RF Signal Generator
9 kHz to 3.0 GHz
Data Sheet
02 | Keysight | N9310A RF Signal Generator - Data Sheet
Denitions and Conditions
Specications” describe the performance of parameters that apply to the full tempera-
ture range of 5 to 45 °C, unless otherwise noted.
Typical” values describe additional product performance information. It is performance
beyond specications that 80 percent of the units exhibit with a 95 percent condence
level over the temperature range 20 to 30 °C. Typical performance does not include mea-
surement uncertainty.
Nominal” values indicate expected performance, or describe product performance that
is useful in the application of the product.
The signal generator will meet its specications when:
It is within its calibration cycle
It has been turned on at least 45 minutes
It has been stored at an ambient temperature within the allowed operating range for
at least two hours before being turned on; if it had previously been stored at a
temperature range inside the allowed storage range, but outside the allowed
operating range
03 | Keysight | N9310A RF Signal Generator - Data Sheet
Supplemental information
Frequency
Range 9 kHz to 3.0 GHz
Resolution 0.1 Hz
Switching speed < 10 ms Typical; Within 0.1 ppm of nal frequency
Frequency reference
Option PFR Standard
Aging rate ± 1 ×10−7 /year ± 1 ×10−6 / year
± 1.5 ×10−7 /2 years
Temperature stability ± 1.5 ×10−8 (20 to 30 °C) ± 1 ×10−6 (5 to 45 °C)
± 5 ×10−8 (5 to 50 °C)
Timebase reference output
Frequency 10 MHz
Amplitude > 0.35 Vrms level into 50 Ω
Connector BNC female
External reference input
Range 2 MHz, 5 MHz, 10 MHz
Amplitude 0.5 to 2 Vrms
Connector and impedance 50 Ω; BNC female
Output
Power 127 to +13 dBm +20 dBm settable
Resolution 0.1 dB
Accuracy < ± 1 dB Fc ≥ 100 kHz, –120 ≤ Level ≤ +13 dBm,
20 to 30 °C
Switching speed < 10 ms Typical; < 0.3 dB deviation
VSWR (typical) < 1.6 1.5 MHz ≤ Fc ≤ 2.5 GHz
< 1.8 2.5 GHz ≤ Fc ≤ 3 GHz
Output connector and impedance N-type; 50 Ω nominal
Reversal power protection
DC voltage 30 V
RF power +36 dBm 1 minute; the warning for reversed power
protection is nominally at +25 dBm
Spectral purity
SSB phase noise < –95 dBc/Hz Typical, Fc = 1 GHz at 20 kHz offset
Residual FM < 30 Hz rms; < 90 Hz peak CW mode, Fc = 1 GHz; BW = 0.3 to 3 kHz
< 20 Hz rms Res FM optimized mode
Harmonics < –30 dBc Level ≤ 0 dBm, Fc ≥ 1 MHz
Non-harmonics < –50 dBc Level ≤ 0 dBm, ≥ 10 kHz from carrier
Specications
04 | Keysight | N9310A RF Signal Generator - Data Sheet
Supplemental information
Characteristic SSB phase noise
Sweep modes RF and LF
LF sweep range 20 Hz to 80 kHz
RF sweep range 9 kHz to 3 GHz
Sweep points 2 to 1,001
Dwell time 10 ms to 1 s
Amplitude
Sweep range 127 to +13 dBm
Sweep points 2 to 1,001
Dwell time 10 ms to 1 s
Simultaneous modulation 1
AM I/Q FM ØM Pulse
Internal External Internal External Internal External
AM Internal − −
External − − − −
I/Q − − • • •
FM Internal • • • •
External • • • •
ØM • • • •
Pulse Internal − −
External − −
1. The N9310A has one external modulation input connector. The simultaneous external modulations are applied to the same input signal.
Specications (continued)
fc = 1,000 MHz fc = 2,000 MHz
05 | Keysight | N9310A RF Signal Generator - Data Sheet
Specications (continued)
Supplemental information
Amplitude modulation (Fc ≥ 100 kHz)
Operating modes Internal, external AC
Range 0 to 100% Envelope peak < maximum specied power
Resolution 0.1%
Rates 20 Hz to 20 kHz
Accuracy < ± (5% of setting +0.2%) 1 kHz, 0 dBm and 80% modulation,
0.3 to 3 kHz bandwidth
Distortion < 2% 1 kHz, 0 dBm and 80% modulation,
0.5 to 15 kHz bandwidth
External input MOD IN connector
Sensitivity 0.5 Vpeak Input voltage for 100% modulation depth
Input impedance BNC; > 100 kΩ Nominal
Frequency modulation (Fc ≥ 100 kHz)
Operating modes Internal, external AC
Frequency deviation 20 Hz to 100 kHz
Resolution < 1% Minimum 1 Hz
Rates 20 Hz to 80 kHz
Distortion 1% 1 kHz rate, 0.3 to 3 kHz bandwidth,
deviation = 50 kHz
Deviation accuracy < ± (5% of FM deviation +300 Hz) 1 kHz, 0 dBm and 50 kHz deviation,
0.3 to 3 kHz bandwidth
Carrier frequency deviation < 200 Hz Relative to carrier; external mode
External input MOD IN connector
Sensitivity 0.5 Vpeak Input voltage for 100 kHz modulation deviation
Input impedance BNC; > 100 kΩ Nominal
Phase modulation (Fc ≥ 100 kHz)
Operating modes Internal
Phase deviation 0 to 10 rad Rate ≤ 10 kHz
0 to 5 rad 10 kHz < rate ≤ 20 kHz
Resolution < 1%
Rates 300 Hz to 20 kHz
Deviation accuracy < ± (5% of FM deviation +0.2 rad) 1 kHz rate, 0.3 to 3 kHz bandwidth
Distortion < 1.5% 1 kHz rate, 0.3 to 3 kHz bandwidth,
deviation = 5 rad
Input impedance BNC; > 100 kΩ Nominal
Pulse modulation
Operating modes Internal, external
On/Off ratio ≥ 40 dB
Rise/Fall time < 3 μs
Pulse width 100 μs to 1 s Internal, external
Pulse period 200 μs to 2 s Internal
Time resolution 1 μs
Input connector and voltage level BNC female; TTL
06 | Keysight | N9310A RF Signal Generator - Data Sheet
Specications (continued)
Supplemental information
Internal modulation source Provides a modulation signal for AM, FM, phase modulation, and LF out
Waveform Sine
Frequency range 20 Hz to 80 kHz
Resolution 0.1 Hz
Accuracy 0.005% Typical
LF out (Internal modulation source)
Amplitude 0 to 3 Vpeak Level to high impedance
Output voltage resolution < 1% 1 mV minimum resolution
Frequency response < ± 0.2 dB 20 Hz to 20 kHz
Total harmonic distortion < 0.1% Typical; 20 Hz to 20 kHz, 30 kHz low pass lter
Connector and impedance BNC female; < 1Ω Front panel
Precision frequency reference (option PFR)
Output frequency 10 MHz
Accuracy ± [(time since last adjustment × aging rate) + temperature stability+ calibration accuracy 2] 3
Temperature Stability
20 to 30 °C ± 1.5 ×10−8
5 to 50 °C ± 5 ×10−8
Aging
1 year ± 1 × 10−7
2 years ± 1.5 × 10−7
Achievable Initial Calibration Accuracy ± 4 × 10−8
Output level > +4 dBm
Connector BNC female, 50 Ω nominal, rear panel
Calibration connection Mini USB port, real panel
I/Q modulation (Option 001 only)
Operating mode External I/Q inputs
VSWR < 1.5
Full scale input 22
I + Q = 0.5 Vrms
Modulation frequency range DC to 20 MHz At 3 dB points
Carrier suppression 40 dBc Typical; modulation frequency = 10 kHz
QPSK EVM 3% Typical; 1 Msps; 0.22 RRC lter
GMSK phase error 1.2 °rms Typical; 1 Msps; BT = 0.5
Connector and impedance BNC female; 50 Ω Rear panel
2. Calibration accuracy depends on how accurately the frequency standard was adjusted to 10 MHz. If the adjustment procedure is followed, the calibration
accuracy is given by the specication of the achievable initial calibration accuracy.
3. The specication applies after the generator has been powered on for four hours.
07 | Keysight | N9310A RF Signal Generator - Data Sheet
Supplemental information
USB connector
USB host interface 3 x A plug V 1.1 protocol
USB device interface 1 x B plug V 1.1 protocol
General
Recommended calibration cycle 2-year Keysight Technologies, Inc. has veried that the
stability of this product’s architecture justies a
longer calibration interval of 2 years.
Power requirement 100 to 240 Vac; 50 to 60 Hz Auto-ranging
Power consumption 65 W
Temperature range 5 to 45 °C Operating
–20 to 70 °C Storage
Weight 9.2 kg Nominal
Dimensions 132.5 x 320 x 400 mm H x W x D
Display
Resolution 640 x 480
Size 165.1 mm (6.5 in) diagonal (nominal)
Data storage
Internal 16 MB nominal
External Supports USB 2.0-compatible memory devices
EMC
Complies with European EMC Directive 2004/108/EC
IEC/EN 61326-1 or IEC/EN 61326-2-1
CISPR Pub 11 group 1, class A
AS/NZS CISPR 11:2004
ICES/NMB-001:2004
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Safety
Complies with European Low Voltage Directive 2006/95/EC
IEC/EN 61010-1 2nd edition
Canada: CSA C22.2 No. 61010-1-04
USA: UL 61010-1 2nd edition
Audio noise
Acoustic noise emission Geraeuschemission
LpA < 70 dB LpA < 70 dB
Operator position Am Arbeitsplatz
Normal position Normaler Betrieb
Per ISO 7779 Nach DIN 45635 t.19
Environmental stress
Samples of this product have been type tested in accordance with the Keysight Environmental Test Maunal and veried to be robust against the
environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration,
altitude, and power line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3
Specications (continued)
08 | Keysight | N9310A RF Signal Generator - Data Sheet
This information is subject to change without notice.
© Keysight Technologies, 2017
Published in USA, December 1, 2017
5990-8116EN
www.keysight.com
www.keysight.com/find/n9310a
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada (877) 894 4414
Brazil 55 11 3351 7010
Mexico 001 800 254 2440
United States (800) 829 4444
Asia Pacic
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 11 2626
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 6375 8100
Europe & Middle East
Austria 0800 001122
Belgium 0800 58580
Finland 0800 523252
France 0805 980333
Germany 0800 6270999
Ireland 1800 832700
Israel 1 809 343051
Italy 800 599100
Luxembourg +32 800 58580
Netherlands 0800 0233200
Russia 8800 5009286
Spain 800 000154
Sweden 0200 882255
Switzerland 0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom 0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-9-7-17)
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
Evolving Since 1939
Our unique combination of hardware, software, services, and people can help you
reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight.
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
www.keysight.com/find/emt_product_registration
Register your products to get up-to-date product information and
find warranty information.
Keysight Services
www.keysight.com/find/service
Keysight Services can help from acquisition to renewal across your
instrument’s lifecycle. Our comprehensive service offerings—one-
stop calibration, repair, asset management, technology refresh,
consulting, training and more—helps you improve product quality
and lower costs.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure
your instruments are operating to specification, so you can rely on
accurate measurements.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and
product breadth, combined with channel partner convenience.

Navigation menu