Carl Zeissc`m=C=c`m=ticççí=`çåíêçä=m~åÉäUser ManualG-30-1706-enVersion 1.12009-01-15
FCP & FCP WL Foot Control PanelVersion 1.1Page 2 G-30-1706-en About this manual The user manual is part of the scope of delivery.• Carefully read it before using the device.• Keep it at the site of use of the foot control panel.• Store it for the entire service life of the foot control panel.• Pass it on to every subsequent owner or user of the foot control panel.Orientation aids – The chapter overview at the beginning of the user manual provides a sum-mary of all subjects.– The contents of each chapter are specified in detail at the beginning of each chapter.– A keyword index at the end of the manual facilitates the search for specific terms.Applicable area This user manual applies to foot control panels with the following identifica-tion:– Rating label numbers: 304970-9100-000 304970-9200-000Information on the manufacturer:Subject to change in design and scope of delivery and as a result of ongoing technical development. Printed in Germany.© Carl Zeiss Surgical GmbH 2009Carl Zeiss Surgical GmbHA Carl Zeiss Meditec Company73446 OberkochenGermanyFax: + 49 (0) 7364 - 20 4823Email: surgical@meditec.zeiss.comInternet: www.meditec.zeiss.com
FCP & FCP WLFoot Control Panel Version 1.1G-30-1706-en Page 3Chapter overviewChapter: Safety measures 5Chapter:  Design and function 19Chapter:  Preparations for use 31Chapter: Operation 41Chapter:  What to do in the event of malfunctions 45Chapter:  Care and maintenance 51Chapter: System data 55Chapter: Index 61
FCP & FCP WL Foot Control Panel Version 1.1Page 4 G-30-1706-en
Version 1.1G-30-1706-en Page 5FCP & FCP WLFoot Control PanelSafety measuresWe would like to provide you with information about safety aspects which must be observed when handling this foot control panel. This chapter con-tains a summary of the most important information concerning matters rele-vant to instrument safety. Key to the symbols in this user manual .............................................6Hazard symbols ...........................................................................................6Information symbols....................................................................................7Directives and standards ...................................................................7Target group .....................................................................................8Field of application ............................................................................8Intended use ...............................................................................................8Normal use..................................................................................................8Typical misuse .............................................................................................8Notes for the operator.......................................................................9Duties of the operator .................................................................................9Requirements to be met by the user ..........................................................12Warranty and liability.................................................................................13Requirements for operation.............................................................14Before every use........................................................................................14During use.................................................................................................14After every use ..........................................................................................15Warning labels and notes ................................................................16
Safety measures FCP & FCP WL Foot Control PanelVersion 1.1Page 6 G-30-1706-enKey to the symbols in this user manualHazard symbolsThe following safety information has been incorporated in the user manual. Please note this information and be particularly careful in these cases.WARNING Indicates a hazard which can lead to death or severe injury if it is not pre-vented.CAUTION Indicates a hazard which can cause slight and medium injury if it is not pre-vented.NOTE Indicates a hazard which can cause damage to material if it is not pre-vented.
Version 1.1G-30-1706-en Page 7FCP & FCP WLFoot Control Panel Key to the symbols in this user manualInformation symbolsThe following information symbols are used in this user manual:–Listing9Prerequisite for an action • Prompt for actionÆResult of an actionAdditional information and tips. No warnings of hazards are provided.Directives and standardsThe foot control panel described in this manual has been designed and tested in accordance with Carl Zeiss safety standards as well as German and interna-tional standards. This guarantees a high degree of safety.The foot control panel has been designed in compliance with the require-ments of:– EN (European standard)– IEC (International Electrotechnical Commission)– UL (Underwriters Laboratories)In accordance with Directive 93/42/EEC for medical devices, the complete quality management system of the company Carl Zeiss Surgical GmbH, 73446 Oberkochen, Germany, has been certified by DQS Deutsche Gesellschaft zur Zertifizierung von Managementsystemen GmbH, a notified body, under reg-istration number 250758 MP23.
Safety measures FCP & FCP WL Foot Control PanelVersion 1.1Page 8 G-30-1706-enTarget groupThis user manual is intended for physicians, nurses and other medical staff who prepare, operate or maintain the foot control panel after appropriate training and in accordance with the instructions given in this manual. Installa-tion and service work not described in this manual must only be performed by specialists from Carl Zeiss.Field of applicationIntended useThe foot control panel is an accessory for all types of Carl Zeiss surgical micro-scopes. It enables operation of the motor-driven and controllable functions of the microscope by foot control.Normal useThis foot control panel permits either cordless or corded operation of up to 14 different functions of a surgical microscope, including its options, by foot control.Typical misuseThe foot control panel must only be operated in combination with devices from Carl Zeiss.
Version 1.1G-30-1706-en Page 9FCP & FCP WLFoot Control Panel Notes for the operatorNotes for the operatorThe correct use of the foot control panel is absolutely vital for safe operation. Therefore, please thoroughly familiarize yourself with the content of this user manual before starting up the foot control panel. Please also observe the user manuals of the other devices used. Further information is available from our service department or from authorized representatives.Duties of the operatorNotes on thefoot control panel• Make sure that the installation conditions and the use of the foot control panel meet microsurgical requirements:–low vibration– clean environment– avoidance of extreme mechanical stress.• Observe the legal regulations for accident prevention and occupational health and safety applicable in the country concerned.• Do not operate the components contained in the delivery package–in explosion-risk areas,– if inflammable anesthetics or volatile solvents such as alcohol, benzine or similar chemicals are present at a distance of less than 25 cm.• Switch off the connected foot control panel and remove the batteries if you notice any smoke, sparks or unusual noise. Do not use the foot con-trol panel until it has been repaired by our service team.• Do not force cable connections. If the male and female parts do not read-ily connect, make sure that they are appropriate for one another. If any of the connectors are damaged, have our service representative repair them. • Over long distances (e.g. removal, return for repair, etc.), the foot control panel must always be transported in the original packaging or in special return packaging. For details, please contact your dealer or the Carl Zeiss service team.• Use the foot control panel only for the application described.
Safety measures FCP & FCP WL Foot Control PanelVersion 1.1Page 10 G-30-1706-en• Operate the foot control panel only with the components included in the delivery package. If you wish to use other components, make sure that Carl Zeiss or the manufacturer of the components has verified and con-firmed that these components meet the respective safety standards and can be used without risk.• If a failure occurs which you cannot correct with the aid of the chapter "What to do in the event of malfunctions“, attach a sign to the foot con-trol panel stating it is out of order and contact our service representative.• The foot control panel must not be modified without the manufacturer's approval.• The foot control panel is a high-grade technological product. To ensure optimum performance and safe working order, we recommend having it checked on a regular basis by our service representative or trained special-ist staff.Notes on EMC (Electro-magnetic Compatibility)The foot control panel complies with the EMC requirements of IEC 60601-1-2. For operating the foot control panel, observe the EMC precau-tions specified below.• Only use options approved by Carl Zeiss for this foot control panel. • Do not use any portable or mobile RF communication equipment in the vicinity of the foot control panel as this may lead to an impairment of the foot control panel's function.• The foot control panel meets the RFI requirements of Class B. It cannot be ruled out, however, that interferences may occur in RF receivers (e.g. TV or radio sets) in the vicinity. If such interferences are noticed, please in-form the Carl Zeiss Service Dept.Information on exposure to radio frequency radiation in accordance with FCCThe emitted output power of the component is far below the FCC limit values for radio frequency exposure. Nevertheless, the component should be used in such a way that potential contact with persons during normal operation is kept to a minimum.NOTE ClassThis Class B digital component complies with Canadian standard ICES-003.
Version 1.1G-30-1706-en Page 11FCP & FCP WLFoot Control Panel Notes for the operatorNOTE Limit values and what to do in the event of interference with digital equipmentThis component has been tested and found to comply with the limits for a Class B digital device in accordance with Part 15 of FCC regulations. These limits have been stipulated to provide adequate protection against harmful exposure when the component is operated in residential areas. This compo-nent generates and uses radio frequency energy and may emit such energy. If not installed and used in accordance with the relevant instructions, it may cause interference to radio communication. However, there is no guarantee that interference will not occur in a particular installation. If this component causes interference with radio and TV reception, which can be determined by turning the component off and on, the user should try to correct the in-terference by one or several of the following measures:– Relocate or reorient the receiving antenna.– Move the component further away from the receiver.– Plug the component into a different outlet so that it is not on the same circuit as the receiver.– Consult an experienced radio/TV technician.NOTE Part 15 of FCC regulationsThis component complies with Part 15 of FCC regulations [and with the In-dustry Canada RSS-210 standard].Operation of the component is subject to the following two conditions:– the component must not cause any harmful interference and – the component must accept any interference received, including interfer-ence that may cause undesired operation.NOTE Modification or conversionAny modification or conversion not explicitly approved by Carl Zeiss will in-validate the FCC license for the operation of this component.
Safety measures FCP & FCP WL Foot Control PanelVersion 1.1Page 12 G-30-1706-enRequirements to be met by the user• The foot control panel must only be used by staff who have undergone appropriate training and instruction. It is the duty of the customer or insti-tution operating the component to provide training and instruction for the relevant staff.• Please keep the user manual where it is easily accessible at all times for the persons operating the system.• Modifications and repairs of the foot control panel or any components op-erated together with the foot control panel may only be performed by our service representative or by other authorized persons.
Version 1.1G-30-1706-en Page 13FCP & FCP WLFoot Control Panel Notes for the operatorWarranty and liabilityWarranty and liability depend on the applicable contractual stipulations.NOTE Loss of warrantyThe manufacturer is not liable for any damage caused by unauthorized per-sons tampering with the foot control panel. Furthermore, this will forfeit any rights to claim under warranty.
Safety measures FCP & FCP WL Foot Control PanelVersion 1.1Page 14 G-30-1706-enRequirements for operationOur service representative or an expert authorized by us will install the foot control panel. Please ensure that the following requirements are met for fur-ther operation:9The connecting components have been properly connected. The screw connections have been firmly tightened.9All cables and plugs are in perfect condition.Before every use• Make sure that all ”Requirements for operation” are fulfilled.• Go through the checklist.• Reattach any covers or caps that were removed from the foot control pan-el. Close any existing openings with the corresponding caps.During use• Constantly monitor the foot control panel during use.
Version 1.1G-30-1706-en Page 15FCP & FCP WLFoot Control Panel Requirements for operationAfter every use• Put the foot control panel in its rest position by attaching its bracket e.g. to floor stand (1). ÆThe foot control panel is shut down in the rest position to save energy.The energy-saving rest position extends the service life of the batteries in the cordless version of the foot control panel.1
Safety measures FCP & FCP WL Foot Control PanelVersion 1.1Page 16 G-30-1706-enWarning labels and notes1Type labelA - Label for the corded version of the foot control panelB - Label for the cordless version of the foot control panel– Manufacturer (company name)– Serial number (SN)–Product name– Order number (REF)– IP classification (degree of protection against the harmful ingress of water and solid particles)2"Observe the user manual" labelObserve the user manual or further applicable documents.3"Observe disposal regulations" labelElectrical or electronic devices must not be disposed of as normal domestic waste. For more information on the disposal of electrical and electronic devices, please see the chapter "Maintenance and care".4"Year of manufacture" labelThis label indicates the year of manufacturer of the system.5Radio equipment approval identification(for FCP WL only)CAUTION Note the warning labels and notes!• If you notice that any label on your foot control panel is missing or has become illegible, please contact us or one of our authorized representa-tives. We will supply a replacement.1A1BIPX8IPX8
Version 1.1G-30-1706-en Page 17FCP & FCP WLFoot Control Panel Warning labels and notes6Ready for use status indicator7Cable connection status indicator8Radio link status indicator9Battery status indicator 1 A1 B23679854IPX8IPX8
Safety measures FCP & FCP WL Foot Control PanelVersion 1.1Page 18 G-30-1706-en
Version 1.1G-30-1706-en Page 19FCP & FCP WLFoot Control PanelDesign and functionVersions of the foot control panel ...................................................21Components of the foot control panel.............................................22Standard assignment in the longitudinal configuration ...................26Standard assignment in the transverse configuration......................28
Design and function FCP & FCP WL Foot Control PanelVersion 1.1Page 20 G-30-1706-en
Version 1.1G-30-1706-en Page 21FCP & FCP WLFoot Control Panel Versions of the foot control panelVersions of the foot control panel1Cordless foot control panel with 14 functions2Corded foot control panel with 14 functions12
Design and function FCP & FCP WL Foot Control PanelVersion 1.1Page 22 G-30-1706-enComponents of the foot control panel1Connecting cableConnects the foot control panel to the appropriate connector on a sus-pension system or wall-mounted control panel.2BracketAllows you to hang up the foot control panel on the floor stand or on the wall when not in use.3ControlsButtons operated with your foot.4IndicatorThe adjustable indicator is required for the cordless version. It indicates the assignment to a surgical microscope after pairing.5JoystickPermits you to control e.g. an X-Y coupling or the motorized lateral and front-to-back tilt of the surgical microscope.6Rocker switchThe two rocker switches are used to operate the ”Zoom” and ”Focus” functions. For this, each of the rocker switches contains two switching el-ements, allowing you to control the functions by toe and heel movement without the need to shift your foot. You can optionally assign the ”Zoom” and ”Focus” functions to the switching elements of the rocker switches. Switchover is performed auto-matically on the suspension system.ALongitudinal configuration:The left switching elements of the rocker switches are used to operate the ”Zoom” function. The right switching elements of the rocker switches are used to operate the ”Focus” function. For more information, please see page 26.BTransverse configuration:The switching elements of the front rocker switch are used to operate the ”Focus” function. The switching elements of the back rocker switch are used to operate the ”Zoom” function. For more information, please see page 28. 7Foot restThe bridge  between the two rocker switches (6) serves as a support to rest your foot on.8Status indicatorsIndicate different statuses by shining/not shining.
Version 1.1G-30-1706-en Page 23FCP & FCP WLFoot Control Panel Components of the foot control panel5ABZoom FocusFocusZoom126763333334333 38
Design and function FCP & FCP WL Foot Control PanelVersion 1.1Page 24 G-30-1706-enFunctions of the status indicators– During cordless operation, you have to activate the "ready for use", "radio link" and "battery" status indicators by simultaneously pressing a button each on the left and right side of the foot control panel. The status indi-cators are only active while the buttons are pressed.– During corded operation via the connecting cable, the "ready for use" and "cable connection" status indicators are permanently lit.1Ready for use status indicator– Shining green: the foot control panel is ready for use, no error is present.– Shining amber: the foot control panel detects a temporary error and eliminates is automatically.– Shining red: the foot control panel is not ready for use, an error is present.2Cable connection status indicator– Shining green: the connecting cable has been plugged in, power sup-ply is present.– Not shining: the foot control panel operates in the cordless mode.3Radio link status indicator– Shining green: very good radio link.– Shining amber: reduced radio link, e.g. due to slight interference, shielding or too large a distance between the suspension system and foot control panel.– Shining red: no radio link, e.g. due to non-existent pairing with the suspension system, see page 34, shielding or too large a distance be-tween the suspension system and foot control panel.– Flashing amber: while pairing with the suspension system is in progress.4Battery status indicator– Shining green: batteries are OK.– Shining amber: batteries will not last much longer.• Obtain replacement batteries and keep them ready at hand.– Shining red: batteries are depleted and need to be replaced immedi-ately.
Version 1.1G-30-1706-en Page 25FCP & FCP WLFoot Control Panel Components of the foot control panelMeaning of the acoustic signalsSignal Cause Remedybeep Confirms a function --beep-beep Warning • Put the foot control panel in its rest position (vertical position), see page 38.• Check the status indica-tors, see page 24.beep-beep-beep Problem see "Troubleshooting"1243
Design and function FCP & FCP WL Foot Control PanelVersion 1.1Page 26 G-30-1706-enStandard assignment in the longitudinal configurationThe longitudinal configuration refers to the focus and zoom functions.JoystickThe joystick permits control in the following four directions: 1front,2back,3left,4right.5SwitchFunction depending on surgical microscope used6SwitchFunction depending on surgical microscope used7Focus downFocusing on a lower plane, the working distance is increased8Focus upFocusing on a higher plane, the working distance is reduced9SwitchFunction depending on surgical microscope used10 SwitchFunction depending on surgical microscope used11 Zoom inIncreasing magnification, the field of view is reduced12 Zoom outReducing magnification, the field of view is increased13 SwitchFunction depending on surgical microscope used14 SwitchFunction depending on surgical microscope used
Version 1.1G-30-1706-en Page 27FCP & FCP WLFoot Control Panel Standard assignment in the longitudinal configuration67811121314 510 91243
Design and function FCP & FCP WL Foot Control PanelVersion 1.1Page 28 G-30-1706-enStandard assignment in the transverse configurationThe transverse configuration refers to the focus and zoom functions.JoystickThe joystick permits control in the following four directions: 1front,2back,3left,4right.5Switch,Function depending on surgical microscope used6SwitchFunction depending on surgical microscope used7Focus downFocusing on a lower plane, the working distance is increased8Focus upFocusing on a higher plane, the working distance is reduced9SwitchFunction depending on surgical microscope used10 Zoom outReducing magnification, the field of view is increased11 Zoom inIncreasing magnification, the field of view is reduced12 SwitchFunction depending on surgical microscope used13 SwitchFunction depending on surgical microscope used14 SwitchFunction depending on surgical microscope used
Version 1.1G-30-1706-en Page 29FCP & FCP WLFoot Control Panel Standard assignment in the transverse configuration 16791281314 511 10243
Design and function FCP & FCP WL Foot Control PanelVersion 1.1Page 30 G-30-1706-en
Version 1.1G-30-1706-en Page 31FCP & FCP WLFoot Control PanelPreparations for useInserting the batteries for cordless operation..................................32Pairing with the suspension system.................................................34Connecting the connection cable.....................................................36Exchanging the joystick head ..........................................................37Bringing the foot control panel into its working position................38Adapter plate on the operating chair ..............................................39
Preparations for use FCP & FCP WL Foot Control PanelVersion 1.1Page 32 G-30-1706-enInserting the batteries for cordless operation• Turn the two locks (1) on the battery compartment from the 12 o'clock position to the 3 o'clock position.• Open the cover of the battery compartment (2).• Insert three C type batteries in the battery compartment, ensuring that the polarity is correct. The polarity is shown at the bottom of the battery com-partment (3).• Close the cover of the battery compartment (2).• Turn the two locks (1) on the battery compartment from the 3 o'clock po-sition back to the 12 o'clock position.NOTE The wrong type of batteries may damage the foot control panel or impair its function.• Only use C type 1.5 V alkaline-manganese batteries, also see the chapter "Technical data". • Do not use any rechargeable batteries. The operating duration with re-chargeable batteries has not been defined.
Version 1.1G-30-1706-en Page 33FCP & FCP WLFoot Control Panel Inserting the batteries for cordless operation123
Preparations for use FCP & FCP WL Foot Control PanelVersion 1.1Page 34 G-30-1706-enPairing with the suspension systemPairing means the fixed relative assignment of the suspension system and foot control panel.Pairing must be performed for cordless operation.• Put the foot control panel in a vertical position (1) in the immediate vicinity of the suspension system. • Start the pairing process on the suspension system, see user manual for the suspension system. Æ"Radio link" status indicator (2) flashes amber.• Press one of the buttons (3) on the foot control panel for about 5 seconds.Æ"Radio link" status indicator (2) on the foot control panel starts flashing.• Release button (3) on the foot control panel.Æ"Radio link" status indicator (2) continues flashing for a few seconds.Successful pairingÆA beep indicates that pairing has been successfully completed.ÆThe "radio link" status indicator (2) shines green for about 1 second.You can also check whether pairing was successful by simultaneously pressing two buttons on the foot control panel. "Radio link" status indi-cator (2) will be lit.Pairing failedÆThree beeps indicate that pairing has not been successfully completed.Æ"Radio link" status indicator (2) shines red for about 1 second.• Repeat the pairing process as described above.NOTE The foot control panel may be disabled.• If pairing is performed incorrectly or not at all, the foot control panel may be disabled, or activation of a control may trigger functions on a different suspension system not assigned to the foot control panel.
Version 1.1G-30-1706-en Page 35FCP & FCP WLFoot Control Panel Pairing with the suspension systemCAUTION Risk of confusion when using multiple foot control panels• The use of several foot control panels may lead to confusion and, as a re-sult, to malfunction.• After pairing, mark the associated suspension system with a number (stickers provided in the delivery package) and set indicator (4) on the foot control panel to the number assigned to the suspension system.• Make sure that the assigned identification is unique in the entire surgery area.• Check the function of the foot control panel on the basis of the checklist on page 42.123333334
Preparations for use FCP & FCP WL Foot Control PanelVersion 1.1Page 36 G-30-1706-enConnecting the connection cableThe connecting cable is required for corded operation of the foot control panel.Only use – cable 304970-8730-000 with a 3 m length or – cable 304970-8760-000 with a 6 m length• Insert 6-pin connector (1) in socket (3) on the foot control panel.• Insert 4-pin connector (2) in socket (4) on the suspension system.CAUTION Tripping hazard caused by loops of the connecting cable• Check the routing of the connecting cable.• Remove any loops and route the connecting cable so as to avoid any trip-ping hazards.13214
Version 1.1G-30-1706-en Page 37FCP & FCP WLFoot Control Panel Exchanging the joystick headExchanging the joystick headThe delivery package of the foot control panel includes an extension for the joystick and a joystick knob. You can use these items to suit your personal work style.• Remove joystick extension (1) by pulling it upward. • Push joystick knob (2) into position as far as it will go.12
Preparations for use FCP & FCP WL Foot Control PanelVersion 1.1Page 38 G-30-1706-enBringing the foot control panel into its working position• Bring the foot control panel into its working position (1). – Working position: 0° ... 10°– Rest position: 45 ° ... 90°In its rest position (2), the foot control panel is completely disabled. It shuts down to save energy. The energy-saving rest position extends the service life of the batteries in the cordless FCP version.120°10°90°45°
Version 1.1G-30-1706-en Page 39FCP & FCP WLFoot Control Panel Adapter plate on the operating chairAdapter plate on the operating chair To permit improved ergonomic use, Carl Zeiss offers an adapter plate for the operating chair. The installation will be performed by Zeiss service staff.1Operating chair 304731-9901-0002Adapter plate 000000-1117-104 12
Preparations for use FCP & FCP WL Foot Control PanelVersion 1.1Page 40 G-30-1706-en
Version 1.1G-30-1706-en Page 41FCP & FCP WLFoot Control PanelOperationChecklist for cordless foot control panel .........................................42Checklist for corded foot control panel ...........................................43
Operation FCP & FCP WL Foot Control PanelVersion 1.1Page 42 G-30-1706-enChecklist for cordless foot control panel• Always check the function of the foot control panel before surgery (with-out patient!) using the following checklist:Prerequisites 9The foot control panel is in its working position.9The batteries have been inserted with correct polarity.9The suspension system has been powered on.9The foot control panel has been paired with the suspension system.9The number in the indicator of the foot control panel is identical to that on the suspension system.9The freely programmable buttons have been configured.• Check all functions of the foot control panel.ÆThe suspension system and the surgical microscope must perform the de-fined functions.CAUTION Failure of functions!If a function fails, do not use the foot control panel for safety reasons!• If possible, correct the malfunction (see "What to do in the event of mal-functions“) or contact the Zeiss service department.CAUTION Unintended operation of the foot control panel!A vertically adjustable operating table may accidentally activate buttons on the foot control panel.• Therefore, make sure during installation and use that any unintended ac-tivation of controls is not possible.NOTE Use of drapesThe use of a drape on the foot control panel reduces its friction on the floor, thus impairing slip resistance.
Version 1.1G-30-1706-en Page 43FCP & FCP WLFoot Control Panel Checklist for corded foot control panelChecklist for corded foot control panel• Always check the function of the foot control panel before surgery (with-out patient!) using the following checklist:Prerequisites 9The suspension system has been powered on.9The connecting cable is undamaged and ready for operation.9The connecting cable has been connected to the suspension system.9The freely programmable buttons have been configured.• Check all functions of the foot control panel.ÆThe suspension system and the surgical microscope must perform the de-fined functions. CAUTION Failure of functions!If a function fails, do not use the foot control panel for safety reasons!• If possible, correct the malfunction (see "What to do in the event of mal-functions“) or contact the Zeiss service department.CAUTION Unintended operation of the foot control panel!A vertically adjustable operating table may accidentally activate buttons on the foot control panel.• Therefore, make sure during installation and use that any unintended ac-tivation of controls is not possible.NOTE Use of drapesThe use of a drape on the foot control panel reduces its friction on the floor, thus impairing slip resistance.
Operation FCP & FCP WL Foot Control PanelVersion 1.1Page 44 G-30-1706-en
Version 1.1G-30-1706-en Page 45FCP & FCP WLFoot Control PanelWhat to do in the event of malfunctionsWhat to do in an emergency ...........................................................46Failure of the cordless foot control panel...................................................46Troubleshooting in the cordless foot control panel .........................47For your safety ..........................................................................................47Malfunction of the foot control panel........................................................47Troubleshooting in the corded foot control panel ...........................49For your safety ..........................................................................................49Malfunction of the foot control panel........................................................49
What to do in the event of malfunctions FCP & FCP WL Foot Control PanelVersion 1.1Page 46 G-30-1706-enWhat to do in an emergencyFailure of the cordless foot control panelThe suspension system or the surgical microscope does not respond to the ac-tivation of any of the controls on the foot control panel. • Insert 6-pin connector (1) of the connecting cable in socket (3) on the foot control panel.• Insert 4-pin connector (2) of the connecting cable in socket (4) on the sus-pension system. ÆThe foot control panel now operates in the corded mode.13214
Version 1.1G-30-1706-en Page 47FCP & FCP WLFoot Control Panel Troubleshooting in the cordless foot control panelTroubleshooting in the cordless foot control panelFor your safetyThis device is a high-grade technological product. To ensure optimum per-formance and safe working order, we recommend having it checked by our service representative as part of regular scheduled maintenance.• If a failure occurs which you cannot correct with the aid of the chapter "What to do in the event of malfunctions“, attach a sign to the device stat-ing it is out of order and contact our service representative.Malfunction of the foot control panel The status indicators and acoustic signals help you diagnose malfunctions, see page 24 f.Problem Possible cause Remedy SeeTemporary failure of function Batteries are depleted Replace batteries page 32Use of rechargeable batteries Replace batteries page 32Failure of individual button functionsReconfigure button function - only possible if functions can be configured on suspension systemContact service dept.----Failure / malfunction of radio link in cordless foot control panelConnect cable page 36Position switch always de-tects rest positionConnect cable page 36Interference in radio link Connect cable page 36Weak radio signal Connect cable page 36No pairing with suspension systemPerform pairing with suspen-sion systempage 34
What to do in the event of malfunctions FCP & FCP WL Foot Control PanelVersion 1.1Page 48 G-30-1706-enUnintended activation of functionButton got jammed after acti-vationPut foot control panel in its rest positionReconfigure button function - only possible if functions can be configured on suspension systemContact service dept.page 15----Foot control panel sends in-correct activation signalPut foot control panel in its rest positionReconfigure button function - only possible if functions can be configured on suspension systempage 15--Wrong foot control panel Check whether the identifica-tions on the suspension system and in the indicator of the foot control panel are identical.--Perform pairing with the sus-pension systempage 34Problem Possible cause Remedy See
Version 1.1G-30-1706-en Page 49FCP & FCP WLFoot Control Panel Troubleshooting in the corded foot control panelTroubleshooting in the corded foot control panelFor your safetyThis device is a high-grade technological product. To ensure optimum per-formance and safe working order, we recommend having it checked by our service representative as part of regular scheduled maintenance.• If a failure occurs which you cannot correct with the aid of the chapter "What to do in the event of malfunctions“, attach a sign to the device stat-ing it is out of order and contact our service representative.Malfunction of the foot control panel The status indicators and acoustic signals help you diagnose malfunctions, see page 24 f.Problem Possible cause Remedy SeeFoot control panel is inopera-tiveConnecting cable not con-nected to suspension systemConnect the connecting cable page 36Temporary failure of function Failure of individual button functionsReconfigure button function - only possible if functions can be configured on suspension systemContact service dept.----
What to do in the event of malfunctions FCP & FCP WL Foot Control PanelVersion 1.1Page 50 G-30-1706-enUnintended activation of functionButton got jammed after acti-vationDisconnect cable from sus-pension system.Reconfigure button function - only possible if functions can be configured on suspension systemContact service dept.page 15page 36--Foot control panel sends in-correct activation signalDisconnect cable from sus-pension system.Reconfigure button function - only possible if functions can be configured on suspension systempage 36--Problem Possible cause Remedy See
Version 1.1G-30-1706-en Page 51FCP & FCP WLFoot Control PanelCare and maintenanceCare of the device............................................................................52Cleaning....................................................................................................52Disinfection ...............................................................................................52Environmental protection measures ................................................53Note on disposal .......................................................................................53
Care and maintenance FCP & FCP WL Foot Control PanelVersion 1.1Page 52 G-30-1706-enCare of the deviceCleaningContamination of the foot control panel– If possible, clean the foot control panel and its components directly after use.– Contaminations should not be allowed to dry on the objects, as this would make cleaning and disinfecting more difficult.Cleaning mechanical surfacesAll mechanical surfaces of the foot control panel and its components can be cleaned by wiping them with a damp cloth. Do not use any aggressive or abra-sive cleaning agents.Clean off any residue using a mixture of 50% ethyl alcohol and 50% distilled water plus a dash of household dish-washing liquid.DisinfectionIt may be necessary to disinfect the foot control panel to permit its use in the OR, for example. We recommend using MELISEPTOL disinfectant from B. Braun, Melsungen AG. MELISEPTOL is available from Carl Zeiss, and you can also obtain it locally in many countries from representatives of B. Braun, Melsungen AG.CAUTION Observe the manufacturer's instructions for use!• Use MELISEPTOL in accordance with the manufacturer's specifications and observe the instructions for use provided with the product.• Wear disposable plastic gloves to prevent skin contact with the disinfect-ant.
Version 1.1G-30-1706-en Page 53FCP & FCP WLFoot Control Panel Environmental protection measuresEnvironmental protection measuresNote on disposalUser information on the disposal of electrical and electronic devicesThis symbol means that the product must not be disposed of as normal do-mestic waste.The correct disposal of electrical or electronic devices helps to protect the en-vironment and to prevent potential hazards to the environment and/or human health which may occur as a result of improper handling of the devices con-cerned.For detailed information on the disposal of the product, please contact your local dealer or the device manufacturer or its legal successor. Please also note the manufacturer's current information on the Internet. In the event of resale of the product or its components, the seller is required to inform the buyer that the product must be disposed of in accordance with the applicable na-tional regulations currently in force.For customers in the European UnionPlease contact your dealer or supplier if you wish to dispose of electrical or electronic devices.Information on disposal in countries outside the European UnionThis symbol is only applicable in the European Union. For the disposal of elec-trical and electronic devices, please observe the relevant national legislation and other regulations applicable in your country.
Care and maintenance FCP & FCP WL Foot Control PanelVersion 1.1Page 54 G-30-1706-enDisposal of batteriesPlease observe the local directives relating to the disposal of batteries, or con-tact your local sales organization for further information.The batteries must no be disposed of as normal domestic waste. Use an ap-propriate facility for the disposal of batteries.WARNING Risk of explosion!• Never throw batteries into a fire.
Version 1.1G-30-1706-en Page 55FCP & FCP WLFoot Control PanelSystem dataTechnical data .................................................................................56Mechanical data ........................................................................................56Electrical data............................................................................................56Ordering data ..................................................................................58Approval verification .......................................................................59Ambient conditions .........................................................................60
System data FCP & FCP WL Foot Control PanelVersion 1.1Page 56 G-30-1706-enTechnical dataMechanical dataElectrical dataElectrical data, cordless operationComponent FeatureDimensions L x W x H approx. 410 x 243 x 128 mmWeight approx. 2.4 kgComponent FeaturePower supply 3 x C type 1.5 V alkaline-manganese batteriesCurrent consumption Max. 100 mAService life of batteries Approx. 2 to 6 months - depending on the oper-ating duration of the foot control panelElectrical standard Complying with IEC 60601-1; UL 60601-1; Degree of protection IPX8Cable connection Remote socket for connecting cableThe device has been designed for continuous operation.
Version 1.1G-30-1706-en Page 57FCP & FCP WLFoot Control Panel Technical dataElectrical data, corded operationComponent FeaturePower supply 15 VDC ± 10 % via the suspension systemCurrent consumption Max. 200 mAElectrical standard Complying with IEC 60601-1; UL 60601-1;Degree of protection IPX8Cable connection Remote socket for connecting cableThe device has been designed for continuous operation.
System data FCP & FCP WL Foot Control PanelVersion 1.1Page 58 G-30-1706-enOrdering dataFoot control panel with 14 functions - cordless and cordedDisinfection - MeliseptolDisposable glovesDescription Cat. No.Foot control panel with 14 functions, corded 304970-9100-000Foot control panel with 14 functions, cordless 304970-9200-000Joystick knob 304970-8807-000Joystick extension 304970-8808-000C type 1.5 V alkaline-manganese battery3 batteries required000000-0035-244Connecting cable, 3 m 304970-8730-000Connecting cable, 6 m 304970-8760-000Adapter plate for operating chair 000000-1117-104Description Cat. No.1 l MELISEPTOL in vario bottle 000000-0103-907MELISEPTOL HBV spray, 250 ml 000000-0103-910MELISEPTOL HBV cloths 000000-0103-911Description Cat. No.Size 1 (large) size 8-9  000000-0117-736Size 2 (medium) size 7-8 000000-0117-737
Version 1.1G-30-1706-en Page 59FCP & FCP WLFoot Control Panel Approval verificationApproval verificationDescription IdentificationARIB The foot control panel meets the requirements of ARIB (Association of Radio Industries and Busi-nesses, Japan)EMC requirements The foot control panel meets the EMC require-ments in accordance with IEC 60601-1-2.The foot control panel meets the RFI requirements of Class B.FCC The foot control panel meets the requirements of Part 15 of FCC (Federal Communications Commis-sion, USA)Radio equipment The foot control panel meets the requirements of EC Directive 1999/5/ECThe system is marked withMedical device The system meets the essential requirements stipu-lated in Annex I to Directive 93/42/EEC governing medical devices.
System data FCP & FCP WL Foot Control PanelVersion 1.1Page 60 G-30-1706-enAmbient conditionsFor operationFor transportation and storageFeature Admissible rangeTemperature + 10 °C ... + 40 °CRelative humidity 30 % ... 75 %Air pressure 700 hPa ... 1060 hPaFeature Admissible rangeTemperature - 40 °C ... + 70 °CRel. humidity(without condensa-tion)10 % ... 90 %Air pressure 500 hPa ... 1060 hPa
FCP & FCP WLFoot Control Panel IndexVersion 1.1G-30-1706-en Seite 61IndexAAcoustic signal  .........................................................................................25Approval verification .................................................................................59BBatteries, depleted ....................................................................................47Batteries, disposal .....................................................................................54Batteries, inserting ....................................................................................32Battery compartment ................................................................................32Battery type ..............................................................................................32Bracket .....................................................................................................22CChecklist for corded foot control panel .....................................................43Checklist for cordless foot control panel  ...................................................42Cleaning ...................................................................................................52Connecting cable ......................................................................................22Connecting cable, not connected  .............................................................49Controls ...................................................................................................22Current consumption ................................................................................56EElectrical standard  ....................................................................................56Electromagnetic compatibility ...................................................................10Electromagnetic interference  ....................................................................10EMC requirements of IEC 60601-1-2 ........................................................10FFailure of a function  ...........................................................................42, 43Failure of the cordless foot control panel ..................................................46Foot rest ...................................................................................................22IIndicator .............................................................................................22, 35Intended use  ..............................................................................................8
Index FCP & FCP WL Foot Control PanelVersion 1.1Seite 62 G-30-1706-enJJoystick ...............................................................................................22, 37LLabels on the foot control panel  ...............................................................16Limit class  .................................................................................................10Longitudinal configuration ..................................................................22, 26MMalfunction when using multiple foot control panels ................................35NNormal use  .................................................................................................8Notes on EMC  ..........................................................................................10PPairing ......................................................................................................34Power supply  ............................................................................................56Problem ..............................................................................................47, 49RRadio frequency radiation in acc. with FCC  ...............................................10Radio link ..................................................................................................47Radio signal  ..............................................................................................47Rest position .......................................................................................15, 38Rocker switch  ...........................................................................................22SSafety .........................................................................................................7Safety standards  .........................................................................................7Service life of batteries ..............................................................................56Status indicators  .................................................................................22, 24TTarget group ...............................................................................................8Technical data  ..........................................................................................56Transverse configuration .....................................................................22, 28Troubleshooting in the corded foot control panel  .....................................49Troubleshooting in the cordless foot control panel  ...................................47
FCP & FCP WLFoot Control Panel IndexVersion 1.1G-30-1706-en Seite 63UUse of multiple foot control panels ...........................................................35WWorking position ................................................................................38, 42
Index FCP & FCP WL Foot Control PanelVersion 1.1Seite 64 G-30-1706-en
Version 1.1G-30-1706-en Page 65FCP & FCP WLFoot Control Panel(Blank page, for your notes . . .)