Keysight M9217A
PXIe Single-Slot 2-Channel Isolated Digitizer
Data Sheet
Introduction
The Keysight Technologies, Inc. M9217A is a PXI Express based, single-slot 2-channel, isolated
digitizer that supports up to ±256 V input voltage at 20 MSa/s sampling rate with 16-bit resolution.
With its built-in peak voltage protection of up to 400 V, the M9217A is ideal for analyzing high-voltage
and transient signals as seen in automotive, mechatronics and aerospace defense applications.
Key Features
±256 V input voltage
2-channel isolated, with ADCs per channel, simultaneously sampled channels
20 MSa/s sampling rate at 16-bit resolution
Built-in 400 V peak voltage protection
Single PXI express slot
PXI trigger types supported: PXI TTL / trigger and PXI Star trigger
IVI-COM, IVI-C, LabVIEW G drivers
03 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet
±256 V input voltage
Input channels with the ability to measure waveforms up to 256 V are ideal for analyzing
high-voltage and transient signals as seen in automotive and aerospace defense
applications. Most oscilloscopes and PXI digitizers have a maximum input range less
than 40 Vpk. This also reduces the need to add expensive input signal attenuation and
signal conditioning circuitry, saving test development time and money.
2-channel isolated, with ADCs per channel
In addition, since the ±256 V input voltage range is combined with 16-bit analog-to-
digital converters (ADCs), the isolated front-end and low input offset allows voltages
from as low as 250 mV to as high as 256 V to be measured simultaneously.
The M9217A’s individually isolated 2-channel inputs have been designed for high
performance with an A/D converter per channel to ensure the signals you measure
are accurately digitized without distortion or additional noise. Channel input range is
congurable from ±250 mV up to ±256 V, with a oating voltage up to ±40 V to accom-
modate differential waveform acquisition.
20 MSa/s sampling rate at 16-Bit resolution
The M9217A’s ability to support up to 20 MSa/s sampling rate at 16-bit resolution
enables sampling of transient waveforms like injector waveforms (Figure 1).
toff Vpk
tpw
Ringing
Vdson
ton
ton < 0.1 us
Vdson ~ 600 mV - 800 mV
toff ~ 10 us
Vpk ~ 70 V - 146 V
200 V (peak)
Figure 1. Sample injector waveform
04 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet
Built-in 400 V peak voltage protection
To support high voltage and transient signals in automotive, mechatronics and aerospace
and defense applications, the M9217A has a built-in peak voltage protection against
overvoltage spikes of up to 400 V.
32 MSa/channel with up to 1024 segmented memory
Coupled with the memory segmentation feature of up to 1024 records, the M9217A’s
32M sample memory size enables multiple bursts of data to be digitized without the
need for re-initialization between bursts. The record size is congured by selecting the
total number of samples, including pre-trigger samples.
Support for multiple trigger types
The M9217A supports different trigger types like PXI trigger bus, PXI Star, external input/
output, channel and software trigger. Congurable trigger delay and trigger hold-off
allow you to better dene where record data is collected relative to the trigger event.
The congurable trigger delay feature allows precise positioning of acquisition relative to
the trigger event, while the trigger hold-off feature helps avoid false triggers.
The PXI TTL trigger output enables the M9217A digitizer to synchronize to other devices.
Multiple M9217A digitizers can be synchronized for higher channel count.
IO Libraries
Keysight input/output or IO Libraries Suite offers fast and easy connection to instru-
ments. The latest suite version extends that capability to include modular instruments.
New support for PXI such as the Keysight IO Libraries Suite helps you display all of
the modules in your system. From here, you can view information about the installed
software or start the soft front panel of the module. Launch the soft front panel of the
module directly from Keysight Connection Expert. Figure 2 shows the M9217A soft front
panel.
Figure 2. M9217A soft front panel
05 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet
Drivers
The M9217A digitizer is supplied with a comprehensive portfolio of module drivers, doc-
umentation, examples, and software tools to help you quickly develop test systems with
your software platform of choice. The module comes with IVICOM, IVI-C, and LabVIEW
G software drivers that work in the most popular test and measurement development
environments, including LabVIEW and LabWindows/CVI from National Instruments,
Microsoft C/C++, C#, and VB.NET.
Easy software integration
The module software support provides context-sensitive help, complete documentation
and code examples that allow a quick module set up and basic acquisition functionalities.
These code examples can be easily modied, so that the card can be quickly integrated
into a test system. Included are application code examples for LabVIEW, LabWindows/
CVI, Visual Studio C, C++, and C# and Visual Basic.
Calibration Intervals
The M9217A is factory-calibrated and shipped with a calibration certicate. Calibration is
recommended every year in order to verify product performance.
M9217A specications
2-channel isolated digitizer with ADCs per channel and utilizing one PXI express slot.
Max sample rate 20 MSa/s
Sample resolution 16 bits
Input conguration Isolated inputs ( each channel independently isolated)
Isolation voltage (low to chassis) +40 V
Maximum input range (high to low) +256 Vpk
Input impedance 1 MW // 40 pF
Input coupling DC or AC
AC cutoff frequency < 10 Hz
Input ranges ±256 V, ±128 V, ±64 V, ±32 V, ±16 V, ±8 V,
±4V, ±2 V, ±1 V, ±500 mV, ±250 mV
Over voltage protection Yes
Maximum overvoltage protection ±400 Vppk
Analog bandwidth (-3 dB) 20 MHz typical
06 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet
Trigger
Each trigger event causes acquisition of data into a single record at the congured sample rate. The
number of data records is congurable from 1 to 1024.
Source Description
Immediate Trigger at initiation of acquisition,
External SMD connector, TTL input edge (rising edge)
PXI_Trig[0:7], PXI_STAR PXI connector XJ4 , TTL input edge (rising edge)
DSTARB PXIe connector XJ3, TTL input edge (rising edge)
Software Instrument command
Channel/edge Selectable level, rising/falling, hysteresis.
Channel/window Selectable high and low levels, leaving/entering.
Sampling
Programmable sample rates 1 kSa/s, 2 kSa/s, 5 kSa/s, 10 kSa/s, 50 kSa/s, 100 kSa/s, 200 kSa/s,
500 kSa/s, 1 MSa/s, 2 MSa/s, 5 MSa/s, 10 MSa/s, 20 MSa/s
External event output Trigger
Output signal TTL ( rising edge )
Impedance 25 or 50
Pulse width 1 µs
Trigger modes
Record length 2 to 32 million samples
Pre trigger 0 to record length -1
Post trigger Record length-pre-trigger
Timestamp triggered event Continuous running timestamp
Timestamp resolution 10 ns
Trigger delay 0 – 3600 s with 50 ns resolution
Trigger hold-off 0 – 10 s with 50 ns resolution
Trigger latency150 ns
1. Latency between edge trigger detection and rst trigger sample.
Accuracy
DC accuracy
Total specication (% of reading + % of range)
23 °C ±5%
Range ±% of reading ±% of range
250 mV 0.1 0.3
500 mV 0.1 0.2
1 V, 2 V 0.1 0.12
4 V, 64 V 0.1 0.3
8 V, 128 V 0.1 0.2
16 V, 32 V, 256 V 0.1 0.12
100,000 reading average @ 1 MSa/s
Integral nonlinearity ±5 LSB
Differential nonlinearity ±1 LSB
07 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet
Dynamic characteristic (typical)
Source SFDR dBc THD dBc
250 mV 71 79
500 mV 77 83
1 V 81 85
2 V 85 82
4 V 70 80
Noise power ( typical)
Range Noise voltage1
Vrms
Equivalent2
S/N dB
250 mV 0.5 mV 51
0.5 mV 0.55 mV 55
1 V 0.6 mV 60
2 V 0.9 mV 63
4 V 4 mV 56
8 V 6.5 mV 58
16 V 8.5 mV 61
32 V 11 mV 65
64 V 60 mV 56
128 V 94 mV 58
256 V 120 mV 62
1. 50 Ω source resistance.
2. For -1 dB of full scale range
AC atness (DC-4 MHz) typical
250 mV, 500 mV,
1 V, 2 V
< 0.3 dB relative to 20 kHz
4 V, 8 V, 16 V, 32 V < 0.5 dB relative to 20 kHz
128 V < 1.2 dB relative to 20 kHz
Waveform memory
Data memory standard132 Msa/ch
Random access to readings in multiple record mode Capture multiple records from single trigger
source
1. Nominal values. Specic sample max is 30,000,000
Waveform measurements
Voltage Peak-to-peak, Minimum, Maximum, High (Top), Low (Base), Amplitude, RMS,
Cycle RMS, Average, Cycle Average, Overshoot, Preshoot
Time Rise, Fall, Frequency, Period, Positive Width, Negative Width, Positive Duty
Cycle, Negative Duty Cycle
Utilities
Calibration cycle 1 year
Measured for input of 1 Mhz
08 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet
This information is subject to change without notice.
© Keysight Technologies, 2017
Published in USA, December 1, 2017
5992-0815EN
www.keysight.com
www.keysight.com/nd/automotive
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada (877) 894 4414
Brazil 55 11 3351 7010
Mexico 001 800 254 2440
United States (800) 829 4444
Asia Pacic
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 11 2626
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 6375 8100
Europe & Middle East
Austria 0800 001122
Belgium 0800 58580
Finland 0800 523252
France 0805 980333
Germany 0800 6270999
Ireland 1800 832700
Israel 1 809 343051
Italy 800 599100
Luxembourg +32 800 58580
Netherlands 0800 0233200
Russia 8800 5009286
Spain 800 000154
Sweden 0200 882255
Switzerland 0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom 0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-9-7-17)
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
Evolving Since 1939
Our unique combination of hardware, software, services, and people can help you
reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight.
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.keysight.com/find/emt_product_registration
Register your products to get up-to-date product information and
find warranty information.
Keysight Services
www.keysight.com/find/service
Keysight Services can help from acquisition to renewal across your
instrument’s lifecycle. Our comprehensive service offerings—one-
stop calibration, repair, asset management, technology refresh,
consulting, training and more—helps you improve product quality
and lower costs.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure
your instruments are operating to specification, so you can rely on
accurate measurements.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and
product breadth, combined with channel partner convenience.

Navigation menu