BIO CONTROL MEDICAL 0020 PROGRAMMING WAND User Manual RU 53 001 Revk Physician Programmer IFU clean

BIO CONTROL MEDICAL (B.C.M.) LTD. PROGRAMMING WAND RU 53 001 Revk Physician Programmer IFU clean

Users Manual

 BIOCONTROL MEDICAL BCM LTD INSTRUCTIONS FOR USE   DOC NO. RU-53-001  REV. K SUBJECT:  UNIFIED PHYSICIAN PROGRAMMER Page 1 of 40   MODIFICATION INSTRUCTION NO.:  MODIFICATION DATE:  10/03/10 MODIFICATION DESCRIPTION:     Update to PP software and application instructions per ECO 172 and CardioFit 8.17 changes   ISSUED BY  TITLE  DATE  SIGNATURE DUDU RIGLER  DIRECTOR OF CLINICAL AFFAIRS  10/03/10    APPROVAL LIST NAME  TITLE  DATE  SIGNATURE TSACHI CZACZKES  DIRECTOR OF ENGINEERING    SHAI AYAL  DIRECTOR OF DEVELOPMENT    TAMIR SHEFER  DIRECTOR OF QUALITY ASSURANCE    RAMI BIRAN  VP REGULATORY & CLINICAL AFFAIRS     VERSION  NAME  DATE  ECO NO:  MODIFICATION DESCRIPTION A  ARI KEREN-YAAR   06/08/2006  ---  First Issue B  ARI KEREN-YAAR  31/10/2006  ---  Editorial Changes C  ARI KEREN-YAAR   11/12/2006  ---  Addition of buttons; add applicable icons and authorized representative D  YITZHAK SINAI   01/02/2007  136  Update of device and software to 7.14 E  LEVY ITSIK  15/03/2007  ---  Changes According TO ITL Remarks. F  ROMANO AVINOAM  05/05/2008  153  Update to software 1.0.105 G  LEVY ITSIK  15/08/2008  KEMA 2103063 Adding battery lifetime contour; list of abbreviations; clarification of terms and warnings H  DUDU RIGLER  28/12/2008  KEMA 2103063-17.9.08 Addition of follow-up requirements; RF and FCC symbols and text Addition of ICD compliance; updating to software 1.0.14; updating troubleshooting and EMC requirements. I  DUDU RIGLER  24/03/09  167  Layout Review; Editorial Changes; Type BF applied part.
  J  DUDU RIGLER  4/06/09 151 (additional items not identified before) + 165 Update 8.16 version and software 2.0.31:  (1) impedance measurement details (DI#002 in 5.3.3); (2) indication on inhibition due to magnet  application (DI#004 in 5.3.1) (3) estimation on HRR during on time (DI#005, USRM#11 in 5.3.2) (4) histogram of pt. HR (DI#006 in 5.3.2, 5.3.3, 6.6) (5) Explanation on low and high HR thresholds (DI#007/008 in 5.3.3) (6) Clarification of parameter setting in the existence of CRM implantable devices (in 7)  (7) Remove US IDE indications and labeling (in 2) (8) Interrogation includes the last measured RR interval (DI#165/009 in 5.3.1, 5.3.2) (9) Update GUI according to usability report (USRM#1/#4)  K  DUDU RIGLER  10/03/10  172  Layout Review; Editorial Changes; Update 8.176 version and software 2.0.33: (1) Use of USB programming Wand  (2) Communication indicators, USRM#38 (3) Inserting the printer in the system content (4) Training indication per USRM#2 (5) Update GUI screens (6) Impedance rage and accuracy (7) Reduce PPB setting for use without CRMs. (8) Update battery longevity information following 3-years of shelf life (9) Change to FCC class A (10)  Update troubleshooting
      Neurostimulator For Heart Failure Therapy   PHYSICIAN PROGRAMMER MODEL 5300   INSTRUCTIONS FOR USE (IFU)         CAUTION – Investigational Device. Limited By Federal Law To Investigational Use      0344
   Physician Programmer   IFU     RU-53-001 k                                                                                                                              4  Explanation of symbols in labeling Symbol  Explanation  Symbol  Explanation  Contents  Warning: Patient Might Be Compromised SN Serial Number   Caution: Possible damage to device  Product Literature  Note: Important Information  Consult Instructions For Use  Do Not Use If Package Damaged  Type BF Applied Part  Manufacturer  European Approval Mark  Date Of Manufacturing   On   Storage Temperature Range   OFF  Storage & Transport Relative Humidity Range  Stepping Prohibited  Storage & Transport Atmospheric Pressure Range  Sitting Prohibited  Authorized Representative In The European Community  Do Not Dispose Of. Contact BCM For Recycling.  Complies with FCC part 15  Contains RF transmitters
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            5  CONTENTS:  1.  Description  7 2.  Intended Use / Indication  8 3.  Precautions  8 4.  Warnings  9 5.  Setting Up The System  10 5.1  Charging the Physician Programmer  10 5.2  Getting started  10 5.3  General System Interface Description  11 5.3.1  The "General" Tab  12 5.3.2 The "Statistic" tab; Therapy evaluation.  15 5.3.3 The "Setting" Tab  18 6. Therapy Setting and Optimization Program  26 6.1 General  27 6.2 Therapy Optimization Goal  27 6.3 Communication rules with the patient  27 6.4 Visit Activities Flow  27 6.5 Parameter setting process  28 6.6 Instantaneous HRR evaluation and PPB/ON time change  29 7. Interactions With Cardiac Implantable Devices  30 7.1 Implanting CardioFit with another cardiac implanted device in the same patient  30 7.2 Programming device parameters  30 7.3 Patient follow-up  32 8 Maintanance, Handling, and Storage  32 8.1  Maintenance  32 8.2  Handling  33 8.3  Storage  33 9. Troubleshooting  34 10. Product Information and Support  36 11.  Disclaimer Of Warranty  36 Appendix A. Electromagnetic environment – guidance IEC 60601-1-2  37
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            6  List of Abbreviations  Abbreviation  Explanation BCM  BioControl Medical CIS  CardioFit™ Implanted Stimulator  CRT-D  Cardiac Resynchronization Therapy - Implantable Cardioverter - Defibrillators CSL  CardioFit™ Stimulation Lead EMI  Electromagnetic interference HR / BPM  Heart Rate / Beats Per Minute ICD  Implantable Cardioverter-Defibrillators, Single, or Dual chamber  IHRR  Instantaneous Heart Rate Reduction ISE  Intracardiac Sensing Lead NYHA  New York Heart Association PPB  Pulse Per Beat – The number of pulses provided by the stimulator per heartbeat PRI  Pulse Repetition Interval – The distance between pulses when PPT > 1 QTP  Quasitrapezoidal Pulse – Shape of the default pulse RF  Radio Frequency TD  Technical Dossier PP  Physician Programmer VF  Ventricular Fibrillation VT  Ventricular Tachycardia
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            7  1.   DESCRIPTION  Introduction The  model  5300  Physician  Programmer  (PP)  is  used  to  interrogate  and  program  the BioControl  model  5000  CardioFit™  neurostimulator  (the  stimulator).  The  UniPP  can retrieve data  from  and  set programmable  parameters  in  the  CardioFit™  stimulator via wireless  communication.  Programming  is  done  using  the  CardioFit™  Physician Programmer software installed on a PC compatible laptop and a programming Wand.   The Physician Programmer uses the programming Wand to transmit a wireless signal in a  format  that  is  required  for  communication  with  the  stimulator  (Figure  1).  During communication, the Wand is placed over the skin next to the stimulator position.  Figure 1 – Model 5300 Physician Programmer   Specifications and user information regarding the wireless technology The CardioFit system uses a short-range (up to 5 cm) inductive coupling communication scheme in frequency of 27 kHz. The system is ready to communicate when the green light is blinking on the programming wand. See warnings and precautions that should be exercised when using the system under Sections 3 and 4 below.  Screen Printer Programming Wand  Keyboards
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            8  Contents of package The model 5300 Physician Programmer consists of: a.  One Laptop PC installed Model 5310 with BioControl CardioFit™ Unified Physician Programmer software version 2.0; b.  One Programming Wand (Model 5320); c.  One Portable Printer - Model 5330; d.  1 X Product Literatures   2.   INTENDED USE / INDICATION  The model 5300 Physician Programmer is intended for use only with CardioFit™ system, and is subjected to the same indications for use. In Europe, please refer to RU-05-001. When using the device under clinical investigation, please refer to the study protocol.  3.   PRECAUTIONS  •  Do not immerse any part of the Physician Programmer in liquid. •  Physician  Programmer  (excluding  Wand)  should  be  at  a  distance  of  at  least  1.5 meters from the patient during operation.  •  Operator should not touch the PP and patient simultaneously.  •  The  use  of  non-BioControl  components  with  BioControl  components  (other  than supplied  and/or  approved  by  BioControl)  may  result  in  patient  injury,  loss  of stimulation or components damage. •  Avoid using the PP and the BioControl stimulator near sensitive electronic equipment. The low-level wireless signals transmitted by the two devices may interfere with the electronic equipment. •  Compatibility with other (non-BioControl) implanted devices should be verified prior to implantation (see section 7) •  This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant  to Part 15 of the FCC  rules.  These limits are designed to provide reasonable protection against harmful interference in a residential installation. •  This equipment generates; uses and can radiate radio frequency energy and, if not installed  and  used  in  accordance  with  the  instructions,  may  cause  harmful interference  to  radio  communications.    However,  there  is  no  guarantee  that interference will not occur in a particular installation.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            9   4.   WARNINGS •  Physician,  or  other  qualified  healthcare  provider  should  be  trained  in  using  the programmer. Training programs are available from the manufacturer. •  No modification of this equipment is allowed. •  Do not connect any equipment to Laptop except dedicated Programming Wand, and formatted USB Memory Stick. •  To avoid the risk of electric shock, this equipment must only be connected to a supply main with protective earth. •  Pediatric use – safety and effectiveness of Physician Programmer for pediatric use, has not been established. •  When using the programming wand in a sterile field, place it into a sterile sleeve. The wand is not and cannot be sterilized. •  Wireless signals from the Physician Programmer may interfere with the performance of other devices. •  Portable and mobile RF communications equipment can affect the CardioFit. •  Keep cellular phones at least 15 cm away from the device.  The cellular phone is a source of EMI and could affect device’s operation. •  Using the programming wand over a metallic field might decrease its operating range.    CAUTION: This device complies with Part 15 of the FCC Rules.  Operation is subject to the following two conditions: (1) This device may not cause harmful interference; and  (2) This device must accept any interference received, including interference that may cause undesired operation.  FCC WARNINGS: Modifications not expressly approved by the manufacturer could void the user authority  to  operate  the  equipment  under  FCC  Rules  received,  including interference that may cause undesired operation.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            10  5.   SETTING UP THE SYSTEM  5.1   Charging the Physician Programmer      1.  Charge the PP prior to use and make sure it is fully charged.  5.2   Getting started 1.  Verify that the Wand is connected to the programming Wand USB port of the PP.  2.  Turn “ON” the PP. The CardioFit  Physician Programmer "Welcome" screen  will  be automatically uploaded.   Figure 2: Welcome Screen  The  green  led  (ready)  of  the  Wand  will  light,  indicating  proper  initialization  of  the system.  WARNING: Programming the CardioFit stimulator  should  be  done  only in  controlled environments, such an operating room or a clinic
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            11   3.  Position  the  programming  Wand  over  the  patient's  clothes,  directly  over  the implantable stimulator’s location.    WARNING: Avoid direct contact with patient skin – place above thin clothing or use a thin wrapper around the wand to reduce the risk of skin irritation and/or cross contamination between patients  The blue led (communication) of the wand should blink at a constant frequency of 0.5 Hz (every 2 seconds). Such blinking shows proper communication and readiness for receiving and transmitting signals.   4.  Check the UniPP clock. In case of incorrect time, click “CHANGE" button and adjust the time using the keyboard.  5.3   General System Interface Description  Select the “INTERROGATE” (F1) button on the "Welcome" screen. This initial interrogation sets up the Physician Programmer screen  with the  panes and programmable  buttons relevant to the version of the device interrogated. The Programmer interface consists of 4 tabs, which can be freely navigated by pressing on the tab name on the left.  General tab: Shows general device settings and allows recording of patient and system information  Statistics tab: Shows device heart rate statistics and allows for statistics counter reset Settings tab: Main tab used for programming device stimulation parameters. Technician tab: Shows device technical parameters
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            12  5.3.1  The "General" Tab   Figure 3: The General tab following device Interrogation The general tab is divided into 3 panes:  The left pane includes patient and system information: •  Patient Information   Patient name – An editable text view. Up to 14 characters  Patient date of birth  – An editable date field   Date of implantation  – An editable date field   •  CardioFit Implantable Stimulator (CIS)  information (automatically  recognized by the device)  Serial number - Displays the device’s serial number  Firmware version - Device’s software version   Hardware version - Device’s hardware version   Mechanics version- Device’s mechanical version   •  Lead information  Lead serial number  - An editable text view up to 32 characters  Lead model - An editable text view up to 32 characters  Lead size - An editable text view up to 32 characters
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            13  •  Sensing lead information   Serial number - An editable text view up to 32 characters.  Model - An editable text view up to 32 characters.  •  Notes An editable text view which can be used at the physician’s discretion  The right pane includes: •  An overview of current device settings including Activity status: On time, Off time, On cycles, Off cycles, Duty cycle, PPB, Current •  Battery status and estimated running time.  For a detailed description of these parameters, please see section 5.3.3, The Settings tab.  •  Advisory Notes Advisory  notes  will  appear  in  case  the  CIS  has  experienced  one  of  the  following conditions: •  Battery voltage is too low •  Indication of insufficient discharge of stimulation electrode was given •  Indication from runaway protection mechanism was given •  CardioFit performed reset •  Bad memory shadow indication was given •  Magnet was applied to disable stimulation •  High/low impedance indication was given . Actions that should be taken following each advisory note are described in section 9  Pressing the “Clear Advisory Notes” button will clear advisory notes  The bottom pane includes 4 buttons:  Interrogate [F1] Device Interrogation provides current device settings. Interrogation checks the parameters of the device and presents the current values of all programmable buttons and includes last measured RR interval. The interrogate function does not change parameters stored in the stimulator  Send [F3] Press this button in order to communicate and apply the parameter setting changes. It will also perform interrogation and update output parameters such as counters.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            14  Print Report  This button generates a hardcopy report with device parameters. After pressing the ‘PRINT REPORT’ button, the image file will be ready for printing. Single click on the Power button will turn on the printer. Double clicks will turn it off.   Figure 4: A Sample Print Report    Exit Exits the Physician Programmer.    NOTE: Before exiting the PP and ending the visit with the patient, it is recommended to save the device logs onto a USB memory stick for further off- line analysis of the therapy when it is required. Plug the memory stick to the USB port and press Ctrl+Alt+L button on the UniPP Keyboard     WARNING: Format the USB memory stick prior to use with the Physician Programmer  These 4 buttons (Interrogate [F1], Send [F3], Print Report, and Exit) appear also in the setting tab and have similar functions.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            15  5.3.2 The "Statistic" tab; Therapy evaluation.   The UniPP calculates and analyses the effect of the stimulation parameters on the patient's heart, thus enabling evaluation of optimal therapy setting. Measures are presented numerically and graphically. Statistics are accumulated by the CIS. The statistics can be reset manually.  The "Statistic" tab information is typically used at the beginning of the patient visit, in order to estimate the effect of the last prescribed therapy. At the end of the visit, the "Statistic" tab should be revisited in order to reset the data and enable estimation of therapy effect at the next visit.   Figure 5: The Statistic tab  The Statistics tab is divided into 3 panes:  The left Statistics pane includes the Heart Rate (HR) numerical information: • Last reset – Date of the last statistical information reset • Duration – Time period over which the statistics are calculated and presented • Average HR – The average HR of all heart beats detected by the device • Average HR at On time - The average HR of the heart beats detected within the pre-specified HR  detection thresholds (low and high HR) and  calculated  by the device while stimulation was delivered
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            16  • Average HR at Off time - The average HR of the heart beats detected within the pre-specified  detection  limits  (low  and  high  HRs)  and  calculated  by  the  device  while stimulation was not given • Low HR events at On time - A counter registering how many detection events (R waves) occurred at a rate below that defined as Low HR Threshold.  • High HR events at On time - A counter registering how many detection events (R waves) occurred at a rate above that defined as High HR Threshold.  • Low HR events at Off time - A counter registering how many detection events (R waves) occurred at a rate below that defined as Low HR Threshold. • High HR events at Off time - A counter registering how many detection events (R waves) occurred at a rate above that defined as High HR Threshold.  • Actual  Duty  cycle  –  The  ratio  (in  percentage)  between  the  period  at  which stimulation  was  administrated  and  the  period  at  which  stimulation  was  not administrated. • Momentary HR – The value of the last RR interval measured by the device in bpm  The  right  HR  Reduction  pane  includes  graphical  presentation  of  the  average instantaneous heart rate reduction (iHRR) for the time duration indicated in the left pane. The graph provides visual evaluation of the effect of stimulation on the heart rate. The three left points are the average HR for the last 2 beats before the stimulation is switched from  off  to  on  and  the  average  HR  for  the  first  beat  after  switching  from  off  to  on respectively. The three right points are the average HR for the 2 last beats before the stimulation is switched from on to off and the average HR for the first beat after switching from  on to off respectively. The heart rates shown are averaged over  all  on/off cycles that occurred since the last reset. The iHRR value is presented below the figure. It is calculated from the mean value of the first  two  points  (off  stimulation)  minus  the  mean  of  the  following  two  points  (on stimulation). Two  blue  dashed  lines  are  drawn  at  the  values  of  the average  HR  at  On  time  and average HR at Off time.   Zooming in/out Marking the desired area in the graph to be zoomed-in while left clicking the “mouse” will provide focused area of interest. Repeat the procedure for higher magnification if required. Right click will reset the axis scale.    NOTE: The iHRR figure statistic is reliable only if the interval duration (since last reset counters) is greater  than 24 hours   The bottom Heart rate histogram pane shows a graphical presentation of the heart rate histogram for the time duration indicated in the left pane and the functional button.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            17  Retrieve statistics Updates the statistics parameters including update of the HR reduction figure as well as the HR histogram,   Click  the 'PRINT REPORT'  button  after  pressing  the 'RETRIEVE  STATISTICS'  button for patient  history  records.  This  report  documents  the  therapy  performance  since last visit (since last reset counters activation).   Reset counters Resetting the CardioFit statistics counters.  After pressing the button, a message box appears and indicates successful reset action.   Click  the  'PRINT REPORT'  button  after  pressing  the  'RESET COUNTERS' button  for patient  history  records.  This  report  documents  the  end  of  visit  therapy  parameter setting.     NOTE: Pressing “Reset counters” button will not update the screen with reset counters; for the updated numbers please use “retrieve statistics”.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            18   5.3.3 The "Setting" Tab  This screen allows the operator to control the stimulation therapy delivered by the device to  the  patient.  Adjustment  of  HR  detection  and  Vagus  nerve  stimulation  therapy programming and changes are performed through the Setting tab.  The top pane contains a switch screen of IEGM and HR streaming. Clicking the relevant tab on the top of the screen switches between the screens. Description of buttons and panes below the switch screen is found under the HR screen.  IEGM screen The CardioFit detects right ventricle signal in order to synchronize stimulation with the cardiac cycle. The max detection sensitivity = 2mV.   Figure 6: The IEGM Setting tab  Update IEGM (F9) The Update IEGM (F9) button is located at the right pane of the setting screen Pressing the button will update the three windows that compose the IEGM screen. Each window represents one second. Please note: the three windows are not continuous in time.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            19  Zooming in/out The zoom control buttons are located at the left side of the IEGM screen marked by + and –. Pressing  the  +  results  in  IEGM  magnifying,  each  pressing  will  provide  higher magnification, allow better view of the IEGM and ease threshold setting. Pressing the – button will decrease the IEGM magnification.  Stimulation annotation line Appears as a black line below the IEGM and the threshold graph, When stimulation is being delivered an upward pointing bar will be presented, when stimulation is being inhibited a downward pointing bar will be presented, in addition, a letter representing the inhibition cause will appear below the bar as follow: Inhibition letter legends: (also available when pressing the "?" button) O – Off time, Off cycle M – Magnet has been applied. F – Fault (lead status out of range, low battery, bad shadow, device reset) R – Runaway protection has been activated. D – Discharge, in case the electrode has not been sufficiently discharged. H – High HR, short RR. L – Low HR, Low daily Avg. HR I – Impedance not measured P – Pulse S - Sense  IEGM max measured amplitude Below each IEGM window screen, the “Peak” R wave value is presented.  Detection Threshold The threshold for R wave detection appears  on the grid  of the display  as a  bright purple/magenta line that can be set higher or lower by left clicking and dragging or by entering the numeric value.  An optimal threshold setting should be at two-thirds of the R wave peak. If the “visible” box is not checked, the threshold will not be shown. The "Sensitivity" indicator shows the guaranteed sensitivity at the chosen detection threshold  Invert signal Marking the check box will invert the heart signal; use this button when the negative projection of the signal is much greater then the positive.  HR screen The HR screen presents the heart rate in bpm during the last 128 RR intervals with marks for stimulations. The figure enables on-line evaluation of the heart rate response to stimulation.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            20   Figure 7: The HR Setting tab  Update HR [F8] This button is located at the right pane of the setting screen Pressing this button will display a graph of heart rate of the last 128 RR intervals. Every such heartbeat is marked by a green dot on the graph. Stimulated R waves will be marked in larger dots.   Zooming in/out Marking the desired area in the graph to be zoomed-in while left clicking the “mouse” will provide focused area of interest. Repeat the procedure for higher magnification if required. Right click will reset the axis scale.  The bottom pane of the “Setting” tab includes stimulation parameters settings:   Stimulation Settings:  Active Check box may be un-checked in order to inactivate the device. In the ‘non active’ mode the CIS is not sensing or counting and not stimulating. When the box is checked the CIS is activated.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            21   Current  Sets the amplitude of the current provided by the stimulation lead in milliamps (mA). Value can only be selected from a table appearing when button is left-clicked, as a safety measure. Current should not be increased by more than 0.2mA at one time. The default value is 0mA, max value is 6mA.  PPB Pulses Per Beat will set a constant number of pulses per single heartbeat. Max  PPB  value  is  3  in  regular  use  and  8  in  implantations  (only  appear  in  the CardioFit-X version).  The following table specifies stimulation interval duration from at each specified PPB   PPB  actual PPB  Delay From R [ms]  Pulse Repetition Interval [ms] 1  1  94   8 2  2  8   86  3  3  8   39   Blanking  Determines the duration of blanking period following stimulation. The purpose of blanking during which there is no detection, is to prevent T wave detection, as well as to prevent the stimulation itself from causing misdetection. Range: 8 ~ 1000 ms in 7-8ms steps. It appears in the IEGM as uncolored line area.  The default value is 102ms.  Max detectable heart rate The number appearing is the maximum regular beats, or more precisely - Intracardiac signals, per minute that may be detected. This value is the product of several variables, including: PPB, and blanking. The ‘Max detectable HR’ field becomes red if the programmed ‘High HR Threshold’ is greater than 90% of the max detection rate.   WARNING: If the High HR Threshold is higher than the Max detectable HR, the device may not be able to prevent stimulation when it is resulted from detection of noisy signal
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            22  Stimulation Limits:  High HR Threshold The heart rate above which no stimulation is provided no matter what the mode, cycle or stimulation parameters are. Stimulations are inhibited when 3 out of 4 of the last RR intervals are shorter than this value. The default value is 130 BPM.  Low HR Threshold The heart rate below which no stimulation is provided no matter what the mode, cycle or stimulation parameters are. Stimulations are inhibited when 3 out of 4 of the last RR intervals are longer than this value. The default value is 56 BPM.  Daily Average RR This feature cancels the stimulation Cycle mode. It automatically use the calculated last 24 hours average heart rate by the device and enable stimulation only when the measured  heart  rate  is  higher  than  the  average  value.  The  value  is  updated automatically every 24 hours. At the first time used, the value should be set manually. When  the  default  ‘Off’  value  is  presented  the  stimulation  limitation  ignores  this feature.  Intermittent Mode Settings:  Treatment mode  The CardioFit™ stimulation algorithm consists of four combined time intervals, the On/Off Time and the On/Off Cycles.    CardioFit administers stimulation in an intermittent, periodic manner. The period is determined by the "on time" and "off time" programmable parameters. A "stimulation cycle" is the time period consisting of a single "off" period followed by a single "on" period. The "On cycle" and "Off cycles" programmable parameters give further control of the intermittent administration of stimulation by allowing a long period in which no stimulation is administered. CardioFit will administer "On cycle” normal stimulation cycles,  to  be  followed  by  a  period  of  "Off  cycles"  stimulation  cycles  in  which  no stimulations will be administered.  For example: See Figure 8, where the following parameters are programmed:  on time=4 sec, off time=1 sec, on cycle=2, off cycle=3.  Based  on  these  parameters,  the  stimulation  cycle=off  time  +  on  time  =  5  sec. Therefore,  10  seconds  (2  cycles)  of  intermittent  stimulation  are  followed  by  15 seconds of no stimulations, etc.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            23   Figure 8: On/Off time and On/Off Cycle operation parameters.  Duty Cycle The duty cycle indicator calculates and present the actual activation time according to the last programmed settings. The indicator becomes red if the calculated value exceeds 30%. The duty cycle should not exceed 30%.  )()( OffCycleOnCycleOffTimeOnTimeOnCycleOnTimeDutyCycle +⋅+⋅=    NOTE: Values for the treatment mode parameters can only be chosen from a table appearing when button is left-clicked, as a safety measure  Lead Status Calculates  the  CSL  lead  impedance  and  provide  an  indication  status  in  20% accuracy. Working range is 300 – 2500 Ω, 4 – 10µF.   WARNING: Lead status ‘Short’ or ‘High’ may indicate on lead break or damaged insulation and need to be reported immediately to BioControl representative
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            24  The result will be displayed on the Lead status button and will be written in the log files as a comment. To perform the test press the ‘lead status’, the test take a minute and is part of a routine On Time cycle, a progress bar indicate on the test status. In order to perform the test, four conditions must be fulfilled. The conditions are:  On time ≥ 1 On cycle ≥ 1 Active On R wave detection    NOTE: If one of the conditions is not fulfilled the Programmer will issue a warning: "Impedance measurement was not performed. Verify that stimulation is not inhibited".  The lead status shown will be from the last successful test. Stimulation will not be delivered if the lead status results is not “OK”       Battery Status  Battery capacity indicator is shown as a blue strip containing a percentage measure  of full capacity, Battery Voltage is provided in Volt units in parenthesis. Battery expected remaining time is expressed in months.    NOTE: If less than 3 months remain for battery operation (assuming no change in setting) a message will appear in the gray space below “Please schedule device replacement”.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            25   Figure 9: Predicted device life time as function of current and PPT  The nominal predicted lifetime of the CardioFit™ Implantable Stimulator battery in the maximal therapeutic setting is 4.5 years and the minimum is 3.0 years. The nominal predicted lifetime is calculated under the following conditions: Stimulation electrode Impedance: 1740Ω; Capacitance: 4.7µF Stimulation current: 5mA, 3 pulses per beat,  Average HR: 100bpm, Stimulation duty cycle: 25%.  Print Report See section 5.3.1  Add Remarks Adds a line in the log file, on which user remarks may be written.   Emergency button [F12] This button function is to switch the device to an active off mode without the need of pressing the "send" button, use this button when there is need to terminate the device function immediately.  Exit Exits the Physician Programmer and returns to Windows.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            26     NOTE: When a button is pressed or a new parameter is typed a light blue box will appear around it, until the button “Send” [F3] has been pressed. It is possible to program several parameters one after the other and then press the “Send’ button.    Range and default of programmed parameters list:  Parameter  Range  Default On Time [s]  0…(1)…10…(5)…255  0 Off Time [s]  0…(1)…10…(5)…255  0 PPB  1, 2, 3  1 Current [mA]  0…(0.1)…6  0 On Cycles  0…(1)…127  1 Off Cycles  0…(1)…127  0 Low Stimulation Limit [BPM]  30…(4)…320  56 High Stimulation Limit [BPM]  Off, 30…(4)…320  150 Daily Average HR  Off, 30.1 - 7680  0ff Blanking [ms]  8…(8)…1000  102 Detection Threshold [mV]  1.88 to 39.69  39.69 Detection Thresholds sensitivity [mV]   ≥2   39.69  6.  THERAPY SETTING AND OPTIMIZATION PROGRAM   WARNING: The Physician Programmer should be used only by authorized and trained personal     NOTE: Make sure the programming wand is correctly positioned over the stimulator, and a green light+ blue blinking light appears. Hold the programming wand in place until the transmitted parameters had been programmed.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            27     NOTE: After each time a button is pressed or a parameter is typed the “send” button on the screen should be pressed, or alternatively press the keyboard “F3” button.  6.1  General This section provides guidance and recommendation regarding use of the CardioFit Unified Physician Programmer (UniPP) for device activation and parameter setting optimization in patients implanted with the CardioFit™ System during the adjustment period that all implanted subjects undergo in order to provide them with optimal CardioFit therapy. This period is defined as "optimization period". The optimization period starts at weeks 3-4 post operative (depending on patient recovery from system implantation and wound healing), and lasts around 4 to 6 weeks. In order to optimally set the therapy, approximately 6 to 8 setting sessions are required during that period depends on patient adjustment and tolerability to the stimulation.  6.2  Therapy Optimization Goal The goal of the optimization is to adjust. the stimulation parameters to the patient specific physiological  characteristics in order to achieve the optimal  synchronized vagus  nerve stimulation (with the depolarization rate of the ventricles), CardioFit therapy settings. Optimal setting criteria are: •  Instantaneous Heart Rate Reduction ("iHRR") greater than 0.5 bpm  in an interval ≥ 24-hour; and  •  Minimal discomfort to the patient  6.3  Communication rules with the patient •  Prior to each session consult with the patient’s physician regarding his or her condition.  Physical examination is required prior to each optimization session.  •  Ask  the  patient  for  his/her  condition  -  presence  of  adverse  events  or unusual feelings since last visit and during the visit should be noted. •  It  is  recommended  to  use  standard  stimulation  sensation  scale  to standardize patient comfort with the programmed therapy setting level.  6.4  Visit Activities Flow •  Physical examination should be performed prior to parameter change.  •  Review the last patient clinical outcome o  Stimulation discomfort characteristic - Interview patient for his response to therapy and general clinical feeling. Ask patient to determine the stimulation discomfort level since last visit, if any, using the stimulation sensation pain scale  o  Adverse events
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            28  o  Electrocardiogram  from  most  updated  12-lead  ECG/Holter. Identify  any  new  significant  abnormalities  (Heart  rate, conduction, arrhythmias, etc.) •  Interrogate  the  device  and  review  the  information  appeared  in  the "General" tab of the UniPP. Identify any parameter settings discrepancy or system malfunctions. •  Switch to the "Statistic" tab and press the "retrieve statistic" button. Review the  last  interval  heart  rate  outcome  and  compare  it  with  the  previous results.  •  Set therapy parameters •  After  the  last  parameter  setting  change,  follow  the  patient  for  any discomfort for at least 30 minutes before release.  6.5  Parameter setting process •  Use the "Setting" tab in the UniPP for setting parameters •  Verify that "active" button is marked •  HR wave detection: Press the IEGM tab to view IEGM and than press the "Update IEGM" [F9]. Verify correct signal threshold for detection. Detection  thresholds  should  be  set  to  the  2/3  amplitude  of  R  wave.  Minimum sensitivity is 2 mV. •  At  the  first  activation  of  the  device  set  the  parameters  as  indicated  in following table. Parameter  Value On Time [s]  10 Off Time [s]  15 PPB  1 Current [mA]  0.1 On Cycles  30 Off Cycles  20 Low Stimulation Limit [BPM]  56 High Stimulation Limit [BPM]  130 Daily Average HR  Off Blanking [ms]  70  •  Increase  current  in  steps of  0.2 mA  and press  "Send [F3]"  button.  Allow patient to adjust the change for 5 minutes. •  When  the  patient  detects  stimulation,  ask  him/her  to  determine  the stimulation sensation level. If the sensation is intolerable, wait for at least 15  minutes  before  proceeding.  In  case  that  sensation  level  does  not decrease,  reduce  current  in  0.1  mA  steps  until  comfortable  sensation  of stimulation  is  achieved.  Set  the  final  parameters  for  the  optimization session accordingly.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            29  •  Continue with the same procedure at each optimization session up to 5.5 mA and continuously evaluate the effect of the stimulation setting on iHRR.  6.6  Instantaneous HRR evaluation and PPB/ON time change iHRR can be observed at the "Statistics" tab: HR statistics, HR reduction figure and HR distribution histogram for the last interval and in the "Setting" tab: HR figure. •  ON Time settings change: o  Review the HR reduction figure appears in the "Statistic" tab after pressing the "Retrieve" button at the beginning of each session. o  Increase ON time, if the HR value at the last measuring point of the Stimulation ON time is lower or equal to the HR value of the previous measuring point and the average HR. o  Decrease ON time, if the HR value at the last measuring point of the Stimulation ON time is higher than the HR value of the previous measuring point and the average HR. o  Please note that whenever On time is modified, the Off time should  be  modified  accordingly  in  order  to  keep  the  overall Duty Cycle constant. •  PPB setting change; When significant iHRR in an interval ≥ 24-hours is observed switch setting to 2 PPB. o  The change may require reduction of 0.5mA in pulse current  o  Re-evaluate  the  iHRR  in  the  following  visit.  In  case  the measured iHRR value decreased following the change, revert to the last setting with the 1 PPB. •  It is mandatory to print report at the beginning and completion of visit, all reports should be reviewed and signed by the patient physician.     NOTE: Press the EXIT key after parameter setting is completed
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            30  7.  INTERACTIONS WITH CARDIAC IMPLANTABLE DEVICES  7.1  Implanting CardioFit with another cardiac implanted device in the same patient  This section provides the procedure that is to be followed when implanting CardioFit in a patient  with  an  implanted  cardiac  device  (e.g.  pacemaker,  bi-ventricular  pacemaker and/or  defibrillator)  under  the  approved  use  of  the  CardioFit  within  the  specific geography. Following this procedure shall minimize the risk of interaction between the two implanted devices and will allow safe delivery of therapy from both of them.          Before implantation: •  Before implanting one or both of the devices, ensure that both devices can communicate with their respective programmers: o  Place both devices at the same distance at which they will be implanted in the patient:    Do not remove the devices out of their sterile packages    If one of the devices is already implanted, place the other device (still inside the sterile package) on the patient on the area where it is to be implanted   Initiate communication with both devices and ensure communication is error free   If one device’s programming wand interferes with the other devices programming wand, repeat the test with only one programming wand in the device area – both devices should be able to communicate with their programmer while the other programmer is not near •  If both devices can communicate proceed with implantation  7.2  Programming device parameters After both devices are implanted, use the following guidelines to set pacing/sensing parameters in the devices. These guidelines minimize the possibility of CardioFit and the other device sensing each other’s stimulations as cardiac events.  Note: It is possible and common for a device to be both anti-bradycardia and anti-tachycardia (e.g. a CRT/D). In this case guidelines for both types of modes should be followed    NOTE: Please note that the term ‘ICD’ refers to: Single chamber Implantable Cardioverter-Defibrillators, Dual chamber Implantable Cardioverter-Defibrillators and Cardiac Resynchronization Therapy- Implantable Cardioverter-Defibrillators (CRT-D)
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            31  For several parameters below, conversion from beats per minute (bpm) to milliseconds (ms) is needed. Use the following formula:  ms = 1000*60/bpm bpm = 1000*60/ms  •  Never use uni-polar sensing in a device implanted with CardioFit •  Set the CardioFit sensing threshold to be at least 50% of the R-wave amplitude •  Ensure that the other device has a refractory period of at least 135 ms following a sensed ventricular event o  The 135 ms minimal refractory period should be set in all leads used for sensing by the other device. For devices that sense both in the atrium and ventricle, both the atrial refractory period after ventricular sensing (e.g. Post-Ventricular Atrial Blanking) and the ventricular refractory period after ventricular sensing should be greater than 135 ms  •  If the other device is an anti-bradycardia device: o  CardioFit Low HR threshold shall be set to be above the other device’s basic pacing rate    CAUTION: In case that CardioFit low HR threshold is set below the other device’s basic pacing rate, the CardioFit will stimulate on paced beats. The physician should take this into account when programming CardioFit stimulation parameters, or should consider abolition of CardioFit implantation  o  Other device is using bi-ventricular pacing   The other device’s and CardioFit’s parameters shall be set to values such that the other device’s V-V delay is smaller than the CardioFit maximal detectable heart rate Example: Other device’s V-V delay = 50 ms 130 ms = 1200 bpm CardioFit Maximal detectable heart rate should be smaller than 1200 bpm •  If the other device is an anti-tachycardia device o  The other device’s and CardioFit’s parameters shall be set to values such that the CardioFit High HR threshold is well below the other device’s VT zone Example: Other device’s VT zone = 180 bpm CardioFit High HR threshold should be below 160 bpm o  The other device’s and CardioFit’s parameters shall be set to values such that the CardioFit High HR threshold is well below the other device’s anti-tachycardia pacing interval
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            32  Example: Other device’s anti-tachycardia pacing interval= 200 ms 200 ms = 300 bpm CardioFit High HR threshold should be below 280 bpm    7.3  Patient follow-up During patient follow-up special attention should be paid to ensure that the above guidelines are still in effect. Also, any irregular events should be scrutinized to ensure that they did not occur due to device-device interactions. Such events may include: •  Defibrillation shocks •  Pacemaker detection problems (double-counting) •  CardioFit detection problems  •  Communication problems It is recommended that any record of the IEGM from such an event be inspected to ensure that no device-device interaction had occurred.    NOTE: Both devices, ICD and the CardioFit cannot be interrogated simultaneously    WARNING: It is advised to ensure proper sensing by the CardioFit of intrinsic and paced beats   8  MAINTANANCE, HANDLING, AND STORAGE  8.1   Maintenance Except for PP battery charging and cleaning of the Programming Wand the Physician Programmer does not require maintenance:  •  Do not sterilize the programming wand, use only disinfecting pre-moistened cloth and wipe the Wand surface. •  Return the UniPP to BioControl if repair or replacement is required. •  If required, the Physician Programmer components can be cleaned with a moistened cloth.   NOTE: For additional PC laptop and printer operation and maintenance instructions  please refer to the vendors manuals included in this package
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            33  8.2   Handling •  Never immerse the programming wand in liquid. •  Place thermal A4 paper in printer .  8.3   Storage •  Do not store the Physician Programmer where it might be exposed to water or other liquids.  Moisture can damage its functionality! •  Store the Physician Programmer between 0ºC and 45ºC. Temperature outside this range can damage the Physician Programmer. •  Relative Humidity (RH): Between 10% and 95% RH.  •  Atmospheric Pressure: Between –400m and 3000m (700-1060 hPa). •  Store the Physician Programmer in a secure place.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            34   9. TROUBLESHOOTING  The following table presents possible errors and suggested solutions:   Error  Feature  Suggestion Blue light not blinking Check the green light – if the green light is off take the following actions: 1.  Start the Physician programmer. 2.  Verify that the Physician Programmer and programming wand are properly connected. 3.  disconnect the programmer from mains  If the green light is on take the following actions: 1.  Reposition wand 2.  Ask patient to change his/her position. 3.  Check physician programmer operation on a neutral CardioFit™-X device. 4.  Change the programming wand. 5.  change location together with the patient 6.  In case the problem did not resolved contact BioControl Medical (BCM) Ltd.  Communication error  Blue light blinking in high frequency 1.  Check if the CardioFit programming wand is located in the proximity of another programming wand (ICD, CRT). 2.  Try to isolate the CardioFit programming wand from others electrical devices 3.  Contact BioControl Medical (BCM) Ltd. Physician Programmer not responding or a persistent malfunction not related to communication. 1.  Restart the program. 2.  If the problem is not resolved, restart the system. 3.  Contact BioControl Medical (BCM) Ltd.     CAUTION: In case of unsuccessful Turn-off of the neurostimulator while using the Physician Programmer, use safety magnet to turn the neurostimulator OFF. Please contact BioControl Medical personnel, immediately.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            35     The following table presents possible error messages:   Message  Description  Action "A different CardioFit was interrogated, do you wish to proceed?"   After   interrogation, the serial number of the device is different from the last SN. Press YES in case the new device (patient) interrogation was scheduled or press NO and interrogate again the right device in the case of a mistake.   "There is no communication"  Indicates a communication error.  See the table above. "Unfamiliar device or old device version or communication error" Unexpected model or version was read, may indicate an old device version. Re-interrogates the device, if the massage still appears contact BioControl Medical. "Unable to write to a log file"  In case of an error in writing to a log file.  Make sure all log files are closed. "Unable to read from the log file"  A problem in reading the log file.  Make sure all log files are closed. “Disk on key is not connected, operation canceled! “ Disk on key should be connected in order to perform this action. Connect a Disk-On-Key to the programmer. "The operation cannot be performed when device is not active" The desired action requires an active CardioFit. Appears when updating IEGM or HR while the device is set to “active off”, set the device to “Active on”. "Please schedule device replacement" When CardioFit longevity is less than 3 months.  Schedule device replacement.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            36  10. PRODUCT INFORMATION AND SUPPORT  For technical support for the CardioFit™ system Physician Programmer or any of its accessories, please contact BioControl:  BioControl Medical (B.C.M) Ltd. 3 Geron St. Yehud, 56100           ISRAEL              TEL: + 972-3-6322126    FAX: +972-3-6322125     Email: info@biocontrol.co.il   11.  DISCLAIMER OF WARRANTY  Although the Physician Programmer hereafter referred to as “Product” has been carefully designed, manufactured and tested prior to use, the Product may fail to perform its intended function satisfactorily for a variety of reasons. Warnings contained in the Product labeling provide more detailed information and are considered an integral part of this disclaimer. BioControl, therefore, disclaims all warranties, both expressed and implied, with respect to the Product. BioControl shall not be liable for any incidental or consequential damages caused by any use, defect or failure of the Product, whether the claim is based on warranty, contract, tort or otherwise.  The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this disclaimer of warranty is held by any court of competent jurisdiction to be illegal, unenforceable or in conflict with applicable law, the validity of the remaining portion of the disclaimer of warranty shall not be affected and all rights and obligations shall be construed and enforced as if this disclaimer of warranty did not contain the particular part or term held to be invalid.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            37  Appendix A. Electromagnetic environment – guidance IEC 60601-1-2  Guidance and manufacturer’s declaration – electromagnetic emissions The Physician Programmer model 5300 & CardioFit-X Ver 8.17 are intended for use in the electromagnetic environment specified below. The customer or the user of the Physician Programmer model 5300 & CardioFit-X ver 8.17 should assure that it is used in such an environment. Emissions test  Compliance  Electromagnetic environment - guidance RF emissions CISPR 11  Group 2  The Physician Programmer model 5300 & CardioFit-X ver 8.17 must emit electromagnetic energy in order to perform their intended function. Nearby electronic equipment may be affected. RF emissions CISPR 11  Class B Harmonic emissions IEC 61000-3-2  Class D Voltage fluctuations/flicker emissions IEC 61000-3-3 Complies The Physician Programmer model 5300 & CardioFit-X ver 8.17 are suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.9  Guidance and manufacturer’s declaration – electromagnetic immunity The Physician Programmer model 5300 & CardioFit-X Ver 8.17 are intended for use in the electromagnetic environment specified below. The customer or the user of the Physician Programmer model 5300 & CardioFit-X Ver 8.17 should assure that it is used in such an environment. Immunity Test  IEC 60601 test level  Compliance level  Electromagnetic environment – guidance Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact  ±8 kV air ±6 kV contact  ±8 kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. Electrical fast transient/burst  IEC 61000-4-4. ±2 kV for power supply lines  ±1 kV for input/output lines ±2 kV for power supply lines  ±1 kV for input/output lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000-4-5  ±1 kV differential mode  ±2 kV common mode ±1 kV differential mode  ±2 kV common mode Mains power quality should be that of a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11. <5 % UT  (>95 % dip in UT) for 0,5 cycle  40 % UT  (60 % dip in UT) for 5 cycles  70 % UT (30 % dip in UT) for 25 cycles  <5 % UT (>95 % dip in UT) for 5 sec <5 % UT  (>95 % dip in UT) for 0,5 cycle  40 % UT  (60 % dip in UT) for 5 cycles  70 % UT (30 % dip in UT) for 25 cycles  <5 % UT (>95 % dip in UT) for 5 sec Mains power quality should be that of a typical commercial or hospital environment. If the user of the Physician Programmer model 5300 & CardioFit-X Ver 8.17 requires continued operation during power mains interruptions, it is recommended that the Physician Programmer model 5300 & CardioFit-X Ver 8.17 be powered from an uninterruptible power supply or a battery Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 3 A/m  3 A/m  Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE UT is the a.c. mains voltage prior to application of the test level.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            38   Guidance and manufacturer’s declaration – electromagnetic immunity The Physician Programmer model 5300 & CardioFit-X Ver 8.17 are intended for use in the electromagnetic environment specified below. The customer or the user of the Physician Programmer model 5300 & CardioFit-X Ver 8.17 should assure that it is used in such an environment. Immunity Test  IEC 60601 test level  Compliance level  Electromagnetic environment – guidance       Portable and mobile RF communications equipment should be used no closer to any part of the Physician Programmer model 5300 & CardioFit-X Ver 8.17, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.   Recommended separation distance Conducted RF IEC 61000-4-6  3 Vrms 150 kHz to 80 MHz  3 V  d = 1.2 √P Radiated RF IEC 61000-4-3  3 V/m 80 MHz to 2,5 GHz  3 V/m  d = 1.2 √P  80 MHz to 800 MHz d = 2.3 √P  800 MHz to 2,5 GHz        where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).  Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b  Interference may occur in the vicinity of equipment marked with the following symbol:  NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Physician Programmer model 5300 & CardioFit-X Ver 8.17 are used exceeds the applicable RF compliance level above, the Physician Programmer model 5300 & CardioFit-X Ver 8.17 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Physician Programmer model 5300 & CardioFit-X Ver 8.17 b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
   Unified Physicain Programmer   IFU     RU-53-001 k                                                                                                                            39   Recommended separation distances between portable and mobile RF communications equipment and the Physician Programmer model 5300 & CardioFit-X Ver 8.17 The Physician Programmer model 5300 & CardioFit-X Ver 8.17 are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Physician Programmer model 5300 & CardioFit-X Ver 8.17can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Physician Programmer model 5300 & CardioFit-X Ver 8.17 as recommended below, according to the maximum output power of the communications equipment.  Separation distance according to frequency of transmitter m Rated maximum output power of transmitter W  150 kHz to 80 MHz d = 1.2 √P 80 MHz to 800 MHz d = 1.2 √P 800 MHz to 2,5 GHz d = 2.3 √P 0.01  0.12  0.12  0.23 0.1  0.38  0.38  0.73 1  1.2  1.2  2.3 10  3.8  3.8  7.2 100  12  12  23 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.  NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.  NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
              BioControl Medical BCM Ltd.  3 Geron Street, P.O.B 2713 Yehud 56217, ISRAEL Tel: +972 3 6322126  Fax: +972 3 6322125 Email: info@biocontrol.co.il  The  system  and  its  use  for  treatment  of  cardiological  pathologies  is protected  under  one  or  more  of  United  States  patents  nos.  6,600,954, 6,684,105,    7561922,  7627384,  7634317  and  7,346,398,  as  well  as corresponding international patents. Other patents are pending.   CAUTION - Investigational Device. Limited By Federal Law To Investigational Use. Hereby,  BioControl  Medical  BCM  Ltd.,  declares  that  CardioFit™ Neurostimulator  is  in  compliance  with  the  essential  requirements  and other relevant provisions of Directive 1999/5/EC    © BioControl Medical BCM Ltd. All rights reserved  MedNet GmbH Borkstrasse 10 48163 Münster, Germany Tel: +49 2513 22660, Fax: +49 2513 226622 www.medneteurope.com  Year of authorization: 2008  0344

Navigation menu