653B Resistivity Meter 653
User Manual: 653B
Open the PDF directly: View PDF .
Page Count: 16
Download | |
Open PDF In Browser | View PDF |
Model 653B Resistivity Meter Instruction Manual Manual No. 101418648, Revision E Instrument No. 101582036 Model 653B Resistivity Meter Instruction Manual ©2014 Fann Instrument Company Houston, Texas, USA All rights reserved. No part of this work covered by the copyright hereon may be reproduced or copied in any form or by any means (graphic, electronic, or mechanical) without first receiving the written permission of Fann Instrument Company, Houston, Texas, USA. Printed in USA. The information contained in this document includes concepts, methods, and apparatus which may be covered by U.S. Patents. Fann Instrument Company reserves the right to make improvements in design, construction, and appearance of our products without prior notice. FANN® and the FANN logo are registered trademarks of Fann Instrument Company in the United States and/or other countries. All other trademarks mentioned in the operating instructions are the exclusive property of the respective manufacturers. Contact FANN Phone TELEPHONE: 281-871-4482 TOLL FREE: 800-347-0450 FAX: 281-871-4358 Mail Fann Instrument Company P.O. Box 4350 Houston, Texas, 77210 USA Location Fann Instrument Company 14851 Milner Road, Gate 5 Houston, Texas, 77032, USA Online www.fann.com fannmail@fann.com 101418648 Revision E, January 2014 2 Model 653B Instruction Manual Table of Contents 1 Introduction ...............................................................................................................4 1.1 Document Conventions ....................................................................................4 2 Safety .......................................................................................................................6 2.1 Battery Safety ...................................................................................................6 3 Features and Specifications ......................................................................................7 4 Operation ..................................................................................................................9 4.1 Fluid Samples ...................................................................................................9 4.2 Semisolid Samples ......................................................................................... 10 5 Troubleshooting and Maintenance .......................................................................... 13 5.1 Calibration ...................................................................................................... 13 5.2 Battery Replacement ...................................................................................... 13 5.3 Cleaning ......................................................................................................... 14 6 Parts List.................................................................................................................15 7 Warranty and Returns ............................................................................................. 16 7.1 Warranty ......................................................................................................... 16 7.2 Returns ........................................................................................................... 16 List of Figures Figure 3-1 Resistivity Meter, Model 653B ........................................................................ 8 Figure 4-1 Resistivity Chart for NaCl Solutions .............................................................. 12 List of Tables Table 3-1 Resistivity Meter, Model 653B Specifications .................................................. 7 Table 3-2 Range of Environmental Conditions ................................................................ 7 Table 6-1 Model 653B Resistivity Meter Included Parts, P/N 101582036 ..................... 15 Table 6-2 Optional Equipment ....................................................................................... 15 101418648 Revision E, January 2014 3 Model 653B Instruction Manual 1 Introduction The Model 653B Resistivity Meter is a portable electronic meter that measures the resistivity of fluids, slurries, and semisolids according to API Recommended Practice 13B-1. The resistivity range for this meter is 0.01 to 10 ohm meters. This ohm-meter reading can be converted to ppm sodium chloride by using the sodium chloride resistivity chart provided in this instruction manual. Field and laboratory personnel rely on this instrument to evaluate formation characteristics from electric logs. Resistivity is the ability of a material to resist conduction; conductivity is the reciprocal of resistivity. 1.1 Document Conventions The following icons are used as necessary in this instruction manual. NOTE. Notes emphasize additional information that may be useful to the reader. CAUTION. Describes a situation or practice that requires operator awareness or action in order to avoid undesirable consequences. MANDATORY ACTION. Gives directions that, if not observed, could result in loss of data or in damage to equipment. WARNING! Describes an unsafe condition or practice that if not corrected, could result in personal injury or threat to health. 101418648 Revision E, January 2014 4 Model 653B Instruction Manual ELECTRICITY WARNING! Alerts the operator that there is risk of electric shock. HOT SURFACE! Alerts the operator that there is a hot surface and that there is risk of getting burned if the surface is touched. EXPLOSION RISK! Alerts the operator that there is risk of explosion. 101418648 Revision E, January 2014 5 Model 653B Instruction Manual 2 Safety Safe laboratory practices and procedures should be observed while operating and maintaining the Model 653B. 2.1 Battery Safety On nine volt batteries, the positive and negative terminals are on the same end, next to each other. If these terminals touch anything metal, such as steel wool, coins, or a paper clip, a short circuit could occur, creating heat, sparks, a fire, or an explosion. Nine volt batteries can explode if they are stored in a hot location, charged improperly, or if the terminals contact metal, creating a short circuit. Some batteries contain toxic or corrosive material, which can leak if the battery is damaged. • Store unused 9-volt batteries in their original packaging, a sealed plastic container, or securely cover both terminals to prevent short circuits. • Do not charge rechargeable batteries before storing them. • Store batteries at or below 70oF (21oC). 101418648 Revision E, January 2014 6 Model 653B Instruction Manual 3 Features and Specifications The Model 653B can measure resistivity from 0.01 to 10 ohm-meters. The ohmmeter reading can be converted into parts per million sodium chloride (or equivalent calcium chloride, potassium chloride or other carbonate or hydroxyl salts) using the resistivity chart (Figure 4-1). This instrument can be operated by its 9V battery. Refer to Figure 3-1. The Model 653B has these features: • Direct reading in ohm-meters • Built-in temperature probe • Push Button for calibration (black) Table 3-1 Resistivity Meter, Model 653B Specifications Category Specification Meter Range 0.01 to 10 ohm-meters Meter Dimensions (Height x Width x Length) 3.1 x 3.7 x 6.2 inches Carrying Case Dimensions (Height x Width x Length) 9 x 3.5 x 7.9 inches Weight 3.25 lb (1.5 kg) Power Supply 9V alkaline battery 8.0 x 9.5 x 15.8 centimeters 22.9 x 8.9 x 20.1 centimeters Table 3-2 Range of Environmental Conditions Maximum Altitude Temperature Range Maximum Relative Humidity (RH) 101418648 6562 ft (2000 m) 41oF to 104oF (5oC to 40oC) 80% RH at 87.8oF (31oC) or less 50% RH at 104oF (40oC) Revision E, January 2014 7 Model 653B Instruction Manual Figure 3-1 Resistivity Meter, Model 653B 101418648 Revision E, January 2014 8 Model 653B Instruction Manual 4 Operation This resistivity meter can measure resistivity of fluid samples and semisolid samples (e.g., filter cakes or mud solids). Procedures for both types of samples are listed here. The resistivity chart (Figure 4-1) is used to determine the quantity of sodium chloride (in combination with distilled water or other salt-free aqueous medium) that is required to produce a solution with the same resistivity as the sample. The concentration for carbonate salts, calcium salts, and hydroxyl salts can be found in conductance tables for aqueous solutions. 4.1 Fluid Samples This procedure explains how to measure resistivity of liquids, filtrates, and drilling fluids. Refer to Figure 3-1. 1. Remove the sample cell and fill it with the sample. Fill and discharge the sample cell several times before the final fill. To fill use the suction ball to pull sample into the Lucite cell. Make sure there are no bubbles in the sample. 2. Reattach the sample cell onto the pins on the meter. Be sure the sample fills the area between the two metal posts in the cell. If the sample is not at room temperature, wait for the temperatures of the sample and the sell to reach equilibrium. 3. Press black switch to calibrate the meter. Use the ADJ control to set the meter to the ADJ position. 4. Press both the black and red switches at the same time. Then read the resistivity directly from the meter in ohm-meters. 101418648 Revision E, January 2014 9 Model 653B Instruction Manual 5. Read the temperature of the sample as shown on the thermometer in the cell wall. 6. Remove the sample cell and clean it with distilled water. If necessary, use a pipe cleaner and mild soap to clean the cell bore. Do not use solvents to clean the cell. 7. Using the resistivity chart (Figure 4-1), convert the resistivity in ohm-meters to concentration of chlorides. 4.2 Semisolid Samples This procedure explains how to measure resistivity of semisolid samples, such as filter cakes and mud solids. Refer to Figure 3-1. 1. Prepare samples of uniform moisture content. 2. Remove the sample cell from the resistivity meter. 3. Completely fill the slot on the outer surface of the cell with semisolid sample. 4. Reattach the sample cell onto the pins on the meter. Be sure the sample fills the area between the two metal posts in the cell. If the sample is not at room temperature, wait for the temperatures of the sample and the sell to reach equilibrium. 5. Press black switch to calibrate the meter. Use the ADJ control to set the meter to the ADJ position. 6. Press both the black and red switches at the same time. Then read the resistivity directly from the meter in ohm-meters. 7. Read the temperature of the sample as shown on the thermometer in the cell wall. 8. Remove the sample cell and clean it with distilled water. If necessary, use a pipe cleaner and mild soap to clean the slot. 101418648 Revision E, January 2014 10 Model 653B Instruction Manual Do not use solvents to clean the cell. 9. Using the resistivity chart (Figure 4-1), convert the resistivity in ohm-meters to concentration of chlorides. 101418648 Revision E, January 2014 11 Model 653B Instruction Manual Figure 4-1 Resistivity Chart for NaCl Solutions 101418648 Revision E, January 2014 12 Model 653B Instruction Manual 5 Troubleshooting and Maintenance Calibration and regular maintenance procedures are described in this section. If more extensive maintenance or service of the instrument is required, please contact Fann Instrument Company. 5.1 Calibration Press black switch to calibrate the meter. Use the ADJ control to set the meter to the ADJ position. A thorough calibration requires a calibration kit. A calibration kit (P/N 210179) is available as a separate purchase. See the parts list for details. 5.2 Battery Replacement When the meter cannot be calibrated with the zero control, replace the 9 Volt battery (P/N 205643). Observe the polarity marking on the printed circuit board. If the battery is inserted incorrectly, it will become damaged. 101418648 Revision E, January 2014 13 Model 653B Instruction Manual 5.3 Cleaning Always clean the cell immediately after making a measurement. Do not use solvents to clean the cell. Avoid scratching the surface of the cell. 1. Purge the cell with distilled water until it is clean. 2. If additional cleaning is necessary, use a pipe cleaner and mild soap. 3. After extensive use, follow these steps to thoroughly clean the cell. a. Soak the cell bore in cleaning solution (P/N 210181) for 2 to 3 hours. b. Scrub it with a pipe cleaner. c. Rinse with distilled water. d. Dry the cell with a clean pipe cleaner. 101418648 Revision E, January 2014 14 Model 653B Instruction Manual 6 Parts List Table 6-1 Model 653B Resistivity Meter Included Parts, P/N 101582036 Part No. Description 205643 208758 210174 210170 210168 204330 101418648 205764 210441 205119 205120 210071 206043 BATTERY, 9 VOLT CARRYING CASE CELL (COMPLETE) CIRCUIT BOARD ASSEMBLY ELECTRODE CONTACT (2 REQUIRED) KNOB INSTRUCTION MANUAL METER, 0 TO 50 MICRO AMPS PIPE CLEANERS, 20/PKG BLACK PUSHBUTTON SWITCH RED PUSHBUTTON SWITCH RUBBER BULB THERMOMETER Table 6-2 Optional Equipment Part No. 210179 210181 210182 Description RESISTIVITY CALIBRATION KIT* CLEANING SOLUTION, 4 OZ. STANDARD RESISTIVITY SOLUTION, 4 0Z. *Calibration Kit includes pipe cleaners, cleaning solution, and standard resistance solution. 101418648 Revision E, January 2014 15 Model 653B Instruction Manual 7 Warranty and Returns 7.1 Warranty Fann Instrument Company warrants only title to the equipment, products and materials supplied and that the same are free from defects in workmanship and materials for one year from date of delivery. THERE ARE NO WARRANTIES, EXPRESS OR IMPLIED OF MERCHANTABILITY, FITNESS OR OTHERWISE BEYOND THOSE STATED IN THE IMMEDIATELY PRECEDING SENTENCE. Fann's sole liability and Customer's exclusive remedy in any cause of action (whether in contract, tort, breach of warranty or otherwise) arising out of the sale, lease or use of any equipment, products or materials is expressly limited to the replacement of such on their return to Fann, or at Fann's option, to the allowance to Customer of credit for the cost of such items. In no event shall Fann be liable for special, incidental, indirect, consequential or punitive damages. Notwithstanding any specification or description in its catalogs, literature or brochures of materials used in the manufacture of its products, Fann reserves the right to substitute other materials without notice. Fann does not warrant in any way equipment, products, and material not manufactured by Fann, and such will be sold only with the warranties, if any, that are given by the manufacturer thereof. Fann will only pass through to Customer the warranty granted to it by the manufacturer of such items. 7.2 Returns For your protection, items being returned must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Fann will not be responsible for damage resulting from careless or insufficient packing. Before returning items for any reason, authorization must be obtained from Fann Instrument Company. When applying for authorization, please include information regarding the reason the items are to be returned. Our correspondence address: Fann Instrument Company P.O. Box 4350 Houston, Texas USA 77210 Telephone: Toll Free: FAX: 281-871-4482 800-347-0450 281-871-4446 Email fannmail@fann.com 101418648 Our shipping address: Fann Instrument Company 14851 Milner Road, Gate 5 Houston, Texas USA 77032 Revision E, January 2014 16
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : No Author : Marsha Tomlin Category : Drilling Fluids Comments : MODEL 653B RESISTIVITY METER, REV E,JANUARY 2014.UPDATE TO MATCH MODEL 88 MANUAL, EDITS, PARTS LISTS, WARRANTY STMT & NEW PICTURES. Company : Halliburton Create Date : 2014:01:31 13:29:48-06:00 Keywords : Model 653B, resistivity meter, 101418648, 101582036 Modify Date : 2014:02:03 10:12:22-06:00 Source Modified : D:20140131192813 Subject : Fann Instrument Company Instruction Manual for Equipment Tag Ad Hoc Review Cycle ID : -52690876 Tag Author Email : fedderly@earthlink.net Tag Author Email Display Name : Martha Wiles Tag Email Subject : template Tag Previous Ad Hoc Review Cycle ID: 1065231200 Tag Reviewing Tools Shown Once : Tagged PDF : Yes XMP Toolkit : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26 Metadata Date : 2014:02:03 10:12:22-06:00 Creator Tool : Acrobat PDFMaker 10.0 for Word Document ID : uuid:193d9f3a-ff5f-4aca-bce6-a88d8c69e326 Instance ID : uuid:0f8e9cc9-1ef4-4622-ab11-19460a5afe66 Format : application/pdf Title : Model 653B Resistivity Meter Description : Fann Instrument Company Instruction Manual for Equipment Creator : Marsha Tomlin Producer : Adobe PDF Library 10.0 Headline : Fann Instrument Company Instruction Manual for Equipment Page Layout : OneColumn Page Mode : UseOutlines Page Count : 16EXIF Metadata provided by EXIF.tools