小型環境_e_ D250 (115V) Espec SH 641 Datasheet

User Manual: D250 (115V)

Open the PDF directly: View PDF PDF.
Page Count: 14

Download小型環境_e_ D250 (115V) Espec-SH-641 Datasheet
Open PDF In BrowserView PDF
®

E stablished 1981

Advanced Test Equipment Rentals
www.atecorp.com 800-404-ATEC (2832)

Bench-Top Type Temperature
(& Humidity) Chamber
SH・SU

CAT.NO.E01111-K1209

Characteristics
Full variation to select from
With its 3 temperature range patterns
(to −60, −40 or −20℃), and 2 volumes
(20 and 60L) selection, the Bench-top
chambers offer 10 models in temperature
or temperature/ humidity configuration,
to fulfill customers' needs.

Light and small structure
The chamber structure is optimized to
save space while offering a wide test
area (22.5L for 1st model). Its relatively
l ig ht we ig ht a l lows s e t u p a l mo s t
everywhere.

Stacked chambers (on stands)

Full-size chamber performance
in a compact design
The SH-661 model achieves −60 to
+150℃ temperature range and 30 to
95% rh for humidity in a small structure.
It shows undoubtedly better performance
than any other previous bench-top
model.

A refrigeration system that
reduces energy
consumption by 55%
Our exclusive refrigerator system with
capacity control achieves 55% reduction
in power consumption compared to our
previous model. (SH/SU-221・241 models)
Test area (60L model)

Easy wiring access
The chamber features free access to the
test area. Cable ports are provided on
both right and left side to allow easy
wiring of specimens for measurement
or voltage application.

3

Characteristics
Layout arrangement thanks to
additional stand (option)
L and H stands are equipped with casters,
and allow easy storage to optimize your
testing zone. L stand is lower to fit under
H stand and stack chambers. (L=Low,
H=High)
* It is recommended to secure stands to the
floor with provided anchor bolts.

Easy water supply
Water is supplied from the front side
tank (equipped with a level sensor).
Auxiliary water tank circuit (option) is
also available for extended operation
time. (SH model)

Right-opening door (option)
The door opening can be reversed to fit
your installation space.

Stands (option)

Viewing window (option)
An optional viewing window can be
added to check specimens and wiring
during testing.
( W215 × H 215m m fo r 20L mo d el ,
W215×H315mm for 60L model)
* Standard performance of the chamber will
not be met when adding this option.

Water supply tank

Right-opening door (option)

Viewing window (option)

4

Characteristics
Ion migration evaluation
The Bench-top chamber can be
combined with ESPEC's Ion Migration
Evaluation System (AMI) to ensure
even more accurate evaluation results.

Paperless recorder (option)
Portable type, to record temperatures
from various sources, such as test area
temperature. Recording is possible on
Compact Flash Card or via USB port.

Material labeling for
easy recycling
Example: Bench-top Type Temperature (& Humidity) Chamber connected
with the Ion Migration Evaluation System (sold separately)

Paperless recorder Portable type (option)

5

Cable port plug (Material marked)

Plastic molded components are labeled
and easily detachable to make recycling
e a s ie r fo r f u t u r e d i s p o s a l of t h e
equipment.

Control operation

Key input setting
E a s y s e t t i n g of t e m p e r a t u r e a n d
humidity values, but also upper and
lower temperature (& humidity) limit
alarms, by mechanical key input.

Description of program function
Program function

Pattern
Temp.
3

8

4
5

2

Step program

7

1

6

9

Program operation up to 9 steps
Besides const ant set ting, prog ram
o p e r a t io n i s ava i l a ble by s e t t i n g
different patterns (9 step/ pattern) of
heat up and pull down temperatures, to
meet requi rements of temperat u re
characteristic test or temperature (&
humidity) cycling. The instrumentation
additionally offers various functions.

Time
Temp.
3

Temperature
gradient program

2

Remote control from your PC
Using "WEB Manager" (converter sold
separately) and an intranet to connect
the chamber and a computer makes it
possible to edit profiles and perform
monitoring from the computer.

7
8

4

1

6

9

5

*1,2
Time

1.HOLD
Temp.

2.CONST

Retains value
of last step

Termination
program

Flexible communication interface
An RS-485 connector is equipped as
standard, but RS-232C or GP-IB is also
available as option.

*1,2

END

END

Operation at
constant temp.

END

3.OFF

Time

Chamber
OFF

Time

Time

*1 Sets a program repetition frequency between a range of 1 and 99.
*2 Selects HOLD, CONST or OFF when a program is over.

Temperature (& humidity) program indicator controller
Setting

Mechanical key input

Display

7 segment LED display

Operating mode

Program operation, constant operation

Control

PID control

Setting and indication

Temp. : 0.1ºC
Humid. : 1%rh (SH only)
Time : 1 minute

Setting and
indication ranges

Temp. : −25 to +155ºC (SH-221, SU-221)
−45 to +155ºC (SH-241/641, SU-241/641)
−65 to +155ºC (SH-261/661, SU-261/661)
Humid. : 0 to 100%rh (SH only)
Time : 0 to 99 hours 59 minutes, 100 to 999 hours

Indication accuracy *

Temp. : 0.5ºC (Typ.)
Humid. : ±2%rh (Typ.) (SH only)
Time : within 30 sec. per month

Program memory capacity 9 steps per pattern (1 to 99 times)
Interface

RS-485

Auxiliary functions

Input-burnout detection
Upper and lower temp. (& humid.) limit alarm
Self-diagnostic (watchdog timer)
Alarm indication
Power cut protection
Refrigerator capacity automatic control

Battery

Lithium battery ×1

* At ambient temperature +23ºC

6

SH
Model

SH−221

SH−241

System
−20 to +150℃
(−4 to +302℉)

Temp. range

Temp. performance *1

−40 to +150℃
(−40 to +302℉)

Temp. Heat up rate
rate of
change Pull down rate

−60 to +150℃
(−76 to +302℉)

−40 to +150℃
(−40 to +302℉)

−60 to +150℃
(−76 to +302℉)

3.2℃ /min.

3.2℃ /min.

3.2℃ /min.

2.9℃ /min.

2.9℃ /min.

2.1℃ /min.

2.1℃ /min.

2.1℃ /min.

1.7℃ /min.

1.7℃ /min.

Heat up time

From −20 to +150℃ From −40 to +150℃ From −60 to +150℃ From −40 to +150℃ From −60 to +150℃
within 55 min.
within 60 min.
within 70 min.
within 70 min.
within 80 min.

Temp. extreme
achievement time
Pull down time

From +20 to −20℃
within 20 min.

From +20 to −40℃
within 50 min.

From +20 to −60℃
within 70 min.

From +20 to −40℃
within 60 min.

From +20 to −60℃
within 90 min.

−20℃

−40℃

−60℃

−40℃

−60℃

Humid. range

30 to 95%rh (Refer to diagram on page 9)

Humid. fluctuation

±3.0%rh

Heater

Nichrome strip wire heater

Refrigeration unit

Humidifier
Construction

SH−661

Temp. gradient /
2.5℃ (−20 to +100℃)
2.5℃ (−40 to +100℃) 2.5℃ (−60 to +100℃) 2.5℃ (−40 to +100℃) 2.5℃ (−60 to +100℃)
Temp. variation in space 4.0℃ (+100.1 to +150℃) 4.0℃ (+100.1 to +150℃) 4.0℃ (+100.1 to +150℃) 4.0℃ (+100.1 to +150℃) 4.0℃ (+100.1 to +150℃)

Lowest attainable temp.

Sheathed heater

System

Mechanical single-stage refrigeration system

Mechanical cascade refrigeration system

Cooler

Plate fin cooler

Refrigerator

Hermetically sealed compressor, Air-cooled condenser, Expansion mechanism: Capillary tube system

Refrigerator capacity

400W

[Unit 1: 400W ×1, Unit 2: 400W ×1]

Refrigerant

R404A

R23, R404A

Capacity

22.5 L

64 L

Chamber total load
resistance

20 kg

Inside dimensions
mm (inch) *2

W300×H300×D250
(W11.81×H11.81×D9.84)

Outside dimensions
mm (inch) *2

W440×H630×D695
(W17.32×H24.80×D27.36)

Weight

83 kg (78 for 100V type)

Power supply *3

W400×H400×D400
(W15.75×H15.75×D15.75)

W440×H690×D785
(W17.32×H27.18×D30.91)

W540×H730×D920
(W21.26×H28.74×D36.22)

105 kg

130 kg

+5 to +35℃ (+41 to +95℉)

Allowable ambient conditions
Utility requirements

SH−641

±0.3℃ (−20 to +100℃) ±0.3℃ (−40 to +100℃) ±0.3℃ (−60 to +100℃) ±0.3℃ (−40 to +100℃) ±0.3℃ (−60 to +100℃)
±0.5℃ (+100.1 to +150℃) ±0.5℃ (+100.1 to +150℃) ±0.5℃ (+100.1 to +150℃) ±0.5℃ (+100.1 to +150℃) ±0.5℃ (+100.1 to +150℃)

Temp. fluctuation

Humid.
performance *1

SH−261

Balanced Temperature & Humidity Control system (BTHC system)

100V AC 1φ50/60Hz

13.5 A

15.0 A

18.0 A

115V AC 1φ60Hz

13.0 A

14.5 A

—

200V AC 1φ50/60Hz

—

10.5 A

220V AC 1φ50/60Hz

8.0 A

9.5 A

10.0 A

*4

8.0 A

9.5 A

10.0 A

230V AC 1φ50Hz *4

7.5 A

9.5 A

10.0 A

220V AC 1φ60Hz

Noise level

*5

Exhaust heat quantity

55 dB

59 dB

61 dB

3500 kJ/h (836 kcal/h)

4000 kJ/h (955 kcal/h)

5040 kJ/h (1204 kcal/h)

*1 The performance values are based on IEC 60068-3-5:2001, JTM K07:2007 for the temperature chamber, IEC 60068-3-6:2001, JTM K09:2009 for the humidity
chamber. Performance figures are given for a +23℃ ambient temperature, 65% rh, rated power supply and no specimens inside the test area. However, the lowest
attainable temperature is given for a max. ambient temperature of +30℃. Heat-up time is the achieved time from lowest temperature to highest temperature within
temperature range.
*2 Excluding protrusions.
*3 At ambient temperature +23℃. For SH-641/661, make sure to check the power capability of your installation beforehand.
*4 Compliance with CE Marking.
*5 Measurements are to be taken in an anechoic room at a height of 1.2m from the floor and a distance of 1m from the chamber front panel
(ISO 1996-1:2003 _ A-weighted sound pressure level)

7

SU
Model

SU−221

SU−241

System
−20 to +150℃
(−4 to +302℉)

Temp. range

Temp. performance *1

−40 to +150℃
(−40 to +302℉)

Temp. gradient /
2.5℃ (−20 to +100℃)
Temp. variation in space 4.0℃ (+100.1 to +150℃)
Temp. Heat up rate
rate of
change Pull down rate

SU−661

−60 to +150℃
(−76 to +302℉)

−40 to +150℃
(−40 to +302℉)

−60 to +150℃
(−76 to +302℉)

2.5℃ (−40 to +100℃)
2.5℃ (−60 to +100℃)
2.5℃ (−40 to +100℃)
2.5℃ (−60 to +100℃)
4.0℃ (+100.1 to +150℃) 4.0℃ (+100.1 to +150℃) 4.0℃ (+100.1 to +150℃) 4.0℃ (+100.1 to +150℃)

3.2℃ /min.

3.2℃ /min.

3.2℃ /min.

2.9℃ /min.

2.9℃ /min.

2.1℃ /min.

2.1℃ /min.

2.1℃ /min.

1.7℃ /min.

1.7℃ /min.

Heat up time

From −20 to +150℃ From −40 to +150℃ From −60 to +150℃ From −40 to +150℃ From −60 to +150℃
within 55 min.
within 60 min.
within 70 min.
within 70 min.
within 80 min.

Temp. extreme
achievement time
Pull down time

From +20 to −20℃
within 20 min.

From +20 to −40℃
within 50 min.

From +20 to −60℃
within 70 min.

From +20 to −40℃
within 60 min.

From +20 to−60℃
within 90 min.

−20℃

−40℃

−60℃

−40℃

−60℃

Lowest attainable temp.

Refrigeration unit

Heater

Nichrome strip wire heater

System

Mechanical single-stage refrigeration system

Mechanical cascade refrigeration system

Cooler

Plate fin cooler

Refrigerator

Hermetically sealed compressor, Air-cooled condenser, Expansion mechanism: Capillary tube system

Refrigerator capacity

400W

[Unit 1: 400W ×1, unit 2: 400W ×1]

Refrigerant

R404A

R23, R404A

Capacity

22.5 L

64 L

Chamber total load
resistance

20 kg

Inside dimensions
mm (inch) *2

W300×H300×D250
(W11.81×H11.81×D9.84)

Outside dimensions
mm (inch) *2
Weight

Power supply

*3

W400×H400×D400
(W15.75×H15.75×D15.75)

W440×H560×D695
(W17.32×H24.8×D27.36)

W440×H620×D785
(W17.32×H24.41×D30.91)

W540×H660×D920
(W21.26×H25.98×D36.22)

78 kg (73 for 100V type)

100 kg

123 kg

+5 to +35℃ (+41 to +95℉)

Allowable ambient conditions
Utility requirements

SU−641

±0.3℃ (−20 to +100℃) ±0.3℃ (−40 to +100℃) ±0.3℃ (−60 to +100℃) ±0.3℃ (−40 to +100℃) ±0.3℃ (−60 to +100℃)
±0.5℃ (+100.1 to +150℃) ±0.5℃ (+100.1 to +150℃) ±0.5℃ (+100.1 to +150℃) ±0.5℃ (+100.1 to +150℃) ±0.5℃ (+100.1 to +150℃)

Temp. fluctuation

Construction

SU−261

Balanced Temperature Control system (BTC system)

100V AC 1φ50/60Hz

11.0 A

15.0 A

18.0 A

115V AC 1φ60Hz

10.5 A

14.5 A

—

200V AC 1φ50/60Hz
220V AC 1φ50/60Hz
220V AC 1φ60Hz

*4

230V AC 1φ50Hz *4
Noise level *5
Exhaust heat quantity

—
7.0 A

10.5 A
9.5 A

10.0 A

7.0 A

9.5 A

10.0 A

6.5 A

9.5 A

10.0 A

55 dB

59 dB

61 dB

3500 kJ/h (836 kcal/h)

4000 kJ/h (955 kcal/h)

5040 kJ/h (1204 kcal/h)

*1 The performance values are based on IEC 60068-3-5:2001, JTM K07:2007 for the temperature chamber. Performance figures are given for a +23℃ ambient temperature,
65%rh, rated power supply and no specimens inside the test area. However, the lowest attainable temperature is given for a max. ambient temperature of +30 ℃.
Heat-up time is the achieved time from lowest temperature to highest temperature within temperature range.
*2 Excluding protrusions.
*3 At ambient temperature +23℃. For SU-641/661, make sure to check the power capability of your installation beforehand.
*4 Compliance with CE Marking.
*5 Measurements are to be taken in an anechoic room at a height of 1.2m from the floor and a distance of 1m from the chamber front panel
(ISO 1996-1:2003 _ A-weighted sound pressure level)

8

TEMPERATURE & HUMIDITY CONTROL RANGE (SH)

FITTINGS

Relative Humidity (%rh)

●

100
90
80
70
60
50
40
30
20
10
0

95%rh

●

●

●

●

●

●

●

●

10

20

15℃

30

40

50

60

Temperature (℃)

70

80

90

100

●

Temperature-Humidity recorder terminal
Specimen power supply control terminal
External alarm terminal
External output terminal
Cable port
Power cable
Drain hose
Water supply tank (SH only)
Humidifying tray drain plug (SH only)
Drain socket for water sensor box (SH only)

85℃

* At ambient temperature +23℃

ACCESSORIES
●

MODEL
S

-

1
Temperature range

2: −20 to +150℃
4: −40 to +150℃
6: −60 to +150℃
●

Capacity 2: 22.5L
6: 64.0L
H: Temperature & humidity chamber
U: Temperature chamber

●

●

●

●

SAFETY DEVICES

●

●

●

●

●

●

●

●

●

●

●

●

●

●

●

●

9

Leakage breaker
Thermal fuse
Boil dry protector (SH only)
Control circuit short-circuit protective fuse
Overheat and overcool protector
Temperature switch for air circulator
Specimen power supply control terminals
Refrigerator overload relay
Refrigerator high pressure switch
Chamber door switch
Upper and lower temperature (& humidity) limit alarms
(Built-into temperature [humidity] controller)
Burn-out detection circuit
(Built-into temperature [humidity] controller)
Watchdog timer
(Built-into temperature [humidity] controller)
Refrigerator automatic delay circuit
(Built-into temperature [humidity] controller)

●

Shelf
1
Capacity
SH/ SU-221, 241, 261 : 500g
SH/ SU-641, 661
: 5kg
Max. number of shelves
SH/ SU-221, 241, 261 : 5 (Shelf pitch 35mm)
SH/ SU-641, 661
: 5 (Shelf pitch 50mm)
Connector
2P for connecting terminal for temp. & humid. recorder
SH:2/ SU:1
6P for connecting signal terminal
1
Cable port plug (φ25mm)
2
Cartridge fuse (Class A, 250V, 3A)
1
Socket adapter (100V, 115V on 20L type only)
1
Wet-bulb wick (Box of 24pcs)
1 (SH only)
Humidifying tray drain hose 2m
1 (SH only)
Drain hose for water sensor box 0.3m
1 (SH only)
User's manual (CD-R, booklet)
1 set

70
(2.8)

70
(2.8)

30 150
(5.9)

540 (21.3)
70
(2.8)

300 (11.8)

695 (27.4)

920 (36.2)

250 (9.8)

●

440 (17.3)

300 (11.8)

70
(2.8)

70
(2.8)

70
(2.8)

SH-221・241

440 (17.3)

510 (20.0)
690 (27.2)

300 (11.8)

440 (17.3)

510 (20.0)
630 (24.8)

440 (17.3)

400 (15.7)

540 (21.3)

610 (24.0)
730 (28.7)

●
250 (9.8)

300 (11.8)

30 90
(3.5)

30 90
(3.5)

●

400 (15.7)

DIMENSIONS
unit:mm (inch)

SH-261

35 (1.4)
785 (30.9)
35 (1.4)

250 (9.8)
250 (9.8)

SH-641・661

400 (15.7)

10 (0.4)

400 (15.7)

10

30

400 (15.7)

11
30

540 (21.2)

400 (15.7)
70
(2.8)

300 (11.8)

695 (25.7)

SU-641・661

920 (36.2)

10 (0.4)
250 (9.8)

●

440 (17.3)

250 (9.8)
300 (11.8)

70
(2.8)

70
(2.8)

70
(2.8)

SU-221・241

436.5 (17.1)

590 (23.2)
620 (24.4)

300 (11.8)

440 (17.3)

530 (20.8)
560 (22.0)

440 (17.3)

400 (15.7)

70
(2.8)

70
(2.8)

●
250 (9.8)

300 (11.8)

536.5 (21.1)

630 (24.8)
660 (25.9)

30

●

400 (15.7)

DIMENSIONS
unit:mm (inch)

SU-261

35 (1.4)
785 (30.9)

250 (9.8)

35(1.4)

OPTIONS
Paperless recorder (Portable type)

Temperature recorder (Portable type)

Viewing window

Records temperature of each section
such as the temperature inside the
chamber.
Data saving cycle: 5 sec
Portable type
External recording media:
CF memory card (256MB)
USB port
Language support: ENG, JPN

[SRJ25]
Temperature range: −100 to +200℃
Effective recording chart width:
100mm
Number of inputs: Temperature 6

SH/SU-221, 241, 261 W215×H215mm
SH/SU-641, 661
W215×H315mm

[Temperature type]
Temperature range: −100 to +200℃
Number of inputs (Initial setting):
Temperature 1
(5 more channels can be turned ON)
[Temperature and humidity type]
Temperature range: − 50 to +150℃
−100 to +150℃
Humidity range: 0 to 100%rh
Number of inputs (Initial setting):
Temperature 1 / Humidity 1
(4 more channels can be turned ON)

Temperature & humidity recorder (Portable type)
Effective recording chart width:
100mm
Number of inputs: Temperature 5
Humidity 1
[SRJ14]
Temperature range: −100 to +150℃
Humidity range: 0 to 100%rh (for SH)

When equipped with a viewing
window, chamber performance
are as follow:
Temperature uniformity :
±1.0℃ (−20 to +100℃)
±1.5℃ (−100.1 to +150℃)
Humidity uniformity : ±5.0%rh
Temp. heat up rate:
Std rate +10 min. (or less)
Temp. pull down rate:
Std rate +10 min. (or less)
(+20 to −35℃)
Std rate +30 min. (or less)
(+20 to −60℃)

[SRJ12]
Temperature range: −50 to +150℃/
Humidity range: 0 to 100%rh (for SH)

W215×H215mm

Inner door
Wet-bulb temperature detecting terminal

Paperless recorder Portable type

For the test area wet-bulb temperature
detection. A ther mocouple type T
(C o p p e r / C o p p e r - N i c k e l ) o r o f
equivalent thermoelectric capacity is
used. Connector included.

Glass door provided inside the chamber,
t o obse r ve t he c ond it ion s of t he
specimens.
* A wiper is included for SH models.
* This option is not compatible with a viewing
window or right-opening door.
* Used for observation of the specimens inside
the chamber.

* Not available for SU.

Thermocouple
Attached to specimens to measure
specimen temperature.
・2m
・4m
・6m

SU model

* Thermocouple type T without ball
(Copper / Copper-Nickel)
* Same as accessory items

12

OPTIONS
Right-opening door

Auxiliary water tank circuit (SH)

Stand

Door can be changed for a reverse one
(hinges on the right, handle on the left)

Used to automatically refill the standard
tank with the auxiliary water tank.
Required water quality:
Conductivity 0.1 to 10μS/cm
Water supply pressure:
4.9 to 19.6KPa (Gauge)

Enhances the mobility of the chamber,
a n d f a c i l i t a t e s l o a d / u n l o a d of
specimens.
L stand can fit under the H stand to
save space.

* This option is not compatible with inner glass
door option.

Auxiliary water tank (SH)
Used to refill the standard tank.

Tray for auxiliary tank (SH)
P revent s water lea kage du r i ng
standard tank refill process.

Additional cable port

* Be sure to secure the stand to the floor with
the provided anchor bolts.

・ L stand
SH/SU-221, 241:
W440×H450×D750mm (25kg)
SH/SU-261:
W440×H390×D750mm (21.5kg)
SH/SU-641, 661:
W540×H350×D860mm (28kg)
・ H stand
SH/SU-221, 241, 261:
W630×H1140×D815mm (36kg)
SH/SU-641, 661:
W730×H1140×D925mm (40kg)

Shelf

Provided in addition / replacement of
the standard cable ports.
・ 25mm diameter
・ 50mm diameter
・ 100mm diameter
・ Flat cable port (W100×H25mm)
* Each cable port is equipped with a silicone
sponge rubber plug.
* Basic specification of the chamber may not be
met when equipped with a cable port.

Equivalent to standard accessory:
〈SH/SU-221, 241, 261〉
Effective size: W200×D150mm
Load resistance: 500gr.
〈SH/SU-641, 661〉
Effective size: W300×D300mm
Load resistance: 5kg.

Specimen basket
Equivalent to standard accessory.
Material: Stainless steel (5 mesh)

25mm diameter type

L stand

flat type

Cable port rubber plug
Prevents air leakage from the cable
port.

Two chambers stacked
(L and H stands)

■Some photographs listed in this catalog contain Japanese display.
■This product can be customized for your needs. For further information, please contact us.

13

OPTIONS
Interface
・GPIB
・RS-232C
* Select one, instead of standard RS-485

Communication cables

・RS-485
・GPIB
・RS-232C

5m/ 10m/ 30m
2m/ 4m
1.5m/ 3m/ 6m

Safety precautions
Do not use specimens which are explosive or inflammable, or which contain such
substances. To do so could be hazardous, as this may lead to fire or explosion.

●

Do not place corrosive materials in the chamber. If corrosive substances or liquid is
used, the life of the unit may be significantly shortened specifically because of the
corrosion of stainless steel, resin and silicone materials.

●

●

Do not place life forms or substances that exceed allowable heat generation.

●

Be sure to read the user's manual before operation.

14

http://www.espec.co.jp/english

Head Office
3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan
Tel : 81-6-6358-4741
Fax : 81-6-6358-5500

ESPEC NORTH AMERICA, INC.
Tel : 1-616-896-6100 Fax : 1-616-896-6150
ESPEC EUROPE GmbH
Tel : 49-89-1893-9630 Fax : 49-89-1893-96379
ESPEC (CHINA) LIMITED
Tel : 852-2620-0830
Fax : 852-2620-0788
ESPEC ENVIRONMENTAL EQUIPMENT (SHANGHAI) CO., LTD.
Head Office
Tel : 86-21-51036677 Fax : 86-21-63372237
BEIJING Branch
Tel : 86-10-64627025 Fax : 86-10-64627036
TIANJIN Branch
Tel : 86-22-26210366 Fax : 86-22-26282186
GUANGZHOU Branch
Tel : 86-20-83317826 Fax : 86-20-83317825
SHENZHEN Branch
Tel : 86-755-83674422 Fax : 86-755-83674228
SUZHOU Branch
Tel : 86-512-68028890 Fax : 86-512-68028860
ESPEC TEST TECHNOLOGY (SHANGHAI) CO., LTD.
Tel : 86-21-68798008 Fax : 86-21-68798088
ESPEC SOUTH EAST ASIA SDN.BHD.
Tel : 60-3-8945-1377 Fax : 60-3-8945-1287

Quality Management System Assessed
and Registered
ESPEC CORP. has been assessed by and
registered in the Quality Management System
based on the International Standard ISO
9001:2008 (JIS Q 9001:2008) through the
Japanese Standards Association (JSA).
*Registration : ESPEC CORP.
(Overseas subsidiaries not included)

TS9B24C03 (The contents of this catalog is as of September, 2012.)

Environmental Management System Assessed
and Registered
ESPEC CORP.

●

Specifications are subject to change without notice due to design improvements.

●

Corporate names and trade names mentioned in this catalog are trademarks or registered trademarks.



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.7
Linearized                      : No
Author                          : tsucky
Create Date                     : 2012:09:10 10:11:47+09:00
Modify Date                     : 2015:05:01 09:59:45-07:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Creator Tool                    : Adobe InDesign CS3_J (5.0.4)
Metadata Date                   : 2015:05:01 09:59:45-07:00
Producer                        : Acrobat Distiller 9.5.2 (Macintosh)
Format                          : application/pdf
Creator                         : tsucky
Title                           : 小型環境_e_.indd
Document ID                     : uuid:31bdfea2-b283-4791-b088-d6c23b1a328b
Instance ID                     : uuid:8b552605-f400-44dc-bea1-f26697d5c385
Page Layout                     : TwoPageRight
Page Count                      : 14
EXIF Metadata provided by EXIF.tools

Navigation menu