NI PXI/PCI 5152 Specifications 5152specs

User Manual: 5152specs

Open the PDF directly: View PDF PDF.
Page Count: 26

NI PXI/PCI-5152 Specifications
8-Bit 2 GS/s Digitizer
Unless otherwise noted, the following conditions were used for each
specification: all filter settings, all impedance selections, and sample
clock set to 1 GS/s. Real-Time Interleaved Sampling (TIS) mode provides
a 2 GS/s real-time sample rate for a single channel.
Typical values are representative of an average unit operating at room
temperature. Specifications are subject to change without notice. For the
most recent NI 5152 specifications, visit ni.com/manuals.
To access the NI 5152 documentation, including the NI High-Speed
Digitizers Getting Started Guide, go to Start»All Programs»National
Instruments»NI-SCOPE»Documentation.
Hot Surface If the NI 5152 has been in use, it may exceed safe handling temperatures and
cause burns. Allow the NI 5152 to cool before removing it from the PXI chassis.
Contents
Vertical ................................................................................................... 2
Analog Input (Channel 0 and Channel 1) ....................................... 2
Horizontal ............................................................................................... 9
Sample Clock .................................................................................. 9
Phase-Locked Loop (PLL) Reference Clock .................................. 10
PFI 0 (Sample Clock and Reference Clock Input,
Front Panel Connector) ................................................................ 11
PFI 1 (Reference Clock Output, Front Panel Connector) ............... 11
Trigger .................................................................................................... 11
Reference (Stop) Trigger ................................................................ 11
TClk Specifications ................................................................................ 15
Waveform Specifications ....................................................................... 16
Calibration .............................................................................................. 17
Power ..................................................................................................... 17
Software ................................................................................................. 18
Environment ........................................................................................... 19
NI PXI-5152 ................................................................................... 19
NI PCI-5152 .................................................................................... 20
NI PXI/PCI-5152 Specifications 2 ni.com
Safety, Electromagnetic Compatibility, and CE Compliance .................20
Safety ...............................................................................................20
Electromagnetic Compatibility ........................................................21
CE Compliance ................................................................................21
Waste Electrical and Electronic Equipment (WEEE) .....................21
Physical ...................................................................................................22
Front Panel Connectors ...................................................................22
Dimensions and Weigh.....................................................................23
NI PXI-5152 .............................................................................23
NI PCI-5152 .............................................................................24
Where to Go for Support ........................................................................25
Vertical
Analog Input (Channel 0 and Channel 1)
Specification Value Comments
Number of
Channels
Two (simultaneously sampled)
Connectors BNC
Impedance and Coupling
Input
Impedance
50 Ω ±1.5%
1MΩ ±0.75% in parallel with a typical capacitance of 22 pF
Software
selectable.
Input Coupling AC, DC, GND
© National Instruments Corporation 3 NI PXI/PCI-5152 Specifications
Voltage Levels
Full Scale (FS)
Input Range
and
Programmable
Vertical Offset
50 Ω 1MΩ
Range
(Vpk-pk)
Offset
(Volts)
Range
(Vpk-pk)
Offset
(Volts)
0.1 ±10.1 ±1
0.2 ±10.2 ±1
0.4 ±10.4 ±1
1±1 1 ±1
2±6 2 ±10
4±5 4 ±10
10 ±210 ±10
Maximum
Input Overload
50 Ω1MΩ
7V
rms with |Peaks| 10 V |Peaks| 42 V
Accuracy
Resolution 8 bits
DC Accuracy
(Programmable
Vertical Offset
= 0 Volts)
Range (Vpk-pk)50 Ω and 1 MΩ Within ±5 °C of
self-calibration
temperature.
0.1 to 1 ± (1.26% of Input +
1.0% of FS + 500 μV)
2 to 10 ± (1.26% of Input +
1.0% of FS + 5 mV)
Programmable
Vertical Offset
Accuracy
±0.9% of offset setting Within ±5 °C of
self-calibration
temperature.
DC Drift Range (Vpk-pk)50 Ω and 1 MΩ Use DC drift to
calculate errors
when
temperature
changes more
than ±5 °C
since the last
self-calibration.
0.1 to 1 ± (0.052% of Input + 100 μV)
per °C
2 to 10 ± (0.052% of Input + 1.0 mV)
per °C
Specification Value Comments
NI PXI/PCI-5152 Specifications 4 ni.com
Crosstalk,
Typical CH 0 to/from CH 1* Ext Trig to CH 0 or CH 1**
* Measured on
one channel
with test signal
applied to
another
channel, with
same range
setting on both
channels.
** 10 Vpk-pk
signal applied
to external
trigger channel.
Applies to all
ranges on CH 0
and CH1.
< –80 dB at 10 MHz
< 60dB at 100MHz
< –80 dB at 10 MHz
< –80 dB at 100 MHz
Bandwidth and Transient Response
Bandwidth
(–3 dB)
Range
(Vpk-pk)50 Ω1MΩ
Filter off.
Bandwidth for
0to 3C.
Reduce by
0.25% per °C
above 30 °C.
All ranges
except 0.1
340 MHz typical
300 MHz min.
300 MHz typical
260 MHz min.
0.1 165 MHz typical
135 MHz min.
135 MHz typical
110 MHz min.
Rise/Fall Time,
Typical
Range
(Vpk-pk)50 Ω1MΩ
Filter off.
* 50 Ω
terminator
connected to
front panel
BNC connector.
All ranges
except 0.1 1.2 ns 1.4 ns*
0.1 2.4 ns 2.8 ns*
Bandwidth
Limit Filter 20 MHz Noise Filter
AC Coupling
Cutoff (–3 dB),
Typical
50 Ω1MΩ
50 Ω source
assumed.
106 kHz 12 Hz
Specification Value Comments
© National Instruments Corporation 5 NI PXI/PCI-5152 Specifications
Figure 1. NI 5152 Frequency Response, 50 Ω, 1 V (Typical)
Figure 2. NI 5152 Frequency Response, 50 Ω, 1 V (Typical)
1
0
–1
–2
–3
–4
–5
–6
–7
–8
–9
–10
1 10 100 600
Frequency (MHz)
Noise Filter
Amplitude (db relative to 50 kHz)
Full Bandwidth
1
0
–1
–2
–3
–4
–5
–6
–7
–8
–9
–10
0 50 100 150 200 250 300 350 400 450 500 550 600
Frequency (MHz)
Full Bandwidth 2 GS/s
(Time Interleaved
Sampling Mode)
Full Bandwidth 1 GS/s
Amplitude (db relative to 50 kHz)
NI PXI/PCI-5152 Specifications 6 ni.com
Figure 3. NI 5152 Step Response, 50 Ω, 10 Vpk-pk through 0.2 Vpk-pk Range, Typical
Figure 4. NI 5152 Step Response, 50 Ω, 0.1 Vpk-pk Range, Typical
–5 –2.5 0 2.5 5 7.5 10
Time (ns)
Volts
–5 –2.5 0 2.5 5 7.5 10
Time (ns)
Volts
© National Instruments Corporation 7 NI PXI/PCI-5152 Specifications
Figure 5. NI 5152 Typical Dynamic Performance, 50 Ω, 1 Vpk-pk Range,
9.425 MHz, –1 dBFS Input Signal
Figure 6. NI 5152 TIS Typical Dynamic Performance, 50 Ω, 1 Vpk-pk Range,
9.425 MHz, –1 dBFS Input Signal
Specification Value Comments
Spectral Characteristics
ENOB Noise Filter On Noise Filter Off 1V
pk-pk range,
10 MHz, –1 dBFS
input signal.
Includes the 2nd
through the 5th
harmonics.
7.3 7.1
Signal to Noise
and Distortion
(SINAD), Typical
45 dB 43 dB
0
–10
–20
–30
–40
–50
–60
–70
–80
–90
–100
0 100 200 300 400 500
Frequency (MHz)
Amplitude (dBFS)
8-Bit SFDR Limit
0
–10
–20
–30
–40
–50
–60
–70
–80
–90
–100
0 200 400 600 800 1000
Frequency (MHz)
Amplitude (dBFS)
8-Bit SFDR Limit
NI PXI/PCI-5152 Specifications 8 ni.com
Specification Val ue Comments
RMS Noise Range
(Vpk-pk)Noise Filter On Noise Filter Off
50 Ω terminator
connected to input.
0.1 240 μVrms
(0.24% FS)
320 μVrms
(0.32% FS)
0.2 480 μVrms
(0.24% FS)
600 μVrms
(0.30% FS)
0.4 960 μVrms
(0.24% FS)
1.12 mVrms
(0.28% FS)
12.4 mVrms
(0.24% FS)
2.6 mVrms
(0.26% FS)
24.8 mVrms
(0.24% FS)
6.0 mVrms
(0.30% FS)
49.6 mVrms
(0.24% FS)
11.2 mVrms
(0.28% FS)
10 24 mVrms
(0.24% FS)
26 mVrms
(0.26% FS)
Skew
Channel to Channel
Skew, Typical
< 100 ps
© National Instruments Corporation 9 NI PXI/PCI-5152 Specifications
Horizontal
Sample Clock
Specification Value Comments
Sources Internal: Onboard Clock (internal VCSO)*
External: PFI 0 (front panel SMB connector)
* Internal Sample
Clock is locked to
the Reference
Clock or derived
from the onboard
VCSO.
Onboard Clock (Internal VCSO)
Sample Rate Range
Real-Time
Sampling,
Single-Shot
TIS** Mode,
Single-Shot
Random
Interleaved
Sampling
(RIS)*** Mode
* Divide by n
decimation used
for all rates less
than 1 GS/s.
** TIS is a type of
real-time sampling
that is sometimes
called ping-pong.
*** RIS is a type of
equivalent-time
sampling.
15.26 kS/s
to 1 GS/s*
2GS/s
(Single channel
only)
2GS/s to
20 GS/s in
increments of
1GS/s
(Repetitive
waveforms only)
Timebase
Accuracy
Not Phase-Locked to
Reference Clock
Phase-Locked to
Reference Clock
1 GHz ±25 ppm within
±5 °C of external
calibration
temperature
Equal to the Reference
Clock accuracy
Sample Clock
Delay Range
±1 Sample Clock period
Sample Clock
Delay/Adjustment
Resolution
5ps
NI PXI/PCI-5152 Specifications 10 ni.com
Phase-Locked Loop (PLL) Reference Clock
External Sample Clock
Sources PFI 0 (front panel SMB connector)
Frequency Range 350 MHz to 1 GHz Divide by
ndecimation
available where
1 n 65,535.
For more
information about
Sample Clock and
decimation, refer to
the NI High-Speed
Digitizers Help.
Duty Cycle
Tolerance
45% to 55%
Specification Va lue Comments
Sources NI PXI-5152 NI PCI-5152
PXI_CLK10
(PXI backplane
connector)
PFI 0 (front panel SMB
connector)
RTSI 7
PFI 0 (front panel SMB
connector)
Frequency Range 1 MHz to 20 MHz in 1 MHz increments
Default of 10 MHz
The PLL Reference Clock frequency must be accurate
to ±50 ppm.
Duty Cycle
Tolerance
45% to 55%
Exported
Reference Clock
Destinations
NI PXI-5152 NI PCI-5152
PXI_Trig <0..7>
(backplane connector)
PFI 1 (front panel SMB
connector)
RTSI <0..7>
PFI 1 (front panel SMB
connector)
Specification Value Comments
© National Instruments Corporation 11 NI PXI/PCI-5152 Specifications
PFI 0 (Sample Clock and Reference Clock Input, Front Panel Connector)
PFI 1 (Reference Clock Output, Front Panel Connector)
Trigger
Reference (Stop) Trigger
Specification Val ue Comments
Input Voltage
Range
Sine wave: 0.65 Vpk-pk to 2.8 Vpk-pk
(0 dBm to 13 dBm)
Maximum Input
Overload
7V
rms with |Peaks| 10 V
Impedance 50 Ω
Coupling AC
Specification Val ue Comments
Output Impedance 50 Ω
Logic Type 3.3 V CMOS
Maximum Drive
Current
±24 mA
Specification Va lue Comments
Trigger
Types
Edge, Window, Hysteresis, Digital, Immediate,
and Software
Refer to the
following
sections
and to the
NI High-Speed
Digitizers Help
for more
information
about what
sources are
available for
each trigger
type.
Trigger Sources NI PXI-5152 NI PCI-5152
CH 0, CH 1,
TRIG, PFI <0..1>
PXI_Trig <0..6>,
PXI Star Trigger, and
Software
CH 0, CH 1,
TRIG, PFI <0..1>,
RTSI <0..6>, and Software
NI PXI/PCI-5152 Specifications 12 ni.com
Time Resolution TDC Onboard Clock External Clock TDC = Time to
Digital
Conversion
Circuit.
On 5ps N/A
Off 1ns External Clock Period
Rearm Time TDC Rearm Time Holdoff set to 0.
On 8μs
Off 1μs
Holdoff From Rearm time up to [(232 1) × Sample Clock Period]
Trigger Delay From 0 up to [(235 – 1) – posttrigger samples] ×
(1/sample rate), in seconds
Analog Trigger (Edge, Window, and Hysteresis Trigger Types)
Sources CH 0, CH 1, TRIG (front panel BNC connectors)
Trigger Level
Range
CH 0, CH 1 TRIG (External Trigger) DC to 300 MHz
100% FS ±5V
Voltage
Resolution
8 bits (1 in 256)
Edge Trigger
Sensitivity
CH 0, CH 1 TRIG (External Trigger) DC to 300 MHz
10% FS 0.5 Vpk-pk
Trigger Level
Accuracy,
Typical
CH 0, CH 1 TRIG (External Trigger)
±5% FS up to 10 MHz ±1V (±10% FS)
up to 10 MHz)
Trigger Jitter 10 ps rms typical, 20 ps rms maximum Within ±5 °C of
self-calibration
temperature.
Trigger Filters Low Frequency (LF)
Reject
High Frequency (HF)
Reject
50 kHz 50 kHz
Specification Va lue Comments
© National Instruments Corporation 13 NI PXI/PCI-5152 Specifications
Digital Trigger (Digital Trigger Type)
Sources NI PXI-5152 NI PCI-5152
PXI_Trig <0..6> (backplane
connector)
PFI <0..1> (front panel
SMB connectors)
PXI Star Trigger (backplane
connector)
RTSI <0..6>
PFI <0..1> (front panel
SMB connector)
External Trigger Input (Front Panel Connector)
Connector BNC
Impedance 1 MΩ in parallel with a typical capacitance of 22 pF
Coupling AC, DC
AC Coupling
Cutoff (–3 dB)
12 Hz
Input Voltage
Range
± 5V
Maximum Input
Overload
|Peaks| 42 V
PFI 0 and PFI 1 (Programmable Function Interface, Front Panel Connectors)
Connector SMB jack
Direction Bidirectional
As an Input (Trigger)
Destination Start Trigger (Acquisition Arm)
Reference (Stop) Trigger
Arm Reference
Advance Trigger
Input Impedance 150 kΩ
VIH 2.0 V
VIL 0.8 V
Specification Va lue Comments
NI PXI/PCI-5152 Specifications 14 ni.com
Maximum Input
Overload
0.5 V to 5.5 V
Maximum
Frequency
25 MHz
As an Output (Event)
Sources Start Trigger (Acquisition Arm)
Reference (Stop) Trigger
End of Record
Done (End of Acquisition)
Probe Compensation (1 kHz, 50% duty cycle square wave,
PFI 1 only)
Output
Impedance
50 Ω
Logic Type 3.3 V CMOS
Maximum Drive
Current
±24 mA
Maximum
Frequency
25 MHz
Specification Va lue Comments
© National Instruments Corporation 15 NI PXI/PCI-5152 Specifications
TClk Specifications
National Instruments TClk synchronization method and the NI-TClk driver
are used to align the sample clocks on any number of SMC-based modules
in a chassis. For more information about TClk synchronization, refer to the
NI-TClk Synchronization Help, which is located within the NI High-Speed
Digitizers Help.
Specifications are valid for any number of modules installed in one
NI PXI-1042 chassis.
All parameters set to identical values for each SMC-based module.
Sample Clock set to 1 GS/s and all filters are disabled.
For other configurations, including multichassis systems, contact
NI Technical Support at ni.com/support.
Note Although you can use NI-TClk to synchronize nonidentical modules, these
specifications apply only to synchronizing identical modules.
Specification Value Comments
Intermodule SMC Synchronization Using NI-TClk for Identical Modules (Typical)
Skew 500 ps Caused by clock and analog
path delay differences.
No manual adjustment
performed.
Skew After Manual
Adjustment
5ps For information about
manual adjustment, refer to
the Synchronization
Repeatability Optimization
topic in the NI-TClk
Synchronization Help.
For additional help with the
adjustment process, contact
NI Technical Support at
ni.com/support.
Sample Clock
Delay/Adjustment
Resolution
5ps
NI PXI/PCI-5152 Specifications 16 ni.com
Waveform Specifications
Specification Value Comments
Onboard Memory
Size
Real-Time and RIS
Modes Real-Time TIS Mode
* PXI version only
8MB Standard
(8 MS per channel)
8MB Standard (8 MS)
64 MB Option
(64 MS per channel)
64 MB Option (64 MS)
256 MB Option
(256 MS per channel)
256 MB Option (256 MS)
512 MB Option*
(512 MS per channel)
512 MB Option* (512 MS)
Minimum Record
Length
1 Sample
Number of
Pretrigger Samples
Zero up to full record length Single-record
mode and
multiple-record
mode.
Number of
Posttrigger
Samples
Zero up to full record length Single-record
mode and
multiple-record
mode.
Maximum Number
of Records in
Onboard Memory
Memory
Option
Real-Time
Sampling Mode
* It is possible to
exceed these
numbers if you
fetch records while
acquiring data. For
more information,
refer to the
NI High-Speed
Digitizers Help.
8 MB per channel 32,768*
64 MB per channel 100,000*
256 MB per channel 100,000*
Allocated Onboard
Memory per
Record
[(Record length × 1 byte/sample) + 240 bytes]
rounded up to next multiple of 128 bytes
© National Instruments Corporation 17 NI PXI/PCI-5152 Specifications
Calibration
Power
Specification Va lue Comments
Self-Calibration Self-calibration is done on software command.
The calibration corrects for gain, offset, triggering,
and timing errors for all input ranges.
External Calibration
(Factory Calibration)
The external calibration calibrates the VCSO and the
voltage reference. Appropriate constants are stored
in nonvolatile memory.
Interval for External
Calibration
2 years
Warm-Up Time 15 minutes
Specification Typical Value Comments
+3.3 VDC
NI PXI-5152 NI PCI-5152
1.1 A 2.5 A
+5 VDC 1.9 A 2.4 A
+12 VDC 500 mA 200 mA
–12 VDC 210 mA 0 A
Total Power 21.65 W 22.65 W
NI PXI/PCI-5152 Specifications 18 ni.com
Software
Specification Va lue Comments
Driver
Software
NI-SCOPE 3.2 or later for NI PXI-5152
NI-SCOPE 3.3 or later for NI PCI-5152
NI-SCOPE is an IVI-compliant driver that allows you to
configure, control, and calibrate the NI 5152. NI-SCOPE
provides application programming interfaces for many
development environments.
Application
Software
NI-SCOPE provides programming interfaces,
documentation, and examples for the following
application development environments:
•LabVIEW
• LabWindows/CVI
Measurement Studio
Microsoft Visual C/C++
Microsoft Visual Basic
Interactive
Soft Front
Panel and
Configuration
The Scope Soft Front Panel 2.5 or later supports interactive
control of the NI 5152. The Scope Soft Front Panel is
included on the NI-SCOPE CD.
National Instruments Measurement & Automation Explorer
(MAX) also provides interactive configuration and test tools
for the NI 5152. MAX is included on the NI-SCOPE CD.
© National Instruments Corporation 19 NI PXI/PCI-5152 Specifications
Environment
NI PXI-5152
Note To ensure that the NI PXI-5152 cools effectively, follow the guidelines in
the Maintain Forced-Air Cooling Note to Users included in the hardware kit.
The NI PXI-5152 is intended for indoor use only.
Specification Valu e Comments
Operating
Temperature
0 ºC to +55 ºC in all NI PXI chassis except the following:
0 ºC to +45 ºC when installed in an NI PXI-1000/B or
PXI-101x chassis.
Meets IEC-60068-2-1 and IEC-60068-2-2.
Storage
Temperature
40 ºC to +71 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.
Operating Relative
Humidity
10% to 90%, noncondensing. Meets IEC-60068-2-56.
Storage Relative
Humidity
5% to 95%, noncondensing. Meets IEC-60068-2-56.
Operating Shock 30 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
Storage Shock 50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
Operating Vibration 5 Hz to 500 Hz, 0.31 grms. Meets IEC-60068-2-64.
Storage Vibration 5 Hz to 500 Hz, 2.46 grms. Meets IEC-60068-2-64.
Test profile exceeds requirements of MIL-PRF-28800F,
Class B.
Altitude 2,000 m maximum (at 25 °C ambient temperature)
Pollution Degree 2 —
NI PXI/PCI-5152 Specifications 20 ni.com
NI PCI-5152
Note To ensure that the NI PCI-5152 cools effectively, make sure that the chassis in which
it is used has active cooling that provides at least some airflow across the PCI card cage.
To maximize airflow and extend the life of the device, leave any adjacent PCI slots empty.
Refer to the Maintain Forced Air Cooling Note to Users included in the NI PCI-5152 kit
for important cooling information. The NI PCI-5152 is intended for indoor use only.
Safety, Electromagnetic Compatibility,
and CE Compliance
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control, and
laboratory use:
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Specification Valu e Comments
Operating
Temperature
0 ºC to +45 ºC. Meets IEC-60068-2-1 and IEC-60068-2-2.
Storage
Temperature
40 ºC to +70 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.
Operating Relative
Humidity
10% to 90%, noncondensing. Meets IEC-60068-2-56.
Storage Relative
Humidity
5% to 95%, noncondensing. Meets IEC-60068-2-56.
Storage Shock 50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
Storage Vibration 5 Hz to 500 Hz, 2.46 grms. Meets IEC-60068-2-64.
Test profile exceeds requirements of MIL-PRF-28800F,
Class B.
Altitude 2,000 m maximum (at 25 °C ambient temperature)
Pollution Degree 2 —
© National Instruments Corporation 21 NI PXI/PCI-5152 Specifications
Note For UL and other safety certifications, refer to the product label or visit
ni.com/certification, search by model number or product line, and click
the appropriate link in the Certification column.
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
EN 61326 EMC requirements; Minimum Immunity
EN 55011 Emissions; Group 1, Class A
CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note For EMC compliance, operate this device with RG223/U or equivalent shielded
cable. Operate according to product documentation.
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
73/23/EEC; Low-Voltage Directive (safety)
89/336/EEC; Electromagnetic Compatibility Directive (EMC)
Note Refer to the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, visit ni.com/
certification, search by model number or product line, and click the appropriate link
in the Certification column.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling
center. For more information about WEEE recycling centers and National Instruments
WEEE initiatives, visit ni.com/environment/weee.htm.
NI PXI/PCI-5152 Specifications 22 ni.com
Physical
Front Panel Connectors
Label Function Connector Type
CH 0 Analog input BNC female
CH 1 Analog input BNC female
TRIG External Trigger BNC female
PFI 0 Sample Clock Input,
Reference Clock Input
Digital Trigger
Input/Output
SMB jack
PFI 1 Reference Clock Output,
Digital Trigger
Input/Output
SMB jack
© National Instruments Corporation 23 NI PXI/PCI-5152 Specifications
Dimensions and Weight
NI PXI-5152
Specification Value
Dimensions 3U, One slot, PXI/cPCI Module
21.6 × 2.0 × 13.0 cm
(8.5 × 0.8 × 5.1 in.)
Weight 462 g
(16.3 oz)
13.0 cm
(5.1 in.)
21.6 cm
(8.5 in.)
2.0 cm
(0.8 in.)
NI PXI/PCI-5152 Specifications 24 ni.com
NI PCI-5152
Specification Value
Dimensions 32.20 × 2.0 × 11.3 cm
(12.7 × 0.8 × 4.4 in.)
Weight 445 g
(15.7 oz)
32.20 cm
(12.7 in.)
11.3 cm
(4.4 in.)
2.0 cm
(0.8 in.)
© National Instruments Corporation 25 NI PXI/PCI-5152 Specifications
Where to Go for Support
The National Instruments Web site is your complete resource for technical
support. At ni.com/support you have access to everything from
troubleshooting and application development self-help resources to email
and phone assistance from NI Application Engineers.
A Declaration of Conformity (DoC) is our claim of compliance with the
Council of the European Communities using the manufacturer’s
declaration of conformity. This system affords the user protection for
electronic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting ni.com/certification. If your product supports
calibration, you can obtain the calibration certificate for your product at
ni.com/calibration.
National Instruments corporate headquarters is located at
11500 North Mopac Expressway, Austin, Texas, 78759-3504.
National Instruments also has offices located around the world to help
address your support needs. For telephone support in the United States,
create your service request at ni.com/support and follow the calling
instructions or dial 512 795 8248. For telephone support outside the United
States, contact your local branch office:
Australia 1800 300 800, Austria 43 662 457990-0,
Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599,
Canada 800 433 3488, China 86 21 5050 9800,
Czech Republic 420 224 235 774, Denmark 45 45 76 26 00,
Finland 385 (0) 9 725 72511, France 33 (0) 1 48 14 24 24,
Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737,
Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400,
Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710,
Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466,
New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60,
Poland 48 22 3390150, Portugal 351 210 311 210, Russia 7 495 783 6851,
Singapore 1800 226 5886, Slovenia 386 3 425 42 00,
South Africa 27 0 11 805 8197, Spain 34 91 640 0085,
Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151,
Taiwan 886 02 2377 2222, Thailand 662 278 6777,
Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation.
Refer to the Terms of Use section on ni.com/legal for more information about National
Instruments trademarks. Other product and company names mentioned herein are trademarks or trade
names of their respective companies. For patents covering National Instruments products, refer to the
appropriate location: Help»Patents in your software, the patents.txt file on your CD, or
ni.com/patents.
© 2006-2007 National Instruments Corporation. All rights reserved. 374443C-01 Mar07

Navigation menu