AQUALAB – WATER ACTIVITY METER
INSTALLATION / OPERATIONAL QUALIFICATIONS
13939-01 AquaLab IQ/OQ Form Page 1 of 2
AquaLab Information
Manufacturer: Decagon Devices, Inc.
Model Number:
Serial Number:
Company:
Department:
Room Number:
Contact: Decagon (800) 755-2751
Calibration SOP #:
Calibration Frequency: Daily/Before each use
Installation Qualification
Installation of AquaLab includes placing the instrument on a level surface in a location where the
temperature remains fairly stable. This location should be well away from air conditioner and heater
vents, open windows, outside doors, or other items that may cause rapid temperature fluctuation.
After finding a good location for your AquaLab, plug the power cord to the back of the unit and a
standard AC outlet. The ON/OFF switch is located on the lower right corner of the AquaLab’s back
panel.
AquaLab properly installed at facility
Qualified By: Date Qualified:
(Vendor Representative/Analyst Name)
Operational Qualification
AquaLab’s operation and performance is verified by measuring specially prepared calibration
standards that have a specific molality and water activity. In order to provide the most accurate
readings, your AquaLab should ideally be allowed a warm-up period of at least 15 minutes after
turning it on. Performance Verification Standards come in four water activity levels: 0.250, 0.500,
0.760, and 1.000 aw. The AquaLab dew point sensor will read each standard within ±0.003aw of
the stated value. The volatiles sensor will read each standard within ±0.015aw of the stated value.
Dewpoint Sensor Volatiles Sensor (if applicable)
Standard @ 25°C aw ±0.003 °C aw ±0.015 °C
13.41m LiCl 0.250
0.250
8.57m LiCl 0.500
0.500
6.0m NaCl 0.760
0.760
Steam Distilled H2O 1.000
1.000
AquaLab measured Verification Standards within specification: YES NO
Qualifications met Vendor Acceptance Criteria: YES NO
AQUALAB – WATER ACTIVITY METER
INSTALLATION / OPERATIONAL QUALIFICATIONS
13939-01 AquaLab IQ/OQ Form Page 2 of 2
(Complete the following if “NO” has been checked, otherwise enter N/A)
Explanation of why Acceptance Criteria not met:
Action Plan to meet Acceptance Criteria:
(Check after completion of Action Plan)
Qualification meets Vendor’s Acceptance Criteria after completion of Action Plan: YES NO
Comments: Training on the proper use, calibration and maintenance of AquaLab was provided.
(Signature/Date)
Analyst:
(Print Name) (Sign Name) (Date)