P6209 4 GHz 5X Active Probe Instruction Manual T_P6209 T
User Manual: T_P6209
Open the PDF directly: View PDF
.
Page Count: 74
| Download | |
| Open PDF In Browser | View PDF |
Instruction Manual P6209 4 GHz 5X Active Probe 071-0634-04 Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service. www.tektronix.com Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved. Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077 TEKTRONIX, TEK, TEKPROBE, and SureFoot are registered trademarks, and SureToe, FlexLead, and KlipChip are trademarks of Tektronix, Inc. WARRANTY Tektronix warrants that this product will be free from defects in materials and workmanship for a period of one (1) year from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations. This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product. THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES. Table of Contents General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v vii viii Product Description Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Features and Standard Accessories . . . . . . . . . . . . . . . . . . . . . . Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 2 7 Configuration Probe Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Functional Check Operating Basics Handling the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Maximum Nondestructive Input Voltage . . . . . . . . . . . . . . . . . Input Linear Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . . Ground Lead Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 13 13 14 Helpful Hints Low-inductance Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . SureFoot Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Probe Tip Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 18 19 Specifications P6209 4 GHz Probe Instruction Manual i Table of Contents Theory of Operation Probe Head and Cable Assembly . . . . . . . . . . . . . . . . . . . . . . . Compensation Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Offset Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Probe Identification EEPROM . . . . . . . . . . . . . . . . . . . . . . . . . TEKPROBE SMA Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . VCC, +7 V Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 30 30 30 31 31 Performance Verification Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Equipment Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Output Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DC Attenuation Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Rise Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 34 35 36 38 Adjustments Removing the Compensation Box Bottom Cover . . . . . . . . . . . Adjustment Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Offset Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Offset Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 44 44 46 Maintenance Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preparation for Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 49 49 Troubleshooting Replaceable Parts Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . . ii 55 56 P6209 4 GHz Probe Instruction Manual Table of Contents List of Figures Figure 1: Dynamic and offset limitations . . . . . . . . . . . . . . . . . Figure 2: Probe functional check connections . . . . . . . . . . . . . Figure 3: Ground lead equivalent circuit . . . . . . . . . . . . . . . . . Figure 4: Low-inductance grounding . . . . . . . . . . . . . . . . . . . . Figure 5: Using a SureFoot adapter for grounding . . . . . . . . . . Figure 6: Using a probe tip as a test point . . . . . . . . . . . . . . . . . Figure 7: Typical input impedance and phase versus frequency Figure 8: Typical bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 9: Dimensions of probe head and compensation box . . Figure 10: P6209 simplified schematic diagram . . . . . . . . . . . . Figure 11: TEKPROBE SMA Interface pin assignments . . . . . Figure 12: P6209 output zero setup . . . . . . . . . . . . . . . . . . . . . . Figure 13: P6209 DC attenuation accuracy setup . . . . . . . . . . . Figure 14: Test system rise time setup . . . . . . . . . . . . . . . . . . . Figure 15: Test system and probe rise time setup . . . . . . . . . . . Figure 16: Removing the compensation box bottom cover . . . Figure 17: P6209 adjustment and measurement locations . . . . Figure 18: P6209 offset zero setup . . . . . . . . . . . . . . . . . . . . . . Figure 19: P6209 offset range setup . . . . . . . . . . . . . . . . . . . . . Figure 20: Compensation box test point locations . . . . . . . . . . Figure 21: TEKPROBE SMA Interface pin assignments . . . . . Figure 22: P6209 replaceable parts . . . . . . . . . . . . . . . . . . . . . . Figure 23: P6209 standard accessories . . . . . . . . . . . . . . . . . . . Figure 24: P6209 optional accessories . . . . . . . . . . . . . . . . . . . P6209 4 GHz Probe Instruction Manual 10 11 14 17 18 19 24 24 26 29 31 35 37 38 39 43 44 45 47 52 52 57 58 60 iii Table of Contents iv P6209 4 GHz Probe Instruction Manual General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified. Only qualified personnel should perform service procedures. To Avoid Fire or Personal Injury Connect and Disconnect Properly. Connect the probe outputs to the measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe ground from the circuit under test before disconnecting the probe from the measurement instrument. Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Connect the ground lead of the probe to earth ground only. Do Not Operate Without Covers. Do not operate this product with covers or panels removed. Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel. Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry. P6209 4 GHz Probe Instruction Manual v General Safety Summary Safety Terms and Symbols Terms in This Manual. These terms may appear in this manual: WARNING. Warning statements identify conditions or practices that could result in injury or loss of life. CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property. Terms on the Product. These terms may appear on the product: DANGER indicates an injury hazard immediately accessible as you read the marking. WARNING indicates an injury hazard not immediately accessible as you read the marking. CAUTION indicates a hazard to property including the product. Symbols on the Product. These symbols may appear on the product: CAUTION Refer to Manual vi P6209 4 GHz Probe Instruction Manual Service Safety Summary Only qualified personnel should perform service procedures. Read this Service Safety Summary and the General Safety Summary before performing any service procedures. Do Not Service Alone. Do not perform internal service or adjustments of this product unless another person capable of rendering first aid and resuscitation is present. P6209 4 GHz Probe Instruction Manual vii Service Safety Summary Contacting Tektronix Phone 1-800-833-9200* Address Tektronix, Inc. Department or name (if known) 14200 SW Karl Braun Drive P.O. Box 500 Beaverton, OR 97077 USA Web site www.tektronix.com Sales support 1-800-833-9200, select option 1* Service support 1-800-833-9200, select option 2* Technical support Email: techsupport@tektronix.com 1-800-833-9200, select option 3* 6:00 a.m. - 5:00 p.m. Pacific time * viii This phone number is toll free in North America. After office hours, please leave a voice mail message. Outside North America, contact a Tektronix sales office or distributor; see the Tektronix web site for a list of offices. P6209 4 GHz Probe Instruction Manual Product Description The Tektronix P6209 is a 4 GHz (probe only), 5X active FET probe with 1 pF input capacitance. The probe is designed primarily for use with sampling oscilloscopes, such as the CSA/TDS8000. The P6209 has a low input capacitance and high input resistance, which minimize circuit loading over a wide bandwidth range. The small profile and low-mass head of the P6209 make manual probing of crowded circuits fast and easy. The accessory tips and adapters enable the P6209 to be used on a wide variety of circuit architectures. The P6209 is powered through a TEKPROBE SMA interface between the probe compensation box and the oscilloscope. Options The following options are available when ordering the P6209 probe: H Option D1 - Calibration Data H Option C3 - 3 years Calibration Service H Option D3 - 3 years Calibration Data H Option R3 - 3 years Extended Warranty P6209 4 GHz Probe Instruction Manual 1 Product Description Features and Standard Accessories The P6209 probe is provided with several features and accessories designed to make probing and measurement a simpler task. Please familiarize yourself with these items and their uses. For part number information and quantities of standard and optional accessories, refer to Replaceable Parts on page 55. Probe head assembly. The probe head is designed for ease of use and high performance. Its small size makes it easy to handle in tight areas. Probe tip socket Ground socket Stabilization notch The probe tip socket is sized to easily press onto 0.025 inch pins for direct access. The ground socket provides a short ground path for high fidelity ground connections. The stabilization notch permits you to use adjacent pins to reduce stresses on the probe and pins. TEKPROBE SMA interface. The TEKPROBE SMA interface provides a communication path between the probe and the oscilloscope. Contact pins provide power, offset, and probe characteristic data transfer. See page 31 for more information. Color marker bands 2 Color marker bands. Attach matching pairs of the color marker bands onto the cable at the head and compensation box of each probe. The marker bands enable quick verification of which probe is connected to which instrument channel. P6209 4 GHz Probe Instruction Manual Product Description SureToe probe tip. The SureToe tip is a pointed probe tip useful for probing in dense circuitry. Attach the SureToe adapters the same way as the push-in probe tips. Do not force the tip. Also, be careful not to injure yourself with the sharp probe tip. They can be used with any of the socketed accessory leads. Push-in probe tip. Use the push-in probe tip for general purpose probing by hand. The tip may also be used as a temporary test point. See page 19 for more information. Push-in probe tip The push-in probe tip may also be used with the other socketed leads and adapters. Installing the push-in probe tip. Attach the push-in probe tip by seating the tip into the probe tip socket and pushing the tip in until it is seated. Either end of the tip may be used. Do not force the tip. Also, be careful not to injure yourself with the sharp probe-tip. To remove the tip, gently grab the tip with small pliers and pull the tip out. Right-angle adapter. Use the right-angle adapter for low-profile probing of 0.025 inch square pins. The right-angle adapter allows the probe to lie flat against a circuit board. This enables probing in vertical circuits, such as computer or communications backplanes, or in tight areas, such as between circuit cards. Right angle adapter The right-angle adapter can be used directly with the probe head or attached to the Y-lead adapter or ground leads. The right-angle adapter is attached the same way as the push-in probe tip and can be easily removed by hand. P6209 4 GHz Probe Instruction Manual 3 Product Description Y-lead adapter Y-lead adapter. Use the Y-lead adapter to extend the physical reach of the probe and ground when necessary. The Y-lead adapter accepts any of the probe tips or adapters and can be pushed directly onto 0.025 inch pins. When selecting the grounding connection, maintain as short a ground path as possible. Refer to page 14 for more information. Y-lead adapter To attach the Y-lead adapter, gently press the lead pins into the probe head tip and ground receptacles. Using the black lead for ground is recommended. Three-inch ground lead. Use the three-inch ground lead for general probing. The socketed end of the lead may be connected to any of the probe tips and adapters or fitted onto 0.025 inch pins. To attach the ground lead, press and rotate the lead pin connector into the ground socket on the probe head. The lead may be removed by simply pulling the pin out by hand. When selecting the grounding connection, maintain as short a ground path as possible. Refer to page 14 for more information. 4 P6209 4 GHz Probe Instruction Manual Product Description Customizable ground lead. This ground lead wire can be bent or cut shorter. NOTE: To ease insertion into the ground socket of the probe, cut the tip of this ground lead wire at a 30 to 60 degree angle. Customizable ground lead To maintain signal fidelity while probing, use as short a ground path as possible. Refer to page 14 for more grounding information. Low-inductance ground pogo pin. Use the low-inductance ground pogo pin to substantially reduce ground lead inductance. Because the pogo pin simply touches the ground reference, you can easily move the probe to different points on the circuit under test. Pogo pin To attach, press the pogo pin into the probe head ground socket. When selecting the grounding connection, maintain as short a ground path as possible. Refer to page 14 for more information. Signal-ground adapter. The signal-ground adapter is ideal for use with signal/ground pairs on 0.100 inch header pins. Attach the signal-ground adapter by gently pressing it into the ground socket on the probe head. Signal-ground adapter Be sure to use the stabilization notch whenever possible to avoid slipping and damaging the probe or circuitry under test. P6209 4 GHz Probe Instruction Manual 5 Product Description SMT KlipChip. Use the SMT KlipChip test clips to access fragile, dense circuitry. KlipChip test clips can be connected to the Y-lead or three-inch ground leads. Simply press the lead socket into the KlipChip handle. KlipChip Y-lead adapter The KlipChip body freely turns, allowing better probe orientation. To reduce stress and provide a lower profile on components being tested, the flexible sleeve of the KlipChip bends up to a 35 degree angle. When selecting the grounding connection, maintain as short a ground path as possible. Refer to page 14 for more information. SMA adapter jack. Use the SMA adapter jack with the SMT KlipChips to gain access to calibration signals from SMA connectors on the oscilloscope front panel. Male SMA termination. Protect the probe input circuitry by connecting the termination to the probe input connector when the probe is not in use. 6 P6209 4 GHz Probe Instruction Manual Product Description Optional Accessories The following accessories are available for the probe. See the Replaceable Parts List on page 55 for ordering information. SureFoot probe tips. The SureFoot tip is an integral probe tip and miniature guide that enables fault-free probing of fine-pitch SMD packages. Attach SureFoot adapters the same way as the push-in probe tips. They can be used with any of the socketed accessory leads. The orange, 0.050 inch SureFoot tip is compatible with 50 mil JEDEC packages such as SOIC, PLCC, and CLCC. The blue, 0.025 inch SureFoot tip is compatible with 0.65 mm JEDEC and EIAJ packages. The red, 0.5 mm SureFoot tip is compatible with EIAJ packages. SMK4 Micro KlipChip adapters. Use the Micro KlipChip adapters to probe the leads on integrated circuits that are surface-mounted. SMA-to-probe tip adapter. Use the adapter to connect the probe to SMA cables. The adapter includes a 50-ohm SMA termination. P6209 4 GHz Probe Instruction Manual 7 Product Description 8 P6209 4 GHz Probe Instruction Manual Configuration The P6209 provides the oscilloscope with the probe model number, serial number, and attenuation factor. When connected to the oscilloscope, the display readouts are corrected for the probe attenuation factor. The probe offset control is controlled by the oscilloscope. CAUTION. To prevent damage to your probe, do not apply a voltage outside the Maximum Input Voltage rating. See Specifications on page 21. To prevent electrostatic damage to the instrument and sampling modules, follow the precautions described in the manuals accompanying your instrument. Always use a wrist strap (provided with your probe) when handling the probe and sampling modules, and when making signal connections. Probe Offset The purpose of the offset capability is to permit the input signal to be positioned in the center of the dynamic range of the probe amplifier, to obtain the best transient response. Using the offset to cancel DC signal components enables optimal probe performance. See Figure 1 on page 10 for more information. NOTE. If the offset is not correctly positioned, clipping or distortion of the displayed signal may occur. See your oscilloscope manual for specific instructions on its operation and offset control. P6209 4 GHz Probe Instruction Manual 9 Configuration To set the probe offset, follow these steps: 1. Set the oscilloscope vertical scale to the least sensitive V/div setting (500 mV/div on the CSA/TDS8000). 2. Attach the probe to the circuit. 3. Adjust the probe offset to bring the trace onto the oscilloscope screen. 4. Change the volts/division setting to the desired range, adjusting the offset to keep the trace on screen. NOTE. The P6209 has a ±5.0 V offset range. The linear operating range is ±2.00 V. See Figure 1. Also, see page 13 for more information. Nonoperating range (+30 V maximum nondestructive input voltage ) +7.00 V +5.00 V +5.00 V +3.00 V +2.00 V 0V Maximum AC Signal Amplitude Maximum Offset Range - 2.00 V - 3.00 V - 5.00 V - 5.00 V - 7.00 V Nonoperating range (-- 30 V maximum nondestructive input voltage ) Figure 1: Dynamic and offset limitations 10 P6209 4 GHz Probe Instruction Manual Functional Check After installing the probe on the CSA/TDS8000 oscilloscope, a functional check may be performed using the INTERNAL CLOCK OUTPUT connection on the front panel of the oscilloscope. See Figure 2. SMA adapter jack Figure 2: Probe functional check connections 1. Set the oscilloscope to display the probe channel. 2. Using the Y-lead connector, SMT KlipChips and the SMA adapter jack accessories, connect the probe to the INTERNAL CLOCK OUTPUT connection on the oscilloscope, as shown in Figure 2. 3. In the setup menu, set the trigger source to internal clock and the frequency to 200 kHz. P6209 4 GHz Probe Instruction Manual 11 Functional Check 4. Set the offset to 1V, the vertical scale to 500 mV/div, and the horizontal scale to 2 s/div. 5. The observed waveform should be a 200 kHz square wave with an amplitude of approximately 2 V (the clock output is unterminated), and have an offset of approximately 1 V. If the probe does not pass this functional check, go to Troubleshooting on page 51. 12 P6209 4 GHz Probe Instruction Manual Operating Basics Please follow these operating guidelines to get optimum performance from your P6209 probe. Handling the Probe Exercise care when using and storing the P6209 probe. The probe and cable are susceptible to damage caused by careless use. Always handle the probe using the compensation box and probe head, avoiding undue physical strain to the probe cable, such as kinking, excessive bending, or pulling. Visible dents in the cable will increase signal aberrations. Do not drop the probe or subject it to physical shock. Damage to the probe may result. CAUTION. To prevent damage to your probe, do not apply a voltage outside the Maximum Nondestructive Input Voltage for the probe. Always use a wrist strap (provided with your probe) when handling the probe and sampling modules, and when making signal connections. Maximum Nondestructive Input Voltage Applying voltages above the P6209 design limits may damage the probe tip amplifier. Please refer to Specifications on page 21 for the maximum operating voltage and frequency derating information. Input Linear Dynamic Range The probe head amplifier used by the P6209 has a limited linear operating range. To keep the input linearity error less than 0.1%, you must limit the apparent signal input voltage to ±1.75 V. P6209 4 GHz Probe Instruction Manual 13 Operating Basics Use the DC offset adjustment to maintain the probe within its dynamic range. The nominal offset adjustment range of the P6209 is ±5.0 VDC. For example: to offset a +2 VDC level in a circuit, set the offset to +2 V. NOTE. The probe can tolerate input voltages of ±30 V (≤5 minutes) without damage; however, the linearity error specification does not apply to input voltages outside the linear dynamic range of the probe. See Figure 1 on page 10. Ground Lead Length When you are probing a circuit, you should always use as short a ground lead as possible between the probe head and circuit ground. The series inductance added by the probe tip and ground lead can result in a resonant circuit; this circuit may cause parasitic ringing within the bandwidth of your oscilloscope. When you touch your probe tip to a circuit element, you are introducing a new resistance, capacitance, and inductance into the circuit. Refer to Figure 3. R source Probe R in 20 KΩ V source Probe C in 1 pF L gl (Ground Lead) Figure 3: Ground lead equivalent circuit Ringing and rise time degradation can be masked if the frequency content of the signal degradation is beyond the bandwidth of the oscilloscope. 14 P6209 4 GHz Probe Instruction Manual Operating Basics You can determine if ground lead effects may be a problem in your application if you know the self-inductance (L) and capacitance (C) of your probe and ground lead. Calculate the approximate resonant frequency (f0) at which this parasitic circuit will resonate with the following formula: f0 = 1 2π LC The preceding equation shows that reducing the ground lead inductance will raise the resonant frequency. If your measurements are affected by ringing, your goal is to lower the inductance of your ground path until the resulting resonant frequency is well above the frequency of your measurements. The low-inductance ground contacts described in Accessories can help you reduce the effects of ground lead inductance on your measurements. P6209 4 GHz Probe Instruction Manual 15 Operating Basics 16 P6209 4 GHz Probe Instruction Manual Helpful Hints Follow these helpful hints to make probing easier and noise free. Low-inductance Grounding Placing a ground plane on top of a package being probed can minimize ground lead length and inductance. See Figure 4. Figure 4: Low-inductance grounding Attach a small piece of copper clad on top of the package and connect it to the package ground connection. Use the low-inductance ground lead provided with the P6209 to keep the ground lead length as short as possible. P6209 4 GHz Probe Instruction Manual 17 Helpful Hints This method is very useful when making many measurements on the same package. Using a ground plane on the package makes probing the package easier and avoids adding unnecessary ground lead length and distortion. SureFoot Grounding If you cannot use the recommended low-inductance grounding method, you may ground the probe to the package under test using a SureFoot adapter. Refer to Figure 5. Figure 5: Using a SureFoot adapter for grounding Use a SureFoot adapter at the end of a short ground lead to connect directly to the package ground. This method is preferred over using an adjacent circuit ground because it is the shortest ground path possible. 18 P6209 4 GHz Probe Instruction Manual Helpful Hints Probe Tip Test Points You can solder the push-in probe tip or a 0.025 square pin to a circuit to be used as a temporary test point. See Figure 6. Solder the tip onto a lead or pin with a low-power soldering iron. Press the probe head onto the tip to make a measurement, and then pull the probe head off when you are done. The probe tip may be removed and reused by desoldering it from the circuit and soldering it into another circuit in the future. Solder Figure 6: Using a probe tip as a test point NOTE. Do not use pieces of solid-core copper wire as test points. If the wire breaks off in the probe tip socket, it may be impossible to remove the wire, and it will prevent insertion of other accessory tips. P6209 4 GHz Probe Instruction Manual 19 Helpful Hints 20 P6209 4 GHz Probe Instruction Manual Specifications These specifications apply to a P6209 probe when used with a CSA/TDS8000 oscilloscope, and other oscilloscopes with a TEKPROBE SMA interface. The probe and oscilloscope must first be allowed to warm up for 20 minutes before measurements are taken. CAUTION. Do not apply voltages beyond the nondestructive input voltage range to the probe. Damage to the probe or circuit under test may result. Table 1: Warranted electrical specifications DC Gain Accuracy (probe only) 0.2 ±2% (excludes offset error) Rise Time (small signal, probe only) ≤120 ps (≤ 250 mV, 20--30° C) Output Zero ±7.5 mV or less at output of probe (does not include oscilloscope offset control voltage error) (20--30° C) ±37.5 mV or less displayed on screen with TEKPROBE SMA interface (does not include oscilloscope offset control voltage error) (20--30° C) Delay Time 5.95 ns ±0.2 ns (includes SMA output cable) System Noise 300 V rms or less at output of probe with probe tip grounded P6209 4 GHz Probe Instruction Manual 21 Specifications Table 2: Typical electrical characteristics Bandwidth, (probe only) 4 GHz (See Figure 8) Linear Input Dynamic Range --2.00 V to +2.00 V. (Equivalent to --0.40 V to +0.40 V at the output of the probe.) Linearity ᐔ0.1% over a dynamic range of --1.75 V to +1.75 V (±3.5 mV) Maximum deviation from linear regression expressed as a percentage of the specified full-scale dynamic range ᐔ1.0% over a dynamic range of --2.00 V to +2.00 V (±40 mV) Nondestructive Input Voltage Range --30 V to +30 V (DC + peak AC) (5 minutes maximum continuous exposure) Input Resistance 20 kΩ at DC. (See Figure 7) Input Capacitance 0.85 pF @100 MHz Offset Range --5.0 V to +5.0 V DC Offset Drift 150 V/°C or less at output of probe 0.75 mV/°C or less displayed on screen with TEKPROBE SMA interface DC Offset Scale Accuracy (gain of offset signal path) ᐔ2.0% (of 5X actual probe gain) DC Voltage Measurement Accuracy (referred to input) ᐔ[2% of input + (2% of offset) + 37.5* mV + (3.5 mV or 40 mV†)] * Worst case value. † Worst case values - see linearity specification. 22 P6209 4 GHz Probe Instruction Manual Specifications Calculating DC Voltage Measurement Accuracy: DC Voltage Measurement Accuracy (referred to input) = [2% of input + (2% of offset) + 37.5 mV + 3.5 mV] where: gain error = 2% of input voltage (relative to offset voltage) offset gain error = 2% of effective offset at probe tip output zero = 37.5 mV effective at probe tip (worst case, but does not include any oscilloscope offset control voltage error) linearity error1 = 0.1% of 1.75 V (3.5 V full scale)=3.5 mV High Error Example. To determine the maximum specified DC error when measuring a PECL logic high of approximately 4.1 V with probe offset set to 3.7 V: (2% of input voltage = 8 mV, + 2% of offset voltage = 74 mV, + output zero = 37.5 mV (worst case value), + linearity error = 3.5 mV) = 123 mV Low Error Example. To determine the maximum specified DC error when measuring a LVDS low of approximately 1.0 V without any probe offset (0 V): (2% of input voltage = 20 mV, + 2% of offset voltage = 0 mV, + output zero = 5 mV (effective offset at probe tip due to typical offset value of 1 mV at probe output), + linearity error = 3.5 mV) = 28.5 mV 1 The linearity specification is rated with a segmented range (see page 22). The linearity error is higher at the extreme ends of the dynamic range. P6209 4 GHz Probe Instruction Manual 23 Specifications 90° 20 K 72° 16 K 36° 14 K 18° 12 K 0° 10 K - 18° 8K Phase - 36° 6K - 54° 4K - 72° 2K - 90° 1M 10 M 100 M Frequency 1G 5G Magnuitude (ohms) 54° Phase (degrees) 18 K Magnitude 0 Figure 7: Typical input impedance and phase versus frequency - 10 - 12 - 14 Gain dB - 16 - 18 - 20 - 22 Gain = 20 Log - 24 Ꮛ Ꮠ V OUT V IN - 26 - 28 1 MHz 10 MHz 100 MHz 1 GHz 10 GHz Frequency Figure 8: Typical bandwidth 24 P6209 4 GHz Probe Instruction Manual Specifications Table 3: Environmental characteristics Operating Temperature 0° C to +40° C The environmental exposure procedure used as stated in Tektronix Design Standard 062-2847-00 for Class 5 equipment. Nonoperating Temperature --40° C to +71° C The environmental exposure procedure used as stated in Tektronix Design Standard 062-2847-00 for Class 5 equipment. Humidity 80% RH at 40° C The environmental exposure procedure used as stated in Tektronix Design Standard, 062-2847-00 for Class 5 equipment. Altitude Operating: 15,000 ft. NonOperating: 50,000 ft. P6209 4 GHz Probe Instruction Manual 25 Specifications Table 4: Physical Characteristics Net Shipping Weight 3.1 lb (1.4 kg) Cable Length (probe tip to SMA) (1.37 meters) Figure 9: Dimensions of probe head and compensation box 26 P6209 4 GHz Probe Instruction Manual WARNING The following servicing instructions are for use only by qualified personnel. To avoid injury, do not perform any servicing other than that stated in the operating instructions unless you are qualified to do so. Refer to all safety summaries before performing any service. Theory of Operation There are no user replaceable parts within the probe or the compensation box; however, this theory of operation is provided to assist you in isolating failures to either the probe or the host oscilloscope. Refer to Figure 10 for the simplified schematic. Probe head Compensation box Oscilloscope Probe tip Signal Out Probe tip amplifier Probe ID Out Clock In EEPROM Offset ampifier Offset to probe +5 ±1 V Offset Offset zero -5 Offset gain +7 V Linear regulator +15 V -5 V Ground Probe cable TEKPROBE SMA interface Figure 10: P6209 simplified schematic diagram P6209 4 GHz Probe Instruction Manual 29 Theory of Operation Probe Head and Cable Assembly The probe head assembly contains an active amplifier circuit that buffers and amplifies the input signal. The amplifier receives power and an offset level from the compensation box assembly via the cable assembly. All signal amplification and buffering is performed in the probe head assembly. No further amplification takes place in the compensation box. Compensation Box The compensation box contains the following circuits: H Offset amplifier H Probe identification EEPROM H TEKPROBE SMA interface H VCC, +7 V linear regulator Offset Amplifier The offset amplifier is used to offset the input signal DC component so that it stays at the optimal point of the probe linear dynamic range. For more information on the linear dynamic range characteristic, refer to Input Linear Dynamic Range on page 13. The offset amplifier receives offset information as a ±1 VDC voltage from the oscilloscope. The amplfier then amplifies it to match the probe characteristics and applies it to the probe head circuit. The offset amplifier has two adjustments: offset zero and offset scale. These adjustments rarely need attention; however, detailed adjustment instructions are in the Adjustment section on page 43. Probe Identification EEPROM The probe identification EEPROM is used to configure the oscilloscope to the probe. The EEPROM receives a clock input from 30 P6209 4 GHz Probe Instruction Manual Theory of Operation the oscilloscope, and information about the probe is passed to the oscilloscope. TEKPROBE SMA Interface The TEKPROBE SMA interface provides a communication path between the probe and the oscilloscope. Contact pins provide power, signal, offset, and data transfer for the probe identification EEPROM. Figure 11 shows the probe interface pin functions. Refer to your oscilloscope service documentation for more detailed specifications. GND 5 6 7 +15 V +5 V 4 12 10 8 9 3 11 2 - 15 V -5 V 1 CLK DATA Figure 11: TEKPROBE SMA Interface pin assignments VCC, +7 V Regulator The +15 V input is regulated to +7 V to supply the IC, VCC. P6209 4 GHz Probe Instruction Manual 31 Theory of Operation 32 P6209 4 GHz Probe Instruction Manual Performance Verification Use the following procedures to verify the warranted specifications of the P6209 probe. Before beginning these procedures, refer to page 41 and photocopy the test record and use it to record the performance test results. The recommended calibration interval is one year. These procedures test the following specifications: H Output offset voltage H DC attenuation accuracy H Rise time Equipment Required Refer to Table 5 for a list of the equipment required to verify the performance of your probe. NOTE. These procedures use a CSA/TDS8000 as the host instrument. If you are using a TDS 820, the procedures will vary slightly. Table 5: Equipment required for performance verification and adjustment Item description Performance requirement Oscilloscope TEKPROBE SMA interface Calibration Step Generator Recommended example Tektronix CSA/TDS8000 with a 80E00 Series Sampling Module, or TDS 820 067-1338-XX DC Power Supply1 1 VDC at 1 mA Tektronix PS280 DMM 0.5% accuracy at 2 VDC Tektronix TX3 P6209 4 GHz Probe Instruction Manual 33 Performance Verification Table 5: Equipment required for performance verification and adjustment (Cont.) Item description Performance requirement Recommended example Feedthrough Termination BNC, 50 Ω ±0.05 Ω 011-0129-00 Coaxial cables (2) Male-to-Male SMA, 20 in 174-1427-00 Adapter High Frequency Probe tip adapter w/50 Ω Termination 015-0678-00 Adapter SMA Female-to-Female 015-1012-00 Adapter SMA Female-to-BNC Male 015-0572-00 Adapter SMA Male jack 131-7128-00 Adapter BNC Female-to-Dual Banana 103-0090-00 Adapter Y-lead adapter 196-3457-00 Adapters (2) KlipChip adapter 206-0364-XX 1 Provides DC stimulus if using a TDS 820 Equipment Setup Use this procedure to set up the equipment to test the probe. 1. Connect the probe to the oscilloscope. 2. Connect a Y-lead adapter with KlipChip adapters attached to the probe. 3. Turn on the oscilloscope and enable the channel. 4. Allow 20 minutes for the equipment to warm up. 34 P6209 4 GHz Probe Instruction Manual Performance Verification Output Zero Use this procedure to verify the probe output zero. 1. Connect the test equipment as shown in Figure 12. 2. Ground the probe tip by connecting the two KlipChips together. CSA/TDS8000 Digital multimeter BNC-to-dual banana adapter KlipChip adapter 50 Ω Precision termination P6209 Probe SMA F-to-BNC M adapter Y-lead adapter Figure 12: P6209 output zero setup 3. Set the oscilloscope trigger source to internal clock. 4. Set the oscilloscope channel vertical offset to 0.00 V. 5. Set the multimeter to read DC volts. 6. Record the multimeter reading on the test record. The measured DC level should be 0.00 V ±10 mV for the CSA/TDS8000. P6209 4 GHz Probe Instruction Manual 35 Performance Verification NOTE. The P6209 probe is guaranteed to have less than 7.5 mV offset. However, the CSA/TDS8000 and TDS 820 will have some residual voltage remaining when set to 0.00 V offset that will appear in this test as additional offset at the probe output. DC Attenuation Accuracy Use this procedure to verify the probe DC attenuation accuracy. 1. Set the oscilloscope trigger to internal clock. 2. Set the oscilloscope channel vertical offset to 0.00 V. 3. Set the multimeter to read DC volts. 4. Connect the probe tip and ground to the DC calibrator output of the CSA/TDS8000, using the Y-lead adapter, KlipChips and SMA adapter jack. If the probe is being used with a TDS 820 oscilloscope, a separate DC power supply is needed to provide the stimulus. In this case, connect the probe to the power supply using the Y-lead adapter and KlipChips. Record the results of steps 5 through 11 separately. Use the results to calculate the DC accuracy of the probe in step 12. 5. Set the DC calibration output of the CSA/TDS8000 (or external power supply if using a TDS 820), to +1.000 V. Use the multimeter to verify the DC voltage is as close to +1.000 V as possible. Record this measurement as Vmax. 6. Connect the multimeter as shown in Figure 13 on page 37. 36 P6209 4 GHz Probe Instruction Manual Performance Verification CSA/TDS8000 Digital multimeter SMA adapter jack BNC-to-dual banana adapter 50 Ω Precision termination DC CALIBRATION OUTPUT SMA F-to-BNC M adapter P6209 Probe Figure 13: P6209 DC attenuation accuracy setup 7. Measure and record the multimeter reading as M1. 8. Disconnect the multimeter. 9. Set the DC calibration output of the CSA/TDS8000 (or external power supply if using a TDS 820), to - 1.000 V. Use the multimeter to verify the DC voltage is as close to - 1.000 V as possible. Record this measurement as Vmin. 10. Reconnect the multimeter as shown in Figure 13. 11. Measure and record the multimeter reading as M2. 12. Determine the percent error by using the formula below: %Error = M1–M2 –1 × 100% (Vmax–Vmin) × .2 The calculated error should be ≤2%. 13. Record the results on the test record. P6209 4 GHz Probe Instruction Manual 37 Performance Verification Rise Time This procedure verifies that the probe meets rise time specifications. The probe rise time is calculated from rise times obtained by measuring the test system without the probe attached, and the test system plus the probe. 1. Connect the test equipment as shown in Figure 14. 2. Set the oscilloscope to the following settings: H Trigger to internal clock H Vertical scale to 50 mV/div H Horizontal scale to 100 ps/div H Horizontal record length to 4000 points H Acquisition Mode to average at least 16 samples 3. Adjust the oscilloscope horizontal and vertical position controls to display a signal similar to that shown in Figure 14. CSA/TDS8000 INTERNAL CLOCK OUTPUT SMA cable Generator remote head Calibration step generator Trigger input To power supply Figure 14: Test system rise time setup 38 P6209 4 GHz Probe Instruction Manual Performance Verification 4. Use the oscilloscope measurement capability to calculate and display rise time. Rise time is determined from the 10% and 90% amplitude points on the waveform. Record the rise time as ts. The system rise time (ts ) that you measured in step 4 represents the rise time of the test system without the probe. The following steps instruct you to assemble the test system that includes the probe, as shown in Figure 15. The system and probe rise time (ts+p ) that you measure in step 8 is used to calculate the probe rise time (tp ) in step 9. 5. Connect the test equipment as shown in Figure 15. CSA/TDS8000 INTERNAL CLOCK OUTPUT Calibration step generator SMA cable Trigger input To power supply P6209 Probe Generator remote head SMA F-to-F HF probe tip adapter 50 Ω Termination Figure 15: Test system and probe rise time setup 6. Set the vertical to 50 mV/div and horizontal to 500 ps/div. 7. Adjust the oscilloscope horizontal and vertical position controls to display a signal similar to that shown in Figure 15. P6209 4 GHz Probe Instruction Manual 39 Performance Verification NOTE. Do not touch the probe tip adapter when making calibration measurements. Measurement accuracy is degraded when the probe tip adapter is handled. 8. Use the oscilloscope measurement capability to calculate and display rise time. Rise time is determined from the 10% and 90% amplitude points on the waveform. Record the rise time as ts+p. 9. Calculate the probe only rise time using the following formula: t p = Ꭹt(s+p) − t s 2 2 10. Check that the calculated rise time meets the probe specification. The rise time (tp ) must be ≤120 ps. 11. Record the results on the test record. 40 P6209 4 GHz Probe Instruction Manual Performance Verification Test record Probe Model: Serial Number: Certificate Number: Temperature: RH %: Date of Calibration: Technician: Performance test Minimum Incoming Outgoing Maximum Output zero voltage (at probe output) ± 10 mV (20_ C to 30_ C)1 -- 10 mV ________ ________ + 10 mV DC attenuation accuracy -- 2 % ________ ________ +2% Rise time N/A ________ ________ ≤120 ps 1 CSA/TDS8000 only P6209 4 GHz Probe Instruction Manual 41 Performance Verification 42 P6209 4 GHz Probe Instruction Manual Adjustments The P6209 has two internal controls: offset zero and offset scale. These controls rarely need to be adjusted and only should be changed after a probe performance verification and functional check has been performed on the oscilloscope. To make adjustments to the probe, the compensation box bottom cover must be removed and the equipment allowed to warm up for 20 minutes. Use the optional adjustment tool for making any adjustments. Refer to Table 5 on page 33 for a list of the equipment required to adjust your probe. Removing the Compensation Box Bottom Cover Follow these steps to gain access to the adjustment controls. 1. Using a Pozidrive screwdriver, remove the four screws that attach the bottom cover to the compensation box. Gently remove the cover. Refer to Figure 16. Figure 16: Removing the compensation box bottom cover P6209 4 GHz Probe Instruction Manual 43 Adjustments Adjustment Locations Refer to Figure 17 for the location of the offset adjustments and measurement test points. Offset scale adjust Offset zero adjust J1 Gnd lug R3 Offset test point (white wire) R2 J2 Figure 17: P6209 adjustment and measurement locations Offset Zero Use this procedure to adjust the probe offset zero. 1. Connect the probe power cable to the oscilloscope. 2. Ground the probe tip by connecting the two KlipChip leads together. 3. Set the oscilloscope trigger source to internal clock. 4. Set the channel vertical offset to 0.00 V. 5. Using the (-- ) lead of the DMM, probe the ground lug in the probe comp box. Refer to Figure 17 on page 44. 44 P6209 4 GHz Probe Instruction Manual Adjustments 6. Using the (+) lead of the DMM, carefully probe pin 7 of J2 (white wire) in the probe compensation box. Refer to Figure 17 on page 44. NOTE. The offset voltage measured on pin 7 will be of opposite polarity and approximately one fifth of the offset value displayed on the oscilloscope. 7. Adjust the oscilloscope offset so that the voltage measures 0.000 V on the DMM. Set the offset as close to 0 V as possible. 8. Connect the equipment as shown in Figure 18. 9. Adjust Offset Zero (R2) in the probe compensation box for 0.00 V, ±1 mV on the DMM. Refer to Figure 17 on page 44 for the adjustment location. CSA/TDS8000 Digital multimeter BNC-to-dual banana adapter KlipChip adapter 50 Ω Precision termination P6209 Probe SMA F-to-BNC M adapter Y-lead adapter Figure 18: P6209 offset zero setup P6209 4 GHz Probe Instruction Manual 45 Adjustments Offset Scale Use this procedure to adjust the offset scale of the probe. The offset zero of the probe should be checked before making any adjustment to the offset scale. Refer to Figure 17 on page 44 for the location of the test points and adjustments. 1. Connect the probe power cable to the oscilloscope. 2. Set the oscilloscope trigger source to internal clock. 3. Set the oscilloscope channel vertical offset to +2.5 V. 4. Ground the probe tip by connecting the two KlipChip leads together. 5. Using the (-- ) lead of the DMM, probe the ground lug in the probe compensation box. 6. Using the (+) lead of the DMM, carefully probe pin 7 of J2 (white wire) in the probe compensation box. NOTE. The offset voltage measured on pin 7 will be of opposite polarity and approximately one fifth of the offset value displayed on the oscilloscope. 7. Adjust the oscilloscope offset so that pin 7 of J2 measures - 0.50 V on the DMM. 8. Set the DC power supply to +2.5 V, as measured on the DMM. 9. Connect the equipment as shown in Figure 19. Take care not to disturb the offset level setting. 10. Adjust the Offset Scale (R3) in the probe compensation box so the output measures 0.000 V, ±1 mV on the DMM. 46 P6209 4 GHz Probe Instruction Manual Adjustments CSA/TDS8000 Digital multimeter Power supply 50 Ω Precision termination BNC-to-Dual Banana Adapter -- + SMA F-to-BNC M adapter P6209 Probe Figure 19: P6209 offset range setup NOTE. After the offset adjustments have been made, use the Performance Verification procedures beginning on page 33 to verify your probe meets specifications. P6209 4 GHz Probe Instruction Manual 47 Adjustments 48 P6209 4 GHz Probe Instruction Manual Maintenance This section describes the maintenance procedures for the P6209. Inspection and Cleaning To prevent damage to probe materials, avoid using chemicals that contain benzine, benzene, toluene, xylene, acetone, or similar solvents. Do not immerse the probe or use abrasive cleaners. Dirt may be removed with a soft cloth dampened with a mild detergent and water solution or isopropyl alcohol. Replacement Parts Refer to Replaceable Parts on page 55 for a list of customer replaceable parts. Due to the sophisticated design of the P6209 probe, there are no user replaceable parts within the probe. Preparation for Shipment If the original packaging is unfit for use or not available, use the following packaging guidelines: 1. Use a corrugated cardboard shipping carton having inside dimensions at least one inch greater than the probe dimensions. The box should have a carton test strength of at least 200 pounds. 2. Put the probe into a plastic bag or wrap to protect it from dampness. 3. Place the probe into the box and stabilize it with light packing material. 4. Seal the carton with shipping tape. P6209 4 GHz Probe Instruction Manual 49 Maintenance 50 P6209 4 GHz Probe Instruction Manual Troubleshooting This troubleshooting documentation details isolation to either the probe or the oscilloscope. To troubleshoot the P6209, the probe must be attached to an operating TEKPROBE SMA interface oscilloscope with the bottom cover of the compensation box removed. Refer to page 43 for instructions on opening the compensation box. WARNING. Operating voltages will be present. Observe caution when probing in the compensation box. Be careful not to touch or ground energized components. Work at a Level 1 ESD workstation when performing troubleshooting. The compensation box contains static sensitive devices. P6209 4 GHz Probe Instruction Manual 51 Troubleshooting J1 Refer to Table 6 on page 53 for troubleshooting procedures and Figures 20 and 21 for troubleshooting test points. Ground J2 Offset (white) - 5 V (green) NC DATA (yellow) CLOCK (orange) +5 V (red) +15 V (brown) Figure 20: Compensation box test point locations GND 5 6 7 +15 V +5 V 4 12 10 8 9 CLK 3 11 2 - 15 V -5 V 1 DATA Figure 21: TEKPROBE SMA Interface pin assignments 52 P6209 4 GHz Probe Instruction Manual Troubleshooting Table 6: Troubleshooting guide Symptom Checks Result Action No signal out Check known signal No signal. Continue checks Go to No Probe Power Probe power OK. Replace probe. Distorted or nonlinear signal Is the signal within No linear dynamic range? See page 13. Set channel vertical offset to match signal characteristic. No Offset Adjust (Signal present) Go to Probe Power row Continue checks No Probe Power Probe power OK Check variation of Offset input OK offset input from host instrument. See Figure 20. Offset control voltage at pin 7 of J2 should vary between +1 V and --1 V. Replace probe. Check the power test points in the compensation box. See Figure 20. Power OK Continue checks. Power bad Check for continuity between the TEKPROBE SMA connector pins and the test points in the compensation box. See Figures 20 and 21 on page 52. Try an adjacent channel. Repair host instrument P6209 4 GHz Probe Instruction Manual 53 Troubleshooting 54 P6209 4 GHz Probe Instruction Manual Replaceable Parts This section contains a list of replaceable parts for the P6209 probe. Use this list to identify and order replacement parts. Parts Ordering Information Replacement parts are available from or through your local Tektronix, Inc. service center or representative. Changes to Tektronix instruments are sometimes made to accommodate improved components as they become available and to give you the benefit of the latest circuit improvements. Therefore, when ordering parts, it is important to include the following information in your order: H Part number H Instrument type or model number H Instrument serial number H Instrument modification number, if applicable If a part you order has been replaced with a different or improved part, your local Tektronix service center or representative will contact you concerning any change in the part number. P6209 4 GHz Probe Instruction Manual 55 Replaceable Parts Using the Replaceable Parts List The tabular information in the Replaceable Parts List is arranged for quick retrieval. Understanding the structure and features of the list will help you find the information you need for ordering replacement parts. Item Names In the Replaceable Parts List, an Item Name is separated from the description by a colon (:). Because of space limitations, an Item Name may sometimes appear as incomplete. For further Item Name identification, U.S. Federal Cataloging Handbook H6-1 can be used where possible. Indentation System This parts list is indented to show the relationship between items. The following example is of the indentation system used in the Description column: 1 2 3 4 5 Name & Description Assembly and/or Component Attaching parts for Assembly and/or Component (END ATTACHING PARTS) Detail Part of Assembly and/or Component Attaching parts for Detail Part (END ATTACHING PARTS) Parts of Detail Part Attaching parts for Parts of Detail Part (END ATTACHING PARTS) Attaching parts always appear at the same indentation as the item it mounts, while the detail parts are indented to the right. Indented items are part of, and included with, the next higher indentation. Attaching parts must be purchased separately, unless otherwise specified. Abbreviations Abbreviations conform to American National Standards Institute (ANSI) standard Y1.1 56 P6209 4 GHz Probe Instruction Manual P6209 4 GHz 5X Active Probe 2 010-- 0638-- 00 015-- 1021-- 00 174-- 1120-- 00 016-- 1879-- 01 -2 -3 -4 Tektronix part no. Serial no. Effective Dscont Figure 22: P6209 replaceable parts 22-- 1 Fig. & index no. 1 1 1 1 1 Qty 3 Mfr. code CASE,STORAGE:PLASTIC,W/CND FOAM CA ASSY,RF:COAXIAL,RFD,50 OHM,8.5 L,SMA,MALE TERMN,COAXIAL:SHORT CIRCUIT,SMA,FEMALE TK6108 060D9 26805 PROBE ASSEMBLY:SERVICE REPLACEMENT, SERIALIZED 80009 12345 name & description 4 016-- 1879-- 01 174-- 1120-- 00 2020-- 5353-- 00 010-- 0638-- 00 Mfr. part no. Replaceable Parts 57 58 8 9 2 Figure 23: P6209 standard accessories 1 3 10 4 11 5 12 6 7 Replaceable Parts P6209 4 GHz 5X Active Probe 1 1 016-1772-10 016-1773-10 016-1774-00 ---------- 016-1315-00 196-3482-XX 206-0364-XX 015-- 1020-- 00 131-- 7128-- 00 006-3415-04 071-0634-XX -3 -4 -5 -6 -7 -8 P6209 4 GHz 5X Active Probe -9 - 10 - 11 - 12 1 1 2 1 1 1 1 1 1 1 131-5638-11 1 Qty -2 Serial no. Effective Dscont 196-3456-00 Tektronix part no. 23-- 1 Fig. & index no. MANUAL,TECH:INSTRUCTION,P6209 STRAP,WRIST:3M TYPE 2214, ADJUSTABLE,6 FT COILED CORD CONN,RF JACK:COAX,SMA,50 OHM,MALE,GOLD PL TERMN,COAXIAL:SHORT CIRCUIT,SMA,MALE TIP,PROBE:MICROCKT TEST,SMT KLIPCHIP CUSTOMIZABLE GROUND LEAD, SET OF 5 MARKER KIT,ID:CABLE MARKER BAND,2 EA, VAR COLRS PROBE,TIP,SURETOE:PACKAGE OF 4 (REORDER ST501, PACKAGE OF 12) ACCESSORY KIT:ADAPTER SPRING,PKG OF 10 ACCESSORY KIT:GND PIN LENGTH ADAPTER,PKG OF 10 ACCESSORY KIT:POGO PIN,PKG OF 10 PROBE,TIP:PACKAGE OF 10 LEAD SET:ACCESSORY KIT, INCL THREE 3.0 INCH GROUND LEAD, TWO DIFFERENTIAL LEAD STANDARD ACCESSORIES 12345 name & description 80009 TK0623 74868 0GZV8 80009 80009 80009 80009 80009 80009 80009 80009 80009 Mfr. code 071-0634-XX RTI 8454001829 901-- 9895-- RFX 64SMA-- 50-- 0-- 1 206-0364-XX 196-3482-XX 016-1315-00 ---------- 016-1774-00 016-1773-10 016-1772-10 131-5638-11 196-3456-00 Mfr. part no. Replaceable Parts 59 60 3 4 1 1 SF501 SMK4 003-- 1433-- 02 015-0678-00 SMG50 -3 -4 -5 -6 1 1 1 1 SF502 1 Qty -2 Serial no. Effective Dscont SF503 Tektronix part no. 24--- 1 Fig. & index no. 2 Figure 24: P6209 optional accessories 1 6 P6209 4 GHz 5X Active Probe MICROCKT TEST,SMT KLIPCHIP,PKG OF 20 (SEE 23-- 9) ADAPTER, PROBE:TIP,SMA 2 COMPACT SIZE,TERMINATED INTO 50 OHM TEST FIXTURE TOOL:ADJUSTMENT TOOL,PLASTIC,BLACK TIP,PROBE:MICROCKT TEST,PKG OF 4 ADPTR,SUREFOOT:50 MIL JEDEC,PKG OF 12 ADPTR,SUREFOOT:25 MIL JEDEC,PKG OF 12 ADPTR,SUREFOOT:0.5 MM EIAJ,PKG OF 12 OPTIONAL ACCESSORIES 12345 name & description 5 80009 80009 TK2565 80009 80009 80009 80009 Mfr. code SMG50 015-0678-00 003-- 1433-- 02 SMK4 SF501 SF502 SF503 Mfr. part no. Replaceable Parts Manufacturer UNITREK CORPORATION HUBER & SUHNER INC M/A COM OMNI SPECTRA INC TEKTRONIX INC GENERAL TOOL & SUPPLY CO VISION PLASTICS INC KENT H LANDSBERG CO Mfr. code 060D9 0GZV8 26805 80009 TK0623 TK2565 TK6108 27929 SW 95TH, SUITE 101 26000 SW PARKWAY CENTER DRIVE 2705 NW NICOLAI ST 14150 SW KARL BRAUN DR PO BOX 500 MICROWAVE CONNECTOR DIV 140 4TH AVE 19 THOMPSON DRIVE 3000 COLUMBIA HOUSE BLVD, SUITE 120 Address CROSS INDEX - MFR. CODE NUMBER TO MANUFACTURER WILSONVILLE, OR 97070 WILSONVILLE, OR 97070 PORTLAND, OR 97210 BEAVERTON, OR 97077-- 0001 WALTHAM, MA 02254 ESSEX JUNCTION, VT 05452-- 3408 VANCOUVER, WA 98661 City, state, zip code Replaceable Parts P6209 4 GHz 5X Active Probe 61 Replaceable Parts 62 P6209 4 GHz 5X Active Probe
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : Yes Encryption : Standard V1.2 (40-bit) User Access : Print, Copy, Annotate, Fill forms, Extract, Assemble, Print high-res Create Date : 2003:08:18 13:31:02Z Modify Date : 2003:08:25 19:20:43-06:00 Page Count : 74 Page Mode : UseOutlines Creation Date : 2003:08:18 13:31:02Z Producer : Acrobat Distiller 4.05 for Windows Mod Date : 2003:08:25 19:20:43-06:00 Author : Tektronix, Inc. Metadata Date : 2003:08:25 19:20:43-06:00 Title : P6209 4 GHz 5X Active Probe Instruction Manual Creator : Tektronix, Inc.EXIF Metadata provided by EXIF.tools