Cover Ellipse Service Manual Rev.04

User Manual:

Open the PDF directly: View PDF PDF.
Page Count: 229

DownloadCover  Ellipse Service Manual Rev.04
Open PDF In BrowserView PDF
ELLIPSE
Service Manual

Via E. Barsanti 17/A
00012 Guidonia – ROME (ITALY)
www.ams-analyzers.com

CONFIGURATION SHEET

CHAPTER

DESCRIPTION

REVISION

INDEX

TABLE OF CONTENTS

03

01

INTRODUCTION

03

02

SYSTEM DESCRIPTION

03

03

INSTALLATION

03

04

ELECTRICAL SCHEMES AND DRAWINGS

03

05

DIAGNOSTIC PROGRAM

03

06

SETTINGS AND ADJUSTMENTS

03

07

MAINTENANCE

03

08

HOST COMMUNICATION

04

09

ERROR SIGNALING AND TROUBLESHOOTING

03

Ellipse

Service Manual

Rev.04

01 October 2009

Page 1

INDEX

TABLE OF CONTENTS

Chapter 01

INTRODUCTION

Chapter 02

SYSTEM DESCRIPTION

Chapter 03

INSTALLATION

Chapter 04

ELECTRICAL SCHEMES AND DRAWINGS

Chapter 05

DIAGNOSTIC PROGRAM

Chapter 06

SETTINGS AND ADJUSTMENTS

Chapter 07

MAINTENANCE

Chapter 08

HOST COMMUNICATION

Chapter 09

ERROR SIGNALING AND TROUBLESHOOTING

Ellipse

Service Manual

Rev.03

01 June 2009

Page 1

Chapter 01 – INTRODUCTION

CHAPTER 01
- INTRODUCTION INDEX

1

INTRODUCTION ................................................................................................................. 1

1.1

THE AIM OF THE TECHNICAL MANUAL ………………………………………...………..…..2

1.1.2

MANUAL OUTLINE……………………………..……………………..……………..………………....2

1.2

SYSTEM INTRODUCTION ................................................................................................ 3

1.3

PRECAUTIONARY MEASURES ....................................................................................... 3

1.3.1

CHEMICAL RISKS .............................................................................................................. 5

1.3.2

ELECTRICAL RISKS........................................................................................................... 5

1.3.3

MECHANICAL RISKS......................................................................................................... 4

1.4

GUARANTEE ...................................................................................................................... 5

1.5

TECHNICAL OPERATING FEATURES………………………………………….…. ..... 6

1.5.1

COMPUTER & SOFTWARE FEATURES ........................................................................ .9

1.5.2

OPTIONS............................................................................................................................... 8

1.5.3

DIMENSIOS, WEIGHT & ENVIRONMENT.................................................................... 10

1.5.4

INSTALLATION REQUIREMENTS................................................................................ 10

Ellipse

Service Manual

Rev. 03

01June 2009

Page 1

Chapter 01 – INTRODUCTION

1

INTRODUCTION

1.1 THE AIM OF THE TECHNICAL MANUAL
This manual has been written in order to supply the technical staff, the persons who are responsible
for the maintenance and for resolving instrument failures, a complete and detailed guide of the
Ellipse analyzer, in accordance with the standard UNI EN591 (which requires a manual to be
supplied with vitro diagnostic instruments for professional use).

1.1.2 MANUAL OUTLINE
The technical manual is composed of 10 chapters that not only describe the operative technical
characteristics of the system, but also the reparation and maintenance procedures to use for the
modules that they are composed of.

.

Chapter 1

INTRODUCTION

The manual structure is described, and recommendations are given regarding the general
use of the analyzer. The technical characteristics of the operative system are given in this
chapter.

.

Chapter 2

DESCRIPTION OF THE SYSTEM

Describes the system, in particular, the analytical cycle of the single modules and
electronic boards.

.

Chapter 3

INSTALLATION

Describes the unpacking procedures and the required characteristics for the place of
installation.

.

Chapter 4

ELECTRONICS SCHEMES AND DRAWINGS

Gives the electronics schemes and assembly drawings of the electronics board and the
assembly drawings of the modules that the system is composed of.

Ellipse

Service Manual

Rev. 03

01June 2009

Page 2

Chapter 01 – INTRODUCTION

.

Chapter 5

DIAGNOSTIC PROGRAM

Describes the folders in the Diagnostic Program that are responsible for the checks and the
calibration of the modules in the system.

.

Chapter 6

SETTINGS AND ADJUSTMENTS

Gives the procedures for replacement, adjustment and check of the system modules.

.

Chapter 7

MANTENANCE

Describes the routine maintenance that needs to be done on a regular basis in order to
guarantee the correct functioning of the analyzer.

.

Chapter 8

COMUNICATION WITH HOST

Describes the communication protocol between the analyzer and the Host computer.
•

Chapter 9

ERROR CODES AND GUIDE TO RESOLVE ANOMALIES

Lists the system’s error codes, describes the errors in the results and gives a guide for how to
implement corrective actions.

The technical staff responsible for resolving the structural anomalies is highly suggested to
thoroughly examine the contents of this manual before operating the system.

Ellipse

Service Manual

Rev. 03

01June 2009

Page 3

Chapter 01 – INTRODUCTION

1.2

SYSTEM INTRODUCTION

The “Ellipse” system is a continuous-loading, random access, bench top instrument for performing
chemical and Immunoturbidimetric clinical analysis. It is totally automatic

and computer

controlled.

Via E. Barsanti 17/A
00012 Guidonia – ROME (ITALY)

Ellipse

Service Manual

Rev. 03

01June 2009

Page 4

Chapter 01 – INTRODUCTION

1.3

PRECAUTIONARY MEASURES

1.3.1

CHEMICAL RISKS

The individual operator is responsible for assuring that all possible precautionary measures are
taken against eventual risks associated with the use of the “Ellipse” instrument in clinical laboratory
settings. The manufacturer will provide the reagents kit and specific written information on the use
of each of the reagents.
It is important that the samples be well coagulated and then carefully centrifuged.
Samples which contain fibrinogen clots can obstruct the probe and lead to inexact sampling.
If blood samples containing gel are used, it is suggested that the manufacturer’s recommendations
be followed.
Immediately clean and remove any accidental leakage of reagent or other liquid.

1.3.2

ELECTRICAL RISKS

As with any electrical device, the risk of electric shock exists.
Is therefore necessary to take every precautionary measure possible when working with this, or any
other, electrical instrument to avoid contact with power supply wires, electrical components or
electronic boards.

1.3.3 MECHANICAL RISKS
Several precautionary measures should be taken when operating the analyzer:
avoid wearing very loose clothing or jewelry that could become tangled in the instrument’s moving
parts (e.g. the sample probe); whenever possible, operate the instrument with the main cover panel
lowered.
WARNING: Never attempt to service or substitute any part(s) of the analyzer when the
instrument is turned on.
Any and all technical repairs or servicing must be performed by specialized personnel only.

Ellipse

Service Manual

Rev. 03

01June 2009

Page 5

Chapter 01 – INTRODUCTION

1.4

WARRANTY

AMS guarantees the replacement of all defective components and/or materials for a period of time
not above of 14 months starting from the date of invoicing. Saying guarantee, as well as Technical
Assistance, generally is intended furnished as net ex factory Rome.
This guarantee does not include consumable and instrument parts in contact with liquids. All
components not included in the guarantee are reported in the following table.
Besides guarantee does not cover damage caused by :
- improper use of the ELLIPSE instrument (or however not according to the Producer or
Seller instructions)
- bad transport
-

insufficient (or missing) preventive maintenance by the User

In particular any damages due to the transport must be immediately reported to the carrier when he
delivers.

Ellipse

Service Manual

Rev. 03

01June 2009

Page 6

Chapter 01 – INTRODUCTION

CONSUMABLES AND ACCESSORIES PARTS LIST OUT OF GUARANTEE
Description

Type

Quantity

Reagents containers
Reagents containers

35 ml
6 ml

Samples cups

0.8 ml

12 pieces C101-00190-00
12 "
C101-00191-00
1000 "
AS-65-0002

Short samples cups
Adapter for short samples cups
Reaction sectors
Washing solution bottle
Bottle level sensor
Tubing Kit for peristaltic pump
Tubing Kit – complete
Cleaning solution
Rinse solution
Sampling probe (internal needle)
Complete Sampling probe
Drying Pad
Halogen Lamp (6 V - 10 W)
Interferential filters Kit
Fuse 6,3 A-T
Inlet/outlet fitting for Rinse & Clean conts
Cuvettes protection cover
Reagent protection cover
Reagent plate
Samples rack
Washing station, first or second cannulas (A)
Washing station, third cannula (B)

1000 "
1
"
6
"
1
"
1
"
2
"
1
"
2
"
1
"
1
"
1
"
1
"
1
"
1
"
10 "
1
"
1
"
1
"
1
"
1
"
1
1
1

AS-65-0100
9-01-0609-00
C101-00217-00
9-35-0041-00
9-05-0078-00
65-01835-00
65-01836-00
ASRN0020
ASRN0021
05-00707-00
10-00703-00
10-01920-00
9-35-0016-00
9-65-0029-00
C130-01238-08
01-01224-00
05-01249-00
10-00584-00
10-00585-00
05-01829-00
05-01633-00
05-01633-01

Washing station, fifth cannula without pad (D)

1

05-01919-00

Diluter Micro-Pump

1

"

05-01710-40

Air Micro-Pump (µP 6)
Micro-Pump (µP 2 ÷5)
Predilution rack
Solenoid Valve –2 way
Solenoid Valve –3 way

1
1
1
1
1

"
"
"
"
"

05-01711-20
05-01826-16
05-01735-00
9-35-0035-00
9-35-0036-00

Washing station, fourth cannula (C)

Ellipse

Service Manual

Rev. 03

1 ml

Code

2 lt

250 ml
50 ml

5x20

05-01638-00

01June 2009

Page 7

Chapter 01 – INTRODUCTION

1.5

TECHNICAL OPERATING FEATURES

DESCRIPTION

♦ Fully automatic, random access, continuous loading,
benchtop analyzer for clinical chemistry and
immunoturbidimetric assays;

ASSAY TYPE

♦ End Point, Initial Rate, Kinetic, Bichromatic, Differential;

TEST ENTRY MODE

♦ Selective, Batch, Profiles, STAT

THROUGHPUT

♦ 138 tests per hour

WORKING TEMPERATURE

♦ 37° C
♦ 24 removable containers by 40 ml and 6 ml;
♦ 12 positions for Controls and Standards;

ON LINE REAGENTS

SAMPLE CONTAINERS

SAMPLE LOADING
MINIMUM REACTION VOLUME
MAXIMUM REACTION VOLUME

♦ Positive identification by a Bar Code Reader
♦ Primary tubes (diameter from 8 to 16 mm; height up to
100 mm) cups from 1 to 4 ml;
♦ 4 racks, each having 8 positions for continuous sample
loading
♦ Positive identification by a Bar Code Reader
♦ 220 µl
♦ 550 µl
♦ A single mechanical arm provides all the sampling
operations and is equipped with:

SAMPLING ARM

•

Capacitive liquid level sensing

•

Reagent pre-warming at 37° C

•

Automatic probe washing

♦ Integrated syringe-free module having the following
specifications:

DILUTER

Sample volume:

2 µl ÷ 99 µl (1 µl incr.)

•

Reagent 1 volume: 3 µl ÷ 500 µl (1 µl incr.)

•

Reagent 2 volume: 3 µl ÷ 330 µl (1 µl incr.)

•

Reagent 3 volume: 3 µl ÷ 330 µl (1 µl incr.)

♦ CV < 1 % at 2 µl

PRECISION
READING SYSTEM

Ellipse

•

Service Manual

♦ Direct reading

Rev. 03

01June 2009

Page 8

Chapter 01 – INTRODUCTION

♦ Photometer: double beam, interferential filters
♦ Wavelength: 8 narrow band interferential filters
from 340 nm to 620 nm , plus one available
optional filter position

OPTIC SYSTEM

♦ Light source: 6V/10 W halogen bulb
♦ Linearity range: up to 3,500 Abs
♦ Resolution: 0.0005 Abs
OPTICAL PATH

♦ 6 mm.

WASHING STATION

♦ Composed of five probes that empty, wash and
dry the reaction cuvettes.
♦ 6 singular replaceable racks with 20 cuvettes
each

REACTION PLATE

♦ Cuvettes Q.C. continuously computer controlled
♦ Incubation temperature: 37°C

1.5.1. COMPUTER & SOFTWARE FEATURE
MINIMUM REQUIREMENTS
TYPE
♦ IBM Compatible
♦ Pentium IV 500 MHz, 512 Kb Cache

CPU

♦ RAM 256 Mb
♦ Hard Disk 20 Gb

MEMORY

♦ Floppy Disk 3 1/2” 1.44 Mb
♦ Colour SVGA 15’’ low radiation
MONITOR

Resolution 800 x 600 pixels;
max number of colors 65536 (16 bit)

PRINTER

♦ 80 Columns impact graphic (EPSON LX 300)

INTERFACE

♦ One Bi-directional RS 232C serial ports and one
parallel (one second serial port for the Host link)
♦ Multitasking WINDOWS XP Home edition

SOFTWARE
AVAILABLE LANGUAGES

♦ Disable all the energy saving options
♦ Disable the screen saver
♦ Select English “USA” as language, dot as decimal
symbol and date and time in Regional setting

SETTINGS

Ellipse

♦ Italian, English, Chinese, Czech. Software to be
released soon in these languages: Russian, Portuguese,
French, Polish. Upon request it is possible to release
the software in other languages.

Service Manual

Rev. 03

01June 2009

Page 9

Chapter 01 – INTRODUCTION

NOTE: Even though the computers demonstrate the same technical and operative characteristics,
some of these could have different hardware installed.
This could cause problems for the Ellipse software when running tests (A message appears
indicating “Random” error or blocks the program).
Therefore, if the PC is bought separately/locally, it is highly recommended to test the system at
your offices before preceding with the installation at the final client’s.
Consequently, AMS denies any responsibility for software problems that are due to buying the
computer separately from the instrument.

1.5.2 OPTIONAL MODULES
♦ POSITIVE BARCODE READER

1.5.3 DIMENSIONS, WEIGHT & OPERATING ENVIRONMENT
DIMENSIONS

♦ Height:

53 cm

♦ Depth:

57 cm

♦ Length:

75 cm

♦ 35 Kilos

WEIGHT
OPERATING ENVIRONMENT

♦ Temperature: 18°C ÷ 30°C.
♦ Relative humidity: 20% ÷ 85%

1.5.4 INSTALLATION REQUIREMENTS
♦ Input Voltage 90 ÷ 250 Vac
♦ Input Frequency: 47 ÷ 63 Hz
POWER REQUIREMENTS

♦ Power consumption:
♦ 300 W for the analytical unit
♦ 400 W for the work station
♦ EN 61010-1:1993 +A2:1995
(IN COMPLIANCE WITH THE MAIN

SAFETY REGULATIONS

EUROPEAN DIRECTIVES 73/23/CEE AND
93/68/EEC REGARDING SAFETY)

Ellipse

Service Manual

Rev. 03

01June 2009

Page 10

Chapter 01 – INTRODUCTION

♦ EMC 89/336/EEC – 92/31/EEC

Directives

♦ EN 55011, Class B, Group 1
♦ EN 50081-1:1992 EMC
ELECTROMAGNETIC
COMPATIBILITY

♦ EN 55022
♦ ENV 50140 – ENV 50141
♦ EN 60601-1-2
♦ EN 61000-4

Warning: A steady power supply (+ 10%) must be provided for the instrument.
If it is not, the manufacturer highly recommends the use of:
♦ UPS Uninterruptible Power Supply ( No-break module)
♦ ELECTRONIC STABILIZER

Ellipse

Service Manual

Rev. 03

01June 2009

Page 11

Chapter 02 – SYSTEM DESCRIPTION

CHAPTER 02
– SYSTEM DESCRIPTION INDEX

2

DESCRIPTION OF THE SYSTEM ................................................................................................ 2

2.1

ANALYSES CYCLE ......................................................................................................................... 3

2.1.1 REACTION PLATE ......................................................................................................................... 3
2.1.2 REACTION PLATE CYCLE .......................................................................................................... 3
2.2

SAMPLING ARM - OPERATIONAL SEQUENCE .................................................................... 3

2.3

WASH STATION............................................................................................................................... 6

2.3.1 WASH STATION CYCLE ............................................................................................................... 6
2

DESCRIPTION OF THE SYSTEM .........................................................................................2

2.1

ANALYSES CYCLE................................................................................................................3

2.1.1 REACTION PLATE .................................................................................................................3
2.1.2 REACTION PLATE CYCLE..................................................................................................3
2.2

SAMPLING ARM - OPERATIONAL SEQUENCE...............................................................3

2.3

WASH STATION.....................................................................................................................6

2.3.1 WASH STATION CYCLE.......................................................................................................6
2.4

ELECTRONIC DESCRIPTION...............................................................................................7

2.4.1 INTRODUCTION ....................................................................................................................7
2.4.2 ANALYTICAL CONTROL BOARD (C.P.U.) [P/N: 30-01283-01]......................................7
2.4.3 STEPPER MOTORS DRIVER BOARD [P/N: 30-01284-01]...............................................8
2.4.4 PLATE INTERFACE BOARD (Sx) [P/N: 30-01281-01].......................................................9
2.4.5 ARM INTERFACE BOARD (Dx) [P/N: 30-01282-01] .........................................................9
2.4.6 ELECTROVALVES CONTROL BOARD [P/N: 30-01626-00] ...........................................10
2.4.7 PRE-AMPL/ADC [P/N: 30-00107-00; P/N: 30-00107-03]..................................................10
2.4.8 PHOTOMETER LAMP BOARD [P/N: 30-01576-00] .........................................................11
2.4.9 CONTROL PANEL BOARD [P/N: 30-01850-00] ...............................................................11
2.4.10 LEVEL SENSOR ASSY [P/N: 10-01478-00].......................................................................11

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 1

Chapter 02 – SYSTEM DESCRIPTION

2 DESCRIPTION OF THE SYSTEM
"Ellipse" is a random access, computer controlled, counter-top, clinical analysis instrument. The
system can perform 138 tests per hour and has a machine cycle of 26 seconds. Its execution time
ranges from a minimum of 18 seconds to a maximum of 1032, depending on the analysis method
chosen.
The first time the system is used for laboratory analyses, the operator must configure the system
based on the specific needs of that laboratory; i.e.: the chemistry parameters and the reagents racks,
along with the normal ranges, calibrated and control values, must all be defined.
The daily routine analyses will be carried out according to patient sample arrival in a sequential and
continuous, non-stop manner.
The work list is organized using a loading rack holding up to 8 patient samples. Rack loading is
non-stop.
The racks can accommodate both test tubes and micro cups. The bar code for primary tubes and
reagent containers is an optional feature.
When the system, the analytical unit and the computer, is turned on the color-meter lamp is supplied
with low voltage power (1.2 volts), the sampling arm pre-heater remains turned off, while instead
the reaction plate heater, the reagents refrigerating unit (optional module) and the electronic
components are turned on.
In this phase, the "Stand-by" light, placed on the front panel, will flash until the reaction plate
reaches a temperature of 36° C. When this temperature is reached, the "Stand-by" light will stop
flashing and will remain constantly lit.
In the case of system failure or malfunction, the "Ready" light, situated on the front panel of the
instrument, will light up red.
In order to access the main program, double click on the "Ellipse" icon on the computer desktop.
The main menu - "System Monitor" - will appear.
Whenever any system function is launched, the color-meter lamp and the sampling arm pre-heater
will receive regular power.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 2

Chapter 02 – SYSTEM DESCRIPTION

2.1 ANALYSES CYCLE
2.1.1 REACTION PLATE
The reaction plate of the "Ellipse" system contains 6 disposable racks with 20 reaction cuvettes
each.
The racks can be removed individually.
The basic operating cycle of the reaction plate takes 26 seconds. This cycle includes: optic reading
of the cuvettes in incubation, aspiration and dispensing of the reagents and the samples by the arm,
along with the relative positioning of the plate .
The reactions take place at 37° C. This temperature is maintained constant by a controlled heating
unit placed under the reaction plate .
2.1.2

REACTION PLATE CYCLE

After reagents and samples have been placed in cuvette #1, the reaction plate will rotate 40
positions counter-clockwise, so as to bring the first cuvette to be analyzed in front of the colormeter for reading with either one or two wavelengths, as required.
The plate will then, moving counter-clockwise, carry out all the readings of any other prepared
cuvettes. After having effectuated all the readings, the plate will move counter-clockwise to its
initial position minus one cuvette, ready for a new dispensing.
In this manner, the reaction cuvettes move clockwise 1 - 2 - 3 – 4 for dispensing, washing and for
their relative readings.
2.2

SAMPLING ARM - OPERATIONAL SEQUENCE
1. The sampling arm lifts up from the wash well and carries out a wash cycle;
2. The arm moves toward the specific reagent container, while the diluter aspirates an air
bubble to separate the rinse column from the reagent;
3. The arm lowers itself into the reagent, below the level indicated by the sensor, and aspirates
the required quantity of reagent. If the method requires a Rinse (used in order to reduce the
possibility of negative contamination between the water column and the reagent) an extra
amount of reagent (not used in the analysis) will be aspirated before the quantity of reagent
necessary for the analysis, along with another air bubble for their separation;

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 3

Chapter 02 – SYSTEM DESCRIPTION

4. While the diluter aspirates a second air bubble, the arm rises and then lowers into the wash
well so that it can be washed externally, to minimize cross contamination;
5. The arm moves to the specified sample and aspirates a third air bubble;
6. Once the level sensor has indicated the presence of the liquid, the arm stops and aspirates
the sample;
7. The arm once again is raised, while the diluter aspirates a fourth air bubble to prevent
sample loss;
8. At this point, the arm returns to the wash well in order to clean the outside of the probe and
aspirates a fifth air bubble;
9. The arm moves to the reaction PLATE , dispenses, and mixes the reagent and sample in the
reaction cuvette for incubation and reading;
10. The

Ellipse

arm returns

to the

Service Manual

wash well

Rev. 03

and

carries out

01 June 2009

a

probe

wash cycle.

Page 4

Chapter 02 – SYSTEM DESCRIPTION

SAMPLING SYSTEM
Aspiration of Air
Aspiration of Reagent

Aspiration of Air
(To separate reagent from serum)

To wash well to clean the probe

Aspiration of air

Aspiration of serum
Aspiration of air

To wash well to clean the probe

H2O
AIR

Aspiration of air

RGT
AIR

Dispensing and mixing in cuvette

AIR
SERUM
AIR

To wash well for final wash

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 5

Chapter 02 – SYSTEM DESCRIPTION

2.3 WASH STATION
The reaction plate wash station is made up of a series of five small needles situated on one side of
the reaction PLATE . Said needles are opportunely connected to the valve and pump system for
emptying, washing and drying operations (please see the hydraulic diagram).
2.3.1 WASH STATION CYCLE
The wash station carries out its operations alternating upward and downward movement. In its
downward movement phase the needles are guided in such a manner as to carry out the following
operations:
•

The first needle, using the central cannula, removes the reaction mix while the external
cannula dispenses, shower-like fashion, the wash solution; after that the external
cannula dispenses wash solution and then the liquid is aspirated from the central
cannula;

•

The second needle operates exactly like the first but uses distilled water instead;

•

The third needle operates exactly like the first but dispenses rinse solution into the
cuvette so that an optics check can be performed (if the results are negative, the cuvette
is discarded);

•

The fourth needle aspirates the control water;

•

The fifth needles dries the sides.

All these operations are part of the routine operation of the instrument. Every reaction cuvette is
washed at the end of each round of analysis.
The reusability (optical integrity) of each reaction cuvette is always tested before the next round of
analysis.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 6

Chapter 02 – SYSTEM DESCRIPTION

2.4 ELECTRONIC DESCRIPTION
2.4.1

INTRODUCTION

The Ellipse general interconnection diagram is reported on the document having the code
SC-16-00571-XX. This document shows all the electronic boards and the links among them.
The power supply is connected through a dedicated socket to the power line. It supplies, by two
separate modules, the two requested voltages +24 VDC and +12 VDC.
All the Ellipse electronics boards are following described.
2.4.2

ANALYTICAL CONTROL BOARD (C.P.U.) [P/N: 30-01283-01]

The Analytical Control Board, integrated in the Ellipse system, is the heart of the low level, real
time and processing management. This board, on which is present the micro-controller HITACHI
H8S/2633F (U1), permits to manage all the input and output analogical and digital signals
need to the instrument functionality. The mentioned micro-controller, having 128 pin packaging and
12 ports, contains 16 KB of RAM memory and 256 KB flash memory where the firmware is
installed. The firmware writing on the flash memory is permitted by an external personal computer
by a serial transmission RS232.
The Analytical Control Board is connected to the “Stepper Motors Driver Board” by a 64 pin
frontal clutch connector from which receives all the signals for driving the motors.
The following further boards are directly connected to the Analytical Control Board, through
specific cables:
ƒ

The “Plate Interface Board (Sx)” and “Arm Interface Board (Dx)” for the users
interfacing;

ƒ

The “Pre-Ampl./ADC” for reading the signals from main and reference channels of the
photometer;

ƒ

The “Electrovalves Control Board” for commanding all the hydraulic devices.

Besides, through two distinct and direct serial links, it manages (if present) the bar code reader and
the ISE modules.
On the board are present also the followings two leds:
o The red led (LD1) not used yet;
o The green led (LD2) is flashing when the board is powered and the firmware activated.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 7

Chapter 02 – SYSTEM DESCRIPTION

The power supply, by a specific line of the power cable, supplies the board with +24 VDC and +12
VDC. Further two voltages, +5 VDC and +3,3 VDC are generated on the same board by voltage
regulators.
Components and integrated circuit dedicated to the programming communication with the microcontroller and those for I/O ports protection, are present on the same board.
Finally, the alarmed fan needed for the electronic area cooling, composed by the Analytical Control
Board and the Stepper Motors Driver Board, is powered directly from the same board.
2.4.3

STEPPER MOTORS DRIVER BOARD [P/N: 30-01284-01]

The "Stepper Motors Driver Board" is the interface between the “Analytical Control Board” and the
stepper motors.
Besides the Analytical Control Board, there are two more boards directly linked to the Stepper
Motors Driver Board. They are the Plate Interface Board (Sx) and the Arm Interface Board (Dx)
which transmit (by specific cables) all the command signals to the nine stepper motors present in
the instrument. On the boards are present nine driver components (NMB SDI-C403 or TOSHIBA
TA8435H). Each driver, on which is installed a metal heatsink by a couple of screws, is dedicated
to drive one stepper motor as specified here below:
‰

driver U1 Sample plate motor;

‰

driver U7 Reagent plate motor;

‰

driver U4 Reaction plate motor;

‰

driver U6 Vertical sampling arm motor;

‰

driver U3 Horizontal sampling arm motor;

‰

driver U9 Diluter motor;

‰

driver U2 Photometer filter wheel motor;

‰

driver U5 Washing station motor;

‰

driver U8 Peristaltic pump motor.

By the nine dip switches setting on the board, each motor is driven to ¼ of step unless the filter
wheel motor that is driven to 1/8 of step.
The power supply, by a specific power line cable, supplies two voltages +24 VDC and +12 VDC to
the board. The +5 VDC needed to the driver components is generated on the same board by a
voltage regulator. All the components managing the nine drivers and the relevant signals are
mounted on the same board.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 8

Chapter 02 – SYSTEM DESCRIPTION

2.4.4

PLATE INTERFACE BOARD (SX) [P/N: 30-01281-01]

This board is the hardware interface between the reaction and reading sections of the instrument and
the following boards:
ƒ

the Analytical Control Board which receives from the Plate Interface Board the I/O
signals for managing the relevant devices (the three independents plates home sensors,
the washing station home sensor, the photometer filter wheel home sensor, the reaction
plate N.T.C., the reaction plate heater, the lamp and the fan of the photometer);

ƒ

the Stepper Motors Driver Board from which the Plate Interface Board receives the
command signals for the stepper motors of the following devices: sample plate, reagent
plate, reaction plate, washing station and photometer filter wheel;

ƒ

the Photometer Lamp Board for the photometer halogen lamp checking.

There are also the following two leds:
o the yellow led (LD1) active when the reaction plate resistor is powered;
o the red led (LD2) active when there is a signal from the reaction plate leaking.
The power supply, by a specific power line cable, supplies two voltages +24 VDC and +12 VDC to
the board. The +5 VDC needed is generated on the Plate Interface Board by a voltage regulator. All
the components managing respective devices are mounted on this board.
2.4.5

ARM INTERFACE BOARD (DX) [P/N: 30-01282-01]

This board is the hardware interface between the Sampling sections of the instrument and the
following boards:
ƒ

the Analytical Control Board which receives from the Arm Interface Board the I/O
signals for managing the relevant devices (the vertical sampling arm home sensor, the
horizontal sampling arm home sensor, the diluter home sensor, the diluter electrovalve
and pump, the preheater, the level sensor assembly, the temperature service probe, the
main alarmed fan, the three buttons – indicators of the Control Panel);

ƒ

the Stepper Motors Driver Board from which the Arm Interface Board receives the
command signals for the stepper motors of the following relevant devices: vertical and
horizontal sampling arm, Diluter, peristaltic pump;

ƒ

Ellipse

the Control Panel Board for the three buttons-indicators checking.

Service Manual

Rev. 03

01 June 2009

Page 9

Chapter 02 – SYSTEM DESCRIPTION

There are also the following three led:
o the yellow led (LD1), active when the preheater is powered;
o the red led (LD2), active when there is a signal from the sampling arm leaking sensor;
o the yellow led (LD3), light on when the level sensor is activated.
The power supply, by a specific power line cable, supplies +24 VDC and +12 VDC to the board. +5
VDC needed to the Arm Interface Board is generated on the same board by a voltage regulator. All
the components managing respective devices are mounted on this board.
2.4.6

ELECTROVALVES CONTROL BOARD [P/N: 30-01626-00]

This board communicates with the Analytical Control Board all the I/O signals for managing all the
relevant devices (electrovalves and micropumps related to the washing station, the washing well
and the peristaltic pump; the bottles liquid sensors; the waste sensors; the main cover switch).
The red led (LD1) is active when there is a signal from the washing station leaking sensor;
From the Analytical Control Board the Electrovalves control board receives the +24 VDC . +5
VDC is generated from + 24VDC by a voltage regulator. All the components managing respective
devices are mounted on this board.
2.4.7

PRE-AMPL/ADC [P/N: 30-00107-00; P/N: 30-00107-03]

The Pre-Ampl/ADC board P/N: 30-00107-00 is used for reading the signals from the photometer
main channel, while the Pre-Ampl/ADC P/N: 30-00107-03 has the same function for the
photometer reference channel. They are mounted on the photometer assembly and directly linked to
the Analytical Control Board, by two distinct flat cables, for transmitting all the I/O signals. The
only difference between them is the R1 resistor value. On the boards, the (FD1) photodiode reveals
the signal arriving from the relevant reading channel; the Amplifier (U2) amplifies the signal, which
is then inputted in the serial converter ADS 1250 (U3) for its conversion A/D.
From the Analytical Control Board they receive also two continuous voltages (+VB and –VB).
All the components dedicated to the signal amplification or digital conversion are mounted on the
same boards.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 10

Chapter 02 – SYSTEM DESCRIPTION

2.4.8

PHOTOMETER LAMP BOARD [P/N: 30-01576-00]

This board is linked to the Plate Interface Board (Sx) from which receives the check signal (coming
from the Analytical Control Board) to regulate the light intensity of the photometer halogen lamp
(+6V, 10W). From the interface board receives the continuous voltage (VL) which is then converted
to +6 VDC on the same board, by a voltage regulator, for powering the lamp.
2.4.9

CONTROL PANEL BOARD [P/N: 30-01850-00]

This board is linked to the Arm Interface Board (Sx) to which transmits the I/O signals from the 3
lighting buttons of the Control Panel.
From the interface board receives also the continuous voltage (VB) for the three buttons – indicators
functionality.
2.4.10 LEVEL SENSOR ASSY [P/N: 10-01478-00]
The level sensor assembly is composed by the level sensor board “A” (P/N:30-01392-00) and the
level sensor board “B” (P/N:30-01392-01).
The level sensor assembly, installed on the sampling probe head, is linked to the Arm interface
board. By a specific cable it transmits the relevant signal (the capacity variation produced by the
contact between the sampling probe and the liquid) to the Analytical Control Board.
Level sensor components and integrated circuits are mounted on the level sensor assembly.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 11

Chapter 03 - INSTALLATION

CHAPTER 03 - INSTALLATION
INDEX

3.1 UNPACKING .........................................................................................................................2
3.2 INSTALLATION ...................................................................................................................4
3.2.1 INSTALLATION SITE SPECIFICATIONS ......................................................................4
3.2.2 ELECTRIC CURRENT REQUIREMENTS.......................................................................4
3.2.3 CONNECTIONS TO ACCESSORIES ...............................................................................5
3.2.3.1 POWER SUPPLY/LINE INPUT......................................................................................5
3.2.3.2 CONNECTING THE INSTRUMENT TO THE COMPUTER .......................................6
3.2.4 ATTENTION.......................................................................................................................6
3.2.5 SYMBOLS...........................................................................................................................6
3.2.6 REGULATORY COMPLIANCE........................................................................................6

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 1

Chapter 03 - INSTALLATION

3.1 UNPACKING
The ELLIPSE is packed and delivered in two separate wooden crates: one contains the
analyzer itself and the other the computer, along with its accessories. In the event that the order
not include the PC component, packing and delivery will involve one wooden crate plus a
corrugated cardboard box. The packing has been expressly studied and designed to insure
maximum protection of the contents during shipping and handling. It is therefore extremely
important that the crate(s)/box be carefully examined upon delivery in order to ascertain their
integrity. Special attention should be dedicated to examining the color of the “Shock Watch”
glued to the crates, which must show the color ‘white’. A ‘red’ “Shock Watch” indicates that
the crate(s) have experienced some sort of ‘shock’ during handling, transport and/or delivery.
This fact must be noted by the courier on the delivery note, as must any and all visible external
damage (for example: holes, dents, rips or tears, water marks, etc.) evident at the moment of
delivery. This will simplify matters in the event of any future claims for damages.

Upon arrival of the crate(s)/box, take out the
delivery note and make sure that all the items
on the packing list are included in the crates
and are undamaged. Make sure the series
number on the delivery note/packing list
corresponds to that impressed on the plate on
the right side of the instrument.

ELLIPSE
S/N
P/N
90 - 250Vac
Analyzer
Medical
System

47 - 63Hz

300W

Rome-Italy

Open the crate(s)/box from the top and very
carefully take out:
-

the instrument;

-

the computer and accessories.

MAKE SURE THAT THE UNPACKING IS CARRIED OUT BY TWO PEOPLE.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 2

Chapter 03 - INSTALLATION

Do not discard the delivery crate(s)/box or the packing material until the correct functioning of
the instrument has been ascertained.
Remove all the items from the crate(s)/box very carefully.
Remove the adhesive tape from the cover of the samples and reagents housing, from the front
panels and from the samples and reagents racks.
Before connecting the "Ellipse", remove the protective packing material placed under the
sampling arm and under the wash station group.
Warning: in the event that it is necessary to repack any or all of the delivered item(s), the
following procedures must be carefully followed:
¾ Reposition the protective packing material under the sampling arm and under the wash
station group.
¾ Tape down (using masking tape if possible) the cover of the samples and reagents housing,
the front panels, and the samples and reagents racks.
¾ Remove the probe from the sampling arm and place it inside a cuvette. Then cap the
cuvette and tape the cap down.
¾ Be very careful to not bend the wash station cannulas when repositioning the protective
packing material.
¾ Fill the empty spaces around the accessories packed in the crate using “pluriballs” or other
suitable packing material.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 3

Chapter 03 - INSTALLATION

3.2

INSTALLATION

The ELLIPSE must only be installed by a qualified technician who has been authorised and
trained to do so. During its installation the system will be checked once again to ensure correct
functioning. The persons who are required to operate the ELLIPSE system must have received
the adequate training. This should also include the "know-how" of the normal maintenance for
the instrument. A description of the maintenance will be found in Chapter 7 of this manual.
ELLIPSE is a complex system, and it is therefore extremely important that it is correctly
installed in order to fully guarantee fine performance. If the installation and use directions,
given in this manual, are not correctly followed and/or security indications are not respected,
AMS cannot guarantee correct functioning of the instrument. Apart from this, the security of
the operator could be placed at risk.
3.2.1

INSTALLATION SITE SPECIFICATIONS

Ascertain that the ELLIPSE system is not exposed to direct sunlight, draughts, dust or strong
magnetic fields. In addition, please take note of the following conditions required for the
location of the installation:
USE

In covered and dry place

DEGREE OF POLLUTION

2

INSULATION CLASS

I

INSTALLATION CATEGORY

II

TEMPERATURE
HUMIDITY

between18°- 30°C

ALTITUDE

Max 3000 m
Shelf or table with a minimum surface of 75 x 60 cm
stable and free of vibration
Leave a minimum distance of 10 cm around the
instrument to permit air circulation . Make sure that
the front and rear holes are not blocked by any object

LOCATION
VENTILATION

3.2.2

20% ÷ 85%

ELECTRIC CURRENT REQUIREMENTS

The power voltages to which the instrument is adapted are indicated on the left-hand side
(see fig. 1). It must be plugged into a plug of the correct voltage.

Ellipse

VOLTAGE

100 ÷ 230 Vac 47/63 Hz ± 10%

FUSES

6.3 Amp/T - 5 x 20

Service Manual

Rev. 03

01 June 2009

Page 4

Chapter 03 - INSTALLATION

NOTE:

IT IS ADVISABLE TO MAINTAIN THE MAXIMUM STABILITY OF THE ELECTRICAL CURRENT IN
THE LABORATORY. WHERE THIS IS NOT POSSIBLE OR ASCERTAINABLE, USE OF THE
FOLLOWING SUPPLEMENTARY DEVICES IS RECOMMENDED:

ELECTRONIC STABILIZER
Used to stabilise the electric voltage in the laboratory. Any stabiliser with a power
potential greater than 0.5 KW, currently available on the market, can be used.
NO-BREAK MODULE UPS - (Uninterrupted Power Supply)
This module provides two important functions:
- stabilises the main-line power
- supplies current to the instrument in case of a main-line power failure.
3.2.3 CONNECTION OF THE ACCESSORIES
3.2.3.1 POWER SUPPLY

Fig. 1 - – Plug (use the feeder cable supplied with the instrument).
The sticker indicates the power supply voltage and the values of the fuses.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 5

Chapter 03 - INSTALLATION

3.2.3.2 COMPUTER - INSTRUMENT CONNECTION
The instrument and the Personal Computer are connected by one serial RS232 standard cable
(Cod. 9-35 0055.01), which provides the hardware support for the communication.

3.2.4 ATTENTION
The following label is found at the rear of the instrument.

NOTE:

THE REAR PANELS OF THE INSTRUMENT MUST NEVER BE OPENED WITHOUT HAVING FIRST
SWITCHED THE INSTRUMENT OFF AND DISCONNECTED THE ELECTRICITY CABLE.

THE

MAINTENANCE AND CLEANING PROCEDURES FOUND IN CHAPTER

MANUAL

MUST

BE

RESPECTED

AT

ALL

TIMES.

REMEMBER

TO

07

OF THIS

FOLLOW

THE

DECONTAMINATION PROCEDURE IN CASE OF INSTRUMENT REMOVAL (SEE CHAPTER 07) .

3.2.5 SYMBOLS

ATTENTION: READ THE INSTRUCTIONS IN THE USER MANUAL

TERMINAL OF TOTAL MASS PROTECTION (CONDUCTOR)

3.2.6 REGULATORY COMPLIANCE
The ELLIPSE instrument complies with:
European Directive 98/79/CE for In vitro Diagnostics Devices

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 6

Chapter 04 – ELECTRICAL SCHEMES AND DRAWINGS

CHAPTER 04
- ELECTRICAL SCHEMES AND DRAWINGS INDEX
ELECTRONIC DIAGRAM ELLIPSE

SC-16-00571-XX

HYDRAULIC DIAGRAM ELLIPSE

SI-16-00571-XX

WIRING

ME-50-01692-01

ANALYTICAL CONTROL BOARD

SC02-30-01283-XX

ANALYTICAL CONTROL BOARD

SE-30-01283-XX

STEPPER MOTORS DRIVER BOARD

SC02-30-01284-01

STEPPER MOTORS DRIVER BOARD

SE-30-01284-01

ARM INTERFACE BOARD (DX)

SC02-30-01282-01

ARM INTERFACE BOARD (DX)

SE-30-01282-01

PLATE INTERFACE BOARD (SX)

SC02-30-01281-01

PLATE INTERFACE BOARD (SX)

SE-30-01281-01

PRE-AMPL/ADC

SE-30-00107-XX

PHOTOMETER LAMP BOARD

SE-30-01576-00

LEVEL SENSOR ASSY

SE-10-01478-00

ELECTROVALVES CONTROL BOARD

SC-30-01626-00

ELECTROVALVES CONTROL BOARD

SE-30-01626-00

CONTROL PANEL BOARD

SE-30-01850-00

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 1

Pre-Ampl/ADC
(Sample Channel)

Pre-Ampl/ADC

(Reference Channel)

30-01283-..

(optional)

Electronic Diagram Ellipse

The present document is property of AMS, any use without
authorization will be prosecuted accordingly to the law.
Description

Via E. Barsanti, 17/A
(Via Tiburtina Km 18,300)
00012 Guidonia (Roma)

P/N

20/10/2005

Sheet

Rev.

1 of 1

A. Gagliarducci

R. Cornacchia

L. Massenzi

SC 16-00571-XX D

Doc.

Date

Approved

Checked

Drawn

Rev. D: emissione in riferimento alla RMP n.268

Analytical Control Board

(optional)

BarCode Reader

ISE Module

50-01692-01

30-01283-..

Analytical Control Board

The present document is property of AMS, any use without
authorization will be prosecuted accordingly to the law.
Description

Via E. Barsanti, 17/A
(Via Tiburtina Km 18,300)
00012 Guidonia (Roma)

P/N

20/10/2005

Sheet

Rev.

1 of 1

A. Gagliarducci

R. Cornacchia

L. Massenzi

SC02 30-01283-XX A

Doc.

Date

Approved

Checked

Drawn

Rev. A: emissione in riferimento alla RMP n.268

(Sample Channel)

Pre-Ampl/ADC

(Reference Channel)

Pre-Ampl/ADC

A

B

C

D

VM

H5

TX4

VB

RX4

470uH

470uF/50V
C15 +
L1

VM

J6

J4

11
1

WRL
PC0

CLK
ENG

1D
2D
3D
4D
5D
6D
7D
8D

CLKCU
CLKWH
CLKBV
DIRCU
DIRWH
DIRBV

CLKCA
CLKFO
CLKBH
DIRCA
DIRFO
DIRBH

CLKRE
CLKPP
CLKDO
DIRRE
DIRPP
DIRDO

+ C13

470uF/50V

2
5
6
9
12
15
16
19

A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
A17
A18
A19
A20
A21
A22
A23
A24
A25
A26
A27
A28
A29
A30
A31
A32

PWDRE
PWDPP
PWDDO
ENAPI
ENAPP
ENADO

PWDCU
PWDWH
PWDBV
ENAPI
ENABVWH
ENABVWH

PWDCA
PWDFO
PWDBH
ENAPI
ENABHFO
ENABHFO

5

0.1uF

C24

DTR#

0.1uF

1

DTR

1N4148

D2

1N4148

D1

TX2
RX2

0.1uF

C22

1
2
3
4
5
6
7
8

3V3

0.1uF

C7

4V7

D6

4V7

D5

0.1uF

C36

PVCC

VCC
GND
T1OUT
R1IN
R1OUT
T1IN
T2IN
R2OUT

PVCC

R1
3K3

C5 470pF

0.1uF

16
15
14
13
12
11
10
9

4

RX0
P31
P30
PA1
PA2

C25
0.1uF

PVCC

TX0

R8
0R

LOCK_IN
RESB
RESA
PC0
PA3
PG3 (/CS1)
PF4 (/HWR)
PF5 (/RD)

C18
10uF/25V

-VB

1
2
3
4
5
6
7
8
9
11

18
16
14
12
9
7
5
3
20
10

74HC244

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

C60

R2 220

0.1uF

R36 10K STBY
19
1

PG3
PF5

U5

1
2
3
4

0.1uF

F0
F1
F2
F3
F4
F5
F6
F7
VCC
GND

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

C29 TX4
0.1uF RX4

1
2
3
4
5
6
7
8
C1+
V+
C1C2+
C2VT2OUT
R2IN

U8

0.1uF

R39 10K

G
DIR

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE

U1

VCC
GND
T1OUT
R1IN
R1OUT
T1IN
T2IN
R2OUT

MAX202

H2

1
2

J18

20
10

18
17
16
15
14
13
12
11

18
16
14
12
9
7
5
3
20
10

10
20

2
5
6
9
12
15
16
19

74HC377

GND
VCC

1Q
2Q
3Q
4Q
5Q
6Q
7Q
8Q

74HC244

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

3

16
15
14
13
12
11
10
9

TX1
RX1
P34
P33
P37
P36

0.1uF

C30
0.1uF

2
3
4
5
6
7
8
9

PVCC

R34

LEDV

J3
H4
Diagnostic

J2

H8

PVCC

18
16
14
12
9
7
5
3
20
10

VB

PVCC

PVCC

10K

R38

2

J13

H3

P94

P91

WRL
PC3

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

ENA1

J1

H5

0
Rev.

18
16
14
12
9
7
5
3
20
10

J8

Description

0.1uF

C48

PVCC

0.1uF

C41

PVCC

H2

1
2

J22

J16

SE-30-01283-XX

Document Number

1

J14

C59 +

P97

DACP

10
8
6
4
2

10
8
6
4
2

10
8
6
4
2

9 DRDYRF
CLK
7
G0R
5
G1R
3
1

9 DRDYDA
CLK
7
G0D
5
G1D
3
1
H5x2

9
7
5
3
1

3V3
PVCC
9 DRDYAUX
CLK
7
G0AUX
5
G1AUX
3
1

25
23
21
19
17
15
13
11
9
7
5
3
1

H5x2

9
7
5
3
1

1

Sheet

Drawn

Checked

L.M. 10/09/2004

of

1

R.CORNACCHIA

1

J9

PC3
PC4
CTR2
PF5 (RD)
PD4
PD5
PD6
PD7
RES
RTS
RX2

DACP

P97

H3

H5x2

9
7
5
3
1

H13X2

25
23
21
19
17
15
13
11
9
7
5
3
1

R48 1K

3
2
1

J15

H3

DACB

J17
P96

0

Rev

Checked
Approved
Approved A.G. 10/09/2004

J10

SCLK 10
DINDA 8
-VB
6
DSYNC 4
2
VB

J11

SCLK 10
DINRF 8
-VB
6
DSYNC 4
2
VB

3
2
1
R52 1K

26
24
22
20
18
16
14
12
10
8
6
4
2
J5

J12

SCLK 10
DINAUX 8
-VB
6
DSYNC 4
2
VB

PC2
26
PC1
24
PC0
22
PF4 (HWR)20
PD3
18
PD2
16
PD1
14
PD0
12
P35
10
DTR#
8
TX2
6
PG2 (CS2) 4
2

1uF/50

H3

3
2
1

C55 +

P96

RISAB

1uF/50

H3

3
2
1

Analytical Control Board

RMP 224

PVCC

PVCC

LAMPG
LAMPV
LAMPR
LEDV
LEDR
ENADO
ENAPP
CTR1

CTR2
DSYNC

R42 10K

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

Date: 10/09/2004

Size
A3

10
20

2
5
6
9
12
15
16
19

P95

P92

R41 10K

R47 1K

R51 1K

74HC377

GND
VCC

1Q
2Q
3Q
4Q
5Q
6Q
7Q
8Q

PF3

P73

74HC244

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

Title

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE
U20

2
4
6
8
11
13
15
17
1
19

CLK
ENG
U12

11
1

1D
2D
3D
4D
5D
6D
7D
8D

1

1uF/50

C58 +

SP

1uF/50

C54 +

SB

C+8 10K

3
4
7
8
13
14
17
18

9
8
7
6
5
4
3
2

RR4

Serial IN
RS232

WASTE1
WASTE2
LIQUID1
LIQUID2
LIQUID3
LEAKWH
LEAKB
DEPR

1

VM
EV3PP
EV1H2O
uPBASICP
EV1PP
uP1H2O
SWCOP
WASTE1
LIQUID3
LIQUID1

PE4
PE3
PE5
PE6
PE2
PE1
PE7
PE0

R45 1K

ALLVENT

PVCC

0.1uF

C43

PVCC

PVCC

0.1uF

C37

PVCC

23
21
19
17
15
13
11
9
7
5
3
1

R49 1K

J21
H6

23
21
19
17
15
13
11
9
7
5
3
1

C+8 10K

R44 10K

9
8
7
6
5
4
3
2

1
2
3
4
5
6

24
22
20
18
16
14
12
10
8
6
4
2

J20 H2X12

RR3

LOCK
CTR2
PF2
uPACIDB
SCLK
P35B
ENABHFO
PF7

PF0

PVCC

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

P14
P15
P13
P16
P12
P17
P11
P10

1uF/50

C57 +

TP

1uF/50

C53 +

TB

Electronic
Housing Fan

LED GREEN

BarCode

18
16
14
12
9
7
5
3
20
10

74HC244

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

LD1

1

24
22
20
18
16
14
12
10
8
6
4
2

2

EVBASICP
EVACIDP
uPACIDP

PVCC

VM

74HC244

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

LED RED
LD2

RX0

4K7

4K7

P93

VM

EV2H2O
uPBAND
EV2PP
uP2H2O
LEAKWH
WASTE2
LIQUID4
LIQUID2

0.1uF

C44

C+8 10K

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE
U18

2
4
6
8
11
13
15
17
1
19

9
8
7
6
5
4
3
2

RR2

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE

U19

2
4
6
8
11
13
15
17
1
19

C45

PVCC

R46 1K

AVCC

TX0

R33

PF0
PF1
LIQUID4
PF3
PA0
P35
P77
OPTOWH2

LEDR

VM TX1 RX1

PVCC

0.1uF

C42

PVCC

0.1uF

C38

10
20

DIRCA
DIRCU
DIRRE
DIRFO
DIRWH
DIRBH
DIRBV
DIRDO

0.1uF
R50 1K
P90

AVCC

PB4
PB3
PB5
PB6
PB2
PB7
PB1
PB0
PVCC

PG
PV
PR
SLI
ARMSW
LEAKPI
SWCOP
ALLVENT
ENA0

1

1uF/50

C+8 10K

RR1

TF
C56 +

PVCC

1uF/50

C52 +

TA

PVCC

PVCC

C3

GND
VCC

CLK
ENG

2
5
6
9
12
15
16
19

74HC377
C40 0.1uF

1Q
2Q
3Q
4Q
5Q
6Q
7Q
8Q

1D
2D
3D
4D
5D
6D
7D
8D

U10

11
1

3
4
7
8
13
14
17
18

0.1uF

C4

PVCC

G0D
G1D
G0R
G1R
G0AUX
G1AUX
DIRPP
PWDPP

PVCC

P40
P41
P42
P43
P44
P45
P46
P47

PVCC

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

PVCC

74HC245

VCC
GND

B1
B2
B3
B4
B5
B6
B7
B8

PE7
PE6
PE5
PE4
PE3
PE2
PE1
PE0
P17

PD0

P31
P30
PD7
PD6
PD5
PD4
PD3
PD2
PD1

P32

P34
P33

WRL
PC1

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

H8S/2633F
C51 0.1uF

64
63
62
61
60
59
58
57
56
55
54
53
52
51
50
49
48
47
46
45
44
43
42
41
40
39

PVCC

uP1H2O
uP2H2O
uPBAND
uPBASICP
uPSCAMB
CTRLAMP
RISCPI
BUZZER

LOCKED(ON)

CLK
ENG

1D
2D
3D
4D
5D
6D
7D
8D

U11

11
1

3
4
7
8
13
14
17
18

U16

LOCK

C31

12 FWEBOOT
13 RES
14 MD2
WRL
15 FWE
PC2
16 WRL
17 ENA1
18 ENA0
19 DTR# LOCK_IN
20
PVCC
10

NME1212S

2
4
6
8
11
13
15
17
1
19

18
16
14
12
9
7
5
3
20
10

74HC244
0.1uF

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

A1
A2
A3
A4
A5
A6
A7
A8

U25

2
3
4
5
6
7
8
9

PD0
PD1
PD2
PD3
PD4
PD5
PD6
PD7

R37 10K PA3
P70
P71
PVCC
CLKPP
PWMB
C34 OPTOCA
CLKFO
OPTOCU
CLKWH
0.1uF OPTORE
ENABVWH
OPTOFO
ENAPI
OPTOWH
OPTOBH
PVCC
OPTOBV
OPTODO

0.1uF C27

C19

PVCC

C39

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE

U15

2
4
6
8
11
13
15
17
1
19

3

P34/RXD1/SDA0
P33/TXD1/SCL1
VSS
P32/SCK0/SDA1/IRQ4
PVCC2
P31/RXD0/IRRXD
P30/TXD0/IRTXD
PD7/D15
PD6/D14
PD5/D13
PD4/D12
PD3/D11
PD2/D10
PD1/D9
PVCC1
PD0/D8
VSS
PE7/D7
PE6/D6
PE5/D5
PE4/D4
PE3/D3
PE2/D2
PE1/D1
PE0/D0
P17/TIOCB2/TCKLD/PWM3

PE0
PE1
PE2
0.1uF PE3
PE4
PE5
PE6
PE7

3V3

R35 10K NMI

PVCC

U24 16V8

I0
I1
I2
I3
I4
I5
I6
I7
I8
I9

C28

PVCC

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE

U23

2
4
6
8
11
13
15
17
1
19

0.1uF

C2

C17 +

VB

INPUT
INPUTB
P72
P73
P74
P75
P76
P77

0.1uF
3V3

C1

0.1uF
MAX202

C26

R7
3K3

R6 3K3
2

3K3

R5

R4 3K3
2

C1+
V+
C1C2+
C2VT2OUT
R2IN

U7

PVCC DS1233-10

1
2
3

U6

1

18 P32
16 PC5
14 PC6
12 PC7
PG4
9
PG0
7
PG1
5
WDT
3
20 PVCC
10

74HC244

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

PVCC

RTS

RESA

R43 10K

C23

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE

R40
10K

H2

1
2

J19

U17

DRDYAUX 2
DINDA
4
DINRF
6
DINAUX
8
OPTOPP 11
DRDYDA 13
DRDYRF 15
WDTOVF 17
1
19

PVCC

PVCC

3V3

PVCC

15pF

0.1uF

C8

15pF

C47

C6

RX2
TX2

DIN 64_AB-H

B1
B2
B3
B4
B5
B6
B7
B8
B9
B10
B11
B12
B13
B14
B15
B16
B17
B18
B19
B20
B21
B22
B23
B24
B25
B26
B27
B28
B29
B30
B31
B32

33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64

0.1uF

C35

PVCC

PWDCA
PWDCU
PWDRE
PWDFO
PWDWH
PWDBH
PWDBV
PWDDO

10
20
PVCC
74HC377

GND
VCC

1Q
2Q
3Q
4Q
5Q
6Q
7Q
8Q

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

18 CLKBH
16 EVACIDP
14 EVBASICP
12 EVDOS
CLKBV
9
uPDOS
7
CLKDO
5
uPACIDP
3
PVCC
20 PVCC
10
C33
74HC244
0.1uF

P70
P71
P72
P73
P74
P75
P76
P77

103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128

C9

0.1uF

TP-GND

C46

AN0/P40
AN1/P41
AN2/P42
AN3/P43
AN4/P44
AN5/P45
DA0/AN6/P46
DA1/AN7/P47
AN8/P90
AN9/P91
AN10/P92
AN11/P93
AN12/P94
AN13/P95
DA2/AN14/P96
DA3/AN15/P97
AVSS
CS4/DREQ0/TMCI011/TMRI01P70
CS5/DREQ1/TMCI23/TMRI23/P71
SYNCI/CS6/TEND0/TMO0/P72
CS7/TEND1/TMO1/P73
MRES/TMO2/P74
SCK3/TMO3/P75
RXD3/P76
TXD3/P77
MD0

C10 +

0.1uF

10uF/25V

P40
P41
P42
P43
P44
P45
P46
P47
P90
P91
P92
P93
P94
P95
P96
P97

0.1uF

C50

VB

+

C11

XTAL
Rx1=470
se
X1<5Mhz

4

DTR
RTS

J7

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32

1
2
3
4
5

PVCC

U4

Refer to the Serigrafy

U9

3
4
7
8
13
14
17
18

RES

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE

U14

2
4
6
8
11
13
15
17
1
19

3
2
1

AVCC

PVCC

18 CLKCU
16 EV1PP
14 EV2PP
12 EV3PP
CLKCA
9
EV1H2O
7
CLKRE
5
EV2H2O
3
20 PVCC PVCC
10
C32
74HC244
0.1uF

1Y1
1Y2
1Y3
1Y4
2Y1
2Y2
2Y3
2Y4
VCC
GND

ISE - NO3
Module

1A1
1A2
1A3
1A4
2A1
2A2
2A3
2A4
1OE
2OE

U13

2
4
6
8
11
13
15
17
1
19

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

P10
P11
P12
P13
P14
P15
P16
P17

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

C14

D3
1N5818

0.1uF

OUT
GND
IN

U3

U2

LM7805CT
LM2575

MOLEX

1
2
3
4
5
6

TP1

MD2

3
2
1

PC0
PC1
PC2
PC3

OUT
GND
IN

PC4

470uF/50V
VL
C49 +

1

6,144 Mhz
X1
R3 0R
EXTAL

PF0
PF1
PF2
PF3
PF4
PF5
PF6

10uF/25V

PF7

PC5
PC6
PC7

C12

EXTAL

3V3

OSC1

PB0

5

XTAL
FWE
STBY
NMI
RES

PB1
PB2
PB3
PB4
PB5
PB6
PB7
PA0
PA1
PA2
PA3

LD1086V3.3

WDTOVF
PG4
(CS1)
PG3
(CS2)
PG2
PG1
PG0
P37

P10
P11
P12
P13
P14

1
2
3
4
5

P36
P35

102
101
100
99
98
97
96
95
94
93
92
91
90
89
88
87
86
85
84
83
82
81
80
79
78
77
76
75
74
73
72
71
70
69
68
67
66
65

VREF
AVCC
NC
NC
PF0/BREQ/IRQ2
PF1/BACK/BUZZ
PF2/LCAS/WAIT/BREQ0
PF3/LWR/ADTRG/IRQ3
PF4/HWR
PF5/RD
PF6/AS/LCAS
VSS
PF7/CLK
PVCC1
OSC2
OSC1
VSS
EXTAL
VCC
XTAL
FWE
STBY
NMI
RES
PLLVSS
PLLCAP
PLLVCC
WDTOVF
PG4/CS0
PG3/CS1
PG2/CS2
PG1/CS3/OE/IRQ7
PG0/CAS/IRQ6
P37/TXD4
NC
NC
P36/RXD4
P35/SCK1/SCK4/SCL0/IRQ5

+

VM

C

VB

C

10uF/25V

C

C16
0.1uF

C

MD1
MD2
NC
NC
A0/PC0
A1/PC1
A2/PC2
A3/PC3
VSS
A4/PC4
VCC
A5/PC5
PWM0/A6/PC6
PWM1/A7/PC7
VSS
TIOCA3/A8/PB0
PVCC1
TIOCB3/A9/PB1
TIOCC3/A10/PB2
TIOCD3/A11/PB3
TIOCA4/A12/PB4
TIOCB4/A13/PB5
TIOCA5/A14/PB6
TIOCB5/A15/PB7
A16/PA0
TXD2/A17/PA1
RXD2/A18/PA2
SCK2/A19/PA3
VSS
A20/DACK0/TIOCA0/PO8/P10
A21/DACK1/TIOCB0/PO9/P11
A22/TCKLA/TIOCC0/PO10/P12
A23/TCKLB/TIOCD0/PO11/P13
IRQ0/TIOCA1/PO12/P14
NC
NC
TCKLC/TIOCB1/PO13/P15
IRQ1/PWM2/TIOCA2/PO14/P16

INPUT
WDT
OPTOWH2
CTR2

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38

TF
TP
SP
LIQUID4
DACP

P15
P16

25
23
21
19
17
15
13
11
9
7
5
3
1
26
24
22
20
18
16
14
12
10
8
6
4
2
CTRLAMP
BUZZER

1
2
3
4

LEAKB
PWMB
EV3PP
INPUTB

1
2
3
4
5
6
7
8

RISAB
TB
TA
SB

1
2
3

SLI
ARMSW
CTR1
CTR2
WDT
DACB

1
2
3
4
5

33
31
29
27
25
23
21
19
17
15
13
11
9
7
5
3
1
34
32
30
28
26
24
22
20
18
16
14
12
10
8
6
4
2

OPTOCU
OPTOCA
OPTORE
OPTOFO
OPTOWH
WASTE1
WASTE2
LIQUID1
LIQUID2
LIQUID3
LEAKPI
RISCPI
uPSCAMB
H13X2

LAMPV
LAMPR
LAMPG
H17X2

EVDOS

DSYNC ADC

OPTOBH
OPTOBV
OPTODO
OPTOPP
PR
PG
PV
DEPR
uPACIDB
uPDOS
A

B

C

D

50-01692-01

U5

J7

Stepper Motors Driver Board

The present document is property of AMS, any use without
authorization will be prosecuted accordingly to the law.
Description

Via E. Barsanti, 17/A
(Via Tiburtina Km 18,300)
00012 Guidonia (Roma)

P/N

20/10/2005

Sheet

Rev.

1 of 1

A. Gagliarducci

R. Cornacchia

L. Massenzi

SC02 30-01284-01 A

Doc.

Date

Approved

Checked

Drawn

Rev. A: emissione in riferimento alla RMP n.268

30-01283-..

C34
3n3

5

U7

1

1
2

1
2

1
2

SW DIP-2

SW7

SW DIP-2

SW4

4
3

4
3

4
3

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

SW DIP-2

SW1

GNDD

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

SWRE0
SWRE1

SWCU0
SWCU1

SWCA0
SWCA1

22
17

21
18

19
16

23
20

24
15

VM

R13
1R-2W

D28

D27

D26

D25

0.1uF

C23

R7
1R-2W

D16

D15

D14

D13

0.1uF

C17

R1
1R-2W

D4

D3

D2

RR1

4

C+9 10K PVCC
SWCU1
2
1
SWCU0
3
SWCA1
4
SWCA0
5
SWRE1
6
SWRE0
7
SWFO1
8
SWFO0
9
SWWH0 10

R14
1R-2W

REA
REAN
REB
REBN

R8
1R-2W

CUA
CUAN
CUB
CUBN

R2
1R-2W

CAA
CAAN
CAB
CABN

ENAPP

ENAWH

ENAFO

C35
3n3

C24
0.1uF

C32
3n3

C18
0.1uF

C29
3n3

13

GNDD

U8

1

7
RESMOT2
CLKPP 6
3
DIRPP 5
PWDPP10
4
SWPP0 8
SWPP1 9

13

1
2

1
2

1
2

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

SW DIP-2

SW8

SW DIP-2

SW5

SW DIP-2

4
3

4
3

4
3

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

100uF/50

SW2

GNDD

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

PVCC

U5

1

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

C2

C5

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

100uF/50

100uF/50

GNDD

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

PVCC

U2

1

7
RESMOT2
CLKFO 6
3
DIRFO 5
PWDFO10
4
SWFO0 8
SWFO1 9

13

7
RESMOT2
CLKWH 6
3
DIRWH 5
PWDWH10
4
SWWH0 8
SWWH1 9

C12
0.1uF

PVCC

0.1uF

C40

C8

22
17

21
18

19
16

23
20

VM

SWPP0
SWPP1

SWWH0
SWWH1

SWFO0
SWFO1

22
17

21
18

19
16

23
20

24
15

VM

VM

22
17

21
18

19
16

23
20

24
15

24
15

+

7
RESMOT2
CLKRE 6
3
DIRRE 5
PWDRE10
4
SWRE0 8
SWRE1 9

C7

22
17

21
18

19
16

23
20

24
15

+

13

100uF/50

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

VM

22
17

21
18

19
16

23
20

24
15

D1

0.1uF

C11

VB

470uF/50
+
0.1uF
C39

C38

VL

VB

R15
1R-2W

D32

D31

D30

D29

0.1uF

C25

R9
1R-2W

D20

D19

D18

D17

0.1uF

C19

R3
1R-2W

D8

D7

D6

D5

0.1uF

C13

J4

3

VM

TP1

RR2

ENABH

ENADO

ENABV

C+9 10K PVCC
SWWH1 2
1
SWPP1
3
SWPP0
4
SWBH1
5
SWBH0
6
SWBV1
7
SWBV0
8
SWDO0
9
SWDO1 10

R16
1R-2W

PPA
PPAN
PPB
PPBN

R10
1R-2W

WHA
WHAN
WHB
WHBN

R4
1R-2W

FOA
FOAN
FOB
FOBN

MOLEX
TP-GND

C36
3n3

C26
0.1uF

C33
3n3

C20
0.1uF

C30
3n3

13

13

1
2

1
2

GNDD

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

2

SW DIP-2

SW9

SW DIP-2

SW6

SW DIP-2

SW3

U9

1

7
RESMOT2
CLKDO 6
3
DIRDO 5
PWDDO10
4
SWDO0 8
SWDO1 9

1
2

GNDD

PVCC

U6

1

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

4
3

4
3

4
3

22
17

21
18

19
16

23
20

24
15

C9

SWDO0
SWDO1

SWBV0
SWBV1

SWBH0
SWBH1

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

100uF/50

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

C3

C6

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

100uF/50

J2

100uF/50

GNDD

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

PVCC

U3

1

7
RESMOT2
CLKBH 6
3
DIRBH 5
PWDBH10
4
SWBH0 8
SWBH1 9

13

7
RESMOT2
CLKBV 6
3
DIRBV 5
PWDBV10
4
SWBV0 8
SWBV1 9

C14
0.1uF

PVCC

2

22
17

21
18

19
16

23
20

24
15

+

A

ENARE

C22
0.1uF

GNDD

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

PVCC

U4

1

7
RESMOT2
CLKCU 6
3
DIRCU 5
PWDCU10
4
SWCU0 8
SWCU1 9

13

C4

SDIC403

GNDPA
GNDPB

SA
SB

FB1
FB2

FA1
FA2

VMA
VMB

100uF/50

GNDD

PVCC

U1

1

CK1
RES
CK2
EN
DIR
PD
OSC
M1
M2

VCC

C1

VM

470uF/50
+
C37

PVCC

+

C31
3n3

C16
0.1uF

C28
3n3

7
RESMOT2
CLKCA 6
3
DIRCA 5
PWDCA10
4
SWCA0 8
SWCA1 9

13

PVCC

100uF/50

+
+

B

ENACU

ENACA

C10
0.1uF

RESMOT

R19
10K

3
2
1

+

C

D

U10

OUT
GND
IN

LM2940C

3

1
2
3
4
5
6

4

1

CAA
CAB
CUA
CUB
REA
REB
FOA
FOB
WHA
WHB
19
17
15
13
11
9
7
5
3
1
20
18
16
14
12
10
8
6
4
2
VM

VM

22
17

21
18

19
16

23
20

24
15

VM

0
Rev.

Description

R18
1R-2W

DOA
DOAN
DOB
DOBN

R12
1R-2W

BVA
BVAN
BVB
BVBN

R6
1R-2W

BHA
BHAN
BHB
BHBN

SE-30-01284-01

Document Number

A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15
A16
A17
A18
A19
A20
A21
A22
A23
A24
A25
A26
A27
A28
A29
A30
A31
A32

33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64

PWDRE
PWDPP
PWDDO
ENARE
ENAPP
ENADO

PWDCU
PWDWH
PWDBV
ENACU
ENAWH
ENABV

PWDCA
PWDFO
PWDBH
ENACA
ENAFO
ENABH

DIN 64_AB-H

B1
B2
B3
B4
B5
B6
B7
B8
B9
B10
B11
B12
B13
B14
B15
B16
B17
B18
B19
B20
B21
B22
B23
B24
B25
B26
B27
B28
B29
B30
B31
B32

H8X2

J3

Drawn

Checked

1

Sheet

1

of

1

R.CORNACCHIA

L.M. 10/09/2004

0

Rev

Checked
Approved
Approved A.G. 10/09/2004

J1

1
2
3
4
5
6
7
8
9
10
11
12
13
CLKCU 14
CLKWH 15
CLKBV 16
DIRCU 17
DIRWH 18
DIRBV 19
20
21
22
RESMOT 23
24
25
26
CLKRE 27
CLKPP 28
CLKDO 29
DIRRE 30
DIRPP 31
DIRDO 32

CLKCA
CLKFO
CLKBH
DIRCA
DIRFO
DIRBH

1

Stepper Motors Driver Board

Date: 10/09/2004

Size
A3

RMP 224

R17
1R-2W

D36

D35

D34

D33

0.1uF

C27

R11
1R-2W

D24

D23

D22

D21

0.1uF

C21

Title

H10X2

R5
1R-2W

D12

D11

D10

D9

0.1uF

C15

CAAN
CABN
CUAN
CUBN
REAN
REBN
FOAN
FOBN
WHAN
WHBN

BHA
BHB
BVA
BVB
PPA
PPB
DOA
DOB
15
13
11
9
7
5
3
1
15
13
11
9
7
5
3
1
16
14
12
10
8
6
4
2
16
14
12
10
8
6
4
2
BHAN
BHBN
BVAN
BVBN
PPAN
PPBN
DOAN
DOBN

PVCC

5

+
+

A

B

C

D

30-01283-..

Analytical Control Board

05-01700-20

50-01867-00

Arm Interface Board (Dx)

The present document is property of AMS, any use without
authorization will be prosecuted accordingly to the law.
Description

Via E. Barsanti, 17/A
(Via Tiburtina Km 18,300)
00012 Guidonia (Roma)

P/N

20/10/2005

Sheet

Rev.

1 of 1

A. Gagliarducci

R. Cornacchia

L. Massenzi

SC02 30-01282-01 B

Doc.

Date

Approved

Checked

Drawn

Rev. B: emissione in riferimento alla RMP n.268

50-01692-01

A

B

C

D

1
2

2

1
2
3

J15

MIC5021

C6
100uF/50V
U5
1 VDD Vbst
2 INPUT Gate
3 Ct
Sens4 GND Sens+

R27 10K

10K

R17

10K

3

R41
560R

(see Jumper)

H3
Pre-Heater PWS config.

VM
VPRE
VB

C10
0.1uF

VPRE
PWMB

VB

R2 2K2

VB

D-EVDOS

VM

R47

PVCC

RISANB

H2

J16

Diluter
Valve

H4

C1
E2
K3
A4

J7

Spare

H8
Peristaltic Pump Motor

1
2
3
4
5
6
7
8

J3

-

+

R28 10K

5

VPRE

8
7
6
5

1

1
2
3

LM358/SO

U3A

PVCC

H3

J17

SB

R13 1K

C7
0.01uF

BHA
BHAN
BHB
BHBN

1
2
3
4
5
6
7
8

J1

H2

LED-YELLOW

LD1

PWRES

Q1
IRF530
3
SOURCE

2 DRAIN

3A

F1

H3

J19

1
2
3

Spare

VPRE

1
2

H3

3
2
1

J14

R29
33K

50K

R19

R45 2K2

6

5

CTR1

DACB

10K

R37

4

-

+

CTR2

PWRES

1
2
3

H11

1
2
3
4
5
6
7
8
9
10
11

J13

1
2
3
4
5
6
7
8

J2

TA

DEPR
PVCC

H2

1
2

J12

10K

10K

H3

J23
1
2
3

1N4148

D6

10K

10
8
6
4
2

9
7
5
3
1

6

5

H 5X2

10
8
6
4
2

J27

10K

VB

R36

R22

PVCC

R44
10K

INPUTB

R35

VB

10K

R21

PG

3

R52

PR

1N4148

D7

PVCC

R51

R50

PV

3

Housing Fan
(bottom)

D3 4.7V

R7 10K

OPTOBV

Leak Arm

R43
560R

R6 2K2

VB

R11
10K

PVCC

H4

C1
E2
K3
A4

Vertical Arm Home
Sensor
J6
VB

H8
Vertical Arm Motor

BVA
BVAN
BVB
BVBN

Level Sensor and
Pre-Heater

10K

LD3

R33

7

H3

J20

Spare

LM358/SO

U3B

PVCC

NTC1+

ARMSW

PVCC

SLI
VM

LED-YELLOW

PVCC

10K

R30

R25 10K

PVCC

4

D2 4.7V

R5 10K

D-uPACIDB

VM

R42
560R

R4 2K2

VB

Temp. Service
Probe
J22

H4

C1
E2
K3
A4

Horizontal Arm Home
Sensor
OPTOBH
J5
VB

H8
Horizontal Arm Motor

D-uPDOS

VM

Diluter Pump

D1 4.7V

R3 10K

R46
10k
OPTOPP

RISANB

1 GATE

8

4

8

4

5

470K
1M

2
1
2
1

-

9
7
5
3
1

C9

VB

To Front Panel Green Led

LEAKB

LED-RED

LD2

100uF/50V

+ C1

D4 4.7V

D-LAMPG

R53
750R

To Front Panel Yellow Led

To Front Panel Red Led

R49 750R
D-LAMPR

R48
750R

3
2
1

LM7805CT

OUT
GND
IN

U2

BHAN
BHBN
BVAN
BVBN
PPAN
PPBN
DOAN
DOBN

R9 10K

OPTODO

R1
560R

R18
680R

VB

R8 2K2

VB

D-LAMPV

LM358/SO

7

U4B

H4

C1
E2
K3
A4

J8

Diluter Home
Sensor

0.1uF

PVCC

+

1
2
3
4
5
6
7
8

J4

H8
Diluter Motor

DOA
DOAN
DOB
DOBN

8
4

2

VB

R31
33K

NTC1+

50K

R20

15
13
11
9
7
5
3
1

BHA
BHB
BVA
BVB
PPA
PPB
DOA
DOB

10K

R32

R26 10K

PVCC

100uF/50V

+ C3

15
13
11
9
7
5
3
1

PVCC

H8X2

16
14
12
10
8
6
4
2

J26

C2
0.1uF

16
14
12
10
8
6
4
2

2

2

3

C4
0.1uF

PVCC

-

+

8

1

TP-GND

100uF/50V

C11

LAMPV
LAMPR
LAMPG

EVDOS

Description

TB

TP1

+

VM

1
2
3
4
5
6
7
8
9

OPTOBH
OPTOBV
OPTODO
OPTOPP
PR
PG
PV
DEPR
uPACIDB
uPDOS

RISOUTB
(EV3PP)

uPACIDB
LAMPV
LAMPR
LAMPG
uPDOS
EVDOS

SE-30-01282-01

Document Number

Arm Interface Board (dx)

RMP 224

R34
10K

LM358/SO

0.1uF
U4A

C8

Date: 10/09/2004

Size
A3

Title

0
Rev.

4

1

PPA
PPAN
PPA
PPB
PPBN
PPB

+

2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
32
34

VM

1
2
3
4
5
6

Spare

H9

1
2
3
4
5
6
7
8
9

J24

VM

VB

1

Sheet

Drawn

Checked

1

of

1

R.CORNACCHIA

L.M. 10/09/2004

0

Rev

Checked
Approved
Approved A.G. 10/09/2004

VM

MOLEX

J21

LEAKB
PWMB
RISOUTB
INPUTB

RISAB
TB
TA
SB

VL

D-RISOUTB

D-uPACIDB
D-LAMPV
D-LAMPR
D-LAMPG
D-uPDOS
D-EVDOS

SLI
ARMSW
CTR1
CTR2
WDT
DACB

18
17
16
15
14
13
12
11
10

D-RISOUTB
RISANB
RISOUTB
INPUTB
RISAB
WDT

H17X2

1
3
5
7
9
11
13
15
17
19
21
23
25
27
29
31
33

J25

ULN2803

I1
O1
I2
O2
I3
O3
I4
O4
I5
O5
I6
O6
I7
O7
I8
O8
GND COM

U1

1

A

B

C

D

30-01283-..

Analytical Control Board

Plate Interface Board (Sx)

The present document is property of AMS, any use without
authorization will be prosecuted accordingly to the law.
Description

Via E. Barsanti, 17/A
(Via Tiburtina Km 18,300)
00012 Guidonia (Roma)

P/N

20/10/2005

Sheet

Rev.

1 of 1

A. Gagliarducci

R. Cornacchia

L. Massenzi

SC02 30-01281-01 A

Doc.

Date

Approved

Checked

Drawn

Rev. A: emissione in riferimento alla RMP n.268

50-01692-01

A

B

C

D

H3

3
2
1

J11

1
2

H3

3
2
1

DACP

CTRLAMP

D1 4.7V

INPUT
WDT
OPTOWH2
CTR2

5

100uF/50V

J26

H8

1
2
3
4
5
6
7
8

J2

H4

C1
E2
K3
A4

VB

47K

R51

10K

R54

47K

R25

4

R26
33K

2

3

-

+

2

3

10K

LM358

0.1uF
U2A

WDT
D-CTR2
CTR2
DACP

VB

10K

R28

VM

-

+

R29

H7

1
2
3
4
5
6
7

TP

R47 1K

C11
0.01uF

Spare

10K

R43
560R

SP

1

LM358

0.1uF
U3A

C3

1

VB

R5 2K2

Reaction
Tray Motor

VM J29

R55

PVCC

R27 10K

PVCC

NTC1+

R53
33K

PVCC

MIC5021

8
7
6
5

C13

H4

C1
E2
K3
A4

C10
100uF/50V
U6
1 VDD Vbst
2 INPUT Gate
3 Ct
Sens4 GND Sens+

R52 10K

PVCC

C12
0.1uF

H8

1
2
3
4
5
6
7
8

J3

Reaction Tray Home
Sensor
J8
VB

CUA
CUAN
CUB
CUBN

D2 4.7V

R4 10K

OPTOCA

VM
RISCPI

NTC3+

R42
560R

R3 2K2

Sample Tray
Motor

Sample Tray Home
Sensor
J7
VB

C5
+

VL

CAA
CAAN
CAB
CABN

R2 10K

OPTOFO

Sense

PVCC

PWRES

Reaction tray
Heater

MOLEX-2

J16

MOLEX-3

3
2
1

J15

VL

10K

R45

J19

R41
560R

INPUT

PVCC

NTC3+

To Lamp PWS board

H3

3
2
1

J31

VB

VB

Spare - analog

Photometer
Fan

H4

C1
E2
K3
A4

R1 2K2

Filter Wheel
Motor

Filter Wheel Home
Sensor
J6
VB

H8

1
2
3
4
5
6
7
8

J1

2
DRAIN

FOA
FOAN
FOB
FOBN

8

4

1
GATE

4

8

4

H2
LEAKPI

1
2

J17

LED-YELLOW

LD1

PWRES

H8

1
2
3
4
5
6
7
8

J4

VM

H4

C1
E2
K3
A4

1N4148

D7

1N4148

D8
10K

3

R31
33K

NTC2+

47K

R30

10K

R33

R32 10K

PVCC

R23

R20

R21

PVCC

R44
560R

10K

R22

VB

6

5

R7 2K2

Reagent
Tray Motor

RGT Tray Home Sensor
J9
VB

FUSE 2A
Q1
IRF530
3 SOURCE

F1

D3 4.7V

R6 10K

OPTOCU

REA
REAN
REB
REBN

3

470K
1M

2
1
2
1

-

+

6

5

LM358

U2B

-

R34

10K

LM358

U3B

7

7

C1
E2
K3
A4

R24

TF

OUT
GND
IN

3
2
1

LEAKPI

2

NTC1+

VB

NTC2+

100uF/50V
+
0.1uF
C8

C7

H3

3
2
1

Room Temperature
Sensor
J18

H2

2
1

J14

J27

2

H10X2

PVCC

H4

C1
E2
K3
A4

10K

10K

H2

Description

H3

1
2
3

J33

H3

1
2
3

J24

H4

1
2
3
4

J23

H3

1
2
3

J22

1
2

J21

H3

1
2
3

J20

SE-30-01281-01

Document Number

1

18
17
16
15
14
13
12
11
10
D-CTR2

VB
VL

VM

TP-GND

VM

MOLEX-6

D-uPSCAMB

J28

H4
Spare
uPump

1
2
3
4

J12

TP1

VM

D-uPSCAMB

D-BUZZER

1

Sheet

Drawn

Checked

1

of

1

R.CORNACCHIA

L.M 10/09/2004

0

Rev

Checked
Approved
Approved A.G. 10/09/2004

ULN2803

I1
O1
I2
O2
I3
O3
I4
O4
I5
O5
I6
O6
I7
O7
I8
O8
GND COM

D9 4.7V

R49 10K

OPTOWH2

1
2
3
4
5
6
7
8
9

U1

Plate Interface Board (sx)

RMP 224

LIQUID4

R56 10K

PVCC

LIQUID3

R39 10K

PVCC

10K
LIQUID2

R38

PVCC

LIQUID1

R37 10K

PVCC

WASTE2

R36

WASTE1

R35

560R

R50

R48 2K2

uPSCAMB

BUZZER

CTR2

Date: 10/09/2004

Size
A3

Title

0
Rev.

VB

Spare - Home
Sensor
J32
VB

PVCC

0.1uF

C9

D5 4.7V

R10 10K

OPTOWH

Reaction Tray
Temp. Sensor

560R

R11

PVCC

100uF/50V
+
C6

LED-RED

LD2

2

VB

R9 2K2

Wash Station
Motor

BUZZER

LS1

H4

680R

LM7805

U4

H8

1
2
3
4
5
6
7
8

J5

Wash Station
Home Sensor
J10
VB

D-BUZZER 1

D4 4.7V

PVCC

+

WHA
WHAN
WHB
WHBN

R8 10K

OPTORE

8
4
8
4

5

H13X2

25
23
21
19
17
15
13
11
9
7
5
3
1

26
24
22
20
18
16
14
12
10
8
6
4
2

CAA
CAB
CUA
CUB
REA
REB
FOA
FOB
WHA
WHB
19
17
15
13
11
9
7
5
3
1
20
18
16
14
12
10
8
6
4
2
CAAN
CABN
CUAN
CUBN
REAN
REBN
FOAN
FOBN
WHAN
WHBN

1
1
2
3
4
5
6

+

OPTOCU
OPTOCA
OPTORE
OPTOFO
OPTOWH
WASTE1
WASTE2
LIQUID1
LIQUID2
LIQUID3
LEAKPI
RISCPI
uPSCAMB

CTRLAMP
BUZZER

TF
TP
SP
LIQUID4
DACP

A

B

C

D

A

B

C

D

0805

C8
.1uF

-V

8

+ VS

Photodiode

OUT

SOP8

COM

IN
IN
IN
IN

R4 5.7K
0805

5

2
3
6
7

1

+ VS

3 +

2

C1
.1uF
0805

U5
MC79L05ACD

FD1

6

C9

-VS 10uF

7

LM4040
SOT-23

Z1

6

10K
0805

10uF

.1uF
0805

C12

100K
3296X

C10

TR2

100pF

R5

OPA111
TO99

U2

C2

R1 ? M
1/4W

6

C13
.1uF
0805

.1uF
0805

C16

R3
5.7K
0805

TR1
5K
3296X

R2
5.7K
0805

2

3

-

+

8

0805

C3
.1uF

U4

5

8

6

OPA350
SO8

U6

C11
.1uF
0805

5

10uF

C14
C15
.1uF
0805

10uF

C6

+ VD

4

4

C4
.1uF
0805

C7
.1uF
0805

+ VS

VOUT 1
GND 2-3-6-7
MC78L05ACD
SOP8

U1

VOUT 1
GND 2-3-6-7
+12,5 V
MC78L05ACD
SOP8
+ VS

+12,5 V

Per valore di R1 vedi la relativa soluzione

NOTA :

7

7
5

4
8
1

8

7
1

4
5

+ VD

1
2
3
4
5
6
7
8

10uF

C5

1
2
OSC4P

Y1
OSC-HS

16
15
14
13
12
11
10
9

Friday, February 06, 2004
3

Date:

[MA]00107-00

Document Number

2

C17
.1uF
0805

Description

C18
15pF
0805

10K
0805

10K
0805

DRDY

G1
G0

R10

2

R9

R8
10K
0805

PRE-AMPL/ADC

Emission

+ VD

R7
10K
0805

1
0805

R6

Size
A4

Title

0
Rev.

4
3

DGND
G1
G0
CS
DRDY
CLK
SCLK
DOUT

ADS1250
SOL16

+VIN
-VIN
AGND
VS
VREF
DSYNC
VD
DGND

U3

+ VS

3

1
3
5
7
9

Checked

Sheet

Drawn

2
4
6
8
10

SCLK
DOUT

2
4
6
8
10

1

of

1

1

L.M. 25-05-2002
R.CORNACCHIA

0

Rev

Approved

-V

+12,5 V

A.G. 25-05-2002

Checked

1
3
5
7
9

JP1

Approved

10K
0805

R11

TO
CASE

1

A

B

C

D

A

B

C

5

3
2
1

MOLEX-3

J1

R1 1K

INIT-L

C4
0.1uF

4

4

+

C1
100uF/50V

3
VIN

U1
LT1083/TO

+

C2
10uF/25V

VOUT

D1 1N4002
2
1

ADJ
1

D

5

2

3

2

3

3
1
R2
400R

R6
100R 3296Y

R4 0R

R3
120R

1

3
2N3904

2
R5
1K

INIT-L

C3
100uF/50V

Q1

+

2

Date:

Size
A4

Title

0
Rev.

C5
0.1uF

2

Description

H2

HALOGEN
LAMP

SE-30-01576-00
03/01/2005

Document Number

Photometer Lamp Board

Emission

J2
1
2

Sheet

Drawn

Checked

Approved

1

1

of

1

R.CORNACCHIA

L.M. 03/01/2005

0

Rev

Approved
A.G. 03/01/2005

Checked

1

A

B

C

D

A

B

C

D

VM

2

100nF

C13

100R 1/8W

R9

5

5

VI

10uF/35V

+ C5

1

U4

U5

VSS

TRG

RES

VCC

GND

2

VO

3

VCC

10uF/35V

+ C4

VCC

4

C9

GND

100pF

3

C10
100nF

4

R12
10K

10K

R8

PROBE

10nF

C2

4M7

R16

3

TRG

THR

3

U3

VSS

Q

VCC

100nF

C11

R3
1K

0
Rev.

Description

2

U2

C12 100nF

SE-10-01478-00

Document Number

Level Sensor Assy

Emission

Date: 26/11/2003

Size
A4

Title

R6
270K

R2 220K

R1 220K

2

Sheet

Drawn

Checked

1

R11
10K

1

1

of

1

R.CORNACCHIA

L.M. 26/11/2003

A.G. 26/11/2003

0

Rev

Approved

SIGNAL OUT

10K

R10

Checked

Q1

Approved

R15 10K

1

A

B

C

D

Electrovalves Control Board

The present document is property of AMS, any use without
authorization will be prosecuted accordingly to the law.
Description

Via E. Barsanti, 17/A
(Via Tiburtina Km 18,300)
00012 Guidonia (Roma)

P/N

20/10/2005

Sheet

Rev.

1 of 1

A. Gagliarducci

R. Cornacchia

L. Massenzi

SC 30-01626-00 A

Doc.

Date

Approved

Checked

Drawn

Rev. A: emissione in riferimento alla RMP n.268

Analytical Control Board
30-01283-..

A

B

C

D

1
2

3
2
1

1N4148

2
1

2
1

VM

R3

R6

U2

VIN

VM

5

D-uP2H2O

D-uP2H2O

100uF/50V

C3

CONN CS

EV2

H2

J8

+

VM

1

SWCOP

R12
10K

PVCC

1N4148

D2

H3
Main LID Switch

J2

H2
LEAK w

J3

2

1

2

1

470K

1M

2
1

2
1

2

3

-

+

2

VM

3
2
1

0.1

C1

VM

2
1

VM

4

VM

D-uP1H2O

D-uP1H2O

100uF/50V

C2

LEAKWH

CONN CS

EV4

H2

2
1

J10

+

4

LED-RED

LD1

680R

R5

PVCC

D-uPBAND

D-uPBASICP

VM

6

TLV2461CP

U3

H3

J7

D-uPBASICP

CONN CS

EV3

H2

J9

VOUT

LM7805/TO92

10K

R7

10K

R4

PVCC

7

4

D1

GND

3

5

2
1

VM

VM

R2

10K

WASTE2

10K

WASTE1

2
1

2
1

VM

3

VM

D-EV2H2O

VM

D-EV3PP

D-EV3PP

CONN CS

EV6

H2

2
1

J12

CONN CS

EV8

2
1

VM

PVCC

VM

D-EV1H2O

D-EV1H2O

VM

D-EV2H2O

CONN CS

EV7

H2

2
1

J13

H2

2
1

J14

R10 10K
LIQUID3

WASTE1

PVCC

LIQUID1

PVCC

10K

10K

H2X5
10 9 9
8 7 7
6 5 5
4 3 3
2 1 1

J6

R8

10
8
6
4
2

H4

1
2
3
4

J5

H4

1
2
3
4

J4

R9

PVCC

LIQUID2

R11 10K
LIQUID4

WASTE2

PVCC

D-EV2PP

D-EV1PP

CONN CS

EV5

H2

2
1

J11

PVCC

PVCC

R1

3

2

2

U1

Q1

1

3

SE-30-01626-00

Document Number

1

VM

Sheet

Drawn

Checked

1

1

of

1

R.CORNACCHIA

L.M. 07/09/2004

A.G. 07/09/2004

Checked

VM

D3
UF4004

0

Rev

Approved

D-uP1H2O
D-uP2H2O
D-uPBAND
D-uPBASICP
D-EV1PP
D-EV2PP
D-EV3PP
D-EV1H2O

D-EV2H2O

18
17
16
15
14
13
12
11
10

EV3PP
EV1H2O
uPBASICP
EV1PP
uP1H2O
SWCOP
WASTE1
LIQUID3
LIQUID1

VM

Approved

ULN2803

O1
O2
O3
O4
O5
O6
O7
O8
COM

H2X12
23 23
21 21
19 19
17 17
15 15
13 13
11 11
9 9
7 7
5 5
3 3
1 1

I1
I2
I3
I4
I5
I6
I7
I8
GND

24
22
20
18
16
14
12
10
8
6
4
2

J1

2N2222

1
2
3
4
5
6
7
8
9

24
22
20
18
16
14
12
10
8
6
4
2

Description

R14
10K

R13 220R

uP1H2O
uP2H2O
uPBAND
uPBASICP
EV1PP
EV2PP
EV3PP
EV1H2O

EV2H2O
uPBAND
EV2PP
uP2H2O
LEAKWH
WASTE2
LIQUID4
LIQUID2

VM

Electrovalves Control Board

Emission

EV2H2O

Date: 07/09/2004

Size
A4

Title

0
Rev.

2

1
2

A

B

C

D

A

B

C

D

5

5

VB

1

2

0

0

4

PV
PR
PG

VB

3

2

1

PV

D-LAMPV

10
8
6
4
2
9
7
5
3
1

3

H 5X2

10
8
6
4
2

J1
9
7
5
3
1
D-LAMPV
D-LAMPR
D-LAMPG

VB

PUSHBUTTON-LED Green

0

0

PG

SW2
+

VB

D-LAMPG

PUSHBUTTON-LED Yellow

-

+

SW1

4

Green LED
Red LED
Yellow LED

VB

2

1

-

2

Date:

Size
A4

Title

0
Rev.

25/02/2005

SE-30-01850-00

Document Number

Control Panel Board

Emission
Description

PR

D-LAMPR

PUSHBUTTON-LED Red

0

0

+

SW3

2

Sheet

Drawn

Checked

Approved

1

1

of

1

R.CORNACCHIA

L.M. 25/02/2005

0

Rev

Approved
A.G. 25/02/2005

Checked

1

A

B

C

D

Chapter 05 – DIAGNOSTIC PROGRAM

CHAPTER 05
- DIAGNOSTIC PROGRAM -

INDEX
5 DIAGNOSTIC PROGRAM

3

5.1 STATUS BAR

6

5.2 TEST FOLDER

6

5.3 “ARM” FOLDER

7

5.3.1 VERTICAL COMMAND AREA

8

5.3.2 HORIZONTAL COMMAND AREA

8

5.3.3 STATUS COMMAND AREA

9

5.3.4 ARM SETTING COMMAND AREA

9

5.4 “DILUTER” FOLDER

10

5.4.1 OPERATIONS AREA

10

5.4.2 SENSOR STATE AREA

11

5.4.3 USER AREA

11

5.4.4 STATUS COMMAND AREA

11

5.5 “OPTIC” FOLDER

11

5.5.1 OPERATIONS AREA

12

5.5.2 SENSOR STATE AREA

12

5.5.3 FILTERS AREA

12

5.5.4 DIGITAL CONVERSION AREA

13

5.5.5 STATUS COMMAND AREA

13

5.6 “PLATE” FOLDER

13

5.6.1 OPERATIONS AREA

14

5.6.2 MOTOR ON/OFF AREA

15

5.6.3 STATUS COMMAND AREA

15

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 1

Chapter 05 – DIAGNOSTIC PROGRAM

5.7 “TEMPERATURES” FOLDER

15

5.7.1 TEMPERATURES AREA

16

5.7.2 OFFSET AREA

16

5.7.3 STATUS COMMAND AREA

17

5.8 “WASH” FOLDER

18

5.8.1 OPERATIONS AREA

18

5.8.2 TEST AREA

19

5.8.3 STATUS COMMAND AREA

19

5.9 “USER” FOLDER

20

5.9.1 USER TEST MICRO 2 AREA

20

5.9.2 STATUS COMMAND AREA

21

5.10 “MISCELLANEOUS” FOLDER

21

5.10.1 GENERAL RESET AREA

22

5.10.2 LOG RAW DATA AREA

22

5.10.3 CONTINUOUS TEST RUN

22

5.10.4 STATUS COMMAND AREA

23

5.11 “GLOBAL SETTING” FOLDER

23

5.12 “MACRO SETTING” FOLDER

24

5.13 “CONFIGURATION” FOLDER

25

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 2

Chapter 05 – DIAGNOSTIC PROGRAM

5 DIAGNOSTIC PROGRAM
The diagnostic program enables the operator to perform a complete check of each of the Ellipse’
module functions.
This program has a folder structure, with each folder containing functions pertaining to the specific
module. To launch the program the operator has to click with the left side of the mouse, on the
Diagnostic area located on the lower right side of the screen of the System Monitor (Fig.1).
The Ellipse must be in the stand-by state to access this area.
Fig. 1

Click Here

Once the program starts, the following screen appears (Fig. 2)
Fig. 2

Test
Folders

Optical Sensor
Status
Status Bar

The diagnostic program is subdivided into three distinct areas: test folders, optical sensor status,
status bar. The last two areas are present on every diagnostic window.
The three areas enable the operator to verify multiple functions as specified below:
Ellipse

Service Manual

Rev. 03

01 June 2009

Page 3

Chapter 05 – DIAGNOSTIC PROGRAM

Test Folders:

checks the functionality of the Ellipse’ various sub-systems; The individual
functions are illustrated later on.

Optical Sensor Status: visualizes the sensor status that controls the Sampling Arm, the Diluter, the
Washing Station, the Optical filter wheel, the Reaction plate, the Sample plate, the Reagent plate and
the Cover Lock, as well as the Liquid Reservoir Alarm, the Front panel button and the Level Sensor
Test (Fig. 3)
Fig. 3

Furthermore, there are three types of unused fields for future use: Leak for Arm, Washing Station and
Plate. While the program is running, the fields given in the figure can be either of two distinct colors,
red or green.
The fields:
•

Vertical Arm’s Optical Switch (O.S.)

•

Horizontal Arm’s Optical Switch (O.S.)

•

Diluter’s Optical Switch (O.S.)

•

Wash Station’s Optical Switch (O.S.)

•

Filter Wheel’s Optical Switch (O.S.)

•

Reaction Plate’s Optical Switch (O.S.)

•

Sample Plate’s Optical Switch (O.S.)

•

Reagent Plate’s Optical Switch (O.S.)

turn green when the specific device is in the home position, and turn red when not in the home position.
The field
•

Cover Lock

turns green when the cover is closed, and turns red when open.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 4

Chapter 05 – DIAGNOSTIC PROGRAM

The fields:
•

Arm Leak Alarm (Arm Leaking)

•

Wash Station Leak Alarm (Wash Station Leaking)

•

Plate Leak Alarm (Reaction Plate Leaking)

Turn green when there is no leaking, and red when there is.
The following six fields signal the alarms for the bottles containing the washings liquids and waste
materials:
•

Waste1 Liquid Reservoir Alarm

•

Waste2 Liquid Reservoir Alarm

Turning green when the liquid level is not excessive, and red when the waste bottle indicated by the
alarm is almost full.
•

Liquid 1 Liquid Reservoir Alarm

•

Liquid 2 Liquid Reservoir Alarm

•

Liquid 3 Liquid Reservoir Alarm

•

Liquid 4 Liquid Reservoir Alarm

Turning green if the liquid level is above the predetermined minimum limit, and red if the liquid
quantity is below the predetermined minimum level (the bottle is almost empty).
The following fields signal the state of the button on the Ellipse’s operator panel:
•

Al.Res. Front Panel Button

•

St-By Front Panel Button

•

Ready Front Panel Button

Turning green when the button is not pushed in, and red if it is.

The field
•

Test ON

It is green fixed; if pushed, on the right of it will be showed a red drop that becomes temporarily green
when the liquid is detected.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 5

Chapter 05 – DIAGNOSTIC PROGRAM

5.1 STATUS BAR
Monitors the instrument’s status (Fig.4).
Fig. 4
Time

Firmware Rel.

Al. Res Status Indicator

Date

Instrument Status

Ready Status Indicator
Start/Stop Diagnostics

5.2 TEST FOLDER
The following is a summary of each individual test folder’s function.
The Ellipse’s diagnostic program is subdivided into 10 folders:
Arm:

the adjustment of the arm position can be checked and carried out;

Diluter:

checks if the diluter is functioning correctly;

Optic:

checks if the optical block is functioning correctly;

Plate:

checks if the reaction, samples and reagents plates are functioning correctly;

Temperature:

checks the temperature of the pre-heater and reaction plate;

Wash:

checks if the wash station is functioning correctly;

User:

checks if the user functions are working properly;

Miscellaneous:

performs general checks;

Macro Setting:

the operator can identify which macros have been loaded;

Configuration:

Through the use of a password, only qualified technical personal can adjust
certain parameters (This file is not accessible to the Laboratory operator).

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 6

Chapter 05 – DIAGNOSTIC PROGRAM

5.3

“ARM” FOLDER

Warning: the improper use of the functions described in this folder can damage the sampling probe.

The Arm File is subdivided into four areas:
♦

VERTICAL

♦

HORIZONTAL

♦

STATUS COMMAND

♦

ARM SETTING

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 7

Chapter 05 – DIAGNOSTIC PROGRAM

5.3.1 VERTICAL COMMAND AREA
(5 commands)
Home:

Brings the arm vertically to a home position;

Down Well:

Brings the arm’s axis “z” to the height of the well;

Up:

Raises the arm tenths of a millimeter as preset in the square found immediately
on the right, with relation to the previous Down movement;
the box becomes active when the arm is positioned on a sample, a reagent, a
standard or a cuvette.

Down:

Lowers the arm tenths of a millimeter as pre-determined by the square found
immediately on the right; the box becomes active when the arm is positioned on
a sample, a reagent, a standard or a cuvette.

Go Level:

Brings the probe to either the reagent level, the sample or the standard and reads
the level. Before giving this command, it is necessary to set the parameter for the
probe’s maximum lowering level in the max area (mm);

5.3.2 HORIZONTAL COMMAND AREA
(5 commands)
Back to wash well: Brings the arm back to the home position;
Go Reag.:

Brings the arm to the reagent position specified in the sheet menu next to the
command;

Go Sample:

Brings the arm to the sample position specified in the sheet menu next to the
command;

Go Std.:

Brings the arm to the standard/control position specified in the sheet menu next
to the command;

Go Disp.:

Ellipse

Brings the arm to the dispense position;

Service Manual

Rev. 03

01 June 2009

Page 8

Chapter 05 – DIAGNOSTIC PROGRAM

5.3.3 STATUS COMMAND AREA
Send:

Visualizes the commands sent from the program to the instrument hardware;

Receive:

Visualizes the answers of the instrument’s hardware to the commands sent by the
program;

Sensor State:

Indicates the sensor status relative to the arm’s vertical axis, the internal arm and
the external arm.

5.3.4 ARM SETTING COMMAND AREA
(5 commands)
Step:

Visualizes the number of steps.

+:

Performs the counter-clockwise rotation of the arm (to adjust the position);

- :

Performs the clockwise rotation of the arm (to adjust the position);

Save:

Saves the adjustment data;

Reset:

Completely resets the Arm

In this area the boxes can be activated when the arm is positioned on a sample, a reagent, a standard or
a control or a cuvette.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 9

Chapter 05 – DIAGNOSTIC PROGRAM

5.4 “DILUTER” FOLDER

The Diluter Folder is subdivided into four areas:
♦

OPERATIONS

♦

SENSOR STATE

♦

USER

♦

STATUS COMMAND

5.4.1 OPERATIONS AREA
(3 commands)
Home:

Brings the diluter to the Home position;

Asp:

Aspirates the micro-litres as predetermined in the ul. box (may be set manually
through + and – or by using the cursor located immediately above);

Disp.:

Distributes the micro-litres as predetermined in the ul. box (may be set manually
through + and – or by using the cursor located immediately above);

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 10

Chapter 05 – DIAGNOSTIC PROGRAM

5.4.2 SENSOR STATE AREA
In the Sensor State area the diluter sensor status is indicated.

5.4.3 USER AREA
(1 command)
Probe Wash Test: Activates and deactivates the probe wash system
5.4.4 STATUS COMMAND AREA
Send:

Visualizes the commands sent by the program to the instrument hardware;

Receive:

Visualizes the instrument hardware’s response to the commands sent by the
program.

5.5 “OPTIC” FOLDER

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 11

Chapter 05 – DIAGNOSTIC PROGRAM

The Optic folder is subdivided into five areas:
♦

OPERATIONS

♦

SENSOR STATUS

♦

FILTERS

♦

DIGITAL CONVERSION

♦

STATUS COMMAND

5.5.1 OPERATIONS AREA
(2 commands)
Lamp ON:

Turns on the Optic Lamp;

Lamp OFF:

Turns off the Optic Lamp;

5.5.2 SENSOR STATE AREA
This area indicates the status of the wheel filters.
5.5.3 FILTERS AREA
(10 commands)

Ellipse

Reset :

Positions the wheel filters in home position (dark);

340 nm:

Positions filter n° 1 in front of the reading sensor;

380 nm:

Positions filter n° 2 in front of the reading sensor;

405 nm:

Positions filter n° 3 in front of the reading sensor;;

492 nm:

Positions filter n° 4 in front of the reading sensor;

510 nm:

Positions filter n° 5 in front of the reading sensor;

546 nm:

Positions filter n° 6 in front of the reading sensor;

577 nm:

Positions filter n° 7 in front of the reading sensor;

620 nm:

Positions filter n° 8 in front of the reading sensor;

Spare:

Positions filter n° 9 in front of the reading sensor (Optional);

Service Manual

Rev. 03

01 June 2009

Page 12

Chapter 05 – DIAGNOSTIC PROGRAM

5.5.4 DIGITAL CONVERSION AREA
(2 commands)
START/STOP Sample Channel Conversion: Performs an analogical/digital conversion of the
principal channel;
START/STOP Reference Channel Conversion: Performs an analogical/digital conversion of the
reference channel;
5.5.5

STATUS COMMAND AREA

Send:

Visualizes the commands sent by the program to the instrument hardware;

Receive:

Visualizes the instrument hardware’s response to the commands sent by the
program.

5.6 “PLATE” FOLDER

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 13

Chapter 05 – DIAGNOSTIC PROGRAM

The Plate Folder is subdivided into three areas:
♦

OPERATIONS

♦

ALL MOTORS ON/OFF

♦

STATUS COMMAND

5.6.1

OPERATIONS AREA

Reaction Plate area:
(3 commands)
To Disp.:

Automatically positions the cuvette, indicated on the sheet menu aside, under the
dispensing position

To Optic:

Automatically positions the cuvette, indicated on the sheet menu aside, in front
of the colorimeter

Reset:

Performs the reset of the reaction plate by positioning the # 1 cuvette in the
dispensing position

Sample Plate area:
(2 commands)
To Asp.:

Automatically brings sample, indicated on the sheet menu aside, under the
aspiration position

Reset:

Performs the reset of the sample plate by positioning the # 1 sample in the
aspiration position

Reagent Plate:
(2 commands)
To Asp.:

Automatically brings reagent, indicated on the sheet menu aside, under the
aspiration position

Reset:

Performs the reset of the sample plate by positioning the # 1 reagent in the
aspiration position

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 14

Chapter 05 – DIAGNOSTIC PROGRAM

5.6.2 ALL MOTORS ON/OFF AREA
(1 command)
ALL Motors ON/OFF: Engages/Disengages the motors to allow a manual movement of the modules
(Sampling Arm; Reaction, Reagents and samples Plates, Washing Station, Filter Wheel and Peristaltic
Pump).

5.6.3 STATUS COMMAND AREA
Send:

Visualizes the commands sent by the program to the instrument hardware;

Receive:

Visualizes the instrument hardware’s response to the commands received from
the program.

5.7 “TEMPERATURES” FOLDER

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 15

Chapter 05 – DIAGNOSTIC PROGRAM

The Temperature Folder is subdivided into three areas:
Temperatures
Offset
Status Command
5.7.1 TEMPERATURES AREA

(4 commands)
Temp. Test Run:

To enter in the Temperature Test folder

Plate:

Performs a temperature reading in the reaction plate compartment;

Arm:

Performs a temperature reading in the arm’s pre-heater.

Calib. Sensor:

To carry out the calibration procedure of the thermometric probe included
in the kit P/N 55-01204-00, as illustrated in Chapter Six “Settings and
Adjustments”.

5.7.2 OFFSET AREA
(2 commands)
To correlate the thermometric probe included in the kit P/N 55-01204-00 with another reference
thermometric probe.
Service probe:

The offset value that has been previously inserted and the temperature value
read by the thermometric probe appear in the closed fields.

Cursor:

To enter the offset value ± 3 C°. The offset value will be zero after exiting
from the folder.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 16

Chapter 05 – DIAGNOSTIC PROGRAM

TEMPERATURE TEST AREA

(3 commands)
START:

To perform the test check and to regulate the temperature of the preheater and of the
reaction plate by using the kit P/N 55-01204-00, as illustrated in Chapter 6 “Setting
and Adjustments”.
The number of sample cycles are predefined, whereas the reagent and sample
positions for test execution can be selected by the operator.

Service Probe: The field in which the temperature of the service probe appears.
WASH:

To run a wash cycle on the cuvettes used for testing.

EXIT:

To exit from the Temperature Test box and to run a wash cycle on the cuvettes if
they have been used.

5.7.3 STATUS COMMAND AREA

Ellipse

Send:

Visualizes the commands sent by the program to the instrument hardware;

Receive:

Visualizes the instrument hardware’s response to the commands sent by the program.

Service Manual

Rev. 03

01 June 2009

Page 17

Chapter 05 – DIAGNOSTIC PROGRAM

5.8“WASH” FOLDER

The Wash folder is subdivided into three areas:
♦

OPERATIONS

♦

TEST

♦

STATUS COMMAND

5.8.1

OPERATIONS AREA

(2 commands)
Home:

Brings the washing device to the Home position;

Down:

Makes the washing device go down to the washing position.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 18

Chapter 05 – DIAGNOSTIC PROGRAM

5.8.2 TEST AREA
(5 commands)
Test P5 + E.V. 5:

Performs switching on of the E.V. 5 solenoid valve and turns on the P5 pump
and automatically switching off the E.V. 5 solenoid valve and turns off the P5
pump after a period of four seconds

Test P5:

It turns on the µP5 pump and automatically turns it off after a period of four
seconds.

Test P6–Cuvette Dry:It turns on the µP6 pump and automatically turns it off after a period of four
seconds.
Peristaltic Pump:

It turns on the peristaltic pump and automatically turns it off after one second.

Washing Cycle:

It performs the functional check of the Washing Station , as illustrated in
Chapter Six “Settings and Adjustments”.

5.8.3 STATUS COMMAND AREA
Send:

Visualizes the commands sent by the program to the instrument hardware;

Receive:

Visualizes the instrument hardware’s response to the commands sent by the
program.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 19

Chapter 05 – DIAGNOSTIC PROGRAM

5.9 “USER” FOLDER

The User Folder is subdivided into two areas:
♦

USER TEST MICRO

♦

STATUS COMMAND

5.9.1 USER TEST MICRO AREA
(3 commands)
Alarms ON:

Turns on and off the Alarms indicator lamp;

Ready ON:

Turns on and off the READY indicator lamp;

Test BUZZER:

Performs a complete operational test of the buzzer;

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 20

Chapter 05 – DIAGNOSTIC PROGRAM

5.9.2 STATUS COMMAND AREA
Send:

Visualizes the commands sent by the program to the instrument
hardware;

Receive:

Visualizes the instrument hardware’s response to the commands sent
by the program.

5.10 “MISCELLANEOUS” FOLDER

The Miscellaneous folder is subdivided into four areas:
-

General Reset

-

Log Raw data

-

Continuous Test Run

-

Status Command

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 21

Chapter 05 – DIAGNOSTIC PROGRAM

5.10.1 GENERAL RESET AREA
(1 command)
Reset:

Performs the resetting of the general system.

5.10.2 LOG RAW DATA AREA
(1 command)
Enable:

To save a copy of the files for the technical diagnosis if the instrument
malfunctions. The files are memorized in the folder “Log” inside the folder
“Analyzer”.

5.10.3 CONTINUOUS TEST RUN
(1 command)
Start Test:

Performs an automatic test during which the modules (Sampling Arm, Reaction
plate, Washing Station and Photometer) are moved to all the different positions.

The correct modules positioning is also automatically verified and showed on the following table:

To stop the test push on the “Exit Test” key
Ellipse

Service Manual

Rev. 03

01 June 2009

Page 22

Chapter 05 – DIAGNOSTIC PROGRAM

5.10.4 STATUS COMMAND AREA
Send:

Visualizes the commands sent by the program to the instrument hardware

Receive:

Visualizes the instrument hardware’s response to the commands.

5.11 “GLOBAL SETTING” FOLDER

The Global Setting Folder is for next application.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 23

Chapter 05 – DIAGNOSTIC PROGRAM

5.12 “MACRO SETTING” FOLDER

This folder allows the user to read the macro present in the Ellipse.
To read the macro follow the below procedure:
⇒ Select from the menu sheet “Read Macro” the number of relevant macro;
⇒ Click with the left side of the mouse on the Read area;
⇒ If the Macro requested exists, the information will appear in the designated area.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 24

Chapter 05 – DIAGNOSTIC PROGRAM

5.13 “CONFIGURATION” FOLDER
This folder may be accessed only by qualified technical assistant.
It is necessary for the technical assistant to insert a password to access this folder.

The Configuration Folder is subdivided into four areas:
•

Gap

•

Backlash Dil.

•

Heater

•

Wash. Well level Test

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 25

Chapter 05 – DIAGNOSTIC PROGRAM

Gap: defines the volume, in µL, of the air bubble that separates the Rinse (the liquid column into the
sampling arm tube) from the sampled reagent; this volume must be set in the range from 10 to 24
(Standard value 16).
For the correct setting of the gap volume it is necessary to apply the procedure of BIB (Precision
check of the Analytical plate) and BIC (Volume precision check of the Sampling line).
If BIC’s CV values should exceed the limit specified below, the GAP has to be increased by
steps of 2 µL (starting from 10 µL) until obtaining the expected BIC’s CV value
Precision check of the Analytical plate BIB
Prepare two series of 32 samples each by using a diluted solution of K2Cr2O7 (obtained from 4
gr/l concentrated solution diluted 1:100 with Rinse solution (distilled water + 1000 µL/litre of Brij))
Use the same solution as samples and reagent
Program a new method with the following parameters:
Method name: BIB
Type: End Point
Sample volume K2Cr2O7 0,04 gr/l

3 µL

Reagent volume K2Cr2O7 0,04 gr/l

300 µL

Rinse volume

20 µL

Filter:

340 nm

Incubation time:

116 sec

Calculation factor:

1000

BIB Procedure
1.

Select conical cup in Options

2.

Program the BIB method using the above parameters and configure a reagent rack containing it

3.

Enter in the calibration screen and select RBL for BIB method.

4.

Put the Rinse solution as reagent and standard (fourteenth position of the Crt/Std rack) and Start
the Calibration

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 26

Chapter 05 – DIAGNOSTIC PROGRAM

5.

Program 64 BIB; fill 32 conical samples cup with K2Cr2O7 0,04 gr/l

6.

Load the samples racks into the Analyzer at positions 1 and 2

7.

Substitute the Rinse Solution with K2Cr2O7 0,04 gr/l in the reagent rack

8.

Start the Analyzer. At the end of the last sampling, move the samples racks to positions 3 and 4.

The mean of the reading obtained from each of the 32 sample series must fall within 360 ÷ 400.
The coefficient of variation (CV %) must be lower than 0.7 % for each series of 32 samples.

Precision check of the Sampling volume BIC
Prepare two series of 32 samples each by using K2Cr2O7 4 gr/l (concentrated)
Put Rinse Solution as reagent and K2Cr2O7 solution as sample.
Program a new method with the following parameters:
Method name: BIC
Type: End Point
Sample volume K2Cr2O7 4 gr/l

3 µL

Reagent volume RINSE Solution (distilled water + 1000 µL/litre of Brij) 300 µL
Filter:

340 nm

Incubation time:

116 sec

Calculation factor:

1000

BIC Procedure
1.

Program the BIC method using the above parameters and configure a reagent rack

2.

Enter in the Calibration screen and select RBL for BIC method.

3.

Put the Rinse Solution as reagent and Standard (fourteenth position of the Crt/Std rack) and
Start the Calibration

4.

Program 64 BIC ; fill 32 conical samples cup with K2Cr2O7 4 gr/l (concentrated)

5.

Load the samples racks into the Analyzer at positions 1 and 2

6.

Start the Analyzer. At the end of the last sampling, move the sample racks to positions 3 and 4.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 27

Chapter 05 – DIAGNOSTIC PROGRAM

The mean of the readings obtained from each of the 32 sample series must fall between 350 ÷ 450.
The coefficient of variation (CV %) must be lower than 1.6 % for each series of 32 samples
If BIC’s CV values should exceed the limit specified above, the GAP has to be increased by steps
of 2 µL (starting from 10 µL) until obtaining the expected BIC’s CV value

Backlash Dil.:

Allows to recover the sampling imprecision generated by the motor when it inverts
its rotation; the setting is expressed in µL and obtained by using the following
procedure:

1. Enter in the Diagnostic Arm folder menu and make the Sampling Arm “Reset”
2.

Enter in the Diluter folder, press the “HOME” button and then Push “PROBE WASH TEST”
button three times in order to fill the hydraulic sampling line

3. Remove the plastic cover of the sample racks vane and take out the samples racks
4.

Enter in the ARM folder and push the “GO” button in order to move the probe to sample
position # 1

5. Put a sample cup containing distilled water (about 1ml) under the sampling probe
6. Enter in the DILUTER folder and aspirate 300 µL of distilled water from the cup
7. Remove the sample cup containing distilled water
8. Select 1 µL to dispense and press the “Disp” key repeatedly until water appears on the tip of the
probe.
9. Substract 1 from the number of times you pressed the “Disp” key. The obtained number has to
be introduced as BACKLASH value
10. Push the Save button to memorize the inserted data
11. Push the Reset button into the ARM folder
12. Close the Diagnostic program by clicking on the Diagnostic button

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 28

Chapter 05 – DIAGNOSTIC PROGRAM

Heater:

to adjust the preheater and reaction plate temperatures;

Wash. Well level Test:

By clicking on the “Run Test” button the system performs an automatic
verification and adjustment cycle for the liquid level in the washing well.
A number appears in the box “Level” that corresponds to the rotation time
of the rinse pump in milliseconds, that is necessary to maintain the correct
liquid level in the washing well.
The functioning range is between 350 and 700 milliseconds.

If the

washing well is not sufficiently filled, a message appears to warn the user
to check the hydraulic line.
If the value taken during the test is different from the previous one, it is
necessary to click on “Save” in order to save the new data.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 29

Chapter 06 – SETTINGS AND ADJUSTMENTS

CHAPTER 06
- SETTING AND ADJUSTMENTS INDEX

6

SETTING AND ADJUSTMENTS

6.1

4

SAMPLING ARM

4

6.1.1 ALIGNMENT AND ADJUSTMENTS

4

6.1.2 REPLACEMENT OF THE SAMPLING PROBE

9

6.1.3 REPLACEMENT OF THE SAMPLING ARM

13

6.1.4

16

REPLACEMENT OF THE ARM HOME SENSORS

6.1.4.1 Replacement of the Horizontal Arm Home Sensor

16

6.1.4.2 Replacement of the Vertical Arm Home Sensor

18

6.1.5 REPLACEMENT OF THE SAMPLING ARM MOTORS

19

6.1.5.1 Replacement of the Horizontal Arm Motor

19

6.1.5.2 Replacement of the Vertical Sampling Arm Motor

22

6.1.6 REPLACEMENT OF THE SAMPLING ARM BELTS

24

6.1.6.1 Replacement of the Horizontal Sampling Arm Belt

24

6.1.6.2 Replacement of the Vertical Sampling Arm Belt

26

6.2

27

DILUTER MODULE

6.2.1 REPLACEMENT OF THE DILUTER MODULE

27

6.2.2

32

REPLACEMENT OF THE DILUTER MICROPUMP

6.2.3 REPLACEMENT OF THE DILUTER ELECTROVALVE

35

6.2.4 REPLACEMENT OF THE DILUTER HOME SENSOR

36

6.3

PRE-HEATER AND LEVEL SENSOR

38

6.3.1

SUBSTITUTION OF THE PRE-HEATER AND LEVEL SENSOR ASSEMBLY

38

6.3.2 LEVEL SENSOR – CHECK AND ADJUSTMENT PROCEDURE

40

6.4

44

OPTIC ASSEMBLY

6.4.1 SUBSTITUTION OPTIC ASSEMBLY

44

6.4.2 SUBSTITUTION OF THE OPTIC LAMP

46

6.4.3 ELECTRONIC ADJUSTMENT OF THE OPTIC

47

6.4.4 SUBSTITUTION OF THE OPTO SENSOR

49

6.4.5 SUBSTITUTION OF THE MOTOR

50

6.4.6 SUBSTITUTION OF THE OPTIC FILTERS

51

6.5

52

PLATES

6.5.1 REMOVAL OF THE PLATES ASSEMBLY
Ellipse

Service Manual

52
Rev. 04

01 October 2009

Page 1

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5.2 REPLACEMENT OF THE REAGENTS PLATE MOTOR

58

6.5.3 REPLACEMENT OF THE SAMPLES PLATE MOTOR

62

6.5.4 REPLACEMENT OF THE REACTIONS PLATE MOTOR

65

6.5.5 REPLACEMENT OF THE REAGENTS PLATE BELT

67

6.5.6 REPLACEMENT OF THE SAMPLES PLATE BELT

68

6.5.7 REPLACEMENT OF THE REACTIONS PLATE BELT

70

6.5.8 REPLACEMENT OF THE REAGENTS PLATE PULLEY

72

6.5.9 REPLACEMENT OF THE SAMPLES PLATE PULLEY

73

6.5.10 EPLACEMENT OF THE REACTIONS PLATE BELT

73

6.5.11 REPLACEMENT OF THE REAGENTS PLATE HOME SENSOR

75

6.5.12 REPLACEMENT OF THE SAMPLES PLATE HOME SENSOR

77

6.5.13 REPLACEMENT OF THE REACTIONS PLATE HOME SENSOR

78

6.5.14 ALIGNMENT AND ADJUSTMENT OF THE REAGENTS PLATE

79

6.5.14.1 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate motor
or of the Reagents Plate belt

79

6.5.14.2 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate pulley

80

6.5.14.3 Alignment and adjustment of the Reagents Plate in case of replacement of the Reagents Plate
Home Sensor

82

6.5.15 ALIGNMENT AND ADJUSTMENT OF THE SAMPLE PLATE

83

6.5.15.1 Alignment and adjustment of the Sample Plate in case of replacement of the Samples Plate motor
or of the Samples Plate belt

83

6.5.15.2 Alignment and adjustment of the Samples Plate in case of replacement of the Samples Plate pulley

84

6.5.15.3 Alignment and adjustment of the Samples Plate in case of replacement of the Samples Plate
Home Sensor

85

6.5.16

86

ALIGNMENT AND ADJUSTMENT OF THE REACTIONS PLATE

6.5.16.1 Alignment and adjustment of the Reaction Plate in case of replacement of the Reaction Plate motor
or of the Reaction Plate belt

86

6.5.16.2 Alignment and adjustment of the Reactions Plate in case of replacement of the Reactions Plate pulley

88

6.5.16.3 Alignment and adjustment of the Reactions Plate in case of replacement of the Reactions Plate
Home Sensor

90

6.6

WASHING STATION ASSEMBLY

92

6.6.1

SUBSTITUTION OF THE WASHING STATION ASSEMBLY

92

6.6.2

SUBSTITUTION OF THE WASHING STATION HOME SENSOR

94

6.6.3

SUBSTITUTION OF THE WASHING STATION MOTOR

95

6.6.4

WASHING STATION ASSEMBLY - MECHANICAL ADJUSTMENT

96

6.6.5

WASHING STATION - FUNCTIONAL CHECK

97

6.7

WASH PUMP ASSEMBLY

98

6.7.1

SUBSTITUTION OF THE WASH PUMP ASSEMBLY

98

6.7.2

SUBSTITUTION OF PART(S)

101

6.7.3

FUNCTION CHECK FOR THE WASH PUMP ASSEMBLY

101

6.8

PERISTALTIC PUMP ASSEMBLY

103

6.8.1

SUBSTITUTION OF THE PERISTALTIC PUMP ASSEMBLY

103

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 2

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.8.2

SUBSTITUTION OF THE PERISTALTIC PUMP TUBES

104

6.8.3

SUBSTITUTION OF THE PERISTALTIC PUMP MOTOR

106

6.8.4

FUNCTION CHECK FOR THE PERISTALTIC PUMP

106

6.9

SUBSTITUTION OF THE ELECTRONICS BOARDS

107

6.9.1

ANALYTICAL CONTROL BOARD Cd. 30-01283-01

107

6.9.2

STEPPER MOTORS DRIVER BOARD

108

6.9.3

ARM INTERFACE BOARD

Cd. 30-01282-01

109

6.9.4

PLATE INTERFACE BOARD

Cd. 30-01281-01

110

Ellipse

Service Manual

Cd. 30-01284-01

Rev. 04

01 October 2009

Page 3

Chapter 06 – SETTINGS AND ADJUSTMENTS

6

SETTINGS AND ADJUSTMENTS

6.1 SAMPLING ARM
6.1.1 ALIGNMENT AND ADJUSTMENTS
N.B.: Make sure that the instrument (Ellipse) is turned on before performing this procedure.
1. Launch the “Diagnostic” program.
2. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
3. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal),
Reactions, Reagents and Samples Plates light up in green.
4. Make sure that the Sampling Probe is well-placed on the Sampling Arm (see par. 6.1.2).
5. Make sure that the Sampling Probe is centered with respect to the Washing Well (note: the
centering of the Sampling Probe with respect to the Washing Well must be checked on the
entire stroke of the Sampling Arm). If it not be so, adjust in the following way:
a. remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1);
b. loosen the set screw of the Horizontal Arm Home Sensor flat bar (Fig. 2);
c. rotate the flat bar in the right toward in order to obtain the centering;
d. tighten the set screw of the Horizontal Arm Home Sensor flat bar;
e. select “Arm” folder and click the “Reset” button. Wait until the reset arm procedure has been
completed;
f. verify the accurate centering of the Sampling Arm with respect to the Washing Well. If it not
be so, repeat the steps from 5.b to 5.f .

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 4

Chapter 06 – SETTINGS AND ADJUSTMENTS

Sampling Arm
protective cover
Washing Well
Diluter protective
cover

Sampling Arm
protective cover
Sampling
fastening Arm
screw

Diluter protective
cover fastening screw

Fig. 1 – Protective covers

Slit

Horizontal Arm
Home Sensor

Horizontal Arm Home
Sensor flat bar

Horizontal Arm Home
Sensor fastening screw

Set screw of the Horizontal
Arm Home Sensor flat bar

Preheater fastening screw

Fig. 2 – Horizontal Arm Home Sensor
6. verify the verticality of the Sampling Probe with respect to the Washing Well. If it not be so,
adjust in the following way:

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 5

Chapter 06 – SETTINGS AND ADJUSTMENTS

a. loosen slightly the Screw “A” in order to obtain a longitudinal motion of the Sampling Arm
head (Fig. 3) and/or loosen slightly the Screw “B” (or “Preheater Head fastening screw”) in
order to obtain a latitudinal motion of the Sampling Arm head (Fig. 4);
b. center the Sampling Probe with respect to the Washing Well (checking its verticality for the
entire stroke of the Sampling Arm);
c. tighten the Screw “A” and/or the Screw “B”;
d. verify the accurate verticality of the Sampling Probe with respect to the Washing Well. If it
not be so, repeat the steps from 6.a to 6.d .

Screw “A”

Longitudinal motion

Fig. 3 – Longitudinal motion of the Sampling Arm head

Latitudinal motion

Screw “B”

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 6

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 4 – Latitudinal motion of the Sampling Arm head
7. Make sure that the distance between the top edge of the Washing Well and the tip of the
Sampling Probe is 25 mm.
8. Adjust the Vertical Arm Home Sensor loosing its two fastening screws (Fig. 5).
9. Tighten the two Vertical Arm Home Sensor fastening screws (Fig. 5).
10. Select “Arm” folder and click the “Down Well” button. Wait until the Sampling Probe go
down on the Washing Well.
11.

Click the “Home” button. Wait until the vertical reset arm procedure has been completed.

Vertical Arm
Home Sensor
Vertical Arm Home Sensor
fastening screw
Vertical Arm Home Sensor
anchor bar
Fig. 5 – Vertical Arm Home Sensor

12.

Make sure that the distance between the top edge of the Washing Well and the tip of the
Sampling Probe is 25 mm. If it not be so, repeat the above steps from 8 to 12.

13.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.

14.

Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal),
Reactions, Reagents and Samples Plates light up in green.

15.

Make sure that the Reagents protection cover is permanently locked on the Reagents Plate
(Fig. 6).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 7

Chapter 06 – SETTINGS AND ADJUSTMENTS

16.

Select “Arm” folder.

17.

Click the "Go Std" button, then click the correspondent “Down” button. The arm will move to
position #1 of the Standards Plate, then it will go down until to reach the right height.

18.

Make sure that the Sampling Probe is vertically aligned and centered with respect to position
#1 of the Standards Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
adjust (if necessary) the verticality by using the adjustable anchored bearing (Fig. 7), click the
‘Save’ button, then repeat the above steps from 6.

Reagents protection cover

Fig. 6 – Reagents protection cover on the Reagents Plate

Adjustable
anchored bearing

Horizontal Sampling
Arm Motor

Peristaltic Pump
Motor

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 8

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 7 – Adjustable anchored bearing
19.

Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move
to position #1 of the Reagents Plate, then it will go down until to reach the right height.

20.

Make sure that the Sampling Probe is vertically aligned and centered with respect to position
#1 of the Reagents Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
then click the ‘Save’ button.

21.

Click the "Go Sample" button, then click the correspondent “Down” button. The arm will
move to position #1 of the Sample Rack #1, then it will go down until to reach the right
height.

22.

Make sure that the Sampling Probe is vertically aligned and centered with respect to position
#1 of the Sample Rack #1. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
then click the ‘Save’ button.

23.

Click the "Go Disp." button, then click the correspondent “Down” button. The arm will move
to cuvette #81 of the Reactions Plate, then it will go down until to reach the right height.

24.

Make sure that the Sampling Probe is vertically aligned and centered with respect to cuvette
#81 of the Reactions Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
then click the ‘Save’ button.

25.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general
reset procedure has been completed.

26.

Place the Sampling Arm protective cover screwing its fastening screw (Fig. 1).

N.B.: in case of not enough current centered of the Sampling Probe with respect to the
Washing Well and with respect to the cuvette #81 of the Reactions Plate, repeat the
"Alignment and Adjustments” procedure privileging the centering of the Sampling Probe with
respect to the cuvette #81 of the Reactions Plate.
6.1.2 REPLACEMENT OF THE SAMPLING PROBE
1. Removing of the Sampling Probe:
a. launch the “Diagnostic” program;
b. select "Miscellaneous” folder and request an Instrument General Reset. Wait until the
general reset procedure has been completed;

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 9

Chapter 06 – SETTINGS AND ADJUSTMENTS

c. remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut
(Fig. 8).
2.

Replacement of the Sampling Probe:

a. launch the “Diagnostic” program;
b. select "Miscellaneous” folder and request an Instrument General Reset. Wait until the
general reset procedure has been completed;
c. attach the (straight) Sampling Probe to the Probe holder of the Sampling Arm (Fig. 9 – Step
1);
d. insert the Sampling Probe in the Probe holder and slowly rotate (counter clockwise) the
Sampling Probe ring nut (Fig. 9 – Step 2);
e. keep on rotating the Sampling Probe ring nut until you feel the stop (taking it around
25 mm from the ring nut) (Fig. 9 – Step 3). This is to allow a perfect contact between
Sampling Probe and the Probe holder;
f.

make sure that the Sampling Probe is snug (Fig. 9 – Step 4);

g. take the dynamometric key (Fig. 10) and place the head around the ring nut of the Sampling
Probe (Fig. 11);
h. screw (counter clockwise) the Sampling Probe with the dynamometric key (Fig. 12) until the
correct torque is reached and the key spins freely around the Sampling Probe;
Sampling Arm
head

Sampling Probe
ring nut

CW
Sampling Probe

Fig. 8 - Removing of the Sampling Probe
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 10

Chapter 06 – SETTINGS AND ADJUSTMENTS

Sampling Arm
head

Step 1
Step 2

Probe holder

Sampling Probe
ring nut

CCW

Sampling Probe

Step 3
Step 4

Sampling Probe
ring nut

CCW

Sampling
Probe

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 11

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 9 - Replacement of the Sampling Probe
i. remove the dynamometric key;
j. turn in the “Diagnostic” program;
k. select "Miscellaneous” folder and request an Instrument General Reset. Wait until the
general reset procedure has been completed;
l. make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal),
Reactions, Reagents and Samples Plates light up in green.

N.B.: after having replaced the Sampling Probe, it is necessary to verify the verticality and the
centering of it with respect to the Washing Well. In case of misalignment: perform the
procedure "Alignment and Adjustments”, described into the previous Section 6.1.1, from step
5.

Dynamometric key
head

Fig. 10 – Dynamometric key

Dynamometric key
CCW
Sampling Probe
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 12

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 11 - Dynamometric key head
around the ring nut

Fig. 12 – Dynamometric key rotation

6.1.3 REPLACEMENT OF THE SAMPLING ARM
1. Remove the Sampling Probe from its housing unscrewing (clockwise) its ring nut (Fig. 8).
2. Make sure that the instrument (Ellipse) is turned off.
3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13).
4. Unplug the J1, J2, J5, J6, J13 connectors from the Arm Interface Board (Fig. 14), remove the
Sampling Arm protective cover unscrewing its fastening screw (Fig. 1) and unplug the ground
wire from the Sampling Arm Assembly.
5. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1).
6. Unscrew the two anchored screws (Fig. 15) and pull out the Diluter Module from its hole.
7. Remove the Preheater teflon tube from the manifold of the Diluter Solenoid Valve by
unscrewing the fitting (Fig. 16), paying special attention to liquid leakage.
8. Remove the Preheater clamp, located into the Peristaltic Pump housing, unscrewing its
fastening screw (Fig. 17).
9. Unscrew the two Preheater fastening screws (Fig. 2), loosen the “Screw “B” (Fig. 4) and
remove the Level Sensor and Preheater Assembly from its housing.
10.

Unscrew the two Sampling Arm anchored screws (Fig. 18), pull out the Sampling Arm from
its housing and substitute it with the new one.
To replace the new Sampling Arm, perform the following steps:

11.

Repeat the above steps in inverse order: from 10 to 4.

12.

Perform the procedure "Alignment and Adjustments” described into the previous Section 6.1.1.

13.

Select “Diluter” folder and perform - with the selection of the “Probe Wash Test” button - at
least five successive Probe Wash Test to fill the hydraulic circuit with the liquid.

14.

Verify the absence of liquid leakage from the hydraulic circuit.

15.

Place the instrument front panel screwing its four anchored screws (Fig. 13).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 13

Chapter 06 – SETTINGS AND ADJUSTMENTS

Instrument
front panel

Front panel
anchored screw

Fig. 13 - Front panel anchored screws

Horizontal Mov.
Arm Motor – J1

Vertical Mov. Arm Motor – J2

Vertical Mov. Arm
Home Sensor – J6

Horizontal Mov. Arm
Home Sensor – J5

Preheater and
Level Sensor Assy – J13

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 14

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 14 - Arm Interface Board (Dx) [P/N: 30-01282-01]

Diluter anchored screw

Diluter

Diluter Solenoid
Valve
Manifold

Fig. 15 - Diluter Module (top view)
Fitting

Preheater teflon tube

Fig. 16 – Diluter fitting

Peristaltic Pump
housing

Preheater clamp
Preheater clamp
fastening screw

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 15

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 17 – Preheater clamp

Ground wire

Sampling Arm

Sampling Arm
anchored screw

Fig. 18 – Sampling Arm anchored screws
6.1.4 REPLACEMENT OF THE ARM HOME SENSORS
6.1.4.1 Replacement of the Horizontal Arm Home Sensor
N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure.
1. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1).
2. Loosen the set screw of the Horizontal Arm Home Sensor flat bar and turn the flat bar in order
to unscrew the two Horizontal Arm Home Sensor fastening screws (Fig. 2).
3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13).
4. Unplug the J5 connector from the Arm Interface Board (Fig. 14), take out the entire Arm
Home Sensor Assy and substitute it with the new one (Fig. 19).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 16

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 19 - Arm Home Sensor Assy [P/N: 9-10-0023-60]
To replace the new (Horizontal) Arm Home Sensor Assy, perform the following steps:
5. Repeat steps 4 and 3.
6. Turn on the instrument Ellipse.
7. Launch the “Diagnostic” program.
8. Center (manually) the Sampling Probe with respect to the Washing Well.
9. Position the slit of the Horizontal Arm Home Sensor flat bar (Fig. 2) to the center of the
Horizontal Arm Home Sensor.
10.

Tighten the set screw of the Horizontal Arm Home Sensor flat bar (Fig. 2).

11.

Select “Arm” folder and click the “Reset” button. Wait until the reset arm procedure has been
completed.

12.

Make sure that the Horizontal Arm Home Sensor light up in green.

13.

Make sure that the Sampling Probe is centered with respect to the Washing Well. If it not be
so, adjust in the following way:
a. loosen the set screw of the Horizontal Arm Home Sensor flat bar (Fig. 2);

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 17

Chapter 06 – SETTINGS AND ADJUSTMENTS

b. turn slightly the Horizontal Arm Home Sensor flat bar to the right toward in order to obtain
the centering;
c. tighten the set screw of the Horizontal Arm Home Sensor flat bar (Fig. 2);
d. select “Arm” folder and click the “Reset” button. Wait until the reset arm procedure has been
completed;
e. make sure that the Horizontal Arm Home Sensor light up in green;
f. verify that the Sampling Probe is centered with respect to the Washing Well. If it not be so,
repeat steps from 13.a to 13.f .
14.

perform the procedure "Alignment and Adjustments”, described into the previous
Section 6.1.1, from step 13 to step 26.

6.1.4.2 Replacement of the Vertical Arm Home Sensor
N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure.
1. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1).
2. Unscrew the two Vertical Arm Home Sensor fastening screws (Fig. 5).
3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13).
4. Unplug the J6 connector from the Arm Interface Board (Fig. 14), take out the entire Arm
Home Sensor Assy and substitute it with the new one (Fig. 19).
To replace the new (Vertical) Arm Home Sensor Assy, perform the following steps:
5. Repeat steps 4, 3 and 2.
6. Turn on the instrument Ellipse.
7. Launch the “Diagnostic” program.
8. Select “Arm” folder and click the “Reset” button. Wait until the reset arm procedure has been
completed.
9. Make sure that the Vertical Arm Home Sensor light up in green.
10.

Make sure that the distance between the top edge of the Washing Well and the tip of the
Sampling Probe is 25 mm. If it not be so, adjust in the following way:
a. loosen the two Vertical Arm Home Sensor fastening screws (Fig. 5);
b. move slightly the Vertical Arm Home Sensor to the right vertically toward in order to obtain
the centering;
c. tighten the two Vertical Arm Home Sensor fastening screws (Fig. 5);

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 18

Chapter 06 – SETTINGS AND ADJUSTMENTS

d. select “Plate” folder and click “All Motors OFF” button;
e. move downward (manually) the Sampling Arm;
f. click “All Motors ON” button in the “Plate” folder. Wait until the reset instrument procedure
has been completed;
g. verify that the distance between the top edge of the Washing Well and the tip of the
Sampling Probe is 25 mm. If it not be so, repeat steps from 10.a to 10.g .
11.

Place the Sampling Arm protective cover screwing its fastening screw (Fig. 1).

6.1.5

REPLACEMENT OF THE SAMPLING ARM MOTORS

6.1.5.1

Replacement of the Horizontal Sampling Arm Motor

N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure.
1. Remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut
(Fig. 8).
2. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1).
3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13).
4. Unplug the J1, J2, J5, J6, J13 connectors from the Arm Interface Board (Fig. 14) and the
ground wire from the Sampling Arm Assembly (Fig. 18).
5. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1).
6. Unscrew the two anchored screws (Fig. 15) and pull out the Diluter Module from its hole.
7. Remove the Preheater teflon tube from the manifold of the Diluter Solenoid Valve by
unscrewing the fitting (Fig. 16), paying special attention to liquid leakage.
8. Remove the Preheater clamp, located into the Peristaltic Pump housing, unscrewing its
fastening screw (Fig. 17).
9. Unscrew the two Preheater fastening screws (Fig. 2), loosen the “Screw “B” (Fig. 4) and
remove the Level Sensor and Preheater Assembly from its housing.
10.

Unscrew the two Sampling Arm anchored screws (Fig. 18) and pull out the Sampling Arm
from its housing.

11.

Remove the four Horizontal Sampling Arm Motor fastening screws (Fig. 20).

12.

Remove the belt from the Horizontal Sampling Arm Motor pulley (Fig. 21).

13.

Pull out the Horizontal Sampling Arm Motor from its housing.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 19

Chapter 06 – SETTINGS AND ADJUSTMENTS

Horizontal Sampling
Arm Motor
fastening screw
Horizontal Sampling
Arm Motor

Fig. 20 – Horizontal Sampling Arm Motor

14.

Loosen the two set screws of the Horizontal Sampling Arm Motor pulley (Fig. 21), then extract
the pulley to the motor axis.

15.

Unscrew the four fastening screws of the Horizontal Sampling Arm Motor anchor plate
(Fig. 21), then take out the anchor plate to the motor.

16.

Insert the pulley to the motor axis of the new Arm Motor Assy (Fig. 23).

17.

Position the pulley to 2.00 mm with respect to upper motor plane (Fig. 22), then tighten the
two set screws of the pulley (Fig. 21).

18.

Position correctly the anchor plate on the upper plane of the new motor (Fig. 21), then screw
the four anchor plate fastening screws (Fig. 21).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 20

Chapter 06 – SETTINGS AND ADJUSTMENTS

Upper motor plane
Pulley
Pulley

Set screw

Anchor plate

Anchor plate
fastening screw
Fig. 21 - Horizontal Sampling Arm
Motor anchor plate

19.

Fig. 22 - Horizontal Sampling Arm
Motor scheme

Replace the Horizontal Sampling Arm Motor into its Sampling Arm housing, then insert
(without tighten) the four Horizontal Sampling Arm Motor fastening screws (Fig. 20).

20.

Wrap the belt around the Horizontal Sampling Arm Motor pulley.

21.

Push the motor outward from the arm axis and holding it in this position, so as to maintain the
belt in tension, then tighten the four Horizontal Sampling Arm Motor fastening screws
(Fig. 27).

22.

To replace the Sampling Arm, repeat the above steps in inverse order: from 10 to 1.

N.B.: after having replaced the Horizontal Sampling Arm Motor, it is necessary to align
mechanically the Sampling Probe with respect to the Washing Well and center all the other
positions by following the procedure "Alignment and Adjustments” described into the
previous Section 6.1.1

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 21

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 23 – Arm Motor Assy [P/N: 10-00399-01]

6.1.5.2

Replacement of the Vertical Sampling Arm Motor

N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure.
1. Remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut
(Fig. 8).
2. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1).
3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13).
4. Unplug the J2 connector from the Arm Interface Board (Fig. 14).
5. Loosen the four fastening screws of the belt tension plate that permit to adjust the belt tension
for the vertical Sampling Probe movement (Fig. 24).
6. Remove the belt from the Vertical Sampling Arm Motor pulley (Fig. 25).
7. Unscrew the four Vertical Sampling Arm Motor fastening screws (Fig. 25).

Belt
Belt tension plate

Fastening screw
Fastening screw
Vertical Sampling
Arm Motor pulley
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 22

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 24 – Belt tension plate

Fig. 25 – Vertical Sampling Arm Motor
fastening screws

8. Pull out the Vertical Sampling Arm Motor from its housing, then remove the anchor plate to
the motor.
9. Loosen the two set screws of the Vertical Sampling Arm Motor pulley (Fig. 26), then extract
the pulley to the motor axis.
10.

Insert the pulley to the motor axis of the new Arm Motor Assy (Fig. 23).

11.

Position the pulley to 1.50 mm with respect to upper motor plane (Fig. 26), then tighten the
two set screws of the pulley (Fig. 26).

Set screw

Pulley

Upper motor plane

Fig. 26 – Vertical Sampling Arm Motor scheme
12.

Position the anchor plate on the upper plane of the new motor, then replace the Vertical
Sampling Arm Motor into its Sampling Arm housing (Fig. 25).

13.

Tighten the four Vertical Sampling Arm Motor fastening screws (Fig. 25).

14.

Wrap the belt around the Vertical Sampling Arm Motor pulley.

15. Push upwards the belt tension plate (Fig. 24) and holding it in this position, so as to maintain
the belt in tension, then tighten its four fastening screws (Fig. 24).
16.

To complete, repeat the above steps in inverse order: from 4 to 1.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 23

Chapter 06 – SETTINGS AND ADJUSTMENTS

N.B.: after having replaced the Vertical Sampling Arm Motor, it is necessary to require a
Instrument General Reset and perform the procedure "Alignment and Adjustments”,
described into the previous Section 6.1.1, from step 7.

6.1.6 REPLACEMENT OF THE SAMPLING ARM BELTS
6.1.6.1 Replacement of the Horizontal Sampling Arm Belt
N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure.
1. Remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut
(Fig. 8).
2. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1).
3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 13).
4. Unplug the J1, J2, J5, J6, J13 connectors from the Arm Interface Board (Fig. 14) and the
ground wire from the Sampling Arm Assembly (Fig. 18).
5. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1).
6. Unscrew the two anchored screws (Fig. 15) and pull out the Diluter Module from its hole.
7. Remove the Preheater teflon tube from the manifold of the Diluter Solenoid Valve by
unscrewing the fitting (Fig. 16), paying special attention to liquid leakage.
8. Remove the Preheater clamp, located into the Peristaltic Pump housing, unscrewing its
fastening screw (Fig. 17).
9. Unscrew the two Preheater fastening screws (Fig. 2), loosen the “Screw “B” (Fig. 4) and
remove the Level Sensor and Preheater Assembly from its housing.
10.

Unscrew the two Sampling Arm anchored screws (Fig. 18) and pull out the Sampling Arm
from its housing.

11.

Loosen the four Horizontal Sampling Arm Motor fastening screws (Fig. 27).

12.

Remove the belt from the Horizontal Sampling Arm Motor pulley (Fig. 21) from the horizontal
handling pulley, then substitute it with the new one.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 24

Chapter 06 – SETTINGS AND ADJUSTMENTS

Horizontal Sampling Arm
Motor fastening screw

Belt

Horizontal handling pulley

Fig. 27 – Horizontal Sampling Arm Belt
13.

Wrap the new belt over the Horizontal Sampling Arm Motor pulley and over the horizontal
handling pulley (Fig. 27).

14.

Push the motor outward from the arm axis and holding it in this position, so as to maintain the
belt in tension, then tighten the four Horizontal Sampling Arm Motor fastening screws
(Fig. 27).

15.

To replace the Sampling Arm, repeat the above steps in inverse order: from 10 to 1.

N.B.: after having replaced the Horizontal Sampling Arm Belt, it is necessary to align
mechanically the Sampling Probe with respect to the Washing Well and center all the other
positions by following the procedure "Alignment and Adjustments” described into the
previous Section 6.1.1 .

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 25

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.1.6.2 Replacement of the Vertical Sampling Arm Belt
N.B.: make sure that the instrument (Ellipse) is turned off before performing this procedure.
1. Remove the Sampling Probe from its housing by unscrewing (clockwise) its ring nut
(Fig. 8).
2. Remove the Sampling Arm protective cover unscrewing its fastening screw (Fig. 1).
3. Loosen the four fastening screws of the belt tension plate that permit to adjust the belt tension
for the vertical Sampling Probe movement (Fig. 28).
4. Unscrew the four fastening screws (Fig. 28) of the Vertical Arm Home Sensor anchor bar
(Fig. 5) and remove it to the Sampling Arm.

Fastening screw of the
belt tension plate

Belt tension plate

Fastening screw of the Vertical
Arm Home Sensor anchor bar

Fig. 28 – Fastening screw of the Vertical
Arm Home Sensor anchor bar
5. Unscrew the two clamping belt fastening screw, remove the two small plates (Fig. 29), take off
the belt to the Sampling Arm, then substitute it with the new one.
Ellipse

Service Manual

Rev. 04

01 October 2009

Page26

Chapter 06 – SETTINGS AND ADJUSTMENTS

Belt

Two small plates

Clamping belt
fastening screw

Fig. 29 –Vertical Sampling Arm Belt

6. To replace the new Vertical Sampling Arm Belt, repeat the above steps in inverse order:
from 5 to 1.
N.B.: after having replaced the Vertical Sampling Arm Belt, it is necessary to require a
Instrument General Reset and perform the procedure "Alignment and Adjustments”,
described into the previous Section 6.1.1, from step 7.

6.2
6.2.1

DILUTER MODULE
REPLACEMENT OF THE DILUTER MODULE

1. Make sure the instrument Ellipse is turned off.
2. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1).
3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13).
4. Take off the Rinse Solution cannula (I) from its bottle.
5. Unplug the J4, J8, J16, J17 connectors from the Arm Interface Board (Fig. 30) and the ground
wire from the Diluter Module.
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 27

Chapter 06 – SETTINGS AND ADJUSTMENTS

Preheater and
Level Sensor Assy – J13
Diluter Motor - J4

Diluter Home Sensor – J8

Diluter Electrovalve – J16

Diluter Micropump – J17

Fig. 30 - Arm Interface Board (Dx) [P/N: 30-01282-01]
6. Loosen the two anchored screws (Fig. 31) and pull out the Diluter Module (Fig. 32) from its
hole, making attention not to crash the tubes of the hydraulic circuit.
7. Unscrew the two fitting from the Diluter head (Fig. 31), making attention to the spillage of the
Rinse Solution from the two tubes.
8. Take off the Diluter Module cables from its hole.
9. Unscrew the four Diluter fastening screws (Fig. 31) and pull out the Diluter from its
mechanical support.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page28

Chapter 06 – SETTINGS AND ADJUSTMENTS

Diluter Module
anchored screw
Diluter
fastening screw

Diluter fitting

Fig. 31 – Diluter Module (top view)
10.

Remove the four Diluter fastening screw washer.

11.

Replace the Diluter with one new Diluter.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 29

Chapter 06 – SETTINGS AND ADJUSTMENTS

Diluter Home Sensor
fastening screw

Diluter Home Sensor
[P/N: 9-10-0024-52]

Diluter Motor
[P/N: 10-00399-40]

Fig. 32 – Diluter Module [P/N: 10-00496-01]

12.

To replace, repeat the above steps in inverse order: from 10 to 4.

N.B.: make sure that the two fitting are very screws on the Diluter head in order to guarantee
the hydraulic seal of the circuit. Moreover, make sure that the Diluter Module cables in the
instrument do not hinder the movement of the Pre-Heater in its zone of job.
13.

Turn on the instrument Ellipse.

14.

Make sure that sufficient liquid in the Rinse bottle is present.

15.

Launch the “Diagnostic” program.

16.

Select “Diluter” folder and click the “Home” button. Wait until the reset Diluter procedure has
been completed.

17.

Set up the aspiration volume to the maximum value (980 µl) and click the “Asp” button. Make
sure that the diluter executes the aspiration procedure correctly.

18.

Click the “Disp” button. Make sure that the diluter executes the dispenses procedure correctly.

19.

Click the “Asp” button with the aspiration volume to the maximum value (980 µl).

20.

Click the “Probe Wash Test” button.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 30

Chapter 06 – SETTINGS AND ADJUSTMENTS

N.B.: before proceeding with the Probe Wash Test, make sure that one written on the warning
message is respected.
21.

Click the “OK” button. Wait until the Probe Wash Test has been completed.

22.

Repeat the above steps 20 and 21 at least others five times in order to fill up the hydraulic
circuit completely.

23.

Make sure that inside the Diluter are not present of the air bubbles. If it not be so, repeat the
Probe Wash Test (steps 20 and 21) until the elimination of the air bubbles.

24.

Verify the absence of liquid leakage from the hydraulic circuit.

25.

Click the “Home” button. Wait until the reset Diluter procedure has been completed.

26.

Place the instrument front panel screwing its four anchored screws (Fig. 13).

27.

Place the Diluter protective cover screwing its two fastening screws (Fig. 1).

N.B.: the following steps allows to perform the procedure "Backlash Diluter”. This procedure
in order to recover the sampling imprecision generated by the Diluter Motor when it inverts
its rotation.
28.

Remove the Reagents protection cover on the Reagents Plate and take out the Samples Racks.

29.

Select “Arm” folder and click the “Reset” button. Wait until the reset Sampling Arm procedure
has been completed.

30.

Click the "Go Sample" button. The arm will move to position #1 of the Sample Rack #1.

31.

Put a Sample cup containing distilled water (about 1 ml) under the Sampling Probe.

32.

Select “Diluter” folder, set up the aspiration volume to the 300 µl and click the “Asp” button.
Wait until the Diluter has been aspirated 300 µl of distilled water from the Sample cup.

33.

Remove the Sample cup containing distilled water.

34.

Set up the dispensation volume to the 1 µl and repeatedly click the “Disp” button until water
appears on the tip of the Sampling Probe.

35.

Subtract 1 from the number of times in which it has been selected the “Disp” button. The
obtained number is the Backlash Diluter value.

36.

Select "Configuration" folder, enter the password (“1234”) in the appropriate window and
click the “OK” button.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 31

Chapter 06 – SETTINGS AND ADJUSTMENTS

37.

Select the Backlash Diluter value in the “Backlash Dil.” Pull Down Menu and click the “Save”
button. Wait until the Backlash Diluter value has been saved.

38.

Select “Arm” folder and click the “Reset” button. Wait until the reset Sampling Arm procedure
has been completed.

39.

Put on the Samples Racks and place the Reagents protection cover on the Reagents Plate.

6.2.2 REPLACEMENT OF THE DILUTER MICROPUMP
1. Make sure the instrument Ellipse is turned off.
2. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1).
3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13).
4. Take off the Rinse Solution cannula (I) from its bottle.
5. Unplug the J4, J8, J16, J17 connectors from the Arm Interface Board (Fig. 30) and the ground
wire from the Diluter Module.
6. Loosen the two anchored screws (Fig. 31) and pull out the Diluter Module (Fig. 32) from its
hole, making attention not to crash the tubes of the hydraulic circuit.
7. Take off the Diluter Module cables from its hole.
8. Remove the two tubes from Micropump (Fig. 33).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 32

Chapter 06 – SETTINGS AND ADJUSTMENTS

Diluter Electrovalve
[P/N: 05-00700-48]

Manifold
Fitting
Diluter Micropump
[P/N: 05-01710-40]

Fig. 33 – Diluter Module: Electrovalve and Micropump

Diluter Electrovalve
fastening screws
Electrovalve and
Micropump
mechanical support
Diluter Micropump
fastening screws

Fig. 34 – Electrovalve and Micropump mechanical support
9. Unscrew the two Micropump fastening screws (Fig. 34) and pull out the Micropump from its
mechanical support.
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 33

Chapter 06 – SETTINGS AND ADJUSTMENTS

10.

Replace the Micropump with one new Micropump.

11.

To replace, repeat the above steps in inverse order: from 9 to 4.

12.

Turn on the instrument Ellipse.

13.

Make sure that sufficient liquid in the Rinse bottle is present.

14.

Launch the “Diagnostic” program.

15.

Select “Diluter” folder and click the “Home” button. Wait until the reset Diluter procedure has
been completed.

16.

Set up the aspiration volume to the maximum value (980 µl) and click the “Asp” button. Make
sure that the Diluter executes the aspiration procedure correctly.

17.

Click the “Disp” button. Make sure that the Diluter executes the dispenses procedure correctly.

18.

Click the “Asp” button with the aspiration volume to the maximum value (980 µl).

19.

Click the “Probe Wash Test” button.

N.B.: before proceeding with the Probe Wash Test, make sure that one written on the
warning message is respected.
20.

Click the “OK” button. Wait until the Probe Wash Test has been completed.

21.

Repeat the above steps 20 and 21 at least others five times in order to fill up the hydraulic
circuit completely.

22.

Make sure that inside the Diluter are not present of the air bubbles. If it not be so, repeat the
Probe Wash Test (steps 20 and 21) until the elimination of the air bubbles.

23.

Verify the absence of liquid leakage from the hydraulic circuit.

24.

Click the “Home” button. Wait until the reset Diluter procedure has been completed.

25.

Place the instrument front panel screwing its four anchored screws (Fig. 13).

26.

Place the Diluter protective cover screwing its two fastening screws (Fig. 1).

27.

Select "Configuration" folder, enter the password (“1234”) in the appropriate window and
click the “OK” button.

28.

Select the value 0 in the “Level” Pull Down Menu.

29.

Click the “Run Test” button. Wait until the Washing Well level Test has been completed.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 34

Chapter 06 – SETTINGS AND ADJUSTMENTS

30.

Click the “OK” button in the appropriate window in order to save the new value. Wait until the
new value has been saved.

6.2.3

REPLACEMENT OF THE DILUTER ELECTROVALVE

1. Make sure the instrument Ellipse is turned off.
2. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1).
3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13).
4. Take off the Rinse Solution cannula (I) from its bottle.
5. Unplug the J4, J8, J16, J17 connectors from the Arm Interface Board (Fig. 30) and the ground
wire from the Diluter Module.
6. Loosen the two anchored screws (Fig. 31) and pull out the Diluter Module (Fig. 32) from its
hole, making attention not to crash the tubes of the hydraulic circuit.
7. Take off the Diluter Module cables from its hole.
8. Unscrew the two Electrovalve fastening screws (Fig. 34) and pull out the Electrovalve from its
mechanical support.
9. Remove the two tubes from Electrovalve (Fig. 33).
10.

Unscrew the two fittings from Electrovalve.

11.

Screw the two fittings on the new Electrovalve.

N.B.: make sure that the two fitting are very screws on the new Electrovalve in order to
guarantee the hydraulic seal of the circuit.
12.

To replace, repeat the above steps in inverse order: from 9 to 4.

13.

Turn on the instrument Ellipse.

14.

Make sure that sufficient liquid in the Rinse bottle is present.

15.

Launch the “Diagnostic” program.

16.

Select “Diluter” folder and click the “Home” button. Wait until the reset Diluter procedure has
been completed.

17.

Set up the aspiration volume to the maximum value (980 µl) and click the “Asp” button. Make
sure that the Diluter executes the aspiration procedure correctly.

18.

Click the “Disp” button. Make sure that the Diluter executes the dispenses procedure correctly.

19.

Click the “Asp” button with the aspiration volume to the maximum value (980 µl).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 35

Chapter 06 – SETTINGS AND ADJUSTMENTS

20.

Click the “Probe Wash Test” button.

N.B.: before proceeding with the Probe Wash Test, make sure that one written on the
warning message is respected.
21.

Click the “OK” button. Wait until the Probe Wash Test has been completed.

22.

Repeat the above steps 20 and 21 at least others five times in order to fill up the hydraulic
circuit completely.

23.

Make sure that inside the Diluter are not present of the air bubbles. If it not be so, repeat the
Probe Wash Test (steps 20 and 21) until the elimination of the air bubbles.

24.

Verify the absence of liquid leakage from the hydraulic circuit.

25.

Click the “Home” button. Wait until the reset Diluter procedure has been completed.

26.

Place the instrument front panel screwing its four anchored screws (Fig. 13).

27.

Place the Diluter protective cover screwing its two fastening screws (Fig. 1).

6.2.4

REPLACEMENT OF THE DILUTER HOME SENSOR

1. Make sure the instrument Ellipse is turned off.
2. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 1).
3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13).
4. Unplug the J4, J8, J16, J17 connectors from the Arm Interface Board (Fig. 30) and the ground
wire from the Diluter Module.
5. Loosen the two anchored screws (Fig. 31) and pull out the Diluter Module (Fig. 32) from its
hole, making attention not to crash the tubes of the hydraulic circuit.
6. Take off the Diluter Module cables from its hole.
7. Unscrew the two Diluter Home Sensor fastening screws (Fig. 32) and pull out the Home
Sensor from its hole within the Diluter.
8. Replace the Home Sensor with one new Home Sensor (Fig. 35).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 36

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 35 – Diluter Home Sensor Assy [P/N: 9-10-0024-52]
9. To replace, repeat the above steps in inverse order: from 7 to 4.
10.

Turn on the instrument Ellipse.

11.

Launch the “Diagnostic” program.

12.

Select “Diluter” folder and click the “Home” button. Wait until the reset Diluter procedure has
been completed.

13.

Set up the aspiration volume to the maximum value (980 µl) and click the “Asp” button. Make
sure that the Diluter Home Sensor on the “Diagnostic” program becomes of red color (outside
of the home position).

14.

Click the “Home” button. Wait until the reset Diluter procedure has been completed.

15.

Make sure that the Diluter Home Sensor on the “Diagnostic” program becomes of green color
(home position).

16.

Place the instrument front panel screwing its four anchored screws (Fig. 13).

17.

Place the Diluter protective cover screwing its two fastening screws (Fig. 1).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 37

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.3 PRE-HEATER AND LEVEL SENSOR
6.3.1 SUBSTITUTION OF THE PRE-HEATER AND LEVEL SENSOR ASSEMBLY
1. Turn on the Ellipse system (instrument and computer).
2. Launch the Diagnostic program, select the “Diluter” function and make two or three
“Probe Wash Test” cycles with the Rinse solution cannula out of his bottle, in order to
completely empty the Rinse hydraulic line.
3. Exit from the diagnostic program, pressing the "Diagnostic" key and exit from the analyzer
program pressing “Shutdown” key. Remove the sampling probe from its housing.
4. Unscrew the two fastening screws and remove the Diluter protective cover (Fig. 36).
5. Unscrew the two anchored screws and pull out the Diluter Module from his hole (Fig. 37).
6. Remove the Preheater teflon tube on the Diluter Solenoid Valve by unscrewing the fitting
(Fig. 38).
7. Remove the front panel unscrewing the four anchored screws (Fig. 40) and unplug the J13
connector from the Arm Interface board (Fig. 41).
8. Remove the Preheater clamp located into the Peristaltic Pump housing, unscrewing its
fastening screw (Fig. 44).
9. Unscrew the Preheater and Level sensor fastening screws, remove it from its housing and
substitute it with the new one (Fig. 42).
10.

Reinstall the new assembly, making the vertical alignment of the sampling probe, with
reference to the washing well and then screw the Preheater head and Level sensor fastening
screw (Fig. 39).

11.

After reinstalling the new assembly, repeat the above steps in inverse order: from step 9 to 1
and then, carry out the check procedure for the mechanical functioning described on the next
pages.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 38

Chapter 06 – SETTINGS AND ADJUSTMENTS

WARNING
Do the following operations only after substitution of the Level Sensor and Pre-heater assembly:
a) Remove the white wire from the teflon tube before connecting it to the diluter (the white wire
must be removed only at this stage of the procedure and not before). Be sure of having properly
connected the tube to the diluter.

b) Remove the protecting cap before positioning the sampling probe.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 39

Chapter 06 – SETTINGS AND ADJUSTMENTS

Sampling Arm
protective cover

Washing Well
Diluter protective cover

Fig. 36
6.3.2 LEVEL SENSOR – CHECK AND ADJUSTMENT PROCEDURE
1. Turn on the Ellipse system (instrument and computer) and launch the diagnostic program.
2. Select the "Configuration" function. Digit the password (1234) in the appropriate window and
press “OK”.
3. Press the "Wash. Well level test". By clicking on the “Run Test” button the system performs
an automatic verification and adjustment cycle for the liquid level in the washing well.
A number appears in the box “Level” that corresponds to the rotation time of the rinse pump
in milliseconds, that is necessary to maintain the correct liquid level in the washing well.
4. The functioning range is between 350 and 700 milliseconds. If the washing well is not
sufficiently filled, a message appears to warn the user to check the hydraulic line.
5. If the value taken during the test is different from the previous one, it is necessary to click on
“Save” in order to save the new data.
6. Make sure that no liquid leaks from the probe assembly and/or the dilutor. Make sure that the
liquid flows from the probe freely.
7. Press “Test ON” on the “Level sensor Test” folder, on the right of it will be showed a red
drop that becomes temporarily green when the liquid is detected or when the probe is touched
with your finger (Fig. 43).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 40

Chapter 06 – SETTINGS AND ADJUSTMENTS

Preheater
fastening screw

Anchored screw

Preheater Head
fastening screw
Solenoid Valve
fitting
Fig. 37 Diluter – Top view

Fig. 38 Dilutor

Fig. 39 Sampling Arm

Fig. 40 Instrument front view
Instrument front panel
anchored screw

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 41

Chapter 06 – SETTINGS AND ADJUSTMENTS

Level Sensor and Pre-Heater
assembly Cd 10-00659-00
Preheater & L.S. - J13

Fig. 41 Arm Interface board
Cd. 30-01282-01

Fig. 42 Level Sensor &
Preheater Assembly

Level Sensor Test

Fig. 43

8.

In order to more specifically verify the correct level-sensing functioning, proceed as follows:

a. Place a cup containing liquid in the first samples rack, position nr 1 (e.g . 500 µl) and launch
the diagnostic program.
b. Select the "Arm" function and press the "Go Sample" key in order to allow the probe to
move into the position corresponding to the sample cup.
c. In the window next to the "Go Level", enter the maximum descent of the probe (e.g.500).
d. Select "Go Level" and make sure that the probe drops down into the cup until it senses the
liquid. In the window to the right of the "Go Level" key the level reached by the probe will
appear, expressed in tenths of millimeters
e. Repeat step d) several times, each time returning to the Vertical "Home" position. Make sure
that the level indicated remains constant ≤ 10 (1 mm).
f. Repeat steps b) to e) for positions 8 of the first rack using the same sample cup. Make sure
that the difference between all the positions is ≤ 1mm.
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 42

Chapter 06 – SETTINGS AND ADJUSTMENTS

g. Repeat steps b) to f) for samples racks two, three and four.
h. Press the "Back to Wash Well" key and then select "Go Sample" in order to allow the probe
to return to a position in the rack, which does not contain any cup.
i. Select "Go Level" and make sure that the level reached is the maximum indicated in the
given window is 9999.
j. Select the "Reset" key, remove the cups from the racks and exit the program by pressing the
"Diagnostic" key.
k. To exit the analyzer program press “shutdown” key.

Preheater & L.S clamp

Fig. 44

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 43

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.4 OPTIC ASSEMBLY
6.4.1 SUBSTITUTION OPTIC ASSEMBLY
N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.

Stepper Motor assembly Cd. 10-00399-01

Opto coupler assy
Cd. 9-10-0023-00

Pre-Ampl/ADC (Read Channel)
Cd. 30-00107-00

Pre-Ampl/ADC (Ref. Channel)
Cd. 30-00107-03

Fig. 45 Optic Assembly Cd. 10-01688-00

1. Remove the front panel unscrewing the four anchored screws (Fig. 46).
2. Remove the J1, J6, J11, J15 connectors from the Plate Interface board (Fig. 47) and the
ground wire from the Optic Assembly.
3. Unscrew the four fastening screws of the Plate Interface board (Fig. 47).
4. Remove the connectors from JP1 of the Pre-Amplifier ADC boards.
5. Unscrew the four Optic Assembly fastening screws, remove it from its housing and substitute
it with the new one (Fig. 45).
6. Remount the assembly, repeating the above steps, 5 through 1, in reverse order.
7. After remounting, reseat the lamp and carry out the electronic adjustment check procedures as
described in the following pages.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 44

Chapter 06 – SETTINGS AND ADJUSTMENTS

Instrument front panel
anchored screw

Fig. 46 Instrument front view

Filter Wheel Motor – J1

Filter Wheel home sensor – J6

Photometer lamp – J15
Photometer fan– J11
Board fastening screw

Fig. 47 Plate Interface Board Cd. 30-01281-01

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 45

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.4.2 SUBSTITUTION OF THE OPTIC LAMP
N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.
1. Remove the front panel unscrewing the four anchored screws (Fig. 46).
2. Remove the lamp's electrical wires connected to the Lamp Regulator board by loosening the
two screws on the M1 connector (Fig.48).
3. Unscrew the lamp's fastening screws and remove it from its housing.
4. Remount the new lamp, repeating the above steps, 3 through 1, in reverse order.
5. Carry out the electronic adjustment of the Optic as described in the following pages.
Warning: Do not touch the glass portions of the lamp.

Halogen Lamp Cd. 9-35-0016-00

Lamp's fastening screw

M1 Connector
Optic Lamp Board Cd. 30-01576-00
Fig. 48

Ellipse

Trimmer P1

Service Manual

Rev. 04

01 October 2009

Page 46

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.4.3 ELECTRONIC ADJUSTMENT OF THE OPTIC
Warning: make sure that the Optic lamp and the reaction cuvettes are brand-new before
applying the following procedure.
1. Turn on the "Ellipse" system (instrument and computer). Remove the front panel unscrewing
the four anchored screws (Fig.46).
2. Launch the diagnostic program, select the "Plate " function and reset by clicking on the
appropriate key.
3. Make sure that the "Home Sensor" Reaction Plate window lights up green.
4. Place 500 µL of distilled water in cuvette # 1 of the reaction plate.
5. Select Sample plate position number three in order to make accessible the main channel Gain
and Offset holes.
6. Select the "Optic" function and turn the lamp "On" by clicking on the appropriate field.
7. Reset the filter wheel by clicking on the appropriate key.
8. Make sure that "Home Sensor" Filter Wheel window lights up green.
9. Make sure that the Regulator board for Connector M1 shows 6.0 volts ± 0.1. If not, turn the
P1 trimmer (Fig.48).
N.B.: The reaction cuvette must be perfectly clean. If not, carry out a "Cuvette Wash" cycle
(Start Work)
Offset main Channel

Gain reference Channel

Motor Support screw

Offset reference Channel

Gain main Channel

Fig. 49

Ellipse

Service Manual

Optic Assembly – top view

Rev. 04

01 October 2009

Page 47

Chapter 06 – SETTINGS AND ADJUSTMENTS

Warning: Room lighting and daylight can influence the reading. To avoid this interference,
remount the front panel before carrying out any check procedures.
10.

Click on "Start Sample Channel Conversion" key and make sure that the adjacent window
shows a count value of 100 ± 50. If not, adjust the Preamp/ADC (Read Channel) Board
"Offset" trimmer in order to set the desired value (Fig. 49).

11.

After regulating of the main channel offset, it is necessary to click on " Stop Sample Channel
Conversion" key.

12.

Click on "Start Reference Channel Conversion" and make sure that the adjacent window
shows a count value of 100 ± 50. If not, adjust the Preamp/ADC (Ref. Channel) Board
"Offset" trimmer in order to set the desired value (Fig. 49).

13.

After regulating of the reference channel offset, it is necessary to click on " Stop Reference
Channel Conversion" key.

14.

Select the filter positions from 1 to 8 (from 340 nm to 620 nm) and click on” Start/Stop
Sample Channel Conversion" key and determine which filter transmit the highest signal.

15.

Verify in the adjacent window a count value between 55.000 and 60.000. If not, adjust the
Preamp/ADC board "Gain" trimmer in order to set the desired value (Fig. 49).

16.

After regulating the Gain, select all the other filters and clicking on "Start/Stop Sample
Channel Conversion" verify count values between 29.000 and 60.000

Warning: Room lighting and daylight can influence the reading. To avoid this interference,
remount the front panel before carrying out any check procedures.
17.

Select the highest signal filter, clicking on "Start Reference Channel Conversion" and make
sure that the adjacent window shows a count value between 38.000 and 42.000. If not, adjust
the Preamp/ADC board "Gain" trimmer in order to set the desired value (Fig. 49).

18.

After regulating the Gain, select all the other filters and clicking on "Start/Stop Reference
Channel Conversion" verify count values between 19000 and 42000.

19.

After concluding these operations, empty cuvette # 1, remount the front panel and exit the
diagnostic program by clicking on the "Diagnostic" key.

20.

Ellipse

Exit the analyzer program press “shutdown” key.

Service Manual

Rev. 04

01 October 2009

Page 48

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.4.4 SUBSTITUTION OF THE OPTO SENSOR

N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.
1. Remove the front panel unscrewing the four anchored screws (Fig. 46).
2. Remove the J1, J6, J11, J15 connectors from the Reaction Plate Interface board (Fig. 47) and
the ground wire from the Optic Assembly.
3. Unscrew the four fastening screws of the Plate Interface board (Fig. 47).
4. Remove the connectors from JP1 of the Pre-Amplifier ADC boards.
5. Unscrew the four Optic Assembly fastening screws, remove it from its housing (Fig. 45).
6. Remove the lamp's power supply wires connected to the Lamp Regulator board by loosening
the two screws on the M1 connector (Fig. 48).
7. Unscrew the lamp's fastening screws and remove it from its housing.
8. Unscrew the four fastening screws on the motor support in order to access the filter wheel
(Fig. 49).
9. Position the filter wheel in such a manner as to make it possible to see and unscrew the
internal home sensor support screw. Unscrew the external home sensor support screw.
10.

Unscrew the home sensor fastening screws from its support and substitute with the new one
(Fig. 50).

N.B.: when remounting the Home sensor, make sure to center it with respect to the slots.
11.

Remount the assembly, repeating the above steps, 10 through 1, in reverse order.

12.

After remounting, reseat the lamp and carry out the electronic adjustment procedures as
described in the previous pages.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 49

Chapter 06 – SETTINGS AND ADJUSTMENTS

Filter Wheel Motor
Cd. 10-00399-01

Filter Wheel Home sensor
Cd. 9-10-0023-00
Fig. 50

Fig. 51

6.4.5 SUBSTITUTION OF THE MOTOR
N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.
1. Remove the front panel unscrewing the four anchored screws (Fig. 46).
2. Remove the J1, J6, J11, J15 connectors from the Plate Interface board (Fig. 47) and the
ground wire from the Optic Assembly.
3. Unscrew the four fastening screws of the Plate Interface board (Fig .47).
4. Remove the connectors from JP1 of the Pre-Amplifier ADC boards.
5. Unscrew the four Optic Assembly fastening screws, remove it from its housing (Fig. 45).
6. Remove the lamp's power supply wires connected to the Lamp Regulator board by loosening
the two screws on the M1 connector (Fig. 48).
7. Unscrew the lamp's fastening screws and remove it from its housing.
8. Unscrew the four fastening screws on the motor support in order to access the pulley (Fig. 49).
9. Unscrew the four fastening screws on the motor; loosen the pulley fastening screws.
10.

Remove the motor and substitute it with the new one (Fig.48).

11.

Remount the assembly, repeating the above steps, 10 through 1, in reverse order.

Warning: when tightening the motor pulley fastening screws, make sure that the filter wheel
is centered with respect to the opto sensor.
12.

After remounting, reseat the lamp and carry out the electronic adjustment procedures as
described in the previous pages.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 50

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.4.6 SUBSTITUTION OF THE OPTIC FILTERS
Interferential filters Kit Cd. 9-65-0029-00
N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.
1. Remove the front panel unscrewing the four anchored screws (Fig. 46).
2. Remove the J1, J6, J11, J15 connectors from the Plate Interface board (Fig. 47) and the
ground wire from the Optic Assembly.
3. Unscrew the four fastening screws of the Plate Interface board (Fig. 47).
4. Remove the connectors from JP1 of the Pre-Amplifier ADC boards.
5. Unscrew the four Optic Assembly fastening screws and remove it from its housing (Fig. 45).
6. Remove the lamp’s power supply wires connected to the Lamp Regulator board by loosening
the two screws on the M1 connector (Fig. 48).
7. Unscrew the lamp’s fastening screws and remove it from its housing.
8. Unscrew the four fastening screws on the motor support in order to access the filter wheel
(Fig. 49).
9. Unscrew the fastening screw that holds the elastic band of the filters.
10.

Remove the filter from its housing and substitute it with the new one.

11.

Remount the assembly, repeating the above steps, 9 through 1, in reverse order.

12.

After remounting, reseat the lamp and carry out the electronic adjustment procedures as
described in the previous pages.

Warning: Do not touch the glass parts of the filter.
Warning: when substituting an interferential filter due to malfunctioning or functional
anomalies, it is always advisable to substitute the entire set of filters mounted on the
instrument.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 51

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5

PLATES

6.5.1

REMOVAL OF THE PLATES ASSEMBLY

1. Make sure that the instrument (Ellipse) is turned off.
2. Open the analyzer main cover.
3. Remove the Sampling Probe from its housing unscrewing (clockwise) its ring nut.
4. Remove the reactions cuvettes front cover.
5. Remove with accuracy the left reactions cuvettes cover (placed under the Washing Station
cannulas) unscrewing its three screws.
6. Remove the Reagents protection cover and remove the Reagents rack from over the Reagents
Plate.
7. Remove all the Sample Racks from over the Samples Plate.
8. Remove with accuracy all the cuvettes Racks from over the Reactions Plate.
N.B.: before exctracting the Plates Assembly from the analyzer, it is necessary remove the
Sampling Arm, the Diluter Module, the Washing Station and then the Washing Well.
9. To remove the Sampling Arm, perform the relative procedure described into the previous
Section 6.1.3, from step 3 to step 10 (without substitute it with the new one).
10.

To remove the Diluter Module, perform the relative procedure described into the previous
Section 6.2.1, from step 4 to step 6.

11.

To remove the Washing Station, perform the relative procedure described into the successive
Section 6.6.1.

12.

Remove with accuracy the instrument chassis cover unscrewing the six anchored screws
(Fig. 52).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 52

Chapter 06 – SETTINGS AND ADJUSTMENTS

Peristaltic Pump
panel

Chassis cover
anchored screws

Fig. 52 - Chassis cover anchored screws
13.

Remove the two tubes from the overflow elbow fitting and fitting of the Washing Well
(Fig. 52), making attention not to crash them.

14.

Unscrew the Washing Well ring nut (Fig. 52) and pull out the Washing Well from its hole.
Washing Well

Washing Well
ring nut
Fitting

Overflow elbow
fitting
Fig. 6.5.2 - Washing Well

15.

Remove the Peristaltic Pump waste tube to the relative “Y” fitting, making attention to the
spillage of the biohazard liquid from the tube and from the “Y” fitting.

16.

Unplug the J3 connector from the Arm Interface Board and the ground wire from the
Peristaltic Pump mechanical support.

17.

Close with accuracy the instrument main cover.

18.

Open the Electronic Rack panel unscrewing the two quarter-turn fasteners (Fig. 53).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 53

Chapter 06 – SETTINGS AND ADJUSTMENTS

Washing Station
pumps panel

Electronic Rack
panel

Electronic Rack panel
quarter-turn fastener

Fig. 53 - Instrument (posterior view)
19.

Extract the Electronic Rack from its vain (Fig. 53).

Stepper Motors
Driver Board

Analytical Control Board

Fig. 54 - Electronic Rack
20.

Unplug the J6, J8, J9, J10, J11 e J20 connectors and the ground wire from the Analytical
Control Board (Fig. 54).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 54

Chapter 06 – SETTINGS AND ADJUSTMENTS

21.

Unplug the J1, J2 e J4 connectors and the ground wire from the Stepper Motors Driver Board
(Fig. 54).

22.

Insert with accuracy the Electronic Rack in its vain.

23.

Extract the Power Supply Rack from its vain (Fig. 55).

24V Module
12V Module

Power Supply
Cable Power
Supply wiring
Ground connector

Fastening
Screw Module

Fig. 55 - Power Supply Rack
24.

On the Power Supply, unscrew the four fastening screw of the “24V Module” and “12V
Module” and then, remove the relative four cables [JP24+, JP24-, JP12+, JP12-] of the Power
Supply wiring (Fig. 55).

25.

Unscrew the two ground connector fastening screws and remove all its ground wires
(Fig. 55).

26.

Insert with accuracy the Power Supply in its vain.

27.

Close the Electronic Rack panel screwing the two quarter-turn fasteners (Fig. 55).

28.

Lay down delicately the instrument on its posterior side (Fig. 56) and open the instrument
main cover

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 55

Chapter 06 – SETTINGS AND ADJUSTMENTS

Instrument adjustable feet

Plates Assembly nut
(and relative washers)

Fig. 56 - Instrument (inferior surface view)

29.

Loosen the two instrument adjustable feet (Fig. 56).

30.

Holding berthed the Plates Assembly to the instrument, unscrew the four Plates Assembly
nuts and remove the relative washers (Fig. 56).

31.

Close the instrument main cover and then, lay down delicately the instrument on its inferior
side.

32.

Open the instrument main cover, take out with accuracy the Plates Assembly to the
instrument and then, lay down the Plates Assembly outside the instrument.

33.

Close with accuracy the instrument main cover.

34.

Unplug the J28 connector and the ground wire from the Plate Interface Board (Fig. 57).

35.

Unplug the J21, J23 connectors and the ground wire from the Arm Interface Board.

36.

Unplug the two connectors to the Pre-Ampl./ADC (Read Channel) and the Pre-Ampl./ADC
(Reference Channel) placed in the Optic Assembly.

37.

Unscrew only the two Plate Interface Board fastening screws placed on its inferior side.

38.

Unscrew only the two Arm Interface Board fastening screws placed on its inferior side.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 56

Chapter 06 – SETTINGS AND ADJUSTMENTS

39.

Put up the Plates Assembly and remove its Base Plate Assy from the four Plates Assembly
spacing bars (Fig. 6.5.8).
Reactions Plate Motor – J3

Samples Plate Motor – J2

Reagents Plate Motor – J4
Samples Plate Home Sensor – J7

Reactions Plate Home Sensor – J8

Power
Supply – J28

Reagents Plate Home Sensor – J9

Fig. 57 - Plate Interface Board (Sx) [P/N: 30-01281-01]

Plates Assembly

Arm Interface
Board
Plate Interface
Board
Base Plate Assy

Plates Assembly spacing bar

Fig. 58 - Plates Assembly and Base Plate Assy
40.

Lay down the Plates Assembly on its four spacing bars.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 57

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5.2

REPLACEMENT OF THE REAGENTS PLATE MOTOR

N.B.: before replacing of the Reagents Plate Motor, it is necessary remove the Plates
Assembly from the instrument.
1. To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.
2. Unplug the J4 connector from the Plate Interface Board (Fig. 57).
3. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59).

Reagents
Plate
Home Sensor lever
Reagents Plate Motor
fastening screw

Reactions Plate Motor
fastening screw

Reagents Plate
belt
Reagents Plate
pulley

Samples Plate Motor
fastening screw

Fig. 59 - Plates Assembly (inferior view)
4. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the
Reagents Plate Home Sensor lever (Fig. 59) out of the Reagents Plate pulley.
5. Remove the belt from the Reagents Plate Motor pulley (Fig. 59).
6. Remove the belt from the Reagents Plate pulley.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 58

Chapter 06 – SETTINGS AND ADJUSTMENTS

7. Pull upwards the support Reagents Rack (Fig. 60) and remove it to the Plates Assembly.

Support Reagents Rack

Samples Plate

Reactions Plate
Reactions Plate
Home Sensor
Reactions Plate Home Sensor lever
Samples Plate
Home Sensor

Samples Plate Home Sensor lever
Reagent
Tts Plate Home
Sensor

Reagents Plate Home Sensor lever

Fig. 60 - Plates, Plate Home Sensors and Plate Home Sensor levers

Fig. 61 - Support Reagents Rack

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 59

Chapter 06 – SETTINGS AND ADJUSTMENTS

8. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate
(Fig. 59).
9. Unscrew completely the four fastening screws of the Reagents Plate Motor (Fig. 59) and pull
out it from its housing.
10.

Loosen the pulley clamp screw and extract the pulley clamp to the motor pulley
(Fig. 62).

Anchor
plate
fastening screw

Plate Motor

Pulley clamp screw

Anchor plate

Connector Motor

Pulley clamp
Motor pulley

Fig. 6.5.12 - Plate Motor with pulley
11.

Take out the motor pulley from its axis (Fig. 64).

12.

Unscrew the four fastening screws of the Reagents Plate Motor anchor plate (Fig. 62) and
take out the anchor plate to the motor.

13.

Replace the motor with the new Plate Motor Assy (Fig. 63).

Motor axis

Fig. 63 - Plate Motor Assy [P/N: 9-10-0016-00]

Ellipse

Service Manual

Rev. 04

Fig. 64 - Plate Motor

01 October 2009

Page 60

Chapter 06 – SETTINGS AND ADJUSTMENTS

14.

Insert the pulley clamp on the motor pulley (Fig. 62).

15.

Insert the motor pulley (with the motor clamp) to the motor axis of the new Plate Motor Assy
(Fig. 62).

16.

Position the motor pulley to 1.50 mm with respect to upper motor plane (Fig. 65), then tighten
the pulley clamp screw (Fig. 62).

Upper motor plane

Pulley clamp

Stepping
Motor
Motor pulley

Fig. 65 - Plate Motor scheme
17.

Position the anchor plate on the upper plane of the new motor with the two anchor plate slots
on the side of the motor connector (like represented in the (Fig. 62).

18.

Screw the four anchor plate fastening screws (Fig. 62).

19.

Replace the Reagents Plate Motor on its housing with the connector turned towards the
Reagents Plate Home Sensor lever (Fig. 59).

20.

Insert (without tighten) the four Reagents Plate Motor fastening screws (Fig. 59).

21.

Insert the support Reagents Rack in the Plates Assembly (Fig. 61).

22.

Wrap the belt around the Reagents Plate pulley.

23.

Wrap the belt around the Reagents Plate Motor pulley.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 61

Chapter 06 – SETTINGS AND ADJUSTMENTS

24.

Push the motor outward from the Plates pulleys axis and holding it in this position, so as to
maintain the belt in tension, then tighten the four fastening screws of the Reagents Plate
Motor (Fig. 59).

25.

Make sure that the belt thoots are correctly inserted on the Reagents Plate pulley.

26.

Make sure that the belt thoots are correctly inserted on the Reagents Plate Motor pulley.

27.

Insert the Reagents Plate Home Sensor around the Home sensor disk.

28.

Tighten the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered
with respect its index.

29.

Rotate with accuracy the Reagents Plate pulley and check that the belt tension is corrected and
that the Home Sensor disk not to touch the Reagents Home Sensor.

30.

Plug in the J4 connector from the Plate Interface Board (Fig. 57).

31.

To replace the Plates Assembly, perform the procedure described into the previous
Section 6.5.1 in inverse order: from step 40 to step 1.

N.B.: after having replaced the new Reagents Plate Motor and the Plates Assembly, it is
necessary to align mechanically the Reagents Plate by following the procedure described into
the Section 6.5.14.1 .
6.5.3

REPLACEMENT OF THE SAMPLES PLATE MOTOR

N.B.: before replacing of the Samples Plate Motor, it is necessary remove the Plates Assembly
from the instrument.
1. To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.
2. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59).
3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the
Reagents Plate Home Sensor lever out of the Reagents Plate pulley.
4. Remove the belt from the Reagents Plate Motor pulley.
5. Remove the belt from the Reagents Plate pulley.
6. Loosen the two grains of the Reagents Plate pulley (Fig. 66) and extract the pulley to the
plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66).
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 62

Chapter 06 – SETTINGS AND ADJUSTMENTS

7. Pull upwards the support Reagents Rack (Fig. 61) and remove it to the Plates Assembly.
8. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate
(Fig. 59).
9. Unplug the J2 connector from the Plate Interface Board (Fig. 57).
10.

Loosen the four fastening screws of the Samples Plate Motor (Fig. 59).

11.

Remove the belt from the Samples Plate Motor pulley (Fig. 66).

12.

Remove the belt from the Samples Plate pulley.

13.

Unscrew completely the four fastening screws of the Samples Plate Motor (Fig. 59) and pull
out it from its housing.

14.

Loosen the pulley clamp screw and extract the pulley clamp to the motor pulley
(Fig. 62).

15.

Take out the motor pulley from its axis (Fig. 64).

16.

Unscrew the four fastening screws of the Samples Plate Motor anchor plate (Fig. 62) and take
out the anchor plate to the motor.

17.

Replace the motor with the new Plate Motor Assy (Fig. 63).

18.

Insert the pulley clamp on the motor pulley (Fig. 62).

19.

Insert the motor pulley (with the motor clamp) to the motor axis of the new Plate Motor Assy
(Fig. 62).

20.

Position the motor pulley to 1.50 mm with respect to upper motor plane (Fig. 65), then tighten
the pulley clamp screw (Fig. 62).

21.

Position the anchor plate on the upper plane of the new motor with the two anchor plate slots
on the side of the motor connector (like represented in the Fig. 62).

22.

Screw the four anchor plate fastening screws (Fig. 62).

23.

Replace the Samples Plate Motor on its housing with the connector turned towards the Optic
Assembly (Fig. 59).

24.

Insert (without tighten) the four Samples Plate Motor fastening screws (Fig. 59).

25.

Wrap the belt around the Samples Plate pulley.

26.

Wrap the belt around the Samples Plate Motor pulley.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 63

Chapter 06 – SETTINGS AND ADJUSTMENTS

27.

Push the motor outward from the Plates pulleys axis and holding it in this position, so as to
maintain the belt in tension, then tighten the four fastening screws of the Samples Plate Motor
(Fig. 59).

28.

Make sure that the belt thoots are correctly inserted on the Samples Plate pulley.

29.

Make sure that the belt thoots are correctly inserted on the Samples Plate Motor pulley.

30.

Rotate with accuracy the Samples Plate pulley and check that the belt tension is corrected.

31.

Plug in the J2 connector from the Plate Interface Board (Fig. 57).

32.

Insert the Reagents Plate pulley (Fig. 66) on the plates axis, making attention not to fold (or to
detach) its Home Sensor disk [Fig. 66] (and with the Home Sensor disk slit aligned with
respect to the slits of the other two Home Sensor disks).

33.

Tighten the two grains of the Reagents Plate pulley (Fig. 66) only after to have verified that
the three Plate Motor pulleys are locked between of them.

34.

Wrap the belt around the Reagents Plate pulley.

35.

Wrap the belt around the Reagents Plate Motor pulley.

36.

Push the motor outward from the Plates pulleys axis and holding it in this position, so as to
maintain the belt in tension, then tighten the four fastening screws of the Reagents Plate
Motor (Fig. 59).

37.

Make sure that the belt thoots are correctly inserted on the Reagents Plate pulley.

38.

Make sure that the belt thoots are correctly inserted on the Reagents Plate Motor pulley.

39.

Insert the Reagents Plate Home Sensor around the Home sensor disk.

40.

Tighten the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered
with respect its index.

41.

Rotate with accuracy the Reagents Plate pulley and check that the belt tension is corrected and
that the Home Sensor disk not to touch the Reagents Home Sensor.

42.

To replace the Plates Assembly, perform the procedure described into the previous
Section 6.5.1 in inverse order: from step 40 to step 1.

N.B.: after having replaced the new Samples Plate Motor, the Reagents Plate Motor and the
Plates Assembly, it is necessary to align mechanically the Samples Plate and the Reagents
plate by following the procedure described into the Section 6.5.15.1
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 64

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5.4

REPLACEMENT OF THE REACTIONS PLATE MOTOR

N.B.: before replacing of the Reactions Plate Motor, it is necessary remove the Plates
Assembly from the instrument.
1. To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.
2. Perform the procedure described into the previous Section 6.5.3 from step 2 to step 8.
3. Unscrew the Samples Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the
Samples Plate Home Sensor lever out of the Samples Plate pulley.
4. Loosen the four fastening screws of the Samples Plate Motor (Fig. 59).
5. Remove the belt from the Samples Plate Motor pulley.
6. Remove the belt from the Samples Plate pulley.
7. Loosen the two grains of the Samples Plate pulley (Fig. 66) and extract the pulley to the plates
axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66).
8. Loosen the four fastening screws of the Reactions Plate Motor (Fig. 59).
9. Remove the belt from the Reactions Plate Motor pulley.
10.

Remove the belt from the Reactions Plate pulley.

11.

Unplug the J3 connector from the Plate Interface Board (Fig. 57).

12.

Unscrew completely the four fastening screws of the Reactions Plate Motor (Fig. 59) and pull
out it from its housing.

13.

Loosen the pulley clamp screw and extract the pulley clamp to the motor pulley
(Fig. 62).

14.

Ellipse

Take out the motor pulley from its axis (Fig. 64).

Service Manual

Rev. 04

01 October 2009

Page 65

Chapter 06 – SETTINGS AND ADJUSTMENTS

15.

Unscrew the four fastening screws of the Reactions Plate Motor anchor plate (Fig. 62) and
take out the anchor plate to the motor.

16.

Replace the motor with the new Plate Motor Assy (Fig. 63).

17.

Insert the pulley clamp on the motor pulley (Fig. 62).

18.

Insert the motor pulley (with the motor clamp) to the motor axis of the new Plate Motor Assy
(Fig. 62).

19.

Position the motor pulley to 1.50 mm with respect to upper motor plane (Fig. 65), then tighten
the pulley clamp screw (Fig. 62).

20.

Position the anchor plate on the upper plane of the new motor with the two anchor plate slots
on the side of the motor connector (like represented in the Fig. 62).

21.

Screw the four anchor plate fastening screws (Fig. 62).

22.

Replace the Reactions Plate Motor on its housing with the connector turned towards the
Reactions Home Sensor lever (Fig. 59).

23.

Insert (without tighten) the four Reactions Plate Motor fastening screws (Fig. 59).

24.

Wrap the belt around the Reactions Plate pulley.

25.

Wrap the belt around the Reactions Plate Motor pulley.

26.

Push the motor outward from the Plates pulleys axis and holding it in this position, so as to
maintain the belt in tension, then tighten the four fastening screws of the Reactions Plate
Motor (Fig. 59).

27.

Make sure that the belt thoots are correctly inserted on the Reactions Plate pulley.

28.

Make sure that the belt thoots are correctly inserted on the Reactions Plate Motor pulley.

29.

Rotate with accuracy the Reactions Plate pulley and check that the belt tension is corrected.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 66

Chapter 06 – SETTINGS AND ADJUSTMENTS

30.

Plug in the J3 connector from the Plate Interface Board (Fig. 57).

31.

Replace the Samples Plate pulley (Fig. 66) and tighten the two its grains only after to have
verified that the two Plate Motor pulleys are locked between of them (and with the Home
Sensor disk slit aligned with respect to the slit of the other Home Sensor disk).

32.

Perform the procedure described into the previous Section 6.5.3 from step 25 to step 29.

33.

Tighten the Samples Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered
with respect its index.

34.

Rotate with accuracy the Samples Plate pulley and check that the belt tension is corrected and
that the Home Sensor disk not to touch the Samples Home Sensor.

35.

Perform the procedure described into the previous Section 6.5.3 from step 32 to step 41.

36.

To replace the Plates Assembly, perform the procedure described into the previous
Section 6.5.1 in inverse order: from step 40 to step 1.

N.B.: after having replaced the new Reactions Plate Motor, the Samples Plate Motor, the
Reagents Plate Motor and the Plates Assembly, it is necessary to align mechanically all the
plates by following the procedure described into the Section 6.5.16.1.

6.5.5 REPLACEMENT OF THE REAGENTS PLATE BELT
N.B.: before replacing of the Reagents Plate Belt, it is necessary remove the Plates Assembly
from the instrument.
1. To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.
2. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59).
3. Remove the belt from the Reagents Plate Motor pulley (Fig. 62).
4. Remove the belt from the Reagents Plate pulley (Fig. 66).
5. Pull upwards the support Reagents Rack (Fig. 61) and remove it to the Plates Assembly.
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 67

Chapter 06 – SETTINGS AND ADJUSTMENTS

6. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate
(Fig. 59).
7. Substitute the belt with the new belt.
8. Insert the support Reagents Rack in the Plates Assembly (Fig. 60).
9. Wrap the new belt around the Reagents Plate pulley.
10.

Wrap the new belt around the Reagents Plate Motor pulley.

11.

Push the motor outward from the Plates pulleys axis and holding it in this position, so as to
maintain the new belt in tension, then tighten the four fastening screws of the Reagents Plate
Motor (Fig. 59).

12.

Make sure that the belt thoots are correctly inserted on the Reagents Plate pulley.

13.

Make sure that the belt thoots are correctly inserted on the Reagents Plate Motor pulley.

14.

Rotate with accuracy the Reagents Plate pulley and check that the belt tension is corrected.

15.

To replace the Plates Assembly, perform the procedure described into the previous Section
6.5.1 in inverse order: from step 40 to step 1.

N.B.: after having replaced the new Reagents Plate belt and the Plates Assembly, it is
necessary to align mechanically the Reagents Plate by following the procedure described into
the Section 6.5.14.1.
6.5.6 REPLACEMENT OF THE SAMPLES PLATE BELT
N.B.: before replacing of the Samples Plate belt, it is necessary remove the Plates Assembly
from the instrument.
1. To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.
2. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59).
3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the
Reagents Plate Home Sensor lever out of the Reagents Plate pulley.
4. Remove the belt from the Reagents Plate Motor pulley (Fig. 62).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 68

Chapter 06 – SETTINGS AND ADJUSTMENTS

5. Remove the belt from the Reagents Plate pulley (Fig. 66).
6. Loosen the two grains of the Reagents Plate pulley (Fig. 66) and extract the pulley to the
plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66).
7. Pull upwards the support Reagents Rack (Fig. 61) and remove it to the Plates Assembly.
8. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate
(Fig. 59).
9. Loosen the four fastening screws of the Samples Plate Motor (Fig. 59).
10.

Remove the belt from the Samples Plate Motor pulley (Fig. 62).

11.

Remove the belt from the Samples Plate pulley (Fig. 66).

12.

Substitute the belt with the new belt.

13.

Insert the support Reagents Rack in the Plates Assembly (Fig. 60).

14.

Wrap the new belt around the Samples Plate pulley.

15.

Wrap the new belt around the Samples Plate Motor pulley.

16.

Push the motor outward from the Plates pulleys axis and holding it in this position, so as to
maintain the belt in tension, then tighten the four fastening screws of the Samples Plate Motor
(Fig. 59).

17.

Make sure that the belt thoots are correctly inserted on the Samples Plate pulley.

18.

Make sure that the belt thoots are correctly inserted on the Samples Plate Motor pulley.

19.

Rotate with accuracy the Samples Plate pulley and check that the belt tension is corrected.

20.

Perform the procedure described into the previous Section 6.5.3 from step 32 to step 41.

21.

To replace the Plates Assembly, perform the procedure described into the previous
Section 6.5.1 in inverse order: from step 40 to step 1.

N.B.: after having replaced the new Samples Plate belt, the Reagents Plate Motor and the
Plates Assembly, it is necessary to align mechanically the Samples Plate and the Reagents
Plate by following the procedure described into the Section 6.5.15.1.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 69

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5.7 REPLACEMENT OF THE REACTIONS PLATE BELT
N.B.: before replacing of the Reactions Plate belt, it is necessary remove the Plates Assembly
from the instrument.
1. To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.
2. Loosen the four fastening screws of the Reagents Plate Motor (Fig. 59).
3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the
Reagents Plate Home Sensor lever out of the Reagents Plate pulley.
4. Remove the belt from the Reagents Plate Motor pulley (Fig. 62).
5. Remove the belt from the Reagents Plate pulley (Fig. 66).
6. Pull upwards the support Reagents Rack (Fig. 61) and remove it to the Plates Assembly.
7. Upset the Plates Assembly carefullying lay down it from the part of the Reactions Plate
(Fig. 59).
8. Loosen the two grains of the Reagents Plate pulley (Fig. 66) and extract the pulley to the
plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66).
9. Loosen the four fastening screws of the Samples Plate Motor (Fig. 59).
10.

Unscrew the Samples Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the
Samples Plate Home Sensor lever out of the Samples Plate pulley.

11.

Remove the belt from the Samples Plate Motor pulley (Fig. 62).

12.

Remove the belt from the Samples Plate pulley (Fig. 66).

13.

Loosen the two grains of the Samples Plate pulley (Fig. 66) and extract the pulley to the plates
axis, making attention not to fold (or to detach) its Home Sensor disk (Fig. 66).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 70

Chapter 06 – SETTINGS AND ADJUSTMENTS

14.

Loosen the four fastening screws of the Reactions Plate Motor (Fig. 59).

15.

Remove the belt from the Reactions Plate Motor pulley (Fig. 62).

16.

Remove the belt from the Reactions Plate pulley (Fig. 66).

17.

Substitute the belt with the new belt.

18.

Insert the support Reagents Rack in the Plates Assembly (Fig. 60).

19.

Wrap the belt around the Reactions Plate pulley.

20.

Wrap the belt around the Reactions Plate Motor pulley.

21.

Push the motor outward from the Plates pulleys axis and holding it in this position, so as to
maintain the belt in tension, then tighten the four fastening screws of the Reactions Plate
Motor (Fig. 59).

22.

Make sure that the belt thoots are correctly inserted on the Reactions Plate pulley.

23.

Make sure that the belt thoots are correctly inserted on the Reactions Plate Motor pulley.

24.

Rotate with accuracy the Reactions Plate pulley and check that the belt tension is corrected.

25.

Replace the Samples Plate pulley and tighten the two its grains only after to have verified that
the two Plate Motor pulleys are locked between of them (and with the Home Sensor disk slit
aligned with respect to the slit of the other Home Sensor disk).

26.

Perform the procedure described into the previous Section 6.5.3 from step 25 to step 29.

27.

Tighten the Samples Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered
with respect its index.

28.

Rotate with accuracy the Samples Plate pulley and check that the belt tension is corrected and
that the Home Sensor disk not to touch the Samples Home Sensor.

29.
Ellipse

Perform the procedure described into the previous Section 6.5.3 from step 32 to step 41.
Service Manual

Rev. 04

01 October 2009

Page 71

Chapter 06 – SETTINGS AND ADJUSTMENTS

30.

To replace the Plates Assembly, perform the procedure described into the previous Section
6.5.1 in inverse order: from step 40 to step 1.

N.B.: after having replaced the new Reactions Plate belt, the Samples Plate pulley, the
Reagents Plate pulley and the Plates Assembly, it is necessary to align mechanically all the
plates by following the procedure described into the Section 6.5.16.1.

6.5.8 REPLACEMENT OF THE REAGENTS PLATE PULLEY
N.B.: before replacing of the Reagents Plate pulley, it is necessary remove the Plates Assembly
from the instrument
1.

To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.

2.

Perform the procedure described into the previous Section 6.5.7 from step 2 to step 13.

3.

Substitute the pulley with the new pulley (Fig. 66).

Samples Plate
pulley
Reactions Plate
pulley

Grain

Home Sensor disk
Reagents Plate
pulley
Home Sensor disk slit
Fig. 66 - Plates pulleys
4.
Ellipse

Perform the procedure described into the previous Section 6.5.3 from step 32 to step 41.
Service Manual

Rev. 04

01 October 2009

Page 72

Chapter 06 – SETTINGS AND ADJUSTMENTS

5.

To replace the Plates Assembly, perform the procedure described into the previous
Section 6.5.1 in inverse order: from step 40 to step 1.

N.B.: after having replaced the new Reagents Plate pulley and the Plates Assembly, it is
necessary to align mechanically the Reagents Plate by following the procedure described into
the Section 6.5.14.2.
6.5.9 REPLACEMENT OF THE SAMPLES PLATE PULLEY
N.B.: before replacing of the Samples Plate pulley, it is necessary remove the Plates Assembly
from the instrument.
1. To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.
2. Perform the procedure described into the previous Section 6.5.4 from step 2 to step 7.
3. Substitute the pulley with the new pulley (Fig. 66).
4. Perform the procedure described into the previous Section 6.5.7 from step 25 to step 29.
5. To replace the Plates Assembly, perform the procedure described into the previous
Section 6.5.1 in inverse order: from step 40 to step 1.
N.B.: after having replaced the new Samples Plate pulley, the Reagents Plate pulley and the
Plates Assembly, it is necessary to align mechanically the Samples Plate and the Reagents
Plate by following the procedure described into the Section 6.5.15.2.

6.5.10 REPLACEMENT OF THE REACTIONS PLATE PULLEY
N.B.: before replacing of the Reactions Plate pulley, it is necessary remove the Plates
Assembly from the instrument
1. To remove the Plates Assembly, perform the relative procedure described into the previous
Section 6.5.1.
2. Perform the procedure described into the previous Section 6.5.7 from step 2 to step 16.
3. Unscrew the Reactions Plate Home Sensor lever adjusting screw (Fig. 67) and rotate the
Reactions Plate Home Sensor lever out of the Reactions Plate pulley.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 73

Chapter 06 – SETTINGS AND ADJUSTMENTS

4. Loosen the two grains of the Reactions Plate pulley (Fig. 67) and extract the pulley to the
plates axis, making attention not to fold (or to detach) its Home Sensor disk (Fig.67).
5. Substitute the pulley with the new pulley (Fig. 66).
6. Replace the new Reactions Plate pulley and tighten the two its grains only after to have
verified that it is locked on the superior surface.
7. Insert the Reactions Plate Home Sensor around the Home Sensor disk.
8. Tighten the Reactions Plate Home Sensor lever adjusting screw (Fig. 67) so that it is centered
with respect to its index.
9. Perform the procedure described into the previous Section 6.5.7 from step 18 to step 29.
10.

To replace the Plates Assembly, perform the procedure described into the previous
Section 6.5.1 in inverse order: from step 40 to step 1.

N.B.: after having replaced the new Reactions Plate pulley, the Samples Plate pulley, the
Reagents Plate pulley and the Plates Assembly, it is necessary to align mechanically all the
plates by following the procedure described into the Section 6.5.16.2 .

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 74

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5.11 REPLACEMENT OF THE REAGENTS PLATE HOME SENSOR
1. Remove the instrument front panel unscrewing its four anchored screws.
2. Unplug the J9 connector from the Plate Interface Board (Fig. 57).
3. Unscrew the Home Sensors screw (Fig. 67).
Reactions Plate Home Sensor
lever adjusting screw
Reactions Plate
Home Sensor lever

Samples Plate Home Sensor
lever adjusting screw
Reagents Plate Home Sensor
lever adjusting screw

Samples Plate
Home Sensor lever

Spacer column
Reagents Plate
Home Sensor
lever

Home Sensors screw

Fig. 67 - Plates Home Sensors levers adjusting screws
4.

Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and remove the
Reagents Plate Home Sensor lever from the Reagents Plate pulley.

5.

Unscrew the two Reagents Plate Home Sensor fastening screws on the its lever (Fig. 68).

6.

Substitute the Home Sensor with the new Home Sensor (Fig. 69).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 75

Chapter 06 – SETTINGS AND ADJUSTMENTS

Home Sensor

Lever

Home Sensor
fastening screw
Fig. 68 - Plate HomeSensor

Fig. 69 – Plate Home Sensor Assy

fastening screws

[P/N: 9-10-0023-00]

7. Replace the new Home Sensor on its lever with the window Home Sensor outward-facing
(Fig. 68) and screw the two Reagents Plate Home Sensor fastening screws.
8. Insert the Reagents Plate Home Sensor around the Home Sensor disk.
9. Replace the Reagents Plate Home Sensor lever on its location and screw the Reagents Plate
Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect to its index.
10.

Screw the Home Sensors screw (Fig. 67).

11.

Plug in the J9 connector in the Plate Interface Board (Fig. 57).

12.

Rotate with accuracy the Reagents Plate pulley and check that the Home Sensor disk not to
touch the Reagents Home Sensor.

13.

Place the instrument front panel screwing its four anchored screws.

N.B.: after having replaced the new Reagents Plate Home Sensor, it is necessary to align
mechanically the Reagents Plate by following the procedure described into the Section
6.5.14.3 .

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 76

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5.12 REPLACEMENT OF THE SAMPLES PLATE HOME SENSOR
1. Remove the instrument front panel unscrewing its four anchored screws.
2. Unscrew the Home Sensors screw (Fig. 67).
3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and remove the
Reagents Plate Home Sensor lever from the Reagents Plate pulley.
4. Unscrew the first spacer of the Home Sensor levers with the relative washer (Fig. 67).
5. Unplug the J7 connector from the Plate Interface Board (Fig. 57).
6. Unscrew the Samples Plate Home Sensor lever adjusting screw (Fig. 67) and remove the
Samples Plate Home Sensor lever from the Samples Plate pulley.
7. Unscrew the two Samples Plate Home Sensor fastening screws on the its lever (Fig. 68).
8. Substitute the Home Sensor with the new Home Sensor (Fig. 69).
9. Replace the new Home Sensor on its lever with the window Home Sensor outward-facing
(Fig. 68) and screw the two Samples Plate Home Sensor fastening screws.
10.

Insert the Samples Plate Home Sensor around the Home sensor disk.

11.

Replace the Samples Plate Home Sensor lever on its location and screw the Samples Plate
Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect to its index.

12.

Replace the spacer washer and screw the first spacer of the Home Sensor levers.

13.

Plug in the J7 connector in the Plate Interface Board (Fig. 57).

14.

Rotate with accuracy the Samples Plate pulley and check that the Home Sensor disk not to
touch the Samples Home Sensor.

15.

Insert the Reagents Plate Home Sensor around the Home Sensor disk.

16.

Replace the Reagents Plate Home Sensor lever on its location and screw the Reagents Plate
Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index.

17.

Screw the Home Sensors screw (Fig. 67).

18.

Rotate with accuracy the Reagents Plate pulley and check that the Home Sensor disk not to
touch the Reagents Home Sensor.

19.

Ellipse

Place the instrument front panel screwing its four anchored screws.

Service Manual

Rev. 04

01 October 2009

Page 77

Chapter 06 – SETTINGS AND ADJUSTMENTS

N.B.: after having replaced the new Samples Plate Home Sensor and the Reagents Plate
Home Sensor, it is necessary to align mechanically the Samples Plate and Reagents Plate by
following the procedure described into the Section 6.5.15.3.
6.5.13 REPLACEMENT OF THE REACTIONS PLATE HOME SENSOR
1. Remove the instrument front panel unscrewing its four anchored screws.
2. Unscrew the Home Sensors screw (Fig. 67).
3. Unscrew the Reagents Plate Home Sensor lever adjusting screw (Fig. 67) and remove the
Reagents Plate Home Sensor lever from the Reagents Plate pulley.
4. Unscrew the first spacer of the Home Sensor levers with the relative washer (Fig. 67).
5. Unscrew the Samples Plate Home Sensor lever adjusting screw (Fig. 67) and remove the
Samples Plate Home Sensor lever from the Samples Plate pulley.
6. Unplug the J8 connector from the Plate Interface Board (Fig. 57).
7. Unscrew the second spacer of the Home Sensor levers with the relative washer (Fig. 57).
8. Unscrew the Reactions Plate Home Sensor lever adjusting screw (Fig. 67) and remove the
Reactions Plate Home Sensor lever from the Reactions Plate pulley.
9. Unscrew the two Reactions Plate Home Sensor fastening screws on the its lever (Fig. 68).
10.

Substitute the Home Sensor with the new Home Sensor (Fig. 69).

11.

Replace the new Home Sensor on its lever with the window Home Sensor outward-facing
(Fig. 68) and screw the two Reactions Plate Home Sensor fastening screws.

12.

Insert the Reactions Plate Home Sensor around the Home sensor disk.

13.

Replace the Reactions Plate Home Sensor lever on its location and screw the Reactions Plate
Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect to its index.

14.

Replace the spacer washer and screw the second spacer of the Home Sensor levers.

15.

Plug in the J8 connector in the Plate Interface Board (Fig. 57).

16.

Rotate with accuracy the Reactions Plate pulley and check that the Home Sensor disk not to
touch the Reactions Home Sensor.

17.

Ellipse

Insert the Samples Plate Home Sensor around the Home Sensor disk.

Service Manual

Rev. 04

01 October 2009

Page 78

Chapter 06 – SETTINGS AND ADJUSTMENTS

18.

Replace the Samples Plate Home Sensor lever on its location and screw the Samples Plate
Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index.

19.

Replace the spacer washer and screw the first spacer of the Home Sensor levers.

20.

Rotate with accuracy the Samples Plate pulley and check that the Home Sensor disk not to
touch the Samples Home Sensor.

21.

Insert the Reagents Plate Home Sensor around the Home sensor disk.

22.

Replace the Reagents Plate Home Sensor lever on its location and screw the Reagents Plate
Home Sensor lever adjusting screw (Fig. 67) so that it is centered with respect its index.

23.

Screw the Home Sensors screw (Fig. 67).

24.

Rotate with accuracy the Reagents Plate pulley and check that the Home Sensor disk not to
touch the Reagents Home Sensor.

25.

Place the instrument front panel screwing its four anchored screws.

N.B.: after having replaced the new Reactions Plate Home Sensor, the Samples Plate Home
Sensor and the Reagents Plate Home Sensor, it is necessary to align mechanically all the
plates by following the procedure described into the Section 6.5.16.3
6.5.14 ALIGNMENT AND ADJUSTMENTS OF THE REAGENTS PLATE
6.5.14.1 Alignment and adjustment of the Reagents Plate in case of replacement of the
Reagents Plate motor or of the Reagents Plate belt
N.B.: in case of replacement of the Reagents Plate motor or of the Reagents Plate belt, it is
necessary only control that the Reagents Plate is centered with respect to the Sampling Arm
1. Make sure the instrument (Ellipse) is turned on.
1. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate.
2. Launch the “Diagnostic” program.
3. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
4. Make sure that the distance between the top edge of the Washing Well and the tip of the
Sampling Probe is 25 mm.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 79

Chapter 06 – SETTINGS AND ADJUSTMENTS

5. Select “Arm” folder.
6. Click the "Go Std" button, then click the correspondent “Down” button. The arm will move to
position #1 of the Standards Plate, then it will go down until to reach the right height.
7. Make sure that the Sampling Probe is vertically aligned and centered with respect to position
#1 of the Standards Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
adjust (if necessary) the verticality by using the adjustable anchored bearing (Fig. 7), click the
‘Save’ button, then repeat the above steps from 6.
8. Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move
to position #1 of the Reagents Plate, then it will go down until to reach the right height.
9. Make sure that the Sampling Probe is vertically aligned and centered with respect to position
#1 of the Reagents Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
then click the ‘Save’ button.
10.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general
reset procedure has been completed.

6.5.14.2 Alignment and adjustment of the Reagents Plate in case of replacement of the
Reagents Plate pulley
N.B.: in case of replacement of the Reagents Plate pulley, align and adjust the Reagents Plate
with respect to the Sampling Arm.
1. Make sure that the instrument (Ellipse) is turned on.
2. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate.
3. Remove the instrument front panel unscrewing its four anchored screws.
4. Launch the “Diagnostic” program.
5. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
6. Make sure that the distance between the top edge of the Washing Well and the tip of the
Sampling Probe is 25 mm.
7. Select “Plate” folder.
8. Click the "ALL Motors OFF" button.
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 80

Chapter 06 – SETTINGS AND ADJUSTMENTS

9. Loosen the two grains of the Reagents Plate pulley (Fig. 66).
10.

Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been
completed.

11.

Select “Arm” folder.

12.

Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move
to position #1 of the Reagents Plate, then it will go down until to reach the right height.

13.

Rotate manually the Reagents Plate until to reach the right centering of the position #1 of the
Reagents Plate with respect to the Sampling Probe, then adjust (if necessary) the positioning
of the Sampling Probe by using the ‘+’ and ‘-‘ keys.

14.

Tighten the grain at sight of the Reagents Plate pulley only after to have verified that the three
Plate Motor pulleys are locked between of them, then click the ‘Save’ button.

15.

Select “Plate” folder and click the "ALL Motors OFF" button.

16.

Rotate manually the Reagents Plate until to have in view the other grain of the Reagents Plate
pulley, then tighten it.

17.

Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been
completed.

18.

Select “Arm” folder.

19.

Click the "Go Std" button, then click the correspondent “Down” button. The arm will move to
position #1 of the Standards Plate, then it will go down until to reach the right height.

20.

Make sure that the Sampling Probe is vertically aligned and centered with respect to
position #1 of the Standards Plate. If it not be so: adjust the positioning by using the ‘+’ and
‘-‘ keys, adjust (if necessary) the verticality by using the adjustable anchored bearing, click
the ‘Save’ button, then repeat the above steps from 15.

21.

Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move
to position #1 of the Reagents Plate, then it will go down until to reach the right height.

22.

Make sure that the Sampling Probe is vertically aligned and centered with respect to position
#1 of the Reagents Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
then click the ‘Save’ button.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 81

Chapter 06 – SETTINGS AND ADJUSTMENTS

23.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general
reset procedure has been completed.

24.

Place the instrument front panel screwing its four anchored screws.

6.5.14.3 Alignment and adjustment of the Reagents Plate in case of replacement of the
Reagents Plate Home Sensor
N.B.: in case of replacement of the Reagents Plate Home Sensor, center the Reagents Plate
with respect to the Sampling Arm adjusting the position of the Reagents Plate Home Sensor
lever.
1. Make sure that the instrument (Ellipse) is turned on.
2. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate.
3. Remove the instrument front panel unscrewing its four anchored screws.
4. Launch the “Diagnostic” program.
5. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
6. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal),
Reactions, Samples and Reagents Plates light up in green.
7. Select “Arm” folder.
8. Click the "Go Reag." button, then click the correspondent “Down” button. The arm will move
to position #1 of the Reagents Plate, then it will go down until to reach the right height.
9. Make sure that the Sampling Probe is vertically aligned and centered with respect to
position #1 of the Reagents Plate. If it not be so:
a.

loosen the Reagents Plate Home Sensor lever adjusting screw (Fig. 67);

b.

turn slightly the Reagents Plate Home Sensor lever to the right toward in order to obtain the
centering;

c.

tighten the Reagents Plate Home Sensor lever adjusting screw;

d.

repeat the steps from 5 to 9.

10.

Click the "Go Std" button, then click the correspondent “Down” button. The arm will move to
position #1 of the Standards Plate, then it will go down until to reach the right height

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 82

Chapter 06 – SETTINGS AND ADJUSTMENTS

11.

Make sure that the Sampling Probe is vertically aligned and centered with respect to
position #1 of the Standards Plate.

12.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general
reset procedure has been completed.

13.

Place the instrument front panel screwing its four anchored screws.

6.5.15 LIGNMENT AND ADJUSTMENTS OF THE SAMPLES PLATE
6.5.15.1 Alignment and adjustment of the Samples Plate in case of replacement of the Samples
Plate motor or of the Samples Plate belt
N.B.: in case of replacement of the Samples Plate motor or of the Samples Plate belt, it is
necessary control that the Samples Plate is centered with respect to the Sampling Arm and
then, align and adjust the Reagents Plate
1. Perform the procedure described into the previous Section 6.5.14.2 from step 1 to step 20.
2. Select “Arm” folder.
3. Click the "Go Sample" button, then click the correspondent “Down” button. The arm will
move to position #1 of the Sample Rack #1, then it will go down until to reach the right
height.
4. Make sure that the Sampling Probe is vertically aligned and centered with respect to
position #1 of the Sample Rack #1. If it not be so: adjust the positioning by using the ‘+’ and
‘-‘ keys and click the ‘Save’ button.
5. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general
reset procedure has been completed.
6. Place the instrument front panel screwing its four anchored screws.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 83

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5.15.2 Alignment and adjustment of the Samples Plate in case of replacement of the
Samples Plate pulley
N.B.: in case of replacement of the Samples Plate pulley, align and adjust the Reagents Plate
and the Samples Plate with respect to the Sampling Arm.
1. Make sure that the instrument (Ellipse) is turned on.
2. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13).
3. Launch the “Diagnostic” program.
4. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
5. Make sure that the distance between the top edge of the Washing Well and the tip of the
Sampling Probe is 25 mm.
6. Select “Plate” folder.
7. Click the "ALL Motors OFF" button.
8. Loosen the two grains of the Samples Plate pulley (Fig. 66).
9. Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been
completed.
10.

Select “Arm” folder.

11.

Click the "Go Sample" button, then click the correspondent “Down” button. The arm will
move to position #1 of the Sample Rack #1, then it will go down until to reach the right
height.

12.

Rotate manually the Samples Plate until to reach the right centering of the position #1 of the
Sample Rack #1 with respect to the Sampling Probe, then adjust (if necessary) the positioning
of the Sampling Probe by using the ‘+’ and ‘-‘ keys.

13.

Tighten the grain at sight of the Samples Plate pulley, then click the ‘Save’ button.

14.

Select “Plate” folder and click the "ALL Motors OFF" button.

15.

Rotate manually the Samples Plate until to have in view the other grain of the Samples Plate
pulley, then tighten it.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 84

Chapter 06 – SETTINGS AND ADJUSTMENTS

16.

Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been
completed.

17.

Select “Arm” folder.

18.

Click the "Go Sample" button, then click the correspondent “Down” button. The arm will
move to position #1 of the Sample Rack #1, then it will go down until to reach the right
height.

19.

Make sure that the Sampling Probe is vertically aligned and centered with respect to
position #1 of the Sample Rack #1.

20.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general
reset procedure has been completed.

21.

Perform the procedure described into the previous Section 6.5.14.2 from step 7 to step 21.

6.5.15.3 Alignment and adjustment of the Samples Plate in case of replacement of the
Samples Plate Home Sensor
N.B.: in case of replacement of the Samples Plate Home Sensor, center the Samples Plate and
the Reagents Plate with respect to the Sampling Arm adjusting the position of the relative
Plate Home Sensor levers.
1. Make sure that the instrument (Ellipse) is turned on.
2. Make sure that the Reagents protection cover is permanently locked on the Reagents Plate.
3. Remove the instrument front panel unscrewing its four anchored screws (Fig. 13).
4. Launch the “Diagnostic” program.
5. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
6. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal),
Reactions, Samples and Reagents Plates light up in green.
7. Select “Arm” folder.
8. Click the "Go Sample" button, then click the correspondent “Down” button. The arm will
move to position #1 of the Sample Rack #1, then it will go down until to reach the right
height.
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 85

Chapter 06 – SETTINGS AND ADJUSTMENTS

9. Make sure that the Sampling Probe is vertically aligned and centered with respect to
position #1 of the Sample Rack #1. If it not be so:
a.

loosen the Samples Plate Home Sensor lever adjusting screw (Fig. 6.5.17);

b.

turn slightly the Samples Plate Home Sensor lever to the right toward in order to obtain the
centering;

c.

tighten the Samples Plate Home Sensor lever adjusting screw;

d. repeat the steps from 5 to 9.
10.

Click the "Go Sample" button, then click the correspondent “Down” button. The arm will
move to position #1 of the Sample Rack #1, then it will go down until to reach the right
height.

11.

Make sure that the Sampling Probe is vertically aligned and centered with respect to
position #1 of the Sample Rack #1.

12.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general
reset procedure has been completed.

13.

Perform the procedure described into the previous Section 6.5.14.3 from step 7 to step 13.

6.5.16

LIGNMENT AND ADJUSTMENTS OF THE REACTIONS PLATE

6.5.16.1 Alignment and adjustment of the Reactions Plate in case of replacement of the
Reactions Plate motor or of the Reactions Plate belt

N.B.: in case of replacement of the Reactions Plate motor or of the Reactions Plate belt, it is
necessary control that the Reactions Plate is centered with respect to the Sampling Arm and
with respect to the Optic Assembly; then, align and adjust the Samples Plate and the Reagents
Plate.

1. Make sure that the instrument (Ellipse) is turned on.
2. Remove the reactions cuvettes front cover.
3. Remove with accuracy the left reactions cuvettes cover (placed under the Washing Station
cannulas) unscrewing its three screws.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 86

Chapter 06 – SETTINGS AND ADJUSTMENTS

4. Take a with Reactions Plate Rack with a opaque cuvette #1 and then, substitute the Reactions
Plate Rack #1 with this new Rack.
5. Make sure that the distance between the top edge of the Washing Well and the tip of the
Sampling Probe is 25 mm.
6. Launch the “Diagnostic” program.
7. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
8. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal),
Reaction, Samples and Reagent Plates light up in green.
9. Make sure that the lamp of the Optic Assembly is turned on.
10.

Select “Optic” folder and click the "Spare" button.

11.

Make sure that the light beam of the lamp is centered with respect to cuvette #1 of the
Reactions Plate. If it not be so: perform the procedure described into the Section 6.5.16.3.

12.

Select “Arm” folder.

13.

Click the "Go Disp." button, then click the correspondent “Down” button. The arm will move
to cuvette #81 of the Reactions Plate, then it will go down until to reach the right height.

14.

Make sure that the Sampling Probe is vertically aligned and centered with respect to
cuvette #81 of the Reactions Plate. If it not be so: adjust the positioning by using the ‘+’ and
‘-‘ keys, then click the ‘Save’ button.

15.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.

16.

Substitute the Reactions Plate Rack #1 with the previous Rack #1.

17.

Replace with accuracy the left reactions cuvettes cover (placed under the Washing Station
cannulas) screwing its three screws.

18.

Replace the reactions cuvettes front cover.

19.

Perform the procedure described into the previous Section 6.5.15.2 from step 6 to step 21.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 87

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.5.16.2 Alignment and adjustment of the Reactions Plate in case of replacement of the
Reactions Plate pulley
N.B.: in case of replacement of the Reactions Plate pulley, align and adjust the Reactions
Plate, the Samples Plate and the Reagents Plate with respect to the Sampling Arm.
1. Make sure the instrument (Ellipse) is turned on.
2. Remove the instrument front panel unscrewing its four anchored screws.
3. Remove the reactions cuvettes front cover.
4. Remove with accuracy the left reactions cuvettes cover (placed under the Washing Station
cannulas) unscrewing its three screws.
5. Take a with Reactions Plate Rack with a opaque cuvette #1 and then, substitute the Reactions
Plate Rack #1 with this new Rack.
6. Launch the “Diagnostic” program.
7. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
8. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal),
Reaction, Samples and Reagent Plates light up in green.
9. Select “Plate” folder.
10.

Click the "ALL Motors OFF" button.

11.

Make sure that the two Reagents pulley grains are locked.

12.

Make sure that the three Plate Home Sensor levers (Fig. 67) are centered with respect to the
relative index.

13.

Loosen the two grains of the Reactions Plate pulley (Fig. 66).

14.

Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been
completed.

15.

Select “Optic” folder and click the "Spare" button.

16.

Make sure that the lamp of the Optic Assembly is turned on.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 88

Chapter 06 – SETTINGS AND ADJUSTMENTS

17.

Rotate manually the Reactions Plate until to reach the right centering of the cuvette #1 of the
Reactions Plate Rack #1 with respect to the light beam of the Optic Assembly.

18.

Tighten the grain at sight of the Reactions Plate pulley.

19.

Select “Plate” folder and click the "ALL Motors OFF" button.

20.

Rotate manually the Reactions Plate until to have in view the other grain of the Reactions
Plate pulley, then tighten it.

21.

Click the "ALL Motors ON" button. Wait until the reset instrument procedure has been
completed.

22.

Select “Optic” folder and click the "Spare" button.

23.

Make sure that the cuvette #1 of the Reactions Plate Rack #1 is centered with respect to the
light beam of the Optic Assembly. If it not be so:
a. loosen the Reactions Plate Home Sensor lever adjusting screw (Fig. 67);
b. turn slightly the Reactions Plate Home Sensor lever to the right toward in order to obtain the
centering;
c. tighten the Reactions Plate Home Sensor lever adjusting screw;
d. select “Plate” folder and click the "Reset" button of the Reactions Plate;
e. repeat the step 23.

24.

Make sure that the Reagents protection cover is permanently locked on the Reagents Plate.

25.

Perform the procedure described into the previous Section 6.1.1 from step 4 to step 12.

26.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.

27.

Select “Arm” folder.

28.

Click the "Go Disp." button, then click the correspondent “Down” button. The arm will move
to cuvette #81 of the Reactions Plate, then it will go down until to reach the right height.

29.

Make sure that the Sampling Probe is vertically aligned and centered with respect to cuvette
#81 of the Reactions Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
then click the ‘Save’ button.

30.

Ellipse

Substitute the Reactions Plate Rack #1 with the previous Rack #1.

Service Manual

Rev. 04

01 October 2009

Page 89

Chapter 06 – SETTINGS AND ADJUSTMENTS

31.

Replace with accuracy the left reactions cuvettes cover (placed under the Washing Station
cannulas) screwing its three screws.

32.

Replace the reactions cuvettes front cover.

33.

Perform the procedure described into the previous Section 6.5.15.2 from step 4 to step 21.

6.5.16.3 Alignment and adjustment of the Reactions Plate in case of replacement of the
Reactions Plate Home Sensor
N.B.: in case of replacement of the Reactions Plate Home Sensor, center the Reactions Plate,
the Samples Plate and the Reagents Plate with respect to the Sampling Arm adjusting the
position of the relative Plate Home Sensor levers
1. Make sure the instrument (Ellipse) is turned on.
2. Remove the instrument front panel unscrewing its four anchored screws.
3. Remove the reactions cuvettes front cover.
4. Remove with accuracy the left reactions cuvettes cover (placed under the Washing Station
cannulas) unscrewing its three screws.
5. Take a with Reactions Plate Rack with a opaque cuvette #1 and then, substitute the Reactions
Plate Rack #1 with this new Rack.
6. Launch the “Diagnostic” program.
7. Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.
8. Make sure that all the Home Sensors of the Sampling Arm (Vertical and Horizontal),
Reaction, Samples and Reagent Plates light up in green.
9. Select “Optic” folder and click the "Spare" button.
10.

Make sure that the cuvette #1 of the Reactions Plate Rack #1 is centered with respect to the
light beam of the Optic Assembly. If it not be so:
a. loosen the Reactions Plate Home Sensor lever adjusting screw (Fig. 67);
b. turn slightly the Reactions Plate Home Sensor lever to the right toward in order to obtain the
centering;

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 90

Chapter 06 – SETTINGS AND ADJUSTMENTS

c. tighten the Reactions Plate Home Sensor lever adjusting screw;
d. select “Plate” folder and click the "Reset" button of the Reactions Plate;
e. repeat the step 10.
11.

Make sure that the Reagents protection cover is permanently locked on the Reagents Plate.

12.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the reset
instrument procedure has been completed.

13.

Select “Arm” folder.

14.

Click the "Go Disp." button, then click the correspondent “Down” button. The arm will move
to cuvette #81 of the Reactions Plate, then it will go down until to reach the right height.

15.

Make sure that the Sampling Probe is vertically aligned and centered with respect to cuvette
#81 of the Reactions Plate. If it not be so: adjust the positioning by using the ‘+’ and ‘-‘ keys,
then click the ‘Save’ button.

16.

Substitute the Reactions Plate Rack #1 with the previous Rack #1.

17.

Replace with accuracy the left reactions cuvettes cover (placed under the Washing Station
cannulas) screwing its three screws.

18.

Replace the reactions cuvettes front cover.

19.

Select "Miscellaneous” folder and request an Instrument General Reset. Wait until the general
reset procedure has been completed.

20.

Ellipse

Perform the procedure described into the previous Section 6.5.15.3 from step 7 to step 13

Service Manual

Rev. 04

01 October 2009

Page 91

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.6 WASHING STATION ASSEMBLY
6.6.1 SUBSTITUTION OF THE WASHING STATION ASSEMBLY
N.B.: make sure that the Ellipse instrument is turned off before performing this substitution
procedure.
1. Unscrew the fastening screw and remove the Washing Station protective cover (Fig. 70).
2. Remove the top cover of the Washing Station unscrewing the four fastening screws (Fig. 71).
3. Disconnect the eight tubes attached to the probes on the Washing Station (Fig. 72).
4. Unscrew the two Washing Station fastening screws (Fig. 71) and pull out the Washing Station
from its housing.
5. Remove the instrument front panel unscrewing the four anchored screws (Fig. 73).
6. Remove the J5 and J10 connectors from the Plate Interface Board (Fig. 74).
7. Remove the ground wire from the Washing Station.
8. Remove the Washing Station Assembly and replace it with the new one.
9. Remount, repeating the above steps, 7 through 1, in reverse order.
10. After substituting and remounting, carry out the mechanical adjustment and functional check
procedures as described in the next pages.

Fastening screw

Washing Station
probes holder

Washing Station protective cover

Fig. 70 - Washing Station protective cover
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 92

Chapter 06 – SETTINGS AND ADJUSTMENTS

Top Cover fastening screws
Tubing

Spring

Washing Station
fastening screws
Fig. 72 - Washing Station probes

Fig. 71 - Washing Station Assembly
[P/N: 10-01299-00]

Instrument
front panel

Front panel
anchored screws

Fig. 73 - Front panel anchored screws

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 93

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.6.2 SUBSTITUTION OF THE WASHING STATION HOME SENSOR
N.B.: make sure that the Ellipse instrument is turned off before performing this substitution
procedure.
1. Unscrew the fastening screw and remove the Washing Station protective cover (Fig. 70).
2. Remove the instrument front panel unscrewing the four anchored screws (Fig. 73).
3. Remove the two fastening screws from the Home Sensor Assy (Fig. 75), unplug the J10
connector from the Plate Interface Board (Fig. 74), take out the entire assy and replace it with
the new one.
4. To remount, repeat the above steps in inverse order: from 3 to 1.
5. After substituting and remounting, carry out the mechanical adjustment and functional check
procedures as described in the next pages.

Washing Station Motor – J5

Washing Station Home Sensor – J10
01

Fig. 74 - Plate Interface Board [P/N: 30-01281-01]

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 94

Chapter 06 – SETTINGS AND ADJUSTMENTS

Washing Station Home Sensor
Home Sensor fastening screw
Adjusting set screw

Tubing
guide
Washing Station Motor
Fig. 76 - Washing Station Motor
[ Assy P/N: 10-00399-01]

Fig. 75 - Washing Station Home Sensor
[ Assy P/N: 9-10-0023-60]

6.6.3 SUBSTITUTION OF THE WASHING STATION MOTOR
N.B.: make sure that the Ellipse instrument is turned off before performing this substitution
procedure.
1. Unscrew the fastening screw and remove the Washing station protective cover (Fig. 70).
2. Unscrew the two Washing Station fastening screws (Fig. 71) and pull out the Washing station
from its housing.
3. Remove the instrument front panel unscrewing the four anchored screws (Fig. 73).
4. Remove the J5 connector from the Plate Interface Board (Fig. 74).
5. Loosen the two flywheel and motor joint screws.
6. Unscrew the four fastening screws on the motor (Fig. 76), take out it and replace it with the
new one.
7. Remount, repeating the above steps, 6 through 1, in reverse order.
8. After substituting and remounting, carry out the mechanical adjustment and functional check
procedures as described in the next pages.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 95

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.6.4 WASHING STATION ASSEMBLY - MECHANICAL ADJUSTMENT
1. Make sure that the cuvette #1 of the Reactions Plate is centered with respect to the light beam
of the Photometer reading channel.
2. Turn on the Ellipse instrument, unscrew the fastening screw of the Washing Station protective
cover (Fig. 70) and remove the cover from its place.
3. Launch the Diagnostic program, select “Wash” folder and click the “Home” button. Wait until
the reset Washing Station procedure has been completed.
4. Make sure that the Home Sensor of the Washing Station light up in green.
5. Click the “Down to Adjust” button in the “Wash” folder. The Washing Station will go down
until to reach the right height for the adjusting.
6. Make sure that the five probes of the Washing Station are centered with respect to the five
correspondent cuvettes of the Reactions Plate. If it not be so, adjust in the following way:
a. loosen the two Washing Station fastening screws (Fig. 71);
b. shift horizontally – carefully – the Washing Station in order to obtain the
centering of the five probes with respect to the five correspondent cuvettes;
c. tighten the two Washing Station fastening screws (Fig. 71);
d. click the “Home” button in the “Wash” folder and wait until the reset
Washing Station procedure has been completed;
e. make sure that the Home Sensor of the Washing Station light up in green;
f. click the “Down to Adjust” button in the “Wash” folder and wait until the
request movement has been completed;
g. repeat the above step 6.
7. Click the “Home” button in the “Wash” folder. Wait until the reset Washing Station
procedure has been completed.
8. Click the “Down (Full)” button in the “Wash” folder. The Washing Station will go down until
to reach the cuvettes of the Reactions Plate.
9. Make sure that the five probes of the Washing Station are centered with respect to the five
correspondent cuvettes of the Reactions Plate and, in particular, that the Drying Pad run freely
along the walls of the correspondent cuvette.
10. Make sure that the five probes of the Washing Station are slightly raised. If it not be so, adjust
in the following way:
a. click the “Home” button in the “Wash” folder and wait until the reset Washing
Station procedure has been completed;
b. make sure that the Home Sensor of the Washing Station light up in green;
c. loosen the two fastening screws of the Washing Station Home Sensor (Fig. 75);
d. move slightly the Washing Station Home Sensor to the right vertically toward in
order to obtain the positioning of the Washing Station probes to the correct
height;
e. tighten the two fastening screws of the Washing Station Home Sensor (Fig. 75);
f. click the “Home” button in the “Wash” folder and wait until the reset Washing
Station procedure has been completed;
g. make sure that the Home Sensor of the Washing Station light up in green;
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 96

Chapter 06 – SETTINGS AND ADJUSTMENTS

h. ick the “Down (Full)” button in the “Wash” folder and wait until the Washing
Station has been reach the cuvettes of the Reactions Plate;
i. repeat the above step 10.
11. Click the “Home” button in the “Wash” folder. Wait until the reset Washing Station procedure
has been completed.
12. Make sure that the Home Sensor of the Washing Station light up in green.
13. Replace the Washing Station protective cover on the Washing Station and screw its fastening
screw (Fig. 70).
Additional information: make sure that the Sampling Probe is centered with respect to
the Reactions Plate cuvette. If it not be so, perform the relative adjusting procedure.
Adjusting
set screw

If the horizontal backlash of the Washing Station probes holder (Fig. 70)
is excessive, then adjust it acting on the adjusting set screw (Fig. 75).

6.6.5 WASHING STATION - FUNCTIONAL CHECK
1. Turn on the Ellipse system and launch the “Analyzer” program.
2. Launch the Diagnostic program and select the "Wash " folder.
3. Click on the “Washing Cycle” button and perform the functional check of the Washing
Station through the following steps:
4. Select, in the “Cycle #” field, the number of washing cycles to be performed from 1 to 10 and
press the “Washing Cycle” button.
Attention: if there are doubts about the right washing system functionality, select only
one washing cycle in order to avoid cuvettes overflow.
5. Remove the reaction rack cover and the left side reaction rack cover unscrewing the three
anchored screws.
6. Put one reaction rack below the Washing Station, permitting at the cannulas to reach the
bottom of the cuvettes (as shown in Fig. 77).
Reaction rack

Fig. 77 - Washing Station and reaction rack
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 97

Chapter 06 – SETTINGS AND ADJUSTMENTS

7. By referring to the “Washing Test” window, push the “OK” button to start the Washing Cycle.
8. Check that the cuvettes are correctly filled and emptied (Fig. 77):

• Dispensed volumes shall be: Æ 300 µL ± 50 into the first cuvette, two cycles
Æ 300 µL ± 50 into the second cuvette, two cycles
Æ 300 µL ± 50 into the third cuvette, two cycles plus one third
cycle where the dispensed volume shall be 400 µL ± 50
If not, check and clean in case the relevant hydraulic line including micro-pump and valve,
taking reference in the Hydraulic Diagram (SI-16-00571-01).
9. Remove the reaction rack below the Washing Station and remount the reaction rack covers.
10. Exit from the Diagnostic program by pressing the “Diagnostic” button.
11. Exit from the “Analyzer” program by pressing the “Shutdown” button.

6.7 WASH PUMP ASSEMBLY
6.7.1 SUBSTITUTION OF THE WASH PUMP ASSEMBLY
N.B.: make sure that the Ellipse instrument is turned off before performing this substitution
procedure.
Rear door fastening screw

Fig. 78 Instrument rear view

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 98

Chapter 06 – SETTINGS AND ADJUSTMENTS

Micro-Pump P2 Assy
Cd. 05-01826-16

Anchored screw

µP6
µP5

µP4

µP3

µP2

Micro-Pump P6 Assy
Cd. 05-01711-20
Electro-Valve EV2 (2 way)
Cd. 09-35-0035-00
Fig. 79 Wash Pump Assembly Cd. 05-01488-00

Unscrew the four fastening screws to remove the door located on the instrument rear side to access
the Wash Pump Assembly (Fig. 78) .
Disconnect the air tubes from µP6 and the tubes arriving on the Electro-Valves and Micro-Pumps
(Fig. 79).
Unscrew the four anchored screws indicated which hold the Wash Pump Assembly (Fig. 79).
Disconnect connectors J1, J2 e J6 from the Hydraulics Interface board (Fig. 81).
Remove the Wash Pump Assembly from its site and replace it with the new one.
Remount, repeating the above steps, 4 through 1, in reverse order.
After mounting the new module, carry out the function check as described in the next pages.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 99

Chapter 06 – SETTINGS AND ADJUSTMENTS

N.B:. If power loss is experienced in the filling or emptying of liquid (after a period of
instrument activity), perform the following operations before substituting or servicing any
part(s):
•

Disconnect the liquid input tube from the malfunctioning pump.

•

Insert a small amount of liquid using a syringe from the disconnected ends in order to wet the
membranes (Fig. 80).
Remove the syringe and reconnect the tube(s); perform functioning check as described in the
next points.

Fig. 80
Hydraulics Interface board
Cd. 30-01626-00

J6
J1
J2

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 100

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 81 Wash Pump Assembly (Rear view)
6.7.2 SUBSTITUTION OF PART(S)
N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.
1. Unscrew the four fastening screws to remove the door located on the instrument rear side to
access the Wash Pump Assembly (Fig. 78) .
2. For each part to be replaced, if necessary, disconnect the relative electrical power connection,
the hydraulic tubes and the fastening screws.
3. If necessary, unscrew the four anchored screws of the assembly and turn it ahead (Fig.81).
4. Remount, repeating the above steps, 3 through 1, in reverse order.
5. After mounting the new part, carry out the function check as described in the next pages.
6.7.3 FUNCTION CHECK FOR THE WASH PUMP ASSEMBLY
1. Turn on the "Ellipse" system and launch the “Analyzer” program.
2. Launch the diagnostic program and select the "Wash " function.
3. Click on the “Washing Cycle” key and perform the functional check of the Washing Station
through the following steps:
4. Select, in the “Cycle #” field, the number of washing cycles to be performed from 1 to 10
and press the “Washing Cycle” key.
Attention: if there are doubts about the right washing system functionality, select only one
washing cycle in order to avoid cuvettes overflow
5. Remoe the reaction rack cover and the left side reaction rack cover unscrewing the three

anchored screws.
6. Put one reaction rack below the washing station, permitting to the cannulas to reach the
cuvettes button (Fig. 82).

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 101

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 82 Washing station
7.

By referring to the window here below, push the “OK” key to start the Washing Cycle

Check that the cuvettes are correctly filled and emptied (Fig. 82)
•

Dispensed volumes shall be:

300 µL ± 50 into the first cuvette, two cycles
300 µL ± 50 into the second cuvette, two cycles
300 µL ± 50 into the third cuvette, two cycles plus one third
cycle where the dispensed volume shall be 400 µL ± 50

If not, check and clean in case the relevant hydraulic line including micro-pump and valve, taking
reference in the Hydraulic Diagram (SI 16-00571-01).
Remove the reaction rack below the washing station and remount the reaction rack covers.
Exit the Diagnostic program by pressing the “Diagnostic” key.
Exit the Analyzer program press “Shutdown” key.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 102

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.8 PERISTALTIC PUMP ASSEMBLY
6.8.1 SUBSTITUTION OF THE PERISTALTIC PUMP ASSEMBLY

Peristaltic pump fastening screw

Peristaltic pump door

Hinged guide tube
Fig. 83

Instrument right view

Fig. 84

fastening fitting

Peristaltic Pump Assy
Cd. 05-01489-00

1. Open the door located on the right side of the instrument Fig. 83.
2. Unhook the hinged guide by pushing its up side toward the left and up, as reported in Fig. 85.
3. Pull out the tubes off their relative nipples unscrewing the fastening fittings (Fig. 84).
4. Remove the front panel unscrewing the four anchored screws (Fig. 87).
5. Unplug the J3 connector from the Sampling Interface Board. (Fig. 88).
6. Unscrew the two fastening screws (Fig. 84), remove the Peristaltic Pump module and
substitute it with the new one.
7. To remount, repeat the above steps in reverse order: from 6 to 1.
8. After mounting the new module, carry out the function check for the peristaltic pump module
as described in the next pages.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 103

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.8.2 SUBSTITUTION OF THE PERISTALTIC PUMP TUBES
1. Open the door located on the right side of the instrument Fig. 83.
2. Unhook the hinged guide by pushing its up side toward the left and up, as reported in Fig. 85.
3. Pull out the tubes off their relative nipples unscrewing the fastening fittings (Fig. 84) and
substitute with new ones.
4. Remount, repeating the above steps in reverse order: from 3 to 1.
•

Always use original replacement parts; never lubricate the peristaltic pump roller bearings.

•

After prolonged instrument inactivity: verify the efficiency of the peristaltic pump tubes.

Please note: if the instrument is not able to completely empty the washing well, check the condition
of the tube and make sure that there are not obstruction into the hydraulic lines.

Peristaltic Pump Rotor

Pulley screw

Peristaltic Pump Motor

Fig. 85 P.Pump Rotor Cd.05-01512-00

Ellipse

Service Manual

Fig.86 P.Pump Motor Cd. 10-00399-01

Rev. 04

01 October 2009

Page 104

Chapter 06 – SETTINGS AND ADJUSTMENTS

Fig. 87 Instrument front view
Instrument front panel
anchored screw
J 3 - Peristaltic pump Motor

Fig. 88 Arm Interface board Cd. 30-01282-01

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 105

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.8.3 SUBSTITUTION OF THE PERISTALTIC PUMP MOTOR
1. Open the door located on the right side of the instrument Fig. 83.
2. Unhook the hinged guide by pushing its up side toward the left and up, as reported in Fig. 85.
3. Pull out the tubes off their relative nipples unscrewing the fastening fittings (Fig. 84).
4. Remove the front panel unscrewing the four anchored screws (Fig. 87).
5. Unscrew the two fastening screws (Fig. 84), unplug the J3 connector from the Sampling
Interface Board (Fig. 88) and remove the Peristaltic Pump module from his housing.
6. Unscrew the four motor fastening screws and loosen the two motor pulley screws which hold
the peristaltic pump rotor(Fig. 85).
7. Take out the motor and substitute it with the new one.
8. To remount, repeat the above steps in reverse order: from 6 to 1.
9. After mounting the new motor, carry out the function check for the peristaltic pump module
as described in the next pages.
6.8.4 FUNCTION CHECK FOR THE PERISTALTIC PUMP
1. Turn on the Ellipse system (instrument and computer) and launch the “Analyzer” program.
Select "Start Work" and perform one or more cuvette washing cycles. Make sure that the
sampling probe washing well empties completely without any leakage of liquid. If leakage
should occur, replace the Peristaltic Pump tubes as illustrated in the previous points.
2. Close the "Ellipse" program by clicking on "Shutdown".

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 106

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.9 SUBSTITUTION OF THE ELECTRONICS BOARDS
6.9.1 ANALYTICAL CONTROL BOARD Cd. 30-01283-01
N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.

Instrument front panel
anchored screw

Rear door anchored screw
Fig. 89 Instrument rear view

Fig. 90 Instrument front view

1. Unscrew the two anchored screws to open the door located on the rear side of the instrument
to access the Electronic Assy (Fig. 89) .
2. Take out the Electronic Assy as showed (Fig. 3), disconnect connectors J1, J6, J8, J9, J10,
J11, J13, J20 and the ground wire from the Analytical Control board (Fig. 91).
3. Unscrew the four fastening screws on the Analytical Control board, disconnect it from the
Stepper Motors Driver Board and replace it with the new one.
4. Remount, repeating the above steps, 3 through 1, in reverse order.
5. After mounting the new board, turn on the Ellipse system and launch the “Diagnostic
program” in order to transfer all the previous instrument settings to the new Analytical control
board.
6. To exit the diagnostic program, press the “Diagnostic” key.
7. To exit the analyzer program, press the “Shutdown” key.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 107

Chapter 06 – SETTINGS AND ADJUSTMENTS

J1- Stepper Motors Driver Board

J6

J8

J9

J11
J10
J20
J1
J13

J7-Analytical Control board

Board fastening screw

Fig. 91 Electronic Assy
J4

J3

J2

Fig. 92 Electronic Assy

Board fastening screw

6.9.2 STEPPER MOTORS DRIVER BOARD Cd. 30-01284-01
N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.
1. Unscrew the two anchored screws to open the door located on the rear side of the instrument
to access the Electronic Assy (Fig. 89) .
2. Take out the Electronic Assy as showed (Fig. 91) and disconnect connectors J2, J3, J4
(Fig.92).
3. Unscrew the two fastening screws on the Stepper Motors Driver Board, disconnect it from the
Analytical Control board and replace it with the new one.
4. Remount, repeating the above steps, 3 through 1, in reverse order.
Ellipse

Service Manual

Rev. 04

01 October 2009

Page 108

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.9.3 ARM INTERFACE BOARD

Cd. 30-01282-01

N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.
1. Remove the front panel unscrewing the four anchored screws (Fig. 90).
2. Unplug the J1, J2, J3, J4, J5, J6, J8, J13, J16, J17, J21, J23, J25, J26 J27connectors and the
ground wire from the Arm Interface board (Fig. 93).
3. Unscrew the four fastening screws of the Arm Interface board and replace it with the new one.
4. Remount, repeating the above steps, 3 through 1, in reverse order.

J1

J2

J3

J4

J16

J17

J5

J6

J13

J8
J27
J21
J23
J26

J25

Fig. 93

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 109

Chapter 06 – SETTINGS AND ADJUSTMENTS

6.9.4 PLATE INTERFACE BOARD

Cd. 30-01281-01

N.B.: Make sure the instrument (Ellipse) is turned off before performing this substitution
procedure.
1. Remove the front panel unscrewing the four anchored screws (Fig. 90).
2. Unplug the J1, J2, J3, J4, J5, J6, J7, J8, J9, J10, J11, J14, J15, J16, J26, J27, J28 connectors
and the ground wire from the Plate Interface board (Fig. 94).
3. Unscrew the four fastening screws of the Plate Interface board and replace it with the new
one.
4.

Remount, repeating the above steps, 3 through 1, in reverse order.

J1

J2

J3

J4

J8

J9

J5

J6
J14
J7

J10

J16

J11

J28

J15
J26

J27

Fig. 94 Plate Interface board

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 110

Chapter 07 - MAINTENANCE

CHAPTER 07
- MAINTENANCE INDEX

7.

MAINTENANCE…………………………………………………………………….…..2

7.1

PREVENTIVE MAINTENANCE……………………………………………………….2

7.2

LIST OF PARTS SUBJECT TO WEAR AND USAGE………………………………...4

7.3

SAMPLING PROBE - CLEANING PROCEDURE…………………………………….5

7.4

WASH STATION CANNULAS - CLEANING PROCEDURE..…………..…………...5

7.5

WASH SOLUTION BOTTLES - CLEANING……………………………..…………...6

7.5.1 WASH SOLUTION BOTTLES - CLEANING PROCEDURE…………………...…….6
7.6

TIP WASHING ………...……………………………………………………………......7

7.6.1

TIP - CLEANING PROCEDURE………………………………………………..………7

7.6.2 PROCEDURE FOR REPLACING THE TIP……………………………………………9
7.7

HYDRAULIC CIRCUIT WASHING ………...…………………………..….………..10

7.7.1

HYDRAULIC CIRCUIT - CLEANING PROCEDURE…….…………………….…..10

7.8

CHANGING THE PERISTALTIC PUMP TUBES…………….…………………..….11

7.8.1 PROCEDURE FOR REPLACING THE PERISTALTIC PUMP TUBES………..…...11
7.9

CHANGING THE PHOTOMETER LAMP………………………………………….. .12

7.9.1 PROCEDURE FOR REPLACING THE PHOTOMETER LAMP……………….….. .12
7.10

CHANGING THE REACTION CUVETTES………………………….…………….. .14

7.10.1 PROCEDURE FOR REPLACING THE REACTION CUVETTES………..………….14
7.11

CHANGING THE TUBES KIT…………………………………………………………16

7.11.1 PROCEDURE FOR REPLACING THE TUBES KIT…………………..………...……16
7.12

TABLE C – LIST OF THOSE MAINTENANCE PROCEDURES THAT CAN BE
PERFORMED BY THE USER AND/OR BY THE MAINTENANCE

TECHNICIAN

…………………………………………………………….........................……………..17
7.13

Ellipse

DECONTAMINATION PROCEDURE………………………………………………...19

Service Manual

Rev. 03

01 June 2009

Page 1

Chapter 07 - MAINTENANCE

7

MAINTENANCE

This chapter contains all those routine operations, which concern instrument maintenance. Said
procedures, listed and described below, should be carefully and scrupulously followed in order to
guarantee the manufacturer’s specifications and the perfect working order of the instrument over
time.

7.1

PREVENTIVE MAINTENANCE

MAINTENANCE SCHEDULE
Table A, illustrated below, lists all those procedures to be carried out by the user/operator and the
relative frequency schedule. Strict adherence to said schedule will guarantee the optimal operative
efficiency of the instrument.

TABLE A – MAINTENANCE SCHEDULE
FREQUENCY

PROCEDURE

NOTES

Check the levels of all the wash
DAILY – Before launching
solutions (Rinse, Water and See Table B
“Start Work”
Cleaning)
DAILY – Before launching Check the levels of Reagents,
See Table B
“Start Work”
Standards and Controls
Check the levels of the Waste
DAILY – Before launching
Bottles and, if necessary, empty
“Start Work”
them
See the User’s Manual,
DAILY – Before launching
Carry out a WBL cycle
Chapter 03– Description
“Start Work”
of Instrument Software
Clean the Sampling Probe using
DAILY – After Shutdown
See Procedure 7.3
either paper towelling or gauze

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 2

Chapter 07 - MAINTENANCE

TABLE B – MAINTENANCE SCHEDULE
FREQUENCY
EVERY TWO WEEKS
EVERY TWO WEEKS

PROCEDURE

NOTES

Clean the Wash Station’s four
See Procedure 7.4
cannulas
Clean the wash solution bottles
See Procedure 7.5
(Rinse, Water and Cleaning)

ONCE A MONTH

Clean the Tip

See Procedure 7.6

ONCE A MONTH

Clean the Hydraulic Circuit

See Procedure 7.7

EVERY THREE MONTHS

Change the Reaction Cuvettes

See Procedure 7.10

EVERY SIX MONTHS

Change the Peristaltic Pump Tubes

See Procedure 7.8

EVERY SIX MONTHS

Change the Tip

See Procedure 7.6

ONCE A YEAR

Change all the tubes (Tube Kit)

See Procedure 7.11

EVERY 2000 HOURS

Change the Photometer lamp

See Procedure 7.9

•

N. B.: the above-described maintenance schedule refers to that situation in which the
workload of the Analyzer is approximately 500 tests per day. The interval frequency
may vary according to the individual instrument’s daily workload.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 3

Chapter 07 - MAINTENANCE

7.2

LIST OF PARTS SUBJECT TO WEAR AND USAGE

Description
Reagents containers
Reagents containers
Samples cups
Short samples cups
Adapter for short samples cups
Reaction sectors
Washing solution bottle
Bottle level sensor
Tubing Kit for peristaltic pump
Tubing Kit – complete
Cleaning solution
Rinse solution
Sampling probe (internal needle)
Complete Sampling probe
Drying Pad
Halogen Lamp (6 V - 10 W)
Interferential filters Kit
Fuse 6,3 A-T
Inlet/outlet fitting for Rinse & Clean conts

Type
35 ml
6 ml
0.8 ml
1 ml

2 lt

250 ml
50 ml

5x20

Cuvettes protection cover
Reagent protection cover
Reagent plate
Samples rack
Washing station, first or second cannulas (A)
Washing station, third cannula (B)
Washing station, fourth cannula (C)
Washing station, fifth cannula without pad (D)
Diluter Micro-Pump
Air Micro-Pump (µP 6)
Micro-Pump (µP 2 ÷5)
Predilution rack
Solenoid Valve –2 way
Solenoid Valve –3 way

Ellipse

Service Manual

Rev. 03

Quantity
12
12
1000
1000
1
6
1
1
3
1
2
1
1
1
1
1
1
10

Code
C101-00190-00
C101-00191-00
AS-65-0002
AS-65-0100
9-01-0609-00
C101-00217-00
9-35-0041-00
9-05-0078-00
65-01835-00
65-01836-00
ASRN0020
ASRN0021
05-00707-00
10-00703-00
01-01920-00
9-35-0016-00
9-65-0029-00
C130-01238-08

1

01-01224-00

1
1
1
1
1
1
1
1
1
1
1
1
1
1

05-01249-00
10-00584-00
10-00585-00
05-01829-00
05-01633-00
05-01633-01
05-01638-00
05-01919-00
05-01710-40
05-01711-20
05-01826-16
05-01735-00
9-35-0035-00
9-35-0036-00

01 June 2009

Page 4

Chapter 07 - MAINTENANCE

7.3

7.4

7.5

SAMPLING PROBE – CLEANING PROCEDURE
1)

Turn off the Analyzer

2)

Use only lint-free paper towelling or gauze

3)

Dampen the gauze or paper towelling with distilled water and clean the outside of the
sampling probe. Wipe the probe from the top downwards only! This is to avoid that
any bits of cloth, paper or lint fibres accidentally enter the probe itself.

4)

The manufacturer suggests that once weekly the above-described cleaning procedure
be performed using, instead of only simple distilled water, a 5% sodium hypochlorite
solution to dampen the gauze and then be repeated using distilled water.

WASH STATION CANNULAS – CLEANING PROCEDURE
1)

Turn off the Analyzer.

2)

Place a sheet of paper under the wash station cannulas in order to keep any extraneous
material from accidentally falling into the cuvettes.

3)

Use only lint-free paper towelling or gauze.

4)

Dampen the gauze or paper towelling with distilled water and clean the outside of the
cannulas. Wipe the cannulas from the top downwards only! This is to avoid that any
bits of cloth, paper or lint fibres accidentally enter the cannulas themselves.

5)

The manufacturer suggests that once weekly the above-described cleaning procedure
be performed using, instead of only simple distilled water, a 5% sodium hypochlorite
solution to dampen the gauze and then be repeated using distilled water.

WASH SOLUTION BOTTLES – CLEANING

During normal use and over time, mold and dust can build up inside the wash solution bottles.
For this reason, it is extremely important that they be periodically washed. Said cleaning must be
thorough and meticulous in order to insure that every trace of mold or residue be removed.
How often the bottles must be cleaned depends on their use and on the quality of the distilled water
used in that particular laboratory. However, the manufacturer recommends thorough washing at least
once every two weeks.
It is extremely important that the user not underestimate the risks associated with mold and dust
particles. They are to be regarded as a serious hazard as they can be the cause of instrument
malfunction.
The wash solution bottles are located on the right side of the analyzer.
Ellipse

Service Manual

Rev. 03

01 June 2009

Page 5

Chapter 07 - MAINTENANCE

7.5.1 WASH SOLUTION BOTTLES - CLEANING PROCEDURE
1) Turn off the Analyzer.
2) Pull the level sensor connectors out from the bottle caps.
3) Take the caps off the bottles and empty them.
4) Fill each bottle with a 5% sodium hypochlorite solution.
5) Clean the inside of each bottle using a bottlebrush in order to remove all traces of mold
and/or residue.
6) Leave the sodium hypochlorite solution stand in the bottles for at least ten minutes.
7) Empty the bottles, rinse them repeatedly and well with tap water, and then twice more
using distilled water.
8) Dry the bottles.
9) Fill the bottles with their proper solutions.
10) Replace the bottles in their respective housings.
11) Close the bottles and reconnect the level sensors.
12) Carry out two ‘Wash Cuvettes’ cycles and two ‘WBL’ cycles. Check system efficiency
by comparing the WBL values and the test results obtained against the laboratory’s
quality control values.

7.6

TIP WASHING

The Tip is used to dry the cuvettes after they have been washed. This drying process is carried out via
aspiration and therefore, over time, the Tip will necessarily absorb various contaminating particles.
The manufacturer suggests that the Tip be replaced every six months. Said frequency may vary
depending on the workload of the individual laboratory and the operating conditions/environment of
the single instrument.
The Tip must be washed regularly in order to guarantee proper functioning and must be replaced with
a new Tip as necessary.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 6

Chapter 07 - MAINTENANCE

7.6.1 TIP - CLEANING PROCEDURE
1.

Turn off the Analyzer.

2.

Unscrew the fastening screw and remove the Washing station protective cover (Fig. 1).

3.

Remove the top cover of the washing station unscrewing the four fastening screws (Fig. 2).

4.

Remove the cannula that contains the Tip (the 5th cannula) and the spring (Fig. 3).

5.

Disconnect the aspiration tube (Fig. 4).

6.

Once the cannula has been removed, immerge it in a 5% solution of sodium hypochlorite for at
least 15 minutes (Fig. 5).

7.

Attach a 10 ml syringe to the cannula and aspirate and dispense the hypochlorite solution
through the cannula (and the Tip) until the latter is completely clean (Fig. 6). This aspirating
and dispensing forces the liquid through the Tip fibers in both directions

8.

Once the Tip is clean, repeat the aspirating and dispensing cycle 10 more times using distilled
water, then disconnect the syringe.

9.

Re-connect the aspiration tube to the cannula (Fig. 7).

10.

Re-position the cannula in its housing being careful to not kink the aspiration tube

11.

Re-position the spring on the cannula’s cylindrical axel.

12.

Re-position the top cover of the washing station, making sure that the cylindrical tops of the
cannulas fit into the corresponding holes on it.

13.

Manually move the wash station downwards and centre the tip with the cuvette turning it.

14.

Remount the Washing station protective cover (Fig. 1).
Washing Station
protective cover

Fig.1
Ellipse

Instrument Top view
Service Manual

Rev. 03

01 June 2009

Page 7

Chapter 07 - MAINTENANCE

Spring

5th cannula

Fastening screw
Fig. 2

Fig. 3

Aspiration tube

Fig. 4

Fig. 5

Sodium hypochlorite

Connect the syringe to the cannula

Fig. 6

Ellipse

Fig.

Service Manual

Rev. 03

01 June 2009

Page 8

Chapter 07 - MAINTENANCE

7.6.2

PROCEDURE FOR REPLACING THE TIP

1.

Turn off the Analyzer.

2.

Unscrew the fastening screw and remove the Washing station protective cover (Fig. 1).

3.

Remove the top cover of the washing station unscrewing the four fastening screws
(Fig. 2).

4.

Remove the cannula that contains the Tip (the 5th cannula) and the spring (Fig. 3).

5.

Disconnect the aspiration tube (Fig. 4), remove the used Tip and replace it with a new one. Press
lightly to push the new Tip into place – be careful to not press too hard as this could deform the
Tip (Fig. 8).

6.

Insert the cannula containing the new Tip into a cuvette, pushing it down inside until it takes on
the shape of the inside of the cuvette (Fig. 9).

7.

Remove the cuvette, connect the aspiration tube to the cannula and reposition the cannula back
into its proper housing in the wash station. Be careful to not kink the aspiration tube while doing
so.

8.

Reposition the spring in its proper housing. Remount the wash station coverlid making sure that
the cylindrical tops of the cannulas fit into the corresponding holes on the wash station coverlid.

9.

Manually move the wash station downwards and centre the tip with the cuvette turning it.

10. Remount the Washing station protective cover (Fig. 1).

Push the new Tip into the cuvette

Pull to remove the used Tip
Fig. 8

Ellipse

Fig. 9

Service Manual

Rev. 03

01 June 2009

Page 9

Chapter 07 - MAINTENANCE

7.7

HYDRAULIC CIRCUIT WASHING

During normal use and over time, mold and dust can build up inside the wash bottles and can have a
negative effect on the hydraulic circuit, compromising the correct functioning of the micro-pump and
valves. This, in turn, can lead to inefficiency in the sampling probe and cuvette washing system.
For this reason, it is extremely important that the hydraulic circuit be periodically washed. Said
cleaning must be thorough and meticulous in order to assure that every trace of mold or residue be
removed.
How often the hydraulic circuit must be washed depends on the operating conditions/environment of
the single instrument and the quality of the distilled water used in that particular laboratory. The
manufacturer recommends thorough washing at least once a month.
It is extremely important that the user not underestimate the risks associated with mold and dust
particles. They are to be regarded as a serious hazard as they can be the cause of instrument
malfunction.
The hydraulic circuit input cannulas are located inside the bottles on the left side of the Analyzer.

7.7.1 HYDRAULIC CIRCUIT - CLEANING PROCEDURE
1)

Turn on the Analyzer.

2)

Prepare a bottle containing 500 ml of a 5% sodium hypochlorite solution

3)

Insert the four aspiration cannulas, located inside the liquids bottles, into the bottle containing
the sodium hypochlorite solution.

4)

Have the instrument carry out a ‘WBL’ cycle and then a ‘Wash cuvettes’ cycle.

5)

Wait for fifteen minutes. Clean and dry the three cannulas and then insert them into a bottle
containing distilled water.

6)

Have the instrument carry out a ‘WBL’ cycle and then a ‘Wash cuvettes’ cycle.

7)

Insert the four aspiration cannulas back into their respective bottles. Said bottles should have,
in the meantime, been cleaned and filled with a fresh supply of the required solution.

8)

Have the instrument carry out a ‘WBL’ cycle and then a ‘Wash cuvettes’ cycle. Check system
efficiency by comparing the WBL values and the test results obtained against the laboratory’s
quality control values.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 10

Chapter 07 - MAINTENANCE

7.8

CHANGING THE PERISTALTIC PUMP TUBES

The manufacturer recommends that the Peristaltic Pump tubes be replaced every six months. Said
frequency may vary depending on the workload of the individual laboratory. The quality and
reliability of these tubes is fundamental to a correct emptying of the cuvettes.

7.8.1 PROCEDURE FOR REPLACING THE PERISTALTIC PUMP TUBES
1. Open the panel located on the right side of the instrument pushing down the looking door(Fig. 1).
2. Unhook the hinged guide by pushing the pin toward to the left (Fig. 2).
3. Pull the tubes off their relative nipples and substitute with new ones (Fig. 2).
4. After having unlooked the hinged guide, disconnect the tubes from their nipples
5. Remount, repeating the above steps, 4 through 2, in reverse order. Turn clock wise the peristaltic
pump rotor, in manual way, in order to be sure that tubes are right positioned inside the guide.
Looking door

Pin

Tubes to be replaced

Hinged guide lock

Peristaltic Pump Rotor
Fig. 1 Instrument right side

Nipples
Fig. 2

Peristaltic Pump Assembly

6. Turn on the Analyzer and wait until the instrument has reached its proper operating temperature.
7. Have the instrument carry out a ‘Wash cuvettes’ cycle and then a ‘WBL’ cycle. Check system
efficiency by comparing the WBL values and the test results obtained against the laboratory’s
quality control values.
8. Make sure that there is no leakage and then close back the panel.
•

Always use original replacement parts; never lubricate the peristaltic pump roller bearings.

•

After prolonged instrument inactivity: verify the efficiency of the peristaltic pump tubes.
Please note: if the instrument is not able to completely empty the washing well, check the condition
of the tube and make sure that there are not obstruction into the hydraulic lines.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 11

Chapter 07 - MAINTENANCE

7.9

CHANGING THE PHOTOMETER LAMP

The manufacturer suggests that the lamp be replaced after approximately 2000 hours of use.
Figure 2 illustrates the photometer lamp, its support base and its power supply wires. There is a small
hole on the lamp base useful for its mechanical alignment. The lamp is mounted on the photometer,
which is located on the left-hand side of the reaction plate.

7.9.1 PROCEDURE FOR REPLACING THE PHOTOMETER LAMP
1)

Turn off the Analyzer.

2)

Remove the front panel unscrewing the four anchored screws (Fig. 2).

3)

Disconnect the power supply wires from the Lamp Regulator Board by loosening the clamp
screws on the M1 connector (Fig. 3).

4)

Unscrew the lamp’s fastening screw and remove the lamp from its housing (Fig. 3)

5)

Replace the old lamp with a new one making sure that the pin is correctly inserted in the
alignment hole (Fig. 3). Remount the assembly by repeating the above steps 4 through 1 in
inverse order.

6)

Turn the Analyzer on and wait until the instrument has reached its proper operating
temperature.

7)

Have the instrument carry out a ‘WBL’ cycle. Check system efficiency by comparing the WBL
values and the test results obtained against the laboratory’s quality control values.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 12

Chapter 07 - MAINTENANCE

Washing Station
protective cover

Fig.1

Diluter protective cover

Instrument Top view

Instrument front panel
anchored screw

Fig. 2

Ellipse

Instrument front view

Service Manual

Rev. 03

01 June 2009

Page 13

Chapter 07 - MAINTENANCE

Halogen Lamp Cd. 9-35-0016-00
Lamp's fastening screw
M1 Connector
Photometer Lamp Board Cd. 30-01576-00
Trimmer P1
Fig. 3

WARNING! DO NOT TOUCH THE GLASS PART OF THE LAMP WITH YOUR FINGERS!
IF NECESSARY, USE A CLEAN CLOTH TO REMOVE DUST, OR ALCOHOL TO REMOVE MORE
STUBBORN DIRT.

7.10 CHANGING THE REACTION CUVETTES
Over time and through normal use the perfect transparency of the cuvettes diminishes. This less-thanperfect transparency has a negative impact on the quality of the optical readings. The manufacturer
suggests that the cuvettes be replaced after three months of use. The cuvettes racks are located inside
the reaction plate.

7.10.1 PROCEDURE FOR REPLACING THE REACTION CUVETTES
1.

Turn on the Analyzer and wait until the instrument has reached its proper operating
temperature.

2.

Click on the change cuvettes rack button

, located under the work program (Please see

Chapter 03 – Description of Instrument Software).
3.

A pull-down menu will appear. Click on the rack number 1 and then on OK, in order to have
the selected rack accessible (Fig. 1)

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 14

Chapter 07 - MAINTENANCE

Fig. 1

Reaction Rack cover

4.

Remove the reaction rack cover from its housing (Fig. 1).

5.

Turn the two locking pins making free the reaction rack.

6.

Remove the reaction rack from its housing and replace it with a new one (Fig. 2).

7.

Make sure to reinsert the new rack VERTICALLY. Turn the two locking pins in order to fix
the rack. Be especially careful to not touch the external surface of the cuvettes dedicated to
photometric reading.

8.

Repeat the procedure from point 2 to 7 in order to replace the reaction racks from 2 to 6.

9.

Have the instrument carry out a ‘WBL’ cycle. Check system efficiency by comparing the WBL
values and the test results obtained against the laboratory’s quality control values.

Reaction Rack

Fig. 2

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 15

Chapter 07 - MAINTENANCE

locking pin

Fig. 3

7.11

CHANGING THE TUBES KIT

Over time and through normal use, the tubes become worn.
The manufacturer suggests that the tubes be replaced at least once a year. These tubes are located
inside the instrument to link the Washing Station, the Peristaltic Pump and the Washing Well to the
Washing Pump Assembly.

7.11.1 PROCEDURE FOR REPLACING THE TUBES KIT
2.

Turn off the Analyzer.

3.

Unscrew the four screws to remove the panel located on the rear side of the instrument to
access the washing pump assembly (Fig. 1 and Fig 2).

4.

Replace the tubes following the indications provided in the hydraulic diagram
(SI-00571-XX).

5.

Turn on the Analyzer and wait until the instrument has reached its proper operating
temperature.

6.

Have the instrument carry out a ‘Wash cuvettes’ cycle and then a ‘WBL’ cycle. Make sure that
there is no leakage.

7.

Ellipse

Check system efficiency by comparing the WBL values and the test results obtained against the
laboratory’s quality control values.

Service Manual

Rev. 03

01 June 2009

Page 16

Chapter 07 - MAINTENANCE

Fig. 1

Instrument rear view

Fig.2

Pumps module

7.12 TABLE C – LIST OF THOSE MAINTENANCE PROCEDURES THAT
CAN BE PERFORMED BY THE USER AND/OR BY THE
MAINTENANCE TECHNICIAN
PROCEDURE

FREQUENCY (*)

EFFECTUATED BY

CLEANING THE SAMPLING PROBE

DAILY

OPERATOR

CLEANING THE FOUR WASH STATION
CANNULAS

EVERY TWO
WEEKS

OPERATOR

CLEANING THE WASH SOLUTION BOTTLES

EVERY TWO
WEEKS

OPERATOR

REPLACING REACTION CUVETTES

EVERY THREE
MONTHS

OPERATOR

CLEANING THE TIP

ONCE A MONTH

OPERATOR

REPLACING THE TIP

AS NEEDED

OPERATOR

REPLACING THE PERISTALTIC PUMP TUBES

EVERY SIX
MONTHS

OPERATOR

REPLACING THE PHOTOMETER BULB

AFTER 2000
HOURS OF USE

OPERATOR

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 17

Chapter 07 - MAINTENANCE

PROCEDURE

FREQUENCY (*)

EFFECTUATED BY

REPLACING THE SAMPLING PROBE

AS NEEDED

OPERATOR

CLEANING THE HYDRAULIC CIRCUIT

AS NEEDED

OPERATOR

REPLACING THE TUBES (Tubes Kit)

ONCE A YEAR

MAINTENANCE
TECH

REPLACING THE PRE-HEATER AND THE
SENSOR LEVEL

AS NEEDED

MAINTENANCE
TECH

MECHANICAL ALIGNMENT OF THE SAMPLING
ARM

AS NEEDED

MAINTENANCE
TECH

REPLACING OR ADJUSTING THE OPTIC
SENSORS

AS NEEDED

MAINTENANCE
TECH

REPLACING THE BELT

AS NEEDED

MAINTENANCE
TECH

REPLACING A MOTOR

AS NEEDED

MAINTENANCE
TECH

ALIGNING THE WASH STATION/REACTION
PLATE

AS NEEDED

MAINTENANCE
TECH

ALIGNING THE REACTION OR REAGENT OR
SAMPLE PLATE

AS NEEDED

MAINTENANCE
TECH

REPLACING THE PHOTOMETER

AS NEEDED

MAINTENANCE
TECH

ADJUSTING THE PHOTOMETER

AS NEEDED

MAINTENANCE
TECH

REPLACING THE ELECTRONIC BOARDS/CARDS
AND THE MECHANICAL MODULES

AS NEEDED

MAINTENANCE
TECH

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 18

Chapter 07 - MAINTENANCE

N. B.: the above-indicated frequency intervals may vary according to the individual instrument’s
daily workload.
7.13 DECONTAMINATION PROCEDURE
Before replacing any instrument parts, repairing any defective items or performing any instrument
maintenance procedure(s), the operator or maintenance technician must carry out the belowdescribed decontamination procedure of the instrument part(s) involved in the operation(s).
This procedure can be performed on:
the entire Analyzer
those part(s) of the instrument subject to possible contamination
Material necessary
-

an ESOFENOL solution diluted to 6% (60 cc in one liter of distilled water).

ESOFENOL is

an antibacterial and antiviral substance.
-

Rubber gloves

-

Mask

-

Lab coat

External surfaces and individual parts
Spray the solution all over the instrument, paying particular attention to wetting:
-

the sampling arm

-

the reaction plate (including the cuvettes)

-

the reagent and standard and control plate

-

the sample plate (including the racks)

-

the instrument chassis

Allow the solution to stand for approximately 30 minutes
Wipe the solution off the instrument and the various components using a sponge dampened
with distilled water
Carry out a decontamination of the internal hydraulic circuit.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 19

Chapter 07 - MAINTENANCE

Hydraulic circuit (Entire Instrument)
Fill a container with a 5% sodium hypochlorite solution.
Disconnect the three silicon aspiration tubes from the nipples on the caps of their respective
tanks.
Immerge all three tubes into the container filled with the 5% sodium hypochlorite solution and
have the instrument carry out two ‘Wash cuvettes’ cycles and then two ‘WBL’ cycles.
Remove the three tubes from the 5% sodium hypochlorite solution and immerge them in another
container containing the instrument Rinse Solution.
Have the instrument carry out one ‘Wash cuvettes’ cycle and then one ‘WBL’ cycle.
Reconnect the three aspiration tubes to the nipples on their respective tanks.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 20

Chapter 08 – Host Communication

CHAPTER 08
– HOST COMMUNICATION –
INDEX

8.1 COMMUNICATION WITH THE HOST COMPUTER .............................................................. 2
8.1.1 COMMUNICATION PARAMETERS ...................................................................................... 2
8.2 PROTOCOL SPECIFICATIONS.................................................................................................. 3
8.3 Header Record (H) ........................................................................................................................ 4
8.4 Patient Record (P) ......................................................................................................................... 6
8.5 Test Order Record (O) .................................................................................................................. 8
8.6 Results Record (R) ...................................................................................................................... 11
8.7 Message Terminator Record (L) ................................................................................................. 13
8.8 Request Information Record (Q): ............................................................................................... 13

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 1

Chapter 08 – Host Communication

8.1 COMMUNICATION WITH THE HOST COMPUTER
The Ellipse can be connected to a host computer for the purpose of facilitating results print-out and
patient management.
In order to enable communication between the Ellipse and the host computer, select the Host Link
field under Options in the Parameters menu.
To activate communication between the Ellipse and the host computer, select Host–Tx (please see
the software description in Chapter 03 of the User’s Manual)

8.1.1 COMMUNICATION PARAMETERS
Communication parameters can be modified in the Parameters.ini file into the Analyzer 140 folder
for every need.
The Ellipse PC is linked to the HOST computer using RS-232C serial connector having as default
communication parameters the following specifications:
o
o
o
o
o
o

Transmission method
Baud Rate
Data bits:
Parity
Stop bit
Connector

:
:
:
:
:
:

Asynchronous, half duplex
9600 Bit/sec.
8
None
1
9 pin type D (male output from the Ellipse computer)

1-CF(DCD)

(DSR)CC-6

2-BB(RD)

(RTS)CA-7

3-BA(TD)

(CTS)CB-8

4-CD(DTR)

(RI)CE-9

5-AB(GND)

Serial connector

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 2

Chapter 08 – Host Communication

ELLIPSE PC
DB 9 F

HOST PC
DB 25 F

HOST PC
DB 9 F

3 Txd

Rxd

3

2

2 Rxd

Txd

2

3

5 SG

SG

7

5

8.2 PROTOCOL SPECIFICATIONS
This part of Help (Protocol Specifications) contains information for the laboratory computer and
analyzer. This exchange of data follows specific ASTM protocols:
E 1381-95 Standard Specification for Low-Level Protocol to Transfer Messages
between Clinical Laboratory Instruments and Computer Systems;
E 1394-97 Standard Specification for Transferring Information between Clinical
Instruments and Computer Systems.
ASTM uses a number of different terms to indicate the way it groups data.
•

Field: an individual piece of data often referred to as a data field or a data element.

•

Record: a number of logically related data fields grouped together to form one part of a
complete message.

•

Repeat field: a data field of the same type as the one immediately preceding it. A delimiter
separates one instance of a repeat field from the next.

•

Component field: part of data field that might contain more than one piece of data.

The default communication configuration for the Analyzer is the following: "9600,N, 8,1".
ASTM uses record types that are common and familiar to all laboratory personnel. It uses the
following record types:
•

Header Record (H): contains identifying information about the sending station,
conventions that the device uses for field recognition, and the date and time of send station
transmission.

Ellipse

Service Manual

Rev. 04

01 October 2009

Page 3

Chapter 08 – Host Communication

•

Patient Record (P): contains patient information and identification number.

•

Test Order Record (O): contains information about the assay or requests themselves and
includes other data.
Result Record (R): contains information about the outcome of individual tests for an

•

individual patient and follows a sample program record. The results contain the actual
measurements derived from the test and a comparison of the individual result to certain
ranges specified as norms for the laboratory.
Message Terminator Record (L): although the ASTM protocol supports three additional

•

record types - a Request for Information Record, a Scientific Record and a Manufacturer's
Information Record - the Analyzer is not implementing these in the first release and will
ignore them.
Request Information Record (Q): is used by either clinical instruments or computer

•

systems for a remote request for information from its reciprocal system.
The instrument does not send or accept comment records.

8.3 HEADER RECORD (H)
Field

Field Title

Down
Load

Up
Load

Max
Len

Description and Valid Values

1

Record Type
ID

R

A

1

This is a required field that contains an
“H” identifying it as a header record.

2

Delimiters

I

A

4

The Analyzer System uses only the four
default values shown here. Delimiters
may not be duplicated. The field
delimiter follows the escape character to
separate the delimiter specification from
a subsequent field in the header record.
Using default values, the first six
characters of the header record will
appear using the following characters:
H I\^&I

Ellipse

Service Manual

Field Delimiter

I

Repeat Delimiter

\

Component Delimiter

^

Rev. 04

01 October 2009

Page 4

Chapter 08 – Host Communication

Escape Delimiter

3
4
5
6
7
8
9
10

11

12
13
14

Legend:

Ellipse

Message
Control ID
Access
Password
Sender Name
or ID
Sender Street
Address
Reserved Field

I

N

I

N

I

A

I

N

I

N

Sender Tel.
Number
Characteristics
of Sender
Receiver ID

I

N

I

N

I

N

Comments or
Special
Instructions
Processing

I

N

I

N

ASTM Version
No.
Date and Time

I

N

I

A

R
O
I
A

Required
Optional
Ignored
Always

Service Manual

10

&

‘SHAnalyzer’: This is the name of the
device that is sending the data.

14

D
U
N
S

Date and Time of transmission:
formatted as YYYYMMDDHHMMSS.
For example: 3:35 PM on March 1, 1995
would be represented using the
following characters:
19950301153500.
Down Load
Up Load
Never
Sometimes
Rev. 04

01 October 2009

Page 5

Chapter 08 – Host Communication

Example Header Record Layouts (H)
Download
Host

H I \ ^ I I I HOST I I I I I I I I I19950301153500
Upload
H | \ ^ & | | | SHAnalyzer | | | | | | | | | 19950301154000

Analyzer System

8.4 PATIENT RECORD (P)
Down
Load

Up
Load

Max
Len

Record Type ID

R

A

1

2

Sequence Number

R

A

3

3

Practice Assigned
Patient ID

R

A

15

4

I

N

5

Laboratory
Assigned
Patient ID
Patient ID No. 3

I

N

6

Patient Name

O

S

7

I

N

8

Mother's Maiden
Name
Birth Date

O

9

Patient Sex

10

Patient
Race/Ethnic
Origin

Fi
el
d

Field Title

1

Ellipse

Description and Valid Values

This is a required field that contains a
“P” identifying it as a patient record.
This field starts with a “1” for the
patient and is incremented by 1 for each
additional patient within the
transmission.
This field can be assigned by the
instrument with no corresponding
download.

36

This field has two components:
• Last Name (up to 20 characters)
• First Name (up to 15 characters).

S

8

R

A

1

Formatted as YYYYMMDD: For
example, a birth date of December 1,
1980 would be represented as:
19801201
The valid values are:
• M for Male
• F for Female

I

N

Service Manual

The Analyzer System will ignore this
field at launch.

Rev. 04

01 October 2009

Page 6

Chapter 08 – Host Communication

11

Patient Address

O

S

12

Reserved Field

I

N

13

I

N

I

N

15

Patient Tel
Number
Attending
Physician ID
Special Field 1

I

N

16

Special Field 2

I

N

17

Patient Height

I

N

18

Patient Weight

I

N

19

Patient Known or
Suspected
Diagnosis

I

N

20

Patient Active
Medications
Patient's Diet
Practice Field No.
1
Practice Field No.
2
Admission Date
and Discharge
Date (if desired)

I

N

I
I

N
N

I

N

O

S

14

21
22
23
24

25

Admission Status

I

N

26

Location

O

S

27

Nature of
Alternative
Diagnostic Code
and classifiers
Alternative
Diagnostic Code
and classification

I

N

I

N

29

Patient Religion

I

N

30

Marital Status

I

N

28

Ellipse

Service Manual

60

For Analyzer, this is a four- component
field:
• Address (25 characters)
• City (25 characters)
• State (2 characters e.g.: NY, IT)
• Zip (5 characters)

8

Admission date only.
Formatted as YYYYMMDD.

20

Rev. 04

01 October 2009

Page 7

Chapter 08 – Host Communication

31

Isolation Status

I

N

32

Language

I

N

33

Hospital Service

I

N

34

Hospital
Institution
Dosage Category

I

N

I

N

35

Legend:

R
O
I
A

Required
Optional
Ignored
Always

D
U
N
S

Down Load
Up Load
Never
Sometimes

Example Patient Record (P)
Download
Host

P | 1 | B108K | | | MW5910^Smith | | 19861002 | M | | Park Avenue^New
York^NY^10002 | | | | | | | | | | | | | 20020923 | | Hematology | | | | | | | | | |

Analyzer
System

P | 1 | B108K | | | MW5910^Smith | | 19861002 | M | | Park Avenue^New
York^NY^10002 | | | | | | | | | | | | | 20020923 | | Hematology | | | | | | | | |

8.5 TEST ORDER RECORD (O)
Field

Field Title

Down
Load

Up
Load

Max
Length

Description and Valid Values

1

Record Type
ID

R

A

1

This is required field that contains
an “O” identifying it as an order

2

Sequence
Number

R

A

3

This field starts with “1” for the first
Test Order Record and is
incremented by 1 for each additional
Test Order Record within the
record.
This will be reset to “1” whenever
another patient record is transmitted.

3

Ellipse

Specimen ID

R

Service Manual

A

15

Rev. 04

Although the operator can manually
edit this field at any time, the value
of this field is usually assigned by
the laboratory computer before
down loading. The Analyzer uses
and reports its results based on the
assigned specimen ID.
01 October 2009

Page 8

Chapter 08 – Host Communication

4

Instrument
Specimen ID

I

N

5

Universal Test
ID

I

N

I

N

I

N

R

A

6

Priority

I

N

7

Request
Ordered
Date/Time
Specimen
Collected
Date/Time
Collection End
Time

I

N

I

N

I

N

10

Collection
Volume/Units

I

N

11

Collector ID

I

N

12

Action Code

I

N

13

Danger Code

I

N

14

Relevant
Clinical Info.

I

N

15

Date/Time
Specimen
Received

I

N

8

9

Ellipse

Service Manual

9

Rev. 04

This is a four-component field:
• Universal Test ID Code (not
used)
• Universal Test ID Name (not
used)
• Universal Test ID Type (not
used)
• Manufacturer's or local code
(6 characters):
This is the code defined in the
Analyzer.

01 October 2009

Page 9

Chapter 08 – Host Communication

16

Specimen
Type

R

A

17

Ordering
Physician

I

N

18

Physician Tel.
Number

I

N

19

User Field No.
1

I

N

20

User Field No.
2

I

N

21

Lab Field No.
1

I

N

22

Lab Field No.
2

I

N

23

Date /Time
Result
Reported Last
or Modified
Instrument
Charge

I

N

I

N

25

Instrument
Section ID

I

N

26

Record Type

I

A

27

Reserved Field

I

N

28

Location or
Ward of
Specimen
Collection
Nosocomial
Infection Flag

I

N

I

N

30

Specimen
Service

I

N

31

Specimen
Institution

I

N

24

29

Ellipse

Service Manual

1

This is a numeric field indicating the
type of specimen:
The Imm. System uses the following
ASCII characters:
0= Serum
1= Urine

1

The field indicates the direction of
the transmission:
O - Down Loading
F - Up Loading

Rev. 04

01 October 2009

Page 10

Chapter 08 – Host Communication

Legend
:

R
O
I
A

Required
Optional
Ignored
Always

D
U
N
S

Down Load
Up Load
Never
Sometimes

Example Test Order Record Layouts (O)
Download
Host

O | 1 | AR102 | | ^^^GLU | | | | | | | | | | | 0 | | | | | | | | | | O | | | | |

Analyzer System

O | 1 | AR102 | | ^^^GLU | | | | | | | | | | | 0 | | | | | | | | | | F | | | | |

8.6 RESULTS RECORD (R)
Field

Field Title

Down
Load

Up
Load

Max
Len

1

Record Type ID

A

1

This is a required field that contains an
“R” identifying it as a Results Record.

2

Sequence
Number

A

3

3

Universal Test ID

This field starts with “1” for the first
result and is incremented by 1 for each
additional result within the record.
This will be reset to “1” when the
results from another test order record
are transmitted to the laboratory
computer.
This is a four-component field:
• Universal Test ID Code
(not used)
• Universal Test ID Name
(not used)
• Universal Test ID Type
(not used)
• Local Manufacturer's or local
code (6 characters) this is the
code defined in the Analyzer.

I

N

I

N

I

N

R

4

Ellipse

Data or
Measurement
value

Service Manual

9

A

A

10

Rev. 04

‘Data’ is a 10-character, floating point
field that includes the decimal point.
The number of precision point digits
will vary according to the test and is
configurable on the Analyzer.
01 October 2009

Page 11

Chapter 08 – Host Communication

5

Units of Measure

A

6

This is a field for up to 6 characters
that the operator defines for analytic
measurement.

6

Reference Ranges

A

21

This field has two components; one
giving the lower limit and the other the
upper limit of the range. The format
for this field is N^N.

7

Result Abnormal
Flags

A

2

This field indicates the normal status
of the result. The following codes are
valid values:
L - Below Low normal
H - Above High normal
LL - Below Panic normal
HH - Above Panic normal
< - Below absolute low (under
linearity)
> - Above absolute high (over
linearity)
N - Normal
A - Abnormal
E – Edited

8

9

Nature of
Abnormality
Testing
Result Status

N

A

1

The Imm. System currently
implements only two valid values:
F - final results;
V - operator verified/approved result.

10

Date of Change in
Instrument
Normative Values
or Units
Operator ID

N

12

Date/Time Test
Started

N

13

Date/Time Test
Completed

A

11

Ellipse

Service Manual

N

14

Rev. 04

Date and Time of test completion:
formatted as
YYYYMMDDHHMMSS.

01 October 2009

Page 12

Chapter 08 – Host Communication

14

Instrument ID

N
R
O
I
A

Legend:

Required
Optional
Ignored
Always

D
U
N
S

Down Load
Up Load
Never
Sometimes

Example Result Record Layouts (R)
Upload
R | 1 | ^^^GLU | 70.97 | UL | 70^105 | N | | F | | | | 20020923114302 |

Analyzer System

8.7 MESSAGE TERMINATOR RECORD (L)
Field

Field Title

Down
Load

Up
Load

Max
Len

1

Record Type
ID

R

A

1

2

Sequence
Number
Termination
Code

R

A

1

R

A

1

3

Description and Valid Values
This is a required field that contains
an “L” identifying it as an Message
Terminator Record.
For a message terminator, this
message should always be “1”.
This indicates the cause of
termination. The following codes are
valid values for the Analyzer :
Null or N-normal termination

R
O
I
A

Legend:

Required
Optional
Ignored
Always

D Down Load
U Up Load
N Never
S Sometimes

Example Message Terminator Record Layout (L)
Host

LI1IN

Analyzer System

LI1IN

8.8 REQUEST INFORMATION RECORD (Q):
Field
1

Ellipse

Field Title
Record Type ID

Service Manual

Down
Load

Up
Load

Max
Len

Description and Valid Values

A

1

This is a required field that
contains a “Q” identifying it as a
request.

Rev. 04

01 October 2009

Page 13

Chapter 08 – Host Communication

2

Sequence Number

A

1

It is always “1”.

3

Starting Range ID
Number

A

31

This field can either be:
"ALL" to mean all
demographics and tests being
ordered should be sent to the
instrument at this time,
or can have two components:
• Computer system patient
ID No. (up to 15
characters);
• Computer system
specimen ID No. (up to 15
characters).

4
5
6

Ending Range ID
Number
Universal Test ID

N
N
N

11

Nature of Request
Time Limits
Beginning Request
Results Date and
Time
Ending Request
Results Date and
Time
Requesting
Physician Name
Requesting
Physician
Telephone Number
User Field No. 1

12

User Field No. 2

N

13

Request
Information Status
Codes

A

7

8

9
10

N

N

N
N

N
1

It is always "O" (requesting test
orders and demographics only).

Example Request Information Record Layouts (Q)
Download To
Analyzer System

Ellipse

H | \^& | | | SHAnalyzer | | | | | | | | | 20020927100402
Q | 1 | ALL | | | | | | | | | | O
L|1|N

Service Manual

Rev. 04

01 October 2009

Page 14

Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING

CHAPTER 09 - ERROR SIGNALING AND
TROUBLESHOOTING
INDEX

9.1

ERROR SEGNALING ........................................................................................................2

9.1.1 SYSTEM ERRORS .............................................................................................................2
9.1.2 RESULT FLAGS.................................................................................................................3
9.1.3 DESCRIPTION OF RESULT FLAGS................................................................................4
9.2 TROUBLESHOOTING GUIDE .........................................................................................7

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 1

Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING

9.1

ERROR SIGNALING

This chapter is dedicated to a description of the error signalling which may occur during the
programming or carrying out of the various operations. Error signals can be divided into the
following two groups:
➢

System Errors

➢

Result Flags

9.1.1 SYSTEM ERRORS
Whenever a system error is detected, it is signalled via the activation of the below-listed
warning lights/buttons:
➢ Warning Light: a triangle-shape located in the upper, right-hand portion of the System

Monitor screen. It lights up yellow when activated.
If this yellow triangle does light up, the operator need simply click on it to access the relative
visual text warning message indicating the cause of the signalled anomaly (said window will
open in that area dedicated to viewing data).
Following is a list of possible visual text “Warning!” messages:
•

Liquid Alarm I (Rinse Solution)

•

Liquid Alarm II (Distilled Water)

•

Liquid Alarm III (Cleaning Solution)

•

Liquid Alarm IV (Optional Solution)

•

Temperature Out-of-Range

•

Host Serial Port cannot be opened

➢ Fatal Error: an “X”-shape located in the upper, right-hand portion of the System Monitor

mask. It lights up red when activated.
If this red “X” lights up, the operator need simply click on it to access the relative visual text
message explaining the cause of the Fatal Error (said window will open in that area of the mask
dedicated to viewing data).
Following is a list of possible Fatal Error visual text messages:
•

Horizontal Arm Error

•

Sample Tray Error

•

Filter Error

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 2

Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING

•

ADC Error – Sample Channel

•

ADC Error – Reference Channel

•

Reaction Tray Position Error

•

Plate Error (Home sensor not found)

•

Vertical Arm error

•

Reagent Tray Error

•

Wash System Error

•

Diluter Error

•

Interlock Open while Instrument is running

•

Macro send Error

•

Overflow Rx Buffer Micro

•

Time out Error (Generic error)

•

Invalid command error

•

Invalid parameter error

•

Check sum error

9.1.2 RESULT FLAGS
Result Flags are categorized under the following group headings, Each group is identified by a
symbol, as listed here-below:
X

PHYSICAL ERRORS

R

CONCENTRATION ERRORS

C

CALIBRATION ERRORS

A

OPTIC DENSITY ERRORS

E

RESULTS EDITED MANUALLY

?

PROGRAMMING ERRORS

Every Result Flag, signalling an error, is accompanied by a symbol representing the group it is
part of. The operator need only click on the small red square ( ) next to the Result Flag symbol
to access the visual text message explaining the cause(s) of the signalled error.
In the central column of the following tables, the user will find those symbols which signal the
type of error encounter, as used in the print-out of the final report. These symbols can be
modified by the operator, in the Parameters section, under print options.
Ellipse

Service Manual

Rev. 03

01 June 2009

Page 3

Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING

WARNING! The use of the Error Symbols in the print-out of the results (inclusion and/or
exclusion) IS UNDER THE DIRECT AND SOLE RESPONSIBILITY OF THE USER

9.1.3 DESCRIPTION OF RESULT FLAGS

¾

X Physical Errors

Temperature Error:

T

Reaction temperature (Reaction Plate) is
out-of-range.
Either no sample or sample serum quantity
below minimum or above maximum level
for the declared container.

No Sample

S

No Reagent:

R

Either no reagent or reagent level below
minimum or above maximum level for the
declared container.

No Rack:

*

No Rack present during sampling.

¾ R Concentration Errors
Very Low and Very High:

L-H

Flags determined by test results out-of-range as
setup in the Methods.

Low Alert and High Alert:

A

Flags determined by test results out-of-range as
setup in the Methods.

Low and High Linearity Limit:

G

Flags determined by test results out-of-range as

Calculation Error:

C

Concentration calculation error due to
foreseeable causes (Asymptote).

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 4

Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING

¾

C Calibration Errors

RBL missing:

*

No Reagent Blank Level.

Calibration missing:

*

No Standard or no Calibration curve.

STD Replicate insufficient:

*

Insufficient number of valid Standard Replicates.

STD Replicate outside CV%:

*

Coefficient of Variation Percentage in the Standard
Replicates over the set value.

Invalid Calibration:

*

Calibration curve not valid – either because it is
not monotonic or because the Fit is above the set
value.

¾

A Optic Density Errors

Inversion:

I

Reaction direction not in line with that set-up.

End Point Limit:

P

Values over the limits setup in the Methods
Parameters

Depletion Limit:

D

Values over the limits setup in the Methods
Parameters.

First Limit:

*

Values over the limits setup in the Methods
Parameters.

FIT:

F

Values over the limits setup in the Methods
Parameters.

RBL out-of-range:

*

Reagent Blank Levels outside the range Setup.

Sample outside Standard:

#

Sample absorbance outside the calibration curve.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 5

Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING

¾

E

Results Edited Manually

Results Edited:

¾

?

E

This symbol automatically appears whenever the
operator has manually modified the obtained
results. This operation annuls all the symbols
indicating errors which, in this case, will not be
viewed.

Programming Errors

All those software errors which are deemed unforeseeable are indicated using this symbol.

¾

Result Asterisks

In the event that the following situations and/or error conditions occur, the system will notify
the user via a visualization of two asterisks (**) in the Results field:
PHYSICAL ERRORS
1 – very little or no sample
2 – very little or no reagent
3 – no rack
CONCENTRATION ERRORS
4 – calculation error
5 – signalling depletion limit
PROGRAMMING ERROR
6 – all cases
In the following situations and/or error conditions, the results field will contain a “0”:
1 – if the result is less than zero
2 – if the result is equal to zero
OPTIC DENSITY ERRORS
3 – Flag signalling inversion

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 6

Chapter 09 - ERROR SIGNALING AND TROUBLESHOOTING

9.2 TROUBLESHOOTING GUIDE
™ SOLUTION

PROBLEM

➢ POSSIBLE CAUSE

Repeatability of results insufficient

➢ Sample Probe dirty
™ Clean the Sample Probe as described in Chapter 07
Maintenance
➢ Hydraulic leak and/or air bubble in the hydraulic circuit
(sampling)
™ Check Sample Probe fit
™ Check the fit of the hydraulic tubes and their connection:
if necessary, substitute the tubes and/or adjust their
connections/fittings
➢ Wash solution is contaminated. If the wash solution
contains contaminating particles (e.g.: mold, dust, lint), these
micro-particles can cause errors during the WBL running.
™ Change the wash solution
™ Clean the Wash Solution Bottle(s) and carry out the
Hydraulic Circuit wash procedure as described in
Chapter 07 – Maintenance
➢ Deterioration of the Reagent(s)
™ Substitute the bad reagent(s)
➢ Reaction cuvettes not dried correctly after being washed
™ Check the Tip used to dry the cuvettes after washing to
make sure it is in good working condition. If necessary,
clean the Tip or change it, following the procedure
described in Chapter 07 – Maintenance
➢ Light bulb not stable
™ Light bulb nearing the end of its 2000-hour life cycle
duration, or premature deterioration. In both cases,
change the light bulb following the substitution
procedure described in Chapter 07 – Maintenance.

Insufficient volume/quantity of the ➢ This type of problem can present itself either when the
various Rinse, Cleaning and
instrument is first turned on due to a lack of liquid in one or
Distilled Water solutions.
more bottles, or during test running whenever the involved
liquid has been finished
™ Refill the required bottle. If the instrument is running
tests, it will automatically Pause. Wait until sampling is
suspended, then refill the involved bottle with the
required liquid (Rinse/Cleaning/ Distilled Water). Press
Start to continue running the rest of the programmed
tests. This operation will not adversely affect the already
run tests.

Ellipse

Service Manual

Rev. 03

01 June 2009

Page 7



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Create Date                     : 2009:06:12 09:14:41Z
Modify Date                     : 2010:01:19 09:15:48+01:00
Page Count                      : 229
Creation Date                   : 2009:06:12 09:14:41Z
Mod Date                        : 2010:01:19 09:15:48+01:00
Producer                        : Acrobat Distiller 5.0.5 (Windows)
Author                          : gda3
Metadata Date                   : 2010:01:19 09:15:48+01:00
Creator                         : gda3
Title                           : Microsoft Word - Cover .doc
Has XFA                         : No
EXIF Metadata provided by EXIF.tools

Navigation menu