration than provided by Alcon is responsible for continued compliance to the requirements of clause 16 of IEC 60601-1:2005 (as amended). If in doubt, consult the Technical Services department or your local Alcon representative.  Follow local governing ordinances and recycling plans regarding disposal or recycling of device components and packaging.1. “Age and sex prole of patients having cataract surgery between 1986 and 2003”Philip O'Reilly, FRCSI (Ophth), U. Mahmound, FRCOphth, P. Hayes, FRCOphth, P. Tormey, FRCOphth, S. Beatty, MD.Journal of Cataract Refractive Surgery 2005; 31:2162-2166
1.4  8065751772  User Information – Environmental Considerations    The equipment that you have purchased requires the use of natural resources for its production and operation. This equipment may also contain hazardous substances which could have potential effect on the environment and human health if disposed of improperly.    In order to avoid the entry of any such substances into our environment, and to promote natural resource conservation, please install, maintain, and operate the equipment in accordance with the instructions. Information on the location of hazardous substances, resource consumption and emissions of the equipment can be found throughout this Operator's Manual. Please use the appropriate take-back systems. Such take-back systems reuse or recycle many of the materials in your end-of-life equipment in a benecial way. Please contact your local Alcon ofce for assistance in take-back options through Alcon or other providers.  The crossed-bin symbol located on this equipment reminds you to use take-back systems, while also emphasizing the requirement to collect waste equipment separately, and not dispose of it as unsorted municipal waste. The Pb notation, if present, indicates that the labeled device contains greater than 0.004% lead.  If you need more information on the collection, reuse or recycle systems available to you, please contact your local or regional waste administration, or contact your local Alcon ofce for more information.   Universal Precautions    Universal precautions shall be observed by all people who come in contact with the instrument and/or accessories to help prevent their exposure to blood-borne pathogens and/or other potentially infectious materials. In any circumstance, wherein the exact status of blood or body uids/tissues encountered are unknown, it shall be uniformly considered potentially infectious and handled in accordance with OSHA or your own national guidelines.  EMC Statements    It is important to install and use the equipment in accordance with the instructions in order to prevent harmful interference with other devices in the vicinity. If this equipment causes harmful interference to other devices (determined by turning equipment off and on), the user is encouraged to try to correct interference by one or more of the following measures:   •  Reorient or relocate the other device(s).   •  Increase the distance between the equipment.   •  Connect this equipment into an outlet on a circuit different from that to which the other device(s) is connected.  •  Consult the manufacturer or your Alcon eld service engineer for help.Pb
8065751772 1.5Table 1-1 Guidance and Manufacturer's Declaration - Electromagnetic Emissions - The Centurion® Vision System is intended for use in the electromagnetic environment specied below. The customer or the user of the Centurion®  Vision System should assure that it is used in such an environment.Emissions TestRF emissionsCISPR 11RF emissionsCISPR 11Harmonic emissionsIEC 61000-3-2Voltage uctuations/Flicker emissionsIEC 61000-3-3Electromagnetic Environment-GuidanceThe Centurion®  Vision System uses RF energy only for its internal function.  Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.The Centurion®  Vision System is suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network that supplies buildings used for domestic purposes.The  EMC  Statement  provides  guidance  on  steps  to  take  in  case  of electromagnetic interference.ComplianceGroup 1Class  AClass AComplies  Users should be aware of known RF sources, such as radio or TV stations and hand-held or mobile two-way radios, and consider them when installing a medical device or system. Portable and mobile RF communications equipment such as cellular telephones can affect medical electrical equipment (see Table 1-3 for recommended separation distances).  Be aware that adding accessories or components, or modifying the medical device or system, may degrade the EMI performance. Consult with qualied personnel regarding changes to the system conguration.WARNINGS!The  use  of  accessories,  transducers,  and  cables  other  than  those  specied,  with the exception of transducers and cables sold by the manufacturer of the system as replacement parts for internal components, may result in increased emissions or decreased immunity of the system.The system should not be used adjacent to, or stacked with, other equipment; and that if adjacent to or stacked use is necessary, the system should be observed to verify normal operation in the conguration in which it will be used.MAGNETIC  AND  ELECTRICAL  INTERFERENCE  -  Magnetic  and  electrical  elds  are capable of interfering with the proper performance of the device. For this reason make sure  that  all  external  devices  operated  in  the  vicinity  of  the  device  comply  with  the relevant EMC requirements. X-ray equipment, magnetic resonance tomography (MRT), nuclear magnetic resonance (NMR), or magnetic resonance imaging (MRI) devices are possible sources of interference as they may emit higher levels of electromagnetic radiation. See the Magnetic Resonance Unsafe icon in Figure 1-2.
1.6  8065751772Table 1-2  Guidance and Manufacturer's Declaration - Electromagnetic Immunity - The Centurion® Vision System is intended for use in the electromagnetic environment specied below. The customer or the user of the Centurion®  Vision System should assure that it is used in such an environment.Note:  UT is the a.c. mains voltage prior to application of the test level.Note 1:  At 80 MHz and 800 MHz, the higher frequency range applies.Note 2:  These guidelines may not apply in all situations.  Electromagnetic propagation is affected by  absorption  and  reection  from structures, objects, and people.a  Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy.  To access the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered.  If the measured eld strength in the location in which the (equipment or system) is used exceeds the applicable RF compliance level above, the (equipment or system) should be observed to verify normal operation.  If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Centurion® Vision System.b  Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.Immunity TestElectrostatic discharge (ESD)IEC 61000-4-2Electrical fasttransient/burstIEC 61000-4-4SurgeIEC 61000-4-5Voltage dips, short interruptions, and voltage variations on power supply input linesIEC 61000-4-11Power frequency (50/60 Hz) magnetic eldIEC 61000-4-8Conducted RFIEC 61000-4-6Radiated RFIEC 61000-4-3IEC 60601 Test Level•  ±6 kV contact•  ±8 kV air•  ±2 kV for power supply lines•  ±1 kV for input/output lines•  ±1 kV differential mode•  ±2 kV common mode•  < 5 % UT (> 95 % dip in UT) for 0.5 cycle•  40 % UT (60 % dip in UT) for 5 cycles•  70 % UT (30 % dip in UT) for 25 cycles•  < 5 % UT (> 95 % dip in UT) for 5 s3 A/m3 Vrms150 kHz to 80 MHz3 V/m80 MHz to 2.5 GHzCompliance Level•  ±6 kV contact•  ±8 kV air•  ±2 kV for power supply lines•  ±1 kV for input/output lines•  ±1 kV differential mode•  ±2 kV common mode•  < 5 % UT (> 95 % dip in UT) for 0.5 cycle•  40 % UT (60 % dip in UT) for 5 cycles•  70 % UT (30 % dip in UT) for 25 cycles•  < 5 % UT (> 95 % dip in UT) for 5 s3 A/m3 Vrms3 V/mElectromagnetic Environment-GuidanceFloors should be wood, concrete, or ceramic tile. If oors are covered with synthetic material, the relative humidity should be at least 30 %.Mains power quality should be that of a typical hospital (including ambulatory surgery center) environment. To avoid pre-mature shutdown due to fast transients avoid powering the Centurion®  Vision System on the same branch circuit with sources that can generate fast transients (inductive switching; e.g., high current motors).Mains power quality should be that of a typical hospital (including ambulatory surgery center) environment.Mains power quality should be that of a typical hospital (including  ambulatory  surgery  center)  environment. If  the  use  of  the  Centurion®  Vision  System  requires continued operation during power mains interruptions, it is  recommended  that the Centurion®  Vision  System  be powered from an uninterruptible power supply or a battery.Power  frequency  magnetic  elds  should  be  at  levels characteristic  of  a  typical  location  in  a  typical  hospital (including ambulatory surgery center) environment.Portable and mobile RF communications equipment should be used no closer to any part of the Centurion®  Vision System, including cables, than the recommended separation distance calculated from the  equation  applicable  to  the frequency to the transmitter.Recommended separation distance:d = 1.2√Pd = 1.2 √ P  80 MHz to 800 MHzd = 2.3 √ P  800 MHz to 2.5 GHzwhere    P    is  the  maximum  output  power  rating  to  the transmitter  in  watts  (W)  according  to  the  transmitter manufacturer  and  d    is  the  recommended  separation distance in meters (m).Field strength from xed RF transmitters, as determined by an electromagnetic site surveya, should be less than the compliance level in each frequency rangeb.Interference may occur in the vicinity ofequipment marked with following symbol.
8065751772 1.7Table 1-3  Recommended  Separation  Distances  Between  Portable  and  Mobile  RF  Communications Equipment and the Centurion® Vision System - The Centurion® Vision System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Centurion® Vision System can help prevent electromagnetic interference  by  maintaining  a  minimum  distance  between  portable  and  mobile  RF communications equipment (transmitters) and the Centurion® Vision System as recommended below, according to the maximum output power of the communications equipment.For transmitters rates at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.Note 1 -  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.Note 2 -  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reection from structures, objects, and people.800 MHz to 2.5 GHzd = 2.3 √ P 0.230.732.37.32380 MHz to 800 MHzd = 1.2 √ P 0.120.381.23.812150 kHz to 80 MHzd = 1.2 √ P 0.120.381.23.812Rated maximum outputpower of transmitter(W) 0.010.1110100Separation distance according to frequency of transmitter(m)
1.8  8065751772  Equipment Contains Radio Transmitters  •  ZigBee Radio Modular (Communication link with Footswitch, HUD and Media Center)    -  Frequency or frequency band of transmission: 2.405 – 2.480 GHz    -  Type and frequency characteristics of the modulation: OQPSK (Offset quadrature phase-shift keying)    -  The Effective Radiated Power (ERP): 12.91 dBm (19.54 mW)  •  Wireless LAN device (Optional)   - Frequency or frequency band of transmission: 2.412 – 2.484 GHz and 5.180 - 5.700 GHz    -  Type and frequency characteristics of the modulation: OFDM, DSSS, CCK, DQPSK, DBPSK, 64 QAM, 16 QAM    -  The Effective Radiated Power (ERP): 17.09 dBm (51.17 mW)  •  Wireless Footswitch Charger    -  Frequency or frequency band of charging transmission: 50 kHz    -  Frequency or frequency band communication transmission: 115 kHz    -  Type and frequency characteristics of the modulation: FSK (Frequency Shift Keying)    -  The Effective Radiated Power (ERP): -14.89 dBm (53.18 μW)  USA – Federal Communications Commission (FCC)    This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.CAUTION  Change or modications made to this equipment (including antenna) not expressly approved by Alcon may void the FCC authorization to operate this equipment.    FCC Radiation Exposure StatementCAUTION  To ensure that the radio transmitter complies with current FCC regulations limiting both maximum output RF power and human exposure to radio frequency radiation, a separate distance of at least 20 cm must be maintained between the unit’s antenna and the body of the user and any nearby persons at all times, and unit’s antenna must not be co-located or operating in conjunction with any other antenna or transmitter.
8065751772 1.9 Canada – Industry of Canada (IC)    This device complies with Industry Canada licence-exempt RSS standards. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Cet appareil est conforme aux normes d’Industrie Canada RSS exemptes de licence. Son fonctionnement est soumis aux deux conditions suivantes: (1) Cet appareil ne doit pas provoquer d’interférences nuisibles, et (2) cet appareil doit accepter toute interférence, y compris les interférences pouvant provoquer un fonctionnement indésirable de l’appareil.  Transmitter Antenna:    Under Industry Canada regulations, this radio transmitter may only operate using an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically radiated power (e.i.r.p.) is not more than that necessary for successful communication. Conformément à la réglementation de l’industrie du Canada, cet émetteur de radio ne peut être utilisé qu’avec un type d’antenne approuvé pour l’émetteur par Industrie Canada et seulement avec une valeur de gain inferieur ou égale au gain maximum approuvé par Industrie Canada. Pour réduire les risques potentiels d’interférence à autrui, le type d’antenne et son gain doivent être choisis de sorte que la puissance isotrope rayonnée équivalente (p.i.r.e.) ne dépasse pas la valeur qui est nécessaire pour une communication réussi.    The radio transmitters, contained in the system, have been approved by Industry Canada to operate with the antenna types listed below with the maximum permissible gain and required antenna impedance for each antenna type indicated. Antenna types not included in this list, having a gain greater than the maximum gain indicated for that type, are strictly prohibited for use with this device. Les émetteurs radio, contenues dans le système, ont été approuvés par Industrie Canada pour fonctionner avec les types d’antenne énumérés ci-dessous et ayant un gain admissible maximal et l’impédance requise pour chaque type d’antenne. Les types d’antenne non inclus dans cette liste, ou dont le gain est supérieur au gain maximal indiqué, sont strictement interdits pour l’exploitation de l’émetteur. Approved Antennas  Part Number  Manufacturer  Description  Min. Separation  WPC25A  Taoglas  2.4 GHz Ceramic Patch Antenna  20 cm  GW.71.5153  Taoglas  2.4 / 5.8 GHz Dipole Antenna  20 cm  ANTB98-061A0  Sansei Denki  Sleeve antenna 2.4 / 5 GHz  20 cm  Exposure of Humans to RF Fields:    This device complies with the RF exposure limits for humans as called out in RSS-102. Cet appareil est conforme aux limites d’exposition RF pour les êtres humains comme elles le sont notiées dans la norme RSS-102.
1.10  8065751772Table 1-4  Information on the Location of Hazardous Substances in the Centurion® Vision System - The Centurion® Vision System contains hazardous substances which could have potential effect on the environment and human health if disposed of improperly.  Material Location  Hazardous Substances ContainedPrinted Circuit Board Assembly  Lead, Polybrominated Biphenyls (PBB)Other Electrical / Electronic Device  Lead, Polybrominated Biphenyls (PBB)Cable Assembly  LeadPower Supply  Lead, Polybrominated Biphenyls (PBB)Host PC Module  Lead, Polybrominated Biphenyls (PBB)Liquid Crystal Display  LeadBattery  Lead, Lithium, Zn/MnO2IV Pole Assembly  Lead, Polybrominated Biphenyls (PBB)Remote Control  LeadFluidics Assembly  LeadPneumatic Assembly  Lead  Europe – R&TTE Directive 99/5/EC  This device complies with the requirements of the Council Directive 99/5/EC (R&TTE).CAUTION    The radio equipment is intended to be used in all EU and AFTA countries. Outdoor use may be restricted to certain frequencies and/or may require a license for operation. Contact local Authority for procedure to follow.    NOTE: Combinations of power levels and antennas resulting in a radiated power of above 100 mW equivalent isotropic radiated power (e.i.r.p) are considered as not compliant with the above mentioned directive and are not allowed for use within the European community and countries that have adopted the European R&TTE directive 1999/5/EC.    For more details on legal combinations of power levels and antennas, contact Alcon Compliance. Japan    This device complies with ARIB STD-66 Radio Standard in Japan.
8065751772 1.11WARNINGS AND CAUTIONS Many of these warnings are stated elsewhere in this manual; however, for easy reference they are repeated in greater detail here. If additional information is required, please contact your local Alcon service representative, or the Technical Services Department. There are no user serviceable components inside the Centurion® Vision System console or footswitch. Refer all service issues to your factory-trained Alcon service engineer.WARNINGS!The Centurion® Vision System battery can only be serviced by a factory-trained Alcon service engineer. Access by untrained personnel can lead to injury.A qualied technician must perform a visual inspection of the following components every twelve months:  •  Warning Labels (see section one of this manual)  •  Power Cord  •  FusesIn case of a deciency, do not use the system; call Alcon Technical Services.A qualied technician must check ground continuity and leakage current every twelve months to ensure they are within the limits of the applicable standards (for example: EN60601-1/IEC60601-1). Values must be recorded, and if they are above the limits of the applicable standards, or 50 % above initial measurement, do not use the system; call Alcon Technical Services.If the Centurion® Vision System is used at the 220 V - 240 V range in the United States or Canada, it should be used on a center-tapped, 240 V single phase circuit.Console isolation from mains is achieved through a two pole power switch. Turn OFF power switch or unplug the power cord from wall outlet to achieve isolation from mains.•  Do not use the Centurion® Vision System near ammable anesthetics.Do not exceed maximum capacity of drain bag (500 ml). Excessive pressure can result from exceeding drain bag maximum capacity and potentially result in a hazardous condition for the patient.Inadvertent actuation of Prime or Tune while a handpiece is in the eye can create a hazardous condition that may result in patient injury.Keep clear of display base when raising display from stored position to prevent skin, hair, and /or clothing from being trapped at the base.The maximum allowable load on the instrument tray is 20 lb. (9 kg).Place the instrument tray in the stored position prior to transportation to avoid a situation that could cause the system to tip.Console might overbalance when it is pushed and its wheels are immobilized (blocked).Route  the  footswitch  cable,  power  cord  and  any  other  cables  connected  to  the Centurion® Vision System to avoid tripping.
1.12  8065751772WARNINGS!Appropriate use of Centurion® Vision System parameters and accessories is important for successful procedures. Use of low vacuum limits, low ow rates, low bottle  heights,  high  power  settings,  extended  power  usage,  power  usage  during occlusion  conditions  (beeping  tones),  failure  to  sufciently  aspirate  viscoelastic prior to using power, excessively tight incisions,  and  combinations  of  the  above actions may result in signicant temperature increases at incision site and inside the eye, and lead to severe thermal eye tissue damage.Good clinical practice dictates testing for adequate irrigation, aspiration ow, reux, and operation as applicable for each handpiece prior to entering eye.Ensure that the tubings are not occluded during any phase of operation.If the handpiece test chamber is collapsed after tuning, there is a potential of low irrigation ow through the handpiece and may result in a uidic imbalance. This, in turn, may cause a shallowing or collapsing of the anterior chamber.Avoid setting the patient above the FMS unless PEL is used. Operating with the patient above the FMS without PEL adjustment will result in a lower irrigation pressure than indicated on the display, and possible underventing.Use of BSS® irrigating uid bags other than those approved by Alcon for use in the active uidics system can result in patient injury or system damage.Use of appropriate technique and settings is important to minimize fragments and turbulence.Do not remove the FMS during the surgical procedure.In the event of a system error release footswitch to the up position.Improper handling or removal of dual irrigation handpiece tip from eye may cause draining of the uidics system.CAUTIONS•  Modication of the equipment is NOT allowed without prior authorization from the manufacturer. If this equipment is modied, appropriate inspection and testing must be conducted to ensure continued safe use of the equipment.•  Avoid spilling BSS® irrigating solution, or moisture of any kind, around the electrical handpiece connectors.•  Do not spray any liquid (i.e. cleaning solution or water) upward into the console vents.•  Do not push or pull the unit by the display, the tray, or the IV pole. Wrapping around the rear and sides of the system is a handle provided for moving the instrument. The unit should be pulled and not pushed, especially over elevator and door thresholds.
8065751772 1.13 Handpiece Care  Phaco handpieces are surgical instruments and must be handled with care. The handpiece tip should not touch any solid object while in operation. Immediately following surgery the handpiece must be thoroughly cleaned. Be sure handpiece connector is completely dry before connecting it to console. For cleaning and sterilization procedures, see the Directions for Use (DFU) supplied with the handpiece.WARNINGS!If  in  the  medical  opinion  of  the  physician  a  patient  with  a  prion  related  disease undergoes a high risk procedure, the instrument should be destroyed or be processed according to local requirements.Use of a phaco handpiece in the absence of irrigation ow and/or in the presence of reduced or lost aspiration ow and/or sideways orientation of the Kelman® and OZil® 1 2  tips can cause excessive heating and potential thermal injury to adjacent eye tissues.Appropriate use of Centurion® Vision System parameters and accessories is important for  successful  procedures.  Use  of  low  vacuum  limits,  low  ow  rates,  low  bottle heights, high power settings, extended power usage, power usage during occlusion conditions (beeping tones), failure to sufciently aspirate viscoelastic prior to using power, excessively tight incisions, and combinations of the above actions may result in signicant temperature increases at incision site and inside the eye, and lead to severe thermal eye tissue damage.Use of an ultrasonic handpiece other than an OZil® torsional handpiece, or use of a handpiece repaired without Alcon authorization, is not permitted, and may result in patient injury, including potential shock hazard to patient and/or operator.The U/S tips supplied in the Centurion® Vision System pack are only to be used on an OZil® torsional handpiece. Each U/S tip is intended to be used only once  per case, and then disposed of according to local governing ordinances.Mismatching U/S tips and infusion sleeves may create potentially hazardous uidic imbalances.Directing  energy  toward  non-lens  material,  such  as  iris  or  capsule,  may  cause mechanical and/or thermal tissue damage.Perform visual inspection of accessories for burs or bent tips prior to use.Use of appropriate technique and settings is important to minimize fragments and turbulence.
1.14  8065751772CAUTIONSNever ultrasonically clean the phaco handpiece; irreparable damage may result.Prior to sterilization, the phaco handpiece should always have the connector end cap secured and placed in the sterilization tray. This will prevent damage to the connectors and handpieces during handling, and especially during autoclaving.The phaco handpiece and INTREPID® AutoSert® IOL Injector must be at room temperature just before use. Allow the handpiece to air cool for at least 15 minutes after autoclaving; never immerse the handpiece in liquid when hot.Do not operate the phaco handpiece unless the tip is immersed in BSS® sterile irrigating solution or distilled water or is in surgical use. Irreparable damage to the handpiece and tip can result if run dry.Ensure that test chamber is lled with BSS® sterile irrigating solution before tuning the phaco handpiece. Tuning a handpiece dry may result in premature tip failure and breakage.Quenching a hot handpiece in water can cause damage and will void warranty.Be sure handpiece is completely dry before connecting it to console. Damage to handpiece and console may result if plugged in when wet. Handpiece Tips    Ensure that handpiece tip is fully tightened to the handpiece. If not securely attached, an error may be generated and/or inadequate tuning will occur. Ensure that the tip is not too tight so that it can be removed after use.    Use of a tool other than tip wrenches supplied by Alcon may cause damage to the tip and/or handpiece.WARNING!Poor clinical performance will result if tip is not secured tightly to the handpiece.During any ultrasonic procedure, metal particles may result from inadvertent touching of the ultrasonic tip with a second instrument. Another potential source of metal particles resulting from any ultrasonic handpiece may be the result of ultrasonic energy causing micro abrasion of the ultrasonic tip.
8065751772 1.15 Ultraow® II (I/A) Handpiece  Prior to each procedure inspect the two O-rings where the tip screws onto the Ultraow® II I/A handpiece. If damaged or missing, replace the o-rings. If in doubt, contact Alcon's Technical Services Department.WARNINGS!Use of non-Alcon surgical reusable or disposable I/A handpieces that do not meet Alcon surgical specications, or use of an Alcon handpiece not specied for use with the Centurion® Vision System, may result in a uidic imbalance. This, in turn, may cause a shallowing or collapsing of the anterior chamber.Exceeding the recommended level of 100 mmHg (133 hPa) with a 0.5 mm or larger I/A tip may  cause  anterior chamber shallowing  and/or  incarceration or tearing  of posterior capsule.I/A tips are not to be used with a phaco handpiece. Recommended Vacuum Range for I/A Tips  It is important that only the proper size I/A tip be used when operating with maximum vacuum. Only 0.2 mm or 0.3 mm I/A tips should be used with vacuum limits above 100 mmHg (133 hPa). I/A adjustable vacuum range is 0 - 700 mmHg (0 - 933 hPa). Centurion® Vitrectomy Probe    The vitrectomy probe, a guillotine vitreous cutter, is intended for single use only.  Vitrectomy cutting performance may vary at high altitudes. Consult Alcon Technical Service for additional information.WARNINGS!Do not test or operate vitrectomy probe unless tip of probe is immersed in BSS® sterile  irrigating  solution  or  distilled  water  or  is  in  surgical  use.  Irreparable damage to the probe and tip can result if run dry.Perform visual inspection of accessories for burs or bent tips prior to use.Connect pneumatic tubing connectors from vitrectomy probe to console prior to initiating  prime of probe. Initiating prime of the vitrectomy  probe,  or  running  the vitrectomy system, with one or both pneumatic connectors disconnected may cause the ow of non-sterile air over the sterile eld for a brief moment.Do not use vitrectomy probes that are not approved for use on Centurion® system.After lling and testing, and before surgical use, verify that the probe is properly actuating  and  aspirating.  This  may  require  lowering  cut  rate  to  achieve  good visualization. The port should always remain in open position in footpedal position 1. If cutting port is partially closed while in position 1, replace the probe. Prior to entry into the eye, and  with  tip of probe  in sterile irrigating solution,  the surgeon should step on the footpedal for visual verication that the probe is cutting:•  If the cutter is observed to not fully close, or does not move when the probe is actuated, replace the probe.•  If cutting port is partially closed while idle, replace the probe.•  If air bubbles are observed in the aspiration line or exiting the probe tip during priming, replace the probe.•  If a reduction of cutting capability or vacuum is observed during the surgical procedure, stop immediately and replace the probe.
1.16  8065751772 INTREPID® AutoSert® IOL InjectorCAUTIONS• Do not ultrasonically clean the AutoSert® IOL Injector connector. Ultrasonic cleaning will cause irreparable damage.•  Use care when handling AutoSert® IOL Injector, particularly when cleaning. Always clean handpiece over a surface cushioned with a pad or rubber mat.•  Be sure handpiece cable connector is dry before connecting it to the console.•  Do not disconnect cable connector from Centurion® system console until the IOL Injector plunger is fully retracted.•  Do not immerse the AutoSert® IOL Injector in any uid when the plunger is not retracted.•  As  part  of  a  properly  maintained  surgical  environment,  it  is  recommended  that a backup IOL injector be made available in the event the AutoSert® IOL injector handpiece does not perform as expected.WARNINGS!•  The AutoSert® IOL Injector is non-sterile and must be cleaned and sterilized prior to, and immediately after, each use.•  Never immerse the AutoSert® IOL Injector in liquid after autoclaving; allow it to air cool for at least 15 minutes. Quenching could result in a potentially hazardous condition for the patient.•  The AutoSert® IOL Injector delivery system is for the implantation of Alcon qualied AcrySof® foldable IOLs. Unqualied lenses shall not be used with the system. See DFU, or contact your Alcon representative, for qualied lens/cartridge combinations.•  The cartridge/IOL combination listed in the DFU, along with Alcon settings, has been validated per section 5 of BS EN ISO 11979-3:2006. Appropriate use of AutoSert® IOL Injector settings is important for successful IOL implantation. Inappropriate use of settings may lead to a potentially hazardous condition for the patient. Aspiration/Vacuum Adjustments    Adjusting aspiration rates or vacuum limits above the preset values may result in aspiration levels (volumes) exceeding irrigation inow.WARNING!Adjusting aspiration rates or vacuum limits above the preset values, or lowering the IOP or IV pole below the preset values, may cause chamber shallowing or collapse which may result in patient injury.
8065751772 1.17 Presurgical Check-out Tests  Presurgical check-out tests must be performed as outlined in the Operating Instructions section. If an Event message is displayed on the front panel, refer to the Troubleshooting section of this manual. If the problem persists, DO NOT PROCEED.WARNINGS!When lling handpiece test chamber, if stream of uid is weak or absent, good uidics response will be jeopardized. Good clinical practice dictates the testing for adequate irrigation and aspiration ow prior to entering the eye.Ensure that tubings are not occluded or pinched during any phase of operation.Perform visual inspection of accessories for burrs or bent tips prior to use. IV PoleWARNINGS!Keep clear of the IV pole when it is in motion to prevent skin, hair, and/or clothing from being trapped in the IV pole mechanism. The IV pole moves during power on/off, priming, and bottle height adjustment.IV pole rises automatically. To avoid stretching drip chamber tubing, and possibly pulling drip chamber out of bottle, tubing must hang freely with no interference.When out of use, remove uid bottle from IV pole and ip bottle hanger into its storage position to avoid injury.Empirical numbers for bottle heights are not a replacement for competent surgical technique. The surgeon should visually and physically monitor intraocular pressure. Footswitch  If required, the footswitch may be wiped with alcohol, mild soap and water, or any germicidal solution that is compatible with the plastic parts.WARNING!Route the footswitch cable properly to avoid tripping.CAUTIONSDo not clean the footswitch using solvents, abrasives, or any cleaner that is not compatible with plastic parts made of GE Cycoloy CU 6800 and LEXAN 920A. Damage may result.Never pick up or move the footswitch by the cable. Dropping or kicking the footswitch can cause irreparable damage.
1.18  8065751772 Occlusion Tones  Two different occlusion tones (intermittent beeping tones during occlusion) indicate that the vacuum is near or at its preset limit, and aspiration ow is reduced or stopped to avoid exceeding the limit. The rst type, the I/A occlusion tone, sounds when occlusion occurs during aspiration only (in the absence of ultrasonic power). The I/A occlusion tone is a lower, intermittent single beep. The second type of occlusion tone, the phaco occlusion tone, is a higher, intermittent double beep, and sounds when occlusion occurs during application of ultrasonic power.  The I/A occlusion and phaco occlusion tones indicate that the vacuum has reached its maximum allowed preset value. The I/A occlusion tone can be turned off, while the phaco occlusion tone cannot be turned off.WARNINGS!The phaco occlusion bell indicates no aspiration ow. Use of high U/S settings and/or prolonged use may lead to thermal injury.Use of the phaco handpiece in the absence of irrigation ow and/or in the presence of reduced or lost aspiration ow can cause excessive heating and potential thermal injury to adjacent eye tissues.In the event of a persistent loss of aspiration during the application of U/S power, remove U/S power via footswitch control.  Vacuum Tone    A vacuum tone is provided. The pitch will vary relative to the amount of vacuum. A high vacuum can indicate that little to no ow is occuring. This tone can be reduced in volume, but not turned off.WARNINGS!A moderate to high vacuum tone may indicate little to no ow is occuring. Use of the phaco handpiece in the absence of irrigation ow and/or in the presence of reduced or lost aspiration ow can cause excessive heating and potential thermal injury to adjacent eye tissues.Do not exceed maximum capacity of drain bag (500 ml). Excessive pressure can result from exceeding drain bag maximum capacity and potentially result in a hazardous condition for the patient.In the event of a persistent loss of aspiration during the application of U/S power, remove U/S power via footswitch control. Cautery, Diathermy, Coagulation Denition The Centurion® Vision System uses the word “Coagulation” in place of Cautery or Diathermy, based on the following denition:  Coagulation - Isolated, bipolar, high frequency current supplied to conductors (e.g. forceps). Current passes between these electrodes, halting bleeding. (Abbreviated “Coag” in some of the text of this operator’s manual.)
8065751772 1.19  Coagulation Function  Listed below are general precautions to be followed when using the Coagulation function:  •  To ensure safe operation of the coagulation function, only approved cables and accessories must be used (See your Alcon representative). Coagulation performance can be guaranteed only when using Alcon components or Alcon-endorsed components.  •  To reduce the risk of accidental burns, caution should always be taken when operating high-frequency surgical equipment.  •  Interference produced by the operation of high-frequency surgical equipment may adversely inuence the operation of other electronic equipment.  •  Accessories should be checked regularly; electrode cables should particularly be checked for possible damage to the insulation.  •  Operation of the coagulation step is limited to extraocular uses only.  •  The lowest power level in coagulation step should always be selected for the intended purpose.  •  Skin-to-skin contact (for example between the arms and body of the patient) should be avoided, for example by insertion of dry gauze.  •  When HF (high frequency) surgical equipment and physiological monitoring equipment are used simultaneously on the same patient, any monitoring electrodes should be placed as far as possible from the surgical electrodes. Needle monitoring electrodes are not recommended.  •  In all cases, monitoring systems incorporating high frequency current-limiting devices are recommended.  •  The cables to the surgical electrodes should be positioned in such a way that contact with the patient or other leads is avoided.  •  Temporarily unused active electrodes should be stored so that they are isolated from the patient.  •  The use of ammable anaesthetics or oxidizing gases such as nitrous oxide (N2O) and oxygen should be avoided if a surgical procedure is carried out in the region of the thorax or the head, unless these agents are sucked away.  • Non-ammable agents should be used for cleaning and disinfection wherever possible.  •  Flammable agents used for cleaning or disinfecting, or as solvents of adhesives, should be allowed to evaporate before the application of HF surgery. Some materials, for example cotton, wool and gauze, when saturated with oxygen may be ignited by sparks produced in normal use of the HF surgical equipment.  •  Accessories should have a rated voltage equal to or greater than the maximum coagulation output voltage.WARNINGS!Do not use the coagulation function on patients with pacemakers or implanted debrillatory devices. If electrosurgery is used on patients with implanted cardiac pacemakers  or  debrillatory  devices  or  pacemaker  electrodes,  be  aware  that irreparable  damage  to  the  pacemaker  or  debrillatory  device  and  its  function may occur and lead to ventricular brillation. Please check with the pacemaker or debrillatory device manufacturers for their recommendations.Failure  of  the  HF  surgical  equipment  (coagulation  circuitry)  could  result  in an unintended increase of output power.
1.20  8065751772 VideOverlay SystemWARNINGS!Do not remove VideOverlay cover; there are no user-serviceable parts inside. Refer servicing to qualied service personnel.Do not simultaneously touch the VideOverlay enclosure and the patient.CAUTIONS•  Do not use multiple portable socket outlets with this system.•  Use only the Alcon-supplied serial cable to connect the Centurion® Vision System to the VideOverlay System.
8065751772 1.21 Consumable Packs    Consumable items used with the Centurion® Vision System during surgery are designed to be used once and then discarded, unless labeled otherwise.   All Centurion® packs contain Directions for Use (DFU). It is important to read and understand the DFU’s prior to use.  In all cases, the instrument setup instructions contained in the manual should be thoroughly understood prior to using any of the pack congurations.  NOTE: If an inconsistency exists between the instructions in the operator’s manual and the Directions For Use (DFU) supplied with a consumable pack or accessory, follow the DFU.WARNINGS!Mismatch of consumable components and use of settings not specially adjusted for a particular combination of consumable components may create a patient hazard.Do not use packs that have exceeded the expiration date.Sterile disposable medical devices should not be reused! These components have been designed for one time use only; do not reuse.Potential risk from reuse or reprocessing the following products labeled for single use include:•  Bipolar Coagulation Instruments - Thermal injury or electrical shock caused by a damaged bipolar instrument, and foreign particle introduction into the eye.•  Fluid Management Components - Fluid path leaks or obstruction resulting in reduced uidics performance, and foreign particle introduction into the eye.•  Phacoemulsication Tips - Reduced tip cutting performance, presence of tip burrs, uid path obstruction, and foreign particle introduction into the eye.•  Vitreous Cutting Instruments - Reduced vitreous cutting performance, uid path obstruction, and foreign particle introduction into the eye.The equipment used in conjunction with the Alcon disposables constitutes a complete surgical system. Use of disposables other than Alcon disposables may affect system performance and create potential hazards, and if it is determined to have contributed to the malfunction of the equipment under contract, could result in the voidance of the contract and/or invoicing at prevailing hourly rates.Perform visual inspection of accessories for burs or bent tips prior to use.Do not remove the FMS during the procedure.Do not use any Vitrectomy probes that have not been approved for use on the Centurion® Vision System.
1.22  8065751772PRODUCT SERVICE  For product service, please contact Alcon’s Technical Services Department at the number provided below.  Operators experiencing problems with the system should refer to the Operating Instructions and Troubleshooting sections of this manual. A problem which persists should be referred to the Alcon Technical Services Department or your local authorized service representative.  For optimum performance, it is the user’s responsibility to schedule preventive maintenance service on the system and its accessories a minimum of one time per year. Additional preventive maintenance may be required based upon system use. Alcon’s Field Service Engineers are trained and equipped to provide the highest quality of workmanship.  Safety performance should be veried by the user (e.g., qualied service personnel) at least twice a year. Ground resistance, leakage current, and dielectric withstand voltage must be checked to appropriate national standard.  To avoid unnecessary shipping, please contact your Alcon Technical Services Department prior to return of any system or accessories. If return of the equipment is deemed necessary, a Return Material Authorization will be issued with appropriate shipping instructions.Alcon Technical Services Department15800 Alton ParkwayIrvine, California 92618-3818(800) 832-7827, or (949) 753-1393
8065751772 1.23LIMITED WARRANTY  Alcon will repair or replace at its option, any system or accompanying accessories found to be defective in material and/or workmanship for a period of one (1) year from the date of initial installation. This warranty applies to the original purchaser of the system, when said system is properly installed, maintained, and operated in accordance with published instructions.  Alcon shall not be obligated to provide services under this warranty for damage to or destruction of systems covered where such damage or destruction is a result of or caused by re or explosion of any origin, riot, civil commotion, aircraft, war, or any Act of God including, but not limited to lightning, windstorm, hail, ood or an earthquake.  This warranty does not cover damage resulting from service repair or other alteration by any person other than an Alcon-authorized service person, and any warranties provided by Alcon with respect to this equipment shall become void and of no further force and effect if this equipment is serviced by anyone other than Alcon-authorized service personnel. In particular, Alcon shall have no obligation to replace, repair or credit customer’s account for the cost of the equipment, which has been subject to service or other alteration by persons other than Alcon-authorized service personnel.  The express warranty above is the sole warranty obligation of Alcon, and the remedy provided above is in lieu of any and all other remedies. There are no other agreements, guarantees, or warranties – oral or written, expressed or implied – including , without limitation, warranties of merchantability or tness for a particular purpose. Alcon shall have no liability whatsoever for any incidental or consequential damages arising out of any defect, improper use, or unauthorized service or repair.WARNING!The  consumable  products  used  in  conjunction  with  ALCON® instrument products constitute a complete surgical system. Use of consumable products and handpieces other than those manufactured by Alcon may affect system performance and create potential hazards. If it is determined that consumable products or handpieces not manufactured by Alcon have contributed to the malfunction of the equipment during warranty period, service will be provided at prevailing hourly rates.
8065751772 1.25Figure 1-2  ICONS USED WITH CENTURION® VISION SYSTEM - Icons identifying modes, functions, etc., that are used with the Centurion® Vision System are identied in this chart. The icons shown on this page are for reference only.Eject FMSLamp to illuminate instrument trayConnector forCENTURION® OZil® handpieceConnector forINFINITI® OZil® handpieceConnector forINTREPID® AutoSert® IOL InjectorConnector forVitrectomy probe tubingConnector forAutocapsulorhexis handpieceConnector forCoagulation handpieceConnector forCabled FootswitchMagnetic Resonance UnsafeNon-ionizing electromagnetic radiationDate of ManufactureManufacturerOSHA recognized NRTL, TUV SUDAmerica mark, providing electrical safetycertification to North Americanrequirements for Medical Devices.Type BF equipment, providing both theattributes of basic insulation and "floated"isolation.Follow Instructions for Use(white figure on blue background)WARNING: The console mightoverbalance when it is pushed and itswheels are immobilized (blocked)(black symbol behind lined out red circle)WARNING: Dangerous Voltage(black symbols on yellow background)GENERAL WARNING (black symbols on yellow background)Equipotential ground connectionAC VoltagePower stand-by statefor a part of equipmentON (POWER)OFF (POWER)Fuse Size, Type, and RatingUse appropriate take-back system(see Environmental Considerations in this manual)Pb notation, if present, indicates leadcontent greater than 0.004%.Catalog NumberSerial NumberPbSNREFT10.0AH/250VMR
1.26  8065751772Figure 1-3  LABELING ON CENTURION® VISION SYSTEM - Labels used on the Centurion® Vision System are illustrated here. The labels on this page are intended for reference only.215-2216-001 REV P0Label printed on rear panel of Centurion® Vision SystemLabel located on rear panel of Centurion® Vision SystemLabel printed on power input moduleLabel located on rear panel of Centurion® Vision SystemLabel located on back of remote controlLabel located on bottom of footswitch Label located on instrument trayT10.0AH/250VWARNING:FOR CONTINUEDPROTECTION AGAINST RISK OF FIRE,REPLACE ONLY WITH SAME TYPEAND RATING OF FUSE.IPX8IPX1®© 2012 Novartis220-240V        50/60 Hz  6A100-120V  50/60 Hz  10AFor applicable patents, please see the ABOUTscreen on the monitor during operation.ALCON LABORATORIES, INC.6201 SOUTH FREEWAYFORT WORTH, TX 76134-2099 USAMADE IN USA®®DANGER:DANGER:CAUTION:CAUTION:RISK OF EXPLOSION IF USED IN THEPRESENCE OF FLAMMABLEANESTHETICS.RISQUE D’EXPLOSION. NE PASEMPLOYER EN PRESENCED’ANESTHESIQUES INFLAMMABLES.GROUNDING RELIABILITY CAN ONLYBE ACHIEVED WHEN EQUIPMENT ISCONNECTED TO AN EQUIVALENTRECEPTACLE MARKED HOSPITALGRADE.RISK OF BURNS AND FIRE - DO NOTUSE NEAR CONDUCTIVE MATERIALS.RENEW ELECTRODE CABLES UPONEVIDENCE OF DETERIORATION.This device complies with part 15 of the FCCrules. Operation is subject to the following twoconditions: (1) This device may not causeharmful interference, and (2) this device mustaccept any interference received, includinginterference that may cause undesiredoperationFCC ID: VMCNGPFSW1IC:      7345A-NGPFSW1Contains:FCC ID: RI7XE61IC:    5131A-XE61FCC ID: N6C-SDMANIC:    4908B-SDMAN215-2985-001 REV P0
1.28  8065751772LAST PAGE OF THIS SECTIONTHIS PAGE INTENTIONALLY BLANK