Biomedical Systems BRHA02 HANDHELD MOBILE TELEMETRY DEVICE User Manual Operations Manual

Biomedical Systems Corporation HANDHELD MOBILE TELEMETRY DEVICE Operations Manual

Operations Manual

Biomedical Systems77 Progress Parkway • Saint Louis, MO 63043www.biomedsys.com/truvueToll Free 1-800-877-6334Physician’s Operation Manual
2
3TruVue Mobile Telemetry Monitoring SystemTABLE OF CONTENTSIndications for Use ........................................................................................................................4System Overview ...........................................................................................................................5  Patient Devices ...................................................................................................................................6  Data and Monitoring Center..........................................................................................................7Service Overview ...........................................................................................................................8  Ordering TruVue .................................................................................................................................8  Initiating Monitoring ........................................................................................................................9  Concluding Monitoring ................................................................................................................ 10  Breaks in Monitoring ..................................................................................................................... 10Operation and Performance Specications .......................................................................11  Sensor Operation ............................................................................................................................ 11  Sensor Performance Specications ..........................................................................................13  Handheld Operation ...................................................................................................................... 15  Handheld Performance Specications ................................................................................... 24  Algorithm Operation and Performance ................................................................................. 26Description of Device Symbols ..............................................................................................30Summary of Caution Statements ..........................................................................................31
PHYSICIAN’S OPERATION MANUALTruVue Indications of Use4TruVue Indications for UseCaution:  Federal Law (USA) restricts this device to sale by or on the order of a physician.Indications for Use:  The TruVue® System is intended for use by patients who experience transient events     that may suggest cardiac arrhythmia.  Patients who require monitoring of eect of drugs to control ventricular rate in various    atrial arrhythmias (e.g. atrial brillation).  Patients with symptoms that may be due to cardiac arrhythmias. These may include     but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope     of unknown etiology in which arrhythmias are suspected or need to be excluded; and     c) dyspnea (shortness of breath).  Patients recovering from cardiac surgery who are indicated for outpatient arrhythmia     monitoring.  ECG data recorded by the device can be analyzed by other processing systems, such as    the BMS Century Holter system to provide Holter style reports.Contraindications :The TruVue System is contraindicated for those patients requiring attended, In-hospital monitoring for life threatening arrhythmias.This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.NOTICE:  Changes or modications to this equipment not expressly approved by Biomedical Systems may void the FCC authorization to operate this equipment.  Sensor FCC ID: YCVBRSA01 Handheld FCC ID: YCVBRHA01 / YCVBRHA02Copyright © 2012 Biomedical Systems. All Rights Reserved.Patent Pending. Biomedical Systems reserves the right to change specications at any time without notice.
5PHYSICIAN’S OPERATION MANUALTruVue System OverviewSystem OverviewThe TruVue system is a wireless ECG analysis and monitoring system used for the diagnosis of cardiac arrhythmia in ambulatory patients. ECG data is acquired from the patient on a body worn sensor, stored and then transmitted to a data center through a handheld device carried with the patient. No action is required by the patient to transmit ECG data. At the data center, all ECG is stored and then analyzed for arrhythmia. Portions of the ECG containing arrhythmic events are transmitted to our monitoring center for human conrmation before being compiled into a report and transmitted to the physician. The system also allows for real time 2-way communications of voice and data between the patient and the monitoring center.  5 TruVue  System Overview   System Overview The TruVue system is a wireless ECG analysis and monitoring system used for the diagnosis of cardiac arrhythmia in ambulatory patients. ECG data is acquired at the patient on a body worn sensor, stored and then transmitted to a data center through a handheld device carried with the patient. No action is re-quired by the patient to transmit ECG data. At the data center, all ECG is stored and then analyzed for arrhythmia. Portions of the ECG containing arrhythmic events are transmitted to monitoring center for human confirmation before being compiled into a report and transmitted to the physician. The system also allows for real time 2-way communications of voice and data between the patient and the monitor-ing center or physician. CAUTION: The TruVue system is not an emergency response device. The patient should call 911 and/or their local emergency medical service if they feel they are having a medical emergency. Note: The TruVue system does not provide interpretative statements. Interpretation and clinical diag-nosis is the responsibility of the physician. PHYSICIAN’S OPERATION MANUAL
6PHYSICIAN’S OPERATION MANUALTruVue System OverviewPatient DevicesThe patient devices consist of a body worn Sensor, a Handheld device that provides communication and the patient user interface, and a charger for the Handheld. The sensor acquires and stores 2 channels of full disclosure ECG data covering the entire monitor-ing period (up to 30 days). While aquiring ECG data, the sensor also continuously transmits the full disclosure data to the handheld over a radio link with a range up to approximately 100 feet.The handheld continuously transmits the full disclosure ECG data over the cellular network to the 24/7 attended BMS monitoring center, where the ECG is analyzed. Any detected arrhythmias are conrmed by our certied monitoring technicians before being reported to the physician.If the patient is symptomatic, they can enter their symptoms on the handheld. These symptoms are immediately transmitted to the monitoring center for review and correlation with the ECG data.Text messages and voice calls can be placed to the patient handheld any time the device is in cellular coverage. TruVue HandheldTruVue Sensor   Applied Sensor
7PHYSICIAN’S OPERATION MANUALTruVue System OverviewData and Monitoring CenterFull disclosure ECG data transmitted from the handheld is stored in the BMS monitoring center, where arrhythmia analysis algorithms analyze for:• Pause / Asystole• Tachycardia• Bradycardia• Atrial Fibrillation• Idioventricular Rhythms• Supraventricular Tachycardia• Ventricular Tachycardia• Ventricular FibrillationWhen one of the above arrhythmias is detected, a certied monitoring technician conrms the arrhythmia and prepares and annotates a sample to be included on a physician report. A report is sent immediately to the physician if the arrhythmia meets the immediate report criteria specied for the patient, or sent on a daily summary report per physicians orders.A daily or weekly summary report is prepared per the prescribing physician’s preference that can include:• Heart Rate Trend graph• Atrial Fibrillation Burden graph• Samples of any arrhythmias detected, or ECG samples at the high    and low HR if there were no arrhythmiasReports can be faxed, mailed, and/or viewed and printed on-line.  Prior to printing your patient’s report, you may enter any comments or interpretations on the report.The TruVue system allows you to view your patient’s monitoring record at any time, including all reports, samples and full disclosure ECG data since the inception of the monitoring period.
8PHYSICIAN’S OPERATION MANUALTruVue Service OverviewOrdering TruVueThe TruVue Mobile Telemetry system is provided as a service by Biomedical Systems.  You or your staff may order this service for your patients by logging on to our on-line web application, Global Cardio, and completing the on-line patient enrollment form.  If you do not have the Global Cardio application installed, please contact Biomedical Systems to arrange installation.NOTE:  When ordering TruVue for your patients, all physician orders require the following information to be provided to Biomedical Systems:• Patient name (rst, last, and middle initial)• Patient I.D. and Date of Birth• Patient demographics (home address, telephone number, cell phone number, etc.)• Patient primary and secondary insurance information (ID #, group #, address, telephone #)Upon receipt of an order for the TruVue service Biomedical Systems will: A) Conrm the insurance coverage for the patient.B)  Contact the patient and conrm the delivery address for the device kitC) Congure the device for your patient and ship the device kit and all consumables required for      the  entire monitoring period.Please discuss with your patients:• Reason for ordering mobile telemetry• Importance of proper hook-up and securing electrodes to skin• Change sensor battery every 24-hours and charge handheld unit during times of sleep• Anticipated monitoring duration• Instruct them to go to the nearest emergency room or call 911 in the event of a life-threatening    EmergencyIn addition to discussing the above with your patient, we encourage you to provide the patient with contact information for Biomedical Systems.
9Initiating Monitoring When the patient receives the device, Biomedical Systems will speak with them to walk themthrough the hook up and verify the proper operation of the system. Our certied monitoring technicians will:A) Conrm the identity of the patient.B) Review proper device operation with the patient. C) Instruct the patient on the proper application of electrodes and how to begin monitoring.D) Take a baseline recording and verify proper operation of the device.Electrode Site Preparation and Proper Positioning  White Electrode                  Black Electrode  (RA)                      (LA)                         Red Electrode  Figure 1                     (LL)CAUTION:  Shave any hair that is in the area the electrodes are placed. 1.  Shave area where electrodes will be        placed (if applicable).  Wipe each area with alcohol    in a circular motion and let dry.   (See Figure 1 for electrode placement) 2.   Remove the sensor from the box.  Snap       each lead wire onto an electrode. (See Figure 2)3.   Remove backing from the       electrode attached to the black      snap and place it on the left      side  of your upper chest just       below your clavicle as shown in       Figure 1.                     Figure 2PHYSICIAN’S OPERATION MANUALTruVue Service Overview
10PHYSICIAN’S OPERATION MANUALTruVue Service Overview4.   Remove backing from the electrode attached to the  red snap and place it on the lower left portion of your  chest as shown in Figure 1.5.   Remove backing from the electrode attached to the  white snap and place it on the right side of your upper  chest just below your clavicle as shown in Figure 1.CAUTION:  Press rmly all around electrode patches to secure them rmly to skin.Concluding MonitoringWhen monitoring is nished, Biomedical Systems will contact the patient and arrange for the device to be returned to us.  Our monitoring staff will prepare a summary report for your review.If you reach a diagnosis for your patient prior to the end-monitoring date or wish to extend the monitoring period past the date, please contact the Biomedical Systems monitoring center.Breaks in MonitoringThe monitoring period can be suspended and resumed later if the patient requires a hospitilization or a break in service for any other reason, such as out of the country travel.  During a monitoring break you will not receive any daily reports.  Electrode Site Placement and Proper Positioning- Continued
11PHYSICIAN’S OPERATION MANUALOperation and Performance SpecicationsECG Acquisition and StorageThe Sensor acquires two channels of ECG through a three wire shielded cable connected to standard Holter monitoring electrodes. Standard leads II and III of the Einthoven triangle are sampled at 1000 samples per second (SPS) with +/- 40 mV of dynamic range with .05 to 150 Hz band pass. The data is then ltered and down sampled to 250 (SPS) before being stored on the  sensor. The sensor retains up to 30 days of  ECG data. ECG data is stored with the patient ID and an error detecting code.ECG TransmissionThe Sensor transmits the acquired ECG data to the handheld over an encrypted Bluetooth link with a range of  up to approximately 100 feet. The range of this link can vary depending on environmental factors. If the sensor goes out of range of the handheld the patient will be alerted. The handheld and sensor are paired together prior to providing the kit to the patient and will only communicate with each other. Neither the sensor or handheld will communicate with other devices (they are “non-discoverable” and “non-connectable” per the Bluetooth specication). The ECG is protected from data corruption by an error detecting code that “travels” with ECG data throughout the TruVue system, ensuring that no corruption of the data occurs during transmission to and storage at the monitoring center.The Sensor can be placed in “airplane mode” through the handheld user interface. This turns off all radios so the patient can continue to collect ECG data (but not transmit it to the handheld) in areas where wireless devices are not allowed. The stored ECG is transmitted to the monitoring center when the radios are turned back on. If the radios are turned off when the handheld is powered up the patient is prompted to turn them on again.User InterfaceThe Sensor will alert the patient with a speaker tone and a ashing LED when the battery is low, if a lead falls off, or if the sensor is out of range of the handheld. The patient can silence an alert temporarily by using the large pushbutton on the sensor if they choose. AlgorithmWhen communications between the Handheld and the Biomedical Systems data center are interrupted for any reason the sensor runs a rhythm analysis algorithm that detects potentially signicant arrhythmias that have a high heart rate or ventricular rhythm. If a potential arrhythmia is detected then an alert is presented to the patient that instructs them to move to an area with cellular coverage so the ECG data can be transmitted to the data center for analysis and conrmation of the rhythm.Powering the sensorThe sensor is powered by inserting the battery in the battery compartment. It is always on - there is no separate on/off switch. The patient replaces the battery in the sensor once a day with the supplied lithium  AAA battery.Sensor Operation
12PHYSICIAN’S OPERATION MANUALLeadset:The lead wires are permanently attached to the sensor hardware and are not user replaceable.CAUTION: Do not attempt to remove the lead wires from the sensor.CAUTION: Inspect the leadwires for any fraying and/or cracking in the insulation prior to use.Note: BMS will perform this check before providing the equipment to the patient.Lanyard:A lanyard (neck strap) is attached to the sensor for the convenience of the patient and to prevent the sensor and lead wires from dangling.  The lanyard can be removed from the sensor if desired.Battery:The sensor is powered by a 1.5V AAA lithium battery. CAUTION: Use only the supplied  lithium AAA batteries that are provided with the patient kit.Do not use rechargeable batteries.CAUTION: Do not store sensor with the battery in place for extended periods of time.  Remove the battery after each monitoring period.Cleaning:The sensor may be cleaned with Isopropyl Alcohol. Do not submerse the sensor in any liquid.Handling precautions:To ensure proper operation of the sensor please follow these handling precautions:CAUTION: Do not drop the sensor or handheld unit. CAUTION: Do not pull or yank on the sensor lead wires. CAUTION: Do not expose sensor  or handheld to excessive dust or to extreme temperatures. CAUTION: Do not immerse or otherwise allow any liquid to enter the sensor or handheld.  CAUTION: Do not store the sensor or handheld units in direct sunlight or near corrosive liquids.CAUTION: Do not allow sensor to get wet. When Showering or Bathing:• Remove lead wires attached to sensor from the electrodes.• Place sensor (attached to lanyard) in a dry secure place.  • Remove electrodes (patches) from skin even if they have already been changed in past 24 hours.  • After showering or bathing, dry skin thoroughly.     Do not apply powder or lotion of any kind to chest area.• Wipe skin with alcohol in area where electrodes (patches) will be placed.     Replace electrodes (patches) as previously instructed.Sensor OperationOperation and Performance Specications
13PHYSICIAN’S OPERATION MANUALSensor Performance SpecicationsStandardsThe sensor complies with the following medical device standards:-AAMI EC 38-1998, Ambulatory Electrocardiographs.-EN60601 -1 Medical electrical equipment, Part 1: General requirements for safety-EN60601 -1 Medical electrical equipment, Part 1-2: Electromagnetic compatibilitySensor Performance Specications    Parameter    Notes  Min.  Typ.  Max.  Unit  Physical:  Length        3.1    in.   Width        1    in.  Thickness        .8    in.  Weight    With AAA Battery    60    gm  ECG Cable Lengths  Dual channel 3 electrode     18    in.          Parameter    Test Conditions  Min.  Typ.  Max.  Unit   Environmental:      Operating Temperature    0    45  ºC Storage Temperature    -10    60  ºC Relative Humidity     10    95  % Shock-Unpackaged Unit    36      in. Water Resistance    Not Water ResistantComplies with AAMI-EC38and EN60601-1Operation and Performance Specications
14PHYSICIAN’S OPERATION MANUALSensor Performance Specications - Continued  Parameter  Test Conditions  Min.  Typ.  Max.  Unit   Electrical:  Battery Voltage  1 -AAA Lithium Energizer  0.9  1.5  3.0  Volts       Battery Current  At 1.5V Battery Voltage, all      400  mA      circuits turned on  Lithium Battery  Lithium-Ion Battery Not User  2.0    3.1  Volts  Voltage  Replaceable  VREF Voltage    1.22  1.25  1.0  Volts  Reference  VREFAD Voltage    2.45  2.55  2.60  Volts  Reference  Input Impedence  @ 5 Hz  1.0  1.5  1.6  MW  CMRR  @ 60 Hz  86      dB   CMR Range   AC + DC  -1.5    +1.5  Volts  Differential Range  AC  -40    +40  mV      DC+80 mV AC @ 5 Hz  -500    +500  mV  Fast Baseline Reset-  After Removing Overloading  0.45  0.5  0.55  Hz  3 db Frequency  Signal  Bandwidth  +1 dB referenced to 15  0.05    150  Hz      Hz  Low Pass Filter Gain  @ 250 Hz  -18    -17  dB  Pacemaker Pulse  1 microsecond max pulse rise and fall  Detection  times  Pacemaker Pulse    0.2    2.5  msec  Width  Pacemaker Pulse    1.0    250  mV  Amplitude Communications    Frequency  Bluetooth Class 2,3    2.4    gHz  Communications  Bluetooth SPP Prole, non    2.2    Ver  Protocol  discoverable  Output power      0    dB User Interface  Complies with AAMI-EC38      and EN60601-1  Pushbutton  Used for silencing alerts  LED    For device alerts   Speaker  For device alerts    400-2500    HzComplies with AAMI-EC38and EN60601-1Operation and Performance Specications
15PHYSICIAN’S OPERATION MANUALHandheld OperationCommunicationsThe Handheld communicates with the sensor over an encrypted Bluetooth link with a range of up to approximately 100 feet. The range of this link can vary depending on environmental factors. If the sensor and handheld lose communication the patient will be notied with a short beep on the sensor.The Handheld transmits ECG data to the BMS data center over the cellular network. Transmissions are bursted with a maximum latency of 2.5 minutes when the Handheld is in coverage on the network. When out of coverage of the cellular network the handheld commands the sensor to run the potential arrhythmia detector algorithm (described under the Sensor Operation section).ECG data is transmitted to the data center without modication and is protected from corruption by an error detecting code embedded in the ECG data.Text messages can be sent from the BMS monitoring center for display on the patient’s handheld device.The Handheld can also receive a voice call from the monitoring center in the event that monitoring staff needs to speak with the patient and they cannot be reached at their regular phone numbers. Only the monitoring center can initiate a voice call, the handheld will only accept incoming calls from the monitoring center, and the patient cannot initiate an outgoing call. CAUTION: The handheld is a cellular phone. Follow your implantable device manufacturers recommendations on the use of cellular phones with your implant. Operation and Performance Specications
PHYSICIAN’S OPERATION MANUALTruVue Indications of Use16Handheld OperationPowering the handheld:The handheld is powered by an internal rechargeable Lithium Ion battery. The on/off button is located on the right side of the unit and the handheld is powered on by pressing and holding the power button for approximately 5 seconds. The provided wall charger charges the battery. The handheld can be powered on whenever the wall charger is attached, regardless of whether the battery is depleted or not.The handheld battery will typically power the handheld 16 hours without recharging. The patient should leave the handheld attached to the wall charger while they are sleeping.CAUTION: Do not attempt to replace the handheld battery.CAUTION: Use only supplied wall charger with the handheld.Cleaning:The handheld may be cleaned with Isopropyl Alcohol. Do not submerse the handheld in any liquid.Handling precautions:To ensure proper operation of the handheld please follow these handling precautions.CAUTION: Do not drop the handheld unit. CAUTION: Always use the carrying case when carrying the handheld on your body.CAUTION: Do not expose sensor or handheld to excessive dust or to extreme temperatures.CAUTION: Do not immerse or otherwise allow any liquid to enter the sensor or handheld.  CAUTION: Do not store the sensor or handheld unit in direct sunlight or near corrosive liquids.Operation and Performance Specications
17PHYSICIAN’S OPERATION MANUALUser Interface:A graphical user interface is incorporated for display of messages to the patient, input of symptoms and control of the system.  A 5-way set of navigation keys and two soft keys are used to navigate the userinterface.  An LED illuminates to indicate the status of the device when it is on and to indicate the charging status when the unit is off . The handheld incorporates a loudspeaker and vibrator for alerting the patient.Operating modes:The TruVue device kit operates in two primary modes, monitoring and pre-monitoring. The unit is provided to the patient in pre-monitoring mode and is activated into monitoring mode by a BMS monitoring technician during the hook-up call from the patient. In pre-monitoring mode the patient kit does not record, store or transmit ECG data.User Interface in Pre-monitoring mode:When the patient receives the unit and turns it on they will proceed through the following sets of screens that instruct the patient to call the BMS monitoring center for hook-up instructions:During the hook-up call, the BMS monitoring center technician will perform the hook-up (see “Initiating Monitoring” in the “Service Overview” section) and provide the patient with a code that enables the transition of the unit into monitoring mode. At this point the devices are actively monitoring the patient.Operation and Performance Specications
18User Interface in Monitoring ModeMain screenThe main screen is the top level screen for the user interface and is displayed when the device is turned on and whenever the LCD wakes up from power saving mode. It consists of the status bar, the task wheel and softkey area. The status bar displays various indicators of device operation. The task wheel scrolls using the up and down arrow keys and allows the patient to select various tasks to perform by selecting the center key. The softkey area contains two indicators that change depending on what state or screen the device is in.    Status Bar           Task Wheel     Softkeys   Task Wheel OptionsPHYSICIAN’S OPERATION MANUALRecord symptomsand activity levelRead Text messages fromthe monitoringcenterTurn wirelessradios on oro.  Place thedevice in“ight mode”View large battery level indicators forthe handheldand sensorAdjust the volume and vibrator for boththe handheld and sensorView deviceinformationsuch as serial number, SW versions, etc.Operation and Performance Specications
19PHYSICIAN’S OPERATION MANUALUser Interface in Monitoring ModeStatus Bar IndicatorsUpload ActivitySensor BatterySensor ConnectionCell StrengthHandheldBatteryFull High Low Empty Absent ChargingBattery Level IconsCell Strength Icons0 1 2 3 4 Cell OHandheld OnlySensor ConnectedSensor DisconnectedSensor ConnectionUpload ActivityUpload InactiveUpload ActivityOperation and Performance Specications
20PHYSICIAN’S OPERATION MANUALUser Interface in Monitoring:Record Symptoms Screens:Pressing the center key when the task wheel is on the Record Symptoms task allows the patient to enter both their current symptoms and their current activity level. This information is uploaded to the monitoring center and is available for correlation with the patient’s ECG at the time they recorded the symptom.Messages Screens:Pressing the center key when the task wheel is on the Messages task allows the patient to read textmessages sent from the monitoring center. Messages can be entered through GlobalCardio. The indicator on the main screen changes to indicate that there are unread messages waiting. The handheld can store up to 4 messages. If a 5th message is received then the earliest read message will be deleted.Unread message waiting All messages have been readOperation and Performance Specications
21PHYSICIAN’S OPERATION MANUALUser Interface in Monitoring Mode:Wireless Settings Screens:Pressing the center key when the task wheel is on the Wireless Settings  task allows the patient to turn off all wireless radios in the system for a time.  This is a useful feature if the patient is on an airplane or in some other area where cellular phones are not allowed. The sensor will continue to record all ECG data while the handheld radio is off. When the radios are turned back on, all stored data will be transmitted to the monitoring center.When the device is turned back on, the user is always prompted to turn on the radios if the handheld is in “airplane” mode.The user is also prompted every two hours to turn the radios on through an alert message. The potential arrhythmia detection algorithm does not run when the device is in “airplane” mode, since it is assumed that the patient is unable to transmit any data due to their physical location.Operation and Performance Specications
22PHYSICIAN’S OPERATION MANUALUser Interface in Monitoring Mode:Battery Level Screen:Pressing the center key when the task wheel is on the  Battery Levels task allows the patient to view the large battery level indicators.  Volume Screens:Pressing the center key when the task wheel is on the Volume task allows the patient to adjust the volume and vibrate levels on the handheld.   The sensor volume can only be set when the sensor is connected to the handheld.  To set to vibrate only, the volume slider is moved to 0 and the vibrate indicator is automatically checked.  Vibration can also be selected in addition to volume. Operation and Performance Specications
23PHYSICIAN’S OPERATION MANUALUser Interface in Monitoring Mode:Device Info Screens:Pressing the center key when the task wheel is on the Device Info task allows the patient to view the following device information:    • Cellular Network Signal Strength• Handheld and Sensor Serial Numbers • Patient ID• Handheld and Sensor Software Versions• Wireless Setting  (Normal or Off)Lock/Unlock Screens:To conserve battery life, there is a lock and unlock screen feature on the Handheld unit.  It can be activated manually, or it will automatically go into lock mode after 2 minutes if you have not pressed a key on the key pad.  If you want to lock the screen manually, press the “Lock” key.  Press and hold the Conrm key for 2 secondsto lock the screen. The patient’s ECG will continue to be sent to our monitoring center for review. To unlock the keypad, press any key other than the left softkey to display the lock screen and then press the “Unlock” key.Operation and Performance Specications
24PHYSICIAN’S OPERATION MANUALHandheld Performance SpecicationsStandardsThe handheld complies with the following medical device standards:-AAMI EC 38-1998, Ambulatory Electrocardiographs.-EN60601 -1 Medical electrical equipment, Part 1: General requirements for safety-EN60601 -1 Medical electrical equipment, Part 1-2: Electromagnetic compatibilityParameter Notes Min. Typ. Max. UnitPhysical:Length 5 in.Width 2.25 in.Thickness .8 in.Weight 150 gmUser InterfaceDisplay 240X320 QVGA OLEDKeypad 5 way navigation + 2 softkeysReceiver/mic For voice callLoudspeaker For device alertsEnvironmental: Complies with AAMI-EC38and EN60601-1Operating Temperature0 45 ºCStorage Temperature-10 60 ºCRelative Humidity10 95 %Shock-Unpackaged unit Per AAMI-EC38 36 in.Shock-Packaged Unit Per AAMI-EC38Water Resistance IPX 0Operation and Performance Specications
25PHYSICIAN’S OPERATION MANUALHandheld Performance SpecicationsParameterLithium Battery Voltage Lithium-Ion Battery Not User Replaceable4.2 VoltsLithium Battery Current 1 AmpsCharger EN60601 Approved directplug-in Class II AC adapterpower supply rated 100-240V~100 240 VoltsCommunicationsCellular EGSM/GPRS/EGPRS900/1800/850/1900 MHzBluetooth Bluetooth Class 1Communications Protocol Bluetooth SPP Prole , non discoverable2.2 VerOperation and Performance Specications
26PHYSICIAN’S OPERATION MANUALAlgorithm Operation and PerformanceSignicant Arrhythmia:The TruVue arrhythmia detection algorithm continuously processes ECG transmitted from the patient devices and detects the following rate, rhythm, and morphology based arrhythmias:When an arrhythmia is detected, it is agged for immediate review by a BMS certied cardiactechnician. The technician conrms the arrhythmia and prepares a report for immediate transmission if the arrhythmia meets the Signicant Arrhythmia criteria specied by the physician.Representative Samples, Trend and Arrhythmia Burden:In addition to detecting signicant arrhythmias, the TruVue analysis algorithm runs an additional scan of each 24 Hour ECG period.  During this second pass, the algorithm collects additional arrhythmia examples, minimum and maximum HR strips, HR trend data, and AF burden information.This data is presented to the monitoring staff for validation and inclusion on the Daily or Weekly report.  It may also be reviewed on-line at any time in Global Cardio. Representative samples are a set of rhythm strips that represent the patient’s condition for that day.  The algorithm identies the following samples:1. If no arrhythmia occurred, the algorithm identies the lowest noise, highest heart rate and lowest noise, lowest heart rate samples.2. If arrhythmias did occur (for each signicant arrhythmia class dened above) the algorithm identies the most serious, lowest noise sample. The HR trend graph presents the patient heart rate represented as a moving average over every 8 beats.The Atrial Fibrillation burden graph presents the amount of time the patient was in AF with 10 minute resolution. If the patient was in Atrial Fibrillation for over 30 seconds during any 10 minute period then that period is marked as AF. This graph can selectably be presented for the current day or for the entire monitoring period to date.  Tachycardia  Bradycardia  Pause/Asystole  Atrial Fibrillation  Idioventricular Rhythm  Supraventricular Tachycardia  Ventricular Tachycardia  Ventricular Fibrillation Operation and Performance Specications
27PHYSICIAN’S OPERATION MANUALAlgorithm Operation and PerformanceSignicant Arrhythmia Criteria: Any technician-conrmed serious arrhythmia will be transmitted immediately on a Signicant Event report by the physician’s preferred method of notication.The default criteria are:Arrhythmia Default Criteria Criteria RangePause/Asystole > 3 seconds 1-5 secondsBradycardia < 40 bpm,  >30 sec 20-80 bpmTachycardia > 180 bpm, sustained for 15+ beats 120-300 bpmSupraventricular Tachycardia > 150 bpm,  > 30 sec 100-200 bpm, 5-60 secVentricular Tachycardia Rate: > 130 bpm(4 or more beats)Rate: 80-150 bpmBeats: 3-10Idioventricular Rhythm > 15 beats, HR < dened VT Rate 5-50 beatsVentricular Fibrillation AlwaysAtrial Fibrillation First onset for patientThen rate > 150 or < 40 bpm1-10 OnsetsRate: 20-220 bpmPatient Initiated Always sentThe default criteria above can be used or the physician can specify the criteria to be used as long as it falls within the criteria range specied above. The criteria can be specied for a particular patient, for all the physician’s patients, for a particular ofce location or for the entire practice.Operation and Performance Specications
28PHYSICIAN’S OPERATION MANUALAlgorithm Operation and PerformanceBeat Detection and Classication:The TruVue algorithm can discriminate between  normal and ventricular beat morphologies. For each beat complex the algorithm determines the R-point for HR calculation.The beat detection performance (as tested under ANSI/AAMI-EC 57:1998, Testing and ReportingPerformance Results of Cardiac Rhythm and ST Segment Measurement Algorithms) is:Sensitivity, % Positive Predictivity, %QRS Detection (MIT DB) 99.91 99.87QRS Detection (AHA DB) 99.72 99.82QRS Detection (NST DB) 95.17 87.57V-morphology Detection (MIT DB) 91.83 93.56V-morphology Detection (AHA DB) 76.24 92.08V-morphology (NST-DB) 88.90 46.78Heart Rate Averaging:The heart rate is averaged over 8 R-R intervals (HR = 480/duration of 8 consecutive RR intervals in seconds) and becomes the basis for rate based arrhythmia detection following the beat classication step. The HR calculation had a mean RMS error of 1.735 as tested per EC-57 on the MIT database.Atrial Fibrillation Detection Algorithm:The Atrial brillation algorithm detects the irregularity of R-R intervals and examines the signal for utter waves. When a certain irregularity is detected, the algorithm performs additional checks to determine if the underlying rhythm is bigeminy or trigeminy and looks at the presence of utter waves as a secondary indicator. Atrial Fibrillation detection performance as tested per EC-57 is:Sensitivity, % Positive Predictivity, %Atrial Fibrillation detection - all events(MIT-DB)92 100Atrial Fibrillation detection - eventslonger than 30 seconds (MIT-DB)100 100Operation and Performance Specications
29PHYSICIAN’S OPERATION MANUALAlgorithm Operation and PerformanceVentricular Fibrillation Detection:The TruVue algorithm can detect VF rhythms with the following performance as tested under EC-57Sensitivity, % Positive Predictivity, %Ventricular Fibrillation Detection(MIT DB)100 100Ventricular Fibrillation Detection(AHA DB)90 100Operation and Performance SpecicationsVentricular Fibrillation Detection(CU DB)97 73
PHYSICIAN’S OPERATION MANUALTruVue Indications of Use30Description of Device SymbolsThis equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.These limits are designed to provide reasonable protection against harmful interference in a residential installation.  This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.  However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:  * Reorient or relocate the receiving antenna. * Increase the separation between the equipment and receiver. * Connect the equipment into an outlet on a circuit different from that to which the    receiver is connected. * Consult the dealer or an experienced radio/TV technician for help. Type BF Electrical IsolationRead Manual FirstDC CurrentRadiofrequency radiation exposure information:For body worn operation, this phone has been tested and meets the FCC RF exposure guidelines when used with the Biomedical Systems accessories supplied or designated for this product.  Use of other accessories may not ensure compliance with FCC RF exposure guidelines.
31PHYSICIAN’S OPERATION MANUALSummary of Caution statements:CAUTION: Do not attempt to remove the lead wires from the sensor.CAUTION: Inspect the leadwires for any fraying and/or cracking in the insulation prior to use.CAUTION: Do not drop the sensor or handheld unit. CAUTION: Always use the carrying case when carrying the handheld on your body.CAUTION: Do not pull or yank on the sensor lead wires CAUTION: Do not expose sensor  or handheld to excessive dust or to extreme temperatures CAUTION: Do not immerse or otherwise allow any liquid to enter the sensor or handheld.  CAUTION: Do not get the sensor or handheld wet.  CAUTION: Do not store the sensor or handheld unit in direct sunlight or near corrosive liquidsCAUTION: Do not attempt to replace the handheld battery.CAUTION: Use only supplied wall charger with the handheld.CAUTION: The handheld is a cellular phone. Follow your implantable device manufacturers recommendations on the use of the cellular phones with your implant. CAUTION: Press rmly all around electrode patches to secure them rmly to skin.  CAUTION: Shave any hair that is in the area the electrodes are placed. CAUTION: The TruVue system is not an emergency response device. The patient should call 911 and/or their local emergency medical service if they feel they are having a medical emergency.CAUTION:  Federal Law (USA) restricts this device to sale by or on the order of  a physician.Note: The TruVue system does not provide interpretative statements. Interpretation and clinical diagnosis is the responsibility of the physician.Summary of Caution Statements
MFG-LB-201.002Authorized Representative:  Tessa De SmetBiomedical Systems SA/NV1945 Chausee de Wavre, B1160Brussels, Belgium(Belgium) +11 32 2 661 20 70, Fax +11 32 2 661 20 710843