Carl Zeiss Surgical 304970-8200 Bluetooth Module User Manual G 30 1706 bk

Carl Zeiss Surgical GmbH Bluetooth Module G 30 1706 bk

User Manual

Download: Carl Zeiss Surgical 304970-8200 Bluetooth Module User Manual G 30 1706 bk
Mirror Download [FCC.gov]Carl Zeiss Surgical 304970-8200 Bluetooth Module User Manual G 30 1706 bk
Document ID1065142
Application IDHL489JlgcdqEuVbQaOiOJQ==
Document DescriptionUser Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize173.17kB (2164680 bits)
Date Submitted2009-02-05 00:00:00
Date Available2009-02-05 00:00:00
Creation Date2009-01-15 10:28:33
Producing SoftwareAcrobat Distiller 8.1.0 (Windows)
Document Lastmod2009-01-19 11:26:38
Document TitleG_30_1706.bk
Document CreatorFrameMaker 5.5.6p145
Document Author: zobnd

Carl Zeiss
c`m=C=c`m=ti
cççí=`çåíêçä=m~åÉä
User Manual
G-30-1706-en
Version 1.1
2009-01-15
FCP & FCP WL Foot Control Panel
About this manual
Orientation aids
Applicable area
The user manual is part of the scope of delivery.
•
Carefully read it before using the device.
•
Keep it at the site of use of the foot control panel.
•
Store it for the entire service life of the foot control panel.
•
Pass it on to every subsequent owner or user of the foot control panel.
–
The chapter overview at the beginning of the user manual provides a summary of all subjects.
–
The contents of each chapter are specified in detail at the beginning of
each chapter.
–
A keyword index at the end of the manual facilitates the search for specific
terms.
This user manual applies to foot control panels with the following identification:
–
Rating label numbers: 304970-9100-000 304970-9200-000
Information on the manufacturer:
Carl Zeiss Surgical GmbH
A Carl Zeiss Meditec Company
73446 Oberkochen
Germany
Fax: + 49 (0) 7364 - 20 4823
Email: surgical@meditec.zeiss.com
Internet: www.meditec.zeiss.com
Subject to change in design and scope of delivery and as a result of ongoing
technical development. Printed in Germany.
© Carl Zeiss Surgical GmbH 2009
Page 2
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Chapter overview
Version 1.1
G-30-1706-en
Chapter:
Safety measures
Chapter:
Design and function
19
Chapter:
Preparations for use
31
Chapter:
Operation
41
Chapter:
What to do in the event of malfunctions
45
Chapter:
Care and maintenance
51
Chapter:
System data
55
Chapter:
Index
61
Page 3
FCP & FCP WL Foot Control Panel
Page 4
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Safety measures
We would like to provide you with information about safety aspects which
must be observed when handling this foot control panel. This chapter contains a summary of the most important information concerning matters relevant to instrument safety.
Key to the symbols in this user manual ............................................. 6
Hazard symbols ...........................................................................................6
Information symbols....................................................................................7
Directives and standards ................................................................... 7
Target group ..................................................................................... 8
Field of application ............................................................................ 8
Intended use ...............................................................................................8
Normal use..................................................................................................8
Typical misuse .............................................................................................8
Notes for the operator....................................................................... 9
Duties of the operator .................................................................................9
Requirements to be met by the user ..........................................................12
Warranty and liability.................................................................................13
Requirements for operation............................................................. 14
Before every use ........................................................................................14
During use.................................................................................................14
After every use ..........................................................................................15
Warning labels and notes ................................................................ 16
Version 1.1
G-30-1706-en
Page 5
Safety measures
FCP & FCP WL Foot Control Panel
Key to the symbols in this user manual
Hazard symbols
The following safety information has been incorporated in the user manual.
Please note this information and be particularly careful in these cases.
WARNING
Indicates a hazard which can lead to death or severe injury if it is not prevented.
CAUTION
Indicates a hazard which can cause slight and medium injury if it is not prevented.
NOTE
Page 6
Indicates a hazard which can cause damage to material if it is not prevented.
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Key to the symbols in this user manual
Information symbols
The following information symbols are used in this user manual:
–
Listing
9 Prerequisite for an action
•
Prompt for action
Æ Result of an action
Additional information and tips.
No warnings of hazards are provided.
Directives and standards
The foot control panel described in this manual has been designed and tested
in accordance with Carl Zeiss safety standards as well as German and international standards. This guarantees a high degree of safety.
The foot control panel has been designed in compliance with the requirements of:
–
EN (European standard)
–
IEC (International Electrotechnical Commission)
–
UL (Underwriters Laboratories)
In accordance with Directive 93/42/EEC for medical devices, the complete
quality management system of the company Carl Zeiss Surgical GmbH, 73446
Oberkochen, Germany, has been certified by DQS Deutsche Gesellschaft zur
Zertifizierung von Managementsystemen GmbH, a notified body, under registration number 250758 MP23.
Version 1.1
G-30-1706-en
Page 7
Safety measures
FCP & FCP WL Foot Control Panel
Target group
This user manual is intended for physicians, nurses and other medical staff
who prepare, operate or maintain the foot control panel after appropriate
training and in accordance with the instructions given in this manual. Installation and service work not described in this manual must only be performed by
specialists from Carl Zeiss.
Field of application
Intended use
The foot control panel is an accessory for all types of Carl Zeiss surgical microscopes. It enables operation of the motor-driven and controllable functions of
the microscope by foot control.
Normal use
This foot control panel permits either cordless or corded operation of up to
14 different functions of a surgical microscope, including its options, by foot
control.
Typical misuse
The foot control panel must only be operated in combination with devices
from Carl Zeiss.
Page 8
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Notes for the operator
Notes for the operator
The correct use of the foot control panel is absolutely vital for safe operation.
Therefore, please thoroughly familiarize yourself with the content of this user
manual before starting up the foot control panel. Please also observe the user
manuals of the other devices used. Further information is available from our
service department or from authorized representatives.
Duties of the operator
Notes on the
foot control panel
Version 1.1
G-30-1706-en
•
Make sure that the installation conditions and the use of the foot control
panel meet microsurgical requirements:
–
low vibration
–
clean environment
–
avoidance of extreme mechanical stress.
•
Observe the legal regulations for accident prevention and occupational
health and safety applicable in the country concerned.
•
Do not operate the components contained in the delivery package
–
in explosion-risk areas,
–
if inflammable anesthetics or volatile solvents such as alcohol, benzine
or similar chemicals are present at a distance of less than 25 cm.
•
Switch off the connected foot control panel and remove the batteries if
you notice any smoke, sparks or unusual noise. Do not use the foot control panel until it has been repaired by our service team.
•
Do not force cable connections. If the male and female parts do not readily connect, make sure that they are appropriate for one another. If any of
the connectors are damaged, have our service representative repair them.
•
Over long distances (e.g. removal, return for repair, etc.), the foot control
panel must always be transported in the original packaging or in special
return packaging. For details, please contact your dealer or the Carl Zeiss
service team.
•
Use the foot control panel only for the application described.
Page 9
Safety measures
Notes on EMC (Electromagnetic Compatibility)
FCP & FCP WL Foot Control Panel
•
Operate the foot control panel only with the components included in the
delivery package. If you wish to use other components, make sure that
Carl Zeiss or the manufacturer of the components has verified and confirmed that these components meet the respective safety standards and
can be used without risk.
•
If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the foot control panel stating it is out of order and contact our service representative.
•
The foot control panel must not be modified without the manufacturer's
approval.
•
The foot control panel is a high-grade technological product. To ensure
optimum performance and safe working order, we recommend having it
checked on a regular basis by our service representative or trained specialist staff.
The foot control panel complies with the EMC requirements of
IEC 60601-1-2. For operating the foot control panel, observe the EMC precautions specified below.
•
Only use options approved by Carl Zeiss for this foot control panel.
•
Do not use any portable or mobile RF communication equipment in the
vicinity of the foot control panel as this may lead to an impairment of the
foot control panel's function.
•
The foot control panel meets the RFI requirements of Class B. It cannot be
ruled out, however, that interferences may occur in RF receivers (e.g. TV
or radio sets) in the vicinity. If such interferences are noticed, please inform the Carl Zeiss Service Dept.
Information on exposure to radio frequency radiation in accordance
with FCC
The emitted output power of the component is far below the FCC limit values
for radio frequency exposure. Nevertheless, the component should be used in
such a way that potential contact with persons during normal operation is
kept to a minimum.
NOTE
Class
This Class B digital component complies with Canadian standard ICES-003.
Page 10
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
NOTE
Notes for the operator
Limit values and what to do in the event of interference with digital
equipment
This component has been tested and found to comply with the limits for a
Class B digital device in accordance with Part 15 of FCC regulations. These
limits have been stipulated to provide adequate protection against harmful
exposure when the component is operated in residential areas. This component generates and uses radio frequency energy and may emit such energy.
If not installed and used in accordance with the relevant instructions, it may
cause interference to radio communication. However, there is no guarantee
that interference will not occur in a particular installation. If this component
causes interference with radio and TV reception, which can be determined
by turning the component off and on, the user should try to correct the interference by one or several of the following measures:
NOTE
–
Relocate or reorient the receiving antenna.
–
Move the component further away from the receiver.
–
Plug the component into a different outlet so that it is not on the same
circuit as the receiver.
–
Consult an experienced radio/TV technician.
Part 15 of FCC regulations
This component complies with Part 15 of FCC regulations [and with the Industry Canada RSS-210 standard].
Operation of the component is subject to the following two conditions:
NOTE
–
the component must not cause any harmful interference and
–
the component must accept any interference received, including interference that may cause undesired operation.
Modification or conversion
Any modification or conversion not explicitly approved by Carl Zeiss will invalidate the FCC license for the operation of this component.
Version 1.1
G-30-1706-en
Page 11
Safety measures
FCP & FCP WL Foot Control Panel
Requirements to be met by the user
Page 12
•
The foot control panel must only be used by staff who have undergone
appropriate training and instruction. It is the duty of the customer or institution operating the component to provide training and instruction for the
relevant staff.
•
Please keep the user manual where it is easily accessible at all times for the
persons operating the system.
•
Modifications and repairs of the foot control panel or any components operated together with the foot control panel may only be performed by our
service representative or by other authorized persons.
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Notes for the operator
Warranty and liability
Warranty and liability depend on the applicable contractual stipulations.
NOTE
Loss of warranty
The manufacturer is not liable for any damage caused by unauthorized persons tampering with the foot control panel. Furthermore, this will forfeit any
rights to claim under warranty.
Version 1.1
G-30-1706-en
Page 13
Safety measures
FCP & FCP WL Foot Control Panel
Requirements for operation
Our service representative or an expert authorized by us will install the foot
control panel. Please ensure that the following requirements are met for further operation:
9 The connecting components have been properly connected. The screw
connections have been firmly tightened.
9 All cables and plugs are in perfect condition.
Before every use
•
Make sure that all ”Requirements for operation” are fulfilled.
•
Go through the checklist.
•
Reattach any covers or caps that were removed from the foot control panel. Close any existing openings with the corresponding caps.
•
Constantly monitor the foot control panel during use.
During use
Page 14
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Requirements for operation
After every use
•
Put the foot control panel in its rest position by attaching its bracket e.g.
to floor stand (1).
Æ The foot control panel is shut down in the rest position to save energy.
The energy-saving rest position extends the service life of the batteries in
the cordless version of the foot control panel.
Version 1.1
G-30-1706-en
Page 15
Safety measures
FCP & FCP WL Foot Control Panel
Warning labels and notes
CAUTION
Note the warning labels and notes!
•
1A
If you notice that any label on your foot control panel is missing or has
become illegible, please contact us or one of our authorized representatives. We will supply a replacement.
1 Type label
A - Label for the corded version of the foot control panel
B - Label for the cordless version of the foot control panel
IPX8
1B
IPX8
–
Manufacturer (company name)
–
Serial number (SN)
–
Product name
–
Order number (REF)
–
IP classification (degree of protection against the harmful ingress of
water and solid particles)
2 "Observe the user manual" label
Observe the user manual or further applicable documents.
3 "Observe disposal regulations" label
Electrical or electronic devices must not be disposed of as normal domestic
waste. For more information on the disposal of electrical and electronic
devices, please see the chapter "Maintenance and care".
4 "Year of manufacture" label
This label indicates the year of manufacturer of the system.
5 Radio equipment approval identification
(for FCP WL only)
Page 16
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Warning labels and notes
6 Ready for use status indicator
7 Cable connection status indicator
8 Radio link status indicator
9 Battery status indicator
1A
IPX8
1B
IPX8
Version 1.1
G-30-1706-en
Page 17
Safety measures
Page 18
FCP & FCP WL Foot Control Panel
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Design and function
Versions of the foot control panel ................................................... 21
Components of the foot control panel............................................. 22
Standard assignment in the longitudinal configuration ................... 26
Standard assignment in the transverse configuration...................... 28
Version 1.1
G-30-1706-en
Page 19
Design and function
Page 20
FCP & FCP WL Foot Control Panel
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Versions of the foot control panel
Versions of the foot control panel
1 Cordless foot control panel with 14 functions
2 Corded foot control panel with 14 functions
Version 1.1
G-30-1706-en
Page 21
Design and function
FCP & FCP WL Foot Control Panel
Components of the foot control panel
1 Connecting cable
Connects the foot control panel to the appropriate connector on a suspension system or wall-mounted control panel.
2 Bracket
Allows you to hang up the foot control panel on the floor stand or on the
wall when not in use.
3 Controls
Buttons operated with your foot.
4 Indicator
The adjustable indicator is required for the cordless version. It indicates the
assignment to a surgical microscope after pairing.
5 Joystick
Permits you to control e.g. an X-Y coupling or the motorized lateral and
front-to-back tilt of the surgical microscope.
6 Rocker switch
The two rocker switches are used to operate the ”Zoom” and ”Focus”
functions. For this, each of the rocker switches contains two switching elements, allowing you to control the functions by toe and heel movement
without the need to shift your foot.
You can optionally assign the ”Zoom” and ”Focus” functions to the
switching elements of the rocker switches. Switchover is performed automatically on the suspension system.
A Longitudinal configuration:
The left switching elements of the rocker switches are used to operate the
”Zoom” function. The right switching elements of the rocker switches are
used to operate the ”Focus” function. For more information, please see
page 26.
B Transverse configuration:
The switching elements of the front rocker switch are used to operate the
”Focus” function. The switching elements of the back rocker switch are
used to operate the ”Zoom” function. For more information, please see
page 28.
7 Foot rest
The bridge between the two rocker switches (6) serves as a support to
rest your foot on.
8 Status indicators
Indicate different statuses by shining/not shining.
Page 22
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Components of the foot control panel
Focus
Zoom
Focus
Zoom
Version 1.1
G-30-1706-en
Page 23
Design and function
FCP & FCP WL Foot Control Panel
Functions of the status indicators
–
During cordless operation, you have to activate the "ready for use", "radio
link" and "battery" status indicators by simultaneously pressing a button
each on the left and right side of the foot control panel. The status indicators are only active while the buttons are pressed.
–
During corded operation via the connecting cable, the "ready for use" and
"cable connection" status indicators are permanently lit.
1 Ready for use status indicator
–
Shining green: the foot control panel is ready for use, no error is
present.
–
Shining amber: the foot control panel detects a temporary error and
eliminates is automatically.
–
Shining red: the foot control panel is not ready for use, an error is
present.
2 Cable connection status indicator
–
Shining green: the connecting cable has been plugged in, power supply is present.
–
Not shining: the foot control panel operates in the cordless mode.
3 Radio link status indicator
–
Shining green: very good radio link.
–
Shining amber: reduced radio link, e.g. due to slight interference,
shielding or too large a distance between the suspension system and
foot control panel.
–
Shining red: no radio link, e.g. due to non-existent pairing with the
suspension system, see page 34, shielding or too large a distance between the suspension system and foot control panel.
–
Flashing amber: while pairing with the suspension system is in
progress.
4 Battery status indicator
–
Shining green: batteries are OK.
–
Shining amber: batteries will not last much longer.
•
–
Page 24
Obtain replacement batteries and keep them ready at hand.
Shining red: batteries are depleted and need to be replaced immediately.
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Components of the foot control panel
Meaning of the acoustic signals
Signal
Cause
Remedy
beep
Confirms a function
--
beep-beep
Warning
•
Put the foot control panel
in its rest position (vertical
position), see page 38.
•
Check the status indicators, see page 24.
beep-beep-beep Problem
Version 1.1
G-30-1706-en
see "Troubleshooting"
Page 25
Design and function
FCP & FCP WL Foot Control Panel
Standard assignment in the longitudinal configuration
The longitudinal configuration refers to the focus and zoom functions.
Joystick
The joystick permits control in the following four directions:
1 front,
2 back,
3 left,
4 right.
5 Switch
Function depending on surgical microscope used
6 Switch
Function depending on surgical microscope used
7 Focus down
Focusing on a lower plane, the working distance is increased
8 Focus up
Focusing on a higher plane, the working distance is reduced
9 Switch
Function depending on surgical microscope used
10 Switch
Function depending on surgical microscope used
11 Zoom in
Increasing magnification, the field of view is reduced
12 Zoom out
Reducing magnification, the field of view is increased
13 Switch
Function depending on surgical microscope used
14 Switch
Function depending on surgical microscope used
Page 26
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Standard assignment in the longitudinal configuration
Version 1.1
G-30-1706-en
14
13
12
11
10
Page 27
Design and function
FCP & FCP WL Foot Control Panel
Standard assignment in the transverse configuration
The transverse configuration refers to the focus and zoom functions.
Joystick
The joystick permits control in the following four directions:
1 front,
2 back,
3 left,
4 right.
5 Switch,
Function depending on surgical microscope used
6 Switch
Function depending on surgical microscope used
7 Focus down
Focusing on a lower plane, the working distance is increased
8 Focus up
Focusing on a higher plane, the working distance is reduced
9 Switch
Function depending on surgical microscope used
10 Zoom out
Reducing magnification, the field of view is increased
11 Zoom in
Increasing magnification, the field of view is reduced
12 Switch
Function depending on surgical microscope used
13 Switch
Function depending on surgical microscope used
14 Switch
Function depending on surgical microscope used
Page 28
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Standard assignment in the transverse configuration
14
13
12
11
Version 1.1
G-30-1706-en
10
Page 29
Design and function
Page 30
FCP & FCP WL Foot Control Panel
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Preparations for use
Inserting the batteries for cordless operation.................................. 32
Pairing with the suspension system................................................. 34
Connecting the connection cable..................................................... 36
Exchanging the joystick head .......................................................... 37
Bringing the foot control panel into its working position................ 38
Adapter plate on the operating chair .............................................. 39
Version 1.1
G-30-1706-en
Page 31
Preparations for use
FCP & FCP WL Foot Control Panel
Inserting the batteries for cordless operation
NOTE
Page 32
The wrong type of batteries may damage the foot control panel or
impair its function.
•
Only use C type 1.5 V alkaline-manganese batteries, also see the chapter
"Technical data".
•
Do not use any rechargeable batteries. The operating duration with rechargeable batteries has not been defined.
•
Turn the two locks (1) on the battery compartment from the 12 o'clock
position to the 3 o'clock position.
•
Open the cover of the battery compartment (2).
•
Insert three C type batteries in the battery compartment, ensuring that the
polarity is correct. The polarity is shown at the bottom of the battery compartment (3).
•
Close the cover of the battery compartment (2).
•
Turn the two locks (1) on the battery compartment from the 3 o'clock position back to the 12 o'clock position.
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Inserting the batteries for cordless operation
Version 1.1
G-30-1706-en
Page 33
Preparations for use
FCP & FCP WL Foot Control Panel
Pairing with the suspension system
NOTE
The foot control panel may be disabled.
•
If pairing is performed incorrectly or not at all, the foot control panel may
be disabled, or activation of a control may trigger functions on a different
suspension system not assigned to the foot control panel.
Pairing means the fixed relative assignment of the suspension system and foot
control panel.
Pairing must be performed for cordless operation.
•
Put the foot control panel in a vertical position (1) in the immediate vicinity
of the suspension system.
•
Start the pairing process on the suspension system, see user manual for
the suspension system.
Æ "Radio link" status indicator (2) flashes amber.
•
Press one of the buttons (3) on the foot control panel for about 5 seconds.
Æ "Radio link" status indicator (2) on the foot control panel starts flashing.
•
Release button (3) on the foot control panel.
Æ "Radio link" status indicator (2) continues flashing for a few seconds.
Successful pairing
Æ A beep indicates that pairing has been successfully completed.
Æ The "radio link" status indicator (2) shines green for about 1 second.
You can also check whether pairing was successful by simultaneously
pressing two buttons on the foot control panel. "Radio link" status indicator (2) will be lit.
Pairing failed
Æ Three beeps indicate that pairing has not been successfully completed.
Æ "Radio link" status indicator (2) shines red for about 1 second.
•
Page 34
Repeat the pairing process as described above.
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
CAUTION
Pairing with the suspension system
Risk of confusion when using multiple foot control panels
•
The use of several foot control panels may lead to confusion and, as a result, to malfunction.
•
After pairing, mark the associated suspension system with a number
(stickers provided in the delivery package) and set indicator (4) on the
foot control panel to the number assigned to the suspension system.
•
Make sure that the assigned identification is unique in the entire surgery
area.
•
Check the function of the foot control panel on the basis of the checklist
on page 42.
3 4
Version 1.1
G-30-1706-en
Page 35
Preparations for use
FCP & FCP WL Foot Control Panel
Connecting the connection cable
CAUTION
Tripping hazard caused by loops of the connecting cable
•
Check the routing of the connecting cable.
•
Remove any loops and route the connecting cable so as to avoid any tripping hazards.
The connecting cable is required for corded operation of the foot control
panel.
Only use
–
cable 304970-8730-000 with a 3 m length or
–
cable 304970-8760-000 with a 6 m length
•
Insert 6-pin connector (1) in socket (3) on the foot control panel.
•
Insert 4-pin connector (2) in socket (4) on the suspension system.
Page 36
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Exchanging the joystick head
Exchanging the joystick head
The delivery package of the foot control panel includes an extension for the
joystick and a joystick knob. You can use these items to suit your personal
work style.
•
Remove joystick extension (1) by pulling it upward.
•
Push joystick knob (2) into position as far as it will go.
Version 1.1
G-30-1706-en
Page 37
Preparations for use
FCP & FCP WL Foot Control Panel
Bringing the foot control panel into its working position
•
Bring the foot control panel into its working position (1).
–
Working position: 0° ... 10°
–
Rest position: 45 ° ... 90°
In its rest position (2), the foot control panel is completely disabled. It shuts
down to save energy. The energy-saving rest position extends the service life
of the batteries in the cordless FCP version.
90°
45°
10°
0°
Page 38
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Adapter plate on the operating chair
Adapter plate on the operating chair
To permit improved ergonomic use, Carl Zeiss offers an adapter plate for the
operating chair. The installation will be performed by Zeiss service staff.
1 Operating chair 304731-9901-000
2 Adapter plate 000000-1117-104
Version 1.1
G-30-1706-en
Page 39
Preparations for use
Page 40
FCP & FCP WL Foot Control Panel
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Operation
Checklist for cordless foot control panel ......................................... 42
Checklist for corded foot control panel ........................................... 43
Version 1.1
G-30-1706-en
Page 41
Operation
FCP & FCP WL Foot Control Panel
Checklist for cordless foot control panel
•
CAUTION
Always check the function of the foot control panel before surgery (without patient!) using the following checklist:
Failure of functions!
If a function fails, do not use the foot control panel for safety reasons!
•
CAUTION
If possible, correct the malfunction (see "What to do in the event of malfunctions“) or contact the Zeiss service department.
Unintended operation of the foot control panel!
A vertically adjustable operating table may accidentally activate buttons on
the foot control panel.
•
NOTE
Therefore, make sure during installation and use that any unintended activation of controls is not possible.
Use of drapes
The use of a drape on the foot control panel reduces its friction on the floor,
thus impairing slip resistance.
Prerequisites
9 The foot control panel is in its working position.
9 The batteries have been inserted with correct polarity.
9 The suspension system has been powered on.
9 The foot control panel has been paired with the suspension system.
9 The number in the indicator of the foot control panel is identical to that
on the suspension system.
9 The freely programmable buttons have been configured.
•
Check all functions of the foot control panel.
Æ The suspension system and the surgical microscope must perform the defined functions.
Page 42
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Checklist for corded foot control panel
Checklist for corded foot control panel
•
CAUTION
Always check the function of the foot control panel before surgery (without patient!) using the following checklist:
Failure of functions!
If a function fails, do not use the foot control panel for safety reasons!
•
CAUTION
If possible, correct the malfunction (see "What to do in the event of malfunctions“) or contact the Zeiss service department.
Unintended operation of the foot control panel!
A vertically adjustable operating table may accidentally activate buttons on
the foot control panel.
•
NOTE
Therefore, make sure during installation and use that any unintended activation of controls is not possible.
Use of drapes
The use of a drape on the foot control panel reduces its friction on the floor,
thus impairing slip resistance.
Prerequisites
9 The suspension system has been powered on.
9 The connecting cable is undamaged and ready for operation.
9 The connecting cable has been connected to the suspension system.
9 The freely programmable buttons have been configured.
•
Check all functions of the foot control panel.
Æ The suspension system and the surgical microscope must perform the defined functions.
Version 1.1
G-30-1706-en
Page 43
Operation
Page 44
FCP & FCP WL Foot Control Panel
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
What to do in the event of malfunctions
What to do in an emergency ........................................................... 46
Failure of the cordless foot control panel ...................................................46
Troubleshooting in the cordless foot control panel ......................... 47
For your safety ..........................................................................................47
Malfunction of the foot control panel........................................................47
Troubleshooting in the corded foot control panel ........................... 49
For your safety ..........................................................................................49
Malfunction of the foot control panel........................................................49
Version 1.1
G-30-1706-en
Page 45
What to do in the event of malfunctions
FCP & FCP WL Foot Control Panel
What to do in an emergency
Failure of the cordless foot control panel
The suspension system or the surgical microscope does not respond to the activation of any of the controls on the foot control panel.
•
Insert 6-pin connector (1) of the connecting cable in socket (3) on the foot
control panel.
•
Insert 4-pin connector (2) of the connecting cable in socket (4) on the suspension system.
Æ The foot control panel now operates in the corded mode.
Page 46
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Troubleshooting in the cordless foot control panel
Troubleshooting in the cordless foot control panel
For your safety
This device is a high-grade technological product. To ensure optimum performance and safe working order, we recommend having it checked by our
service representative as part of regular scheduled maintenance.
•
If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the device stating it is out of order and contact our service representative.
Malfunction of the foot control panel
The status indicators and acoustic signals help you diagnose malfunctions, see
page 24 f.
Problem
Possible cause
Remedy
See
Replace batteries
page 32
Use of rechargeable batteries Replace batteries
page 32
Temporary failure of function Batteries are depleted
Failure of individual button
functions
Version 1.1
G-30-1706-en
Reconfigure button function - -only possible if functions can
be configured on suspension
system
Contact service dept.
--
Failure / malfunction of radio
link in cordless foot control
panel
Connect cable
page 36
Position switch always detects rest position
Connect cable
page 36
Interference in radio link
Connect cable
page 36
Weak radio signal
Connect cable
page 36
No pairing with suspension
system
Perform pairing with suspension system
page 34
Page 47
What to do in the event of malfunctions
FCP & FCP WL Foot Control Panel
Problem
Possible cause
Remedy
See
Unintended activation of
function
Button got jammed after acti- Put foot control panel in its
vation
rest position
page 15
Reconfigure button function - -only possible if functions can
be configured on suspension
system
-Contact service dept.
Foot control panel sends incorrect activation signal
Put foot control panel in its
rest position
page 15
Reconfigure button function - -only possible if functions can
be configured on suspension
system
Wrong foot control panel
Check whether the identifica- -tions on the suspension
system and in the indicator of
the foot control panel are
identical.
Perform pairing with the sus- page 34
pension system
Page 48
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Troubleshooting in the corded foot control panel
Troubleshooting in the corded foot control panel
For your safety
This device is a high-grade technological product. To ensure optimum performance and safe working order, we recommend having it checked by our
service representative as part of regular scheduled maintenance.
•
If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions“, attach a sign to the device stating it is out of order and contact our service representative.
Malfunction of the foot control panel
The status indicators and acoustic signals help you diagnose malfunctions, see
page 24 f.
Problem
Possible cause
Remedy
See
Foot control panel is inopera- Connecting cable not contive
nected to suspension system
Connect the connecting cable page 36
Temporary failure of function Failure of individual button
functions
Reconfigure button function - -only possible if functions can
be configured on suspension
system
Contact service dept.
Version 1.1
G-30-1706-en
--
Page 49
What to do in the event of malfunctions
FCP & FCP WL Foot Control Panel
Problem
Possible cause
Remedy
See
Unintended activation of
function
Button got jammed after acti- Disconnect cable from susvation
pension system.
page 15
Reconfigure button function - page 36
only possible if functions can
be configured on suspension
system
Foot control panel sends incorrect activation signal
Contact service dept.
--
Disconnect cable from suspension system.
page 36
Reconfigure button function - -only possible if functions can
be configured on suspension
system
Page 50
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Care and maintenance
Care of the device............................................................................ 52
Cleaning ....................................................................................................52
Disinfection ...............................................................................................52
Environmental protection measures ................................................ 53
Note on disposal .......................................................................................53
Version 1.1
G-30-1706-en
Page 51
Care and maintenance
FCP & FCP WL Foot Control Panel
Care of the device
Cleaning
Contamination of the foot control panel
–
If possible, clean the foot control panel and its components directly after
use.
–
Contaminations should not be allowed to dry on the objects, as this would
make cleaning and disinfecting more difficult.
Cleaning mechanical surfaces
All mechanical surfaces of the foot control panel and its components can be
cleaned by wiping them with a damp cloth. Do not use any aggressive or abrasive cleaning agents.
Clean off any residue using a mixture of 50% ethyl alcohol and 50% distilled
water plus a dash of household dish-washing liquid.
Disinfection
CAUTION
Observe the manufacturer's instructions for use!
•
Use MELISEPTOL in accordance with the manufacturer's specifications
and observe the instructions for use provided with the product.
•
Wear disposable plastic gloves to prevent skin contact with the disinfectant.
It may be necessary to disinfect the foot control panel to permit its use in the
OR, for example.
We recommend using MELISEPTOL disinfectant from B. Braun, Melsungen
AG.
MELISEPTOL is available from Carl Zeiss, and you can also obtain it locally in
many countries from representatives of B. Braun, Melsungen AG.
Page 52
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Environmental protection measures
Environmental protection measures
Note on disposal
User information
on the disposal of electrical and electronic devices
This symbol means that the product must not be disposed of as normal domestic waste.
The correct disposal of electrical or electronic devices helps to protect the environment and to prevent potential hazards to the environment and/or human
health which may occur as a result of improper handling of the devices concerned.
For detailed information on the disposal of the product, please contact your
local dealer or the device manufacturer or its legal successor. Please also note
the manufacturer's current information on the Internet. In the event of resale
of the product or its components, the seller is required to inform the buyer
that the product must be disposed of in accordance with the applicable national regulations currently in force.
For customers in the European Union
Please contact your dealer or supplier if you wish to dispose of electrical or
electronic devices.
Information on disposal in countries outside the European Union
This symbol is only applicable in the European Union. For the disposal of electrical and electronic devices, please observe the relevant national legislation
and other regulations applicable in your country.
Version 1.1
G-30-1706-en
Page 53
Care and maintenance
FCP & FCP WL Foot Control Panel
Disposal of batteries
WARNING
Risk of explosion!
•
Never throw batteries into a fire.
Please observe the local directives relating to the disposal of batteries, or contact your local sales organization for further information.
The batteries must no be disposed of as normal domestic waste. Use an appropriate facility for the disposal of batteries.
Page 54
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
System data
Technical data ................................................................................. 56
Mechanical data ........................................................................................56
Electrical data ............................................................................................56
Ordering data .................................................................................. 58
Approval verification ....................................................................... 59
Ambient conditions ......................................................................... 60
Version 1.1
G-30-1706-en
Page 55
System data
FCP & FCP WL Foot Control Panel
Technical data
Mechanical data
Component
Feature
Dimensions L x W x H
approx. 410 x 243 x 128 mm
Weight
approx. 2.4 kg
Electrical data
Electrical data, cordless operation
Component
Feature
Power supply
3 x C type 1.5 V alkaline-manganese batteries
Current consumption
Max. 100 mA
Service life of batteries Approx. 2 to 6 months - depending on the operating duration of the foot control panel
Electrical standard
Complying with IEC 60601-1; UL 60601-1;
Degree of protection IPX8
Cable connection
Remote socket for connecting cable
The device has been designed for continuous operation.
Page 56
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Technical data
Electrical data, corded operation
Component
Feature
Power supply
15 VDC ± 10 % via the suspension system
Current consumption
Max. 200 mA
Electrical standard
Complying with IEC 60601-1; UL 60601-1;
Degree of protection IPX8
Cable connection
Remote socket for connecting cable
The device has been designed for continuous operation.
Version 1.1
G-30-1706-en
Page 57
System data
FCP & FCP WL Foot Control Panel
Ordering data
Foot control panel with 14 functions - cordless and corded
Description
Cat. No.
Foot control panel with 14 functions, corded
304970-9100-000
Foot control panel with 14 functions, cordless
304970-9200-000
Joystick knob
304970-8807-000
Joystick extension
304970-8808-000
C type 1.5 V alkaline-manganese battery
000000-0035-244
3 batteries required
Connecting cable, 3 m
304970-8730-000
Connecting cable, 6 m
304970-8760-000
Adapter plate for operating chair
000000-1117-104
Disinfection - Meliseptol
Description
Cat. No.
1 l MELISEPTOL in vario bottle
000000-0103-907
MELISEPTOL HBV spray, 250 ml
000000-0103-910
MELISEPTOL HBV cloths
000000-0103-911
Disposable gloves
Page 58
Description
Cat. No.
Size 1 (large) size 8-9
000000-0117-736
Size 2 (medium) size 7-8
000000-0117-737
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Approval verification
Approval verification
Description
Identification
ARIB
The foot control panel meets the requirements of
ARIB (Association of Radio Industries and Businesses, Japan)
EMC requirements
The foot control panel meets the EMC requirements in accordance with IEC 60601-1-2.
The foot control panel meets the RFI requirements
of Class B.
FCC
The foot control panel meets the requirements of
Part 15 of FCC (Federal Communications Commission, USA)
Radio equipment
The foot control panel meets the requirements of
EC Directive 1999/5/EC
The system is marked with
Medical device
Version 1.1
G-30-1706-en
The system meets the essential requirements stipulated in Annex I to Directive 93/42/EEC governing
medical devices.
Page 59
System data
FCP & FCP WL Foot Control Panel
Ambient conditions
For operation
Feature
Admissible range
Temperature
+ 10 °C ... + 40 °C
Relative humidity
30 % ... 75 %
Air pressure
700 hPa ... 1060 hPa
For transportation and storage
Page 60
Feature
Admissible range
Temperature
- 40 °C ... + 70 °C
Rel. humidity
(without condensation)
10 % ... 90 %
Air pressure
500 hPa ... 1060 hPa
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Index
Index
Acoustic signal .........................................................................................25
Approval verification .................................................................................59
Batteries, depleted .................................................................................... 47
Batteries, disposal .....................................................................................54
Batteries, inserting .................................................................................... 32
Battery compartment ................................................................................32
Battery type ..............................................................................................32
Bracket .....................................................................................................22
Checklist for corded foot control panel .....................................................43
Checklist for cordless foot control panel ...................................................42
Cleaning ...................................................................................................52
Connecting cable ......................................................................................22
Connecting cable, not connected .............................................................49
Controls ...................................................................................................22
Current consumption ................................................................................56
Electrical standard .................................................................................... 56
Electromagnetic compatibility ...................................................................10
Electromagnetic interference .................................................................... 10
EMC requirements of IEC 60601-1-2 ........................................................10
Failure of a function ...........................................................................42, 43
Failure of the cordless foot control panel ..................................................46
Foot rest ...................................................................................................22
Indicator .............................................................................................22, 35
Intended use ..............................................................................................8
Version 1.1
G-30-1706-en
Seite 61
Index
FCP & FCP WL Foot Control Panel
Joystick ...............................................................................................22, 37
Labels on the foot control panel ...............................................................16
Limit class .................................................................................................10
Longitudinal configuration ..................................................................22, 26
Malfunction when using multiple foot control panels ................................35
Normal use .................................................................................................8
Notes on EMC ..........................................................................................10
Pairing ......................................................................................................34
Power supply ............................................................................................56
Problem ..............................................................................................47, 49
Radio frequency radiation in acc. with FCC ...............................................10
Radio link ..................................................................................................47
Radio signal ..............................................................................................47
Rest position .......................................................................................15, 38
Rocker switch ...........................................................................................22
Safety .........................................................................................................7
Safety standards .........................................................................................7
Service life of batteries ..............................................................................56
Status indicators .................................................................................22, 24
Target group ...............................................................................................8
Technical data ..........................................................................................56
Transverse configuration .....................................................................22, 28
Troubleshooting in the corded foot control panel .....................................49
Troubleshooting in the cordless foot control panel ...................................47
Seite 62
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
Index
Use of multiple foot control panels ...........................................................35
Working position ................................................................................38, 42
Version 1.1
G-30-1706-en
Seite 63
Index
Seite 64
FCP & FCP WL Foot Control Panel
Version 1.1
G-30-1706-en
FCP & FCP WLFoot Control Panel
(Blank page, for your notes . . .)
Version 1.1
G-30-1706-en
Page 65
Carl Zeiss Surgical GmbH
A Carl Zeiss Meditec Company
73446 Oberkochen
Germany
Fax.: + 49 (0)7364 - 20 4823
Email: surgical@meditec.zeiss.com
Internet: www.meditec.zeiss.com

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Page Mode                       : UseOutlines
XMP Toolkit                     : Adobe XMP Core 4.0-c316 44.253921, Sun Oct 01 2006 17:14:39
Creator Tool                    : FrameMaker 5.5.6p145
Modify Date                     : 2009:01:19 11:26:38+01:00
Create Date                     : 2009:01:15 10:28:33+01:00
Metadata Date                   : 2009:01:19 11:26:38+01:00
Format                          : application/pdf
Title                           : G_30_1706.bk
Creator                         : zobnd
Producer                        : Acrobat Distiller 8.1.0 (Windows)
Document ID                     : uuid:2fe95985-d796-42fd-ba9c-5a8dd25b60b8
Instance ID                     : uuid:c6ae6a8b-01e1-45f9-9266-b77d7a929f9a
Page Count                      : 66
Author                          : zobnd
EXIF Metadata provided by EXIF.tools
FCC ID Filing: WCJ304970-8200

Navigation menu