Form_Aqua Lab IQ OQ Form Aqua

User Manual: Form_AquaLab IQ-OQ Form

Open the PDF directly: View PDF PDF.
Page Count: 2

DownloadForm_Aqua Lab IQ-OQ Form Aqua
Open PDF In BrowserView PDF
AQUALAB – WATER ACTIVITY METER
INSTALLATION / OPERATIONAL QUALIFICATIONS
AquaLab Information
Manufacturer:
Model Number:
Serial Number:
Company:
Department:
Room Number:
Contact:
Calibration SOP #:
Calibration Frequency:

Decagon Devices, Inc.

Decagon (800) 755-2751
Daily/Before each use

Installation Qualification
Installation of AquaLab includes placing the instrument on a level surface in a location where the
temperature remains fairly stable. This location should be well away from air conditioner and heater
vents, open windows, outside doors, or other items that may cause rapid temperature fluctuation.
After finding a good location for your AquaLab, plug the power cord to the back of the unit and a
standard AC outlet. The ON/OFF switch is located on the lower right corner of the AquaLab’s back
panel.



AquaLab properly installed at facility

Qualified By:

Date Qualified:
(Vendor Representative/Analyst Name)

Operational Qualification
AquaLab’s operation and performance is verified by measuring specially prepared calibration
standards that have a specific molality and water activity. In order to provide the most accurate
readings, your AquaLab should ideally be allowed a warm-up period of at least 15 minutes after
turning it on. Performance Verification Standards come in four water activity levels: 0.250, 0.500,
0.760, and 1.000 aw. The AquaLab dew point sensor will read each standard within ±0.003aw of
the stated value. The volatiles sensor will read each standard within ±0.015aw of the stated value.
Standard @ 25°C
13.41m LiCl
0.250
0.250
8.57m LiCl
0.500
0.500
6.0m NaCl
0.760
0.760
Steam Distilled H2O
1.000
1.000

Dewpoint Sensor
°C
aw ±0.003

Volatiles Sensor (if applicable)
°C
aw ±0.015

AquaLab measured Verification Standards within specification:
 YES
 YES  NO
Qualifications met Vendor Acceptance Criteria:

13939-01 AquaLab IQ/OQ Form

Page 1 of 2

 NO

AQUALAB – WATER ACTIVITY METER
INSTALLATION / OPERATIONAL QUALIFICATIONS

(Complete the following if “NO” has been checked, otherwise enter N/A)

Explanation of why Acceptance Criteria not met:

Action Plan to meet Acceptance Criteria:

(Check after completion of Action Plan)

Qualification meets Vendor’s Acceptance Criteria after completion of Action Plan:  YES
Comments:

 NO

Training on the proper use, calibration and maintenance of AquaLab was provided.

(Signature/Date)

Analyst:
(Print Name)

13939-01 AquaLab IQ/OQ Form

(Sign Name)

Page 2 of 2

(Date)



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Author                          : 
Create Date                     : 2011:03:02 09:13:49-08:00
Modify Date                     : 2016:06:08 15:28:26-07:00
XMP Toolkit                     : Adobe XMP Core 5.6-c015 84.158975, 2016/02/13-02:40:29
Creator Tool                    : PScript5.dll Version 5.2.2
Metadata Date                   : 2016:06:08 15:28:26-07:00
Producer                        : Acrobat Distiller 9.4.2 (Windows)
Format                          : application/pdf
Creator                         : 
Title                           : 
Document ID                     : uuid:690c736c-c379-46c2-ba85-c2f4d0d4a4e0
Instance ID                     : uuid:fcdeb9b2-f234-e84c-b278-e5361aa0c46a
Page Count                      : 2
EXIF Metadata provided by EXIF.tools

Navigation menu