1.2 SPECIFICATIONS Fluke PG7202 High Pressure Gas Piston Gauge User Manual
Fluke-PG7601-Piston-Gauge-Manual-Brochure Fluke-PG7601-Piston-Gauge-Manual-Brochure
Fluke-PG7202-High-Pressure-Gas-Piston-Gauge-Manual-Brochure Fluke-PG7202-High-Pressure-Gas-Piston-Gauge-Manual-Brochure
User Manual: Fluke-PG7202-High-Pressure-Gas-Piston-Gauge-User-Manual
Open the PDF directly: View PDF .
Page Count: 214
Download | |
Open PDF In Browser | View PDF |
PG7000™ Piston Gauges PG7102™, PG7202™, PG7302™, PG7601™ (Ver. 3.0 and Higher) Operation and Maintenance Manual © 2011 Fluke Calibration Warning High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be released unexpectedly and with extreme force. High pressure systems should be assembled and operated only by personnel who have been instructed in proper safety practices. © 2011 Fluke Calibration All rights reserved. Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the express written permission of Fluke Calibration, 4765 East Beautiful Lane, Phoenix, Arizona 85044-5318 USA. Fluke Calibration makes sincere efforts to ensure the accuracy and quality of its published materials; however, no warranty, expressed or implied, is provided. Fluke Calibration disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual or products described in it. Mention of any product or brand does not constitute an endorsement by Fluke Calibration of that product or brand. This manual was originally composed in English and was subsequently translated into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version and other language versions, the English version predominates. Products described in this manual are manufactured under international patents and one or more of the following U.S. patents: 6,701,791, 5,142,483, 5,257,640, 5,331,838, 5,445,035. Other U.S. and international patents pending. Fluke Calibration, FCAL, DH, DHI, PG7000, PG7102, PG7202, PG7302, PG7601, COMPASS, CalTool are trademarks, registered and otherwise, of Fluke Corporation. LabVIEW is registered trademark of National Instruments Corporation. Swagelok is a registered trademark of the Swagelok Company. Document No. 3152117 110325 Printed in the USA © 2011 Fluke Calibration Page II Table of Contents Table of Contents ................................................................. III Tables ................................................................................ VII Figures .............................................................................. VIII About This Manual ................................................................ IX 1. Introduction ..................................................................... 1 1.1 1.2 Product Overview ................................................................................................................................... 1 Specifications ......................................................................................................................................... 2 1.2.1 1.2.1.1 1.2.1.2 1.2.2 1.2.2.1 1.2.2.2 1.2.2.3 1.2.3 1.2.4 1.2.4.1 1.2.4.2 1.2.4.3 1.3 Terminal and Platform Front and Rear Panels ................................................................................... 10 1.3.1 1.3.1.1 1.3.1.2 1.3.2 2. General Specifications............................................................................................................................. 2 Embedded Features ............................................................................................................................... 3 Ambient and Instrument Condition Measurements ................................................................................. 4 Piston-Cylinder Modules ......................................................................................................................... 5 PC-7100/7600 ........................................................................................................................................ 5 PC-7200 ................................................................................................................................................. 6 PC-7300 ................................................................................................................................................. 7 Mass Sets ................................................................................................................................................. 7 Pressure Measurements .......................................................................................................................... 8 PC-7100/7600 ........................................................................................................................................ 8 PC-7200 ................................................................................................................................................. 9 PC-7300 ................................................................................................................................................. 9 Terminal Front and Rear Panels............................................................................................................ 10 PG Terminal Front Panel ...................................................................................................................... 10 PG terminal Rear Panel ....................................................................................................................... 11 Platform Rear Panels ............................................................................................................................. 11 Installation ..................................................................... 13 2.1 Unpacking And Inspection .................................................................................................................. 13 2.1.1 2.1.1.1 2.1.1.2 2.1.1.3 2.1.1.4 2.1.2 2.1.2.1 2.1.2.2 2.1.2.3 2.2 2.3 Removing from Packaging .................................................................................................................... 13 Platform ............................................................................................................................................... 13 Mass Set .............................................................................................................................................. 13 Piston-Cylinder Module(s) .................................................................................................................... 14 Automated Mass Handler ..................................................................................................................... 14 Inspecting Contents ............................................................................................................................... 14 Platform ............................................................................................................................................... 14 Mass Set .............................................................................................................................................. 19 Piston-Cylinder Module(s) .................................................................................................................... 21 Site Requirements ................................................................................................................................ 22 Setup ..................................................................................................................................................... 23 2.3.1 2.3.1.1 2.3.1.2 2.3.1.3 2.3.2 2.3.3 Preparing for Operation ......................................................................................................................... 23 Setting Up the Platform ........................................................................................................................ 23 System Pressure Interconnections ....................................................................................................... 24 Setting Up a Mass Set.......................................................................................................................... 24 installing a Piston-Cylinder Module into the Platform ......................................................................... 25 Switching a PG7202 Between Gas Operation and Oil Operation ........................................................ 26 Page III © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 2.4 Power Up and Verification ................................................................................................................... 28 2.4.1 2.4.2 2.4.3 2.4.4 2.4.5 2.4.6 2.4.7 2.4.8 2.4.8.1 2.4.9 2.4.10 2.4.11 2.5 3. Power UP ................................................................................................................................................ 28 Check that On-Board Piston-Cylinder Module and Mass Set Information are Correct ...................... 28 Set Local Gravity Value ......................................................................................................................... 28 Setup Pressure Equation Variable Input Sources ................................................................................ 28 Check Proper Operation of Ambient Condition Measurements .......................................................... 29 apply pressure to THE piston-cylinder module.................................................................................... 29 Check Proper Behavior of Motorized Piston Rotation ......................................................................... 30 Check Proper Operation of Piston Behavior Measurements............................................................... 30 Verify Vacuum Reference (PG7601 Only) ............................................................................................ 30 Check Automated Pressure Generation (If Present) ............................................................................ 31 Check/Set Security Level ....................................................................................................................... 31 Additional Precautions to Take Before Making Pressure Measurements .......................................... 31 Short Term Storage .............................................................................................................................. 32 General Operation ........................................................... 33 3.1 Fundamental Operating Principles ..................................................................................................... 33 3.1.1 3.2 3.3 3.4 Keypad Layout And Protocol .............................................................................................................. 35 Sounds .................................................................................................................................................. 36 Pressure Ready/Not Ready Indication ................................................................................................ 36 3.4.1 3.4.2 3.4.3 3.5 3.6 3.7 3.8 3.9 Piston Position Ready/Not Ready ......................................................................................................... 36 Piston Rotation Ready/Not Ready ......................................................................................................... 37 Vacuum Reference Ready/Not Ready (PG7601 Only) .......................................................................... 38 Piston Position ..................................................................................................................................... 38 Mass Loading Protocol ........................................................................................................................ 39 Main Run Screen .................................................................................................................................. 42 General Function/Menu Flow Chart .................................................................................................... 43 Direct Function Keys............................................................................................................................ 44 3.9.1 3.9.2 3.9.3 3.9.3.1 3.9.4 3.9.4.1 3.9.4.2 3.9.5 3.9.5.1 3.9.5.2 3.9.6 3.9.7 3.9.8 3.9.8.1 3.9.9 3.9.9.1 3.9.9.2 3.9.9.3 3.9.9.4 3.9.9.5 3.9.9.6 3.9.10 3.9.11 3.9.11.1 3.9.11.2 3.9.11.3 3.9.12 3.9.13 3.10 Gas Operated, Liquid Lubricated Piston-Cylinder Operating Principle (PG7202).............................. 34 Direct Function Keys Summary ............................................................................................................ 44 [P-C] ........................................................................................................................................................ 45 [UNIT] ...................................................................................................................................................... 46 Customizing Pressure Units Available Under the UNIT Function .......................................................... 47 [MODE].................................................................................................................................................... 48 Differential Measurement Mode (PG7601 Only) ................................................................................... 49 HIGH LINE Differential measurement mode (PG7102, PG7302 and PG7202 ONLY) ........................... 56 [SYSTEM] ................................................................................................................................................ 68 First System Run Screen...................................................................................................................... 68 Second System Run Screen ................................................................................................................ 69 [AMBIENT] .............................................................................................................................................. 70 [HEAD] .................................................................................................................................................... 71 [ROTATE] ................................................................................................................................................ 73 <2PRE-DECEL> .................................................................................................................................. 74 [GEN] (OPTIONAL) ................................................................................................................................. 75 <2target>.............................................................................................................................................. 76 <3raise> ............................................................................................................................................... 77 <4UL> .................................................................................................................................................. 77 <5tol> ................................................................................................................................................... 77 <6refloat> ............................................................................................................................................. 78 <7Vol> ................................................................................................................................................. 78 [RES] ....................................................................................................................................................... 79 [ENTER/SET P] from Run Screen .......................................................................................................... 80 [ENTER/SET P] in Pressure to Mass Mode .......................................................................................... 81 [ENTER/SET P] in Mass to Pressure Mode .......................................................................................... 83 Commands for Zero Pressure, Ending a Test ....................................................................................... 83 [P OR M].................................................................................................................................................. 84 [ ] and [ ], [←] ................................................................................................................................. 84 [SETUP] Menu....................................................................................................................................... 86 3.10.1 3.10.2 3.10.3 © 2011 Fluke Calibration <1select> ................................................................................................................................................ 87 <2view> ................................................................................................................................................... 88 <3edit> .................................................................................................................................................... 89 Page IV TABLE OF CONTENTS 3.11 [SPECIAL] Menu ................................................................................................................................... 91 3.11.1 3.11.1.1 3.11.1.2 3.11.1.3 3.11.1.4 3.11.1.5 3.11.1.6 3.11.1.7 3.11.1.8 3.11.1.9 3.11.1.10 3.11.1.11 3.11.1.12 3.11.1.13 3.11.1.14 3.11.1.15 3.11.2 3.11.3 3.11.3.1 3.11.3.2 3.11.3.3 3.11.3.4 3.11.4 3.11.4.1 3.11.4.2 3.11.4.3 3.11.4.4 3.11.4.5 3.11.5 3.11.5.1 3.11.5.2 3.11.5.3 3.11.5.4 3.11.5.5 3.11.6 3.11.7 3.11.8 3.11.8.1 3.11.8.2 3.11.8.3 3.11.8.4 3.11.9 3.11.9.1 3.11.9.2 3.11.9.3 3.11.9.4 3.11.9.5 3.11.9.6 4. <1PC/MS> ............................................................................................................................................... 92 Create Piston-Cylinder Module ............................................................................................................. 93 Edit Piston-Cylinder Module ................................................................................................................. 95 View Piston-Cylinder Module ................................................................................................................ 95 Delete Piston-Cylinder Module ............................................................................................................. 96 Select the active piston-cylinder module ............................................................................................... 96 Add Mass Set ....................................................................................................................................... 97 Edit mass set ......................................................................................................................................101 View Mass Set ....................................................................................................................................101 Delete Mass Set ..................................................................................................................................101 Select Mass Set ..................................................................................................................................102 Add Mass Loading Bell........................................................................................................................102 Edit Mass Loading Bell ........................................................................................................................104 View mass loading bell ........................................................................................................................104 delete mass loading bell ......................................................................................................................104 Select Mass Loading Bell ....................................................................................................................105 <2presu> ................................................................................................................................................105 <3HEAD> ...............................................................................................................................................105 <3head>, <1fluid> ...............................................................................................................................106 <3head>, <2unit> ................................................................................................................................106 <3head>, <3atm> ................................................................................................................................107 <3head>, <4piston> ............................................................................................................................107 <4Prefs>.................................................................................................................................................108 <4PREFS>, <1ScrSVR> .....................................................................................................................108 <4PREFS>, <2Sound>........................................................................................................................108 <4PREFS>, <3Time> ..........................................................................................................................109 <4PREFS>, <4ID> ..............................................................................................................................109 <4prefs>, <5level> ..............................................................................................................................110 <5remote> ..............................................................................................................................................113 COM1, COM2 AND COM3 (RS232) ....................................................................................................113 IEEE-488.............................................................................................................................................113 RS232 Self Test ..................................................................................................................................114 External Barometer (RPM) Communications (COM2) ..........................................................................114 External Vacuum Gauge Communications (COM2) (PG7601 Only) ....................................................116 <6GL>.....................................................................................................................................................118 <7cal>.....................................................................................................................................................119 <8AMH> .................................................................................................................................................119 <2control>, <1up/down> ......................................................................................................................120 <2control>, <2discreet>.......................................................................................................................120 <2control>, <3loadall> .........................................................................................................................120 <2control>, <4unloadall> .....................................................................................................................121 <9reset> .................................................................................................................................................121 <9reset>, <1sets> ...............................................................................................................................121 <9reset>, <2units> ..............................................................................................................................122 <9reset>, <3com> ...............................................................................................................................122 <9reset>, <4cal> .................................................................................................................................122 <9RESET>, <5SETUPS> ....................................................................................................................123 <9reset>, <6all> ..................................................................................................................................123 Remote Operation ......................................................... 125 4.1 4.2 Overview ............................................................................................................................................. 125 Interfacing ........................................................................................................................................... 125 4.2.1 4.2.1.1 4.2.1.2 4.2.2 4.3 Commands .......................................................................................................................................... 127 4.3.1 4.3.2 4.3.3 4.3.3.1 4.3.4 4.3.4.1 4.3.4.2 4.4 RS232 Interface .....................................................................................................................................125 COM1 .................................................................................................................................................125 COM2 AND COM3 ..............................................................................................................................126 IEEE-488 (GPIB).....................................................................................................................................126 Command Syntax ..................................................................................................................................127 COMMAND summary ............................................................................................................................127 Error Messages .....................................................................................................................................129 AMH errors..........................................................................................................................................130 Command Descriptions ........................................................................................................................130 IEEE Std. 488.2 Common and Status Commands ...............................................................................130 PG7000 commands.............................................................................................................................133 Status System ..................................................................................................................................... 160 4.4.1 4.4.1.1 4.4.1.2 Status Reporting System ......................................................................................................................160 Status Byte Register............................................................................................................................160 Standard Event Register .....................................................................................................................161 Page V © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 4.5 High Line Differential Mode Programming Examples ..................................................................... 162 4.5.1 4.5.2 5. M a i n t e n a n c e , Ad j u s t m e n t s a n d C a l i b r a t i o n . . . . . . . . . . . . . . . . . . . . . . 1 6 5 5.1 5.2 Introduction ........................................................................................................................................ 165 Platform ............................................................................................................................................... 166 5.2.1 5.2.1.1 5.2.1.2 5.2.1.3 5.2.1.4 5.2.1.5 5.2.1.6 5.2.2 5.2.3 5.2.4 5.2.5 5.3 5.3.2.2 5.3.2.3 5.3.3 5.3.4 5.3.5 5.3.6 5.3.6.1 5.4 Cleaning.................................................................................................................................................189 Recalibration .........................................................................................................................................189 Updating Mass Set Files .....................................................................................................................189 Reloading Embedded Software into PG7000 Flash Memory .......................................................... 189 Disassembly and Reassembly of PG7000 ........................................................................................ 190 5.6.1 5.6.2 5.6.3 Platform .................................................................................................................................................190 Terminal .................................................................................................................................................190 AMH Automated Mass Handler Removal .............................................................................................190 Troubleshooting............................................................ 191 6.1 7. Disassembly, Cleaning and Maintenance ............................................................................................173 Disassembly and Reassembly .............................................................................................................174 Disassembly and Reassembly of gas operated, gas lubricated piston-cylinder modules (PC-7100/7600) ...............................................................................................................................174 Disassembly And Reassembly of Gas Operated, Liquid Lubricated Piston-Cylinder Modules (PC-7200) ....................................................................................................178 Disassembly and Reassembly of Oil Operated, Oil Lubricated Piston-Cylinder Modules (PC-7300) ....................................................................................................180 Filling or Emptying Gas Operated, Liquid Lubricated Piston-Cylinder Module Reservoir with Liquid ............................................................................................................................182 Cleaning Piston-Cylinders ....................................................................................................................183 Lubricating Piston-Cylinder Modules ..................................................................................................185 Recalibration .........................................................................................................................................188 updating piston-cylinder module files ...................................................................................................188 Mass Sets ............................................................................................................................................ 189 5.4.1 5.4.2 5.4.2.1 5.5 5.6 Calibration/Adjustment of On-Board Measurement Functions ..........................................................166 Principles ............................................................................................................................................166 Barometric Pressure Sensor ...............................................................................................................166 AMBIENT TEMPERATURE sensor .....................................................................................................167 Relative Humidity sensor .....................................................................................................................168 piston-cylinder module temperature sensor .........................................................................................168 Reference Vacuum Gauge (PG7601 Only)..........................................................................................170 Piston Position Detection Adjustment .................................................................................................170 Emptying Oil Run-Off Tray (PG7202 and PG7302 Only) .....................................................................171 Purge Mounting Post Liquid Run Off (PG7202 Only) ..........................................................................172 Drive Belt Replacement ........................................................................................................................172 Piston-Cylinder Modules ................................................................................................................... 173 5.3.1 5.3.2 5.3.2.1 6. Recommended sequence for a host program to remotely set a new high line pressure and enable high line differential mode.................................................................................162 Recommended Sequence for a Host Program to Remotely Enable High Line Differential Mode using the Last Line Pressure Setting ....................................................164 Overview ............................................................................................................................................. 191 Appendix ...................................................................... 195 7.1 Conversion of Numerical Values ...................................................................................................... 195 7.1.1 7.2 7.2.1 7.2.2 7.2.3 7.2.3.1 7.2.3.2 7.3 7.4 Pressure ................................................................................................................................................195 Defined Pressure Calculations .......................................................................................................... 195 PG7102, PG7202 and PG7302 ...............................................................................................................197 PG7601...................................................................................................................................................198 Fluid Heads............................................................................................................................................199 Fluid Head Components ......................................................................................................................199 Overall Fluid Head Correction .............................................................................................................200 Glossary .............................................................................................................................................. 201 Limited Warranty and Limitation of Liability .................................................................................... 203 © 2011 Fluke Calibration Page VI Tables Table 1. PG7102 Parts List ........................................................................................................................ 15 Table 2. PG7202 Parts List ........................................................................................................................ 16 Table 3. PG7302 Parts List ........................................................................................................................ 17 Table 4. PG7601 Parts List ........................................................................................................................ 18 Table 5. Manual Mass Set Parts List (excluding 80 and 100 kg)............................................................... 19 Table 6. Manual Mass Set Parts List (80 and 100 kg) ............................................................................... 19 Table 7. AMH-38 Mass Set Parts List ........................................................................................................ 19 Table 8. AMH-100 Mass Set Parts List ...................................................................................................... 19 Table 9. Mass Set Compositions ............................................................................................................... 20 Table 10. Mass Set Compatibility .............................................................................................................. 20 Table 11. PC-7100/7600 Piston-Cylinder Modules Parts List.................................................................... 21 Table 12. PC-7200 Piston-Cylinder Modules Parts List ............................................................................. 21 Table 13. PC-7300 Piston-Cylinder Modules Parts List............................................................................ 22 Table 14. Summary of PG7000 Direct Function Key Operations .............................................................. 44 Table 15. Pressure Units of Measure Available ......................................................................................... 47 Table 16. Valve Settings for Setting Differential Mode Static Pressure ..................................................... 52 Table 17. Valve Settings to Apply PG7000 Pressure to the RPM for Differential Mode Offsetting ........... 53 Table 18. Valve Settings for Operating in Differential Mode ...................................................................... 55 Table 19. SETUP File Choices, Factory Preferred Choice and Normal Value .......................................... 87 Table 20. Security Levels - Functions NOT Executed Per Function/Level .............................................. 111 Table 21. COM1, COM2 and COM3 Available Settings .......................................................................... 113 Table 22. COM1 DB-9F Pin Designation ................................................................................................. 126 Table 23. COM2 and COM3 DB-9M Pin Designation .............................................................................. 126 Table 24. Command Summary ................................................................................................................ 127 Table 25. Error Messages ........................................................................................................................ 129 Table 26. Status Byte Register ................................................................................................................ 160 Table 27. Standard Event Register .......................................................................................................... 161 Table 28. Mounting Post Wire Colors, Description and Location ............................................................ 170 Table 29. PG7000 Troubleshooting Checklist ......................................................................................... 191 Table 30. Pressure Unit of Measure Conversions ................................................................................... 195 Table 31. PG7000 Defined Pressure Calculation Variables .................................................................... 196 Table 32. DHI Authorized Service Providers ........................................................................................... 204 Page VII © 2011 Fluke Calibration TABLES AND FIGURES Figures Figure 1. PG Terminal Front Panel ............................................................................................................ 10 Figure 2. PG Terminal Rear Panel ............................................................................................................. 11 Figure 3. PG Platform Rear Panel ............................................................................................................. 11 Figure 4. Piston-Cylinder Module Installation ............................................................................................ 26 Figure 5. Piston Gauge Operating Principle .............................................................................................. 33 Figure 6. Gas Operated, Liquid Lubricated Piston-Cylinder (PC-7200) Operating Principle .................... 35 Figure 7. PG7000 Keypad Layout .............................................................................................................. 35 Figure 8. Piston Stroke and Zones ............................................................................................................ 39 Figure 9. Run Screen Flow Chart .............................................................................................................. 43 Figure 10. Differential Mode Controller Schematic .................................................................................... 51 Figure 11. High Line Differential Mode Schematic .................................................................................... 59 Figure 12. Status Byte Register ............................................................................................................... 160 Figure 13. PG7202 Mounting Post Drain ................................................................................................. 172 Figure 14. 10 and 20 kPa/kg Gas Piston-Cylinder Module (Expanded View) ......................................... 175 Figure 15. 10 kPa/kg Piston Insertion Tool .............................................................................................. 176 Figure 16. Gas Piston-Cylinder Module Sleeve Nut Tool ........................................................................ 176 Figure 17. 50, 100 and 200 kPa/kg Gas Piston-Cylinder Modules (Expanded View) ............................. 177 Figure 18. Gas Operated, Liquid Lubricated Piston-Cylinder Module (Expanded View) ........................ 179 Figure 19. Oil Piston-Cylinder Module (Expanded View) ......................................................................... 181 Figure 20. Filling Gas Operated, Liquid Lubricated Piston-Cylinder Module Reservoir (PC-7200)........ 183 Figure 21. Gas Operated, Gas Lubricated Piston-Cylinder Module Lubrication Chart ............................. 186 Figure 22. Gas Operated, Liquid Lubricated Piston-Cylinder Module Lubrication Chart ......................... 187 Figure 23. Oil Operated Piston-Cylinder Module Lubrication Chart......................................................... 188 © 2011 Fluke Calibration Page VIII About This Manual This manual provides the user with the information necessary to operate various PG7000 Piston Gauges. It also includes a great deal of additional information provided to help you optimize PG7000 use and take full advantage of its many features and functions. This manual covers four PG7000 models: PG7102, PG7202, PG7302 and PG7601. The four models have many features and characteristics in common as well as individual differences. When discussing features that are common to all four models, they are referred to collectively as PG7000. When providing information pertaining to a specific model, that model is referred to by its specific model number. Before using the manual, take a moment to familiarize yourself with the Table of Contents structure. All first time PG7000 users should read Sections 1 and 2. Section 3 provides a comprehensive description of general PG7000 operating principles. Section 4 covers remote communication with an external computer. Section 5 provides maintenance and calibration information. Section 6 is a quick troubleshooting guide. Use the information in Section 6 to troubleshoot unexpected PG7000 behavior based on the symptoms of that behavior. Certain words and expressions have specific meaning as they pertain to PG7000s. The Glossary (see Section 7) is useful as a quick reference for the definition of specific words and expressions as they are used in this manual. Note For those of you who “don’t read manuals”, go directly to section Error! Reference source not found., initial setup, to set up your PG7000. Then go to section 2.4, power up and verification. This will get you running quickly with minimal risk of causing damage to yourself or your PG7000. THEN… when you have questions or start to wonder about all the great features you might be missing, get into the manual! Manual Conventions Caution “Caution” is used in throughout the manual to identify conditions or actions that could cause harm to the PG7000 or to the devices that are connected to it. Warning “Warning” is used in throughout the manual to identify actions that could pose a hazard to the user of the PG7000. Note “Note” is used throughout the manual to identify operating and applications advice and additional explanations. [ ] Indicates direct function keys (e.g., [RANGE]). < > Indicates molbox1+ screen displays (e.g., <1yes>). Page IX © 2011 Fluke Calibration ABOUT THIS MANUAL Notes © 2011 Fluke Calibration Page X 1. Introduction 1.1 Product Overview PG7000 Piston Gauges are reference level pressure standards that operate on the piston gauge principle. Pressure is defined by balancing it against the force exerted by a known mass accelerated by gravity on the effective area of a piston-cylinder. A PG7000 piston gauge consists of the PG7000 Platform, one or several piston-cylinder modules, a mass set. An automated mass handling system is available. A PG7000 system normally also includes the means to generate and adjust pressures and to interconnect the system components and a device being ® calibrated or tested. The pressure generation component can be manual or automated. COMPASS for Pressure software may also be included to assist in executing test sequences, acquiring test data and producing test reports. There are four PG7000 Platforms: PG7102, PG7202, PG7302 and PG7601. These have a common PG7000 presentation and features. They are distinguished by their normal operating medium (oil and/or gas) and the capability to define pressures relative to a vacuum reference. • PG7102 - Gas operated with gas lubricated piston-cylinder modules (PC-7100/7600 modules) - Maximum pressure is 11 MPa (1 600 psi) - Does not support definition of pressure against a vacuum reference • PG7202 - • PG7302 - Oil operated (PC-7300 modules) - Maximum pressure is 500 MPa (72 500 psi) • PG7601 - Gas operated, gas lubricated piston-cylinder modules (PC-7100/7600 modules) - Maximum pressure is 7 MPa (1 000 psi) - Supports definition of pressure against a vacuum reference Gas operated, liquid lubricated piston-cylinder modules (PC-7200 modules) Oil operated piston-cylinder modules (PC-7300 modules) Maximum pressure is 110 MPa (16 000 psi) when operated with a PC-7200 module Maximum pressure is 200 MPa (30 000 psi) when operated with a PC-7300 module Does not support definition of pressure against a vacuum reference PG7000 platforms, piston-cylinder modules, mass sets and mass handling systems are designed to maximize metrological performance and ease of operation. They include many features that enhance the fundamental precision and stability of pressure measurements as well as simplifying use and reducing operator influence on the measurements. Extensive monitoring and controlling capability and advanced local and remote user interfaces are integrated into PG7000 Platforms. Operator interaction with PG7000 and its extensive capabilities and peripherals is accomplished through a single display and keypad on the PG Terminal or from a computer via a single standard RS232 or IEEE-488 interface. Page 1 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 1.2 Specifications 1.2.1 General Specifications Power Requirements 85 to 264 VAC, 50/60 Hz, 22 VA max. consumption. Operating Temperature Range 15 to 35 °C Operating Humidity Range 5 to 95% R.H., non-condensing Weight Instrument platform with no mass loaded. PG7102 PG7202 PG7302 PG7601 PG Terminal 13 kg (29 lb) 13 kg (29 lb) 13 kg (29 lb) 17 kg (37 lb) 1.4 kg (3 lb) Dimensions Instrument Platform 36 cm H x 40 cm W x 35 cm D (14.5 in. x 15.8 in. x 13.8 in.) (Height: Top of mounting post with piston-cylinder module installed for PG7102/PG7202/PG7302; top of bell jar for PG7601.) PG Terminal 12 cm H x 15 cm W x 20 cm D (4.7 in. H x 5.9 in. W x 7.9 in. D) Microprocessors Instrument Platform PG Terminal Motorola 68302 Hitachi 64180 Communication Ports RS232 IEEE-488 COM1: Host computer COM2: External barometer or vacuum gauge (PG7601) and pass through communications COM3: Automated pressure generator/controller Host computer Overall Pressure Ranges Actual range depends on piston-cylinder and mass set selection PG7102 Gauge: Absolute: Differential: 5 kPa to 11 MPa (0.7 to 1 600 psi) 105 kPa to 11 MPa (15 to 1 600 psi) DP + static pressure < 11 MPa (1 600 psi) PG7202 Gauge (gas): 100 kPa to 110 MPa (15 to 16 000 psi) Absolute (gas):200 kPa to 110 MPa (30 to 16 000 psi) Gauge (oil): (PG7302 module): 1 to 200 MPa (150 to 30 000 psi) Absolute (oil): (PG7302 module) 1.1 to 200 MPa (165 to 30 000 psi) Differential: DP + static pressure < 110 MPa (16 000 psi) Note PC-7200 can be operated in oil up to 200 MPa (30 000 psi) when used with PC-7300 oil operated pistoncylinders. PG7302 Gauge: Absolute: 20 kPa to 500 MPa (3 to 75 000 psi) 120 kPa to 500 MPa (20 to 75 000 psi) PG7601 Gauge: Absolute: Differential: 4 kPa to 7 MPa (0.6 to 1 000 psi) 7 kPa to 7 MPa (0.7 to 1 000 psi) - 90 to 350 kPa (-13 to 50 psi) at 15 to 200 kPaa (2 to 30 psia) static pressure PG7102 Gas: PG7202 Gas: any non-corrosive Oil: Di2-EthylHexyl Sebacate (synthetic oil) PG7302 Oil: Di2-EthylHexyl Sebacate (synthetic oil) PG7601 Gas: Operating Media © 2011 Fluke Calibration Page 2 air, helium, nitrogen air, helium, nitrogen 1. INTRODUCTION Maximum Mass Load PG7102 100 kg, while not exceeding 11 MPa (1 600 psi) PG7202 100 kg, while not exceeding 110 MPa (16 000 psi) when operated with PC-7200 piston-cylinder modules or 200 MPa (30 000 psi) when operated with PC-7300 piston-cylinder modules PG7302 100 kg PG7601 38 kg Pressure Connections PG7102 Test port: DH200 PG7202 Test port: Drain port: DH500 DH500 PG7302 Test port: DH500 PG7601 Test port: Bell Jar Vent Port: DH200 DH200 Vacuum Reference Pump Down Port: KF25 (KF40 available on optional AMH automated mass handler) External Vacuum Port: Optional KF25 on bell jar (KF40 available on optional AMH automated mass handler) Note DH200 and DH500 are gland and collar type fittings for 1/4 in. (6.35 mm) coned and left hand threaded tubes. DH200 is equivalent to AE SF250C, HIP LF4, etc. DH500 is equivalent to AE F250C, HIP HF4, etc. CE Conformance Available, must be specified. 1.2.1.1 Embedded Features • Local control with 2 x 20 vacuum fluorescent display and 4 x 4 function driven keypad. • Real time (1 second update rate) display and measurement of ambient (pressure, temperature, humidity) and instrument (piston-cylinder temperature, piston position, piston drop rate, piston rotation rate, piston rotation decay rate, reference vacuum) conditions. • Real time (1 second update rate) mass-to-pressure and pressure-tomass calculations taking into consideration all environmental and operational variables. • Full gas and liquid fluid head corrections including DUT head correction and piston position head correction. • Adjustable mass loading resolution (0.01 g to 0.1 kg). • Audible prompts of instrument status (piston movement, Ready/Not Ready indication) with override capability. • Integrated automated mass handling option (AMH-38 or AMH-100). • Interfacing and automatic exploitation of external barometer via RS232. • Interfacing and automatic exploitation of any external vacuum gauge via RS232 (PG7601 only). • Automated differential mode to define low differential pressures at various static pressures between vacuum and two atmospheres. • Automated high line differential mode to define differential pressure at high line pressure. • Storage and one step activation of metrological data on up to 18 pistoncylinder modules, (3) mass sets and (3) mass loading bells. Page 3 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL • Continuous pressure Ready/Not Ready indication based on measured conditions. • Motorized, intelligent piston drive system based measured rotation rate with operator alert and manual override. • Integrated automated pressure control with standard Fluke Calibration pressure controllers. • Full RS232 and IEEE-488 communications with multi-level commands to set and read all instrument functions. 1.2.1.2 Ambient and Instrument Condition Measurements Temperature Ambient Range Piston Cylinder Module o 0 to 40 C 0 to 40 oC Resolution 0.1 0.01 Measurement Uncertainty ±1 ± 0.1 Barometric Pressure with Internal Sensor Range Resolution Measurement Uncertainty 70 to 110 kPa 10 Pa ± 140 Pa Barometric pressure can also be read automatically with any RS232 device such as a FCAL RPM. Relative Humidity Range Resolution Measurement Uncertainty 5 to 95 % RH 1 % RH ± 10 % RH Piston Position Range Resolution Measurement Uncertainty ± 4.5 mm 0.1 mm ± 0.2 mm Piston Rotation (Rate and deceleration) Range Resolution 2 to 99 rpm 1 rpm Vacuum (PG7601 only) Range Resolution Measurement Uncertainty © 2011 Fluke Calibration 0 to 20 Pa 0.01 Pa ± 0.1 Pa or 10 % of reading, whichever is greater Page 4 1. INTRODUCTION 1.2.2 Piston-Cylinder Modules All piston-cylinders are integrated modules including mounting hardware delivered in individual shipping and storage bullet cases. 1.2.2.1 PC-7100/7600 Gas operated, gas lubricated piston-cylinder characteristics. Used in PG7102 and PG7601 platforms. PC-7100/7600-10, TC PC-7100/7600-10-L Operation Gas operated, gas lubricated Piston Material Tungsten carbide Cylinder Material Tungsten carbide Nominal Diameter Nominal Area Mounting System 35 mm 1 000 mm2 Simple free deformation PC-7100/7600-20 Operation Gas operated, gas lubricated Piston Material Tungsten carbide Cylinder Material Tungsten carbide Nominal Diameter Nominal Area Mounting System 25 mm 500 mm2 Simple free deformation PC-7100/7600-50 Operation Gas operated, gas lubricated Piston Material Tungsten carbide Cylinder Material Tungsten carbide Nominal Diameter Nominal Area Mounting system 16 mm 200 mm2 Negative free deformation PC-7100/7600-100 Operation Gas operated, gas lubricated Piston Material Tungsten carbide Cylinder Material Tungsten carbide Nominal Diameter Nominal Area Mounting System 11 mm 98 mm2 Negative free deformation PC-7100/7600-200 Operation Gas operated, gas lubricated Piston Material Tungsten carbide Cylinder Material Tungsten carbide Nominal Diameter Nominal Area Mounting System 8 mm 50 mm2 Negative free deformation Page 5 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 1.2.2.2 PC-7200 Gas operated, liquid lubricated piston-cylinder module characteristics. Used in PG7202 platform. Note Though not recommended for day-to-day operation, PC-7200 modules can also be filled completely with oil and operated with oil as the test medium (see Section 2.3.3). PC-7200-100 Operation Lubricating Liquid Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Gas operated, liquid lubricated Synturion 6 (Krytox® optional) Tungsten carbide 11.2 mm 98.1 mm2 Negative free deformation PC-7200-200 Operation Lubricating Liquid Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Gas operated, liquid lubricated Synturion 6 (Krytox® optional) Tungsten carbide 7.9 mm 49.0 mm2 Negative free deformation PC-7200-500 Operation Lubricating Liquid Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Gas operated, liquid lubricated Di-2-ethylhexyl Sebacate (Krytox® optional) Tungsten carbide 5.0 mm 19.6 mm2 Negative free deformation PC-7200-1 Operation Lubricating Liquid Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Gas operated, liquid lubricated Di-2-ethylhexyl Sebacate (Krytox® optional) Tungsten carbide 3.5 mm 9.8 mm2 Negative free deformation PC-7200-2 Operation Lubricating Liquid Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System © 2011 Fluke Calibration Page 6 Gas operated, liquid lubricated Di-2-ethylhexyl Sebacate (Krytox® optional) Tungsten carbide 2.5 mm 4.9 mm2 Negative free deformation 1. INTRODUCTION 1.2.2.3 PC-7300 Oil operated, oil lubricated piston-cylinder module characteristics. Used in PG7302 and PG7202 platforms (1 MPa/kg and higher only in PG7202). Note PC-7300 modules PC-7300-1, -2 and -5 may also be used in a PG7202 platform. PC-7300-100 Operation Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Oil operated, oil lubricated Tungsten carbide 11.2 mm 98.1 mm2 Simple free deformation PC-7300-200 Operation Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Oil operated, oil lubricated Tungsten carbide 7.9 mm 49.0 mm2 Simple free deformation PC-7300-500 Operation Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Oil operated, oil lubricated Tungsten carbide 5.0 mm 19.6 mm2 Simple free deformation Operation Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Oil operated, oil lubricated Tungsten carbide 3.5 mm 9.8 mm2 Simple free deformation Operation Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Oil operated, oil lubricated Tungsten carbide 2.5 mm 4.9 mm2 Simple free deformation Operation Piston and Cylinder Material Nominal Diameter Nominal Area Mounting System Oil operated, oil lubricated Tungsten carbide 1.6 mm 2.0 mm2 Simple free deformation PC-7300-1 PC-7300-2 PC-7300-5 1.2.3 Mass Sets All masses are delivered in molded, reusable, transit cases with custom inserts. Masses > 50g Material Finish Adjustment Tolerance Uncertainty of Measured Values Masses < 50g 304L non-magnetic stainless steel Electropolished ± 20 ppm of nominal value (manual mass sets, AMH automated mass handler mass sets do not have fixed adjustment tolerances) ± 5 ppm or 1 mg, whichever is greater ± 1 mg Page 7 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Note Masses designated “tare” are delivered without reported measured values and are intended only for use on the “tare” PG7000 in high line differential pressure measurement mode. 1.2.4 Pressure Measurements 1.2.4.1 PC-7100/7600 Note For uncertainty in piston-cylinder effective area and typical measurement uncertainty in pressure defined by the piston gauge, see the piston-cylinder calibration report and current revision of Technical Note 7920TN01. PC-7100/7600-10 PC-7100/7600-10-L Sensitivity1 Reproducibility2 Typical Drop Rate (35 kg) 0.02 Pa + 0.5 ppm ± 2 ppm 0.2 mm/min PC-7100/7600-20 Sensitivity1 Reproducibility2 Typical Drop Rate (35 kg) 0.04 Pa + 0.5 ppm ± 2 ppm 0.3 mm/min PC-7100/7600-50 Sensitivity1 Reproducibility2 Typical Drop Rate (35 kg) 0.1 Pa + 0.5 ppm ± 2 ppm 0.5 mm/min PC-7100/7600-100 Sensitivity1 Reproducibility2 Typical Drop Rate (35 kg) 0.2 Pa + 0.5 ppm ± 3 ppm 0.7 mm/min PC-7100/7600-200 Sensitivity1 Reproducibility2 Typical Drop Rate (35 kg) 1 0.4 Pa + 0.5 ppm ± 3 ppm 1.0 mm/min Sensitivity: The smallest variation in input detectable in output. Note Piston-cylinder modules designated “tare” are delivered without reported values and are intended only for use on the “tare” PG7000 in high line differential pressure measurement mode. © 2011 Fluke Calibration Page 8 1. INTRODUCTION 1.2.4.2 PC-7200 Note For uncertainty in piston-cylinder effective area and typical measurement uncertainty in pressure defined by the piston gauge, see the piston-cylinder calibration report and current revision of Technical Note 7920TN01. PC-7200-100 Sensitivity1 Reproducibility2 2 Pa + 1 ppm ± 5 ppm Typical Drop Rate (50 kg) 0.10 mm/min Sensitivity1 4 Pa + 1 ppm PC-7200-200 Reproducibility2 Typical Drop Rate (50 kg) ± 5 ppm 0.15 mm/min PC-7200-500 Sensitivity1 Reproducibility2 Typical Drop Rate (50 kg) 10 Pa + 1 ppm ± 5 ppm 0.20 mm/min PC-7200-1 Sensitivity1 Reproducibility2 Typical Drop Rate (50 kg) 20 Pa + 1 ppm ± 5 ppm 0.25 mm/min PC-7200-2 Sensitivity1 Reproducibility2 Typical Drop Rate (50 kg) 40 Pa + 1 ppm ± 5 ppm 0.50 mm/min 1 Sensitivity: The smallest variation in input detectable in output. 2 Reproducibility: Combined long term stability of piston-cylinder effective area and masses. 1.2.4.3 PC-7300 Note For uncertainty in piston-cylinder effective area and typical measurement uncertainty in pressure defined by the piston gauge, see the piston-cylinder calibration report and current revision of Technical Note 7920TN01. PC-7300-100 Sensitivity1 Reproducibility2 2 Pa + 1 ppm ± 5 ppm Typical Drop Rate (50 kg) 0.02 mm/min Sensitivity1 4 Pa + 1 ppm PC-7300-200 Reproducibility2 Typical Drop Rate (50 kg) ± 5 ppm 0.04 mm/min PC-7300-500 Sensitivity1 Reproducibility2 Typical Drop Rate (50 kg) Page 9 10 Pa + 1 ppm ± 5 ppm 0.10 mm/min © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL PC-7300-1 Sensitivity1 Reproducibility2 Typical Drop Rate (50 kg) 20 Pa + 1 ppm ± 5 ppm 0.20 mm/min PC-7300-2 Sensitivity1 Reproducibility2 Typical Drop Rate (50 kg) 40 Pa + 1 ppm ± 5 ppm 0.40 mm/min PC-7300-5 Sensitivity1 Reproducibility2 Typical Drop Rate (50 kg) 1.3 100 Pa + 1 ppm ± 5 ppm 1.00 mm/min 1 Sensitivity: The smallest variation in input detectable in output. 2 Reproducibility: Combined long term stability of piston-cylinder effective area and masses. Terminal and Platform Front and Rear Panels 1.3.1 Terminal Front and Rear Panels 1.3.1.1 PG Terminal Front Panel The front panel assembly provides a 2 x 20 vacuum fluorescent display and a 4 x 4 membrane keypad for local user interface. The terminal front panel assembly is the same for all PG7000 models. Fluorescent display Keypad Figure 1. PG Terminal Front Panel © 2011 Fluke Calibration Page 10 1. INTRODUCTION 1.3.1.2 PG terminal Rear Panel The rear panel assembly provides the communications connection to the PG7000 Platform and the power connection module. The terminal rear panel assembly is the same for all PG7000 models. 1. Power switch 2. Fuse 3. Power receptacle 4. Connector for cable to PG7000 (25-pin) 5. Cooling fan Figure 2. PG Terminal Rear Panel 1.3.2 Platform Rear Panels The PG7000 Platform rear panels provide the connection to the PG Terminal, remote communication connections and pressure connection ports. The rear panels of all PG7000 models are identical except for the pressure connections (see Figure 3, # 7). COM2 (RS232) - External Barometer, External Vacuum Gauge (PG7601 only) and Pass Through Communications COM3 (RS232) - Automated Pressure Generation/Control Component Temperature - Humidity Probe COM1 (RS232) - Remote Host Communications IEEE-488 - Remote Host Communications AMH Connection Pressure Ports: PG7102 - TEST port: DH200 PG7202 - TEST and DRAIN ports: DH500 PG7302 - TEST port: DH500 PG7601 - TEST and VACUUM vent ports: DH200 Vacuum pull down port on front left side: KF25 PG7000 Terminal Port Figure 3. PG Platform Rear Panel Page 11 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Notes © 2011 Fluke Calibration Page 12 2. Installation 2.1 Unpacking And Inspection 2.1.1 Removing from Packaging A typical PG7000 system includes the PG7000 Platform (see Section 2.1.1.1), a mass set, (see Section 2.1.1.2), one or more piston-cylinder modules (see Section 2.1.1.3) and other accessories such as an AMH automated mass handler and/or pressure generation and control components (see the accessory Operation and Maintenance Manual or Instruction Sheet). 2.1.1.1 Platform Caution The mass loading bell is a metrological element that is part of the mass set. Like all of the masses, it is preferable not to handle it with bare hands. Protective gloves are provided in the accessory kit of each PG7000 Platform. The PG7000 Platform is shipped in a reusable, molded shipping and storage case. Open the PG7000 shipping and storage case (it is the large, 66 cm x 53 cm x 47 cm case). Remove the PG Terminal and accessories from upper packing insert. Inspect and inventory the accessories (see Section 2.1.2). Remove the upper packing insert. Carefully lift the PG7000 Platform from its position in the lower packing insert. Note the orientation so that the same orientation will be used when PG7000 is repacked. Reinstall the upper packing insert into the shipping and storage case and store in a safe place. 2.1.1.2 Mass Set Caution The stability over time of PG7000 pressure measurements is a function of the stability of the masses loaded on the piston. Precautions should be taken in handling the masses to minimize influences that may change their mass. This includes always wearing protective gloves when handling the masses to avoid contaminating them with body oils and perspiration. Protective gloves are provided in the accessory kits of PG7000 Platforms. The mass set accessories are shipped in a separate corrugated container. Open the corrugated container and inspect and inventory the accessories. The PG7000 masses are shipped in reusable, molded shipping and storage cases. The PG7000 masses should be removed from their shipping cases and inventoried when actually setting up the PG7000 system. Page 13 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 2.1.1.3 Piston-Cylinder Module(s) The piston-cylinder modules are shipped in Acetal bullet cases that are packed in corrugated containers with custom foam inserts. Open the corrugated containers and remove the piston-cylinder modules and accessories. The bullet cases screw open by turning the lid counterclockwise. Caution When reinstalling an oil (PC-7300) or liquid lubricated (PC-7200) piston-cylinder module in its bullet case, be sure to empty out any liquid that may have collected in the hole in the bottom of the case. Excess liquid will not compress, making it difficult to fully close the case and could result in damaging it. 2.1.1.4 Automated Mass Handler See the AMH-38/AMH-100 Operation and Maintenance Manual. 2.1.2 Inspecting Contents Check that all items are present and have NO visible signs of damage. A parts list of items supplied is provided in Section 2.1.2.1 for PG7000, Section 2.1.2.2 for mass sets, and Section 2.1.2.3 for piston-cylinder modules. 2.1.2.1 Platform Each PG7000 Platform is delivered complete with accessories as listed by part number in Tables 1 through 4. © 2011 Fluke Calibration Page 14 2. INSTALLATION Table 1. PG7102 Parts List DESCRIPTION Platform PG7102 PG7102 P/N 3069572 NON-CE P/N 3072317 CE 3117734 3117752 Manual Mass Bell 3071537 Terminal 3069735 PG Terminal to Platform Cable Non-CE (DB25M - DB25F, ≈ 1.8 meters) 3068724 CE (DB25M - DB25F, ≈ 1.5 meters) 3072235 Power Cable 3133781 (Black) TH Probe Assembly 2106009 Accessory Kit 3117741 NIP, SS, DH200, 2.75 in. 3068377 ADPT, SS, DH200 F x 1/8 in. NPT F 3068547 O-ring, Buna 2-242 (2 ea.) 3135041 Storage Cover, 7600 Type 3135594 Allen Wrench, 2.5 mm 3136044 Allen Wrench, 3 mm 3135703 Allen Wrench, 5 mm 3136098 Spanner Wrench (Metrological) 3068940 Krytox GPL205/6 0.5 oz. 2493420 Gift Kit with Gloves 3123777 ADPT, DH200 M x 1/8 in. swage 3069062 Documentation Calibration Report (PG Platform) Calibration Report (Mass Bell) Technical Data PG7000 Operation & Maintenance Manual Documentation CD 3153005 (Gray) 3152121 3152121 3152139 3152117 3139043 Page 15 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Table 2. PG7202 Parts List DESCRIPTION Platform PG7202 PG7202 #3070404 NON-CE #3072395 CE 3119996 3120027 Manual Mass Bell 3071537 Terminal 3069735 PG Terminal to Platform Cable Non-CE (DB25M - DB25F, ≈ 1.8 meters) 3068724 CE (DB25M - DB25F, ≈ 1.5 meters) 3072235 Power Cable 3133781 (Black) TH Probe Assembly 2106009 Accessory Kit 3120011 DH500 M x 1/8 in. NPT F 3142684 O-ring, Buna 2-242 (2 ea.) 3135041 Storage Cover, 7600 Type 3135594 Allen Wrench, 2.5 mm 3136044 Allen Wrench, 3 mm 3135703 Allen Wrench, 5 mm 3136098 Wrench, 5/8 in. 3139417 Collar, SS, DH500 3068607 Krytox GPL205/6 0.5 oz. 2493420 Gift Kit with Gloves 3123777 Documentation Calibration Report (PG) Calibration Report (Mass Bell) Technical Data PG7000 Operation & Maintenance Manual Documentation CD © 2011 Fluke Calibration 3152121 3152121 3152139 3152117 3139043 Page 16 3153005 (Gray) 2. INSTALLATION Table 3. PG7302 Parts List PG7302 P/N 3069747 NON-CE DESCRIPTION Platform PG7302 P/N 3072339 CE 3118073 3118086 Manual Mass Bell 3071537 Terminal 3069735 PG Terminal to Platform Cable Non-CE (DB25M - DB25F, ≈ 1.8 meters) 3068724 CE (DB25M - DB25F, ≈ 1.5 meters) 3072235 Power Cable 3133781 (Black) TH Probe Assembly 2106009 Accessory Kit 3120011 DH500 M x 1/8 in. NPT F 3153005 (Gray) 3142684 O-ring, Buna 2-242 (2 ea.) 3135041 Storage Cover, 7600 Type 3135594 Allen Wrench, 2.5 mm 3136044 Allen Wrench, 3 mm 3135703 Allen Wrench, 5 mm 3136098 Wrench, 5/8 in. 3139417 Collar, SS, DH500 3068607 Krytox GPL205/6 0.5 oz. 2493420 Gift Kit with Gloves 3123777 Documentation Calibration Report (PG) Calibration Report (Mass Bell) Technical Data PG7000 Operation & Maintenance Manual Documentation CD 3152121 3152121 3152139 3152117 3139043 Page 17 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Table 4. PG7601 Parts List DESCRIPTION Platform PG7601 PG7601 P/N 3069028 NON-CE P/N 3072258 CE 3117525 3117540 Manual Mass Bell 3071603 Bell Jar and Seal 3068933 and 3068634 Terminal 3069735 PG Terminal to Platform Cable Non-CE (DB25M - DB25F, ≈ 1.8 meters) 3068724 CE (DB25M - DB25F, ≈ 1.5 meters) 3072235 Power Cable 3133781 (Black) TH Probe Assembly 2106009 Accessory Kit 3117533 NIP, SS, DH200, 2.75 in. 3068377 ADPT, SS, DH200 F x 1/8 in. NPT F 3068547 O-ring, Buna 2-242 (2 ea.) 3135041 Storage Cover, 7600 Type 3135594 Allen Wrench, 2.5 mm 3136044 Allen Wrench, 3 mm 3135703 Allen Wrench, 5 mm 3136098 Spanner Wrench (Metrological) 3068940 Krytox GPL205/6 .5 oz. 2493420 Gift Kit with Gloves 3123777 ADPT, DH200 M x 1/8 in. swage 3069062 Valve, Vacuum Relief 3124573 Documentation Calibration Report (PG) Calibration Report (Mass Bell) Technical Data PG7000 Operation & Maintenance Manual Documentation CD © 2011 Fluke Calibration 3152121 3152121 3152139 3152117 3139043 Page 18 3153005 (Gray) 2. INSTALLATION 2.1.2.2 Mass Set PG7000 mass sets are composed of different combinations of individual masses and accessories depending on the specific mass set ordered (see Tables 5 - 9). Table 5. Manual Mass Set Parts List (excluding 80 and 100 kg) DESCRIPTION Mass Set PART NO. Refer to Table 9 3068969 3068991 35 kg set 1 ea. 1 ea. 40 kg set 1 ea. 1 ea. 45 kg set 1 ea. 1 ea. 55 kg set 1 ea. 2 ea. Reusable Molded Transit Case with Foam Inserts Mass Set Tray and Spindle 3147461 and 3148764 Dust Covers 3138017 and 3138130 Calibration Report 3152121 Table 6. Manual Mass Set Parts List (80 and 100 kg) DESCRIPTION Mass Set PART NO. Refer to Table 9 Reusable Molded Transit Case with Foam Inserts 3068969 3068984 80 kg set 1 ea. 2 ea. 100 kg set 1 ea. 3 ea. Mass Set Tray and Spindle 3147461 and 3148764 Dust Covers 3138017 and 3138127 Calibration Report 3152121 Table 7. AMH-38 Mass Set Parts List DESCRIPTION Mass Set PART NO. Refer to Table 9 3123990 3069004 13 kg set (MS-AMH-13) 1 ea. 1 ea. 25 kg set (MS-AMH-25) 1 ea. 1 ea. 1 ea. 1 ea. Reusable Molded Transit Case with Foam Inserts 39 kg set (MS-AMH-38) Calibration Report 3152121 Table 8. AMH-100 Mass Set Parts List DESCRIPTION Mass Set PART NO. Refer to Table 9 3123990 3068984 40 kg set (MS-AMH-40) 1 ea. 1 ea. 60 kg set (MS-AMH-60) 1 ea. 2 ea. 80 kg set (MS-AMH-80) 1 ea. 2 ea. 100 kg set (MS-AMH-100) 1 ea. 3 ea. Reusable Molded Transit Case with Foam Inserts Calibration Report 3152121 Page 19 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Table 9. Mass Set Compositions MASS SET COMPOSITION DESIGNATION PART # NOMINAL TOTAL MASS (kg) kg 5 kg 2 kg 1 kg 0.5 kg 0.2 kg 0.1 kg MAKE-UP MASS (kg) 10 MS-7001-35 3069850 35 - 5 2 1 1 2 1 1 (4.5) - MS-7002-35 3069861 35 - 5 2 1 1 2 1 1 (4) - MS-7002-40 3070021 40 - 6 2 1 1 2 1 1 (4) - MS-7002-45 3069980 45 - 7 2 1 1 2 1 1 (4) - MS-7002-55 3069877 55 - 9 2 1 1 2 1 1 (4) - MS-7002-80 3070000 80 6 1 2 1 1 2 1 1 (9) - MS-7002-100 3070017 100 8 1 2 1 1 2 1 1 (9) - DESIGNATION PART # NOMINAL TOTAL MASS (kg) MASS SET COMPOSITION 10 kg 6.4 kg 6.2 kg 3.2 kg 1.6 kg 0.8 kg 0.4 kg 0.2 kg 0.1 kg BELL, SHAFT, BINARY MASS CARRIER (3 PARTS) 1 1 1 MS-AMH-13 3071491 13 - - 1 1 1 1 1 MS-AMH-25 3071484 25 - - 3 1 1 1 1 1 1 1 MS-AMH-38 3071433 38 - - 5 1 1 1 1 1 1 1 MS-AMH-40 3071528 40 3 1 - 1 1 1 1 1 1 1 MS-AMH-60 3071519 60 5 1 - 1 1 1 1 1 1 1 MS-AMH-80 3071504 80 7 1 - 1 1 1 1 1 1 1 MS-AMH-100 3071440 100 9 1 - 1 1 1 1 1 1 1 Table 10. Mass Set Compatibility DESIGNATOR NOMINAL TOTAL MASS (kg) MS-7001-35 35 MS-7002-35 35 • • • MS-7002-40 40 • • • MS-7002-45 45 • • • MS-7002-55 55 • • • MS-7002-80 80 •1 •1 • MS-7002-100 100 •1 •1 • MS-AMH-13 13 • MS-AMH-25 25 • MS-AMH-38 38 • MS-AMH-40 40 • • • MS-AMH-60 60 •1 •1 • MS-AMH-80 80 •1 •1 • MS-AMH-100 100 •1 •1 • 1 PG7102 PG7202 PG7302 PG7601 • These mass sets, on certain piston-cylinder sizes, can cause the maximum working pressure of the PG Platform to be exceeded. Do not exceed the following maximum working pressures: 1. PG7201: 11 MPa (1 600 psi) 2. PG7202: When using PC-7200 piston-cylinder modules: 110 MPa (16 000 psi) When using PC-7300 piston-cylinder modules: 200 MPa (30 000 psi) 3. PG7302:500 MPa (72 500 psi) © 2011 Fluke Calibration Page 20 2. INSTALLATION Note The mass loading bell and piston make up part of the total mass load. The mass loading bell for loading manual mass sets is delivered with the PG7000 platform. The mass loading bell for AMH mass sets is delivered with the mass set. Piston-cylinder modules are purchased and delivered separately. 2.1.2.3 Piston-Cylinder Module(s) Table 11. PC-7100/7600 Piston-Cylinder Modules Parts List 10 kPa PC-7100/ 7600-10-L 10 kPa PC-7100/ 760010 TC 20 kPa PC-7100/ 7600-20 50 kPa PC-7100/ 7600-50 100 kPa PC-7100/ 7600-100 200 kPa PC-7100/ 7600-200 3171975 3070095 3071581 3070109 3071615 3070111 Piston-Cylinder Module 3125106 3122937 3122116 3124088 3122234 3124194 Hermetic Acetal Bullet Case 3070203 3070203 3070203 3070203 3070203 3070203 Accessory Kit 3125242 3122928 3122229 3124345 3124345 3124345 O-rings 3134867 3136458 3134867 3136458 3134867 3136458 3134880 3136458 3134880 3136458 3134880 3136458 Insertion Tool 3071793 3071841 N/A N/A N/A N/A 3152121 3152121 3152121 3152121 3152121 3152121 Piston-Cylinder Kit Calibration Report Table 12. PC-7200 Piston-Cylinder Modules Parts List 100 kPa PC-7200-100 200 kPa PC-7200-200 500 kPa PC-7200-500 1 MPa PC-7200-1 2 MPa PC-7200-2 3070419 3070428 3070437 3070443 3070455 Piston-Cylinder Module 3120245 3120250 3120292 3120303 3120326 Hermetic Acetal Bullet Case 3071852 3071852 3071852 3071852 3071852 Accessory Kit 3120277 3120277 3120315 3120315 3120315 (2) 3134158 (2) 3134158 (2) 3134158 (2) 3134158 (2) 3134158 2721018 2721018 2721018 2721018 2721018 2644003 2644003 2644003 2644003 2644003 N/A N/A 3148417 3148417 3148417 Piston-Cylinder Kit O-rings Anti-Extrusion Ring Syringe Syringe tips Synturion 6 fluid Sebacate fluid Calibration Report 3139439 3139439 3139439 3139439 3139439 (3) 3139421 (3) 3139421 (3) 3139421 (3) 3139421 (3) 3139421 3120289 3120289 - - - - - 3120590 3120590 3120590 3152121 3152121 3152121 3152121 3152121 Page 21 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Table 13. PC-7300 Piston-Cylinder Modules Parts List 100 kPa PC-7300-100 200 kPa PC-7300-200 500 kPa PC-7300-500 1 MPa PC-7300-1 2 MPa PC-7300-2 5 MPa PC-7300-5 3070039 3070042 3070056 3070063 3070074 3070088 Piston-Cylinder Module 3118918 3118976 3119032 3119116 3119178 3119229 Hermetic Acetal Bullet Case 3071865 3071865 3071865 3071865 3071865 3071865 Accessory Kit 3119102 3119102 3119102 3119366 3119366 3119366 Piston-Cylinder Kit O-rings Calibration Reports 2.2 2527053 2527053 2527053 1785497 1785497 1785497 3134022 3134022 3134022 (5) 927863 (5) 927863 (5) 927863 3152121 3152121 3152121 3152121 3152121 3152121 Site Requirements The exact PG7000 system installation is affected by the elements other than the PG7000 Platform that make up the PG7000 system. When selecting and preparing a site to set up the PG7000 system, the following should be considered: • Ambient conditions: To achieve optimum metrological performance, ambient conditions should be controlled and maintained within the following: ♦ Temperature: 19 to 23 °C, minimize rate of change of temperature. ♦ Relative Humidity: 10 to 60 %RH (non-condensing). ♦ Ambient Pressure: Minimize external influences that will cause barometric instability. ♦ Air Currents: Do not install the PG7000 Platform under a source of vertical air currents such as an overhead air conditioning duct. These can blow on the mass load and add unquantified forces. ♦ Vibration: Minimize local vibration. Excessive vibration will reduce the stability of the pressures defined by PG7000 (vibration affects the floating piston). Excessive high frequency vibration, for example from a vacuum pump on the same table as the PG7000, may affect piston sensitivity. • Bench stability: Up to 100 kg may be loaded and unloaded onto the PG7000 Platform. The bench on which the PG7000 sits should not deflect significantly under the mass load changes. This can be verified by setting the PG7000 Platform on the bench, leveling it, loading and unloading the complete mass set while observing whether the level setting changes. • Location of other components: Plan the space required and a convenient layout for the complete PG7000 system including the PG Terminal, mass set, pressure generation/control component(s), test instrument connection and computer (if present). If using a PPC, MPC, GPC or OPG to generate/control pressure, see its Operation and Maintenance Manual for information on installing it. If a Fluke Calibration interconnections kit is being used to interconnect the components, see its instruction sheet. • Electrical and pressure supplies: Plan the supply of electrical power to the PG Terminal and to the pressure generation/control component(s), if needed. If using a PPC, MPC, GPC or OPG to generate/control pressure, see its Operation and Maintenance Manual for information on the pressures source(s) it needs and how to connect them. Gas supplied to a PC-7100/7600 piston-cylinder module must be clean and dry (instrument grade minimum, high purity preferred) to avoid contaminating the piston-cylinder gap. • Reference vacuum supply (PG7601 only): Plan for the vacuum connection to the platform or the optional AMH automated mass handler and the location of the reference vacuum pump. • Bell jar placement (PG7601 only): Plan a location for the bell jar when it is removed from the platform to load and unload masses. A small shelf is often used for this purpose. • AMH automated mass handler (optional) placement: If an AMH is being used, plan electrical and pneumatic connections to it as well as a location to place it when it is removed from the platform (see the AMH-38/AMH-100 Operation and Maintenance Manual). © 2011 Fluke Calibration Page 22 2. INSTALLATION 2.3 Setup 2.3.1 Preparing for Operation Note Before setting up the PG7000 system, see Section 2.2 for information on site requirements. To prepare PG7000 for check out and operation: Set up the PG7000 Platform (see Section 2.3.1.1). If an optional AMH-100 mass set is being used, set it up with its mass set (see the AMH38/AMH-100 Operation and Maintenance Manual). Make the system pressure interconnections (see Section 2.3.1.2). If a manual mass set is being used, set up the manual mass set (see Section 2.3.1.3). 2.3.1.1 Setting Up the Platform To set up the PG7000 Platform proceed as follows: Place the PG7000 Platform on the site table in the desired orientation. Though the rear panel is usually in the back, any orientation can be used. Place the PG7000 Terminal at the desired location. Connect the PG7000 Terminal to the PG7000 Platform using the 25-pin cable supplied. Connect the PG7000 Temperature - Humidity Probe per Figure 3. Connect electrical power (85 to 264 VAC, 50/60 Hz) to the PG7000 Terminal using the power cable supplied. Any grounded power cable with a standard IEC320-313 connection may be used. (PG7601 Only) - Install the vacuum vent valve kit on the vacuum vent port on the rear of the PG7000 Platform. Refer to the instruction sheet provided with the vent valve assembly. Connect the reference vacuum source and shutoff valve to the reference vacuum port. Take measures to assure that vacuum oil cannot return to the PG7601. If an optional AMH-38 automated mass handler is being used, the reference vacuum may be connected to the larger KF40 vacuum connection on the AMH-38 vacuum chamber. If an external barometer and/or vacuum gauge is/are being used, establish communications between the barometer/vacuum gauge and the PG7000 Platform by connecting the external device RS232 port to the PG7000 Platform Com2 port and setting up PG7000 to read and use an external barometer and/or vacuum gauge (see Section 3.11.5.4, 3.11.5.5). Set the external barometer head height (see Section 3.11.3.3). If an automated pressure generation/control component is being used, establish communications between the automated pressure generation/control component and the PG7000 Platform by connecting the generation/control component RS232 port to the PG7000 Platform COM3 port and setting up PG7000 to use an automated pressure generation/control component (see Section 3.9.9). If an AMH automated mass handling system is being used, set up the AMH mass set and the AMH mass handler following the instructions in the AMH38/AMH-100 Operation and Maintenance Manual. Level the platform using the PG7000 Platform’s two leveling feet and the level mounted on the front of the platform. (Or on the optional AMH mass handling system.) Page 23 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 2.3.1.2 System Pressure Interconnections Interconnect the PG7000 Platform, pressure generation/control components and a test connection. The pressure connection on the PG7000 TEST port is: • PG7102/PG7601: DH200 (DH200 is equivalent to AE SF250C,HIP LF4, etc.). Adapters to convert the DH200 connection to 1/8 in. NPT female and 1/8 in. swage are provided with PG7102 and PG7601 Platform accessories. • PG7202: DH500 (DH500 is equivalent to AE F250C, HIP HF4, etc.). Connect the gas test medium pressure control/generation component to the TEST port. If the PG7202 is to be operated in oil, connect the oil pressure control/generation component to the DRAIN port (the gas or oil component is disconnected when not in use, see Section 2.3.3). • PG7302: DH500 (DH500 is equivalent to AE F250C, HIP HF4, etc.). Note If using a standard interconnections kit such as PK-7600PPC/MPC P/N 3069508, PK-7600-PPC/MPC-DIF P/N 3070127 or PK-7100-MPCD-DIF P/N 3070226, see the instruction sheet provided with the kit for installation instructions. If an OPG1 or GPC1 generator/controller is being used, pressure interconnecting hardware is included with it. See its Operation and Maintenance Manual for connection instructions. 2.3.1.3 Setting Up a Mass Set Note If installing an MS-AMH-xxx mass set for use with an AMH automated mass handling system, see the AMH-38/AMH-100 Operation and Maintenance Manual. To install a PG7000 manual mass set, place the mass loading tray (provided with the mass set accessories) at the desired location then install the individual masses on the mass loading tray. See the AMH-38/AMH-100 Operation and Maintenance Manual for instructions on installation of an AMH mass set for automated mass handling. Caution It is VERY IMPORTANT that the individual masses be installed on the mass loading tray in proper sequence. This will ensure that PG7000 mass loading instructions are executed properly (carefully follow the instructions provided in Section 2.3.1.3, Installing Masses on the Mass Loading Tray). Installing Masses on the Mass Loading Tray (Manual Mass Set) PG7000 masses are shipped in reusable, molded shipping and storage cases. One of the cases contains a 4.0, 4.5 or 5 kg mass and the masses of 2 kg and under, the other case(s) contain(s) the main masses of 10 or 5 kg each. Each mass is packed in a sealed plastic bag and then placed in a protective shipping insert. Caution The stability over time of PG7000 pressure measurements is a function of the stability of the masses loaded on the piston. Precautions should be taken in handling the masses to minimize influences that may change their mass. This includes always wearing protective gloves when handling the masses to avoid contaminating them with body oils and perspiration. Protective gloves are provided in the accessory kits of PG7000 Platforms. © 2011 Fluke Calibration Page 24 2. INSTALLATION To install the masses on the mass loading tray, proceed as follows: Open the shipping cases. Install the main masses: The main masses (a series of 10 kg masses if the mass set is > 55 kg, a series of 5 kg masses if the mass set is < 60 kg) are installed horizontally aligned on the mass loading spindle. The main masses are sequentially numbered starting with he number 1. The main mass with the highest sequential number is installed first at the bottom of the stack (i.e., the first mass loaded on the tray). The rest of the main masses should be stacked upwards in descending order ending with main mass sequential number 1. Be careful NOT to confuse the makeup mass (refer to next item) with main mass 1. Install the make up mass: The make up mass is a single mass sequentially numbered 1. It has the same diameter as the main masses. It is a 9 kg mass if the main masses are 10 kg. It is 4 or 4.5 kg if the main masses are 5 kg. The make up mass is placed on top of the main mass stack. It is always the top of the stack. Install the fractionary masses: The fractionary masses are all the masses of lower value than the main masses and makeup mass. These are masses of 5 kg and under for mass sets with 10 kg main masses. They are masses of 2 kg and under for mass sets with 5 kg main masses. Fractionary masses of 1 to 5 kg are discs with a central hole. Fractionary masses of 100 to 500 g are solid, small diameter pucks. Fractionary masses of 50 g and under are grams masses packed and stored in their own separate storage case. Fractionary mass discs and pucks are installed vertically in the corresponding slots in the mass loading tray. Use a consistent setup for the sequence number when there are two masses (e.g., always load sequential number 1 in the front). 2.3.2 installing a Piston-Cylinder Module into the Platform To operate the PG7000 Platform, a piston-cylinder module must be installed in its mounting post. To install a piston-cylinder module in the PG7000 Platform, proceed as follows: Remove the PG7000 Platform mounting post plug. Unscrew the ORANGE plastic mounting post plug that is installed in the PG7000 Platform mounting post. Rotate counterclockwise to remove. Remove the piston-cylinder module from its bullet case. Select a piston-cylinder module. Open the piston-cylinder module bullet case by rotating its lid counterclockwise. Remove the piston-cylinder module from the bullet case base by unthreading it from the case. Hold the piston-cylinder module body by the knurled area and rotate it counterclockwise. Note PC-7200 gas operated, liquid lubricated piston-cylinder modules are delivered with their lubricating liquid reservoir drained. The reservoir must be filled prior to using the piston-cylinder module. If installing a PC-7200 gas operated, liquid lubricated piston-cylinder module, see Section 5.3.3 for instructions on how to fill the module’s liquid lubrication reservoir prior to installation, then continue the procedure from this point. Caution When reinstalling an oil or liquid lubricated gas piston-cylinder module in its bullet case, be sure to empty out any liquid that may have collected in the hole in the bottom of the case. The liquid will not compress, making it difficult to fully close the case and could result in damaging it. Page 25 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Place the piston-cylinder module in the PG7000 Platform mounting post. Place the piston-cylinder module (thread down) into the PG7000 Platform mounting post (see Figure 4 below). Note If installing the piston-cylinder module into a PG7302 or a PG7202 to be operated with oil, first use the oil generation control component to fill the counter bore in the mounting post with oil until it just starts to overflow. Purge the air from under the piston-cylinder module (PC-7300 modules only in PG7302 or PG7202). Rotate the piston-cylinder module clockwise until all threads are engaged and there is no gap between the piston-cylinder module and the mounting post. Back the piston-cylinder module off slightly by rotating it counterclockwise 3/4 turn. Slowly supply additional oil to the mounting post from the pressure generation/control component. Watch the oil run off tube on the bottom right of the platform. As soon as oil appears or can be seen flowing, stop the oil supply. Screw the piston-cylinder module into the PG7000 Platform mounting post. Rotate the piston-cylinder module clockwise until all threads are engaged and there is NO gap between the pistoncylinder module and the PG7000 mounting post. Slight resistance will be encountered in the second half of travel as the piston-cylinder module O-rings seat in the mounting post. Figure 4. Piston-Cylinder Module Installation Caution • Low torque manual rotation is all that should be required to fully seat the piston-cylinder module into the PG7000 mounting post. Never force the piston-cylinder module into the mounting post. • Always maintain PG7202 and PG7302 piston- cylinder modules vertical with the piston cap up. Do not invert the assembly, as this might allow liquid to run up into the piston head and into the adjustment mass and cap. Liquid contamination of the piston head and cap changes the mass of the piston assembly and could lead to out of tolerance pressure definitions at low mass loads. If liquid contaminates the adjustment mass and cap, disassemble the module and clean it (see Sections 5.3.1, 5.3.4, 5.3.2.2). 2.3.3 Switching a PG7202 Between Gas Operation and Oil Operation PG7202 can be operated with gas as the pressurized medium using PC-7200 gas operated, liquid lubricated modules or with oil as the pressurized medium using oil operated PC-7300 modules. © 2011 Fluke Calibration Page 26 2. INSTALLATION Note PC-7200 gas operated, liquid lubricated piston-cylinder modules can be operated in gas oil filled with oil and operated in oil. Oil operation of PC7202 modules is not recommended for routine operation. It can be useful in very specific crossfloating circumstances, particularly when establishing a calibration link between independent gas operated and oil operated piston gauges. When switching PC-7200 module between oil and gas and oil operation, see Section 5.3.3 for information on emptying excess oil from the module. Switching a PG7202 from gas to oil operation Note Only the high pressure PC-7300 oil modules may be used in the PG7202 platform (PC-7300-1, -2, -5) Disconnect the gas generation/control system from the TEST port. Disconnect the tube at the DH500 TEST port connection on the back of the PG7202. Loosely install a DH500 plug in the TEST port. Connect the oil generation/control system to the DRAIN port. Connect a tube from the oil generation/control system to the PG7202 DRAIN port and tighten it (torque DH500 glands to 15 Nm (12 ft.lb)). Fill the PG7202 mounting post with oil. Hand tighten the DH500 plug in the TEST port. Use the oil generation/control component to fill the PG7202 mounting post with oil. Fill to the edge of the second step in the mounting post. Place a paper towel under the TEST port plug and loosen the DH500 plug allowing oil to run out until it is at the level of the first step in the mounting post and there is no air in the run off oil. If there is still air in the run off, repeat the operation. After filling the mounting post, tighten the DH500 plug (torque DH500 glands to 15 Nm (12 ft.lb)). Install a PC-7300 oil operated piston-cylinder module in the PG7202 platform. Install the module and purge the air from underneath it (see Section2.3.2). Operate with oil as the test medium. The PG7202 can now be operated with oil as the test medium. Caution The maximum working pressure of the PG7202 platform when used in oil with a PC-7300 module is 200 MPa (30 000 psi). The maximum pressure when using a PC-7200 module is 110 MPa (16 000 psi). Do not exceed this limit. Switching a PG7202 from oil to gas operation Remove the piston-cylinder module. Disinstall the PC-7300 oil operated piston-cylinder module for the PG7202 platform. Disconnect the oil generation/control system from the DRAIN port. Place a paper towel under the DRAIN port and disconnect the tube to the oil generation/control system. Let all the oil run out of the PG7202 platform. Remove the DH500 plug from the TEST port. Place a paper towel under the TEST port and remove the DH500 plug. Let all the oil run out of the PG7202 platform.. Install a PC-7200 gas operated, liquid lubricated piston-cylinder module in the PG7202 platform. Install the piston-cylinder module (see Section 2.3.2). Connect the gas generation/control system to the TEST port. Connect a tube from the oil generation/control system to the PG7202 TEST port and tighten it (torque DH500 glands to 15 Nm (12 ft.lb)). Prepurge the oil from the PG7202 mounting post. Hold a paper towel lightly over the DRAIN port opening. Leaving the drain port open, use the gas generation/control component to flow enough gas through the mounting post to just lift the piston. Page 27 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Purge the PG7202 mounting post. Purge the PG7202 mouting post using the standard purging procedure (see Section 5.2.4) and finish by tightening the plug in the DRAIN port. Operate with gas as the test medium. The PG7202 can now be operated with gas as the test medium. Caution The maximum working pressure of the PG7202 platform when used with a PC-7200 module is 110 MPa (16 000 psi). Do not exceed this limit. 2.4 Power Up and Verification 2.4.1 Power UP Turn the PG7000 power ON by pressing the power ON/OFF switch on the rear panel of the PG Terminal. Observe the PG terminal display as the terminal connects with the PG7000 Platform, tests, initializes and goes to the main run screen (see Section 3.7). If <….Searching…..> displays for more than 5 seconds, the communications between the PG7000 and the PG Terminal are failing. Check that the PG7000 to PG Terminal cable is properly installed. If PG7000 fails to reach the main run screen, service may be required. Record the sequence of operations and displays observed and contact a Fluke Calibration Authorized Service Provider (see Table 32). 2.4.2 Check that On-Board Piston-Cylinder Module and Mass Set Information are Correct PG7000 uses stored piston-cylinder and mass set metrological information to calculate the reference pressures it defines (see Section 3.1). For the pressure values to be correct, the stored metrological information on the piston-cylinder, mass set and mass loading bell must be correct. Before using PG7000 for accurate pressure definition, the validity of the stored information should be verified. This consists of comparing the piston-cylinder, mass set and mass loading bell information stored in PG7000 to the information in the current piston-cylinder and mass set calibration reports. To verify the PG7000 piston-cylinder, mass set and mass loading bell information, use the piston-cylinder and mass set viewing capabilities accessed by pressing [SPECIAL], <1PC/MS> (see Section 3.11.1). Compare all the information contained in the PG7000 piston-cylinder, mass set and mass loading bell files to the information on the current piston-cylinder, mass set and mass loading bell calibration reports. 2.4.3 Set Local Gravity Value PG7000 uses the value of local acceleration due to gravity (gl) in its calculation of the reference pressure it defines (see Section 3.1). The correct value of local acceleration due to gravity at the site of PG7000 use must be entered. This is accomplished by pressing [SPECIAL], <6gl> (see Section 3.11.6) and editing the value of local gravity. 2.4.4 Setup Pressure Equation Variable Input Sources PG7000 uses many variables in calculating defined pressures. The sources of the variables are determined by the current SETUP file. SETUP files are viewed, created, edited and selected using the SETUP function accessed by pressing [SETUP] (see Section 3.10). A factory SETUP file is available and already selected on a new PG7000, but the operator may desire to customize it. The factory setup file assumes that PG7000’s internal measurement values will be used whenever possible. © 2011 Fluke Calibration Page 28 2. INSTALLATION 2.4.5 Check Proper Operation of Ambient Condition Measurements PG7000 automatically measures ambient conditions and uses these conditions in its pressure calculations. To verify that the ambient condition measurements are operating properly proceed as follows: • Display current ambient condition readings: Press [AMBIENT]. The ambient conditions run screen is displayed (see Section 3.9.6). • Verify proper ambient condition readings: Compare the ambient condition values displayed to the actual values of ambient conditions. Refer to the ambient condition measurement specifications when evaluating the ambient readings (see Section 1.2.1.2). Note • The unit of measure in which ambient pressure is displayed is the same as the unit selected by pressing [UNIT] (see Section 3.9.3). Units of measure in which other ambient condition values are expressed cannot be changed. • PG7000 allows the source of ambient condition values used in reference pressure calculations to be specified. The source may be PG7000’s onboard measurements, default values or operator entered values. See Section 3.10 for information on specifying the source of ambient condition values used by PG7000 in reference pressure calculations. 2.4.6 apply pressure to THE piston-cylinder module Note This section assumes that the PG7000 system has already been set up, including pressure interconnection (see Section 2.3). Caution Before applying pressure to the PG7000 system, be sure that all pressure vessels and connections are rated for the pressure levels that will be applied and that all connections have been properly tightened. Continuing with the PG7000 set up and check out requires applying pressure to the pistoncylinder module and floating the piston. Proceed as follows: Turn OFF automated piston rotation (if present) and automated pressure generation (if present). This will prevent the automated rotation and pressure generation features (if present) from interfering during verification of these features (see Sections 3.9.8 and 3.9.9 for information on automated piston rotation and pressure generation). Load mass on the piston. Install the mass loading bell on the piston. Then load the make up mass (9, 4.5 or 4 kg depending on the mass set) (see Section 2.3.1.3). If the PG7000 platform is equipped with AMH automated mass handling, press [ENTER] and enter a low value of pressure or mass to cause mass to be loaded onto the piston. Float the piston. Use the pressure generation/control component of the PG7000 system to apply pressure under the piston through the PG7000 Platform TEST port. The piston will float at a pressure approximately equal to the piston mass to pressure conversion factor multiplied by the mass load in kg. The piston-cylinder conversion factor is marked on the top of the piston cap and is in either kPa or MPa per kilogram [kPa/kg or MPa/kg]. Page 29 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Note If the piston cannot be floated because it immediately sinks down from a float position, there is a leak in the pressure system. Identify and eliminate leaks until the piston, once floated, falls at a rate less than the nominal fall rate given in the specifications of the piston-cylinder module being used (see Section 1.2.2). 2.4.7 Check Proper Behavior of Motorized Piston Rotation The motorized rotation feature can be set to engage and disengage automatically as needed to maintain the floating piston rotating within a set rotation rate range when floating, and to stop rotation before mass manipulation to set a different pressure occurs (see Section 3.9.8). Motorized piston rotation can also be activated manually by pressing [ ]. Stopping piston rotation can be activated by pressing and holding [ ] and then pressing [←] (see Section 3.9.13). Turn automated rotation OFF by pressing [ROTATE] and selecting <1off>. Load at least the mass bell on the piston, float the piston and press [ ]. When the piston is floating, the motorized rotation system should engage and cause the mass bell and piston to begin rotating. Within 5 to 30 seconds, depending on the mass load, the rotation rate should reach 80 rpm for PG7601 or about 50 rpm for PG7102 or PG7302. Current rotation rate can be observed by pressing [SYSTEM] (see Section 3.9.5). 2.4.8 Check Proper Operation of Piston Behavior Measurements Float the piston and rotate it (see Sections 2.4.6 and 2.4.7). Press [SYSTEM] once to reach the first SYSTEM run screen. Verify that the piston position, piston fall/rise rate, piston rotation rate and piston rotation decay rate are indicating correctly (see Section 3.9.5). The piston position reading system may be calibrated using an on-board procedure if necessary (see Section 5.2.2). Calibration of piston position is recommended when installing a new PG7000 and regularly after installation. Press [SYSTEM] again to reach the second SYSTEM run screen. Verify that the piston-cylinder temperature and temperature rate of change are indicating correctly (see Section 3.9.5). If checking a PG7601 Platform, verify that the vacuum reference measurement is operating correctly (see Section 2.4.8.1). 2.4.8.1 Verify Vacuum Reference (PG7601 Only) PG7601 includes provisions for establishing and measuring a vacuum reference. To verify the vacuum reference capability: Install a piston-cylinder module (see Section 2.3.2). Install the bell jar on the PG7601 Platform (the bell jar aligns on the PG7601 vacuum plate and seals itself). If using AMH-38 automated mass handling, install the AMH-38 mass handler (see the AMH-38/AMH-100 Operation and Maintenance Manual). Apply a vacuum through the reference vacuum port (KF25 port on front left side of platform or KF40 if using the optional AMH-38 automated mass handler). Press [SYSTEM] twice to observe the value of vacuum read by the PG7601’s built-in vacuum gauge. If a vacuum pump of adequate capacity has been correctly connected to the reference vacuum port, the vacuum read by the PG7601 built-in vacuum gauge should go to 4 Pascal [Pa] or lower in less than five minutes on the first pump down and two to three minutes on immediately subsequent pump downs. © 2011 Fluke Calibration Page 30 2. INSTALLATION If this performance is NOT achieved: 2.4.9 • The vacuum pump may be inadequate. • The connection of the vacuum pump to the PG7601 platform of AMH-38 may have leaks or excessive restrictions. • The PG7601 built-in vacuum gauge or external vacuum gauge may be incorrect. • There may be a leak in the PG7601. Check Automated Pressure Generation (If Present) Caution Before applying pressure to the PG7000 system, be sure that all pressure vessels and connections are rated for the pressure levels that will be applied and that all connections have been properly tightened. To check automated pressure generation/control: Verify that the automated pressure generation/control component is properly connected to the system (see Section 2.3.1.2). When operation has returned to the main run screen, turn ON automated pressure generation control, if available, by pressing [GEN] and selecting <1on> (see Section 3.9.9). Press [P OR M], <1pressure> to select pressure entry mode (see Section 3.9.12). Press [ENTER] and enter a pressure value to be generated and follow the mass loading instruction (see Section 3.6). Verify the pressure generation/control component properly generates pressure and floats the PG7000 piston. 2.4.10 Check/Set Security Level PG7000 has a security system based on User Levels. By default, the security system is set to low and NO password is required to change the security level. See Section 3.11.4.5 for information on the security system. As part of the PG7000 startup, set your desired security level and a password. Caution PG7000 is delivered with the security level set to low to avoid inadvertent altering of critical internal settings but with access to changing security levels unrestricted. It is recommended that the low security level be maintained at all times and password protection be implemented if control over setting of security levels is desired. 2.4.11 Additional Precautions to Take Before Making Pressure Measurements Before using PG7000 to make accurate pressure measurements, consider the following: • Select/activate the correct piston-cylinder module, mass set and mass loading bell (see Sections 3.9.2, 3.11.1.10, 3.11.1.15). • If using an automated pressure generation/control component with automated pressure generation, set the pressure controller’s upper limit (UL) (see Section 3.9.9.3). • Enter the correct value of local gravity at the site of use (see Section 3.11.6). • Consider head corrections (see Sections 3.9.7 and 3.11.3). • Level the PG7000 Platform properly (see Section 2.3.1.1). Page 31 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 2.5 • Select the correct pressure unit of measure and measurement mode (see Sections 3.9.3 and 3.9.5). • Verify that the settings for the sources of variables to be used by PG7000 in its calculations of reference pressures are those desired (see Section 3.10). • Verify that the piston-cylinder module is correctly cleaned and operating properly (see Section 5.3). • PG7202 only: Section 5.3.3). • Verify that there are NO leaks in the pressure system. Ensure that the piston-cylinder module liquid reservoir is filled (see Short Term Storage The following is recommended for short term storage of PG7000. Remove all masses from the PG7000 Platform (manual mass set only). Vent all circuits to atmosphere. Turn OFF power using the power switch on the rear of the PG7000 Terminal. Cover the PG7000 Platform and mass set with the dust covers included in the platform and mass set accessories (not used with AMH mass handling system). © 2011 Fluke Calibration Page 32 3. General Operation 3.1 Fundamental Operating Principles PG7000s operate on the principle of the piston gauge in which pressure is defined by balancing it against a known force on a known area (see Figure 5). The known area is defined by a vertically mounted piston rotating in a cylinder and the known force is applied to the piston by loading it with known mass subjected to acceleration due to gravity. When the force applied by the pressure and the force applied by the mass accelerated by gravity are in equilibrium, the piston floats and the pressure under the piston remains constant. The pressure can be calculated following the equation in Figure 5 (see also, Section 7.2). Generally, the pressurized fluid under the piston also lubricates the gap between the piston and the cylinder. PG7202 uses a unique gas operated, liquid lubricated piston-cylinder (see Section 3.1.1). Figure 5. Piston Gauge Operating Principle The PG7000 Platform is designed to mount a variety of piston-cylinder sizes, allow pressure to be applied under the piston and allow masses to be loaded on top of the piston. There are different PG7000 models depending upon whether the pressure medium is oil or gas and whether a vacuum reference is needed. The measurement uncertainty in the pressure defined by the piston gauge depends on the uncertainty in the effective area of the piston-cylinder and the force applied by the mass accelerated by gravity. PG7000 stores the calibrated values of the piston-cylinders and masses it uses in on-board files (see Section 3.11.1). To determine the effective area of the piston-cylinder and the force applied by the masses under actual operating conditions, a number of influences on these values must be quantified and taken into consideration. For this reason, PG7000 includes extensive features to monitor the behavior and conditions of the piston-cylinder as well as ambient conditions that affect pressure definition (see Sections 3.9.5 and 3.9.6). PG7000 uses the piston-cylinder, mass and ambient condition information to calculate the pressure defined by a given mass load or the mass load needed to define a given pressure (see Section 7.2). The source of each value used by PG7000 in its calculations can be selected by the user between PG7000’s internal measurements, default values or user entered values. These sources are defined in SETUP files (see Section 3.10). Page 33 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Once PG7000 has been set up, it is used in day-to-day operation either to define pressures applied to a device or system under test or to measure a stable pressure. To interface with the PG Terminal, the operator: Selects the appropriate piston-cylinder to cover the pressure range (see Section 3.11.1.5). Selects the desired pressure unit of measure (see Section 3.9.3). Selects the desired pressure measurement mode (gauge, absolute, differential) (see Section 3.9.4). Sets the head difference between the PG7000 and the device under test (see Section 3.9.7). Selects mass to pressure or pressure to mass operating mode (see Section 3.9.12). Enters a pressure to define or a current mass load to calculate (see Section 3.9.11.1 and 3.9.11.2). Loads masses by hand or with optional AMH automated mass handler, floats piston and defines pressure. 3.1.1 Gas Operated, Liquid Lubricated Piston-Cylinder Operating Principle (PG7202) PG7202 provides very high pressure gas pressure operation using a unique gas operated, liquid lubricated piston-cylinder system. The principle of operation of the gas operated, liquid lubricated piston-cylinder is simple but very effective (see Figure 6). The measured gas pressure, Pg, is applied to the bottom of the piston and to the top of a liquid reservoir located around the cylinder. The reservoir is connected to the gap between the piston and the cylinder through lateral holes near the bottom of the cylinder, allowing liquid from the reservoir to enter the gap. The pressure of the liquid in the gap, Pl, is equal to the gas pressure Pg, plus the liquid head, h. Therefore, the liquid pressure in the gap is always higher than the gas pressure by the amount of the liquid head regardless of the gas pressure value. Since h is small and the space between the piston and cylinder is typically < 1 micron, the bleed of liquid from the bottom of the cylinder towards the gas pressure is extremely small. The mounting post of the 7202 piston gauge is configured so this minute amount of liquid drops directly into a sump that is deadened (see Section 5.2.4, 3.1.1) and not in the flow path of test gas into and out of the system. Though molecules of the liquid may migrate through the gas, no significant contamination of the test system occurs. Because the liquid reservoir is contained in the piston-cylinder module, piston-cylinders can be removed and installed in the PG7202 piston gauge platform with no loss of liquid from the reservoir. All PC-7200 piston-cylinder modules can be delivered using Krytox, a fluorinated synthetic fluid, to lubricate the piston-cylinder in applications where the system must remain perfectly free of hydrocarbons (e.g. when calibrating instrumentation for oxygen service). Note PC-7200 gas operated, liquid lubricated piston-cylinder modules are delivered with their lubricating liquid reservoir drained. The reservoir must be filled prior to using the piston-cylinder module and then regularly when used (see Section 5.3.3) © 2011 Fluke Calibration Page 34 3. GENERAL OPERATION Figure 6. Gas Operated, Liquid Lubricated Piston-Cylinder (PC-7200) Operating Principle 3.2 Keypad Layout And Protocol PG7000 has a 4 x 4 keypad for local operator access to direct functions, function menus and for data entry. The Function/Data keys allow very commonly used functions to be accessed directly from the main run screen by a single keystroke. The name of the function is on the bottom half of the key (see Section 3.9.1). These keys enter numerical values when editing. The Editing and Execution keys are for execution, suspending execution, backing up in menus and editing entries. The Menu/Data keys provide access to function menus from the main run screen. The menu name is on the bottom half of the key. The SETUP menu is for more frequently used functions. The SPECIAL menu is for functions that are NOT generally used as a part of day to day operation. These keys enter numerical values when editing. Figure 7. PG7000 Keypad Layout Key press confirmation is provided by both tactile and audible feedback. A single beep confirms a valid entry. A descending two note tone signals an invalid entry. The audible valid entry feedback can be suppressed or modified by pressing [SPECIAL] and selecting <5prefs>, <2sound> (see Section 3.11.4.2). Pressing the [ENTER/SET P] key generally causes execution or forward movement in the menu tree. [ENTER/SET P] is also used to enter a command to set a pressure. Pressing the [ESCAPE] key generally allows movement back in the menu tree and/or causes execution to cease or suspend without changes being implemented. Pressing [ESCAPE] repeatedly eventually returns to the main run screen. From the main run screen, pressing [ESCAPE] allows momentary viewing of the PG7000 identification screen. Page 35 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Pressing the [+/-] key changes a numerical sign when editing. It also toggles through multiple screens when available. Pressing the [←] and [→] keys when editing allows reverse and forward cursor movement when editing data entry. These keys are also used to scroll through choices. Menu selections can be made by pressing the number of the selection directly or by pressing [←] and [→] to place the cursor on the number of the desired selection and pressing [ENTER]. Note Some screens go beyond the two lines provided by the display. This is indicated by a flashing arrow in the second line of the display. Press [←] and [→] to move the cursor to access the lines that are NOT visible or directly enter the number of the hidden menu choice if you know it. 3.3 Sounds The PG Terminal is equipped with a variable frequency tone device to provide audible feedback and alarms. Some sounds can be modified and all sounds can be suppressed (see Section 3.11.4.2). Sounds are used for the following indications: Valid key press Invalid key press Piston left end of stroke high or low 3.4 Brief high frequency beep. Three rapid, low frequency beeps. Three rapid valid key press beeps. Piston was at low stop or high stop and just entered spring zone (see Section 3.5). Pressure Ready/Not Ready Indication The three characters on the top line, far left, of the main run screen provide a pressure Ready/Not Ready indication. This indication is intended to give the user a clear and objective indication of when PG7000 conditions are such that the value of pressure it is defining is valid and in tolerance. There are three Ready/Not Ready indication characters to indicate the status of the three main Ready/Not Ready criteria. The Ready/Not Ready indication characters are shown on the first line, top left hand side, of the main run screen. 1. Piston position and vertical movement. 2. Piston rotation. 3. Reference vacuum (PG7601 in absolute by vacuum measurement mode only). *** 100.4755 kPa g h + 0.1 mm 10.00564 kg For each Ready/Not Ready indication character, <*> indicates a Ready condition. Therefore, <***> or <** > indicates that all conditions necessary for an in tolerance pressure definition are present. Any indication other than <*> indicates Not Ready. See Sections 3.4.1, 3.4.2 and 3.4.3 for details on each of the three Ready/Not Ready indicating characters. Note The criteria used to distinguish between Ready and Not Ready conditions can be customized by the user (see Sections 3.4.1, 3.4.2, 3.4.3). 3.4.1 Piston Position Ready/Not Ready The piston position Ready/Not Ready character indicates Ready or Not Ready based on the position of the piston in its vertical stroke (see Section 3.5) and an automated measure generation status (see Section 3.9.9). This ensures that pressure definitions will be made with the piston not more than a certain distance from mid-float position and that measurements will not be made while the action of an automated pressure controller may influence the pressure. The piston position Ready/Not Ready character is the first character from the left on the top line of the main run screen. © 2011 Fluke Calibration Page 36 3. GENERAL OPERATION The piston position Ready/Not Ready criterion is determined by the current SETUP file and can be customized by the user (see Section 3.10). Piston position Ready/Not Ready character indications include: <*> Piston position Ready (within the position limits specified in the current SETUP file) (see Section 3.5). <↓> Piston position Not Ready, low (below the position limits specified in the current SETUP file, see Section 3.10). The <↓> flashes if the piston is not at the bottom stop position to alert the user that this indicator is Not Ready. <↑> Piston position Not Ready, high (above the position limits specified in the current SETUP file) (see Section 3.10). The <↑> flashes if the piston is not at the top stop position to alert the user that this indicator is Not Ready. Piston position not known (current specified mass load is less than the load of the piston + bell). The bell must be installed for PG7000 piston position measurement to operate correctly so piston position values shown are not valid when the bell is not installed. Tare PG7000 is Not Ready (PG7102 and PG7202 only). Indicates that piston position or rotation rate is Not Ready on the tare PG7000 in high line differential mode (see Section 3.9.4.2). >Note 3.4.2 • Piston position always indicates Not Ready if automated pressure generation is adjusting pressure, regardless of actual piston position. • In PG7102 and PG7202 high line differential pressure mode (see Section 3.9.4.2), the piston position Ready/Not Ready indicator applies to the pistons of both the tare and the reference PG7000. For Ready to be indicated, both pistons must be within the ready limits for piston position. Piston Rotation Ready/Not Ready The piston rotation Ready/Not Ready character indicates Ready or Not Ready based on the rotation rate of the piston. The piston rotation Ready/Not Ready character is the second character from the left on the top line of the main run screen. The piston rotation rate Ready/Not Ready criterion is specific to the currently active piston-cylinder module and can be edited by the user (see Sections 3.11.1.1). Piston rotation rate Ready/Not Ready character indications include: <*> Rotation rate Ready: Rotation rate is within the rotation rate limits specified in the current piston-cylinder module file) (see Section 3.11.1.1). <<> > Rotation rate Not Ready, low: Rotation rate is less than the lower rotation rate limit specified in the current piston-cylinder module file (see Section 3.11.1.1) or motorized rotation system is currently engaged. Note that the low rotation limit is automatically reduced when the mass loaded on the piston is less than 3 kg. The <<> flashes if the piston is floating to alert the user that this indicator is Not Ready. Rotation rate not known: Current specified mass load is less than the load of the (piston + bell). The bell must be installed for PG7000 piston rotation rate measurement to operate correctly so piston rotation rate cannot be measured when the bell is not installed. Page 37 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Note At mass loads less then 3 kg the low rotation limit is automatically reduced to an operational minimum to maximize free rotation time. In PG7102 and PG7202 high line differential pressure mode (see Section 3.9.4.2), the piston rotation rate Ready/Not Ready indicator applies to the pistons of both the tare and the reference PG7000. For Ready to be indicated, both pistons must be within the ready limits for piston rotation rate. 3.4.3 Vacuum Reference Ready/Not Ready (PG7601 Only) The vacuum reference Ready/Not Ready character is only used with PG7601, as it is the only PG7000 model with vacuum reference capability. It is only active when operating in absolute by vacuum mode (see Section 3.9.4). The vacuum reference Ready/Not Ready character indicates Ready or Not Ready based on the value of reference vacuum when making measurements in absolute by vacuum mode. This ensures that definitions of absolute pressure with a vacuum reference will be made with the vacuum under the PG7601 bell jar lower than a specified value. When the vacuum is not low enough, vacuum measurement errors may be excessive. The vacuum reference Ready/Not Ready character is the third character from the left on the top line of the main run screen. The vacuum reference Ready/Not Ready criterion is determined by the current SETUP file and can be customized by the user (see Section 3.10). The vacuum reference criterion is a fixed value that can be customized by the user when the vacuum reference selection in the SETUP file is internal. If the selection in the SETUP file is NOT internal, the vacuum reference Ready/Not Ready character always indicates Ready and the value cannot be customized. Vacuum reference Ready/Not Ready character indications include: <> (Blank) Vacuum reference Ready/Not Ready is not in use. This is not a PG7601; or if PG7601, current measurement mode is not absolute by vacuum. <*> Vacuum reference Ready. Vacuum value is below limit specified in the current SETUP file if source is internal OR source is default or user and the vacuum value is not a measured value (see Section 3.10). < > > Vacuum reference Not Ready. Current SETUP file source for vacuum is internal and vacuum value is above the limit specified (see Section 3.10). The <>> flashes if the piston is floating to alert the user that this indicator is Not Ready. 3.5 Piston Position When operating a PG7000 piston gauge, reference pressure values are defined by loading known mass values on the piston and adjusting the pressure to float the piston. Piston position is measured and displayed real time on the MAIN run screen (see Section 3.7) and in the first SYSTEM run screen (see Section 3.9.5). Piston position is used as a criterion for the Ready/Not Ready indication as valid measurements can only be made when the piston is in the correct position (see Section 3.4.1). The full piston stroke is ± 4.5 mm from the midstroke position. positions and zones as illustrated in Figure 8. These zones are: The stroke is divided into different • High and low stops: The piston is at the minimum or maximum end of stroke. The pressure applied to the piston is higher (at high stop) or lower (at low stop) than that corresponding to the mass loaded on the piston. The high and low stop positions are fixed. • High and low spring zones: The combination of pressure and the force of the high or low stop springs have caused the piston to leave the stop. The pressure applied to the piston is within the equivalent of 2 kg of the pressure corresponding to the mass loaded on the piston. The high and low spring zones indicate the piston is about to float. The spring zone positions are fixed. © 2011 Fluke Calibration Page 38 3. GENERAL OPERATION • High and low measurement zones: The piston is within the zone in which a Ready measurement can be made (see Section 3.4.1). The pressure applied to the piston is the pressure corresponding to the mass loaded on the piston. The default value of the high and low measurement zones is ± 2.5 mm around midstroke position. This value can be adjusted in the SETUP file (see Section 3.10). • Midstroke: The piston is at the middle of its stroke. The bottom of the piston (or its equivalent for hollow pistons) is at the reference level marked on the mounting post (see Section 3.9.7). There is equal stroke available to the high and low stops. Note The piston position monitoring system is driven by the internal ring in the base of the mass loading bell. When the bell is not installed, piston position cannot be measured. If the current mass load is less than the mass of the piston + bell, PG7000 will display < ---- > where piston position is normally displayed. If the current mass load is greater than the mass of the piston + bell but the bell is not actually installed, the piston position indication is not valid. Figure 8. Piston Stroke and Zones 3.6 Mass Loading Protocol PURPOSE To provide the operator with mass loading instructions and allow mass entry in convenient nominal values which PG7000 can accurately convert to measured mass values to determine the actual mass load. PRINCIPLE All PG7000 manual mass set masses of 0.1 kg and above, including those in the mass set, the mass loading bell and the piston assembly, are adjusted so that their true mass is within ± 20 ppm of their nominal value. The nominal value is marked on each mass. Each mass’s actual individual value is measured more accurately than ± 20 ppm and reported in the mass set, mass loading bell or pistoncylinder module calibration report. These measured actual values are used by PG7000 in all of its defined pressure calculations. The nominal mass values, and the sequential numbers of the nominal masses when there are several of the same value, are used by PG7000 to describe the mass to load or the mass that is loaded. Following a few simple mass loading rules allows PG7000 to accurately transform actual mass values into nominal mass values and vice-versa so that mass loading instructions to the operator and the operator mass entries can be made in simple nominal mass instructions while representing actual individual mass values. Since the nominal mass values written on the masses and the actual values of the individual masses are different, in pressure to mass mode (see Section 3.9.12) the actual mass value loaded on the piston will be different from the mass loading instruction. In the same manner, in mass to pressure mode the nominal value of the mass load that the operator enters is different from the actual mass loaded on the piston. The mass value shown in the MAIN run screen (see Section 3.7) is always the actual mass value. The mass loading instruction given in pressure to mass mode and the mass value Page 39 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL entered by the operator in mass to pressure mode is always the nominal mass value. If PG7000 mass loading protocol is followed, the conversion from nominal mass to actual mass occurs correctly allowing very simple nominal mass loading and accounting for the operator while using very accurate measured mass values for metrological calculations. OPERATION Note To avoid wear to masses and to the piston-cylinder module end of stroke stops, the piston should not be rotating when masses are loaded or unloaded. Before loading masses, stop piston rotation by hand or using the motorized rotation system (see Section 3.9.13). PG7000 instructions to the operator of mass to load, and operator entries of the mass that is loaded, are always expressed in terms of nominal mass within 0.1 kg and in terms of grams for values under 0.1 kg. This protocol is also followed when using the AMH automated mass handler. In pressure to mass mode, the instruction of the mass to load to achieve the requested pressure is formatted: Load nominal mass: kkk.k kg and gg.gg g In mass to pressure mode, the entry of the mass currently loaded on the piston is formatted: Nominal mass load: kkk.k kg and gg.gg g kkk.k and gg.gg must be expressed and loaded following the mass loading rules below: Mass Loading Rules See Sections 2.3.1.3, Installing Masses On The Mass Loading Tray and 3.6. for PG7000 mass set principles and protocol information. Refer to display examples immediately above to identify kkk.k and gg.gg. See EXAMPLES: Mass Loading below for mass loading examples. PG7000 mass loading entries and instructions are always formatted: kkk.k kg and nn.nn g • The mass set being used must be selected as the active mass set (see Section 3.11.1.10) and must be properly set up prior to use (see Section 3.11.1.6). • kkk.k is made up of the piston, the mass loading bell and mass set masses of 0.1 kg and greater. When using an AMH mass set, the binary mass carrier and mass lifting shaft are also included. • kkk.k always includes the mass loading bell and then the makeup mass, if possible. • kkk.k is made up of the largest masses possible rather than a combination of smaller masses. EXAMPLE: To load 5 kg on a PG7601: USE the piston (0.2 kg) + the mass loading bell (0.3 kg) + the 4.5 kg makeup mass. DO NOT USE the piston + the mass loading bell + 2 kg #1 + 2 kg #2 + 0.5 kg #1. When several masses of the same nominal value are included in kkk.k, they are loaded in numerical sequence, low to high. EXAMPLE: When loading three 5 kg masses: USE 5 kg mass #1, #2 and #3. DO NOT USE 5 kg mass #4, #3, #1. This rule will be followed automatically if masses are installed and used as described in Section 3.6. • gg.gg is made up of mass from the gram trim mass set (masses of 50 g to 0.01 g). These masses can be loaded in any order. © 2011 Fluke Calibration Page 40 3. GENERAL OPERATION Caution When PG7000 provides mass loading instructions and calculates the true mass of the mass currently loaded, it assumes that the mass set in use has been set up correctly (see Section 2.3.1.3). For PG7000 mass loading protocol to operate properly, the mass set in use must be EXACTLY the mass set that has been defined by the add and/or edit mass function (see Section 3.11.1.6). Note In PG7102 and PG7202 High Line Differential mode (see Section 3.9.4.2), there are two types of mass loading instructions: line pressure setting and differential pressure setting. Both follow conventional PG7000 mass loading protocol. The line pressure mass instructions are always given with 0.1 kg resolution. This is to preserve the smaller masses for setting the differential pressure. The differential pressure mass loading instructions follow conventional mass loading protocol using the masses that remain after the line pressure has been set. It may be necessary to adjust the line pressure value to assure that adequate masses remain to set the desired differential pressure. For example, if the line pressure requires loading 10.6 kg, the 0.5 and 0.1 kg masses will be used to set the line pressure and will not be available if they are needed to set the differential pressure. If this situation occurs, consider adjusting the line pressure so that it is set using a whole number of kilograms, in this example 10.0 or 11.0 kg. EXAMPLES: Manual Mass Loading • Mass loading instruction using a PG7601 and a 35 kg mass set: 12.3 kg and 32.33 g • Mass loading instruction using PG7302 and an 80 kg mass set: 77.6 kg and 10.45 g Load: Load: piston (0.2 kg) piston (0.2 kg) bell (0.3 kg) bell (0.8 kg) 4.5 kg #1 (makeup mass) 9 kg #1 (makeup mass) 5 kg #1 10 kg #1 through #6 2 kg #1 5 kg #1 0.2 kg #1 2 kg #1 0.1 kg #1 0.5 kg #1 32.33 g from trim mass set 0.1 kg #1 10.45 g from trim mass set Page 41 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 3.7 Main Run Screen The PG7000 main run screen is its home display that is reached upon powering up and from which other functions and menus are accessed. It is the top level of all menu structures. The main run screen is where PG7000 is left in normal operation. It displays the pressure defined by PG7000, Ready/Not Ready condition and a variety of other information. Note The screen described below is called the MAIN run screen. The SYSTEM and AMBIENT screens (see Sections 3.9.5 and 3.9.6) are also run screens in the sense that all the other functions may be accessed directly from them and pressure or mass entries may be made from them. See Section 3.8 for a description of PG7000’s main menu/function structure. 3. : Current pressure unit of measure (see Section 3.9.3). Current measurement mode. for absolute, for gauge, for differential or high line differential (see Section 3.9.4). Automated pressure generation status. if automated pressure generation is ON. The flashes if automated pressure generation is ON and active, blank if automated pressure generation is not ON (see Section 3.9.9). 4. prvPRESSURE1 unit aGh +N.N mm NN.NNNNN kg 1. 5. : Three Ready/Not Ready indication characters (from left to right: piston position, piston rotation, vacuum reference) (see Section 3.4). Each character indicates as follows: : Piston Position Ready/Not Ready. <*> when Ready, <↑> when Not Ready due to piston above measuring zone, <↓> when Not Ready due to piston below measuring zone, > if piston position not available or out of range. Flashes if condition is Not Ready and piston is floating (see Section 3.4.1). Indicates Not Ready when automated pressure generation is adjusting pressure.
: Piston Rotation Ready/Not Ready. <*> when Ready, <<> when Not Ready due to piston rotation rate too low or motorized rotation engaged, > if piston position not available or out of range. Flashes if condition is Not Ready and piston is floating. (see Section 3.4.2). : Vacuum Reference Ready/Not Ready (PG7601 only). <*> when Ready, if Not Ready due to vacuum not below ready limit. Flashes if condition is Not Ready and piston is floating (see Section 3.4.3). 2. DUT head correction status. if a head correction is currently being applied, blank if head height is zero (see Section 3.9.7). 7. : Current mass load in actual mass (assuming mass loading instructions have been followed and/or mass loading entries are correct (see Section 3.6). 8. <± N.N mm>: Current piston position in millimeters from midstroke position. Positive values are above midstroke, negative values below. Indicates < ---- > if piston position not available or out of range (see Section 3.5). Note : Numerical value and sign of the pressure defined by PG7000 with current mass loaded on current piston when all Ready/Not Ready indication characters indicate Ready. © 2011 Fluke Calibration 6. Page 42 • When a number is too large to display in the allocated screen space, PG7000 displays <********> or . • PG7000 has a screen saver function that causes the display to dim if NO key is pressed for 10 minutes. Pressing a key restores full power to the display. The screen saver activation time can be changed or screen saving can be completely disabled (see Section 3.4.5.1). PG7000™ OPERATION AND MAINTENANCE MANUAL 3.8 General Function/Menu Flow Chart Local operator interface with PG7000 is provided by PG Terminal keypad and display. Normal PG7000 operation is organized around run screens from which functions and menus are accessed. The run screens include: • MAIN run screen: Displays Ready/Not Ready, defined pressure, piston position and mass load. • (2) SYSTEM run screens: (PG7601 only). • Display piston position, rotation, temperature and reference vacuum AMBIENT run screen: Displays ambient pressure temperature, relative humidity and local gravity. The flow chart below outlines the operating protocol of the run screens and associated key presses. POWER UP INTRO SCREEN or RESET ≈ 5 SEC DELAY ESC MAIN RUN SCREEN ESC +/- ESC SYSTEM AMBIENT SYSTEM 1 RUN SCREEN AMBIENT RUN SCREEN ESC or SYSTEM SYSTEM 2 RUN SCREEN TO PREVIOUS RUN SCREEN OTHER FUNCTION KEYS RES P-C UNIT MODE SYSTEM AMBIENT ESC ENT SPECIAL SETUP HEAD ROTATE GEN Figure 9. Run Screen Flow Chart Page 43 © 2010 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 3.9 Direct Function Keys 3.9.1 Direct Function Keys Summary Local operator interface with PG7000 is provided by PG Terminal keypad and display. To minimize the use of multi-layered menu structures, the keypad’s numerical keys also provide direct access to the most commonly used functions. The function accessed is labeled on the bottom half of the each key. Direct function keys are active whenever PG7000 is in a run screen (MAIN, SYSTEM or AMBIENT). Table 14 summarizes the operation of the direct function keys. Section 3.8 provides a flow chart of PG Terminal general operation. See corresponding manual Sections to 3.9.1 to 3.11.9. Note It may be useful to keep a copy of Table 14 near the PG7000, especially when first becoming acquainted with its operation. Table 14. Summary of PG7000 Direct Function Key Operations DIRECT FUNCTION KEYS ARE ACTIVE FROM ANY RUN SCREEN (main, ambient, system). See corresponding manual sections for full detail. © 2011 Fluke Calibration Menu of less frequently used internal functions and settings including resets, user preferences, internal calibration and remote interface setup. Set the resolution of PG7000 mass loading commands (0.01 g to 0.1 kg). Edit and select files that determine the sources of the values for the variables PG7000 uses in calculations of reference pressures. Access SYSTEM run screens (2) which display current measurements of piston behavior, piston-cylinder temperature and reference vacuum if present. Toggles between the two screens. Select pressure to mass or mass to pressure operation. Access AMBIENT run screen which displays current values of barometric pressure, ambient temperature, ambient humidity and local gravity as specified in the active SETUP file. Activate motorized control manually. Activates when pressed, deactivates when released. Adjust height of DUT fluid head correction. Set to zero to defeat head correction. From a run screen: Access mass or pressure setting commands. From other screens: Select menu choice, enter value, confirm selection. Select active piston-cylinder module (range). Turn automatic activation of motorized rotation ON/OFF. Set unit of measure in which pressures are defined. Choice of units available from this key can be customized. Turn automated pressure generation/control ON/OFF, adjust control parameters, set upper limit for automated pressure control. Has no effect if an automated pressure generation/control component is not part of the PG7000 system. Set pressure measurement mode (gauge, absolute, differential). Page 44 3. GENERAL OPERATION 3.9.2 [P-C] PURPOSE To view and/or change the active piston-cylinder module. piston-cylinder module is equivalent to changing the range. In most cases, changing the Note • [P-C] is for selecting the active piston-cylinder module only. Pistoncylinder modules can be added and deleted and their characteristics can be edited by pressing [SPECIAL] and selecting <1PC/MS> (see Section 3.11.1). • The active mass set and mass loading bell are selected by pressing [SPECIAL] and selecting <1PC/MS> (see Section 3.11.1). PRINCIPLE To make valid pressure and mass load calculations, PG7000 must know the exact characteristics of the piston-cylinder module that is currently in use. See Section 7.2 for detailed information on PG7000 pressure and mass calculations. Most PG7000s are used with more than one piston-cylinder module. Detailed characteristics on up to 18 piston-cylinder modules are stored in files. These files can be added, deleted, viewed and edited by pressing [SPECIAL], 1pc/ms (see Section 3.11.1). The P-C function provides rapid access, from any run screen, to viewing the piston-cylinder modules available and selecting one to be active. OPERATION Pressing [P-C] activates the piston-cylinder module viewing and selecting function. Pressing the [P-C] key again or [+/-] while in the P-C function steps through displays of available piston-cylinder modules. When [P-C] is first pressed, a summary of the characteristics of the active piston-cylinder module is displayed, for example: 1. Nominal pressure to mass conversion coefficient of the pistoncylinder module that is currently selected (active). This value is calculated from the effective area of the piston-cylinder and is in kPa (if < 1 MPa) or MPa. Upgraded Type 5000 pistoncylinders may be displayed in psi. 2. Serial number of the active piston-cylinder module. 3. Current pressure unit of measure. 4. Nominal pressure in current pressure units resulting from loading all the mass of the active mass set. 5. Nominal pressure in current pressure units resulting from loading the piston and mass bell only. Active 10 kPa/kg 247 0.7 to 50 psi Pressing [P-C] again or [+/-] causes the screen to step through the other available piston-cylinder modules in the sequence that they were added. 1. Nominal pressure to mass conversion coefficient of the pistoncylinder module. This value is calculated from the effective area of the piston-cylinder and is always in kPa (if < MPa) or MPa. 2. Serial number of the piston-cylinder module. 3. Current pressure unit of measure. 4. Nominal maximum pressure using the piston-cylinder module. Pressure, in current pressure units, resulting from loading all the mass of the active mass set. 5. Nominal minimum pressure using the piston-cylinder module. Pressure, in current pressure units, resulting from loading the piston and mass bell only. Select 200kPa/kg 382 14 to 1000 psi Pressing [ENTER] while in the P-C function causes PG7000 to select the currently displayed piston-cylinder module as the active piston-cylinder module. Page 45 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Pressing [ESCAPE] while in the P-C function returns to the main run screen without changing piston-cylinder modules. Note • The pressure unit of measure in which the range offered by the pistoncylinder module using the active mass set is displayed can be changed by pressing [UNIT] (see Section 3.9.3). • When the current pressure unit of measure is an altitude unit, pistoncylinder module range is expressed in kPa if the altitude unit is meters (m) or psi if the altitude unit is feet (ft). 3.9.3 [UNIT] PURPOSE To specify the unit of measure in which defined pressures are expressed. PRINCIPLE PG7000 supports a wide variety of pressure units of measure. Internally, PG7000 always operates in Pascal [Pa] (the SI unit of pressure). Values of pressure are represented in other units by the application of conversion factors to convert from Pascal. See Section 7.1.1 for Unit of Measure Conversions. The pressure unit of measure selection (e.g., psi, kPa, etc.) is separate from the pressure measurement mode selection (gauge, absolute or differential). See Section 3.9.4 for information on changing the measurement mode. OPERATION To change the pressure unit of measure, press the [UNIT] function key from any run screen. The display is (default units depend on model): 1kPa 2Mpa 3Pa 4bar 5psi 6kcm2 The cursor is on the number corresponding to the active unit. To change the pressure unit, select the desired unit. The display returns to the previous run screen with the selected unit active. If the pressure unit selected is inWa, the reference temperature for water density must be specified. When inWa is selected as the unit, the next display is: Select inWa ref temp 4°C 20°C 20°F Select the desired reference temperature for water density using the [←] or [→] key to move the cursor. Pressing [ENTER] returns to the main run screen with inWa based on water density at the selected reference temperature as the active pressure unit. The current inWa reference temperature can be viewed by observing the position of the cursor in the reference temperature screen. Note • See Section 7.1 for tables of the pressure unit of measure conversion factors used by PG7000. • The UNIT function provides rapid access to a choice of up to six units. The choice of units can be customized from a wider selection by the user (see Section 3.9.3.1). The default units of the UNIT function depend on the PG7000 model. The default units are: • PG7102 and PG7601: 1kPa 2MPa 3Pa 4bar 5psi 6kcm2 • PG7202 and PG7302: 1MPa 2kPa 3bar 4psi 5kcm2 • To restore the default UNIT function settings, use the Unit Reset (see Section 3.11.9.2). © 2011 Fluke Calibration Page 46 3. GENERAL OPERATION • Certain internal and/or metrological functions (e.g., vacuum reference pressure) are always represented in Pascal [Pa], regardless of the active pressure unit of measure. • When the current pressure unit of measure is an altitude unit, atmospheric pressure in the AMBIENT run screen is expressed in kPa if the altitude unit is meters (m) or psi if the altitude unit is feet (ft). 3.9.3.1 Customizing Pressure Units Available Under the UNIT Function PURPOSE To customize the selection of pressure units that are available for selection from the [UNIT] function key. PRINCIPLE The UNIT function provides a choice of different pressure units of measure depending on the PG7000 model. The units that are available by default are those indicated in the Note in Section 3.9.3. However, PG7000 supports many other pressure units of measure. Other units can be made available for selection and units can be deleted by customizing the UNIT function. This allows PG7000 to offer a very wide selection of units while simplifying day to day operation. The typical user will customize the [UNIT] function key to support the six most commonly used units. OPERATION To customize the [UNIT] function key, from the main run screen press [SETUP] and select <2PresU>. The display is: 1. The UNIT number in the [UNIT] menu that is to be changed. Set up user unit #6 Enter the number of the unit position that you would like to change. The display becomes: Unit #6 1SI 2other 3altitude 4user Select the desired pressure unit category. SI units include units based on SI such as mmHg. Select the desired unit from the unit list (see Table 15). Table 15. Pressure Units of Measure Available <1SI> <1Pa> <2Kpa> <3MPa> <4mbar> <5bar> <6mmHg> <7mmWa> <2Other> <1psi> <2psf> <3inHg> <4inWa> <5kcm2> <6Torr> <7mTor> <8none> <3altitude>* <4User>** <1m> <2ft> <1user> * <3altitude>: PG7601 in absolute by vacuum mode only. Not available in PG7102, PG7202 and PG7302. ** <4user>: User defined unit. As there are no altitude units on PG7102, PG7202 and PG7302, user is <3user>. Page 47 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL The UNIT function display is not required to display six units. To delete the current unit from the UNIT screen and show no unit, select <2other>, <6none> for that unit number. If <4user> was selected, the user unit must be defined. The display is: 1. Define user unit: 1.000000 Units/Pa Entry field. Enter the number of user units per Pascal [Pa] in the entry field. Pressing [ENTER] defines the user unit and returns to the screen. Note 3.9.4 • See Section 7.1.1 for the pressure unit of measure conversion factors used by PG7000. • The user defined unit can be assigned a user defined label using the UDU remote command (see Section 4.3.4.2). [MODE] PURPOSE To specify the measurement mode (gauge, absolute, differential) in which PG7000 defines reference pressures. Note Pressure measurement mode and unit of measure are not the same thing. See Section 3.9.3 for information on the pressure unit of measure. PRINCIPLE Pressure defined relative to absolute zero or vacuum is generally referred to as absolute pressure. Pressure defined relative to atmospheric pressure is generally referred to as gauge pressure, with pressure below atmosphere called negative gauge pressure. Pressure defined relative to another pressure that may be, but is not necessarily atmospheric pressure, is generally referred to as differential pressure. PG7000s (depending on the model) can define absolute, gauge, negative gauge and differential pressures. These are referred to as measurement modes. The specific PG7000 measurement modes are as follows: • gauge: Defines gauge pressure by leaving the PG7000 mass load open to atmosphere. This measurement mode is supported by all PG7000 models. It does not allow negative gauge pressures. The minimum gauge pressure is the pressure resulting from loading the combined mass of the piston and the bell on piston-cylinder effective area. • absolute by adding atmospheric pressure (absolute by ATM): Defines absolute pressure by defining a pressure in the same manner as in gauge mode and adding the value of atmospheric pressure measured by a barometer. This measurement mode is supported by all PG7000 models. The barometer can be PG7000’s on-board sensor or a remote RS232 barometer. The uncertainty on the value of atmospheric pressure measured by the barometer must be considered but this value can become relatively small as pressure goes up. For example, if using a ± 0.01 % barometer, the added uncertainty at 1 MPa (150 psi) is 10 ppm. © 2011 Fluke Calibration Page 48 3. GENERAL OPERATION This mode is more convenient and less costly than absolute by vacuum since no vacuum reference needs to be established. However, it does not allow absolute pressures under atmosphere and the minimum absolute pressure is atmospheric pressure plus the pressure resulting from loading the combined mass of the piston and the bell on the piston-cylinder effective area. • absolute by vacuum (absolute by vac) (PG7601 only): Defines absolute pressure by measuring relative to an evacuated bell jar. This mode is time consuming as the vacuum under the bell jar must be made and broken to adjust the mass load for each pressure to be set. This mode is required for setting absolute pressures under atmospheric pressure and for lowest uncertainty under about 1 MPa (150 psi). • differential (dif) (PG7601 only): Defines differential pressures at an absolute static pressure between vacuum and two atmospheres by defining an absolute pressure relative to an evacuated bell jar and subtracting static pressure monitored by a digital pressure indicator. An offsetting technique ensures that only the digital indicator’s resolution and very short term repeatability influence the measurement results. Allows positive and negative differential pressure (including gauge pressures) with one common hardware setup. Covers pressures very near and at zero without limitations due to piston-cylinder size and mass loads (see Section 3.9.4.1). • high line differential (HLdif) (PG7102 and PG7202 only): Defines differential pressures and at gauge static pressure between the lowest gauge pressure supported by the piston-cylinder and the maximum gauge pressure supported by the piston-cylinder. Requires the use of a PG7202 or PG7102 as the “reference” and a second gas operated PG7000 as the “tare”. Differential pressures are defined by setting a common line pressure on both PG7000s and then adding the differential pressure to the reference PG7000 (see Section 3.9.4.2). OPERATION To change the pressure measurement mode, press [MODE] from any run screen. The resulting display depends on the PG7000 Model: The cursor is on the number corresponding to the current measurement mode. Measurement mode: 1gauge 2abs 3HLdif Measurement mode: 1g 2avac 3aatm 4dif PG7102, PG7202, PG7302 [MODE] Screen PG7601 [MODE] Screen For PG7102, PG7202 and PG7302, selecting <2abs> accesses absolute by ATM mode (see PRINCIPLE above). For PG7102, PG7202 and PG7302 selecting <3HLdif> accesses high line differential mode (see PRINCIPLE above and Section 3.9.4.2). For PG7601, selecting <2avac> accesses absolute by vacuum mode. Selecting <3aatm> accesses absolute by ATM mode. Selecting <4dif> accesses differential mode (see PRINCIPLE above and Section 3.9.4.1). Making a measurement mode selection returns to the previous run screen with the selected mode active. Note When using an AMH automated mass handler, be sure to apply drive vacuum to the AMH when switching operation to absolute by vacuum mode (see the AMH-38/AMH-100 Operation and Maintenance Manual). 3.9.4.1 Differential Measurement Mode (PG7601 Only) PURPOSE To define gauge pressures near and under atmospheric pressure that cannot be covered in conventional gauge mode; to define pressures at a static pressure near atmospheric pressure but other than atmospheric pressure. Page 49 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Note Differential mode operation is only supported by PG7601. Differential mode operation requires that the PG7601 SETUP AtmP selection be for an external barometer connected to COM2 (see Section 3.10, 3.11.5.4). Differential mode operation also requires the PG7000 Differential Mode Interconnections Kit P/N 3070127, which includes the Differential Mode Controller. PRINCIPLE Differential mode is designed to allow PG7601 to define pressures which piston gauges have typically not covered or for which they have been difficult to use. This includes positive and negative gauge pressures right around zero, for example ± 2.5 kPa (10 in. H2O); as well as low differential pressures at low absolute static pressures, for example 0 to 2.5 kPa (20 Torr) at 15 kPa absolute (2.2 psia) static pressure. The principal of differential mode is to define a differential pressure as the difference between an absolute pressure relative to vacuum and a second absolute pressure (the static pressure) measured by a precision Reference Pressure Monitor (RPM). An offsetting procedure “tares” the RPM so that the uncertainty contributed to the differential pressure by the RPM’s measurement is a function of the RPM resolution and short term repeatability rather than its absolute accuracy and long term stability. For differential mode operation, the static pressure is set to the desired value as measured by the RPM (atmospheric pressure for gauge pressure measurements). Then RPM offset at the static pressure is determined by applying the static pressure as defined by the PG7601 following: RPMoffset = PGRPMstart - RPMindicate Where: PGRPMstart = RPMindicate = Pressure applied by PG7601 to RPM (nominally equal to RPMstart) RPM reading when PGRPMstart is applied to RPM test port Once the RPM offset has been determined, differential pressures are defined by subtracting the static pressure measured by the RPM from an absolute pressure defined by the PG7601 following: PGdif = PGabs - RPMcurrent - RPMoffset Where: Pgdif Pgabs RPMcurrent RPMoffset = = = = Differential pressure defined by PG7601 Absolute pressure defined by PG7601 Current RPM indication of static pressure Disagreement between PG7601 and RPM at the operating static pressure For differential mode operation, the PG7000 Differential Mode Interconnections Kit should be used to provide the DUT high and low test connections and support the RPM offsetting procedure (see Figure 9 and Section 3.9.4.1). A precision Reference Pressure Monitor, such as a Dhi RPM, must be selected for the AtmP source in SETUP (see Section 3.10, 3.11.5.4). PG7601 differential measurement mode manages the data acquisition and handling to support differential mode operation without operator effort. Differential mode operation requires setting the static pressure at which differential measurements are to be made and regular determination of the RPM offset. These functions are supported under [MODE], <4dif>. © 2011 Fluke Calibration Page 50 3. GENERAL OPERATION 1. PG/STATIC P CONTROL Port 2. PG Port 3. TEST HIGH Port (Quick Connection) 4. TEST LOW Port (Quick Connection) 5. 1/8 in. NPT F or 4T Quick Connector 6. STATIC P Port 7. STATIC P SUPPLY Port 8. CONTROL SELECTION Valve 9. PG SHUTOFF Valve 10. TEST BYPASS Valve 11. Reference Pressure Monitor (RPM) (Optional) 12. STATIC P SELECTION Valve 13. Static Pressure Exhaust (Internal) Figure 10. Differential Mode Controller Schematic Note Technical Note 9940TN02 provides more detailed information on differential mode principles and includes a complete uncertainty analysis. Consult Fluke Calibration or visit www.dhinstruments.com. OPERATION To select differential mode operation and access differential mode functions press [MODE] and select <4dif>. Differential mode operation includes: • Selecting differential mode, setting static pressure and finding the RPM offset (see Section 3.9.4.1, Selecting Differential Mode, Setting Static Pressure, Finding RPM Offset (<1run>). • Operating in differential mode (see Section 3.9.4.1, Operating In Differential Mode). • Viewing differential mode static pressure and RPM offset (see Section 3.9.4.1, Viewing Static Pressure and RPM Offset (<2view>). • Selecting Differential Mode, Setting Static Pressure, Finding RPM Offset (<1run>) Page 51 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Note To operate in differential mode and/or execute the RPM offset determination procedure, the AtmP selection under SETUP must be external (COM2) (see Section 3.10). If this is not the case, an error message will be displayed ( ) when attempting to activate differential mode. To select differential mode operation and/or execute the RPM offset determination procedure, press [MODE] and select <4dif>, <1run>. The display is: 1. Current value of RPM offset. Always in Pascal [Pa]. Offset: 9.03 Pa New offset? 1no 2yes Select <1no> to use the existing RPM offset and return the previous run screen in differential mode (see Section 3.9.4.1, Operating in Differential Mode). Note The RPM offset should be redetermined any time the static pressure value is changed, as well as at the beginning of each calibration or test sequence. If the difference in the value of the RPM offset in subsequent determinations is < 0.1 Pa, then less frequent offset determinations may be considered. Select <2yes> to proceed with a new determination of the RPM offset (see Section 3.9.4.1, PRINCIPAL. The display is: 1. Current pressure reading of the RPM in the current units of. Set RPM P & [ENT] 96.57785 kPa a Table 16. Valve Settings for Setting Differential Mode Static Pressure VALVE VALVE SETTING Control Selection Static P PG Shutoff Closed Test Bypass Open Static P Selection Off If static pressure is atmospheric pressure VENT or OFF With the Differential Mode Controller valves (see Figure 9) set as indicated in Table 16 use the system’s pressure control component to set the desired static pressure value as read by the RPM. If the desired static pressure is atmospheric pressure (for example, for low gauge and negative gauge differential pressures), set the STATIC P SELECTION valve to VENT. If a REFERENCE VOLUME is connected to the STATIC P port and the desired static pressure is less than atmospheric pressure, consider setting the STATIC P SELECTION VALVE to SUPPLY. This will connect the pressure setting vacuum source directly to the reference volume and reduce pull down time. © 2011 Fluke Calibration Page 52 3. GENERAL OPERATION Caution Be sure the TEST BYPASS valve is in the OPEN position before adjusting the static pressure. Failure to do so will cause differential pressure to be generated across the HIGH and LOW TEST ports which may overpressure the DUT. When the static pressure, as indicated by the RPM, has been set to the desired value, press [ENTER]. PG7000 reads the RPM and calculates the mass to load on PG7000 to set the pressure value indicated by the RPM. The display is: Load mass & vac: 9.2 kg & 32.47 g Note When using an AMH automated mass handler, the mass value to set the value indicated by the RPM is always loaded with resolution of 0.1 kg. This is to avoid having to break the reference vacuum to load trim masses. The differential pressure values are loaded with the resolution specified in [RES] (see Section 3.9.10). This is the standard nominal mass loading in instruction display of pressure to mass mode (see Section 3.9.11.1). The mass loading instruction is always given with 0.01 g resolution regardless of the current mass loading resolution setting (see Section 3.9.10). Load the mass value following PG7000 mass loading protocol (see Section 3.6). If an AMH-38 automated mass handler is being used, the mass is loaded automatically and default mass loading reasolution of 0.1 kg is used. Set the PG7000 Differential Mode Controller (see Figure 9) to apply the pressure defined by the PG7000 to the RPM (see Table 17). Table 17. Valve Settings to Apply PG7000 Pressure to the RPM for Differential Mode Offsetting VALVE VALVE SETTING Control Selection PG PG Shutoff Open Test Bypass Open Static P Selection Off If static pressure is atmospheric pressure VENT or OFF Once the valves have been properly set and the mass has been loaded, install the bell jar and establish vacuum under the bell jar. Then press [ENTER]. The display is: 1. Current pressure reading of the RPM in the current units of measure. 2. Standard main run screen piston position indication (see Section 3.7). 3. Current RPM offset (pressure applied by PG7000 - RPM indication). Indicates ******* if the value is > 9999.99 Pa. Pressure unit of measure is always Pascal [Pa]. Page 53 <<* 99.9785 kPa a h -3.5 Off: -7.89 Pa © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Float and rotate the PG7000 piston. When the PG7000 is ready (see Section 3.4), press [ENTER]. Wait about 6 seconds while the PG7000 takes and averages multiple readings. The display is: 1. RPM offset currently in use. 2. New RPM offset just measured. Old offset: -7.72 Pa New offset: -7.89 Pa To record the new RPM offset and go to the previous run screen in differential mode with the new RPM offset active, press [ENTER]. To return to the screen, press [ESCAPE]. See Section 3.9.4.1, Operating in Differential Mode, for information on differential mode operation. Note • The value of the RPM offset typically should be inside of ± 20 Pa. An RPM offset value outside of this range may indicate a poorly calibrated RPM or a problem with the offset determination procedure. • The difference between the RPM offset at the beginning and at the end of a test will be reflected as zero drift of the differential pressures defined during the test. To quantify this value, perform the RPM offset procedure at the beginning and at the end of a test and evaluate the difference between the starting and ending RPM offset values. Operating in Differential Mode Caution Differential mode is often used to calibrate DUTs with very low differential pressure ranges. These can easily be over pressured and damaged by inadvertently applying excessive pressure to one port. OPEN the MPC1-D BYPASS valve to zero the differential across the DUT when there is a risk of overpressure. From a practical standpoint, operating in differential mode is nearly identical to operating in absolute by vacuum mode (see Section 3.9.11). Unless and AMH-38 automated mass handler is used, vacuum under the bell jar must be broken to make mass load changes and reestablished to set a pressure. Operation is in either pressure to mass or mass to pressure mode (see Section 3.9.12). To operate in differential mode, set the PG7000 Differential Mode Controller (see Figure 9) to apply pressure defined by PG7000 to the high side of the DUT (see Table 18). © 2011 Fluke Calibration Page 54 3. GENERAL OPERATION Table 18. Valve Settings for Operating in Differential Mode VALVE Control Selection VALVE SETTING To adjust pressure under PG7000 piston: PG To adjust static pressure: STATIC P PG Shutoff Piston is floating or near floating: OPEN Any other time: CLOSED Test Bypass To make differential pressure measurements: CLOSED To set or check DUT zero: OPEN Static P Selection OFF If static pressure is atmospheric pressure: VENT or OFF Note For the highest quality differential mode measurements, use consistent conditions for static pressure and PG7000 vacuum reference pressure at each point. Set the PG7000 Differential Mode Interconnections Kit CONTROL valve to STATIC P to readjust the static pressure to the same value at each point if necessary. This value should be the same as the value at which the RPM offset was determined. Typical Differential Mode Operational Sequence See Figure 9 for a schematic of the Differential Mode Controller and its valves. Set RPM offset at desired static pressure (see Section 3.9.4.1, Selecting Differential Mode, Setting Static Pressure, Finding RPM Offset (<1run>)). Select pressure to mass or mass to pressure mode (see Section 3.9.12). Put the Differential Mode Controller PG SHUTOFF valve in CLOSED position. If desired, read DUT output at zero differential pressure with TEST BYPASS valve open. Put TEST BYPASS valve in CLOSED position. If desired, put CONTROL valve in STATIC P position and use system control component to adjust static pressure to starting value. Press [ENTER] and enter a pressure or mass value. Load mass as instructed on PG7000. Install bell jar on PG7000, shut the PG7000 vacuum vent valve, open vacuum reference shutoff valve. Wait for vacuum under bell jar to reach Ready condition. If using AMH-38 automated mass handler and mass loading resolution of 0.1kg, mass can be changed without breaking vacuum. Put CONTROL valve in PG position and use system control component to float PG7000 piston. Slowly put PG SHUTOFF valve into OPEN position. Readjust pressure to float the PG7000 piston if necessary. If desired, put CONTROL valve in STATIC P position and use system control component to adjust static pressure to starting value. When PG7000 indicates Ready on all Ready/Not Ready indicators, take DUT reading at differential pressure indicated on the top line of the PG7000 display. Slowly put PG SHUTOFF valve in CLOSED position. Shut vacuum reference shutoff valve, open vacuum vent valve. Wait for pressure under bell jar to return to ambient. Remove bell jar. This step is not necessary is using AMH-38 automated mass handler and mass loading resolution of 0.1 kg. Repeat Steps through Page 55 for each desired differential pressure point. © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Setting Zero Differential Pressure See Figure 9 for a schematic of the Differential Mode Controller and its valves. Zero differential pressure can be set by PG7000 or simply by setting the Differential Mode Controller TEST BYPASS valve to the OPEN position which connects the HIGH and LOW TEST ports together. The TEST BYPASS valve OPEN position is a “true” zero. The zero differential pressure value set by PG7000 has the same uncertainty as any other differential pressure point (see Section 1.2.4.1). Comparing the DUT reading at zero set by TEST BYPASS in the OPEN position and at zero set by PG7000 in differential mode like any other differential pressure, can help quantify the quality of the current RPM offset value. If the difference between the two zero readings is excessive, the RPM offset value should be redetermined (see Section 3.9.4.1, Selecting Differential Mode, Setting Static Pressure, Finding RPM Offset (<1run>)). Also, consider correcting all the DUT readings by the difference in the zero reading found in the two different conditions to correct for the defect in the RPM offset. Viewing Static Pressure and RPM Offset (<2view>) To view the current RPM offset and the static pressure value at which the offset was recorded, press [MODE] and select <4dif>, <2view>. This display is: 1. RPM offset value that is currently active. 2. PG7000 pressure at which RPM offset was recorded. Offset: -6.33 Pa @99.8755 kPa a Press [ESCAPE] to return to the menu. Press [ENTER] to return to the previous run screen. 3.9.4.2 HIGH LINE Differential measurement mode (PG7102, PG7302 and PG7202 ONLY) PURPOSE To define differential pressures at elevated line pressures up to 11 MPa (1 600 psi) with PG7102 or 110 MPa (16 000 psi) with PG7202 in gas and up to 500 MPa (72 500 psi) in oil. Note High line differential (HL dif) mode operation is only supported by PG7102, PG7202 and PG7302. HL dif mode operation requires the use of a two identical PG7102, PG7202 or PG7302s in tandem. Both PG7000s must have nominally identical pistoncylinders and one of the PG7000s must have a “tare” pistoncylinder. The HL dif mode setup for pressure up to 110 MPa (1 600 psi) in gas normally also includes an MPC1-D-1000 or MPC1-D-3000 manual pressure controller and the PK-7100MPCD-DIF Interconnections Kit P/N 3070226. The HL dif mode setup for pressure greater than 110 MPa uses GPC1 (gas), OPG1 or MPG1 (oil) and custom interconnecting hardware. © 2011 Fluke Calibration Page 56 3. GENERAL OPERATION PRINCIPLE Note Technical Note 0080TN03 provides more detailed information on high line differential mode principles including a complete uncertainty analysis. Consult Fluke Calibration or visit our website, www.dhinstruments.com to obtain a copy. High line differential mode is designed to allow two PG7102, PG7202 or PG7302 piston gauges to be used together to define differential pressures relative to line pressures significantly above atmospheric pressure. This capability is most often used to test or calibrate differential pressure devices designed to operate at elevated line pressures under their normal operating line pressure conditions. The main challenge of defining differential pressures at elevated line pressures comes from the very high ratio of line pressure to differential pressure. Relatively small errors and instabilities in the line pressure are very large relative to the differential pressure. The principal of high line differential mode is to “crossfloat” two PG7000 piston gauges so that they define a common line pressure. After the crossfloat is completed, a BYPASS valve is closed isolating one PG7000 from the other. The “tare” PG7000 maintains the line pressure. Mass is added to the “reference” PG7000 to define differential pressures “on top of” the line pressure. The very high precision of the PG7000 gas operated piston gauges allows them to set and maintain a common line pressure with uncertainty much lower than the overall measurement uncertainty on either piston gauge. The two PG7000s are crossfloated at the line pressure prior to making differential measurement to minimize the contribution of line pressure errors to differential pressure. The crossfloating procedure consists of making mass adjustments on the tare PG7000 so that both pistons, when connected together at the line pressure, fall at their “natural” drop rate. PG7000’s embedded piston fall rate measurements and calculations capabilities are used to assist the operator in performing the crossfloat. The PG7000 that is used only to maintain the line pressure on the low side of the device under test (DUT) is designated the tare PG7000. The tare PG7000 is a standard PG7000 but it is normally used with a tare piston-cylinder and a tare mass set (a standard piston-cylinder and mass set may also be used). A tare pistoncylinder effective area is entered as the nominal effective area and its characteristics include k(P), which must be entered in the piston-cylinder definition file (see Section 3.11.1.1). A tare mass set has the same configuration as a standard mass set but the exact values of the masses are not measured and the mass set is set up using nominal values for each mass measured value (see Section 3.11.1.6). The tare piston mass is adjusted to be slightly under the nominal piston mass. This assures that the tare side will always be the light side when crossfloated with the reference side. If a tare piston-cylinder and mass set is not used on the tare PG7000, the tare side may not be the light side in a crossfloat. In this case, the tare mass load on the reference side should be increased slightly (generally < 1 g) to ensure that it is heavier, taking care not to confuse the added tare mass with the subsequent differential mass load. The PG7000 that is used to apply the added pressure to create the differential pressures on the high side of the DUT is designated the reference PG7000. The reference PG7000 is a standard PG7102, PG7202 or PG7302 with a standard piston-cylinder and mass set. The reference must be a PG7102, PG7202 or PG7302. The tare may be a PG7102, PG7202, PG7302 or a PG7601. Page 57 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL The tare and reference PG7000s must have nominally identical piston-cylinders. The two PG7000s operate together with the reference acting as “master” and the tare as “slave”. For communications between the two to occur, COM2 of the reference must be connected to COM1 of the tare with a valid RS232 cable and proper settings must be made on both interfaces (see Section 3.11.5.1). For high line differential mode operation up to 11 MPa (1 600 psi) in gas, an MPC1-D and the PK-7100-MPCD-DIF Interconnections Kit P/N 401645 should be used to provide the DUT high and low test connections and support line pressure and differential pressure setting procedures (see Figure 10). For operation at pressure greater than 11 MPa, GPC1 for gas, OPG1 or MPG1 for oil, and custom interconnecting hardware are used in a functionally similar setup. PG7000 high line differential measurement mode manages the data acquisition and handling necessary to support differential mode operation. Differential mode operation requires setting the line pressure including crossfloating the two PG7000 piston gauges and then setting differential pressures at the line pressure. These two independent operations are supported under [MODE], <3HLdif>, <1run>. Caution © 2011 Fluke Calibration • High line differential mode relies upon the very high sensitivity of PG7000 gas operated piston-cylinders to set and stabilize low differential pressures relative to very high line pressures. To meet the full performance potential of PG7000 high line differential mode operation, external influences on the piston gauges must be minimized. Air currents and vibrations are the most significant possible influences. Do not operate near an active air conditioning or heating duct, avoid opening and closing doors or any movement of personnel around the system. Consider putting the PG7000s in an electrostatic free enclosure if the environment cannot be adequately controlled. • See Section 7.2.1 for information on the exact calculations used by PG7000 to obtain the differential pressures defined in high line differential mode. Page 58 3. GENERAL OPERATION Tare PG7000 (PG7102, PG7302, or PG7601) PG7202, MPC1-D or other VENT Valve Differential Device Under Test (DUT Reference PG7000 (PG7102, PG7202 or PG7302) MPC1-D or other Hi Variable Volume (Hi VV) MPC1-D or other HI/LO BYPASS Valve MPC1-D or other Lo Variable Volume (Lo VV) MPC1-D or other INLET Valve Figure 11. High Line Differential Mode Schematic Note Technical Note 0080TN03 provides more detailed information on high line differential mode principles using PG7102 and PG7601 piston gauges and includes a complete uncertainty analysis. Consult Fluke Calibration or visit our website, www.dhinstruments.com to obtain a copy. OPERATION Note High line differential mode does not support operation with AMH automated mass handler(s). To select high line differential mode operation and access differential mode functions press [MODE], <3HLdif> on the reference PG7000. Note When operating in high line differential pressure mode, the reference PG7000 is “master” and the tare PG7000is “slave”. All operator interaction is with the PG Terminal of the reference PG7000. The tare’s display is for information only. The tare PG7000 keypad is inactive. Page 59 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL High line differential mode operation includes: • Entering differential mode (see Section 3.9.4.2, Entering High Line Differential Mode ([MODE], <3HLdif>, <1run>)). • Setting a line pressure (see Section 3.9.4.2, Setting a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>)). • Setting differential pressures at a high line pressure (see Section 3.9.4.2, Setting Differential Pressures at a High Line Pressure). • Viewing current line pressure and natural fall rates for the line pressure (see Section 3.9.4.2, Viewing Line Pressure, Starting Piston-Cylinder Temperatures and Natural Piston Fall Rates ([MODE], <3HLdif>, <2view>). Entering High Line Differential Mode ([MODE], <3HLdif>, <1run>) To enter high line differential mode, press [MODE] ,<3HLdif>, <1run> on the reference PG7000’s PG Terminal. The reference PG7000 initializes high line differential mode. To successfully enter high line differential mode the reference PG7000 must be able to communicate with the tare PG7000, the piston-cylinders in both PG7000s must be nominally identical and the piston-cylinder in the tare PG7000 must have a non-zero value for k(P) in its piston-cylinder module file. When initializing high line differential mode, the reference PG7000 tests for these conditions and provides error messages as follows: © 2011 Fluke Calibration • : COM2 of the reference PG7000 must be connected to COM1 of the tare PG7000 using a valid RS232 cable and both interfaces must be properly set so that communications between the two PG7000s can occur (see Sections 3.11.5.1, 3.11.5.4). If the reference is unable to communicate with the tare, this error message is displayed. If this error message is observed, correct the communications problem and retry (see Section 3.11.5.1). • : The piston-cylinder modules in the two PG7000s must be nominally identical (have the same nominal mass to pressure conversion coefficient). If the reference PG7000 finds the two piston-cylinder modules to be different, this error message is displayed. If this error message is observed, correct the situation by changing piston-cylinder selection or adjusting one pistoncylinder’s definition (see Sections 3.9.2, 3.11.1.2). • : The piston-cylinder module selected in the tare PG7000 must have a non-zero value for k(P) in the active piston-cylinder module file. If the k(P) value is zero, this error message is displayed. If this error message is observed, correct the situation by providing a non-zero value for k(P) in the piston-cylinder module file of the tare piston-cylinder (see Section 3.11.1.2). Page 60 3. GENERAL OPERATION After high line differential mode is initialized, the display of the tare PG7000 is: 1. Standard Ready/Not Ready indicators for the tare PG7000. 2. Last line pressure set in the pressure unit of measure that was active when the line pressure was set. The unit of measure remains the same, even if the reference PG7000’s unit of measure is changed, until a new line pressure is set. 3. Label indicating that the mass load includes whatever trim masses were loaded on the tare piston in the crossfloating process to reach equilibrium. 4. Tare mass load for the last line pressure set. 5. Current piston position of the tare PG7000. < 1000.70 psi LP - 3.7 30.3 kg + trim While in high line differential pressure mode, the tare PG Terminal keypad is inactive. After high line differential mode is initialized, the display of the reference PG7000 is: 1. Current (last set) value of line pressure in current pressure unit of measure. Line P: 1000 psi New line P? 1no 2yes Select <1no> to use the existing line pressure settings and return to the previous run screen in high line differential measurement mode (see Section 3.9.4.2, Setting Differential Pressures at a High Line Pressure). Select <2yes> to set a new line pressure value or repeat the line pressure setting procedure at the current line pressure value (see Section 3.9.4.2, Setting a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>)). Note The line pressure setting procedure must be executed each time the line pressure is changed and should be executed at the beginning of each differential pressure calibration sequence even if the line pressure is the same as the line pressure previously used. Select <2yes> to go to the line setting procedure. Setting a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>) Caution To meet the full performance potential of PG7000 high line differential mode operation, external influences on the piston gauges must be reduced or eliminated. Air currents and vibrations are the most significant possible influences. Do not operate near an active air conditioning or heating duct, avoid opening and closing doors or any movement of personnel around the system. Caution Leaks at any point in the test system are highly detrimental to measurement results in high line differential mode. Thoroughly leak check the system before operation and correct any leaks detected. Page 61 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Setting a line pressure has two steps: Setting the line pressure and determining the natural fall rates of the tare and reference pistons. Performing a crossfloat of the pistons to reach equilibrium and minimize differential pressure zero offset. To set a line pressure press [MODE], <3HLdif>, <1run>, <2yes> on the reference PG7000’s PG Terminal. The display is: 1. Edit field for entry of desired line pressure value in current pressure unit of measure. Defaults to last value used. Target line P: 1000 psi Note Values of line pressure are always entered and displayed in gauge mode (i.e., relative to atmospheric pressure). Line pressure mass loading instructions are always given with 0.1 kg resolution. The minimum line pressure is the pressure resulting when the mass of the piston + mass bell is loaded. Enter the desired line pressure value. The next display indicates the mass to load to set the line pressure. Load nominal mass on T & R: 33.3 kg This display corresponds to the standard mass loading instruction display of pressure to mass mode (see Section 3.9.11.1). The mass loading instruction is always given with 0.1 kg resolution regardless of the current mass loading resolution setting (see Section 3.9.10). Load the indicated mass on both the tare and reference PG7000s following PG7000 mass loading protocol using main and fractional masses only (do not use trim masses) (see Section 3.6). Press [ENTER] when ready. Note Line pressure mass instructions are always given with 0.1 kg resolution. This is to preserve the smaller masses for setting the differential pressure. The differential pressure mass loading instructions follow conventional PG7000 mass loading protocol using the masses that remain after the line pressure has been set. It may be necessary to select the line pressure value to assure that adequate masses remain to set the desired differential pressure. For example, if the line pressure requires loading 33.3 kg, a 0.2 and 0.1 kg mass will be used to set the line pressure and will not be available if needed to set the desired differential pressure. If this situation occurs, consider changing the line pressure so that it is set using a whole number of kilograms, in this example 33 or 34 kg. Open bypass, float T Close bypass,[ENTER] The display is: © 2011 Fluke Calibration Page 62 3. GENERAL OPERATION With the system’s BYPASS valve OPEN (see Figure 10), use the INLET and/or VENT valve and the high side variable volume to adjust the pressure to float the piston of the tare PG7000. This should cause the reference PG7000 piston to be at its bottom stop. Once the tare piston is floating, close the system’s BYPASS valve and press [ENTER]. The next display is: 1. Tare piston position. 2. Reference piston position. +1.2 mm –3.6 Set T & R to + 1.0&[ENT] Use the high and/or low side variable volume to set both pistons to a position just above + 1.0 mm. Ensure that both pistons are rotating. As both pistons fall through the + 1.0 position, press [ENTER]. PG7000 pauses for 5 seconds and then measures the natural fall rates of both pistons for 30 seconds. 1. Indication of tare piston position. 2. Indication of reference piston position. 3. 30 second count down. T+1.0 mm R+1.0 Findingrates 29 Wait for the 30 second timer to count down while the measurements are made. Be sure not to interfere with the free movement of the piston-cylinders during the countdown. After the countdown completes, the results are displayed: 1. Average fall rate of the taring piston. 2. Average fall rate of the reference piston. 3. Difference between the reference and taring piston fall rates (R - T). This is the “natural fall rate difference”. T-0.9 mm/min R-1.1 ∆-0.2 Save 1yes 2no Note The “natural fall rate difference” is the difference between the fall rates of the two pistons when they are floating naturally at the line pressure. When performing the crossfloat between the pistons in the next step of line pressure setting, the objective will be to adjust the mass of the tare piston until the “crossfloat fall rate difference” measured with the system’s BYPASS valve open is equal to the “natural fall rate difference” measured with the BYPASS closed. Select <2no> to repeat the “natural fall rate difference” measurement. Select <2yes> to accept the “natural fall rate difference” value and continue to the second step of line pressure setting which is crossfloating the two piston-cylinders. The display is: Page 63 Open bypass and [ENTER] to xfloat © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Open the BYPASS valve and press [ENTER] to proceed with the crossfloat. The display is: 1. Tare piston position. 2. Reference piston position. 3. Difference between the “natural fall rate difference” and the “crossfloat fall rate difference” in mm/min. This is the “equilibrium fall rate”. T-1.2 mm R+0.5 ∆+1.2 B/P open,[ENT] Note The “equilibrium fall rate” is the difference between the “natural fall rate difference” and the “crossfloat fall rate difference”: Equilibrium fall rate = (Rnat - Tnat) - (Rxfloat -Txfloat) When the “equilibrium fall rate” is zero, the two piston-cylinders are at equilibrium at the line pressure. Crossfloat the two pistons to find equilibrium by making mass adjustments on the tare piston. Finding equilibrium is an iterative process which is assisted by PG7000. Pressing [ENTER] starts an automated fall rate measurement sequence that measures the fall rates more precisely than the real time indication. Following a 30 second countdown or when one of the pistons has moved more than 1 mm, PG7000 displays the measured “equilibrium fall rate” and suggests the mass adjustment needed on the tare piston to achieve an “equilibrium fall rate” of zero. This process can be repeated as many times as desired. Note If a standard piston-cylinder and mass set (rather than a tare piston-cylinder and mass set) is being used in the tare PG7000, the tare piston may be the heavy one at the start of the crossfloat (see Section 3.9.4.2, PRINCIPLE). In this case, rather than reduce mass on the tare PG7000, add a small amount of mass to the reference PG7000 before starting the crossfloat so that the tare piston will be the lighter one (generally < 1 g is needed). This will simplify the crossfloating procedure and subsequent differential mass loading. Be sure not to confuse the additional tare mass on the reference piston with the subsequent differential mass loads. Using trim masses (50 g and below) from the trim mass set, adjust the mass load on the tare PG7000 to establish equilibrium (“equilibrium fall rate” near zero). When the mass adjustments are complete, close the BYPASS valve and put both pistons at a position just above 0.0 mm. Ensure that both pistons are rotating. As the pistons fall through the 0.0 mm point, OPEN the BYPASS valve, then press [ENTER]. After a 5 second delay for stabilization, PG7000 begins the fall rate measurement. The display is: 1. © 2011 Fluke Calibration Indication of tare piston position. 2. Indication of reference piston position. 3. 30 second count down. 4. Average “equilibrium fall rate” since start of the countdown. Page 64 T +0.1 mm R +0.0 ∆-12.2 Xfloating 30 3. GENERAL OPERATION PG7000 is measuring the drop rate of the two pistons and calculating the “equilibrium fall rate”. Be sure to not interfere with the free movement of the piston-cylinders or to introduce any outside interference while the countdown occurs. The fall rate measurement continues until either piston has moved 1 mm or 30 seconds have elapsed. The measurement can also be concluded by pressing [ENTER]. When the fall rate measurement has concluded, the results are displayed: 1. Average “equilibrium fall rate” measured in last crossfloat. 2. Suggested mass change (in grams) on the tare (T) piston to achieve equilibrium based on equilibrium fall rate just measured. ∆-10.3 +0.050 g on T Repeat 1yes 2no Evaluate the quality of the equilibrium. Check the value of the “equilibrium fall rate”. Consider that the closer the rate is to zero, the better the equilibrium; consider that the suggested mass change on the tare piston times the mass to pressure conversion coefficient of the piston approximates the pressure value of the defect in the equilibrium. Check the difference between the DUT output with the BYPASS open and BYPASS closed to evaluate the zero error due to the equilibrium. Select <1yes> to return to the crossfloat screen. If you are satisfied with the quality of the equilibrium, select <2no> to go to the run screen in high line differential pressure mode and set differential pressures at this line pressure (see Section 3.9.4.2, Setting Differential Pressures at a High Line Pressure). Note • The value of the average “equilibrium fall rate” measured by the crossfloat should typically be inside of ± 5 mm/min and/or the suggested mass adjustment should be inside of 50 mg. • The device under test (DUT) can often be used to evaluate the quality of the equilibrium between the two pistons and to correct for the zero error caused by the defect in equilibrium. Observe the change in the DUT output when the system BYPASS valve is opened and closed with the pistons floating. With the system BYPASS valve open, by definition, the differential pressure applied to the DUT is zero. The change in the DUT output observed when the BYPASS valve is closed with the pistons floating is the zero error due to the defect in the crossfloat equilibrium. Consider correcting all of the DUT readings by the value of this offset to correct for the crossfloat zero error. When using the offset, keep in mind that, due to differential evolution of the two PG7000 piston-cylinder temperatures, the differential pressure at zero differential mass load may not be zero. Setting Differential Pressures at a High Line Pressure Caution • To meet the full performance potential of PG7000 high line differential mode operation, external influences on the piston gauges must be reduced or eliminated. Air currents and vibrations are the most significant possible influences. Do not operate near an active air conditioning or heating duct, avoid opening and closing doors or any movement of personnel around the system. Page 65 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL • It is not possible to operate in differential mode with leaks at any point in the test system. Thoroughly leak check the system before operation and correct any leaks detected. • High line differential mode is often used to calibrate DUTs whose differential pressure range is very low relative to the line pressure. These DUTs can easily be over pressured by inadvertently applying excessive pressure to one port. When operating in high line differential pressure mode, OPEN the system BYPASS valve to zero the differential across the DUT when there is any risk of overpressure. Before operating in differential mode the line pressure must be set (see Section 3.9.4.2, Setting a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>)). Once the line pressure is set, from a practical standpoint, operating in high line differential mode is very similar to operating in gauge mode (see Section 3.9.11). All operator interaction is with the PG Terminal of the reference PG7000. The PG Terminal of the tare PG7000 is for display only. The tare PG7000 PG Terminal keypad is locked out during high line differential mode operation. Operation is in either pressure to mass or mass to pressure mode (see Section 3.9.12). To operate in differential mode the system BYPASS valve must be closed (see Figure 10). All mass loading and unloading to define differential pressures is made on the reference PG7000. Mass loading and unloading instructions follow conventional PG7000 mass loading protocol (see Section 3.6) but the differential mass loading instructions are “on top of” the line pressure mass load. The mass load corresponding to the line pressure on the tare and reference PG7000s must remain exactly the same throughout the differential measurements. While in high line differential mode, the piston position and piston rotation Ready/Not Ready indicators on the reference PG7000 reflect the status of both the tare and the reference PG7000s. For the indicator to indicate Ready, both the tare and the reference PG7000s must meet their current Ready/Not Ready criteria (see Section 3.4). A in a Ready/Not Ready status indicator indicates that the tare PG7000 is Not Ready. Note © 2011 Fluke Calibration • When operating in high line differential mode, the measurement mode indicating character directly to the right of the pressure unit of measure in the main run screen is (see Section 37). • Zero differential pressure (BYPASS valve closed) can be set by the system. In some cases, the differential mass to load to set zero is not zero. This is normal and can occur due to differential changes in the temperature of the tare and reference piston-cylinders after the original line pressure crossfloat. A differential piston-cylinder module temperature change while running in high line differential mode affects the line pressure causing an offset to the differential pressure. The offset to the differential pressure causes a non-zero differential mass load to be needed to set zero differential pressure (see Section 7.2.1). Conversely, and for the same reason, a differential mass load of zero may not result in a differential pressure of zero. Page 66 3. GENERAL OPERATION Typical Sequence to Set Differential Pressures at a High Line Pressure Set the desired line pressure (see Section 3.9.4.2, Setting a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>)). Select pressure to mass or mass to pressure mode (see Section 3.9.12). If desired, read DUT output at zero differential pressure with BYPASS valve OPEN. Then, put the BYPASS valve in CLOSED position. Float both PG7000 pistons before operating the BYPASS valve. Read DUT output at zero as defined by the PG7000s if desired (this value can be used as an offset correction on subsequent DUT differential pressure outputs). Press [ENTER] and enter a pressure or mass value. All mass loading and unloading to define differential pressures is performed on the reference PG7000. Follow normal PG7000 mass loading protocol (see Section 3.6). Note, however, that the mass loading instructions are “on top of” the mass already loaded to define the line pressure. The mass corresponding to the line pressure must remain exactly the same on both the tare and reference PG7000s throughout the differential pressure measurements. Float the tare and reference pistons. For best results put both pistons slightly above the 0.0 point, for example +0.3 mm. When the reference PG7000 indicates Ready on both Ready/Not Ready indicators (see Section 3.4), take the DUT r ead i n g at t h e d if f er en t ia l pr es s ur e in d ic at ed o n th e t o p l i n e of the reference display. For best results log DUT readings and average them over time while the PG7000 pistons fall through the 0.0 mm point. Averaging for 10 to 30 seconds allows the random pressure noise from piston rotation to be integrated and eliminated. Repeat Steps through for each desired differential pressure point. If the last differential pressure is zero, consider reading the DUT output with the BYPASS valve CLOSED. Then OPEN the BYPASS valve to set “true” zero differential pressure. Viewing Line Pressure, Starting Piston-Cylinder Temperatures and Natural Piston Fall Rates ([MODE], <3HLdif>, <2view>) To view the last line pressure setting, the starting piston-cylinder module temperatures, the natural piston fall rates and the “natural fall rate difference” recorded at that line pressure, press [MODE] and select <3HLdif>, <2view>. The display is: 1. Last line pressure value set (line pressure is always in gauge mode). 2. Temperature of the reference piston-cylinder logged when the line pressure crossfloat was completed. 3. Temperature of the tare piston-cylinder T 21.24 ºC R 21.47 module logged when the line pressure crossfloat was completed. The starting temperatures are the temperatures logged at the time the line pressure setting crossfloat was completed. These temperatures are used in calculating corrections to the differential pressure to take into account differential temperature changes in the piston after completion of the crossfloat (see Section 7.2.1 for differential pressure calculation information). Page 67 Line P: 1000 psi © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Press [ENTER] to continue to the next view screen or [ESCAPE] to go to the previous screen. The next view screen is: 1. Average fall rate of the tare piston recorded at the last line pressure. 2. Average fall rate of the reference piston recorded at the last line pressure. 3. Difference between the reference and tare piston average fall rates (R - T). This is the “natural fall rate difference”. T-0.91 mm/min R-1.10 ∆ -0.2 Save 1yes 2no Note The “natural fall rate difference” is the difference between the fall rates of the two pistons when they are floating naturally at the line pressure. The “natural fall rate” can change slightly as operating conditions change so the “natural fall rate” determination and piston crossfloat procedure should be performed at the beginning of each sequence of differential pressure definitions. To return to the menu, press [ESCAPE]. To return to the previous run screen, press [ENTER]. 3.9.5 [SYSTEM] PURPOSE To access the two SYSTEM run screens which display current values of PG7000 piston behavior, piston-cylinder temperature and vacuum reference (PG7601 only) measurements. OPERATION To access the SYSTEM run screens, press [SYSTEM] from any other run screen. There are two SYSTEM run screens. Pressing [SYSTEM] or [±] when in a SYSTEM screen toggles between the first and second SYSTEM screens. See Sections 3.9.5.1 and 3.9.5.2 for detailed information on the contents of the first and second SYSTEM run screens. Note The SYSTEM screens are run screens. This means that other functions can be accessed from the SYSTEM screens and the active SYSTEM screen will be returned to when leaving functions. The MAIN screen and AMBIENT screen are also run screens (see Section 3.8). 3.9.5.1 First System Run Screen OPERATION The first SYSTEM run screen provides real time display of piston rotation rate, decay in piston rotation rate, piston position and piston fall rate. To access the first SYSTEM run screen, press [SYSTEM] from any other run screen. © 2011 Fluke Calibration Page 68 3. GENERAL OPERATION The first SYSTEM run screen displays: 1. 2. : Numerical value of current piston rate of rotation. The unit of measure is rotations per minute [rpm] and cannot be changed. Flashes when Not Ready and piston is floating (see Section 3.4.2). Indicates < ---- > when information is unavailable or out of range. <±nn/min>: Numerical value of current decay in piston rotation rate (deceleration). The unit of measure is rotations per minute [rpm/min] and cannot be changed. < ---- > when information is unavailable or out of range. nnn rpm ±n.nn mm ±nn/min ±n.nn/min 3. <±n.nn/min>: Sign and numerical value of current piston vertical rate of displacement. A negative value indicates piston falling. A positive value indicates piston rising. The unit of measure is millimeters per minute [mm/min] and cannot be changed. Indicates < ---- > when information is unavailable or out of range. 4. <±n.nn mm>: Sign and numerical value of current position of the piston within the piston stroke (see Section 3.5). The unit of measure is millimeters [mm] away from mid-stroke position and cannot be changed. Indicates when the piston is at the high stop (all the way up) and when the piston is at the low stop (all the way down). Flashes when Not Ready and piston is floating (see Section 3.4.1). Indicates < ---- > when information is unavailable or out of range. Pressing [ESCAPE] in the first SYSTEM run screen returns operation to the MAIN run screen. Pressing [SYSTEM] or [±] toggles between the first and second run screen. All function keys are active from the first SYSTEM run screen and operation returns to that screen when leaving functions the were accessed from it. Note • The measurement systems for piston behavior indications rely on movement of the mass loading bell. Piston behavior indications (piston position, piston rotation rate) are not valid when the mass bell is not loaded on the piston. • See Section 3.5 for information on the piston stroke and measurement zone. 3.9.5.2 Second System Run Screen OPERATION The second SYSTEM run screen displays the values of piston-cylinder temperature and temperature rate of change measured by PG7000. PG7601 also displays reference vacuum and vacuum rate of change. If internal or external measurement is specified for the vacuum values in SETUP (see Section 3.10), the second SYSTEM run screen provides a real time display of the values measured. To access the second SYSTEM run screen, press [SYSTEM] or [±] from the first SYSTEM run screen. To access the first SYSTEM run screen, press [SYSTEM] from any run screen. Page 69 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL The second SYSTEM run screen displays: 1. 2. 3. 4. : Current piston-cylinder temperature. Source of value can be internal measurement, default or user depending on current SETUP selection (see Section 3.10). The unit of measure is degrees Centigrade [±°C] and cannot be changed. Indicates < ---> when information is unavailable or out of range. n.nn°C nnn.n Pa ±n.nn/min ±nnn.n/min <+n.nn/min>: Sign and numerical value of current rate of change of temperature. A negative value indicates temperature decreasing. A positive value indicates temperature increasing. The unit of measure is degrees Centigrade per minute [°C/min] and cannot be changed. Indicates < ---- > when information is unavailable or out of range. Indicates < ---- > when “user” or “default” is the current SETUP selection for piston-cylinder temperature source (see Section 3.10). <+nnn.n/min>: (Used for PG7601 only.) Sign and numerical value of current rate of change of the vacuum reference pressure. A negative value indicates pressure decreasing. A positive value indicates pressure increasing. The unit of measure is Pascal per minute [Pa/min] and cannot be changed. Indicates < ---- > when information is unavailable or out of range. if the PG7000 model is not PG7601 or if user or default is the current SETUP selection for reference vacuum (see Section 3.10). : (Used for PG7601 only.) Current vacuum reference value. Can be internal measurement, external measurement, default or user depending on current SETUP selection. Flashes when Not Ready and piston is floating (see Section 3.4.3). The unit of measure is Pascal [Pa] and cannot be changed. Indicates < >20 Pa > if current SETUP selection is internal or external and current measurement is out of range or greater than 20 Pascal. if PG7000 model is not PG7601. Pressing [ESCAPE] in the second SYSTEM run screen returns operation to the MAIN run screen. Pressing [SYSTEM] or [±] toggles between the first and second SYSTEM run screen. All function keys are active from the second SYSTEM run screen and operation returns to that screen when leaving functions that were accessed from it. Note The current selection in SETUP determines the source of the values used by PG7000 for piston-cylinder temperature and vacuum reference values. If the SETUP setting is user or default, the SYSTEM screen displays the user or default value, not PG7000’s on-board measurement(s). 3.9.6 [AMBIENT] PURPOSE Access the AMBIENT run screen which displays the current ambient condition values being used by PG7000 for calculations of reference pressures. PRINCIPLE PG7000 uses ambient condition values to calculate the reference pressures that it defines (see Section 7.2). The source of the ambient condition values is specified in the current SETUP file (see Section 3.10). The AMBIENT run screen displays the current ambient condition values. If the SETUP selection for the ambient condition is internal measurement, then the AMBIENT run screen provides a real time display of the measurement of PG7000’s on-board sensor for that variable. OPERATION To access the AMBIENT run screen, press [AMBIENT] from any other run screen (MAIN or SYSTEM). © 2011 Fluke Calibration Page 70 3. GENERAL OPERATION Note The AMBIENT screen is a run screen. This means that other functions can be accessed from the AMBIENT screen and the active AMBIENT screen will be returned to when leaving functions. The MAIN screen and SYSTEM screens are also run screens (see Sections 3.7 and 3.9.5). The AMBIENT run screen displays: 1. : Current numerical value and pressure unit of atmospheric pressure. Source of value can be internal measurement, remote barometer, default or user depending on current SETUP selection (see Section 3.10). The unit of measure is determined by the UNIT function setting (see Section 3.9.3). ATM head is applied to internal or external barometer reading to correct atmospheric pressure to the PG7000 reference level (see Section 3.11.3.3). Indicates < --- > when information is unavailable or out of range. Indicates if the current SETUP selection is a remote barometer and PG7000 communication with the barometer times out. nnnnnnn uuuu nn.n°C nn%RH n.nnnnnn/s2 2. : Current ambient temperature. Source of value can be internal measurement, default or user depending on current SETUP selection (see Section 3.10). The unit of measure is degrees Centigrade [ºC] and cannot be changed. Indicates < ---- > when information is unavailable or out of range. 3. : Value of local acceleration due to gravity. Can be default or user depending on current SETUP selection (see Section 3.10). The unit of measure is meters per second squared (m/s2) and cannot be changed. : Current ambient relative humidity. Can be internal measurement, default or user depending on current SETUP selection (see Section 3.10). The unit of measure is percent relative humidity (%RH) and cannot be changed. Indicates < -- > if current SETUP selection is internal and current measurement is unavailable or out of range. 4. Pressing [ESCAPE] in the AMBIENT run screen returns operation to the MAIN run screen. All function keys are active from the AMBIENT run screen and operation returns to that screen when leaving functions that were accessed from it. Note • The current selection in SETUP determines the source of the values used by PG7000 for atmospheric pressure, ambient temperature and relative humidity. If the SETUP setting for these values is user or default, the AMBIENT screen displays the user or default value, not PG7000’s onboard measurement(s). • To change the ambient pressure units of measure, see Section 3.9.3. When the current pressure unit of measure is an altitude unit, atmospheric pressure in the AMBIENT run screen is expressed in kPa if the altitude unit is meters (m) or psi if the altitude unit is feet (ft). 3.9.7 [HEAD] PURPOSE To cause a pressure value, representing the fluid head resulting from a difference in height, to be added to the pressure defined by PG7000 at its reference level. To set the height of the DUT head. PRINCIPLE The pressure defined by the PG7000’s floating piston is the pressure at the bottom of the piston. This is referred to as the PG7000 reference level. The height of the bottom of the piston with the piston in mid-stroke position is marked reference level on the PG7000 piston-cylinder module mounting post. Frequently, when performing a calibration or test, the device or system under test is at a different height than the PG7000 reference level. This difference in height (referred to as DUT head) can cause a significant difference between the pressure defined by the PG7000 at its reference level and the pressure actually applied to the device under test located at a different height. In this case, it is useful to make a head correction to the pressure defined by the PG7000 at its reference level in order to accurately predict the pressure actually applied at a different height. The HEAD function allows this head correction Page 71 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL to be applied automatically for a variety of fluids based on operator entry of the height difference. The fluid used and the HEAD function units of measure are set by pressing [SPECIAL] and selecting <3head> (see Section 3.11.3). PG7000 can accurately determine head pressures for gases (nitrogen, helium and air) and liquids (Di-2 ethylhexyl Sebacate oil, water and a user defined liquid) as the pressurized medium. Note • When gas is the test fluid, use of the HEAD function is most important at low absolute pressures. In this case, specifying the head height within ± 0.2 in. (5 mm) is adequate to ensure that, even in the worst case, the uncertainty on the head correction will be insignificant relative to the tolerance on the PG7000 measurement. Use of the HEAD function to ensure tolerance measurements is particularly critical when a liquid is the test fluid, due to the high density of liquids. To determine when and how precisely a head correction for liquids must be made, 0.03 psi/inch (90 Pa/cm) may be used as an estimation of the liquid head value. • Regardless of the head function’s setting, corrections are automatically applied to the calculated reference pressure to compensate for the deviation between the current piston position and the mid-stroke position (see Section 3.11.3.4). • The pistons of certain gas operated piston-cylinder modules are hollow. Due to their irregular shape, for these pistons, the natural reference level is not at the bottom of the piston. So that, in practice, the actual reference level is the same for all piston-cylinder modules, a reference level offset is applied when a hollow piston is used. The reference level offset is included in the piston-cylinder file (see Section 3.11.1.1) and corrects the reference level back to the reference point marked on the mounting post. OPERATION To access the HEAD function, press [HEAD]. The display is: 1. Test fluid currently specified for the head correction. 2. Entry field for head height (1 to 999 cm or in.). Edit DUT head height 95 cm N2 Entering a value of zero turns the HEAD function OFF. Entering a value other than zero turns the HEAD function ON using the height entered. Pressing [ESCAPE] returns to the main run screen with NO change to the current head setting. (+) Reference Line REFERENCE LEVEL (-) © 2011 Fluke Calibration Page 72 3. GENERAL OPERATION Note • The reference height of PG7000 pressure definition is the bottom of the piston in mid-stroke float position. This position is marked on the pistoncylinder module mounting post and on the optional AMH, mass automated handler. The DUT head height should be entered as a positive value if the device or system under test is higher than the PG7000 reference level and negative if it is lower. • To change units of DUT head height between inches and centimeters and to change the test fluid, press [SPECIAL] and select <3Head> (see Section 3.11.3). • When the HEAD function is ON (DUT head value different from 0), the application of a head correction is indicated by in the right side of the top line of the MAIN run screen (see Section 3.7). When the HEAD function is OFF, the is NOT shown. PG7000’s also have a separate head correction to compensate for the deviation between the current piston position and mid-stroke (see Section 3.11.3.4). This PISTON head can be turned ON and OFF (see Section 3.11.3.4). 3.9.8 [ROTATE] PURPOSE To turn automatic control of motorized piston rotation (acceleration and brake) ON and OFF. Note See Section 3.9.13 for information on manual control of motorized piston rotation. PRINCIPLE The motorized piston rotation system is used to start or increase piston rotation rate when the piston is floating. It is also used to stop piston rotation when necessary, for example before manipulating mass to set a new pressure. The system operates by engaging a motor driven belt around the bottom of the mass loading bell to accelerate or brake the rotation rate of the mass bell and piston it is loaded on. The motorized rotation system can engage with the piston at any position in its stroke and at any rotation speed with minimal impact on piston position and the defined pressure. When the motorized rotation system disengages, the piston is completely free. With automatic motorized rotation ON, the motorized rotation system engages and disengages automatically as needed when the piston is floating to maintain the piston rotation rate above the minimum rate Ready limit (see Section 3.4.2). The rotation rate is measured by PG7000 on board sensors and the rotation rate limits are set in the file of the active piston cylinder (see Section 3.11.1.1). Whenever the piston is floating, the motorized rotation system will attempt to maintain the piston rotation rate within the Ready limits (except under the cutoff mass load of 3 kg, at which the low limit is reduced to minimum to maximize free rotation time. The piston rotation Ready/Not Ready indication character indicates Not Ready to alert the operator when the motorized rotation system is about to engage. The rotation system will not engage when the current mass load is less than the mass of the piston + mass loading bell. The motorized rotation system is also used to brake and stop rotation when starting a new pressure point. If the piston is floating and rotating when [ENTER/SET P] is pressed, the motorized rotation system engages at a speed near the rotation rate of the piston and then brakes it to a stop. Piston rotation is stopped to avoid loading and unloading masses on the rotating piston and to avoid stopping rotation by friction between the piston and the piston end of stroke stops. With automatic motorized rotation OFF, the motorized rotation system engages only when actuated by the operator. Pressing [ ] accelerates rotation or [ ] followed by [←] stops rotation (see Section 3.9.13). Page 73 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Automatic motorized rotation is generally left ON for normal operation. It is turned OFF in situations where rotation system actuation independent of operator initiative is undesired (e.g. when performing a crossfloat intercomparison with another piston gauge). OPERATION To access the AutoRotate [ROTATE]. The display is: functions, press AutoRotate OFF 1on 2pre-decel or indicates the current state of AutoRotate. Use <1on> or <1off> to change the state. When automated rotation is OFF, the PG7000 motorized rotation system will only engage if the operator presses [ ] or [ ] followed by [←] (see Section 3.9.13). When automated rotation is ON, the automated rotation system engages automatically. When the piston is floating, it engages as needed to maintain the piston rotation rate above the Ready limit (see Section 3.4.2). When [ENTER/SET P] is pressed, it engages to stop rotation before masses are loaded and/or the pressure is adjusted (this function can be turned ON and OFF, see Section 3.9.8.1). Manual control using [ ] and [ ] followed by [←] is also still active. Note • Automatic motorized piston rotation is dependent on PG7000’s measurements of piston position and rotation rate. These measurements are only available when the mass loading bell is installed on the piston. Automatic piston rotation is suspended when the current PG7000 mass load does not include the mass loading bell. • Proper operation of the rotation system is dependent on PG7000’s measurements of piston position. When using AutoGen, be sure that the piston position indication system is properly adjusted (see Section 5.2.2). • When PG7601 is operating in absolute by vacuum mode using the internal vacuum sensor to measure reference vacuum, automatic motorized piston rotation will not engage until the reference vacuum value is within the Ready limit (see Section 3.4.3). 3.9.8.1 <2PRE-DECEL> PURPOSE To turn ON and OFF a function that causes the automated rotation system to begin the piston rotation deceleration when [ENTER/SET P] is pressed rather than at the time mass is to be loaded. On a PG7000 Platform equipped with motorized rotation, when AutoRotate is on, the piston rotation deceleration function is used to stop rotation before masses are loaded or pressure is adjusted. As the deceleration function can take up to one minute to execute, it can be initiated the moment that [ENTER/SET P] is pressed to enter a new pressure or mass target. The deceleration function then runs while the new target value is being entered. However, when the next target does not require changing main masses, it may not be necessary to stop piston rotation. In these cases, the running the deceleration function is probably not desired and it is not beneficial for piston deceleration to initiate it when [ENTER/SET P] is pressed. For this reason, the function to start piston deceleration when [ENTER/SET P] is pressed can be turned ON and OFF. If pre-deceleration is ON and AutoRotate is ON, piston rotation deceleration always initiates immediately when [ENTER/SET P] is pressed. If pre-deceleration is OFF and AutoRotate is ON, piston rotation deceleration occurs after entry of the pressure or mass target value and only if the new target requires changing main masses with a manual mass set or operation AMH if an automated mass handling system is active. © 2011 Fluke Calibration Page 74 3. GENERAL OPERATION OPERATION To turn ON and OFF the function that causes stopping piston rotation to initiate when [ENTER/SET P] is pressed, press [ROTATE], <2pre-decel>. The cursor is on the choice corresponding to the current state. Select <2on> for piston rotation deceleration to initiate when [ENTER/SET P] is pressed. Select <1off> for deceleration to initiate only after entry of a new pressure or mass target and only if main masses need to be moved or AMH automated mass handling needs to be operated. The default is <2on>. 3.9.9 [GEN] (OPTIONAL) PURPOSE To turn ON and OFF automated pressure generation/control and view and edit automated pressure control settings. Requires that an automated pressure generation/control component be included in the PG7000 system and properly configured (see Section 2.4.9). PRINCIPLE PG7000s support automated pressure generation/control components. These components, when properly configured and interfaced with the PG7000 platform, are controlled by the PG7000 platform to automatically set and adjust pressure to float the piston. Automated pressure generation and control components are interfaced via the PG7000 platform’s COM3 RS232 port (see Section 3.11.5.1). Once an automated pressure generation/control component has been properly configured and interfaced with the PG7000 platform, the functions under [GEN] are used to turn automated pressure generation/control ON and OFF and to set operating parameters associated with automated pressure control. With automated pressure generation/control ON, PG7000 uses the automated control component to float and refloat its piston when a pressure or mass value is entered. [ENTER/SET P] is pressed to initiate a new command. The automated control function is suspended when any function key is pressed, when entering remote mode or if automated pressure control is turned OFF. After a target pressure or mass has been entered (locally or remotely) and the required mass has been loaded, the GEN function controls the pressure control component as needed to float the PG7000 piston at the piston float target (see Section 3.9.9.1) and refloat it if necessary. If the piston moves beyond the high or low piston position Ready limit the GEN function refloats it to the piston float target. The piston float target and piston position ready limits are user adjustable (see Section 3.10 <6READY). See Section 3.5, Figure 8 for a description piston position stroke zones and limits. With automated pressure generation/control OFF, PG7000 leaves the automated pressure generation/control component idle and does not attempt to use it. If an automated pressure control component is included in the PG7000 system, automated pressure generation/control is generally left ON for normal operation. It is turned OFF in situations where pressure control independent of operator initiative is undesired (e.g. when performing a crossfloat intercomparison with another piston gauge). The [GEN] menu includes: 1. Turning the GEN function ON and OFF. 2. The adjustable piston float target that defines the position to which the piston is set when floated (see Section 3.9.9.1). 3. A choice to have the piston raised to the top of the stroke before manipulating mass using an AMH automated mass handler (see Section 3.9.9.2). 4. Viewing and setting the UPPER LIMIT of the automated pressure generation component to avoid accidental overpressure (see Section 3.9.9.2). 5. Viewing and setting the assumed pressure controller tolerance used to determine pressure setting limits when floating the PG7000 piston (see Section 3.9.9.4). 6. A choice to not readjust piston position if the piston is already floating within the Ready position limit after a new target is executed (see Section 3.9.9.5). Page 75 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 7. Viewing and setting the volume of the system to which the PG7000 system is connected (PG7302 and PG7307 only) (see Section 3.9.9.6). OPERATION To access the GEN functions, press [GEN]. display is: The AutoGen OFF 1on 2target 3raise 4UL 5tol 6refloat or indicates the current state of AutoGen. Use <1on> or <1off> to change the state. If automated pressure control is OFF, PG7000 attempts to turn automated pressure control ON when <1on> is pressed. To do so, communication must be established with a valid automated pressure control component over its COM3 RS232 port (see Section 3.11.5.1). If PG7000 is not able to establish communication with a valid pressure control component, is displayed momentarily. Correct the communications error with the pressure control component and try again. If PG7000 is able to establish communications with a valid automated pressure control component,
is displayed momentarily and automated control is turned ON. The PG7000 automated control function will attempt to float the piston within the piston position ready limits (see Section 3.4.1) after any mass to pressure or pressure to mass command once mass loading has been confirmed. It will continue to attempt to float its piston until [ENTER] is pressed to initiate a new command, a function key is pushed to interrupt AugoGen, or automated pressure control is turned OFF using [GEN]. If automated pressure control is ON and <1off> is pressed, is displayed momentarily and automated control is turned OFF. The PG7000 automated control function is inactive. Note • Automated pressure generation/control ON is indicated by a in the right of the top line of the PG Terminal main run screen. The flashes when automated generation is active (see Section 3.7). • Proper operation of the automated pressure generation/control function is dependent on PG7000’s measurements of piston position. When using AutoGen, be sure that the piston position indication system is properly adjusted (see Section 5.2.2). • Proper operation of the automated pressure generation/control function is dependent on PG7000’s measurements of piston position and rotation rate. These measurements are only available when the mass loading bell is installed on the piston. Automated pressure generation/control is automatically turned off when the PG7000 entered pressure or mass load does not include the mass loading bell. • Automated pressure generation/control is not available when operating in PG7000 high line differential pressure mode (see Section 3.9.4.2). Caution To help protect against accidental overpressure, when using automated pressure generation/control, set the upper limit (UL) of the pressure control system using [GEN], <4UL> (see Section 3.9.9.3). 3.9.9.1 <2target> PURPOSE To adjust the distance from midstroke to which the piston position must be set before the automated generation system considers the piston floating process complete (see Sections 3.9.9, 3.5). © 2011 Fluke Calibration Page 76 3. GENERAL OPERATION Note With PG7202, AutoGen pressure control using the PPCH-G pressure control does not stop when the target piston position is reached. The PPCH-G thermal pressure control unit (TPCU) is used continuously to attempt to maintain the piston at the target position. The default target piston position for a PG7202 is mid-float position (0.0 mm). OPERATION To adjust the AutoGen piston float target press [GEN], <2target>. Edit the value of the target as desired. The default is + 1 mm (0 mm for PG7202). The entry should not exceed + 2.5 mm. 3.9.9.2 <3raise> PURPOSE To turn ON and OFF a function that causes the automated generation component (see Section 3.9.9.) to raise the PG7000 piston to the top of its stroke before automated mass manipulation by an AMH automated mass handler. This can avoid a large pressure change which may occur in certain circumstances when AMH lifts the mass load off of the piston. OPERATION To turn ON and OFF the raise piston before mass loading fuction, press [GEN], <3raise>. The cursor is on the choice corresponding to the current state. Select <1no> for the piston not to be raised or <2yes> to raise the piston. The default is <1no>. 3.9.9.3 <4UL> PURPOSE To read and/or set the UPPER LIMIT (UL) of the automated pressure generation component used by AutoGen. This function is used to protect against accidential overpressure when using the PG7000 automated pressure generation function (see Section 3.9.9 and the generation component’s Operation and Maintenance Manual, UL section) OPERATION To view or set the UPPER LIMIT of the automated pressure control component used by AutoGen, press [GEN], <4UL>. If AutoGen is not currently ON, the control component UPPER LIMIT cannot be accessed and an error message is displayed. If AutoGen is ON, the current UPPER LIMIT of the automated control component is displayed and can be edited. The automated pressure control component will abort pressure generation and beep repeatedly if its UPPER LIMIT is exceeded. 3.9.9.4 <5tol> PURPOSE To read and/or set the pressure measuring tolerance of the pressure controller used by AutoGen to automate pressure control. Note This function is used only when the pressure controller is a PPC3. Page 77 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL PRINCIPLE The control tolerance function can be used to reduce the time required to set pressure and float the PG7000 piston when the pressure controller used is significantly more accurate than the default tolerance value. The tolerance can also be described as the degree of agreement between the pressure control’s pressure measurement and the pressure value set by the PG7000 when its piston is floating. In general, making the tolerance smaller reduces the time required to float the piston and making the tolerance wider increases the time. However, if the tolerance is set too small so that it does not correctly reflect the degree of agreement between the pressure controller pressure measurement and the PG7000’s floating piston, overshoot of the piston float point or inability to float the piston will result. The pressure controller tolerance determines, when applicable: a) the distance from the actual pressure target that pressure is set before the AutoGen function begins to seek movement of the piston. b) the amount of pressure overshoot that must occur before “forced rotation” is used to overcome possible friction between the piston and the cylinder. Note This function is used only when the pressure controller is a PPC3. OPERATION To view or set the pressure control tolerance, press [GEN], <5tol>. The current value of the pressure control tolerance is displayed. The value is in % of full scale of the pressure controller. If the pressure controller is AutoRanged, the AutoRange maximum pressure is used as full scale (see the pressure controller’s Operation and Maintenance Manual). The default tolerance value is 0.05% of full scale. Edit the tolerance value if desired. Do not make the pressure tolerance less than the worse case agreement between the pressure controller pressure measurement and the pressure indicated by the PG7000 when its piston is floating. 3.9.9.5 <6refloat> PURPOSE To turn ON and OFF a function that causes the automated generation component (see Section 3.9.9.) to refloat the piston to the target piston position after a new pressure or mass target is entered, even if the piston is already floating within the piston position Ready limits. When Refloat is ON, the piston is always refloated to the target piston position after a new pressure or mass target command. This gives the full stroke of the piston to drop before refloat is necessary. When Refloat is OFF, the time required to activate the pressure control component and refloat the piston is eliminated if it is not needed. This can result in very rapid pressure setting when the piston is still floating after a mass load change. OPERATION To turn ON and OFF the Refloat function, press [GEN], <6refloat>. The cursor is on the choice corresponding to the current state. Select <1no> for the piston NOT to be controlled to the target position after a pressure or mass command if the piston is already in the Ready piston position limits. Select <2yes> for the piston to always to be controlled to the raise the piston. The default is <2yes>. 3.9.9.6 <7Vol> PURPOSE To read and/or set the volume of the system to which the PG7000 system is connected. © 2011 Fluke Calibration Page 78 3. GENERAL OPERATION Note This function is used available only with PG7302 and PG7307. PRINCIPLE The piston floating routines of PG7302 or PG7307 oil operated piston gauge using a PPCH automated pressure controller are highly dependent upon the volume of the test system into which pressure is being controlled. The VOL function is used to specify the test volume so that the PG7302 or PG7307 may scale the PPCH rates properly for the volume. Operating with an improperly specified volume will cause the piston floating routine to be either very slow (specified volume to small) or to overshoot (specified volume too large). Note • If the test system volume is not known, the PPCH volume determination function may be used to measure it. This function is run directly from the PPCH front panel (see the PPCH Operation and Maintenance Manual, {INTERNAL], <1CONFIG>)., • The test volume can be read or set remotely (see Section 4.3.4, ). OPERATION To view or set the test volume, press [GEN], <7vol>. The current volume value is displayed. The value is in cubic centimeres (cc). The default volume is 30 cc. The maximum volume is 300 cc. Edit the volume value if desired. 3.9.10 [RES] PURPOSE To set the resolution with which PG7000 loads mass in response to pressure or mass commands (see Section 3.9.12). PRINCIPLE PG7000 piston-cylinders are sized such that there is a whole number, nominal relationship between mass loaded on the piston in kilograms [kg] and the pressure at which the piston will float in kilo Pascal [kPa] or Mega Pascal [MPa]. This relationship is called the pressure to mass conversion coefficient and is expressed as kPa/kg or MPa/kg. The pressure to mass conversion coefficient is marked on the cap of each piston. PG7000 mass sets are made up of masses in multiples and submultiples of the kilogram making it simple to load mass values rounded to 0.01 g, 1 g, 10 g or 0.1 kg. When using PG7000 to define pressure, the desired pressure value is entered (see Section 3.9.11) and PG7000 prompts the user with the mass value to be loaded. Due to the many variables that influence the exact pressure to mass relationship for a piston-cylinder (even though there is nominally a whole number mass to pressure relationship) the mass value to load to reach exactly the pressure requested is always an odd value. Therefore, defining the exact pressure value requested always requires loading mass with 0.01 g resolution. When it is acceptable for the pressure values defined to not be exactly the nominal pressure value of the point, operation can be simplified and mass loading errors can be reduced by loading mass with a lower level of resolution and using the pressure that the lower level resolution mass load generates. For example, on a piston-cylinder with a nominal pressure to mass relationship of 10 kPa/kg, defining a pressure of exactly 100 kPa, nominally requires loading 10 kg of mass. However, once all the influences on the measurement are taken into consideration, the actual mass to load to define exactly 100 kPa will not be 10.00000 kg, it will be a value near 10 kg such as 9.99731 kg. This value is difficult to load, as it requires relatively complex mass accounting Page 79 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL and the manipulation of very small sub-gram masses. To avoid handling a difficult, odd mass value, one might instead decide to load 10 kg and use whatever pressure results as the reference pressure. In this example, loading 10 kg rather than 9.99731 kg would result in defining 100.0269 kPa rather than 100.0000 kPa. The pressure defined is only very slightly different from the nominal value and there is no additional uncertainty if that value is used. The savings in time and reduction of possible mass loading errors are significant. The PG7000 RES function is used to cause PG7000 to automatically calculate mass loads to a whole number value starting at 0.01 g and increasing in powers of 10 to 0.1 kg. This function is very useful to make operation more convenient and less error prone when it is not imperative that the pressure defined by PG7000 be exactly the nominal pressure of the test or calibration sequence. Note The mass loading resolution of AMH automated mass handling systems is 0.1 kg. The default mass loading resolution when AMH is initialized is 0.1 kg. If resolution finer than 0.1 kg is set when AMH is active, the AMH loads the required mass value wit 0.1 kg resolution and the operator is prompted with an instruction to load the trim mass under 0.1 kg. For fully automated operation, the mass loading resolution should always be set to 0.1 kg. OPERATION To access the resolution function, press [RES]. The display is: Mass loading rsltn: 0.01 g < and > Press the [←] and [→] keys to select the desired level of resolution. [←] decreases resolution and [→] increases resolution. Press [ENTER] to set the selected resolution and return to the main run screen. The resolution range is from 0.01 g to 0.1 kg in powers of 10. Note • The RES setting has no affect in mass to pressure mode. The RES setting only affects the resolution of the mass commands that result from pressure entries in pressure to mass mode (see Section 3.9.12). • In PG7000 high line differential pressure mode (see Section 3.9.4.2), line pressures setting is not affected by the RES setting; line pressures are always set with 0.1 kg resolution. Differential pressure mass loading resolution is determined by the RES setting. 3.9.11 [ENTER/SET P] from Run Screen PURPOSE To enter and execute pressure to mass or mass to pressure commands (see Section 3.9.12). PRINCIPLE PG7000 can calculate and display the mass to be loaded to achieve an entered pressure value (pressure to mass mode), or the pressure resulting from an entered mass load (mass to pressure mode). The P OR M function is used to set pressure to mass or mass to pressure mode (see Section 3.9.12). Pressing [ENTER/SET P] from any run screen (MAIN, SYSTEM or AMBIENT) accesses the pressure or mass entry screen which allows the command value to be entered and proceeds through the sequence to set or read the pressure defined by PG7000. OPERATION To access the pressure or mass entry screen, press [ENTER/SET P] from any run screen. The sequence after [ENTER/SET P] has been pressed varies between mass to pressure and pressure to mass mode. The mode is selected by pressing [P OR M] (see Section 3.9.12). See Section 3.9.11.1 for details on [ENTER/SET P] in pressure to mass mode and Section 3.9.11.2 for mass to pressure mode. See immediately below for typical operational sequences in gauge and absolute modes. See Section 3.9.4.1, Operating in Differential Mode, for typical differential mode operational sequence. © 2011 Fluke Calibration Page 80 3. GENERAL OPERATION Typical Gauge and Absolute by ATM Mode Operational Sequence Press [MODE] and select gauge or absolute by ATM mode as desired (see Section 3.9.4). Press [P OR M] and select pressure to mass or mass to pressure mode (see Section 3.9.12). Press [ENTER/SET P] and enter a pressure or mass value. If the piston is floating and AutoRoate is ON, the braking function engages to stop piston rotation (see Section 3.9.8). Load mass as instructed (see Section 3.6). If an AMH automated mass handling system is active, the mass is loaded automatically with resolution of 0.1 kg. Use the system pressure control component to float the PG7000 piston. If the AutoGen function is ON, the automated pressure control component floats the piston automatically (see Section 3.9.9). When PG7000 indicates Ready on all Ready/Not Ready indicators (see Section 3.4), take a DUT reading at the pressure indicated on the top line of PG7000 display. Repeat Steps through for each desired pressure value. Typical Absolute by Vacuum Mode Operational Sequence (PG7601 Only) Press [MODE] and absolute by vacumm (avac) mode (see Section 3.9.4). Press [P OR M] and select pressure to mass or mass to pressure mode (see Section 3.9.12). Press [ENTER/SET P] and enter a pressure or mass value. If AutoRoate is ON, the braking function engages to stop piston rotation (see Section 3.9.8). Load mass as instructed (see Section 3.6). If an AMH automated mass handling system is active, the mass is loaded automatically to resolution of 0.1 kg. Install bell jar on PG7000, shut PG7000 vacuum vent valve, open vacuum reference shutoff valve. Wait for vacuum under bell jar to reach Ready condition (see Section 3.4.3). Use system pressure control component to float the PG7000 piston. If the AutoGen function is ON, the automated pressure control component floats the piston automatically (see Section 3.9.9). When PG7000 indicates Ready on all Ready/Not Ready indicators (see Section 3.4), take a DUT reading at absolute pressure indicated on the top line of the PG7000 display. Shut vacuum reference shutoff valve, open vacuum vent valve. Wait for pressure under bell jar to return to ambient. Remove bell jar. If an AMH automated mass handling system is active and mass loading resolution is 0.1 kg, the vacuum does not need to be broken and reestablished at each increment since the masses are moved automatically. Repeat Steps through for each desired differential pressure point. 3.9.11.1 [ENTER/SET P] in Pressure to Mass Mode PURPOSE To enter and execute a pressure to mass command in pressure to mass mode (see Section 3.9.12). OPERATION Put the PG7000 in pressure to mass operation mode (see Section 3.9.12), then press [ENTER/SET P] in any run screen. If automated rotation is on, the screen shows until piston deceleration is complete. Note Pressing [ENTER] in the run screen causes automated pressure generation to be suspended if ON (see Section 3.9.9) and AutoRotate to stop piston rotation if ON (see Section 3.9.8). When [ENTER] is pressed to confirm mass entry, automated pressure generation and/or motorized rotation resume. Page 81 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL The display is: 1. Current measurement mode (see Section 3.9.4). 2. Current pressure unit of measure (see Section 3.9.3). 3. Entry field for the target value of pressure to be set. Target pressure: 100.0000 kPa a Use the numerical keys and editing keys to enter the target pressure value desired. Press [ENTER/SET P] to process the target pressure value. If the pressure value entered cannot be executed, an explanatory error message is displayed momentarily and operation returns to the target pressure entry screen with the previous target pressure value displayed. If the target pressure value entered is valid, the value is processed and operation proceeds to the mass loading instruction screen. The display is: Load nominal mass: 10.0 kg and 3.17 g 1. Mass to load using trim masses. 2. Nominal mass to load using main and fractional masses. Note If an AMH automated mass handling system is active and mass loading resolution is set to 0.1 kg, the AMH loads the mass automatically. As it does, the steps of its operation are displayed. If resolution is set to higher than 0.1 kg, the operator is prompted to load the trim mass value. With PG7601 in absolute by vacuum measurement mode (see Section 3.9.4), the instruction is indicating that the nominal mass value should be loaded and then the bell jar should be installed and vacuum established under the bell jar. Load the nominal mass value following the protocol described in Section 3.6 and press [ENTER/SET P]. When [ENTER/SET P] is pressed confirming that the nominal mass value has been loaded, operation returns to the previous run screen with the new pressure target and mass value active. Caution To make “in tolerance” measurements, it is imperative that all mass loading instructions be executed following the protocol describe in Section 3.6. This ensures that the actual mass value resulting from a nominal mass loading command will be correct. Failure to load masses following the PG7000 mass loading protocol is likely to result in out of tolerance mass load determinations and pressure definitions. Note © 2011 Fluke Calibration • The resolution with which the pressure to mass mode mass loading instruction is given depends on the resolution set in the RES function. The RES function makes it possible to avoid loading high resolution mass values when it is not imperative that the pressure defined be exactly the nominal pressure requested (see Section 3.9.10). • The pressure to mass loading instruction is given in nominal mass while the main run screen displays the true mass loaded. For this reason, the nominal mass loading instruction and the true mass displayed in the main run screen are slightly different values. This is normal operation (see Section 3.6). Page 82 3. GENERAL OPERATION 3.9.11.2 [ENTER/SET P] in Mass to Pressure Mode PURPOSE To enter and execute a mass to pressure command in mass to pressure mode (see Section 3.9.12). OPERATION Put the PG7000 in mass to pressure operation mode (see Section 3.9.12), then press [ENTER/SET P] in any run screen. If automated rotation is on, the screen shows until piston deceleration is complete. Note Pressing [ENTER] in the run screen causes automated pressure generation to be suspended if ON (see Section 3.9.9) and AutoRotate to stop piston rotation if ON (see Section 3.9.8). When [ENTER] is pressed to confirm mass entry, automated pressure generation and/or motorized rotation resume. The display is: 1. Edit field for total trim mass currently loaded. 2. Edit field for nominal mass of main and fractional masses currently loaded. Load nominal mass: 10.0 kg and 3.17 g Use the numerical and editing keys to enter the nominal mass to be loaded on the piston following the mass loading protocol described in Section 3.6. Press [ENTER/SET P] to process the mass value. If the mass value entered cannot be executed, an explanatory error message is displayed momentarily and operation returns to the mass entry screen with the previous nominal mass value displayed. If the mass value entered is valid, the value is processed and operation proceeds to the previous run screen with the new mass value active. Caution To make “in tolerance” measurements, it is imperative that the value of mass loaded on the piston be the NOMINAL mass following the protocol describe in Section 3.6. This ensures that PG7000 will correctly determine the true mass value loaded. Failure to enter nominal mass values following the PG7000 mass loading protocol is likely to result in out of tolerance mass load determination and pressure definitions. Note • The setting of the RES function has no effect on the resolution of mass load entries in mass to pressure mode (see Section 3.9.10). • The mass to pressure mass entry is expressed in nominal mass while the MAIN run screen displays the true mass loaded. For this reason, the nominal mass loading instruction and the true mass displayed in the main run screen are slightly different values. This is normal operation (see Section 3.6). 3.9.11.3 Commands for Zero Pressure, Ending a Test Entering a value of zero as the target presure is a convenient way to end a test and vent the automated pressure control component when AutoGen is ON. Page 83 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Entering a value of zero in pressure to mass mode causes the following sequence to occur: Stop piston rotation if AutoRotate is ON. Vent pressure control component if AutoGen is ON or prompt operator to vent. A special run screen displays reflecting that the PG7000 is at rest and the mass loading bell may not be installed. The display is: 3.9.12 1. > in ready/not ready indicator positions as status of piston is unknown since mass loading bell may not be installed. 2. Unknown pressure except in absolute by atmosphere measurement mode in which the current value of atmospheric pressure is displayed. This is the value of absolute by atmospheric pressure when the system is vented. 3. Unknown/meaningless mass load. 4. Unknown/meaningless piston position. ???----- psi a --- mm --------kg [P OR M] PURPOSE To select between PG7000 pressure to mass or mass to pressure operation mode. PRINCIPLE Piston gauges are generally used either to define desired pressure set points (e.g. when applying reference pressures to a device to be calibrated) or to measure a static pressure (e.g. when performing a crossfloat intercomparison with another piston gauge). PG7000s support these two typical situations with two operating modes: pressure to mass and mass to pressure. In pressure to mass operating mode, the operator enters target pressure values and the PG7000 provides instructions of the mass to load to achieve the desired target pressure. In mass to pressure operating mode, the operator enters the mass currently loaded and the PG7000 determines the pressure resulting from the current mass load. Mass to pressure mode is also useful to determine the true mass resulting from a nominal mass load (see Section 3.6). The P OR M function is used to set the PG7000 operating mode to either pressure to mass or mass to pressure. OPERATION Note For details on pressure to mass and mass to pressure operation, see Section 3.9.11. To access the P OR M function, press [P OR M], the display is: Select entry mode: 1pressure 2mass Selecting <1pressure> activates pressure to mass mode and returns to the previous run screen. Selecting <2mass> activates mass to pressure mode and returns to the previous run screen. 3.9.13 [ ] and [ ], [←] PURPOSE To activate motorized piston rotation manually. PRINCIPLE Motorized rotation engages and disengages to rotate or stop the rotation of the PG7000 piston. © 2011 Fluke Calibration Page 84 3. GENERAL OPERATION Motorized rotation can be set to operate automatically to maintain the piston within Ready condition rotation rate limits when the piston is floating and stop rotation before changing the pressure or mass load (see Section 3.9.8). Motorized rotation can also be engaged manually to accelerate or brake piston rotation at any time under direct operator control. [ ] is used for momentary acceleration of piston rotation. [ ] followed by [←] is used to start a function that stops piston rotation. OPERATION To momentarily engage the motorized piston rotation system and accelerate the piston, press [ ] from any run screen. If PG7000 is equipped with motorized rotation, the motorized rotation system engages and remains engaged until maximum rotation rate has been achieved or the key is released. st The display is a modified version of the 1 system run screen to indicate the piston rotation is being accelerated while showing rotation rate and position: 1. Current piston rotation rate. 2. Current rate of piston vertical displacement. 3. Current piston position. 12 rpm ACCELERATING + 2.05 mm 0.1/min To start the piston braking function press and hold [ ], then press [←]. If PG7000 is equipped with motorized rotation, the motorized rotation system engages and stays engaged until the piston rotation is stopped. Once the braking function starts the keys may be released and the function will complete unless [ESCAPE] is pressed. st The display is a modified version of the 1 system run screen to indicate the piston rotation is being decelerated while showing rotation rate and position: 1. Current piston rotation rate. 2. Current rate of piston vertical displacement. 3. Current piston position. 48 rpm DECELERATING + 2.05 mm 0.1/min To interrupt the piston rotation braking routine, press [ESCAPE]. Note • Most PG7000 platforms produced before 2006 use a solenoid based rotation actuation system. The system is protected against overloading. If the duty cycle becomes excessive, motorized rotation cuts OFF and remains off for a 25 second delay. During the cutoff time, [ ] has no effect. • The motorized rotation system can be set to operate automatically based on current piston position and rotation rate. See Section 3.9.8 for information. • In PG7000 high line differential mode (see Section 3.9.4.2), pressing [ ] on the reference PG7102 engages motorized rotation on both the reference and tare PG7000s. Page 85 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 3.10 [SETUP] Menu PURPOSE To select, view and edit the SETUP files that determine the source of the values that PG7000 uses in its calculation of defined pressure and as criteria for Ready/Not Ready condition determination. PRINCIPLE PG7000 piston gauges perform continuous, real time calculations of the pressure defined by the floating piston under current conditions (see Section 7.2 for the details on the calculations applied). The calculations are used to determine the defined pressure displayed in the MAIN run screen, and to arrive at the mass to load value when a target pressure is entered in pressure to mass mode. The equations that PG7000 uses to calculate pressure include many instrument and ambient variables. PG7000 allows the source of the value used for each variable to be defined. For example, the value of barometric pressure used to calculate current air density and to add to gauge pressure to define absolute pressure in absolute by atmosphere mode could come either from PG7000’s internal sensor, an external barometer connected to PG7000’s COM2, a user entered value or be set to standard atmospheric pressure. The PG7000 SETUP function allows the source (and in some cases the value) of the variables used in the pressure equations to be specified. In order to allow various combinations of sources and/or values to be setup and recalled, SETUP files can be created, stored, edited and recalled. The SETUP files also include certain variables used to determine PG7000 Ready/Not Ready status. Table 19 identifies the variables included in the SETUP file, the factory source setting for each variable and the default value for each variable. Variable source and value selections are recorded in SETUP files. These files can be stored and recalled so that specific combinations of variable sources and/or values can be conveniently recalled. There are 20 SETUP files. File #1 is a factory SETUP file that cannot be edited. It includes the factory default variable choices and values. Note SETUP files, USER values and PG7000’s calculation capabilities can be used to calculate defined pressure in specific conditions independent of actual PG7000 operation. The SETUP function supports the following: • View SETUP files (see Section 3.10.2); • Create/edit SETUP files (see Section 3.10.3); • Select active SETUP file (see Section 3.10.1). © 2011 Fluke Calibration Page 86 3. GENERAL OPERATION Table 19. SETUP File Choices, Factory Preferred Choice and Normal Value VARIABLE SOURCE CHOICES FACTORY DEFAULT CHOICE NORMAL VALUE Atmospheric pressure 1. Internal 2. Normal 3. User 4. COM2 1. Internal 101.325 kPa (14.6959 psi) Ambient temperature 1. Internal 2. Normal 3. User 1. Internal 20 °C Ambient relative humidity 1. Internal 2. Normal 3. User 1. Internal 45 %RH Piston-cylinder temperature 1. Internal 2. Normal 3. User 1. Internal 20 °C Gravity 1. Local 2. Normal 3. User 1. Local 9.806650 m/s2 Vacuum (PG7601 only) 1. Internal 2. Normal 3. User 4. COM2 1. Internal 0 Pa Ready/Not Ready Piston position 1. User 1. User ± 2.5 mm from midfloat position (default) Ready/Not Ready Maximum vacuum reference pressure in absolute by vacuum mode (PG7601 only) 1. User 1. User 5 Pa (default) Note The factory default SETUP choice is the SETUP choice setting for all variables in a new SETUP file. SETUP file #1 is fixed to factory SETUP choices and cannot be altered. OPERATION To access the SETUP menu, press [SETUP] from the main run screen. The display is: Current SETUP: #01 1select 2view 3edit #nn in the upper right hand corner displays the number of the SETUP file that is currently active. Select <1select> to select a different active SETUP file. Select <2view> to view the variable choices and values of the active SETUP file. Select <3edit> to create or edit a SETUP file. See Sections 3.10.1 to 3.10.3 for detailed information on each SETUP function operation. 3.10.1 <1select> PURPOSE To change the active SETUP file number. Page 87 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL Caution Changing the current SETUP file will change the source and in some cases the values of the variables used by PG7000 in calculating defined pressure. SETUP files should be selected and changed only by qualified personnel. Incorrect SETUP file selection may cause out of tolerance calculations and measurements. See Section 3.10, PRINCIPLE, for information on SETUP files and their use. OPERATION To select a SETUP file number to make it the active SETUP file, press [SETUP] and select <1select>. In the following screen enter the number of the desired SETUP file and press [ENTER]. If a valid SETUP file number has been entered, operation returns to the previous run screen with the selected SETUP file number displayed and active. Note • Entering an invalid SETUP file number causes an error message to be displayed briefly and returns to the select SETUP file screen. • If the SETUP file selected specifies COM2 as the source of barometric pressure, PG7000 will attempt to read a barometer on COM2 when initializing the new SETUP file. If PG7000 is unable to read a barometer on COM2, a communications time out message is displayed briefly and operation returns to the select SETUP file screen. See Section 3.11.5.4 for information on configuring COM2 to read an external barometer. 3.10.2 <2view> PURPOSE To view the variable source choices and values of any SETUP file number. OPERATION Note • Selecting <2view> SETUP allows the variable source choices and values of the active SETUP file for viewing only. <2view> does not make the viewed SETUP file active. To select the active SETUP file, press [SETUP] and select <1select>. • See Table 19 for a listing of the source choices for each variable. To view an existing SETUP file, press [SETUP] and select <2view>. The display is: 1. Entry field for number of SETUP file to be viewed. Defaults to SETUP file currently selected. Must be a number from 2 to 10. View SETUP file: #02 Enter the desired SETUP file number and press [ENTER]. If a valid SETUP file number is entered, the display is: View #02 latmP 2ambT 3%RH 4PCT 5g 6READY 7vac Select <1atmP> to view the atmospheric pressure variable source choice and current value. The value is displayed in the current pressure units (see Section 3.9.3). Select <2ambT> to view the ambient temperature variable source choice and the current value. The value is displayed in degrees Centigrade [°C]. Select <3%RH> to view the ambient relative humidity variable source choice and the current value. The value is displayed in %RH. © 2011 Fluke Calibration Page 88 3. GENERAL OPERATION Select <4PCT> to view the piston-cylinder temperature variable source choice and the current value. The value is displayed in degrees Centigrade [°C]. Select <5g> to view the gravity variable source choice and the current value. The value is 2 displayed in meters per second squared [m/s ]. Select <6READY> to view the values of Ready/Not Ready status criteria (see Section 3.4). These include: <1Pposition>: <2vac> (PG7601 only): Piston position Ready/Not Ready limits (see Section 3.4.1). The value is displayed as a band in millimeters about mid-float position [± mm] (see Section 3.5). Maximum reference vacuum pressure limit when operating in absolute by vacuum mode. The value is displayed in Pascal [Pa]. Select <7vac> (PG7601 only): to view the reference vacuum variable source choice and the current value. The value is displayed in Pascal [Pa]. 3.10.3 <3edit> PURPOSE To edit an existing SETUP file and/or to create a new SETUP file. Note See Section 3.10, PRINCIPLE for information on SETUP files and their purpose. OPERATION Note • SETUP file #1 is the factory preferred file and it cannot be edited. • See Table 19 for a listing of the source choices for each variable. To edit an existing SETUP file or create a new SETUP file, press [SETUP] and select <3edit>. The display is: 1. Entry field for number of SETUP file to be edited or created. Defaults to SETUP file currently selected. Must be a number from 2 to 10. Edit SETUP file: #02 Enter the desired SETUP file number and press [ENTER]. If a valid SETUP file number is entered, the display is: Edit #02 latmP 2ambT 3%RH 4PCT 5g 6READY 7vac Selecting an item leads to the variable source choices menu for that item. The cursor is on the source choice that is currently selected in that SETUP file number. Select the desired variable source. The selection causes the variable choice selection to be made and returns to the edit menu. From the variable source choice menus, pressing [ENTER] or [ESCAPE] returns to the edit selection menu. This allows easy, discreet movement between variable source choices when editing a SETUP file. Select <1atmP> to specify the atmospheric pressure variable source for the pressure values that are used to calculate ambient air density and to add to gauge pressure measurements to calculate absolute pressure in absolute by addition of atmosphere mode. Selecting <1atmP> offers the atmospheric pressure variable source choices: <1internal> Use real time readings from PG7000’s on-board barometer. <2normal> Use a fixed value of 101.325 kPa (14.6959) psi. <3user> Use a fixed user entered value. If <3user> is selected, the user value must be entered, in the current pressure unit of measure (see Section 3.9.3). Page 89 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL <4COM2> Use real time readings from the device connected by RS232 to PG7000’s COM2 port. The external barometer must be properly set up to communicate with PG7000 (see Section 3.11.5.4 for information on setting up to read an external barometer). Note For differential mode operation (see Section 3.9.4) the AtmP selection must be <4COM2>. Select <2ambT> to specify the ambient temperature variable source for the temperature values that are used to calculate ambient air density. Selecting <2ambT> offers the temperature variable source choices: <1internal>: Use real time readings from PG7000’s on-board ambient temperature platinum resistance thermometer (in the Temperature - Humidity (TH) Probe on the platform rear panel). <2normal>: Use a fixed value of 20 °C. <3user>: Use a fixed user entered value. If <3user> is selected, the user value must be entered, in degrees Centigrade [°C]. Select <3%RH> to specify the relative humidity variable source for the relative humidity values that are used to calculate ambient air density. Selecting <3%RH> offers the relative humidity variable source choices: <1internal>: Use real time readings from PG7000’s on-board relative humidity sensor (in the Temperature - Humidity (TH) Probe on the platform rear panel). <2normal>: Use a fixed value of 45 %RH. <3user>: Use a fixed user entered value. If <3user> is selected, the user value must be entered, in percent relative humidity [%RH]. Select <4PCT> to specify the piston-cylinder temperature variable source for the temperature values that are used to compensate piston-cylinder effective area for temperature. Selecting <4PCT> offers the temperature variable source choices: <1internal>: Use real time readings from PG7000’s on-board piston-cylinder temperature platinum resistance thermometer (embedded in the piston-cylinder module mounting post). <2normal>: Use a fixed value of 20 °C. <3user>: Use a fixed user entered value. If <3user> is selected, the user value must be entered, in degrees Centigrade [°C]. Select <5g> to specify the value of acceleration due to gravity that is used in calculating the force applied to the piston. Pressing <5g> offers the gravity variable source choices: <1locall>: <2normal>: <3user>: Use the fixed value of local gravity stored under [SPECIAL], <6gl> (see Section 3.11.6). Use a fixed value of 9.806650 m/s2. Use a fixed user entered value different from the current local gravity value stored in the [SPECIAL], <6gl> menu option. If <3user> is selected, the user value must be entered, in meters per second squared [m/s2]. Select <6READY> to edit the values of Ready/Not Ready status criteria (see Section 3.4). These include: <1position>: <2vacl> (PG7601 only): © 2011 Fluke Calibration Piston position limits. Edit the fixed value, in millimeters [mm], of the band around mid-float position within which the piston position is Ready (see Section 3.4.1). This also determines the limit at which the piston is refloated by the GEN function when GEN is ON (see Section 3.9.9). Maximum reference vacuum pressure when operating in absolute by vacuum mode. Edit the value, in Pascal [Pa], under which the reference vacuum must be for a vacuum Ready condition to occur (see Section 3.4.3). Page 90 3. GENERAL OPERATION Select <7vac> to specify the reference vacuum variable source for the value used for the pressure under the bell jar when calculating the absolute pressure defined in absolute by vacuum mode. Pressing <7vac> offers the reference vacuum variable source choices: <1internal>: Use real time readings from PG7000’s on-board vacuum gauge (mounted directly in the vacuum plate, under the mass load). <2normal>: Use a fixed value of 0 Pa. <3user>: <4COM2>: 3.11 Use a fixed user entered value. If <3user> is selected, the user value must be entered, in Pascal [Pa]. Use real time readings from or through the device connected by RS232 to PG7601’s COM2 port. The external vacuum gauge must be properly set up to communicate with PG7601 (see Section 3.11.5.5 for information on setting up to read an external vacuum gauge). [SPECIAL] Menu PURPOSE The [SPECIAL] key accesses a menu of PG7000 functions and settings that are less commonly or not normally used in regular operation. OPERATION 1PC/MS 2presU 3head To access the SPECIAL menu, press [SPECIAL] from the main run screen. This display is: 4prefs 5remote 6gl 7cal 8AMH 9reset Note Some screens, such as the SPECIAL menu, go beyond the two lines provided by the display. This is indicated by a flashing arrow in the second line of the display. Press the [←] and [→] keys to move the cursor to access the lines that are NOT visible or directly enter the number of the hidden menu choice if you know it. Special menu choices include: <1PC/MS>: Edit and view stored piston-cylinder module and mass set metrological information. Select mass set to be used (see Section 3.11.1). <2presU>: Customize the pressure unit of measure choices available under [UNIT] (see Section 3.11.2). <3head>: Change the height unit of measure and the fluid used in DUT head corrections; adjust the barometer head height; turn the automated piston position head correction ON and OFF (PG7302 only) (see Section 3.11.3). <4prefs>: To access a menu of internal PG7000 operational preferences and functions including screen saver, sounds, time/date, instrument ID and user level protection (see Section 3.11.4). <5remote>: Set up/modify PG7000 RS232 (COM1, COM2, COM3) and IEEE-488 interfaces. Test RS232 ports. Set up external barometer communications (see Section 3.11.5). <6gl>: Set the value of local gravity used by PG7000 in reference pressure calculations when gl is specified as the gravity value in the active SETTINGS file (see Section 3.11.6). <7cal>: View the output of and adjust PG7000 internal sensors and measurement systems (see Section 3.11.7). <8AMH>: View the status of and directly control an AMH automated mass handler (see Section 3.11.8). <9reset>: Access and execute various reset options (see Section 3.11.9). Page 91 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 3.11.1 <1PC/MS> PURPOSE To perform the following piston-cylinder module and mass set metrological functions: • View and edit piston-cylinder module metrological information. • Create new piston-cylinder modules. • View and edit mass set and mass loading bell information. • Create new mass sets and mass loading bells. • Select the active mass set and mass loading bell. Note See Section 3.9.2 for information on more direct access to selecting the active piston-cylinder module. Caution The piston-cylinder module and mass set information contained in <1PC/MS> is element specific metrological data. Uninformed or accidental altering of this information may lead to out of tolerance measurements. Piston-cylinder module and mass set information should only be edited by qualified personnel as part of the PG7000 calibration or recalibration process. PRINCIPLE To calculate the pressure defined by its floating piston, PG7000 must have available the specific characteristics of the piston-cylinder module, mass set and mass loading bell that are being used. This information is entered, viewed and edited under [SPECIAL], <1PC/MS>. This is also where the active mass and mass loading bell are selected. The piston-cylinder module and mass set information required by PG7000 is reported on the Calibration Reports that are delivered with the metrological elements. When the PG7000 Platform and metrological elements are delivered together, the metrological elements information is entered by Fluke Calibration at the factory. If the metrological elements are delivered separately from the platform or after a recalibration when new data is available, the information must be entered or edited by the user. OPERATION To access the piston-cylinder module, mass set and mass loading bell information, press [SPECIAL] and select <1PC/MS>. The display is: 1PC 2mass set 3mass bell Select the type of metrological element that you would like to view, edit, create or select. See Sections 3.11.1.1 to 3.11.1.7 for operation of specific piston-cylinder module, mass set and mass bell functions as follows: • Create a new piston-cylinder module (see Section 3.11.1.1). • Edit piston-cylinder module information (see Section 3.11.1.2). • View piston-cylinder module information (see Section 3.11.1.3). • Delete a piston-cylinder module (see Section 3.11.1.4). • Select the active piston-cylinder module (see Section 3.11.1.5). • Create a new mass set (see Section 3.11.1.6). • Edit mass set information (see Section 3.11.1.7). • View mass set information (see Section 3.11.1.8). • Delete a mass set (see Section 3.11.1.9). • Select a mass set (see Section 3.11.1.10). • Add a mass loading bell (see Section 3.11.11). © 2011 Fluke Calibration Page 92 3. GENERAL OPERATION • Edit mass loading bell information (see Section 3.11.12). • View the active mass loading bell (see Section 3.11.1.13). • Delete a mass loading bell (see Section 3.11.1.14). • Select a mass loading bell (see Section 3.11.1.15). 3.11.1.1 Create Piston-Cylinder Module PURPOSE To create a new piston-cylinder module that will be available for selection from the [P-C] function key. Use this function when a new piston-cylinder module has been acquired. Note If the piston-cylinder module and PG7000 Platform were delivered together, the piston-cylinder module has already been entered at the factory. Before creating a new piston-cylinder module, press [P-C] to check if it already exists (see Section 3.9.2). PRINCIPLE The PG7000 add and edit piston-cylinder module functions allow piston-cylinder metrological module variable values to be defined and stored. These values will be used by PG7000 in calculating defined pressure, piston-cylinder range and pressure to mass and mass to pressure values when the piston-cylinder module is made active using [P-C]. The information needed can be found in the calibration report for the piston-cylinder module. Setting up or editing a piston-cylinder module requires specifying, in sequential order, the following: Serial number (S/N) [nnnn]. Effective area (Ae) at 0 pressure and 20 °C [mm2]. Piston temperature coefficient [linear thermal expansivity/°C]. Cylinder temperature coefficient [linear thermal expansivity/°C]. Effective area pressure coefficient [change in Ae/MPa]. Piston-cylinder surface tension effect [N/m]. Reference level offset [mm]. Piston assembly mass [kg]. Piston assembly apparent density [kg/m3]. Rotation rate ready limits [rpm]. k(P) [kg/mm/min]. Calibration report number [nnnnnnnnn]. Calibration report date [yyyymmdd]. OPERATION Note • Adding or editing a piston-cylinder module requires the user to specify a large number of variables in a series of data entry screens described below. For a summary of the piston-cylinder module data requirements in the order they are edited see PRINCIPLE in this section. • PG7000 can store up to 18 piston-cylinder module (PC) files. When <3add> is selected and all the files have already been used, a warning is displayed. A file must be deleted before a file can be added. Page 93 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL To create a new piston-cylinder module press [SPECIAL] , <1PC/MS> and select <1PC>, <3add>. A series of data entry menus will be presented. Edit each menu to reflect the correct information on the piston-cylinder module being added. Press [ENTER] after each entry to proceed to the next menu. Pressing [ESCAPE] at any point offers an query and then returns to the <1PC> menu without saving any of the edited data or [ENTER] returns to the add piston-cylinder module process. The data entry screens to add or edit a piston-cylinder module are, in order: © 2011 Fluke Calibration • Serial number (S/N): Edit the serial number using four digits. This value is marked on the piston cap. The default is 1. • Effective area at 0 pressure and 20 °C: Edit the value of the pistoncylinder effective area at 0 pressure and 20 °C. The value must be 2 entered in square millimeters [mm ]. This value is reported in the calibration report for the piston-cylinder module. The default is 0. • Piston temperature coefficient: Edit the value of the piston material linear thermal expansivity. The value must be entered in terms of relative change per -6 degree centigrade [nn x 10 /°C]. The E-6 exponent is fixed. This value is reported in the calibration report for the piston-cylinder module. The default is 00. • Cylinder temperature coefficient: Edit the value of the cylinder material linear thermal expansivity. The value must be entered in terms -6 -6 of relative change per degree centigrade [nn x 10 /°C]. The 10 exponent is fixed. This value is reported in the calibration report for the piston-cylinder module. The default is 00. • Effective area pressure coefficient: Edit the value of the pistoncylinder change in effective area in terms of relative change per -6 -6 MegaPascal [n.nn x 10 /MPa]. The exponent is 10 for PG7601 and -6 PG7102 and 10 for PG7302. This value is reported in the calibration report for the piston-cylinder module. The default is 0.00. • Piston-cylinder surface tension effect: Edit the value of the surface tension effect of the pressurized fluid on the piston. The value must be entered in Newton per meter [N/m]. This value is reported in the calibration report for the piston-cylinder module. The default is 0.00. The correct value for all gas lubricated and liquid lubricated, gas operated piston-cylinders (PG7102, PG7601 and PG7202) is 0 unless they are operated with oil as the test medium. • Reference level offset: Edit the value of the reference level offset which corrects the piston-cylinder reference level for hollow pistons (5 kPa/kg, 10 kPa/kg, 20 kPa/kg and 50 kPa/kg gas lubricated piston-cylinder modules only) to the reference level marked on the mounting post (see Section 3.9.7). The value must be entered in millimeters [mm]. This value is reported in the calibration report for 5 kPa/kg, 10 kPa/kg, 20 kPa/kg and 50 kPa/kg gas lubricated piston-cylinders; it is 0 for all other piston-cylinders. The default is 0.00. • Piston assembly mass: Edit the value of the mass of the piston assembly (piston + piston head + piston cap). The value must be entered in kilogram [kg]. This value is reported in the calibration report for the piston-cylinder module. The default is 0.200000. • Piston assembly average density: Edit the value of the average density of the piston assembly (piston + piston head + cap). The value 3 must be entered in kilogram per cubic meter [kg/m ]. This value is reported in the calibration report for the piston-cylinder module. The default is 0. Page 94 3. GENERAL OPERATION • Rotation rate ready limits [rpm]: The minimum Ready rotation rate (see Section 3.4.2) and the target maximum rate. The values must be entered in rotations per minute [rpm]. The values (min and max) must be entered in rotations per minute [rpm]. The value automatically defaults to the factory recommended value for the piston-cylinder size and pressurized medium. In most circumstances, the factory default values should be used. • k(P) coefficient [kg/mm/min]: Edit the value of the mass to equilibrium fall rate proportionality coefficient. This value is non-zero only for PG7102 or PC-7600 piston-cylinder modules that are used as the taring piston-cylinder in high line differential measurement mode (see Section 3.9.4.2). In all other cases, this value is zero. The value, when applicable, is specified in the piston-cylinder module technical data. • Calibration report number: Edit the calibration report number to reflect the number of the piston-cylinder module calibration report that is currently valid. The number can be up to nine digits. The default is 1. • Calibration report date: Edit the date of the current calibration report. The date must be expressed in yyyymmdd format. The default is 19800101. After pressing [ENTER] to accept the edited calibration report date the option to save the edited piston-cylinder module information is presented. The display is: Save PC S/N nnnn 1no 2yes Select <2yes> to save the piston-cylinder module information under serial number nnnn and return to the <1PC/MS> menu. Select <1no> to discard all edits and return to the <1PC/MS> menu. 3.11.1.2 Edit Piston-Cylinder Module PURPOSE To edit information contained in an existing piston-cylinder module file. Use this function to change piston-cylinder module data after a recalibration. PRINCIPLE/OPERATION Caution Piston-cylinder module information is element specific metrological data. Uninformed or accidental altering of this information may lead to out of tolerance measurements. Piston-cylinder module information should only be edited by qualified personnel as part of the PG7000 calibration or recalibration process. See Section 3.11.1.1 describing adding a new piston-cylinder module. The principles and procedures to add or edit a piston-cylinder module are identical except for the identification of the piston-cylinder module to be edited. To edit information on an existing piston-cylinder module, press [SPECIAL] , <1PC/MS> and select <1PC>, <2edit>. The display identifies the currently active piston-cylinder module. Press [ENTER] to edit the currently active piston-cylinder module or press [P-C] to toggle through the other piston-cylinder modules available. When the desired pistoncylinder module is identified, press [ENTER] to proceed with editing pistoncylinder module information. Operation of the editing function is identical to the add new piston-cylinder module procedure (see Section 3.11.1.1, OPERATION). 3.11.1.3 View Piston-Cylinder Module Page 95 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL PURPOSE To view the information contained in an existing piston-cylinder module file. PRINCIPLE/OPERATION See Section 3.11.1.1, PRINCIPLE. The piston-cylinder module viewing function allows the piston-cylinder module information to be viewed but not altered. To view information on an existing piston-cylinder module, press [SPECIAL] , <1PC/MS> and select <1PC>, <3view>. The display identifies the currently active piston-cylinder module. Press [ENTER] to view the currently active piston-cylinder module or press [P-C] to toggle through the other piston-cylinder modules available (see Section 3.9.2). When the desired piston-cylinder module is identified, press [ENTER] to proceed with viewing piston-cylinder module information. Successive pressing of [ENTER] steps through the piston-cylinder information screens in the same order as the add and edit functions (see Section 3.11.1, View Piston-Cylinder Module, Principle). 3.11.1.4 Delete Piston-Cylinder Module PURPOSE To delete a piston-cylinder module file so that it is no longer included in the piston-cylinder modules available for selection by pressing [P-C]. OPERATION Caution Once a piston-cylinder module file has been deleted it cannot be recovered. To recreate it, the add piston-cylinder module function must be used (see Section 3.11.1.1) which requires entering all the piston-cylinder module information. Before deleting a piston-cylinder module, consider editing it (see Section 3.11.1.2). To delete a piston-cylinder module press [SPECIAL] , <1PC/MS> and select <1PC>, <4delete>. The display identifies the currently active piston-cylinder module. Press [ENTER] to delete the currently active piston-cylinder module or press [P-C] to toggle through the other piston-cylinder modules available (see Section 3.9.2). When the desired piston-cylinder module is identified, press [ENTER] to proceed with deleting the piston-cylinder module information. Confirmation to delete is required. Select <2yes> to delete the piston-cylinder module. Select <1no> to NOT delete. Note The active piston-cylinder module (piston-cylinder module that is currently selected) cannot be deleted. Before attempting to delete a piston-cylinder module, be sure it is not currently selected. 3.11.1.5 Select the active piston-cylinder module PURPOSE To select the piston-cylinder module that is active and used by PG7000 in its mass loading and defined pressure calculations. Note This function serves the same purpose as pressing [P-C] (see Section 3.9.2). © 2011 Fluke Calibration Page 96 3. GENERAL OPERATION OPERATION To select the active piston-cylinder module press [SPECIAL] , <1PC/MS> and select <1PC>, <5select>. 3.11.1.6 Add Mass Set PURPOSE To create a new mass set that will be available for selection as the active mass set to load on the PG7000 piston. Use this function when a new mass set has been acquired or a mass set was deleted and must be reentered. Note If the mass set and PG7000 Platform were delivered together, the mass set has already been entered at the factory. Before creating a new mass set, check whether it already exists using the view mass set function (see Section 3.11.1.8). PRINCIPLE The PG7000 add and edit mass set functions allow mass set composition and metrological variable values to be defined and stored. These values will be used by PG7000 when the mass set is made active to calculate nominal mass instructions and true mass loads. Up to three mass sets can be created. Setting up or editing a mass set requires defining the density of the masses, the mass set composition and the true mass of each individual mass. This information can be found in the calibration report for the mass set. Mass sets are created and edited in terms of mass groups reflecting the normal composition of a PG7000 mass set. Standard PG7000 mass sets are made up of these groups (see Section 2.1.2.2 for listings of standard PG7000 mass set compositions). The mass groups of a MANUAL mass set include: 1. Make up mass (9, 4.5 or 4 kg): A single mass, sequentially numbered 1. The value of this mass is 9 kg if the mass set main masses are 10 kg and 4 kg or 4.5 kg if the mass set main masses are 5 kg. The purpose of this mass is to arrive at a whole number nominal mass load equal to the sets main masses (10 kg or 5 kg) when added to the piston and mass loading bell. 2. 10 kg masses: The main masses in a standard 80 or 100 kg mass set. Sequentially numbered from 1 up to the total number of 10 kg masses. Mass sets of < 80 kg do not normally include 10 kg masses. 3. 5 kg masses: The main masses in any standard mass set of less than 80 kg. Sequentially numbered from 1 up to the total number of 5 kg masses. Standard 80 and 100 kg mass sets have one 5 kg mass. 4. 2 kg masses: and 2. Every standard mass set has two 2 kg masses numbered 1 5. 1 kg masses: Every standard mass set has one 1 kg mass numbered 1. 6. 0.5 kg masses: Every standard mass set has one 0.5 kg mass number 1. 7. 0.2 kg masses: Every standard mass set has two 0.2 kg masses numbered 1 and 2. 8. 0.1 kg masses: Every standard mass set has one 0.1 kg mass numbered 1. 9. Trim masses: Every standard mass set has a trim mass set made up of masses from 50 to 0.01 g. These masses are NOT defined and entered as part of the mass set. The mass groups of an AMH automated mass handler mass set include: Page 97 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL 1. Main masses (10 kg with AMH-100, 6.2 kg with AMH-38): Sequentially numbered from 1 to a maximum of 9 for AMH-100 or 1 to a maximum of 5 for AMH-38. 2. 6.4 kg mass (AMH-100 only): Single tubular, binary mass. 3. 3.2 kg mass: Single tubular, binary mass. 4. 1.6 kg mass: Single tubular, binary mass. 5. 0.8 kg mass: Single tubular, binary mass. 6. 0.4 kg mass: Single tubular, binary mass. 7. 0.2 kg mass: Single tubular, binary mass. 8. 0.1 kg mass: Single tubular, binary mass. 9. Trim masses: Every standard mass set has a trim mass set made up of masses from 50 to 0.01 g. These masses are NOT defined and entered as part of the mass set. Caution When PG7000 provides mass loading instructions and calculates the true mass of the mass currently loaded, it assumes that the mass set in use has been set up correctly and that masses have been loaded following PG7000 mass loading protocol (see Section 3.6). For PG7000 mass loading protocol to operate properly, the mass set in use must be EXACTLY the mass set that has been defined by the add and/or edit mass function. No extra mass can be included, no mass used can be missing and the sequential numbers of masses in each mass group must be followed with the correct true mass value entered for each individual mass. Using a mass set that is not accurately set up may result in out of tolerance pressure definitions. Note Though the piston and the mass loading bell constitute part of the mass load, they are not included in the definition of a mass set. Also, one PG7000 system may be used with several pistoncylinder modules, mass loading bells and mass sets. Therefore, the piston mass and mass loading bell mass information are not part of the mass set. Piston mass information is defined and stored in the piston-cylinder module file (see Section 3.11.1.1) and mass loading bell information is defined and stored in a separate mass loading bell files (see Section 3.11.1.11). In an AMH mass set, the binary mass carrier and mass lifting shaft are considered part of the bell. OPERATION Caution Mass set information is element specific metrological data. Uninformed or accidental altering of this information may lead to out of tolerance measurements. Mass set information should only be edited by qualified personnel as part of the PG7000 calibration or recalibration process. © 2011 Fluke Calibration Page 98 3. GENERAL OPERATION Note • To use the create or edit mass set function with a manual mass set, it is important to have an understanding of mass set structure and mass groups. Consult Section 3.6, PRINCIPLE, to familiarize yourself with this information before attempting to create or edit a mass set. Before creating or editing an AMH automated mass handler mass set, see the AMH-38/AMH-100 Operation and Maintenance Manual. • PG7000 can store up to 3 MS (mass set) files. When <3add> is selected and all the files have already been used, a warning is displayed. A mass set must be deleted before a mass set can be added. To create a new PG7000 mass set there are three sequential steps: Initialize mass set: Define serial number, density, mass set type (manual or AMH), calibration report number and date. Define individual masses: Edit, add and delete mass groups as necessary to identify and define all masses in the mass set. Save mass set file or abandon changes. Step : Initialize Mass Set Press [SPECIAL] , <1PC/MS> and select <2mass set>, <3add>. Edit the serial number to the number of the mass set being added and press [ENTER]. The display is: Add mass set: S/N 1 Edit the density of the masses being added (all PG7000 mass sets have density of 8 000 kg/m3) and press [ENTER]. The display is: Mass density: 8000 kg/m3 Select <1manual> for a manual mass set or <2AMH> for an automated mass handler mass set. Mass set type: 1manual 2AMH Edit the calibration report number (up to nine digits) to the the number of the current calibration report of the mass set that is being added and press [ENTER]. The display is: Cal report number? 1 Edit the calibration report date to the date of the current calibration report of the mass set that is being added (format must be YYYYMMDD) and press [ENTER]. [ENTER] leads to the second step of mass set adding or editing. Cal report date? 19800101 Pressing [ESCAPE] at any point goes to an abandon edits warning screen. To continue with defining the mass set proceed to Step . Step : Define Individual Masses This step can only be reached by going through Step . Page 99 © 2011 Fluke Calibration PG7000™ OPERATION AND MAINTENANCE MANUAL If the mass set is a MANUAL mass set, the display is: Select mass grp (kg) MKUP 5.00 2.00 1.001 0.50 0.20 0.10 ADD If the mass set is an AMH automated mass handler mass set, the display is: (6.4 is only present for AMH-100) Select mass grp (kg) MAIN 6.40 3.20 1.60 0.80 0.40 0.20 0.10 Note Some screens, such as the SPECIAL menu, go beyond the two lines provided by the display. This is indicated by a flashing arrow in the second line of the display. Press the [←] and [→] keys to move the cursor to access the lines that are NOT visible or directly enter the number of the hidden menu choice if you know it. Refer to Sections 3.11.1.6, PRINCIPLE and 3.6 for information on mass group definitions. To create the mass set, each and every mass in the mass set, no more and no less, (but not the piston and the bell) must be identified and its true mass entered. To select a mass group press the [←] and [→] keys to position the cursor on the mass group to be edited. If the mass is needed but not shown, select to create a new mass group. To delete a mass group, enter <0> as the number of masses in the group. Caution All standard PG7000 MANUAL mass sets have a makeup mass (MKUP). This mass may be 4, 4.5 or 9 kg depending on the mass set (see Section 3.6, PRINCIPLE). It is imperative that the make up mass be defined and entered in the MKUP group. Operation within each mass group is identical, for example, select <5.00>. The display is: 1. Edit field for number of masses in the mass group. # of masses in group (0 deletes grp): 1 Edit the number of masses in group to reflect the number of masses of that nominal value there are in the group and press [ENTER]. The display is: 1. Sequential ID number of this specific mass within the mass group. 2. Edit field for the true value of this specific mass. 5.0 kg mass #1 True mass: 5.00000 Edit the mass value to the true value of the specific mass identified and press [ENTER]. Note For high line differential mode, when setting up the tare mass set of the tare PG7000, the true mass of each mass is entered as equal to the nominal value. Enter the nominal mass values for the true mass values. © 2011 Fluke Calibration Page 100 3. GENERAL OPERATION If there are additional masses in this mass group, the next display is the same as the display immediately above but with the next mass specific mass sequential ID number in the mass group. The screens continue until the true value of all of the masses within the group has been entered. After the last entry, the screen returns to the
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : No Author : Michael D. Foster Comments : Company : DHI Create Date : 2011:03:28 07:36:03-07:00 Modify Date : 2012:02:24 14:50:22-07:00 Source Modified : D:20110328143455 Subject : Tagged PDF : Yes XMP Toolkit : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:08:04 Metadata Date : 2012:02:24 14:50:22-07:00 Creator Tool : Acrobat PDFMaker 9.1 for Word Document ID : uuid:0aba8ab7-5cd9-4237-9234-a0eb0b2ae558 Instance ID : uuid:28c7cf5d-dfb4-4b7f-a229-f0b8176a3b16 Format : application/pdf Title : 1.2 SPECIFICATIONS Description : Creator : Michael D. Foster Producer : Adobe PDF Library 9.0 Keywords : Page Layout : OneColumn Page Count : 214EXIF Metadata provided by EXIF.tools