NI PXI/PCI-5152 Specifications
8-Bit 2 GS/s Digitizer
Unless otherwise noted, the following conditions were used for each
specification: all filter settings, all impedance selections, and sample
clock set to 1 GS/s. Real-Time Interleaved Sampling (TIS) mode provides
a 2 GS/s real-time sample rate for a single channel.
Typical values are representative of an average unit operating at room
temperature. Specifications are subject to change without notice. For the
most recent NI 5152 specifications, visit ni.com/manuals.
To access the NI 5152 documentation, including the NI High-Speed
Digitizers Getting Started Guide, go to Start»All Programs»National
Instruments»NI-SCOPE»Documentation.
Hot Surface If the NI 5152 has been in use, it may exceed safe handling temperatures and
cause burns. Allow the NI 5152 to cool before removing it from the PXI chassis.

Contents
Vertical ................................................................................................... 2
Analog Input (Channel 0 and Channel 1) ....................................... 2
Horizontal ............................................................................................... 9
Sample Clock .................................................................................. 9
Phase-Locked Loop (PLL) Reference Clock .................................. 10
PFI 0 (Sample Clock and Reference Clock Input,
Front Panel Connector) ................................................................ 11
PFI 1 (Reference Clock Output, Front Panel Connector) ............... 11
Trigger .................................................................................................... 11
Reference (Stop) Trigger ................................................................ 11
TClk Specifications ................................................................................ 15
Waveform Specifications ....................................................................... 16
Calibration .............................................................................................. 17
Power ..................................................................................................... 17
Software ................................................................................................. 18
Environment ........................................................................................... 19
NI PXI-5152 ................................................................................... 19
NI PCI-5152 .................................................................................... 20

Safety, Electromagnetic Compatibility, and CE Compliance .................20
Safety ...............................................................................................20
Electromagnetic Compatibility ........................................................21
CE Compliance ................................................................................21
Waste Electrical and Electronic Equipment (WEEE) .....................21
Physical ...................................................................................................22
Front Panel Connectors ...................................................................22
Dimensions and Weigh.....................................................................23
NI PXI-5152 .............................................................................23
NI PCI-5152 .............................................................................24
Where to Go for Support ........................................................................25

Vertical
Analog Input (Channel 0 and Channel 1)
Specification

Value

Comments

Number of
Channels

Two (simultaneously sampled)

—

Connectors

BNC

—

Impedance and Coupling
Input
Impedance
Input Coupling

50 Ω ±1.5%
1 MΩ ±0.75% in parallel with a typical capacitance of 22 pF
AC, DC, GND

NI PXI/PCI-5152 Specifications

Software
selectable.
—

2

ni.com

Specification

Value

Comments

Voltage Levels
Full Scale (FS)
Input Range
and
Programmable
Vertical Offset

Maximum
Input Overload

50 Ω

—

1 MΩ

Range
(Vpk-pk)

Offset
(Volts)

Range
(Vpk-pk)

Offset
(Volts)

0.1

±1

0.1

±1

0.2

±1

0.2

±1

0.4

±1

0.4

±1

1

±1

1

±1

2

±6

2

±10

4

±5

4

±10

10

±2

10

±10

50 Ω

1 MΩ

7 Vrms with |Peaks| ≤10 V

|Peaks| ≤42 V

—

Accuracy
Resolution
DC Accuracy
(Programmable
Vertical Offset
= 0 Volts)

Programmable
Vertical Offset
Accuracy
DC Drift

8 bits

—
Range (Vpk-pk)

50 Ω and 1 MΩ

0.1 to 1

± (1.26% of Input +
1.0% of FS + 500 μV)

2 to 10

± (1.26% of Input +
1.0% of FS + 5 mV)

±0.9% of offset setting

Within ±5 °C of
self-calibration
temperature.

Range (Vpk-pk)

50 Ω and 1 MΩ

0.1 to 1

± (0.052% of Input + 100 μV)
per °C

2 to 10

± (0.052% of Input + 1.0 mV)
per °C

© National Instruments Corporation

Within ±5 °C of
self-calibration
temperature.

3

Use DC drift to
calculate errors
when
temperature
changes more
than ±5 °C
since the last
self-calibration.

NI PXI/PCI-5152 Specifications

Specification
Crosstalk,
Typical

Value

Comments
*

CH 0 to/from CH

1*

Ext Trig to CH 0 or CH

< –80 dB at 10 MHz

< –80 dB at 10 MHz

< –60 dB at 100 MHz

< –80 dB at 100 MHz

1**

Measured on
one channel
with test signal
applied to
another
channel, with
same range
setting on both
channels.

**

10 Vpk-pk
signal applied
to external
trigger channel.
Applies to all
ranges on CH 0
and CH1.
Bandwidth and Transient Response
Bandwidth
(–3 dB)

Range
(Vpk-pk)

1 MΩ

All ranges
except 0.1

340 MHz typical
300 MHz min.

300 MHz typical
260 MHz min.

0.1

165 MHz typical
135 MHz min.

135 MHz typical
110 MHz min.

Rise/Fall Time,
Typical

Bandwidth
Limit Filter

Filter off.

50 Ω

50 Ω

1 MΩ

All ranges
except 0.1

1.2 ns

1.4 ns*

0.1

2.4 ns

2.8 ns*

50 Ω
terminator
connected to
front panel
BNC connector.
*

—

20 MHz Noise Filter

NI PXI/PCI-5152 Specifications

Reduce by
0.25% per °C
above 30 °C.
Filter off.

Range
(Vpk-pk)

AC Coupling
Cutoff (–3 dB),
Typical

Bandwidth for
0 to 30 °C.

50 Ω

1 MΩ

106 kHz

12 Hz

4

50 Ω source
assumed.

ni.com

Amplitude (db relative to 50 kHz)

1
0
–1
–2
–3
–4
–5
–6
–7
–8
–9
–10

Full Bandwidth

Noise Filter

1

10
Frequency (MHz)

100

600

Amplitude (db relative to 50 kHz)

Figure 1. NI 5152 Frequency Response, 50 Ω, 1 V (Typical)
1
0
–1
–2
–3
–4
–5
–6
–7
–8
–9
–10

Full Bandwidth 1 GS/s
Full Bandwidth 2 GS/s
(Time Interleaved
Sampling Mode)

0

50

100 150 200 250 300 350 400 450 500 550
Frequency (MHz)

600

Figure 2. NI 5152 Frequency Response, 50 Ω, 1 V (Typical)

© National Instruments Corporation

5

NI PXI/PCI-5152 Specifications

Volts

–5

–2.5

0

2.5
Time (ns)

5

7.5

10

Volts

Figure 3. NI 5152 Step Response, 50 Ω , 10 Vpk-pk through 0.2 Vpk-pk Range, Typical

–5

–2.5

0

2.5
Time (ns)

5

7.5

10

Figure 4. NI 5152 Step Response, 50 Ω, 0.1 Vpk-pk Range, Typical

NI PXI/PCI-5152 Specifications

6

ni.com

Specification

Value

Comments

Spectral Characteristics
ENOB

Noise Filter On

Noise Filter Off

7.3

7.1

45 dB

43 dB

Signal to Noise
and Distortion
(SINAD), Typical

1 Vpk-pk range,
10 MHz, –1 dBFS
input signal.
Includes the 2nd
through the 5th
harmonics.

Amplitude (dBFS)

0
–10
–20
–30
–40
–50
–60
–70
–80
–90

8-Bit SFDR Limit

–100
0

100

200
300
Frequency (MHz)

400

500

Figure 5. NI 5152 Typical Dynamic Performance, 50 Ω, 1 Vpk-pk Range,
9.425 MHz, –1 dBFS Input Signal

Amplitude (dBFS)

0
–10
–20
–30
–40
–50
–60
–70
–80
–90

8-Bit SFDR Limit

–100
0

200

400
600
Frequency (MHz)

800

1000

Figure 6. NI 5152 TIS Typical Dynamic Performance, 50 Ω, 1 Vpk-pk Range,
9.425 MHz, –1 dBFS Input Signal

© National Instruments Corporation

7

NI PXI/PCI-5152 Specifications

Specification

Value

RMS Noise

Range
(Vpk-pk)

Comments

Noise Filter On

Noise Filter Off

0.1

240 μVrms
(0.24% FS)

320 μVrms
(0.32% FS)

0.2

480 μVrms
(0.24% FS)

600 μVrms
(0.30% FS)

0.4

960 μVrms
(0.24% FS)

1.12 mVrms
(0.28% FS)

1

2.4 mVrms
(0.24% FS)

2.6 mVrms
(0.26% FS)

2

4.8 mVrms
(0.24% FS)

6.0 mVrms
(0.30% FS)

4

9.6 mVrms
(0.24% FS)

11.2 mVrms
(0.28% FS)

10

24 mVrms
(0.24% FS)

26 mVrms
(0.26% FS)

50 Ω terminator
connected to input.

Skew
Channel to Channel
Skew, Typical

< 100 ps

NI PXI/PCI-5152 Specifications

—

8

ni.com

Horizontal
Sample Clock
Specification
Sources

Value

Comments

Internal: Onboard Clock (internal VCSO)*

* Internal

Sample
Clock is locked to
the Reference
Clock or derived
from the onboard
VCSO.

External: PFI 0 (front panel SMB connector)

Onboard Clock (Internal VCSO)
Sample Rate Range
Real-Time
Sampling,
Single-Shot
15.26 kS/s
to 1 GS/s*

Timebase
Accuracy

TIS** Mode,
Single-Shot
2 GS/s
(Single channel
only)

Not Phase-Locked to
Reference Clock
1 GHz ±25 ppm within
±5 °C of external
calibration
temperature

* Divide

Random
Interleaved
Sampling
(RIS)*** Mode

by n
decimation used
for all rates less
than 1 GS/s.

2 GS/s to
20 GS/s in
increments of
1 GS/s

** TIS is a type of
real-time sampling
that is sometimes
called ping-pong.

(Repetitive
waveforms only)

***

Phase-Locked to
Reference Clock

RIS is a type of
equivalent-time
sampling.
—

Equal to the Reference
Clock accuracy

Sample Clock
Delay Range

±1 Sample Clock period

—

Sample Clock
Delay/Adjustment
Resolution

≤5 ps

—

© National Instruments Corporation

9

NI PXI/PCI-5152 Specifications

Specification

Value

Comments

External Sample Clock
Sources

PFI 0 (front panel SMB connector)

Frequency Range

350 MHz to 1 GHz

Duty Cycle
Tolerance

45% to 55%

—
Divide by
n decimation
available where
1 ≤ n ≤ 65,535.
For more
information about
Sample Clock and
decimation, refer to
the NI High-Speed
Digitizers Help.
—

Phase-Locked Loop (PLL) Reference Clock
Specification

Value

Sources

NI PXI-5152

NI PCI-5152

PXI_CLK10
(PXI backplane
connector)
PFI 0 (front panel SMB
connector)
Frequency Range

Comments
—

RTSI 7
PFI 0 (front panel SMB
connector)

1 MHz to 20 MHz in 1 MHz increments
Default of 10 MHz

—

The PLL Reference Clock frequency must be accurate
to ± 50 ppm.
Duty Cycle
Tolerance
Exported
Reference Clock
Destinations

45% to 55%

—

NI PXI-5152

NI PCI-5152

PXI_Trig <0..7>
(backplane connector)

RTSI <0..7>
PFI 1 (front panel SMB
connector)

PFI 1 (front panel SMB
connector)

NI PXI/PCI-5152 Specifications

—

10

ni.com

PFI 0 (Sample Clock and Reference Clock Input, Front Panel Connector)
Specification

Value

Comments

Input Voltage
Range

Sine wave: 0.65 Vpk-pk to 2.8 Vpk-pk
(0 dBm to 13 dBm)

—

Maximum Input
Overload

7 Vrms with |Peaks| ≤10 V

—

Impedance

50 Ω

—

Coupling

AC

—

PFI 1 (Reference Clock Output, Front Panel Connector)
Specification

Value

Comments

Output Impedance

50 Ω

—

Logic Type

3.3 V CMOS

—

Maximum Drive
Current

± 24 mA

—

Trigger
Reference (Stop) Trigger
Specification
Trigger
Types

Value

Comments

Edge, Window, Hysteresis, Digital, Immediate,
and Software

Trigger Sources

NI PXI-5152

NI PCI-5152

CH 0, CH 1,
TRIG, PFI <0..1>
PXI_Trig <0..6>,
PXI Star Trigger, and
Software

© National Instruments Corporation

CH 0, CH 1,
TRIG, PFI <0..1>,
RTSI <0..6>, and Software

11

Refer to the
following
sections
and to the
NI High-Speed
Digitizers Help
for more
information
about what
sources are
available for
each trigger
type.

NI PXI/PCI-5152 Specifications

Specification
Time Resolution

Value

Comments

TDC

Onboard Clock

External Clock

On

5 ps

N/A

Off

1 ns

External Clock Period

TDC = Time to
Digital
Conversion
Circuit.

TDC

Rearm Time

Holdoff set to 0.

On

8 μs

Off

1 μs

Rearm Time

Holdoff

From Rearm time up to [(232 – 1) × Sample Clock Period]

—

Trigger Delay

From 0 up to [(235 – 1) – posttrigger samples] ×
(1/sample rate), in seconds

—

Analog Trigger (Edge, Window, and Hysteresis Trigger Types)
Sources

CH 0, CH 1, TRIG (front panel BNC connectors)

Trigger Level
Range
Voltage
Resolution

CH 0, CH 1

TRIG (External Trigger)

100% FS

±5 V

8 bits (1 in 256)
CH 0, CH 1

TRIG (External Trigger)

10% FS

0.5 Vpk-pk

Trigger Level
Accuracy,
Typical

CH 0, CH 1

TRIG (External Trigger)

±5% FS up to 10 MHz

±1 V (±10% FS)
up to 10 MHz)

≤10 ps rms typical, ≤20 ps rms maximum

Trigger Filters

NI PXI/PCI-5152 Specifications

DC to 300 MHz

—

Edge Trigger
Sensitivity

Trigger Jitter

—

Low Frequency (LF)
Reject

High Frequency (HF)
Reject

50 kHz

50 kHz

12

DC to 300 MHz

—

Within ±5 °C of
self-calibration
temperature.
—

ni.com

Specification

Value

Comments

Digital Trigger (Digital Trigger Type)
Sources

NI PXI-5152

NI PCI-5152

PXI_Trig <0..6> (backplane
connector)

—

RTSI <0..6>
PFI <0..1> (front panel
SMB connector)

PFI <0..1> (front panel
SMB connectors)
PXI Star Trigger (backplane
connector)

External Trigger Input (Front Panel Connector)
Connector

BNC

—

Impedance

1 MΩ in parallel with a typical capacitance of 22 pF

—

Coupling

AC, DC

—

AC Coupling
Cutoff (–3 dB)

12 Hz

—

Input Voltage
Range

±5V

—

Maximum Input
Overload

|Peaks| ≤ 42 V

—

PFI 0 and PFI 1 (Programmable Function Interface, Front Panel Connectors)
Connector

SMB jack

—

Direction

Bidirectional

—

As an Input (Trigger)
Destination

Start Trigger (Acquisition Arm)

—

Reference (Stop) Trigger
Arm Reference
Advance Trigger
Input Impedance

150 kΩ

—

VIH

2.0 V

—

VIL

0.8 V

—

© National Instruments Corporation

13

NI PXI/PCI-5152 Specifications

Specification

Value

Comments

Maximum Input
Overload

–0.5 V to 5.5 V

—

Maximum
Frequency

25 MHz

—

As an Output (Event)
Sources

Start Trigger (Acquisition Arm)

—

Reference (Stop) Trigger
End of Record
Done (End of Acquisition)
Probe Compensation (1 kHz, 50% duty cycle square wave,
PFI 1 only)
Output
Impedance

50 Ω

—

Logic Type

3.3 V CMOS

—

Maximum Drive
Current

± 24 mA

—

Maximum
Frequency

25 MHz

—

NI PXI/PCI-5152 Specifications

14

ni.com

TClk Specifications
National Instruments TClk synchronization method and the NI-TClk driver
are used to align the sample clocks on any number of SMC-based modules
in a chassis. For more information about TClk synchronization, refer to the
NI-TClk Synchronization Help, which is located within the NI High-Speed
Digitizers Help.
•

Specifications are valid for any number of modules installed in one
NI PXI-1042 chassis.

•

All parameters set to identical values for each SMC-based module.

•

Sample Clock set to 1 GS/s and all filters are disabled.

•

For other configurations, including multichassis systems, contact
NI Technical Support at ni.com/support.

Note Although you can use NI-TClk to synchronize nonidentical modules, these

specifications apply only to synchronizing identical modules.

Specification

Value

Comments

Intermodule SMC Synchronization Using NI-TClk for Identical Modules (Typical)
Skew

500 ps

Caused by clock and analog
path delay differences.
No manual adjustment
performed.

Skew After Manual
Adjustment

≤ 5 ps

For information about
manual adjustment, refer to
the Synchronization
Repeatability Optimization
topic in the NI-TClk
Synchronization Help.
For additional help with the
adjustment process, contact
NI Technical Support at
ni.com/support.

Sample Clock
Delay/Adjustment
Resolution

≤ 5 ps

—

© National Instruments Corporation

15

NI PXI/PCI-5152 Specifications

Waveform Specifications
Specification

Value

Onboard Memory
Size

Comments
*

Real-Time and RIS
Modes

PXI version only

Real-Time TIS Mode

8 MB Standard
(8 MS per channel)

8 MB Standard (8 MS)

64 MB Option
(64 MS per channel)

64 MB Option (64 MS)

256 MB Option
(256 MS per channel)

256 MB Option (256 MS)

512 MB Option*
(512 MS per channel)

512 MB Option* (512 MS)

Minimum Record
Length

1 Sample

Number of
Pretrigger Samples

Zero up to full record length

Single-record
mode and
multiple-record
mode.

Number of
Posttrigger
Samples

Zero up to full record length

Single-record
mode and
multiple-record
mode.

Maximum Number
of Records in
Onboard Memory

Allocated Onboard
Memory per
Record

—

Memory
Option

Real-Time
Sampling Mode

8 MB per channel

32,768*

64 MB per channel

100,000*

256 MB per channel

100,000*

[(Record length × 1 byte/sample) + 240 bytes]
rounded up to next multiple of 128 bytes

NI PXI/PCI-5152 Specifications

16

*

It is possible to
exceed these
numbers if you
fetch records while
acquiring data. For
more information,
refer to the
NI High-Speed
Digitizers Help.

—

ni.com

Calibration
Specification

Value

Comments

Self-Calibration

Self-calibration is done on software command.
The calibration corrects for gain, offset, triggering,
and timing errors for all input ranges.

—

External Calibration
(Factory Calibration)

The external calibration calibrates the VCSO and the
voltage reference. Appropriate constants are stored
in nonvolatile memory.

—

Interval for External
Calibration

2 years

—

Warm-Up Time

15 minutes

—

Power
Specification

Typical Value

Comments

NI PXI-5152

NI PCI-5152

+3.3 VDC

1.1 A

2.5 A

+5 VDC

1.9 A

2.4 A

+12 VDC

500 mA

200 mA

–12 VDC

210 mA

0A

Total Power

21.65 W

22.65 W

© National Instruments Corporation

17

—

NI PXI/PCI-5152 Specifications

Software
Specification
Driver
Software

Value
NI-SCOPE 3.2 or later for NI PXI-5152
NI-SCOPE 3.3 or later for NI PCI-5152

Comments
—

NI-SCOPE is an IVI-compliant driver that allows you to
configure, control, and calibrate the NI 5152. NI-SCOPE
provides application programming interfaces for many
development environments.
Application
Software

NI-SCOPE provides programming interfaces,
documentation, and examples for the following
application development environments:

—

• LabVIEW
• LabWindows™/CVI™
• Measurement Studio
• Microsoft Visual C/C++
• Microsoft Visual Basic
Interactive
Soft Front
Panel and
Configuration

The Scope Soft Front Panel 2.5 or later supports interactive
control of the NI 5152. The Scope Soft Front Panel is
included on the NI-SCOPE CD.

—

National Instruments Measurement & Automation Explorer
(MAX) also provides interactive configuration and test tools
for the NI 5152. MAX is included on the NI-SCOPE CD.

NI PXI/PCI-5152 Specifications

18

ni.com

Environment
NI PXI-5152
Note To ensure that the NI PXI-5152 cools effectively, follow the guidelines in
the Maintain Forced-Air Cooling Note to Users included in the hardware kit.
The NI PXI-5152 is intended for indoor use only.

Specification
Operating
Temperature

Value

Comments

0 ºC to +55 ºC in all NI PXI chassis except the following:

—

0 ºC to +45 ºC when installed in an NI PXI-1000/B or
PXI-101x chassis.
Meets IEC-60068-2-1 and IEC-60068-2-2.

Storage
Temperature

–40 ºC to +71 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.

—

Operating Relative
Humidity

10% to 90%, noncondensing. Meets IEC-60068-2-56.

—

Storage Relative
Humidity

5% to 95%, noncondensing. Meets IEC-60068-2-56.

—

Operating Shock

30 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.

—

Storage Shock

50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.

—

Operating Vibration

5 Hz to 500 Hz, 0.31 grms. Meets IEC-60068-2-64.

—

Storage Vibration

5 Hz to 500 Hz, 2.46 grms. Meets IEC-60068-2-64.
Test profile exceeds requirements of MIL-PRF-28800F,
Class B.

—

Altitude

2,000 m maximum (at 25 °C ambient temperature)

—

Pollution Degree

2

—

© National Instruments Corporation

19

NI PXI/PCI-5152 Specifications

NI PCI-5152
Note To ensure that the NI PCI-5152 cools effectively, make sure that the chassis in which
it is used has active cooling that provides at least some airflow across the PCI card cage.
To maximize airflow and extend the life of the device, leave any adjacent PCI slots empty.
Refer to the Maintain Forced Air Cooling Note to Users included in the NI PCI-5152 kit
for important cooling information. The NI PCI-5152 is intended for indoor use only.

Specification

Value

Comments

Operating
Temperature

0 ºC to +45 ºC. Meets IEC-60068-2-1 and IEC-60068-2-2.

—

Storage
Temperature

–40 ºC to +70 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.

—

Operating Relative
Humidity

10% to 90%, noncondensing. Meets IEC-60068-2-56.

—

Storage Relative
Humidity

5% to 95%, noncondensing. Meets IEC-60068-2-56.

—

Storage Shock

50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.

—

Storage Vibration

5 Hz to 500 Hz, 2.46 grms. Meets IEC-60068-2-64.
Test profile exceeds requirements of MIL-PRF-28800F,
Class B.

—

Altitude

2,000 m maximum (at 25 °C ambient temperature)

—

Pollution Degree

2

—

Safety, Electromagnetic Compatibility,
and CE Compliance
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control, and
laboratory use:

NI PXI/PCI-5152 Specifications

•

IEC 61010-1, EN 61010-1

•

UL 61010-1, CSA 61010-1

20

ni.com

Note For UL and other safety certifications, refer to the product label or visit
ni.com/certification, search by model number or product line, and click

the appropriate link in the Certification column.

Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
•

EN 61326 EMC requirements; Minimum Immunity

•

EN 55011 Emissions; Group 1, Class A

•

CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A

Note For EMC compliance, operate this device with RG223/U or equivalent shielded

cable. Operate according to product documentation.

CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
•

73/23/EEC; Low-Voltage Directive (safety)

•

89/336/EEC; Electromagnetic Compatibility Directive (EMC)

Note Refer to the Declaration of Conformity (DoC) for this product for any additional

regulatory compliance information. To obtain the DoC for this product, visit ni.com/
certification, search by model number or product line, and click the appropriate link
in the Certification column.

Waste Electrical and Electronic Equipment (WEEE)
At the end of their life cycle, all products must be sent to a WEEE recycling
center. For more information about WEEE recycling centers and National Instruments
WEEE initiatives, visit ni.com/environment/weee.htm.

EU Customers

© National Instruments Corporation

21

NI PXI/PCI-5152 Specifications

Physical
Front Panel Connectors
Label

Function

Connector Type

CH 0

Analog input

BNC female

CH 1

Analog input

BNC female

TRIG

External Trigger

BNC female

PFI 0

Sample Clock Input,
Reference Clock Input
Digital Trigger
Input/Output

SMB jack

PFI 1

Reference Clock Output,
Digital Trigger
Input/Output

SMB jack

NI PXI/PCI-5152 Specifications

22

ni.com

Dimensions and Weight
NI PXI-5152
Specification
Dimensions

Value
3U, One slot, PXI/cPCI Module
21.6 × 2.0 × 13.0 cm
(8.5 × 0.8 × 5.1 in.)
2.0 cm
(0.8 in.)

13.0 cm
(5.1 in.)

21.6 cm
(8.5 in.)

Weight

462 g
(16.3 oz)

© National Instruments Corporation

23

NI PXI/PCI-5152 Specifications

NI PCI-5152
Specification
Dimensions

Value
32.20 × 2.0 × 11.3 cm
(12.7 × 0.8 × 4.4 in.)
2.0 cm
(0.8 in.)
32.20 cm
(12.7 in.)

11.3 cm
(4.4 in.)

Weight

445 g
(15.7 oz)

NI PXI/PCI-5152 Specifications

24

ni.com

Where to Go for Support
The National Instruments Web site is your complete resource for technical
support. At ni.com/support you have access to everything from
troubleshooting and application development self-help resources to email
and phone assistance from NI Application Engineers.
A Declaration of Conformity (DoC) is our claim of compliance with the
Council of the European Communities using the manufacturer’s
declaration of conformity. This system affords the user protection for
electronic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting ni.com/certification. If your product supports
calibration, you can obtain the calibration certificate for your product at
ni.com/calibration.
National Instruments corporate headquarters is located at
11500 North Mopac Expressway, Austin, Texas, 78759-3504.
National Instruments also has offices located around the world to help
address your support needs. For telephone support in the United States,
create your service request at ni.com/support and follow the calling
instructions or dial 512 795 8248. For telephone support outside the United
States, contact your local branch office:
Australia 1800 300 800, Austria 43 662 457990-0,
Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599,
Canada 800 433 3488, China 86 21 5050 9800,
Czech Republic 420 224 235 774, Denmark 45 45 76 26 00,
Finland 385 (0) 9 725 72511, France 33 (0) 1 48 14 24 24,
Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737,
Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400,
Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710,
Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466,
New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60,
Poland 48 22 3390150, Portugal 351 210 311 210, Russia 7 495 783 6851,
Singapore 1800 226 5886, Slovenia 386 3 425 42 00,
South Africa 27 0 11 805 8197, Spain 34 91 640 0085,
Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151,
Taiwan 886 02 2377 2222, Thailand 662 278 6777,
Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545

© National Instruments Corporation

25

NI PXI/PCI-5152 Specifications

National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation.
Refer to the Terms of Use section on ni.com/legal for more information about National
Instruments trademarks. Other product and company names mentioned herein are trademarks or trade
names of their respective companies. For patents covering National Instruments products, refer to the
appropriate location: Help»Patents in your software, the patents.txt file on your CD, or
ni.com/patents.
© 2006-2007 National Instruments Corporation. All rights reserved.

374443C-01

Mar07



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Encryption                      : Standard V2.3 (128-bit)
User Access                     : Print, Copy, Extract, Print high-res
Page Mode                       : UseOutlines
XMP Toolkit                     : 3.1-702
Producer                        : Acrobat Distiller 7.0.5 (Windows)
Creator Tool                    : FrameMaker 7.2
Modify Date                     : 2007:03:22 16:11:18-05:00
Create Date                     : 1994:04:04 15:32:45Z
Metadata Date                   : 2007:03:22 16:11:18-05:00
Format                          : application/pdf
Title                           : NI PXI/PCI-5152 Specifications
Description                     : 374443C-01, March 2007
Creator                         : Technical Communications
Document ID                     : uuid:41500247-33f7-4809-b168-993d1c4d0812
Instance ID                     : uuid:c8e11139-0b0a-43a8-9d5b-be02a2d4f8ae
Has XFA                         : No
Page Count                      : 26
Page Layout                     : SinglePage
Subject                         : 374443C-01, March 2007
Author                          : Technical Communications
EXIF Metadata provided by EXIF.tools

Navigation menu