Proposed Scope Of Accreditation TRA Cal 2019

User Manual: TRA-Cal-Scope-2019

Open the PDF directly: View PDF PDF.
Page Count: 12

DownloadProposed Scope Of Accreditation TRA-Cal-Scope-2019
Open PDF In BrowserView PDF
Certificate # L2151

Scope of Accreditation
For
Tra-Cal, LLC
7901 Beechcraft Avenue, Suite M & N
Gaithersburg, MD 20879
Todd Chaikin
301-527-9200
In recognition of a successful assessment to ISO/IEC 17025:2005, ANSI/NCSL Z540-1:1994(R2002), and
ANSI/NCSL Z540.3:2006(R2014) sub-clause 5.3 to the following Calibration and Measurement
Capabilities, accreditation has been granted to Tra-Cal, LLC for the following:
Accreditation granted through: May 31, 2019

Calibration
Electrical –
Calibration
Parameter/Equipment

Capacitance1
Range

Expanded Uncertainty of
Measurement (+/-)

1 nF

0.2 pF

0.1 µF

0.0000 29 µF

10 Hz to 10 kHz

190 pF to 1.0999 nF

10 Hz to 3 kHz

1.1 nF to 3.2999 nF

0.028 nF
0.031 nF

10 Hz to 1 kHz

3.3 nF to 10.9999 nF

0.044 nF

10 Hz to 1 kHz

(11 to 32.99) nF

0.21 nF

10 Hz to 1 kHz

(33 to 109.99) nF

0.44 nF

10 Hz to 1 kHz

(110 to 329.99) nF

1.3 nF

(10 to 600) Hz

(0.33 to 1.09) µF

0.007 2 µF

(10 to 300) Hz

(1.1 to 3.29) µF

0.014 µF

(10 to 150) Hz

(3.3 to 10.99) µF

0.045 µF

(10 to 120) Hz

(11 to 32.99) µF

0.19 µF

(10 to 80) Hz

(33 to 109.99) µF

0.69 µF

DC to 50 Hz

(110 to 329.99) µF

2.1 µF

DC to 20 Hz

(0.33 to 1.09) mF

0.009 mF

DC to 6 Hz

(1.1 to 3.29) mF

0.021 mF

DC to 2 Hz

(3.3 to 10.99) mF

0.069 mF

DC to 0.6 Hz

(11 to 32.99) mF

0.32 mF

DC to 0.2 Hz

(33 to 99.99) mF

1.4 mF

Capacitance – Source
1 kHz
1 kHz

Form 28.8—Rev 11 – 3/1/14

Remarks

GR 1404-A Capacitor
GR 1409-T Capacitor
Fluke 5520A Calibrator

Page 1 of 12

Calibration
Parameter/Equipment

Certificate # L2151
Remarks

Range

Expanded Uncertainty of
Measurement (+/-)

Capacitance - Measure
1 kHz

(200 pF to 20 µF)

0.045 µF

Electrical – Current1
Calibration
Parameter/Equipment

Range

Expanded Uncertainty of
Measurement (+/-)

(0 to 2.2) mA
(2.2 to 22) mA
(22 to 220) mA
(0.22 to 2.2) A

0.000 15 mA
0.001 5 mA
0.018 mA
0.24 mA

(2.2 to 11) A

0.007 4 A

(11 to 20) A

0.023A

(20 to 100) A

2.7 A

HP 6260B power
supply

(0 to 100) nA
(0.1 to 1) µA
(1 to 10) µA
(10 to 100) µA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A

0.3 nA
0.000 97 µA
0.000 45µA
0.003 2 µA
0.005 mA
0.007 mA
0.011 mA
0.15 mA

Agilent 3458A 8 ½
Digit Multimeter

DC Current – Source

GenRad 1689
Precision RLC
Digibridge

Remarks

Fluke 5700A Calibrator

Fluke 5520A Calibrator

DC Current – Measure

DC Current - Measure

AC Current - Source

Form 28.8—Rev 11 – 3/1/14

(1 to 10) A
(0 to 25) A
(25 to 50) A
(50 to 100) A
(100 to 750) A
(750 to 1 000) A
(0 to 220) µA
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(0.22 to 2.2) mA
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz

0.025 A
0.29 % of reading
0.34 % of reading
0.34 % of reading
0.36 % of reading
0.31 % of reading
0.2 µA
0.11 µA
0.042 µA
0.18 µA
0.46 µA

Fluke 45 Multimeter
Current Shunt
Monitored with DMM

Fluke 5700A Calibrator

0.002 mA
0.001 1mA
0.000 42 mA
0.001 8 mA
0.004 6 mA

Page 2 of 12

Calibration
Parameter/Equipment

AC Current - Source

AC Current - Source

AC Current - Measure

Form 28.8—Rev 11 – 3/1/14

Range
(2.2 to 22) mA
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(22 to 220) mA
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(0.22 to 2.2) A
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(2.2 to 11) A
40 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
(0 to 100) µA
(10 to 20) Hz
(20 to 45) Hz
45 Hz to 1 kHz
(0.1 to 1.0) mA
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(1 to 10) mA
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(10 to 100) mA
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz

Expanded Uncertainty of
Measurement (+/-)

Certificate # L2151
Remarks

0.021 mA
0.011 mA
0.004 2 mA
0.018 mA
0.046 mA
0.2 mA
0.11 mA
0.047 mA
0.18 mA
0.46 mA

Fluke 5700A Calibrator

0.001 9 A
0.002 2 A
0.025 A
0.007 9 A
0.012 A
0.042 A

Fluke 5700A/5725
Calibrator

0.5 uA
0.22 uA
0.13 uA
0.004 6 mA
0.002 mA
0.001 mA
0.000 77mA
0.000 92 mA
0.005 1 mA
0.008 1 mA
0.049 mA
0.02 mA
0.011 mA
0.007 7 mA
0.011 mA
0.051 mA
0.081 mA

Agilent 3458A 8 ½
Digit DMM

0.49 mA
0.2 mA
0.11 mA
0.077 mA
0.11 mA
0.51 mA
0.81 mA
Page 3 of 12

Calibration
Parameter/Equipment

AC Current - Measure

AC Current - Measure

Electrical – Inductance
Calibration
Parameter/Equipment

Range
(0.1 to 1.0) A
(10 to 20) Hz
(20 to 45) Hz
(45 to 100) Hz
100 Hz to 5 kHz
(5 to 20) kHz
(20 to 50) kHz
(1 to 10) A
45 Hz to 1 kHz

Expanded Uncertainty of
Measurement (+/-)
0.005 A
0.002 4 A
0.001 7 A
0.001 8 A
0.003 9 A
0.012 A
0.34 A

Certificate # L2151
Remarks

Agilent 3458A 8 ½
Digit DMM

Fluke 45 DMM

Range

Expanded Uncertainty of
Measurement (+/-)

Remarks

100 µH

0.29 µH

GR 1482-B

10 mH

0.012 mH

GR 1482-H

1 kHz

100 mH

0.13 mH

GR 1482-L

(0.1 to 1) kHz

500 mH

0.76 mH

GR 1482-N

100 Hz

1H

0.001 3 H

GR 1482-P

1 kHz

1H

0.001 3 H

GR 1482-P

(0.1 to 1) kHz

10 H

0.015 H

GR 1482-T

10 H

0.015 H

GR 1482-T

100 µH to 10 H

0.23 % of reading 0.29 µH

GenRad 1689

Range

Expanded Uncertainty of
Measurement (+/-)

Remarks

1Ω

85 µΩ

L&N 4020B resistor

10 Ω

0.99 mΩ

L&N 4025B resistor

100 Ω

3.8 mΩ

L&N 4030B resistor

1 kΩ

28 mΩ

L&N 4035B resistor

10 kΩ

0.13 Ω

Fluke 742A-10k resistor

100 kΩ

3.4 Ω

L&N 4045B resistor

1 MΩ

38 Ω

L&N 4050B resistor

Inductance – Source
1 kHz
1 kHz

(0.1 to 1) kHz
Inductance - Measure
1 kHz

Electrical – Resistance1
Calibration
Parameter/Equipment

Resistance Source

Form 28.8—Rev 11 – 3/1/14

Page 4 of 12

Calibration
Parameter/Equipment

Resistance Source

Resistance Source

Resistance Measure

Resistance Measure

Form 28.8—Rev 11 – 3/1/14

Range
1
1.9 
10 
19 
100 
190 
1 k
1.9 k
10 k
19 k
100 k
190 k
1 M
1.9 M
10 M
19 M
100 M

Expanded Uncertainty of
Measurement (+/-)

Certificate # L2151
Remarks

0.000 13 
0.000 013 
0.000 094 
0.000 15 
0.000 25 
0.000 41 
0.002 
0.003 3 
0.018 
0.039 
0.22 
0.55 
4.1 
62 kΩ
35 kΩ
53 kΩ
62 kΩ

Fluke 5700A Calibrator

(0 to 11) 
(11 to 33) 
(33 to 110) 
(110 to 330) 
(0.33 to 1.1) k
(1.1 to 3.29) k
(3.3 to 11) k
(11 to 32.99) k
(33 to 110) k
(110 to 330) k
330 k to 1.1 M
(1.1 to 3.3) M
(3.3 to 11) M
(11 to 33) M
(33 to 110) M
(110 to 330) M
(330 to 1 100) M

0.001 2 
0.001 7 
0.001 7 
0.002 9 
0.006 1 
0.034 
0.083 
0.34 
0.84 
3.2 
6.9 
50 
210 
3.5 k
9.7 k
140 kΩ
1.4 MΩ

Fluke 5520A Calibrator

(0 to 1) Ω

81 µΩ

(1 to 10) Ω

0.39 mΩ

(10 to 100) Ω

2.5 mΩ

(0.1 to 1) kΩ

17 mΩ

(1 to 10) kΩ

17 mΩ

(10 to 100) kΩ

1.7 Ω

(0.1 to 1) MΩ

27 Ω

(1 to 10) MΩ

0.61 MΩ

(10 to 100) MΩ

4.2 MΩ

(0.1 to 1) GΩ

8.2 MΩ

Agilent 3458A 8 ½
Digit Multimeter

Agilent 3458A 8 ½
Digit Multimeter

Page 5 of 12

Calibration
Parameter/Equipment

Certificate # L2151
Remarks

Range

Expanded Uncertainty of
Measurement (+/-)

(-200 to 800) °C

0.27 ºC

(-200 to 630) °C

0.2 ºC

Pt 385, 500 Ω

(-200 to 630) °C

0.22 ºC

Pt 385, 1 k Ω

(-200 to 630) °C

0.46 ºC

Pt 3916, 100 Ω

(-200 to 630) °C

0.29 ºC

Pt 3926, 100 Ω

(-200 to 630) °C

0.14 ºC

Cu 427, 10 Ω

(-100 to 260) °C

0.35 ºC

Range

Expanded Uncertainty of
Measurement (+/-)

Remarks

10 MHz to 2.3 GHz
(2.3 to 18.6) GHz
(18.6 to 26.5) GHz

(-20 to 18) dBm
(-20 to 18) dBm
(-20 to 18) dBm

1.1 dB
1.7 dB
2.3 dB

HP 8340B/83554A
Signal Generator

(26.5 to 40) GHz
(40 to 50) GHz

(-25 to 3) dBm
(-25 to 0) dBm

1.2 dB
2.2 dB

HP 83650B Signal
Generator

200 Hz to 10.0 MHz
10MHz to 80.0MHz

(-86.98 to 13.01) dBm
(-86.98 to 13.01) dBm

0.42 dB
0.13 dB

HP 3335A Signal
Generator

RF Power Flatness Source
200 Hz to 10 MHz
10 MHz to 25 MHz
25 MHz to 80 MHz

(-86.98 to 13.01) dBm
(-86.98 to 13.01) dBm
(-86.98 to 13.01) dBm

0.42 dBm
0.14 dBm
0.21 dBm

HP 3335A Signal
Generator

10 MHz to 2.3 GHz
(2.3 to 18.6) GHz
(18.6 to 26.5) GHz

(-20 to 18) dBm
(-20 to 18) dBm
(-20 to 18) dBm

1.1 dB
1.7 dB
2.3 dB

HP8340B Signal
Generator

(26.5 to 40) GHz
(40 to 50) GHz

(-25 to 3) dBm
(-25 to 0) dBm

1.2 dB
2.2 dB

HP 83650B Signal
Generator

RTD Resistance
Simulation
Pt 395, 100 Ω
Pt 385, 200 Ω

Electrical – RF Power
Calibration
Parameter/Equipment

Fluke 5520A Calibrator

RF Power - Source

Form 28.8—Rev 11 – 3/1/14

Page 6 of 12

Calibration
Parameter/Equipment

Range

Expanded Uncertainty of
Measurement (+/-)

(-12 to 0) dB
(-12 to -22) dB
(-22 to -31) dB
(-31 to -40) dB
(-40 to -50) dB
(-50 to -61) dB
(-61 to -71) dB
(-71 to -80) dB
(-80 to -90) dB
(-90 to -100) dB
(-100 to -110) dB
(-110 to -120) dB

0.18 dB
0.21 dB
0.23 dB
0.27 dB
0.29 dB
0.32 dB
0.34 dB
0.51 dB
0.60 dB
0.62 dB
0.65 dB
0.69 dB

150 kHz to 2.6 GHz

0.33 dB + 0.1 dB/dB

(-25 to +20) dBm

100 kHz to 4.2 GHz
10 MHz to 18 GHz
50 MHz to 26.5 GHz
50 MHz to 50 GHz

0.29 dB
0.13 dB
0.23 dB
0.4 dB

Distortion

20 Hz to 20 kHz
(20 to 100) kHz

1.5 dB
3.1 dB

Range

Expanded Uncertainty of
Measurement (+/-)

(0 to 220) mV

6.6 µV

(0.22 to 2.2) V

22 µV

(2.2 to 11) V

0.1 mV

(11 to 22) V

0.23 mV

(22 to 220) V

2.4 mV

(220 to 1 000) V

15 mV

1 µV to 100 mV

3.3 µV

(0.1 to 1) V

14 µV

(1 to 10) V

59 µV

(10 to 100) V

0.8 mV

(100 to 1 000) V

16 mV

(1 to 10) kV

360 V

(10 to 25) kV

400 V

(25 to 40) kV

1.9 kV

RF Relative Attenuation
2.5 MHz to 26.5 GHz

RF Power - Measure
(-127 to -30) dBm

Electrical – Voltage1
Calibration
Parameter/Equipment

DC Voltage - Source

DC Voltage – Measure

Form 28.8—Rev 11 – 3/1/14

Certificate # L2151
Remarks

HP 8902A Measuring
Receiver w/11793A

HP 8902A Measuring
Receiver w/11722A
Sensor
HP 438A Power Meter
w/ 8482A Sensor
w/ 8481A Sensor
w/ 8485A Sensor
w/ 8487A Sensor
HP 8903B

Remarks

Fluke 5700A/5725A
Calibrator

Agilent 3458A 8 ½
Digit Multimeter

Fluke 80K-40 w/8 ½
Digit Multimeter

Page 7 of 12

Calibration
Parameter/Equipment

Range
2.2 mV to 22 mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
22 mV to 220 mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz

AC Voltage Source

AC Voltage Source

Form 28.8—Rev 11 – 3/1/14

Expanded Uncertainty of
Measurement (+/-)
0.002 7 mV
0.001 8 mV
0.001 7 mV
0.003 1 mV
0.004 6 mV
0.007 6 mV
0.011 mV
0.015 mV
0.015 mV
0.006 3 mV
0.003 4 mV
0.011 mV
0.024 mV
0.034 mV
0.047 mV
0.08 mV

220 mV to 2.2 V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz

0.15 mV
0.061 mV
0.028 mV
0.092 mV
0.2 mV
0.28 mV
0.46 mV
0.92 mV

2.2 V to 22 V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz

0.001 5 V
0.000 46 V
0.000 22 V
0.000 36 V
0.000 72 V
0.001 2 V
0.003 1 V
0.006 1 V

22 V to 220 V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz

Certificate # L2151
Remarks

0.015 V
0.004 6 V
0.002 2 V
0.003 6 V
0.007 2 V
0.015 V
0.036 V
0.076 V

Fluke 5700A/5725A
Calibrator

Fluke 5700A/5725A
Calibrator

Page 8 of 12

Calibration
Parameter/Equipment

AC Voltage Source

Range
220 V to 750 V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
0.5 to 1) MHz
750 V
(30 to 50) kHz
(50 to 100) kHz
1100 V
50 Hz to 1 kHz
1100 V
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 30) kHz

AC Voltage –
Measurement

AC Voltage –
Measurement

Form 28.8—Rev 11 – 3/1/14

Expanded Uncertainty of
Measurement (+/-)
0.15 V
0.046 V
0.023 V
0.064 V
0.15 V
0.41 V
1.4 V
3.3 V

Certificate # L2151
Remarks

Fluke 5700A/5725A
Calibrator

0.52 V
2V

Fluke 5725A Calibrator

0.12 V

Fluke 5700A Calibrator

0.11 V
0.21 V
0.76 V

Fluke 5725A Calibrator

10 mV to 100 mV
(1 to 40) Hz
40Hz to 1kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(0.1 to 10) V
(1 to 40) Hz
40Hz to 1kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
300 kHz to 1MHz
(1 to 2) MHz

0.000 085 V
0.000 085 V
0.000 17 V
0.000 36 V
0.001 2 V
0.005 8 V
0.012 V
0.018 V

(10 to 100) V
(1 to 40) Hz
40Hz to 1kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz

0.024 V
0.024 V
0.024 V
0.041 V
0.14 V
0.46 V

(100 to 1 000) V
(1 to 40) Hz
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz

0.46 V
0.46 V
0.69 V
1.4 V
3.5 V

0.003 6 mV
0.002 5 mV
0.003 6 mV
0.012 mV
0.058 mV
0.46 mV
Agilent 3458A 8 ½
Digit Multimeter

Agilent 3458A 8 ½
Digit Multimeter

Page 9 of 12

Calibration
Parameter/Equipment

AC Voltage –
Measurement

Range

Expanded Uncertainty of
Measurement (+/-)

(2 to 20) mV
200 kHz to 1 MHz
(1 to 10) MHz
(10 to 20) MHz

1.7 mV
2.0 mV
2.5 mV

(0.02 to 20) V
200kHz to 1MHz
(1 to 10) MHz
(10 to 20) MHz

0.25 V
0.29 V
0.36 V

Thermocouple Millivolt
Simulation
Type C
Type E

(0 to 2 316) °C

0.98 ºC

(-250 to 1 000) °C

0.44 ºC

Type J

(-210 to 1 200) °C

0.31 ºC

Type K

(-200 to 1 372) °C

0.47 ºC

Type N

(-200 to 1 300) °C

0.47 ºC

Type R

(0 to 1 767) °C

0.66 ºC

Type S

(0 to 1 767) °C

0.55 ºC

Type T

(-250 to 400) °C

0.73 ºC

Type U

(-200 to 600) °C

0.65 ºC

Thermodynamic – Humidity
Calibration
Parameter/Equipment

Range

Expanded Uncertainty of
Measurement (+/-)

Relative Humidity
(23 ± 5) °C

(10 to 80) % RH

2.4 % RH

Thermodynamic – Thermodynamic Sources1
Calibration
Range
Parameter/Equipment
Temperature Measurement
(Ovens, Furnaces,
Freezers, Chambers, Ice
Baths – System Accuracy
Test)

(-45 to 200) ºC

2.2 ºC

(200 to 500) ºC

2.5 ºC

Time and Frequency – Frequency / Period
Calibration
Range
Parameter/Equipment
Frequency – Source2

Form 28.8—Rev 11 – 3/1/14

Expanded Uncertainty of
Measurement (+/-)

Certificate # L2151
Remarks

Fluke 8920A True
RMS Voltmeter

Fluke 5520A Calibrator

Remarks
Veriteq SP-2000
Logger

Remarks

Fluke 51 with Type J
TC probe

Expanded Uncertainty of
Measurement (+/-)

Remarks

10 MHz

7.6 · 10-11 · f

DC to 20 MHz

7.6 · 10-11 · f + 0.58 Hz

HP 58503A GPS
Receiver
HP 3325B Generator

10 MHz to 26.5 GHz

7.6 · 10-11 · f + 1.2 Hz

HP 8340B Generator
Page 10 of 12

Calibration
Parameter/Equipment

Certificate # L2151
Remarks

Range

Expanded Uncertainty of
Measurement (+/-)

1 Hz to 225 MHz

7.6 · 10-11 · f + 0.6 mHz

HP 53131A Counter

10 Hz to 525 MHz

7.6 · 10-11 · f + 0.58 Hz

5351B 50 Ω Counter

(0.5 to 26.5) GHz

7.6 · 10-11 · f + 0.58 Hz

5351B 50 Ω Counter

CW (0.25 to 10) MHz

0.24 kHz

HP 8902A Measuring
Receiver

CW (0.01 to 1.3) GHz

5.9 kHz

Rate (20 Hz to 10 kHz)
Depth (10 to 99) %

CW (0.15 to 10) MHz

3.6 % of reading

Rate (50 Hz to 10 kHz)
Depth (10 to 99) %

CW (0.15 to 10) MHz

2.4 % of reading

Rate (20 Hz to 100 kHz)
Depth (10 to 99) %

CW (0.01 to 1.3) GHz

3.6 % of reading

Rate (50 Hz to 50 kHz)
Depth (10 to 99) %

CW (0.01 to 1.3) GHz

1.3 % of reading

Rate (200 Hz to 10 kHz)

CW (0.15 to 10) MHz

1.3 rad

Rate (200 Hz to 20 kHz)

CW (0.01 to 1.3) GHz

8.8 rad

Frequency - Measure1,2
Frequency Modulation –
Measure
Rate (20 Hz to 10 kHz)
40 kHz Peak Deviation
Rate (20 Hz to 200 kHz)
400 kHz Peak Deviation
Amplitude Modulation –
Measure

HP 8902A Measuring
Receiver

Phase Modulation –
Measure

Time and Frequency – Oscilloscopes1
Calibration
Range
Parameter/Equipment

Expanded Uncertainty of
Measurement (+/-)

(1 to 24.999) mV

0.075 mV

Amplitude

(25 to 109.99) mV

0.17 mV

Square Wave into 1 MΩ
(Peak-to-Peak)

110 mV to 2.1999 V

0.049 V

(2.2 to 10.999) V

0.013 V

(11 to 130) V

0.15 V

(1 to 24.999) mV

0.12 mV

(25 to 109.99) mV

0.36 mV

10 mV to 2.1999 V
(2.2 to 6.6) V

0.005 8 V
0.014 V

Amplitude
Square Wave into 50 Ω
(Peak-to-Peak)
Form 28.8—Rev 11 – 3/1/14

HP 8902A Measuring
Receiver

Remarks

Fluke 5520A-SC600
Calibrator

Fluke 5520A-SC600
Calibrator
Page 11 of 12

Calibration
Parameter/Equipment
Rise / Fall-time
50 Ω load
5 mV to 2.5 V
(Peak-to-Peak)

Bandwidth
relative to 50 kHz
50 Ω load

Timebase
50 Ω Load

Range

Expanded Uncertainty of
Measurement (+/-)

1 kHz to 2 MHz

350 pS

2 MHz to 10 MHz

400 pS

(0.05 to 100) MHz

0.087 mV

(100 to 300) MHz

0.23 mV

(300 to 600) MHz

0.23 mV

(0.05 to 100) MHz

0.058 V

(100 to 300) MHz

0.058 V

(300 to 600) MHz

0.059 V

5 s to 50 mS

0.029 S

20 mS to 2 nS

0.002 9 mS

Certificate # L2151
Remarks
Fluke 5520A-SC600
Calibrator

Fluke 5520A-SC600
Calibrator

Fluke 5520A-SC600
Calibrator

Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range,
expanded uncertainty of measurement and remarks. The expanded uncertainty of measurement is expressed as the standard
uncertainty of the measurement multiplied by a coverage factor of 2 (k=2), corresponding to a confidence level of
approximately 95%.
Notes:
1) Laboratory offers calibration services at the laboratory's own facilities and at the client or other agreed upon facilities.
2) f is defined as the measured / generated frequency.

Approved by:

Date: May 31, 2016
R. Douglas Leonard
Chief Technical Officer

Re-Issued: 5/31/16
Form 28.8—Rev 11 – 3/1/14

Page 12 of 12



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : L-A-B Staff
Create Date                     : 2016:05:31 08:45:40-04:00
Modify Date                     : 2016:06:07 12:52:23-04:00
Language                        : en-US
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Format                          : application/pdf
Creator                         : L-A-B Staff
Title                           : Proposed Scope of Accreditation
Creator Tool                    : Microsoft® Word 2016
Metadata Date                   : 2016:06:07 12:52:23-04:00
Producer                        : Microsoft® Word 2016
Document ID                     : uuid:e8e7f8d6-7528-47f0-9411-97f3f82b9469
Instance ID                     : uuid:518fa864-a90c-4b0b-ab38-52d9ab36f4a2
Page Count                      : 12
EXIF Metadata provided by EXIF.tools

Navigation menu