00809-0100-4727 English Rev. DA Series 8700 Magnetic Flowmeter Flowtubes Product Manual 1 Series 8700 Magnetic Flowmeter Flowtubes NOTICE Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product. Within the United States, Rosemount Inc. has two toll-free assistance numbers. Customer Central: 1-800-999-9307 (7:00 a.m. to 7:00 p.m. CST) Technical support, quoting, and order-related questions. North American 1-800-654-7768 (24 hours a day – Includes Canada) Response Center: Equipment service needs. For equipment service or support needs outside the United States, your local Rosemount representative. The products described in this document are NOT designed for nuclear-qualified applications. For information on Rosemount nuclear-qualified products, contact your local Rosemount Sales Representative. Rosemount, the Rosemount logotype, and SMART FAMILY are registered trademarks of Rosemount Inc. DeltaV is a trademark of the Fisher-Rosemount group of companies. HART is a registered trademark of the HART Communication Foundation. Tefzel and Teflon are registered trademarks of E.I. du Pont de Nemours & Co. Ryton is a registered trademark of the Phillips Petroleum Co. Fluoraz is a registered trademark of Greens, Tweed & Co., Inc. Hastelloy C and Hastelloy C-22 are registered trademarks of Haynes International, FOUNDATION is a trademark of the Fieldbus Foundation. Cover photo: 8711-004AC Fisher-Rosemount satisfies all obligations coming from legislation to harmonize product requirements in the European Union. Rosemount Inc. PR INT IN U. S. A. ED 8200 Market Boulevard Chanhassen, MN 55317 USA Tel 1-800-999-9307 Fax (952) 949-7001 © 2000 Rosemount Inc. Printed in the USA. http://www.rosemount.com SNF-0004 Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings. Table of Contents IMPORTANT Procedures and instructions in this manual may require special precautions to ensure the safety of the personnel performing the operations. Refer to the safety messages at the beginning of each section before performing any operations. SECTION 1 Introduction Manual Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Safety Messages SECTION 2 Installation Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 Step 1: Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Step 2: Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Upstream/Downstream Piping . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Flowtube Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Flow Direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5 Step 3: Installation of Model 8705 and Model 8707 High-Signal . . . 2-5 Gaskets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5 Flange Bolts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 Step 3: Installation of Model 8711 . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8 Gaskets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8 Alignment and Bolting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8 Flange Bolts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9 Step 4: Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10 Step 5: Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 Conduit Ports and Connections . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13 Cable Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15 Flowtube to Model 8712C/U Transmitter Connections . . . . . . . 2-16 High-Signal Flowtube to High-Signal Transmitter Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17 Flowtube to Integral Mount Transmitter Connections . . . . . . . 2-18 Step 6: Process Leak Protection (Model 8705 and Model 8707 High-Signal Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19 Standard Housing Configuration . . . . . . . . . . . . . . . . . . . . . . . . 2-19 Relief Valves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20 Process Leak Containment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21 Step 7: Start-up and Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22 Flowtube Calibration Number . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22 Quick Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22 SECTION 3 Troubleshooting Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Independent Flowtube and Transmitter Replacement . . . . . . . . . . . Flowtube Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Return of Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 3-1 3-2 3-4 i SECTION 4 Specifications: Model 8705 and Model 8707 High-Signal Flowtubes Specifications Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1 Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3 Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4 SECTION 5 Specifications: Model 8711 Flowtube Specifications Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1 Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 APPENDIX A Field-Removable Electrodes Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1 Remove the Electrode Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1 Replace the Electrode Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2 INDEX ii I-1 Section 1 MANUAL SCOPE Introduction The Rosemount® Series 8700 Magnetic Flowmeter System combines separate flowtube and transmitter units. This manual is designed to assist in the installation and operation of Rosemount Model 8705, Model 8707 High-Signal, and Model 8711 Magnetic Flowmeter Flowtubes. Specific transmitter information is located in the product manual for the Model 8712C/U/H (document 00809-0100-4729), Model 8732C (document 00809-0100-4725), or Model 8742C (document 00809-0100-4793) Magnetic Flowmeter Transmitters. Attempting to install and operate Model 8705, Model 8707 High-Signal, or Model 8711 Magnetic Flowmeter Flowtubes without reviewing the instructions contained in this manual could result in personal injury or equipment damage. Section 2: Installation • installation instructions Section 3: Start-Up and Operation • flowtube calibration number • flowtube configuration information. Section 4: Troubleshooting • troubleshooting procedures • electrical circuit diagrams Section 5: Model 8705 and Model 8707 High-Signal Flowtube Specifications • specifications for Model 8705 and Model 8707 • reference tables • dimensional drawings Section 6: Model 8711 Flowtube Specifications • specifications for Model 8711 • reference tables • dimensional drawings Appendix A: Field-Removable Electrodes • instructions for removing and replacing the field-removable electrode assembly 1-1 Rosemount Series 8700 Magnetic Flowmeter Flowtube SYSTEM DESCRIPTION Rosemount Series 8700 Magnetic Flowmeter Systems measure volumetric flow rate by detecting the velocity of a conductive liquid that passes through a magnetic field. Magnetic Flowmeter Systems consist of a flowtube and a transmitter. There are three Rosemount magnetic flowmeter flowtubes: • Flanged Model 8705 • Flanged High-Signal Model 8707 • Wafer-Style Model 8711 There are three Rosemount magnetic flowmeter transmitters: • Model 8712C/U/H • Model 8732C • Model 8742C The flowtube is installed in-line with process piping — either vertically or horizontally. Coils located on opposite sides of the flowtube create the necessary magnetic field. A conductive liquid moving through the magnetic field generates a voltage that is detected by two electrodes. The transmitter drives the coils to generate a magnetic field and electronically conditions the voltage detected by the electrodes. The transmitter then produces output signals proportional to the liquid velocity. SAFETY MESSAGES 1-2 Procedures and instructions in this manual may require special precautions to ensure the safety of the personnel performing the operations. Refer to the safety messages listed at the beginning of each section before performing any operations. Section 2 SAFETY MESSAGES Installation This section covers the steps required to physically install the flowtube. Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Please refer to the following safety messages before performing any operation in this section. Failure to follow these installation guidelines could result in death or serious injury: Installation and servicing instructions are for use by qualified personnel only. Performing any servicing other than that contained in this manual may result in death or serious injury. Do not perform any servicing other than that contained in the operating instructions, unless qualified. The flowtube liner is vulnerable to handling damage. Never place anything through the flowtube for the purpose of lifting or gaining leverage. Liner damage can render the flowtube useless. To avoid possible damage to the flowtube liner ends, do not use metallic or spiral-wound gaskets. If frequent removal is anticipated, take precautions to protect the liner ends. Short spool pieces attached to the flowtube ends are often used for protection. Correct flange bolt tightening is crucial for proper flowtube operation and life. All bolts must be tightened in the proper sequence to the specified torque limits. Failure to observe these instructions could result in severe damage to the flowtube lining and possible flowtube replacement. 2-1 Rosemount Series 8700 Magnetic Flowmeter Flowtube STEP 1: HANDLING Handle all parts carefully to prevent damage. Whenever possible, transport the system to the installation site in the original shipping containers. Teflon®-lined flowtubes are shipped with end covers that protect it from both mechanical damage and normal unrestrained distortion. Remove the end covers just before installation. Flanged 6- through 36-inch flowtubes come with a lifting lug on each flange. The lifting lugs make the flowtube easier to handle when it is transported and lowered into place at the installation site. Flanged ½to 4-inch flowtubes do not have lugs. They must be supported with a lifting sling on either side of the housing. Figure 2-1 shows flowtubes correctly supported for handling and installation. Notice the plywood end pieces are still in place to protect the flowtube liner during transportation. 8705-0281D02A, 0281E02A FIGURE 2-1. Flowtube Support for Handling and Installation ½- through 4-Inch Flowtubes 6- through 36-Inch Flowtubes STEP 2: MOUNTING Physical mounting of a flowtube is similar to installing a typical section of pipe. Conventional tools, equipment, and accessories (bolts, gaskets, and grounding hardware) are required. Calibration Rosemount flowtubes are wet calibrated at the factory. They do not need further calibration during installation. See Safety Messages on page 2-1 for complete warning information. 2-2 Installation To ensure specification accuracy over widely varying process conditions, install the flowtube a minimum of five straight pipe diameters upstream and two pipe diameters downstream from the electrode plane (see Figure 2-2). Upstream/ Downstream Piping FIGURE 2-2. Upstream and Downstream Straight Pipe Diameters 2 Pipe Diameters 8705-0281F02A.EPS 5 Pipe Diameters FLOW The flowtube should be installed in a position that ensures the flowtube remains full during operation. Horizontal or inclined positions are preferred. Figures 2-3, 2-4, and 2-5 show the proper flowtube orientation for the most common installations. Flowtube Orientation The following orientations ensure that the electrodes are in the optimum plane to minimize the effects of entrapped gas. Further information on electrode orientation can be found in Appendix A. As illustrated in Figures 2-4B and 2-5B, avoid downward flows where back pressure does not ensure that the flowtube remains full at all times. FLOW 8705-0005F01A FIGURE 2-3. Horizontal Flowtube Orientation 2-3 Rosemount Series 8700 Magnetic Flowmeter Flowtube FIGURE 2-4. Vertical Flowtube Orientation A 8705-0005G01A, 0005G01B B FLOW FLOW FIGURE 2-5. Incline or Decline Orientation 8705-0005E01A, 0005E01B FLOW FLOW 2-4 A B Installation The flowtube should be mounted so that the FORWARD end of the flow arrow, shown on the flowtube identification tag, points in the direction of flow through the tube (see Figure 2-6). In this mounting configuration, the conduit ports point upstream. Flow Direction 8705-0281H02F FIGURE 2-6. Flow Direction FLOW Flow Direction Arrow STEP 3: INSTALLATION OF MODEL 8705 AND MODEL 8707 HIGH-SIGNAL The following drawings should be used as a guide in the installation of the Model 8705 and Model 8707 High-Signal Flowtubes. Refer to page 2-8 for installation of the Model 8711 Flowtube. Gaskets The flowtube requires a gasket at each of its connections to adjacent devices or piping. The gasket material selected must be compatible with the process fluid and operating conditions. Metallic or spiral-wound gaskets can damage the liner. If the gaskets will be removed frequently, protect the liner ends. A gasket is required on each side of the grounding ring, as shown in Figure 2-7. All other applications (including lining protector and grounding electrode) require only one gasket on each end connection, as shown in Figure 2-8. See Safety Messages on page 2-1 for complete warning information. 2-5 Rosemount Series 8700 Magnetic Flowmeter Flowtube Gasket (Supplied by user) Grounding Ring Gasket (Supplied by user) 8705-0038D FIGURE 2-7. Gasket Placement with Non-attached Grounding Rings Gasket (Supplied by user) Flange Bolts Flowtube sizes and torque values for both ASME B16.5 (ANSI) Class 150 and Class 300 flanges are listed in Table 2-1. Consult factory for other flange ratings and types. Tighten flange bolts in the incremental sequence as shown in Figure 2-9. See Table 5 and Table 7 for bolt sizes and hole diameters. Always check for leaks at the flanges after tightening the flange bolts. Failure to use the correct flange bolt tightening methods can result in severe damage. All flowtubes require a second torquing twenty-four hours after initial flange bolt tightening. See Safety Messages on page 2-1 for complete warning information. 2-6 8705-0040E FIGURE 2-8. Gasket Placement Installation FIGURE 2-9. Flange Bolt Torquing Sequence 8-Bolt 4-Bolt 8701-0870G02A 20-Bolt Torque the flange bolts in increments according to the above numerical sequence. 12-Bolt TABLE 2-1. Flange Bolt Torque Specifications for Model 8705 and Model 8707 High-Signal Flowtubes 14-Bolt Size Code Line Size Class 150 (pound-feet) Class 300 (pound-feet) 005 010 015 020 030 040 060 080 100 120 140 160 180 200 240 300 360 ½-inch (15 mm) 1 inch (25 mm) 1½ inch (40 mm) 2 inch (50 mm) 3 inch (80 mm) 4 inch (100 mm) 6 inch (150 mm) 8 inch (200 mm) 10 inch (250 mm) 12 inch (300 mm) 14 inch (350 mm) 16 inch (400 mm) 18 inch (450 mm) 20 inch (500 mm) 24 inch (600 mm) 30 inch (750 mm) 36 inch (900 mm) 10 10 17 25 45 35 60 80 70 80 100 90 125 125 150 150 200 10 10 22 17 35 50 65 60 65 80 – – – – – – – 2-7 Rosemount Series 8700 Magnetic Flowmeter Flowtube STEP 3: INSTALLATION OF MODEL 8711 The following section should be used as a guide in the installation of the Model 8711 Flowtube. Gaskets The flowtube requires a gasket at each of its connections to adjacent devices or piping. The gasket material selected must be compatible with the process fluid and operating conditions. Metallic or spiral-wound gaskets can damage the liner. If the gaskets will be removed frequently, protect the liner ends. If grounding rings are used, a gasket is required on each side of the grounding ring. Alignment and Bolting 1. On 1½ - through 8-inch (40 through 200 mm) line sizes, place centering rings over each end of the flowtube. The smaller line sizes, 0.15- through 1-inch (4 through 25 mm), do not require centering rings. 2. Insert studs for the bottom side of the flowtube between the pipe flanges. Stud specifications are listed in Table 2-2 Using carbon steel bolts on smaller line sizes, 0.15- through 1-inch (4 through 25 mm), rather than the required stainless steel bolts, will degrade performance. TABLE 2-2. Stud Specifications Nominal Flowtube Size Stud Specifications 0.15 – 1 inch (4 – 25 mm) 316 SST ASTM A193, Grade B8M Class 1 threaded mounted studs 1½ – 8 inch (40 – 200 mm) CS, ASTM A193, Grade B7, threaded mounting studs 3. Place the flowtube between the flanges. Make sure that the centering rings are properly placed in the studs. The studs should be aligned with the markings on the rings that correspond to the flange you are using. 4. Insert the remaining studs, washers, and nuts. 5. Tighten to the torque specifications shown in Table 2-1 Do not overtighten the bolts or the liner may be damaged. NOTE On the 4- and 6-inch PN 10–16, insert the flowtube with rings first and then insert the studs. The slots on this ring scenario are located on the inside of the ring. See Safety Messages on page 2-1 for complete warning information. 2-8 Installation FIGURE 2-10. Gasket Placement with Centering Rings Customer-supplied Gasket 8732-0002A1A Installation, Studs Nuts and Washers Centering Rings FLOW Flange Bolts Flowtube sizes and torque values for both Class 150 and Class 300 flanges are listed in Table 2-3. Tighten flange bolts in the incremental sequence, shown in Figure 2-9. Always check for leaks at the flanges after tightening the flange bolts. All flowtubes require a second torquing 24 hours after initial flange bolt tightening. TABLE 2-3. Torque specifications Size Code Line Size Pound-feet Newton-meter 15F 30F 005 010 015 020 030 040 060 080 0.15 inch (4 mm) 0.30 inch (8 mm) ½-inch (15 mm) 1 inch (25 mm) 1½ inch (40 mm) 2 inch (50 mm) 3 inch (80 mm) 4 inch (100 mm) 6 inch (150 mm) 8 inch (200 mm) 5 5 5 10 15 25 40 30 50 70 6.8 6.8 6.8 13.6 20.5 34.1 54.6 40.1 68.2 81.9 2-9 Rosemount Series 8700 Magnetic Flowmeter Flowtube Grounding the flowtube is one of the most important details of flowtube installation. Proper grounding ensures that only the voltage induced in the magnetic field of the flowtube is measured. Use Table 2-4 to determine which grounding option to follow for proper installation. Attached grounding rings should be grounded equivalently to non-attached grounding rings. STEP 4: GROUNDING NOTE Consult factory for installations requiring cathodic protection or situations where there are high currents or high potential in the process. The flowtube case should always be grounded in accordance with national and local electrical codes. Failure to do so may impair the protection provided by the equipment. The most effective grounding method is direct connection to earth ground with minimal impedance. The Internal Ground Connection (Protective Ground Connection) located in side the junction box is the Internal Ground Connection screw. This screw is identified by the ground symbol: TABLE 2-4. Grounding Installation Grounding Options Type of Pipe No Grounding Options Grounding Rings Grounding Electrodes Lining Protectors Conductive Unlined Pipe See Figure 2-11 Not Required Not Required See Figure 2-12 Conductive Lined Pipe Insufficient Grounding See Figure 2-12 See Figure 2-11 See Figure 2-12 Non-Conductive Pipe Insufficient Grounding See Figure 2-13 See Figure 2-14 See Figure 2-13 2-10 Installation FIGURE 2-11. No Grounding Options or Grounding Electrode in Lined Pipe 8705-0040C Earth Ground FIGURE 2-12. Grounding with Grounding Rings or Lining Protectors Grounding Rings or Lining Protectors 8705-0038C Earth Ground 2-11 Rosemount Series 8700 Magnetic Flowmeter Flowtube FIGURE 2-13. Grounding with Grounding Rings or Lining Protectors 8705-0038E Earth Ground Grounding Rings or Lining Protectors FIGURE 2-14. Grounding with Grounding Electrodes 8705-0040D Earth Ground 2-12 Installation STEP 5: WIRING The conduit connections needed for installation depend on transmitter location. A conduit run between the flowtube and transmitter is not required if the transmitter is integrally mounted on the flowtube. Conduit Ports and Connections Both the flowtube and transmitter junction boxes have ports for ¾-inch NPT conduit connections. These connections should be made in accordance with local or plant electrical codes. Unused ports should be sealed with metal plugs. Housing damage will result if metal plugs are overtightened. Flanged and wafer flowtubes have two conduit ports as shown in Figures 1, 2, 3, and 1. Either one may be used for both the coil drive and electrode cables. Use the stainless steel plug that is provided to seal the unused conduit port. Series 8700 Flowtubes use a pulsed-dc signal generated by Model 8712C/U/H, Model 8732C, or Model 8742C Magnetic Flowmeter Transmitter. Connect the Model 8712H High-Signal Magnetic Flowmeter Transmitter only to the Model 8707 High-Signal Flowtube. This is a pulsed dc magnetic flowmeter. Do not connect ac power to the flowtube or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary. A single dedicated conduit run for the coil drive and electrode cables is needed between a flowtube and a remote transmitter. Bundled cables in a single conduit are likely to create interference and noise problems in your system. Use one set of cables per conduit run. See Figure 2-15 for proper conduit installation diagram and Table 2-5 for recommended cable. Use wire rated for the proper temperature application. For connections in ambient temperatures above 140 °F (60 °C), use a wire rated for 176 °F (80 °C). For ambients greater than 176 °F (80 °C), use a wire rated for 230 °F (110 °C). 2-13 Rosemount Series 8700 Magnetic Flowmeter Flowtube FIGURE 2-15. Conduit Preparation Power Coil Drive and Electrode Cables Outputs Power Outputs Power Coil Drive and Electrode Cables Outputs 8705-0005CDGA, DDGA Outputs Power WRONG TABLE 2-5. Cable Requirements CORRECT Transmitter Input Power Description Part Number Signal Cable (20 AWG) Belden 8762, Alpha 2411 equivalent 08712-0061-0001 Coll Drive Cable (14 AWG) Belden 8720, Alpha 2442 equivalent 08712-0060-0001 Combination Signal and Coil Drive Cable (18 AWG)(1) Belden 9368 equivalent 08712-0750-0001 (1) Combination signal and coil drive cable is not recommended for high-signal magmeter system. For remote mount installations, combination signal and coil drive cable should be limited to less than 100 ft (30 m). Remote transmitter installations require equal lengths of signal and coil drive cables. Integrally mounted transmitters are factory wired and do not require interconnecting cables. Lengths from 5 to 1,000 feet (1.5 to 300 meters) may be specified, and will be shipped with the flowtube. Cable longer than 100 feet (30 meters) is not recommended for high-signal magmeter systems. 2-14 Installation Cable Preparation Prepare the flowtube ends of the coil drive and electrode cables, as shown in Figure 2-16. NOTE The maximum length for unshielded wire in the wiring compartment is ½-inch. Correct cable preparation is important for a successful installation. The cable shield should be stripped back no more than 0.50 inch. If more wire is exposed, electrical noise will increase and create unstable meter readings. FIGURE 2-16. Cable Preparation Detail ch Cable Shield 8705-0041B 0.50 in IMPORTANT Failure to connect the cable shields will result in improper operation. Series 8700 Flowtube Systems require that the cable shields be connected at the flowtube for proper operation. 2-15 Rosemount Series 8700 Magnetic Flowmeter Flowtube Flowtube to Model 8712C/U Transmitter Connections Connect coil drive and electrode cables as shown in Figure 2-17. 8712-04A FIGURE 2-17. Wiring Diagram to Model 8712C/U Transmitter TABLE 2-6. Flowtube to Remote Mount Transmitter 2-16 Rosemount Model 8712C/U Transmitters Rosemount Model 8705/8707/8711 Flowtubes 1 1 2 2 17 17 18 18 19 19 This is a pulsed dc magnetic flowmeter. Do not connect ac power to the flowtube or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary. Installation Connect coil drive and electrode cables as shown in Figure 2-18. High-Signal Flowtube to High-Signal Transmitter Connections 8712-03A FIGURE 2-18. Wiring Diagram to Model 8712H High-Signal Transmitter TABLE 2-7. High-Signal Flowtube to High-Signal Transmitter Rosemount Model 8712H Transmitters Rosemount Model 8707 Flowtubes 1 1 2 2 17 17 18 18 19 19 This is a pulsed dc magnetic flowmeter. Do not connect ac power to the flowtube or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary. 2-17 Rosemount Series 8700 Magnetic Flowmeter Flowtube Connect coil drive and electrode cables as shown in Figure 2-19. Flowtube to Integral Mount Transmitter Connections FIGURE 2-19. Wiring Diagram to Model 8732C/8742C Transmitter 8732-8732B01A Electronics Board TABLE 2-8. Flowtube to Integral Mount Transmitter 2-18 Rosemount Model 8732C/8742C Rosemount Model 8705/8711 Flowtubes 1 1 2 2 17 17 18 18 19 19 W This is a pulsed dc magnetic flowmeter. Do not connect ac power to the flowtube or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary. Installation STEP 6: PROCESS LEAK PROTECTION (MODEL 8705 AND MODEL 8707 HIGH-SIGNAL ONLY) The Model 8705 and Model 8707 High-Signal Flowtube housing is fabricated from carbon steel to perform two separate functions. First, it provides shielding for the flowtube magnetics so that external disturbances cannot interfere with the magnetic field and thus affect the flow measurement. Second, it provides the physical protection to the coils and other internal components from contamination and physical damage that might occur in an industrial environment. The housing is completely welded and gasket-free. The three housing configurations are identified by the W0, W1, or W3 in the model number option code when ordering. Below are brief descriptions of each housing configuration, which are followed by a more detailed overview. • Code W0 — sealed, welded coil housing (standard configuration) • Code W1 — sealed, welded coil housing with a relief valve capable of venting fugitive emissions to a safe location (additional plumbing from the flowtube to a safe area, installed by the user, is required to vent properly) • Code W3 — sealed, welded coil housing with separate electrode compartments capable of venting fugitive emissions (additional plumbing from the flowtube to a safe area, installed by the user, is required to vent properly) Standard Housing Configuration The standard housing configuration is identified by a code W0 in the model number. This configuration does not provide separate electrode compartments with external electrode access. In the event of a process leak, these models will not protect the coils or other sensitive areas around the flowtube from exposure to the pressure fluid. FIGURE 2-20. Standard Housing — Configuration Sealed Welded Housing (Option Code W0) ¾–14 NPT Conduit Connection 8705-1002A05D (no relief valve) 2-19 Rosemount Series 8700 Magnetic Flowmeter Flowtube Relief Valves The first optional configuration, identified by the W1 in the model number option code, uses a completely welded coil housing. This configuration does not provide separate electrode compartments with external electrode access. This optional housing configuration provides a relief valve in the housing to prevent possible overpressuring caused by damage to the lining or other situations that might allow process pressure to enter the housing. The relief valve will vent when the pressure inside the flowtube housing exceeds 5 psi. Additional piping (provided by the user) may be connected to this relief valve to drain any process leakage to safe containment (see Figure 2-21). FIGURE 2-21. Coil-Housing Configuration — Standard Welded Housing With Relief Valve (Option Code W1) Optional: Use drain port to plumb to a safe area (Supplied by user) ¾–14 NPT Conduit Connection 8705-0021A05B ¼'' NPT – 5 psi Pressure Relief Valve 2-20 Installation Process Leak Containment The second optional configuration, identified as option code W3 in the model number, divides the coil housing into three compartments: one for each electrode and one for the coils. Should a damaged liner or electrode fault process fluid to migrate behind the electrode seals, the fluid is contained in the electrode compartment. The sealed electrode compartment prevents the process fluid from entering the coil compartment where it would damage the coils and other internal components. The electrode compartments are designed to contain the process fluid at full line pressure. An o-ring sealed cover provides access to each of the electrode compartments from outside the flowtube; drainports are provided in each cover for the removal of fluid. NOTE The electrode compartment could contain full line pressure and it must be depressurized before the cover is removed. FIGURE 2-22. Housing Configuration — Sealed Electrode Compartment (Option Code W3) Fused Glass Seal O-Ring Seal ¼–NPT Grounding Electrode Port Optional: Use drain port to plumb to a safe area (Supplied by user) 8705-0007ADGB Sealed Electrode Compartment If necessary, capture any process fluid leakage, connect the appropriate piping to the drainports, and provide for proper disposal (see Figure 2-22). 2-21 Rosemount Series 8700 Magnetic Flowmeter Flowtube STEP 7: START-UP AND OPERATION Start-up and operation is dependent upon the transmitter selected to complement the flowtube. For transmitter start-up information, refer to the Model 8712C/U/H manual (document 00809-0100-4729), the Model 8732C manual (document 00809-0100-4725), or the Model 8742C manual (document 00809-0100-4793). Flowtube Calibration Number A unique flowtube calibration number, imprinted on the flowtube tag, enables any flowtube to be used with any Rosemount transmitter without further calibration. Rosemount flow lab tests determine individual flowtube output characteristics. The characteristics are identified by a 16-digit calibration number. The 16-digit calibration number can be programmed into the Model 8712C/U/H or Model 8732C transmitter using the Local Operator Interface (LOI) or the Model 275 HART® Communicator. Refer to the appropriate transmitter product manual for complete instructions. In a FOUNDATION™ fieldbus environment, the Model 8742C can be configured using the DeltaV ™ configuration tool or another FOUNDATION fieldbus configuration device. The calibration number is more than a correction factor, or K- factor, for the flowtube. The first five digits represent the low frequency gain. The ninth through thirteenth digits represent the high frequency gain. Both numbers are normalized from an ideal number of 10000. Standard configurations use the low frequency gain, but in noisy applications it may be worthwhile to switch to the higher frequency. An additional transmitter procedure, called Auto Zero, is required to perform at the higher coil drive frequency. The full procedure can be found in the appropriate transmitter manual. The seventh and eigth digits represent the zero offset at both frequencies where the nominal value is 50. Empty pipe functionality is a transmitter feature that is controlled by the flowtube calibration number. To turn off this feature, change the fourteenth digit to the number 7. Quick Start-up To initiate a basic transmitter start-up, only four parameters are required: • Tube calibration number • Tube size • Units • Analog output range (URV) Refer to the Model 8712C/U/H, Model 8732C, or Model 8742C product manuals for complete transmitter start-up instructions. 2-22 Section 3 SAFETY MESSAGES Troubleshooting Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Refer to the following safety messages before performing any operation in this section. Failure to follow these installation guidelines could result in death or serious injury: Installation and servicing instructions are for use by qualified personnel only. Performing any servicing other than that contained in this manual may result in death or serious injury. Do not perform any servicing other than that contained in the operating instructions, unless qualified. Mishandling products exposed to a hazardous substance may result in death or serious injury. If the product being returned was exposed to a hazardous substance as defined by OSHA, a copy of the required Material Safety Data Sheet (MSDS) for each hazardous substance identified must be included with the returned goods. INDEPENDENT FLOWTUBE AND TRANSMITTER REPLACEMENT Rosemount flow lab tests determine individual flowtube characteristics and account for them with the 16-digit calibration number. Flowtube interchangability reduces the need for spare transmitters. It also ensures factory-calibrated accuracy without additional calibration procedures or equipment. The unique calibration number is loaded into the transmitters, enabling the interface of a Rosemount flowtube to communicate with a Rosemount transmitter. A description of the flowtube calibration number is included on page 2-22. 3-1 Rosemount Series 8700 Magnetic Flowmeter Flowtube FLOWTUBE TROUBLESHOOTING The Model 8712C/U/H, Model 8732C, or Model 8742C Magnetic Flowmeter Transmitters perform self diagnostics on the entire magnetic flowmeter system: the transmitter, the flowtube, and the interconnecting wiring. While most of the diagnostics are related to the transmitter microprocessor, some tests diagnose specific flowtube problems. NOTE Before performing any of the flowtube tests, cut off power and disconnect all connections inside the flowtube junction box. If a problem with the flowtube is identified, the following chart can assist in troubleshooting the flowtube. Take all readings from inside the junction box with a multimeter. Readings taken at the transmitter terminals may provide incorrect or inconclusive information and should be avoided. A flowtube circuit diagram is shown in Figure 3-1. 3-2 Troubleshooting TABLE 3-1. Flowtube Troubleshooting Test A–Flowtube Coil Step 1 Disconnect power from the transmitter by removing the fuse. Disconnect wires 1 and 2 from the transmitter. Step 2 ➤ Measure the resistance across wires 1 and 2 going to the flowtube, using the lowest ohms scale. The reading should be between 2 and 18 V. Step 3 ➤ A reading outside this range indicates that the coils or cables may be open or shorted. Test B– Coil Shield to Coil Step 1 Disconnect power from the transmitter by removing the fuse. Disconnect wires 1, 2, and ground from the transmitter. Step 2 ➤ Measure the resistance from the coil shield (ground) to wires 1 and 2 using the highest scale. Both readings should be overrange. Step 3 ➤ Any reading on the scale indicates that the coils are shorted to the housing. Test C– Electrode Shield to Electrode (See Note Below) Step 1 Disconnect power from the transmitter by removing the fuse. Disconnect wires 17, 18, and 19 from the transmitter. Test with process in the flowtube (either flow or no flow). Step 2 ➤ Measure the resistance from wire 17 to 18 and 17 to 19. This reading will change as you hold the leads on the wires, so use the initial reading. These readings should both be between 1 kV and 3 MV and close to each other. Step 3 ➤ A reading near 68kV or 0V indicates a possible shorted electrode. A stable reading indicates a shorted electrode. A high reading indicates a possible coated electrode, non-conductive process, or electrode not in contact with process. Test D– Positive to Negative Electrode (See Note Below) Step 1 Disconnect power from the transmitter by removing the fuse. Disconnect wires 18, and 19 from the transmitter. Test with process in the flowtube (either flow or no flow). Step 2 ➤ Measure the resistance between wires 18 and 19. This reading should be in the range between 100 kV and 2MV. Step 3 ➤ An overrange reading indicates a coated electrode, non-conductive process, or electrode not in contact with the process. Test E– Coils to Electrode Step 1 Disconnect power from the transmitter by removing the fuse. Disconnect all wires from the transmitter. Step 2 ➤ Measure the resistance between wires 1 or 2 and 18 and between 1 or 2 and 19. Both readings should be in overrange. Step 3 ➤ Any reading on scale indicates a resistive path from the coils to electrode. 3-3 Rosemount Series 8700 Magnetic Flowmeter Flowtube FIGURE 3-1. Flowtube Circuit Diagram 68.1kV (not applicable for flowtubes with N0 hazardous certification approval option code) 68.1kV 8712-0007E04A Flowtube Housing RETURN OF MATERIALS To expedite the return process outside the United States, contact the nearest Fisher-Rosemount representative. The North American Response Center (1-800-654-7768) will assist you with any needed information or materials and is available for users within the United States and Canada, 24 hours a day. The center will ask for product model and serial numbers, and will provide a Return Material Authorization (RMA) number. The center will also ask for the name of the process material the product was last exposed to. Mishandling products exposed to a hazardous substance may result in death or serious injury. A copy of the Material Safety Data Sheet may be required, depending on the nature of the hazardous substance as defined by OSHA. The North American Response Center will provide the additional information and procedures necessary to return goods exposed to hazardous substances. See Safety Messages on page 3-1 for complete warning information. 3-4 Section 4 Specifications: Model 8705 and Model 8707 High-Signal Flowtubes SPECIFICATIONS Functional Specifications Service Conductive liquids and slurries Line Sizes 1 /2–36 inch (15–900 mm) for Model 8705 3–36 inch (80–600 mm) for Model 8707 Interchangeability Model 8705 flowtubes are interchangeable with Model 8712C/U, Model 8732, and Model 8742C Transmitters. Model 8707 High-Signal Flowtubes are interchangeable with Model 8712H High-Signal Transmitters. System accuracy is maintained regardless of line size or optional features. Each flowtube nameplate has a sixteen-digit calibration number that can be entered into a transmitter through the Local Operator Interface (LOI) or the HART Communicator on the Model 8712C/U/H and the Model 8732C. In a FOUNDATION™ fieldbus environment, the Model 8742C can be configured using the DeltaV™ fieldbus configuration tool or another FOUNDATION fieldbus configuration device. No further calibration is necessary. Upper Range Limit 30 ft/s (10 m/s) Process Temperature Limits Teflon (PTFE) Lining –20 to 350 °F (–29 to 177 °C) Tefzel (ETFE) Lining –20 to 300 °F (–29 to 149 °C) Polyurethane Lining 0 to 140 °F (–18 to 60 °C) Neoprene Lining 0 to 185 °F (–18 to 85 °C) Linatex Lining (Not available for Model 8707) 0 to 158 °F (–18 to 70°C) Ambient Temperature Limits –30 to 150 °F (–34 to 65 °C) Ambient Conditions Overvoltage category I. Pollution Degree 2 4-1 Rosemount Series 8700 Magnetic Flowmeter Flowtubes Pressure Limits See Table 1 and Table 2 for flange limits. (Verify that the process temperature does not exceed liner material specifications.) Pressure and Vacuum Limits Full vacuum at maximum lining material temperature; consult factory for vacuum applications that require Teflon (PTFE) lining material and line sizes greater than 6 inches (150 mm) or larger TABLE 4-1. Flowtube Temperature vs. Pressure Limits for ASME B16.5 Class Flanges (1/2- to 24-inch line sizes)(1) Pressure Flange Material Flange Rating @ -20 to 100 °F (-29 to 38 °C) @ 200 °F (93 °C) @ 300 °F (149 °C) @ 350 °F (177 °C) Carbon Steel Class 150 Class 300 Class 150 Class 300 255 psi 740 psi 275 psi 720 psi 260 psi 675 psi 235 psi 600 psi 230 psi 655 psi 205 psi 530 psi 215 psi 645 psi 190 psi 500 psi 304 Stainless Steel (1) 30- and 36-inch AWWA C207 Table 2 Class D rated to 150 psi at 150 °F (66 °C). TABLE 4-2. Flowtube Temperature vs. Pressure Limits for DIN Flanges (15 to 600 mm line sizes) Pressure Flange Material Carbon Steel 304 Stainless Steel Flange Rating @ -29 to 50 °C (-20 to 122 °F) @ 100 °C (212 °F) @ 150 °C (302 °F) @ 175°C (347 °F) PN 10(1) PN 16 PN 25 PN 40 PN 10(1) PN 16 PN 25 PN 40 10 bar 16 bar 25 bar 40 bar 10 bar 16 bar 25 bar 31.1 bar 10 bar 16 bar 25 bar 40 bar 8.4 bar 13.5 bar 21.2 bar 27.5 bar 9.6 bar 15.2 bar 24 bar 37.1 bar 7.6 bar 12.2 bar 19.1 bar 25.8 bar 9 bar 14.2 bar 22.5 bar 34.5 bar 7.2 bar 11.6 bar 18.2 bar 25.1 bar (1) Minimum temperature is -10 °C (14 °F). Submergence Protection IP 68. Continuous to 30 feet (10 meters) Enclosure Rating NEMA 4x. CSA 4 Electrical Rating (Model 8707 only) Coil drive: 185 V pulse dc, 6 Hz, 250 W Electrode: 5 V 1 W Hazardous Location Certifications N0 Factory Mutual (FM) Approval Dust-ignition proof for Class II/III, Division 1 Groups E, F, and G; Non-incendive for Class I, Division 2 Groups A, B, C, and D; T5 temperature code; For non-flammable process fluid service only AND Canadian Standards Association (CSA) Approval Suitable for use in Class I, Division 2 Groups A, B, C, and D; Dust-ignition proof for Class II/III, Division 1, Groups E, F, and G hazardous locations AND CE Marking (Model 8705 only) 4-2 Specifications: Model 8705 and Model 8707 High-Signal Flowtubes N5 Factory Mutual (FM) Approval Dust-ignition proof for Class II/III, Division 1 Groups E, F, and G; Non-incendive for Class I, Division 2 Groups A, B, C, and D; T5 temperature code; For flammable process fluid service KD KEMA/CENELEC (Model 8705 only) EEx e ia IIC T3...T6 (See Table 3) AND CE Marking (Model 8705 only) TABLE 4-3. Relation Between Ambient Temperature, Process Temperature, and Temperature Class(1) Meter Size Maximum Ambient Maximum Process (inches) Temperature Temperature 1 /2 1 1 11/2 11/2 2 2 2 3-4 3-4 3-4 3-4 6 6 6 6 8 - 36 8 - 36 8 - 36 8 - 36 149 °F (65 °C) 149 °F (65 °C) 95 °F (35 °C) 149 °F (65 °C) 140 °F (60 °C) 149 °F (65 °C) 149 °F (65 °C) 104 °F (40 °C) 149 °F (65 °C) 149 °F (65 °C) 131 °F (55 °C) 104 °F (40 °C) 149 °F (65 °C) 149 °F (65 °C) 149 °F (65 °C) 140 °F (60 °C) 149 °F (65 °C) 149 °F (65 °C) 149 °F (65 °C) 149 °F (65 °C) 240 °F (116 °C) 248 °F (120 °C) 95 °F (35 °C) 257 °F (125 °C) 140 °F (60 °C) 257 °F (125 °C) 167 °F (75 °C) 104 °F (40 °C) 266 °F (130 °C) 167 °F (75 °C) 194 °F (90 °C) 104 °F (40 °C) 175 °F (79 °C) 167 °F (75 °C) 230 °F (110 °C) 140 °F (60 °C) 284 °F (140 °C) 240 °F (116 °C) 176 °F (80 °C) 149 °F (65 °C) Temperature Class T3 T3 T4 T3 T4 T3 T4 T5 T3 T4 T5 T6 T3 T4 T5 T6 T3 T4 T5 T6 (1) This table is applicable for KEMA/CENELEC approval only. Conductivity Limits Process liquid must have a conductivity of 5 microsiemens/cm (5 micromhos/cm) or greater for Model 8705. Process liquid must have a conductivity of 50 microsiemens/cm (50 micromhos/cm) for Model 8707 (excludes the effect of interconnecting cable length in remote mount transmitter installations). Performance Specifications (System specifications are given using the frequency output and with the unit at referenced conditions.) Accuracy Model 8705 with Model 8712C/U, Model 8732C, or Model 8742C ±0.5% of rate from 1 to 30 ft/s (0.3 to 10 m/s); includes combined effects of linearity, hysteresis, repeatability, and calibration uncertainty; accuracy is ±0.005 ft/s (±0.0015 m/s) from low-flow cutoff to 1.0 ft/s (0.3 m/s) Model 8707 with Model 8712C/U/H, Model 8732C, or Model 8742C ±0.5% of rate from 3 to 30 ft/s (1 to 10 m/s); include combined effects of linearity, hysteresis, repeatability and calibration uncertainty; accuracy is ±0.015 ft/s (±0.0045 m/s) from low-flow cutoff to 3.0 ft/s (1 m/s) 4-3 Rosemount Series 8700 Magnetic Flowmeter Flowtubes Vibration Effect IEC 770 Pipeline Installation Conditions Mounting Position Effect None when installed to ensure flowtube remains full. Physical Specifications Non-Wetted Materials Flowtube Housing Welded, AISI Type 304 SST or Type 316L SST Flanges Carbon steel, AISI Type 304 SST, or Type 316L SST Paint Polyurethane Process Wetted Materials Lining Teflon (PTFE), Tefzel (ETFE), polyurethane, neoprene, Linatex Electrodes 316L SST, Hastelloy C-276, tantalum, 90% platinum-10% iridium, titanium Process Connections ASME B16.5 (ANSI) Class 150, Class 300, or Class 600 0.5- to 24-inch AWWA C207 Table 2 Class D 30- and 36-inch DIN PN 10, 16, 25, and 40 PN10: Not available for flange sizes from 15 to 150 mm PN16: Not available for flange sizes from 15 to 80 mm PN 25: Not available for flange sizes from 15 to 150 mm PN40: Available for all flange sizes AISI Type 304 SST Sanitary Tri-Clover 3-A approved quick disconnect ferrule-mounted to ASME B16.5 (ANSI) Class 150 flange; 0.5- to 3-inch Electrical Connections Two ¾–14 NPT connections with number 8 screw terminals are provided in the terminal enclosure for electrical wiring. Grounding Electrode A grounding electrode is installed similarly to the measurement electrodes through the flowtube lining. It is available in all electrode materials. Grounding Rings Grounding rings are installed between the flange and the tube face on both ends of the flowtube. They have an I.D. slightly larger than the flowtube I.D. and an external tab to attach ground wiring. Grounding rings are available in 316L SST, Hastelloy-C, titanium, and tantalum. 4-4 Specifications: Model 8705 and Model 8707 High-Signal Flowtubes Lining Protectors Lining protectors are installed between the flange and the tube face on both ends of the flowtube. The leading edge of lining material is protected by the lining protector; lining protectors cannot be removed once they are installed. Lining protectors are available in 316L SST, Hastelloy-C, and titanium. Flowtube Dimensions See Table 5, Table 7, Table 6 See Figure ?-??, Figure ?-??, and Figure ?-?? Weight See Table 4 TABLE 4-4. Flowtube Weight CF = Consult Factory Nominal Line Size(1) Inches (mm) Flowtube Flange Rating ASME B16.5 (ANSI) Flowtube Weight lb (kg) DIN ½ (15) ½ (15) 150 300 PN 40 20 22 (9) (10) 1 (25) 1 (25) 150 300 PN 40 20 22 (9) (10) 1½ (40) 1½ (40) 150 300 PN 40 22 24 (10) (11) 2 (50) 2 (50) 150 300 PN 40 26 28 (12) (13) 3 (80) 3 (80) 150 300 PN 40 40 47 (18) (21) 4 (100) 4 (100) 150 300 PN 16 48 65 (22) (30) 6 (150) 6 (150) 150 300 PN 16 81 93 (37) (42) 8 (200) 8 (200) 150 300 PN 10 110 162 (50) (74) 10 (250) 10 (250) 150 300 PN 10 220 (98) 300 (136) 12 (300) 12 (300) 150 300 PN 10 330 (150) 435 (197) 14 (350) 16 (400) 150 150 PN 10 PN 10 370 (168) 500 (227) 18 (450) 20 (500) 150 150 PN 10 PN 10 600 (272) 680 (308) 24 (600) 150 PN 10 1,000 (454) 30 (750) 36 (900) 125 125 CF CF 1,400 (637) 1,975 (898) (1) 30- and 36-inch AWWA C207 Table 2 Class D rated to 150 psi at 150 °F (66 °C). 4-5 Rosemount Series 8700 Magnetic Flowmeter Flowtubes TABLE 4-5. Model 8705 and Model 8707 Dimensions in Inches (Millimeters). Refer to Dimensional Drawings, Figure 1, Figure 2, and Figure 3 Line Size(1) and Flange Rating (ASME B 16.5) Liner Face Diameter “A” Process Flange Rad. “B” Overall Flowtube Length “L”(2) Body Height “C” Body Width ’“D” Centerline to Conduit “E” Bolt Hole Circle Diameter Bolt Hole Diameter Number and Size of Bolts 0.5–150 0.5 –300 1.38 (35) 1.38 (35) 1.75 (44) 1.88 (48) 7.88 (200) 7.88 (200) 8.75 (222) 8.75 (222) 6.88 (175) 6.88 (175) 5.16 (131) 5.16 (131) 2.38 (60) 2.62 (67) 0.62 (16) 0.62 (16) 4–1/2 4–1/2 1 –150 1 –300 2.00 (51) 2.00 (51) 2.13 (54) 2.44 (62) 7.88 (200) 7.88 (200) 8.75 (222) 8.75 (222) 7.34 (186) 7.34 (186) 5.16 (131) 5.16 (131) 3.12 (79) 3.50 (89) 0.62 (16) 0.75 (19) 4–1/2 4–5/8 1.5 –150 1.5 –300 2.88 (73) 2.88 (73) 2.50 (64) 3.06 (78) 7.88 (200) 7.88 (200) 9.52 (242) 9.52 (242) 7.05 (179) 7.05 (179) 5.57 (141) 5.57 (141) 3.88 (99) 4.50 (114) 0.62 (16) 0.88 (22) 4–1/2 4–3/4 2 –150 2 –300 3.62 (92) 3.62 (92) 3.00 (76) 3.25 (83) 7.88 (200) 7.88 (200) 9.52 (242) 9.52 (242) 7.47 (190) 7.47 (190) 5.57 (141) 5.57 (141) 4.75 (121) 5.00 (127) 0.75 (19) 0.75 (19) 4–5/8 8–5/8 3 –150 3 –300 5.00 (127) 5.00 (127) 3.75 (95) 4.13 (105) 7.88 (200) 8.63 (219) 11.52 (293) 11.52 (293) 9.57 (243) 9.57 (243) 6.57 (167) 6.57 (167) 6.00 (152) 6.62 (168) 0.75 (19) 0.88 (22) 4–5/8 8–3/4 4 –150 4 –300 6.19 (157) 6.19 (157) 4.50 (114) 5.00 (127) 9.84 (250) 10.88 (276) 12.22 (310) 12.22 (310) 10.01 (254) 10.01 (254) 6.92 (176) 6.92 (176) 7.50 (191) 7.88 (200) 0.75 (19) 0.88 (22) 8–5/8 8–3/4 6 –150 6 –300 8.50 (216) 8.50 (216) 5.50 (140) 6.25 (159) 11.81 (300) 13.06 (332) 14.39 (366) 14.39 (366) 10.41 (264) 10.41 (264) 8.05 (204) 8.05 (204) 9.50 (241) 10.62 (270) 0.88 (22) 0.88 (22) 8–3/4 12–3/4 8 –150 8 –300 10.62 (270) 10.62 (270) 6.75 (171) 7.50 (191) 13.78 (350) 15.60 (396) 16.33 (415) 16.33 (415) 11.38 (289) 11.38 (289) 9.02 (229) 9.02 (229) 11.75 (298) 13.00 (330) 0.88 (22) 1.00 (25) 8–3/4 12–7/8 10 –150 10 –300 12.75 (324) 12.75 (324) 8.00 (203) 8.75 (225) 15.00 (381) 17.13 (435) 19.11 (485) 19.11 (485) 17.00 (432) 17.00 (432) 10.44 (265) 10.44 (265) 14.25 (362 15.25 (387) 1.00 (25) 1.12 (28) 12–7/8 16–1 12 –150 12 –300 15.00 (381) 15.00 (381) 9.50 (241) 10.25 (260) 18.00 (457) 20.14 (512) 21.27 (540) 21.27 (540) 19.16 (487) 19.16 (487) 11.52 (293) 11.52 (293) 17.00 (432) 17.75 (451) 1.00 (25) 1.25 (32) 12–7/8 16–11/8 14 –150 14 –300 16.25 (413) 16.25 (413) 10.50 (267) 11.50 (292) 21.00 (533) 23.25 (591) 23.39 (594) 23.39 (594) 21.28 (541) 21.28 (541) 12.58 (320) 12.58 (320) 18.75 (476) 20.25 (514) 1.12 (28) 1.25 (32) 12–1 20–11/8 16 –150 16 –300 18.50 (470) 18.50 (470) 11.75 (298) 12.75 (324) 24.00 (610) 26.25 (667) 25.41 (645) 25.41 (645) 23.30 (592) 23.30 (592) 13.59 (345) 13.59 (345) 21.25 (540) 22.50 (572) 1.12 (28) 1.38 (35) 16–1 20–11/4 18 –150 18 –300 21.00 (533) 21.00 (533) 12.50 (318) 14.00 (356) 27.00 (686) 30.12 (765) 27.93 (709) 27.93 (709) 25.82 (656) 25.82 (656) 14.85 (377) 14.85 (377) 22.75 (578) 24.75 (629) 1.25 (32) 1.38 (35) 16–11/8 24–11/4 20 –150 20 –300 23.00 (584) 23.00 (584) 13.75 (349) 15.25 (387) 30.00 (762) 33.25 (845) 29.95 (761) 29.95 (761) 27.84 (707) 27.84 (707) 15.86 (403) 15.86 (403) 25.00 (635) 27.00 (686) 1.25 (32) 1.38 (35) 20–11/8 24–11/4 24 –150 24 –300 27.25 (692) 27.25 (692) 16.00 (406) 18.00 (457) 36.00 (914) 39.64 (1007) 34.50 (876) 34.50 (876) 32.39 (823) 32.39 (823) 18.14 (461) 18.14 (461) 29.50 (749) 32.00 (813) 1.37 (35) 1.62 (41) 20–11/4 24–11/2 30 36 33.80 (859) 40.27 (1023) 19.38 (492) 23.00 (584) 37.25 (946) 40.75 (1035) 40.41 (1026) 48.29 (1227) 38.50 (928) 46.38 (1178) 21.31 (541) 25.25 (641) 36.00 (914) 42.75 (1086) 1.38 (35) 1.63 (41) 28–11/4 32–11/2 Dimensions with ASME B16.5 (ANSI) Flanges (1) 30- and 36-inch AWWA C207 Table 2 Class D rated to 150 psi at 150 °F. (2) When grounding rings (2 rings per meter) are specified, add 0.25 inch (6.35 mm) for 0.50- through 14-inch (15 through 350 mm) flowtubes, add 0.50 inch (12.7 mm) for 16-inch (400 mm) and larger. When lining protectors are specified, add 0.25 inch (6.35 mm) for ½- through 12-inch (15 through 300 mm) flowtubes, add 0.50 inch (12.7 mm) for 14- through 36-inch (350 through 900 mm) flowtubes. TABLE 4-6. Flowtube Dimensions in inches (millimeters) Line Size and Flange Rating Nominal Tri-Clamp Process Flange Rad. Body Height “C” Max Diameter “B” “E” Overall Flowtube Length “L” 0.5–150 lb. 1.00 (25) 1.75 (44) 8.38 (213) 5.16 (131) 13.78 (350) 1–150 lb. 1.50 (40) 2.13 (54) 8.38 (213) 5.16 (131) 13.78 (350) 1.5–150 lb. 2.00 (50) 2.50 (64) 9.00 (229) 5.56 (141) 13.78 (350) 2–150 lb. 3.00 (80) 3.00 (76) 9.00 (229) 5.56 (141) 13.78 (350) 3–150 lb. 4.00 (100) 3.75 (95) 12.00 (305) 6.57 (167) 13.78 (350) Dimensions with ASME B16.5 (ANSI) Flanges and Tri-Clamp Adapters. 4-6 Centerline to Conduit Specifications: Model 8705 and Model 8707 High-Signal Flowtubes TABLE 4-7. Model 8705 Flowtube Dimensions with DIN Flanges in Millimeters (Inches) Liner Face Diameter “A” Process Flange Rad. “B” Overall Flowtube Length “L”(2) Body Height “C” Body Width “D” with Port Centerline to Conduit “E” Bolt Hole Circle Diameter Bolt Hole Diameter Number of Bolts 15 mm PN 10–40 45 (1.77) 47 (1.87) 200 (7.88) 222 (8.75) 175 (6.88) 131 (5.16) 65 (2.56) 14 (0.55) 4 25 mm PN 10–40 68 (2.68) 58 (2.27) 200 (7.88) 222 (8.75) 186 (7.34) 131 (5.16) 85 (3.35) 14 (0.55) 4 40 mm PN 10–40 88 (3.46) 75 (2.96) 200 (7.87) 242 (9.52) 179 (7.05) 141 (5.57) 110 (4.33) 18 (0.71) 4 50 mm PN 10–40 102 (4.02) 83 (3.25) 200 (7.87) 242 (9.52) 190 (7.47) 141 (5.57) 125 (4.92) 18 (0.71) 4 80 mm PN 10–40 138 (5.43) 100 (3.94) 200 (7.87) 293 (11.52) 243 (9.57) 167 (6.57) 160 (6.30) 18 (0.71) 8 100 mm PN 10–16 158 (6.22) 110 (4.33) 250 (9.84) 310 (12.22) 254 (10.01) 176 (6.92) 180 (7.09) 18 (0.71) 8 100 mm PN 25–40 162 (6.38) 117 (4.63) 250 (9.84) 310 (12.22) 254 (10.01) 176 (6.92) 190 (7.48) 22 (0.87) 8 Line Size(1) and Flange Rating 150 mm PN 10–16 212 (8.35) 142 (5.61) 300 (11.81) 366 (14.39) 264 (10.41) 204 (8.05) 240 (9.45) 22 (0.87) 8 150 mm PN 25 218 (8.58) 150 (5.91) 300 (11.81) 366 (14.39) 264 (10.41) 204 (8.05) 240 (9.45) 22 (0.87) 8 150 mm PN 40 218 (8.58) 150 (5.91) 332 (13.06) 366 (14.39) 264 (10.41) 204 (8.05) 240 (9.45) 22 (0.87) 8 200 mm PN 10 268 (10.55) 170 (6.70) 351 (13.81) 415 (16.33) 289 (13.38) 229 (9.02) 295 (11.61) 22 (0.87) 8 200 mm PN 16 268 (10.55) 170 (6.70) 351 (13.81) 415 (16.33) 289 (13.38) 229 (9.02) 295 (11.61) 22 (0.87) 8 200 mm PN 25 278 (10.94) 180 (7.09) 350 (13.78) 415 (16.33) 289 (13.38) 229 (9.02) 310 (12.20) 26 (1.02) 12 200 mm PN 40 285 (11.22) 187 (7.38) 396 (15.60) 415 (16.33) 289 (13.38) 229 (9.02) 320 (12.60) 30 (1.18) 12 250 mm PN 10 320 (12.60) 197 (7.70) 381 (15.00) 485 (19.11) 432 (17.00) 265 (10.44) 350 (13.78) 22 (0.87) 12 250 mm PN 16 320 (12.60) 202 (7.97) 381 (15.00) 485 (19.11) 432 (17.00) 265 (10.44) 355 (13.98) 26 (1.02) 12 250 mm PN 25 335 (13.19) 213 (8.39) 381 (15.00) 485 (19.11) 432 (17.00) 265 (10.44) 370 (14.67) 30 (1.18) 12 250 mm PN 40 345 (13.58) 225 (8.86) 435 (17.13) 485 (19.11) 432 (17.00) 265 (10.44) 385 (15.16) 33 (1.30) 12 300 mm PN 10 370 (14.57) 223 (8.76) 457 (18.00) 540 (21.27) 487 (19.16) 265 (10.44) 400 (15.75) 22 (0.87) 12 300 mm PN 16 378 (14.88) 230 (9.06) 457 (18.00) 540 (21.27) 487 (19.16) 293 (11.52) 410 (16.14) 26 (1.02) 12 300 mm PN 25 395 (15.55) 242 (9.55) 457 (18.00) 540 (21.27) 487 (19.16) 293 (11.52) 430 (16.93) 30 (1.18) 16 300 mm PN 40 410 (16.14) 258 (10.12) 512 (20.14) 540 (21.27) 487 (19.16) 293 (11.52) 450 (17.72) 33 (1.30) 16 350 mm PN 10 430 (16.93) 252 (9.94) 534 (21.03) 594 (23.39) 541 (21.28) 293 (11.52) 460 (18.11) 22 (0.87) 16 350 mm PN 16 438 (17.24) 260 (10.24) 534 (21.03) 594 (23.39) 541 (21.28) 320 (12.58) 470 (18.50) 26 (1.02) 16 350 mm PN 25 450 (17.72) 277 (10.93) 534 (21.03) 594 (23.39) 541 (21.28) 320 (12.58) 490 (19.29) 33 (1.30) 16 350 mm PN 40 465 (18.31) 290 (11.42) 591 (23.25) 594 (23.39) 541 (21.28) 320 (12.58) 510 (20.08) 36 (1.42) 16 400 mm PN 10 482 (18.98) 282 (11.12) 610 (24.00) 645 (25.04) 592 (23.30) 345 (13.59) 515 (20.28) 26 (1.02) 16 400 mm PN 16 490 (19.29) 290 (11.42) 610 (24.00) 645 (25.04) 592 (23.30) 345 (13.59) 525 (20.67) 30 (1.18) 16 400 mm PN 25 505 (19.88) 310 (12.21) 610 (24.00) 645 (25.04) 592 (23.30) 345 (13.59) 550 (21.65) 36 (1.42) 16 400 mm PN 40 535 (21.06) 330 (12.99) 667 (26.25) 645 (25.04) 592 (23.30) 345 (13.59) 585 (23.03) 39 (1.54) 16 450 mm PN 10 532 (20.94) 308 (12.13) 686 (27.00) 709 (27.93) 656 (25.82) 377 (14.85) 565 (22.24) 26 (1.02) 20 450 mm PN 16 550 (21.65) 320 (12.60) 686 (27.00) 709 (27.93) 656 (25.82) 377 (14.85) 585 (23.03) 30 (1.18) 20 450 mm PN 40 560 (22.05) 343 (13.50) 765 (30.12) 709 (27.93) 656 (25.82) 377 (14.85) 610 (24.02) 30 (1.18) 20 500 mm PN 10 585 (23.03) 335 (13.19) 762 (30.00) 761 (29.95) 707 (27.84) 403 (15.86) 620 (24.41) 26 (1.02) 20 500 mm PN 16 610 (24.02) 358 (14.08) 762 (30.00) 761 (29.95) 707 (27.84) 403 (15.86) 650 (25.59) 33 (1.30) 20 500 mm PN 25 615 (24.21) 365 (14.37) 762 (30.00) 761 (29.95) 707 (27.84) 403 (15.86) 660 (25.98) 36 (1.42) 20 500 mm PN 40 615 (24.21) 378 (14.88) 845 (33.25) 761 (29.95) 707 (27.84) 403 (15.86) 670 (26.38) 42 (1.65) 20 600 mm PN 10 685 (26.97) 390 (15.36) 914 (36.00) 885 (34.85) 823 (32.39) 461 (18.14) 725 (28.54) 30 (1.18) 20 600 mm PN 16 725 (28.54) 420 (16.54) 914 (36.00) 877 (34.51) 823 (32.39) 461 (18.14) 770 (30.31) 36 (1.42) 20 600 mm PN 25 720 (28.35) 423 (16.64) 914 (36.00) 877 (34.51) 823 (32.39) 461 (18.14) 770 (30.31) 39 (1.54) 20 600 mm PN 40 735 (18.94) 445 (17.52) 1,007 (39.64) 886 (34.88) 823 (32.39) 461 (18.14) 795 (31.30) 48 (1.88) 20 Dimensions with DIN Flanges (1) Consult factory for larger line sizes. (2) When grounding rings (2 rings per meter) are specified, add 6.35 mm (0.25 in.) for 15 mm through 350 mm (½- through 14 in.) flowtubes or 12.7 mm (0.50 in.) for 400 mm (16 in.) and larger. When lining protectors are specified, add 6.35 mm (0.25 in.) for 15 mm through 300 mm (½- through 12-in.) flowtubes, 12.7 mm (0.50 in.) for 350 mm through 900 mm (14- through 36-in.) flowtubes. 4-7 Rosemount Series 8700 Magnetic Flowmeter Flowtubes Figure 4-1. Dimensional Drawing of Model 8705 and Model 8707 Flowtubes, Typical of 1/2- through 4-inch (15 through 100 mm) Line Sizes with Option Code W1, Housing Configuration 0.66 (17) 4.75 (121) 2.00 (51) 1.00 (25) 1.35 (34) 2.10 (53) 2.10 (53) 3 /4–14 NPT Conduit Connection “E” “C” Relief Valve 8705-1002A05A,1002B05A1 “B” “A” “L” NOTES Dimensions are in inches (millimeters). See Table 5 and 7 for variable dimensions. Figure 4-2. Dimensional Drawing of Model 8705 and Model 8707 Flowtubes, Typical of 6- through 36-inch (150 through 900 mm) Line Sizes with Option Code W3, Housing Configuration 0.66 (17) 3 /4–14 NPT Conduit Connection 2.00 (51) 1.00 (25) 4.75 (121) 1.35 (34) 2.10 (53) 2.10 (53) “E” “C” “A” “D” NOTES Dimensions are in inches (millimeters). Consult factory for larger line sizes. See Table 5 and 7 for variable dimensions. 4-8 Separate Electrode Compartments “L” 8705-1002A06A,1002B06A “B” Specifications: Model 8705 and Model 8707 High-Signal Flowtubes Figure 4-3. Dimensional Drawing of Model 8705 Sanitary Flowtubes, Typical of 1/2- through 3-inch (15 through 86 mm) Line Sizes with Option Code W0, Housing Configuration 2.00 (51) 1.00 (25) 4.75 (121) 0.66 (17) 1.35 (34) “E” 2.10 (53) 3 /4–14 NPT Conduit Connection “C” “L” NOTES Dimensions are in inches (millimeters). See Table 7 for variable dimensions. 8705-1002A05B, 1002B05C “B” 4-9 Section 5 Specifications: Model 8711 Flowtube SPECIFICATIONS Functional Specifications Service Conductive liquids and slurries Line Sizes 0.15- through 8-inch (4 through 200 mm) Interchangeability Model 8711 Flowtubes are interchangeable with Model 8712C/U, Model 8732, and Model 8742C Transmitters. System accuracy is maintained regardless of line size or optional features. Each flowtube nameplate has a sixteen-digit calibration number that can be entered into a transmitter through the Local Operator Interface (LOI) or the HART Communicator on the Model 8712C/U/H and the Model 8732C. In a FOUNDATION fieldbus environment, the Model 8742C can be configured using the DeltaV fieldbus configuration tool or another FOUNDATION fieldbus configuration device. No further calibration is necessary. Upper Range Limit 30 ft/s (10 m/s) Process Temperature Limits Tefzel (ETFE) Lining –20 to 300 °F (–29 to 149 °C) for 0.5- through 8-inch (15–200 mm) line sizes –20 to 200 °F (–29 to 93 °C) for 0.15- and 0.3-inch (4 and 8 mm) line sizes Teflon (PTFE) Lining -20 to 350 °F (-29 to 177 °C) Ambient Temperature Limits –30 to 150 °F (–34 to 65 °C) Ambient Conditions Overvoltage Category I. Pollution Degree 2 Pressure and Vacuum Limits Line sizes 0.5- through 8-inch (15 through 200 mm): full vacuum to 740 psi (51.0 bar); line sizes 0.15- through 0.30-inch (4 through 8 mm): full vacuum to 288 psi (19.8 bar); consult factory for vacuum applications that require Teflon (PTFE) lining material Enclosure Rating NEMA 4x. CSA Type 4 5-1 Rosemount Series 8700 Magnetic Flowmeter Flowtubes Hazardous Location Certifications N0 Factory Mutual (FM) Approval Dust-ignition proof for Class II/III, Division 1 Groups E, F, and G; Non-incendive for Class I, Division 2 Groups A, B, C, and D; T5 temperature code; For non-flammable process fluid service only AND Canadian Standards Association (CSA) Approval Suitable for use in Class I, Division 2 Groups A, B, C, and D; Dust-ignition proof for Class II/III, Division 1, Groups E, F, and G hazardous locations AND CE Marking. N5 Factory Mutual (FM) Approval; Dust-ignition proof for Class II/III, Division 1 Groups E, F, and G hazardous locations; Non-incendive for Class I, Division 2 Groups A, B, C, and D; T5 temperature code; For flammable process fluid service E5 Factory Mutual (FM) Approval Explosion Proof for Class I, Division 1, Groups C and D; T6 temperature code; flammable process fluid service; Available for remote mount transmitter or integral mount Model 8732C transmitters CD KEMA/CENELEC Approval EEx e ia IIC T3...T6 (See Table 5-1) Conductivity Limits Process liquid must have a conductivity of 5 microsiemens/cm (5 micromhos/cm) or greater for Model 8711. Excludes the effect of interconnecting cable length in remote mount transmitter installations TABLE 5-1. Relation Between Ambient Temperature, Process Temperature, and Temperature Class(1) Meter Size (inches) 1 /2 1 1 11/2 11/2 2 2 2 3-4 3-4 3-4 3-4 6 6 6 6 8 - 36 8 - 36 8 - 36 8 - 36 Maximum Ambient Temperature Maximum Process Temperature Temperature Class 149 °F (65 °C) 149 °F (65 °C) 95 °F (35 °C) 149 °F (65 °C) 140 °F (60 °C) 149 °F (65 °C) 149 °F (65 °C) 104 °F (40 °C) 149 °F (65 °C) 149 °F (65 °C) 131 °F (55 °C) 104 °F (40 °C) 149 °F (65 °C) 149 °F (65 °C) 149 °F (65 °C) 140 °F (60 °C) 149 °F (65 °C) 149 °F (65 °C) 149 °F (65 °C) 149 °F (65 °C) 240 °F (116 °C) 248 °F (120 °C) 95 °F (35 °C) 257 °F (125 °C) 140 °F (60 °C) 257 °F (125 °C) 167 °F (75 °C) 104 °F (40 °C) 266 °F (130 °C) 167 °F (75 °C) 194 °F (90 °C) 104 °F (40 °C) 175 °F (79 °C) 167 °F (75 °C) 230 °F (110 °C) 140 °F (60 °C) 284 °F (140 °C) 240 °F (116 °C) 176 °F (80 °C) 149 °F (65 °C) T3 T3 T4 T3 T4 T3 T4 T5 T3 T4 T5 T6 T3 T4 T5 T6 T3 T4 T5 T6 (1) This table applicable for KEMA/CENELEC approval only. 5-2 Specifications: Model 8711 Flowtube Performance Specifications (System specifications are given using the frequency output and with the unit at referenced conditions.) Accuracy Model 8711 with Model 8712C/U, Model 8732C, or Model 8742C Transmitters ±0.5% of rate from 3 to 30 ft/s (1 to 10 m/s); include combined effects of linearity, hysteresis, repeatability and calibration uncertainty; accuracy is ±0.015 ft/s (±0.0045 m/s) from low-flow cutoff to 3.0 ft/s (1 m/s) Vibration Effect IEC 770 Pipeline Installation Conditions Mounting Position Effect None when installed to ensure flowtube remains full. Physical Specifications Non-Wetted Materials Flowtube 303 SST (ASTM A-743) Coil Housing Investment cast steel (ASTM A-27) Paint Polyurethane Process-Wetted Materials Lining Tefzel (ETFE), Teflon (PTFE) Electrodes 316L SST, Hastelloy C-276, tantalum, 90% platinum—10% iridium, titanium Process Connections Mounts between these Flange Configurations ASME B16.5 (ANSI): Class 150, 300.vv DIN: PN 10 and 25 BS: 10 Table D, E, and F 5-3 Rosemount Series 8700 Magnetic Flowmeter Flowtubes Studs, Nuts, and Washers(1) ASME B16.5 (ANSI) Line sizes 0.15- through 1-inch (4 through 25 mm): 316 SST, ASTM A193, Grade B8M, Class 1 threaded mounting studs; ASTM A194, Grade 8M heavy hex nuts; SAE per ANSI B18.2.1, Type A, Series N flat washers Line sizes 1.5- through 8-inch (40 through 200 mm): CS, ASTM A193, Grade B7, Class 1 threaded mounting studs; ASTM A194, Grade 2H heavy hex nuts; SAE per ANSI B18.2.1, Type A, Series N flat washers; all items clear, chromate zinc-plated DIN Line sizes 4 through 25 mm (0.15- through 1-inch): 316 SST ASTM A193, Grade B8M Class 1 threaded mounting studs; ASTM A194, Grade 8M, DIN 934 H=D, metric heavy hex nuts; 316 SST, A4, DIN 125 flat washers Line sizes 40 through 200 mm (1.5- through 8-inch): CS, ASTM A193, Grade B7 threaded mounting studs; ASTM A194, Grade 2H, DIN 934 H=D, metric heavy hex nuts; CS, DIN 125 flat washers; all items yellow zinc-plated Electrical Connections Two ¾–14 NPT connections with number 8 screw terminals are provided in the terminal enclosure for electrical wiring. Grounding Electrode A grounding electrode is installed similarly to the measurement electrodes through the flowtube lining. It is available in all electrode materials. Grounding Rings Grounding rings are installed between the flange and the tube face on both ends of the flowtube. They have an I.D. slightly larger than the flowtube I.D. and an external tab to attach ground wiring. Grounding rings are available in 316L SST, Hastelloy C-276, titanium, and tantalum. Flowtube Dimensions and Weight See Table 1 TABLE 1. Flowtube Dimensions and Weight Nominal Line Size Inches (mm) 0.15 0.30 0.5 1 1.5 2 3 4 6 8 (4) (8) (15) (25) (40) (50) (80) (100) (150) (200) Flowtube Housing Dimensions “A” Max. 4.00 4.00 4.00 4.31 4.42 4.64 5.26 5.87 6.97 8.00 (102) (102) (102) (109) (112) (118) (134) (149) (177) (2003) “B” 5.44 5.44 5.44 6.06 7.41 7.94 9.19 10.41 12.60 14.66 (1) 5-4 Flowtube Length Inside Diameter “D” “C” (138) (138) (138) (154) (188) (202) (233) (264) (320) (372) 3.56 3.56 3.56 4.50 3.28 3.91 5.16 6.38 8.56 10.63 (90) (90) (90) (114) (83) (99) (131) (162) (217) (270) 2.17 2.17 2.17 2.17 2.73 3.26 4.68 5.88 6.87 8.86 (55) (55) (55) (55) (69) (83) (119) (149) (174) (225) .165 .287 .595 .959 1.50 1.95 2.98 3.90 5.825 7.87 (4) (7) (15) (24) (38) (50) (76) (99) (148) (200) Weight lb (kg) 4 4 4 5 5 7 13 22 35 60 (2) (2) (2) (2) (2) (3) (6) (10) (16) (27) 0.15 and 0.30 inch (4 and 80 mm) flowtubes mount between 1/2-inch (13 mm) flange. Specifications: Model 8711 Flowtube Figure 5-1. Model 8711 Dimensional Drawings (1.5-inch through 8-inch line sizes) 0.66 (17) 1.35(34) 2.10 (53) “A” “B” “C” 4.75 (121) 0.5 (13) NOTE Dimensions are in inches (millimeters). See Table 1 for variable dimensions. “D” 8711-1012B04A, 1012A04A Flow 5-5 Rosemount Series 8700 Magnetic Flowmeter Flowtubes Figure 5-2. Model 8711 Dimensional Drawings (0.15-inch through 1-inch line sizes). 4.5 (121) 0.87 (22) 2.00 (51) 0.66 (17) 1.35 (34) 2.10 (53) 3.37 (86) “A” “B” ¾–14 NPT Conduit Connection (2 places) “C” 4.75 (121) 0.50 (13) Grounding Clamp NOTE Dimensions are in inches (millimeters). See Table 1 for variable dimensions. 5-6 “D” 8711-1012B03A, 1012A03A Flow Appendix A Field-Removable Electrodes SAFETY MESSAGES Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Failure to follow these installation guidelines could result in death or serious injury: Installation and servicing instructions are for use by qualified personnel only. Performing any servicing other than that contained in this manual may result in death or serious injury. Do not perform any servicing other than that contained in the operating instructions. The field-removable electrode option allows the user to remove the electrode assembly with the flowtube still mounted in the line. Frequently, this option will be used for cleaning the electrode head when coating is of concern. The flowtube should be drained of any process fluid prior to disassembly of the electrodes. To avoid personal injury, care should be taken when handling electrodes that have been in contact with corrosive process fluids. Take care to avoid rotating the electrode when removing it to avoid damage to the o-ring. Some resistance may be experienced due to the tight o-ring fit. REMOVE THE ELECTRODE ASSEMBLY Use the following procedure to remove the electrode assembly from the flowtube. 1. Drain the flowtube of any process fluid prior to disassembly of the electrodes. 2. Remove the screws that secure the electrode cover. 3. Remove the electrode cover and o-ring. It is generally recommended that new o-rings be installed on the electrode and the electrode cover during reassembly. 4. Remove the electrode lead screw that secures the signal wire to the electrode. 5. Remove the electrode retaining nut. 6. Take the electrode from the electrode housing by pulling it straight out, with firm pressure, along the axis of the electrode. A-1 Rosemount Series 8700 Magnetic Flowmeter Flowtube REPLACE THE ELECTRODE ASSEMBLY Use the following procedure to replace the electrode assembly into the flowtube. 1. Lubricate the o-ring. 2. Install the o-ring on the electrode. 3. Insert the electrode into the electrode housing. Push straight in until the electrode is seated. Avoid rotating the electrode or the electrode housing as this could result in leakage. 4. Secure the electrode into the housing with the retaining nut and lock washer. Tighten the retaining nut to 15 in/oz of torque. Failure to tighten the fasteners can cause loss of liquid tight seal and result in damage to the unit. 5. Secure the signal wire to the electrode with the electrode lead screw. 6. Install the o-ring into the electrode cover. 7. Secure the electrode cover to the flowtube with the screws. FIGURE A-1. The Field-Replaceable Electrode Lock Washer Housing Nut Lock Washer Housing Nut O-rIng Electrode Lead Screw Spring Lockwasher A-2 Insulator Cap 8705-1002B03A Electrode Index Flowtube Orientation . . . . . . . . . . . . 2-3 A Accuracy Ensuring . . . . . . . . . . . . . . . . . . . . . 2-3 Model 8705/8707 . . . . . . . . . . . . . . 5-3 Model 8711 . . . . . . . . . . . . . . . . . . . 6-3 C Cable Preparation . . . . . . . . . . . . . . 2-15 Calibration . . . . . . . . . . . . . . . . . 2-2, 2-22 Canadian Standards Association Approval Model 8705/8707 . . . . . . . . . . .5-2, 6-2 Conductivity Model 8705/8707 . . . . . . . . . . . . . . 5-3 Model 8711 . . . . . . . . . . . . . . . . . . . 6-2 Conduit Ports and Connections . . 2-13 G Gaskets . . . . . . . . . . . . . . . . . . . . . 2-5, 2-8 Spiral-wound . . . . . . . . . . . . . . . . . . 2-1 Grounding . . . . . . . . . . . . . . . . . . . . . 2-10 Grounding Electrodes . . . . . . . . . . 2-10 Grounding Rings . . . . . . . . . . . . . . 2-10 Lining Protectors . . . . . . . . . . . . . 2-10 Grounding Rings . . . . . . . . . . . . . . . . . 2-5 H Handling . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Hazardous Locations Certifications Model 8705/8707 . . . . . . . . . . . . . . . 5-2 Model 8711 . . . . . . . . . . . . . . . . . . . 6-2 D I Dedicated Conduit . . . . . . . . . . . . . . 2-13 Dimensional Drawing Model 8705/8707 . . . . . . . . . . .5-8, 5-9 Model 8711 . . . . . . . . . . . . . . . .6-5, 6-6 Dimensions Model 8705/8707 . . . . . . . . . . . . . . 5-5 Model 8711 . . . . . . . . . . . . . . . . . . . 6-4 Direction . . . . . . . . . . . . . . . . . . . . . . . 2-4 Downstream/Upstream Piping . . . . 2-3 Drawings Model 8705 . . . . . . . . . . . . . . . . . . . 2-5 Model 8707 High-Signal . . . . . . . . 2-5 Model 8711 . . . . . . . . . . . . . . . . . . . 2-8 Independent Replacement . . . . . . . . 4-1 Input Power . . . . . . . . . . . . . . . . . . . . 2-14 Installation . . . . . . . . . . . . . . . . . . . . . . 2-1 Installation Drawings Model 8705 . . . . . . . . . . . . . . . . . . . 2-5 Model 8707 High-Signal . . . . . . . . . 2-5 Model 8711 . . . . . . . . . . . . . . . . . . . 2-8 E Electrical Considerations Input Power . . . . . . . . . . . . . . . . . 2-14 Electrical Rating . . . . . . . . . . . . . . . . 5-2 Empty pipe functionality . . . . . . . . 2-22 Enclosure Rating . . . . . . . . . . . . . . . . 5-2 F Factory Mutual Approval Model 8705/8707 . . . . . . . 5-2, 5-3, 6-2 Flange Bolts . . . . . . . . . . . . . . . . . . . . 2-6 Flanges Class 150 . . . . . . . . . . . . . . . . .2-6, 2-9 Class 300 . . . . . . . . . . . . . . . . .2-6, 2-9 Flow Direction . . . . . . . . . . . . . . .2-4, 2-5 Flowtube Calibration Number . . . 2-22 16-digit Calibration Number . . . . 2-22 empty pipe functionality . . . . . . . 2-22 K KEMA/CENELEC Approval Model 8705/8707 . . . . . . . . . . . . . . . 5-3 Model 8711 . . . . . . . . . . . . . . . . . . . 6-2 L Lining Protectors . . . . . . . . . . . . . . . 2-10 M Model 8712C/U Transmitter Connections . . . . . . . 2-16 Mounting . . . . . . . . . . . . . . . . . . . . . . . 2-2 Model 8705/8707 . . . . . . . . . . . . . . . 5-4 Model 8711 . . . . . . . . . . . . . . . . . . . 6-3 O Operation . . . . . . . . . . . . . . . . . . . . . . 2-22 Orientation . . . . . . . . . . . . . . . . . . . . . . 2-3 P Piping . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Positioning Flowtube Horizontal/Inclined . . . . . . . . . . . . . 2-3 I-1 Rosemount Series 8700 Magnetic Flowmeter Flowtubes ASME/ANSI . . . . . . . . . . . . . . . 6-3 conductivity limits . . . . . . . . . . 6-2 dimensional drawings . . . 6-5, 6-6 electrical connections . . . . . . . 6-4 electrodes . . . . . . . . . . . . . . . . . 6-3 enclosure rating . . . . . . . . . . . . 6-1 flowtube dimensions . . . . . . . . 6-4 functional specifications . . . . . 6-1 grounding electrode . . . . . . . . . 6-4 grounding rings . . . . . . . . . . . . 6-4 hazardous locations . . . . . . . . . 6-2 interchangeability . . . . . . . . . . 6-1 line sizes . . . . . . . . . . . . . . . . . . 6-1 lining . . . . . . . . . . . . . . . . . . . . 6-3 mounting position effect . . . . . 6-3 non-wetted materials . . . . . . . 6-3 performance specifications . . . 6-3 physical specifications . . . . . . . 6-3 pressure and vacuum limits . . 6-1 process conditions . . . . . . . . . . 6-3 process temperature limits . . . 6-1 process-wetted materials . . . . 6-3 service . . . . . . . . . . . . . . . . . . . 6-1 upper range limit . . . . . . . . . . . 6-1 vibration effect . . . . . . . . . . . . . 6-3 weight . . . . . . . . . . . . . . . . . . . . 6-4 Pressure limits . . . . . . . . . . . . . . . . . . . . . . . . 5-2 Model 8705/8707 . . . . . . . . . . . . . . 5-2 Process Leak Containment . . . . . . . . . . . . . . . . 2-21 Protection . . . . . . . . . . . . . . . . . . . 2-19 Q Quick Start-up . . . . . . . . . . . . . . . . . 2-22 R Relief Valves . . . . . . . . . . . . . . . . . . . . A-2 Return of Materials . . . . . . . . . . . . . . 4-4 S Safety . . . . . . . . . . . . . . . . . . . . . . . . . . A-1 Specifications Model 8705 and Model 8707 . . . . . 5-1 accuracy . . . . . . . . . . . . . . . . . . 5-3 ambient temperature limits . . 5-1 ASME/ANSI . . . . . . . . . . . . . . 5-4 conductivity limits . . . . . . . . . 5-3 dimensional drawing . . . .5-8, 5-9 electrical connections . . . . . . . 5-4 electrodes . . . . . . . . . . . . . . . . 5-4 enclosure rating . . . . . . . . . . . 5-2 flowtube dimensions . . . . . . . . 5-5 functional specifications . . . . . 5-1 grounding electrode . . . . . . . . 5-4 grounding rings . . . . . . . . . . . 5-4 hazardous locations . . . . . . . . 5-2 interchangeability . . . . . . . . . 5-1 line sizes . . . . . . . . . . . . . . . . . 5-1 lining . . . . . . . . . . . . . . . . . . . . 5-4 lining protectors . . . . . . . . . . . 5-5 mounting position effect . . . . . 5-4 non-wetted materials . . . . . . . 5-4 performance specifications . . . 5-3 physical specifications . . . . . . 5-4 pressure limits . . . . . . . . . . . . 5-2 process conditions . . . . . . . . . . 5-4 process temperature limits . . 5-1 process wetted materials . . . . 5-4 service . . . . . . . . . . . . . . . . . . . 5-1 submergence protection . . . . . 5-2 upper range limit . . . . . . . . . . 5-1 vacuum limits . . . . . . . . . . . . . 5-2 vibration effect . . . . . . . . . . . . 5-4 weight . . . . . . . . . . . . . . . . . . . 5-5 Model 8707 electrical rating . . . . . . . . . . . . Model 8711 . . . . . . . . . . . . . . . . . . . accuracy . . . . . . . . . . . . . . . . . . ambient conditions . . . . . . . . . ambient temperature limits . . I-2 5-2 6-1 6-3 6-1 6-1 Spiral-wound Gaskets . . . . . . . . . . . . 2-1 Start-up . . . . . . . . . . . . . . . . . . . . . . . . 2-22 T Temperature . . . . . . . . . . . . . . . . . . . 2-13 Model 8705/8707 . . . . . . . . . . . . . . . 5-1 Model 8711 . . . . . . . . . . . . . . . . . . . 6-1 Transmitter Connections High-Signal . . . . . . . . . . . . . . . . . . 2-17 Model 8712C/U . . . . . . . . . . . . . . . 2-16 Model 8732C . . . . . . . . . . . . . . . . . 2-18 Transporting System . . . . . . . . . . . . . 2-2 Troubleshooting . . . . . . . . . . . . . . . . . 4-1 Chart . . . . . . . . . . . . . . . . . . . . . . . . 4-2 U Upstream/Downstream Piping . . . . 2-3 V Vacuum Limits . . . . . . . . . . . . . . . . . . 5-2 Vibration Model 8705/8707 . . . . . . . . . . . . . . . 5-4 Model 8711 . . . . . . . . . . . . . . . . . . . 6-3 W Warning . 1-1, 2-1, 2-13, 2-16, 2-17, 2-18, 4-1 Index Weight Model 8705/8707 . . . . . . . . . . . . . . 5-5 Model 8711 . . . . . . . . . . . . . . . . . . . 6-4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . Cable Preparation . . . . . . . . . . . . Conduit Ports and Connections . Dedicated Conduit . . . . . . . . . . . . Temperature . . . . . . . . . . . . . . . . 2-13 2-15 2-13 2-13 2-13 I-3 This document is available electronically at www.rosemount.com. Fisher-Rosemount Groeneveldselaan 6-8 3909 AZ 3900 AB Veenendaal The Netherlands Tel 31 (0) 318 549 549 Fax 31 (0) 318 549 559 Singapore PTE LTD 1 Pandan Crescent Singapore 128461 Tel 31 (65) 777-8211 Fax 31 (65) 777-0947 email: AP.RMT-Specialist@frco.com PR Fisher-Rosemount Flow 8200 Market Boulevard Chanhassen, MN 55317 USA Tel 1-800-999-9307 Fax (952) 949-7001 © 2000 Rosemount Inc. INT ED Rosemount Inc. IN U. S. A. http://www.rosemount.com ¢00809-0100-4727X¤ 00809-0100-4727 Rev. DA 4/00
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.2 Linearized : Yes Create Date : 1910:00:50 21:45:24 Producer : Acrobat Distiller 3.01 for Windows Author : Kenneth Brandt Title : Creator : FrameMaker 5.5.2L11a Modify Date : 2000:05:04 13:23:06 Page Count : 53 Page Mode : UseOutlinesEXIF Metadata provided by EXIF.tools