Beijing GODA Instruments GDRD5X6XC Pulse Radar Level Instrument User Manual FCC version12 02

Beijing GODA Instruments Co., LTD. Pulse Radar Level Instrument FCC version12 02

Users Manual

Download: Beijing GODA Instruments GDRD5X6XC Pulse Radar Level Instrument User Manual FCC version12 02
Mirror Download [FCC.gov]Beijing GODA Instruments GDRD5X6XC Pulse Radar Level Instrument User Manual FCC version12 02
Document ID2514805
Application IDzqaVHrCdb2iCb7UZQ2pXHA==
Document DescriptionUsers Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeYes
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize116.34kB (1454296 bits)
Date Submitted2015-01-27 00:00:00
Date Available2015-01-27 00:00:00
Creation Date2014-12-02 13:27:55
Producing SoftwareCorel PDF Engine Version 1.0.0.458
Document Lastmod2014-12-02 13:27:55
Document TitleFCC version12_02
Document CreatorCorelDRAW 版本 12.0

Level Measurement Expert
Pulse Radar Level Instrument
Beijing GODA Instruments Co.,Ltd.
北京古大仪表有限公司
Table of Contents
1 For your safety......
...............
2 Measurement Principle . ...............
3 Product Overview....................
4 Mounting Requirement.................
. ...1
...2
....3
...7
5 Electrical Connection.................. . . . 1 3
6 Adjustment Instructions................. . .15
7 Dimensional Drawings.................. ..18
8 Technical Specifications................. . . 2 2
9 Selection & Ordering Information........... . 2 7
10 Application Questionnaire . ........... . .. 3 3
1. For your safety
Authorised personal
All operations described in this operating instructions manual must be carried out ouly by trained
specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be
worn.
Appropriate use
Operational reliability is ensured only if the instrument is properly used according to the
specifications in the operating instructions manual as well as possible supplementary instructions.
FCC Certification
This device complied with Part 15 of the FCC Rules. Operation is subject to the following
two conditions:
(1)this device may not cause harmful interference, and
(2)this device must accept any interference received, including interference that may
cause undesired operation.
Changes or modifications not expressly approved by the manufacturer could void the user's
authority to operate the equipment.
Warning: User must keep a safety distance of at least 20cm from the antenna.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
-- Reorient or relocate the receiving antenna.
-- Increase the separation between the equipment and receiver.
-- Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
-- Consult the dealer or an experienced radio/TV technician for
help.
2 . Measurement Principle
Principle
The extremely narrow microwave pulse emitted by the antenna on radar level instrument
can travel at the speed of light and part of its energy, which is reflected off the surface of
target medium, is received by the very same antenna. The time lapse between pulse emission
and reception by the antenna is proportional to the distance between the surface of target
medium and the reference point on antenna. However, due to the fact that the electromagnetic
wave is transmitted at extremely high speed, which leads to the tiny time lapse (nanosecond
level) and makes it difficult to be identified, GDRD series of radar level instrument have
adopted a special demodulation technology, enabling itself to detect the time lapse between
pulse emission and reception correctly, and eventually generate accurate measurement result.
Features
The pulse radar level instrument, adopted K band range as transmission frequency, which
make this series have specialties as below: Small beam angle, which centralize energy, make
GDRD high ability of anti-jamming, hence high accuracy and reliable. Small antenna size,
easy to mount and easy to equip extra dust protection. Small blind zone, good accuracy
e v e n f o r s m a l l v e s s e l s . S h o r t e r w a v e - l e n g t h , s u i t a b l e f o r s m a l l p o w e r.
Equipped with advanced microprocessor and unique EchoDiscovery echo processing
technology, the radar level instrument can be used under various hazardous process conditions
The pulse radar level instrument, with pulses as its working tool and extremely low
emission power, can be mounted on various metal or nonmetal vessels, harmless towards
the environment and human beings.
3 Product Overview
GDRD55
GDRD61
Liquid
Application:
Max Measurement Range::
Measurement Accuracy:
Common
±3mm
30m
Enhanced 70m
±10mm
Liquid
Level measurement in liquids,
especially highly erosive liquids
10m
±5mm
Process Temperature:
(-40~100) o C
(-40~130)o C
Process Pressure:
Ordinary pressure
(-0.1~0.3)MPa
Signal Output:
Power:
(4~20) mA/HART
2-wire(DC24V)
Display module:
None
Housing:
Pa66
Process Connection:
Lifting frame or 1’NPT
(4~20)mA/HART
2-wire(DC24V)
Flange Accessories:
Antenna:
GDRD57
GDRD56
Liquid
Level measurement in liquids, under certain
temperature and pressure, mildly erosive liquids
30m
±3mm
(-40~80)o C
(-40~130)o C
(-60~250)o C
(-60~400)o C
Normal
(-0.1~4)MPa
(-0.1~40)MPa
Liquid
Level measurement of highly erosive
medium under certain pressure/
temperature limit and suitable for
20m
±3mm
(-40~150)o C
(-0.1~0.5)MPa
(4~20)mA/HART
2-wire(DC24V)
(4~20)mA/HART
2-wire(DC24V)
Optional
Optional
H/I/J/K
L/M/P
T/V
Pulse Radar Level Instrument
GDRD59
GDRD58
Solid
strong dew/dust/crystal
Solid
Normal Temperature/Normal Pressure
70m
±15mm
15m
±10mm
(-40~80)o C
(-40~120)o C
(-60~250)o C
(-60~400)o C
Normal
(-0.1~4)MPa
(-0.1~40)MPa
(-40~80)o C
Normal
(4~20)mA/HART
2-wire(DC24V)
(4~20)mA/HART
2-wire(DC24V)
Optional
Optional
H/I/J/K
L/M/P
L/M/N
T/V
Housing
Serial number
Aluminium
Material
Specialty
Process Connection
Serial number
Material
PFA
Stainless Steel
Stainless Steel\\
(Huff)
Pressure
(-0.1~0.3)MPa
(-0.1~4)MPa
(-0.1~0.5)MPa
(-40~130)oC
Temperature
(-60~150)oC
Stainless Steel
(-60~130)oC
Stainless Steel Flange
(-0.1~4)MPa
(-0.1~40)MPa
(-60~250)oC
(-60~400)oC
Flange Accessories
Serial number
(PTFE/PP)Flange
Material
Specialty
Rust tolerated
Stainless Steel Flange
Stainless Steel Gimbal Flange
High temp./High Pressure
High temp./Normal Pressure
Antenna
Serial number
Material
Specification
Specialty
Φ44/Length86
Φ44/Length108
Stainless steel
Φ48/Length140
Φ78/Length227
Φ98/Length288
Φ123/Length620
PFA
Dn80
DN100
Normal Temperature Temperature tolerated/ Rust tolerated/
Pressure tolerated
Pressure tolerated
Stainless steel(PFA
shield)
Φ98/300
Φ123/625
Normal Temperature/
Normal Pressure
Stainless steel
Φ198
Φ246
Temperature tolerated/
Pressure tolerated
Pulse Radar Level Instrument
4 . Mounting Requirements
Basic Requirements
There is a certain existing beam angle while the antenna transmitting microwave pulses. There should be no barriers
between the lower edge of antenna and surface of measured medium. Therefore it is highly recommended to avoid
facilities inside vessels, such asladders, limit switches, heating spirals, struts and etc, during the mounting process.
“False echo learning” must be carried out during the installation in this case. Furthermore, microwave beams must
NOT intersect the filling streams. Be cautions during the installation: the highest level of target medium must NOT
enter into blanking zone; the instrument must keep certain distance to vessel walls; every possible measure needs
to be taken to position the instrument so that the direction of antenna emission is perpendicular to the surface of
measured medium. The installation of instruments in explosion proof area must abide by relevant local or federal
safety regulations. Stainless steel housing should be used for intrinsically safe explosion proof version, which is also
applicable in explosion proof areas. The instrument must be connected with ground in this case.
Illustrations
43
The reference plane is the
thread or flange surface
100%
1. Blanking Zone(menu1.9)
2. Empty(menu1.8)
3. Max. Adjustment(menu1.2)
4. Min. Adjustment(menu1.1)
0%
Note:The highest level of measured medium must not enter into vlanking zone while
radar level measurement instrument is in operation.
Mounting Position
>200mm
Minium distance of 500mm
between instrument and vessel
wall during installation
1.Reference Plane
2.Center of Vessel or Symmetrical
Axis
2
1.Reference Plane
2.Center of Vessel or Symmetrical
Axis
>200mm
The best mounting position for a
conical vessel with flat top is the
center of its top, as the effective
measurement can reach the
bottom of vessel.
Installation with Gimbal
Pulse Radar Level Instrument
Damp-proof
In order to avoid dampness under
outdoor or humid indoor conditions
or for those instruments mounted
on cooling/heating vessels, seal
rings used on cables should be
screwed tight, plus the cable must
be bended downward outside
cable entry, indicated on the
diagram below
Antenna Extension
>10mm
The transducer end must at
least protrude 10mm out of
socket.
GDRD56 Antenna Extension
If the senser is mounted in a
socket extension that is too long,
strong false echoes are generated
which enterfere with the measurement.
Make sure that the horn antenna
protrudes out of the socket piece.
d
1½ "
50mm(2")
80mm(3")
100mm(4")
150mm(6")
h m ax
2 00m m
2 50m m
3 00m m
5 00m m
8 00m m
G D R D 5 7 E xtensio n p ipe de m n s t r a t i on
d
50mm(2")
80mm(3")
100mm(4")
h max
100mm
150mm
250mm
Rights and Wrongs in Mounting
A大 约 1 3 D
10
1.Correct
2.Wrong:Instruments are mounted
in the center of concave or arched
vassel tops, which results in
multiple echoes.
1.Wrong: Mount the instrument
in/above filling stream, which
results in the measurement of
filling stream not the target medium.
2.Correct:
Pulse Radar Level Instrument
Reflection board installation
If there are barriers in vessels, it is
required to mount baffle-board, by
doing this, the echo reflected by the
barrier will be reflected out. And “False
Echo Storage” will be applied.
Agitator
If there are agitators in vessels,
instrument must be mounted as
far away from agitators as possible.
Once installation completed, a
"false echo learning" should be
carried out while agitators in
motion to eliminate negative
influence caused by false echo
of agitators. You are advised to
opt for installation with standpipe
if foam or wave is generated due
to the action of agitators.
11
Installation with Standpipe
By using standpipe, the influence of foams can be reduced.
Side through Standpipe
installation with Standpipe
ventilation hole
You are advised to opt for installation
with standpipe (or bypass tube) to avoid
the influence on measurement caused
by barriers inside vessels or foam
generation.
It is advised to install antenna inside
of the standpipe to avoid the error
caused by foam. The minimum inner
diameter of standpipe should be 50mm.
Avoid large cracks or welding seam
when connecting standpipe. False
echo storage must be carried out as
well in this case.
Note: You must NOT mount instrument inside standpipe while
measuring adhesive medium.
12
Pulse Radar Level Instrument
5 Electrical Connection
Power Supply
20mA/HART(2-Wire) Power supply and current signal are carried by the same two-wire connection cable.
This equipment is not allowed to be connected to public utility power lines.
See the Technical Specifications of this guide for detailed requirement on power
supply. A safety barrier should be placed between power supply and instrument for
intrinsically safe version.
See the Technical Specifications of this guide for detailed requirement on power supply.
Earth-connected current output can be used for standard version of level instruments,
while the explosion proof version must be operated with a floating current output. Both
instruments and earth terminals should be connected with ground firmly and securely.
Normally you can either choose to connect with the earth terminal on vessel or adjacent
ground in case of plastic vessels.
Cable Connection
General Introduction
Standard 2-wire cable with outside diameter of 5...9mm, which assures the seal effect
of cable entry, can be used as feeder cable. You are recommended to use screened
Shielding & Grounding
The two ends of shielded cable must be connected with earth terminal. The shielded
cable must be connected with inner earth terminal directly inside the transducer, while
the outside earth terminal on housing must be connected with ground. In the event of
earth-connected current, the shielding side of shielded cable must be connected to
ground potential via a ceramic capacitor (e.g. : 1μF 1500V) in order to dampen the low
frequency grounding current and avoid the disturbance caused by high frequency signals
Wiring Diagram
2-wire
2-wire wiring used for HART
1)Power Supply and Signal Output
13
Pulse Radar Level Instrument
Explosion Proof Connection
This product is an intrinsic safety explosion proof version (Exia ⅡC T6 Ga) with stainless steel housing and
plastic-encapsulated internal structure aimed to prevent sparks resulted from transducer and circuit malfunction
from leaking out. It is applicable for the non-contact continuous level measurement of flammable medium under
the level of explosion proof inferior to Exia ⅡC T6 Ga. You are required to use series (intrinsic safety explosion
proof: [Exia] ⅡC,voltage of power supply: 24V DC±10%,short-circuit current: 114mA,operating current:
4...20mA) of safety barriers, which are supplementary to this product, for the power supply of this product.
All connection cables must be screened with max. length of 500m. Stray capacitor≤0.1μF/Km,stray inductance
≤1mH/Km. The level measurement instrument must be connected to ground potential and unapproved
supplementary devices are not allowed to use.
FBS-2
1 2 3
None-Explosion-Proof Zone
4 5 6
Explosion Proof Zone
(4~20)mA
(4~20)mA IN (4~20)mA OUT
24V OUT
24V IN
1 2 3
4 5 6
choose either one
CO MW AY
Adjust with GODAware
14
GODAware
6 Adjustment Instructions
Adjustment Methods
Three adjustment methods available for GDRD series :
1.Display/Adjustment Module
2.Adjustment software GODAware
3.HART handheld programmer
ViewPoint is a pluggable display/adjustment module. The adjustment can be done through operating with four
buttons on ViewPoint. Optional menu operation languages are available for selection. ViewPoint is only used for
display after adjustment in that the measurement results can be seen clearly through the glass window.
Display/Adjustment Module
dB
12 0
11 0
10 0
90
80
70
60
50
40
30
20
10
0 . 00
2. 0 0
3.0 0
4 . 00
5 .0 0
LCD
Adjustment Keypad
[ O K ]Keypad
-Enter programming mode;
-Confirm programming options;
-Confirm modifications to parameters.
]Keypad
-Choose programming options;
-Choose the digit of parameters to edit;
-Display the contents of parameters.
]Keypad
-Modify parameter values.
[ B K ]Keypad
-Programming mode exit;
-Return to higher menu level.
Shortcut
[ B K ]Display Echo wave
15
Pulse Radar Level Instrument
GODAware
Connect with another unit throuth HART .
- DC
Choost Either One
GODAware
C O M WAY
1 RS232 Connect Cable/USB port
GDRD series
HATR pont adapter used on COMWAY convertor
250 ohm Resistance
COMWAY Convertor
Connect with another unit throuth I 2C .
GODAware
COMW A Y
- DC
1 RS232 Connect Cable/USB port
2 GDRD series
3 I2C adapter pont used on MOMWAY convertor
4 COMWAY Convertor
16
HART Handheld Programmer
Adjust GDRD
series with HART Handheld Programmer
- DC
Choost Either One
1 HART Handheld Programmer
2 GDRD series
3 250 ohm Resistance
17
Pulse Radar Level Instrument
7 Dimension(Unit:mm)
Housing
Material: AL/PBT/ 316L
GDRD61
Nylon Sheath 1
Sealing Hose 2
Sealing Flat Gasket 3
Upper Cover 4
Note:
To install from 4 to 1.
To remove from 1 to 4.
22
337.5
GDR D55 Threaded Vision
86.0
44
G112A(112NPT)
18
G DRD 56 Threaded Vision
Huff
Φ48
Φ78
Φ98
Φ123
H(316L) H(316L Shiele)
140
227
288
300
620
625
G DRD 57 Flange Version
DN80
PN1.6 Φ200 Φ160 Φ132 174
DN100 PN1.6 Φ220 Φ180 Φ156 260
GDRD58 Threaded Vision
Huff
Φ48
Φ78
Φ98
Φ123
H(316L) H(316LShiele)
140
227
288
620
300
625
19
Pulse Radar Level Instrument
GDRD58 Gimbal Flange
Huff
n×d
n×d
H1(316L) H1(316L Shield)
140
227
288
300
620
625
H1
H1
Φ48
Φ78
Φ98
Φ123
DN 100/4"
220mm 180mm
11.5mm 8xΦ18mm
DN 125/5"
250mm 210mm
11.5mm 8xΦ18mm
GDRD58/59 Threaded Vision
20
GDRD58/59 Gimbal Flange
10mm 4×φ18mm
10mm 4×φ18mm
10mm 4×φ22mm
10mm 4×φ22mm
10mm 4×φ26mm
21
Pulse Radar Level Instrument
8 Technical Specifications
General Parameters
series/Type
Process
Connection
Material
GDRD55
GDRD56
ThreadG1½A
Thread1½NPT
PP/PFA
GDRD57
ThreadG1½A
Thread1½NPT
Flange 316L
Stainless Steel 316L
PFA
PFA
GDRD58
GDRD59
GDRD61
ThreadG1½A
ThreadG1½A
Lifting Frame
Flange 316L
Flange 316L
Thread 1"NPT
Thread1½NPT
Thread1½NPT
Stainless Steel316L
PFA
Stainless Steel316L
PFA
Aluminum
PP
Housing
Aluminium,Stainless Steel 316L
Seal ring between housing and housing cover
Silicone
ViewPoint window on housing
Polycarbonate
Ground terminal
Stainless Steel
Weight
Weight
-GDRD61
-GDRD55
-GDRD56
-GDRD57
-GDRD58
-GDRD59
1.6kg
1kg(Depend on process connections and housings)
2kg(Depend on process connections and housings)
3kg(Depend on process connections and housings)
7kg(Depend on process connections and housings)
2kg(Depend on process connections and housings)
Power
Standard Version
Power Consumption
Ripple Allowed
(21.6~26.4)V DC
max.22.5mA
-<100Hz
-(100~100K)Hz
Uss<1V
2-wire
Parameters on Cable
Cable Entry/Plug
Spring Connection Terminal
Output
Output Signal
Resolution
Fault Signal
-2-wire load resistance
Integration Time
22
Uss<10mV
One cable entry of M20x1.5(cable diameter of 5~9m),
one binding of M20x1.5
Applicable for cables with cross section of 2.5mm2
4...20mA/HART
1.6μA
Constant current output: 20.5mA;
22mA;3.9mA
See diagram below
0...40sec, adjustable
Characteristic
parameter
Blanking Distance
Max Measurement Distance
End of Antenna
-GDRD55
-GDRD56
-GDRD57
-GDRD58
-GDRD59
-GDRD61/61L
Measurement Interval
Adjustment Time
10m(liquid)
30m(liquid)
20m(liquid)
70m (solid)
15m (solid)
30m/70m(liquid)
About 1sec(Depend on parameter settings)
About 1sec(Depend on parameter settings)
Resolution of Display
1mm
Accuracy
See the diagram below
Temperature for Storage/Transport
(-40~100)oC
Process Temperature (Probe)
Relative Humidity
Pressure
Vibration Proof
-GDRD55
-GDRD56
-GDRD57
-GDRD58
-GDRD59
-GDRD61
<95%
Max.40MPa
(-40~130)oC
(-60~400)oC
(-40~150)oC
(-60~400)oC
(-40~200)oC
(-40~100)oC
Mechanical vibration10m/s 10m2/s ,10~150Hz
23
Pulse Radar Level Instrument
GDRD61
Accuracy
See the diagram below
10mm
3mm
0.5m
30m
-3mm
-10mm
GDRD55
Accuracy See the diagram left
10mm
5mm
0.5m
10m
-5mm
-10mm
1) The generation of accurate measurement results needs longer time than usual in the event of drastic level changes(mx. Error 10%).
GDRD56
Accuracy
See the accuracy illustration diagram below
10mm
3mm
0.5m
- 3 mm
- 1 0m m
24
30m
GDRD57
Accuracy
See the accuracy illustration diagram below
10mm
3mm
1.0m
20m
-3mm
-10mm
GDRD58
Accuracy
See the accuracy illustration diagram below
30mm
15mm
1.0m
70m
- 1 5m m
-30mm
25
GDRD59
Accuracy
See the accuracy illustration diagram below
15mm
10mm
1.0m
-10mm
- 1 5m m
26
30m
Pulse Radar Level Instrument
9 Selection & Ordering Information
GDRD55
Explosion Proof Approval
P Standard
I Intrinsically Safe(Exia IIC T6 Ga)
Shape of Antenna/Material
B (R)Airproof HornΦ44mm/L86
N (R)Airproof HornΦ44mm/L108
Process Connection/Material
GP (F)Thread G1½A
NP (F)Thread 1½NPT
Length of Vessel Socket
A NO
Electronic
B (4~20)mA/HART2-Wire
Housing/Protection
A Aluminium/IP67
Cable Entry
M M20x1.5
½NPT
Display/Programming
A Yes
X No
27
GDRD56
Explosion Proof Approval
P Standard
I Intrinsically Safe(Exia IIC T6 Ga)
Shape of Antenna/Material
(T)Horn
(T)Horn
(T)Horn
(Z)Horn
Φ78mm/L227/Stainless Steel 316L
Φ98mm/L288/Stainless Steel 316L
Φ123mm/L620/Stainless Steel 316L
Φ78mm/Flexible/Stainless Steel 316L
Process Connection/Material
GP (H)thread G1½A/Stainless Steel 316L
GA (H)thread 1½NPT/Stainless Steel 316L
GB (G)thread G1½A/PP
GC (J)thread G1½A/Stainless Steel 316L/temperature(-60~250)oC
GE (I)thread G1½A/Stainless Steel 316L(Huff)
Flange/Material
Material
Spec Code
DN50
DN80
DN100
DN125
PP(L)
FA
GA
HA
IA
PTFE(L)
FB
GB
HB
IB
Stainless Steel(M)
FC
GC
HC
IC
F0 No
Seal/Process Temperature
2 Viton(-60~150)oC
3 Kalrez(-60~250)oC
4 Graphite(-60~400)oC
Electronic
B (4~20)mA/HART 2-Wire
Housing/Protection
A Aluminium/IP67
Cable Entry
M M20x1.5
½NPT
Display/Programming
A Yes
X No
28
Pulse Radar Level Instrument
GDRD57
Explosion Proof Approval
P Explosion Proof Approval
I Intrinsically Safe(Exia IIC T6 Ga)
Shape of Antenna/Material
C (U)Stainless Steel&PFA Flange DN80
D (U)Stainless Steel&PFA Flange Dn100
Electronic
B (4~20)mA/HART 2-Wire
Housing/Protection
A Aluminium/IP67
Cable Entry
M20x1.5
½NPT
Display/Programming
A Yes
X No
29
GDRD58
Explosion Proof Approval
P Explosion Proof Approval
I Intrinsically Safe(Exia IIC T6 Ga)
Shape of Antenna/Material
(T)Horn Φ78mm/L227/Stainless Steel316L
(T)Horn Φ98mm/L288/Stainless Steel316L
(T)Horn Φ123mm/L620/Stainless Steel316L
(V)Horn Φ98mm/L300/Stainless Steel316L/PFA Shield
(V)Horn Φ123mm/L625/Stainless Steel316L/PFA Shield
(W)ParaboloidΦ198mm/Stainless Steel316L
(W)ParaboloidΦ246mm/Stainless Steel316L
(Z)Horn Φ78mm/Flexible/Stainless Steel316L
Process Connection/Material
GP (H) Thread G1½A/Stainless Steel316L
GA (H) Thread 1½NPT/Stainless Steel316L
GB (G) Thread G1½A/PP
GC (J) Thread G1½A/Stainless Steel316L/ Temperature (-60~250)oC
GE (I) Thread G1½A/Stainless Steel316L(Huff)
GF (E) Thread G1½A/Stainless Steel316L/Temperature (-60~250)oC
Flange/Material
Material
Spec Code
DN50
DN80
DN100
DN125
PP(L) PTFE(L) Stainless Steel(M)Gimbal Flange(PP)(N)Gimbal Flange(Stainless Steel)(P)
FA
GA
HA
IA
FB
GB
HB
IB
FC
GC
HC
IC
HD
ID
F0 No
Seal/Process Temperature
2 Viton(-60~150)oC
3 Kalrez(-60~250)oC
4 Graphite(-60~400)oC
Electronic
B (4~20)mA/HART 2-Wire
Housing/Protection
A Aluminium/IP67
Cable Entry
M M20x1.5
½NPT
Display/Programming
A Yes
X No
30
HE
IE
Pulse Radar Level Instrument
GDRD59
Explosion Proof Approval
P Explosion Proof Approval
I Intrinsically Safe(Exia IIC T6 Ga)
Shape of Antenna/Material
(T)Horn Φ78mm/L227/Stainless Steel316L
(T)Horn Φ98mm/288/Stainless Steel316L
(T)Horn Φ123mm/L620/Stainless Steel316L
(V)Horn Φ98mm/L300/Stainless Steel316L/PFA Shield
(V)Horn Φ123mm/L625/Stainless Steel316L/PFA Shield
(W)Parabolic Φ198mm/Stainless Steel316L
(W)Parabolic Φ246mm/Stainless Steel316L
(Z)Horn Φ78mm/Flexible
Process Connection/Material
GP (H) Thread G1½A/Stainless Steel316L
GA (H) Thread 1½NPT/Stainless Steel316L
GB (G)thread G1½A/PP
GE (I) Thread G1½A/Stainless Steel316L(Huff)
GF (E) Thread G1½A/Stainless Steel316L/Temperature (-60~250)oC
Flange/Material
Material
Spec Code
DN50
DN80
DN100
DN125
DN150
DN200
DN250
PP(L)
FA
GA
HA
IA
JA
KA
LA
PTFE(L)
Stainless Steel(M)
FB
GB
HB
IB
JB
KB
LB
FC
GC
HC
IC
JC
KC
LC
Gimbal Flange(Stainless Steel)(P)
HE
IE
JE
KE
LE
F0 No
Seal/Process Temperature
2 Viton(-60~80)oC
3 Kalrez(-60~250) C
Electronic
B (4~20)mA/HART 2-Wire
Housing/Protection
A Aluminium/IP67
Cable Entry
M20x1.5
½NPT
Display/Programming
A Yes
X No
31
GDRD61
Explosion Proof Approval
Standard(max measurement range of 30m)
Intrinsically Safe Exia IIC T6 Ga
Enhanced(max measurement range of 70m)
Shape of Antenna/Material
Horn Φ78mm/L221/PA66
Process Connection/Material
GB THREAD 1″NPT PP
GD LIFTING FRAME
Electronic
(4~20)MA/HART
Housing/Protection
IP68
Programming
A Yes
No
Sunshield
A Yes
X No
Cable
A Standard twin-core shielded (leagth of 10m)
B Length of twin-core shielded(length:X m)
C Length of 7-core shielded lead(length:X m)
32
Pulse Radar Level Instrument
10 Application Questionnaire
Approvals
Standard Version
Intrinsically Safe Version(Exia IIB T6 Ga)
Intrinsically Safe Version(Exia IIC T6 Ga)
Intrinsically Safe Version+Explosion Proof(Exdia [ia Ga] IIC T6 Gb)
Measured Medium
Name
Condition
Liquid
Temperature:
Min.
Surface
Flat
Dielectric Constant
Solid
Form
Mass
Norm.
°
Turbulent
εr<3
εr>3
Particle
Max.
°
Agitated
Dust)
°C
Vorte
Atmosphere
Atmosphere
Form
Atmosphere Pressure
Foam
Dust
Min.
Deposit
Norm.
Vapour
Max.
Vessel
Shape of Top
Flat
Height
Critical Information
Arch
Conical
Horizontal
Diameter
Nozzle Length:
Nozzle Diameter:
Measurement Range:
Process Connection
Thread(
G3/4 A
NPT
G1A
)
Swivelling Holder
3/4
Flange(DN=
G 1 A、M 1 0 5 x 2
G1½A
1 ½ N PT
G2A )
Installation
Mode:
Top
Side
Filling Stream inlet position and installation position (Please specify in the diagram below)
Circular Vessel
Square Vessel
Communication
( 4~20) mA/ HART
Display
Yes
No
Customer Information
Please give brief explanation on the application of
instrument:
Contact:
Company:
Address:
P. C.:
Email:
Tel:
Fax:
Date:
33
Beijing GODA Instruments Co.,Ltd.
Beijing GODA Instruments Co., Ltd.
Add: Management Address:3RD Floor,Block B,
Hongxianhong base,Beiqijia Town,Changping
District,Beijing.
Tel:(010)89759341 89759342 89759343
Fax:(010)89759327
PC: 102209
Website: www.godacn.com
89759344
Email:sales@godacn.com
Sales Dept Of Beijing GODA Instruments Co., Ltd.
Add: Room1303, Block D, Ocean International Center,
No 62 Middle Of east 4th ring, Chaoyang Dist.,
Beijing,100025, China
Tel:(010)59648788(Ext.8)
Fax:(010)59648789
PC: 100025
Edition: GD/200902

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : Yes
Page Count                      : 36
Create Date                     : D:20141202132755