Creative Industry PC-60NW Fingertip Oximeter User Manual

Shenzhen Creative Industry Co., Ltd. Fingertip Oximeter Users Manual

Users Manual

User Manual for Fingertip Oximeter I   Instructions to User Dear Users, Thank you very much for purchasing our product. Please read the manual very carefully before using this device. Failure to follow these instructions can cause measuring abnormality or damage to the oximeter. The manual is published in English and we have the ultimate right to explain the Manual. No part of this manual may be photocopied, reproduced or translated into another language without the prior written consent. We reserve the right to improve and amend it at any time without prior notice. This device complies with Part 15 of the FCC Rules.  Operation is subject to the following two conditions: (1)   this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Version of the Manual: Ver  1 . 0  Revised date: November 1, 2011 All rights reserved.                  3502-2290047
User Manual for Fingertip Oximeter II   Warnings 0 Check the device to make sure that there is no visible damage that may affect user’s safety or measurement performance with regard to sensors and clips. It is recommended that the device should be inspected minimally once a week. When there is obvious damage, stop using the device. 0 Special attention should be paid while the oximeter is used continuously under the ambient temperature exceeds 37°, burning hurt may occur because of over-heating on the sensor. 0 An uncomfortable or painful feeling may appear   if using the oximeter continuously on the same place for a long time, especially for poor microcirculation patients.   0 It is recommended that the oximeter should not be applied to the same location for longer than 2 hours. If any abnormal condition is found, please change the position of oximeter.   0 Avoid placing the device on the same limb which is wrapped with a cuff for blood pressure measurement or during venous infusion.
User Manual for Fingertip Oximeter III   0 DO NOT clip this device on edema or tender tissue.  0 The light (the infrared light is invisible) emitted   from the device is harmful to the eyes, so service technician or testee should not stare at the light. 0 The oximeter is not a treatment device. 0 When disposing of the monitor and its    accessories, the local law should be followed. Instructions for Operation y The finger should be put in properly and correctly. y  Do not shake the finger. Keep at ease during measurement. y   Do not put wet finger directly into sensor. y   Avoid placing the device on the same limb which is wrapped with a cuff for blood pressure measurement or during venous infusion. y   Do not let anything block the emitting light from device. y  Vigorous exercise and electrosurgical device interference may affect the measuring accuracy.   y   The orientation-sensor works on the basis of the gravity. A small movable metal ball is built in the orientation-sensor for detecting the orientation of
User Manual for Fingertip Oximeter IV   the oximeter. When you want to change the oximeter’s display direction, if you move the oximeter too slowly, the movable metal ball will also move slowly because of not enough acceleration. Consequently the response of orientation detection would be delayed. Please move the oximeter with a bit of force if you want to change the display direction (such as bend/extend your finger quickly), so an acceleration is provided to the orientation-sensor for quick sensing the orientation change.   y  Using enamel or other makeup on the nail may affect the accuracy of measurement, too long fingernail may casue failure of measurement or unaccuracy measurement result. y Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture. y Existence of high intensive light sources, such as fluorescence light, ruby lamb, infrared heater or strong sunshire, etc. may cause unaccuracy of measurement result. Please put an opaque cover on the sensor or change the measuring site. y If the first reading appears with poor waveform (irregular or not smooth), then the reading is
User Manual for Fingertip Oximeter V   unlikely true, the more stable value is expected by waiting for a while, or a restart is needed when necessary. Note: Due to the working principle of orientation sensor used in oximeter, there is a small metal ball which is movable within its compartment of the orientation-sensor. Therefore you can hear a slight “clatter” sound when you wave or shake the oximeter. It is normal and not caused by unwanted part.
User Manual for Fingertip Oximeter VI   Content 1 Overview ....................................................................... 1 1.1 Appearance............................................................. 1 1.2 Name and Model.................................................... 2 1.3 Intended Use........................................................... 2 1.4 Structure and Conformation................................ 2 1.5 Features .................................................................. 2 2 Battery Installation...................................................... 3 3 Operation ...................................................................... 4 3.1 Directly measurement ........................................... 4 3.2 Alarm and alarm silence....................................... 7 3.3 Menu Screen........................................................... 8 3.4 External SpO2 Probe Connection ...................... 10 3.5 Data transmission .................................................11 4. Technical Specifications............................................ 12 5. Accessories ................................................................. 15 6. Repair and Maintenance.......................................... 15
User Manual for Fingertip Oximeter VII   6.1 Oximeter Maintenance........................................ 15 6.2 Battery Maintenance........................................... 17 6.3 Cleaning and Disinfecting Instruction............... 17 7. Troubleshooting......................................................... 18 Appendix........................................................................ 20 A Key of Symbols........................................................ 20 B SpO2 Common Knowledge...................................... 21
User Manual for Fingertip Oximeter 1   1 Overview 1.1 Appearance  Figure 1 Front/Rear View
User Manual for Fingertip Oximeter 2   1.2 Name and Model Name: Fingertip Oximeter Model: PC-60NW 1.3 Intended Use This Fingertip Oximeter is intended for measuring the pulse rate and functional oxygen saturation (SpO2) through patient’s finger. It is applicable for spot-checking SpO2 and pulse rate of adult and pediatric patients in homes and clinics.   1.4 Structure and Conformation It consists of main unit and photoelectric sensor, and additional data upload connector.   1.5 Features •  Wireless data transmission can communicate with PC/mobile phone/PDA. •  External pediatric SPO2 probe available •  Large true color OLED display of SpO2, PR Pulse Bar, PI & Plethysmogram •  Automatic change display direction •  Automatic power on/off
User Manual for Fingertip Oximeter 3   •  Audible & visible alarm function   •  Pulse beep with pitch tone, pulse beep on/off and alarm limits can be set via setup menu. •  Shift parameter display between PR and PI •  2AAA alkaline batteries with low power consumption •  Low battery voltage indication 2 Battery Installation  Figure 2 Battery Installation 1. Refer to Figure2, insert two AAA size batteries into the battery compartment properly.   2. Replace the cover.   Please make sure that the batteries are correctly installed, or incorrect installation may cause the device not to work.
User Manual for Fingertip Oximeter 4   3 Operation   3.1 Directly measurement   1. Open the clip as shown in Figure 3.    Figure 3 Put finger into the Oximeter 2. Put finger into the rubber cushions of the clip (make sure the finger is in the correct position), and then clip the finger. 3. The device will power on automatically in 2 seconds, and start to display software version number. 4. Next enter into data display screen (as shown in Figure 4). User can read the values and view the waveform from display screen.     Figure4A           Figure4B
User Manual for Fingertip Oximeter 5   Screen Description:  “%SpO2”: SpO2 symbol; “99”: SpO2 value;  “PR”: Pulse rate icon; “65”: Pulse rate value;  “”: Pulse beat symbol;  “”: Pulse intensity histogram.  “PI%”:  Perfusion index icon; “1.4”: Perfusion index value;      “ ”:Wireless symbol   ) Change display direction: Tilt the oximeter to change display direction. It is better for user to read value conveniently. ) Shift parameter display between PR and PI during measurement: Short time press Display Key to shift the 4A and 4B. When shown as 4B, the display will shift 4A automatically after 20 seconds without operation.
User Manual for Fingertip Oximeter 6   )  Wireless icon“ ”: The color of “ ”   Definition "Wireless" function is disabled “”displays gray The device fails to setup a wireless connection with the surrounding host. “ ” flashes blue The device is being to establish a wireless connection with the surrounding host. “ ” long lights blue Successful wireless connection between the device and a host is established. No display“ ”icon  Hardware failure of wireless transmission function.
User Manual for Fingertip Oximeter 7   When the device fails to try establishing wireless connection within 3 minutes, the icon “ ” will become gray and the "Wireless" function is disabled automatically. You have to enable it next time manually. Notes: Notes: The pulse beep has the pitch-tone feature (when SpO2 value is higher than 90%), that means, the beeping tone changes according to the SpO2 value. 3.2 Alarm and alarm silence When measuring, if SpO2 value or pulse rate value exceeds the preset alarm limit, the device will alarm automatically and the value which exceeds limit on the screen will flash. The detailed information refers to chapter 4. Apply for belowed methods to relief alarm sound when alarm event happens: 1. When SpO2 value and PR value get normally. 2. Press Display Key to mute. If this alarm event continues, the oximeter will resume alarm sound automatically 2 minutes later. 3. Remove the finger from the oximeter or SpO2 probe.
User Manual for Fingertip Oximeter 8   3.3 Menu Screen  Figure 5 Menu Screen Longtime press display key could enter the menu screen. Menu screen description: “Wireless”:the wireless on-off button. Transmitting data to PC when it is on. “on” and “off” can be optional. The factory default is “on”.   “SpO2 alm Lo”:SpO2 alarm: Lower limit.The user can modify the value of 85~99, the step is “1”, the default is 90. “PR alm Hi”:Pulse Rate alarm upper limit. The user can modify the value of 100~240, the step is “5”, the default is 120.
User Manual for Fingertip Oximeter 9   “PR alm Lo”:Pulse Rate alarm lower limit. The user can modify the value of 30~60, the step is “1”, the default is 50. “Pulse beep”:Pulse beep button. When SpO2  value (90~99) changes, the pitch tone changes accordingly.  “Save, exit menu”:long time pressing this item to store and exit from the setup menu then enter the display screen. “Restore default”:Restore default setting. Refer to Figure 5 for each default value. On setup menu screen: 1. Short time press Display Key to choose the setting item; 2. Longtime press Display Key to active the setting item, then short time press it to modify the setting parameter; 3. Next, longtime press Display Key to confirm the modification and exit from this setting item.   4. At last, move the setting item to “Save, exit menu”, and long time pressing Display Key to store the modification and exit from the setup menu.
User Manual for Fingertip Oximeter 10   3.4 External SpO2 Probe Connection 1. Connect the external SpO2 probe to SpO2 sensor connector in the following way. Make sure the side with “Arrow” faces upwards.  Figure 6 Probe Connection Note: when the external SpO2 probe is connected well, the built-in finger clip sensor will be disabled. The measurement is detected from the external SpO2 probe.    2. The finger should be put in SpO2 probe properly and correctly. 3. The oximeter will power on automatically 2 seconds later, then display software version number. 4. Other operation is similar to chapter 3.1 directly measurement.
User Manual for Fingertip Oximeter 11   3.5 Data transmission This oximeter has the function of wireless data transmission. The user could effectively transmit the data to computer through the wireless communication module. Refer to the 《Oximeter Data Manager》for detailed information.
User Manual for Fingertip Oximeter 12   4. Technical Specifications A.  Technique: dual-wavelength LED sensor, LED sensor wavelength: Red light: 663 nanometers, Infrared light: 890 nanometers. Maximal optical output power: less than 1.5mW maximum average. B.  SpO2 measurement Measuring range: 70%~100% Measuring accuracy:  not greater than 3% for SpO2 range from 70% to 100% Note: Accuracy defined as root-mean-square value of deviation according to ISO 9919.     SpO2 alarm low limit range:  85%~99% (default 90%) C.   Pulse Rate measurement Measuring range: 30bpm~240bpm Measuring accuracy: ±2bpm or ±2% (whichever is greater)  Pulse Rate alarm range:   high limit: 100~240bpm (default 120bpm)
User Manual for Fingertip Oximeter 13   low limit: 30~60bpm (default 50bpm) D.     Perfusion Index(PI) Display Range: 0.2%~20%   E.    Audible &visual alarm function     When measuring, if SpO2 value or pulse rate value exceeds the preset alarm limit, the device will alarm automatically and the value which exceeds limit on the screen will flash. The oximeter will shut down automatically in 8 seconds with no signal. F.  Display mode:Color OLED Display G.  Power supply requirement:  2 x LR03 (AAA) alkaline batteries   Supply voltage: 3.0VDC  Operating current: ≤40mA  H.  Environment requirement    Operating Temperature:   5 ~40°C Operating Humidity:      30~80% Atmospheric pressure:    70~106kPa I.   The performance under low perfusion condition The accuracy of SpO2 and PR measurement still meet the precision described above when the modulation amplitude is as low as 0.6%.
User Manual for Fingertip Oximeter 14   J.   Resistance to interference of surrounding light: The difference between the SpO2 value measured in the condition of indoor natural light and that of darkroom is less than ±1%. K.  Dimensions: 60 mm (L) × 33 mm (W) × 30 mm (H) Net Weight: 35g (including battery) L.  Classification: The type of protection against electric shock: Internally powered equipment. The degree of protection against electric shock: Type BF applied parts. The degree of protection against harmful ingress of liquids: Ordinary equipment without protection against ingress of water. Electro-Magnetic Compatibility:GroupⅡ, Class B
User Manual for Fingertip Oximeter 15   5. Accessories A.  A lanyard B.  Two batteries C.  A pouch D.    An External SpO2 Probe (optional) E.    A User Manual F.  Quality Certificate G.    Installation CD (optional) Note: The accessories are subject to change. Detailed items and quantity see the Packing List. 6. Repair and Maintenance 6.1 Oximeter Maintenance The service life (not a warranty) of this device is 5 years. In order to ensure its long service life, please pay attention to the use of maintenance. y   Please change the batteries when the low-voltage indicator lightens. y  Please clean the surface of the oximeter before use. Use soft cloth with alcohol to wipe the oximeter first, and then let it dry in air or wipe it dry.
User Manual for Fingertip Oximeter 16   y   Please take out the batteries if the oximeter will not be used for a long time. y   The recommended storage environment of the device:  ambient temperature: -20ºC ~60ºC,   relative humidity 10%~95%,   atmospheric pressure: 50kPa~107.4kPa.   y  The oximeter is calibrated in the factory before sale, there is no need to calibrate it during its life cycle. However, if it is necessary to verify its accuracy routinely, the user can do the verification by means of SpO2 simulator, or it can be done by the local third party test house.   y  Necessary servicing must be performed by qualified service engineers ONLY. Users are not permitted to maintain it by themselves.   High-pressure sterilization cannot be used on the device.    Do not immerse the device in liquid.
User Manual for Fingertip Oximeter 17   6.2 Battery Maintenance y   Keep the both sides of coin cell clean. y   Low temperature may decrease the performance of coin cell, and low battery indicator may appear early. In such case, please put coin cell into pocket for warm before use, thus bring it back to normal condition. y  Do not let any conductive metal (such as tweezers) contact both sides of coin cell simultaneously to avoid short circuit. y  Charge the coin cell for 8~10 hours each time; ambient temperature should be 5~40℃. y  If the coin cell is full after charging, but its performance decreases apparently, it means the coin cell is exhausted, please change a new one.   6.3 Cleaning and Disinfecting Instruction y   Surface-clean sensor with a soft cloth by wetting with a solution such as 75% isopropyl alcohol, if low-level disinfection is required, use a 1:10 bleach solution.   y  Then surface-clean by soft cloth wet with clean water and let air dry or wipe it dry. Caution: Do not sterilize by irradiation steam, or ethylene oxide. Do not use the sensor if it is damaged.
User Manual for Fingertip Oximeter 18   7. Troubleshooting Trouble  Possible Reason Solution Display direction doesn’t change or changes insensitively. Maybe the oximeter is not used for a long time, the movable metal ball within the orientation-sensor can not move freely. Please shake the oximeter with a certain force to make the movable metal ball move freely. If the problem still exists,   maybe the orientation-sensor is not working properly. Please contact the local service center. The SpO2 and Pulse Rate display instable 1. The finger is not placed far enough inside. 2. The finger is shaking or the patient is moving.1. Place the finger correctly inside and try again.  2. Let the patient keep calm.
User Manual for Fingertip Oximeter 19   Can not turn on the device 1. The batteries are drained or almost drained.   2. The batteries are not inserted properly.  3. The device is malfunctioning. 1. Change batteries. 2. Reinstall batteries. 3. Please contact the local service center. No display   1. The device will power off automatically when it gets no signal for 8 seconds. 2. The batteries are almost drained. 1. Normal. 2. Change batteries. No display of the wireless icon  “ ” Hardware failure of wireless transmission  function. Please contact the local service center.
User Manual for Fingertip Oximeter 20   Appendix  A Key of Symbols Symbol Description  With Type BF applied part    Warning ─ See User Manual %SpO2  The pulse oxygen saturation PR  Pulse rate (beats per minute)  Pulse rate icon  Wireless icon  Low battery voltage SN  Serial number
User Manual for Fingertip Oximeter 21   B SpO2 Common Knowledge   1. Meaning of SpO2 SpO2 is the saturation percentage of oxygen in the blood, so called O2 concentration in the blood; it is defined by the percentage of oxyhemoglobin (HbO2) in the total hemoglobin of the arterial blood. SpO2 is an important physiological parameter to reflect the respiration function; it is calculated by the following method: SpO2 = HbO2/ (HbO2 +Hb)×100% HbO2  are the oxyhemoglobins (oxygenized hemoglobin), Hb are those hemoglobins which release oxygen. 2. Principle of Measurement Based on Lamber-Beer law, the light absorbance of a given substance is directly proportional with its density or concentration. When the light with certain wavelength emits on human tissue, the measured intensity of light after absorption, reflecting and attenuation in tissue can reflect the structure character of the tissue by which the light passes. Due to that oxygenated hemoglobin (HbO2)
User Manual for Fingertip Oximeter 22   and deoxygenated hemoglobin (Hb) have different absorption character in the spectrum range from red to infrared light (600nm~1000nm wavelength), by using these characteristics, SpO2 can be determined. SpO2 measured by this oximeter is the functional oxygen saturation -- a percentage of the hemoglobin that can transport oxygen. In contrast, hemoximeters report fractional oxygen saturation – a percentage of all measured hemoglobin, including dysfunctional hemoglobin, such as carboxyhemoglobin or metahemoglobin.        Clinical application of pulse oximeters: SpO2  is an important physiological parameter to reflect the respiration and ventilation function, so SpO2  monitoring used in treatment has become more popular. (For example, such as monitoring patients with serious respiratory disease, patients under anesthesia during operation and premature and neonatal infants) The status of SpO2 can be determined in timely manner by measurement and will allow finding the hypoxemia patient earlier, thereby preventing or reducing accidental death caused by hypoxia effectively.
User Manual for Fingertip Oximeter 23   3.Factors affecting SpO2 measuring accuracy (interference reason)      Intravascular dyes such as indocyanine green or methylene blue    Exposure to excessive illumination, such as surgical lamps, bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight.  Vascular dyes or external used color-up product such as nail enamel or color skin care    Excessive patient movement  Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line    Exposure to the chamber with High pressure oxygen   There is an arterial occlusion proximal to the sensor  Blood vessel contraction caused by peripheral vessel hyperkinesias or body temperature decreasing
User Manual for Fingertip Oximeter 24   4. Factors causing low SpO2 Measuring value (pathology reason)    Hypoxemia disease, functional lack of HbO2  Pigmentation or abnormal oxyhemoglobin level  Abnormal oxyhemoglobin variation  Methemoglobin disease  Sulfhemoglobinemia or arterial occlusion exists near sensor  Obvious venous pulsations  Peripheral arterial pulsation becomes weak  Peripheral blood supply is not enough
FCC Notice FCC Statement  This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct.  The interference by one or more of the following measures: ●  Reorient or relocate the receiving antenna. ●  Increase the separation between the equipment and receiver. ●  Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. ●  Consult the dealer or an experienced radio/TV technician for help  This device complies with Part 15 of FCC Rules.     Operation is subject to the following two conditions: 1. This device may not cause harmful interference, and 2. This device must accept any interference received, including interference that may cause undesired operation.  Note: “Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.”