TSI PARTICLE
TECHNOLOGY
PARTICLE INSTRUMENTS

UNDERSTANDING, ACCELERATED

TSI PARTICLE
TECHNOLOGY
A leading developer of aerosol research instrumentation
since 1966, TSI® Incorporated offers a line of particle
instru­ments that is second to none. Our products for sizing,
counting, characterizing, gen­erating, and dispersing aerosol
particles are well known—and well respected—all over the
world. We are dedicated to providing our customers with
the most innovative particle technology available.
This catalog contains many proven instruments as well as
our newest product offerings. The scope of this product
line shows how TSI leads when it comes to performance
and depth. Our intent is to offer a TSI in­strument to handle
most every particle research need. So browse through the
following pages, then let us know how we can help you
achieve your research goals.

TABLE OF
CONTENTS
PARTICLE SIZERS

AIR MONITORING

Scanning Mobility Particle Sizer™ Spectrometers (Series 3938)

2

Ultrafine Particle Monitor (3031)				

21

NanoScan Nanoparticle Sizer (3910)				

3

Environmental Sampling System (3031200)			

21

Optical Particle Sizer (3330)					4
Aerodynamic Particle Sizer® Spectrometer (3321)			

5

PARTICLE SIZERS & IMAGING SYSTEMS

Aerosol Diluter (3302A)					6

Phase Doppler Particle Analyzer (PDPA)				

22

Impactor Inlet (3306)					6

Global Sizing Velocimeter (GSV)				

22

Laser Aerosol Spectrometer (3340)				

7

Fast Mobility Particle Sizer™ Spectrometer (3091)			

8

AEROSOL GENERATORS & DISPERSERS
Vibrating Orifice Aerosol Generator (3450)			

PARTICLE COUNTERS & DETECTORS
Water-Based Condensation Particle Counters (3788, 3787, 3783)
Condensation Particle Counters (3776, 3775, 3772)			

24

Electrostatic Classifiers (3082)					24
9-10

Condensation Monodisperse Aerosol Generator (3475)		

25

10

Electrospray Aerosol Generator (3482)				

25

Nano Enhancer (3777)					10

Small-Scale Powder Disperser (3433)				

26

Handheld Condensation Particle Counter (3007)			

11

Fluidized Bed Aerosol Generator (3400A)			

26

Nanoparticle Surface Area Monitor (3550)			

11

Atomizers (3076, 3079A, 9302, 9306)				

27

Aerosol Electrometer (3068B)					11

Large-Particle Aerosol Generator (8108)				

27

DustTrak™ Aerosol Monitor (8530, 8532, 8533, 8534)		

12

Oil Droplet Generator (9307-6)					27

SidePak™ Personal Aerosol Monitor (AM510)			

12

Submicrometer Monodisperse Aerosol Generation System (3940A)

28

Supermicrometer Monodisperse Aerosol Generation System (3941)

28

AIR FILTER & RESPIRATOR TESTERS
Automated Filter Testers (8130, 8130-EN, 8127, 8127-EN, 3160)

14

INSTRUMENTS & ACCESSORIES

Component Filter-Testing Systems (3150)			

15

Nanometer Aerosol Sampler (3089)				

29

Laser Photometer (8587A)					16

Mass Flowmeter (4000 Series)				

29

Large-Particle Aerosol Generator (8108)				

16

Diffusion Dryer (3062, 3062-NC)				

29

Respirator Leak Tester (8119)					16

Aerosol Neutralizers (3012, 3054, 3077, 348002)			

30

Advanced Aerosol Neutralizer (3088)				

30

ENGINE EMISSIONS

Filtered Air Supply (3074B)					31

Engine Exhaust Particle Sizer™ Spectrometer (3090)		

17

Flow Splitter (3708)						31

Engine Exhaust Condensation Particle Counter (3790A)		

18

Particle Size Selector (376060)				

Electrical Aerosol Detector (3070A)				

18

Vacuum Pumps (3032, 3033)					32

Rotating Disk Thermodiluters and Thermal Conditioner Air Supply

19

Process Aerosol Monitor (3375)				

(379020A, 379030)
Nanoparticle Emission Tester (3795)				

20

31
32

APPLICATIONS
Collectively, our line of particle instruments spans the size range from
0.002 to 2000 micrometers. This unique and comprehensive family of
products is used all over the world in a variety of interesting applications.
+ Basic aerosol research
+ Nanotechnology
+ Environmental studies
+ Bio-aerosol detection
+ Pharmaceutical research
+ Health effects studies
+ Inhalation toxicology
+ Filter testing
+ Instrument calibration and standards
+ Indoor-air-quality testing
+ Industrial hygiene research
+ Energy and combustion studies
+ Diesel- and gasoline-engine emissions measurements
+ Climate change research
+ Particle formation and growth studies
Ask your TSI representative for information about specific applications,
instrument operation, specifications, or new instruments not included in
this catalog. To request additional literature or to place an order, call:

US & CANADA: 1-800-874-2811
EUROPE: +49 241 523030
Additional contact information appears on the back cover. Visit our web
site for the most current product information available:

www.tsi.com

OPERATIONAL RANGES FOR
TSI PARTICLE INSTRUMENTS
Particle Diameter
0.001 µm
1 nm

0.1 µm
100 nm

0.01 µm
10 nm

100 µm
100,000 nm

10 µm
10,000 nm

1 µm
1,000 nm

1,000 µm
1,000,000 nm

2,000 µm
2,000,000 nm

Particle Sizers
Pages 1-8, 17

Series 3938 Scanning Mobility Particle Sizer™ Spectrometers
3090 Engine Exhaust Particle Sizer™ Spectrometer
3091 Fast Mobility Particle Sizer™ Spectrometer
3330 Optical Particle Sizer
3321 Aerodynamic Particle Sizer® Spectrometer

Phase Doppler Particle Analyzers
3340 Laser Aerosol Spectrometer
3031 Ultrafine Particle Monitor

Global Sizing Velocimeter

3910 NanoScan

Exposure Monitors
Page 11-12

3550 Nanoparticle Surface Area Monitor
8520 DustTrak™ Aerosol Monitor
AM510 SidePak™ Personal Monitor

Particle Counters & Detectors
Pages 9-10

3788 Ultrafine Water-Based Condensation Particle Counter
3787 Water-Based Condensation Particle Counter
3776 Ultrafine Condensation Particle Counter
3775 Condensation Particle Counter
3772 Condensation Particle Counter
3790 Condensation Particle Counter for PMP
3007 Handheld Condensation Particle Counter
3070A Electrical Aerosol Detector
3068B Aerosol Electrometer
3783 EPC™ Environmental Particle Counter™ Monitor
3777 Nano Enhancer

Aerosol Generators and Dispersers
Pages 23-28

3482 Electrospray Aerosol Generator
Series 3082 Electrostatic Classifiers
3076/3079/9302/9306 Atomizers
3475 Condensation Monodisperse Aerosol Generator
8108 Large-Particle Aerosol Generator
3400 Fluidized Bed Aerosol Generator
3433 Small-Scale Powder Disperser
3450 Vibrating Orifice Aerosol Generator

0.001 µm
1 nm

0.01 µm
10 nm

0.1 µm
100 nm

1 µm
1,000 nm

10 µm
10,000 nm

Particle Diameter

100 µm
100,000 nm

1,000 µm
1,000,000 nm

2,000 µm
2,000,000 nm

ENGINE EMISSIONS

PARTICLE SIZERS

3938N87

The most comprehensive selection of instruments for sizing submicrometer and supermicrometer particles.

SMPS

Sizer

Particle size
range (µm)

3938L72

0.01 to 1.0

3938L75

0.01 to 1.0

3938N75

0.004 to 0.15

3085A

3938NL75

0.004 to 1.0

3081A & 3085A

3938L76

0.01 to 1.0

3081A

3938N76

0.0025 to 0.15

3938NL76

0.0025 to 1.0

3938L87

0.01 to 1.0

3938N87

0.005 to 0.15

3938NL87

0.005 to 1.0

3938L88

0.01 to 1.0

Particle
concentration
(#/cm3)

Resolution
(total channels
measured)

Channels
per decade

Key feature

DMA

3081A

1 to 10

7**

10 to 600
(selectable)

Varies by
model, 192
channels from
0.001 to
1.0 µm,
collectively

4, 8, 16, 32,
64
(selectable)

Highestresolution;
individual
components
provide
greatest
flexibility

3085A

CPC

3775
butanol

3776

3081A & 3085A
3081A
3085A

3787

3081A & 3085A

3938N88

0.0025 to 0.15

3085A

3938NL88

0.0025 to 1.0

3081A & 3085A

3938E77

0.001 to 0.05

3086

3938EL77

0.001 to 1.0

3081A & 3086

NanoScan

3910

0.01 to 0.42

1 to 106

60

13

8

Portable

FMPS

3091

0.0056 to 0.56

Varies by
size*

1

32

16

Fast
distributions

APS

3321

0.37 to 20

0.001 to 104

1 sec to 18
hours
(variable)

52

32

Aerodynamic
size

OPS

3330

0.3 to 10

1 to 3000

1 sec to 24
hours (variable)

up to 16

LAS

3340

0.09 to 7.5

1 to 18,000a

1 sec to 60
hours (variable)

up to 99

Variable

Condensing
liquid

3772

water

3081A

n/a not applicable
* FMPS: 100 to 107 at 5.6 nm, 1 to 105 at 560 nm (1-sec average)
**	Upper end of concentration determined by Aerosol Neutralizers
Models 3077, 3077A, 3088 specifications

1

Measurement
time (sec)

3788

3777
&
3772

diethylene
glycol &
butanol

Built in

Built
in

isopropyl
alcohol

n/a

n/a

n/a

High res
optical
Optical to
smaller sizes

† Optical detection down to 0.37 µm; aerodynamic sizing down to 0.5 µm
‡	Up to 52 channels of aerodynamic size, up to 64 channels of fluorescence
intensity, up to 16 channels of scattered-light intensity
a
Depends on sample flow

SCANNING MOBILITY PARTICLE
SIZER™ SPECTROMETERS

Model 3938

Upgrade your older SMPS™ system.

Our most versatile submicrometer particle sizers provide

Customers with series SMPS 3936 spectrometers can upgrade to a Series

the highest resolution and accuracy available.

3938 system. Keep the CPC and aerosol neutralizer, then let us modify

Collectively, our Series 3938 Scanning Mobility Particle Sizer (SMPS™)

your old DMA for use on a Model 3082 plat­form! Purchase of the Series

spectrometers measure particles from 1 to 1,000 nm. They display data

3938 software upgrade is required. Call your TSI representative for

using up to 167 actual size channels (up to 64 channels per decade). A

additional requirements.

continuous, fast-scanning measurement technique eliminates gaps in
particle-size-distribution data and allows measurements to be completed

Thermodiluter and Data Merge Software

in as few as 10 seconds. SMPS spectrometers are capable of measuring a

The Rotating Disk Thermodiluters complement TSI particle sizers and

very wide concentration range, from 1 to 10,000,000 particles/cm .

counters when used for particle emission measurements, especially

3

when sampling, diluting, and conditioning exhaust particles from diesel
Component SMPS systems feature the new Electrostatic Classifier Model

and spark-ignition engines. See page 19 for a description of the

3082 with your choice of Differential Mobility Analyzer (DMA), and a

thermodiluters and accessories.

large selection of Con­den­sa­tion Parti­cle Counters (CPCs). The versatility
afforded by individual components enables you to select a system that

The Data Merge Model 390069 Software module works with our Aerosol

best fits your sizing requirements. All components can be operated as

Instru­ment Manager software and simplifies the tedious task of merging

stand-alone instruments for experiments involving monodisperse aerosol

SMPS and Aerody­namic Particle Sizer® (APS™) data files. Data sets can be

generation or counting the total number of particles.

averaged, merged, and fitted to multimodal distributions, and graphs and
tables can be generated easily.

SMPS spectrometers provide additional flexibility by allowing you to
adjust the sample flow rate (from 0.2 to 3.0 L/min). This effectively en­
ables you to select the particle size range of interest. The Electrostatic
Classifier has temperature- and pressure-corrected flow rates for
measurements made at elevations other than sea level. The end result
is an un­matched, proven solution for re­search involving combustion,
atmospheric aerosols, indoor air quality, filter testing, and much more.
A new color touch screen allows easy operation and provides impressive
file-management capabilities. Data can be weighted by number, surface,
volume, or mass. An export function allows easy transport of files
to spreadsheet or other applications for customized data handling.
Additional software capabilities in­clude multiple-scan averaging, a buffer

SMPS component systems give you the highestresolution particle size data.

for comparing data sets, multiple-charge correction for up to 10 charges,
programmable start/stop times, automatic file storage and printout
options, and impactor- and CPC-efficiency correction factors.

SMPS Accessories (available separately)
379020A		

Rotating Disk Thermodiluter (page 19)

390069		

Data Merge Software (page 5)

3089		

Nanometer Aerosol Sampler (page 29)

3088 		

Soft X-Ray Neutralizer (page 30)

Data Merge model 390069 Software allows you to
merge and fit SMPS and APS data files to create and
display a wide particle size range.

2

NANOSCAN NANOPARTICLE SIZER

Model 3910
Affordable, portable nanoparticle sizer for particles down to 10 nm.
The TSI NanoScan SMPS opens the door to routine nanoparticle size measurements. This
revolutionary sizer fits a TSI SMPS™ Spectrometer into an affordable, portable package that is about
the size of a basketball. Easy to use, lightweight, and battery-powered, NanoScan SMPS enables
investigators to collect valuable nanoparticle size data from more sites, including size distributions
down to 10 nm for concentrations up to 1,000,000 particles/cm3 in selectable SCAN or SINGLE
measurement modes. Derived from TSI core technologies, the NanoScan SMPS is an innovative,
cost-effective solution for real-time nanoparticle size measurements.
Data collection begins with a touch of the instrument display - no need for a dedicated computer to
set up the instrument or save data. The user interface is intuitive and easy for new users to operate.
NanoScan SMPS displays real-time number, surface area, or mass size distributions, concentrations,
and statistics. From the front panel, users can program start time, number of samples, and other
parameters. A full suite of instrument diagnostic data can be viewed from the Setup Screen.
In addition to nanoparticle size distributions, the NanoScan SMPS can collect second-by-second
concentration data at a single mobility diameter. If the nanoparticle source of concern generates 50
nm particles, it is possible to easily monitor 50 nm mobility diameter with 1 second time resolution
to keep a real-time record of concentration levels.
Combine the NanoScan SMPS and the Optical Particle Sizer Model 3330 to measure three orders
of size magnitude from 10 nm to 10 μm using affordable MIM software in concert with portable,
real-time instruments.

3

Screen shot of NanoScan SMPS during nanoparticle size
distribution measurement.

OPTICAL PARTICLE SIZER

Model 3330
Easy-to-use particle sizer for particles 0.3 - 10 μm in size.
Optical Particle Sizer 3330 (OPS) is a light, portable unit that uses single-particle counting
technology to provide fast, accurate measurement of particle concentration from 0 to 3,000
particles/cm3 and particle size distribution for 0.3 – 10 μm in up to 16 channels. Backed by over 40
years of aerosol instrumentation design experience, the OPS uses state-of-the-art optics with 120°
light collection and sophisticated electronics processing resulting in precision, high-quality data.
The affordable, easy-to-use package features a color touch screen with an intuitive user interface.

OPS Model 3330 status screen.

Rigorous factory calibration standards ensure measurement accuracy.
In addition to improving the core measurement fundamentals of OPCs, the Model 3330 includes the
ability to enter the index of refraction and shape factor of the aerosol into the software to provide
more accurate size distributions. The size boundaries can be adjusted using both real and imaginary
components of refractive index. A unique density for every size channel can also be entered into
the software to further improve mass concentration measurements.
The Model 3330 is manufactured at TSI’s ISO 9001 certified facility. It is calibrated using NIST
traceable PSL spheres and TSI’s accredited Electrostatic Classifier and Condensation Particle
Counters. PSL is the industry-wide calibration aerosol of choice because it has properties close to
many real world aerosols and is traceable to national standards throughout the world. Each OPS

Mass size distribution for 0.49 µm ± 9 nm classified PSL spheres.

that leaves the factory is built for longevity, backed by TSI’s commitment to quality, and supported
by our worldwide network of committed TSI professionals.
OPS Accessories (available separately)
3332-10		

10:1 Diluter

3332-100		

100:1 Diluter

8535		

Environmental Enclosure

Environmental Enclosure

4

AERODYNAMIC PARTICLE SIZER® SPECTROMETER

APS software has ad­vanced instrument-control and datahandling features. Above, a graph of aerodynamic diameter is
displayed with particle-size and statistics tables.

Model 3321
The only way to determine a particle’s true airborne
behavior is to measure its aerodynamic diameter.
Aerodynamic measurements account for differences in particle size, shape, and density. This
is crucial when determining if a particle will penetrate a filter, be removed by a cyclone, or be
deposited in the lung. The Aerodynamic Particle Sizer (APS™) spectrometer has been used
successfully for over 20 years in laboratory and field applications to provide high-resolution, realtime aerodynamic measurements in the range from 0.5 to 20 µm. Our latest models also measure
light-scattering intensity in the equivalent optical size range of 0.37 to 20 µm. By providing paired
data for each particle, the APS opens up exciting new possibilities for aerosol scientists interested in
studying the makeup of an aerosol.

Three graphs showing aerodynamic diameter, side scatter, and
correlated data are displayed on this screen simultaneously.

The Model 3321 APS spectrometer uses a patented, double-crest optical system for unmatched

United States Patent Number 5,561,515

sizing accuracy. It also includes a redesigned nozzle configuration and improved signal processing.
The result is greater small-particle sizing efficiency, im­proved accuracy of mass-weighted
distributions, and near elimination of false background counts. The Aerosol Instrument Manager®
software provides advanced data-handling capabilities.
APS Accessories (available separately)
3302A		

Aerosol Diluter (page 6)

3306		

Impactor Inlet (page 6)

3433		

Small-Scale Powder Disperser (page 26)

390069		

Data Merge Software (below)

Data Merge Software
The Model 390069 Data Merge Software module works with our Aerosol Instru­ment Manager
software and simplifies the tedious task of merging Scanning Mobility Parti­cle Sizer™ (SMPS™) and
APS data files. Data sets can be averaged, merged, and fitted to multimodal distributions, and graphs
and tables can be generated easily.

5

AEROSOL DILUTER

IMPACTOR INLET

3302A

Model 3306

Conditions high-concentration aerosols

Collects a size-segregated sample for mass or chemical

for use with the APS spectrometer.

analysis while making APS measurements.

This diluter reduces particle concentrations of high-concentration

An accessory for our Aerodyna­mic Par­ticle Sizer spectro­meter (page 5),

aerosols, providing a representative sample that meets the input

the 3306 combines a single-stage impactor with a filter. It takes a

requirements of the Aerodyna­mic Par­ticle Sizer® (APS™) spectro­meter

size-segregated sample and directs a diluted (80:1), representative portion

(page 5). The 3302A achieves dilution ratios of 100:1 and 20:1 using easy-

of the initial test aerosol into the particle sizer for measurement. The in­let

to-change capillary tubes. Two diluters in a tan­­­dem configuration provide

aero­sol passes through a single-stage impactor (2.5 or 4.7 µm, 50% cut

dilution ratios as high as 10,000:1.

size) and is collected with an after-filter for later mass or chem­ical
analysis. Model 3306 includes two inlet throats: one for standard

Engineered to provide very low particle loss in the 0.5 to 10 µm size

applications; the other for pharmaceutical research. (APS sold separately)

range, the Aerosol Diluter is totally self-contained and requires no outside
power or compressed gas. Durable con­struc­tion and simple maintenance

Accessory (available separately)

procedures translate into years of trouble-free operation. Model 3302A

3033

Vacuum Pump (page 32)

works with both old and new APS spectrometers. (APS sold separately)

6

LASER AEROSOL
SPECTROMETER

Model 3340
Higher-sensitivity, high-resolution general-purpose particle sizer.

Monotonic Intensity Response

TSI’s Laser Aerosol Spectrometer 3340 is a high-sensitivity, high-resolution,

102

general-purpose aerosol particle sizer. This ‘turn on and measure’ instrument

Wide Angle Scattering

Complete size distributions are measured in less than a second by simply
powering up and sampling.
The 3340 is a handy tool to have for monitoring your filters or your processes,
measuring in a lab, or sampling on field campaigns. It’s just as useful as your
primary aerosol tool, but the 3340 can also provide valuable supplemental
information to aerosol measurements based on other sizing techniques
(i.e., SMPS, APS, CPC, FMPS, etc.).

Scattered Signal [a.u.]

allows users to easily measure a size range of 0.09 (90 nm) to 7.5 µm.
100

10-2

Narrow Angle Scattering

10-4

22-4°

10-6

120-35°

10

10

-1

0

Particle Size [μm]

Patented IntraCavity Laser
Sealed
High Reflector

Recessed
Parallel Window

External
High Reflector
Aerosol

Standard
Laser 5mW
>1W Intracavity Power

7

FAST MOBILITY PARTICLE SIZER® SPECTROMETER

Model 3091
Measures size distribution and number concentration of rapidly changing,
submicrometer aerosol particles in real time.
The Fast Mobility Particle Sizer (FMPS™) spectrometer measures particles in the range from 5.6 to
560 nm, offering a total of 32 channels of resolution (16 channels per decade). This submicrometer
particle sizer uses an electrical mobility measurement technique similar to that used in the SMPS
spectrometers. However, instead of a CPC, the Model 3091 FMPS spectrometer uses multiple, lownoise electrometers for particle detection. This produces particle-size-distribution measurements
with 1-second resolution, providing the ability to visualize particle events and changes in particle
size distribution in real time.
The Model 3091 operates at a high flow rate (10 L/min) to minimize diffusion losses of ultrafine and
nanoparticles. It operates at ambient pressure to prevent evaporation of volatile and semivolatile
particles. It requires no consumables. Plus, it uses an efficient, unipolar charger to eliminate the need
for a radioactive neutralizer.
The FMPS is easy to transport, set up, and operate. It can be configured to measure single or multiple
runs continuously for up to 12 hours. Its large, color VGA display and built-in control knob provide
easy access to instrument functions, set-up menus, and data displays. Software highlights include
a variety of graphing options, including 3-D playback of size distribution and concentration versus
time, data export capabilities, and the ability to input individual effective densities per channel to
calculate a continuous output of total particulate mass.
All of these features make the FMPS spectrometer appropriate for a variety of applications,
especially particle formation and growth studies, indoor-air-quality measurements, environmental
research, inhalation toxicology studies, urban canyon studies, and transient emission studies from

FMPS software enables you to view a variety of parameters.
Here, the same data are pictured in a bimodal concentration
histogram and a 3-D size distribution vs. time.

stacks, boilers, and wood burners.

Developed by TSI Incorporated under license from Airel, Ltd. of Tartu, Estonia.
U.S. Patent No. 7,230,431

8

ENGINE EMISSIONS

PARTICLE COUNTERS AND DETECTORS

3775

3783

3772

3787

3788

3772

3776

3777

TSI leads when it comes to instruments for particle counting, aerosol detection, and particle concentration measurement.

ULTRAFINE CONDENSATION
PARTICLE COUNTER

CONDENSATION
PARTICLE COUNTER

Model 3776

Model 3775

Detects particles down to 2.5 nm with extended single

This CPC detects particles down to 4 nm at

particle counting up to 300,000 particles/cm .

concentrations up to 10,000,000 particles/cm3.

The UCPC is designed primarily for researchers interested in airborne

The Model 3775 CPC is a general-purpose counter that accurately

particles smaller than 20 nm. It uses a special sheath-air flow design that

measures concentration over a wide range. It features two detection

improves response time and increases counting efficiency. It features

modes: extended single-particle counting with continuous, live-time

single particle counting with continuous, live-time coincidence correction

coincidence correction and photometric counting for concentrations up

for accurate measurements, an anti-spill design, water-removal capability,

to 10,000,000 particles/cm3. The Model 3775 also features an improved

built-in data logging and storage capabilities, a removable memory card,

transition between the two counting modes, anti-spill design, water-

and USB and Ethernet connectors. This UCPC is ideally suited for

removal capability, data logging and storage capabilities, a removable

atmospheric and climate research, particle formation and growth studies,

memory card, and USB and Ethernet ports. We recommend this CPC for

combustion and engine research, and nanotechnology applications.

basic aerosol research, environmental monitoring, health effects studies,

3

inhalation and exposure studies, combustion research, and more.
Model

3788

3787

3783

3776

3775

3772

3007

Minimum Particle Size (nm, 50% efficiency)

2.5

5

7

2.5

4

10

10

1.3

Upper Concentration Limit (particles/cm3)

4 x 105

2.5 x 107

106

3 x 105

107

104

105

3 x 105

Response Time (sec, 95% response)

~0.25†

~0.25†

<3

<1†

<4†

<3

<9

<5

Aerosol Flow Rate (cm /min)

300

1,000

120

50

300

1,000

100

2,500

n-Butyl Alcohol

Isopropanol

Diethlene
Glycol

Yes

No

Yes

3

Condensing Liquid

Water

SMPS‡ Compatibility

Yes

No

Data Logging Capability

Yes

No

AIM Software
Display Type

Yes*

Yes
Touchscreen w/ Graphics

Vacuum Source
† High-flow mode
‡ Series 3936 Scanning Mobility Particle Sizer spectrometers
* Includes data logging capability and AIM software when used in conjunction with 3772.

9

Yes

Yes*
Color VGA

Internal

3777

LCD

LCD

LCD

External

Internal

External

CONDENSATION
PARTICLE COUNTER

NANO
ENHANCER

Model 3772

Model 3777

A compact, full-featured CPC that

Particle detection down to 1nm.

detects particles down to 10 nm.

The 1nm Nano Enhancer Model 3777 is designed

The Model 3772 uses single particle counting

for researchers interested in the formation

with continuous, live-time coincidence

of aerosol particles, in size ranges below

correction to accurately measure concentrations

conventional condensation particle counters

up to 10,000 particles/cm3. It is ideally suited for

(CPCs). The Model 377 Nano Enhancer enables

applications that do not require measurement

the detection of particle down to 1nm, making

of high concentrations, including basic aerosol

it ideal for a wide variety of studies including

research, filter and air-cleaner testing, particle

particle formation and growth, atmospheric

counter calibration, particle shedding and

and climate aerosols and nanotechnology

component testing, and more. This CPC also

research. The Model 3777 has been optimized

features an anti-spill design, water-removal

for use with TSI’s One-DMA Differential Mobility

capability, built-in data logging and storage,

Analyzer model 3086, creating a Scanning

removable memory card, and USB and Ethernet

Mobility Particle Sizer (SMPS) spectrometer

ports. It requires an external vacuum pump

capable of sizing over the 1nm to 50nm size

(sold separately).

range, with minimal diffusion losses.

EPC™ ENVIRONMENTAL
PARTICLE COUNTER

GENERAL PURPOSE
WATER-BASED CPC

NANO WATER-BASED
CONDENSATION PARTICLE
COUNTER N-WCPC

Model 3783

Model 3787

Model 3788

Real-time, Ultrafine Particulate (UFP)

A fast SMPS™ spectrometer-compatible

World’s fastest CPC – counting

number concentrations.

WCPC that detects down to 5 nanometers.

down to 2.5 nanometers.

Targeted at unattended monitorning, the EPC™

TSI’s Model 3787 General Purpose Water-Based

TSI’s N-WCPC Model 3788 is designed for

monitor provides quantitative UFP

Condensation Particle Counter (GP-WCPC) can

investigators interested in detecting the

concentration data to supplement mass-based

detect airborne particles down to 5 nm in

smallest nanoparticles down to 2.5 nm. This

and FRM equivalent methods. This freshly

diameter using state-of-the-art, water-based

sophisticated particle counter uses state-of-the-

engineered, water-based condensation particle

condensation particle counting technology.

art, water-based condensation technology and

counter detects ultrafine particles down to 7 nm

With a high sample flow rate of 0.6 LPM,

features the highest activation energy and

for single-particle counting up to 10,000,000

the versatile Model 3787 is a good choice for

lowest detectable particle size of the WCPC

particles/cm . The EPC™ monitor provides

low concentration measurements, yet it is also

family. With less than 0.1 second rise time, the

robust field perfomance and requires minimal

capable of measuring up to 250,000 particles/cm3 Model 3788 is the fastest CPC commercially

maintenance, making it a good choice for indoor

using exclusively single particle counting. The

available and its high sample flow rate provides

and outdoor ambient monitoring as well as

carefully designed flow path, high flow rate, and

low diffusion losses and low Poisson noise.

many other applications that require 24/7

sophisticated electronics used in the Model 3787

Other important elements include a

operation.

result in a rise time of < ⅓ second.

thermodynamically optimized growth region

3

and enhanced optical and detection design for
CPC Accessories (available separately)

impressive signal to noise ratios. The Model

376060

3788 can be used as a stand-alone counter

Particle Size Selector (page 31)

376061	
Additional diffusion screens for
Particle Size Selector (Qty. 12)
3032	Vacuum Pump for 3772 and 3783

measuring up to 400,000 particles/cm3 (using
single particle counting) or as a component of a
TSI SMPS™ spectrometer.

CPCs (page 32)

10

HANDHELD
CONDENSATION
PARTICLE COUNTER

NANOPARTICLE SURFACE
AREA MONITOR

AEROSOL ELECTROMETER

Model 3007

Model 3550*

Model 3068B

Portable, battery-powered CPC that

Measures lung-deposited surface

Measures net charge or concentration

detects particles down to 10 nm.

area of inhaled particles.

of aerosol.

The Model 3007 is a hand-held CPC intended for

Recent research has shown that the surface

Our Aerosol Electrometer has been updated

measuring ultrafine particles in a wide variety

area of nanoparticles is highly correlated

to make use of new technologies that were

of applications. Its small size and ergonomic

to exposure-related adverse health effects.

developed for our Electrical Aerosol Detector and

design make it the best choice for short-term

The Model 3550 indicates the surface area

Engine Exhaust Particle Sizer™ spectrometer.

outdoor and indoor-air-quality monitoring,

of nanoparticle aerosols that deposit in the

It measures the total net charge on aerosol

nanoparticle work area surveys, and mobile

lung. The reported measurements correspond

particles from 0.002 to 5 µm at a user-selectable

aerosol research. This highly portable CPC

to the ICRP lung deposition curves for the

flow rate from 1 to 10 L/min. When paired with

weighs only 1.7 kg (3.8 pounds)!

tracheobronchial (TB) and alveolar (A) regions

a TSI Electrostatic Classifier, the electrometer

of the human respiratory tract. The Model 3550

measures the number concentration of

measures particles in the size range of 10 to

monodisperse aerosol. This configuration is used

1,000 nm. This is an important instrument for

primarily for calibrating and testing particle

research in the fields of inhalation toxicology,

instruments like CPCs. Data are presented on

health effects, and epidemiology, and for

the front-panel display in real time and also as

measuring and monitoring workplace exposure.

an analog voltage output for data recording.

Model 3550 offers user-selectable measurement

The electrometer is compatible with the Aerosol

modes, comprehensive software, continuous

Instru­ment Mana­ger® software. An external

operation, a wide dynamic range, and high

vacuum pump is required (sold separately).

time resolution.
*U.S. Patents 6,544,484 and 7,812,306

11

DUSTTRAK™ II & DRX AEROSOL MONITORS

SIDEPAK™ PERSONAL AEROSOL MONITOR

Models 8530, 8532, 8533 & 8534

Model AM510

Simply leaves everyone else in the dust.

A personal, battery-operated laser photometer

The DustTrak™ II and DRX Aerosol Monitors are battery-operated, data-

that measures and records worker exposure to dust.

logging, light-scattering laser photometers that give you real-time aerosol

The SidePak™ Personal Aerosol Monitor is a rugged, light­weight, belt-

mass readings. These monitors measure aerosol contaminants such as

mounted laser photometer that weighs as little as 454 grams (16 ounces).

dust, smoke, fumes and mists. They use a sheath air system that isolates

A built-in sampling pump allows the use of a variety of size-selective in­let

the aerosol in the optics chamber to keep the optics clean for improved

conditioners. Model AM510 includes TSI’s Smart Bat­tery Management

reliability and low maintenance. Suitable for clean office settings as well

System™, which includes long-running NiMH or alkaline battery packs and

as harsh industrial workplaces, construction and environmental sites, and

indicates run time in minutes re­maining. An easy-to-read display shows

other outdoor applications.

real-time aerosol mass concentration and 8-hour, time-weighted average
(TWA). Data logging and long battery life make the AM510 ideal for work

Ideal applications include industrial/occupational hygiene surveys, indoor-

shift or extended sampling.

air-quality investigations, outdoor environmental monitoring, fugitive
emissions monitoring, site perimeter monitoring, dust control operations,
baseline trending and screening, engineering control evaluations, point
source monitoring, engineering studies, remote monitoring, and
process monitoring.

SidePak AM510
Particle Size Range

DustTrak II
Model 8530

DustTrak II
Model 8532

DustTrak DRX
Model 8533

0.1 to 10 µm

Aerosol Concentration
Range

0.001 to 20 mg/m3

Measurement Output

Single Mass Concentration
(PM10, Respirable,
PM2.5, PM1)

Gravimetric Reference
Filter

No

Data Logging

31,000 data points,
21.5 days at 1 min

Type

Personal

0.001 to 400 mg/m3

0.1 to ~15 µm
0.001 to 400 mg/m3

Single Mass Concentration
(PM10, Respirable, PM2.5, or PM1)
Yes

DustTrak DRX
Model 8534

No

0.001 to 150 mg/m3

0.001 to 150 mg/m3

Simultaneous Size-Segregated Mass Fractions for PM1,
PM2.5, Respirable, PM10 and Total. All Displayed.
Yes

No

5 MB of on-board memory (>60,000 data points); 45 days at 1 minute logging interval.
Desktop

Handheld

Desktop

Handheld

12

ENGINE EMISSIONS

AIR FILTER & RESPIRATOR TESTERS

3160

8127/8130

Our automated air filter and respirator testers are known for simple, reliable operation.
Component systems are also available.
8127a
Measurement Application

Loading and Quality Control Tests

Maximum Efficiency
Aerosol Type

b

99.999%
DOP, PAO, DEHS, Paraffin,
and other Oils

Aerosol Generation

3160
Penetration vs. Particle Size
99.999999%

DOP, PAO, DEHS, Paraffin,
and other Oils or NaCl

Atomizer

DOP, PAO, DEHS, and
other Oils or NaCl
Atomizer with Classifier

Count Median Diameterc

0.2 μm (Oil)

0.2 μm (Oil) or
0.075 μm (NaCl)

N/A

Geometric Standard
Deviationc

<1.6 (Oil)

<1.6 (Oil) or < 1.86 (NaCl)

<1.3

Flow Rate

15 to 100 L/min

Resistance

5 to 100 L/min

0-150 mm H2O (0-1470 Pa)
Condensation Particle
Counter

Particle Detection

Light Scattering Photometer

Typical Test Lengthd

10 sec

30 sec to 20 min

Data Reporting

Alpha-Numeric Display and RS-232

PC with Integrated
Software

Operation

Stand Alone Tester/Automated Production Lines

Stand Alone Tester

Compliance

US 42 CFR part 84, EN 143, JMOL, ISO 23328-1

EN 1822 parts 3 and 5

a) EN versions (for equivalent results to EN 143
standard) available (8127-1-EN, 8130-1-EN)
b) Aerosol abbreviations: DOP (dioctyl phthalate), PAO
(polyalpha olefin), DEHS (di-ethylhexyl sebacate)

13

8130a

c) EN version CMD and GSD are different.
See 8127/8130 spec sheet for more info
d) Efficiencies higher than 99.9999%
require longer than typical testing times

AUTOMATED FILTER TESTER

AUTOMATED FILTER TESTER

Models 8130 & 8130-EN

Model 3160

Filter testers for commercial respirator

Determines penetration vs. particle size of filters

and military mask testing.

and filter media.

Model 8130 is designed for simple, fast, automated operation. That is why

Model 3160 is the most advanced automated tester available for

it excels in quality-control and production-testing applications. It can also

challenging filters and filter media with submicrometer aerosols. It can be

be used for certification testing of all respirator filter categories in

used to test both low- and high-efficiency filters and filter media, up to

accordance with US 42 CFR part 84, as well as for other standards around

99.999999% efficient (eight 9’s), or penetrations down to 0.000001%.

the globe. For customers who need to test according to European standard

The 3160 uses a bank of atomizers and the TSI Electrostatic Classifier

EN 143, TSI now offers the Model 8130-EN tester.

to challenge a filter or filter media with size-controlled, monodisperse
particles. Two Condensation Particle Counters (CPCs) simultaneously

Models 8130 and 8130-EN can measure efficiencies up to 99.999%, or

count the upstream and downstream particles and computer software

penetrations as low as 0.001%, using either polydisperse oil or sodium

calculates the penetration value. Filters can be sequentially challenged

chloride aerosol. Both testers produce oil aerosol without heat, so a

with up to 20 different monodisperse particle sizes from 15 to 800 nm.

variety of oils can be used. Dual light-scattering photometers provide

The penetration value for each particle size is calculated. At the end of

simultaneous upstream and downstream particle concentration

a test, the 3160 generates a curve of penetration vs. particle size and

measurements to ensure fast, repeatable results. A touch-panel display

produces a summary of test results, including the most penetrating

makes these testers extremely easy to use, so very little operator

particle size (MPPS). Test results can be automatically saved in a

training is needed. On-screen menus enable operators to easily change

Microsoft® Access® data base and downloaded into Microsoft® Excel®.

test parameters to accommodate special requirements. An internal
microprocessor controls the tester and calculates and displays filter

Model 3160 complies with EN 1822-3 and provides the most complete

penetration, resistance value, and flow rate. Additionally, it performs self-

information on filter penetration available from any filter tester. You’ll find

diagnostic checks during operation and automatically zeros the

it invaluable for product development and quality control.

tester between measurements to ensure accurate, stable,
reproducible results.

AUTOMATED FILTER TESTER

Models 8127 & 8127-EN
Oil-only filter testers for commercial respirator
and military mask testing.
Model 8127 includes all of the features of our Model 8130, except it is an
oil-only tester. It can be used for quality control and certification testing in
accordance with US 42 CFR part 84, as well as for other standards around
the globe. For customers who need to test according to European standard
EN 143, TSI now offers the Model 8127-EN tester.
Models 8127 and 8127-EN can measure efficiencies up to 99.999%,
or penetrations as low as 0.001%, using polydisperse oil aerosol. Both
testers produce oil aerosol without heat, so a variety of oils can be used.
Dual light-scattering photometers provide simultaneous upstream and
downstream particle concentration measurements to ensure accurate,
fast, repeatable results.

14

COMPONENT FILTER-TESTING SYSTEMS

Model 3150
Off-the-shelf solutions to your custom air filter testing needs.
For the wide range of flow rates and particle sizes needed to meet the
requirements of the many filter-testing standards and research needs, an
automated tester is not always practical. Building a filter test system from
components is often the best way to satisfy your measurement needs.
TSI has developed the CFTS as an easy-to-use system to integrate all the
necessary parts of this type of system.
The Component Filter Test System (CFTS) Model 3150 from TSI consists
of software and a hardware module to provide an off-the-shelf solution
for all your custom filter-testing needs. Pre-configured to work with
TSI’s world-class detectors and sizers, the CFTS provides a platform for
numerous filter test applications.
Because filters are used for a wide variety of applications and are tested
to many different filter test standards, you need a filter-testing system
that is flexible enough to change to meet these various standards. The
CFTS system is designed for this flexibility. It measures flow using a
variety of techniques and controls blowers to achieve the required flow.
It has multiple ports used to read temperature, pressure, and relative
humidity, then make flow corrections and log sensor data for test reports.
Whether designing a new test duct or updating an existing system, the
CFTS provides an easy-to-use system with the flexibility to meet your
filter-testing needs.
CFTS is structured as a core platform which controls the filter test. Its
graphical user interface is used to define the test layout and procedures as
well as being your interface for running the test. The CFTS has drivers to
communicate with the particle instruments and is the interface for
reading sensors and controlling flow. During testing, it collects, saves, and
exports the data needed for test reports. No longer is custom software
required for every component change.

15

CFTS software screen shot.

LASER PHOTOMETER

LARGE-PARTICLE
AEROSOL GENERATOR

RESPIRATOR
LEAK TESTER

Model 8587A

Model 8108

Model 8119

A compact, reliable photometer for

Generates high-concentration aerosols

Turns Models 8127 and 8130 into

customized filter testing.

up to 10 µm in diameter.

respirator leak testers.

This photometer features a reliable laser diode

The Large-Particle Aerosol Generator produces

This accessory enables Models 8127 and 8130

that produces constant laser power, so aerosol

highly concentrated aerosol over a wide

to perform fast, easy, after-maintenance leak

concentration measurements remain stable

particle-size range, from 0.1 to 10 µm. Designed

tests on full-face and half-mask air-purifying

over a long period of time. A sheath-air design

for easy cleaning, this generator produces

respirators. After installing the 8119, simply

keeps the optics clean for low background

potassium chloride (KCl) and other aerosols.

mount the respirator on the headform and scan

levels and minimal maintenance requirements.

Model 8108 meets the requirements of ASHRAE

the respirator with the hand-held probe. Test

The 8587A uses an internal switching valve to

52.2 for KCl test aerosol and can also be used

critical areas like eye-lens seals and exhaust

measure both the upstream and downstream

for ISO/TS 11155-1:2001 filter efficiency testing.

valves. Audible and visual alarms are triggered

aerosol concentration. A special high-speed

It is an ideal aerosol generator for fractional

if the aerosol concentration inside the mask

“purge” mode shortens the purge time when

efficiency testing of general ventilation filters

exceeds a preset level.

switching between upstream and downstream

and automotive cabin-air filters.

measurements. A simple command set can be
incorporated into your LabVIEW® program to
give you complete flexibility in test protocol and
database management. All of these features
combine to make the 8587A ideally suited for
custom filter testing applications.

16

ENGINE EMISSIONS

ENGINE EMISSIONS
ENGINE EXHAUST PARTICLE SIZER™ SPECTROMETER

Model 3090

The EEPS operates at ambient pressure to

EEPS software allows users to display

The best tool for measuring transient

prevent evaporation of volatile and semivolatile

measurements in a variety of graphical and

particle emissions and characterizing

particles, requires no consumables, and uses

tabular formats, including 3-D viewing of size

exhaust after-treatment devices.

an efficient, unipolar charger to eliminate the

distribution and concentration versus time.

The Engine Exhaust Particle Sizer (EEPS™)

need for a radioactive neutralizer. Users can

These can be replayed for a unique “movie” view

spectrometer measures the size distribution

select from multiple matrices tailored to specific

of the entire en­gine cycle, or you can zoom in

of engine particle emissions in the range from

aerosols for more accurate measurement.

on a period of interest. The software in­cludes a
data export capability and allows users to input

5.6 to 560 nm with the fastest time resolution
available (10 times per second!). Users can

Ease of operation is a key feature of this

individual effective densities per particle size

visualize and study the dynamic behavior of

instrument. All components, including the

channel to calculate a continuous output of total

emissions that occur during transient test

vacuum source, are housed in a single cabinet

particulate mass.

cycles, such as changes in engine speed, torque,

that weighs just 32 kg (~70 lbs). Just turn on the

or load. They may also measure emissions that

power and allow the instrument to warm up.

EEPS Accessories (available separately)

occur during the first few seconds of a cold start

A microprocessor corrects for volumetric flow

379020A-30

Rotating Disk Diluter with

or during regeneration of a particle trap or diesel

and barometric pressure automatically. This

		

Thermal Conditioner Air

particulate filter (DPF).

will maintain calibration and provide accurate

		

Supply (page 19)

particle size distribution information. The EEPS
Measurements are displayed with high size

also features an external “start” input trigger

resolution (32 total channels, 16 channels per

for remote operation, two analog inputs to log

decade). The EEPS spectrometer operates over

and correlate other engine parameters, and four

a wide particle concentration range, which

user-configurable analog outputs to integrate

makes it well-suited for measuring upstream

emission measurements with the test cell

and downstream of a particle trap or DPF to

host computer.

determine soot loading and removal efficiency.

17

The EEPS spectrometer was developed by TSI Incorporated
under license from Airel, Ltd. of Tartu, Estonia. Additional
assistance was provided by the University of Minnesota Center
for Diesel Research.
U.S. Patent No. 7,230,431

ENGINE EXHAUST CONDENSATION
PARTICLE COUNTER

Figure 1: Proposed Regulation 83 particle sampling and measurement.

EECPC Accessoies (sold separately)
3032		

Vacuum Pump, 115 VAC (page 32)

3032-EC		

Vacuum Pump, Europe only (page 32)

Model 3790A

3032-1		

Vacuum Pump, 230 VAC (page 32)

The particle number (PN) concentration benchmark

379020A-30

Rotating Disk Diluter with Thermal Conditioner Air

for ECE Regulations 83 and 49.

		

Supply (page 19)

The Engine Exhaust Condensation Particle Counter (EECPC) accurately
measures PN concentration of exhaust emissions. In fact, the
GRPE Particle Measurement Programme (PMP) concluded that PN
measurements using a CPC plus thermodilution are 20 times more
sensitive and much less variable than the traditional method (i.e.,
gravimetric filter analysis). As a result, the measurement of solid PN

ELECTRICAL AEROSOL DETECTOR

emissions has been included in Regulation 83 (Euro 5) for certification of
new passenger vehicles with diesel engines, and later for Regulation 49
(EURO 6) for heavy-duty engines.
Model 3790 EECPC is fully compliant for light-duty vehicle certification in
accordance with all Regulation 83 requirements. The Model 3790 EECPC
incorporates a wide assortment of design improvements and features
such as anti-spill design, condensate removal, removable saturator for
ease of maintenance, built-in microprocessor with USB, RS-232 and
Ethernet communication interfaces, touch-panel membrane keys and a
display that enables instrument set-up, viewing particle concentration
and count data, interrogating instrument status, and data storage
capabilities. The EECPC includes our Aerosol Instrument Manager®
software. An external vacuum pump is required – sold separately.
ECE Regulations 83 and 49 mandate that only the number concentration
of solid particles are measured. TSI offers the Rotating Disk Diluter with
Thermal Conditioner Air Supply to eliminate or suppress the formation
of volatile compounds so that the EECPC only counts solid particles.
Together, the Models 3790 and 379020A-30 fulfill the proposed solid PN
sampling and measurement requirements, as shown in Figure 1.

Model 3070A
A simple, low-cost tool for measuring
particle emissions second-by-second.
The Electrical Aerosol Detector (EAD) is a robust instrument that provides
highly repeatable, real-time measurements in the range from 10 to
1000 nm. It operates over a wide concentration range and requires no
working fluid or consumables. When combined with a TSI Rotating Disk
Thermodiluter, the EAD becomes a real-time solid PN detector. These
features make the EAD ideally suited to evaluate particle emissions during
transient engine operation, to characterize efficiency of exhaust aftertreatment devices, and for on-board emissions testing.

18

ROTATING DISK THERMODILUTERS AND
THERMAL CONDITIONER AIR SUPPLY

Models 379020A & 379030

Thermodiluters

Dilute and condition combustion sources to preserve the

379020A		

Rotating Disk Thermodiluter

original particle size distribution and number concentration

379020A-30

Rotating Disk Diluter with Thermal Conditioner

for accurate measurements.

		Air Supply

TSI Model 379020A Rotating Disk Thermodiluter is highly regarded in the
field of particle emission measurement, especially for sampling, diluting,

Thermodiluter Accessories (sold separately)

and conditioning exhaust emissions from diesel and spark-ignition

379030		

Thermal Conditioner Air Supply

engines, as well as for performing stack emission studies.

1137021

Hardware to run 3090 with 379020A

379025		

Housing for Diluter Head

The Model 379020A features a separate diluter head and control unit,

379027		

Cyclone (PM2.5)

which allows the sample to be diluted and thermally conditioned at the

379032A		

Digital Control Unit

point of measurement (i.e., tailpipe, CVS tunnel, or stack). It also features a
variable dilution ratio that is adjustable from 15:1 to 3000:1 and selectable
heated diluter temperatures up to 150°C to avoid measurement of
condensed volatile materials.
In order to measure only the solid particle emissions, it is necessary to
further thermally condition the sample to eliminate the semi-volatile
and volatile fractions. TSI offers the Model 379030 Thermal Conditioner
Air Supply that can be combined with the Model 379020A. It uses an
evaporation tube to heat the sample up to 400ºC, which effectively
eliminates volatile compounds that may have formed in the exhaust as it
cools or becomes diluted. Rotating Disk Thermodiluters and accessories
can be used with all of our submicrometer particle sizers, especially the
EEPS™ and SMPS™ spectrometers, all of our CPCs,
and the EAD.

19

These instruments are produced in Switzerland by Matter Aerosol.

NANOPARTICLE EMISSION TESTER

General Purpose Test Mode Display Output: Live strip
chart of total solid particle number concentration with
1 hz frequency and cumulative statistics.

Model 3795

A Sophisticated Instrument in a Simple Package

Portable, accurate and Regulatory-compliant.

The Nanoparticle Emissions Tester combines a traditionally complex

The TSI Nanoparticle Emission Tester (NPET) Model 3795 is a

system of components into a portable, robust measurement tool. The five

portable, accurate instrument capable of measuring total solid

key design components include:

particle number emissions from a variety of diesel-powered
sources, including construction machinery, buses, stationary

a. Sampling probe

gensets, and more. Featuring a robust, user-friendly design,
the NPET can be used in the field by researchers, regulatory

b. Preconditioner

inspectors, and maintenance personnel alike.
c. Recirculating dilution flow conditioner: A silica desiccant dryer and
Official Certification Testing

two high capacity HEPA filters.

The Nanoparticle Emissions Tester fully meets Swiss Regulation 941.242 for
the periodic certification of diesel-powered machinery equipped with a DPF.

d. Volatile particle remover: A catalytic stripper heated to 350° C
evaporates, and oxidizes volatile components.

Solid Particle Measurements
e. Particle counter: An isopropanol-based CPC provides accurate, direct

Sampling from combustion sources is often challenging due to the
presence of volatile material. Volatile components are extremely sensitive

measurement of particle number emissions.

to sampling conditions and can grow existing particles and form new
particles through condensation. By evaporating and oxidizing volatile
components and particles, the NPET Model 3795 measures only the
remaining solid particles.

Inlet

Total
Flow

10:1
Diluter

a

1 μm Cyclone &
Water Trap

d

Sample
Flow

e

Dilution Flow

Orifice ) (

Filter

Filter

c

Pump

Filter

Filter

b

Condensation
Particle Counter

Catalytic
Stripper

Bypass Flow

Pump

Dryer

20

ENGINE EMISSIONS

AIR MONITORING
ULTRAFINE PARTICLE MONITOR

ENVIRONMENTAL SAMPLING SYSTEM

Model 3031

Model 3031200

Provides continuous size distribution and number

Provides representative sampling and proper conditioning of

concentration data for long-term air quality monitoring.

ambient submicrometer aerosol for accurate size distribution

The Model 3031 Ultrafine Particle (UFP) Monitor has been specifically

and particle number concentrations.

designed for long-term air-quality monitoring networks. It operates

The Model 3031200 Environmental Sampling System is an accessory

continuously, 24-hours a day, for months on end, with minimal

for use with a wide variety of TSI particle sizers and counters, including

maintenance and requires no working fluids. The UFP Monitor has no

the Ultrafine Particle Monitor, Scanning Mobility Particle Sizer™

radioactive source, so there are no special licensing requirements and no

spectrometers, and Condensation Particle Counters. The Model 3031200

restrictions on its use or where it can be installed. The Model 3031 fits

consists of a standard PM10 inlet, a sharp-cut PM1 cyclone, a flow splitter,

into a standard 19-inch, rack-mount cabinet, which allows it to be easily

and a Nafion® dryer. Combine these components with your choice of

installed into existing roadside and urban-air-quality monitoring stations.

appropriate-length sampling tubes and vacuum source for easy setup in

A bench-top version of this instrument, the Model 3031-1, is available for

the field.

laboratory or mobile applications.

21

PARTICLE SIZERS & IMAGING SYSTEMS

PHASE DOPPLER PARTICLE ANAYLZER

GLOBAL SIZING VELOCIMETER

Model PDPA

Model GSV

A noninvasive instrument that measures particle,

For simultaneous size and velocity measurements.

droplet, or bubble size.

GSV is an imaging-based system that provides simultaneous

Phase Doppler technique allows for the sizing of spherical particles

measurements of the size and velocity of individual droplets, bubbles,

(typically spray droplets, bubbles, and hollow or solid spheres). Along with

or spherical particles in an area. It uses a laser light sheet to illuminate

size information, this technique obtains particle velocity. Typical limits for

a region of the particle field. Imaging-based measurement methods

common configurations are 0.5 µm on the lower end and 5 mm on

feature the significant advantage of providing thousands of individual

the upper end.

droplet or particle size/velocity measurements from a single image. The
simultaneous nature of the measurements makes advanced analysis

The Phase Doppler Particle Analyzer (PDPA) uses the principles of light-

(such as size and velocity correlations) simple, making it an attractive

scattering interferometry. It requires no calibration because particle size

measurement solution for a variety of industrial and research applications.

and velocity are dependent only on the laser wavelength and optical
configuration. PDPA technique does not depend upon scattered-light
intensity and is not subject to beam attenuation or deflection that occurs
in dense particle and combustion environments. Consequently, the
technique works well in dense spray and combustion measurements.
PDPA systems are used in a wide variety of disciplines around the world,
including combustion research, aircraft icing research, inkjet printer
development, industrial sprays research, spray drying, characterization of
metered-dose inhalers, droplet size and velocity measurements of medical
nasal sprays, and characterization of paint sprays.

22

AEROSOL GENERATORS AND DISPERSERS

3450

3082L

Collectively, our generators and dispersers produce particles in the range of 0.002 to 200 micrometers.
Monodisperse Generators

Powder Generators

Polydisperse Generators

Model

3450

Series
3080

Particle size
range (µm)

1 to 200†

0.002 to
1.0

0.1 to 8.0

0.002 to
> 0.1

0.5 to 40‡

1.0 to 50‡

Particle
concentration

<103/cm3

<105/
cm3

>106/cm3

>107/cm3

10 to 100
mg/m3

0.3 to 40§
mg/m3

>107/cm3

>108/cm3

>107/cm3

>107/cm3

>107 cm3

104/cm3

Nominal flow
rate (L/min)

30 to 80**

≤2

3.5 to 4

0.25 to 3

5 to 15

5

3.0 to 3.5

1.0 to 4.2

6.5

6.5 to 39††

30/1000

140

3475

3482

3400A

3433*

3076*

3079A*

9302*

9306*

9307/
9307-6

0.01 to 2.0 (nominal 0.3-µm count mean diameter)

*May be used to disperse monodisperse solid particles such as PSL. †Large sizes require optional orifices. ‡Aerodynamic diameter. §High concentrations require optional high-speed motor.
**Includes dilution air; nominal dispersion air flow rate is 1.5 L/min. ††Flow rate at 25-psig inlet pressure; higher pressures and flow rates possible

23

8108
0.1 to
10.0

VIBRATING ORIFICE AEROSOL GENERATOR

ELECTROSTATIC CLASSIFIERS

Model 3450

Model 3082

Creates uniform, mono­disperse particles in the

Primary-standard instruments that produce highly

range of 1 to 200 µm with unmatched accuracy.

monodisperse, submicrometer aerosols.

The Vi­brat­ing Orifice Aer­o­­sol Generator (VOAG) is a highly accurate source

The Series 3082 Electrostatic Classifiers are primary-standard aerosol

of mo­no­disperse particles in the range of 1 to 200 µm. Us­ing a variety

instruments that give you highly monodisperse, submicrometer aerosol

of solutes and solvents, the VOAG cre­ates solid or liquid aerosol particles

from a polydisperse source. Our classifiers have been used in a variety of

uniform in size, shape, density, and surface characteristics.The VOAG

aerosol-generation or particle-sizing applications with highly

produces uniform particles by con­trol­­ling the breakup of a liquid jet. It

repeatable results.

delivers a con­sistent volume of liquid us­ing a constant-flow sy­ringe pump.
Its consistency and ac­curacy make this aerosol gen­erator an effective

Electrostatic Classifiers neutralize polydisperse aerosol and use a

choice for applications like basic aero­sol re­search, in­stru­ment calibration,

differential mobility analyzer (DMA) to classify and strip out a narrow,

and product design and development. Stan­dard orifices are included for

predictable size. TSI provides a choice of three DMA col­umns. You can

small particle sizes; additional ori­fices are available for producing

purchase any column and interchange them on the same platform,

larger particles.

giving you unprecedented versatility. The platform is available
separately for use with your own DMA.

VOAG Accessories (available separately)
3054		

Aerosol Neutralizer (page 30)

Particles produced with our Electro­static Classifiers range in size from

3074B		

Filtered Air Supply (page 31)

0.001 to 1.0 µm. For monodisperse aerosol generation, simply set the

393520		

10-µm orifice (one included with instrument)

software for the desired particle size.

393530		

20-µm orifice (two included with instrument)

393540		

35-µm orifice

Electrostatic Classifiers

393550		

50-µm orifice

3082		Classifier Platform with out Aerosol Neutralizer,

393560		

100-µm orifice

393590		

0.5-µm filters, 13-mm diameter (quantity of 100,

3081A		

Long DMA (0.01 to 1 µm, included with 3080L)

		

one set included with instrument)

3085A		

Nano DMA (0.002 to 0.15 µm, included

but no DMA or Impactor Inlet

		with 3080N)
The Model 3450 VOAG is part of the Model 3941 Supermicrometer Mono­dis­perse Aerosol
Generation System described on page 28.

Electrostatic Classifiers are included in our SMPS™ systems. Model 3080L is part of the Model 3940
Submi­cro­meter Monodisperse Aerosol Generation System. DMA columns are interchangeable.
For restrictions, consult local authorities on the use of Aerosol Neutralizers. The Nano DMA was
developed in cooperation with the Uni­ver­sity of Minnesota Particle Technology Laboratory and
Gerhard Mercator Universi­ty. Refer to United States Patent Number 6,230,572.
A micro-orifice inside the generation
head vibrates during operation, producing
highly uniform particles in solid or liquid
form. (photo courtesy of Prof. Richard
K. Chang)

DMAs contain highly polished surfaces for high-resolution par­ticle size classification (left). The
Series 3080 offers a choice of two DMAs (right).

24

CONDENSATION MONODISPERSE
AEROSOL GENERATOR

ELECTROSPRAY AEROSOL GENERATOR

Model 3475

Model 3482

Generates high-concentration, monodisperse

Produces monodisperse particles as small as 2 nanometers.

aerosols quickly and accurately.

The Electrospray Aerosol Generator (EAG) uses a patented technique to

The Condensation Monodis­perse Aerosol Gen­erator (CMAG) is a

produce high concentrations of monodisperse, submicrometer particles in

condensation-type instrument that produces high-concentration,

the range from 2 to >100 nm in diameter. The EAG produces such small,

monodisperse aerosol particles. It is well-suited for challenging HEPA

uniform particles by pushing a charged liquid solution or suspension

and ULPA filters, seeding wind tunnels, conducting inhalation

through a capillary tube and exerting an electrical field on the liquid at

studies, or other applications requiring mono­disperse particles

the capillary tip. The electrical field pulls the liquid from the capillary,

in high concentrations.

forming individual droplets. Air and CO2 mixed with the droplets evaporate
the liquid, and the remaining particles are neutralized by an ionizer. The

The CMAG generates liquid or solid particles from a variety of oils, waxes,

result is a neutralized, mono­disperse aerosol. Applications for the EAG

and other materials, in concentrations great­er than 10 particles/cm .

include instrument calibration, nano­meter-sized powder dispersion,

It generates monodisperse particles in the range of 0.1 to 8 µm and

macromolecular analysis, and nano-aerosol studies.

6

3

operates at a flow rate of 3.5 to 4 L/min. Particles can be fluorescently
or radioactively labeled. A coil heater inside the saturator and a

EAG Accessories

condensation-nuclei bypass filter provide rapid re­sponse when changing

3482-spump

Syringe Pump

aerosol size. The CMAG can operate for long periods without interruption.

3982-Auto(C)

Autosampler and HPLC Pump

Aerosol may be monitored for size and concentration using the optional

3982-autopump

Standalone HPLC Pump

Pro­cess Aerosol Monitor.
CMAG Accessories (available separately)
3375		

Process Aerosol Monitor (page 32)

Please specify voltage requirements. Models 3475 and 3375 are produced in Germany by
Topas GmbH.

1

2

3

The Electrospray produces particles down to 2 nm. A viewport allows you to watch the capillary
tip during operation: (1) no liquid flow, (2) with liquid flow but no electrical field, and (3) with liquid
flow and an electrical field. The third view illustrates stable operation.

25

SMALL-SCALE POWDER DISPERSER

FLUIDIZED BED AEROSOL GENERATOR

Model 3433

Model 3400A

Disperses small quantities of powder

Disperses powders in stable concentrations

for powder sizing applications.

for dust experiments or particle seeding.

The Small-Scale Powder Disperser (SSPD) aerosolizes very small

The Fluidized Bed Aerosol Generator (FBAG) is our general-purpose

quantities of powder by lifting particles from a turn­table using a venturi

powder disperser. It prepares any dry, free-flowing powder for dispersion

aspirator. The turntable rotates at variable speeds, controlling aerosol

in a gas. It disperses powders that range from 0.5 to 40 µm, with

concentrations from 0.3 to 4.0 mg/m3 (up to 40 mg/m3 with optional

concentrations from 10 to 100 mg/m3. Un­surpassed constant output and

high-speed motor). Shear forces created in the SSPD are sufficient to

con­centration make the FBAG useful for inhalation toxicology studies,

deagglomerate most dry particles in the size range of 1.0 to 50 µm.

laser-velocimeter seeding, and filter testing.

A selection of optional turn­tables is available to meet special needs,

FBAG Accessories (available separately)

in­cluding one designed to redisperse airborne material collected on 25

3012		

Aerosol Neutralizer (page 30)

and 35-mm membrane filters used during environmental or exposure

3074B		

Filtered Air Supply (page 31)

monitoring. Because it disperses only a small amount of sample, the SSPD

1502574		

Replacement bronze beads

is great for studies involving rare, expensive, or hazardous materials. It
also effectively disperses polystyrene latex (PSL) spheres used to calibrate
measurement instruments. If you want to measure the size distribution
of powders, the SSPD can be paired with our Aerodynamic Particle Sizer®
spectrometer (page 5).
SSPD Accessories (available separately)
3074B		

Filtered Air Supply (page 31)

1030737		

Standard turntable (one included with instrument)

1030770		

Turntable preloaded with 5-, 7-, 10-, 15-, 20-,

		

and 30-µm PSL particles

1030771		

Membrane-filter turntable

1030772		

V-groove turntable

1030779		

High-speed motor for output up to 40 mg/m3

A variety of turntables provides flexi­bility when
using the SSPD.

A bed of bronze beads in­side the Model
3400A breaks up powder agglomerates
before dispersion.

26

ATOMIZERS

LARGE-PARTICLE AEROSOL GENERATOR

Models 3076, 3079A, 9302 & 9306

Model 8108

Generate polydisperse, high-concentration aerosols.

Generates high-concentration aerosols

TSI offers four highly dependable instruments for nebulizing a liquid

up to 10 µm in diameter.

solution or suspension:

The Large-Particle Aerosol Generator produces highly concentrated

Constant Output Atomizer Model 3076. Generates aero­sols of constant

aerosol over a wide particle-size range, from 0.1 to 10 µm. Designed for

particle size in concentrations over 10 particles/cm (nominal). Its

easy cleaning, this generator produces potassium chloride (KCl) and other

nominal aerosol flow rate is 3.0 to 3.5 L/min. Stainless-steel components

aerosols. Model 8108 meets the requirements of ASHRAE 52.2 and can

make this collison-type atomizer suitable for biological and medical

also be used for ISO/TS 11155-1:2001 filter efficiency testing. It’s an ideal

research, material synthesis, filter testing, instrument calibration, and

aerosol generator for fractional efficiency testing of general ventilation

basic research.

filters and automotive cabin-air filters.

7

3

Portable Atomizer Model 3079A. This rugged, compact atomizer
generates particles in concentrations over 108 particles/cm3 and offers
an adjustable flow rate from 1.0 to 4.2 L/min. A built-in, low-noise

OIL DROPLET GENERATOR

compressor provides compressed air, and the atomizer head is made
entirely of stainless steel. Operating components are protected by a hood,
making this atomizer highly portable and suitable for acceptance tests.
Single-Jet Atomizer Model 9302. Our simplest atomizer includes a builtin pressure regulator for controlling air from an external source.
It produces particles in concentrations over 107 particles/cm3
at a nominal flow rate of 6.5 L/min.
Six-Jet Atomizer Model 9306. Features the highest flow rate of any TSI
atomizer and a built-in dilution system. Users may select up to six jets,
each producing particle concentrations greater than 107 particles/cm3 at
6.5 L/min (nominal at 25 psig pressure). Built-in dilution air controlled by a
valve and rotameter allows you to vary the output particle concentration.
All four atomizers produce a mean droplet diameter of 0.3 µm with a
geometric standard deviation of less than 2.0. They are suitable for work
with a wide range of solutions and suspensions, including polystyrene
latex (PSL) spheres, dioctyl phthalate (DOP), silicon oil, salt or sugar
solutions, and methylene blue.
Please specify voltage requirements when ordering Model 3079. Model 3079 is produced in
Germany by Topas GmbH.

27

Model 9307-6
Generates large quantities of oil droplets for seeding.
TSI’s Model 9307 is a general purpose atomizer that uses a Laskin
nozzle to produce large quantities of oil or salt particles. An internal
impactor plate helps to generate particles with a reasonably narrow size
distribution, while a valve and pressure gauge arrangement provides an
easy way to control inlet air pressure, allowing for larger particle output
volume if necessary.

SUBMICROMETER MONODISPERSE AEROSOL
GENERATION SYSTEM

SUPERMICROMETER MONODISPERSE
AEROSOL GENERATION SYSTEM

Model 3940A

Model 3941

A complete system for generating monodisperse,

Everything you need to generate monodisperse

submicrometer particles.

particles as large as 200 micrometers.

The Submicrometer Mo­no­disperse Aerosol Gen­­eration System gives you

Our Supermicrometer Monodisperse Aerosol Generation System enables

the ability to produce mono­disperse particles from 0.01 to 1.0 µm in

you to produce the most monodisperse particles possible in a laboratory.

dia­meter. The system includes:

It generates particles from 1 to 200 µm in diameter, from nearly any

+ 3082 Electrostatic Classifier Platform (page 24)

material that can be put into solution with a volatile solvent. System

+ 3081A Long DMA (page 24)

components include:

+ 3077A Aerosol Neutralizer (page 30)

+ 3450 Vibrating Orifice Aerosol Generator (page 24)

+ 3012 Aerosol Neutralizer (page 30)

+ 3054 Aerosol Neutralizer (page 30)

+ 3074B Filtered Air Supply (page 31)

+ 3074B Filtered Air Supply (page 31)

+ 3076 Constant Output Atomizer (page 27)

+ 393520 10-µm orifice

+ 3062 Diffusion Dryer (page 29)

+ 393530 20-µm orifice

+ 1035900 Inlet Impactor

+ 393540 35-µm orifice

+ Interconnecting hardware

+ 393550 50-µm orifice
+ 393560 100-µm orifice
+ Interconnecting hardware

28

INSTRUMENTS & ACCESSORIES
NANOMETER AEROSOL
SAMPLER

MASS FLOWMETER

DIFFUSION DRYER

Model 3089

Model 4000 Series

Model 3062, 3062-NC

Deposits 2- to 100-nm particles on TEM

Monitors experimental flows, simply

Removes moisture from sample aerosols.

grids, AFM substrates, or glass slides.

and reliably.

Our Diffusion Dryer includes a removable

The Nanometer Aerosol Sampler (NAS) allows

Our Mass Flowmeters are perfect for monitoring

ex­tractor for collecting large water droplets.

you to sample charged particles, like those from

flows in your experimental setups. They

Desiccant surrounding the aerosol flowpath

the output of a Differential Mobility Analyzer

measure mass flow rates up to 20 std. L/min,

removes excess moisture by diffusional capture.

(DMA), onto sample substrates for analysis.

with exceptional accuracy at all flow rates. Plus,

Because aerosol never comes in contact with

You control the spot size of the deposition

they measure flow volume, gas temperature,

the desiccant, there is minimal particle loss.

using two electrode sizes and get a uniform

and absolute pressure. A low pressure-drop

Regenerate the desiccant simply by removing it

deposition spot size that is optimal for your

minimizes the impact of flow circuit back-

from the Diffusion Dryer and baking it at 120°C.

analysis system.

pressure on your test instruments. Data appears

Maximum flow rate is 4 L/min. The -NC version

on the LCD display in real time. Models 4140 &

has identical specifications, but no cobalt.

4143 include an analog output for recording the
flow rate and an RS-232 computer connection
for operational control, data logging, and data
display. An in-line HEPA filter is included to
protect the flowmeter from particles in the
aerosol sample and help maintain calibration.
Models 4140 & 4143 are calibrated for air and
have tube ends with a 0.5-inch outside diameter.

*Please specify voltage requirements.

29

*Please specify voltage requirements.

AEROSOL NEUTRALIZERS

ADVANCED AEROSOL
NEUTRALIZER

3054
3012

3077

Models 3012, 3054, 3077, 348002

Model 3088

Minimize particle losses and coagulation by electrostatic charges, or charge

A nonradioactive aerosol neutralizer.

particles properly for size analysis or air-filter measurements.

TSI’s Advanced Aerosol Neutralizer 3088 offers

Aerosol particles dispersed by nebulization, combustion, or powder dispersion are usually

an alternative to the traditional radioactive

electrostatically charged and are subject to high losses during transport. To re­duce transport losses

neutralizers frequently required for aerosol

and ensure that instruments operating on the electrostatic principle work properly, particles must

measurement applications. Due to increasingly

be neutralized. TSI Aerosol Neutralizers use a radioactive source (85Kr or 210 Po) to perform this

stringent local, state, and national regulations,

function. The radioactive source ionizes the surrounding atmosphere creating positive and negative

obtaining licensing to acquire and use

ions. Par­ticles carrying a high charge can discharge by interacting with ions of opposite polarity.

radioactive sources is often difficult and in some

After a short time, the particles reach charge equilibrium. TSI recommends models 3012A, 3054A,

cases prohibited. Fully compliant with US FDA,

or 3077A for aerosols with higher charge levels or when operating at higher flow rates or

CDRH* standards, the Model 3088 (U.S. Patent

high concentrations.

7,796,727) provides an attractive alternative,
with sizing performance virtually identical to

Aerosol Neutralizers
3012		

For general-purpose applications with high flow rates (up to 50 L/min)

3012A		

Same as above, but with five times the activity

3054		

For use with Model 3450 Vibrating Orifice Aerosol Generator (page 24)

TSI’s Aerosol Neutralizer Model 3077A.
*US FDA, CDRH – United States Food and Drug Administration,
Center for Devices and Radiological Health

3054A		 Same as above, but with twice the activity
3077		 For general-purpose applications with low flow rates (up to 5 L/min); standard
		 with Series 3936 Scanning Mobility Particle Sizer™ spectrometers (page 1)
3077A		

Same as above, but with five times the activity; included with Model 3034

		

Scanning Mobility Particle Sizer spectrometer

348002		

Included with Model 3480 Electrospray Aerosol Generator (page 25)

*Provide end-user name and address when ordering Aerosol Neutralizers. TSI has been issued license number 1154-200-62 by
the Minnesota Department of Health to sell and distribute these Aerosol Neutralizers. Users in the United States need not apply
for additional U.S. Government licenses to handle these products. However, some state and local governments may require special
licenses, and some organizations may have special handling procedures. Check all local requirements.

30

FILTERED AIR SUPPLY

FLOW SPLITTER

PARTICLE SIZE SELECTOR

Model 3074B

Model 3708

Model 376060

Cleans, dries, and regulates

Routes sample from one source

Allows selection of different

compressed air for aerosol generation

to several instruments.

cutoff sizes for CPCs.

and other applications.

The Flow Splitter directs an aerosol sample

The Particle Size Selector (PSS) allows you

The Filtered Air Supply removes oil or other

to as many as four destinations at once. Need

to control the lower size cutoff of a TSI

liquid droplets from the incoming air using

only two or three flow paths? Then simply

Condensation Particle Counter (CPC). The PSS

two prefilters. It also removes any remaining

block the unused outlet ports. This accessory is

is a separating device that selectively removes

moisture in the air­stream by passing the air

especially useful when performing instrument

small particles from an aerosol by diffusion.

through an advanced membrane dryer (no more

comparison or calibration experiments.

Simply add or remove diffusion screens to

drying of dessicant material!). Plus, it removes

Smooth flow transitions provide equal flow

change the lower cutoff size. The cutoff shifts

fine particles using a high-efficiency filter at

distribution. Stainless-steel construction and

toward larger sizes as more screens are added.

the outlet. This full-featured compressed-

an electropolished interior prevent the aerosol

air conditioner allows you to make pressure

from being contaminated. The Flow Splitter has

The PSS includes 11 screens and, therefore, can

adjustments using an included gas-regulator

a ⅜-inch straight-tube inlet and four ¼-inch

be configured for 11 cutoff sizes. An extra set of

valve. It handles a maximum flow rate of 60 L/

outlets (outside diameters). Maximum total flow

12 screens may be ordered to ex­pand the cutoff

min at a dewpoint as low as 2°C. Maximum inlet

rate is 30 L/min.

range further. Specific cutoff sizes vary based

pressure is 1,000 kPa (150 psig).

on CPC operating flow rate.
PSS Accessory (available separately)
376061		

Set of 12 additional 		

		diffusion screens
The technique of using a Condensation Par­ticle Counter with
diffusion screens to select specific particle size ranges is covered
in United States Patent Number 5,072,626.

31

VACUUM PUMPS

PROCESS AEROSOL MONITOR

Models 3032/3033

Model 3375

Use these top-quality pumps when

Monitors high-concentration, monodisperse aerosols.

you need a portable vacuum source.

The Process Aerosol Monitor (PAM) measures high-concentration,

Model 3032 is a diaphragm-type pump that produces flow rates up to 5 L/

monodisperse aerosols like those produced by our Model 3475

min. Oversized, permanently lubricated bearings promote longer life and

Condensation Monodisperse Aerosol Gen­erator (CMAG) or other Sinclair-

maintenance-free operation. The pump operates in any position. We offer

LaMer-type generators. It monitors aero­sols on-line, measuring particle

this small, reliable pump for use with our Model 3790, 3783, and 3772

size and concentration in real time.

Condensa­tion Particle Counters, or Series 3936 Scanning Mobility Particle
Sizer™ spectrometers containing a 3772 CPC.

The PAM is a com­pact, robust de­vice that connects directly to the
generator outlet. It validates particle size and concentration and helps

Model 3033 produces flows up to 60 L/min. This high-quality, rotary-vane

confirm that the generator is working properly. It is also suitable for

pump contains self-sealing, compound-carbon vanes that self-adjust as

monitoring an aerosol while making adjustments to the generator’s

they wear. Therefore, it always operates at top efficiency. Permanently

operating parameters.

lubricated ball bearings make the 3033 virtually maintenance-free.
This is the pump we recommend for use with our Model 3068B Aerosol

This instrument arrives precalibrated for diethyl-hexyl sebacate (DEHS).

Electrometer or when using multiple CPCs that require an external

It includes an RS-232 serial interface and a dyna­mic link library (DLL)

vacuum source. This pump is also suitable for TSI Model 3306 Im­pactor

routine and spreadsheet program. This enables the PAM to communicate

Inlet when used with our supermicrometer particle sizers.

with a computer so you can store data on a hard disk—a good idea for
calibration purposes!

Vacuum Pumps
3032

Flow rates up to 5 L/min, 115 V

3032-1

Flow rates up to 5 L/min, 230 V

Please specify voltage requirements. Model 3375 are produced in Germany by Topas GmbH.

3032-EC Flow rates up to 5 L/min, 230 V (Europe only)
3033

Flow rates up to 60 L/min, 115 V (North America only, customers

in other parts of the world must contact the factory for model number and
power ratings.)
TSI recommends these pumps only for use with specific TSI particle instruments. Please specify
voltage requirements.

32

33

TSI Incorporated serves a global market by investigating, identifying,

and safety, indoor air quality, fluid dynamics, and biohazard detection.

and solving measurement problems. As an industry leader in the

With headquarters based in the U.S. and field offices throughout

design and production of precision instruments, TSI partners with

Europe and Asia, TSI has established a worldwide presence in the

research institutions and customers around the world to set the

markets we serve. Every day, our dedicated employees turn research

standard for measurements relating to aerosol science, air flow, health

into reality.

34

Ordering

General Information

To order, contact your nearest representative or sales office. If you don’t

TSI Incorporated manufactures innovative instruments for use in industry

know which office handles your territory, then contact our corporate

and research. The particle instruments described in this catalog represent

headquarters. Our staff will answer any questions you may have or

only one of our product families. TSI offers a broad array of sensors and

they will put you in contact with the appropriate sales office. Contact

instrumentation systems used in a variety of measurement applications

information can also be found on the TSI website. When ordering, specify

around the globe.

the model number, instrument name, accessory models and names, and
voltage requirements.

Headquartered in Shoreview, Minnesota, TSI has sales and representative
offices all over the world. For more information on TSI particle instruments,

Customer Service

use the contact information shown below or visit particle.tsi.com. For

TSI Customer Service Specialists are available to answer your questions

information on TSI instruments not discussed in this catalog, go to the

about installation or operation:

main TSI web page at www.tsi.com

US & Canada: 1-800-874-2811
Europe: +49 241 523030
or www.tsi.com

A sincere effort was made to ensure that all information in this catalog was current at the time of
publication. However, specifications, features, and availability are subject to change. Please check
with your TSI representative for the latest information. Prototype or early instru­ments are depicted
in some photographs. Final products may vary from those pictured.
TSI, the TSI logo, Aero­dynamic Particle Sizer, APS, Aerosol In­strument Manager, Engine Ex­haust
Particle Sizer, EEPS, Fast Mobility Particle Sizer, FMPS, Scanning Mobility Particle Sizer, SMPS,
DustTrak and SidePak, are trademarks of TSI Incorporated.
Microsoft and Windows are trademarks of Microsoft Corporation.
LabVIEW is a trademark of National Instru­ments Corporation.

Aero­dynamic Particle Sizer, Aerosol Instrment Manager, are registered trademarks, and
APS, EEPS, Engine Exhaust Particle Sizer are trademarks

UNDERSTANDING, ACCELERATED

TSI Incorporated - Visit our website www.tsi.com for more information.
USA
UK
France
Germany

Tel: +1 800 874 2811
Tel: +44 149 4 459200
Tel: +33 1 41 19 21 99
Tel: +49 241 523030

P/N 1931001 Rev L

India
China
Singapore

©2015 TSI Incorporated

Tel: +91 80 67877200
Tel: +86 10 8219 7688
Tel: +65 6595 6388
Printed in U.S.A.

Kenelec Scientific Pty Ltd
23 Redland Drive, Mitcham VIC 3132
Call: +61 (0) 3 9873 1022
www.kenelec.com.au
National Freecall no: 1033 73 22 33
E sales@kenelec.com.au



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.7
Linearized                      : No
Encryption                      : Standard V4.4 (128-bit)
User Access                     : Print, Copy, Extract, Print high-res
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.1.0-jc003
Format                          : application/pdf
Producer                        : Adobe PDF Library 15.0; modified using iTextSharp™ 5.5.13 ©2000-2018 iText Group NV (AGPL-version)
Create Date                     : 2015:10:29 13:40:47-05:00
Creator Tool                    : Adobe InDesign CC 2015 (Windows)
Modify Date                     : 2018:02:23 14:03:08-06:00
Title                           : Particle_Catalog_Web.pdf
Creator                         : Ektron Automation
Page Layout                     : SinglePage
Page Count                      : 40
Trapped                         : False
Author                          : Ektron Automation
EXIF Metadata provided by EXIF.tools

Navigation menu