Cover Ellipse Service Manual Rev.04
User Manual:
Open the PDF directly: View PDF .
Page Count: 229
Download | |
Open PDF In Browser | View PDF |
ELLIPSE Service Manual Via E. Barsanti 17/A 00012 Guidonia – ROME (ITALY) www.ams-analyzers.com CONFIGURATION SHEET CHAPTER DESCRIPTION REVISION INDEX TABLE OF CONTENTS 03 01 INTRODUCTION 03 02 SYSTEM DESCRIPTION 03 03 INSTALLATION 03 04 ELECTRICAL SCHEMES AND DRAWINGS 03 05 DIAGNOSTIC PROGRAM 03 06 SETTINGS AND ADJUSTMENTS 03 07 MAINTENANCE 03 08 HOST COMMUNICATION 04 09 ERROR SIGNALING AND TROUBLESHOOTING 03 Ellipse Service Manual Rev.04 01 October 2009 Page 1 INDEX TABLE OF CONTENTS Chapter 01 INTRODUCTION Chapter 02 SYSTEM DESCRIPTION Chapter 03 INSTALLATION Chapter 04 ELECTRICAL SCHEMES AND DRAWINGS Chapter 05 DIAGNOSTIC PROGRAM Chapter 06 SETTINGS AND ADJUSTMENTS Chapter 07 MAINTENANCE Chapter 08 HOST COMMUNICATION Chapter 09 ERROR SIGNALING AND TROUBLESHOOTING Ellipse Service Manual Rev.03 01 June 2009 Page 1 Chapter 01 – INTRODUCTION CHAPTER 01 - INTRODUCTION INDEX 1 INTRODUCTION ................................................................................................................. 1 1.1 THE AIM OF THE TECHNICAL MANUAL ………………………………………...………..…..2 1.1.2 MANUAL OUTLINE……………………………..……………………..……………..………………....2 1.2 SYSTEM INTRODUCTION ................................................................................................ 3 1.3 PRECAUTIONARY MEASURES ....................................................................................... 3 1.3.1 CHEMICAL RISKS .............................................................................................................. 5 1.3.2 ELECTRICAL RISKS........................................................................................................... 5 1.3.3 MECHANICAL RISKS......................................................................................................... 4 1.4 GUARANTEE ...................................................................................................................... 5 1.5 TECHNICAL OPERATING FEATURES………………………………………….…. ..... 6 1.5.1 COMPUTER & SOFTWARE FEATURES ........................................................................ .9 1.5.2 OPTIONS............................................................................................................................... 8 1.5.3 DIMENSIOS, WEIGHT & ENVIRONMENT.................................................................... 10 1.5.4 INSTALLATION REQUIREMENTS................................................................................ 10 Ellipse Service Manual Rev. 03 01June 2009 Page 1 Chapter 01 – INTRODUCTION 1 INTRODUCTION 1.1 THE AIM OF THE TECHNICAL MANUAL This manual has been written in order to supply the technical staff, the persons who are responsible for the maintenance and for resolving instrument failures, a complete and detailed guide of the Ellipse analyzer, in accordance with the standard UNI EN591 (which requires a manual to be supplied with vitro diagnostic instruments for professional use). 1.1.2 MANUAL OUTLINE The technical manual is composed of 10 chapters that not only describe the operative technical characteristics of the system, but also the reparation and maintenance procedures to use for the modules that they are composed of. . Chapter 1 INTRODUCTION The manual structure is described, and recommendations are given regarding the general use of the analyzer. The technical characteristics of the operative system are given in this chapter. . Chapter 2 DESCRIPTION OF THE SYSTEM Describes the system, in particular, the analytical cycle of the single modules and electronic boards. . Chapter 3 INSTALLATION Describes the unpacking procedures and the required characteristics for the place of installation. . Chapter 4 ELECTRONICS SCHEMES AND DRAWINGS Gives the electronics schemes and assembly drawings of the electronics board and the assembly drawings of the modules that the system is composed of. Ellipse Service Manual Rev. 03 01June 2009 Page 2 Chapter 01 – INTRODUCTION . Chapter 5 DIAGNOSTIC PROGRAM Describes the folders in the Diagnostic Program that are responsible for the checks and the calibration of the modules in the system. . Chapter 6 SETTINGS AND ADJUSTMENTS Gives the procedures for replacement, adjustment and check of the system modules. . Chapter 7 MANTENANCE Describes the routine maintenance that needs to be done on a regular basis in order to guarantee the correct functioning of the analyzer. . Chapter 8 COMUNICATION WITH HOST Describes the communication protocol between the analyzer and the Host computer. • Chapter 9 ERROR CODES AND GUIDE TO RESOLVE ANOMALIES Lists the system’s error codes, describes the errors in the results and gives a guide for how to implement corrective actions. The technical staff responsible for resolving the structural anomalies is highly suggested to thoroughly examine the contents of this manual before operating the system. Ellipse Service Manual Rev. 03 01June 2009 Page 3 Chapter 01 – INTRODUCTION 1.2 SYSTEM INTRODUCTION The “Ellipse” system is a continuous-loading, random access, bench top instrument for performing chemical and Immunoturbidimetric clinical analysis. It is totally automatic and computer controlled. Via E. Barsanti 17/A 00012 Guidonia – ROME (ITALY) Ellipse Service Manual Rev. 03 01June 2009 Page 4 Chapter 01 – INTRODUCTION 1.3 PRECAUTIONARY MEASURES 1.3.1 CHEMICAL RISKS The individual operator is responsible for assuring that all possible precautionary measures are taken against eventual risks associated with the use of the “Ellipse” instrument in clinical laboratory settings. The manufacturer will provide the reagents kit and specific written information on the use of each of the reagents. It is important that the samples be well coagulated and then carefully centrifuged. Samples which contain fibrinogen clots can obstruct the probe and lead to inexact sampling. If blood samples containing gel are used, it is suggested that the manufacturer’s recommendations be followed. Immediately clean and remove any accidental leakage of reagent or other liquid. 1.3.2 ELECTRICAL RISKS As with any electrical device, the risk of electric shock exists. Is therefore necessary to take every precautionary measure possible when working with this, or any other, electrical instrument to avoid contact with power supply wires, electrical components or electronic boards. 1.3.3 MECHANICAL RISKS Several precautionary measures should be taken when operating the analyzer: avoid wearing very loose clothing or jewelry that could become tangled in the instrument’s moving parts (e.g. the sample probe); whenever possible, operate the instrument with the main cover panel lowered. WARNING: Never attempt to service or substitute any part(s) of the analyzer when the instrument is turned on. Any and all technical repairs or servicing must be performed by specialized personnel only. Ellipse Service Manual Rev. 03 01June 2009 Page 5 Chapter 01 – INTRODUCTION 1.4 WARRANTY AMS guarantees the replacement of all defective components and/or materials for a period of time not above of 14 months starting from the date of invoicing. Saying guarantee, as well as Technical Assistance, generally is intended furnished as net ex factory Rome. This guarantee does not include consumable and instrument parts in contact with liquids. All components not included in the guarantee are reported in the following table. Besides guarantee does not cover damage caused by : - improper use of the ELLIPSE instrument (or however not according to the Producer or Seller instructions) - bad transport - insufficient (or missing) preventive maintenance by the User In particular any damages due to the transport must be immediately reported to the carrier when he delivers. Ellipse Service Manual Rev. 03 01June 2009 Page 6 Chapter 01 – INTRODUCTION CONSUMABLES AND ACCESSORIES PARTS LIST OUT OF GUARANTEE Description Type Quantity Reagents containers Reagents containers 35 ml 6 ml Samples cups 0.8 ml 12 pieces C101-00190-00 12 " C101-00191-00 1000 " AS-65-0002 Short samples cups Adapter for short samples cups Reaction sectors Washing solution bottle Bottle level sensor Tubing Kit for peristaltic pump Tubing Kit – complete Cleaning solution Rinse solution Sampling probe (internal needle) Complete Sampling probe Drying Pad Halogen Lamp (6 V - 10 W) Interferential filters Kit Fuse 6,3 A-T Inlet/outlet fitting for Rinse & Clean conts Cuvettes protection cover Reagent protection cover Reagent plate Samples rack Washing station, first or second cannulas (A) Washing station, third cannula (B) 1000 " 1 " 6 " 1 " 1 " 2 " 1 " 2 " 1 " 1 " 1 " 1 " 1 " 1 " 10 " 1 " 1 " 1 " 1 " 1 " 1 1 1 AS-65-0100 9-01-0609-00 C101-00217-00 9-35-0041-00 9-05-0078-00 65-01835-00 65-01836-00 ASRN0020 ASRN0021 05-00707-00 10-00703-00 10-01920-00 9-35-0016-00 9-65-0029-00 C130-01238-08 01-01224-00 05-01249-00 10-00584-00 10-00585-00 05-01829-00 05-01633-00 05-01633-01 Washing station, fifth cannula without pad (D) 1 05-01919-00 Diluter Micro-Pump 1 " 05-01710-40 Air Micro-Pump (µP 6) Micro-Pump (µP 2 ÷5) Predilution rack Solenoid Valve –2 way Solenoid Valve –3 way 1 1 1 1 1 " " " " " 05-01711-20 05-01826-16 05-01735-00 9-35-0035-00 9-35-0036-00 Washing station, fourth cannula (C) Ellipse Service Manual Rev. 03 1 ml Code 2 lt 250 ml 50 ml 5x20 05-01638-00 01June 2009 Page 7 Chapter 01 – INTRODUCTION 1.5 TECHNICAL OPERATING FEATURES DESCRIPTION ♦ Fully automatic, random access, continuous loading, benchtop analyzer for clinical chemistry and immunoturbidimetric assays; ASSAY TYPE ♦ End Point, Initial Rate, Kinetic, Bichromatic, Differential; TEST ENTRY MODE ♦ Selective, Batch, Profiles, STAT THROUGHPUT ♦ 138 tests per hour WORKING TEMPERATURE ♦ 37° C ♦ 24 removable containers by 40 ml and 6 ml; ♦ 12 positions for Controls and Standards; ON LINE REAGENTS SAMPLE CONTAINERS SAMPLE LOADING MINIMUM REACTION VOLUME MAXIMUM REACTION VOLUME ♦ Positive identification by a Bar Code Reader ♦ Primary tubes (diameter from 8 to 16 mm; height up to 100 mm) cups from 1 to 4 ml; ♦ 4 racks, each having 8 positions for continuous sample loading ♦ Positive identification by a Bar Code Reader ♦ 220 µl ♦ 550 µl ♦ A single mechanical arm provides all the sampling operations and is equipped with: SAMPLING ARM • Capacitive liquid level sensing • Reagent pre-warming at 37° C • Automatic probe washing ♦ Integrated syringe-free module having the following specifications: DILUTER Sample volume: 2 µl ÷ 99 µl (1 µl incr.) • Reagent 1 volume: 3 µl ÷ 500 µl (1 µl incr.) • Reagent 2 volume: 3 µl ÷ 330 µl (1 µl incr.) • Reagent 3 volume: 3 µl ÷ 330 µl (1 µl incr.) ♦ CV < 1 % at 2 µl PRECISION READING SYSTEM Ellipse • Service Manual ♦ Direct reading Rev. 03 01June 2009 Page 8 Chapter 01 – INTRODUCTION ♦ Photometer: double beam, interferential filters ♦ Wavelength: 8 narrow band interferential filters from 340 nm to 620 nm , plus one available optional filter position OPTIC SYSTEM ♦ Light source: 6V/10 W halogen bulb ♦ Linearity range: up to 3,500 Abs ♦ Resolution: 0.0005 Abs OPTICAL PATH ♦ 6 mm. WASHING STATION ♦ Composed of five probes that empty, wash and dry the reaction cuvettes. ♦ 6 singular replaceable racks with 20 cuvettes each REACTION PLATE ♦ Cuvettes Q.C. continuously computer controlled ♦ Incubation temperature: 37°C 1.5.1. COMPUTER & SOFTWARE FEATURE MINIMUM REQUIREMENTS TYPE ♦ IBM Compatible ♦ Pentium IV 500 MHz, 512 Kb Cache CPU ♦ RAM 256 Mb ♦ Hard Disk 20 Gb MEMORY ♦ Floppy Disk 3 1/2” 1.44 Mb ♦ Colour SVGA 15’’ low radiation MONITOR Resolution 800 x 600 pixels; max number of colors 65536 (16 bit) PRINTER ♦ 80 Columns impact graphic (EPSON LX 300) INTERFACE ♦ One Bi-directional RS 232C serial ports and one parallel (one second serial port for the Host link) ♦ Multitasking WINDOWS XP Home edition SOFTWARE AVAILABLE LANGUAGES ♦ Disable all the energy saving options ♦ Disable the screen saver ♦ Select English “USA” as language, dot as decimal symbol and date and time in Regional setting SETTINGS Ellipse ♦ Italian, English, Chinese, Czech. Software to be released soon in these languages: Russian, Portuguese, French, Polish. Upon request it is possible to release the software in other languages. Service Manual Rev. 03 01June 2009 Page 9 Chapter 01 – INTRODUCTION NOTE: Even though the computers demonstrate the same technical and operative characteristics, some of these could have different hardware installed. This could cause problems for the Ellipse software when running tests (A message appears indicating “Random” error or blocks the program). Therefore, if the PC is bought separately/locally, it is highly recommended to test the system at your offices before preceding with the installation at the final client’s. Consequently, AMS denies any responsibility for software problems that are due to buying the computer separately from the instrument. 1.5.2 OPTIONAL MODULES ♦ POSITIVE BARCODE READER 1.5.3 DIMENSIONS, WEIGHT & OPERATING ENVIRONMENT DIMENSIONS ♦ Height: 53 cm ♦ Depth: 57 cm ♦ Length: 75 cm ♦ 35 Kilos WEIGHT OPERATING ENVIRONMENT ♦ Temperature: 18°C ÷ 30°C. ♦ Relative humidity: 20% ÷ 85% 1.5.4 INSTALLATION REQUIREMENTS ♦ Input Voltage 90 ÷ 250 Vac ♦ Input Frequency: 47 ÷ 63 Hz POWER REQUIREMENTS ♦ Power consumption: ♦ 300 W for the analytical unit ♦ 400 W for the work station ♦ EN 61010-1:1993 +A2:1995 (IN COMPLIANCE WITH THE MAIN SAFETY REGULATIONS EUROPEAN DIRECTIVES 73/23/CEE AND 93/68/EEC REGARDING SAFETY) Ellipse Service Manual Rev. 03 01June 2009 Page 10 Chapter 01 – INTRODUCTION ♦ EMC 89/336/EEC – 92/31/EEC Directives ♦ EN 55011, Class B, Group 1 ♦ EN 50081-1:1992 EMC ELECTROMAGNETIC COMPATIBILITY ♦ EN 55022 ♦ ENV 50140 – ENV 50141 ♦ EN 60601-1-2 ♦ EN 61000-4 Warning: A steady power supply (+ 10%) must be provided for the instrument. If it is not, the manufacturer highly recommends the use of: ♦ UPS Uninterruptible Power Supply ( No-break module) ♦ ELECTRONIC STABILIZER Ellipse Service Manual Rev. 03 01June 2009 Page 11 Chapter 02 – SYSTEM DESCRIPTION CHAPTER 02 – SYSTEM DESCRIPTION INDEX 2 DESCRIPTION OF THE SYSTEM ................................................................................................ 2 2.1 ANALYSES CYCLE ......................................................................................................................... 3 2.1.1 REACTION PLATE ......................................................................................................................... 3 2.1.2 REACTION PLATE CYCLE .......................................................................................................... 3 2.2 SAMPLING ARM - OPERATIONAL SEQUENCE .................................................................... 3 2.3 WASH STATION............................................................................................................................... 6 2.3.1 WASH STATION CYCLE ............................................................................................................... 6 2 DESCRIPTION OF THE SYSTEM .........................................................................................2 2.1 ANALYSES CYCLE................................................................................................................3 2.1.1 REACTION PLATE .................................................................................................................3 2.1.2 REACTION PLATE CYCLE..................................................................................................3 2.2 SAMPLING ARM - OPERATIONAL SEQUENCE...............................................................3 2.3 WASH STATION.....................................................................................................................6 2.3.1 WASH STATION CYCLE.......................................................................................................6 2.4 ELECTRONIC DESCRIPTION...............................................................................................7 2.4.1 INTRODUCTION ....................................................................................................................7 2.4.2 ANALYTICAL CONTROL BOARD (C.P.U.) [P/N: 30-01283-01]......................................7 2.4.3 STEPPER MOTORS DRIVER BOARD [P/N: 30-01284-01]...............................................8 2.4.4 PLATE INTERFACE BOARD (Sx) [P/N: 30-01281-01].......................................................9 2.4.5 ARM INTERFACE BOARD (Dx) [P/N: 30-01282-01] .........................................................9 2.4.6 ELECTROVALVES CONTROL BOARD [P/N: 30-01626-00] ...........................................10 2.4.7 PRE-AMPL/ADC [P/N: 30-00107-00; P/N: 30-00107-03]..................................................10 2.4.8 PHOTOMETER LAMP BOARD [P/N: 30-01576-00] .........................................................11 2.4.9 CONTROL PANEL BOARD [P/N: 30-01850-00] ...............................................................11 2.4.10 LEVEL SENSOR ASSY [P/N: 10-01478-00].......................................................................11 Ellipse Service Manual Rev. 03 01 June 2009 Page 1 Chapter 02 – SYSTEM DESCRIPTION 2 DESCRIPTION OF THE SYSTEM "Ellipse" is a random access, computer controlled, counter-top, clinical analysis instrument. The system can perform 138 tests per hour and has a machine cycle of 26 seconds. Its execution time ranges from a minimum of 18 seconds to a maximum of 1032, depending on the analysis method chosen. The first time the system is used for laboratory analyses, the operator must configure the system based on the specific needs of that laboratory; i.e.: the chemistry parameters and the reagents racks, along with the normal ranges, calibrated and control values, must all be defined. The daily routine analyses will be carried out according to patient sample arrival in a sequential and continuous, non-stop manner. The work list is organized using a loading rack holding up to 8 patient samples. Rack loading is non-stop. The racks can accommodate both test tubes and micro cups. The bar code for primary tubes and reagent containers is an optional feature. When the system, the analytical unit and the computer, is turned on the color-meter lamp is supplied with low voltage power (1.2 volts), the sampling arm pre-heater remains turned off, while instead the reaction plate heater, the reagents refrigerating unit (optional module) and the electronic components are turned on. In this phase, the "Stand-by" light, placed on the front panel, will flash until the reaction plate reaches a temperature of 36° C. When this temperature is reached, the "Stand-by" light will stop flashing and will remain constantly lit. In the case of system failure or malfunction, the "Ready" light, situated on the front panel of the instrument, will light up red. In order to access the main program, double click on the "Ellipse" icon on the computer desktop. The main menu - "System Monitor" - will appear. Whenever any system function is launched, the color-meter lamp and the sampling arm pre-heater will receive regular power. Ellipse Service Manual Rev. 03 01 June 2009 Page 2 Chapter 02 – SYSTEM DESCRIPTION 2.1 ANALYSES CYCLE 2.1.1 REACTION PLATE The reaction plate of the "Ellipse" system contains 6 disposable racks with 20 reaction cuvettes each. The racks can be removed individually. The basic operating cycle of the reaction plate takes 26 seconds. This cycle includes: optic reading of the cuvettes in incubation, aspiration and dispensing of the reagents and the samples by the arm, along with the relative positioning of the plate . The reactions take place at 37° C. This temperature is maintained constant by a controlled heating unit placed under the reaction plate . 2.1.2 REACTION PLATE CYCLE After reagents and samples have been placed in cuvette #1, the reaction plate will rotate 40 positions counter-clockwise, so as to bring the first cuvette to be analyzed in front of the colormeter for reading with either one or two wavelengths, as required. The plate will then, moving counter-clockwise, carry out all the readings of any other prepared cuvettes. After having effectuated all the readings, the plate will move counter-clockwise to its initial position minus one cuvette, ready for a new dispensing. In this manner, the reaction cuvettes move clockwise 1 - 2 - 3 – 4 for dispensing, washing and for their relative readings. 2.2 SAMPLING ARM - OPERATIONAL SEQUENCE 1. The sampling arm lifts up from the wash well and carries out a wash cycle; 2. The arm moves toward the specific reagent container, while the diluter aspirates an air bubble to separate the rinse column from the reagent; 3. The arm lowers itself into the reagent, below the level indicated by the sensor, and aspirates the required quantity of reagent. If the method requires a Rinse (used in order to reduce the possibility of negative contamination between the water column and the reagent) an extra amount of reagent (not used in the analysis) will be aspirated before the quantity of reagent necessary for the analysis, along with another air bubble for their separation; Ellipse Service Manual Rev. 03 01 June 2009 Page 3 Chapter 02 – SYSTEM DESCRIPTION 4. While the diluter aspirates a second air bubble, the arm rises and then lowers into the wash well so that it can be washed externally, to minimize cross contamination; 5. The arm moves to the specified sample and aspirates a third air bubble; 6. Once the level sensor has indicated the presence of the liquid, the arm stops and aspirates the sample; 7. The arm once again is raised, while the diluter aspirates a fourth air bubble to prevent sample loss; 8. At this point, the arm returns to the wash well in order to clean the outside of the probe and aspirates a fifth air bubble; 9. The arm moves to the reaction PLATE , dispenses, and mixes the reagent and sample in the reaction cuvette for incubation and reading; 10. The Ellipse arm returns to the Service Manual wash well Rev. 03 and carries out 01 June 2009 a probe wash cycle. Page 4 Chapter 02 – SYSTEM DESCRIPTION SAMPLING SYSTEM Aspiration of Air Aspiration of Reagent Aspiration of Air (To separate reagent from serum) To wash well to clean the probe Aspiration of air Aspiration of serum Aspiration of air To wash well to clean the probe H2O AIR Aspiration of air RGT AIR Dispensing and mixing in cuvette AIR SERUM AIR To wash well for final wash Ellipse Service Manual Rev. 03 01 June 2009 Page 5 Chapter 02 – SYSTEM DESCRIPTION 2.3 WASH STATION The reaction plate wash station is made up of a series of five small needles situated on one side of the reaction PLATE . Said needles are opportunely connected to the valve and pump system for emptying, washing and drying operations (please see the hydraulic diagram). 2.3.1 WASH STATION CYCLE The wash station carries out its operations alternating upward and downward movement. In its downward movement phase the needles are guided in such a manner as to carry out the following operations: • The first needle, using the central cannula, removes the reaction mix while the external cannula dispenses, shower-like fashion, the wash solution; after that the external cannula dispenses wash solution and then the liquid is aspirated from the central cannula; • The second needle operates exactly like the first but uses distilled water instead; • The third needle operates exactly like the first but dispenses rinse solution into the cuvette so that an optics check can be performed (if the results are negative, the cuvette is discarded); • The fourth needle aspirates the control water; • The fifth needles dries the sides. All these operations are part of the routine operation of the instrument. Every reaction cuvette is washed at the end of each round of analysis. The reusability (optical integrity) of each reaction cuvette is always tested before the next round of analysis. Ellipse Service Manual Rev. 03 01 June 2009 Page 6 Chapter 02 – SYSTEM DESCRIPTION 2.4 ELECTRONIC DESCRIPTION 2.4.1 INTRODUCTION The Ellipse general interconnection diagram is reported on the document having the code SC-16-00571-XX. This document shows all the electronic boards and the links among them. The power supply is connected through a dedicated socket to the power line. It supplies, by two separate modules, the two requested voltages +24 VDC and +12 VDC. All the Ellipse electronics boards are following described. 2.4.2 ANALYTICAL CONTROL BOARD (C.P.U.) [P/N: 30-01283-01] The Analytical Control Board, integrated in the Ellipse system, is the heart of the low level, real time and processing management. This board, on which is present the micro-controller HITACHI H8S/2633F (U1), permits to manage all the input and output analogical and digital signals need to the instrument functionality. The mentioned micro-controller, having 128 pin packaging and 12 ports, contains 16 KB of RAM memory and 256 KB flash memory where the firmware is installed. The firmware writing on the flash memory is permitted by an external personal computer by a serial transmission RS232. The Analytical Control Board is connected to the “Stepper Motors Driver Board” by a 64 pin frontal clutch connector from which receives all the signals for driving the motors. The following further boards are directly connected to the Analytical Control Board, through specific cables: The “Plate Interface Board (Sx)” and “Arm Interface Board (Dx)” for the users interfacing; The “Pre-Ampl./ADC” for reading the signals from main and reference channels of the photometer; The “Electrovalves Control Board” for commanding all the hydraulic devices. Besides, through two distinct and direct serial links, it manages (if present) the bar code reader and the ISE modules. On the board are present also the followings two leds: o The red led (LD1) not used yet; o The green led (LD2) is flashing when the board is powered and the firmware activated. Ellipse Service Manual Rev. 03 01 June 2009 Page 7 Chapter 02 – SYSTEM DESCRIPTION The power supply, by a specific line of the power cable, supplies the board with +24 VDC and +12 VDC. Further two voltages, +5 VDC and +3,3 VDC are generated on the same board by voltage regulators. Components and integrated circuit dedicated to the programming communication with the microcontroller and those for I/O ports protection, are present on the same board. Finally, the alarmed fan needed for the electronic area cooling, composed by the Analytical Control Board and the Stepper Motors Driver Board, is powered directly from the same board. 2.4.3 STEPPER MOTORS DRIVER BOARD [P/N: 30-01284-01] The "Stepper Motors Driver Board" is the interface between the “Analytical Control Board” and the stepper motors. Besides the Analytical Control Board, there are two more boards directly linked to the Stepper Motors Driver Board. They are the Plate Interface Board (Sx) and the Arm Interface Board (Dx) which transmit (by specific cables) all the command signals to the nine stepper motors present in the instrument. On the boards are present nine driver components (NMB SDI-C403 or TOSHIBA TA8435H). Each driver, on which is installed a metal heatsink by a couple of screws, is dedicated to drive one stepper motor as specified here below: driver U1 Sample plate motor; driver U7 Reagent plate motor; driver U4 Reaction plate motor; driver U6 Vertical sampling arm motor; driver U3 Horizontal sampling arm motor; driver U9 Diluter motor; driver U2 Photometer filter wheel motor; driver U5 Washing station motor; driver U8 Peristaltic pump motor. By the nine dip switches setting on the board, each motor is driven to ¼ of step unless the filter wheel motor that is driven to 1/8 of step. The power supply, by a specific power line cable, supplies two voltages +24 VDC and +12 VDC to the board. The +5 VDC needed to the driver components is generated on the same board by a voltage regulator. All the components managing the nine drivers and the relevant signals are mounted on the same board. Ellipse Service Manual Rev. 03 01 June 2009 Page 8 Chapter 02 – SYSTEM DESCRIPTION 2.4.4 PLATE INTERFACE BOARD (SX) [P/N: 30-01281-01] This board is the hardware interface between the reaction and reading sections of the instrument and the following boards: the Analytical Control Board which receives from the Plate Interface Board the I/O signals for managing the relevant devices (the three independents plates home sensors, the washing station home sensor, the photometer filter wheel home sensor, the reaction plate N.T.C., the reaction plate heater, the lamp and the fan of the photometer); the Stepper Motors Driver Board from which the Plate Interface Board receives the command signals for the stepper motors of the following devices: sample plate, reagent plate, reaction plate, washing station and photometer filter wheel; the Photometer Lamp Board for the photometer halogen lamp checking. There are also the following two leds: o the yellow led (LD1) active when the reaction plate resistor is powered; o the red led (LD2) active when there is a signal from the reaction plate leaking. The power supply, by a specific power line cable, supplies two voltages +24 VDC and +12 VDC to the board. The +5 VDC needed is generated on the Plate Interface Board by a voltage regulator. All the components managing respective devices are mounted on this board. 2.4.5 ARM INTERFACE BOARD (DX) [P/N: 30-01282-01] This board is the hardware interface between the Sampling sections of the instrument and the following boards: the Analytical Control Board which receives from the Arm Interface Board the I/O signals for managing the relevant devices (the vertical sampling arm home sensor, the horizontal sampling arm home sensor, the diluter home sensor, the diluter electrovalve and pump, the preheater, the level sensor assembly, the temperature service probe, the main alarmed fan, the three buttons – indicators of the Control Panel); the Stepper Motors Driver Board from which the Arm Interface Board receives the command signals for the stepper motors of the following relevant devices: vertical and horizontal sampling arm, Diluter, peristaltic pump; Ellipse the Control Panel Board for the three buttons-indicators checking. Service Manual Rev. 03 01 June 2009 Page 9 Chapter 02 – SYSTEM DESCRIPTION There are also the following three led: o the yellow led (LD1), active when the preheater is powered; o the red led (LD2), active when there is a signal from the sampling arm leaking sensor; o the yellow led (LD3), light on when the level sensor is activated. The power supply, by a specific power line cable, supplies +24 VDC and +12 VDC to the board. +5 VDC needed to the Arm Interface Board is generated on the same board by a voltage regulator. All the components managing respective devices are mounted on this board. 2.4.6 ELECTROVALVES CONTROL BOARD [P/N: 30-01626-00] This board communicates with the Analytical Control Board all the I/O signals for managing all the relevant devices (electrovalves and micropumps related to the washing station, the washing well and the peristaltic pump; the bottles liquid sensors; the waste sensors; the main cover switch). The red led (LD1) is active when there is a signal from the washing station leaking sensor; From the Analytical Control Board the Electrovalves control board receives the +24 VDC . +5 VDC is generated from + 24VDC by a voltage regulator. All the components managing respective devices are mounted on this board. 2.4.7 PRE-AMPL/ADC [P/N: 30-00107-00; P/N: 30-00107-03] The Pre-Ampl/ADC board P/N: 30-00107-00 is used for reading the signals from the photometer main channel, while the Pre-Ampl/ADC P/N: 30-00107-03 has the same function for the photometer reference channel. They are mounted on the photometer assembly and directly linked to the Analytical Control Board, by two distinct flat cables, for transmitting all the I/O signals. The only difference between them is the R1 resistor value. On the boards, the (FD1) photodiode reveals the signal arriving from the relevant reading channel; the Amplifier (U2) amplifies the signal, which is then inputted in the serial converter ADS 1250 (U3) for its conversion A/D. From the Analytical Control Board they receive also two continuous voltages (+VB and –VB). All the components dedicated to the signal amplification or digital conversion are mounted on the same boards. Ellipse Service Manual Rev. 03 01 June 2009 Page 10 Chapter 02 – SYSTEM DESCRIPTION 2.4.8 PHOTOMETER LAMP BOARD [P/N: 30-01576-00] This board is linked to the Plate Interface Board (Sx) from which receives the check signal (coming from the Analytical Control Board) to regulate the light intensity of the photometer halogen lamp (+6V, 10W). From the interface board receives the continuous voltage (VL) which is then converted to +6 VDC on the same board, by a voltage regulator, for powering the lamp. 2.4.9 CONTROL PANEL BOARD [P/N: 30-01850-00] This board is linked to the Arm Interface Board (Sx) to which transmits the I/O signals from the 3 lighting buttons of the Control Panel. From the interface board receives also the continuous voltage (VB) for the three buttons – indicators functionality. 2.4.10 LEVEL SENSOR ASSY [P/N: 10-01478-00] The level sensor assembly is composed by the level sensor board “A” (P/N:30-01392-00) and the level sensor board “B” (P/N:30-01392-01). The level sensor assembly, installed on the sampling probe head, is linked to the Arm interface board. By a specific cable it transmits the relevant signal (the capacity variation produced by the contact between the sampling probe and the liquid) to the Analytical Control Board. Level sensor components and integrated circuits are mounted on the level sensor assembly. Ellipse Service Manual Rev. 03 01 June 2009 Page 11 Chapter 03 - INSTALLATION CHAPTER 03 - INSTALLATION INDEX 3.1 UNPACKING .........................................................................................................................2 3.2 INSTALLATION ...................................................................................................................4 3.2.1 INSTALLATION SITE SPECIFICATIONS ......................................................................4 3.2.2 ELECTRIC CURRENT REQUIREMENTS.......................................................................4 3.2.3 CONNECTIONS TO ACCESSORIES ...............................................................................5 3.2.3.1 POWER SUPPLY/LINE INPUT......................................................................................5 3.2.3.2 CONNECTING THE INSTRUMENT TO THE COMPUTER .......................................6 3.2.4 ATTENTION.......................................................................................................................6 3.2.5 SYMBOLS...........................................................................................................................6 3.2.6 REGULATORY COMPLIANCE........................................................................................6 Ellipse Service Manual Rev. 03 01 June 2009 Page 1 Chapter 03 - INSTALLATION 3.1 UNPACKING The ELLIPSE is packed and delivered in two separate wooden crates: one contains the analyzer itself and the other the computer, along with its accessories. In the event that the order not include the PC component, packing and delivery will involve one wooden crate plus a corrugated cardboard box. The packing has been expressly studied and designed to insure maximum protection of the contents during shipping and handling. It is therefore extremely important that the crate(s)/box be carefully examined upon delivery in order to ascertain their integrity. Special attention should be dedicated to examining the color of the “Shock Watch” glued to the crates, which must show the color ‘white’. A ‘red’ “Shock Watch” indicates that the crate(s) have experienced some sort of ‘shock’ during handling, transport and/or delivery. This fact must be noted by the courier on the delivery note, as must any and all visible external damage (for example: holes, dents, rips or tears, water marks, etc.) evident at the moment of delivery. This will simplify matters in the event of any future claims for damages. Upon arrival of the crate(s)/box, take out the delivery note and make sure that all the items on the packing list are included in the crates and are undamaged. Make sure the series number on the delivery note/packing list corresponds to that impressed on the plate on the right side of the instrument. ELLIPSE S/N P/N 90 - 250Vac Analyzer Medical System 47 - 63Hz 300W Rome-Italy Open the crate(s)/box from the top and very carefully take out: - the instrument; - the computer and accessories. MAKE SURE THAT THE UNPACKING IS CARRIED OUT BY TWO PEOPLE. Ellipse Service Manual Rev. 03 01 June 2009 Page 2 Chapter 03 - INSTALLATION Do not discard the delivery crate(s)/box or the packing material until the correct functioning of the instrument has been ascertained. Remove all the items from the crate(s)/box very carefully. Remove the adhesive tape from the cover of the samples and reagents housing, from the front panels and from the samples and reagents racks. Before connecting the "Ellipse", remove the protective packing material placed under the sampling arm and under the wash station group. Warning: in the event that it is necessary to repack any or all of the delivered item(s), the following procedures must be carefully followed: ¾ Reposition the protective packing material under the sampling arm and under the wash station group. ¾ Tape down (using masking tape if possible) the cover of the samples and reagents housing, the front panels, and the samples and reagents racks. ¾ Remove the probe from the sampling arm and place it inside a cuvette. Then cap the cuvette and tape the cap down. ¾ Be very careful to not bend the wash station cannulas when repositioning the protective packing material. ¾ Fill the empty spaces around the accessories packed in the crate using “pluriballs” or other suitable packing material. Ellipse Service Manual Rev. 03 01 June 2009 Page 3 Chapter 03 - INSTALLATION 3.2 INSTALLATION The ELLIPSE must only be installed by a qualified technician who has been authorised and trained to do so. During its installation the system will be checked once again to ensure correct functioning. The persons who are required to operate the ELLIPSE system must have received the adequate training. This should also include the "know-how" of the normal maintenance for the instrument. A description of the maintenance will be found in Chapter 7 of this manual. ELLIPSE is a complex system, and it is therefore extremely important that it is correctly installed in order to fully guarantee fine performance. If the installation and use directions, given in this manual, are not correctly followed and/or security indications are not respected, AMS cannot guarantee correct functioning of the instrument. Apart from this, the security of the operator could be placed at risk. 3.2.1 INSTALLATION SITE SPECIFICATIONS Ascertain that the ELLIPSE system is not exposed to direct sunlight, draughts, dust or strong magnetic fields. In addition, please take note of the following conditions required for the location of the installation: USE In covered and dry place DEGREE OF POLLUTION 2 INSULATION CLASS I INSTALLATION CATEGORY II TEMPERATURE HUMIDITY between18°- 30°C ALTITUDE Max 3000 m Shelf or table with a minimum surface of 75 x 60 cm stable and free of vibration Leave a minimum distance of 10 cm around the instrument to permit air circulation . Make sure that the front and rear holes are not blocked by any object LOCATION VENTILATION 3.2.2 20% ÷ 85% ELECTRIC CURRENT REQUIREMENTS The power voltages to which the instrument is adapted are indicated on the left-hand side (see fig. 1). It must be plugged into a plug of the correct voltage. Ellipse VOLTAGE 100 ÷ 230 Vac 47/63 Hz ± 10% FUSES 6.3 Amp/T - 5 x 20 Service Manual Rev. 03 01 June 2009 Page 4 Chapter 03 - INSTALLATION NOTE: IT IS ADVISABLE TO MAINTAIN THE MAXIMUM STABILITY OF THE ELECTRICAL CURRENT IN THE LABORATORY. WHERE THIS IS NOT POSSIBLE OR ASCERTAINABLE, USE OF THE FOLLOWING SUPPLEMENTARY DEVICES IS RECOMMENDED: ELECTRONIC STABILIZER Used to stabilise the electric voltage in the laboratory. Any stabiliser with a power potential greater than 0.5 KW, currently available on the market, can be used. NO-BREAK MODULE UPS - (Uninterrupted Power Supply) This module provides two important functions: - stabilises the main-line power - supplies current to the instrument in case of a main-line power failure. 3.2.3 CONNECTION OF THE ACCESSORIES 3.2.3.1 POWER SUPPLY Fig. 1 - – Plug (use the feeder cable supplied with the instrument). The sticker indicates the power supply voltage and the values of the fuses. Ellipse Service Manual Rev. 03 01 June 2009 Page 5 Chapter 03 - INSTALLATION 3.2.3.2 COMPUTER - INSTRUMENT CONNECTION The instrument and the Personal Computer are connected by one serial RS232 standard cable (Cod. 9-35 0055.01), which provides the hardware support for the communication. 3.2.4 ATTENTION The following label is found at the rear of the instrument. NOTE: THE REAR PANELS OF THE INSTRUMENT MUST NEVER BE OPENED WITHOUT HAVING FIRST SWITCHED THE INSTRUMENT OFF AND DISCONNECTED THE ELECTRICITY CABLE. THE MAINTENANCE AND CLEANING PROCEDURES FOUND IN CHAPTER MANUAL MUST BE RESPECTED AT ALL TIMES. REMEMBER TO 07 OF THIS FOLLOW THE DECONTAMINATION PROCEDURE IN CASE OF INSTRUMENT REMOVAL (SEE CHAPTER 07) . 3.2.5 SYMBOLS ATTENTION: READ THE INSTRUCTIONS IN THE USER MANUAL TERMINAL OF TOTAL MASS PROTECTION (CONDUCTOR) 3.2.6 REGULATORY COMPLIANCE The ELLIPSE instrument complies with: European Directive 98/79/CE for In vitro Diagnostics Devices Ellipse Service Manual Rev. 03 01 June 2009 Page 6 Chapter 04 – ELECTRICAL SCHEMES AND DRAWINGS CHAPTER 04 - ELECTRICAL SCHEMES AND DRAWINGS INDEX ELECTRONIC DIAGRAM ELLIPSE SC-16-00571-XX HYDRAULIC DIAGRAM ELLIPSE SI-16-00571-XX WIRING ME-50-01692-01 ANALYTICAL CONTROL BOARD SC02-30-01283-XX ANALYTICAL CONTROL BOARD SE-30-01283-XX STEPPER MOTORS DRIVER BOARD SC02-30-01284-01 STEPPER MOTORS DRIVER BOARD SE-30-01284-01 ARM INTERFACE BOARD (DX) SC02-30-01282-01 ARM INTERFACE BOARD (DX) SE-30-01282-01 PLATE INTERFACE BOARD (SX) SC02-30-01281-01 PLATE INTERFACE BOARD (SX) SE-30-01281-01 PRE-AMPL/ADC SE-30-00107-XX PHOTOMETER LAMP BOARD SE-30-01576-00 LEVEL SENSOR ASSY SE-10-01478-00 ELECTROVALVES CONTROL BOARD SC-30-01626-00 ELECTROVALVES CONTROL BOARD SE-30-01626-00 CONTROL PANEL BOARD SE-30-01850-00 Ellipse Service Manual Rev. 03 01 June 2009 Page 1 Pre-Ampl/ADC (Sample Channel) Pre-Ampl/ADC (Reference Channel) 30-01283-.. (optional) Electronic Diagram Ellipse The present document is property of AMS, any use without authorization will be prosecuted accordingly to the law. Description Via E. Barsanti, 17/A (Via Tiburtina Km 18,300) 00012 Guidonia (Roma) P/N 20/10/2005 Sheet Rev. 1 of 1 A. Gagliarducci R. Cornacchia L. Massenzi SC 16-00571-XX D Doc. Date Approved Checked Drawn Rev. D: emissione in riferimento alla RMP n.268 Analytical Control Board (optional) BarCode Reader ISE Module 50-01692-01 30-01283-.. Analytical Control Board The present document is property of AMS, any use without authorization will be prosecuted accordingly to the law. Description Via E. Barsanti, 17/A (Via Tiburtina Km 18,300) 00012 Guidonia (Roma) P/N 20/10/2005 Sheet Rev. 1 of 1 A. Gagliarducci R. Cornacchia L. Massenzi SC02 30-01283-XX A Doc. Date Approved Checked Drawn Rev. A: emissione in riferimento alla RMP n.268 (Sample Channel) Pre-Ampl/ADC (Reference Channel) Pre-Ampl/ADC A B C D VM H5 TX4 VB RX4 470uH 470uF/50V C15 + L1 VM J6 J4 11 1 WRL PC0 CLK ENG 1D 2D 3D 4D 5D 6D 7D 8D CLKCU CLKWH CLKBV DIRCU DIRWH DIRBV CLKCA CLKFO CLKBH DIRCA DIRFO DIRBH CLKRE CLKPP CLKDO DIRRE DIRPP DIRDO + C13 470uF/50V 2 5 6 9 12 15 16 19 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A20 A21 A22 A23 A24 A25 A26 A27 A28 A29 A30 A31 A32 PWDRE PWDPP PWDDO ENAPI ENAPP ENADO PWDCU PWDWH PWDBV ENAPI ENABVWH ENABVWH PWDCA PWDFO PWDBH ENAPI ENABHFO ENABHFO 5 0.1uF C24 DTR# 0.1uF 1 DTR 1N4148 D2 1N4148 D1 TX2 RX2 0.1uF C22 1 2 3 4 5 6 7 8 3V3 0.1uF C7 4V7 D6 4V7 D5 0.1uF C36 PVCC VCC GND T1OUT R1IN R1OUT T1IN T2IN R2OUT PVCC R1 3K3 C5 470pF 0.1uF 16 15 14 13 12 11 10 9 4 RX0 P31 P30 PA1 PA2 C25 0.1uF PVCC TX0 R8 0R LOCK_IN RESB RESA PC0 PA3 PG3 (/CS1) PF4 (/HWR) PF5 (/RD) C18 10uF/25V -VB 1 2 3 4 5 6 7 8 9 11 18 16 14 12 9 7 5 3 20 10 74HC244 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND C60 R2 220 0.1uF R36 10K STBY 19 1 PG3 PF5 U5 1 2 3 4 0.1uF F0 F1 F2 F3 F4 F5 F6 F7 VCC GND DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 C29 TX4 0.1uF RX4 1 2 3 4 5 6 7 8 C1+ V+ C1C2+ C2VT2OUT R2IN U8 0.1uF R39 10K G DIR 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE U1 VCC GND T1OUT R1IN R1OUT T1IN T2IN R2OUT MAX202 H2 1 2 J18 20 10 18 17 16 15 14 13 12 11 18 16 14 12 9 7 5 3 20 10 10 20 2 5 6 9 12 15 16 19 74HC377 GND VCC 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q 74HC244 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND 3 16 15 14 13 12 11 10 9 TX1 RX1 P34 P33 P37 P36 0.1uF C30 0.1uF 2 3 4 5 6 7 8 9 PVCC R34 LEDV J3 H4 Diagnostic J2 H8 PVCC 18 16 14 12 9 7 5 3 20 10 VB PVCC PVCC 10K R38 2 J13 H3 P94 P91 WRL PC3 DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 ENA1 J1 H5 0 Rev. 18 16 14 12 9 7 5 3 20 10 J8 Description 0.1uF C48 PVCC 0.1uF C41 PVCC H2 1 2 J22 J16 SE-30-01283-XX Document Number 1 J14 C59 + P97 DACP 10 8 6 4 2 10 8 6 4 2 10 8 6 4 2 9 DRDYRF CLK 7 G0R 5 G1R 3 1 9 DRDYDA CLK 7 G0D 5 G1D 3 1 H5x2 9 7 5 3 1 3V3 PVCC 9 DRDYAUX CLK 7 G0AUX 5 G1AUX 3 1 25 23 21 19 17 15 13 11 9 7 5 3 1 H5x2 9 7 5 3 1 1 Sheet Drawn Checked L.M. 10/09/2004 of 1 R.CORNACCHIA 1 J9 PC3 PC4 CTR2 PF5 (RD) PD4 PD5 PD6 PD7 RES RTS RX2 DACP P97 H3 H5x2 9 7 5 3 1 H13X2 25 23 21 19 17 15 13 11 9 7 5 3 1 R48 1K 3 2 1 J15 H3 DACB J17 P96 0 Rev Checked Approved Approved A.G. 10/09/2004 J10 SCLK 10 DINDA 8 -VB 6 DSYNC 4 2 VB J11 SCLK 10 DINRF 8 -VB 6 DSYNC 4 2 VB 3 2 1 R52 1K 26 24 22 20 18 16 14 12 10 8 6 4 2 J5 J12 SCLK 10 DINAUX 8 -VB 6 DSYNC 4 2 VB PC2 26 PC1 24 PC0 22 PF4 (HWR)20 PD3 18 PD2 16 PD1 14 PD0 12 P35 10 DTR# 8 TX2 6 PG2 (CS2) 4 2 1uF/50 H3 3 2 1 C55 + P96 RISAB 1uF/50 H3 3 2 1 Analytical Control Board RMP 224 PVCC PVCC LAMPG LAMPV LAMPR LEDV LEDR ENADO ENAPP CTR1 CTR2 DSYNC R42 10K DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 Date: 10/09/2004 Size A3 10 20 2 5 6 9 12 15 16 19 P95 P92 R41 10K R47 1K R51 1K 74HC377 GND VCC 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q PF3 P73 74HC244 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND Title 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE U20 2 4 6 8 11 13 15 17 1 19 CLK ENG U12 11 1 1D 2D 3D 4D 5D 6D 7D 8D 1 1uF/50 C58 + SP 1uF/50 C54 + SB C+8 10K 3 4 7 8 13 14 17 18 9 8 7 6 5 4 3 2 RR4 Serial IN RS232 WASTE1 WASTE2 LIQUID1 LIQUID2 LIQUID3 LEAKWH LEAKB DEPR 1 VM EV3PP EV1H2O uPBASICP EV1PP uP1H2O SWCOP WASTE1 LIQUID3 LIQUID1 PE4 PE3 PE5 PE6 PE2 PE1 PE7 PE0 R45 1K ALLVENT PVCC 0.1uF C43 PVCC PVCC 0.1uF C37 PVCC 23 21 19 17 15 13 11 9 7 5 3 1 R49 1K J21 H6 23 21 19 17 15 13 11 9 7 5 3 1 C+8 10K R44 10K 9 8 7 6 5 4 3 2 1 2 3 4 5 6 24 22 20 18 16 14 12 10 8 6 4 2 J20 H2X12 RR3 LOCK CTR2 PF2 uPACIDB SCLK P35B ENABHFO PF7 PF0 PVCC DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 P14 P15 P13 P16 P12 P17 P11 P10 1uF/50 C57 + TP 1uF/50 C53 + TB Electronic Housing Fan LED GREEN BarCode 18 16 14 12 9 7 5 3 20 10 74HC244 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND LD1 1 24 22 20 18 16 14 12 10 8 6 4 2 2 EVBASICP EVACIDP uPACIDP PVCC VM 74HC244 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND LED RED LD2 RX0 4K7 4K7 P93 VM EV2H2O uPBAND EV2PP uP2H2O LEAKWH WASTE2 LIQUID4 LIQUID2 0.1uF C44 C+8 10K 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE U18 2 4 6 8 11 13 15 17 1 19 9 8 7 6 5 4 3 2 RR2 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE U19 2 4 6 8 11 13 15 17 1 19 C45 PVCC R46 1K AVCC TX0 R33 PF0 PF1 LIQUID4 PF3 PA0 P35 P77 OPTOWH2 LEDR VM TX1 RX1 PVCC 0.1uF C42 PVCC 0.1uF C38 10 20 DIRCA DIRCU DIRRE DIRFO DIRWH DIRBH DIRBV DIRDO 0.1uF R50 1K P90 AVCC PB4 PB3 PB5 PB6 PB2 PB7 PB1 PB0 PVCC PG PV PR SLI ARMSW LEAKPI SWCOP ALLVENT ENA0 1 1uF/50 C+8 10K RR1 TF C56 + PVCC 1uF/50 C52 + TA PVCC PVCC C3 GND VCC CLK ENG 2 5 6 9 12 15 16 19 74HC377 C40 0.1uF 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q 1D 2D 3D 4D 5D 6D 7D 8D U10 11 1 3 4 7 8 13 14 17 18 0.1uF C4 PVCC G0D G1D G0R G1R G0AUX G1AUX DIRPP PWDPP PVCC P40 P41 P42 P43 P44 P45 P46 P47 PVCC DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 PVCC 74HC245 VCC GND B1 B2 B3 B4 B5 B6 B7 B8 PE7 PE6 PE5 PE4 PE3 PE2 PE1 PE0 P17 PD0 P31 P30 PD7 PD6 PD5 PD4 PD3 PD2 PD1 P32 P34 P33 WRL PC1 DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 H8S/2633F C51 0.1uF 64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49 48 47 46 45 44 43 42 41 40 39 PVCC uP1H2O uP2H2O uPBAND uPBASICP uPSCAMB CTRLAMP RISCPI BUZZER LOCKED(ON) CLK ENG 1D 2D 3D 4D 5D 6D 7D 8D U11 11 1 3 4 7 8 13 14 17 18 U16 LOCK C31 12 FWEBOOT 13 RES 14 MD2 WRL 15 FWE PC2 16 WRL 17 ENA1 18 ENA0 19 DTR# LOCK_IN 20 PVCC 10 NME1212S 2 4 6 8 11 13 15 17 1 19 18 16 14 12 9 7 5 3 20 10 74HC244 0.1uF 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND A1 A2 A3 A4 A5 A6 A7 A8 U25 2 3 4 5 6 7 8 9 PD0 PD1 PD2 PD3 PD4 PD5 PD6 PD7 R37 10K PA3 P70 P71 PVCC CLKPP PWMB C34 OPTOCA CLKFO OPTOCU CLKWH 0.1uF OPTORE ENABVWH OPTOFO ENAPI OPTOWH OPTOBH PVCC OPTOBV OPTODO 0.1uF C27 C19 PVCC C39 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE U15 2 4 6 8 11 13 15 17 1 19 3 P34/RXD1/SDA0 P33/TXD1/SCL1 VSS P32/SCK0/SDA1/IRQ4 PVCC2 P31/RXD0/IRRXD P30/TXD0/IRTXD PD7/D15 PD6/D14 PD5/D13 PD4/D12 PD3/D11 PD2/D10 PD1/D9 PVCC1 PD0/D8 VSS PE7/D7 PE6/D6 PE5/D5 PE4/D4 PE3/D3 PE2/D2 PE1/D1 PE0/D0 P17/TIOCB2/TCKLD/PWM3 PE0 PE1 PE2 0.1uF PE3 PE4 PE5 PE6 PE7 3V3 R35 10K NMI PVCC U24 16V8 I0 I1 I2 I3 I4 I5 I6 I7 I8 I9 C28 PVCC 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE U23 2 4 6 8 11 13 15 17 1 19 0.1uF C2 C17 + VB INPUT INPUTB P72 P73 P74 P75 P76 P77 0.1uF 3V3 C1 0.1uF MAX202 C26 R7 3K3 R6 3K3 2 3K3 R5 R4 3K3 2 C1+ V+ C1C2+ C2VT2OUT R2IN U7 PVCC DS1233-10 1 2 3 U6 1 18 P32 16 PC5 14 PC6 12 PC7 PG4 9 PG0 7 PG1 5 WDT 3 20 PVCC 10 74HC244 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND PVCC RTS RESA R43 10K C23 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE R40 10K H2 1 2 J19 U17 DRDYAUX 2 DINDA 4 DINRF 6 DINAUX 8 OPTOPP 11 DRDYDA 13 DRDYRF 15 WDTOVF 17 1 19 PVCC PVCC 3V3 PVCC 15pF 0.1uF C8 15pF C47 C6 RX2 TX2 DIN 64_AB-H B1 B2 B3 B4 B5 B6 B7 B8 B9 B10 B11 B12 B13 B14 B15 B16 B17 B18 B19 B20 B21 B22 B23 B24 B25 B26 B27 B28 B29 B30 B31 B32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 0.1uF C35 PVCC PWDCA PWDCU PWDRE PWDFO PWDWH PWDBH PWDBV PWDDO 10 20 PVCC 74HC377 GND VCC 1Q 2Q 3Q 4Q 5Q 6Q 7Q 8Q 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND 18 CLKBH 16 EVACIDP 14 EVBASICP 12 EVDOS CLKBV 9 uPDOS 7 CLKDO 5 uPACIDP 3 PVCC 20 PVCC 10 C33 74HC244 0.1uF P70 P71 P72 P73 P74 P75 P76 P77 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 C9 0.1uF TP-GND C46 AN0/P40 AN1/P41 AN2/P42 AN3/P43 AN4/P44 AN5/P45 DA0/AN6/P46 DA1/AN7/P47 AN8/P90 AN9/P91 AN10/P92 AN11/P93 AN12/P94 AN13/P95 DA2/AN14/P96 DA3/AN15/P97 AVSS CS4/DREQ0/TMCI011/TMRI01P70 CS5/DREQ1/TMCI23/TMRI23/P71 SYNCI/CS6/TEND0/TMO0/P72 CS7/TEND1/TMO1/P73 MRES/TMO2/P74 SCK3/TMO3/P75 RXD3/P76 TXD3/P77 MD0 C10 + 0.1uF 10uF/25V P40 P41 P42 P43 P44 P45 P46 P47 P90 P91 P92 P93 P94 P95 P96 P97 0.1uF C50 VB + C11 XTAL Rx1=470 se X1<5Mhz 4 DTR RTS J7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 1 2 3 4 5 PVCC U4 Refer to the Serigrafy U9 3 4 7 8 13 14 17 18 RES 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE U14 2 4 6 8 11 13 15 17 1 19 3 2 1 AVCC PVCC 18 CLKCU 16 EV1PP 14 EV2PP 12 EV3PP CLKCA 9 EV1H2O 7 CLKRE 5 EV2H2O 3 20 PVCC PVCC 10 C32 74HC244 0.1uF 1Y1 1Y2 1Y3 1Y4 2Y1 2Y2 2Y3 2Y4 VCC GND ISE - NO3 Module 1A1 1A2 1A3 1A4 2A1 2A2 2A3 2A4 1OE 2OE U13 2 4 6 8 11 13 15 17 1 19 DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7 P10 P11 P12 P13 P14 P15 P16 P17 PB0 PB1 PB2 PB3 PB4 PB5 PB6 PB7 C14 D3 1N5818 0.1uF OUT GND IN U3 U2 LM7805CT LM2575 MOLEX 1 2 3 4 5 6 TP1 MD2 3 2 1 PC0 PC1 PC2 PC3 OUT GND IN PC4 470uF/50V VL C49 + 1 6,144 Mhz X1 R3 0R EXTAL PF0 PF1 PF2 PF3 PF4 PF5 PF6 10uF/25V PF7 PC5 PC6 PC7 C12 EXTAL 3V3 OSC1 PB0 5 XTAL FWE STBY NMI RES PB1 PB2 PB3 PB4 PB5 PB6 PB7 PA0 PA1 PA2 PA3 LD1086V3.3 WDTOVF PG4 (CS1) PG3 (CS2) PG2 PG1 PG0 P37 P10 P11 P12 P13 P14 1 2 3 4 5 P36 P35 102 101 100 99 98 97 96 95 94 93 92 91 90 89 88 87 86 85 84 83 82 81 80 79 78 77 76 75 74 73 72 71 70 69 68 67 66 65 VREF AVCC NC NC PF0/BREQ/IRQ2 PF1/BACK/BUZZ PF2/LCAS/WAIT/BREQ0 PF3/LWR/ADTRG/IRQ3 PF4/HWR PF5/RD PF6/AS/LCAS VSS PF7/CLK PVCC1 OSC2 OSC1 VSS EXTAL VCC XTAL FWE STBY NMI RES PLLVSS PLLCAP PLLVCC WDTOVF PG4/CS0 PG3/CS1 PG2/CS2 PG1/CS3/OE/IRQ7 PG0/CAS/IRQ6 P37/TXD4 NC NC P36/RXD4 P35/SCK1/SCK4/SCL0/IRQ5 + VM C VB C 10uF/25V C C16 0.1uF C MD1 MD2 NC NC A0/PC0 A1/PC1 A2/PC2 A3/PC3 VSS A4/PC4 VCC A5/PC5 PWM0/A6/PC6 PWM1/A7/PC7 VSS TIOCA3/A8/PB0 PVCC1 TIOCB3/A9/PB1 TIOCC3/A10/PB2 TIOCD3/A11/PB3 TIOCA4/A12/PB4 TIOCB4/A13/PB5 TIOCA5/A14/PB6 TIOCB5/A15/PB7 A16/PA0 TXD2/A17/PA1 RXD2/A18/PA2 SCK2/A19/PA3 VSS A20/DACK0/TIOCA0/PO8/P10 A21/DACK1/TIOCB0/PO9/P11 A22/TCKLA/TIOCC0/PO10/P12 A23/TCKLB/TIOCD0/PO11/P13 IRQ0/TIOCA1/PO12/P14 NC NC TCKLC/TIOCB1/PO13/P15 IRQ1/PWM2/TIOCA2/PO14/P16 INPUT WDT OPTOWH2 CTR2 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 TF TP SP LIQUID4 DACP P15 P16 25 23 21 19 17 15 13 11 9 7 5 3 1 26 24 22 20 18 16 14 12 10 8 6 4 2 CTRLAMP BUZZER 1 2 3 4 LEAKB PWMB EV3PP INPUTB 1 2 3 4 5 6 7 8 RISAB TB TA SB 1 2 3 SLI ARMSW CTR1 CTR2 WDT DACB 1 2 3 4 5 33 31 29 27 25 23 21 19 17 15 13 11 9 7 5 3 1 34 32 30 28 26 24 22 20 18 16 14 12 10 8 6 4 2 OPTOCU OPTOCA OPTORE OPTOFO OPTOWH WASTE1 WASTE2 LIQUID1 LIQUID2 LIQUID3 LEAKPI RISCPI uPSCAMB H13X2 LAMPV LAMPR LAMPG H17X2 EVDOS DSYNC ADC OPTOBH OPTOBV OPTODO OPTOPP PR PG PV DEPR uPACIDB uPDOS A B C D 50-01692-01 U5 J7 Stepper Motors Driver Board The present document is property of AMS, any use without authorization will be prosecuted accordingly to the law. Description Via E. Barsanti, 17/A (Via Tiburtina Km 18,300) 00012 Guidonia (Roma) P/N 20/10/2005 Sheet Rev. 1 of 1 A. Gagliarducci R. Cornacchia L. Massenzi SC02 30-01284-01 A Doc. Date Approved Checked Drawn Rev. A: emissione in riferimento alla RMP n.268 30-01283-.. C34 3n3 5 U7 1 1 2 1 2 1 2 SW DIP-2 SW7 SW DIP-2 SW4 4 3 4 3 4 3 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB SW DIP-2 SW1 GNDD CK1 RES CK2 EN DIR PD OSC M1 M2 VCC SWRE0 SWRE1 SWCU0 SWCU1 SWCA0 SWCA1 22 17 21 18 19 16 23 20 24 15 VM R13 1R-2W D28 D27 D26 D25 0.1uF C23 R7 1R-2W D16 D15 D14 D13 0.1uF C17 R1 1R-2W D4 D3 D2 RR1 4 C+9 10K PVCC SWCU1 2 1 SWCU0 3 SWCA1 4 SWCA0 5 SWRE1 6 SWRE0 7 SWFO1 8 SWFO0 9 SWWH0 10 R14 1R-2W REA REAN REB REBN R8 1R-2W CUA CUAN CUB CUBN R2 1R-2W CAA CAAN CAB CABN ENAPP ENAWH ENAFO C35 3n3 C24 0.1uF C32 3n3 C18 0.1uF C29 3n3 13 GNDD U8 1 7 RESMOT2 CLKPP 6 3 DIRPP 5 PWDPP10 4 SWPP0 8 SWPP1 9 13 1 2 1 2 1 2 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB SW DIP-2 SW8 SW DIP-2 SW5 SW DIP-2 4 3 4 3 4 3 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB 100uF/50 SW2 GNDD CK1 RES CK2 EN DIR PD OSC M1 M2 VCC PVCC U5 1 CK1 RES CK2 EN DIR PD OSC M1 M2 VCC C2 C5 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB 100uF/50 100uF/50 GNDD CK1 RES CK2 EN DIR PD OSC M1 M2 VCC PVCC U2 1 7 RESMOT2 CLKFO 6 3 DIRFO 5 PWDFO10 4 SWFO0 8 SWFO1 9 13 7 RESMOT2 CLKWH 6 3 DIRWH 5 PWDWH10 4 SWWH0 8 SWWH1 9 C12 0.1uF PVCC 0.1uF C40 C8 22 17 21 18 19 16 23 20 VM SWPP0 SWPP1 SWWH0 SWWH1 SWFO0 SWFO1 22 17 21 18 19 16 23 20 24 15 VM VM 22 17 21 18 19 16 23 20 24 15 24 15 + 7 RESMOT2 CLKRE 6 3 DIRRE 5 PWDRE10 4 SWRE0 8 SWRE1 9 C7 22 17 21 18 19 16 23 20 24 15 + 13 100uF/50 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB VM 22 17 21 18 19 16 23 20 24 15 D1 0.1uF C11 VB 470uF/50 + 0.1uF C39 C38 VL VB R15 1R-2W D32 D31 D30 D29 0.1uF C25 R9 1R-2W D20 D19 D18 D17 0.1uF C19 R3 1R-2W D8 D7 D6 D5 0.1uF C13 J4 3 VM TP1 RR2 ENABH ENADO ENABV C+9 10K PVCC SWWH1 2 1 SWPP1 3 SWPP0 4 SWBH1 5 SWBH0 6 SWBV1 7 SWBV0 8 SWDO0 9 SWDO1 10 R16 1R-2W PPA PPAN PPB PPBN R10 1R-2W WHA WHAN WHB WHBN R4 1R-2W FOA FOAN FOB FOBN MOLEX TP-GND C36 3n3 C26 0.1uF C33 3n3 C20 0.1uF C30 3n3 13 13 1 2 1 2 GNDD CK1 RES CK2 EN DIR PD OSC M1 M2 VCC 2 SW DIP-2 SW9 SW DIP-2 SW6 SW DIP-2 SW3 U9 1 7 RESMOT2 CLKDO 6 3 DIRDO 5 PWDDO10 4 SWDO0 8 SWDO1 9 1 2 GNDD PVCC U6 1 CK1 RES CK2 EN DIR PD OSC M1 M2 VCC 4 3 4 3 4 3 22 17 21 18 19 16 23 20 24 15 C9 SWDO0 SWDO1 SWBV0 SWBV1 SWBH0 SWBH1 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB 100uF/50 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB C3 C6 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB 100uF/50 J2 100uF/50 GNDD CK1 RES CK2 EN DIR PD OSC M1 M2 VCC PVCC U3 1 7 RESMOT2 CLKBH 6 3 DIRBH 5 PWDBH10 4 SWBH0 8 SWBH1 9 13 7 RESMOT2 CLKBV 6 3 DIRBV 5 PWDBV10 4 SWBV0 8 SWBV1 9 C14 0.1uF PVCC 2 22 17 21 18 19 16 23 20 24 15 + A ENARE C22 0.1uF GNDD CK1 RES CK2 EN DIR PD OSC M1 M2 VCC PVCC U4 1 7 RESMOT2 CLKCU 6 3 DIRCU 5 PWDCU10 4 SWCU0 8 SWCU1 9 13 C4 SDIC403 GNDPA GNDPB SA SB FB1 FB2 FA1 FA2 VMA VMB 100uF/50 GNDD PVCC U1 1 CK1 RES CK2 EN DIR PD OSC M1 M2 VCC C1 VM 470uF/50 + C37 PVCC + C31 3n3 C16 0.1uF C28 3n3 7 RESMOT2 CLKCA 6 3 DIRCA 5 PWDCA10 4 SWCA0 8 SWCA1 9 13 PVCC 100uF/50 + + B ENACU ENACA C10 0.1uF RESMOT R19 10K 3 2 1 + C D U10 OUT GND IN LM2940C 3 1 2 3 4 5 6 4 1 CAA CAB CUA CUB REA REB FOA FOB WHA WHB 19 17 15 13 11 9 7 5 3 1 20 18 16 14 12 10 8 6 4 2 VM VM 22 17 21 18 19 16 23 20 24 15 VM 0 Rev. Description R18 1R-2W DOA DOAN DOB DOBN R12 1R-2W BVA BVAN BVB BVBN R6 1R-2W BHA BHAN BHB BHBN SE-30-01284-01 Document Number A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A20 A21 A22 A23 A24 A25 A26 A27 A28 A29 A30 A31 A32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 PWDRE PWDPP PWDDO ENARE ENAPP ENADO PWDCU PWDWH PWDBV ENACU ENAWH ENABV PWDCA PWDFO PWDBH ENACA ENAFO ENABH DIN 64_AB-H B1 B2 B3 B4 B5 B6 B7 B8 B9 B10 B11 B12 B13 B14 B15 B16 B17 B18 B19 B20 B21 B22 B23 B24 B25 B26 B27 B28 B29 B30 B31 B32 H8X2 J3 Drawn Checked 1 Sheet 1 of 1 R.CORNACCHIA L.M. 10/09/2004 0 Rev Checked Approved Approved A.G. 10/09/2004 J1 1 2 3 4 5 6 7 8 9 10 11 12 13 CLKCU 14 CLKWH 15 CLKBV 16 DIRCU 17 DIRWH 18 DIRBV 19 20 21 22 RESMOT 23 24 25 26 CLKRE 27 CLKPP 28 CLKDO 29 DIRRE 30 DIRPP 31 DIRDO 32 CLKCA CLKFO CLKBH DIRCA DIRFO DIRBH 1 Stepper Motors Driver Board Date: 10/09/2004 Size A3 RMP 224 R17 1R-2W D36 D35 D34 D33 0.1uF C27 R11 1R-2W D24 D23 D22 D21 0.1uF C21 Title H10X2 R5 1R-2W D12 D11 D10 D9 0.1uF C15 CAAN CABN CUAN CUBN REAN REBN FOAN FOBN WHAN WHBN BHA BHB BVA BVB PPA PPB DOA DOB 15 13 11 9 7 5 3 1 15 13 11 9 7 5 3 1 16 14 12 10 8 6 4 2 16 14 12 10 8 6 4 2 BHAN BHBN BVAN BVBN PPAN PPBN DOAN DOBN PVCC 5 + + A B C D 30-01283-.. Analytical Control Board 05-01700-20 50-01867-00 Arm Interface Board (Dx) The present document is property of AMS, any use without authorization will be prosecuted accordingly to the law. Description Via E. Barsanti, 17/A (Via Tiburtina Km 18,300) 00012 Guidonia (Roma) P/N 20/10/2005 Sheet Rev. 1 of 1 A. Gagliarducci R. Cornacchia L. Massenzi SC02 30-01282-01 B Doc. Date Approved Checked Drawn Rev. B: emissione in riferimento alla RMP n.268 50-01692-01 A B C D 1 2 2 1 2 3 J15 MIC5021 C6 100uF/50V U5 1 VDD Vbst 2 INPUT Gate 3 Ct Sens4 GND Sens+ R27 10K 10K R17 10K 3 R41 560R (see Jumper) H3 Pre-Heater PWS config. VM VPRE VB C10 0.1uF VPRE PWMB VB R2 2K2 VB D-EVDOS VM R47 PVCC RISANB H2 J16 Diluter Valve H4 C1 E2 K3 A4 J7 Spare H8 Peristaltic Pump Motor 1 2 3 4 5 6 7 8 J3 - + R28 10K 5 VPRE 8 7 6 5 1 1 2 3 LM358/SO U3A PVCC H3 J17 SB R13 1K C7 0.01uF BHA BHAN BHB BHBN 1 2 3 4 5 6 7 8 J1 H2 LED-YELLOW LD1 PWRES Q1 IRF530 3 SOURCE 2 DRAIN 3A F1 H3 J19 1 2 3 Spare VPRE 1 2 H3 3 2 1 J14 R29 33K 50K R19 R45 2K2 6 5 CTR1 DACB 10K R37 4 - + CTR2 PWRES 1 2 3 H11 1 2 3 4 5 6 7 8 9 10 11 J13 1 2 3 4 5 6 7 8 J2 TA DEPR PVCC H2 1 2 J12 10K 10K H3 J23 1 2 3 1N4148 D6 10K 10 8 6 4 2 9 7 5 3 1 6 5 H 5X2 10 8 6 4 2 J27 10K VB R36 R22 PVCC R44 10K INPUTB R35 VB 10K R21 PG 3 R52 PR 1N4148 D7 PVCC R51 R50 PV 3 Housing Fan (bottom) D3 4.7V R7 10K OPTOBV Leak Arm R43 560R R6 2K2 VB R11 10K PVCC H4 C1 E2 K3 A4 Vertical Arm Home Sensor J6 VB H8 Vertical Arm Motor BVA BVAN BVB BVBN Level Sensor and Pre-Heater 10K LD3 R33 7 H3 J20 Spare LM358/SO U3B PVCC NTC1+ ARMSW PVCC SLI VM LED-YELLOW PVCC 10K R30 R25 10K PVCC 4 D2 4.7V R5 10K D-uPACIDB VM R42 560R R4 2K2 VB Temp. Service Probe J22 H4 C1 E2 K3 A4 Horizontal Arm Home Sensor OPTOBH J5 VB H8 Horizontal Arm Motor D-uPDOS VM Diluter Pump D1 4.7V R3 10K R46 10k OPTOPP RISANB 1 GATE 8 4 8 4 5 470K 1M 2 1 2 1 - 9 7 5 3 1 C9 VB To Front Panel Green Led LEAKB LED-RED LD2 100uF/50V + C1 D4 4.7V D-LAMPG R53 750R To Front Panel Yellow Led To Front Panel Red Led R49 750R D-LAMPR R48 750R 3 2 1 LM7805CT OUT GND IN U2 BHAN BHBN BVAN BVBN PPAN PPBN DOAN DOBN R9 10K OPTODO R1 560R R18 680R VB R8 2K2 VB D-LAMPV LM358/SO 7 U4B H4 C1 E2 K3 A4 J8 Diluter Home Sensor 0.1uF PVCC + 1 2 3 4 5 6 7 8 J4 H8 Diluter Motor DOA DOAN DOB DOBN 8 4 2 VB R31 33K NTC1+ 50K R20 15 13 11 9 7 5 3 1 BHA BHB BVA BVB PPA PPB DOA DOB 10K R32 R26 10K PVCC 100uF/50V + C3 15 13 11 9 7 5 3 1 PVCC H8X2 16 14 12 10 8 6 4 2 J26 C2 0.1uF 16 14 12 10 8 6 4 2 2 2 3 C4 0.1uF PVCC - + 8 1 TP-GND 100uF/50V C11 LAMPV LAMPR LAMPG EVDOS Description TB TP1 + VM 1 2 3 4 5 6 7 8 9 OPTOBH OPTOBV OPTODO OPTOPP PR PG PV DEPR uPACIDB uPDOS RISOUTB (EV3PP) uPACIDB LAMPV LAMPR LAMPG uPDOS EVDOS SE-30-01282-01 Document Number Arm Interface Board (dx) RMP 224 R34 10K LM358/SO 0.1uF U4A C8 Date: 10/09/2004 Size A3 Title 0 Rev. 4 1 PPA PPAN PPA PPB PPBN PPB + 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 VM 1 2 3 4 5 6 Spare H9 1 2 3 4 5 6 7 8 9 J24 VM VB 1 Sheet Drawn Checked 1 of 1 R.CORNACCHIA L.M. 10/09/2004 0 Rev Checked Approved Approved A.G. 10/09/2004 VM MOLEX J21 LEAKB PWMB RISOUTB INPUTB RISAB TB TA SB VL D-RISOUTB D-uPACIDB D-LAMPV D-LAMPR D-LAMPG D-uPDOS D-EVDOS SLI ARMSW CTR1 CTR2 WDT DACB 18 17 16 15 14 13 12 11 10 D-RISOUTB RISANB RISOUTB INPUTB RISAB WDT H17X2 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 J25 ULN2803 I1 O1 I2 O2 I3 O3 I4 O4 I5 O5 I6 O6 I7 O7 I8 O8 GND COM U1 1 A B C D 30-01283-.. Analytical Control Board Plate Interface Board (Sx) The present document is property of AMS, any use without authorization will be prosecuted accordingly to the law. Description Via E. Barsanti, 17/A (Via Tiburtina Km 18,300) 00012 Guidonia (Roma) P/N 20/10/2005 Sheet Rev. 1 of 1 A. Gagliarducci R. Cornacchia L. Massenzi SC02 30-01281-01 A Doc. Date Approved Checked Drawn Rev. A: emissione in riferimento alla RMP n.268 50-01692-01 A B C D H3 3 2 1 J11 1 2 H3 3 2 1 DACP CTRLAMP D1 4.7V INPUT WDT OPTOWH2 CTR2 5 100uF/50V J26 H8 1 2 3 4 5 6 7 8 J2 H4 C1 E2 K3 A4 VB 47K R51 10K R54 47K R25 4 R26 33K 2 3 - + 2 3 10K LM358 0.1uF U2A WDT D-CTR2 CTR2 DACP VB 10K R28 VM - + R29 H7 1 2 3 4 5 6 7 TP R47 1K C11 0.01uF Spare 10K R43 560R SP 1 LM358 0.1uF U3A C3 1 VB R5 2K2 Reaction Tray Motor VM J29 R55 PVCC R27 10K PVCC NTC1+ R53 33K PVCC MIC5021 8 7 6 5 C13 H4 C1 E2 K3 A4 C10 100uF/50V U6 1 VDD Vbst 2 INPUT Gate 3 Ct Sens4 GND Sens+ R52 10K PVCC C12 0.1uF H8 1 2 3 4 5 6 7 8 J3 Reaction Tray Home Sensor J8 VB CUA CUAN CUB CUBN D2 4.7V R4 10K OPTOCA VM RISCPI NTC3+ R42 560R R3 2K2 Sample Tray Motor Sample Tray Home Sensor J7 VB C5 + VL CAA CAAN CAB CABN R2 10K OPTOFO Sense PVCC PWRES Reaction tray Heater MOLEX-2 J16 MOLEX-3 3 2 1 J15 VL 10K R45 J19 R41 560R INPUT PVCC NTC3+ To Lamp PWS board H3 3 2 1 J31 VB VB Spare - analog Photometer Fan H4 C1 E2 K3 A4 R1 2K2 Filter Wheel Motor Filter Wheel Home Sensor J6 VB H8 1 2 3 4 5 6 7 8 J1 2 DRAIN FOA FOAN FOB FOBN 8 4 1 GATE 4 8 4 H2 LEAKPI 1 2 J17 LED-YELLOW LD1 PWRES H8 1 2 3 4 5 6 7 8 J4 VM H4 C1 E2 K3 A4 1N4148 D7 1N4148 D8 10K 3 R31 33K NTC2+ 47K R30 10K R33 R32 10K PVCC R23 R20 R21 PVCC R44 560R 10K R22 VB 6 5 R7 2K2 Reagent Tray Motor RGT Tray Home Sensor J9 VB FUSE 2A Q1 IRF530 3 SOURCE F1 D3 4.7V R6 10K OPTOCU REA REAN REB REBN 3 470K 1M 2 1 2 1 - + 6 5 LM358 U2B - R34 10K LM358 U3B 7 7 C1 E2 K3 A4 R24 TF OUT GND IN 3 2 1 LEAKPI 2 NTC1+ VB NTC2+ 100uF/50V + 0.1uF C8 C7 H3 3 2 1 Room Temperature Sensor J18 H2 2 1 J14 J27 2 H10X2 PVCC H4 C1 E2 K3 A4 10K 10K H2 Description H3 1 2 3 J33 H3 1 2 3 J24 H4 1 2 3 4 J23 H3 1 2 3 J22 1 2 J21 H3 1 2 3 J20 SE-30-01281-01 Document Number 1 18 17 16 15 14 13 12 11 10 D-CTR2 VB VL VM TP-GND VM MOLEX-6 D-uPSCAMB J28 H4 Spare uPump 1 2 3 4 J12 TP1 VM D-uPSCAMB D-BUZZER 1 Sheet Drawn Checked 1 of 1 R.CORNACCHIA L.M 10/09/2004 0 Rev Checked Approved Approved A.G. 10/09/2004 ULN2803 I1 O1 I2 O2 I3 O3 I4 O4 I5 O5 I6 O6 I7 O7 I8 O8 GND COM D9 4.7V R49 10K OPTOWH2 1 2 3 4 5 6 7 8 9 U1 Plate Interface Board (sx) RMP 224 LIQUID4 R56 10K PVCC LIQUID3 R39 10K PVCC 10K LIQUID2 R38 PVCC LIQUID1 R37 10K PVCC WASTE2 R36 WASTE1 R35 560R R50 R48 2K2 uPSCAMB BUZZER CTR2 Date: 10/09/2004 Size A3 Title 0 Rev. VB Spare - Home Sensor J32 VB PVCC 0.1uF C9 D5 4.7V R10 10K OPTOWH Reaction Tray Temp. Sensor 560R R11 PVCC 100uF/50V + C6 LED-RED LD2 2 VB R9 2K2 Wash Station Motor BUZZER LS1 H4 680R LM7805 U4 H8 1 2 3 4 5 6 7 8 J5 Wash Station Home Sensor J10 VB D-BUZZER 1 D4 4.7V PVCC + WHA WHAN WHB WHBN R8 10K OPTORE 8 4 8 4 5 H13X2 25 23 21 19 17 15 13 11 9 7 5 3 1 26 24 22 20 18 16 14 12 10 8 6 4 2 CAA CAB CUA CUB REA REB FOA FOB WHA WHB 19 17 15 13 11 9 7 5 3 1 20 18 16 14 12 10 8 6 4 2 CAAN CABN CUAN CUBN REAN REBN FOAN FOBN WHAN WHBN 1 1 2 3 4 5 6 + OPTOCU OPTOCA OPTORE OPTOFO OPTOWH WASTE1 WASTE2 LIQUID1 LIQUID2 LIQUID3 LEAKPI RISCPI uPSCAMB CTRLAMP BUZZER TF TP SP LIQUID4 DACP A B C D A B C D 0805 C8 .1uF -V 8 + VS Photodiode OUT SOP8 COM IN IN IN IN R4 5.7K 0805 5 2 3 6 7 1 + VS 3 + 2 C1 .1uF 0805 U5 MC79L05ACD FD1 6 C9 -VS 10uF 7 LM4040 SOT-23 Z1 6 10K 0805 10uF .1uF 0805 C12 100K 3296X C10 TR2 100pF R5 OPA111 TO99 U2 C2 R1 ? M 1/4W 6 C13 .1uF 0805 .1uF 0805 C16 R3 5.7K 0805 TR1 5K 3296X R2 5.7K 0805 2 3 - + 8 0805 C3 .1uF U4 5 8 6 OPA350 SO8 U6 C11 .1uF 0805 5 10uF C14 C15 .1uF 0805 10uF C6 + VD 4 4 C4 .1uF 0805 C7 .1uF 0805 + VS VOUT 1 GND 2-3-6-7 MC78L05ACD SOP8 U1 VOUT 1 GND 2-3-6-7 +12,5 V MC78L05ACD SOP8 + VS +12,5 V Per valore di R1 vedi la relativa soluzione NOTA : 7 7 5 4 8 1 8 7 1 4 5 + VD 1 2 3 4 5 6 7 8 10uF C5 1 2 OSC4P Y1 OSC-HS 16 15 14 13 12 11 10 9 Friday, February 06, 2004 3 Date: [MA]00107-00 Document Number 2 C17 .1uF 0805 Description C18 15pF 0805 10K 0805 10K 0805 DRDY G1 G0 R10 2 R9 R8 10K 0805 PRE-AMPL/ADC Emission + VD R7 10K 0805 1 0805 R6 Size A4 Title 0 Rev. 4 3 DGND G1 G0 CS DRDY CLK SCLK DOUT ADS1250 SOL16 +VIN -VIN AGND VS VREF DSYNC VD DGND U3 + VS 3 1 3 5 7 9 Checked Sheet Drawn 2 4 6 8 10 SCLK DOUT 2 4 6 8 10 1 of 1 1 L.M. 25-05-2002 R.CORNACCHIA 0 Rev Approved -V +12,5 V A.G. 25-05-2002 Checked 1 3 5 7 9 JP1 Approved 10K 0805 R11 TO CASE 1 A B C D A B C 5 3 2 1 MOLEX-3 J1 R1 1K INIT-L C4 0.1uF 4 4 + C1 100uF/50V 3 VIN U1 LT1083/TO + C2 10uF/25V VOUT D1 1N4002 2 1 ADJ 1 D 5 2 3 2 3 3 1 R2 400R R6 100R 3296Y R4 0R R3 120R 1 3 2N3904 2 R5 1K INIT-L C3 100uF/50V Q1 + 2 Date: Size A4 Title 0 Rev. C5 0.1uF 2 Description H2 HALOGEN LAMP SE-30-01576-00 03/01/2005 Document Number Photometer Lamp Board Emission J2 1 2 Sheet Drawn Checked Approved 1 1 of 1 R.CORNACCHIA L.M. 03/01/2005 0 Rev Approved A.G. 03/01/2005 Checked 1 A B C D A B C D VM 2 100nF C13 100R 1/8W R9 5 5 VI 10uF/35V + C5 1 U4 U5 VSS TRG RES VCC GND 2 VO 3 VCC 10uF/35V + C4 VCC 4 C9 GND 100pF 3 C10 100nF 4 R12 10K 10K R8 PROBE 10nF C2 4M7 R16 3 TRG THR 3 U3 VSS Q VCC 100nF C11 R3 1K 0 Rev. Description 2 U2 C12 100nF SE-10-01478-00 Document Number Level Sensor Assy Emission Date: 26/11/2003 Size A4 Title R6 270K R2 220K R1 220K 2 Sheet Drawn Checked 1 R11 10K 1 1 of 1 R.CORNACCHIA L.M. 26/11/2003 A.G. 26/11/2003 0 Rev Approved SIGNAL OUT 10K R10 Checked Q1 Approved R15 10K 1 A B C D Electrovalves Control Board The present document is property of AMS, any use without authorization will be prosecuted accordingly to the law. Description Via E. Barsanti, 17/A (Via Tiburtina Km 18,300) 00012 Guidonia (Roma) P/N 20/10/2005 Sheet Rev. 1 of 1 A. Gagliarducci R. Cornacchia L. Massenzi SC 30-01626-00 A Doc. Date Approved Checked Drawn Rev. A: emissione in riferimento alla RMP n.268 Analytical Control Board 30-01283-.. A B C D 1 2 3 2 1 1N4148 2 1 2 1 VM R3 R6 U2 VIN VM 5 D-uP2H2O D-uP2H2O 100uF/50V C3 CONN CS EV2 H2 J8 + VM 1 SWCOP R12 10K PVCC 1N4148 D2 H3 Main LID Switch J2 H2 LEAK w J3 2 1 2 1 470K 1M 2 1 2 1 2 3 - + 2 VM 3 2 1 0.1 C1 VM 2 1 VM 4 VM D-uP1H2O D-uP1H2O 100uF/50V C2 LEAKWH CONN CS EV4 H2 2 1 J10 + 4 LED-RED LD1 680R R5 PVCC D-uPBAND D-uPBASICP VM 6 TLV2461CP U3 H3 J7 D-uPBASICP CONN CS EV3 H2 J9 VOUT LM7805/TO92 10K R7 10K R4 PVCC 7 4 D1 GND 3 5 2 1 VM VM R2 10K WASTE2 10K WASTE1 2 1 2 1 VM 3 VM D-EV2H2O VM D-EV3PP D-EV3PP CONN CS EV6 H2 2 1 J12 CONN CS EV8 2 1 VM PVCC VM D-EV1H2O D-EV1H2O VM D-EV2H2O CONN CS EV7 H2 2 1 J13 H2 2 1 J14 R10 10K LIQUID3 WASTE1 PVCC LIQUID1 PVCC 10K 10K H2X5 10 9 9 8 7 7 6 5 5 4 3 3 2 1 1 J6 R8 10 8 6 4 2 H4 1 2 3 4 J5 H4 1 2 3 4 J4 R9 PVCC LIQUID2 R11 10K LIQUID4 WASTE2 PVCC D-EV2PP D-EV1PP CONN CS EV5 H2 2 1 J11 PVCC PVCC R1 3 2 2 U1 Q1 1 3 SE-30-01626-00 Document Number 1 VM Sheet Drawn Checked 1 1 of 1 R.CORNACCHIA L.M. 07/09/2004 A.G. 07/09/2004 Checked VM D3 UF4004 0 Rev Approved D-uP1H2O D-uP2H2O D-uPBAND D-uPBASICP D-EV1PP D-EV2PP D-EV3PP D-EV1H2O D-EV2H2O 18 17 16 15 14 13 12 11 10 EV3PP EV1H2O uPBASICP EV1PP uP1H2O SWCOP WASTE1 LIQUID3 LIQUID1 VM Approved ULN2803 O1 O2 O3 O4 O5 O6 O7 O8 COM H2X12 23 23 21 21 19 19 17 17 15 15 13 13 11 11 9 9 7 7 5 5 3 3 1 1 I1 I2 I3 I4 I5 I6 I7 I8 GND 24 22 20 18 16 14 12 10 8 6 4 2 J1 2N2222 1 2 3 4 5 6 7 8 9 24 22 20 18 16 14 12 10 8 6 4 2 Description R14 10K R13 220R uP1H2O uP2H2O uPBAND uPBASICP EV1PP EV2PP EV3PP EV1H2O EV2H2O uPBAND EV2PP uP2H2O LEAKWH WASTE2 LIQUID4 LIQUID2 VM Electrovalves Control Board Emission EV2H2O Date: 07/09/2004 Size A4 Title 0 Rev. 2 1 2 A B C D A B C D 5 5 VB 1 2 0 0 4 PV PR PG VB 3 2 1 PV D-LAMPV 10 8 6 4 2 9 7 5 3 1 3 H 5X2 10 8 6 4 2 J1 9 7 5 3 1 D-LAMPV D-LAMPR D-LAMPG VB PUSHBUTTON-LED Green 0 0 PG SW2 + VB D-LAMPG PUSHBUTTON-LED Yellow - + SW1 4 Green LED Red LED Yellow LED VB 2 1 - 2 Date: Size A4 Title 0 Rev. 25/02/2005 SE-30-01850-00 Document Number Control Panel Board Emission Description PR D-LAMPR PUSHBUTTON-LED Red 0 0 + SW3 2 Sheet Drawn Checked Approved 1 1 of 1 R.CORNACCHIA L.M. 25/02/2005 0 Rev Approved A.G. 25/02/2005 Checked 1 A B C D Chapter 05 – DIAGNOSTIC PROGRAM CHAPTER 05 - DIAGNOSTIC PROGRAM - INDEX 5 DIAGNOSTIC PROGRAM 3 5.1 STATUS BAR 6 5.2 TEST FOLDER 6 5.3 “ARM” FOLDER 7 5.3.1 VERTICAL COMMAND AREA 8 5.3.2 HORIZONTAL COMMAND AREA 8 5.3.3 STATUS COMMAND AREA 9 5.3.4 ARM SETTING COMMAND AREA 9 5.4 “DILUTER” FOLDER 10 5.4.1 OPERATIONS AREA 10 5.4.2 SENSOR STATE AREA 11 5.4.3 USER AREA 11 5.4.4 STATUS COMMAND AREA 11 5.5 “OPTIC” FOLDER 11 5.5.1 OPERATIONS AREA 12 5.5.2 SENSOR STATE AREA 12 5.5.3 FILTERS AREA 12 5.5.4 DIGITAL CONVERSION AREA 13 5.5.5 STATUS COMMAND AREA 13 5.6 “PLATE” FOLDER 13 5.6.1 OPERATIONS AREA 14 5.6.2 MOTOR ON/OFF AREA 15 5.6.3 STATUS COMMAND AREA 15 Ellipse Service Manual Rev. 03 01 June 2009 Page 1 Chapter 05 – DIAGNOSTIC PROGRAM 5.7 “TEMPERATURES” FOLDER 15 5.7.1 TEMPERATURES AREA 16 5.7.2 OFFSET AREA 16 5.7.3 STATUS COMMAND AREA 17 5.8 “WASH” FOLDER 18 5.8.1 OPERATIONS AREA 18 5.8.2 TEST AREA 19 5.8.3 STATUS COMMAND AREA 19 5.9 “USER” FOLDER 20 5.9.1 USER TEST MICRO 2 AREA 20 5.9.2 STATUS COMMAND AREA 21 5.10 “MISCELLANEOUS” FOLDER 21 5.10.1 GENERAL RESET AREA 22 5.10.2 LOG RAW DATA AREA 22 5.10.3 CONTINUOUS TEST RUN 22 5.10.4 STATUS COMMAND AREA 23 5.11 “GLOBAL SETTING” FOLDER 23 5.12 “MACRO SETTING” FOLDER 24 5.13 “CONFIGURATION” FOLDER 25 Ellipse Service Manual Rev. 03 01 June 2009 Page 2 Chapter 05 – DIAGNOSTIC PROGRAM 5 DIAGNOSTIC PROGRAM The diagnostic program enables the operator to perform a complete check of each of the Ellipse’ module functions. This program has a folder structure, with each folder containing functions pertaining to the specific module. To launch the program the operator has to click with the left side of the mouse, on the Diagnostic area located on the lower right side of the screen of the System Monitor (Fig.1). The Ellipse must be in the stand-by state to access this area. Fig. 1 Click Here Once the program starts, the following screen appears (Fig. 2) Fig. 2 Test Folders Optical Sensor Status Status Bar The diagnostic program is subdivided into three distinct areas: test folders, optical sensor status, status bar. The last two areas are present on every diagnostic window. The three areas enable the operator to verify multiple functions as specified below: Ellipse Service Manual Rev. 03 01 June 2009 Page 3 Chapter 05 – DIAGNOSTIC PROGRAM Test Folders: checks the functionality of the Ellipse’ various sub-systems; The individual functions are illustrated later on. Optical Sensor Status: visualizes the sensor status that controls the Sampling Arm, the Diluter, the Washing Station, the Optical filter wheel, the Reaction plate, the Sample plate, the Reagent plate and the Cover Lock, as well as the Liquid Reservoir Alarm, the Front panel button and the Level Sensor Test (Fig. 3) Fig. 3 Furthermore, there are three types of unused fields for future use: Leak for Arm, Washing Station and Plate. While the program is running, the fields given in the figure can be either of two distinct colors, red or green. The fields: • Vertical Arm’s Optical Switch (O.S.) • Horizontal Arm’s Optical Switch (O.S.) • Diluter’s Optical Switch (O.S.) • Wash Station’s Optical Switch (O.S.) • Filter Wheel’s Optical Switch (O.S.) • Reaction Plate’s Optical Switch (O.S.) • Sample Plate’s Optical Switch (O.S.) • Reagent Plate’s Optical Switch (O.S.) turn green when the specific device is in the home position, and turn red when not in the home position. The field • Cover Lock turns green when the cover is closed, and turns red when open. Ellipse Service Manual Rev. 03 01 June 2009 Page 4 Chapter 05 – DIAGNOSTIC PROGRAM The fields: • Arm Leak Alarm (Arm Leaking) • Wash Station Leak Alarm (Wash Station Leaking) • Plate Leak Alarm (Reaction Plate Leaking) Turn green when there is no leaking, and red when there is. The following six fields signal the alarms for the bottles containing the washings liquids and waste materials: • Waste1 Liquid Reservoir Alarm • Waste2 Liquid Reservoir Alarm Turning green when the liquid level is not excessive, and red when the waste bottle indicated by the alarm is almost full. • Liquid 1 Liquid Reservoir Alarm • Liquid 2 Liquid Reservoir Alarm • Liquid 3 Liquid Reservoir Alarm • Liquid 4 Liquid Reservoir Alarm Turning green if the liquid level is above the predetermined minimum limit, and red if the liquid quantity is below the predetermined minimum level (the bottle is almost empty). The following fields signal the state of the button on the Ellipse’s operator panel: • Al.Res. Front Panel Button • St-By Front Panel Button • Ready Front Panel Button Turning green when the button is not pushed in, and red if it is. The field • Test ON It is green fixed; if pushed, on the right of it will be showed a red drop that becomes temporarily green when the liquid is detected. Ellipse Service Manual Rev. 03 01 June 2009 Page 5 Chapter 05 – DIAGNOSTIC PROGRAM 5.1 STATUS BAR Monitors the instrument’s status (Fig.4). Fig. 4 Time Firmware Rel. Al. Res Status Indicator Date Instrument Status Ready Status Indicator Start/Stop Diagnostics 5.2 TEST FOLDER The following is a summary of each individual test folder’s function. The Ellipse’s diagnostic program is subdivided into 10 folders: Arm: the adjustment of the arm position can be checked and carried out; Diluter: checks if the diluter is functioning correctly; Optic: checks if the optical block is functioning correctly; Plate: checks if the reaction, samples and reagents plates are functioning correctly; Temperature: checks the temperature of the pre-heater and reaction plate; Wash: checks if the wash station is functioning correctly; User: checks if the user functions are working properly; Miscellaneous: performs general checks; Macro Setting: the operator can identify which macros have been loaded; Configuration: Through the use of a password, only qualified technical personal can adjust certain parameters (This file is not accessible to the Laboratory operator). Ellipse Service Manual Rev. 03 01 June 2009 Page 6 Chapter 05 – DIAGNOSTIC PROGRAM 5.3 “ARM” FOLDER Warning: the improper use of the functions described in this folder can damage the sampling probe. The Arm File is subdivided into four areas: ♦ VERTICAL ♦ HORIZONTAL ♦ STATUS COMMAND ♦ ARM SETTING Ellipse Service Manual Rev. 03 01 June 2009 Page 7 Chapter 05 – DIAGNOSTIC PROGRAM 5.3.1 VERTICAL COMMAND AREA (5 commands) Home: Brings the arm vertically to a home position; Down Well: Brings the arm’s axis “z” to the height of the well; Up: Raises the arm tenths of a millimeter as preset in the square found immediately on the right, with relation to the previous Down movement; the box becomes active when the arm is positioned on a sample, a reagent, a standard or a cuvette. Down: Lowers the arm tenths of a millimeter as pre-determined by the square found immediately on the right; the box becomes active when the arm is positioned on a sample, a reagent, a standard or a cuvette. Go Level: Brings the probe to either the reagent level, the sample or the standard and reads the level. Before giving this command, it is necessary to set the parameter for the probe’s maximum lowering level in the max area (mm); 5.3.2 HORIZONTAL COMMAND AREA (5 commands) Back to wash well: Brings the arm back to the home position; Go Reag.: Brings the arm to the reagent position specified in the sheet menu next to the command; Go Sample: Brings the arm to the sample position specified in the sheet menu next to the command; Go Std.: Brings the arm to the standard/control position specified in the sheet menu next to the command; Go Disp.: Ellipse Brings the arm to the dispense position; Service Manual Rev. 03 01 June 2009 Page 8 Chapter 05 – DIAGNOSTIC PROGRAM 5.3.3 STATUS COMMAND AREA Send: Visualizes the commands sent from the program to the instrument hardware; Receive: Visualizes the answers of the instrument’s hardware to the commands sent by the program; Sensor State: Indicates the sensor status relative to the arm’s vertical axis, the internal arm and the external arm. 5.3.4 ARM SETTING COMMAND AREA (5 commands) Step: Visualizes the number of steps. +: Performs the counter-clockwise rotation of the arm (to adjust the position); - : Performs the clockwise rotation of the arm (to adjust the position); Save: Saves the adjustment data; Reset: Completely resets the Arm In this area the boxes can be activated when the arm is positioned on a sample, a reagent, a standard or a control or a cuvette. Ellipse Service Manual Rev. 03 01 June 2009 Page 9 Chapter 05 – DIAGNOSTIC PROGRAM 5.4 “DILUTER” FOLDER The Diluter Folder is subdivided into four areas: ♦ OPERATIONS ♦ SENSOR STATE ♦ USER ♦ STATUS COMMAND 5.4.1 OPERATIONS AREA (3 commands) Home: Brings the diluter to the Home position; Asp: Aspirates the micro-litres as predetermined in the ul. box (may be set manually through + and – or by using the cursor located immediately above); Disp.: Distributes the micro-litres as predetermined in the ul. box (may be set manually through + and – or by using the cursor located immediately above); Ellipse Service Manual Rev. 03 01 June 2009 Page 10 Chapter 05 – DIAGNOSTIC PROGRAM 5.4.2 SENSOR STATE AREA In the Sensor State area the diluter sensor status is indicated. 5.4.3 USER AREA (1 command) Probe Wash Test: Activates and deactivates the probe wash system 5.4.4 STATUS COMMAND AREA Send: Visualizes the commands sent by the program to the instrument hardware; Receive: Visualizes the instrument hardware’s response to the commands sent by the program. 5.5 “OPTIC” FOLDER Ellipse Service Manual Rev. 03 01 June 2009 Page 11 Chapter 05 – DIAGNOSTIC PROGRAM The Optic folder is subdivided into five areas: ♦ OPERATIONS ♦ SENSOR STATUS ♦ FILTERS ♦ DIGITAL CONVERSION ♦ STATUS COMMAND 5.5.1 OPERATIONS AREA (2 commands) Lamp ON: Turns on the Optic Lamp; Lamp OFF: Turns off the Optic Lamp; 5.5.2 SENSOR STATE AREA This area indicates the status of the wheel filters. 5.5.3 FILTERS AREA (10 commands) Ellipse Reset : Positions the wheel filters in home position (dark); 340 nm: Positions filter n° 1 in front of the reading sensor; 380 nm: Positions filter n° 2 in front of the reading sensor; 405 nm: Positions filter n° 3 in front of the reading sensor;; 492 nm: Positions filter n° 4 in front of the reading sensor; 510 nm: Positions filter n° 5 in front of the reading sensor; 546 nm: Positions filter n° 6 in front of the reading sensor; 577 nm: Positions filter n° 7 in front of the reading sensor; 620 nm: Positions filter n° 8 in front of the reading sensor; Spare: Positions filter n° 9 in front of the reading sensor (Optional); Service Manual Rev. 03 01 June 2009 Page 12 Chapter 05 – DIAGNOSTIC PROGRAM 5.5.4 DIGITAL CONVERSION AREA (2 commands) START/STOP Sample Channel Conversion: Performs an analogical/digital conversion of the principal channel; START/STOP Reference Channel Conversion: Performs an analogical/digital conversion of the reference channel; 5.5.5 STATUS COMMAND AREA Send: Visualizes the commands sent by the program to the instrument hardware; Receive: Visualizes the instrument hardware’s response to the commands sent by the program. 5.6 “PLATE” FOLDER Ellipse Service Manual Rev. 03 01 June 2009 Page 13 Chapter 05 – DIAGNOSTIC PROGRAM The Plate Folder is subdivided into three areas: ♦ OPERATIONS ♦ ALL MOTORS ON/OFF ♦ STATUS COMMAND 5.6.1 OPERATIONS AREA Reaction Plate area: (3 commands) To Disp.: Automatically positions the cuvette, indicated on the sheet menu aside, under the dispensing position To Optic: Automatically positions the cuvette, indicated on the sheet menu aside, in front of the colorimeter Reset: Performs the reset of the reaction plate by positioning the # 1 cuvette in the dispensing position Sample Plate area: (2 commands) To Asp.: Automatically brings sample, indicated on the sheet menu aside, under the aspiration position Reset: Performs the reset of the sample plate by positioning the # 1 sample in the aspiration position Reagent Plate: (2 commands) To Asp.: Automatically brings reagent, indicated on the sheet menu aside, under the aspiration position Reset: Performs the reset of the sample plate by positioning the # 1 reagent in the aspiration position Ellipse Service Manual Rev. 03 01 June 2009 Page 14 Chapter 05 – DIAGNOSTIC PROGRAM 5.6.2 ALL MOTORS ON/OFF AREA (1 command) ALL Motors ON/OFF: Engages/Disengages the motors to allow a manual movement of the modules (Sampling Arm; Reaction, Reagents and samples Plates, Washing Station, Filter Wheel and Peristaltic Pump). 5.6.3 STATUS COMMAND AREA Send: Visualizes the commands sent by the program to the instrument hardware; Receive: Visualizes the instrument hardware’s response to the commands received from the program. 5.7 “TEMPERATURES” FOLDER Ellipse Service Manual Rev. 03 01 June 2009 Page 15 Chapter 05 – DIAGNOSTIC PROGRAM The Temperature Folder is subdivided into three areas: Temperatures Offset Status Command 5.7.1 TEMPERATURES AREA (4 commands) Temp. Test Run: To enter in the Temperature Test folder Plate: Performs a temperature reading in the reaction plate compartment; Arm: Performs a temperature reading in the arm’s pre-heater. Calib. Sensor: To carry out the calibration procedure of the thermometric probe included in the kit P/N 55-01204-00, as illustrated in Chapter Six “Settings and Adjustments”. 5.7.2 OFFSET AREA (2 commands) To correlate the thermometric probe included in the kit P/N 55-01204-00 with another reference thermometric probe. Service probe: The offset value that has been previously inserted and the temperature value read by the thermometric probe appear in the closed fields. Cursor: To enter the offset value ± 3 C°. The offset value will be zero after exiting from the folder. Ellipse Service Manual Rev. 03 01 June 2009 Page 16 Chapter 05 – DIAGNOSTIC PROGRAM TEMPERATURE TEST AREA (3 commands) START: To perform the test check and to regulate the temperature of the preheater and of the reaction plate by using the kit P/N 55-01204-00, as illustrated in Chapter 6 “Setting and Adjustments”. The number of sample cycles are predefined, whereas the reagent and sample positions for test execution can be selected by the operator. Service Probe: The field in which the temperature of the service probe appears. WASH: To run a wash cycle on the cuvettes used for testing. EXIT: To exit from the Temperature Test box and to run a wash cycle on the cuvettes if they have been used. 5.7.3 STATUS COMMAND AREA Ellipse Send: Visualizes the commands sent by the program to the instrument hardware; Receive: Visualizes the instrument hardware’s response to the commands sent by the program. Service Manual Rev. 03 01 June 2009 Page 17 Chapter 05 – DIAGNOSTIC PROGRAM 5.8“WASH” FOLDER The Wash folder is subdivided into three areas: ♦ OPERATIONS ♦ TEST ♦ STATUS COMMAND 5.8.1 OPERATIONS AREA (2 commands) Home: Brings the washing device to the Home position; Down: Makes the washing device go down to the washing position. Ellipse Service Manual Rev. 03 01 June 2009 Page 18 Chapter 05 – DIAGNOSTIC PROGRAM 5.8.2 TEST AREA (5 commands) Test P5 + E.V. 5: Performs switching on of the E.V. 5 solenoid valve and turns on the P5 pump and automatically switching off the E.V. 5 solenoid valve and turns off the P5 pump after a period of four seconds Test P5: It turns on the µP5 pump and automatically turns it off after a period of four seconds. Test P6–Cuvette Dry:It turns on the µP6 pump and automatically turns it off after a period of four seconds. Peristaltic Pump: It turns on the peristaltic pump and automatically turns it off after one second. Washing Cycle: It performs the functional check of the Washing Station , as illustrated in Chapter Six “Settings and Adjustments”. 5.8.3 STATUS COMMAND AREA Send: Visualizes the commands sent by the program to the instrument hardware; Receive: Visualizes the instrument hardware’s response to the commands sent by the program. Ellipse Service Manual Rev. 03 01 June 2009 Page 19 Chapter 05 – DIAGNOSTIC PROGRAM 5.9 “USER” FOLDER The User Folder is subdivided into two areas: ♦ USER TEST MICRO ♦ STATUS COMMAND 5.9.1 USER TEST MICRO AREA (3 commands) Alarms ON: Turns on and off the Alarms indicator lamp; Ready ON: Turns on and off the READY indicator lamp; Test BUZZER: Performs a complete operational test of the buzzer; Ellipse Service Manual Rev. 03 01 June 2009 Page 20 Chapter 05 – DIAGNOSTIC PROGRAM 5.9.2 STATUS COMMAND AREA Send: Visualizes the commands sent by the program to the instrument hardware; Receive: Visualizes the instrument hardware’s response to the commands sent by the program. 5.10 “MISCELLANEOUS” FOLDER The Miscellaneous folder is subdivided into four areas: - General Reset - Log Raw data - Continuous Test Run - Status Command Ellipse Service Manual Rev. 03 01 June 2009 Page 21 Chapter 05 – DIAGNOSTIC PROGRAM 5.10.1 GENERAL RESET AREA (1 command) Reset: Performs the resetting of the general system. 5.10.2 LOG RAW DATA AREA (1 command) Enable: To save a copy of the files for the technical diagnosis if the instrument malfunctions. The files are memorized in the folder “Log” inside the folder “Analyzer”. 5.10.3 CONTINUOUS TEST RUN (1 command) Start Test: Performs an automatic test during which the modules (Sampling Arm, Reaction plate, Washing Station and Photometer) are moved to all the different positions. The correct modules positioning is also automatically verified and showed on the following table: To stop the test push on the “Exit Test” key Ellipse Service Manual Rev. 03 01 June 2009 Page 22 Chapter 05 – DIAGNOSTIC PROGRAM 5.10.4 STATUS COMMAND AREA Send: Visualizes the commands sent by the program to the instrument hardware Receive: Visualizes the instrument hardware’s response to the commands. 5.11 “GLOBAL SETTING” FOLDER The Global Setting Folder is for next application. Ellipse Service Manual Rev. 03 01 June 2009 Page 23 Chapter 05 – DIAGNOSTIC PROGRAM 5.12 “MACRO SETTING” FOLDER This folder allows the user to read the macro present in the Ellipse. To read the macro follow the below procedure: ⇒ Select from the menu sheet “Read Macro” the number of relevant macro; ⇒ Click with the left side of the mouse on the Read area; ⇒ If the Macro requested exists, the information will appear in the designated area. Ellipse Service Manual Rev. 03 01 June 2009 Page 24 Chapter 05 – DIAGNOSTIC PROGRAM 5.13 “CONFIGURATION” FOLDER This folder may be accessed only by qualified technical assistant. It is necessary for the technical assistant to insert a password to access this folder. The Configuration Folder is subdivided into four areas: • Gap • Backlash Dil. • Heater • Wash. Well level Test Ellipse Service Manual Rev. 03 01 June 2009 Page 25 Chapter 05 – DIAGNOSTIC PROGRAM Gap: defines the volume, in µL, of the air bubble that separates the Rinse (the liquid column into the sampling arm tube) from the sampled reagent; this volume must be set in the range from 10 to 24 (Standard value 16). For the correct setting of the gap volume it is necessary to apply the procedure of BIB (Precision check of the Analytical plate) and BIC (Volume precision check of the Sampling line). If BIC’s CV values should exceed the limit specified below, the GAP has to be increased by steps of 2 µL (starting from 10 µL) until obtaining the expected BIC’s CV value Precision check of the Analytical plate BIB Prepare two series of 32 samples each by using a diluted solution of K2Cr2O7 (obtained from 4 gr/l concentrated solution diluted 1:100 with Rinse solution (distilled water + 1000 µL/litre of Brij)) Use the same solution as samples and reagent Program a new method with the following parameters: Method name: BIB Type: End Point Sample volume K2Cr2O7 0,04 gr/l 3 µL Reagent volume K2Cr2O7 0,04 gr/l 300 µL Rinse volume 20 µL Filter: 340 nm Incubation time: 116 sec Calculation factor: 1000 BIB Procedure 1. Select conical cup in Options 2. Program the BIB method using the above parameters and configure a reagent rack containing it 3. Enter in the calibration screen and select RBL for BIB method. 4. Put the Rinse solution as reagent and standard (fourteenth position of the Crt/Std rack) and Start the Calibration Ellipse Service Manual Rev. 03 01 June 2009 Page 26 Chapter 05 – DIAGNOSTIC PROGRAM 5. Program 64 BIB; fill 32 conical samples cup with K2Cr2O7 0,04 gr/l 6. Load the samples racks into the Analyzer at positions 1 and 2 7. Substitute the Rinse Solution with K2Cr2O7 0,04 gr/l in the reagent rack 8. Start the Analyzer. At the end of the last sampling, move the samples racks to positions 3 and 4. The mean of the reading obtained from each of the 32 sample series must fall within 360 ÷ 400. The coefficient of variation (CV %) must be lower than 0.7 % for each series of 32 samples. Precision check of the Sampling volume BIC Prepare two series of 32 samples each by using K2Cr2O7 4 gr/l (concentrated) Put Rinse Solution as reagent and K2Cr2O7 solution as sample. Program a new method with the following parameters: Method name: BIC Type: End Point Sample volume K2Cr2O7 4 gr/l 3 µL Reagent volume RINSE Solution (distilled water + 1000 µL/litre of Brij) 300 µL Filter: 340 nm Incubation time: 116 sec Calculation factor: 1000 BIC Procedure 1. Program the BIC method using the above parameters and configure a reagent rack 2. Enter in the Calibration screen and select RBL for BIC method. 3. Put the Rinse Solution as reagent and Standard (fourteenth position of the Crt/Std rack) and Start the Calibration 4. Program 64 BIC ; fill 32 conical samples cup with K2Cr2O7 4 gr/l (concentrated) 5. Load the samples racks into the Analyzer at positions 1 and 2 6. Start the Analyzer. At the end of the last sampling, move the sample racks to positions 3 and 4. Ellipse Service Manual Rev. 03 01 June 2009 Page 27 Chapter 05 – DIAGNOSTIC PROGRAM The mean of the readings obtained from each of the 32 sample series must fall between 350 ÷ 450. The coefficient of variation (CV %) must be lower than 1.6 % for each series of 32 samples If BIC’s CV values should exceed the limit specified above, the GAP has to be increased by steps of 2 µL (starting from 10 µL) until obtaining the expected BIC’s CV value Backlash Dil.: Allows to recover the sampling imprecision generated by the motor when it inverts its rotation; the setting is expressed in µL and obtained by using the following procedure: 1. Enter in the Diagnostic Arm folder menu and make the Sampling Arm “Reset” 2. Enter in the Diluter folder, press the “HOME” button and then Push “PROBE WASH TEST” button three times in order to fill the hydraulic sampling line 3. Remove the plastic cover of the sample racks vane and take out the samples racks 4. Enter in the ARM folder and push the “GO” button in order to move the probe to sample position # 1 5. Put a sample cup containing distilled water (about 1ml) under the sampling probe 6. Enter in the DILUTER folder and aspirate 300 µL of distilled water from the cup 7. Remove the sample cup containing distilled water 8. Select 1 µL to dispense and press the “Disp” key repeatedly until water appears on the tip of the probe. 9. Substract 1 from the number of times you pressed the “Disp” key. The obtained number has to be introduced as BACKLASH value 10. Push the Save button to memorize the inserted data 11. Push the Reset button into the ARM folder 12. Close the Diagnostic program by clicking on the Diagnostic button Ellipse Service Manual Rev. 03 01 June 2009 Page 28 Chapter 05 – DIAGNOSTIC PROGRAM Heater: to adjust the preheater and reaction plate temperatures; Wash. Well level Test: By clicking on the “Run Test” button the system performs an automatic verification and adjustment cycle for the liquid level in the washing well. A number appears in the box “Level” that corresponds to the rotation time of the rinse pump in milliseconds, that is necessary to maintain the correct liquid level in the washing well. The functioning range is between 350 and 700 milliseconds. If the washing well is not sufficiently filled, a message appears to warn the user to check the hydraulic line. If the value taken during the test is different from the previous one, it is necessary to click on “Save” in order to save the new data. Ellipse Service Manual Rev. 03 01 June 2009 Page 29 Chapter 06 – SETTINGS AND ADJUSTMENTS CHAPTER 06 - SETTING AND ADJUSTMENTS INDEX 6 SETTING AND ADJUSTMENTS 6.1 4 SAMPLING ARM 4 6.1.1 ALIGNMENT AND ADJUSTMENTS 4 6.1.2 REPLACEMENT OF THE SAMPLING PROBE 9 6.1.3 REPLACEMENT OF THE SAMPLING ARM 13 6.1.4 16 REPLACEMENT OF THE ARM HOME SENSORS 6.1.4.1 Replacement of the Horizontal Arm Home Sensor 16 6.1.4.2 Replacement of the Vertical Arm Home Sensor 18 6.1.5 REPLACEMENT OF THE SAMPLING ARM MOTORS 19 6.1.5.1 Replacement of the Horizontal Arm Motor 19 6.1.5.2 Replacement of the Vertical Sampling Arm Motor 22 6.1.6 REPLACEMENT OF THE SAMPLING ARM BELTS 24 6.1.6.1 Replacement of the Horizontal Sampling Arm Belt 24 6.1.6.2 Replacement of the Vertical Sampling Arm Belt 26 6.2 27 DILUTER MODULE 6.2.1 REPLACEMENT OF THE DILUTER MODULE 27 6.2.2 32 REPLACEMENT OF THE DILUTER MICROPUMP 6.2.3 REPLACEMENT OF THE DILUTER ELECTROVALVE 35 6.2.4 REPLACEMENT OF THE DILUTER HOME SENSOR 36 6.3 PRE-HEATER AND LEVEL SENSOR 38 6.3.1 SUBSTITUTION OF THE PRE-HEATER AND LEVEL SENSOR ASSEMBLY 38 6.3.2 LEVEL SENSOR – CHECK AND ADJUSTMENT PROCEDURE 40 6.4 44 OPTIC ASSEMBLY 6.4.1 SUBSTITUTION OPTIC ASSEMBLY 44 6.4.2 SUBSTITUTION OF THE OPTIC LAMP 46 6.4.3 ELECTRONIC ADJUSTMENT OF THE OPTIC 47 6.4.4 SUBSTITUTION OF THE OPTO SENSOR 49 6.4.5 SUBSTITUTION OF THE MOTOR 50 6.4.6 SUBSTITUTION OF THE OPTIC FILTERS 51 6.5 52 PLATES 6.5.1 REMOVAL OF THE PLATES ASSEMBLY Ellipse Service Manual 52 Rev. 04 01 October 2009 Page 1 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5.2 REPLACEMENT OF THE REAGENTS PLATE MOTOR 58 6.5.3 REPLACEMENT OF THE SAMPLES PLATE MOTOR 62 6.5.4 REPLACEMENT OF THE REACTIONS PLATE MOTOR 65 6.5.5 REPLACEMENT OF THE REAGENTS PLATE BELT 67 6.5.6 REPLACEMENT OF THE SAMPLES PLATE BELT 68 6.5.7 REPLACEMENT OF THE REACTIONS PLATE BELT 70 6.5.8 REPLACEMENT OF THE REAGENTS PLATE PULLEY 72 6.5.9 REPLACEMENT OF THE SAMPLES PLATE PULLEY 73 6.5.10 EPLACEMENT OF THE REACTIONS PLATE BELT 73 6.5.11 REPLACEMENT OF THE REAGENTS PLATE HOME SENSOR 75 6.5.12 REPLACEMENT OF THE SAMPLES PLATE HOME SENSOR 77 6.5.13 REPLACEMENT OF THE REACTIONS PLATE HOME SENSOR 78 6.5.14 ALIGNMENT AND ADJUSTMENT OF THE REAGENTS PLATE 79 6.5.14.1 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate motor or of the Reagents Plate belt 79 6.5.14.2 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate pulley 80 6.5.14.3 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate Home Sensor 82 6.5.15 ALIGNMENT AND ADJUSTMENT OF THE SAMPLE PLATE 83 6.5.15.1 Alignment and adjustment of the Sample Plate in case of replacement of the Samples Plate motor or of the Samples Plate belt 83 6.5.15.2 Alignment and adjustment of the Samples Plate in case of replacement of the Samples Plate pulley 84 6.5.15.3 Alignment and adjustment of the Samples Plate in case of replacement of the Samples Plate Home Sensor 85 6.5.16 86 ALIGNMENT AND ADJUSTMENT OF THE REACTIONS PLATE 6.5.16.1 Alignment and adjustment of the Reaction Plate in case of replacement of the Reaction Plate motor or of the Reaction Plate belt 86 6.5.16.2 Alignment and adjustment of the Reactions Plate in case of replacement of the Reactions Plate pulley 88 6.5.16.3 Alignment and adjustment of the Reactions Plate in case of replacement of the Reactions Plate Home Sensor 90 6.6 WASHING STATION ASSEMBLY 92 6.6.1 SUBSTITUTION OF THE WASHING STATION ASSEMBLY 92 6.6.2 SUBSTITUTION OF THE WASHING STATION HOME SENSOR 94 6.6.3 SUBSTITUTION OF THE WASHING STATION MOTOR 95 6.6.4 WASHING STATION ASSEMBLY - MECHANICAL ADJUSTMENT 96 6.6.5 WASHING STATION - FUNCTIONAL CHECK 97 6.7 WASH PUMP ASSEMBLY 98 6.7.1 SUBSTITUTION OF THE WASH PUMP ASSEMBLY 98 6.7.2 SUBSTITUTION OF PART(S) 101 6.7.3 FUNCTION CHECK FOR THE WASH PUMP ASSEMBLY 101 6.8 PERISTALTIC PUMP ASSEMBLY 103 6.8.1 SUBSTITUTION OF THE PERISTALTIC PUMP ASSEMBLY 103 Ellipse Service Manual Rev. 04 01 October 2009 Page 2 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.8.2 SUBSTITUTION OF THE PERISTALTIC PUMP TUBES 104 6.8.3 SUBSTITUTION OF THE PERISTALTIC PUMP MOTOR 106 6.8.4 FUNCTION CHECK FOR THE PERISTALTIC PUMP 106 6.9 SUBSTITUTION OF THE ELECTRONICS BOARDS 107 6.9.1 ANALYTICAL CONTROL BOARD Cd. 30-01283-01 107 6.9.2 STEPPER MOTORS DRIVER BOARD 108 6.9.3 ARM INTERFACE BOARD Cd. 30-01282-01 109 6.9.4 PLATE INTERFACE BOARD Cd. 30-01281-01 110 Ellipse Service Manual Cd. 30-01284-01 Rev. 04 01 October 2009 Page 3 Chapter 06 – SETTINGS AND ADJUSTMENTS 6 SETTINGS AND ADJUSTMENTS 6.1 SAMPLING ARM 6.1.1 ALIGNMENT AND ADJUSTMENTS N.B.: Make sure that the instrument (Ellipse) is turned on before performing this procedure. 1. Launch the “Diagnostic” program. 2. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 3. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal), Reactions, Reagents and Samples Plates light up in green. 4. Make sure that the Sampling Probe is well-placed on the Sampling Arm (see par. 6.1.2). 5. Make sure that the Sampling Probe is centered with respect to the Washing Well (note: the centering of the Sampling Probe with respect to the Washing Well must be checked on the entire stroke of the Sampling Arm). If it not be so, adjust in the following way: a. remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1); b. loosen the set screw of the Horizontal Arm Home Sensor flat bar (Fig. 2); c. rotate the flat bar in the right toward in order to obtain the centering; d. tighten the set screw of the Horizontal Arm Home Sensor flat bar; e. select “Arm” folder and click the “Reset” button. Wait until the reset arm procedure has been completed; f. verify the accurate centering of the Sampling Arm with respect to the Washing Well. If it not be so, repeat the steps from 5.b to 5.f . Ellipse Service Manual Rev. 04 01 October 2009 Page 4 Chapter 06 – SETTINGS AND ADJUSTMENTS Sampling Arm protective cover Washing Well Diluter protective cover Sampling Arm protective cover Sampling fastening Arm screw Diluter protective cover fastening screw Fig. 1 – Protective covers Slit Horizontal Arm Home Sensor Horizontal Arm Home Sensor flat bar Horizontal Arm Home Sensor fastening screw Set screw of the Horizontal Arm Home Sensor flat bar Preheater fastening screw Fig. 2 – Horizontal Arm Home Sensor 6. verify the verticality of the Sampling Probe with respect to the Washing Well. If it not be so, adjust in the following way: Ellipse Service Manual Rev. 04 01 October 2009 Page 5 Chapter 06 – SETTINGS AND ADJUSTMENTS a. loosen slightly the Screw “A” in order to obtain a longitudinal motion of the Sampling Arm head (Fig. 3) and/or loosen slightly the Screw “B” (or “Preheater Head fastening screw”) in order to obtain a latitudinal motion of the Sampling Arm head (Fig. 4); b. center the Sampling Probe with respect to the Washing Well (checking its verticality for the entire stroke of the Sampling Arm); c. tighten the Screw “A” and/or the Screw “B”; d. verify the accurate verticality of the Sampling Probe with respect to the Washing Well. If it not be so, repeat the steps from 6.a to 6.d . Screw “A” Longitudinal motion Fig. 3 – Longitudinal motion of the Sampling Arm head Latitudinal motion Screw “B” Ellipse Service Manual Rev. 04 01 October 2009 Page 6 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 4 – Latitudinal motion of the Sampling Arm head 7. Make sure that the distance between the top edge of the Washing Well and the tip of the Sampling Probe is 25 mm. 8. Adjust the Vertical Arm Home Sensor loosing its two fastening screws (Fig. 5). 9. Tighten the two Vertical Arm Home Sensor fastening screws (Fig. 5). 10. Select “Arm” folder and click the “Down Well” button. Wait until the Sampling Probe go down on the Washing Well. 11. Click the “Home” button. Wait until the vertical reset arm procedure has been completed. Vertical Arm Home Sensor Vertical Arm Home Sensor fastening screw Vertical Arm Home Sensor anchor bar Fig. 5 – Vertical Arm Home Sensor 12. Make sure that the distance between the top edge of the Washing Well and the tip of the Sampling Probe is 25 mm. If it not be so, repeat the above steps from 8 to 12. 13. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 14. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal), Reactions, Reagents and Samples Plates light up in green. 15. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate (Fig. 6). Ellipse Service Manual Rev. 04 01 October 2009 Page 7 Chapter 06 – SETTINGS AND ADJUSTMENTS 16. Select “Arm” folder. 17. Click the "Go Std" button, then click the correspondent “Down” button. The arm will move to position #1 of the Standards Plate, then it will go down until to reach the right height. 18. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Standards Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, adjust (if necessary) the verticality by using the adjustable anchored bearing (Fig. 7), click the ‘Save’ button, then repeat the above steps from 6. Reagents protection cover Fig. 6 – Reagents protection cover on the Reagents Plate Adjustable anchored bearing Horizontal Sampling Arm Motor Peristaltic Pump Motor Ellipse Service Manual Rev. 04 01 October 2009 Page 8 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 7 – Adjustable anchored bearing 19. Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move to position #1 of the Reagents Plate, then it will go down until to reach the right height. 20. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Reagents Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, then click the ‘Save’ button. 21. Click the "Go Sample" button, then click the correspondent “Down” button. The arm will move to position #1 of the Sample Rack #1, then it will go down until to reach the right height. 22. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Sample Rack #1. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, then click the ‘Save’ button. 23. Click the "Go Disp." button, then click the correspondent “Down” button. The arm will move to cuvette #81 of the Reactions Plate, then it will go down until to reach the right height. 24. Make sure that the Sampling Probe is vertically aligned and centered with respect to cuvette #81 of the Reactions Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, then click the ‘Save’ button. 25. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed. 26. Place the Sampling Arm protective cover screwing its fastening screw (Fig. 1). N.B.: in case of not enough current centered of the Sampling Probe with respect to the Washing Well and with respect to the cuvette #81 of the Reactions Plate, repeat the "Alignment and Adjustments” procedure privileging the centering of the Sampling Probe with respect to the cuvette #81 of the Reactions Plate. 6.1.2 REPLACEMENT OF THE SAMPLING PROBE 1. Removing of the Sampling Probe: a. launch the “Diagnostic” program; b. select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed; Ellipse Service Manual Rev. 04 01 October 2009 Page 9 Chapter 06 – SETTINGS AND ADJUSTMENTS c. remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut (Fig. 8). 2. Replacement of the Sampling Probe: a. launch the “Diagnostic” program; b. select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed; c. attach the (straight) Sampling Probe to the Probe holder of the Sampling Arm (Fig. 9 – Step 1); d. insert the Sampling Probe in the Probe holder and slowly rotate (counter clockwise) the Sampling Probe ring nut (Fig. 9 – Step 2); e. keep on rotating the Sampling Probe ring nut until you feel the stop (taking it around 25 mm from the ring nut) (Fig. 9 – Step 3). This is to allow a perfect contact between Sampling Probe and the Probe holder; f. make sure that the Sampling Probe is snug (Fig. 9 – Step 4); g. take the dynamometric key (Fig. 10) and place the head around the ring nut of the Sampling Probe (Fig. 11); h. screw (counter clockwise) the Sampling Probe with the dynamometric key (Fig. 12) until the correct torque is reached and the key spins freely around the Sampling Probe; Sampling Arm head Sampling Probe ring nut CW Sampling Probe Fig. 8 - Removing of the Sampling Probe Ellipse Service Manual Rev. 04 01 October 2009 Page 10 Chapter 06 – SETTINGS AND ADJUSTMENTS Sampling Arm head Step 1 Step 2 Probe holder Sampling Probe ring nut CCW Sampling Probe Step 3 Step 4 Sampling Probe ring nut CCW Sampling Probe Ellipse Service Manual Rev. 04 01 October 2009 Page 11 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 9 - Replacement of the Sampling Probe i. remove the dynamometric key; j. turn in the “Diagnostic” program; k. select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed; l. make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal), Reactions, Reagents and Samples Plates light up in green. N.B.: after having replaced the Sampling Probe, it is necessary to verify the verticality and the centering of it with respect to the Washing Well. In case of misalignment: perform the procedure "Alignment and Adjustments”, described into the previous Section 6.1.1, from step 5. Dynamometric key head Fig. 10 – Dynamometric key Dynamometric key CCW Sampling Probe Ellipse Service Manual Rev. 04 01 October 2009 Page 12 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 11 - Dynamometric key head around the ring nut Fig. 12 – Dynamometric key rotation 6.1.3 REPLACEMENT OF THE SAMPLING ARM 1. Remove the Sampling Probe from its housing unscrewing (clockwise) its ring nut (Fig. 8). 2. Make sure that the instrument (Ellipse) is turned off. 3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13). 4. Unplug the J1, J2, J5, J6, J13 connectors from the Arm Interface Board (Fig. 14), remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1) and unplug the ground wire from the Sampling Arm Assembly. 5. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1). 6. Unscrew the two anchored screws (Fig. 15) and pull out the Diluter Module from its hole. 7. Remove the Preheater teflon tube from the manifold of the Diluter Solenoid Valve by unscrewing the fitting (Fig. 16), paying special attention to liquid leakage. 8. Remove the Preheater clamp, located into the Peristaltic Pump housing, unscrewing its fastening screw (Fig. 17). 9. Unscrew the two Preheater fastening screws (Fig. 2), loosen the “Screw “B” (Fig. 4) and remove the Level Sensor and Preheater Assembly from its housing. 10. Unscrew the two Sampling Arm anchored screws (Fig. 18), pull out the Sampling Arm from its housing and substitute it with the new one. To replace the new Sampling Arm, perform the following steps: 11. Repeat the above steps in inverse order: from 10 to 4. 12. Perform the procedure "Alignment and Adjustments” described into the previous Section 6.1.1. 13. Select “Diluter” folder and perform - with the selection of the “Probe Wash Test” button - at least five successive Probe Wash Test to fill the hydraulic circuit with the liquid. 14. Verify the absence of liquid leakage from the hydraulic circuit. 15. Place the instrument front panel screwing its four anchored screws (Fig. 13). Ellipse Service Manual Rev. 04 01 October 2009 Page 13 Chapter 06 – SETTINGS AND ADJUSTMENTS Instrument front panel Front panel anchored screw Fig. 13 - Front panel anchored screws Horizontal Mov. Arm Motor – J1 Vertical Mov. Arm Motor – J2 Vertical Mov. Arm Home Sensor – J6 Horizontal Mov. Arm Home Sensor – J5 Preheater and Level Sensor Assy – J13 Ellipse Service Manual Rev. 04 01 October 2009 Page 14 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 14 - Arm Interface Board (Dx) [P/N: 30-01282-01] Diluter anchored screw Diluter Diluter Solenoid Valve Manifold Fig. 15 - Diluter Module (top view) Fitting Preheater teflon tube Fig. 16 – Diluter fitting Peristaltic Pump housing Preheater clamp Preheater clamp fastening screw Ellipse Service Manual Rev. 04 01 October 2009 Page 15 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 17 – Preheater clamp Ground wire Sampling Arm Sampling Arm anchored screw Fig. 18 – Sampling Arm anchored screws 6.1.4 REPLACEMENT OF THE ARM HOME SENSORS 6.1.4.1 Replacement of the Horizontal Arm Home Sensor N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure. 1. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1). 2. Loosen the set screw of the Horizontal Arm Home Sensor flat bar and turn the flat bar in order to unscrew the two Horizontal Arm Home Sensor fastening screws (Fig. 2). 3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13). 4. Unplug the J5 connector from the Arm Interface Board (Fig. 14), take out the entire Arm Home Sensor Assy and substitute it with the new one (Fig. 19). Ellipse Service Manual Rev. 04 01 October 2009 Page 16 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 19 - Arm Home Sensor Assy [P/N: 9-10-0023-60] To replace the new (Horizontal) Arm Home Sensor Assy, perform the following steps: 5. Repeat steps 4 and 3. 6. Turn on the instrument Ellipse. 7. Launch the “Diagnostic” program. 8. Center (manually) the Sampling Probe with respect to the Washing Well. 9. Position the slit of the Horizontal Arm Home Sensor flat bar (Fig. 2) to the center of the Horizontal Arm Home Sensor. 10. Tighten the set screw of the Horizontal Arm Home Sensor flat bar (Fig. 2). 11. Select “Arm” folder and click the “Reset” button. Wait until the reset arm procedure has been completed. 12. Make sure that the Horizontal Arm Home Sensor light up in green. 13. Make sure that the Sampling Probe is centered with respect to the Washing Well. If it not be so, adjust in the following way: a. loosen the set screw of the Horizontal Arm Home Sensor flat bar (Fig. 2); Ellipse Service Manual Rev. 04 01 October 2009 Page 17 Chapter 06 – SETTINGS AND ADJUSTMENTS b. turn slightly the Horizontal Arm Home Sensor flat bar to the right toward in order to obtain the centering; c. tighten the set screw of the Horizontal Arm Home Sensor flat bar (Fig. 2); d. select “Arm” folder and click the “Reset” button. Wait until the reset arm procedure has been completed; e. make sure that the Horizontal Arm Home Sensor light up in green; f. verify that the Sampling Probe is centered with respect to the Washing Well. If it not be so, repeat steps from 13.a to 13.f . 14. perform the procedure "Alignment and Adjustments”, described into the previous Section 6.1.1, from step 13 to step 26. 6.1.4.2 Replacement of the Vertical Arm Home Sensor N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure. 1. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1). 2. Unscrew the two Vertical Arm Home Sensor fastening screws (Fig. 5). 3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13). 4. Unplug the J6 connector from the Arm Interface Board (Fig. 14), take out the entire Arm Home Sensor Assy and substitute it with the new one (Fig. 19). To replace the new (Vertical) Arm Home Sensor Assy, perform the following steps: 5. Repeat steps 4, 3 and 2. 6. Turn on the instrument Ellipse. 7. Launch the “Diagnostic” program. 8. Select “Arm” folder and click the “Reset” button. Wait until the reset arm procedure has been completed. 9. Make sure that the Vertical Arm Home Sensor light up in green. 10. Make sure that the distance between the top edge of the Washing Well and the tip of the Sampling Probe is 25 mm. If it not be so, adjust in the following way: a. loosen the two Vertical Arm Home Sensor fastening screws (Fig. 5); b. move slightly the Vertical Arm Home Sensor to the right vertically toward in order to obtain the centering; c. tighten the two Vertical Arm Home Sensor fastening screws (Fig. 5); Ellipse Service Manual Rev. 04 01 October 2009 Page 18 Chapter 06 – SETTINGS AND ADJUSTMENTS d. select “Plate” folder and click “All Motors OFF” button; e. move downward (manually) the Sampling Arm; f. click “All Motors ON” button in the “Plate” folder. Wait until the reset instrument procedure has been completed; g. verify that the distance between the top edge of the Washing Well and the tip of the Sampling Probe is 25 mm. If it not be so, repeat steps from 10.a to 10.g . 11. Place the Sampling Arm protective cover screwing its fastening screw (Fig. 1). 6.1.5 REPLACEMENT OF THE SAMPLING ARM MOTORS 6.1.5.1 Replacement of the Horizontal Sampling Arm Motor N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure. 1. Remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut (Fig. 8). 2. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1). 3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13). 4. Unplug the J1, J2, J5, J6, J13 connectors from the Arm Interface Board (Fig. 14) and the ground wire from the Sampling Arm Assembly (Fig. 18). 5. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1). 6. Unscrew the two anchored screws (Fig. 15) and pull out the Diluter Module from its hole. 7. Remove the Preheater teflon tube from the manifold of the Diluter Solenoid Valve by unscrewing the fitting (Fig. 16), paying special attention to liquid leakage. 8. Remove the Preheater clamp, located into the Peristaltic Pump housing, unscrewing its fastening screw (Fig. 17). 9. Unscrew the two Preheater fastening screws (Fig. 2), loosen the “Screw “B” (Fig. 4) and remove the Level Sensor and Preheater Assembly from its housing. 10. Unscrew the two Sampling Arm anchored screws (Fig. 18) and pull out the Sampling Arm from its housing. 11. Remove the four Horizontal Sampling Arm Motor fastening screws (Fig. 20). 12. Remove the belt from the Horizontal Sampling Arm Motor pulley (Fig. 21). 13. Pull out the Horizontal Sampling Arm Motor from its housing. Ellipse Service Manual Rev. 04 01 October 2009 Page 19 Chapter 06 – SETTINGS AND ADJUSTMENTS Horizontal Sampling Arm Motor fastening screw Horizontal Sampling Arm Motor Fig. 20 – Horizontal Sampling Arm Motor 14. Loosen the two set screws of the Horizontal Sampling Arm Motor pulley (Fig. 21), then extract the pulley to the motor axis. 15. Unscrew the four fastening screws of the Horizontal Sampling Arm Motor anchor plate (Fig. 21), then take out the anchor plate to the motor. 16. Insert the pulley to the motor axis of the new Arm Motor Assy (Fig. 23). 17. Position the pulley to 2.00 mm with respect to upper motor plane (Fig. 22), then tighten the two set screws of the pulley (Fig. 21). 18. Position correctly the anchor plate on the upper plane of the new motor (Fig. 21), then screw the four anchor plate fastening screws (Fig. 21). Ellipse Service Manual Rev. 04 01 October 2009 Page 20 Chapter 06 – SETTINGS AND ADJUSTMENTS Upper motor plane Pulley Pulley Set screw Anchor plate Anchor plate fastening screw Fig. 21 - Horizontal Sampling Arm Motor anchor plate 19. Fig. 22 - Horizontal Sampling Arm Motor scheme Replace the Horizontal Sampling Arm Motor into its Sampling Arm housing, then insert (without tighten) the four Horizontal Sampling Arm Motor fastening screws (Fig. 20). 20. Wrap the belt around the Horizontal Sampling Arm Motor pulley. 21. Push the motor outward from the arm axis and holding it in this position, so as to maintain the belt in tension, then tighten the four Horizontal Sampling Arm Motor fastening screws (Fig. 27). 22. To replace the Sampling Arm, repeat the above steps in inverse order: from 10 to 1. N.B.: after having replaced the Horizontal Sampling Arm Motor, it is necessary to align mechanically the Sampling Probe with respect to the Washing Well and center all the other positions by following the procedure "Alignment and Adjustments” described into the previous Section 6.1.1 Ellipse Service Manual Rev. 04 01 October 2009 Page 21 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 23 – Arm Motor Assy [P/N: 10-00399-01] 6.1.5.2 Replacement of the Vertical Sampling Arm Motor N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure. 1. Remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut (Fig. 8). 2. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1). 3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13). 4. Unplug the J2 connector from the Arm Interface Board (Fig. 14). 5. Loosen the four fastening screws of the belt tension plate that permit to adjust the belt tension for the vertical Sampling Probe movement (Fig. 24). 6. Remove the belt from the Vertical Sampling Arm Motor pulley (Fig. 25). 7. Unscrew the four Vertical Sampling Arm Motor fastening screws (Fig. 25). Belt Belt tension plate Fastening screw Fastening screw Vertical Sampling Arm Motor pulley Ellipse Service Manual Rev. 04 01 October 2009 Page 22 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 24 – Belt tension plate Fig. 25 – Vertical Sampling Arm Motor fastening screws 8. Pull out the Vertical Sampling Arm Motor from its housing, then remove the anchor plate to the motor. 9. Loosen the two set screws of the Vertical Sampling Arm Motor pulley (Fig. 26), then extract the pulley to the motor axis. 10. Insert the pulley to the motor axis of the new Arm Motor Assy (Fig. 23). 11. Position the pulley to 1.50 mm with respect to upper motor plane (Fig. 26), then tighten the two set screws of the pulley (Fig. 26). Set screw Pulley Upper motor plane Fig. 26 – Vertical Sampling Arm Motor scheme 12. Position the anchor plate on the upper plane of the new motor, then replace the Vertical Sampling Arm Motor into its Sampling Arm housing (Fig. 25). 13. Tighten the four Vertical Sampling Arm Motor fastening screws (Fig. 25). 14. Wrap the belt around the Vertical Sampling Arm Motor pulley. 15. Push upwards the belt tension plate (Fig. 24) and holding it in this position, so as to maintain the belt in tension, then tighten its four fastening screws (Fig. 24). 16. To complete, repeat the above steps in inverse order: from 4 to 1. Ellipse Service Manual Rev. 04 01 October 2009 Page 23 Chapter 06 – SETTINGS AND ADJUSTMENTS N.B.: after having replaced the Vertical Sampling Arm Motor, it is necessary to require a Instrument General Reset and perform the procedure "Alignment and Adjustments”, described into the previous Section 6.1.1, from step 7. 6.1.6 REPLACEMENT OF THE SAMPLING ARM BELTS 6.1.6.1 Replacement of the Horizontal Sampling Arm Belt N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure. 1. Remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut (Fig. 8). 2. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1). 3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13). 4. Unplug the J1, J2, J5, J6, J13 connectors from the Arm Interface Board (Fig. 14) and the ground wire from the Sampling Arm Assembly (Fig. 18). 5. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1). 6. Unscrew the two anchored screws (Fig. 15) and pull out the Diluter Module from its hole. 7. Remove the Preheater teflon tube from the manifold of the Diluter Solenoid Valve by unscrewing the fitting (Fig. 16), paying special attention to liquid leakage. 8. Remove the Preheater clamp, located into the Peristaltic Pump housing, unscrewing its fastening screw (Fig. 17). 9. Unscrew the two Preheater fastening screws (Fig. 2), loosen the “Screw “B” (Fig. 4) and remove the Level Sensor and Preheater Assembly from its housing. 10. Unscrew the two Sampling Arm anchored screws (Fig. 18) and pull out the Sampling Arm from its housing. 11. Loosen the four Horizontal Sampling Arm Motor fastening screws (Fig. 27). 12. Remove the belt from the Horizontal Sampling Arm Motor pulley (Fig. 21) from the horizontal handling pulley, then substitute it with the new one. Ellipse Service Manual Rev. 04 01 October 2009 Page 24 Chapter 06 – SETTINGS AND ADJUSTMENTS Horizontal Sampling Arm Motor fastening screw Belt Horizontal handling pulley Fig. 27 – Horizontal Sampling Arm Belt 13. Wrap the new belt over the Horizontal Sampling Arm Motor pulley and over the horizontal handling pulley (Fig. 27). 14. Push the motor outward from the arm axis and holding it in this position, so as to maintain the belt in tension, then tighten the four Horizontal Sampling Arm Motor fastening screws (Fig. 27). 15. To replace the Sampling Arm, repeat the above steps in inverse order: from 10 to 1. N.B.: after having replaced the Horizontal Sampling Arm Belt, it is necessary to align mechanically the Sampling Probe with respect to the Washing Well and center all the other positions by following the procedure "Alignment and Adjustments” described into the previous Section 6.1.1 . Ellipse Service Manual Rev. 04 01 October 2009 Page 25 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.1.6.2 Replacement of the Vertical Sampling Arm Belt N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure. 1. Remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut (Fig. 8). 2. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1). 3. Loosen the four fastening screws of the belt tension plate that permit to adjust the belt tension for the vertical Sampling Probe movement (Fig. 28). 4. Unscrew the four fastening screws (Fig. 28) of the Vertical Arm Home Sensor anchor bar (Fig. 5) and remove it to the Sampling Arm. Fastening screw of the belt tension plate Belt tension plate Fastening screw of the Vertical Arm Home Sensor anchor bar Fig. 28 – Fastening screw of the Vertical Arm Home Sensor anchor bar 5. Unscrew the two clamping belt fastening screw, remove the two small plates (Fig. 29), take off the belt to the Sampling Arm, then substitute it with the new one. Ellipse Service Manual Rev. 04 01 October 2009 Page26 Chapter 06 – SETTINGS AND ADJUSTMENTS Belt Two small plates Clamping belt fastening screw Fig. 29 –Vertical Sampling Arm Belt 6. To replace the new Vertical Sampling Arm Belt, repeat the above steps in inverse order: from 5 to 1. N.B.: after having replaced the Vertical Sampling Arm Belt, it is necessary to require a Instrument General Reset and perform the procedure "Alignment and Adjustments”, described into the previous Section 6.1.1, from step 7. 6.2 6.2.1 DILUTER MODULE REPLACEMENT OF THE DILUTER MODULE 1. Make sure the instrument Ellipse is turned off. 2. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1). 3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13). 4. Take off the Rinse Solution cannula (I) from its bottle. 5. Unplug the J4, J8, J16, J17 connectors from the Arm Interface Board (Fig. 30) and the ground wire from the Diluter Module. Ellipse Service Manual Rev. 04 01 October 2009 Page 27 Chapter 06 – SETTINGS AND ADJUSTMENTS Preheater and Level Sensor Assy – J13 Diluter Motor - J4 Diluter Home Sensor – J8 Diluter Electrovalve – J16 Diluter Micropump – J17 Fig. 30 - Arm Interface Board (Dx) [P/N: 30-01282-01] 6. Loosen the two anchored screws (Fig. 31) and pull out the Diluter Module (Fig. 32) from its hole, making attention not to crash the tubes of the hydraulic circuit. 7. Unscrew the two fitting from the Diluter head (Fig. 31), making attention to the spillage of the Rinse Solution from the two tubes. 8. Take off the Diluter Module cables from its hole. 9. Unscrew the four Diluter fastening screws (Fig. 31) and pull out the Diluter from its mechanical support. Ellipse Service Manual Rev. 04 01 October 2009 Page28 Chapter 06 – SETTINGS AND ADJUSTMENTS Diluter Module anchored screw Diluter fastening screw Diluter fitting Fig. 31 – Diluter Module (top view) 10. Remove the four Diluter fastening screw washer. 11. Replace the Diluter with one new Diluter. Ellipse Service Manual Rev. 04 01 October 2009 Page 29 Chapter 06 – SETTINGS AND ADJUSTMENTS Diluter Home Sensor fastening screw Diluter Home Sensor [P/N: 9-10-0024-52] Diluter Motor [P/N: 10-00399-40] Fig. 32 – Diluter Module [P/N: 10-00496-01] 12. To replace, repeat the above steps in inverse order: from 10 to 4. N.B.: make sure that the two fitting are very screws on the Diluter head in order to guarantee the hydraulic seal of the circuit. Moreover, make sure that the Diluter Module cables in the instrument do not hinder the movement of the Pre-Heater in its zone of job. 13. Turn on the instrument Ellipse. 14. Make sure that sufficient liquid in the Rinse bottle is present. 15. Launch the “Diagnostic” program. 16. Select “Diluter” folder and click the “Home” button. Wait until the reset Diluter procedure has been completed. 17. Set up the aspiration volume to the maximum value (980 µl) and click the “Asp” button. Make sure that the diluter executes the aspiration procedure correctly. 18. Click the “Disp” button. Make sure that the diluter executes the dispenses procedure correctly. 19. Click the “Asp” button with the aspiration volume to the maximum value (980 µl). 20. Click the “Probe Wash Test” button. Ellipse Service Manual Rev. 04 01 October 2009 Page 30 Chapter 06 – SETTINGS AND ADJUSTMENTS N.B.: before proceeding with the Probe Wash Test, make sure that one written on the warning message is respected. 21. Click the “OK” button. Wait until the Probe Wash Test has been completed. 22. Repeat the above steps 20 and 21 at least others five times in order to fill up the hydraulic circuit completely. 23. Make sure that inside the Diluter are not present of the air bubbles. If it not be so, repeat the Probe Wash Test (steps 20 and 21) until the elimination of the air bubbles. 24. Verify the absence of liquid leakage from the hydraulic circuit. 25. Click the “Home” button. Wait until the reset Diluter procedure has been completed. 26. Place the instrument front panel screwing its four anchored screws (Fig. 13). 27. Place the Diluter protective cover screwing its two fastening screws (Fig. 1). N.B.: the following steps allows to perform the procedure "Backlash Diluter”. This procedure in order to recover the sampling imprecision generated by the Diluter Motor when it inverts its rotation. 28. Remove the Reagents protection cover on the Reagents Plate and take out the Samples Racks. 29. Select “Arm” folder and click the “Reset” button. Wait until the reset Sampling Arm procedure has been completed. 30. Click the "Go Sample" button. The arm will move to position #1 of the Sample Rack #1. 31. Put a Sample cup containing distilled water (about 1 ml) under the Sampling Probe. 32. Select “Diluter” folder, set up the aspiration volume to the 300 µl and click the “Asp” button. Wait until the Diluter has been aspirated 300 µl of distilled water from the Sample cup. 33. Remove the Sample cup containing distilled water. 34. Set up the dispensation volume to the 1 µl and repeatedly click the “Disp” button until water appears on the tip of the Sampling Probe. 35. Subtract 1 from the number of times in which it has been selected the “Disp” button. The obtained number is the Backlash Diluter value. 36. Select "Configuration" folder, enter the password (“1234”) in the appropriate window and click the “OK” button. Ellipse Service Manual Rev. 04 01 October 2009 Page 31 Chapter 06 – SETTINGS AND ADJUSTMENTS 37. Select the Backlash Diluter value in the “Backlash Dil.” Pull Down Menu and click the “Save” button. Wait until the Backlash Diluter value has been saved. 38. Select “Arm” folder and click the “Reset” button. Wait until the reset Sampling Arm procedure has been completed. 39. Put on the Samples Racks and place the Reagents protection cover on the Reagents Plate. 6.2.2 REPLACEMENT OF THE DILUTER MICROPUMP 1. Make sure the instrument Ellipse is turned off. 2. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1). 3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13). 4. Take off the Rinse Solution cannula (I) from its bottle. 5. Unplug the J4, J8, J16, J17 connectors from the Arm Interface Board (Fig. 30) and the ground wire from the Diluter Module. 6. Loosen the two anchored screws (Fig. 31) and pull out the Diluter Module (Fig. 32) from its hole, making attention not to crash the tubes of the hydraulic circuit. 7. Take off the Diluter Module cables from its hole. 8. Remove the two tubes from Micropump (Fig. 33). Ellipse Service Manual Rev. 04 01 October 2009 Page 32 Chapter 06 – SETTINGS AND ADJUSTMENTS Diluter Electrovalve [P/N: 05-00700-48] Manifold Fitting Diluter Micropump [P/N: 05-01710-40] Fig. 33 – Diluter Module: Electrovalve and Micropump Diluter Electrovalve fastening screws Electrovalve and Micropump mechanical support Diluter Micropump fastening screws Fig. 34 – Electrovalve and Micropump mechanical support 9. Unscrew the two Micropump fastening screws (Fig. 34) and pull out the Micropump from its mechanical support. Ellipse Service Manual Rev. 04 01 October 2009 Page 33 Chapter 06 – SETTINGS AND ADJUSTMENTS 10. Replace the Micropump with one new Micropump. 11. To replace, repeat the above steps in inverse order: from 9 to 4. 12. Turn on the instrument Ellipse. 13. Make sure that sufficient liquid in the Rinse bottle is present. 14. Launch the “Diagnostic” program. 15. Select “Diluter” folder and click the “Home” button. Wait until the reset Diluter procedure has been completed. 16. Set up the aspiration volume to the maximum value (980 µl) and click the “Asp” button. Make sure that the Diluter executes the aspiration procedure correctly. 17. Click the “Disp” button. Make sure that the Diluter executes the dispenses procedure correctly. 18. Click the “Asp” button with the aspiration volume to the maximum value (980 µl). 19. Click the “Probe Wash Test” button. N.B.: before proceeding with the Probe Wash Test, make sure that one written on the warning message is respected. 20. Click the “OK” button. Wait until the Probe Wash Test has been completed. 21. Repeat the above steps 20 and 21 at least others five times in order to fill up the hydraulic circuit completely. 22. Make sure that inside the Diluter are not present of the air bubbles. If it not be so, repeat the Probe Wash Test (steps 20 and 21) until the elimination of the air bubbles. 23. Verify the absence of liquid leakage from the hydraulic circuit. 24. Click the “Home” button. Wait until the reset Diluter procedure has been completed. 25. Place the instrument front panel screwing its four anchored screws (Fig. 13). 26. Place the Diluter protective cover screwing its two fastening screws (Fig. 1). 27. Select "Configuration" folder, enter the password (“1234”) in the appropriate window and click the “OK” button. 28. Select the value 0 in the “Level” Pull Down Menu. 29. Click the “Run Test” button. Wait until the Washing Well level Test has been completed. Ellipse Service Manual Rev. 04 01 October 2009 Page 34 Chapter 06 – SETTINGS AND ADJUSTMENTS 30. Click the “OK” button in the appropriate window in order to save the new value. Wait until the new value has been saved. 6.2.3 REPLACEMENT OF THE DILUTER ELECTROVALVE 1. Make sure the instrument Ellipse is turned off. 2. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1). 3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13). 4. Take off the Rinse Solution cannula (I) from its bottle. 5. Unplug the J4, J8, J16, J17 connectors from the Arm Interface Board (Fig. 30) and the ground wire from the Diluter Module. 6. Loosen the two anchored screws (Fig. 31) and pull out the Diluter Module (Fig. 32) from its hole, making attention not to crash the tubes of the hydraulic circuit. 7. Take off the Diluter Module cables from its hole. 8. Unscrew the two Electrovalve fastening screws (Fig. 34) and pull out the Electrovalve from its mechanical support. 9. Remove the two tubes from Electrovalve (Fig. 33). 10. Unscrew the two fittings from Electrovalve. 11. Screw the two fittings on the new Electrovalve. N.B.: make sure that the two fitting are very screws on the new Electrovalve in order to guarantee the hydraulic seal of the circuit. 12. To replace, repeat the above steps in inverse order: from 9 to 4. 13. Turn on the instrument Ellipse. 14. Make sure that sufficient liquid in the Rinse bottle is present. 15. Launch the “Diagnostic” program. 16. Select “Diluter” folder and click the “Home” button. Wait until the reset Diluter procedure has been completed. 17. Set up the aspiration volume to the maximum value (980 µl) and click the “Asp” button. Make sure that the Diluter executes the aspiration procedure correctly. 18. Click the “Disp” button. Make sure that the Diluter executes the dispenses procedure correctly. 19. Click the “Asp” button with the aspiration volume to the maximum value (980 µl). Ellipse Service Manual Rev. 04 01 October 2009 Page 35 Chapter 06 – SETTINGS AND ADJUSTMENTS 20. Click the “Probe Wash Test” button. N.B.: before proceeding with the Probe Wash Test, make sure that one written on the warning message is respected. 21. Click the “OK” button. Wait until the Probe Wash Test has been completed. 22. Repeat the above steps 20 and 21 at least others five times in order to fill up the hydraulic circuit completely. 23. Make sure that inside the Diluter are not present of the air bubbles. If it not be so, repeat the Probe Wash Test (steps 20 and 21) until the elimination of the air bubbles. 24. Verify the absence of liquid leakage from the hydraulic circuit. 25. Click the “Home” button. Wait until the reset Diluter procedure has been completed. 26. Place the instrument front panel screwing its four anchored screws (Fig. 13). 27. Place the Diluter protective cover screwing its two fastening screws (Fig. 1). 6.2.4 REPLACEMENT OF THE DILUTER HOME SENSOR 1. Make sure the instrument Ellipse is turned off. 2. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1). 3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13). 4. Unplug the J4, J8, J16, J17 connectors from the Arm Interface Board (Fig. 30) and the ground wire from the Diluter Module. 5. Loosen the two anchored screws (Fig. 31) and pull out the Diluter Module (Fig. 32) from its hole, making attention not to crash the tubes of the hydraulic circuit. 6. Take off the Diluter Module cables from its hole. 7. Unscrew the two Diluter Home Sensor fastening screws (Fig. 32) and pull out the Home Sensor from its hole within the Diluter. 8. Replace the Home Sensor with one new Home Sensor (Fig. 35). Ellipse Service Manual Rev. 04 01 October 2009 Page 36 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 35 – Diluter Home Sensor Assy [P/N: 9-10-0024-52] 9. To replace, repeat the above steps in inverse order: from 7 to 4. 10. Turn on the instrument Ellipse. 11. Launch the “Diagnostic” program. 12. Select “Diluter” folder and click the “Home” button. Wait until the reset Diluter procedure has been completed. 13. Set up the aspiration volume to the maximum value (980 µl) and click the “Asp” button. Make sure that the Diluter Home Sensor on the “Diagnostic” program becomes of red color (outside of the home position). 14. Click the “Home” button. Wait until the reset Diluter procedure has been completed. 15. Make sure that the Diluter Home Sensor on the “Diagnostic” program becomes of green color (home position). 16. Place the instrument front panel screwing its four anchored screws (Fig. 13). 17. Place the Diluter protective cover screwing its two fastening screws (Fig. 1). Ellipse Service Manual Rev. 04 01 October 2009 Page 37 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.3 PRE-HEATER AND LEVEL SENSOR 6.3.1 SUBSTITUTION OF THE PRE-HEATER AND LEVEL SENSOR ASSEMBLY 1. Turn on the Ellipse system (instrument and computer). 2. Launch the Diagnostic program, select the “Diluter” function and make two or three “Probe Wash Test” cycles with the Rinse solution cannula out of his bottle, in order to completely empty the Rinse hydraulic line. 3. Exit from the diagnostic program, pressing the "Diagnostic" key and exit from the analyzer program pressing “Shutdown” key. Remove the sampling probe from its housing. 4. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 36). 5. Unscrew the two anchored screws and pull out the Diluter Module from his hole (Fig. 37). 6. Remove the Preheater teflon tube on the Diluter Solenoid Valve by unscrewing the fitting (Fig. 38). 7. Remove the front panel unscrewing the four anchored screws (Fig. 40) and unplug the J13 connector from the Arm Interface board (Fig. 41). 8. Remove the Preheater clamp located into the Peristaltic Pump housing, unscrewing its fastening screw (Fig. 44). 9. Unscrew the Preheater and Level sensor fastening screws, remove it from its housing and substitute it with the new one (Fig. 42). 10. Reinstall the new assembly, making the vertical alignment of the sampling probe, with reference to the washing well and then screw the Preheater head and Level sensor fastening screw (Fig. 39). 11. After reinstalling the new assembly, repeat the above steps in inverse order: from step 9 to 1 and then, carry out the check procedure for the mechanical functioning described on the next pages. Ellipse Service Manual Rev. 04 01 October 2009 Page 38 Chapter 06 – SETTINGS AND ADJUSTMENTS WARNING Do the following operations only after substitution of the Level Sensor and Pre-heater assembly: a) Remove the white wire from the teflon tube before connecting it to the diluter (the white wire must be removed only at this stage of the procedure and not before). Be sure of having properly connected the tube to the diluter. b) Remove the protecting cap before positioning the sampling probe. Ellipse Service Manual Rev. 04 01 October 2009 Page 39 Chapter 06 – SETTINGS AND ADJUSTMENTS Sampling Arm protective cover Washing Well Diluter protective cover Fig. 36 6.3.2 LEVEL SENSOR – CHECK AND ADJUSTMENT PROCEDURE 1. Turn on the Ellipse system (instrument and computer) and launch the diagnostic program. 2. Select the "Configuration" function. Digit the password (1234) in the appropriate window and press “OK”. 3. Press the "Wash. Well level test". By clicking on the “Run Test” button the system performs an automatic verification and adjustment cycle for the liquid level in the washing well. A number appears in the box “Level” that corresponds to the rotation time of the rinse pump in milliseconds, that is necessary to maintain the correct liquid level in the washing well. 4. The functioning range is between 350 and 700 milliseconds. If the washing well is not sufficiently filled, a message appears to warn the user to check the hydraulic line. 5. If the value taken during the test is different from the previous one, it is necessary to click on “Save” in order to save the new data. 6. Make sure that no liquid leaks from the probe assembly and/or the dilutor. Make sure that the liquid flows from the probe freely. 7. Press “Test ON” on the “Level sensor Test” folder, on the right of it will be showed a red drop that becomes temporarily green when the liquid is detected or when the probe is touched with your finger (Fig. 43). Ellipse Service Manual Rev. 04 01 October 2009 Page 40 Chapter 06 – SETTINGS AND ADJUSTMENTS Preheater fastening screw Anchored screw Preheater Head fastening screw Solenoid Valve fitting Fig. 37 Diluter – Top view Fig. 38 Dilutor Fig. 39 Sampling Arm Fig. 40 Instrument front view Instrument front panel anchored screw Ellipse Service Manual Rev. 04 01 October 2009 Page 41 Chapter 06 – SETTINGS AND ADJUSTMENTS Level Sensor and Pre-Heater assembly Cd 10-00659-00 Preheater & L.S. - J13 Fig. 41 Arm Interface board Cd. 30-01282-01 Fig. 42 Level Sensor & Preheater Assembly Level Sensor Test Fig. 43 8. In order to more specifically verify the correct level-sensing functioning, proceed as follows: a. Place a cup containing liquid in the first samples rack, position nr 1 (e.g . 500 µl) and launch the diagnostic program. b. Select the "Arm" function and press the "Go Sample" key in order to allow the probe to move into the position corresponding to the sample cup. c. In the window next to the "Go Level", enter the maximum descent of the probe (e.g.500). d. Select "Go Level" and make sure that the probe drops down into the cup until it senses the liquid. In the window to the right of the "Go Level" key the level reached by the probe will appear, expressed in tenths of millimeters e. Repeat step d) several times, each time returning to the Vertical "Home" position. Make sure that the level indicated remains constant ≤ 10 (1 mm). f. Repeat steps b) to e) for positions 8 of the first rack using the same sample cup. Make sure that the difference between all the positions is ≤ 1mm. Ellipse Service Manual Rev. 04 01 October 2009 Page 42 Chapter 06 – SETTINGS AND ADJUSTMENTS g. Repeat steps b) to f) for samples racks two, three and four. h. Press the "Back to Wash Well" key and then select "Go Sample" in order to allow the probe to return to a position in the rack, which does not contain any cup. i. Select "Go Level" and make sure that the level reached is the maximum indicated in the given window is 9999. j. Select the "Reset" key, remove the cups from the racks and exit the program by pressing the "Diagnostic" key. k. To exit the analyzer program press “shutdown” key. Preheater & L.S clamp Fig. 44 Ellipse Service Manual Rev. 04 01 October 2009 Page 43 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.4 OPTIC ASSEMBLY 6.4.1 SUBSTITUTION OPTIC ASSEMBLY N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. Stepper Motor assembly Cd. 10-00399-01 Opto coupler assy Cd. 9-10-0023-00 Pre-Ampl/ADC (Read Channel) Cd. 30-00107-00 Pre-Ampl/ADC (Ref. Channel) Cd. 30-00107-03 Fig. 45 Optic Assembly Cd. 10-01688-00 1. Remove the front panel unscrewing the four anchored screws (Fig. 46). 2. Remove the J1, J6, J11, J15 connectors from the Plate Interface board (Fig. 47) and the ground wire from the Optic Assembly. 3. Unscrew the four fastening screws of the Plate Interface board (Fig. 47). 4. Remove the connectors from JP1 of the Pre-Amplifier ADC boards. 5. Unscrew the four Optic Assembly fastening screws, remove it from its housing and substitute it with the new one (Fig. 45). 6. Remount the assembly, repeating the above steps, 5 through 1, in reverse order. 7. After remounting, reseat the lamp and carry out the electronic adjustment check procedures as described in the following pages. Ellipse Service Manual Rev. 04 01 October 2009 Page 44 Chapter 06 – SETTINGS AND ADJUSTMENTS Instrument front panel anchored screw Fig. 46 Instrument front view Filter Wheel Motor – J1 Filter Wheel home sensor – J6 Photometer lamp – J15 Photometer fan– J11 Board fastening screw Fig. 47 Plate Interface Board Cd. 30-01281-01 Ellipse Service Manual Rev. 04 01 October 2009 Page 45 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.4.2 SUBSTITUTION OF THE OPTIC LAMP N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. 1. Remove the front panel unscrewing the four anchored screws (Fig. 46). 2. Remove the lamp's electrical wires connected to the Lamp Regulator board by loosening the two screws on the M1 connector (Fig.48). 3. Unscrew the lamp's fastening screws and remove it from its housing. 4. Remount the new lamp, repeating the above steps, 3 through 1, in reverse order. 5. Carry out the electronic adjustment of the Optic as described in the following pages. Warning: Do not touch the glass portions of the lamp. Halogen Lamp Cd. 9-35-0016-00 Lamp's fastening screw M1 Connector Optic Lamp Board Cd. 30-01576-00 Fig. 48 Ellipse Trimmer P1 Service Manual Rev. 04 01 October 2009 Page 46 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.4.3 ELECTRONIC ADJUSTMENT OF THE OPTIC Warning: make sure that the Optic lamp and the reaction cuvettes are brand-new before applying the following procedure. 1. Turn on the "Ellipse" system (instrument and computer). Remove the front panel unscrewing the four anchored screws (Fig.46). 2. Launch the diagnostic program, select the "Plate " function and reset by clicking on the appropriate key. 3. Make sure that the "Home Sensor" Reaction Plate window lights up green. 4. Place 500 µL of distilled water in cuvette # 1 of the reaction plate. 5. Select Sample plate position number three in order to make accessible the main channel Gain and Offset holes. 6. Select the "Optic" function and turn the lamp "On" by clicking on the appropriate field. 7. Reset the filter wheel by clicking on the appropriate key. 8. Make sure that "Home Sensor" Filter Wheel window lights up green. 9. Make sure that the Regulator board for Connector M1 shows 6.0 volts ± 0.1. If not, turn the P1 trimmer (Fig.48). N.B.: The reaction cuvette must be perfectly clean. If not, carry out a "Cuvette Wash" cycle (Start Work) Offset main Channel Gain reference Channel Motor Support screw Offset reference Channel Gain main Channel Fig. 49 Ellipse Service Manual Optic Assembly – top view Rev. 04 01 October 2009 Page 47 Chapter 06 – SETTINGS AND ADJUSTMENTS Warning: Room lighting and daylight can influence the reading. To avoid this interference, remount the front panel before carrying out any check procedures. 10. Click on "Start Sample Channel Conversion" key and make sure that the adjacent window shows a count value of 100 ± 50. If not, adjust the Preamp/ADC (Read Channel) Board "Offset" trimmer in order to set the desired value (Fig. 49). 11. After regulating of the main channel offset, it is necessary to click on " Stop Sample Channel Conversion" key. 12. Click on "Start Reference Channel Conversion" and make sure that the adjacent window shows a count value of 100 ± 50. If not, adjust the Preamp/ADC (Ref. Channel) Board "Offset" trimmer in order to set the desired value (Fig. 49). 13. After regulating of the reference channel offset, it is necessary to click on " Stop Reference Channel Conversion" key. 14. Select the filter positions from 1 to 8 (from 340 nm to 620 nm) and click on” Start/Stop Sample Channel Conversion" key and determine which filter transmit the highest signal. 15. Verify in the adjacent window a count value between 55.000 and 60.000. If not, adjust the Preamp/ADC board "Gain" trimmer in order to set the desired value (Fig. 49). 16. After regulating the Gain, select all the other filters and clicking on "Start/Stop Sample Channel Conversion" verify count values between 29.000 and 60.000 Warning: Room lighting and daylight can influence the reading. To avoid this interference, remount the front panel before carrying out any check procedures. 17. Select the highest signal filter, clicking on "Start Reference Channel Conversion" and make sure that the adjacent window shows a count value between 38.000 and 42.000. If not, adjust the Preamp/ADC board "Gain" trimmer in order to set the desired value (Fig. 49). 18. After regulating the Gain, select all the other filters and clicking on "Start/Stop Reference Channel Conversion" verify count values between 19000 and 42000. 19. After concluding these operations, empty cuvette # 1, remount the front panel and exit the diagnostic program by clicking on the "Diagnostic" key. 20. Ellipse Exit the analyzer program press “shutdown” key. Service Manual Rev. 04 01 October 2009 Page 48 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.4.4 SUBSTITUTION OF THE OPTO SENSOR N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. 1. Remove the front panel unscrewing the four anchored screws (Fig. 46). 2. Remove the J1, J6, J11, J15 connectors from the Reaction Plate Interface board (Fig. 47) and the ground wire from the Optic Assembly. 3. Unscrew the four fastening screws of the Plate Interface board (Fig. 47). 4. Remove the connectors from JP1 of the Pre-Amplifier ADC boards. 5. Unscrew the four Optic Assembly fastening screws, remove it from its housing (Fig. 45). 6. Remove the lamp's power supply wires connected to the Lamp Regulator board by loosening the two screws on the M1 connector (Fig. 48). 7. Unscrew the lamp's fastening screws and remove it from its housing. 8. Unscrew the four fastening screws on the motor support in order to access the filter wheel (Fig. 49). 9. Position the filter wheel in such a manner as to make it possible to see and unscrew the internal home sensor support screw. Unscrew the external home sensor support screw. 10. Unscrew the home sensor fastening screws from its support and substitute with the new one (Fig. 50). N.B.: when remounting the Home sensor, make sure to center it with respect to the slots. 11. Remount the assembly, repeating the above steps, 10 through 1, in reverse order. 12. After remounting, reseat the lamp and carry out the electronic adjustment procedures as described in the previous pages. Ellipse Service Manual Rev. 04 01 October 2009 Page 49 Chapter 06 – SETTINGS AND ADJUSTMENTS Filter Wheel Motor Cd. 10-00399-01 Filter Wheel Home sensor Cd. 9-10-0023-00 Fig. 50 Fig. 51 6.4.5 SUBSTITUTION OF THE MOTOR N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. 1. Remove the front panel unscrewing the four anchored screws (Fig. 46). 2. Remove the J1, J6, J11, J15 connectors from the Plate Interface board (Fig. 47) and the ground wire from the Optic Assembly. 3. Unscrew the four fastening screws of the Plate Interface board (Fig .47). 4. Remove the connectors from JP1 of the Pre-Amplifier ADC boards. 5. Unscrew the four Optic Assembly fastening screws, remove it from its housing (Fig. 45). 6. Remove the lamp's power supply wires connected to the Lamp Regulator board by loosening the two screws on the M1 connector (Fig. 48). 7. Unscrew the lamp's fastening screws and remove it from its housing. 8. Unscrew the four fastening screws on the motor support in order to access the pulley (Fig. 49). 9. Unscrew the four fastening screws on the motor; loosen the pulley fastening screws. 10. Remove the motor and substitute it with the new one (Fig.48). 11. Remount the assembly, repeating the above steps, 10 through 1, in reverse order. Warning: when tightening the motor pulley fastening screws, make sure that the filter wheel is centered with respect to the opto sensor. 12. After remounting, reseat the lamp and carry out the electronic adjustment procedures as described in the previous pages. Ellipse Service Manual Rev. 04 01 October 2009 Page 50 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.4.6 SUBSTITUTION OF THE OPTIC FILTERS Interferential filters Kit Cd. 9-65-0029-00 N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. 1. Remove the front panel unscrewing the four anchored screws (Fig. 46). 2. Remove the J1, J6, J11, J15 connectors from the Plate Interface board (Fig. 47) and the ground wire from the Optic Assembly. 3. Unscrew the four fastening screws of the Plate Interface board (Fig. 47). 4. Remove the connectors from JP1 of the Pre-Amplifier ADC boards. 5. Unscrew the four Optic Assembly fastening screws and remove it from its housing (Fig. 45). 6. Remove the lamp’s power supply wires connected to the Lamp Regulator board by loosening the two screws on the M1 connector (Fig. 48). 7. Unscrew the lamp’s fastening screws and remove it from its housing. 8. Unscrew the four fastening screws on the motor support in order to access the filter wheel (Fig. 49). 9. Unscrew the fastening screw that holds the elastic band of the filters. 10. Remove the filter from its housing and substitute it with the new one. 11. Remount the assembly, repeating the above steps, 9 through 1, in reverse order. 12. After remounting, reseat the lamp and carry out the electronic adjustment procedures as described in the previous pages. Warning: Do not touch the glass parts of the filter. Warning: when substituting an interferential filter due to malfunctioning or functional anomalies, it is always advisable to substitute the entire set of filters mounted on the instrument. Ellipse Service Manual Rev. 04 01 October 2009 Page 51 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5 PLATES 6.5.1 REMOVAL OF THE PLATES ASSEMBLY 1. Make sure that the instrument (Ellipse) is turned off. 2. Open the analyzer main cover. 3. Remove the Sampling Probe from its housing unscrewing (clockwise) its ring nut. 4. Remove the reactions cuvettes front cover. 5. Remove with accuracy the left reactions cuvettes cover (placed under the Washing Station cannulas) unscrewing its three screws. 6. Remove the Reagents protection cover and remove the Reagents rack from over the Reagents Plate. 7. Remove all the Sample Racks from over the Samples Plate. 8. Remove with accuracy all the cuvettes Racks from over the Reactions Plate. N.B.: before exctracting the Plates Assembly from the analyzer, it is necessary remove the Sampling Arm, the Diluter Module, the Washing Station and then the Washing Well. 9. To remove the Sampling Arm, perform the relative procedure described into the previous Section 6.1.3, from step 3 to step 10 (without substitute it with the new one). 10. To remove the Diluter Module, perform the relative procedure described into the previous Section 6.2.1, from step 4 to step 6. 11. To remove the Washing Station, perform the relative procedure described into the successive Section 6.6.1. 12. Remove with accuracy the instrument chassis cover unscrewing the six anchored screws (Fig. 52). Ellipse Service Manual Rev. 04 01 October 2009 Page 52 Chapter 06 – SETTINGS AND ADJUSTMENTS Peristaltic Pump panel Chassis cover anchored screws Fig. 52 - Chassis cover anchored screws 13. Remove the two tubes from the overflow elbow fitting and fitting of the Washing Well (Fig. 52), making attention not to crash them. 14. Unscrew the Washing Well ring nut (Fig. 52) and pull out the Washing Well from its hole. Washing Well Washing Well ring nut Fitting Overflow elbow fitting Fig. 6.5.2 - Washing Well 15. Remove the Peristaltic Pump waste tube to the relative “Y” fitting, making attention to the spillage of the biohazard liquid from the tube and from the “Y” fitting. 16. Unplug the J3 connector from the Arm Interface Board and the ground wire from the Peristaltic Pump mechanical support. 17. Close with accuracy the instrument main cover. 18. Open the Electronic Rack panel unscrewing the two quarter-turn fasteners (Fig. 53). Ellipse Service Manual Rev. 04 01 October 2009 Page 53 Chapter 06 – SETTINGS AND ADJUSTMENTS Washing Station pumps panel Electronic Rack panel Electronic Rack panel quarter-turn fastener Fig. 53 - Instrument (posterior view) 19. Extract the Electronic Rack from its vain (Fig. 53). Stepper Motors Driver Board Analytical Control Board Fig. 54 - Electronic Rack 20. Unplug the J6, J8, J9, J10, J11 e J20 connectors and the ground wire from the Analytical Control Board (Fig. 54). Ellipse Service Manual Rev. 04 01 October 2009 Page 54 Chapter 06 – SETTINGS AND ADJUSTMENTS 21. Unplug the J1, J2 e J4 connectors and the ground wire from the Stepper Motors Driver Board (Fig. 54). 22. Insert with accuracy the Electronic Rack in its vain. 23. Extract the Power Supply Rack from its vain (Fig. 55). 24V Module 12V Module Power Supply Cable Power Supply wiring Ground connector Fastening Screw Module Fig. 55 - Power Supply Rack 24. On the Power Supply, unscrew the four fastening screw of the “24V Module” and “12V Module” and then, remove the relative four cables [JP24+, JP24-, JP12+, JP12-] of the Power Supply wiring (Fig. 55). 25. Unscrew the two ground connector fastening screws and remove all its ground wires (Fig. 55). 26. Insert with accuracy the Power Supply in its vain. 27. Close the Electronic Rack panel screwing the two quarter-turn fasteners (Fig. 55). 28. Lay down delicately the instrument on its posterior side (Fig. 56) and open the instrument main cover Ellipse Service Manual Rev. 04 01 October 2009 Page 55 Chapter 06 – SETTINGS AND ADJUSTMENTS Instrument adjustable feet Plates Assembly nut (and relative washers) Fig. 56 - Instrument (inferior surface view) 29. Loosen the two instrument adjustable feet (Fig. 56). 30. Holding berthed the Plates Assembly to the instrument, unscrew the four Plates Assembly nuts and remove the relative washers (Fig. 56). 31. Close the instrument main cover and then, lay down delicately the instrument on its inferior side. 32. Open the instrument main cover, take out with accuracy the Plates Assembly to the instrument and then, lay down the Plates Assembly outside the instrument. 33. Close with accuracy the instrument main cover. 34. Unplug the J28 connector and the ground wire from the Plate Interface Board (Fig. 57). 35. Unplug the J21, J23 connectors and the ground wire from the Arm Interface Board. 36. Unplug the two connectors to the Pre-Ampl./ADC (Read Channel) and the Pre-Ampl./ADC (Reference Channel) placed in the Optic Assembly. 37. Unscrew only the two Plate Interface Board fastening screws placed on its inferior side. 38. Unscrew only the two Arm Interface Board fastening screws placed on its inferior side. Ellipse Service Manual Rev. 04 01 October 2009 Page 56 Chapter 06 – SETTINGS AND ADJUSTMENTS 39. Put up the Plates Assembly and remove its Base Plate Assy from the four Plates Assembly spacing bars (Fig. 6.5.8). Reactions Plate Motor – J3 Samples Plate Motor – J2 Reagents Plate Motor – J4 Samples Plate Home Sensor – J7 Reactions Plate Home Sensor – J8 Power Supply – J28 Reagents Plate Home Sensor – J9 Fig. 57 - Plate Interface Board (Sx) [P/N: 30-01281-01] Plates Assembly Arm Interface Board Plate Interface Board Base Plate Assy Plates Assembly spacing bar Fig. 58 - Plates Assembly and Base Plate Assy 40. Lay down the Plates Assembly on its four spacing bars. Ellipse Service Manual Rev. 04 01 October 2009 Page 57 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5.2 REPLACEMENT OF THE REAGENTS PLATE MOTOR N.B.: before replacing of the Reagents Plate Motor, it is necessary remove the Plates Assembly from the instrument. 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Unplug the J4 connector from the Plate Interface Board (Fig. 57). 3. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59). Reagents Plate Home Sensor lever Reagents Plate Motor fastening screw Reactions Plate Motor fastening screw Reagents Plate belt Reagents Plate pulley Samples Plate Motor fastening screw Fig. 59 - Plates Assembly (inferior view) 4. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the Reagents Plate Home Sensor lever (Fig. 59) out of the Reagents Plate pulley. 5. Remove the belt from the Reagents Plate Motor pulley (Fig. 59). 6. Remove the belt from the Reagents Plate pulley. Ellipse Service Manual Rev. 04 01 October 2009 Page 58 Chapter 06 – SETTINGS AND ADJUSTMENTS 7. Pull upwards the support Reagents Rack (Fig. 60) and remove it to the Plates Assembly. Support Reagents Rack Samples Plate Reactions Plate Reactions Plate Home Sensor Reactions Plate Home Sensor lever Samples Plate Home Sensor Samples Plate Home Sensor lever Reagent Tts Plate Home Sensor Reagents Plate Home Sensor lever Fig. 60 - Plates, Plate Home Sensors and Plate Home Sensor levers Fig. 61 - Support Reagents Rack Ellipse Service Manual Rev. 04 01 October 2009 Page 59 Chapter 06 – SETTINGS AND ADJUSTMENTS 8. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate (Fig. 59). 9. Unscrew completely the four fastening screws of the Reagents Plate Motor (Fig. 59) and pull out it from its housing. 10. Loosen the pulley clamp screw and extract the pulley clamp to the motor pulley (Fig. 62). Anchor plate fastening screw Plate Motor Pulley clamp screw Anchor plate Connector Motor Pulley clamp Motor pulley Fig. 6.5.12 - Plate Motor with pulley 11. Take out the motor pulley from its axis (Fig. 64). 12. Unscrew the four fastening screws of the Reagents Plate Motor anchor plate (Fig. 62) and take out the anchor plate to the motor. 13. Replace the motor with the new Plate Motor Assy (Fig. 63). Motor axis Fig. 63 - Plate Motor Assy [P/N: 9-10-0016-00] Ellipse Service Manual Rev. 04 Fig. 64 - Plate Motor 01 October 2009 Page 60 Chapter 06 – SETTINGS AND ADJUSTMENTS 14. Insert the pulley clamp on the motor pulley (Fig. 62). 15. Insert the motor pulley (with the motor clamp) to the motor axis of the new Plate Motor Assy (Fig. 62). 16. Position the motor pulley to 1.50 mm with respect to upper motor plane (Fig. 65), then tighten the pulley clamp screw (Fig. 62). Upper motor plane Pulley clamp Stepping Motor Motor pulley Fig. 65 - Plate Motor scheme 17. Position the anchor plate on the upper plane of the new motor with the two anchor plate slots on the side of the motor connector (like represented in the (Fig. 62). 18. Screw the four anchor plate fastening screws (Fig. 62). 19. Replace the Reagents Plate Motor on its housing with the connector turned towards the Reagents Plate Home Sensor lever (Fig. 59). 20. Insert (without tighten) the four Reagents Plate Motor fastening screws (Fig. 59). 21. Insert the support Reagents Rack in the Plates Assembly (Fig. 61). 22. Wrap the belt around the Reagents Plate pulley. 23. Wrap the belt around the Reagents Plate Motor pulley. Ellipse Service Manual Rev. 04 01 October 2009 Page 61 Chapter 06 – SETTINGS AND ADJUSTMENTS 24. Push the motor outward from the Plates pulleys axis and holding it in this position, so as to maintain the belt in tension, then tighten the four fastening screws of the Reagents Plate Motor (Fig. 59). 25. Make sure that the belt thoots are correctly inserted on the Reagents Plate pulley. 26. Make sure that the belt thoots are correctly inserted on the Reagents Plate Motor pulley. 27. Insert the Reagents Plate Home Sensor around the Home sensor disk. 28. Tighten the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index. 29. Rotate with accuracy the Reagents Plate pulley and check that the belt tension is corrected and that the Home Sensor disk not to touch the Reagents Home Sensor. 30. Plug in the J4 connector from the Plate Interface Board (Fig. 57). 31. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Reagents Plate Motor and the Plates Assembly, it is necessary to align mechanically the Reagents Plate by following the procedure described into the Section 6.5.14.1 . 6.5.3 REPLACEMENT OF THE SAMPLES PLATE MOTOR N.B.: before replacing of the Samples Plate Motor, it is necessary remove the Plates Assembly from the instrument. 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59). 3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the Reagents Plate Home Sensor lever out of the Reagents Plate pulley. 4. Remove the belt from the Reagents Plate Motor pulley. 5. Remove the belt from the Reagents Plate pulley. 6. Loosen the two grains of the Reagents Plate pulley (Fig. 66) and extract the pulley to the plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66). Ellipse Service Manual Rev. 04 01 October 2009 Page 62 Chapter 06 – SETTINGS AND ADJUSTMENTS 7. Pull upwards the support Reagents Rack (Fig. 61) and remove it to the Plates Assembly. 8. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate (Fig. 59). 9. Unplug the J2 connector from the Plate Interface Board (Fig. 57). 10. Loosen the four fastening screws of the Samples Plate Motor (Fig. 59). 11. Remove the belt from the Samples Plate Motor pulley (Fig. 66). 12. Remove the belt from the Samples Plate pulley. 13. Unscrew completely the four fastening screws of the Samples Plate Motor (Fig. 59) and pull out it from its housing. 14. Loosen the pulley clamp screw and extract the pulley clamp to the motor pulley (Fig. 62). 15. Take out the motor pulley from its axis (Fig. 64). 16. Unscrew the four fastening screws of the Samples Plate Motor anchor plate (Fig. 62) and take out the anchor plate to the motor. 17. Replace the motor with the new Plate Motor Assy (Fig. 63). 18. Insert the pulley clamp on the motor pulley (Fig. 62). 19. Insert the motor pulley (with the motor clamp) to the motor axis of the new Plate Motor Assy (Fig. 62). 20. Position the motor pulley to 1.50 mm with respect to upper motor plane (Fig. 65), then tighten the pulley clamp screw (Fig. 62). 21. Position the anchor plate on the upper plane of the new motor with the two anchor plate slots on the side of the motor connector (like represented in the Fig. 62). 22. Screw the four anchor plate fastening screws (Fig. 62). 23. Replace the Samples Plate Motor on its housing with the connector turned towards the Optic Assembly (Fig. 59). 24. Insert (without tighten) the four Samples Plate Motor fastening screws (Fig. 59). 25. Wrap the belt around the Samples Plate pulley. 26. Wrap the belt around the Samples Plate Motor pulley. Ellipse Service Manual Rev. 04 01 October 2009 Page 63 Chapter 06 – SETTINGS AND ADJUSTMENTS 27. Push the motor outward from the Plates pulleys axis and holding it in this position, so as to maintain the belt in tension, then tighten the four fastening screws of the Samples Plate Motor (Fig. 59). 28. Make sure that the belt thoots are correctly inserted on the Samples Plate pulley. 29. Make sure that the belt thoots are correctly inserted on the Samples Plate Motor pulley. 30. Rotate with accuracy the Samples Plate pulley and check that the belt tension is corrected. 31. Plug in the J2 connector from the Plate Interface Board (Fig. 57). 32. Insert the Reagents Plate pulley (Fig. 66) on the plates axis, making attention not to fold (or to detach) its Home Sensor disk [Fig. 66] (and with the Home Sensor disk slit aligned with respect to the slits of the other two Home Sensor disks). 33. Tighten the two grains of the Reagents Plate pulley (Fig. 66) only after to have verified that the three Plate Motor pulleys are locked between of them. 34. Wrap the belt around the Reagents Plate pulley. 35. Wrap the belt around the Reagents Plate Motor pulley. 36. Push the motor outward from the Plates pulleys axis and holding it in this position, so as to maintain the belt in tension, then tighten the four fastening screws of the Reagents Plate Motor (Fig. 59). 37. Make sure that the belt thoots are correctly inserted on the Reagents Plate pulley. 38. Make sure that the belt thoots are correctly inserted on the Reagents Plate Motor pulley. 39. Insert the Reagents Plate Home Sensor around the Home sensor disk. 40. Tighten the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index. 41. Rotate with accuracy the Reagents Plate pulley and check that the belt tension is corrected and that the Home Sensor disk not to touch the Reagents Home Sensor. 42. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Samples Plate Motor, the Reagents Plate Motor and the Plates Assembly, it is necessary to align mechanically the Samples Plate and the Reagents plate by following the procedure described into the Section 6.5.15.1 Ellipse Service Manual Rev. 04 01 October 2009 Page 64 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5.4 REPLACEMENT OF THE REACTIONS PLATE MOTOR N.B.: before replacing of the Reactions Plate Motor, it is necessary remove the Plates Assembly from the instrument. 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Perform the procedure described into the previous Section 6.5.3 from step 2 to step 8. 3. Unscrew the Samples Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the Samples Plate Home Sensor lever out of the Samples Plate pulley. 4. Loosen the four fastening screws of the Samples Plate Motor (Fig. 59). 5. Remove the belt from the Samples Plate Motor pulley. 6. Remove the belt from the Samples Plate pulley. 7. Loosen the two grains of the Samples Plate pulley (Fig. 66) and extract the pulley to the plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66). 8. Loosen the four fastening screws of the Reactions Plate Motor (Fig. 59). 9. Remove the belt from the Reactions Plate Motor pulley. 10. Remove the belt from the Reactions Plate pulley. 11. Unplug the J3 connector from the Plate Interface Board (Fig. 57). 12. Unscrew completely the four fastening screws of the Reactions Plate Motor (Fig. 59) and pull out it from its housing. 13. Loosen the pulley clamp screw and extract the pulley clamp to the motor pulley (Fig. 62). 14. Ellipse Take out the motor pulley from its axis (Fig. 64). Service Manual Rev. 04 01 October 2009 Page 65 Chapter 06 – SETTINGS AND ADJUSTMENTS 15. Unscrew the four fastening screws of the Reactions Plate Motor anchor plate (Fig. 62) and take out the anchor plate to the motor. 16. Replace the motor with the new Plate Motor Assy (Fig. 63). 17. Insert the pulley clamp on the motor pulley (Fig. 62). 18. Insert the motor pulley (with the motor clamp) to the motor axis of the new Plate Motor Assy (Fig. 62). 19. Position the motor pulley to 1.50 mm with respect to upper motor plane (Fig. 65), then tighten the pulley clamp screw (Fig. 62). 20. Position the anchor plate on the upper plane of the new motor with the two anchor plate slots on the side of the motor connector (like represented in the Fig. 62). 21. Screw the four anchor plate fastening screws (Fig. 62). 22. Replace the Reactions Plate Motor on its housing with the connector turned towards the Reactions Home Sensor lever (Fig. 59). 23. Insert (without tighten) the four Reactions Plate Motor fastening screws (Fig. 59). 24. Wrap the belt around the Reactions Plate pulley. 25. Wrap the belt around the Reactions Plate Motor pulley. 26. Push the motor outward from the Plates pulleys axis and holding it in this position, so as to maintain the belt in tension, then tighten the four fastening screws of the Reactions Plate Motor (Fig. 59). 27. Make sure that the belt thoots are correctly inserted on the Reactions Plate pulley. 28. Make sure that the belt thoots are correctly inserted on the Reactions Plate Motor pulley. 29. Rotate with accuracy the Reactions Plate pulley and check that the belt tension is corrected. Ellipse Service Manual Rev. 04 01 October 2009 Page 66 Chapter 06 – SETTINGS AND ADJUSTMENTS 30. Plug in the J3 connector from the Plate Interface Board (Fig. 57). 31. Replace the Samples Plate pulley (Fig. 66) and tighten the two its grains only after to have verified that the two Plate Motor pulleys are locked between of them (and with the Home Sensor disk slit aligned with respect to the slit of the other Home Sensor disk). 32. Perform the procedure described into the previous Section 6.5.3 from step 25 to step 29. 33. Tighten the Samples Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index. 34. Rotate with accuracy the Samples Plate pulley and check that the belt tension is corrected and that the Home Sensor disk not to touch the Samples Home Sensor. 35. Perform the procedure described into the previous Section 6.5.3 from step 32 to step 41. 36. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Reactions Plate Motor, the Samples Plate Motor, the Reagents Plate Motor and the Plates Assembly, it is necessary to align mechanically all the plates by following the procedure described into the Section 6.5.16.1. 6.5.5 REPLACEMENT OF THE REAGENTS PLATE BELT N.B.: before replacing of the Reagents Plate Belt, it is necessary remove the Plates Assembly from the instrument. 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59). 3. Remove the belt from the Reagents Plate Motor pulley (Fig. 62). 4. Remove the belt from the Reagents Plate pulley (Fig. 66). 5. Pull upwards the support Reagents Rack (Fig. 61) and remove it to the Plates Assembly. Ellipse Service Manual Rev. 04 01 October 2009 Page 67 Chapter 06 – SETTINGS AND ADJUSTMENTS 6. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate (Fig. 59). 7. Substitute the belt with the new belt. 8. Insert the support Reagents Rack in the Plates Assembly (Fig. 60). 9. Wrap the new belt around the Reagents Plate pulley. 10. Wrap the new belt around the Reagents Plate Motor pulley. 11. Push the motor outward from the Plates pulleys axis and holding it in this position, so as to maintain the new belt in tension, then tighten the four fastening screws of the Reagents Plate Motor (Fig. 59). 12. Make sure that the belt thoots are correctly inserted on the Reagents Plate pulley. 13. Make sure that the belt thoots are correctly inserted on the Reagents Plate Motor pulley. 14. Rotate with accuracy the Reagents Plate pulley and check that the belt tension is corrected. 15. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Reagents Plate belt and the Plates Assembly, it is necessary to align mechanically the Reagents Plate by following the procedure described into the Section 6.5.14.1. 6.5.6 REPLACEMENT OF THE SAMPLES PLATE BELT N.B.: before replacing of the Samples Plate belt, it is necessary remove the Plates Assembly from the instrument. 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59). 3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the Reagents Plate Home Sensor lever out of the Reagents Plate pulley. 4. Remove the belt from the Reagents Plate Motor pulley (Fig. 62). Ellipse Service Manual Rev. 04 01 October 2009 Page 68 Chapter 06 – SETTINGS AND ADJUSTMENTS 5. Remove the belt from the Reagents Plate pulley (Fig. 66). 6. Loosen the two grains of the Reagents Plate pulley (Fig. 66) and extract the pulley to the plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66). 7. Pull upwards the support Reagents Rack (Fig. 61) and remove it to the Plates Assembly. 8. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate (Fig. 59). 9. Loosen the four fastening screws of the Samples Plate Motor (Fig. 59). 10. Remove the belt from the Samples Plate Motor pulley (Fig. 62). 11. Remove the belt from the Samples Plate pulley (Fig. 66). 12. Substitute the belt with the new belt. 13. Insert the support Reagents Rack in the Plates Assembly (Fig. 60). 14. Wrap the new belt around the Samples Plate pulley. 15. Wrap the new belt around the Samples Plate Motor pulley. 16. Push the motor outward from the Plates pulleys axis and holding it in this position, so as to maintain the belt in tension, then tighten the four fastening screws of the Samples Plate Motor (Fig. 59). 17. Make sure that the belt thoots are correctly inserted on the Samples Plate pulley. 18. Make sure that the belt thoots are correctly inserted on the Samples Plate Motor pulley. 19. Rotate with accuracy the Samples Plate pulley and check that the belt tension is corrected. 20. Perform the procedure described into the previous Section 6.5.3 from step 32 to step 41. 21. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Samples Plate belt, the Reagents Plate Motor and the Plates Assembly, it is necessary to align mechanically the Samples Plate and the Reagents Plate by following the procedure described into the Section 6.5.15.1. Ellipse Service Manual Rev. 04 01 October 2009 Page 69 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5.7 REPLACEMENT OF THE REACTIONS PLATE BELT N.B.: before replacing of the Reactions Plate belt, it is necessary remove the Plates Assembly from the instrument. 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59). 3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the Reagents Plate Home Sensor lever out of the Reagents Plate pulley. 4. Remove the belt from the Reagents Plate Motor pulley (Fig. 62). 5. Remove the belt from the Reagents Plate pulley (Fig. 66). 6. Pull upwards the support Reagents Rack (Fig. 61) and remove it to the Plates Assembly. 7. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate (Fig. 59). 8. Loosen the two grains of the Reagents Plate pulley (Fig. 66) and extract the pulley to the plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66). 9. Loosen the four fastening screws of the Samples Plate Motor (Fig. 59). 10. Unscrew the Samples Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the Samples Plate Home Sensor lever out of the Samples Plate pulley. 11. Remove the belt from the Samples Plate Motor pulley (Fig. 62). 12. Remove the belt from the Samples Plate pulley (Fig. 66). 13. Loosen the two grains of the Samples Plate pulley (Fig. 66) and extract the pulley to the plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66). Ellipse Service Manual Rev. 04 01 October 2009 Page 70 Chapter 06 – SETTINGS AND ADJUSTMENTS 14. Loosen the four fastening screws of the Reactions Plate Motor (Fig. 59). 15. Remove the belt from the Reactions Plate Motor pulley (Fig. 62). 16. Remove the belt from the Reactions Plate pulley (Fig. 66). 17. Substitute the belt with the new belt. 18. Insert the support Reagents Rack in the Plates Assembly (Fig. 60). 19. Wrap the belt around the Reactions Plate pulley. 20. Wrap the belt around the Reactions Plate Motor pulley. 21. Push the motor outward from the Plates pulleys axis and holding it in this position, so as to maintain the belt in tension, then tighten the four fastening screws of the Reactions Plate Motor (Fig. 59). 22. Make sure that the belt thoots are correctly inserted on the Reactions Plate pulley. 23. Make sure that the belt thoots are correctly inserted on the Reactions Plate Motor pulley. 24. Rotate with accuracy the Reactions Plate pulley and check that the belt tension is corrected. 25. Replace the Samples Plate pulley and tighten the two its grains only after to have verified that the two Plate Motor pulleys are locked between of them (and with the Home Sensor disk slit aligned with respect to the slit of the other Home Sensor disk). 26. Perform the procedure described into the previous Section 6.5.3 from step 25 to step 29. 27. Tighten the Samples Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index. 28. Rotate with accuracy the Samples Plate pulley and check that the belt tension is corrected and that the Home Sensor disk not to touch the Samples Home Sensor. 29. Ellipse Perform the procedure described into the previous Section 6.5.3 from step 32 to step 41. Service Manual Rev. 04 01 October 2009 Page 71 Chapter 06 – SETTINGS AND ADJUSTMENTS 30. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Reactions Plate belt, the Samples Plate pulley, the Reagents Plate pulley and the Plates Assembly, it is necessary to align mechanically all the plates by following the procedure described into the Section 6.5.16.1. 6.5.8 REPLACEMENT OF THE REAGENTS PLATE PULLEY N.B.: before replacing of the Reagents Plate pulley, it is necessary remove the Plates Assembly from the instrument 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Perform the procedure described into the previous Section 6.5.7 from step 2 to step 13. 3. Substitute the pulley with the new pulley (Fig. 66). Samples Plate pulley Reactions Plate pulley Grain Home Sensor disk Reagents Plate pulley Home Sensor disk slit Fig. 66 - Plates pulleys 4. Ellipse Perform the procedure described into the previous Section 6.5.3 from step 32 to step 41. Service Manual Rev. 04 01 October 2009 Page 72 Chapter 06 – SETTINGS AND ADJUSTMENTS 5. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Reagents Plate pulley and the Plates Assembly, it is necessary to align mechanically the Reagents Plate by following the procedure described into the Section 6.5.14.2. 6.5.9 REPLACEMENT OF THE SAMPLES PLATE PULLEY N.B.: before replacing of the Samples Plate pulley, it is necessary remove the Plates Assembly from the instrument. 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Perform the procedure described into the previous Section 6.5.4 from step 2 to step 7. 3. Substitute the pulley with the new pulley (Fig. 66). 4. Perform the procedure described into the previous Section 6.5.7 from step 25 to step 29. 5. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Samples Plate pulley, the Reagents Plate pulley and the Plates Assembly, it is necessary to align mechanically the Samples Plate and the Reagents Plate by following the procedure described into the Section 6.5.15.2. 6.5.10 REPLACEMENT OF THE REACTIONS PLATE PULLEY N.B.: before replacing of the Reactions Plate pulley, it is necessary remove the Plates Assembly from the instrument 1. To remove the Plates Assembly, perform the relative procedure described into the previous Section 6.5.1. 2. Perform the procedure described into the previous Section 6.5.7 from step 2 to step 16. 3. Unscrew the Reactions Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the Reactions Plate Home Sensor lever out of the Reactions Plate pulley. Ellipse Service Manual Rev. 04 01 October 2009 Page 73 Chapter 06 – SETTINGS AND ADJUSTMENTS 4. Loosen the two grains of the Reactions Plate pulley (Fig. 67) and extract the pulley to the plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig.67). 5. Substitute the pulley with the new pulley (Fig. 66). 6. Replace the new Reactions Plate pulley and tighten the two its grains only after to have verified that it is locked on the superior surface. 7. Insert the Reactions Plate Home Sensor around the Home Sensor disk. 8. Tighten the Reactions Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect to its index. 9. Perform the procedure described into the previous Section 6.5.7 from step 18 to step 29. 10. To replace the Plates Assembly, perform the procedure described into the previous Section 6.5.1 in inverse order: from step 40 to step 1. N.B.: after having replaced the new Reactions Plate pulley, the Samples Plate pulley, the Reagents Plate pulley and the Plates Assembly, it is necessary to align mechanically all the plates by following the procedure described into the Section 6.5.16.2 . Ellipse Service Manual Rev. 04 01 October 2009 Page 74 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5.11 REPLACEMENT OF THE REAGENTS PLATE HOME SENSOR 1. Remove the instrument front panel unscrewing its four anchored screws. 2. Unplug the J9 connector from the Plate Interface Board (Fig. 57). 3. Unscrew the Home Sensors screw (Fig. 67). Reactions Plate Home Sensor lever adjusting screw Reactions Plate Home Sensor lever Samples Plate Home Sensor lever adjusting screw Reagents Plate Home Sensor lever adjusting screw Samples Plate Home Sensor lever Spacer column Reagents Plate Home Sensor lever Home Sensors screw Fig. 67 - Plates Home Sensors levers adjusting screws 4. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and remove the Reagents Plate Home Sensor lever from the Reagents Plate pulley. 5. Unscrew the two Reagents Plate Home Sensor fastening screws on the its lever (Fig. 68). 6. Substitute the Home Sensor with the new Home Sensor (Fig. 69). Ellipse Service Manual Rev. 04 01 October 2009 Page 75 Chapter 06 – SETTINGS AND ADJUSTMENTS Home Sensor Lever Home Sensor fastening screw Fig. 68 - Plate HomeSensor Fig. 69 – Plate Home Sensor Assy fastening screws [P/N: 9-10-0023-00] 7. Replace the new Home Sensor on its lever with the window Home Sensor outward-facing (Fig. 68) and screw the two Reagents Plate Home Sensor fastening screws. 8. Insert the Reagents Plate Home Sensor around the Home Sensor disk. 9. Replace the Reagents Plate Home Sensor lever on its location and screw the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect to its index. 10. Screw the Home Sensors screw (Fig. 67). 11. Plug in the J9 connector in the Plate Interface Board (Fig. 57). 12. Rotate with accuracy the Reagents Plate pulley and check that the Home Sensor disk not to touch the Reagents Home Sensor. 13. Place the instrument front panel screwing its four anchored screws. N.B.: after having replaced the new Reagents Plate Home Sensor, it is necessary to align mechanically the Reagents Plate by following the procedure described into the Section 6.5.14.3 . Ellipse Service Manual Rev. 04 01 October 2009 Page 76 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5.12 REPLACEMENT OF THE SAMPLES PLATE HOME SENSOR 1. Remove the instrument front panel unscrewing its four anchored screws. 2. Unscrew the Home Sensors screw (Fig. 67). 3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and remove the Reagents Plate Home Sensor lever from the Reagents Plate pulley. 4. Unscrew the first spacer of the Home Sensor levers with the relative washer (Fig. 67). 5. Unplug the J7 connector from the Plate Interface Board (Fig. 57). 6. Unscrew the Samples Plate Home Sensor lever adjusting screw (Fig. 67) and remove the Samples Plate Home Sensor lever from the Samples Plate pulley. 7. Unscrew the two Samples Plate Home Sensor fastening screws on the its lever (Fig. 68). 8. Substitute the Home Sensor with the new Home Sensor (Fig. 69). 9. Replace the new Home Sensor on its lever with the window Home Sensor outward-facing (Fig. 68) and screw the two Samples Plate Home Sensor fastening screws. 10. Insert the Samples Plate Home Sensor around the Home sensor disk. 11. Replace the Samples Plate Home Sensor lever on its location and screw the Samples Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect to its index. 12. Replace the spacer washer and screw the first spacer of the Home Sensor levers. 13. Plug in the J7 connector in the Plate Interface Board (Fig. 57). 14. Rotate with accuracy the Samples Plate pulley and check that the Home Sensor disk not to touch the Samples Home Sensor. 15. Insert the Reagents Plate Home Sensor around the Home Sensor disk. 16. Replace the Reagents Plate Home Sensor lever on its location and screw the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index. 17. Screw the Home Sensors screw (Fig. 67). 18. Rotate with accuracy the Reagents Plate pulley and check that the Home Sensor disk not to touch the Reagents Home Sensor. 19. Ellipse Place the instrument front panel screwing its four anchored screws. Service Manual Rev. 04 01 October 2009 Page 77 Chapter 06 – SETTINGS AND ADJUSTMENTS N.B.: after having replaced the new Samples Plate Home Sensor and the Reagents Plate Home Sensor, it is necessary to align mechanically the Samples Plate and Reagents Plate by following the procedure described into the Section 6.5.15.3. 6.5.13 REPLACEMENT OF THE REACTIONS PLATE HOME SENSOR 1. Remove the instrument front panel unscrewing its four anchored screws. 2. Unscrew the Home Sensors screw (Fig. 67). 3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and remove the Reagents Plate Home Sensor lever from the Reagents Plate pulley. 4. Unscrew the first spacer of the Home Sensor levers with the relative washer (Fig. 67). 5. Unscrew the Samples Plate Home Sensor lever adjusting screw (Fig. 67) and remove the Samples Plate Home Sensor lever from the Samples Plate pulley. 6. Unplug the J8 connector from the Plate Interface Board (Fig. 57). 7. Unscrew the second spacer of the Home Sensor levers with the relative washer (Fig. 57). 8. Unscrew the Reactions Plate Home Sensor lever adjusting screw (Fig. 67) and remove the Reactions Plate Home Sensor lever from the Reactions Plate pulley. 9. Unscrew the two Reactions Plate Home Sensor fastening screws on the its lever (Fig. 68). 10. Substitute the Home Sensor with the new Home Sensor (Fig. 69). 11. Replace the new Home Sensor on its lever with the window Home Sensor outward-facing (Fig. 68) and screw the two Reactions Plate Home Sensor fastening screws. 12. Insert the Reactions Plate Home Sensor around the Home sensor disk. 13. Replace the Reactions Plate Home Sensor lever on its location and screw the Reactions Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect to its index. 14. Replace the spacer washer and screw the second spacer of the Home Sensor levers. 15. Plug in the J8 connector in the Plate Interface Board (Fig. 57). 16. Rotate with accuracy the Reactions Plate pulley and check that the Home Sensor disk not to touch the Reactions Home Sensor. 17. Ellipse Insert the Samples Plate Home Sensor around the Home Sensor disk. Service Manual Rev. 04 01 October 2009 Page 78 Chapter 06 – SETTINGS AND ADJUSTMENTS 18. Replace the Samples Plate Home Sensor lever on its location and screw the Samples Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index. 19. Replace the spacer washer and screw the first spacer of the Home Sensor levers. 20. Rotate with accuracy the Samples Plate pulley and check that the Home Sensor disk not to touch the Samples Home Sensor. 21. Insert the Reagents Plate Home Sensor around the Home sensor disk. 22. Replace the Reagents Plate Home Sensor lever on its location and screw the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index. 23. Screw the Home Sensors screw (Fig. 67). 24. Rotate with accuracy the Reagents Plate pulley and check that the Home Sensor disk not to touch the Reagents Home Sensor. 25. Place the instrument front panel screwing its four anchored screws. N.B.: after having replaced the new Reactions Plate Home Sensor, the Samples Plate Home Sensor and the Reagents Plate Home Sensor, it is necessary to align mechanically all the plates by following the procedure described into the Section 6.5.16.3 6.5.14 ALIGNMENT AND ADJUSTMENTS OF THE REAGENTS PLATE 6.5.14.1 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate motor or of the Reagents Plate belt N.B.: in case of replacement of the Reagents Plate motor or of the Reagents Plate belt, it is necessary only control that the Reagents Plate is centered with respect to the Sampling Arm 1. Make sure the instrument (Ellipse) is turned on. 1. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate. 2. Launch the “Diagnostic” program. 3. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 4. Make sure that the distance between the top edge of the Washing Well and the tip of the Sampling Probe is 25 mm. Ellipse Service Manual Rev. 04 01 October 2009 Page 79 Chapter 06 – SETTINGS AND ADJUSTMENTS 5. Select “Arm” folder. 6. Click the "Go Std" button, then click the correspondent “Down” button. The arm will move to position #1 of the Standards Plate, then it will go down until to reach the right height. 7. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Standards Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, adjust (if necessary) the verticality by using the adjustable anchored bearing (Fig. 7), click the ‘Save’ button, then repeat the above steps from 6. 8. Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move to position #1 of the Reagents Plate, then it will go down until to reach the right height. 9. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Reagents Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, then click the ‘Save’ button. 10. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed. 6.5.14.2 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate pulley N.B.: in case of replacement of the Reagents Plate pulley, align and adjust the Reagents Plate with respect to the Sampling Arm. 1. Make sure that the instrument (Ellipse) is turned on. 2. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate. 3. Remove the instrument front panel unscrewing its four anchored screws. 4. Launch the “Diagnostic” program. 5. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 6. Make sure that the distance between the top edge of the Washing Well and the tip of the Sampling Probe is 25 mm. 7. Select “Plate” folder. 8. Click the "ALL Motors OFF" button. Ellipse Service Manual Rev. 04 01 October 2009 Page 80 Chapter 06 – SETTINGS AND ADJUSTMENTS 9. Loosen the two grains of the Reagents Plate pulley (Fig. 66). 10. Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been completed. 11. Select “Arm” folder. 12. Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move to position #1 of the Reagents Plate, then it will go down until to reach the right height. 13. Rotate manually the Reagents Plate until to reach the right centering of the position #1 of the Reagents Plate with respect to the Sampling Probe, then adjust (if necessary) the positioning of the Sampling Probe by using the ‘+’ and ‘-‘ keys. 14. Tighten the grain at sight of the Reagents Plate pulley only after to have verified that the three Plate Motor pulleys are locked between of them, then click the ‘Save’ button. 15. Select “Plate” folder and click the "ALL Motors OFF" button. 16. Rotate manually the Reagents Plate until to have in view the other grain of the Reagents Plate pulley, then tighten it. 17. Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been completed. 18. Select “Arm” folder. 19. Click the "Go Std" button, then click the correspondent “Down” button. The arm will move to position #1 of the Standards Plate, then it will go down until to reach the right height. 20. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Standards Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, adjust (if necessary) the verticality by using the adjustable anchored bearing, click the ‘Save’ button, then repeat the above steps from 15. 21. Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move to position #1 of the Reagents Plate, then it will go down until to reach the right height. 22. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Reagents Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, then click the ‘Save’ button. Ellipse Service Manual Rev. 04 01 October 2009 Page 81 Chapter 06 – SETTINGS AND ADJUSTMENTS 23. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed. 24. Place the instrument front panel screwing its four anchored screws. 6.5.14.3 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate Home Sensor N.B.: in case of replacement of the Reagents Plate Home Sensor, center the Reagents Plate with respect to the Sampling Arm adjusting the position of the Reagents Plate Home Sensor lever. 1. Make sure that the instrument (Ellipse) is turned on. 2. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate. 3. Remove the instrument front panel unscrewing its four anchored screws. 4. Launch the “Diagnostic” program. 5. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 6. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal), Reactions, Samples and Reagents Plates light up in green. 7. Select “Arm” folder. 8. Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move to position #1 of the Reagents Plate, then it will go down until to reach the right height. 9. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Reagents Plate. If it not be so: a. loosen the Reagents Plate Home Sensor lever adjusting screw (Fig. 67); b. turn slightly the Reagents Plate Home Sensor lever to the right toward in order to obtain the centering; c. tighten the Reagents Plate Home Sensor lever adjusting screw; d. repeat the steps from 5 to 9. 10. Click the "Go Std" button, then click the correspondent “Down” button. The arm will move to position #1 of the Standards Plate, then it will go down until to reach the right height Ellipse Service Manual Rev. 04 01 October 2009 Page 82 Chapter 06 – SETTINGS AND ADJUSTMENTS 11. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Standards Plate. 12. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed. 13. Place the instrument front panel screwing its four anchored screws. 6.5.15 LIGNMENT AND ADJUSTMENTS OF THE SAMPLES PLATE 6.5.15.1 Alignment and adjustment of the Samples Plate in case of replacement of the Samples Plate motor or of the Samples Plate belt N.B.: in case of replacement of the Samples Plate motor or of the Samples Plate belt, it is necessary control that the Samples Plate is centered with respect to the Sampling Arm and then, align and adjust the Reagents Plate 1. Perform the procedure described into the previous Section 6.5.14.2 from step 1 to step 20. 2. Select “Arm” folder. 3. Click the "Go Sample" button, then click the correspondent “Down” button. The arm will move to position #1 of the Sample Rack #1, then it will go down until to reach the right height. 4. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Sample Rack #1. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys and click the ‘Save’ button. 5. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed. 6. Place the instrument front panel screwing its four anchored screws. Ellipse Service Manual Rev. 04 01 October 2009 Page 83 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5.15.2 Alignment and adjustment of the Samples Plate in case of replacement of the Samples Plate pulley N.B.: in case of replacement of the Samples Plate pulley, align and adjust the Reagents Plate and the Samples Plate with respect to the Sampling Arm. 1. Make sure that the instrument (Ellipse) is turned on. 2. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13). 3. Launch the “Diagnostic” program. 4. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 5. Make sure that the distance between the top edge of the Washing Well and the tip of the Sampling Probe is 25 mm. 6. Select “Plate” folder. 7. Click the "ALL Motors OFF" button. 8. Loosen the two grains of the Samples Plate pulley (Fig. 66). 9. Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been completed. 10. Select “Arm” folder. 11. Click the "Go Sample" button, then click the correspondent “Down” button. The arm will move to position #1 of the Sample Rack #1, then it will go down until to reach the right height. 12. Rotate manually the Samples Plate until to reach the right centering of the position #1 of the Sample Rack #1 with respect to the Sampling Probe, then adjust (if necessary) the positioning of the Sampling Probe by using the ‘+’ and ‘-‘ keys. 13. Tighten the grain at sight of the Samples Plate pulley, then click the ‘Save’ button. 14. Select “Plate” folder and click the "ALL Motors OFF" button. 15. Rotate manually the Samples Plate until to have in view the other grain of the Samples Plate pulley, then tighten it. Ellipse Service Manual Rev. 04 01 October 2009 Page 84 Chapter 06 – SETTINGS AND ADJUSTMENTS 16. Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been completed. 17. Select “Arm” folder. 18. Click the "Go Sample" button, then click the correspondent “Down” button. The arm will move to position #1 of the Sample Rack #1, then it will go down until to reach the right height. 19. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Sample Rack #1. 20. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed. 21. Perform the procedure described into the previous Section 6.5.14.2 from step 7 to step 21. 6.5.15.3 Alignment and adjustment of the Samples Plate in case of replacement of the Samples Plate Home Sensor N.B.: in case of replacement of the Samples Plate Home Sensor, center the Samples Plate and the Reagents Plate with respect to the Sampling Arm adjusting the position of the relative Plate Home Sensor levers. 1. Make sure that the instrument (Ellipse) is turned on. 2. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate. 3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13). 4. Launch the “Diagnostic” program. 5. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 6. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal), Reactions, Samples and Reagents Plates light up in green. 7. Select “Arm” folder. 8. Click the "Go Sample" button, then click the correspondent “Down” button. The arm will move to position #1 of the Sample Rack #1, then it will go down until to reach the right height. Ellipse Service Manual Rev. 04 01 October 2009 Page 85 Chapter 06 – SETTINGS AND ADJUSTMENTS 9. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Sample Rack #1. If it not be so: a. loosen the Samples Plate Home Sensor lever adjusting screw (Fig. 6.5.17); b. turn slightly the Samples Plate Home Sensor lever to the right toward in order to obtain the centering; c. tighten the Samples Plate Home Sensor lever adjusting screw; d. repeat the steps from 5 to 9. 10. Click the "Go Sample" button, then click the correspondent “Down” button. The arm will move to position #1 of the Sample Rack #1, then it will go down until to reach the right height. 11. Make sure that the Sampling Probe is vertically aligned and centered with respect to position #1 of the Sample Rack #1. 12. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed. 13. Perform the procedure described into the previous Section 6.5.14.3 from step 7 to step 13. 6.5.16 LIGNMENT AND ADJUSTMENTS OF THE REACTIONS PLATE 6.5.16.1 Alignment and adjustment of the Reactions Plate in case of replacement of the Reactions Plate motor or of the Reactions Plate belt N.B.: in case of replacement of the Reactions Plate motor or of the Reactions Plate belt, it is necessary control that the Reactions Plate is centered with respect to the Sampling Arm and with respect to the Optic Assembly; then, align and adjust the Samples Plate and the Reagents Plate. 1. Make sure that the instrument (Ellipse) is turned on. 2. Remove the reactions cuvettes front cover. 3. Remove with accuracy the left reactions cuvettes cover (placed under the Washing Station cannulas) unscrewing its three screws. Ellipse Service Manual Rev. 04 01 October 2009 Page 86 Chapter 06 – SETTINGS AND ADJUSTMENTS 4. Take a with Reactions Plate Rack with a opaque cuvette #1 and then, substitute the Reactions Plate Rack #1 with this new Rack. 5. Make sure that the distance between the top edge of the Washing Well and the tip of the Sampling Probe is 25 mm. 6. Launch the “Diagnostic” program. 7. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 8. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal), Reaction, Samples and Reagent Plates light up in green. 9. Make sure that the lamp of the Optic Assembly is turned on. 10. Select “Optic” folder and click the "Spare" button. 11. Make sure that the light beam of the lamp is centered with respect to cuvette #1 of the Reactions Plate. If it not be so: perform the procedure described into the Section 6.5.16.3. 12. Select “Arm” folder. 13. Click the "Go Disp." button, then click the correspondent “Down” button. The arm will move to cuvette #81 of the Reactions Plate, then it will go down until to reach the right height. 14. Make sure that the Sampling Probe is vertically aligned and centered with respect to cuvette #81 of the Reactions Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, then click the ‘Save’ button. 15. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 16. Substitute the Reactions Plate Rack #1 with the previous Rack #1. 17. Replace with accuracy the left reactions cuvettes cover (placed under the Washing Station cannulas) screwing its three screws. 18. Replace the reactions cuvettes front cover. 19. Perform the procedure described into the previous Section 6.5.15.2 from step 6 to step 21. Ellipse Service Manual Rev. 04 01 October 2009 Page 87 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.5.16.2 Alignment and adjustment of the Reactions Plate in case of replacement of the Reactions Plate pulley N.B.: in case of replacement of the Reactions Plate pulley, align and adjust the Reactions Plate, the Samples Plate and the Reagents Plate with respect to the Sampling Arm. 1. Make sure the instrument (Ellipse) is turned on. 2. Remove the instrument front panel unscrewing its four anchored screws. 3. Remove the reactions cuvettes front cover. 4. Remove with accuracy the left reactions cuvettes cover (placed under the Washing Station cannulas) unscrewing its three screws. 5. Take a with Reactions Plate Rack with a opaque cuvette #1 and then, substitute the Reactions Plate Rack #1 with this new Rack. 6. Launch the “Diagnostic” program. 7. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 8. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal), Reaction, Samples and Reagent Plates light up in green. 9. Select “Plate” folder. 10. Click the "ALL Motors OFF" button. 11. Make sure that the two Reagents pulley grains are locked. 12. Make sure that the three Plate Home Sensor levers (Fig. 67) are centered with respect to the relative index. 13. Loosen the two grains of the Reactions Plate pulley (Fig. 66). 14. Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been completed. 15. Select “Optic” folder and click the "Spare" button. 16. Make sure that the lamp of the Optic Assembly is turned on. Ellipse Service Manual Rev. 04 01 October 2009 Page 88 Chapter 06 – SETTINGS AND ADJUSTMENTS 17. Rotate manually the Reactions Plate until to reach the right centering of the cuvette #1 of the Reactions Plate Rack #1 with respect to the light beam of the Optic Assembly. 18. Tighten the grain at sight of the Reactions Plate pulley. 19. Select “Plate” folder and click the "ALL Motors OFF" button. 20. Rotate manually the Reactions Plate until to have in view the other grain of the Reactions Plate pulley, then tighten it. 21. Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been completed. 22. Select “Optic” folder and click the "Spare" button. 23. Make sure that the cuvette #1 of the Reactions Plate Rack #1 is centered with respect to the light beam of the Optic Assembly. If it not be so: a. loosen the Reactions Plate Home Sensor lever adjusting screw (Fig. 67); b. turn slightly the Reactions Plate Home Sensor lever to the right toward in order to obtain the centering; c. tighten the Reactions Plate Home Sensor lever adjusting screw; d. select “Plate” folder and click the "Reset" button of the Reactions Plate; e. repeat the step 23. 24. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate. 25. Perform the procedure described into the previous Section 6.1.1 from step 4 to step 12. 26. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 27. Select “Arm” folder. 28. Click the "Go Disp." button, then click the correspondent “Down” button. The arm will move to cuvette #81 of the Reactions Plate, then it will go down until to reach the right height. 29. Make sure that the Sampling Probe is vertically aligned and centered with respect to cuvette #81 of the Reactions Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, then click the ‘Save’ button. 30. Ellipse Substitute the Reactions Plate Rack #1 with the previous Rack #1. Service Manual Rev. 04 01 October 2009 Page 89 Chapter 06 – SETTINGS AND ADJUSTMENTS 31. Replace with accuracy the left reactions cuvettes cover (placed under the Washing Station cannulas) screwing its three screws. 32. Replace the reactions cuvettes front cover. 33. Perform the procedure described into the previous Section 6.5.15.2 from step 4 to step 21. 6.5.16.3 Alignment and adjustment of the Reactions Plate in case of replacement of the Reactions Plate Home Sensor N.B.: in case of replacement of the Reactions Plate Home Sensor, center the Reactions Plate, the Samples Plate and the Reagents Plate with respect to the Sampling Arm adjusting the position of the relative Plate Home Sensor levers 1. Make sure the instrument (Ellipse) is turned on. 2. Remove the instrument front panel unscrewing its four anchored screws. 3. Remove the reactions cuvettes front cover. 4. Remove with accuracy the left reactions cuvettes cover (placed under the Washing Station cannulas) unscrewing its three screws. 5. Take a with Reactions Plate Rack with a opaque cuvette #1 and then, substitute the Reactions Plate Rack #1 with this new Rack. 6. Launch the “Diagnostic” program. 7. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 8. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal), Reaction, Samples and Reagent Plates light up in green. 9. Select “Optic” folder and click the "Spare" button. 10. Make sure that the cuvette #1 of the Reactions Plate Rack #1 is centered with respect to the light beam of the Optic Assembly. If it not be so: a. loosen the Reactions Plate Home Sensor lever adjusting screw (Fig. 67); b. turn slightly the Reactions Plate Home Sensor lever to the right toward in order to obtain the centering; Ellipse Service Manual Rev. 04 01 October 2009 Page 90 Chapter 06 – SETTINGS AND ADJUSTMENTS c. tighten the Reactions Plate Home Sensor lever adjusting screw; d. select “Plate” folder and click the "Reset" button of the Reactions Plate; e. repeat the step 10. 11. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate. 12. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset instrument procedure has been completed. 13. Select “Arm” folder. 14. Click the "Go Disp." button, then click the correspondent “Down” button. The arm will move to cuvette #81 of the Reactions Plate, then it will go down until to reach the right height. 15. Make sure that the Sampling Probe is vertically aligned and centered with respect to cuvette #81 of the Reactions Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys, then click the ‘Save’ button. 16. Substitute the Reactions Plate Rack #1 with the previous Rack #1. 17. Replace with accuracy the left reactions cuvettes cover (placed under the Washing Station cannulas) screwing its three screws. 18. Replace the reactions cuvettes front cover. 19. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general reset procedure has been completed. 20. Ellipse Perform the procedure described into the previous Section 6.5.15.3 from step 7 to step 13 Service Manual Rev. 04 01 October 2009 Page 91 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.6 WASHING STATION ASSEMBLY 6.6.1 SUBSTITUTION OF THE WASHING STATION ASSEMBLY N.B.: make sure that the Ellipse instrument is turned off before performing this substitution procedure. 1. Unscrew the fastening screw and remove the Washing Station protective cover (Fig. 70). 2. Remove the top cover of the Washing Station unscrewing the four fastening screws (Fig. 71). 3. Disconnect the eight tubes attached to the probes on the Washing Station (Fig. 72). 4. Unscrew the two Washing Station fastening screws (Fig. 71) and pull out the Washing Station from its housing. 5. Remove the instrument front panel unscrewing the four anchored screws (Fig. 73). 6. Remove the J5 and J10 connectors from the Plate Interface Board (Fig. 74). 7. Remove the ground wire from the Washing Station. 8. Remove the Washing Station Assembly and replace it with the new one. 9. Remount, repeating the above steps, 7 through 1, in reverse order. 10. After substituting and remounting, carry out the mechanical adjustment and functional check procedures as described in the next pages. Fastening screw Washing Station probes holder Washing Station protective cover Fig. 70 - Washing Station protective cover Ellipse Service Manual Rev. 04 01 October 2009 Page 92 Chapter 06 – SETTINGS AND ADJUSTMENTS Top Cover fastening screws Tubing Spring Washing Station fastening screws Fig. 72 - Washing Station probes Fig. 71 - Washing Station Assembly [P/N: 10-01299-00] Instrument front panel Front panel anchored screws Fig. 73 - Front panel anchored screws Ellipse Service Manual Rev. 04 01 October 2009 Page 93 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.6.2 SUBSTITUTION OF THE WASHING STATION HOME SENSOR N.B.: make sure that the Ellipse instrument is turned off before performing this substitution procedure. 1. Unscrew the fastening screw and remove the Washing Station protective cover (Fig. 70). 2. Remove the instrument front panel unscrewing the four anchored screws (Fig. 73). 3. Remove the two fastening screws from the Home Sensor Assy (Fig. 75), unplug the J10 connector from the Plate Interface Board (Fig. 74), take out the entire assy and replace it with the new one. 4. To remount, repeat the above steps in inverse order: from 3 to 1. 5. After substituting and remounting, carry out the mechanical adjustment and functional check procedures as described in the next pages. Washing Station Motor – J5 Washing Station Home Sensor – J10 01 Fig. 74 - Plate Interface Board [P/N: 30-01281-01] Ellipse Service Manual Rev. 04 01 October 2009 Page 94 Chapter 06 – SETTINGS AND ADJUSTMENTS Washing Station Home Sensor Home Sensor fastening screw Adjusting set screw Tubing guide Washing Station Motor Fig. 76 - Washing Station Motor [ Assy P/N: 10-00399-01] Fig. 75 - Washing Station Home Sensor [ Assy P/N: 9-10-0023-60] 6.6.3 SUBSTITUTION OF THE WASHING STATION MOTOR N.B.: make sure that the Ellipse instrument is turned off before performing this substitution procedure. 1. Unscrew the fastening screw and remove the Washing station protective cover (Fig. 70). 2. Unscrew the two Washing Station fastening screws (Fig. 71) and pull out the Washing station from its housing. 3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 73). 4. Remove the J5 connector from the Plate Interface Board (Fig. 74). 5. Loosen the two flywheel and motor joint screws. 6. Unscrew the four fastening screws on the motor (Fig. 76), take out it and replace it with the new one. 7. Remount, repeating the above steps, 6 through 1, in reverse order. 8. After substituting and remounting, carry out the mechanical adjustment and functional check procedures as described in the next pages. Ellipse Service Manual Rev. 04 01 October 2009 Page 95 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.6.4 WASHING STATION ASSEMBLY - MECHANICAL ADJUSTMENT 1. Make sure that the cuvette #1 of the Reactions Plate is centered with respect to the light beam of the Photometer reading channel. 2. Turn on the Ellipse instrument, unscrew the fastening screw of the Washing Station protective cover (Fig. 70) and remove the cover from its place. 3. Launch the Diagnostic program, select “Wash” folder and click the “Home” button. Wait until the reset Washing Station procedure has been completed. 4. Make sure that the Home Sensor of the Washing Station light up in green. 5. Click the “Down to Adjust” button in the “Wash” folder. The Washing Station will go down until to reach the right height for the adjusting. 6. Make sure that the five probes of the Washing Station are centered with respect to the five correspondent cuvettes of the Reactions Plate. If it not be so, adjust in the following way: a. loosen the two Washing Station fastening screws (Fig. 71); b. shift horizontally – carefully – the Washing Station in order to obtain the centering of the five probes with respect to the five correspondent cuvettes; c. tighten the two Washing Station fastening screws (Fig. 71); d. click the “Home” button in the “Wash” folder and wait until the reset Washing Station procedure has been completed; e. make sure that the Home Sensor of the Washing Station light up in green; f. click the “Down to Adjust” button in the “Wash” folder and wait until the request movement has been completed; g. repeat the above step 6. 7. Click the “Home” button in the “Wash” folder. Wait until the reset Washing Station procedure has been completed. 8. Click the “Down (Full)” button in the “Wash” folder. The Washing Station will go down until to reach the cuvettes of the Reactions Plate. 9. Make sure that the five probes of the Washing Station are centered with respect to the five correspondent cuvettes of the Reactions Plate and, in particular, that the Drying Pad run freely along the walls of the correspondent cuvette. 10. Make sure that the five probes of the Washing Station are slightly raised. If it not be so, adjust in the following way: a. click the “Home” button in the “Wash” folder and wait until the reset Washing Station procedure has been completed; b. make sure that the Home Sensor of the Washing Station light up in green; c. loosen the two fastening screws of the Washing Station Home Sensor (Fig. 75); d. move slightly the Washing Station Home Sensor to the right vertically toward in order to obtain the positioning of the Washing Station probes to the correct height; e. tighten the two fastening screws of the Washing Station Home Sensor (Fig. 75); f. click the “Home” button in the “Wash” folder and wait until the reset Washing Station procedure has been completed; g. make sure that the Home Sensor of the Washing Station light up in green; Ellipse Service Manual Rev. 04 01 October 2009 Page 96 Chapter 06 – SETTINGS AND ADJUSTMENTS h. ick the “Down (Full)” button in the “Wash” folder and wait until the Washing Station has been reach the cuvettes of the Reactions Plate; i. repeat the above step 10. 11. Click the “Home” button in the “Wash” folder. Wait until the reset Washing Station procedure has been completed. 12. Make sure that the Home Sensor of the Washing Station light up in green. 13. Replace the Washing Station protective cover on the Washing Station and screw its fastening screw (Fig. 70). Additional information: make sure that the Sampling Probe is centered with respect to the Reactions Plate cuvette. If it not be so, perform the relative adjusting procedure. Adjusting set screw If the horizontal backlash of the Washing Station probes holder (Fig. 70) is excessive, then adjust it acting on the adjusting set screw (Fig. 75). 6.6.5 WASHING STATION - FUNCTIONAL CHECK 1. Turn on the Ellipse system and launch the “Analyzer” program. 2. Launch the Diagnostic program and select the "Wash " folder. 3. Click on the “Washing Cycle” button and perform the functional check of the Washing Station through the following steps: 4. Select, in the “Cycle #” field, the number of washing cycles to be performed from 1 to 10 and press the “Washing Cycle” button. Attention: if there are doubts about the right washing system functionality, select only one washing cycle in order to avoid cuvettes overflow. 5. Remove the reaction rack cover and the left side reaction rack cover unscrewing the three anchored screws. 6. Put one reaction rack below the Washing Station, permitting at the cannulas to reach the bottom of the cuvettes (as shown in Fig. 77). Reaction rack Fig. 77 - Washing Station and reaction rack Ellipse Service Manual Rev. 04 01 October 2009 Page 97 Chapter 06 – SETTINGS AND ADJUSTMENTS 7. By referring to the “Washing Test” window, push the “OK” button to start the Washing Cycle. 8. Check that the cuvettes are correctly filled and emptied (Fig. 77): • Dispensed volumes shall be: Æ 300 µL ± 50 into the first cuvette, two cycles Æ 300 µL ± 50 into the second cuvette, two cycles Æ 300 µL ± 50 into the third cuvette, two cycles plus one third cycle where the dispensed volume shall be 400 µL ± 50 If not, check and clean in case the relevant hydraulic line including micro-pump and valve, taking reference in the Hydraulic Diagram (SI-16-00571-01). 9. Remove the reaction rack below the Washing Station and remount the reaction rack covers. 10. Exit from the Diagnostic program by pressing the “Diagnostic” button. 11. Exit from the “Analyzer” program by pressing the “Shutdown” button. 6.7 WASH PUMP ASSEMBLY 6.7.1 SUBSTITUTION OF THE WASH PUMP ASSEMBLY N.B.: make sure that the Ellipse instrument is turned off before performing this substitution procedure. Rear door fastening screw Fig. 78 Instrument rear view Ellipse Service Manual Rev. 04 01 October 2009 Page 98 Chapter 06 – SETTINGS AND ADJUSTMENTS Micro-Pump P2 Assy Cd. 05-01826-16 Anchored screw µP6 µP5 µP4 µP3 µP2 Micro-Pump P6 Assy Cd. 05-01711-20 Electro-Valve EV2 (2 way) Cd. 09-35-0035-00 Fig. 79 Wash Pump Assembly Cd. 05-01488-00 Unscrew the four fastening screws to remove the door located on the instrument rear side to access the Wash Pump Assembly (Fig. 78) . Disconnect the air tubes from µP6 and the tubes arriving on the Electro-Valves and Micro-Pumps (Fig. 79). Unscrew the four anchored screws indicated which hold the Wash Pump Assembly (Fig. 79). Disconnect connectors J1, J2 e J6 from the Hydraulics Interface board (Fig. 81). Remove the Wash Pump Assembly from its site and replace it with the new one. Remount, repeating the above steps, 4 through 1, in reverse order. After mounting the new module, carry out the function check as described in the next pages. Ellipse Service Manual Rev. 04 01 October 2009 Page 99 Chapter 06 – SETTINGS AND ADJUSTMENTS N.B:. If power loss is experienced in the filling or emptying of liquid (after a period of instrument activity), perform the following operations before substituting or servicing any part(s): • Disconnect the liquid input tube from the malfunctioning pump. • Insert a small amount of liquid using a syringe from the disconnected ends in order to wet the membranes (Fig. 80). Remove the syringe and reconnect the tube(s); perform functioning check as described in the next points. Fig. 80 Hydraulics Interface board Cd. 30-01626-00 J6 J1 J2 Ellipse Service Manual Rev. 04 01 October 2009 Page 100 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 81 Wash Pump Assembly (Rear view) 6.7.2 SUBSTITUTION OF PART(S) N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. 1. Unscrew the four fastening screws to remove the door located on the instrument rear side to access the Wash Pump Assembly (Fig. 78) . 2. For each part to be replaced, if necessary, disconnect the relative electrical power connection, the hydraulic tubes and the fastening screws. 3. If necessary, unscrew the four anchored screws of the assembly and turn it ahead (Fig.81). 4. Remount, repeating the above steps, 3 through 1, in reverse order. 5. After mounting the new part, carry out the function check as described in the next pages. 6.7.3 FUNCTION CHECK FOR THE WASH PUMP ASSEMBLY 1. Turn on the "Ellipse" system and launch the “Analyzer” program. 2. Launch the diagnostic program and select the "Wash " function. 3. Click on the “Washing Cycle” key and perform the functional check of the Washing Station through the following steps: 4. Select, in the “Cycle #” field, the number of washing cycles to be performed from 1 to 10 and press the “Washing Cycle” key. Attention: if there are doubts about the right washing system functionality, select only one washing cycle in order to avoid cuvettes overflow 5. Remoe the reaction rack cover and the left side reaction rack cover unscrewing the three anchored screws. 6. Put one reaction rack below the washing station, permitting to the cannulas to reach the cuvettes button (Fig. 82). Ellipse Service Manual Rev. 04 01 October 2009 Page 101 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 82 Washing station 7. By referring to the window here below, push the “OK” key to start the Washing Cycle Check that the cuvettes are correctly filled and emptied (Fig. 82) • Dispensed volumes shall be: 300 µL ± 50 into the first cuvette, two cycles 300 µL ± 50 into the second cuvette, two cycles 300 µL ± 50 into the third cuvette, two cycles plus one third cycle where the dispensed volume shall be 400 µL ± 50 If not, check and clean in case the relevant hydraulic line including micro-pump and valve, taking reference in the Hydraulic Diagram (SI 16-00571-01). Remove the reaction rack below the washing station and remount the reaction rack covers. Exit the Diagnostic program by pressing the “Diagnostic” key. Exit the Analyzer program press “Shutdown” key. Ellipse Service Manual Rev. 04 01 October 2009 Page 102 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.8 PERISTALTIC PUMP ASSEMBLY 6.8.1 SUBSTITUTION OF THE PERISTALTIC PUMP ASSEMBLY Peristaltic pump fastening screw Peristaltic pump door Hinged guide tube Fig. 83 Instrument right view Fig. 84 fastening fitting Peristaltic Pump Assy Cd. 05-01489-00 1. Open the door located on the right side of the instrument Fig. 83. 2. Unhook the hinged guide by pushing its up side toward the left and up, as reported in Fig. 85. 3. Pull out the tubes off their relative nipples unscrewing the fastening fittings (Fig. 84). 4. Remove the front panel unscrewing the four anchored screws (Fig. 87). 5. Unplug the J3 connector from the Sampling Interface Board. (Fig. 88). 6. Unscrew the two fastening screws (Fig. 84), remove the Peristaltic Pump module and substitute it with the new one. 7. To remount, repeat the above steps in reverse order: from 6 to 1. 8. After mounting the new module, carry out the function check for the peristaltic pump module as described in the next pages. Ellipse Service Manual Rev. 04 01 October 2009 Page 103 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.8.2 SUBSTITUTION OF THE PERISTALTIC PUMP TUBES 1. Open the door located on the right side of the instrument Fig. 83. 2. Unhook the hinged guide by pushing its up side toward the left and up, as reported in Fig. 85. 3. Pull out the tubes off their relative nipples unscrewing the fastening fittings (Fig. 84) and substitute with new ones. 4. Remount, repeating the above steps in reverse order: from 3 to 1. • Always use original replacement parts; never lubricate the peristaltic pump roller bearings. • After prolonged instrument inactivity: verify the efficiency of the peristaltic pump tubes. Please note: if the instrument is not able to completely empty the washing well, check the condition of the tube and make sure that there are not obstruction into the hydraulic lines. Peristaltic Pump Rotor Pulley screw Peristaltic Pump Motor Fig. 85 P.Pump Rotor Cd.05-01512-00 Ellipse Service Manual Fig.86 P.Pump Motor Cd. 10-00399-01 Rev. 04 01 October 2009 Page 104 Chapter 06 – SETTINGS AND ADJUSTMENTS Fig. 87 Instrument front view Instrument front panel anchored screw J 3 - Peristaltic pump Motor Fig. 88 Arm Interface board Cd. 30-01282-01 Ellipse Service Manual Rev. 04 01 October 2009 Page 105 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.8.3 SUBSTITUTION OF THE PERISTALTIC PUMP MOTOR 1. Open the door located on the right side of the instrument Fig. 83. 2. Unhook the hinged guide by pushing its up side toward the left and up, as reported in Fig. 85. 3. Pull out the tubes off their relative nipples unscrewing the fastening fittings (Fig. 84). 4. Remove the front panel unscrewing the four anchored screws (Fig. 87). 5. Unscrew the two fastening screws (Fig. 84), unplug the J3 connector from the Sampling Interface Board (Fig. 88) and remove the Peristaltic Pump module from his housing. 6. Unscrew the four motor fastening screws and loosen the two motor pulley screws which hold the peristaltic pump rotor(Fig. 85). 7. Take out the motor and substitute it with the new one. 8. To remount, repeat the above steps in reverse order: from 6 to 1. 9. After mounting the new motor, carry out the function check for the peristaltic pump module as described in the next pages. 6.8.4 FUNCTION CHECK FOR THE PERISTALTIC PUMP 1. Turn on the Ellipse system (instrument and computer) and launch the “Analyzer” program. Select "Start Work" and perform one or more cuvette washing cycles. Make sure that the sampling probe washing well empties completely without any leakage of liquid. If leakage should occur, replace the Peristaltic Pump tubes as illustrated in the previous points. 2. Close the "Ellipse" program by clicking on "Shutdown". Ellipse Service Manual Rev. 04 01 October 2009 Page 106 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.9 SUBSTITUTION OF THE ELECTRONICS BOARDS 6.9.1 ANALYTICAL CONTROL BOARD Cd. 30-01283-01 N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. Instrument front panel anchored screw Rear door anchored screw Fig. 89 Instrument rear view Fig. 90 Instrument front view 1. Unscrew the two anchored screws to open the door located on the rear side of the instrument to access the Electronic Assy (Fig. 89) . 2. Take out the Electronic Assy as showed (Fig. 3), disconnect connectors J1, J6, J8, J9, J10, J11, J13, J20 and the ground wire from the Analytical Control board (Fig. 91). 3. Unscrew the four fastening screws on the Analytical Control board, disconnect it from the Stepper Motors Driver Board and replace it with the new one. 4. Remount, repeating the above steps, 3 through 1, in reverse order. 5. After mounting the new board, turn on the Ellipse system and launch the “Diagnostic program” in order to transfer all the previous instrument settings to the new Analytical control board. 6. To exit the diagnostic program, press the “Diagnostic” key. 7. To exit the analyzer program, press the “Shutdown” key. Ellipse Service Manual Rev. 04 01 October 2009 Page 107 Chapter 06 – SETTINGS AND ADJUSTMENTS J1- Stepper Motors Driver Board J6 J8 J9 J11 J10 J20 J1 J13 J7-Analytical Control board Board fastening screw Fig. 91 Electronic Assy J4 J3 J2 Fig. 92 Electronic Assy Board fastening screw 6.9.2 STEPPER MOTORS DRIVER BOARD Cd. 30-01284-01 N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. 1. Unscrew the two anchored screws to open the door located on the rear side of the instrument to access the Electronic Assy (Fig. 89) . 2. Take out the Electronic Assy as showed (Fig. 91) and disconnect connectors J2, J3, J4 (Fig.92). 3. Unscrew the two fastening screws on the Stepper Motors Driver Board, disconnect it from the Analytical Control board and replace it with the new one. 4. Remount, repeating the above steps, 3 through 1, in reverse order. Ellipse Service Manual Rev. 04 01 October 2009 Page 108 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.9.3 ARM INTERFACE BOARD Cd. 30-01282-01 N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. 1. Remove the front panel unscrewing the four anchored screws (Fig. 90). 2. Unplug the J1, J2, J3, J4, J5, J6, J8, J13, J16, J17, J21, J23, J25, J26 J27connectors and the ground wire from the Arm Interface board (Fig. 93). 3. Unscrew the four fastening screws of the Arm Interface board and replace it with the new one. 4. Remount, repeating the above steps, 3 through 1, in reverse order. J1 J2 J3 J4 J16 J17 J5 J6 J13 J8 J27 J21 J23 J26 J25 Fig. 93 Ellipse Service Manual Rev. 04 01 October 2009 Page 109 Chapter 06 – SETTINGS AND ADJUSTMENTS 6.9.4 PLATE INTERFACE BOARD Cd. 30-01281-01 N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution procedure. 1. Remove the front panel unscrewing the four anchored screws (Fig. 90). 2. Unplug the J1, J2, J3, J4, J5, J6, J7, J8, J9, J10, J11, J14, J15, J16, J26, J27, J28 connectors and the ground wire from the Plate Interface board (Fig. 94). 3. Unscrew the four fastening screws of the Plate Interface board and replace it with the new one. 4. Remount, repeating the above steps, 3 through 1, in reverse order. J1 J2 J3 J4 J8 J9 J5 J6 J14 J7 J10 J16 J11 J28 J15 J26 J27 Fig. 94 Plate Interface board Ellipse Service Manual Rev. 04 01 October 2009 Page 110 Chapter 07 - MAINTENANCE CHAPTER 07 - MAINTENANCE INDEX 7. MAINTENANCE…………………………………………………………………….…..2 7.1 PREVENTIVE MAINTENANCE……………………………………………………….2 7.2 LIST OF PARTS SUBJECT TO WEAR AND USAGE………………………………...4 7.3 SAMPLING PROBE - CLEANING PROCEDURE…………………………………….5 7.4 WASH STATION CANNULAS - CLEANING PROCEDURE..…………..…………...5 7.5 WASH SOLUTION BOTTLES - CLEANING……………………………..…………...6 7.5.1 WASH SOLUTION BOTTLES - CLEANING PROCEDURE…………………...…….6 7.6 TIP WASHING ………...……………………………………………………………......7 7.6.1 TIP - CLEANING PROCEDURE………………………………………………..………7 7.6.2 PROCEDURE FOR REPLACING THE TIP……………………………………………9 7.7 HYDRAULIC CIRCUIT WASHING ………...…………………………..….………..10 7.7.1 HYDRAULIC CIRCUIT - CLEANING PROCEDURE…….…………………….…..10 7.8 CHANGING THE PERISTALTIC PUMP TUBES…………….…………………..….11 7.8.1 PROCEDURE FOR REPLACING THE PERISTALTIC PUMP TUBES………..…...11 7.9 CHANGING THE PHOTOMETER LAMP………………………………………….. .12 7.9.1 PROCEDURE FOR REPLACING THE PHOTOMETER LAMP……………….….. .12 7.10 CHANGING THE REACTION CUVETTES………………………….…………….. .14 7.10.1 PROCEDURE FOR REPLACING THE REACTION CUVETTES………..………….14 7.11 CHANGING THE TUBES KIT…………………………………………………………16 7.11.1 PROCEDURE FOR REPLACING THE TUBES KIT…………………..………...……16 7.12 TABLE C – LIST OF THOSE MAINTENANCE PROCEDURES THAT CAN BE PERFORMED BY THE USER AND/OR BY THE MAINTENANCE TECHNICIAN …………………………………………………………….........................……………..17 7.13 Ellipse DECONTAMINATION PROCEDURE………………………………………………...19 Service Manual Rev. 03 01 June 2009 Page 1 Chapter 07 - MAINTENANCE 7 MAINTENANCE This chapter contains all those routine operations, which concern instrument maintenance. Said procedures, listed and described below, should be carefully and scrupulously followed in order to guarantee the manufacturer’s specifications and the perfect working order of the instrument over time. 7.1 PREVENTIVE MAINTENANCE MAINTENANCE SCHEDULE Table A, illustrated below, lists all those procedures to be carried out by the user/operator and the relative frequency schedule. Strict adherence to said schedule will guarantee the optimal operative efficiency of the instrument. TABLE A – MAINTENANCE SCHEDULE FREQUENCY PROCEDURE NOTES Check the levels of all the wash DAILY – Before launching solutions (Rinse, Water and See Table B “Start Work” Cleaning) DAILY – Before launching Check the levels of Reagents, See Table B “Start Work” Standards and Controls Check the levels of the Waste DAILY – Before launching Bottles and, if necessary, empty “Start Work” them See the User’s Manual, DAILY – Before launching Carry out a WBL cycle Chapter 03– Description “Start Work” of Instrument Software Clean the Sampling Probe using DAILY – After Shutdown See Procedure 7.3 either paper towelling or gauze Ellipse Service Manual Rev. 03 01 June 2009 Page 2 Chapter 07 - MAINTENANCE TABLE B – MAINTENANCE SCHEDULE FREQUENCY EVERY TWO WEEKS EVERY TWO WEEKS PROCEDURE NOTES Clean the Wash Station’s four See Procedure 7.4 cannulas Clean the wash solution bottles See Procedure 7.5 (Rinse, Water and Cleaning) ONCE A MONTH Clean the Tip See Procedure 7.6 ONCE A MONTH Clean the Hydraulic Circuit See Procedure 7.7 EVERY THREE MONTHS Change the Reaction Cuvettes See Procedure 7.10 EVERY SIX MONTHS Change the Peristaltic Pump Tubes See Procedure 7.8 EVERY SIX MONTHS Change the Tip See Procedure 7.6 ONCE A YEAR Change all the tubes (Tube Kit) See Procedure 7.11 EVERY 2000 HOURS Change the Photometer lamp See Procedure 7.9 • N. B.: the above-described maintenance schedule refers to that situation in which the workload of the Analyzer is approximately 500 tests per day. The interval frequency may vary according to the individual instrument’s daily workload. Ellipse Service Manual Rev. 03 01 June 2009 Page 3 Chapter 07 - MAINTENANCE 7.2 LIST OF PARTS SUBJECT TO WEAR AND USAGE Description Reagents containers Reagents containers Samples cups Short samples cups Adapter for short samples cups Reaction sectors Washing solution bottle Bottle level sensor Tubing Kit for peristaltic pump Tubing Kit – complete Cleaning solution Rinse solution Sampling probe (internal needle) Complete Sampling probe Drying Pad Halogen Lamp (6 V - 10 W) Interferential filters Kit Fuse 6,3 A-T Inlet/outlet fitting for Rinse & Clean conts Type 35 ml 6 ml 0.8 ml 1 ml 2 lt 250 ml 50 ml 5x20 Cuvettes protection cover Reagent protection cover Reagent plate Samples rack Washing station, first or second cannulas (A) Washing station, third cannula (B) Washing station, fourth cannula (C) Washing station, fifth cannula without pad (D) Diluter Micro-Pump Air Micro-Pump (µP 6) Micro-Pump (µP 2 ÷5) Predilution rack Solenoid Valve –2 way Solenoid Valve –3 way Ellipse Service Manual Rev. 03 Quantity 12 12 1000 1000 1 6 1 1 3 1 2 1 1 1 1 1 1 10 Code C101-00190-00 C101-00191-00 AS-65-0002 AS-65-0100 9-01-0609-00 C101-00217-00 9-35-0041-00 9-05-0078-00 65-01835-00 65-01836-00 ASRN0020 ASRN0021 05-00707-00 10-00703-00 01-01920-00 9-35-0016-00 9-65-0029-00 C130-01238-08 1 01-01224-00 1 1 1 1 1 1 1 1 1 1 1 1 1 1 05-01249-00 10-00584-00 10-00585-00 05-01829-00 05-01633-00 05-01633-01 05-01638-00 05-01919-00 05-01710-40 05-01711-20 05-01826-16 05-01735-00 9-35-0035-00 9-35-0036-00 01 June 2009 Page 4 Chapter 07 - MAINTENANCE 7.3 7.4 7.5 SAMPLING PROBE – CLEANING PROCEDURE 1) Turn off the Analyzer 2) Use only lint-free paper towelling or gauze 3) Dampen the gauze or paper towelling with distilled water and clean the outside of the sampling probe. Wipe the probe from the top downwards only! This is to avoid that any bits of cloth, paper or lint fibres accidentally enter the probe itself. 4) The manufacturer suggests that once weekly the above-described cleaning procedure be performed using, instead of only simple distilled water, a 5% sodium hypochlorite solution to dampen the gauze and then be repeated using distilled water. WASH STATION CANNULAS – CLEANING PROCEDURE 1) Turn off the Analyzer. 2) Place a sheet of paper under the wash station cannulas in order to keep any extraneous material from accidentally falling into the cuvettes. 3) Use only lint-free paper towelling or gauze. 4) Dampen the gauze or paper towelling with distilled water and clean the outside of the cannulas. Wipe the cannulas from the top downwards only! This is to avoid that any bits of cloth, paper or lint fibres accidentally enter the cannulas themselves. 5) The manufacturer suggests that once weekly the above-described cleaning procedure be performed using, instead of only simple distilled water, a 5% sodium hypochlorite solution to dampen the gauze and then be repeated using distilled water. WASH SOLUTION BOTTLES – CLEANING During normal use and over time, mold and dust can build up inside the wash solution bottles. For this reason, it is extremely important that they be periodically washed. Said cleaning must be thorough and meticulous in order to insure that every trace of mold or residue be removed. How often the bottles must be cleaned depends on their use and on the quality of the distilled water used in that particular laboratory. However, the manufacturer recommends thorough washing at least once every two weeks. It is extremely important that the user not underestimate the risks associated with mold and dust particles. They are to be regarded as a serious hazard as they can be the cause of instrument malfunction. The wash solution bottles are located on the right side of the analyzer. Ellipse Service Manual Rev. 03 01 June 2009 Page 5 Chapter 07 - MAINTENANCE 7.5.1 WASH SOLUTION BOTTLES - CLEANING PROCEDURE 1) Turn off the Analyzer. 2) Pull the level sensor connectors out from the bottle caps. 3) Take the caps off the bottles and empty them. 4) Fill each bottle with a 5% sodium hypochlorite solution. 5) Clean the inside of each bottle using a bottlebrush in order to remove all traces of mold and/or residue. 6) Leave the sodium hypochlorite solution stand in the bottles for at least ten minutes. 7) Empty the bottles, rinse them repeatedly and well with tap water, and then twice more using distilled water. 8) Dry the bottles. 9) Fill the bottles with their proper solutions. 10) Replace the bottles in their respective housings. 11) Close the bottles and reconnect the level sensors. 12) Carry out two ‘Wash Cuvettes’ cycles and two ‘WBL’ cycles. Check system efficiency by comparing the WBL values and the test results obtained against the laboratory’s quality control values. 7.6 TIP WASHING The Tip is used to dry the cuvettes after they have been washed. This drying process is carried out via aspiration and therefore, over time, the Tip will necessarily absorb various contaminating particles. The manufacturer suggests that the Tip be replaced every six months. Said frequency may vary depending on the workload of the individual laboratory and the operating conditions/environment of the single instrument. The Tip must be washed regularly in order to guarantee proper functioning and must be replaced with a new Tip as necessary. Ellipse Service Manual Rev. 03 01 June 2009 Page 6 Chapter 07 - MAINTENANCE 7.6.1 TIP - CLEANING PROCEDURE 1. Turn off the Analyzer. 2. Unscrew the fastening screw and remove the Washing station protective cover (Fig. 1). 3. Remove the top cover of the washing station unscrewing the four fastening screws (Fig. 2). 4. Remove the cannula that contains the Tip (the 5th cannula) and the spring (Fig. 3). 5. Disconnect the aspiration tube (Fig. 4). 6. Once the cannula has been removed, immerge it in a 5% solution of sodium hypochlorite for at least 15 minutes (Fig. 5). 7. Attach a 10 ml syringe to the cannula and aspirate and dispense the hypochlorite solution through the cannula (and the Tip) until the latter is completely clean (Fig. 6). This aspirating and dispensing forces the liquid through the Tip fibers in both directions 8. Once the Tip is clean, repeat the aspirating and dispensing cycle 10 more times using distilled water, then disconnect the syringe. 9. Re-connect the aspiration tube to the cannula (Fig. 7). 10. Re-position the cannula in its housing being careful to not kink the aspiration tube 11. Re-position the spring on the cannula’s cylindrical axel. 12. Re-position the top cover of the washing station, making sure that the cylindrical tops of the cannulas fit into the corresponding holes on it. 13. Manually move the wash station downwards and centre the tip with the cuvette turning it. 14. Remount the Washing station protective cover (Fig. 1). Washing Station protective cover Fig.1 Ellipse Instrument Top view Service Manual Rev. 03 01 June 2009 Page 7 Chapter 07 - MAINTENANCE Spring 5th cannula Fastening screw Fig. 2 Fig. 3 Aspiration tube Fig. 4 Fig. 5 Sodium hypochlorite Connect the syringe to the cannula Fig. 6 Ellipse Fig. Service Manual Rev. 03 01 June 2009 Page 8 Chapter 07 - MAINTENANCE 7.6.2 PROCEDURE FOR REPLACING THE TIP 1. Turn off the Analyzer. 2. Unscrew the fastening screw and remove the Washing station protective cover (Fig. 1). 3. Remove the top cover of the washing station unscrewing the four fastening screws (Fig. 2). 4. Remove the cannula that contains the Tip (the 5th cannula) and the spring (Fig. 3). 5. Disconnect the aspiration tube (Fig. 4), remove the used Tip and replace it with a new one. Press lightly to push the new Tip into place – be careful to not press too hard as this could deform the Tip (Fig. 8). 6. Insert the cannula containing the new Tip into a cuvette, pushing it down inside until it takes on the shape of the inside of the cuvette (Fig. 9). 7. Remove the cuvette, connect the aspiration tube to the cannula and reposition the cannula back into its proper housing in the wash station. Be careful to not kink the aspiration tube while doing so. 8. Reposition the spring in its proper housing. Remount the wash station coverlid making sure that the cylindrical tops of the cannulas fit into the corresponding holes on the wash station coverlid. 9. Manually move the wash station downwards and centre the tip with the cuvette turning it. 10. Remount the Washing station protective cover (Fig. 1). Push the new Tip into the cuvette Pull to remove the used Tip Fig. 8 Ellipse Fig. 9 Service Manual Rev. 03 01 June 2009 Page 9 Chapter 07 - MAINTENANCE 7.7 HYDRAULIC CIRCUIT WASHING During normal use and over time, mold and dust can build up inside the wash bottles and can have a negative effect on the hydraulic circuit, compromising the correct functioning of the micro-pump and valves. This, in turn, can lead to inefficiency in the sampling probe and cuvette washing system. For this reason, it is extremely important that the hydraulic circuit be periodically washed. Said cleaning must be thorough and meticulous in order to assure that every trace of mold or residue be removed. How often the hydraulic circuit must be washed depends on the operating conditions/environment of the single instrument and the quality of the distilled water used in that particular laboratory. The manufacturer recommends thorough washing at least once a month. It is extremely important that the user not underestimate the risks associated with mold and dust particles. They are to be regarded as a serious hazard as they can be the cause of instrument malfunction. The hydraulic circuit input cannulas are located inside the bottles on the left side of the Analyzer. 7.7.1 HYDRAULIC CIRCUIT - CLEANING PROCEDURE 1) Turn on the Analyzer. 2) Prepare a bottle containing 500 ml of a 5% sodium hypochlorite solution 3) Insert the four aspiration cannulas, located inside the liquids bottles, into the bottle containing the sodium hypochlorite solution. 4) Have the instrument carry out a ‘WBL’ cycle and then a ‘Wash cuvettes’ cycle. 5) Wait for fifteen minutes. Clean and dry the three cannulas and then insert them into a bottle containing distilled water. 6) Have the instrument carry out a ‘WBL’ cycle and then a ‘Wash cuvettes’ cycle. 7) Insert the four aspiration cannulas back into their respective bottles. Said bottles should have, in the meantime, been cleaned and filled with a fresh supply of the required solution. 8) Have the instrument carry out a ‘WBL’ cycle and then a ‘Wash cuvettes’ cycle. Check system efficiency by comparing the WBL values and the test results obtained against the laboratory’s quality control values. Ellipse Service Manual Rev. 03 01 June 2009 Page 10 Chapter 07 - MAINTENANCE 7.8 CHANGING THE PERISTALTIC PUMP TUBES The manufacturer recommends that the Peristaltic Pump tubes be replaced every six months. Said frequency may vary depending on the workload of the individual laboratory. The quality and reliability of these tubes is fundamental to a correct emptying of the cuvettes. 7.8.1 PROCEDURE FOR REPLACING THE PERISTALTIC PUMP TUBES 1. Open the panel located on the right side of the instrument pushing down the looking door(Fig. 1). 2. Unhook the hinged guide by pushing the pin toward to the left (Fig. 2). 3. Pull the tubes off their relative nipples and substitute with new ones (Fig. 2). 4. After having unlooked the hinged guide, disconnect the tubes from their nipples 5. Remount, repeating the above steps, 4 through 2, in reverse order. Turn clock wise the peristaltic pump rotor, in manual way, in order to be sure that tubes are right positioned inside the guide. Looking door Pin Tubes to be replaced Hinged guide lock Peristaltic Pump Rotor Fig. 1 Instrument right side Nipples Fig. 2 Peristaltic Pump Assembly 6. Turn on the Analyzer and wait until the instrument has reached its proper operating temperature. 7. Have the instrument carry out a ‘Wash cuvettes’ cycle and then a ‘WBL’ cycle. Check system efficiency by comparing the WBL values and the test results obtained against the laboratory’s quality control values. 8. Make sure that there is no leakage and then close back the panel. • Always use original replacement parts; never lubricate the peristaltic pump roller bearings. • After prolonged instrument inactivity: verify the efficiency of the peristaltic pump tubes. Please note: if the instrument is not able to completely empty the washing well, check the condition of the tube and make sure that there are not obstruction into the hydraulic lines. Ellipse Service Manual Rev. 03 01 June 2009 Page 11 Chapter 07 - MAINTENANCE 7.9 CHANGING THE PHOTOMETER LAMP The manufacturer suggests that the lamp be replaced after approximately 2000 hours of use. Figure 2 illustrates the photometer lamp, its support base and its power supply wires. There is a small hole on the lamp base useful for its mechanical alignment. The lamp is mounted on the photometer, which is located on the left-hand side of the reaction plate. 7.9.1 PROCEDURE FOR REPLACING THE PHOTOMETER LAMP 1) Turn off the Analyzer. 2) Remove the front panel unscrewing the four anchored screws (Fig. 2). 3) Disconnect the power supply wires from the Lamp Regulator Board by loosening the clamp screws on the M1 connector (Fig. 3). 4) Unscrew the lamp’s fastening screw and remove the lamp from its housing (Fig. 3) 5) Replace the old lamp with a new one making sure that the pin is correctly inserted in the alignment hole (Fig. 3). Remount the assembly by repeating the above steps 4 through 1 in inverse order. 6) Turn the Analyzer on and wait until the instrument has reached its proper operating temperature. 7) Have the instrument carry out a ‘WBL’ cycle. Check system efficiency by comparing the WBL values and the test results obtained against the laboratory’s quality control values. Ellipse Service Manual Rev. 03 01 June 2009 Page 12 Chapter 07 - MAINTENANCE Washing Station protective cover Fig.1 Diluter protective cover Instrument Top view Instrument front panel anchored screw Fig. 2 Ellipse Instrument front view Service Manual Rev. 03 01 June 2009 Page 13 Chapter 07 - MAINTENANCE Halogen Lamp Cd. 9-35-0016-00 Lamp's fastening screw M1 Connector Photometer Lamp Board Cd. 30-01576-00 Trimmer P1 Fig. 3 WARNING! DO NOT TOUCH THE GLASS PART OF THE LAMP WITH YOUR FINGERS! IF NECESSARY, USE A CLEAN CLOTH TO REMOVE DUST, OR ALCOHOL TO REMOVE MORE STUBBORN DIRT. 7.10 CHANGING THE REACTION CUVETTES Over time and through normal use the perfect transparency of the cuvettes diminishes. This less-thanperfect transparency has a negative impact on the quality of the optical readings. The manufacturer suggests that the cuvettes be replaced after three months of use. The cuvettes racks are located inside the reaction plate. 7.10.1 PROCEDURE FOR REPLACING THE REACTION CUVETTES 1. Turn on the Analyzer and wait until the instrument has reached its proper operating temperature. 2. Click on the change cuvettes rack button , located under the work program (Please see Chapter 03 – Description of Instrument Software). 3. A pull-down menu will appear. Click on the rack number 1 and then on OK, in order to have the selected rack accessible (Fig. 1) Ellipse Service Manual Rev. 03 01 June 2009 Page 14 Chapter 07 - MAINTENANCE Fig. 1 Reaction Rack cover 4. Remove the reaction rack cover from its housing (Fig. 1). 5. Turn the two locking pins making free the reaction rack. 6. Remove the reaction rack from its housing and replace it with a new one (Fig. 2). 7. Make sure to reinsert the new rack VERTICALLY. Turn the two locking pins in order to fix the rack. Be especially careful to not touch the external surface of the cuvettes dedicated to photometric reading. 8. Repeat the procedure from point 2 to 7 in order to replace the reaction racks from 2 to 6. 9. Have the instrument carry out a ‘WBL’ cycle. Check system efficiency by comparing the WBL values and the test results obtained against the laboratory’s quality control values. Reaction Rack Fig. 2 Ellipse Service Manual Rev. 03 01 June 2009 Page 15 Chapter 07 - MAINTENANCE locking pin Fig. 3 7.11 CHANGING THE TUBES KIT Over time and through normal use, the tubes become worn. The manufacturer suggests that the tubes be replaced at least once a year. These tubes are located inside the instrument to link the Washing Station, the Peristaltic Pump and the Washing Well to the Washing Pump Assembly. 7.11.1 PROCEDURE FOR REPLACING THE TUBES KIT 2. Turn off the Analyzer. 3. Unscrew the four screws to remove the panel located on the rear side of the instrument to access the washing pump assembly (Fig. 1 and Fig 2). 4. Replace the tubes following the indications provided in the hydraulic diagram (SI-00571-XX). 5. Turn on the Analyzer and wait until the instrument has reached its proper operating temperature. 6. Have the instrument carry out a ‘Wash cuvettes’ cycle and then a ‘WBL’ cycle. Make sure that there is no leakage. 7. Ellipse Check system efficiency by comparing the WBL values and the test results obtained against the laboratory’s quality control values. Service Manual Rev. 03 01 June 2009 Page 16 Chapter 07 - MAINTENANCE Fig. 1 Instrument rear view Fig.2 Pumps module 7.12 TABLE C – LIST OF THOSE MAINTENANCE PROCEDURES THAT CAN BE PERFORMED BY THE USER AND/OR BY THE MAINTENANCE TECHNICIAN PROCEDURE FREQUENCY (*) EFFECTUATED BY CLEANING THE SAMPLING PROBE DAILY OPERATOR CLEANING THE FOUR WASH STATION CANNULAS EVERY TWO WEEKS OPERATOR CLEANING THE WASH SOLUTION BOTTLES EVERY TWO WEEKS OPERATOR REPLACING REACTION CUVETTES EVERY THREE MONTHS OPERATOR CLEANING THE TIP ONCE A MONTH OPERATOR REPLACING THE TIP AS NEEDED OPERATOR REPLACING THE PERISTALTIC PUMP TUBES EVERY SIX MONTHS OPERATOR REPLACING THE PHOTOMETER BULB AFTER 2000 HOURS OF USE OPERATOR Ellipse Service Manual Rev. 03 01 June 2009 Page 17 Chapter 07 - MAINTENANCE PROCEDURE FREQUENCY (*) EFFECTUATED BY REPLACING THE SAMPLING PROBE AS NEEDED OPERATOR CLEANING THE HYDRAULIC CIRCUIT AS NEEDED OPERATOR REPLACING THE TUBES (Tubes Kit) ONCE A YEAR MAINTENANCE TECH REPLACING THE PRE-HEATER AND THE SENSOR LEVEL AS NEEDED MAINTENANCE TECH MECHANICAL ALIGNMENT OF THE SAMPLING ARM AS NEEDED MAINTENANCE TECH REPLACING OR ADJUSTING THE OPTIC SENSORS AS NEEDED MAINTENANCE TECH REPLACING THE BELT AS NEEDED MAINTENANCE TECH REPLACING A MOTOR AS NEEDED MAINTENANCE TECH ALIGNING THE WASH STATION/REACTION PLATE AS NEEDED MAINTENANCE TECH ALIGNING THE REACTION OR REAGENT OR SAMPLE PLATE AS NEEDED MAINTENANCE TECH REPLACING THE PHOTOMETER AS NEEDED MAINTENANCE TECH ADJUSTING THE PHOTOMETER AS NEEDED MAINTENANCE TECH REPLACING THE ELECTRONIC BOARDS/CARDS AND THE MECHANICAL MODULES AS NEEDED MAINTENANCE TECH Ellipse Service Manual Rev. 03 01 June 2009 Page 18 Chapter 07 - MAINTENANCE N. B.: the above-indicated frequency intervals may vary according to the individual instrument’s daily workload. 7.13 DECONTAMINATION PROCEDURE Before replacing any instrument parts, repairing any defective items or performing any instrument maintenance procedure(s), the operator or maintenance technician must carry out the belowdescribed decontamination procedure of the instrument part(s) involved in the operation(s). This procedure can be performed on: the entire Analyzer those part(s) of the instrument subject to possible contamination Material necessary - an ESOFENOL solution diluted to 6% (60 cc in one liter of distilled water). ESOFENOL is an antibacterial and antiviral substance. - Rubber gloves - Mask - Lab coat External surfaces and individual parts Spray the solution all over the instrument, paying particular attention to wetting: - the sampling arm - the reaction plate (including the cuvettes) - the reagent and standard and control plate - the sample plate (including the racks) - the instrument chassis Allow the solution to stand for approximately 30 minutes Wipe the solution off the instrument and the various components using a sponge dampened with distilled water Carry out a decontamination of the internal hydraulic circuit. Ellipse Service Manual Rev. 03 01 June 2009 Page 19 Chapter 07 - MAINTENANCE Hydraulic circuit (Entire Instrument) Fill a container with a 5% sodium hypochlorite solution. Disconnect the three silicon aspiration tubes from the nipples on the caps of their respective tanks. Immerge all three tubes into the container filled with the 5% sodium hypochlorite solution and have the instrument carry out two ‘Wash cuvettes’ cycles and then two ‘WBL’ cycles. Remove the three tubes from the 5% sodium hypochlorite solution and immerge them in another container containing the instrument Rinse Solution. Have the instrument carry out one ‘Wash cuvettes’ cycle and then one ‘WBL’ cycle. Reconnect the three aspiration tubes to the nipples on their respective tanks. Ellipse Service Manual Rev. 03 01 June 2009 Page 20 Chapter 08 – Host Communication CHAPTER 08 – HOST COMMUNICATION – INDEX 8.1 COMMUNICATION WITH THE HOST COMPUTER .............................................................. 2 8.1.1 COMMUNICATION PARAMETERS ...................................................................................... 2 8.2 PROTOCOL SPECIFICATIONS.................................................................................................. 3 8.3 Header Record (H) ........................................................................................................................ 4 8.4 Patient Record (P) ......................................................................................................................... 6 8.5 Test Order Record (O) .................................................................................................................. 8 8.6 Results Record (R) ...................................................................................................................... 11 8.7 Message Terminator Record (L) ................................................................................................. 13 8.8 Request Information Record (Q): ............................................................................................... 13 Ellipse Service Manual Rev. 04 01 October 2009 Page 1 Chapter 08 – Host Communication 8.1 COMMUNICATION WITH THE HOST COMPUTER The Ellipse can be connected to a host computer for the purpose of facilitating results print-out and patient management. In order to enable communication between the Ellipse and the host computer, select the Host Link field under Options in the Parameters menu. To activate communication between the Ellipse and the host computer, select Host–Tx (please see the software description in Chapter 03 of the User’s Manual) 8.1.1 COMMUNICATION PARAMETERS Communication parameters can be modified in the Parameters.ini file into the Analyzer 140 folder for every need. The Ellipse PC is linked to the HOST computer using RS-232C serial connector having as default communication parameters the following specifications: o o o o o o Transmission method Baud Rate Data bits: Parity Stop bit Connector : : : : : : Asynchronous, half duplex 9600 Bit/sec. 8 None 1 9 pin type D (male output from the Ellipse computer) 1-CF(DCD) (DSR)CC-6 2-BB(RD) (RTS)CA-7 3-BA(TD) (CTS)CB-8 4-CD(DTR) (RI)CE-9 5-AB(GND) Serial connector Ellipse Service Manual Rev. 04 01 October 2009 Page 2 Chapter 08 – Host Communication ELLIPSE PC DB 9 F HOST PC DB 25 F HOST PC DB 9 F 3 Txd Rxd 3 2 2 Rxd Txd 2 3 5 SG SG 7 5 8.2 PROTOCOL SPECIFICATIONS This part of Help (Protocol Specifications) contains information for the laboratory computer and analyzer. This exchange of data follows specific ASTM protocols: E 1381-95 Standard Specification for Low-Level Protocol to Transfer Messages between Clinical Laboratory Instruments and Computer Systems; E 1394-97 Standard Specification for Transferring Information between Clinical Instruments and Computer Systems. ASTM uses a number of different terms to indicate the way it groups data. • Field: an individual piece of data often referred to as a data field or a data element. • Record: a number of logically related data fields grouped together to form one part of a complete message. • Repeat field: a data field of the same type as the one immediately preceding it. A delimiter separates one instance of a repeat field from the next. • Component field: part of data field that might contain more than one piece of data. The default communication configuration for the Analyzer is the following: "9600,N, 8,1". ASTM uses record types that are common and familiar to all laboratory personnel. It uses the following record types: • Header Record (H): contains identifying information about the sending station, conventions that the device uses for field recognition, and the date and time of send station transmission. Ellipse Service Manual Rev. 04 01 October 2009 Page 3 Chapter 08 – Host Communication • Patient Record (P): contains patient information and identification number. • Test Order Record (O): contains information about the assay or requests themselves and includes other data. Result Record (R): contains information about the outcome of individual tests for an • individual patient and follows a sample program record. The results contain the actual measurements derived from the test and a comparison of the individual result to certain ranges specified as norms for the laboratory. Message Terminator Record (L): although the ASTM protocol supports three additional • record types - a Request for Information Record, a Scientific Record and a Manufacturer's Information Record - the Analyzer is not implementing these in the first release and will ignore them. Request Information Record (Q): is used by either clinical instruments or computer • systems for a remote request for information from its reciprocal system. The instrument does not send or accept comment records. 8.3 HEADER RECORD (H) Field Field Title Down Load Up Load Max Len Description and Valid Values 1 Record Type ID R A 1 This is a required field that contains an “H” identifying it as a header record. 2 Delimiters I A 4 The Analyzer System uses only the four default values shown here. Delimiters may not be duplicated. The field delimiter follows the escape character to separate the delimiter specification from a subsequent field in the header record. Using default values, the first six characters of the header record will appear using the following characters: H I\^&I Ellipse Service Manual Field Delimiter I Repeat Delimiter \ Component Delimiter ^ Rev. 04 01 October 2009 Page 4 Chapter 08 – Host Communication Escape Delimiter 3 4 5 6 7 8 9 10 11 12 13 14 Legend: Ellipse Message Control ID Access Password Sender Name or ID Sender Street Address Reserved Field I N I N I A I N I N Sender Tel. Number Characteristics of Sender Receiver ID I N I N I N Comments or Special Instructions Processing I N I N ASTM Version No. Date and Time I N I A R O I A Required Optional Ignored Always Service Manual 10 & ‘SHAnalyzer’: This is the name of the device that is sending the data. 14 D U N S Date and Time of transmission: formatted as YYYYMMDDHHMMSS. For example: 3:35 PM on March 1, 1995 would be represented using the following characters: 19950301153500. Down Load Up Load Never Sometimes Rev. 04 01 October 2009 Page 5 Chapter 08 – Host Communication Example Header Record Layouts (H) Download Host H I \ ^ I I I HOST I I I I I I I I I19950301153500Upload H | \ ^ & | | | SHAnalyzer | | | | | | | | | 19950301154000 Analyzer System 8.4 PATIENT RECORD (P) Down Load Up Load Max Len Record Type ID R A 1 2 Sequence Number R A 3 3 Practice Assigned Patient ID R A 15 4 I N 5 Laboratory Assigned Patient ID Patient ID No. 3 I N 6 Patient Name O S 7 I N 8 Mother's Maiden Name Birth Date O 9 Patient Sex 10 Patient Race/Ethnic Origin Fi el d Field Title 1 Ellipse Description and Valid Values This is a required field that contains a “P” identifying it as a patient record. This field starts with a “1” for the patient and is incremented by 1 for each additional patient within the transmission. This field can be assigned by the instrument with no corresponding download. 36 This field has two components: • Last Name (up to 20 characters) • First Name (up to 15 characters). S 8 R A 1 Formatted as YYYYMMDD: For example, a birth date of December 1, 1980 would be represented as: 19801201 The valid values are: • M for Male • F for Female I N Service Manual The Analyzer System will ignore this field at launch. Rev. 04 01 October 2009 Page 6 Chapter 08 – Host Communication 11 Patient Address O S 12 Reserved Field I N 13 I N I N 15 Patient Tel Number Attending Physician ID Special Field 1 I N 16 Special Field 2 I N 17 Patient Height I N 18 Patient Weight I N 19 Patient Known or Suspected Diagnosis I N 20 Patient Active Medications Patient's Diet Practice Field No. 1 Practice Field No. 2 Admission Date and Discharge Date (if desired) I N I I N N I N O S 14 21 22 23 24 25 Admission Status I N 26 Location O S 27 Nature of Alternative Diagnostic Code and classifiers Alternative Diagnostic Code and classification I N I N 29 Patient Religion I N 30 Marital Status I N 28 Ellipse Service Manual 60 For Analyzer, this is a four- component field: • Address (25 characters) • City (25 characters) • State (2 characters e.g.: NY, IT) • Zip (5 characters) 8 Admission date only. Formatted as YYYYMMDD. 20 Rev. 04 01 October 2009 Page 7 Chapter 08 – Host Communication 31 Isolation Status I N 32 Language I N 33 Hospital Service I N 34 Hospital Institution Dosage Category I N I N 35 Legend: R O I A Required Optional Ignored Always D U N S Down Load Up Load Never Sometimes Example Patient Record (P) Download Host P | 1 | B108K | | | MW5910^Smith | | 19861002 | M | | Park Avenue^New York^NY^10002 | | | | | | | | | | | | | 20020923 | | Hematology | | | | | | | | | | Analyzer System P | 1 | B108K | | | MW5910^Smith | | 19861002 | M | | Park Avenue^New York^NY^10002 | | | | | | | | | | | | | 20020923 | | Hematology | | | | | | | | | 8.5 TEST ORDER RECORD (O) Field Field Title Down Load Up Load Max Length Description and Valid Values 1 Record Type ID R A 1 This is required field that contains an “O” identifying it as an order 2 Sequence Number R A 3 This field starts with “1” for the first Test Order Record and is incremented by 1 for each additional Test Order Record within the record. This will be reset to “1” whenever another patient record is transmitted. 3 Ellipse Specimen ID R Service Manual A 15 Rev. 04 Although the operator can manually edit this field at any time, the value of this field is usually assigned by the laboratory computer before down loading. The Analyzer uses and reports its results based on the assigned specimen ID. 01 October 2009 Page 8 Chapter 08 – Host Communication 4 Instrument Specimen ID I N 5 Universal Test ID I N I N I N R A 6 Priority I N 7 Request Ordered Date/Time Specimen Collected Date/Time Collection End Time I N I N I N 10 Collection Volume/Units I N 11 Collector ID I N 12 Action Code I N 13 Danger Code I N 14 Relevant Clinical Info. I N 15 Date/Time Specimen Received I N 8 9 Ellipse Service Manual 9 Rev. 04 This is a four-component field: • Universal Test ID Code (not used) • Universal Test ID Name (not used) • Universal Test ID Type (not used) • Manufacturer's or local code (6 characters): This is the code defined in the Analyzer. 01 October 2009 Page 9 Chapter 08 – Host Communication 16 Specimen Type R A 17 Ordering Physician I N 18 Physician Tel. Number I N 19 User Field No. 1 I N 20 User Field No. 2 I N 21 Lab Field No. 1 I N 22 Lab Field No. 2 I N 23 Date /Time Result Reported Last or Modified Instrument Charge I N I N 25 Instrument Section ID I N 26 Record Type I A 27 Reserved Field I N 28 Location or Ward of Specimen Collection Nosocomial Infection Flag I N I N 30 Specimen Service I N 31 Specimen Institution I N 24 29 Ellipse Service Manual 1 This is a numeric field indicating the type of specimen: The Imm. System uses the following ASCII characters: 0= Serum 1= Urine 1 The field indicates the direction of the transmission: O - Down Loading F - Up Loading Rev. 04 01 October 2009 Page 10 Chapter 08 – Host Communication Legend : R O I A Required Optional Ignored Always D U N S Down Load Up Load Never Sometimes Example Test Order Record Layouts (O) Download Host O | 1 | AR102 | | ^^^GLU | | | | | | | | | | | 0 | | | | | | | | | | O | | | | | Analyzer System O | 1 | AR102 | | ^^^GLU | | | | | | | | | | | 0 | | | | | | | | | | F | | | | | 8.6 RESULTS RECORD (R) Field Field Title Down Load Up Load Max Len 1 Record Type ID A 1 This is a required field that contains an “R” identifying it as a Results Record. 2 Sequence Number A 3 3 Universal Test ID This field starts with “1” for the first result and is incremented by 1 for each additional result within the record. This will be reset to “1” when the results from another test order record are transmitted to the laboratory computer. This is a four-component field: • Universal Test ID Code (not used) • Universal Test ID Name (not used) • Universal Test ID Type (not used) • Local Manufacturer's or local code (6 characters) this is the code defined in the Analyzer. I N I N I N R 4 Ellipse Data or Measurement value Service Manual 9 A A 10 Rev. 04 ‘Data’ is a 10-character, floating point field that includes the decimal point. The number of precision point digits will vary according to the test and is configurable on the Analyzer. 01 October 2009 Page 11 Chapter 08 – Host Communication 5 Units of Measure A 6 This is a field for up to 6 characters that the operator defines for analytic measurement. 6 Reference Ranges A 21 This field has two components; one giving the lower limit and the other the upper limit of the range. The format for this field is N^N. 7 Result Abnormal Flags A 2 This field indicates the normal status of the result. The following codes are valid values: L - Below Low normal H - Above High normal LL - Below Panic normal HH - Above Panic normal < - Below absolute low (under linearity) > - Above absolute high (over linearity) N - Normal A - Abnormal E – Edited 8 9 Nature of Abnormality Testing Result Status N A 1 The Imm. System currently implements only two valid values: F - final results; V - operator verified/approved result. 10 Date of Change in Instrument Normative Values or Units Operator ID N 12 Date/Time Test Started N 13 Date/Time Test Completed A 11 Ellipse Service Manual N 14 Rev. 04 Date and Time of test completion: formatted as YYYYMMDDHHMMSS. 01 October 2009 Page 12 Chapter 08 – Host Communication 14 Instrument ID N R O I A Legend: Required Optional Ignored Always D U N S Down Load Up Load Never Sometimes Example Result Record Layouts (R) Upload R | 1 | ^^^GLU | 70.97 | UL | 70^105 | N | | F | | | | 20020923114302 | Analyzer System 8.7 MESSAGE TERMINATOR RECORD (L) Field Field Title Down Load Up Load Max Len 1 Record Type ID R A 1 2 Sequence Number Termination Code R A 1 R A 1 3 Description and Valid Values This is a required field that contains an “L” identifying it as an Message Terminator Record. For a message terminator, this message should always be “1”. This indicates the cause of termination. The following codes are valid values for the Analyzer : Null or N-normal termination R O I A Legend: Required Optional Ignored Always D Down Load U Up Load N Never S Sometimes Example Message Terminator Record Layout (L) Host LI1IN Analyzer System LI1IN 8.8 REQUEST INFORMATION RECORD (Q): Field 1 Ellipse Field Title Record Type ID Service Manual Down Load Up Load Max Len Description and Valid Values A 1 This is a required field that contains a “Q” identifying it as a request. Rev. 04 01 October 2009 Page 13 Chapter 08 – Host Communication 2 Sequence Number A 1 It is always “1”. 3 Starting Range ID Number A 31 This field can either be: "ALL" to mean all demographics and tests being ordered should be sent to the instrument at this time, or can have two components: • Computer system patient ID No. (up to 15 characters); • Computer system specimen ID No. (up to 15 characters). 4 5 6 Ending Range ID Number Universal Test ID N N N 11 Nature of Request Time Limits Beginning Request Results Date and Time Ending Request Results Date and Time Requesting Physician Name Requesting Physician Telephone Number User Field No. 1 12 User Field No. 2 N 13 Request Information Status Codes A 7 8 9 10 N N N N N 1 It is always "O" (requesting test orders and demographics only). Example Request Information Record Layouts (Q) Download To Analyzer System Ellipse H | \^& | | | SHAnalyzer | | | | | | | | | 20020927100402 Q | 1 | ALL | | | | | | | | | | O L|1|N Service Manual Rev. 04 01 October 2009 Page 14 Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING CHAPTER 09 - ERROR SIGNALING AND TROUBLESHOOTING INDEX 9.1 ERROR SEGNALING ........................................................................................................2 9.1.1 SYSTEM ERRORS .............................................................................................................2 9.1.2 RESULT FLAGS.................................................................................................................3 9.1.3 DESCRIPTION OF RESULT FLAGS................................................................................4 9.2 TROUBLESHOOTING GUIDE .........................................................................................7 Ellipse Service Manual Rev. 03 01 June 2009 Page 1 Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING 9.1 ERROR SIGNALING This chapter is dedicated to a description of the error signalling which may occur during the programming or carrying out of the various operations. Error signals can be divided into the following two groups: ➢ System Errors ➢ Result Flags 9.1.1 SYSTEM ERRORS Whenever a system error is detected, it is signalled via the activation of the below-listed warning lights/buttons: ➢ Warning Light: a triangle-shape located in the upper, right-hand portion of the System Monitor screen. It lights up yellow when activated. If this yellow triangle does light up, the operator need simply click on it to access the relative visual text warning message indicating the cause of the signalled anomaly (said window will open in that area dedicated to viewing data). Following is a list of possible visual text “Warning!” messages: • Liquid Alarm I (Rinse Solution) • Liquid Alarm II (Distilled Water) • Liquid Alarm III (Cleaning Solution) • Liquid Alarm IV (Optional Solution) • Temperature Out-of-Range • Host Serial Port cannot be opened ➢ Fatal Error: an “X”-shape located in the upper, right-hand portion of the System Monitor mask. It lights up red when activated. If this red “X” lights up, the operator need simply click on it to access the relative visual text message explaining the cause of the Fatal Error (said window will open in that area of the mask dedicated to viewing data). Following is a list of possible Fatal Error visual text messages: • Horizontal Arm Error • Sample Tray Error • Filter Error Ellipse Service Manual Rev. 03 01 June 2009 Page 2 Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING • ADC Error – Sample Channel • ADC Error – Reference Channel • Reaction Tray Position Error • Plate Error (Home sensor not found) • Vertical Arm error • Reagent Tray Error • Wash System Error • Diluter Error • Interlock Open while Instrument is running • Macro send Error • Overflow Rx Buffer Micro • Time out Error (Generic error) • Invalid command error • Invalid parameter error • Check sum error 9.1.2 RESULT FLAGS Result Flags are categorized under the following group headings, Each group is identified by a symbol, as listed here-below: X PHYSICAL ERRORS R CONCENTRATION ERRORS C CALIBRATION ERRORS A OPTIC DENSITY ERRORS E RESULTS EDITED MANUALLY ? PROGRAMMING ERRORS Every Result Flag, signalling an error, is accompanied by a symbol representing the group it is part of. The operator need only click on the small red square ( ) next to the Result Flag symbol to access the visual text message explaining the cause(s) of the signalled error. In the central column of the following tables, the user will find those symbols which signal the type of error encounter, as used in the print-out of the final report. These symbols can be modified by the operator, in the Parameters section, under print options. Ellipse Service Manual Rev. 03 01 June 2009 Page 3 Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING WARNING! The use of the Error Symbols in the print-out of the results (inclusion and/or exclusion) IS UNDER THE DIRECT AND SOLE RESPONSIBILITY OF THE USER 9.1.3 DESCRIPTION OF RESULT FLAGS ¾ X Physical Errors Temperature Error: T Reaction temperature (Reaction Plate) is out-of-range. Either no sample or sample serum quantity below minimum or above maximum level for the declared container. No Sample S No Reagent: R Either no reagent or reagent level below minimum or above maximum level for the declared container. No Rack: * No Rack present during sampling. ¾ R Concentration Errors Very Low and Very High: L-H Flags determined by test results out-of-range as setup in the Methods. Low Alert and High Alert: A Flags determined by test results out-of-range as setup in the Methods. Low and High Linearity Limit: G Flags determined by test results out-of-range as Calculation Error: C Concentration calculation error due to foreseeable causes (Asymptote). Ellipse Service Manual Rev. 03 01 June 2009 Page 4 Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING ¾ C Calibration Errors RBL missing: * No Reagent Blank Level. Calibration missing: * No Standard or no Calibration curve. STD Replicate insufficient: * Insufficient number of valid Standard Replicates. STD Replicate outside CV%: * Coefficient of Variation Percentage in the Standard Replicates over the set value. Invalid Calibration: * Calibration curve not valid – either because it is not monotonic or because the Fit is above the set value. ¾ A Optic Density Errors Inversion: I Reaction direction not in line with that set-up. End Point Limit: P Values over the limits setup in the Methods Parameters Depletion Limit: D Values over the limits setup in the Methods Parameters. First Limit: * Values over the limits setup in the Methods Parameters. FIT: F Values over the limits setup in the Methods Parameters. RBL out-of-range: * Reagent Blank Levels outside the range Setup. Sample outside Standard: # Sample absorbance outside the calibration curve. Ellipse Service Manual Rev. 03 01 June 2009 Page 5 Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING ¾ E Results Edited Manually Results Edited: ¾ ? E This symbol automatically appears whenever the operator has manually modified the obtained results. This operation annuls all the symbols indicating errors which, in this case, will not be viewed. Programming Errors All those software errors which are deemed unforeseeable are indicated using this symbol. ¾ Result Asterisks In the event that the following situations and/or error conditions occur, the system will notify the user via a visualization of two asterisks (**) in the Results field: PHYSICAL ERRORS 1 – very little or no sample 2 – very little or no reagent 3 – no rack CONCENTRATION ERRORS 4 – calculation error 5 – signalling depletion limit PROGRAMMING ERROR 6 – all cases In the following situations and/or error conditions, the results field will contain a “0”: 1 – if the result is less than zero 2 – if the result is equal to zero OPTIC DENSITY ERRORS 3 – Flag signalling inversion Ellipse Service Manual Rev. 03 01 June 2009 Page 6 Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING 9.2 TROUBLESHOOTING GUIDE SOLUTION PROBLEM ➢ POSSIBLE CAUSE Repeatability of results insufficient ➢ Sample Probe dirty Clean the Sample Probe as described in Chapter 07 Maintenance ➢ Hydraulic leak and/or air bubble in the hydraulic circuit (sampling) Check Sample Probe fit Check the fit of the hydraulic tubes and their connection: if necessary, substitute the tubes and/or adjust their connections/fittings ➢ Wash solution is contaminated. If the wash solution contains contaminating particles (e.g.: mold, dust, lint), these micro-particles can cause errors during the WBL running. Change the wash solution Clean the Wash Solution Bottle(s) and carry out the Hydraulic Circuit wash procedure as described in Chapter 07 – Maintenance ➢ Deterioration of the Reagent(s) Substitute the bad reagent(s) ➢ Reaction cuvettes not dried correctly after being washed Check the Tip used to dry the cuvettes after washing to make sure it is in good working condition. If necessary, clean the Tip or change it, following the procedure described in Chapter 07 – Maintenance ➢ Light bulb not stable Light bulb nearing the end of its 2000-hour life cycle duration, or premature deterioration. In both cases, change the light bulb following the substitution procedure described in Chapter 07 – Maintenance. Insufficient volume/quantity of the ➢ This type of problem can present itself either when the various Rinse, Cleaning and instrument is first turned on due to a lack of liquid in one or Distilled Water solutions. more bottles, or during test running whenever the involved liquid has been finished Refill the required bottle. If the instrument is running tests, it will automatically Pause. Wait until sampling is suspended, then refill the involved bottle with the required liquid (Rinse/Cleaning/ Distilled Water). Press Start to continue running the rest of the programmed tests. This operation will not adversely affect the already run tests. Ellipse Service Manual Rev. 03 01 June 2009 Page 7
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : No Create Date : 2009:06:12 09:14:41Z Modify Date : 2010:01:19 09:15:48+01:00 Page Count : 229 Creation Date : 2009:06:12 09:14:41Z Mod Date : 2010:01:19 09:15:48+01:00 Producer : Acrobat Distiller 5.0.5 (Windows) Author : gda3 Metadata Date : 2010:01:19 09:15:48+01:00 Creator : gda3 Title : Microsoft Word - Cover .doc Has XFA : NoEXIF Metadata provided by EXIF.tools