User’s Guide

s & Assemblies

®

http://www.omega.com
e-mail: info@omega.com

Only model 871A has
CE certification

871A and 872A
Digital Thermometers

omega.com

TM

OMEGA®

OMEGAnet SM On-Line Service
http://www.omega.com

Internet e-mail
info@omega.com

Servicing North America:
USA:

ISO 9001 Certified

Canada:

One Omega Drive, Box 4047
Stamford, CT 06907-0047
Tel: (203) 359-1660
e-mail: info@omega.com
976 Bergar
Laval (Quebec) H7L 5A1
Tel: (514) 856-6928
e-mail: canada@omega.com

FAX: (203) 359-7700

FAX: (514) 856-6886

For immediate technical or application assistance:
USA and Canada:
Mexico and
Latin America:

Sales Service: 1-800-826-6342 / 1-800-TC-OMEGASM
Customer Service: 1-800-622-2378 / 1-800-622-BESTSM
Engineering Service: 1-800-872-9436 / 1-800-USA-WHENSM
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Tel: (95) 800-TC-OMEGASM
En Español: (203) 359-1660 ext: 2203

FAX: (95) 203-359-7807
e-mail: espanol@omega.com

Servicing Europe:

Benelux:

Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 20 6418405
FAX: (31) 20 6434643
Toll Free in Benelux: 06 0993344
e-mail: nl@omega.com

Czech Republic:

Ostravska 767, 733 01 Karvina
Tel: 42 (69) 6311899
e-mail: czech@omega.com

France:

9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400
Toll Free in France: 0800-4-06342
e-mail: france@omega.com

Germany/Austria:

Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49 (07056) 3017
FAX: 49 (07056) 8540
Toll Free in Germany: 0130 11 21 66
e-mail: germany@omega.com

United Kingdom:

25 Swannington Road,
P.O. Box 7, Omega Drive,
Broughton Astley, Leicestershire,
Irlam, Manchester,
LE9 6TU, England
M44 5EX, England
Tel: 44 (1455) 285520
Tel: 44 (161) 777-6611
FAX: 44 (1455) 283912
FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: uk@omega.com

ISO 9002 Certified

FAX: 42 (69) 6311114

FAX: (33) 130-699-120

It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is
constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the
CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.

TABLE OF CONTENTS
MODEL 871A AND MODEL 872A
DIGITAL THERMOMETERS
SECTION

PAGE

SECTION 1 INTRODUCTION .................................................................. 1
1.1
1.2
2.1

General Description ................................................................... 1
Features ..................................................................................... 1
Accessory Items......................................................................... 1

SECTION 2 INSTALLATION .................................................................... 2
2.1
2.2

Unpacking.................................................................................. 2
Battery Installation ..................................................................... 2

SECTION 3 OPERATION ......................................................................... 3
3.1
3.2

Controls and Indicators.............................................................. 3
Operating Procedure.................................................................. 5

SECTION 4 SERVICE INFORMATION .................................................... 5
4.1
4.1.1
4.1.2
4.2
4.2.1
4.2.2
4.2.3
4.2.4
4.3
4.3.1
4.3.2
4.3.3
4.3.4
4.4
4.5
4.5.1
4.5.2
4.5.3
4.5.4

Verification Check ...................................................................... 5
Functionality Check ................................................................... 5
Accuracy Check......................................................................... 6
Calibration for Model 871A ........................................................ 6
Equipment Needed .................................................................... 6
Environmental Conditions.......................................................... 6
Setup.......................................................................................... 7
Calibration Procedure ................................................................ 7
Calibration for Model 872A ........................................................ 8
Equipment Needed .................................................................... 8
Environmental Conditions.......................................................... 8
Setup.......................................................................................... 8
Calibration Procedure ................................................................ 9
Disassembly for Model 871A and Model 872A.........................10
Troubleshooting Guide .............................................................11
DC Voltage checks Model 871A ...............................................12
DC Voltage Checks Model 872A ..............................................12
Waveform Checks Model 871A ................................................13
Waveform Checks Model 872A ................................................13

SECTION 5
5.1

SPECIFICATIONS FOR MODEL 871A
AND MODEL 872A .............................................................13

Parts List ...................................................................................16
i

SECTION
1.1

1

INTRODUCTION

GENERAL DESCRIPTION
The OMEGA® Model 871A and Model 872A Digital Thermometers are
ultrasensitive hand held instruments. Low power LSI devices combined
with a large 0.6" (15 mm) LCD, deliver reliability, extended battery life, and
an extremely readable readout. One nine volt battery will deliver 100 hours
of continuous operation.
The 871A and 872A are operationally identical except for minor differences
related to each model’s thermocouple input: Model 871A is designed for
use with type K (Chromel-Alumel) thermocouples; Model 872A is designed
for use with type J (Iron-Constantan) thermocouples.
Cold junction electronic circuitry in the Model 871A and Model 872A
automatically compensates for ambient temperature changes. A special
integrating A/D converter corrects for thermocouple non-linearity during
each digitization cycle. The resulting conformity to the standard
thermocouple voltage table is within 1ºC over the entire measurement span of
the instrument.
Although the 871A and 872A are ideal for temperature measurement in the
field, they also function well as bench instruments when equipped with the
optional tilt stand. The Model 871A and Model 872A provide dual inputs,
selectable scales (ºF/ºC), 1.0º and 0.1º resolution, and an analog output.
The versatile dual input configuration permits two point temperature
measurement such as those required in process monitoring. The analog
output allows interfacing with chart recorders or data acquisition system’s
functioning as an electronic ice-point reference.

1.2

FEATURES
• Wide temperature range
• Selectable scales (ºF/ºC)
• 0.1º and 1.0º resolution
• 0.25% accuracy

1.2.1

• Dual thermocouple inputs
• Analog output
• Heavy duty construction

ACCESSORY ITEMS
SC-GG-[*]-30-36-SMP-M
KS-871
SC-800
880-A2
U9VL
SMP-[*]-M
SMP-[*]-F
OS-201-1

Thermocouple sensor (supplied with
unit) if purchased separately.
Tilt stand
Soft carrying case
Deluxe carrying case
9 V lithium battery
Subminiature thermocouple connector, male
Subminiature thermocouple connector,
female
2 oz. jar of high thermal conductivity paste
for fast surface measurements

*Insert thermocouple type: J for Model 872, K for Model 871

1

SECTION
2.1

2

INSTALLATION

UNPACKING
Remove the Packing List and verify that all equipment has been received. If there
are any questions about the shipment, please call OMEGA Customer Service
Department at 800-622-2378 or (203) 359-1660.
The following items are included:
Quantity
1
2
1
1
1

Description
871 or 872 Digital Thermometer
Beaded Wire Thermocouples
TAS Transition Adapter
9 Volt Alkaline Battery
Operator’s Manual

Upon receipt of shipment, inspect the container and equipment for any sign of
damage. Take particular note of any evidence of rough handling in transit.
Immediately report any damage to the shipping agent.
NOTE
The carrier will not honor any claims unless all shipping material is
saved for their examination. After examining and removing contents,
save packing material and carton in the event reshipment is necessary.
2.2

BATTERY INSTALLATION

WARNING
Turn the model 871/872A off and disconnect the input probes before
replacing the battery. Put the cover back into place on the battery compartment before resuming use of the instrument.
A 9 V battery is supplied with the instrument but not installed. To install the
battery, remove the cover from the battery compartment by sliding it off in the
direction of the arrow located on the battery cover. The battery connector snaps
on and off the terminals of the battery. Improper installation of the battery will
cause the connecting wires to be severed by excess strain. Proper installation
requires that the battery be positioned in such a manner (see Figure 2-1) that the
leads protruding from the boot of the battery connector face toward the outside
of the battery compartment. If the instrument is going to be stored for a long
period of time or in a high temperature environment, remove the battery to
prevent damage due to leakage.

2

Figure 2-1. Battery Installation

SECTION

3

OPERATION

The Model 871A is designed to work with any type K thermocouple probe or sensor, fitted with an OMEGA miniature thermocouple connector SMP-K-M or industry
equivalent.
The Model 872A is designed to work with any type J thermocouple probe
or sensor, fitted with an OMEGA miniature connector SMP-J-M or
industry equivalent.
3.1

CONTROL AND INDICATORS
Figure 3-1 and Table 3-1 illustrate and describe the controls and indicators on the
871A and 872A.

Figure 3-1. Model 871A Controls and Indicators
*Model 872A has same controls and indicators, only range selections differ.
3

TABLE 3-1
MODEL 871A AND MODEL 872A
CONTROLS AND INDICATORS

ITEM

CONTROL/INDICATOR

FUNCTION

1.

ON/OFF switch

Turns unit on or off.

2.

Scale and range selector

Allows user to select: ºF or ºC scale,
any of four possible thermocouple
ranges, and 1.0º or 0.1º resolution.

3.

Thermocouple selector

Allows user to read either of two
thermocouples by simply turning
the dial.

4.

Thermocouple input jacks

Allows direct connection of two
thermocouples to the unit via an
OMEGA miniature male connector.

5.

Analog output jack,
+mV OUT and -mV OUT

The analog output is a nonlinearized,
cold junction compensated, amplified
thermocouple output. (Not isolated
from selected TC input.) Use a
recorder to monitor temperature or
a sensitive DMM to make higher
resolution reading. Refer to mV
output specifications for mV output
to temperature ratios.

6.

Low battery indicator

When the battery level is low, BAT
will appear in upper left corner of
the display. Battery should then be
replaced to avoid error erroneous
readings due to battery failure.

7.

Overrange/open
thermocouple indication

When an overrange indication or
open thermocouple condition exits,
a 1 followed by blanked digits will
appear in the display.

4

3.2

OPERATING PROCEDURE
1.
2.
3.
4.
5.

Turn power on.
Select scale and range (see Tables 3-2 and 3-3).
Select thermocouple TC1 or TC2.
Connect one or two thermocouples to the input jacks.
Connect a recorder to the analog output to monitor temperatures,
or connect DMM to make high resolution readings, if desired.
TABLE 3-2
MODEL 871A TEMPERATURE RANGE

Range

Span

200°F
2000°F
200°C
1370°C

-40.0°
-40°
-40.0°
-40°

to 199.9°F
to 1999°F
to 199.9°C
to 1370°C

Resolution

0.1°F
1°F
0.1°C
1°C

TABLE 3-3
MODEL 872A TEMPERATURE RANGE

Range

SECTION

Span

200°F
1400°F
200°C
760°C

-40.0°
-40°
-40.0°
-40°

4

SERVICE INFORMATION

4.1

VERIFICATION CHECK

4.1.1

Functionality Check

to 199.9°F
to 1400°F
to 199.9°C
to 760°C

Resolution

0.1°F
1°F
0.1°C
1°C

Attach the GG-TC-K-24-36-SMP beaded wire thermocouple probe
(supplied) to the Model 871A and the GG-TC-J-24-36-SMP beaded wire
thermocouple probe (supplied) to the Model 872A. Hold the sensor end
between two fingers. The reading should change from ambient to between
77º to 104º F (25º to 40ºC).

5

4.1.2

Accuracy Check
1. Prepare a pure-water ice bath (see paragraph 4.2.3, step 5).
2. Connect a thermocouple probe to input 1 of the unit.
3. Immerse the probe into the pure-water ice bath and allow ten
minutes for thermal stabilization.
4. Turn on the instrument and use Tables 4-1 and 4-2 to verify
that the readings on each range are within specification.
TABLE 4-1
MODEL 871A RANGE CHART
Range

Allowable Reading

2000°F
200°F
1370°C
200°C

32°
32.0°
0°
0.0°

±
±
±
±

2°
1.8°
1°
1.0°

TABLE 4-2
MODEL 872A RANGE CHART
Range

Allowable Reading

1400°F
200°F
760°C
200°C

32°
32.0°
0°
0.0°

4.2

CALIBRATION FOR MODEL 871A

4.2.1

Equipment Needed

±
±
±
±

2°
1.8°
1°
1.0°

1. DC Voltage Calibrator with 1 ±0.01% of setting ± 20 µV accuracy.
2. 100: 1 divider: 1.98 kilohm and 20 ohm wirewound resistors with
0.02% ratio accuracy or better.
3. OMEGA TRC III Ice Point Reference Chamber or distilled water ice
bath.
4. OMEGA type K TRP Reference Probe or make a thermocouple transition junction (see paragraph 4.2.3).
4.2.2

Environmental Conditions
To meet all specifications, the conditions for calibration should be an
ambient temperature 23º ± 1ºC; relative humidity less than 80%.
6

4.2.3

Setup
(Refer to Figure 4-1.)
1. To make a thermocouple transition junction, silver solder a piece of
copper wire to each (type K) thermocouple wire and insulate the
junction with a waterproof jacket.
2. Connect the thermocouple wires from the type K transition junction to
an OMEGA miniature thermocouple connector SMP-K-M.
3. Plug in the miniature connector to input TC1 of the Model 871A.
4. Connect the copper wires to the 100:1 divider (see Figure 4-1) with the
copper/wire (positive) junction connected through the 1.98 kilohm
resistor, to the dc calibrator positive terminal.
5. Prepare a distilled water ice bath as follows:
a. Drill a hole in the cap of a Dewar flask or ThermosTM to accommodate
the transition junction.
b. Firmly pack the flask with pea-size ice chips made from distilled
water and then fill the flask with distilled water.*
c. Replace the melted ice with more ice while removing excess water.
d. Place cap on the flask and insert the transition junction into the
flask. Allow 20 minutes for temperature stabilization.
*Using tap water may cause inaccuracies due to contaminants.
*The 20Ω resistor must be held at a uniform
temperature during calibration.

Figure 4-1. Calibration Setup for Model 871A

4.2.4

Calibration Procedure
It is not necessary to disassemble the case for calibration. All adjustments
are accessible by removing the covers shown in Figure 4-2.
1. Remove the battery and trimmer compartment covers from the Model 871A.
2. Turn the Model 871A on and perform the calibration adjustments listed
in Table 4-3.
7

Figure 4-2. Access to Calibration Adjustments
TABLE 4-3
CALIBRATION ADJUSTMENTS FOR MODEL 871A

Step
1
2
3
4
5

Adjustment
Input Offset Null
+ Gain (X1)
+ Gain (X10)
- Gain
ºF Offset

871
Range
220ºC
1370ºC
200ºC
200ºC
200ºF

4.3

CALIBRATION FOR MODEL 872A

4.3.1

Equipment Needed

Calibrator
Setting (V)
0.0000
5.3039
0.6939
-0.1527
0.000

Trimmer
Adjust
R109
R104
R105
R103
R102

Desired
Reading
00.0ºC
1320ºC
170.0ºC
-40.0ºC
32.0ºF

1. DC voltage calibrator with ±0.01% of setting and ± 20 µV accuracy.
2. 100: 1 divider: 1.98 kilohm and 20 ohm wirewound resistors with 0.02%
ratio accuracy or better.
3. OMEGA TRC III Ice Point Reference Chamber or distilled water ice bath.
4. OMEGA type J TRP Reference Probe or make a thermocouple
transition junction (see paragraph 4.3.3).
4.3.2

Environmental Conditions
To meet all specifications, the conditions for calibration should be: an
ambient temperature of 23º ± 1ºC; relative humidity less than 80%.

4.3.3

Setup
1. To make a thermocouple transition junction, silver solder a piece of
copper wire to each (type J) thermocouple wire and insulate the
junction with a waterproof jacket.
2. Connect the thermocouple wires from the type J transition junction to
an OMEGA miniature connector SMP-J-M.

8

4.5.1

3. Plug in the miniature connector to input TC1 of the Model 872A.

1. Battery: VBAT > 7 V
2. Power Supplies:
V + to
(Analog Ground) = +2.8 ± 0.4 V
V + to
(Digital Ground) = +5 ± 1 V
3. Reference Diode: CR101 to
(Analog Ground) = -1.23 ± 0.03 V
4. Negative Reference Divider: (Reference to
Analog Ground)

4. Connect the copper wires to the 100:1 divider (see Figure 4-3) with the
iron wire connected through the 1.98 kilohm resistor, to the dc calibrator positive terminal.

➱➡

DC Voltage Checks For Model 871A

➱

5. Prepare a distilled water bath as outlined in paragraph 4.2.3, step 5.

➱

*The 20Ω resistor must be held at a uniform

R119
Pin

mV*

Pin

mV*

Pin

mV*

2
3
4

-210
-209
-204

5
6
7

-204
-199
-198

8
9
10

-195
-186
-173

*The setting of R104 and the tolerances of the resistors within the network
(R119) can affect these levels by ± 5%.

➱➱

5. Deintegrate Comparators:
U103A PIN 3 to
= 150 ±25mV
U103B, PIN 6 to
= -150 ± 25 mV
6. Cold Junction Voltage:
(Ambient temperature 25º ±3ºC)
-COM to
= 1.0 ±0.6 mV

temperature during calibration.

➱

4.5.2

Figure 4-3. Calibration Setup for Model 872A

DC Voltage Checks For Model 872A

➱➡

1. Battery: VBAT > 7 V
2. Power Supplies:
V + to
(Analog Ground) = +2.8 ± 0.4 V
V + to
(Digital Ground) = +5 + 1 V
3. Reference Diode: CR101 to
(Analog Ground)= -1.23 ±0.03 V
4. Negative Reference Divider: (Reference to
Analog Ground)

➱

➱

R119

4.3.4

Calibration Procedure
It is not necessary to disassemble the case for calibration. All adjustments
are accessible by removing the covers shown in Figure 4.2.
1. Remove the battery and trimmer compartment covers from the Model
872A.
2. Turn the Model 872A on and perform the calibration adjustments listed
in Table 4-4.

Pin

mV*

Pin

mV*

Pin

mV*

TABLE 4-4

2
3
4

-312
-293
-277

5
6
7

-274
-272
-271

8
9
10

-267
-254
-242

CALIBRATION ADJUSTMENTS FOR MODEL 872A

*The setting of R104 and the tolerances of the resistors within the network
(R119) can affect these levels by ±5%.

➱➱

5. Deintegrate Comparators:
U103, Pin 3 to
=190 ±30 mV
U103B, Pin 6 to
= -180 ±30 mV
6. Cold Junction Voltage:
(Ambient temperature 25º ±3ºC)
-COM to
= 1.3 mV ±0.6
12

Step
1
2
3
4
5

Adjustment
Input Offset Null
+ Gain (X1)
+ Gain (X10)
- Gain
ºF Offset

872A
Range
200ºC
760ºC
200ºC
200ºC
200ºF

➱

9

Calibrator
Setting (V)
0.0000
4.2922
1.02217
-.19604
0.000

Trimmer
Adjust
R109
R104
R105
R103
R102

Desired
Reading
00.0ºC
760ºC
190.0ºC
-40.0ºC
32.0ºF

4.4

DISASSEMBLY FOR MODEL 871A AND MODEL 872A
The PC board and LCD assembly are not secured once the case retaining
screws are removed. Be careful not to allow the PC board and LCD
assembly to fall out or shift out of position.
1. Place the unit face down on a bench or other similar surface and
remove the battery compartment cover. Remove and disconnect the
battery. Remove the two #4-40 x 7/8 retaining screws.
2. Grasp the bottom cover at the input jack end and with a lifting
and forward pushing motion (see Figure 4-4 and Figure 4-5), carefully
remove the bottom cover. While removing the cover, feed the
battery connector through the access hole in bottom of the
battery compartment.
3. The component side of the PC board is now exposed and the battery
can be reconnected for troubleshooting. To read the display, some light
downward pressure at the top end of the circuit board may be required
in order to make contact through the conductive elastomer strip,
between the circuit board and the LCD.
4. To remove the PC board from the top cover, grasp the TC switch
assembly and lift until the input jacks become disengaged from the
cover. The PC board can now be removed using a slight clockwise
motion to free the two switch knobs from their normal positions in
the case.
5. The LCD assembly will remain in the top cover when the PC board is
removed. Again, be careful not to allow the LCD assembly to fall
out accidentally.
6. The two switch knobs can be removed from the PC board assembly by
simply pulling them off the switch shafts.
7. The LCD assembly, along with the strip connector, lifts out of the case.
8. To reassemble the unit, reverse the above procedure.

NOTE
Proper alignment of the strip connector is necessary when reinstalling the
LCD assembly. Make sure that the conductor side of the strip connector is
positioned against the plastic support of the assembly.

WARNING
Be sure to replace the on-off switch cover. Common-mode
voltage will be present on the switch, creating a possible hazard if
the cover is not replaced

10

Figure 4-4. Removing the Cover

Figure 4-5. Disassembled Model 871A and Model 872A

4.5

TROUBLESHOOTING GUIDE
The troubleshooting information is intended to be used by qualified electronic maintenance personnel who are familiar with the proper use of standard electronic test equipment.
To gain access to the PC board, refer to the first three paragraphs of the
disassembly instructions. Utilize the Parts List, Schematic and Component
Layout for identifying parts and checkpoint locations. The following checks
should be made with the Models 871A and 872A set to the 200ºC Range.

11

4.5.1

DC Voltage Checks For Model 871A
1. Battery: VBAT > 7 V
2. Power Supplies:
V + to È (Analog Ground) = +2.8 ± 0.4 V
V + to È (Digital Ground) = +5 ± 1 V
3. Reference Diode: CR101 to È (Analog Ground) = -1.23 ± 0.03 V
4. Negative Reference Divider: (Reference to È Analog Ground)
R119
Pin

mV*

Pin

mV*

Pin

mV*

2
3
4

-210
-209
-204

5
6
7

-204
-199
-198

8
9
10

-195
-186
-173

*The setting of R104 and the tolerances of the resistors within the network
(R119) can affect these levels by ± 5%.
5. Deintegrate Comparators:
U103A PIN 3 to È = 150 ±25mV
U103B, PIN 6 to È = -150 ± 25 mV
6. Cold Junction Voltage:
(Ambient temperature 25º ±3ºC)
-COM to È = 1.0 ±0.6 mV
4.5.2

DC Voltage Checks For Model 872A
1. Battery: VBAT > 7 V
2. Power Supplies:
V + to È (Analog Ground) = +2.8 ± 0.4 V
V + to È (Digital Ground) = +5 + 1 V
3. Reference Diode: CR101 to È (Analog Ground)= -1.23 ±0.03 V
4. Negative Reference Divider: (Reference to È Analog Ground)
R119
Pin

mV*

Pin

mV*

Pin

mV*

2
3
4

-312
-293
-277

5
6
7

-274
-272
-271

8
9
10

-267
-254
-242

*The setting of R104 and the tolerances of the resistors within the network
(R119) can affect these levels by ±5%.
5. Deintegrate Comparators:
U103, Pin 3 to È =190 ±30 mV
U103B, Pin 6 to È = -180 ±30 mV
6. Cold Junction Voltage:
(Ambient temperature 25º ±3ºC)
-COM to È = 1.3 mV ±0.6
12

4.5.3

Waveform Checks For Model 871A
Waveform Checks: Referenced To È (Digital Ground)
Clock
A/D Clock
Ref. Clock 1
Ref. Clock 2
Backplane

4.5.3

U111, Pin 3
U112, Pin 5
U102, Pin 9
U102, Pin 14
U101, Pin 20

360 kHz,
40 kHz,
60 kHz,
6 kHz,
50 Hz,

5V pp Square Wave
5V pp Rectangular Wave
5V pp Rectangular Wave
5V pp Rectangular Wave
5V pp Rectangular Wave

Waveform Checks For Model 872A
Waveform Checks: Referenced To È (Digital Ground)
Clock
A/D Clock
Ref. Clock 1
Ref. Clock 2
Backplane

SECTION

5

U111, Pin 3
U112, Pin 5
U102, Pin 9
U102, Pin 14
U101, Pin 20

360 kHz,
40 kHz,
90 kHz,
9 kHz,
50 Hz,

5V pp
5V pp
5V pp
5V pp
5V pp

Square Wave
Rectangular Wave
Rectangular Wave
Rectangular Wave
Rectangular Wave

SPECIFICATIONS FOR MODEL 871A AND
MODEL 872A

GENERAL
THERMOCOUPLE INPUTS:

Two, switch selected

ACCURACY:

±0.25% of reading +1ºC (0.25% ±1.8ºF) for
one year, 64º to 82ºF (10º to 28ºC) ambient

EXTENDED SPAN ACCURACY:

-40º to -55ºC: +2ºC, -0ºC

REPEATABILITY:

±0.2ºC Typical (one hr., constant
ambient temperature)

TEMPERATURE COEFFICIENT:

64º to 82ºF (18º to 28ºC); Less than
±(0.025% of reading +0.1ºC) 1ºC; for
temperatures less than 64ºF (18ºC) or
greater than 82ºF (28ºC)

MAXIMUM ALLOWABLE TC
INPUT OVERLOAD:

35Vdc (42 V peak AC)

NORMAL MODE REJECTION
RATIO:

Greater than 45 dB at 50 and 60 Hz

COMMON MODE REJECTION
RATIO (1KΩ UNBALANCE):

Greater than 120 dB at dc, 50 and 60 Hz

DISPLAY:
MAXIMUM COMMON MODE
VOLTAGE:

3¹⁄₂ digit LCD, 0.6" height. Polarity and
decimal point indication
500 V

13

SPECIFICATION (con’t)
ENVIRONMENTAL LIMITS
OPERATING:

32º to 122º F (0º to 50ºC), less than 80%
relative humidity up to 95ºF (35ºC), less than
70% relative humidity from 95º to 122ºF (35º
to 50ºC); linearly derate 3% R.H./ºC 35º to
60ºC.

STORAGE:

-30º to 140ºF (-35º TO 60ºC), less than 90%
relative humidity up to 95ºF (35ºC); linearly
derate 3% R.H./ºC 35º to 60ºC.

COLD-JUNCTION COMPENSATION: Semiconductor temperature sensor
INPUT CONNECTIONS:

OMEGA subminiature thermocouple
connector (2)

mV OUTPUT CONNECTION:

Dual banana jacks

POWER:

9 V alkaline or carbon-zinc (NEDA 1604)
battery

BATTERY LIFE CONTINUOUS:

100 hrs. typical with alkaline battery; 50 hrs.
typical with carbon-zinc battery

BATTERY INDICATOR:

Display indicates “BAT” when less than
10% of life remains

DIMENSIONS:

H: 7.0" (178 mm) x W: 3.1" (78 mm)
x D: 1.6" (42 mm)

WEIGHT:

10.6 oz. (300 grams)

FOR MODEL 871A
TEMPERATURE SENSOR:

Type K (NiCr-NiAI) thermocouple

RANGES:

200ºF, 2000ºF, 200ºC, 1370ºC

TEMPERATURE SPAN:

-40.0º to 199.9ºF, -40º to 1999ºF
-40.0º to 199.9ºC, -40º to 1370ºC

RESOLUTION:

0.1ºF, 1ºF, 0.1ºC, 1ºC

INPUT CURRENT:

200nA maximum

CONVERSION PERIOD:

400 ms (2.5 readings per sec)

OVERRANGE AND OPEN
THERMOCOUPLE INDICATION:

Display blanked except for overrange digit

14

SPECIFICATION (con’t)
mV OUTPUT:
Non-linearized, cold-junction compensated,
amplified thermocouple output (Not isolated
from selected TC input.)
On ºC Scale:
1mV/ºC on 200º range
0.1mV/ºC on 1370º range
0ºC = OmV output
On º F Scale:
5/9 mV/ºF on 200º range
5/90 mV/ºF on 2000º range
0ºF = OmV output
mV OUTPUT ACCURACY
(18º - 28ºC)

±(3% + 1mV) on 200º ranges;
±(3% + 0.1mV) on High ranges
Includes TC non-linearity, excludes TC
errors. Output resistance: 10k ohms

mV OUTPUT TEMPERATURE
COEFFICIENT (0º - 18ºC & 28º 50ºC):
mV OUTPUT PROTECTION:

Less than (0.1 x accuracy specification)/ºC
35 V dc (42 V peak) max; +mV to -mV
and selected TC input to +mV or -mV.

THERMOCOUPLE
LINEARIZATION:

Multi-scope A/D with 11 piecewise linear
segments between -40º and 1370ºC

FOR MODEL 872A
TEMPERATURE SENSOR:

Type J (Fe-CuNi) thermocouple

RANGES:

200ºF, 1400ºF, 200ºC, 760ºC

TEMPERATURE SPAN:

-40.0º TO 199.9ºF, -40º to 1400ºF
-40.0º TO 199.9ºC, -40º to 760ºC

RESOLUTION:

0.1ºF, 1ºF, 0.1ºC, 1ºC

INPUT CURRENT:

25 nA typical; 50 nA max

CONVERSION PERIOD:

400 ms (2.5 readings/sec.)

OVERRANGE AND OPEN
THERMOCOUPLE INDICATION:

3 least significant digits blanked

15

SPECIFICATION (con’t)
mV OUTPUT:

Non-linearized, cold-junction compensated, amplified thermocouple output
(Not isolated from selected TC input.)
On ºC Scale:
1mV/ºC on 200º range
0.1mVºC on 760º range
0ºC = OmV output
On ºF Scale:
5/9 mV/ºF on 200º range
5/90 mV/ºF on 1400º range
0ºF = OmV output

mV OUTPUT ACCURACY
(18º - 28º C)

mV OUTPUT TEMPERATURE
COEFFICIENT (0º - 18ºC
& 28º - 50°C)

±(5% + 1mV) on 200º ranges;
±(5% + 0.1mV) on High ranges
Includes TC non-linearity, excludes TC
errors. Output resistance: 10k ohms

Less than (0.1 x accuracy specification)/ºC

mV OUTPUT PROTECTION:

35 V dc (42 V peak) max; +mV to -mV and
selected thermocouple input to +mV or -mV.

THERMOCOUPLE LINEARIZATION:

Multi-scope A/D with nine piecewise linear
segments between -40º and 760º C

5.1

PARTS LIST

Circuit
Desig.

Description

Circuit
Desig.

Description

BT101
C101
C102*

Battery, 9V NEDA 1604
Cap, 0.22 µF, 100 V, Polyester
Cap, 0.1 µF, 160 V,
Polypropylene
Cap, 0.1 µF, 100 V,
Polypropylene
Cap, 470pF, 1000 V, Ceramic
Disc
Cap, 220pF, 1000 V, Ceramic
Disc
Cap, 10 µF, 16 V, Aluminum
Electrolytic
Cap, 33pF, 1000 V, Ceramic
Disc
Cap, 47pF, 1000 V, Ceramic
Disc
Cap, 0.1 µF, 16 V, Ceramic
Cap, 0.1 µF, 50 V, Ceramic
Cap, 4.7 µF, 16 V, Aluminum
Electrolytic
Diode, Low Voltage Reference
Liquid Crystal Display, 31⁄2 digit
TC Connector, Female
Jack, Female SMP Connector
Jack, Female
Connector, Battery
Connector, Battery
Transistor, NPN, Switch, 2N3904
Thick Film Resistor Network

R102
R103
R104
R105
R106*

Pot, 100 kΩ
Pot, 2 kΩ
Pot, 1 kΩ
Pot, 500 Ω
Resistor, 180 kΩ, 5%, 1/4W,
Comp
Resistor, 240 kΩ, 5%, 1/4W,
Comp
Resistor, 47 kΩ, 5%, 1/4W,
Comp
Resistor, 470 kΩ, 5%, 1/4W,
Comp
Pot, 2 kΩ
Thick Film Resistor Network
Resistor, 10 kΩ, 10%, 2W,
Comp
Resistor, 10 MΩ, 10%, 1/4W,
Comp
Resistor, 5.1 kΩ, 5%, 1/4W,
Comp
Resistor, 1 MΩ, 5%, 1/4W,
Comp
Resistor, 1 MΩ, 10%, 1/4W,
Comp
Resistor, Selected, 1%, 1/8W,
Mtf.
R115 and U110 are a selected
set

C102**
C103
C104
C105, 106
C107
C108
C109*
C109**
C110**
CR101
DS101
J101, 102
J103, J104**
J104, 105*
J105**
J106*
Q101, 102
R101

R106**
R107
R108
R109
R110
R111
R112
R113
R114*
R114**
R115, 116*
R115**

16

TABLE 3-1

5.1
Circuit
Desig.

Description

Circuit
Desig.

CONTROLS AND INDICATORS

R116**

R116, R122, and U109 are a
selected set
Thick Film Resistor Network
Resistor, 150 kΩ, 5%, 1/4W,
Comp
Resistor, 5.6 kΩ, 5%, 1/4W,
Comp
Thick Film Resistor Network
Resistor, 10 kΩ, 5%, 1/4W,
Comp
Resistor, 10.0 kΩ, 6.1%, Metal
Film
Resistor , Selected, 5%, 1/4W,
Comp
R116, R122, and U109 are a
selected set
Resistor, 324 kΩ, 1%, 1/8W, Mtf.
Resistor, 316 kΩ, 1%, 1/8W, Mtf.
Resistor, 10.0 Ω, 0.1%, Metal
film
Resistor, 10 kΩ, 0.1%, MF
Resistor, 9.9 kΩ, 0.1%, MF
Resistor, 4.7 kΩ, 5%, 1/4 W,
Comp
Switch, Rotary
Switch, Rotary
Switch, SPDT, ON-OFF
TC Cable

U101

R117
R118*

ITEM

CONTROL/INDICATOR

FUNCTION

R118**

1.

ON/OFF switch

Turns unit on or off.

R119
R120

2.

Scale and range selector

Allows user to select: ºF or ºC scale,
any of four possible thermocouple
ranges, and 1.0º or 0.1º resolution.

R121**
R122*†
R122**

3.

4.

5.

6.

7.

PARTS LIST (continued)

MODEL 871A AND MODEL 872A

Thermocouple selector

Thermocouple input jacks

Analog output jack,
+mV OUT and -mV OUT

Low battery indicator

Overrange/open
thermocouple indication

Allows user to read either of two
thermocouples by simply turning
the dial.
Allows direct connection of two
thermocouples to the unit via an
OMEGA miniature male connector.
The analog output is a nonlinearized,
cold junction compensated, amplified
thermocouple output. (Not isolated
from selected TC input.) Use a
recorder to monitor temperature or
a sensitive DMM to make higher
resolution reading. Refer to mV
output specifications for mV output
to temperature ratios.
When the battery level is low, BAT
will appear in upper left corner of
the display. Battery should then be
replaced to avoid error erroneous
readings due to battery failure.

R123*
R123**
R124**
R126*
R127*
R128*
S101
S102
S103
TC-1, 2

U102
U103, U104*
U103, U104**
U105
U106
U107
U108
U109*
U109**
U110*
U110**
U111
U112
U113
U114
Y101

1
3 ⁄2 Digit Single Chip A/D
Converter
Dual Synchronous Up Counter
Low Power JFET Input Op Amp
Dual JFET Input Op Amp
COS/MOS Quad Bilateral Switch
CMOS Quad Exclusive OR Gate
Dual “D” Type Flip-Flop
14 Stage Binary Counter
Linear Op Amp
R116, R222, and U109 are a
selected set
Current Source
R115 and U110 are a
selected set
Quad, 2 Input NOR Gate
COS/MOS Divide by “N”
Counter
Quad, 2 Input, AND Gate
Analog Multiplexer
Resonator, 360 kHz

† Value selected during calibration at factory.
* For Model 871A only
** For Model 872A only
To order parts, specify model and circuit designation.

COMPONENT LAYOUT FOR MODEL 871A
R120
R127

Jumper

When an overrange indication or
open thermocouple condition exits,
a 1 followed by blanked digits will
appear in the display.

COMPONENT LAYOUT FOR MODEL 872A
R118A

R121

Jumper
R124

4

Description

17

SCHEMATIC
MODEL 871A DIGITAL THERMOMETER

18

SCHEMATIC
MODEL 872A DIGITAL THERMOMETER

19

MADE
IN

USA

WARRANTY/DISCLAIMER

OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month
grace period to the normal one (1) year product warranty to cover handling and shipping time.
This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s
Customer Service Department will issue an Authorized Return (AR) number immediately upon
phone or written request. Upon examination by OMEGA, if the unit is found to be defective it will
be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting
from any action of the purchaser, including but not limited to mishandling, improper interfacing,
operation outside of design limits, improper repair, or unauthorized modification. This
WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence
of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration;
improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s
co n t ro l. Co mp o n e n t s w h ich wear are not warranted, including but not limited to
contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any
damages that result from the use of its products in accordance with information provided by
OMEGA, either verbal or written. OMEGA warrants only that the parts
manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER
WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED,
EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total
liability of OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a
“Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in
medical applications or used on humans. Should any Product(s) be used in or with any nuclear
installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes
no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally,
purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever
arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED
RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR WARRANTY RETURNS, please have the
FOR NON-WARRANTY REPAIRS, consult
following information available BEFORE
OMEGA for current repair charges. Have the folcontacting OMEGA:
lowing information available BEFORE
contacting OMEGA:
1. P.O. number under which the product was
PURCHASED,
1. P.O. number to cover the COST
of the repair,
2. Model and serial number of the product under
warranty, and
2. Model and serial number of product, and
3. Repair instructions and/or specific
3. Repair instructions and/or specific problems
problems relative to the product.
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.
This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of
OMEGA ENGINEERING, INC.

Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
TEMPERATURE
M
U
M
U
M
U
M
U
M
U

Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
Wire: Thermocouple, RTD & Thermistor
Calibrators & Ice Point References
Recorders, Controllers & Process Monitors
Infrared Pyrometers

PRESSURE, STRAIN AND FORCE
M
U
M
U
M
U
M
U

Transducers & Strain Gauges
Load Cells & Pressure Gauges
Displacement Transducers
Instrumentation & Accessories

FLOW/LEVEL
M
U
M
U
M
U
M
U

Rotameters, Gas Mass Flowmeters & Flow Computers
Air Velocity Indicators
Turbine/Paddlewheel Systems
Totalizers & Batch Controllers

pH/CONDUCTIVITY

M
U pH Electrodes, Testers & Accessories

M
U Benchtop/Laboratory Meters
M
U Controllers, Calibrators, Simulators & Pumps
M
U Industrial pH & Conductivity Equipment

DATA ACQUISITION
M
U
M
U
M
U
M
U
M
U

Data Acquisition & Engineering Software
Communications-Based Acquisition Systems
Plug-in Cards for Apple, IBM & Compatibles
Datalogging Systems
Recorders, Printers & Plotters

HEATERS
M
U
M
U
M
U
M
U
M
U

Heating Cable
Cartridge & Strip Heaters
Immersion & Band Heaters
Flexible Heaters
Laboratory Heaters

ENVIRONMENTAL
MONITORING AND CONTROL
M
U
M
U
M
U
M
U
M
U
M
U

Metering & Control Instrumentation
Refractometers
Pumps & Tubing
Air, Soil & Water Monitors
Industrial Water & Wastewater Treatment
pH, Conductivity & Dissolved Oxygen Instruments

M618/0896



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Create Date                     : 1999:07:30 06:09:13Z
Modify Date                     : 2005:03:10 13:30:35-06:00
Keywords                        : thermocouple, thermocouples, rtd, temperature control, temperature measurement, resistance temperature device, resistance temperature, temperature sensor, thermocouple probe, thermacouple, thermocuple, thermalcouple, infrared temperature, ir, infrared, meter, panel meter, multimeter, multimeters, multi meter, meter, meters, temperature controller, process control, temperature transmitter, thermowell, thermistor, thermistors, thermocouple wire, electric heater
Page Count                      : 24
Creation Date                   : 1999:07:30 06:09:13Z
Producer                        : Acrobat Distiller 3.01 for Power Macintosh
Author                          : Omega Engineering
Mod Date                        : 2005:03:10 13:30:35-06:00
Metadata Date                   : 2005:03:10 13:30:35-06:00
Creator                         : Omega Engineering
Title                           : Handheld Temperature Meters
EXIF Metadata provided by EXIF.tools

Navigation menu