BIOTRONIK SE and KG TACHBORAX implantable cardioverter defibrillator User Manual
BIOTRONIK SE & Co. KG implantable cardioverter defibrillator
15_TACHBorax UserMan
Inlexa 1 VR-T, DR-T, HF-T ICD-Familie | Tachyarrhythmietherapie | Kardiale Resynchronisationstherapie Gebrauchsanweisung 420651 Revision: A (2015-12-09) © BIOTRONIK SE & Co. KG Alle Rechte vorbehalten. Technische Änderungen vorbehalten. ® Alle verwendeten Produktnamen können Marken oder eingetragene Marken von BIOTRONIK oder dem jeweiligen Eigentümer sein. Index 420651GebrauchsanweisungInlexa 1 VR-T, DR-T, HF-T, BIOTRONIK SE & Co. KG Woermannkehre 1 12359 Berlin · Germany Tel +49 (0) 30 68905-0 Fax +49 (0) 30 6852804 sales@biotronik.com www.biotronik.com 1 Product Description Product Description1403736Technical Manual Intended Medical Use Intended use Inlexa 1 is part of a family of implantable cardioverter-defibrillators (ICDs). The primary objective of the therapy is to prevent sudden cardiac death. Furthermore, the device is capable of treating bradycardia arrhythmias and cardiac resynchronization therapy with multisite ventricular pacing. The implantation of an ICD is a symptomatic therapy with the following objectives: • Termination of spontaneous ventricular fibrillation (VF) through shock delivery • Termination of spontaneous ventricular tachycardia (VT) through antitachycardia pacing (ATP); in case of ineffective ATP or hemodynamically not tolerated VT, with shock delivery • Cardiac resynchronization through multisite ventricular pacing (triple-chamber devices) • Compensation of bradycardia through ventricular (single-chamber devices) or AV sequential pacing (dual- and triple-chamber devices) Diagnosis and therapy forms Required expertise Indications The device monitors the heart rhythm and automatically detects and terminates cardiac arrest resulting from ventricular tachyarrhythmia. All major therapeutic approaches from the field of cardiology and electrophysiology are included. BIOTRONIK Home Monitoring® enables physicians to perform therapy management at any time. In addition to having basic medical knowledge, the user must be thoroughly familiar with the operation and the operation conditions of a device system. • Only qualified medical specialists having this required special knowledge are permitted to use implantable devices. • If users do not possess this knowledge, they must be trained accordingly. Inlexa 1 can treat life-threatening ventricular arrhythmias with antitachycardia pacing and defibrillation. Generally approved differential diagnostics methods, indications, and recommendations for ICD therapy apply to BIOTRONIK devices. See the current guidelines of cardiology associations for guidance. We recommend observing the indications published by the German Cardiac Society (Deutsche Gesellschaft für Kardiologie, Herz- und Kreislaufforschung) and the ESC (European Society of Cardiology). This also applies to the guidelines published by the Heart Rhythm Society (HRS), the American College of Cardiology (ACC), the American Heart Association (AHA), and other national cardiology associations. Single-chamber and dual-chamber Triple-chamber Single-chamber and dual-chamber ICDs are indicated for patients with the following risk: • Sudden cardiac death caused by ventricular arrhythmias Triple-chamber ICDs are indicated for patients with the following risks: • Sudden cardiac death caused by ventricular arrhythmias • Congestive heart failure with ventricular asynchrony 2 Contraindications Known contraindications: • Tachyarrhythmia caused by temporary or reversible irritation, e.g. poisoning, electrolyte imbalance, hypoxia, sepsis or acute myocardial infarction • Such frequent VT or VF that the therapies would cause an unacceptably rapid depletion of the device batteries • VT with few or without clinically relevant symptoms • VT or VF treatable by surgery • Concomitant diseases that would substantially limit a positive prognosis • Accelerated intrinsic rhythm System Overview Device family The complete Inlexa 1device family consists of several device types with a DF-1/IS-1 connection. • Single-chamber: VR-T • Dual-chamber: DR-T • Triple-chamber: HF-T Note: Not all device types are available in every country. Device The device's housing is made of biocompatible titanium, welded from outside and thus hermetically sealed. The ellipsoid shape facilitates implantation in the pectoral muscle area. The connections for bipolar pacing and sensing (and unipolar connections for the triple-chamber device) as well as for shock delivery are found in the device header. The housing serves as a potential antipole during shock delivery or in the case of unipolar lead configuration. DF-1/IS-1 The labeling on each device provides information pertaining to the connector port assignment in the header. VR DR DF-1 SVC DF-1 RV DF-1 SVC IS-1 RA DF-1 RV IS-1 RV IS-1 RV HF Connector Lead Configuration port connector Leads IS-1 LV IS-1 RA IS-1 RV DF-1 SVC DF-1 RV Implantation site Device type RA IS-1 Bipolar Atrium DR, HF (R)V IS-1 Bipolar (Right) ventricle VR, DR, HF RV DF-1 Shock coil Right ventricle VR, DR, HF SVC DF-1 Shock coil Superior vena cava VR, DR, HF LV IS-1 Unipolar, bipolar Left ventricle HF BIOTRONIK leads are sheathed with biocompatible silicone. They can be flexibly maneuvered, are stable long-term, and are equipped for active or passive fixation. They are implanted using a lead introducer set. Some leads are coated with polyurethane which is known to increase the sliding properties for the lead. Leads with steroids reduce inflammatory processes. The fractal design of the electrodes provides for low pacing thresholds. BIOTRONIK provides adapters to connect already implanted leads to new devices. 3 Telemetry Programmer Telemetric communication between the device and the programmer can be carried out following initialization either by applying the programming head (PGH) to the device or by using wireless radio frequency (RF) telemetry in the programmer. BIOTRONIK calls this function SafeSync®. Implantation and follow-up are performed with BIOTRONIK's portable programmer: Programmer software PSW version N.N. and higher There is a programmer with integrated RF telemetry and one with a separate SafeSync Module. Leadless ECG, IEGM, markers and functions are displayed simultaneously on the color display. Using the programmer, the pacing thresholds can be determined and all tests can be performed during in-office follow-up. If necessary, the current software is transferred to the device during implantation. In addition to this, the programmer is used to set mode and parameter combinations, as well as for interrogation and saving of data from the device. Modes The mode setting depends on the individual diagnosis: Device type NBD and NBG codes Modes VR VVI; VVIR; VOO; OFF DR, HF DDD; DDDR; DDDR-ADIR; DDD-ADI; DDI; DDIR; VDD; VDDR; VDI; VDIR VVI; VVIR; AAI; AAIR; VOO; DOO; OFF VVE is the NBD code for the antitachycardia mode of the single-chamber, dualchamber, and triple-chamber devices: Shock in the ventricle Antitachycardia pacing (ATP) in the ventricle Detection via IEGM analysis DDDR is the NBG code for the antibradycardia mode of the dual-chamber devices: Pacing in the atrium and ventricle Sensing in the atrium and ventricle Pulse inhibition and pulse triggering Rate adaptation DDDRV is the NBG code for the antibradycardia mode of the triple-chamber devices: Pacing in the atrium and ventricle Sensing in the atrium and ventricle Pulse inhibition and pulse triggering Rate adaptation Multisite pacing in both ventricles VVIR is the NBG code for the antibradycardia pacing modes of the single-chamber device: Ventricular pacing Sensing in the ventricle Pulse inhibition in the ventricle Rate adaptation 4 BIOTRONIK Home Monitoring® In addition to effective pacing therapy, BIOTRONIK provides a complete therapy management system: • With Home Monitoring, diagnostic and therapeutic information as well as technical data are automatically sent to a stationary or mobile transmitter via an antenna in the device header. The data are encrypted and sent from the transmitter to the BIOTRONIK Service Center via the cellular phone network. • The received data are deciphered and evaluated. Each physician can set the criteria for evaluation to be used for each patient and can configure the time of notification via e-mail, SMS or fax. • A clear overview of the results of this analysis is displayed for the attending physicians on the protected Internet platform Home Monitoring Service Center (HMSC). • Data transmission from the device is performed with a daily device message. • Device messages which indicate special events in the heart or in the device are forwarded immediately. • A test message can be initiated at any time using the programmer to immediately check the Home Monitoring function. Inlexa 1order numbers Not all device types are available in every country: Package contents VR-T DR-T HF-T 405797 405796 40579 The storage package includes the following: • Sterile packaging with device • Serial number label • Patient ID card • Warranty booklet Note: The technical manual pertaining to the device is either included in hard copy form in the storage package or in digital form on the internet. The sterile container includes the following: • Device, blind plugs (if applicable) • Screwdriver 5 Therapeutic and Diagnostic Functions Diagnostic functions • • • • Antitachycardia pacing • • Cardioversion, defibrillation • • Antibradycardia pacing and CRT • • • • • • Storing programs Data from implantation and the most recent interrogations and follow-ups are recorded as well as arrhythmia episodes; they are stored together with other data to assess both the patients' and the device's state at any time. To check the lead for proper functioning, an automatic impedance measurement using subthreshold pacing pulses is performed in the device. Leadless ECG function: For all device types, far-field derivation can be measured without external leads between the right ventricular shock coil and housing, which, depending on the implantation site, corresponds to ECG derivation II or III (Einthoven). Once a telemetry connection has been established during a test procedure in an in-office follow-up, the leadless ECG and the IEGM are displayed with markers. The ICD can treat ventricular tachycardia with antitachycardia pacing (ATP); ATP can also be delivered in the VF zone (ATP One Shot) when the stability criterion indicating that this will be effective before shock delivery (monomorphic rapid VTs) is met. Depending on the device type, the device program contains not only the ICD functions but also all pacemaker functions for 1, 2 or 3 chambers. The heart rhythm is continuously monitored; each arrhythmia is classified according to the heart rate and the adjustable detection criteria. Depending on the preset values, antibradycardia as well as antitachycardia therapy is inhibited or delivered. The ICD can treat ventricular tachyarrhythmia with cardioversion and/or defibrillation. Shock polarity and energy can be programmed individually. Shock energies between 2.0 and 40 J are possible depending on the device family. Before delivery of the shock, the ICD can be set to only deliver a shock when ongoing tachyarrhythmia is confirmed; during this time period the device can identify spontaneous conversion of the tachyarrhythmia and cancel the charging process if necessary. The shock paths can be set between the different shock coils (SVC/RV) and/or the housing. Innovative rate hystereses, automatic sensor functions, and a night program promote the patient's intrinsic rhythm, avoid overdrive pacing, and facilitate adaptation of the device to the individual needs of the patient. Thresholds: atrial as well as ventricular pacing thresholds are automatically determined in the device, automatic threshold monitoring (ATM) for trend analysis. Setting an upper tracking rate for the atrium prevents unspecific atrial pacing, thus reducing the risk of pacemaker-mediated tachycardia. Positive AV hysteresis functions support intrinsic conduction and thus the natural contraction sequence. Negative AV hysteresis functions support the cardiac resynchronization therapy by maintaining pacing in stress situations. For resynchronization of the ventricles, triple-chamber implants have functions for multisite pacing with possible VV delays in either direction. To ensure that no additional surgery is necessary in case of a left-sided increase of pacing threshold or undesired phrenic nerve stimulation, different pacing polarities can be set for the left ventricular lead with a triple-chamber device. There are different therapy programs: • Parameter settings effective for the most common indications in pre-configured programs (Program Consult). • For special indications, individual parameter settings can be stored in up to three therapy programs. 6 Home Monitoring functions • • • The device automatically sends information to the transmitter once a day. It also sends messages related to events, which are immediately forwarded to the Service Center. In addition to this, test messages can be initiated using the programmer. Appointments for Home Monitoring-supported follow-ups can be scheduled via the HMSC. Important medical information in the device messages include the following: — Atrial and ventricular arrhythmias — Parameters relevant to leads in the atrium and ventricle: pacing thresholds, sensing amplitudes, impedances — Current statistics — IEGM online HD® with up to 3 high definition channels 7 General Safety Instructions General Safety Instructions2403736Technical Manual Operating Conditions Technical manuals The following technical manuals provide information about usage of the device systems: — Technical manual for the device — Technical manual for the HMSC — Technical manual for the electrodes — Technical manuals for the programmer and its accessories — Technical manuals for the user interface — Technical manuals for cables, adapters and accessories • Technical manuals are either included in hard copy form in the storage package or in digital form on the internet: manuals.biotronik.com • Follow all relevant technical manuals. • Reserve technical manuals for later use. Care during shipping and storage • • Delivery in shipment mode Temperature Sterile delivery Sterile packaging Single use only Devices must not be stored or transported close to magnets or sources of electromagnetic interference. Note the effects of the storage duration; see Battery Data. The device is delivered in shipment mode to protect the battery; capacitor reforming required during storage could result in controlled extended charge times of the shock capacitors. • The shipment mode is displayed on the programmer after the initial interrogation (it is deactivated during implantation by the first valid (in-range) measurement of the pacing impedance). Extremely low and high temperatures affect the service time of the battery in the device. • Temperatures of +5°C to +45°C are permitted for transport, storage, and use. The device and the screwdriver have been gas-sterilized. Sterility is guaranteed only if the blister and quality control seal have not been damaged. The device and screwdriver are packaged in two separately sealed blisters. The inner blister is also sterile on the outside so that it can be transferred in a sterile state during implantation. The device and screwdriver are intended for single use only. • Do not use the device if the package is damaged. • The device must not be resterilized and reused. 8 Possible Complications General information on medical complications Complications for patients and device systems generally recognized among practitioners also apply to BIOTRONIK devices. • Normal complications may include fluid accumulation within the device pocket, infections, or tissue reactions. Primary sources of complication information include current scientific and technological knowledge. • It is impossible to guarantee the efficacy of antitachycardia therapy, even if the programs have proven successful during tests or subsequent electrophysiological examinations. In rare cases the set parameters may become ineffective. It is possible for therapies to induce or accelerate tachycardia and cause sustained ventricular flutter or fibrillation. Skeletal myopotentials Bipolar sensing and control of sensitivity are adapted by the device to the rate range of intrinsic events so that skeletal myopotentials are usually not recorded. Skeletal myopotentials can nonetheless be classified as intrinsic events especially at very high sensing sensitivity and, depending on the interference, may cause inhibition or antiarrhythmia therapy. In the case of undesired myopotentials, the device switches to asynchronous pacing if the interference rate is exceeded. Possible technical failures Technical failure of a device system cannot be entirely ruled out. Possible causes can include the following: • Lead dislodgement, lead fracture • Insulation defects • Device component failures • Battery depletion • Interrupted telemetry Electromagnetic interference (EMI) Any device can be sensitive to interference if external signals are sensed as intrinsic rhythm or if measurements prevent rate adaptation. • BIOTRONIK devices have been designed so that their susceptibility to EMI is minimal. • Due to the intensity and variety of EMI, there is no guarantee for safety. It is generally assumed that EMI produces only minor symptoms, if any, in patients. • Depending on the pacing mode and the type of interference, sources of interference may lead to pulse inhibition or triggering, an increase in the sensordependent pacing rate or asynchronous pacing. • Under unfavorable conditions, for example during therapeutic or diagnostic procedures, interference sources may induce such a high level of energy into the pacing system that the cardiac tissue surrounding the lead tip is damaged. Device behavior in case of EMI In case of electromagnetic interference, the device switches to asynchronous pacing for as long as the interference rate is exceeded. Static magnetic fields The magnetic sensor in the device detects magnetic fields starting at a magnetic flux density of approximately 1.5 mT. Magnetic fields below 1 mT do not affect the sensor. 9 Possible Risks Procedures to avoid Risky therapeutic and diagnostic procedures The following procedures must be avoided, as they may cause harm to the patient or damage the device and, as a result, put the system functionality at risk: • Transcutaneous electrical nerve stimulation • Hyperbaric oxygen therapy • Applied pressures higher than normal pressure If electrical current from an external source is conducted through the body for diagnostic or therapeutic purposes, then the device can be subjected to interference, which can place the patient at risk. Arrhythmia or ventricular fibrillation can be induced during diathermic procedures such as electrocautery, HF ablation or HF surgery. For example, damaging pressure levels may arise during lithotripsy. For example, excessive warming of body tissue near the device system may occur during therapeutic ultrasound. Influences on the device are not always immediately clear. If risky procedures cannot be avoided, the following should be observed at all times: • Electrically insulate the patient. • Switch off the ICD's detection function. The pacemaker function can remain active. The device may need to be switched to asynchronous modes for this. • Do not introduce energy near the device system. • Additionally check the peripheral pulse of the patient. • Monitor the patient during and after every intervention. External defibrillation The device is protected against the energy that is normally induced by external defibrillation. Nevertheless, any implanted device may be damaged by external defibrillation. Specifically, the current induced in the implanted leads may result in necrotic tissue formation close to the electrode/tissue interface. As a result, sensing properties and pacing thresholds may change. • Place adhesive electrodes anterior-posterior or perpendicular to the axis formed by the device to the heart at least 10 cm away from the device and from implanted leads. Radiation therapy The use of radiation therapy must be avoided due to possible damage to the device and the resulting impaired functional safety. If this type of therapy is to be used anyway, prior risk/benefit analysis is absolutely necessary. The complexity of influencing factors such as different sources of radiation, a variety of devices and therapy conditions makes it impossible to issue directives that guarantee radiation therapy without an impact on the device. The EN 45502 standard pertaining to active implantable medical devices requires the following measures during the administration of therapeutic ionizing radiation: • Adhere to instructions for risky therapy and diagnosis procedures. • Shield device against radiation. • After applying radiation, double-check the device system to make sure it is functioning properly. Note: Please contact BIOTRONIK with questions during the risk/benefit analysis. Magnetic resonance imaging Magnetic resonance imaging must be avoided due to the associated high frequency fields and magnetic flux density: Damage or destruction of the device system by strong magnetic interaction and damage to the patient by excessive warming of the body tissue in the area surrounding the device system. 10 Implantation Implantation3403736Technical Manual Implantation Procedure Having parts ready The following parts that correspond to the requirements of the EC Directive 90/385/ EEC are required: • BIOTRONIK device with blind plug and screwdriver • BIOTRONIK leads and lead introducer set — Single-chamber device: one bipolar ICD lead with 1 or 2 shock coils for the ventricle — Dual-chamber device: one bipolar lead for the atrium and one bipolar ICD lead for the ventricle with 1 or 2 shock coils — Triple-chamber device: an additional unipolar or bipolar LV lead • The lead connections DF-1 and IS-1 are permitted. Use only adapters approved by BIOTRONIK for leads with different lead connections or leads from other manufacturers. • BIOTRONIK programmer (with integrated SafeSync RF telemetry or with separate SafeSync Module) and approved cables • External multi-channel ECG device • Keep spare parts for all sterile components. Keeping an external defibrillator ready To be able to respond to unforeseeable emergencies or possible technical failures of the device: • Keep an external defibrillator and paddles or patch electrodes ready. Unpacking the device WARNING Inadequate therapy due to defective device If an unpacked device is dropped on a hard surface during handling, electronic parts could be damaged. • Use a replacement device. • Return the damaged device to BIOTRONIK. • • • Checking parts Peel the sealing paper off of the outer blister at the marked position in the direction indicated by the arrow. The inner blister must not come into contact with persons who have not sterilized their hands or gloves, nor with non-sterile instruments! Take hold of the inner blister by the gripping tab and take it out of the outer blister. Peel the sealing paper off of the sterile inner blister at the marked position in the direction indicated by the arrow. Damage to any of the parts can result in complications or technical failures. • Check for damage before and after unpacking all parts. • Replace damaged parts. • Upon delivery, the tachyarrhythmia therapy function in the ICD is deactivated. The ICD must only be implanted in this state. • Leads must not be shortened. 11 Implantation site Preventing leakage currents • Depending on lead configuration and the patient's anatomy, the ICD is generally implanted subpectorally on the left side. Leakage currents between the tools and the device must be prevented during implantation. • Electrically insulate the patient. Preventing unintentional shock delivery WARNING Shock delivery with activated ICD There is a risk of unintended shock delivery when handling an activated ICD. • Deactivate ICD therapy before touching the device during implantation, device replacement and explantation. Avoiding damage to the header Set screws and blind plugs (if applicable) must be tightened or loosened with care. • Loosen set screws with the supplied screwdriver. Use only BIOTRONIK screwdrivers with torque control! • Do not forcibly pull out the blind plug! • If lead revision is necessary, re-order sterile screwdrivers from BIOTRONIK. Preventing short circuits in the header WARNING Short circuit due to open lead connector ports Connector ports in the header which are open and thus not electrolyte-proof may cause undesired current flows to the body and penetration of body fluid into the device. • Close unused connector ports with blind plugs. Ensure that connector ports are clean In case of contamination during implantation: • Clean lead connectors with a sterile cloth. • Rinse connector port only with sterile water. Overview: Implanting Connecting the device Prepare the vein. Implant the leads, perform the measurements, and fixate the leads. Form the device pocket. Connect the lead connector to the device. Insert the device. Guide the fixation suture through the opening in the header and fixate the device in the prepared device pocket. Close the device pocket. Check the device with standard tests. The lead connectors are connected to the ports in the header of the device: Disconnect stylets and stylet guides. DF-1/IS-1 connection: • Connect the DF-1 connector for the right-ventricular shock coil to RV. • Connect the DF-1 connector for the supraventricular shock coil to SVC. Or connect a subcutaneous array to SVC. 12 DF-1/IS-1 connection: • Connect the bipolar IS-1 connector for the atrium to RA. • Connect the bipolar IS-1 connector for the right ventricle to RV. • Connect the unipolar or the bipolar IS-1 connector for the left ventricle to LV. Push the lead connector into the header without twisting or bending the connector or conductor until the connector tip (on the DF-1 connector) becomes visible behind the set screw block. This indicator can vary depending on the manufacturer of the lead used. If you cannot easily plug the lead connector into the connection: • Use only sterile water as lubricant. If the lead connector cannot be inserted completely, the set screw may be protruding into the drill hole of the set screw block. • Use the screwdriver to perpendicularly pierce through the slitted point in the center of the silicone plug until it reaches the set screw. • Carefully loosen the set screw without completely unscrewing it, so that it does not become tilted upon retightening. Turn the set screw clockwise until torque control starts (you will hear a clicking sound). Carefully withdraw the screwdriver without retracting the set screw. • In case of IS-1 connections with 2 set screws, tighten both screws! • When you withdraw the screwdriver, the silicone plug automatically seals the access to the screw head safely. Keeping distance between leads WARNING Inadequate therapy When leads are not spaced sufficiently apart or are positioned inappropriately, this can lead to far-field sensing or insufficient defibrillation. • The distance between 2 shock coils must be greater than 6 cm. • Tip and ring electrodes must not have contact with each other. Applying the programming head Establishing telemetry contact Activating ICD therapy The programming head (PGH) features a diagram of the device. This is used to assist in positioning the head to ensure proper telemetry. • Make sure the PGH is positioned correctly. The programmer (or the SafeSync Module) can be no more than 3 m from the device; ideally there should be no hindrances between the patient and the programmer. • Switch on RF telemetry on the programmer. • Apply the programming head for about 2 s until successful initialization is displayed on the programmer: The SafeSync symbol is displayed in the navigator and the signal strength is displayed in the status line. • Remove the programming head. • • • • • Load the software that is suitable for the device type in the programmer. Activate ICD therapy. Shipment mode is permanently deactivated once the leads have been connected and initial measurement of the pacing impedance has been performed successfully. The device data are saved. Take precautionary measures while programming. If the device induces tachycardia while programming ATPs or does not deliver adequate therapy in the DFT test: use emergency shock or an external defibrillator. 13 Precautionary Measures while Programming Performing standard tests and monitoring the patient Critical conditions can occur for the patient even during standard tests due to inadequate parameter settings or interrupted telemetry. • Ensure sufficient patient care even during tests. • After the threshold test, check to determine whether the threshold is clinically and technically justifiable. • Continuously monitor the ECG and the patient's condition. • Cancel testing if necessary. Cancelling telemetry Programmer interference or interrupted telemetry during performance of temporary programs (follow-up tests) can result in inadequate pacing of the patient. This is the case if the programmer can no longer be operated due to a program error or a defective touch screen and therefore the temporary program cannot be terminated. Under these circumstances, it is helpful to cancel telemetry, in which case the device automatically switches to the permanent program. • In the case of telemetry with programming head: lift the PGH by at least 30 cm. • In the case of RF telemetry: switch off and reposition the programmer. • Turn off possible sources of interference. Avoiding critical parameter settings No modes and parameter combinations that pose a risk to the patient should be set. • Prior to setting rate adaptation, determine the patient's capacity for exertion. • Check compatibility and effectiveness of parameter combinations after making settings. • When setting atrial therapies after an AT or AF has been detected, note that no ventricular tachyarrhythmia can be detected for the duration of atrial therapy delivery. Checking for electrodes suitable for the shock path Three different shock paths can be set. Two of these form an electrical path to the housing of the implanted device. • For the RV -> SVC shock path, a second shock coil must be available (dual shock coil). Monitoring the patient when setting asynchronous modes The asynchronous modes V00 and D00 can only be set if tachyarrhythmia sensing is deactivated. This would leave the patient without sensing and therefore without ICD therapy. • Continually monitor the patient. • Keep an external defibrillator ready. Setting sensing Manually set parameters can be unsafe. For example, unsuitable far-field protection may impede sensing of intrinsic pulses. • Note automatic sensitivity control. Preventing device-induced complications BIOTRONIK devices feature several functions to prevent device-induced complications to the greatest extent possible: • Measure the retrograde conduction time. • Set PMT protection. • Set the VA criterion. Preventing conduction of atrial tachycardia BIOTRONIK devices feature several functions to prevent conduction of atrial tachycardia to the ventricle(s): • Set mode switching for indicated patients. • Set the upper rate and the refractory periods to prevent abrupt ventricular rate switching. • Prefer Wenckebach response and avoid 2:1 behavior. • Set all parameters so as to prevent constant changing between atrial and ventricular-controlled modes. 14 Note the reduced pulse amplitude due to a battery voltage drop If the rate and amplitude are set very high and the pulse width is set too long at the same time, the battery voltage may temporarily drop so low that the actual pulse amplitude drops well below the selected level. • Continuously check the pacing efficiency using ECG monitoring. Observe when inducing short-term cardiac arrest To permit TAVI (transcatheter aortic valve implantation), the pressure in the heart must be reduced so that the heart valve can be correctly positioned. Intentional cardiac arrest by high-rate pacing (rapid pacing) should be brief, must be tolerated by the patient and can trigger a life-threatening arrhythmia. • Take all necessary precautionary measures and keep required emergency equipment ready. • Continually monitor the patient by ECG. • Complete the TAVI procedure before high-rate pacing ends. Extend the pacing duration if necessary. • Abort the procedure if it is not successfully completed within the maximum pacing duration so that cardiac arrest can be stopped. • Reactivate ICD therapy at a clinically indicated point in time when the TAVI process is completed. Avoiding AV crosstalk When pacing using atrial ATP parameters, atrial pacing pulses can either be conducted into the ventricle or be sensed such that ventricular pacing is prevented. • Check the settings for the presence of crosstalk. • If necessary, temporarily set VVI and a rate for backup stimulation so that no ventricular pulses are prevented. Observing the shock impedance limit Preventing recurrence after therapy shock The implanted device could be damaged if the shock impedance is too low. • The shock impedance must be > 25 Ω. After a therapy shock, pacing can be performed with a post-shock program if there is no intrinsic rhythm. Permanent program Post-shock program DDD DDI, AAI DDI VDD, VDI VDI VVI and OFF VVI • • The following post-shock program parameters can be adjusted: Post-shock duration, basic rate, rate hysteresis, ventricular pacing, LV T-wave protection, triggering, AV delay (fixed, not dynamic) The default settings for the post-shock program are as follows: A and RV: 7.5 V, 1.5 ms LV: settings from the permanent program Phrenic nerve stimulation that cannot be terminated In rare cases, chronic phrenic nerve stimulation cannot be terminated by reprogramming the available left ventricular pacing configuration or using other measures. • Set a right ventricular mode both in the permanent program as well as the ATP, in the post-shock program and for mode switching if need be. Avoiding risks in the case of exclusive LV pacing Lead dislodgement in the case of exclusive left ventricular pacing could pose the following risks: loss of ventricular pacing and ATP therapy, induction of atrial arrhythmias. • Consider sensing and pacing parameters with reference to loss of therapy. • Exclusive LV pacing is not recommended for patients who depend on the device. • Please note that capture control is not available. • In the case of follow-ups and threshold tests, take loss of synchronized ventricular pacing into consideration. • Mode switching and post shock do not permit exclusive LV pacing. Please note the effects when programming mode switching and the post shock parameters. 15 Recognizing lead failure Considering power consumption and service time Automatic impedance measurement is always switched on. • Impedance values that indicate technical failure of a lead are documented in the event list. RF telemetry requires somewhat more power: Consumption during implantation corresponds to approximately 10 days of service time and consumption during a 20-minute follow-up corresponds to approximately 3 days. • Do not establish unnecessary RF telemetry. • After 5 minutes without input, SafeSync switches to the economy mode. • Check the battery capacity of the device at regular intervals. Magnet Response Application of the programming head when ICD therapy is set If a connected programming head is applied and is communicating with the programmer and ICD therapy is permanently set, detection and therapy remain intact except during the diagnostic tests. If ICD therapy is not set as permanent, no therapy is delivered when the programming head is applied. Programming head application When the programming head is applied, time remains for device interrogation and for manual activation or deactivation of the therapy before the device switches back to the previously set permanent therapy mode. The same applies to programming head application to establish RF telemetry contact. Application of a permanent magnet Applying a permanent magnet interrupts detection and therapy of tachycardia events. After 8 hours of this type of deactivation, the device automatically reactivates the therapy functions to prevent accidental permanent deactivation. • If detection interruptions of longer than 8 hours are required, the magnet has to be briefly removed from the device. The 8 hour countdown restarts when the magnet is applied again. • Use BIOTRONIK magnets: type M-50 permanent magnets. 16 Follow-up Follow-up intervals Follow-up with BIOTRONIK Home Monitoring® Follow-ups must be performed at regular, agreed intervals. • The first follow-up should be carried out by the physician using the programmer (in-office follow-up) approximately 3 months after implantation following the lead ingrowth phase. • The next in-office follow-up should be carried out once a year and no later than 12 months after the last in-office follow-up. Monitoring using the Home Monitoring function does not serve to replace regular in-office appointments with the physician required for other medical reasons. Follow-up supported by Home Monitoring can be used to functionally replace inoffice follow-up under the following conditions: • The patient was informed that the physician must be contacted despite use of the Home Monitoring function if symptoms worsen or if new symptoms arise. • Device messages are transmitted regularly. • The physician decides whether the data transmitted via Home Monitoring with regard to the patient's clinical condition as well as the technical state of the device system are sufficient. If not, an in-office follow-up needs to be carried out. Possible early detection due to information gained via Home Monitoring may necessitate an additional in-office follow-up. For example, the data may indicate at an early stage lead problems or a foreseeable end of service time (ERI). Furthermore, the data could provide indications of previously unrecognized arrhythmias or modification of the therapy by reprogramming the device. Follow-up with the programmer Use the following procedure for in-office follow-up: Record and evaluate the ECG. Interrogate the device. Evaluate the status and automatically measured follow-up data. Check the sensing and pacing functions. Possibly evaluate statistics and IEGM recordings. Manually perform standard tests if necessary. Possibly customize program functions and parameters. Transmit the program permanently to the device. Print and document follow-up data (print report). 10 Finish the follow-up for this patient. 17 Patient Information Patient ID card Prohibitory signs Possible sources of interference A patient ID card is included in delivery. • Provide the patient with the patient ID. • Request that patients contact the physician in case of uncertainties. Premises with prohibitive signs must be avoided. • Draw the patient's attention to prohibitory signs. Electromagnetic interference should be avoided in daily activities. Sources of interference should not be brought into close proximity with the device. • Draw the patient's attention to special household appliances, security checkpoints, anti-theft alarm systems, strong electromagnetic fields, cell phones, and transmitters among other things. • Request patients to do the following: — Use cell phones on the side of their body that is opposite of the device. — Keep the cell phone at least 15 cm away from the device both during use and when stowing. 18 Replacement Indications Possible battery levels • • • • • Elective Replacement Indication (ERI) BOS: Beginning of Service: > 90% charge MOS 1: Middle of Service: 90% to 40% residual charge MOS 2: Middle of Service: < 40% residual charge ERI: Elective Replacement Indication (i.e. RRT: Recommended Replacement Time) EOS: End of Service Elective Replacement Indication can be detected by Home Monitoring. CAUTION Temporally limited therapy If ERI occurs shortly after follow-up and is only detected during the subsequent follow-up, then the remaining service time can be much less than 3 months. • Replace device soon. • • • EOS replacement indication The device can monitor the heart rhythm for at least 3 more months. At least 6 maximum energy shocks can be delivered until EOS occurs. The selected parameters in the software do not change. End of Service can be detected by Home Monitoring. WARNING Patient at risk of death If EOS replacement indication occurs before replacement of the device, then the patient is without therapy. • Replace device immediately. • Monitor patient constantly until immediate replacement of the device! • • VT and VF detection and all therapies are deactivated! The antibradycardia function remains active in the VVI mode: — Ventricular pacing: RV; basic rate 50 bpm; without special pacemaker functions such as hysteresis, etc. — Pulse amplitude of 6 V; pulse width of 1.5 ms — Cycle duration for BIOTRONIK Home Monitoring®: 90 days 19 Explantation and Device Replacement Explantation • • • • Interrogate the device status. Deactivate VT and VF therapies prior to explantation. Remove the leads from the header. Do not simply cut them loose. Use state-of-the-art techniques to remove the device and, if necessary, the leads. Note: Normal oxidation processes may cause ICD housing discolorations. This is neither a device defect nor does it influence device functionality. • Device replacement Explants are biologically contaminated and must be disposed of safely due to risk of infection. If, upon replacing the device, already implanted leads are no longer used but left in the patient, then an additional uncontrolled current path to the heart can result. • Deactivate VT and VF therapies prior to device replacement. • Insulate connector ports that are not used. Basic principles: • The device must not be resterilized and reused. Cremation Disposal Devices must not be cremated. • Explant the device before the cremation of a deceased patient. BIOTRONIK takes back used products for the purpose of environmentally safe disposal. • Clean the explant with a solution of at least 1% sodium hypochlorite. • Rinse off with water. • Fill out explantation form and send to BIOTRONIK together with the cleaned device. 20 Parameters Parameters4403736Technical Manual Bradycardia / CRT General ICD therapy Timing: Basic rate day/night and rate hystereses Parameter Range of values Standard VR DR HF ICD therapy OFF; ON ON Programs Display standard program; Display safe program; Display first interrogated program; Individual 1, 2, 3; Program Consult — Parameter Range of values Standard VR DR HF Basic rate 30 ... (5) ... 100 ... (10) ... 160 bpm 40 bpm Night rate OFF; 30 ... (5) ... 100 bpm OFF Night begins 00:00 ... (00:01) ... 23:59 hh:mm Night ends 60 bpm 22:00 hh: x mm 06:00 hh: mm Rate hysteresis OFF; -5 ... (-5) ... -25 ... (-20) ... -65 bpm OFF Scan/repetitive OFF; ON ON Parameter Range of values Standard VR DR HF AV dynamics Low; Medium; High; Fixed; (Individual) Low Timing: AV delay AV delay (1 or 2) after: – Pacing 15; 40 ... (5) ... 350 ms – – Sensing Either automatic: AV delay after pacing + sense compensation Or: 40 ... (5) ... 350 ms – – At rate 1 50 ... (10) ... 130 bpm 60 bpm – At rate 2 60 ... (10) ... 140 bpm 130 bpm Sense compensation OFF; -5 ... (-5) ... -120 ms -40 ms AV hysteresis mode OFF; Positive; Negative; IRSplus OFF OFF; Positive; Negative OFF AV hysteresis (positive) 70; 110; 150; 200 ms 70 ms AV hysteresis (positive) 70; 110; 150 ms 110 ms AV hysteresis (negative) 10 ... (10) ... 150 ms 50 ms AV scan and repetitive (positive) OFF; ON ON 21 Timing: Post-shock pacing Parameter Range of values Standard VR DR HF Post-shock duration OFF; 10 s; 30 s; 1 min; 2 min; 5 min; 10 min 10 s Post-shock basic rate 30 ... (5) ... 100 ... (10) ... 160 bpm 60 bpm AV delay post shock 50 ... (10) ... 350 ms 140 ms Ventricular post-shock pacing RV; BiV RV Parameter Range of values Standard VR DR HF Timing: Upper rate Upper rate 90 ... (10) ... 160 bpm 130 bpm Atrial upper rate OFF; 175; 200; 240 bpm 200 bpm Timing: Mode switching Parameter Range of values Standard VR DR HF Intervention rate OFF; 120 ... (10) ... 200 bpm 160 bpm Onset criterion 3 ... (1) ... 8 (out of 8) 10 bpm Resolution criterion Modification of basic rate OFF; 5 ... (5) ... 30 bpm After mode VDD(R): VDI(R) VDIR After mode DDD(R): DDI(R) DDIR – Rate OFF; 5 ... (5) ... 50 bpm 10 bpm – Duration 1 ... (1) ... 30 min 1 min Parameter Range of values Standard VR DR HF Mode After mode switching: Timing: Ventricular pacing Permanent RV; BiV; LV BiV Triggering OFF; RVs; RVs+PVC RVs LV T-wave protection OFF; ON ON UTR + 20; 90 ... (10) ... 160 bpm UTR + 20 Maximum trigger rate: – DDD(R) and VDD(R) – DDI(R), VDI(R) and VVI(R) 90 ... (10) ... 160 bpm 130 bpm Initially paced chamber RV; LV LV VV delay after Vp 0 ... (5) ... 100 ms 0 ms 22 Timing: Refractory periods and blanking periods Parameter Range of values Standard VR DR HF PVARP AUTO; 175 ... (25) ... 600 ms 225 ms PVARP extension OFF; ON ON Blanking after atrial pacing 50 ... (10) ... 100 ms 50 ms LV blanking after RV pacing 80 ms RV blanking after LV pacing Far-field protection after Vs OFF; 25 ... (25) ... 225 ms 75 ms Far-field protection after Vp 50 ... (25) ... 225 ms 75 ms Parameter Range of values Standard VR DR HF PMT detection/termination OFF; ON ON VA criterion 250 ... (10) ... 500 ms 350 ms Parameter Range of values Standard VR DR HF Maximum sensor rate 80 ... (10) ... 160 bpm 120 bpm x Sensor gain AUTO; Very low; Low; Medium; High; Very high Medium Sensor threshold Very low; Low; Medium; High; Very high Medium Rate increase 1; 2; 4; 8 bpm/cycle 2 bpm/ cycle Rate decrease 0.1; 0.2; 0.5; 1.0 bpm/ cycle 0.5 bpm/ x cycle Parameter Range of values Standard VR DR HF Pulse amplitude A 0.5 ... (0.25) ... 4.0 ... (0.5) ... 6.0; 7.5 V AUTO Timing: PMT protection Timing: Rate adaptation via accelerometer Pacing: Pulse amplitude and pulse width Pulse amplitude V/RV Pulse amplitude LV Pulse width A 0.4; 0.5 ... (0.25) ... 1.5 ms 0.4 ms Pulse width V/RV Pulse width LV Pacing: Atrial capture control 0.4 ms Parameter Range of values Standard Atrial capture control OFF; ATM ON VR DR HF Threshold test start 2.5 ... (0.5) ... 5.0 V 3.5 V 23 Pacing: Ventricular capture control Lead configuration LV on IS-1 connection Parameter Range of values Standard VR DR HF Ventricular capture control OFF; ATM ON Threshold test start 2.5 ... (0.5) ... 5.0 V 3.5 V VR DR HF Parameter Range of values Standard Pacing polarity LV (IS-1) LV tip -> LV ring LV tip -> RV shock coil LV ring -> LV tip LV ring -> RV shock coil UNIP LV tip -> RV shock coil Sensing polarity LV (IS-1) UNIP; BIPL UNIP 24 Tachycardia Detection Parameter Range of values Standard VR DR HF Interval AT/AF 240 ... 600 ms 300 ms Interval VT1 OFF; 270 ... (10) ... 600 ms OFF ON Interval VT2 OFF; 270 ... (10) ... 500 ms Interval VF OFF; 240 ... (10) ... 400 ms 300 ms Detection counter VT1 10 ... (2) ... 100 28 Detection counter VT2 10 ... (2) ... 80 20 Detection counter VF 6 out-of 8; 8 out-of 12; 18 out of 10 out-of 14; 12 out-of 16; 24 16 out-of 20; 18 out-of 24; 20 out-of 26; 22 out-of 30; 24 out-of 30; 30 out-of 40 Redetection counter VT1 10 ... (2) ... 50 20 Redetection counter VT2 10 ... (2) ... 40 14 Redetection counter VF 6 out-of 8; 8 out-of 12; 8 out 10 out-of 14; 12 out-of 16; of 12 16 out-of 20; 18 out-of 24; 20 out-of 26; 22 out-of 30; 24 out-of 30 SMART detection VT1/VT2 OFF; ON SMART detection ON: – Onset VT1/VT2 4 ... (4) ... 32% 20% – Stability VT1/VT2 8 ... (4) ... 48% 12% 4 ... (4) ... 32% 20% SMART detection OFF: – Onset VT1/VT2 – Stability VT1/VT2 OFF; 8 ... (4) ... 48 ms 24 ms MorphMatch OFF; Monitoring; ON OFF Sustained VT OFF; 1; 2; 3; 5; 10; 20; 30 min OFF Forced termination OFF; 1 ... (1) ... 10 min 1 min 25 Therapy: ATP Parameter Range of values Standard VR DR HF Attempts OFF; 1 ... (1) ... 10 OFF ATP type for VT1/VT2 Burst; Ramp Burst ATP type for VF OFF; Burst; Ramp Burst ATP optimization OFF; ON OFF Number S1 for VT1/VT2 1 ... (1) ... 10 Number S1 for VF S1 decrement for VT1/VT2 5 ... (5) ... 40 ms and for VF 10 ms Scan decrement OFF; 5 ... (5) ... 40 ms OFF Add S1 for VT1/VT2 OFF; ON ON Ventricular pacing for VT1/VT2 RV; LV; BiV RV Ventricular pacing for VF RV R-S1 interval for VT1/VT2 70 ... (5) ... 95% 80% R-S1 interval for VF 85 % Therapy: Shock Parameter Range of values Standard VR DR HF Number of shocks VT1/VT2 0; 1; 2; 6; 8 Number of shocks VF 6; 8 1st Shock for VT1/VT2 OFF; 2 ... (2) ... 20 ... (5) ... 40 J 40 J 2nd Shock for VT1/VT2 OFF; 4 ... (2) ... 20 ... (5) ... 40 J 40 J 3rd - nth shock for VT1/ VT2 OFF; 4*40 J; 6*40 J 6*40 J 1st Shock for VF 2 ... (2) ... 20 ... (5) ... 40 J 40 J 2nd Shock for VF 4 ... (2) ... 20 ... (5) ... 40 J 40 J 3rd - nth shock for VF 4*40 J; 6* 6*40 J For shock in VT1/VT2 and VF: – Confirmation OFF; ON ON – Polarity Normal; Reverse; Alternating Normal – Shock form Biphasic; Biphasic 2 Biphasic – Shock path RV -> ICD+SVC RV -> ICD RV -> SVC RV-> ICD+SVC RV -> ICD 26 Sensing Sensitivity and thresholds Parameter Range of values Standard VR DR HF Sensing A STD; OFF STD Sensing RV STD; TWS; VFS; IND STD Sensing LV STD; OFF; IND STD Upper threshold RV 50; 75% 50% Upper threshold LV 50; 75% 50% Upper threshold duration 110; 150 ... (50) ... 500 ms 350 ms after detection VFS: 110 ms Upper threshold duration after pacing 400 ms Lower threshold RV 25; 50% 25% T-wave suppression after pacing OFF; ON OFF Minimum threshold A 0.2 ... (0.1) ... 2.0 mV 0.4 mV Minimum threshold RV 0.5 ... (0.1) ... 2.5 mV 0.8 mV Minimum threshold LV 0.5 ... (0.1) ... 2.5 ... (0.5) ... 5.0 mV 1.6 mV Diagnostics The following can be set: Parameter Range of values Standard For AT/AF OFF; ON ON ON VR DR HF For SVT OFF; ON Periodic recording When Home Monitoring 90 days OFF: OFF; 30 ... (30) ... 180 days IEGM configuration RA, RV, LV RA, RV, FF FF; RV; LV RA, RV, LV Start resting period 00:00 ... (1:00 AM) ... 23:00 hh:mm 2:00 AM hh:mm Duration of resting period 0.5 ... (0.5) ... 12 h 4h AV delay adjusted in sensing test OFF; 300 ms 300 ms Thoracic impedance (TI) OFF, ON OFF 27 Home Monitoring Parameter Range of values Standard VR DR HF Home Monitoring OFF; ON OFF Time of transmission STD; 00:00 ... (1:00 AM) STD ... 23:00 hh:mm IEGM for therapy episodes OFF; ON ON Ongoing atrial episode OFF; 6; 12; 18 h 12 h Transmission date XX.XX.XXXX Follow-up + 91 days Cycle duration 20 ... (1) ... 366 days 91 days IEGM for monitoring episodes 28 Technical Data Technical Data5403736Technical Manual Mechanical Characteristics Housing Materials in contact with body tissue X-ray identification Devices with header for DF-1 connector: Type Lead connector W x H x D in mm Volume cm3 Mass g VR DF-1 69 x 55 x 12,5 38 71 DR DF-1 69 x 55 x 12,5 39 71 HF DF-1 69 x 58.5 x 12,5 40 73 • • • HK Housing: Titanium Header: Epoxy resin Silicone plugs and blind plugs (if applicable): Silopren or silastic 29 Electrical Characteristics Standards Measuring conditions The specifications are made according to EN 45502-2-2:2008. If not indicated otherwise, all specifications refer to the following conditions: • Ambient temperature: 37ºC ± 2ºC • Pacing/sensing: 500 Ω ±1% • Shock: 50 Ω ±1% Factory settings • • • Arrhythmia zones VT1, VT2, VF: OFF Antibradycardia pacing: OFF Home Monitoring: OFF Telemetry data • • MICS frequencies: 402-405 MHz Maximum power of transmission: < 25 µW (–16 dBm) International radio certification Devices with BIOTRONIK Home Monitoring® are equipped with an antenna for wireless communication. Telemetry information for Canada: This device must neither interfere with meteorological and earth resources technology satellites nor with meteorological stations working in the 400,150 to 406,000 MHZ band, and it must accept any interference received, including interference that may cause undesired operation. • This device will be registered with Industry Canada under the following number: IC: 4708A-TACHBORAX The code IC in front of the certification/ registration number only indicates that the technical requirements for Industry Canada are met. Telemetry information for the USA: This transmitter is authorized by rule under the Medical Device Radiocommunication Service (in part 95 of the FCC Rules) and must not cause harmful interference to stations operating in the 400.150-406.000 MHz band in the Meteorological Aids (i.e., transmitters and receivers used to communicate weather data), the Meteorological Satellite, or the Earth Exploration Satellite Services and must accept interference that may be caused by such stations, including interference that may cause undesired operation. This transmitter shall be used only in accordance with the FCC Rules governing the Medical Device Radiocommunication Service. Analog and digital voice communications are prohibited. Although this transmitter has been approved by the Federal Communications Commission, there is no guarantee that it will not receive interference or that any particular transmission from this transmitter will be free from interference. This device will be registered with Federal Communications Commission under the following number: FCC ID: QRITACHBORAX Telemetry data for Japan: In accordance with Japanese law, this device has been assigned an identification number under the "Ordinance concerning certification of conformity with technical regulations etc. of specified radio equipment", Article 2-1-8. • R: 202-SMD021 30 Pulse form The pacing pulse has the following form: The pulse amplitude reaches its maximum value at the beginning of the pulse (Ua). With increasing pacing duration (tb), the pulse amplitude is reduced dependent on the pacing impedance. Common mode rejection ratio ATP amplitude Rate Common mode rejection ratio Atrium: DR, HF V right: VR, DR, HF V left: HF 16.6 Hz 72 dB 59 dB 57 dB 50 Hz 73 dB 66 dB 57 dB 60 Hz 71 dB 66 dB 62 dB A burst was measured at 500 Ω, an amplitude of 7.5 V (tolerance ±1.5 V), pulse width of 1.5 ms, R-S1 interval of 300 ms and an S1 count of 5: ATP amplitude Automatic sensitivity control Measured minimum Measured maximum Mean value RA 7.57 V 7.49 V 5.1 V RV 7.54 V 7.49 V 5.1 V LV 7.55 V 7.51 V 5.1 V Measurement of actual values and test signal wave shape: standard triangle. Sensitivity Value Tolerance Measured value A: positive 0.2 mV 0.2 ... 0.5 0.23 mV A: negative RV: positive 0.24 mV 0.5 mV 0.3 ... 0.7 RV: negative LV: positive 0.55 mV 0.58 mV 0.5 mV 0.3 ... 0.7 0.52 mV LV: negative Shock energy / peak voltage With shock path: RV to housing + SVC Shock energy (Tolerance) Tolerance Peak voltage Measured value Measured value Shock energy Peak voltage 1 J (0.7 ... 1.18) 100 ... 140 V 0.89 J 125,0 V 20 J (15.9 ... 21.6) 500 ... 550 V 16,62 J 530,4 V 40 J (32.0 ... 43.2) 710 ... 790 V 33,84 J 750,9 V 31 Battery Data Battery characteristics Storage period Calculation of service times The following data is provided by the manufacturers: Manufacturer LITRONIK Batterietechnologie GmbH & Co 01796 Pirna, Germany Battery type LiS 3410 RR System LiMnO2 Battery ID number shown on the programmer Device type VR-T, DR-T, HF-T Battery voltage at ERI 2.85 V Charge time at BOS 8s Charge time at ERI 10 s Usable capacity until ERI: VR, DR, HF 1390 mAh Usable capacity until EOS 1520 mAh The storage period affects the battery service time. • Devices should be implanted within 19 months between the manufacturing date and the use by date (indicated on the package). • If the ICD is implanted shortly before the use by date, the expected service time may be reduced by up to 17 months. • • Calculation of the number of shocks Inlexa 1 VR-T The services times have been calculated as follows – in all chambers depending on the device type: — Pulse amplitude: 2.5 V — Pulse width: 0.4 ms — Pacing impedance: 500 Ω — Basic rate: 60 bpm — Home Monitoring: ON, 1 device message each day and 12 IEGM online HD transmissions per year — Diagnostic functions and recordings: permanently set Capacitor reforming is performed 4 times per year and therefore at least 4 maximum charges for shocks have to be assumed per year even if less than 4 are delivered. Calculation of the number of shocks: Service time [in years] x number of shocks per year Service times with LiS 3410 RR battery: Service time [in years] at number of shocks per year Pacing 12 16 20 0% 10.3 8.3 7.0 6.0 5.3 15% 10.1 8.1 6.8 5.9 5.2 50% 9.5 7.8 6.6 5.7 5.0 100% 8.8 7.3 6.2 5.4 4.8 32 Inlexa 1 DR-T Service times with LiS 3410 RR battery: Service time [in years] at number of shocks per year Pacing Inlexa 1 HF-T 12 16 20 0% 9.4 7.7 6.5 5.7 5.0 15% 9.0 7.4 6.3 5.5 4.9 50% 8.1 6.8 5.9 5.2 4.6 100% 7.1 6.1 5.3 4.7 4.3 Service times with LiS 3410 RR battery: Service time [in years] at number of shocks per year Pacing 12 16 20 0% 8.9 7.4 6.3 5.5 4.9 15% 8.3 7.0 6.0 5.2 4.7 50% 7.2 6.1 5.4 4.8 4.3 100% 6.0 5.3 4.7 4.2 3.9 33 Legend for the Label Label on the package The label icons symbolize the following: Manufacturing date Use by Storage temperature Order number Serial number Product identification number Dangerous voltages! CE mark Contents Follow the instructions for use Sterilized with ethylene oxide Do not resterilize Single use only. Do not reuse! STERILIZE Do not use if packaging is damaged Non-sterile NON STERILE Transmitter with non-ionizing radiation at designated frequency TP2 Compatibility with telemetry protocol version 2 of BIOTRONIK Home Monitoring Device: NBG code and compatible leads Example Factory settings for therapy: OFF Example Screwdriver Example of lead connection: DF-1/IS-1
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : Yes Create Date : 2014:11:12 22:27:29Z Creator : FrameMaker 8.0 Modify Date : 2016:02:12 11:16:30+01:00 Has XFA : No Tagged PDF : Yes XMP Toolkit : Adobe XMP Core 4.2.1-c041 52.342996, 2008/05/07-20:48:00 Creator Tool : FrameMaker 8.0 Metadata Date : 2016:02:12 11:16:30+01:00 Producer : Acrobat Distiller 9.0.0 (Windows) Format : application/pdf Title : untitled Document ID : uuid:89a78610-b966-4db6-b85d-aed835f98544 Instance ID : uuid:69298a63-59bb-4623-a480-ace72424047d Page Layout : OneColumn Page Mode : UseOutlines Page Count : 34EXIF Metadata provided by EXIF.tools