Instruction Manual

P6209
4 GHz 5X Active Probe
071-0634-04

Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.

www.tektronix.com

Copyright © Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and
pending. Information in this publication supercedes that in all previously
published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX, TEK, TEKPROBE, and SureFoot are registered trademarks,
and SureToe, FlexLead, and KlipChip are trademarks of Tektronix, Inc.

WARRANTY
Tektronix warrants that this product will be free from defects in materials and
workmanship for a period of one (1) year from the date of shipment. If any such
product proves defective during this warranty period, Tektronix, at its option,
either will repair the defective product without charge for parts and labor, or will
provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of
the defect before the expiration of the warranty period and make suitable
arrangements for the performance of service. Customer shall be responsible for
packaging and shipping the defective product to the service center designated
by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of
the product to Customer if the shipment is to a location within the country in
which the Tektronix service center is located. Customer shall be responsible for
paying all shipping charges, duties, taxes, and any other charges for products
returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not
be obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than Tektronix representatives to install, repair
or service the product; b) to repair damage resulting from improper use or
connection to incompatible equipment; or c) to service a product that has been
modified or integrated with other products when the effect of such modification
or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS
PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE
PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE
CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS
VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER
TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents
General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

v
vii
viii

Product Description
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Features and Standard Accessories . . . . . . . . . . . . . . . . . . . . . .
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1
2
7

Configuration
Probe Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

9

Functional Check
Operating Basics
Handling the Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maximum Nondestructive Input Voltage . . . . . . . . . . . . . . . . .
Input Linear Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . .
Ground Lead Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

13
13
13
14

Helpful Hints
Low-inductance Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SureFoot Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Probe Tip Test Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

17
18
19

Specifications

P6209 4 GHz Probe Instruction Manual

i

Table of Contents

Theory of Operation
Probe Head and Cable Assembly . . . . . . . . . . . . . . . . . . . . . . .
Compensation Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Offset Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Probe Identification EEPROM . . . . . . . . . . . . . . . . . . . . . . . . .
TEKPROBE SMA Interface . . . . . . . . . . . . . . . . . . . . . . . . . . .
VCC, +7 V Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

30
30
30
30
31
31

Performance Verification
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Equipment Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Output Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DC Attenuation Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Rise Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

33
34
35
36
38

Adjustments
Removing the Compensation Box Bottom Cover . . . . . . . . . . .
Adjustment Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Offset Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Offset Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

43
44
44
46

Maintenance
Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preparation for Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

49
49
49

Troubleshooting
Replaceable Parts
Parts Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Replaceable Parts List . . . . . . . . . . . . . . . . . . . . . . . .

ii

55
56

P6209 4 GHz Probe Instruction Manual

Table of Contents

List of Figures
Figure 1: Dynamic and offset limitations . . . . . . . . . . . . . . . . .
Figure 2: Probe functional check connections . . . . . . . . . . . . .
Figure 3: Ground lead equivalent circuit . . . . . . . . . . . . . . . . .
Figure 4: Low-inductance grounding . . . . . . . . . . . . . . . . . . . .
Figure 5: Using a SureFoot adapter for grounding . . . . . . . . . .
Figure 6: Using a probe tip as a test point . . . . . . . . . . . . . . . . .
Figure 7: Typical input impedance and phase versus frequency
Figure 8: Typical bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9: Dimensions of probe head and compensation box . .
Figure 10: P6209 simplified schematic diagram . . . . . . . . . . . .
Figure 11: TEKPROBE SMA Interface pin assignments . . . . .
Figure 12: P6209 output zero setup . . . . . . . . . . . . . . . . . . . . . .
Figure 13: P6209 DC attenuation accuracy setup . . . . . . . . . . .
Figure 14: Test system rise time setup . . . . . . . . . . . . . . . . . . .
Figure 15: Test system and probe rise time setup . . . . . . . . . . .
Figure 16: Removing the compensation box bottom cover . . .
Figure 17: P6209 adjustment and measurement locations . . . .
Figure 18: P6209 offset zero setup . . . . . . . . . . . . . . . . . . . . . .
Figure 19: P6209 offset range setup . . . . . . . . . . . . . . . . . . . . .
Figure 20: Compensation box test point locations . . . . . . . . . .
Figure 21: TEKPROBE SMA Interface pin assignments . . . . .
Figure 22: P6209 replaceable parts . . . . . . . . . . . . . . . . . . . . . .
Figure 23: P6209 standard accessories . . . . . . . . . . . . . . . . . . .
Figure 24: P6209 optional accessories . . . . . . . . . . . . . . . . . . .

P6209 4 GHz Probe Instruction Manual

10
11
14
17
18
19
24
24
26
29
31
35
37
38
39
43
44
45
47
52
52
57
58
60

iii

Table of Contents

iv

P6209 4 GHz Probe Instruction Manual

General Safety Summary
Review the following safety precautions to avoid injury and prevent
damage to this product or any products connected to it. To avoid
potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.

To Avoid Fire or Personal Injury
Connect and Disconnect Properly. Connect the probe outputs to the
measurement instrument before connecting the probe to the circuit
under test. Disconnect the probe input and the probe ground from the
circuit under test before disconnecting the probe from the measurement instrument.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe
all ratings and markings on the product. Consult the product manual
for further ratings information before making connections to the
product.
Connect the ground lead of the probe to earth ground only.
Do Not Operate Without Covers. Do not operate this product with
covers or panels removed.
Do Not Operate With Suspected Failures. If you suspect there is
damage to this product, have it inspected by qualified service
personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.

P6209 4 GHz Probe Instruction Manual

v

General Safety Summary

Safety Terms and Symbols
Terms in This Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that
could result in injury or loss of life.

CAUTION. Caution statements identify conditions or practices that
could result in damage to this product or other property.

Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you
read the marking.
WARNING indicates an injury hazard not immediately accessible as
you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. These symbols may appear on the product:

CAUTION
Refer to Manual

vi

P6209 4 GHz Probe Instruction Manual

Service Safety Summary
Only qualified personnel should perform service procedures. Read
this Service Safety Summary and the General Safety Summary before
performing any service procedures.
Do Not Service Alone. Do not perform internal service or adjustments
of this product unless another person capable of rendering first aid
and resuscitation is present.

P6209 4 GHz Probe Instruction Manual

vii

Service Safety Summary

Contacting Tektronix
Phone

1-800-833-9200*

Address

Tektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA

Web site

www.tektronix.com

Sales
support

1-800-833-9200, select option 1*

Service
support

1-800-833-9200, select option 2*

Technical
support

Email: techsupport@tektronix.com
1-800-833-9200, select option 3*
6:00 a.m. - 5:00 p.m. Pacific time

*

viii

This phone number is toll free in North America. After office
hours, please leave a voice mail message.
Outside North America, contact a Tektronix sales office or
distributor; see the Tektronix web site for a list of offices.

P6209 4 GHz Probe Instruction Manual

Product Description
The Tektronix P6209 is a 4 GHz (probe only), 5X active FET probe
with 1 pF input capacitance. The probe is designed primarily for use
with sampling oscilloscopes, such as the CSA/TDS8000. The P6209
has a low input capacitance and high input resistance, which
minimize circuit loading over a wide bandwidth range.
The small profile and low-mass head of the P6209 make manual
probing of crowded circuits fast and easy. The accessory tips and
adapters enable the P6209 to be used on a wide variety of circuit
architectures.
The P6209 is powered through a TEKPROBE SMA interface
between the probe compensation box and the oscilloscope.

Options
The following options are available when ordering the P6209 probe:
H Option D1 - Calibration Data
H Option C3 - 3 years Calibration Service
H Option D3 - 3 years Calibration Data
H Option R3 - 3 years Extended Warranty

P6209 4 GHz Probe Instruction Manual

1

Product Description

Features and Standard Accessories
The P6209 probe is provided with several features and accessories
designed to make probing and measurement a simpler task. Please
familiarize yourself with these items and their uses. For part number
information and quantities of standard and optional accessories, refer
to Replaceable Parts on page 55.
Probe head assembly. The probe head is
designed for ease of use and high performance.
Its small size makes it easy to handle in tight
areas.

Probe tip
socket
Ground
socket
Stabilization
notch

The probe tip socket is sized to easily press onto
0.025 inch pins for direct access. The ground
socket provides a short ground path for high
fidelity ground connections.
The stabilization notch permits you to use
adjacent pins to reduce stresses on the probe
and pins.

TEKPROBE SMA interface. The TEKPROBE
SMA interface provides a communication path
between the probe and the oscilloscope. Contact
pins provide power, offset, and probe characteristic data transfer. See page 31 for more information.
Color marker
bands

2

Color marker bands. Attach matching pairs of
the color marker bands onto the cable at the head
and compensation box of each probe. The marker
bands enable quick verification of which probe is
connected to which instrument channel.

P6209 4 GHz Probe Instruction Manual

Product Description

SureToe probe tip. The SureToe tip is a pointed
probe tip useful for probing in dense circuitry.
Attach the SureToe adapters the same way as
the push-in probe tips. Do not force the tip. Also,
be careful not to injure yourself with the sharp
probe tip. They can be used with any of the
socketed accessory leads.
Push-in probe tip. Use the push-in probe tip
for general purpose probing by hand. The tip may
also be used as a temporary test point. See
page 19 for more information.
Push-in probe tip

The push-in probe tip may also be used with the
other socketed leads and adapters.

Installing the push-in probe tip. Attach the
push-in probe tip by seating the tip into the probe
tip socket and pushing the tip in until it is seated.
Either end of the tip may be used.
Do not force the tip. Also, be careful not to injure
yourself with the sharp probe-tip. To remove the
tip, gently grab the tip with small pliers and pull
the tip out.
Right-angle adapter. Use the right-angle
adapter for low-profile probing of 0.025 inch
square pins.
The right-angle adapter allows the probe to lie flat
against a circuit board. This enables probing in
vertical circuits, such as computer or communications backplanes, or in tight areas, such as
between circuit cards.

Right
angle
adapter

The right-angle adapter can be used directly with
the probe head or attached to the Y-lead adapter
or ground leads.
The right-angle adapter is attached the same way
as the push-in probe tip and can be easily
removed by hand.

P6209 4 GHz Probe Instruction Manual

3

Product Description

Y-lead
adapter

Y-lead adapter. Use the Y-lead adapter to
extend the physical reach of the probe and
ground when necessary. The Y-lead adapter
accepts any of the probe tips or adapters and
can be pushed directly onto 0.025 inch pins.
When selecting the grounding connection,
maintain as short a ground path as possible.
Refer to page 14 for more information.

Y-lead
adapter

To attach the Y-lead adapter, gently press the
lead pins into the probe head tip and ground
receptacles. Using the black lead for ground is
recommended.

Three-inch ground lead. Use the three-inch
ground lead for general probing. The socketed
end of the lead may be connected to any of the
probe tips and adapters or fitted onto 0.025 inch
pins.
To attach the ground lead, press and rotate the
lead pin connector into the ground socket on the
probe head. The lead may be removed by simply
pulling the pin out by hand.
When selecting the grounding connection,
maintain as short a ground path as possible.
Refer to page 14 for more information.

4

P6209 4 GHz Probe Instruction Manual

Product Description

Customizable ground lead. This ground lead
wire can be bent or cut shorter.
NOTE: To ease insertion into the ground socket
of the probe, cut the tip of this ground lead wire at
a 30 to 60 degree angle.

Customizable
ground lead

To maintain signal fidelity while probing, use as
short a ground path as possible. Refer to page 14
for more grounding information.

Low-inductance ground pogo pin. Use the
low-inductance ground pogo pin to substantially
reduce ground lead inductance. Because the
pogo pin simply touches the ground reference,
you can easily move the probe to different points
on the circuit under test.
Pogo pin

To attach, press the pogo pin into the probe head
ground socket.
When selecting the grounding connection,
maintain as short a ground path as possible.
Refer to page 14 for more information.

Signal-ground adapter. The signal-ground
adapter is ideal for use with signal/ground pairs
on 0.100 inch header pins.
Attach the signal-ground adapter by gently
pressing it into the ground socket on the probe
head.

Signal-ground
adapter

Be sure to use the stabilization notch whenever
possible to avoid slipping and damaging the
probe or circuitry under test.

P6209 4 GHz Probe Instruction Manual

5

Product Description

SMT KlipChip. Use the SMT KlipChip test clips
to access fragile, dense circuitry.
KlipChip test clips can be connected to the Y-lead
or three-inch ground leads. Simply press the lead
socket into the KlipChip handle.

KlipChip

Y-lead adapter

The KlipChip body freely turns, allowing better
probe orientation. To reduce stress and provide a
lower profile on components being tested, the
flexible sleeve of the KlipChip bends up to a
35 degree angle.
When selecting the grounding connection,
maintain as short a ground path as possible.
Refer to page 14 for more information.

SMA adapter jack. Use the SMA adapter jack
with the SMT KlipChips to gain access to
calibration signals from SMA connectors on the
oscilloscope front panel.
Male SMA termination. Protect the probe input
circuitry by connecting the termination to the
probe input connector when the probe is not in
use.

6

P6209 4 GHz Probe Instruction Manual

Product Description

Optional Accessories
The following accessories are available for the probe. See the
Replaceable Parts List on page 55 for ordering information.
SureFoot probe tips. The SureFoot tip is an
integral probe tip and miniature guide that
enables fault-free probing of fine-pitch SMD
packages. Attach SureFoot adapters the same
way as the push-in probe tips. They can be used
with any of the socketed accessory leads.
The orange, 0.050 inch SureFoot tip is compatible
with 50 mil JEDEC packages such as SOIC,
PLCC, and CLCC.
The blue, 0.025 inch SureFoot tip is compatible
with 0.65 mm JEDEC and EIAJ packages.

The red, 0.5 mm SureFoot tip is compatible with
EIAJ packages.

SMK4 Micro KlipChip adapters. Use the
Micro KlipChip adapters to probe the leads on
integrated circuits that are surface-mounted.

SMA-to-probe tip adapter. Use the adapter to
connect the probe to SMA cables. The adapter
includes a 50-ohm SMA termination.

P6209 4 GHz Probe Instruction Manual

7

Product Description

8

P6209 4 GHz Probe Instruction Manual

Configuration
The P6209 provides the oscilloscope with the probe model number,
serial number, and attenuation factor. When connected to the
oscilloscope, the display readouts are corrected for the probe
attenuation factor.
The probe offset control is controlled by the oscilloscope.
CAUTION. To prevent damage to your probe, do not apply a voltage
outside the Maximum Input Voltage rating. See Specifications on
page 21.

To prevent electrostatic damage to the instrument and sampling
modules, follow the precautions described in the manuals
accompanying your instrument.
Always use a wrist strap (provided with your probe) when handling
the probe and sampling modules, and when making signal
connections.

Probe Offset
The purpose of the offset capability is to permit the input signal to be
positioned in the center of the dynamic range of the probe amplifier,
to obtain the best transient response. Using the offset to cancel DC
signal components enables optimal probe performance. See Figure 1
on page 10 for more information.
NOTE. If the offset is not correctly positioned, clipping or distortion
of the displayed signal may occur. See your oscilloscope manual for
specific instructions on its operation and offset control.

P6209 4 GHz Probe Instruction Manual

9

Configuration

To set the probe offset, follow these steps:
1. Set the oscilloscope vertical scale to the least sensitive V/div
setting (500 mV/div on the CSA/TDS8000).
2. Attach the probe to the circuit.
3. Adjust the probe offset to bring the trace onto the oscilloscope
screen.
4. Change the volts/division setting to the desired range, adjusting
the offset to keep the trace on screen.
NOTE. The P6209 has a ±5.0 V offset range. The linear operating
range is ±2.00 V. See Figure 1. Also, see page 13 for more information.

Nonoperating range (+30 V maximum nondestructive input voltage )
+7.00 V
+5.00 V

+5.00 V
+3.00 V
+2.00 V
0V

Maximum AC Signal Amplitude

Maximum Offset Range

- 2.00 V
- 3.00 V
- 5.00 V

- 5.00 V

- 7.00 V
Nonoperating range (-- 30 V maximum nondestructive input voltage )

Figure 1: Dynamic and offset limitations

10

P6209 4 GHz Probe Instruction Manual

Functional Check
After installing the probe on the CSA/TDS8000 oscilloscope, a
functional check may be performed using the INTERNAL CLOCK
OUTPUT connection on the front panel of the oscilloscope. See
Figure 2.

SMA
adapter jack

Figure 2: Probe functional check connections
1. Set the oscilloscope to display the probe channel.
2. Using the Y-lead connector, SMT KlipChips and the SMA
adapter jack accessories, connect the probe to the INTERNAL
CLOCK OUTPUT connection on the oscilloscope, as shown in
Figure 2.
3. In the setup menu, set the trigger source to internal clock and the
frequency to 200 kHz.

P6209 4 GHz Probe Instruction Manual

11

Functional Check

4. Set the offset to 1V, the vertical scale to 500 mV/div, and the
horizontal scale to 2 s/div.
5. The observed waveform should be a 200 kHz square wave with
an amplitude of approximately 2 V (the clock output is
unterminated), and have an offset of approximately 1 V.
If the probe does not pass this functional check, go to Troubleshooting on page 51.

12

P6209 4 GHz Probe Instruction Manual

Operating Basics
Please follow these operating guidelines to get optimum performance
from your P6209 probe.

Handling the Probe
Exercise care when using and storing the P6209 probe. The probe
and cable are susceptible to damage caused by careless use. Always
handle the probe using the compensation box and probe head,
avoiding undue physical strain to the probe cable, such as kinking,
excessive bending, or pulling. Visible dents in the cable will increase
signal aberrations. Do not drop the probe or subject it to physical
shock. Damage to the probe may result.
CAUTION. To prevent damage to your probe, do not apply a voltage
outside the Maximum Nondestructive Input Voltage for the probe.

Always use a wrist strap (provided with your probe) when
handling the probe and sampling modules, and when making
signal connections.

Maximum Nondestructive Input Voltage
Applying voltages above the P6209 design limits may damage the
probe tip amplifier. Please refer to Specifications on page 21 for the
maximum operating voltage and frequency derating information.

Input Linear Dynamic Range
The probe head amplifier used by the P6209 has a limited linear
operating range. To keep the input linearity error less than 0.1%, you
must limit the apparent signal input voltage to ±1.75 V.

P6209 4 GHz Probe Instruction Manual

13

Operating Basics

Use the DC offset adjustment to maintain the probe within its
dynamic range. The nominal offset adjustment range of the P6209 is
±5.0 VDC. For example: to offset a +2 VDC level in a circuit, set the
offset to +2 V.
NOTE. The probe can tolerate input voltages of ±30 V (≤5 minutes)
without damage; however, the linearity error specification does not
apply to input voltages outside the linear dynamic range of the
probe. See Figure 1 on page 10.

Ground Lead Length
When you are probing a circuit, you should always use as short a
ground lead as possible between the probe head and circuit ground.
The series inductance added by the probe tip and ground lead can
result in a resonant circuit; this circuit may cause parasitic ringing
within the bandwidth of your oscilloscope. When you touch your
probe tip to a circuit element, you are introducing a new resistance,
capacitance, and inductance into the circuit. Refer to Figure 3.

R source
Probe R in
20 KΩ

V source

Probe C in
1 pF

L gl (Ground Lead)

Figure 3: Ground lead equivalent circuit
Ringing and rise time degradation can be masked if the frequency
content of the signal degradation is beyond the bandwidth of the
oscilloscope.

14

P6209 4 GHz Probe Instruction Manual

Operating Basics

You can determine if ground lead effects may be a problem in your
application if you know the self-inductance (L) and capacitance (C)
of your probe and ground lead. Calculate the approximate resonant
frequency (f0) at which this parasitic circuit will resonate with the
following formula:
f0 =

1
2π LC

The preceding equation shows that reducing the ground lead
inductance will raise the resonant frequency. If your measurements
are affected by ringing, your goal is to lower the inductance of your
ground path until the resulting resonant frequency is well above the
frequency of your measurements.
The low-inductance ground contacts described in Accessories can
help you reduce the effects of ground lead inductance on your
measurements.

P6209 4 GHz Probe Instruction Manual

15

Operating Basics

16

P6209 4 GHz Probe Instruction Manual

Helpful Hints
Follow these helpful hints to make probing easier and noise free.

Low-inductance Grounding
Placing a ground plane on top of a package being probed can
minimize ground lead length and inductance. See Figure 4.

Figure 4: Low-inductance grounding
Attach a small piece of copper clad on top of the package and
connect it to the package ground connection. Use the low-inductance
ground lead provided with the P6209 to keep the ground lead length
as short as possible.

P6209 4 GHz Probe Instruction Manual

17

Helpful Hints

This method is very useful when making many measurements on the
same package. Using a ground plane on the package makes probing
the package easier and avoids adding unnecessary ground lead length
and distortion.

SureFoot Grounding
If you cannot use the recommended low-inductance grounding
method, you may ground the probe to the package under test using a
SureFoot adapter. Refer to Figure 5.

Figure 5: Using a SureFoot adapter for grounding
Use a SureFoot adapter at the end of a short ground lead to connect
directly to the package ground. This method is preferred over using
an adjacent circuit ground because it is the shortest ground path
possible.

18

P6209 4 GHz Probe Instruction Manual

Helpful Hints

Probe Tip Test Points
You can solder the push-in probe tip or a 0.025 square pin to a circuit
to be used as a temporary test point. See Figure 6.
Solder the tip onto a lead or pin with a low-power soldering iron.
Press the probe head onto the tip to make a measurement, and then
pull the probe head off when you are done.
The probe tip may be removed and reused by desoldering it from the
circuit and soldering it into another circuit in the future.

Solder

Figure 6: Using a probe tip as a test point
NOTE. Do not use pieces of solid-core copper wire as test points. If
the wire breaks off in the probe tip socket, it may be impossible to
remove the wire, and it will prevent insertion of other accessory tips.

P6209 4 GHz Probe Instruction Manual

19

Helpful Hints

20

P6209 4 GHz Probe Instruction Manual

Specifications
These specifications apply to a P6209 probe when used with a
CSA/TDS8000 oscilloscope, and other oscilloscopes with a
TEKPROBE SMA interface.
The probe and oscilloscope must first be allowed to warm up for
20 minutes before measurements are taken.
CAUTION. Do not apply voltages beyond the nondestructive input
voltage range to the probe. Damage to the probe or circuit under test
may result.
Table 1: Warranted electrical specifications
DC Gain Accuracy (probe only)

0.2 ±2% (excludes offset error)

Rise Time (small signal, probe only)

≤120 ps (≤ 250 mV, 20--30° C)

Output Zero

±7.5 mV or less at output of probe (does not
include oscilloscope offset control voltage
error) (20--30° C)
±37.5 mV or less displayed on screen with
TEKPROBE SMA interface (does not
include oscilloscope offset control voltage
error) (20--30° C)

Delay Time

5.95 ns ±0.2 ns (includes SMA output cable)

System Noise

300 V rms or less at output of probe with
probe tip grounded

P6209 4 GHz Probe Instruction Manual

21

Specifications

Table 2: Typical electrical characteristics
Bandwidth, (probe only)

4 GHz (See Figure 8)

Linear Input Dynamic Range

--2.00 V to +2.00 V.
(Equivalent to --0.40 V to +0.40 V
at the output of the probe.)

Linearity

ᐔ0.1% over a dynamic range of --1.75 V to
+1.75 V (±3.5 mV)

Maximum deviation from linear regression
expressed as a percentage of the specified
full-scale dynamic range

ᐔ1.0% over a dynamic range of --2.00 V to
+2.00 V (±40 mV)

Nondestructive Input Voltage Range

--30 V to +30 V (DC + peak AC) (5 minutes
maximum continuous exposure)

Input Resistance

20 kΩ at DC. (See Figure 7)

Input Capacitance

0.85 pF @100 MHz

Offset Range

--5.0 V to +5.0 V

DC Offset Drift

150 V/°C or less at output of probe
0.75 mV/°C or less displayed on screen with
TEKPROBE SMA interface

DC Offset Scale Accuracy
(gain of offset signal path)

ᐔ2.0% (of 5X actual probe gain)

DC Voltage Measurement Accuracy
(referred to input)

ᐔ[2% of input + (2% of offset) + 37.5* mV +
(3.5 mV or 40 mV†)]

*

Worst case value.

†

Worst case values - see linearity specification.

22

P6209 4 GHz Probe Instruction Manual

Specifications

Calculating DC Voltage Measurement Accuracy:
DC Voltage Measurement Accuracy (referred to input)
= [2% of input + (2% of offset) + 37.5 mV + 3.5 mV]
where:
gain error = 2% of input voltage (relative to offset voltage)
offset gain error = 2% of effective offset at probe tip
output zero = 37.5 mV effective at probe tip (worst case, but
does not include any oscilloscope offset control voltage error)
linearity error1 = 0.1% of 1.75 V (3.5 V full scale)=3.5 mV
High Error Example. To determine the maximum specified DC error
when measuring a PECL logic high of approximately 4.1 V with
probe offset set to 3.7 V:
(2% of input voltage = 8 mV, +
2% of offset voltage = 74 mV, +
output zero = 37.5 mV (worst case value), +
linearity error = 3.5 mV)
= 123 mV
Low Error Example. To determine the maximum specified DC error
when measuring a LVDS low of approximately 1.0 V without any
probe offset (0 V):
(2% of input voltage = 20 mV, +
2% of offset voltage = 0 mV, +
output zero = 5 mV (effective offset at probe tip due to typical
offset value of 1 mV at probe output), +
linearity error = 3.5 mV)
= 28.5 mV
1

The linearity specification is rated with a segmented range (see page 22).
The linearity error is higher at the extreme ends of the dynamic range.

P6209 4 GHz Probe Instruction Manual

23

Specifications

90°

20 K

72°

16 K

36°

14 K

18°

12 K

0°

10 K

- 18°

8K

Phase

- 36°

6K

- 54°

4K

- 72°

2K

- 90°

1M

10 M

100 M
Frequency

1G

5G

Magnuitude (ohms)

54°
Phase (degrees)

18 K

Magnitude

0

Figure 7: Typical input impedance and phase versus frequency
- 10
- 12
- 14

Gain dB

- 16
- 18
- 20
- 22

Gain = 20 Log

- 24

Ꮛ Ꮠ
V OUT
V IN

- 26
- 28
1 MHz

10 MHz

100 MHz

1 GHz

10 GHz

Frequency

Figure 8: Typical bandwidth

24

P6209 4 GHz Probe Instruction Manual

Specifications

Table 3: Environmental characteristics
Operating Temperature

0° C to +40° C
The environmental exposure procedure used as
stated in Tektronix Design Standard 062-2847-00 for
Class 5 equipment.

Nonoperating Temperature

--40° C to +71° C
The environmental exposure procedure used as
stated in Tektronix Design Standard 062-2847-00 for
Class 5 equipment.

Humidity

80% RH at 40° C
The environmental exposure procedure used as
stated in Tektronix Design Standard, 062-2847-00 for
Class 5 equipment.

Altitude

Operating: 15,000 ft.
NonOperating: 50,000 ft.

P6209 4 GHz Probe Instruction Manual

25

Specifications

Table 4: Physical Characteristics
Net Shipping Weight

3.1 lb (1.4 kg)

Cable Length (probe tip to SMA)

(1.37 meters)

Figure 9: Dimensions of probe head and compensation box

26

P6209 4 GHz Probe Instruction Manual

WARNING
The following servicing instructions are for use only by
qualified personnel. To avoid injury, do not perform any
servicing other than that stated in the operating instructions
unless you are qualified to do so. Refer to all safety
summaries before performing any service.

Theory of Operation
There are no user replaceable parts within the probe or the
compensation box; however, this theory of operation is provided to
assist you in isolating failures to either the probe or the host
oscilloscope. Refer to Figure 10 for the simplified schematic.
Probe head

Compensation box

Oscilloscope

Probe tip
Signal Out
Probe tip
amplifier

Probe ID Out
Clock In

EEPROM
Offset ampifier

Offset to probe
+5

±1 V Offset
Offset
zero
-5

Offset
gain
+7 V
Linear
regulator

+15 V

-5 V
Ground
Probe
cable

TEKPROBE
SMA interface

Figure 10: P6209 simplified schematic diagram

P6209 4 GHz Probe Instruction Manual

29

Theory of Operation

Probe Head and Cable Assembly
The probe head assembly contains an active amplifier circuit that
buffers and amplifies the input signal. The amplifier receives power
and an offset level from the compensation box assembly via the
cable assembly.
All signal amplification and buffering is performed in the probe head
assembly. No further amplification takes place in the compensation
box.

Compensation Box
The compensation box contains the following circuits:
H Offset amplifier
H Probe identification EEPROM
H TEKPROBE SMA interface
H VCC, +7 V linear regulator
Offset Amplifier
The offset amplifier is used to offset the input signal DC component
so that it stays at the optimal point of the probe linear dynamic
range. For more information on the linear dynamic range characteristic, refer to Input Linear Dynamic Range on page 13.
The offset amplifier receives offset information as a ±1 VDC voltage
from the oscilloscope. The amplfier then amplifies it to match the
probe characteristics and applies it to the probe head circuit.
The offset amplifier has two adjustments: offset zero and offset scale.
These adjustments rarely need attention; however, detailed
adjustment instructions are in the Adjustment section on page 43.
Probe Identification EEPROM
The probe identification EEPROM is used to configure the
oscilloscope to the probe. The EEPROM receives a clock input from

30

P6209 4 GHz Probe Instruction Manual

Theory of Operation

the oscilloscope, and information about the probe is passed to the
oscilloscope.
TEKPROBE SMA Interface
The TEKPROBE SMA interface provides a communication path
between the probe and the oscilloscope. Contact pins provide power,
signal, offset, and data transfer for the probe identification
EEPROM.
Figure 11 shows the probe interface pin functions. Refer to your
oscilloscope service documentation for more detailed specifications.
GND
5
6
7

+15 V
+5 V

4
12
10

8
9

3

11
2

- 15 V
-5 V

1

CLK

DATA

Figure 11: TEKPROBE SMA Interface pin assignments
VCC, +7 V Regulator
The +15 V input is regulated to +7 V to supply the IC, VCC.

P6209 4 GHz Probe Instruction Manual

31

Theory of Operation

32

P6209 4 GHz Probe Instruction Manual

Performance Verification
Use the following procedures to verify the warranted specifications
of the P6209 probe. Before beginning these procedures, refer to
page 41 and photocopy the test record and use it to record the
performance test results. The recommended calibration interval is
one year.
These procedures test the following specifications:
H Output offset voltage
H DC attenuation accuracy
H Rise time

Equipment Required
Refer to Table 5 for a list of the equipment required to verify the
performance of your probe.
NOTE. These procedures use a CSA/TDS8000 as the host instrument.
If you are using a TDS 820, the procedures will vary slightly.
Table 5: Equipment required for performance verification and adjustment
Item description

Performance
requirement

Oscilloscope

TEKPROBE SMA interface

Calibration Step Generator

Recommended example
Tektronix CSA/TDS8000
with a 80E00 Series
Sampling Module, or
TDS 820
067-1338-XX

DC Power Supply1

1 VDC at 1 mA

Tektronix PS280

DMM

0.5% accuracy at 2 VDC

Tektronix TX3

P6209 4 GHz Probe Instruction Manual

33

Performance Verification

Table 5: Equipment required for performance verification and adjustment (Cont.)
Item description

Performance
requirement

Recommended example

Feedthrough Termination

BNC, 50 Ω ±0.05 Ω

011-0129-00

Coaxial cables (2)

Male-to-Male SMA, 20 in

174-1427-00

Adapter

High Frequency Probe tip
adapter w/50 Ω Termination

015-0678-00

Adapter

SMA Female-to-Female

015-1012-00

Adapter

SMA Female-to-BNC Male

015-0572-00

Adapter

SMA Male jack

131-7128-00

Adapter

BNC Female-to-Dual
Banana

103-0090-00

Adapter

Y-lead adapter

196-3457-00

Adapters (2)

KlipChip adapter

206-0364-XX

1

Provides DC stimulus if using a TDS 820

Equipment Setup
Use this procedure to set up the equipment to test the probe.
1. Connect the probe to the oscilloscope.
2. Connect a Y-lead adapter with KlipChip adapters attached to the
probe.
3. Turn on the oscilloscope and enable the channel.
4. Allow 20 minutes for the equipment to warm up.

34

P6209 4 GHz Probe Instruction Manual

Performance Verification

Output Zero
Use this procedure to verify the probe output zero.
1. Connect the test equipment as shown in Figure 12.
2. Ground the probe tip by connecting the two KlipChips together.
CSA/TDS8000

Digital
multimeter

BNC-to-dual
banana adapter

KlipChip
adapter

50 Ω Precision
termination
P6209 Probe

SMA F-to-BNC M
adapter

Y-lead
adapter

Figure 12: P6209 output zero setup
3. Set the oscilloscope trigger source to internal clock.
4. Set the oscilloscope channel vertical offset to 0.00 V.
5. Set the multimeter to read DC volts.
6. Record the multimeter reading on the test record.
The measured DC level should be 0.00 V ±10 mV for
the CSA/TDS8000.

P6209 4 GHz Probe Instruction Manual

35

Performance Verification

NOTE. The P6209 probe is guaranteed to have less than 7.5 mV
offset. However, the CSA/TDS8000 and TDS 820 will have some
residual voltage remaining when set to 0.00 V offset that will appear
in this test as additional offset at the probe output.

DC Attenuation Accuracy
Use this procedure to verify the probe DC attenuation accuracy.
1. Set the oscilloscope trigger to internal clock.
2. Set the oscilloscope channel vertical offset to 0.00 V.
3. Set the multimeter to read DC volts.
4. Connect the probe tip and ground to the DC calibrator output of
the CSA/TDS8000, using the Y-lead adapter, KlipChips and SMA
adapter jack.
If the probe is being used with a TDS 820 oscilloscope, a separate
DC power supply is needed to provide the stimulus. In this case,
connect the probe to the power supply using the Y-lead adapter
and KlipChips.
Record the results of steps 5 through 11 separately. Use the results to
calculate the DC accuracy of the probe in step 12.
5. Set the DC calibration output of the CSA/TDS8000 (or external
power supply if using a TDS 820), to +1.000 V. Use the
multimeter to verify the DC voltage is as close to +1.000 V as
possible. Record this measurement as Vmax.
6. Connect the multimeter as shown in Figure 13 on page 37.

36

P6209 4 GHz Probe Instruction Manual

Performance Verification

CSA/TDS8000
Digital multimeter

SMA adapter jack

BNC-to-dual
banana adapter

50 Ω Precision
termination

DC
CALIBRATION
OUTPUT

SMA F-to-BNC M
adapter
P6209 Probe

Figure 13: P6209 DC attenuation accuracy setup
7. Measure and record the multimeter reading as M1.
8. Disconnect the multimeter.
9. Set the DC calibration output of the CSA/TDS8000 (or external
power supply if using a TDS 820), to - 1.000 V. Use the
multimeter to verify the DC voltage is as close to - 1.000 V as
possible. Record this measurement as Vmin.
10. Reconnect the multimeter as shown in Figure 13.
11. Measure and record the multimeter reading as M2.
12. Determine the percent error by using the formula below:

%Error =

M1–M2
–1 × 100%
(Vmax–Vmin) × .2

The calculated error should be ≤2%.
13. Record the results on the test record.

P6209 4 GHz Probe Instruction Manual

37

Performance Verification

Rise Time
This procedure verifies that the probe meets rise time specifications.
The probe rise time is calculated from rise times obtained by
measuring the test system without the probe attached, and the test
system plus the probe.
1. Connect the test equipment as shown in Figure 14.
2. Set the oscilloscope to the following settings:
H Trigger to internal clock
H Vertical scale to 50 mV/div
H Horizontal scale to 100 ps/div
H Horizontal record length to 4000 points
H Acquisition Mode to average at least 16 samples
3. Adjust the oscilloscope horizontal and vertical position controls
to display a signal similar to that shown in Figure 14.
CSA/TDS8000

INTERNAL
CLOCK
OUTPUT
SMA cable
Generator
remote head

Calibration
step generator

Trigger
input

To power
supply

Figure 14: Test system rise time setup

38

P6209 4 GHz Probe Instruction Manual

Performance Verification

4. Use the oscilloscope measurement capability to calculate and
display rise time. Rise time is determined from the 10% and 90%
amplitude points on the waveform. Record the rise time as ts.
The system rise time (ts ) that you measured in step 4 represents the
rise time of the test system without the probe.
The following steps instruct you to assemble the test system that
includes the probe, as shown in Figure 15. The system and probe rise
time (ts+p ) that you measure in step 8 is used to calculate the probe
rise time (tp ) in step 9.
5. Connect the test equipment as shown in Figure 15.
CSA/TDS8000

INTERNAL
CLOCK
OUTPUT

Calibration
step generator

SMA cable

Trigger
input
To power
supply

P6209 Probe

Generator remote head
SMA F-to-F
HF probe tip adapter
50 Ω Termination

Figure 15: Test system and probe rise time setup
6. Set the vertical to 50 mV/div and horizontal to 500 ps/div.
7. Adjust the oscilloscope horizontal and vertical position controls
to display a signal similar to that shown in Figure 15.

P6209 4 GHz Probe Instruction Manual

39

Performance Verification

NOTE. Do not touch the probe tip adapter when making calibration
measurements. Measurement accuracy is degraded when the probe
tip adapter is handled.
8. Use the oscilloscope measurement capability to calculate and
display rise time. Rise time is determined from the 10% and 90%
amplitude points on the waveform. Record the rise time as ts+p.
9. Calculate the probe only rise time using the following formula:
t p = Ꭹt(s+p) − t s
2

2

10. Check that the calculated rise time meets the probe specification.
The rise time (tp ) must be ≤120 ps.
11. Record the results on the test record.

40

P6209 4 GHz Probe Instruction Manual

Performance Verification

Test record
Probe Model:
Serial Number:
Certificate Number:
Temperature:
RH %:
Date of Calibration:
Technician:
Performance test

Minimum

Incoming

Outgoing

Maximum

Output zero voltage
(at probe output)
± 10 mV (20_ C to 30_ C)1

-- 10 mV

________

________

+ 10 mV

DC attenuation accuracy

-- 2 %

________

________

+2%

Rise time

N/A

________

________

≤120 ps

1

CSA/TDS8000 only

P6209 4 GHz Probe Instruction Manual

41

Performance Verification

42

P6209 4 GHz Probe Instruction Manual

Adjustments
The P6209 has two internal controls: offset zero and offset scale.
These controls rarely need to be adjusted and only should be changed
after a probe performance verification and functional check has been
performed on the oscilloscope.
To make adjustments to the probe, the compensation box bottom
cover must be removed and the equipment allowed to warm up for
20 minutes. Use the optional adjustment tool for making any
adjustments. Refer to Table 5 on page 33 for a list of the equipment
required to adjust your probe.

Removing the Compensation Box Bottom Cover
Follow these steps to gain access to the adjustment controls.
1. Using a Pozidrive screwdriver, remove the four screws that attach
the bottom cover to the compensation box. Gently remove the
cover. Refer to Figure 16.

Figure 16: Removing the compensation box bottom cover

P6209 4 GHz Probe Instruction Manual

43

Adjustments

Adjustment Locations
Refer to Figure 17 for the location of the offset adjustments and
measurement test points.
Offset scale
adjust

Offset zero
adjust
J1

Gnd lug

R3

Offset test point
(white wire)

R2

J2

Figure 17: P6209 adjustment and measurement locations

Offset Zero
Use this procedure to adjust the probe offset zero.
1. Connect the probe power cable to the oscilloscope.
2. Ground the probe tip by connecting the two KlipChip leads
together.
3. Set the oscilloscope trigger source to internal clock.
4. Set the channel vertical offset to 0.00 V.
5. Using the (-- ) lead of the DMM, probe the ground lug in the probe
comp box. Refer to Figure 17 on page 44.

44

P6209 4 GHz Probe Instruction Manual

Adjustments

6. Using the (+) lead of the DMM, carefully probe pin 7 of J2 (white
wire) in the probe compensation box. Refer to Figure 17 on
page 44.
NOTE. The offset voltage measured on pin 7 will be of opposite
polarity and approximately one fifth of the offset value displayed on
the oscilloscope.
7. Adjust the oscilloscope offset so that the voltage measures
0.000 V on the DMM. Set the offset as close to 0 V as possible.
8. Connect the equipment as shown in Figure 18.
9. Adjust Offset Zero (R2) in the probe compensation box for
0.00 V, ±1 mV on the DMM. Refer to Figure 17 on page 44 for
the adjustment location.
CSA/TDS8000

Digital
multimeter

BNC-to-dual
banana adapter

KlipChip
adapter

50 Ω Precision
termination
P6209 Probe

SMA F-to-BNC M
adapter

Y-lead
adapter

Figure 18: P6209 offset zero setup

P6209 4 GHz Probe Instruction Manual

45

Adjustments

Offset Scale
Use this procedure to adjust the offset scale of the probe. The offset
zero of the probe should be checked before making any adjustment
to the offset scale. Refer to Figure 17 on page 44 for the location of
the test points and adjustments.
1. Connect the probe power cable to the oscilloscope.
2. Set the oscilloscope trigger source to internal clock.
3. Set the oscilloscope channel vertical offset to +2.5 V.
4. Ground the probe tip by connecting the two KlipChip leads
together.
5. Using the (-- ) lead of the DMM, probe the ground lug in the probe
compensation box.
6. Using the (+) lead of the DMM, carefully probe pin 7 of J2 (white
wire) in the probe compensation box.
NOTE. The offset voltage measured on pin 7 will be of opposite
polarity and approximately one fifth of the offset value displayed on
the oscilloscope.
7. Adjust the oscilloscope offset so that pin 7 of J2 measures
- 0.50 V on the DMM.
8. Set the DC power supply to +2.5 V, as measured on the DMM.
9. Connect the equipment as shown in Figure 19. Take care not to
disturb the offset level setting.
10. Adjust the Offset Scale (R3) in the probe compensation box so
the output measures 0.000 V, ±1 mV on the DMM.

46

P6209 4 GHz Probe Instruction Manual

Adjustments

CSA/TDS8000
Digital
multimeter
Power supply
50 Ω Precision
termination

BNC-to-Dual
Banana Adapter

--

+

SMA F-to-BNC M
adapter

P6209 Probe

Figure 19: P6209 offset range setup
NOTE. After the offset adjustments have been made, use the
Performance Verification procedures beginning on page 33 to verify
your probe meets specifications.

P6209 4 GHz Probe Instruction Manual

47

Adjustments

48

P6209 4 GHz Probe Instruction Manual

Maintenance
This section describes the maintenance procedures for the P6209.

Inspection and Cleaning
To prevent damage to probe materials, avoid using chemicals that
contain benzine, benzene, toluene, xylene, acetone, or similar
solvents.
Do not immerse the probe or use abrasive cleaners.
Dirt may be removed with a soft cloth dampened with a mild
detergent and water solution or isopropyl alcohol.

Replacement Parts
Refer to Replaceable Parts on page 55 for a list of customer
replaceable parts. Due to the sophisticated design of the P6209
probe, there are no user replaceable parts within the probe.

Preparation for Shipment
If the original packaging is unfit for use or not available, use the
following packaging guidelines:
1. Use a corrugated cardboard shipping carton having inside
dimensions at least one inch greater than the probe dimensions.
The box should have a carton test strength of at least 200 pounds.
2. Put the probe into a plastic bag or wrap to protect it from
dampness.
3. Place the probe into the box and stabilize it with light packing
material.
4. Seal the carton with shipping tape.

P6209 4 GHz Probe Instruction Manual

49

Maintenance

50

P6209 4 GHz Probe Instruction Manual

Troubleshooting
This troubleshooting documentation details isolation to either the
probe or the oscilloscope.
To troubleshoot the P6209, the probe must be attached to an
operating TEKPROBE SMA interface oscilloscope with the bottom
cover of the compensation box removed. Refer to page 43 for
instructions on opening the compensation box.
WARNING. Operating voltages will be present. Observe caution when
probing in the compensation box. Be careful not to touch or ground
energized components.
Work at a Level 1 ESD workstation when performing troubleshooting. The compensation box contains static sensitive devices.

P6209 4 GHz Probe Instruction Manual

51

Troubleshooting

J1

Refer to Table 6 on page 53 for troubleshooting procedures and
Figures 20 and 21 for troubleshooting test points.

Ground

J2

Offset (white)
- 5 V (green)
NC
DATA (yellow)
CLOCK (orange)
+5 V (red)
+15 V (brown)

Figure 20: Compensation box test point locations
GND
5
6
7

+15 V
+5 V

4
12
10

8
9

CLK

3

11
2

- 15 V
-5 V

1

DATA

Figure 21: TEKPROBE SMA Interface pin assignments

52

P6209 4 GHz Probe Instruction Manual

Troubleshooting

Table 6: Troubleshooting guide
Symptom

Checks

Result

Action

No signal out

Check known signal

No signal.

Continue checks

Go to No Probe
Power

Probe power OK.

Replace probe.

Distorted or
nonlinear signal

Is the signal within
No
linear dynamic
range? See page 13.

Set channel vertical
offset to match signal
characteristic.

No Offset Adjust
(Signal present)

Go to Probe Power
row

Continue checks

No Probe Power

Probe power OK

Check variation of
Offset input OK
offset input from host
instrument. See
Figure 20. Offset
control voltage at pin
7 of J2 should vary
between +1 V and
--1 V.

Replace probe.

Check the power
test points in the
compensation box.
See Figure 20.

Power OK

Continue checks.

Power bad

Check for continuity
between the
TEKPROBE SMA
connector pins and
the test points in the
compensation box.
See Figures 20 and
21 on page 52.
Try an adjacent
channel.
Repair host
instrument

P6209 4 GHz Probe Instruction Manual

53

Troubleshooting

54

P6209 4 GHz Probe Instruction Manual

Replaceable Parts
This section contains a list of replaceable parts for the P6209 probe.
Use this list to identify and order replacement parts.

Parts Ordering Information
Replacement parts are available from or through your local
Tektronix, Inc. service center or representative.
Changes to Tektronix instruments are sometimes made to accommodate improved components as they become available and to give you
the benefit of the latest circuit improvements. Therefore, when
ordering parts, it is important to include the following information in
your order:
H Part number
H Instrument type or model number
H Instrument serial number
H Instrument modification number, if applicable
If a part you order has been replaced with a different or improved
part, your local Tektronix service center or representative will
contact you concerning any change in the part number.

P6209 4 GHz Probe Instruction Manual

55

Replaceable Parts

Using the Replaceable Parts List
The tabular information in the Replaceable Parts List is arranged for
quick retrieval. Understanding the structure and features of the list
will help you find the information you need for ordering replacement
parts.
Item Names
In the Replaceable Parts List, an Item Name is separated from the
description by a colon (:). Because of space limitations, an Item
Name may sometimes appear as incomplete. For further Item Name
identification, U.S. Federal Cataloging Handbook H6-1 can be used
where possible.
Indentation System
This parts list is indented to show the relationship between items.
The following example is of the indentation system used in the
Description column:
1 2 3 4 5
Name & Description
Assembly and/or Component
Attaching parts for Assembly and/or Component
(END ATTACHING PARTS)
Detail Part of Assembly and/or Component
Attaching parts for Detail Part
(END ATTACHING PARTS)
Parts of Detail Part
Attaching parts for Parts of Detail Part
(END ATTACHING PARTS)

Attaching parts always appear at the same indentation as the item it
mounts, while the detail parts are indented to the right. Indented
items are part of, and included with, the next higher indentation.
Attaching parts must be purchased separately, unless otherwise
specified.
Abbreviations
Abbreviations conform to American National Standards Institute
(ANSI) standard Y1.1

56

P6209 4 GHz Probe Instruction Manual

P6209 4 GHz 5X Active Probe
2

010-- 0638-- 00

015-- 1021-- 00

174-- 1120-- 00

016-- 1879-- 01

-2

-3

-4

Tektronix
part no.

Serial no.
Effective Dscont

Figure 22: P6209 replaceable parts

22-- 1

Fig. &
index no.

1

1

1

1

1

Qty

3

Mfr.
code

CASE,STORAGE:PLASTIC,W/CND FOAM

CA ASSY,RF:COAXIAL,RFD,50 OHM,8.5 L,SMA,MALE

TERMN,COAXIAL:SHORT CIRCUIT,SMA,FEMALE

TK6108

060D9

26805

PROBE ASSEMBLY:SERVICE REPLACEMENT, SERIALIZED 80009

12345 name & description

4

016-- 1879-- 01

174-- 1120-- 00

2020-- 5353-- 00

010-- 0638-- 00

Mfr. part no.

Replaceable Parts

57

58
8

9

2

Figure 23: P6209 standard accessories

1

3

10

4

11

5

12

6

7

Replaceable Parts

P6209 4 GHz 5X Active Probe

1
1

016-1772-10

016-1773-10

016-1774-00

----------

016-1315-00

196-3482-XX

206-0364-XX

015-- 1020-- 00

131-- 7128-- 00

006-3415-04

071-0634-XX

-3

-4

-5

-6

-7

-8

P6209 4 GHz 5X Active Probe

-9

- 10

- 11

- 12

1

1

2

1

1

1

1

1

1

1

131-5638-11

1

Qty

-2

Serial no.
Effective Dscont

196-3456-00

Tektronix
part no.

23-- 1

Fig. &
index no.

MANUAL,TECH:INSTRUCTION,P6209

STRAP,WRIST:3M TYPE 2214, ADJUSTABLE,6 FT COILED
CORD

CONN,RF JACK:COAX,SMA,50 OHM,MALE,GOLD PL

TERMN,COAXIAL:SHORT CIRCUIT,SMA,MALE

TIP,PROBE:MICROCKT TEST,SMT KLIPCHIP

CUSTOMIZABLE GROUND LEAD, SET OF 5

MARKER KIT,ID:CABLE MARKER BAND,2 EA, VAR COLRS

PROBE,TIP,SURETOE:PACKAGE OF 4 (REORDER ST501,
PACKAGE OF 12)

ACCESSORY KIT:ADAPTER SPRING,PKG OF 10

ACCESSORY KIT:GND PIN LENGTH ADAPTER,PKG OF 10

ACCESSORY KIT:POGO PIN,PKG OF 10

PROBE,TIP:PACKAGE OF 10

LEAD SET:ACCESSORY KIT, INCL THREE 3.0 INCH
GROUND LEAD, TWO DIFFERENTIAL LEAD

STANDARD ACCESSORIES

12345 name & description

80009

TK0623

74868

0GZV8

80009

80009

80009

80009

80009

80009

80009

80009

80009

Mfr.
code

071-0634-XX

RTI 8454001829

901-- 9895-- RFX

64SMA-- 50-- 0-- 1

206-0364-XX

196-3482-XX

016-1315-00

----------

016-1774-00

016-1773-10

016-1772-10

131-5638-11

196-3456-00

Mfr. part no.

Replaceable Parts

59

60
3

4

1
1

SF501

SMK4

003-- 1433-- 02

015-0678-00

SMG50

-3

-4

-5

-6

1

1

1

1

SF502

1

Qty

-2

Serial no.
Effective Dscont

SF503

Tektronix
part no.

24--- 1

Fig. &
index no.

2

Figure 24: P6209 optional accessories

1

6

P6209 4 GHz 5X Active Probe

MICROCKT TEST,SMT KLIPCHIP,PKG OF 20 (SEE 23-- 9)

ADAPTER, PROBE:TIP,SMA 2 COMPACT
SIZE,TERMINATED INTO 50 OHM TEST FIXTURE

TOOL:ADJUSTMENT TOOL,PLASTIC,BLACK

TIP,PROBE:MICROCKT TEST,PKG OF 4

ADPTR,SUREFOOT:50 MIL JEDEC,PKG OF 12

ADPTR,SUREFOOT:25 MIL JEDEC,PKG OF 12

ADPTR,SUREFOOT:0.5 MM EIAJ,PKG OF 12

OPTIONAL ACCESSORIES

12345 name & description

5

80009

80009

TK2565

80009

80009

80009

80009

Mfr.
code

SMG50

015-0678-00

003-- 1433-- 02

SMK4

SF501

SF502

SF503

Mfr. part no.

Replaceable Parts

Manufacturer

UNITREK CORPORATION

HUBER & SUHNER INC

M/A COM OMNI SPECTRA INC

TEKTRONIX INC

GENERAL TOOL & SUPPLY CO

VISION PLASTICS INC

KENT H LANDSBERG CO

Mfr. code

060D9

0GZV8

26805

80009

TK0623

TK2565

TK6108

27929 SW 95TH, SUITE 101

26000 SW PARKWAY CENTER DRIVE

2705 NW NICOLAI ST

14150 SW KARL BRAUN DR
PO BOX 500

MICROWAVE CONNECTOR DIV 140 4TH AVE

19 THOMPSON DRIVE

3000 COLUMBIA HOUSE BLVD, SUITE 120

Address

CROSS INDEX - MFR. CODE NUMBER TO MANUFACTURER

WILSONVILLE, OR 97070

WILSONVILLE, OR 97070

PORTLAND, OR 97210

BEAVERTON, OR 97077-- 0001

WALTHAM, MA 02254

ESSEX JUNCTION, VT
05452-- 3408

VANCOUVER, WA 98661

City, state, zip code

Replaceable Parts

P6209 4 GHz 5X Active Probe

61

Replaceable Parts

62

P6209 4 GHz 5X Active Probe



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
Encryption                      : Standard V1.2 (40-bit)
User Access                     : Print, Copy, Annotate, Fill forms, Extract, Assemble, Print high-res
Create Date                     : 2003:08:18 13:31:02Z
Modify Date                     : 2003:08:25 19:20:43-06:00
Page Count                      : 74
Page Mode                       : UseOutlines
Creation Date                   : 2003:08:18 13:31:02Z
Producer                        : Acrobat Distiller 4.05 for Windows
Mod Date                        : 2003:08:25 19:20:43-06:00
Author                          : Tektronix, Inc.
Metadata Date                   : 2003:08:25 19:20:43-06:00
Title                           : P6209 4 GHz 5X Active Probe Instruction Manual
Creator                         : Tektronix, Inc.
EXIF Metadata provided by EXIF.tools

Navigation menu