Form_Aqua Lab IQ OQ Form Aqua
User Manual: Form_AquaLab IQ-OQ Form
Open the PDF directly: View PDF .
Page Count: 2
Download | |
Open PDF In Browser | View PDF |
AQUALAB – WATER ACTIVITY METER INSTALLATION / OPERATIONAL QUALIFICATIONS AquaLab Information Manufacturer: Model Number: Serial Number: Company: Department: Room Number: Contact: Calibration SOP #: Calibration Frequency: Decagon Devices, Inc. Decagon (800) 755-2751 Daily/Before each use Installation Qualification Installation of AquaLab includes placing the instrument on a level surface in a location where the temperature remains fairly stable. This location should be well away from air conditioner and heater vents, open windows, outside doors, or other items that may cause rapid temperature fluctuation. After finding a good location for your AquaLab, plug the power cord to the back of the unit and a standard AC outlet. The ON/OFF switch is located on the lower right corner of the AquaLab’s back panel. AquaLab properly installed at facility Qualified By: Date Qualified: (Vendor Representative/Analyst Name) Operational Qualification AquaLab’s operation and performance is verified by measuring specially prepared calibration standards that have a specific molality and water activity. In order to provide the most accurate readings, your AquaLab should ideally be allowed a warm-up period of at least 15 minutes after turning it on. Performance Verification Standards come in four water activity levels: 0.250, 0.500, 0.760, and 1.000 aw. The AquaLab dew point sensor will read each standard within ±0.003aw of the stated value. The volatiles sensor will read each standard within ±0.015aw of the stated value. Standard @ 25°C 13.41m LiCl 0.250 0.250 8.57m LiCl 0.500 0.500 6.0m NaCl 0.760 0.760 Steam Distilled H2O 1.000 1.000 Dewpoint Sensor °C aw ±0.003 Volatiles Sensor (if applicable) °C aw ±0.015 AquaLab measured Verification Standards within specification: YES YES NO Qualifications met Vendor Acceptance Criteria: 13939-01 AquaLab IQ/OQ Form Page 1 of 2 NO AQUALAB – WATER ACTIVITY METER INSTALLATION / OPERATIONAL QUALIFICATIONS (Complete the following if “NO” has been checked, otherwise enter N/A) Explanation of why Acceptance Criteria not met: Action Plan to meet Acceptance Criteria: (Check after completion of Action Plan) Qualification meets Vendor’s Acceptance Criteria after completion of Action Plan: YES Comments: NO Training on the proper use, calibration and maintenance of AquaLab was provided. (Signature/Date) Analyst: (Print Name) 13939-01 AquaLab IQ/OQ Form (Sign Name) Page 2 of 2 (Date)
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.5 Linearized : No Author : Create Date : 2011:03:02 09:13:49-08:00 Modify Date : 2016:06:08 15:28:26-07:00 XMP Toolkit : Adobe XMP Core 5.6-c015 84.158975, 2016/02/13-02:40:29 Creator Tool : PScript5.dll Version 5.2.2 Metadata Date : 2016:06:08 15:28:26-07:00 Producer : Acrobat Distiller 9.4.2 (Windows) Format : application/pdf Creator : Title : Document ID : uuid:690c736c-c379-46c2-ba85-c2f4d0d4a4e0 Instance ID : uuid:fcdeb9b2-f234-e84c-b278-e5361aa0c46a Page Count : 2EXIF Metadata provided by EXIF.tools