5990 7580_pxi Dm_2 17 12_d3a A_M9183A A M9183A

User Manual: A_M9183A

Open the PDF directly: View PDF PDF.
Page Count: 14

Download5990-7580_pxi-dm_2-17-12_d3a A_M9183A A M9183A
Open PDF In BrowserView PDF
Agilent PXI Digital
Multimeters

Data Sheet

6½ Digit, High Performance
M9182A
M9183A

• 6½ digit resolution
• High throughput: up to 15,000 rdgs/sec,
66 µs single reading interval
• DCV basic one year accuracy: 40 ppm
• Ten standard measurements: DCV,
ACV, DCI, ACI, 2- and 4-wire resistance,
frequency, period, capacitance, and
temperature
• Floating isolation (CAT II) up to 300 V

TM

Discover the Alternatives...
... Agilent Modular Products

O VERV IE W
Agilent’s M9182A and M9183A 6½ digit high performance
PXI digital multimeters (DMMs) offer fast throughput, flexible measurements, and trustworthy results. The M9182A
provides ten built-in measurement types with all the accuracy and stability you would expect from an Agilent 6½
digit DMM. The M9183A provides the same capabilities as
the M9182A, with market-leading measurement speed of
up to 15,000 readings per second, additional ranges, and a
DC source.

F eatures

I n du s tr ies a n d A p p l i c a t i o n s

• DCV, ACV, DCI, ACI, 2- and 4-wire resistance, frequency,
period, capacitance, and temperature

• 6½ digit resolution
• Up to 15,000 readings per second, 66 µs single reading
interval time (M9183A)
• Up to 4,500 readings per second, 222 µs single reading
interval time (M9182A)
• Basic 1 year DCV accuracy of 40 ppm

• Aerospace and defense
• Automotive electronics test

• Pulse width, duty cycle, and totalizer/event counter
(M9183A)

• Industrial electronics test

• DC voltage and current source (M9183A)

• Medical device test

• External trigger in and DMM out, to synchronize with
external multiplexers and instruments

• Semiconductor and component test

• Analog threshold trigger with pre- and post- levels for
measuring the correct signal
• Floating isolation (CAT II) to 300 V
• Software drivers to support most common programming
environments
• Chassis connector compatibility: PXI-1 (J-1 only), PXIe
hybrid slot

Customer Values
• Measurements you can trust
• Higher test throughput due to the lowest latency
• Application development in the environment of your
choice reduces development time
• Customer supportable calibration procedures as well as
calibration services available from Agilent

PXI-DMM M918xA Feature Summary
All three products are 6½ digit PXI DMMs that take DCV, ACV, DCI, ACI, 2- and 4-wire resistance measurements.
DMM

Description

DCV basic Maximum
1 year
reading rate at
accuracy 4½ digits

Other
measurements

Triggering

DC source

M9181A*

Basic features
PXI DMM
High
performance
PXI DMM
Enhanced
performance
PXI DMM

90 ppm

150 rdgs/sec

None

Immediate

n/a

40 ppm

4,500 rdgs/sec

M9182A

M9183A

40 ppm

Temperature,
capacitance,
frequency, period
15,000 rdgs/sec Temperature,
capacitance,
frequency, period,
offset compensated
Ω, pulse width, duty
cycle, totalizer/
event counter

Immediate, analog n/a
threshold, PXI trigger
bus, external
Immediate, analog ± 10 V
threshold, PXI trigger ± (1.2 µA to 12 mA)
bus, external

*This table and more information about the M9181A can be found in the M9181A data sheet, literature number 5990-9035EN.
2

E ASY SET UP … T ES T … A N D M A I N T E N A N C E
H a r d w a r e P l a t f o rm
Compliance

Easy software integration

The M9182A and M9183A 6½ digit DMMs are PXI
compliant, using either a cPCI (J1), PXI-1 (J1) or PXIe
Hybrid slot. Designed to benefit from fast data interfaces,
an M9182A or M9183A DMM can be integrated with other
test and automation modules in a PXI, CompactPCI, or
Hybrid chassis. The PXI format offers high performance
in a small, rugged package. It is an ideal deployment
platform for many automated test systems. A wide array
of complementary PXI products are currently available.
Products from Agilent include switches, multiplexers,
digitizers, waveform generators, and local oscillators.

Application code examples are included for LabVIEW,
Visual Basic, C/C++, C#, and MATLAB – demonstrating
DMM set up and basic functionality. These application
code examples can be used to help you integrate the DMM
module into your measurement system.

Soft front panel
The soft front panel provides easy to use instrument
control. The M9182A and M9183A graphical user interface
guides developers through module setup so users can
quickly configure the DMM.

S oftw a r e P l a t f o r m
IO Libraries Suite
Agilent IO Libraries Suite offers fast and easy instrument
connections. IO Libraries Suite 16.1 supports PXI, helping
you display all of the modules in your system, view
information about installed software and allows you to
more easily find the right driver and start module soft front
panels directly with Agilent Connection Expert.
National Instruments IO libraries are also supported, and
may be used along with Agilent IO libraries.

Figure 1. M9183A soft front panel

Drivers

One notable feature of the soft front panel is the Driver Call
Log which allows the user to see the driver calls for each
button pushed. The user can then incorporate the driver
calls into the their application program – enabling fast and
easy program development.

Agilent’s digital multimeters come complete with software
drivers for Windows® XP, Windows Vista, and
Windows 7 (32 and 64 bit). These software drivers work
in the most popular test and measurement development
environments including: LabVIEW, Visual Studio® (C, C++,
C#, Visual Basic), MATLAB, and VEE.

Calibration
Each M9182A and M9183A DMM is factory calibrated
and shipped with an ISO-9002, NIST-traceable calibration
certificate.
Calibration is required once per year. A documented
calibration process allows you to do in-rack calibration
using standard calibration sources. Alternatively, Agilent
and 3rd party calibration labs offer calibration services for
the M9182A and M9183A DMMs.

3

T E CH NIC A L S P EC IFIC A T I O N S A N D C H A R A C T E R I S T I C S
M9182A and M9183A: Accuracy specifications ±(% of reading + % of range) 1,2
Function

Range 3

DC voltage

True RMS,
AC voltage 4,5
(Fast RMS off)

True RMS,
AC voltage4,5
(Fast RMS on)

Frequency

24 hour
23 °C ± 1 °C

90 day
23 °C ± 5 °C

1 year
23 °C ± 5 °C

200.0000 mV
2.000000 V
20.00000 V
200.0000 V
300.0000 V

0.0030 + 0.0005
0.0020 + 0.0002
0.0040 + 0.0006
0.0030 + 0.0002
0.0130 + 0.0002

0.0040 + 0.0008
0.0030 + 0.0002
0.0050 + 0.0007
0.0040 + 0.0001
0.0230 + 0.0003

0.0050 + 0.0010
0.0040 + 0.0003
0.0070 + 0.0008
0.0050 + 0.0003
0.0250 + 0.0003

200.0000 mV6 10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz - 100 kHz

3.00 + 0.18
0.37 + 0.08
0.13 + 0.05
0.25 + 0.08
1.90 + 0.18

3.10 + 0.19
0.38 + 0.09
0.14 + 0.06
0.26 + 0.10
1.95 + 0.19

3.20 + 0.22
0.40 + 0.10
0.15 + 0.06
0.27 + 0.12
2.00 + 0.20

2.000000 V

10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz - 100 kHz

3.00 + 0.10
0.37 + 0.07
0.05 + 0.05
0.32 + 0.06
1.90 + 0.08

3.10 + 0.11
0.38 + 0.08
0.06 + 0.06
0.33 + 0.07
2.00 + 0.09

3.20 + 0.13
0.40 + 0.09
0.07 + 0.06
0.35 + 0.08
2.10 + 0.10

20.00000 V

10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz - 100 kHz

3.00 + 0.07
0.37 + 0.06
0.06 + 0.05
0.18 + 0.09
1.30 + 0.15

3.10 + 0.08
0.38 + 0.07
0.07 + 0.06
0.20 + 0.11
1.40 + 0.18

3.30 + 0.10
0.40 + 0.08
0.07 + 0.07
0.22 + 0.13
1.50 + 0.20

200.0000 V
& 300.0000 V

10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz - 100 kHz

3.00 + 0.07
0.43 + 0.06
0.07 + 0.05
0.28 + 0.07
1.30 + 0.09

3.10 + 0.08
0.44 + 0.07
0.08 + 0.07
0.30 + 0.08
1.60 + 0.12

3.30 + 0.08
0.45 + 0.08
0.09 + 0.08
0.32 + 0.10
2.40 + 0.13

200.0000 mV6 350 Hz - 800 Hz
800 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz - 100 kHz

0.60 + 0.08
0.13 + 0.05
0.55 + 0.08
5.30 + 0.18

0.65 + 0.09
0.14 + 0.06
0.60 + 0.10
5.40 + 0.19

0.70 + 0.10
0.15 + 0.06
0.63 + 0.12
5.60 + 0.20

2.000000 V

350 Hz - 800 Hz
800 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz - 100 kHz

0.93 + 0.07
0.07 + 0.05
0.62 + 0.06
5.10 + 0.08

0.96 + 0.08
0.08 + 0.06
0.65 + 0.07
5.20 + 0.09

1.00 + 0.09
0.08 + 0.06
0.70 + 0.08
5.30 + 0.10

20.00000 V

350 Hz - 800 Hz
800 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz - 100 kHz

0.93 + 0.06
0.07 + 0.05
0.31 + 0.09
2.00 + 0.15

0.96 + 0.07
0.07 + 0.06
0.33 + 0.11
2.20 + 0.18

1.00 + 0.08
0.07 + 0.07
0.35 + 0.13
2.40 + 0.20

200.0000 V
& 300.0000 V

350 Hz - 800 Hz
800 Hz - 10 kHz
10 kHz - 50 kHz
50 kHz - 100 kHz

1.00 + 0.06
0.07 + 0.05
0.34 + 0.07
2.50 + 0.09

1.10 + 0.07
0.07 + 0.07
0.45 + 0.08
2.80 + 0.12

1.10 + 0.08
0.08 + 0.08
0.50 + 0.10
3.20 + 0.13

4

T E CH NIC A L S P EC IFIC A T I O N S A N D C H A R A C T E R I S T I C S
M9182A and M9183A: Accuracy specifications ±(% of reading + % of range) 1,2
Function

Range 3

Frequency,
test current or
burden voltage

24 hour
23 °C ± 1 °C

90 day
23 °C ± 5 °C

1 year
23 °C ± 5 °C

Resistance 7

20.00000 Ω
(M9183A only)
200.0000 Ω
2.000000 kΩ
20.00000 kΩ
200.0000 kΩ
2.000000 MΩ
20.00000 MΩ
200.0000 MΩ
(M9183A , 2-wire only)

10 mA

0.004 + 0.004

0.009 + 0.004

0.014 + 0.005

1 mA
1 mA
100 µA
10 µA
1 µA
100 nA
4 nA

0.004 + 0.002
0.003 + 0.002
0.003 + 0.002
0.006 + 0.002
0.018 + 0.002
0.120 + 0.002
0.800 + 0.013

0.010 + 0.002
0.008 + 0.002
0.008 + 0.002
0.010 + 0.002
0.030 + 0.003
0.130 + 0.003
1.000 + 0.015

0.013 + 0.003
0.012 + 0.002
0.012 + 0.002
0.016 + 0.003
0.040 + 0.004
0.200 + 0.003
1.300 + 0.025

200.0000 nA
(M9183A only)
2.000000 µA
(M9183A only)
20.00000 µA
(M9183A only)
200.0000 µA
(M9183A only)
2.000000 mA
20.00000 mA
200.0000 mA
2.000000 A

< 100 µV

0.130 + 0.020

0.160 + 0.023

0.170 + 0.030

< 100 µV

0.050 + 0.004

0.080 + 0.003

0.210 + 0.008

< 100 µV

0.050 + 0.002

0.080 + 0.003

0.130 + 0.004

< 2.5 mV

0.052 + 0.100

0.070 + 0.150

0.100 + 0.200

< 25 mV
< 250 mV
< 55 mV
< 520 mV

0.020 + 0.015
0.020 + 0.002
0.020 + 0.025
0.100 + 0.003

0.030 + 0.020
0.035 + 0.003
0.030 + 0.030
0.150 + 0.004

0.040 + 0.028
0.045 + 0.003
0.040 + 0.040
0.200 + 0.005

2.000000 mA 9

10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 1 kHz
1 kHz - 10 kHz

2.70 + 0.20
0.90 + 0.20
0.04 + 0.08
0.12 + 0.20

2.90 + 0.20
0.90 + 0.20
0.08 + 0.15
0.14 + 0.20

2.90 + 0.20
1.00 + 0.20
0.12 + 0.20
0.22 + 0.20

20.00000 mA

10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 1 kHz
1 kHz - 10 kHz

1.80 + 0.15
0.60 + 0.15
0.07 + 0.05
0.21 + 0.15

2.60 + 0.15
0.90 + 0.15
0.15 + 0.10
0.30 + 0.20

2.80 + 0.15
1.00 + 0.15
0.16 + 0.15
0.40 + 0.20

200.0000 mA

10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 1 kHz
1 kHz - 10 kHz

1.80 + 0.20
0.60 + 0.20
0.15 + 0.08
0.30 + 0.15

2.70 + 0.20
0.90 + 0.20
0.17 + 0.09
0.35 + 0.18

2.80 + 0.20
1.00 + 0.20
0.20 + 0.11
0.40 + 0.20

2.000000 A

10 Hz - 20 Hz
20 Hz - 47 Hz
47 Hz - 1 kHz
1 kHz - 10 kHz

1.80 + 0.20
0.66 + 0.30
0.30 + 0.19
0.40 + 0.20

2.50 + 0.23
0.80 + 0.30
0.33 + 0.19
0.45 + 0.23

2.70 + 0.25
0.90 + 0.30
0.35 + 0.20
0.50 + 0.25

200 mV to 300 V

1Hz - 20 Hz
20 Hz - 130 Hz
130 Hz - 640 Hz
640 Hz - 2.5 kHz
2.5 kHz - 40 kHz
40 kHz - 200 kHz
200 kHz - 300 kHz

0.08 + 0.01
0.03 + 0.01
0.03 + 0.01
0.03 + 0.01
0.03 + 0.01
0.05 + 0.01
0.07 + 0.01

0.08 + 0.01
0.03 + 0.01
0.03 + 0.01
0.03 + 0.01
0.03 + 0.01
0.05 + 0.01
0.07 + 0.01

0.08 + 0.01
0.03 + 0.01
0.03 + 0.01
0.03 + 0.01
0.03 + 0.01
0.05 + 0.01
0.07 + 0.01

DC current

True RMS,
AC current 8

Frequency or
period 10, 14

5

T E CH NIC A L S P EC IFIC A T I O N S A N D C H A R A C T E R I S T I C S
M9182A and M9183A: Accuracy specifications ±(% of reading + % of range) 1,2
Function

Range

Full scale reading 24 hour
or resolution
23 °C ± 1 °C

90 day
23 °C ± 5 °C

1 year
23 °C ± 5 °C

Single shot
duty cycle 11
[M9183A only]

2 - 100 Hz
100 Hz - 1 kHz
1 - 10 kHz

0.02 %
0.20 %
2.00 %

0.03 ± 0.03
0.03 ± 0.30
0.03 ± 3.00

0.03 ± 0.03
0.03 ± 0.30
0.03 ± 3.00

0.03 ± 0.03
0.03 ± 0.30
0.03 ± 3.00

Single shot
pulse width 12, 14
[M9183A only]

14 μs - 62.5 ms

1 µs

.01 ± 4 µs

01 ± 4 µs

01 ± 4 µs

Capacitance 13

1000.0 pF
10.000 nF
100.00 nF
1.0000 µF
10.000 µF
100.00 µF
1.0000 mF
10.000 mF

1199.9 pF
11.999 nF
119.99 nF
1.1999 µF
11.999 µF
119.99 µF
1.1999 mF
11.999 mF

1.00 + 0.10
1.20 + 0.05
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.20 + 0.10
2.00 + 0.10

1.00 + 0.10
1.20 + 0.05
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.20 + 0.10
2.00 + 0.10

1.00 + 0.10
1.20 + 0.05
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.00 + 0.10
1.20 + 0.10
2.00 + 0.10

[M9183A and
M9182A]

1. Specifications are for 1 hour warm up, within 1 hour self-cal, aperture ≥ 0.5 sec. Slow AC filter for AC measurements only.
2. For temperatures outside the range of 23 °C ± 5 °C, but within 0 °C to 50 °C, add 0.1 × accuracy specification per °C.
3. 20% over range on all ranges except 300 V range, 10% over range for 300 V range.
4. Minimum input specified: 5 mV or 1% of range, whichever is larger.
5. Signal is limited to 8x106 Volt Hz product. For example, at 32 kHz, the highest input is 250 V.
6. For inputs from 5 mV to 15 mV, add 100 µV to the specification.
7. Specifications are for 4-wire resistance measurements, for 2-wire, add 1 mΩ additional error to the specification; for offset
compensated ohms (M9183A only), add 0.02% of range.
8. Minimum input specified: 60 µA or 1.5% of range, whichever is larger; for inputs < 5% of full scale, add 0.02% of range.
9. For inputs from 60 to 120 µA, add 10 µA to the specification; for inputs < 5% of full scale, add 0.1% of range.
10. Minimum amplitude greater of: 100 mV, or 5 % of range for 1 Hz to 2.5 kHz, or 25 % of range for 2.5 kHz to 200 kHz , or 40% of range
for 200 kHz to 300 kHz..
11. Specifications are % of reading (0.03) ± adder.
12. Specifications are % of reading + time.
13. Specifications apply to input signals ≥ 5% of range.
14. Maximum wait time for duty cycle and period is 5 seconds.

Defin itions f o r s p e c i f i c a t i o ns
Specification (spec): Represents warranted performance of a calibrated instrument that has been stored for a minimum
of two hours within the operating temperature range of 0 to 55 °C, unless otherwise stated, and after a one hour warm-up
period. The specifications include measurement uncertainty. Data represented in this document are specifications unless
otherwise noted.
Typical (typ): Represents characteristic performance, which 80% of the instruments manufactured will meet. This data is
not warranted, does not include measurement uncertainty, and is valid only at room temperature (approximately 25 °C).
Nominal (nom): The expected mean or average performance, or an attribute whose performance is by design, such as the
50 Ω connector. This data is not warranted and is measured at room temperature (approximately 25 °C).
Measured (meas): An attribute measured during the design phase for purposes of communicating expected performance,
such as amplitude drift vs. time. This data is not warranted and is measured at room temperature (approximately 25 °C).
Note: All graphs contain measured data from several units at room temperature unless otherwise noted.

6

T E CH NIC A L S P EC IFIC A T I O N S A N D C H A R A C T E R I S T I C S
M9182A and M9183A Sensitivity (nom)
Function

Lowest Range

Sensitivity

DCV

200.0000 mV

0.1 µV

ACV

200.0000 mV

0.1 µV

Resistance (M9183A)

20.00000 Ω

10 µΩ

Resistance (M9182A)

200.0000 Ω

100 µΩ

DCI (M9183A)

200.0000 nA

0.1 pA

DCI (M9182A)

2.000000 mA

10 nA

ACI

2.000000 mA

1 nA

Capacitance

1000.0 pF

0.1 pF

M9182A and M9183A temperature accuracy (spec) 1
Temperature
function

Type

R0 (Ω)

Sensitivity

Range/max
temperature

1 year
23 °C ± 5 °C

RTD temperature
measurement 2,3

pt385

100 Ω, 200 Ω

0.01 °C

-150 to 650 °C

± 0.06 °C

500 Ω, 1 kΩ

0.01 °C

-150 to 650 °C

± 0.03 °C

Less than 12 Ω

0.01 °C

-100 to 200 °C

± 0.18 °C at ≤ 20 °C
± 0.05 °C otherwise

Higher than 90 Ω 0.01 °C

-100 to 200 °C

± 0.10 °C at ≤ 20 °C
± 0.05 °C otherwise

B

NA

0.01 °C

2200 °C

± 0.38 °C

E

NA

0.01 °C

1200 °C

± 0.035 °C

J

NA

0.01 °C

2000 °C

± 0.06 °C

K

NA

0.01 °C

3000 °C

± 0.07 °C

N

NA

0.01 °C

3000 °C

± 0.10 °C

R

NA

0.01 °C

2700 °C

± 0.25 °C

S

NA

0.01 °C

3500 °C

± 0.35 °C

T

NA

0.01 °C

550 °C

± 0.06 °C

2.25 kΩ

NA

0.01 °C

-80 to 150 °C

± 0.1 °C

5 kΩ

NA

0.01 °C

-80 to 150 °C

± 0.1 °C

10 kΩ

NA

0.01 °C

-80 to 150 °C

± 0.1 °C

Cu (Copper)

Thermocouple
temperature
measurement 4,5

Thermistor

3

1. Specifications are for one hour warm up, within one hour self-cal, aperture ≥ 0.5 sec.
2. 4-wire RTD measurement, R0 variable 1 Ω to 7 k Ω.
3. For total measurement accuracy, add temperature probe error.
4. For total measurement accuracy, add thermocouple error and cold junction compensation.
5. DMM linearization temperature range may be greater than that of the thermocouple device.

7

T E CH NIC A L S P EC IFIC A T I O N S A N D C H A R A C T E R I S T I C S
Source-Measure [(spec) unless otherwise stated)]
M9183A source DC voltage, measure DC voltage
Parameter

Closed Loop

Open Loop

DC voltage source (output) range

+10 to -10 V

+10 to -10 V

DC source sink current at 5 V output

5 mA

5 mA

DAC resolution (nom)

18 bits /12 bits

12 bits

DC voltage source accuracy 1 Year, 23°± 5° C

0.01% ± 0.001%

0.5% ± 0.2%

Settling time (typ)

100 ms/1 ms

1 ms

Source resistance (nom)

200 ohms

200 ohms

M9183A source DC voltage, measure DC current
DC voltage source (output) range

-10.000 to +10.000 V

DC current measurement range

0 to ± 20 mA

Voltage resolution (nom)

5 mV

Voltage source accuracy 1 Year, (23 °C ± 5 °C)

1,2,3,4

2.0% ± 0.4%

Settling time (typ)

100 ms

DC current measurement accuracy

0.1% + 0.005%

M9183A source DC current, measure DC voltage
DC voltage measurement range

0 to ± 2.0 V

Current output

Compliance voltage

Minimum Level

Source Accuracy
1 year, (23 °C ± 5 °C) 1,2,3

< 1.25 µA

4.2 V

10 nA

1% + 1%

< 12.5 µA

4.2 V

50 nA

1% + 1%

< 125 µA

4.2 V

100 nA

1% + 0.5%

< 1.25 mA

4.2 V

1 µA

1% + 0.5%

< 12.5 mA

1.2 V

10 µA

1% + 0.5%

1. Specifications are for one hour warm up, within one hour self-cal, slow AC filter.
2. For temperatures outside the range of 23 °C ± 5 °C, but within 0 °C to 50 °C, add 0.1 × accuracy specification per °C.
3. Repetitive reading at an aperture of 133 ms or higher.
4. If DCV source > 6 V, then add current * 4 ohms; if DCV source > 8 V, then add current * 30 ohms.
8

T E CH NIC A L S P EC IFIC A T I O N S A N D C H A R A C T E R I S T I C S
Triggering Characteristics
The M9182A and M9183A have advanced triggering capabilities that exceed those found on other digital
multimeters. Advanced triggering enables you to obtain the signal you need and accurately measure it, in a variety
of applications.

External hardware trigger
Trigger input voltage level range (at DIN 7 connector)

+3 to +15 V activates the trigger

Minimum trigger pulse width

Aperture + 50 µs

Trigger input impedance

3 kΩ

Edge

Selectable positive or negative edge

PXI bus trigger inputs
Trigger input voltage level range (via PXI backplane)

CMOS level (see PXI standard)

Minimum trigger pulse width

Aperture + 50 µs

Edge

Selectable positive or negative edge

Trigger features
Trigger sources

Trigger delay
[Auto delay (default delay) ensures 1st
reading accuracy in most configurations)

Immediate, PXI trigger , external DIN
connector, analog threshold trigger
source
Measurement delay

50 µs to 15 s

Resolution

1 µs to 65 ms and 16 µs above 65 ms

Reading storage
Multi Sample Mode

Circular buffer - 80 readings
Aperture range

(DCV and DCI functions only)

2.5 µs to 160 ms (M9183A)
130 µs to 160 ms (M9182A)

Maximum read interval range

1sec for aperatures ≥ 625 µs, else 65 ms

Reading per trigger

Up to 80 readings (maximum of 78
pre-triggers or a maximum of 80 post
triggers)

Trigger sources

PXI trigger, external DIN connector, and
analog threshold (accuracy within 5% of
range)

9

M EAS U R EM E N T CH A RA CT E R I S TICS
Resolution vs. Aperture and Reading Rate for DCV, DCI, Ω
Measurement
aperture

Maximum
readings per
second

Resolution

10 ms

98

6½ digits (22 bits)

625 µs

1,200

5½ digits (18 bits)

14,000

130 µs

4,500

4½ digits (14 bits)

10,000

2.5 µs

15,000
4½ digits (14 bits)
(M9183A only) 1

8,000

Aperture reading rate
(< 1 NPLC)
16,000
12,000

6,000
4,000
2,000

1. 15,000 readings/second represent a typical maximum with a
measurement aperture of 2.5 µs. Results will vary, depending on
what PC hardware, PXI hardware, and driver are used.

10 mS

625 µS

130 µS
M9182A

2.5 µS
M9183A

Transaction Speed
Transactional I/O speed is a single reading measurement.
This is important when you are taking many single
measurements with the DMM. The M9183A delivers the
highest transactional measurement speed in its class.
These fast readings, up to 15,000 readings per second with
a read interval rate of 66 µs, provides the lowest latency,
translating into higher test-system throughput and lower
cost of test per unit tested.

Variable delay can be programmed to allow fully settled
readings in most configurations.

Time frame of a single measurement
Minimum read interval = 66 µs (M9183A)
Minimum read interval = 222 μs (M9182A)

System Reading and Throughput Rates
Switching
ranges within
a function

Aperture (A)

Range change
time (ms)

Switch
between
functions

Aperture (A)

Function
change time
(ms)

DCV

A ≤ 20 ms

(A × 0.2) + 15

DCV

A < 16 ms

15.6

A > 20 ms

A + 15.6

A ≥ 16 ms

A + 25

Resistance (2wire or 4-wire)

A < 33 ms

(A × 0.05) + 15.5

A < 16.66 ms

7.8

A ≥ 33 ms

A + 13

DCI (200 mA or
2 A to any other
range)

A ≤ 40 ms

4.2

40 ms < A < 66.66 ms 7.8

A > 40 ms

15.7

A ≥ 66.66 ms

(A × 0.51) + 45

DCI (all other
ranges)

All apertures

DCV to
capacitance

A < 33.33 ms

23.4

A ≥ 33.33 ms

(A × 0.65) + 50

Capacitance

All apertures

Resistance to
capacitance

A ≤ 33.33 ms

23.4

Resistance to
DCI

1
12

10

16.66 ms ≤ A < 40 ms A × 0.65

33.33 ms < A < 80 ms (A × 2) + 35
80 ms ≤ A < 160 ms

23.4

A ≥ 160 ms

160

M EAS U R EM E N T CH A RA CT E R I S TICS
DC voltage
Measurement method

Delta-sigma A/D conversion

Input resistance

200 mV, 2.0 V ranges: >10 GΩ with typical leakage of < 50 pA;
20 V, 200 V, 300 V ranges: 10.0 MΩ

Input isolation

330 VDC, 250 VAC from Earth ground

Input overvoltage protection

330 VDC all ranges

DCV noise rejection

Normal mode rejection at 50, 60, or 400 Hz ± 0.5%; > 95 dB (apertures ≥
0.160 s); CMRR (1 kΩ lead imbalance) ≥ 120 dB

True RMS AC voltage
Measurement method

AC coupled (10 Hz to 100 kHz) true RMS — measures the AC component of
an input waveform that consists of AC and DC components.

Crest factor

Maximum crest factor of 4 at full scale, 7 at 10% of range

Input impedance

1 MΩ, in parallel with < 300 pF

Settling time

< 0.5 sec to within 0.1% of final value
Fast RMS: < 0.05 sec to within 0.1% of final value

Peak input

8 x 106 volt Hz product (example: 250 V @ 32 kHz)

Input overvoltage protection

330 VAC all ranges

ACV noise rejection

Common mode rejection at 50 Hz or 60 Hz; 1 kΩ imbalance in either lead
> 60 dB

Resistance
Measurement method

Selectable 2-wire or 4-wire. Current source referenced to LO output

Offset compensation (M9183A only)

All ranges, use with apertures > 5 ms
Voffset + (I*R) < 2.2 V for ranges ≥ 2 kΩ
Voffset + (I*R) < 0.22 V for ranges < 2 kΩ

Maximum test voltage

240 mV for 20 Ω and 200 Ω ranges; 2.4 V for 20 kΩ to 20 MΩ ranges;
1.0 V for 200 MΩ range (M9183A only)

Maximum lead resistance (4-wire)

50 kΩ for 200 kΩ, 2.0 MΩ, and 20 MΩ ranges; 5 kΩ for 20 kΩ range
500 Ω for 200 Ω and 2 kΩ ranges; 50 Ω for 20 Ω range

Input protection

330 V on all ranges

DC current
Shunt resistance

10 Ω for 2 mA and 20 mA, 0.1 Ω for 200 mΩ and 2 A;
Virtual zero shunt for 200 μA, 20 μA, 2 μA, and 200 nA range (M9183A only)

Input protection

Protected with 2.5 A, 250 V fast blow fuse

True RMS AC current
Measurement method

AC coupled true RMS measurement (measures the AC component only.)
analog RMS DC converter.

Shunt resistance

10 Ω for 2 mA and 20 mA, 0.1 Ω for 200 mA and 2 A

Input protection

Protected with 2.5 A, 250 V fast blow fuse

11

M EAS U R EM E N T CH A RA CT E R I S TICS
Frequency and period
Measurement method

Direct (conventional) counting

Input impedance

1 MΩ with < 300 pF

Sensitivity (130 Hz)

.001 Hz

Totalizer (M9183A only)
Active edge polarity

Positive or negative transition

Maximum count

10,000,000,000

Allowed rate

1 to 30,000 events per second

Threshold

Set threshold DAC

Accuracy

±2 counts

Capacitance
Measurement method

Differential charge balance: variable currents used to stimulate dV/dt
response.

Connection type

2-wire

Environmental and physical characteristics
Temperature range

Operating -10° to 55 °C
Non-operating -40 ° to +85 °C

Relative humidity

Operating to 80% at 40 °C
Storage to 95% at 40 °C

Connectors

V HI*, 2-wire Ω IN, DCV OUT
V LO*, 2-wire Ω IN, DCV OUT
I HI*, 4-wire Ω IN
I LO*, 4-wire Ω IN
Sync OUT
External Trigger IN
Trigger and Sync common

Safety

Complies with IEC 61010-1, CAT II 300 V, pollution degree 2

EMC

Complies with EN61326-1 Industrial Environment

Warm-up time

1 hour

Sheathed banana jack
Sheathed banana jack
Sheathed banana jack
Sheathed banana jack
DIN 7, pin 2
DIN 7, pin 7
DIN 7, pin 4

* Do not connect the V HI and V LO or I HI and I LO connectors to anything when the measurement function does not utilize those
connectors.

Physical characteristics

Power dissipation:

Dimensions

3U/1-slot PXI/CompactPCI standard

+5 V

Total power

Weight

0.5 kg (1 lb.)

300 mA

1.5 W max

12

C O N FI G U RA T I O N
Hardware

Calibration and Warranty

Model

Description

Advantage services: calibration and warranty

M9182A

PXI 6½ digit multimeter, high performance

M9183A

PXI 6½ digit multimeter, enhanced
performance

Agilent Advantage Services is committed to your
success throughout your equipment’s lifetime.

DMM units
include:

Getting started guide, software drivers,
user and service documentation (on CD
ROM), Agilent I/O libraries

Calibration

Software
Model

Description

Supported
Operating
Systems

Microsoft Windows XP (32-bit),
Microsoft Windows Vista (32/64-bit)
Microsoft Windows 7 (32/64-bit)

Standard
Compliant
Drivers

IVI-C, IVI-COM, LabVIEW

Yearly calibration, for 3 years

R-50C-011-5

Yearly calibration, for 5 years

R-50C-021-3

Yearly ANSI Z540-1-1994 calibration,
for 3 years

R-50C-021-5

Yearly ANSI Z540-1-1994 calibration,
for 5 years

Warranty
Standard warranty is 1 year
R-51B-001-3C 1 year return-to-Agilent warranty
extended to 3 years

Supported
VisualStudio (VB.NET, C#, C/C++)®,
Application
LabVIEW, MATLAB, VEE
Development
Environments
(ADE)
Agilent IO
Libraries

R-50C-011-3

R-51B-001-5C 1 year return-to-Agilent warranty
extended to 5 years

Accessories

Includes: VISA Libraries, Agilent
Connection Expert, IO Monitor

Model

Description

34138A

Test lead set

Related products

13

Model

Description

M9018A

18-slot PXIe chassis

M9021A

PCIe cable interface to an external
system controller

M9036A

Embedded controller

M9101A

PXI high-density multiplexer,
64 channels, reed relays

M9103A

PXI high-density multiplexer,
99 channels, armature relays

M9120A

PXI high-density matrix switch, 4x32,
armature relays

M9121A

PXI high-density matrix switch, 4x64,
reed relays

M9181A

PXI digital multimeter, 6½ digit, basic
features

T h e Mo d u l ar Tang r am
The four-sided geometric symbol that appears throughout this
document is called a tangram. This seven-piece puzzle originated
in China a few centuries ago. The goal is to create shapes—from
simple to complex—that form an identifiable silhouette. As with
a tangram, the possibilities may seem infinite as you begin to
create a new test system. With a set of clearly defined elements—
architecture, hardware, software—Agilent can help you create the
system you need, from simple to complex.

Discover the Alternatives …
... Agilent Modular Products

Agilent Advantage Services is committed to your
success throughout your equipment’s lifetime. We share
measurement and service expertise to help you create the
products that change our world. To keep you competitive,
we continually invest in tools and processes that speed up
calibration and repair, reduce your cost of ownership, and
move us ahead of your development curve.
www.agilent.com/find/advantageservices

www.agilent.com/quality
PXI is a registered US trademark of PXI Systems Alliance. Visual Studio,
Visual C, C++, and C#, and Visual Basic are registered trademarks of
Microsoft Corporation. Windows and MS Windows are U.S. registered
trademarks of Microsoft Corporation. MATLAB is a U.S. registered
trademark of The Math Works, Inc.
Product specifications and descriptions in this document subject to change
without notice.
© Agilent Technologies, Inc. 2012 Printed in USA, February 17, 2012
5990-7580EN

w w w .ag i l ent.co m
www.agilent.com/find/modular
www.agilent.com/find/pxi-switch
For more information on Agilent Technologies’ products, applications
or services, please contact your local Agilent office. The complete list
is available at: www.agilent.com/find/contactus

Americas
Canada
Brazil
Mexico
United States

(877) 894 4414
(11) 4197 3600
01800 5064 800
(800) 829 4444

Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand

1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008

Europe & Middle East
Austria
Belgium
Denmark
Finland
France
Germany
Ireland
Israel
Italy
Netherlands
Spain
Sweden
Switzerland
United Kingdom

43 (0) 1 360 277 1571
32 (0) 2 404 93 40

45 45 80 12 15
358 (0) 10 855 2100
0825 010 700*
*0.125 €/minute
49 (0) 7031 464 6333
1890 924 204
972-3-9288-504/544
39 02 92 60 8484
31 (0) 20 547 2111
34 (91) 631 3300
0200-88 22 55
0800 80 53 53

44 (0) 118 927 6201

For other unlisted Countries: www.agilent.com/find/contactus
Revised: January 6, 2012



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : Yes
XMP Toolkit                     : 3.1-701
Producer                        : Acrobat Distiller 7.0 (Windows)
Create Date                     : 2012:02:17 12:31:30-08:00
Creator Tool                    : Adobe InDesign CS5 (7.0)
Modify Date                     : 2012:02:17 12:31:30-08:00
Format                          : application/pdf
Title                           : 5990-7580_pxi-dm_2-17-12_d3a.indd
Creator                         : jfox
Document ID                     : uuid:cfb230ea-608f-4058-8818-be8518510ddd
Instance ID                     : uuid:49481157-121a-4245-9e33-21a1a33600e0
Page Count                      : 14
Author                          : jfox
EXIF Metadata provided by EXIF.tools