Andon Health 926 Wireless Blood Pressure Monitor User Manual MODEL KD 595

Andon Health Co., Ltd Wireless Blood Pressure Monitor MODEL KD 595

User Manual

DN:KD-926-SMSY01 V1.0                                  Operation Guide      Operation Guide (V1.0)    Product/project name:Blood Pressure Monitor Model    name:   KD-926         Project    number:                                              Drafted      by:                Date                 Reviewed by:                      Date                   Approved by:                        Date
DN:KD-926-SMSY01 V1.0                                  Operation Guide  1/16 MODEL KD-926 Wireless Blood Pressure Monitor   (ELECTRONIC SPHYGMOMANOMETER) OPERATION GUIDE __________________________________________________________________________ INDEX  IMPORTANT INFORMATION ......................................................................... 2 CONTENTS AND DISPLAY INDICATORS ..................................................... 2 INTENDED USE .............................................................................................. 3 CONTRAINDICATION ..................................................................................... 3 PRODUCT DESCRIPTION .............................................................................. 3 SPECIFICATIONS ........................................................................................... 4 NOTICE ........................................................................................................... 4 SETUP AND OPERATING PROCEDURES .................................................... 6 1. DOWNLOAD THE FREE APP .................................................................... 6 2. BATTERY LOADING AND AC ADAPTER LOADING ................................ 6 3. CLOCK AND DATE ADJUSTMENT ........................................................... 7 4. CONNECTING THE CUFF TO THE MONITOR .......................................... 8 5. APPLYING THE CUFF ................................................................................ 8 6. BODY POSTURE DURING MEASUREMENT ............................................ 9 7. TAKING YOUR BLOOD PRESSURE READING ........................................ 9 8. DISPLAYING STORED RESULTS ........................................................... 10 9. SYNCHRONIZING STORED RESULTS ................................................... 10 10. DELETING MEASUREMENTS FROM THE MEMORY ........................... 11 11. ASSESSING HIGH BLOOD PRESSURE FOR ADULTS ....................... 11 12. TECHNICAL ALARM DESCRIPTION ..................................................... 12 13. TROUBLESHOOTING (1) ....................................................................... 12 14. TROUBLESHOOTING (2) ....................................................................... 13 MAINTENANCE ............................................................................................ 14 EXPLANATION OF SYMBOLS ON UNIT ..................................................... 14 WARRANTY INFORMATION ........................................................................ 15 SERVICE CENTER ....................................................................................... 15 IMPORTANT INFORMATION REQUIRED BY THE R&TTE ........................ 15 ELECTROMAGNETIC COMPATIBILITY INFORMATION ............................ 16
DN:KD-926-SMSY01 V1.0                                  Operation Guide  2/16 IMPORTANT INFORMATION  NORMAL BLOOD PRESSURE FLUCTUATION  All physical activity, excitement, stress, eating, drinking, smoking, body posture and many other activities or factors (including taking a blood pressure measurement) will influence blood pressure value. Because of this, it is mostly unusual to obtain identical multiple blood pressure readings.    Blood pressure fluctuates continually day and night. The highest value usually appears in the daytime and lowest one usually at midnight. Typically, the value begins to increase at around 3:00 AM, and reaches to highest level in the daytime while most people are awake and active.  Considering the above information, it is recommended that you measure your blood pressure at approximately the same time each day.  Too frequent measurements may cause injury due to blood flow interference, please always relax a minimum moment of 1 to 1.5 minutes between measurements to allow the blood circulation in your arm to recover. It is rare that you obtain identical blood pressure readings each time.  CONTENTS AND DISPLAY INDICATORS
DN:KD-926-SMSY01 V1.0                                  Operation Guide  3/16  INTENDED USE  Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm.The cuff circumference is limited to 22cm-42cm(approx.8 21/32″-16 17/32″).  CONTRAINDICATION    It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer.  PRODUCT DESCRIPTION  Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse rate can be measured automatically and non-invasively. The LCD display will show blood pressure and pulse rate. The most recent 2×60 measurements can be stored in the memory with date and time stamp. The Electronic Sphygmomanometer corresponds to the below standards:    IEC 60601-1:2005 +A1:2012(E)/EN 60601-1:2006/A11: 2011 (Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance), IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests), IEC80601-2-30:2009+AMD1:2013/EN 80601-2-30:2010/A1:2015(Medical electrical equipment –Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers)EN 1060-1: 1995 +    A2: 2009 (Non-invasive sphygmomanometers - Part 1: General requirements), EN 1060-3: 1997 + A2: 2009 (Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems).
DN:KD-926-SMSY01 V1.0                                  Operation Guide  4/16 SPECIFICATIONS 1.  Product name: Wireless Blood Pressure Monitor 2.  Model: KD-926 3.  Classification: Internally powered, Type BF applied part,IP20,No AP or APG, Continuous operation 4.  Machine size: Approx. 145 mm ×90 mm × 51mm 5.  Cuff circumference: 8-21/32” to 16-17/32”(22cm-42cm) 6.  Weight: Approx. 234g (8 1/4oz.) (exclude cuff) 7.  Measuring method: Oscillometric method, automatic inflation and measurement 8.  Memory volume: 2×60 times with time and date stamp 9.  Power source: batteries: 4x1.5V SIZE AA, DC 6V 600mA 10. Measurement range: Cuff pressure: 0-300mmHg Systolic:          60-260mmHg Diastolic:        40-199mmHg Pulse rate:    40-180 beats/minute 11. Accuracy:   Pressure:     ±3mmHg Pulse rate:   ±5% 12. Environmental temperature for operation: 10°C -40°C(50°F -104°F) 13. Environmental humidity for operation:   ≤85%RH 14. Environmental temperature for storage and transport: -20℃~50℃(-4℉~122℉) 15. Environmental humidity for storage and transport:  ≤85%RH 16. Environmental pressure: 80kPa-105kPa 17. Battery life: Approx 300 times. 18. Wireless Connection: Smart Bluetooth Frequency Band: 2.400~2.4835GHz 19. All components belonging to the pressure measuring system, including: Pump,Valve, LCD, Cuff, Sensor   Note: These specifications are subject to change without notice.  NOTICE   1.  Read all of the information in the operation guide and any other literature in the box before operating the unit. 2.  Stay still, calm and rest for 5 minutes before blood pressure measurement. 3.  The cuff should be placed at the same level as your heart. 4.  During measurement, neither speak nor move your body and arm. 5.  Measuring on same arm for each measurement. 6.  Please always relax at least 1 or 1.5 minutes between measurements to allow the blood circulation  in  your  arm  to  recover.  Prolonged  over-inflation  (cuff  pressure  exceed  300 mmHg or maintained above15 mmHg for longer than 3 minutes) of the bladder may cause ecchymoma of your arm. 7.  Consult your physician if you have any doubt about below cases:
DN:KD-926-SMSY01 V1.0                                  Operation Guide  5/16 1) The application of the cuff over a wound or inflammation diseases; 2) The application of the cuff on any limb where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present; 3) The application of the cuff on the arm on the side of a mastectomy; 4) Simultaneously used with other monitoring medical equipments on the same limb; 5) Need to check the blood circulation of the user. 8.  This Electronic Sphygmomanometer is designed for adults and should never be used on infants or  young children. Consult your physician or  other health care  professionals before use on older children. 9.  Do not use this unit in a moving vehicle, This may result in erroneous measurement. 10. Blood  pressure  measurements  determined  by  this  monitor  are  equivalent  to  those obtained by a trained observer using the cuff/stethoscope auscultation method, within the limits  prescribed  by  the  American  National  Standard  Institute,  Electronic  or  automated sphygmomanometers. 11. Information regarding potential electromagnetic or other interference between the blood pressure  monitor  and  other  devices  together  with  advice  regarding  avoidance  of  such interference please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION.                                      12. If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the procedure of blood pressure measurement, a signal of   will be displayed. Under this condition, the Electronic Sphygmomanometer can keep function, but the results may not be accurate, it’s suggested that you consult with your physician for accurate assessment.     There are 2 conditions under which the signal of IHB will be displayed: 1) The coefficient of variation (CV) of pulse period >25%. 2) The difference of adjacent pulse period≥0.14s, and the number of such pulse takes more than 53 percentage of the total number of pulse.   13. Please do not use the cuff other than supplied by the manufacturer, otherwise it may bring biocompatible hazard and might result in measurement error. 14.  The monitor might not meet its performance specifications or cause safety hazard if stored or used outside the specified temperature and humidity ranges in specifications. 15.  Please do not share the cuff with other infective person to avoid cross-infection. 16. This equipment has been tested and found to comply with the limits for a Class B digital device,  pursuant  to  part  15  of  the  FCC  Rules.  These  limits  are  designed  to  provide reasonable  protection  against  harmful  interference  in  a  residential  installation.  This equipment generates, uses and can radiate radio frequency energy and, if not installed and  used in accordance  with  the  instructions, may  cause  harmful  interference to  radio communications.  However,  there  is  no  guarantee  that  interference  will  not  occur  in  a particular  installation.  If  this  equipment  does  cause  harmful  interference  to  radio  or television reception, which can be determined by turning the equipment off and on, the user  is  encouraged  to  try  to  correct  the  interference  by  one  or  more  of  the  following measures: —Reorient or relocate the receiving antenna. —Increase the separation between the equipment and receiver. —Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
DN:KD-926-SMSY01 V1.0                                  Operation Guide  6/16 —Consult the dealer or an experienced radio/TV technician for help. 17. Medical AC adapter which output is DC 6.0V 600mA and complied with IEC 60601-1/EN 60601-1/UL 60601-1 and IEC 60601-1-2/EN 60601-1-2/UL 60601-1-2 is suitable for this monitor. Please note that the monitor jack size: hole Φ5.5mm, center pin Φ2.0mm. Please pay attention to polarity. 18. Measurements are not possible in patients with a high frequency of arrhythmias. 19. The device is not intended for use  on neonates,  children or pregnant women. (Clinical testing has not been conducted on neonates, children or pregnant women.) 20. Motion, trembling, shivering may affect the measurement reading. 21. The device would not apply to the patients with poor peripheral circulation, noticeably low blood pressure, or low body temperature (there will be low blood flow to the measurement position). 22. The device would not apply to the patients who use an artificial heart and lung (there will be no pulse) 23. Consult  your  physician  before  using  the  device  for  any  of  the  following  conditions: common  arrhythmias  such  as  atrial  or  ventricular  premature  beats  or  atrial  fibrillation, arterial sclerosis, poor perfusion, diabetes, pre-eclampsia, enal diseases.  SETUP AND OPERATING PROCEDURES 1.  DOWNLOAD THE FREE APP Prior to first use, download and install the Andon App from the App Store(iOS device) or Google Play(Android device). Use keyword search terms ”mHealth”. 2.  BATTERY LOADING AND AC ADAPTER LOADING a.  Open battery cover at the back of the monitor. b. Load four “AA” size batteries. Please pay attention to polarity. c.  Close the battery cover. When LCD shows battery symbol , replace all batteries with new ones.     Rechargeable batteries are not suitable for this monitor.   Remove the batteries if  the monitor will not  be used  for  a month or more to avoid relevant damage of battery leakage.   Avoid the battery fluid to get in your eyes. If it should get in your eyes, immediately rinse with plenty of clean water and contact a physician. d.  If you use the AC adapter, please make sure the monitor turn off or no batteries. Put the connector plug of the adapter into the socket as the picture, Then plug the adapter to AC outlet. When disconnect the AC Adapter:                 Remove the AC Adapter from the electrical outlet;   Remove the AC Adapter plug from the monitor socket.  Do not plug or unplug the power cord into the electrical outlet with wet hands.  Do not overload power outlets. Plug the device into the appropriate voltage outlet.   If the AC adapter is abnormal, please change the adapter.   Do not pull out the adapter when you are using the monitor.
DN:KD-926-SMSY01 V1.0                                  Operation Guide  7/16           Do not use any other type of AC adapter as it may harm the monitor. The monitor, the batteries and the cuff, must be disposed of according to local regulations at the end of their usage.  3.  CLOCK AND DATE ADJUSTMENT a.  Once you install the battery,the Blood Pressure Monitor will enter Clock and Date Adjustment Mode. When you turn off the monitor, it will enter Clock Mode, and LCD will display time and date. b.  While the monitor is in Clock Mode , adjustment can be reached by pressing both the “START/STOP” and “MEM” button for 1 seconds in Clock Mode to change the time.See picture3. c.  In  Clock  and  Date  Adjustment  Mode  ,  the  time  format  will  blink  at  first  ,  see picture3-1 .The default time format is 24h(Europe Version) and the default clock and date is 2016-1-1 12:00. d. Press the button “START/STOP” repeatedly, the year (first usage: default is 2016, range is 2016~2099), month, day, hour and minute will blink in turn, see picture3- 2&3-3 & 3-4 & 3-5 & 3-6. While the number is blinking, press the button “MEM” to increase  the  number,  keep  on  pressing  the  button  "MEM",  the  number  will increase faster.                Picture 3              Picture 3-1        Picture 3-2          Picture 3-3                             Picture 3-4          Picture 3-5          Picture 3-6 e.  During  adjusting  clock  and  date,  the  monitor  will  go  back  to  Clock  Mode automatically when no button will be pressed within 60 seconds. f. You can turn off the monitor by pressing “START/STOP” button when the minute is blinking, then the time and date is confirmed. Note: 3.1  The clock format could be set by user.
DN:KD-926-SMSY01 V1.0                                  Operation Guide  8/16 3.2  Table 1 instructs the conversion relations between 24 hour format and 12 hour format. Table 1 24 hour format 12 hour format 24 hour format 12 hour format 0:00 12:00 AM 12:00 12:00 PM 1:00 1:00 AM 13:00 1:00 PM 2:00 2:00 AM 14:00 2:00 PM 3:00 3:00 AM 15:00 3:00 PM 4:00 4:00 AM 16:00 4:00 PM 5:00 5:00 AM 17:00 5:00 PM 6:00 6:00 AM 18:00 6:00 PM 7:00 7:00 AM 19:00 7:00 PM 8:00 8:00 AM 20:00 8:00 PM 9:00 9:00 AM 21:00 9:00 PM 10:00 10:00 AM 22:00 10:00 PM 11:00 11:00 AM 23:00 11:00 PM   4.  CONNECTING THE CUFF TO THE MONITOR  Insert the cuff tubing connector into the air port in the side of the monitor. Make sure that the connector is completely inserted to avoid air leakage during blood pressure measurements.  Avoid compression or restriction of the connection tubing during measurement,which may cause inflation error, or harmful injury due to continuous cuff pressure.   5.  APPLYING THE CUFF  a.  Pulling the cuff end    through the medal loop (the cuff is packaged like this already), turn it outward (away from your body) and tighten it and close the Velcro fastener. b.  Place the cuff around a bare left arm 1-2cm above the elbow joint.  c.  While seated, place palm upside in front of you on a flat surface such as a desk or table. Position the air tube in the middle of your arm in line with your middle finger. d.  The cuff should fit comfortably, yet snugly around your arm.You should be able to insert one finger between your arm and the cuff. Note:    Please refer to the cuff circumference range in “SPECIFICATIONS” to make sure that the appropriate cuff is used.  Measure on the left arm each time.
DN:KD-926-SMSY01 V1.0                                  Operation Guide  9/16  Do not move your arm, body, or the monitor and do not move the rubber tube during measurement.  Stay quiet, calm for 5 minutes before blood pressure measurement.  Please keep the cuff clean. If the cuff becomes dirty, remove it from the monitor and clear it by hand in a mild detergent, then rinse it thoroughly in cold water. Never dry the cuff in clothes dryer or iron it. Clean the cuff after the usage of every 200 times is recommended.      6.  BODY POSTURE DURING MEASUREMENT  Sitting Comfortably Measurement a. Be seated with your feet flat on the floor, and don’t cross your legs.   b.  Place palm upside in front of you on a flat surface such as desk or table. c.  The middle of the cuff should be at the level of the right atrium of the heart.    Lying Down Measurement a.  Lie on your back.   b.  Place your left arm straight along your side with your palm upside. c.  The cuff should be placed at the same level as your heart.  7.  TAKING YOUR BLOOD PRESSURE READING  a.  After applying the cuff and your body is in a comfortable position, press the “START/STOP” button. All display characters are shown for self-test. You can check the LCD display according to the right picture. Please contact the service center if segment is missing.See picture 7-1. b.  Then the current memory bank ( or ) is blink. See picture 7-2. Press “MEM” button to change over to other bank. See picture 7-3. Confirm your selection by pressing “START” button. The current bank can also be confirmed automatically after 5 seconds with no operation.            Picture 7-1            Picture 7-2            Picture 7-3            Picture 7-4            Picture 7-5 c.  After selecting the memory bank, the monitor starts to seek zero pressure. See picture 7-4. d.  Then the cuff will be slowly inflated. The blood pressure and pulse will be measured during inflation. Inflation will stop as soon as the blood pressure and pulse rate have been calculated and displayed on the LCD. The irregular
DN:KD-926-SMSY01 V1.0                                  Operation Guide  10/16 heartbeat symbol (if any) and blood pressure classification indicator will blink on the LCD. See picture 7-5. The result will automatically be stored in the Memory bank of the monitor. e.  After  measurement,  the  monitor  will  turn  off  automatically  after  1  minute  of  no operation. f. During  measurement,  you  can  press  the  “START/STOP”  button  to  turn  off  the monitor manually. Note: Please consult a health care professional for interpretation of pressure measurements.  8.  DISPLAYING STORED RESULTS  a.  In Clock Mode, press “MEM” button, the monitor will display sign of current group.The amount of results in current user memory zone will be displayed. See picture 8.Press “START/STOP” button to switch group, press “MEM” to confirm current group.  b.  Then LCD will display the average value of the last three results in this bank. See picture 8-1. If no result stored, LCD will show picture 8-2. c. Press “MEM” button, the LCD will display the average value of all results in the current user memory zone. See picture 8-3. If no result stored in the current user memory zone, LCD will display “0” for blood pressure and pulse rate. See picture 8-4. d. Press “MEM” button again, the most recent result will be displayed with date and time stamp.  See  picture  8-5.  Irregular  heartbeat  symbol  (if  any)  and  blood  pressure classification indicator will blink at the same time. If the monitor has no result stored in the current user memory zone, the LCD will display “0” for blood pressure and pulse rate. See picture 8-6. e. Press “MEM” button again to review the next result. In this way, repeatedly pressing the “MEM” button displays the respective results measured previously.                                       Picture8                      Picture 8-1                    Picture 8-2           Picture 8-3                                                 Picture 8-4                  Picture 8-5              Picture 8-6      Note:    When the monitor displaying the measurement, the classification color indicator
DN:KD-926-SMSY01 V1.0                                  Operation Guide  11/16 can  be  shown  different  color  according  to  the  systolic  pressure  and  diastolic pressure. Refer to the “ASSESSING HIGH BLOOD PRESSURE FOR ADULTS” section.    9.  SYNCHRONIZING STORED RESULTS a.  In  Clock  Mode,  press  “MEM”  button,  the  monitor  will  wait  Bluetooth  connect  and Bluetooth  symbol  flashing  after  confirming  the  group.  See  picture  8-1.  Bluetooth symbol  will  stop  flashing  when  Bluetooth  is  connected.  Send  symbol  flashing  until Synchronize finished. See picture 8-3. Send symbol flashing means it is synchronizing the memory at this time.   b. If  an  accident  happens,  synchronous  unfinished,  unfinished  symbol  flashing.  See picture 8-5.   c.  When Bluetooth is disconnected, the monitor will turn off automatically after 1 minute of no operation. You can also press the “START/STOP” button to turn off the monitor manually.  10. DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying (exclude average value displaying), keeping on pressing button “MEM” for three seconds, all results in the current bank will be deleted . See picture 10. Press the button “START/STOP”, the monitor will be turned off.                                                          Picture 10      11. ASSESSING HIGH BLOOD PRESSURE FOR ADULTS The following guideline for assessing high blood pressure (without regard to age or gender) has  been  established  by  the  World  Health  Organization  (WHO).    Please  note  that  other factors (e.g. diabetes, obesity, smoking, etc.) need to be taken into consideration. Consult with your physician for accurate assessment, and never change your treatment by yourself.
DN:KD-926-SMSY01 V1.0                                  Operation Guide  12/16  Note: It is not intended to provide a basis of any type of rush toward emergency conditions/diagnosis based on the color scheme and that the color scheme is meant only to discriminate between the different levels of blood pressure.    12. TECHNICAL ALARM DESCRIPTION The monitor will show 'HI' or 'Lo' as technical alarm on LCD with no delay if the determined blood pressure (systolic or diastolic) is outside the rated range specified in part SPECIFICACIONS. In this case, you should consult a physician or check if your operation violated the instructions. The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or inactivated. This alarm condition is assigned as low priority according to IEC 60601-1-8. The technical alarm is non-latching and need no reset. The signal displayed on LCD will disappear automatically after about 8 seconds.  13. TROUBLESHOOTING (1) PROBLEM POSSIBLE CAUSE SOLUTION LCD  Display  shows abnormal result The cuff position was not correct or it was not properly tightened Apply  the  cuff  correctly  and  try again. Body posture was not correct during testing Review  the  “BODY  POSTURE DURING  MEASUREMENT” sections  of  the  instructions  and re-test. Speaking,  arm  or  body  movement, angry,  excited  or  nervous  during testing Re-test  when  calm  and  without speaking  or  moving  during  the test. Irregular heartbeat (arrhythmia) It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer.
DN:KD-926-SMSY01 V1.0                                  Operation Guide  13/16  14. TROUBLESHOOTING (2) PROBLEM POSSIBLE CAUSE SOLUTION LCD shows low battery symbol   Low Battery Change the batteries LCD shows “Er 0” Pressure system is unstable before measurement  Don’t move and try again. LCD shows “Er 1” Fail to detect systolic pressure LCD shows “Er 2” Fail to detect diastolic pressure LCD shows “Er 3” Pneumatic system blocked or cuff is too tight during inflation  Apply the cuff correctly and try again LCD shows “Er 4” Pneumatic system leakage or cuff is too loose during inflation LCD shows “Er 5” Cuff pressure above 300mmHg  Measure again after five minutes. If the monitor is still abnormal, please contact the local distributor or the factory. LCD shows “Er 6” More than 3 minutes with cuff pressure above 15 mmHg LCD shows “Er 7” EEPROM accessing error LCD shows “Er 8” Device parameter checking error LCD shows “Er A” Pressure sensor parameter error No response when you press button or load battery. Incorrect operation or strong electromagnetic interference. Take out batteries for five minutes, and then reinstall all batteries. Bluetooth connection unstable Bluetooth connection unsuccessful, monitor is   abnormal, or strong electromagnetic interference is present Reset iOS/Android device. Reset monitor. Make sure the monitor and iOS/Android device are away from other electrical equipment. Please see GENERAL SAFETY AND PRECAUTIONS
DN:KD-926-SMSY01 V1.0                                  Operation Guide  14/16 MAINTENANCE  1.  Do not drop this monitor or subject it to strong impact. 2.  Avoid high temperature and solarization. Do not immerse the monitor in water as this will result in damage to the monitor. 3.  If this monitor is stored near freezing, allow it to acclimate to room temperature before use. 4.  Do not attempt to disassemble this monitor. 5.  If you do not use the monitor for a long time, please remove the batteries. 6.  It  is  recommended  the  performance  should  be  checked  every  2  years  or  after  repair. Please contact the service center. 7.  Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after moistened with water, diluted disinfectant alcohol, or diluted detergent. 8.  No  component  can  be  maintained  by  user  in  the  monitor.  The  circuit  diagrams, component part lists, descriptions, calibration instructions, or other information which will assist  the  user’s  appropriately  qualified  technical  personnel  to  repair  those  parts  of equipment which are designated repairably can be supplied by us. 9.  The monitor can maintain the safety and performance characteristics for a minimum of 10,000  measurements  or  three  years,  and  the  cuff  integrity  is  maintained  after  1,000 open–close cycles of the closure. 10.  It  is  recommended  the  cuff  should  be  disinfected  2  times  every  week  if  needed  (For example, in hospital or in clinique). Wipe the inner side (the side contacts skin) of the cuff by a soft cloth squeezed after moistened with Ethyl alcohol (75-90%), then dry the cuff by airing.  EXPLANATION OF SYMBOLS ON UNIT    Symbol for” THE OPERATION GUIDE MUST BE READ”(The sign background colour: blue.The sign graphical symbol: white)     Symbol for “WARNING”     Symbol for “TYPE BF APPLIED PARTS” (The cuff is type BF applied part)     Symbol for “ENVIRONMENT PROTECTION – Waste electrical products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local Authority or retailer for recycling advice”.  Symbol for “MANUFACTURER”  Symbol for “COMPILES WITH MDD93/42/EEC REQUIREMENTS”                 Symbol for “DATE OF MANUFACTURE”
DN:KD-926-SMSY01 V1.0                                  Operation Guide  15/16 N O . 31,C h a n g jia ng  Ro a d ,N a n ka i D istric t,Tia n jin, P.R.C h in aEC REPA N D O N  H e a lth  C o .,LtdA U TO M A TIC  BLO O D  PRESSU R E M O N ITO RM O D EL: KD -59076V      SN :R0197Lotus Global Co.,Ltd 47 Spenlow House Bermondsey London SE16 4SJ              Symbol for “EUROPEAN REPRESENTATION” SN                Symbol for “SERIAL NUMBER”             Symbol for “KEEP DRY”   WARRANTY INFORMATION Only charge the cost of components and transport.    SERVICE CENTER    ANDON HEALTH CO., LTD. No. 3 Jinping Street, YaAn Road, Nankai District, Tianjin 300190, China. Tel: 86-22-60526081            Lotus Global Co., Ltd.   1 Four Seasons Terrace West Drayton, Middlesex, London, UB7 9GG,United Kingdom  Tel: +0044-20-75868010    Fax: +0044-20-79006187  IMPORTANT INFORMATION REQUIRED BY THE R&TTE  This product is approved in accordance to R&TTE directive transmitter. interference This product complies with Industry Canada. IC: RSS-247 IC NOTICE This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.  The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc and any use of such marks by ANDON HEALTH CO., LTD. is under license. Other trademarks and trade names are those of their respective owners.  This product is approved in accordance to R&TTE directive transmitter. Hereby, ANDON HEALTH CO., LTD., declares that this KD-926 is in compliance with the essential requirements and other relevant provisions of Directive 1999/5/EC. See part Directive 1999/5/EC declaration of conformity   N O . 31,C h a n g jia ng  Ro a d ,N a nka i D istric t,Tia n jin, P.R.C h in aEC REPA N D O N  H e a lth  C o .,LtdA U TO M A TIC  BLO O D  PRESSU R E M O N ITO RM O D EL: KD -59076V      SN :R0197Lotus Global Co.,Ltd 47 Spenlow House Bermondsey London SE16 4SJ
DN:KD-926-SMSY01 V1.0                                  Operation Guide  16/16   ELECTROMAGNETIC COMPATIBILITY INFORMATION Table 1 - Emission Phenomenon Compliance Electromagnetic environment RF emissions  CISPR 11 Group 1, Class B Home healthcare environment Harmonic distortion IEC 61000-3-2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000-3-3 Compliance Home healthcare environment  Table 2 - Enclosure Port Phenomenon Basic EMC standard Immunity test levels Home Healthcare Environment Electrostatic Discharge IEC 61000-4-2 ±8 kV contact ±2kV, ±4kV, ±8kV, ±15kV air Radiated RF EM field IEC 61000-4-3 10V/m 80MHz-2.7GHz 80% AM at 1kHz Proximity  fields  from RF  wireless communications equipment IEC 61000-4-3 Refer to table 3 Rated  power frequency  magnetic fields   IEC 61000-4-8 30A/m 50Hz or 60Hz  Table 3 –  Proximity fields from RF wireless communications equipment Test frequency (MHz) Band (MHz) Immunity test levels Professional healthcare facility environment 385 380-390 Pulse modulation 18Hz, 27V/m 450 430-470 FM,  ±5kHz deviation, 1kHz sine, 28V/m 710 704-787 Pulse modulation 217Hz, 9V/m 745 780 810 800-960 Pulse modulation 18Hz, 28V/m 870 930 1720 1700-1990 Pulse modulation 217Hz, 28V/m 1845 1970 2450 2400-2570 Pulse modulation 217Hz, 28V/m 5240 5100-5800 Pulse modulation 217Hz, 9V/m 5500 5785
DN:KD-926-SMSY01 V1.0                                  Operation Guide  17/16 Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the KD-926 The KD-926 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the KD-926 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the KD-926 as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz   80 MHz to 800 MHz   800 MHz to 2,5 GHz   0,01 0.12 0.12 0.23 0,1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.  Pd 2.1Pd 2.1Pd 3.2
DN:KD-926-SMSY01 V1.0                                  Operation Guide  18/16 FCC Caution.  Any Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.  Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -Reorient or relocate the receiving antenna. -Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -Consult the dealer or an experienced radio/TV technician for help.  The device has been evaluated to meet general RF exposure requirement. The device can be used in portable exposure condition without restriction.