919.1551 Vac Ion Plus 150 Pumps User Manual
User Manual: 919.1551
Open the PDF directly: View PDF .
Download | |
Open PDF In Browser | View PDF |
VacIon Plus 150 pumps Models 919-1510/1520/1540 919-1511/1521/1541 919-1512/1522/1542 919-1550/1560/1580 919-1551/1561/1581 919-1552/1562/1582 Manuale di Istruzioni Bedienungshandbuch Notice de Mode D’Emploi Instruction Manual 87-900-102-01 (E) 05/2011 Notices © Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number Publication Number: 87-900-102-01 (E) Edition Edition 05/2011 Printed in ITALY Agilent Technologies Italia S.p.A. Vacuum Products Division Via F.lli Varian, 54 10040 Leinì (TO) ITALY Warranty The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and nonDOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. Trademarks Windows and MS Windows are U.S. registered trademarks of Microsoft Corporation. Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) VacIon Plus 150 VacIon Plus 150 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 3/88 VacIon Plus 150 4/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Contents Contents 1 Procedura per l’installazione Informazioni Generali 9 10 Preparazione per l’installazione 11 Installazione 12 Uso 14 Procedure di uso 15 Manutenzione 17 Smaltimento 2 17 Anleitung zur Installation Allgemeine Hinweise 19 20 Vor der Installation 21 Installation 22 Gebrauch 24 Bedienungsschritte 25 Wartung 27 Entsorgung 27 3 Procédure pour l’installation Indications Generales 29 30 Préparation de l'installation 31 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5/88 Contents Installation 32 Utilisation 34 Procédure d’utilisation 35 4 Maintenance 37 Mise au rebut 37 Installation Procedure 39 General Information 40 Preparation for Installation 41 Installation 42 Use 44 Operating Procedure 45 Maintenance Disposal 5 47 47 Technical Information 49 VacIon Plus Pump Controllers 50 Description of the VacIon Pump 51 Technical Specification 53 Vaclon Plus 150 Starcell 55 Vaclon Plus 150 Diode 56 Vaclon Plus 150 Noble Diode 57 Outline Drawing 58 Stray Magnetic Field 59 6/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Contents VacIon Plus Pump Installation 62 Typical Installation 65 Inlet Flange Connection 66 Control Unit Connection 69 Safety Interlock 70 Heater Installation 70 Heater Replacement 72 Bakeout Operation 72 Operating Procedure 74 Maintenance 75 Pump Troubleshooting 79 VacIon Plus Pump Replacement Parts and Accessories VacIon Plus Pump Controllers 80 81 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 7/88 Contents 8/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) VacIon Plus 150 pumps Instruction Manual 1 Procedura per l’installazione Informazioni Generali 10 Preparazione per l’installazione 11 Installazione 12 Uso 14 Procedure di uso 15 Manutenzione 17 Smaltimento 17 Traduzione delle istruzioni originali 9/88 1 Procedura per l’installazione Informazioni Generali Informazioni Generali Questa apparecchiatura è destinata ad uso professionale. L'utilizzatore deve leggere attentamente il presente manuale di istruzioni ed ogni altra informazione addizionale fornita dalla Agilent prima dell'utilizzo dell'apparecchiatura. La Agilent si ritiene sollevata da eventuali responsabilità dovute all'inosservanza totale o parziale delle istruzioni, ad uso improprio da parte di personale non addestrato, ad interventi non autorizzati o ad uso contrario alle normative nazionali specifiche. Le pompe della serie VacIon Plus sono pompe ioniche utilizzate comunemente per applicazioni di ultra alto vuoto, grazie alla loro pulizia, capacità di pompare qualsiasi tipo di gas, e del loro funzionamento senza vibrazioni e necessità di manutenzione. Nei paragrafi seguenti sono riportate tutte le informazioni necessarie a garantire la sicurezza dell'operatore durante l'utilizzo dell'apparecchiatura. Informazioni dettagliate sono fornite nell'appendice “Technical information”. Questo manuale utilizza le seguenti convenzioni: AVVERTENZA! I messaggi di avvertenza attirano l'attenzione dell'operatore su una procedura o una pratica specifica che, se non eseguita in modo corretto, potrebbe provocare gravi lesioni personali. ATTENZIONE! NOTA 10/88 I messaggi di attenzione sono visualizzati prima di procedure che, se non osservate, potrebbero causare danni all'apparecchiatura. Le note contengono informazioni importanti estrapolate dal testo. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 1 Procedura per l’installazione Preparazione per l’installazione Preparazione per l’installazione La pompa viene fornita in un imballo protettivo speciale; se si presentano segni di danni, che potrebbero essersi verificati durante il trasporto, contattare l'ufficio vendite locale. Durante l'operazione di disimballaggio, prestare particolare attenzione a non lasciar cadere la pompa e a non sottoporla ad urti o vibrazioni. Non disperdere l'imballo nell'ambiente. Il materiale è completamente riciclabile e risponde alla direttiva CEE 85/399 per la tutela dell'ambiente. ATTENZIONE! Onde evitare problemi di degassamento, non toccare con le mani nude i componenti destinati ad essere esposti al vuoto. Utilizzare sempre i guanti o altra protezione adeguata. NOTA La pompa non può essere danneggiata rimanendo semplicemente esposta all'atmosfera. Si consiglia comunque di mantenerla chiusa fino al momento dell'installazione sul sistema onde evitare eventuale inquinamento da polvere. Figura 1 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 11/88 1 Procedura per l’installazione Installazione Installazione AVVERTENZA! La pompa, a causa del suo peso, deve essere maneggiata tramite appositi attrezzi di sollevamento e spostamento. Non installare e/o utilizzare la pompa in ambienti esposti ad agenti atmosferici (pioggia, gelo, neve), polveri, gas aggressivi, in ambienti esplosivi o con elevato rischio di incendio. Durante il funzionamento, per ottenere le specifiche tecniche dichiarate, la temperatura ambiente deve essere compresa tra 0 °C e +85 °C. ATTENZIONE! La pompa deve essere tenuta sigillata con il suo tubo di ingresso schiacciato finchè non è pronta per essere collegata al sistema. AVVERTENZA! NO! Per evitare lesioni alla persona, non collegare l’alta tensione alla pompa prima che sia installata nel sistema e che tutte le flange di ingresso siano adeguatamente collegate o chiuse. 12/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 1 Procedura per l’installazione Installazione Il funzionamento delle pompe è ottimizzato solo con l’uso delle apposite unità di controllo Agilent (Dual, MidiVac o MiniVac). ATTENZIONE! Il rispetto delle normative di sicurezza nell’uso delle pompe è garantito solo con l’uso delle unità di controllo Agilent. La pompa VacIon Plus può essere installata in qualsiasi posizione. Per convenienza normalmente viene montata in posizione verticale con la flangia di ingresso in alto. Per installazioni con posizione diversa dalla verticale, la pompa deve essere sostenuta tramite i piedini di montaggio posti sul suo fondo. Per informazioni dettagliate sull’installazione della pompa, vedere l’appendice "Technical Information". VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 13/88 1 Procedura per l’installazione Uso Uso Tutte le istruzioni per il corretto funzionamento delle pompe VacIon Plus sono contenute nel manuale dell'unità di controllo. Leggere attentamente tale manuale prima dell'utilizzo. Si raccomanda di portare la pompa ad una pressione di 1x10-4 Torr (mbar) in modo da ottenere un avvio più rapido. A questo scopo è meglio non utilizzare una pompa meccanica sigillata con olio, ma se deve essere utilizzata una tale pompa, si raccomanda di inserire una trappola lungo la linea di vuoto per ridurre la pressione causata dai vapori di acqua e di olio provenienti dalla pompa meccanica. Cercare di ridurre al minimo il tempo in cui la pompa meccanica è aperta verso il sistema e verso la pompa ionica, poiché i suoi vapori si diffondono nel sistema a pressioni inferiori a 1x10-1 Torr (mbar) causando una contaminazione dello stesso. Nei sistemi in cui l’olio deve essere completamente assente, è meglio utilizzare pompe Turbo per la suddetta operazione. Depositi igroscopici e l’assorbimento dell’idrogeno all’interno del composto di titanio possono provocare l’allungamento del tempo di avvio con l’invecchiamento della pompa. Durante l’esposizione all’aria, il deposito del composto di titanio assorbe vapore acqueo; in conseguenza di ciò, al successivo avvio, il riscaldamento della pompa provoca il rilascio del vapore e di parte dell’idrogeno pompato precedentemente, con l’effetto dell’allungamento del tempo di avvio. 14/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 1 Procedura per l’installazione Procedure di uso Procedure di uso Controllare che la polarità dell’unità di controllo sia corretta per la pompa: polarità positiva per le pompe Diode e negativa per le StarCell e le Triode. Fare riferimento al relativo manuale ed osservare la seguente procedura per l’uso della pompa: 1 Tramite una pompa di pre-vuoto portare il sistema da vuoto ad una pressione minima di avvio in accordo alla tabella di pagina 9 ed a seconda del tipo di pompa ionica. 2 Collegare l’unità di controllo ad una apposita fonte di alimentazione ed accenderla. 3 Osservare la tensione, la corrente e la pressione. Se l’avvio è avvenuto ad una pressione di 5x10-2 Torr (mbar), è tipica una tensione di circa 300 - 400 V. Un valore di corrente prossimo alla corrente di corto circuito dell’unità di controllo è indice dell’esistenza di una perdita nella pompa e nel sistema. Un temporaneo incremento della pressione di pre-vuoto è normale durante la fase di avvio. 4 Lasciare la valvola di pre-vuoto aperta durante l’avvio della pompa ionica finchè non si raggiunge una adeguata pressione di avvio. Se la tensione della pompa ionica scende dopo la chiusura della valvola, riaprirla per un pre-pompaggio aggiuntivo. Appena la pressione diminuisce, la tensione cresce nuovamente e la valvola di pre-vuoto può essere chiusa. 5 Quando la tensione ha raggiunto i 2 - 3kV, mettere l’unità di controllo nello stato PROTECT. Il sistema sarà così autoprotetto dall’incremento della pressione oltre 1x10-4 Torr (mbar) quando la pompa non è controllata. Nel caso in cui ci fosse un tale incremento, l’unità di controllo viene spenta automaticamente. 6 La pressione nella pompa può anche essere calcolata leggendo la corrente e convertendo la lettura in pressione tramite l’apposito diagramma pressione-corrente illustrato nell’appendice “Technical Information” di questo manuale. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 15/88 1 Procedura per l’installazione Procedure di uso NOTA I gradini nei diagrammi pressione - corrente sono una caratteristica del modo di funzionamento del Dual. Quando la corrente assorbita dalla pompa raggiunge determinati valori, l'unità di controllo cambia il valore dell'alta tensione in uscita ad un valore più basso. 7 Quando si porta la pompa alla pressione atmosferica, usare azoto secco in modo da evitare l’assorbimento di vapore acqueo da parte delle pareti della pompa. AVVERTENZA! Quando la pompa viene utilizzata per il pompaggio di gas tossici, infiammabili o radioattivi, seguire le appropriate procedure tipiche di ciascun gas. Non usare la pompa in presenza di gas esplosivi. AVVERTENZA! Quando è installato il riscaldatore, non toccare la pompa durante le operazioni di riscaldamento e di raffreddamento. L'elevata temperatura può causare lesioni alle persone. ATTENZIONE! Non avvicinare dispositivi elettronici alla pompa. Il campo magnetico attorno ad essa può provocare dei malfunzionamenti dei dispositivi stessi. 16/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 1 Procedura per l’installazione Manutenzione Manutenzione Le pompe della serie VacIon Plus non richiedono alcuna manutenzione. Qualsiasi intervento deve essere eseguito da personale autorizzato. AVVERTENZA! Prima di effettuare qualsiasi intervento sulla pompa scollegarla dall’alta tensione. Qualora una pompa dovesse essere rottamata, procedere alla sua eliminazione nel rispetto delle normative nazionali specifiche. Smaltimento Significato del logo "WEEE" presente sulle etichette. Il simbolo qui sotto riportato applicato in ottemperanza alla direttiva CE denominata "WEEE". Questo simbolo (valido solo per i paesi della Comunità Europea) indica che il prodotto sul quale è applicato, NON deve essere smaltito insieme ai comuni rifiuti domestici o industriali, ma deve essere avviato ad un sistema di raccolta differenziata. Si invita pertanto l'utente finale a contattare il fornitore del dispositivo, sia esso la casa madre o un rivenditore, per avviare il processo di raccolta e smaltimento, dopo opportuna verifica dei termini e condizioni contrattuali di vendita. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 17/88 1 Procedura per l’installazione Smaltimento 18/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) VacIon Plus 150 pumps Instruction Manual 2 Anleitung zur Installation Allgemeine Hinweise 20 Vor der Installation 21 Installation 22 Gebrauch 24 Bedienungsschritte 25 Wartung 27 Entsorgung 27 Übersetzung der Originalanleitungen 19/88 2 Anleitung zur Installation Allgemeine Hinweise Allgemeine Hinweise Dieses Gerät ist für den professionellen Gebrauch bestimmt. Vor dem Gebrauch soll der Benutzer dieses Handbuch sowie alle weiteren von Agilent mitgelieferten Zusatzdokumentationen genau lesen. Bei vollständiger bzw. teilweiser Nichtbeachtung der enthaltenen Hinweise, unsachgemäßem Gebrauch durch ungeschultes Personal, nicht autorisierten Eingriffen und Mißachtung der nationalen Bestimmungen übernimmt Firma Agilent keinerlei Haftung. Die Pumpen der Serie VacIon Plus sind Ionenpumpen, die aufgrund ihrer Reinheit, ihrer Fähigkeit, alle Arten von Gas zu pumpen, und ihres vibrations- und wartungsfreien Betriebes, allgemein für Ultrahochvakuumanwendungen zum Einsatz kommen. In den folgenden Abschnitten sind alle erforderlichen Informationen für die Sicherheit des Bedieners bei der Anwendung des Geräts aufgeführt. Detaillierte technische Informationen sind im Anhang "Technical Information" enthalten. In dieser Gebrauchsanleitung werden Sicherheitshinweise folgendermaßen hervorgehoben: WARNUNG! Die Warnhinweise lenken die Aufmerksamkeit des Bedieners auf eine spezielle Prozedur oder Praktik, die bei unkorrekter Ausführung schwere Verletzungen hervorrufen könnte. VORSICHT! Die Vorsichtshinweise vor bestimmten Prozeduren machen den Bediener darauf aufmerksam, daß bei Nichteinhaltung Schäden am Gerät entstehen können. HINWEIS Die Hinweise enthalten wichtige Informationen, die aus dem Text hervorgehoben werden. 20/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 2 Anleitung zur Installation Vor der Installation Vor der Installation Die Pumpe wird in einer speziellen Schutzverpackung geliefert. Eventuelle Transportschäden sind der zuständigen örtlichen Verkaufsstelle zu melden. Beim Auspacken vorsichtig vorgehen, damit die Pumpe nicht fällt oder Stößen oder Vibrationen ausgesetzt wird. Das Verpackungsmaterial ist ordnungsgemäß zu entsorgen. Es ist vollständig recyclebar und entspricht der EG-Richtlinie 85/399 für den Umweltschutz. VORSICHT! HINWEIS Um Entgasungsprobleme zu vermeiden, dürfen die Komponenten, die mit dem Vakuum in Kontakt kommen, nicht mit bloßen Händen berührt werden. Stets Handschuhe oder einen anderen geeigneten Schutz tragen. Die Pumpe kann, wenn sie einfach der Atmosphäre ausgesetzt ist, nicht beschädigt werden. Sie sollte jedoch bis zur Installation an der Anlage geschlossen bleiben, um Verunreinigungen durch Staub zu vermeiden. Abbildung 1 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 21/88 2 Anleitung zur Installation Installation Installation WARNUNG! Die Pumpe kann aufgrund ihres Gewichtes mittels geeigneter Hebe- und Förderzeuge gehandhabt werden. Die Pumpe darf nicht in Umgebungen installiert und/oder benutzt werden, die ungeschützt vor Witterungsbedingungen (Regen, Frost, Schnee), Staub und aggressiven Gasen sind, und in denen Explosions- und erhöhte Brandgefahr besteht. Während des Betriebes soll die Umgebungstemperatur zwischen 0 °C und +85 °C betragen, um die angegebenen technischen Merkmale zu gewährleisten. VORSICHT! WARNUNG! Die Pumpe soll versiegelt und mit flachgedrücktem Eintrittsschlauch gehalten werden, bis sie für den Anschluß an das System bereit ist. FALSCH! Um Personenschäden zu vermeiden, darf die Hochspannungsleitung der Pumpe erst angeschlossen werden, wenn die Pumpe im System installiert ist und alle Eintrittsflansche entsprechend angeschlossen oder geschlossen sind. 22/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 2 Anleitung zur Installation Installation Der Pumpenbetrieb wird nur durch den Einsatz speziell dafür vorgesehener Agilent Steuereinheiten (Dual, MidiVac oder MiniVac) optimiert. VORSICHT! Die Einhaltung der Sicherheitsvorschriften beim Gebrauch der Pumpen ist nur bei Verwendung von Agilent Steuereinheiten gewährleistet. Die Pumpe Vaclon Plus kann in einer beliebigen Position installiert werden. Aus praktischen Gründen wird sie in der Regel senkrecht mit nach oben gerichtetem Eintrittsflansch montiert. Bei nicht senkrechten Einbaulagen soll die Pumpe auf Montagefüßen aufliegen, die an ihrer Unterseite angebracht werden. Detaillierte Informationen zur Installation der Pumpe sind dem Anhang “Technical Information” zu entnehmen. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 23/88 2 Anleitung zur Installation Gebrauch Gebrauch Sämtliche Hinweise für den korrekten Betrieb der Pumpen VacIon Plus sind im Handbuch der Steuereinheit enthalten. Dieses Handbuch ist vor der Inbetriebnahme genau durchzulesen. Es wird empfohlen, die Pumpe auf einen Druck von 1x10-4 Torr (mbar) zu bringen, um einen rascheren Anlauf zu gewährleisten. Zu diesem Zweck empfiehlt es sich, keine ölversiegelte mechanische Pumpe zu verwenden. Falls hingegen eine derartige Pumpe benutzt werden muß, soll an der Vakuumleitung eine Falle eingesetzt werden, um den durch die Wasser- und Öldämpfe aus der mechanischen Pumpe erzeugten Druck zu verringern. Es ist zu versuchen, die Zeit auf ein Mindestmaß zu reduzieren, während der die mechanische Pumpe zum System und zur Ionenpumpe geöffnet ist, da sich ihre Dämpfe bei Drücken unter 1x10-1 Torr (mbar) im System verbreiten und dessen Verunreinigung verursachen. Bei Systemen, die vollkommen ölfrei sein müssen, erweisen sich für den obengenannten Vorgang Turbopumpen als geeigneter. Hygroskopische Ablagerungen und die Absorption von Wasserstoff in der Titanverbindung können die Anlaufzeit verlängern und eine kürzere Standzeit der Pumpe verursachen. Die Ablagerungen der Titanverbindungen absorbieren Wasserdampf, wenn sie der Luft ausgesetzt werden. Dadurch bewirkt beim anschließenden Anlaufvorgang die Aufheizung der Pumpe die Abgabe des Dampfes und eines Teils des zuvor gepumpten Wasserstoffs, so daß sich die Anlaufzeit verlängert. 24/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Anleitung zur Installation Bedienungsschritte 2 Bedienungsschritte Es ist zu kontrollieren, daß die Steuereinheit in bezug auf die Pumpe richtig gepolt ist: positive Polarität für Dioden Pumpen und negative für StarCell oder Triode Pumpen. Es ist nach dem diesbezüglichen Handbuch vorzugehen, für den Gebrauch der Pumpe sind die folgenden Bedienungsschritte zu beachten: 1 Mit einer Vorvakuumpumpe das System auf einen Mindestanlaufdruck gemäß Tabelle 9 in Abhängigkeit vom Ionenpumpentyp bringen. 2 Die Steuereinheit an eine entsprechende Versorgungsquelle anschließen und einschalten. 3 Die Spannung, die Stromstärke und den Druck beobachten. Wenn der Anlauf bei einem Druck von 5x10-2 Torr (mbar) erfolgt ist, ist eine Spannung von ca. 300 – 400 V typisch. Ein Stromwert, der sich dem Kurzschlußstromwert der Steuereinheit annähert, weist auf eine Leckstelle an der Pumpe und am System hin. Ein zeitweiliger Anstieg des Vorvakuumdruckes ist während der Anlaufphase normal. 4 Während des Anlaufes der Ionenpumpe soll das Vorvakuumventil geöffnet bleiben, bis ein angemessener Anlaufdruck erreicht ist. Wenn die Spannung der Ionenpumpe nach Schließung des Ventils abfällt, ist das Ventil für eine zusätzliche Vorvakuumpumpung zu öffnen. Sobald der Druck sinkt, steigt die Spannung erneut an und kann das Vorvakuumventil geschlossen werden. 5 Wenn die Spannung 2 – 3 kV erreicht hat, ist die Steuereinheit auf den Status PROTECT zu schalten. Das System verfügt auf diese Weise über einen Selbstschutz vor einem Druckanstieg über 1x10-4 Torr (mbar), wenn die Pumpe nicht kontrolliert ist. Falls ein solcher Anstieg stattfindet, wird die Steuereinheit automatisch ausgeschaltet. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 25/88 2 Anleitung zur Installation Bedienungsschritte 6 HINWEIS Der Pumpendruck kann auch berechnet werden, indem die Stromstärke abgelesen wird und der abgelesene Wert mittels des Druck-Stromstärke-Diagramms im Anhang “Technical Information” des vorliegenden Handbuches in einen Druckwert umgerechnet wird. Die Stufen im Druck-Stromstärke-Diagramm sind charakteris-tisch für die Funktionsweise des Dual. Wenn die Stromauf-nahme der Pumpe ein bestimmtes Ausmaß erreicht, ändert die Steuereinheit die ausgangsseitige Hochspannung auf einen niedrigeren Wert ab. 7 Wenn die Pumpe auf den atmosphärischen Druck gebracht wird, ist trockener Stickstoff zu verwenden, um die Aufnahme von Wasserdampf durch die Pumpenwände zu verhindern. WARNUNG! Wenn die Pumpe zur Förderung von giftigen, leicht entflammbaren oder radioaktiven Gasen benutzt wird, sind die für das jeweilige Gas vorgeschriebenen Vorgänge zu befolgen. Die Pumpe nie bei Vorhandensein von explosivem Gas benutzen. WARNUNG! Wenn die Heizung installiert ist, darf die Pumpe während der Aufheizung und Abkühlung nicht berührt werden. Die hohe Temperatur kann zu Personenschäden führen. VORSICHT! Keine elektronischen Geräte in die Nähe der Pumpe bringen. Das darum befindliche Magnetfeld kann zu Funktionsstörungen der Geräte führen. 26/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 2 Anleitung zur Installation Wartung Wartung Die Pumpen der Serie VacIon Plus erfordern keine Wartung. Sämtliche Eingriffe dürfen nur von autorisiertem Personal vorgenommen werden. WARNUNG! Vor Eingriffen an der Pumpe ist diese von der Hochspannungsquelle zu trennen. Bei eventueller Verschrottung einer Pumpe ist diese entsprechend der einschlägigen nationalen Vorschriften zu entsorgen. Entsorgung Bedeutung des "WEEE" Logos auf den Etiketten. Das folgende Symbol ist in Übereinstimmung mit der EU-Richtlinie WEEE (Waste Electrical and Electronic Equipment) angebracht. Dieses Symbol (nur in den EU-Ländern gültig) zeigt an, dass das betreffende Produkt nicht zusammen mit Haushaltsmüll entsorgt werden darf sondern einem speziellen Sammelsystem zugeführt werden muss. Der Endabnehmer sollte daher den Lieferanten des Geräts - d.h. die Muttergesellschaft oder den Wiederverkäufer - kontaktieren, um den Entsorgungsprozess zu starten, nachdem er die Verkaufsbedingungen geprüft hat. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 27/88 2 Anleitung zur Installation Entsorgung 28/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) VacIon Plus 150 pumps Instruction Manual 3 Procédure pour l’installation Indications Generales 30 Préparation de l'installation31 Installation 32 Utilisation 34 Procédure d’utilisation 35 Maintenance 37 Mise au rebut 37 Traduction de la mode d’emploi originale 29/88 3 Procédure pour l’installation Indications Generales Indications Generales Cet appareillage a été conçu en vue d’une utilisation professionnelle. Il est conseillé à l’utilisateur de lire attentivement cette notice d’instructions ainsi que toute autre indication supplémentaire fournie par Agilent avant d’utiliser l’appareil. Agilent décline toute responsabilité en cas de non respect total ou partiel des instructions fournies, d’opérations non autorisées, d’utilisation impropre par du personnel non formé ou contraires aux réglementations nationales spécifiques. Grâce à leur propreté, à leur capacité de pomper tous les types de gaz, à leur fonctionnement sans vibrations et à l’absence d’entretien, les pompes de la série VacIon Plus sont des pompes ioniques généralement utilisées pour des applications de vide ultra poussé. Les paragraphes suivants fournissent toutes les indications nécessaires à garantir la sécurité de l'opérateur pendant l'utilisation de l'appareillage. Des renseignements plus détaillés se trouvent dans l'appendice "Technical Information". Cette notice utilise les signes conventionnels suivants: AVERTISSEMENT! Les messages d’avertissement attirent l'attention de l'opérateur sur une procédure ou une manœuvre spéciale dont la mauvaise exécution risque de provoquer de graves lésions. ATTENTION! NOTE 30/88 Les messages d'attention apparaissent avant certaines procédures dont le nonrespect peut endommager sérieusement l'appareillage. Les notes contiennent des renseignements importants, extrapolés du texte. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Procédure pour l’installation Préparation de l'installation 3 Préparation de l'installation La pompe est fournie dans un emballage de protection spécial; si l'on constate des signes d’endommagement imputables au transport, contacter aussitôt le revendeur local. Pendant l'opération de déballage, veiller tout particulièrement à ne pas laisser tomber la pompe et à ne lui faire subir aucun choc ni aucune vibration. Ne pas abandonner l'emballage dans la nature. Le matériel est entièrement recyclable et il est conforme à la directive CEE 85/399 en matière de protection de l'environnement. ATTENTION! En vue d'éviter tout problème de dégazage, ne pas toucher à mains nues les éléments devant être exposés au vide. Mettre toujours des gants ou toute autre protection appropriée. NOTE La pompe ne peut être endommagée si elle reste simplement exposée à l'atmosphère. Il est quoi qu’il en soit conseillé de ne pas la retirer de son emballage avant le moment de l'installation, afin d'éviter toute pollution due à la poussière. Figure 1 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 31/88 3 Procédure pour l’installation Installation Installation AVERTISSEMENT! À cause de son poids, il est indispensable de se servir des équipements de levage et de manutention pour déplacer la pompe. Ne pas installer et/ou utiliser la pompe dans des milieux exposés à des agents atmosphériques (pluie, gel, neige), à des poussières, à des gaz de combat ainsi que dans des milieux explosifs ou à fort risque d'incendie. Pendant le fonctionnement, pour respecter les spécifications techniques déclarées la température ambiante doit être comprise entre 0 °C et +85 °C; ATTENTION! AVERTISSEMENT! La pompe doit être conservée scellée avec son tuyan d’entrée aplati jusqu’à ce qu’elle soit prête à être branchée au système. NON! Pour éviter toute lésion aux personnes, ne pas brancher la haute tension à la pompe avant que celle-ci soit installée dans le système et avant que toutes les brides d’entrées soient correctement assemblées ou fermées. Le fonctionnement de la pompe n'est optimisé que si celle-ci est utilisée avec l'une des unités de contrôle Agilent spécifiques (Dual, MidiVac ou MiniVac). 32/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Procédure pour l’installation Installation ATTENTION! 3 Lors de l'utilisation des pompes, le respect des normes de sécurité est impérativement subordonné à l'emploi des unités de contrôle Agilent. La pompe VacIon Plus peut être installée dans toutes les positions. Par facilité elle est généralement montée en position verticale avec la bride d'entrée en partie haute. Pour toute installation différente de la position verticale, la pompe doit reposer sur ses pieds de montage placé sur le fond. Pour plus de détails sur l’installation de la pompe, consulter l’appendice “Technical Information”. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 33/88 3 Procédure pour l’installation Utilisation Utilisation Toutes les instructions pour le fonctionnement correct de la pompe VacIon Plus sont fournies dans la notice de l'unité de contrôle. Il est conseillé de lire attentivement cette notice avant d'utiliser la pompe. Il est recommandé de porter la pompe à une pression de 1x10-4 Torrs (mbars) de façon à obtenir un démarrage plus rapide. Pour ce faire il est préférable de ne pas utiliser une pompe mécanique scellée à l’huile mais au cas où il serait nécessaire d’utiliser une telle pompe, il est recommandé d’introduire un dispositif de retenue le long de la ligne de vide afin de réduire la pression due aux vapeurs d’eau et d’huile engendrées par la pompe mécanique. Essayer de réduire au minimum le temps où la pompe mécanique est ouverte vers le système et vers la pompe ionique car ses vapeurs se répandent dans le système à des pressions inférieures à 1x10-1 Torrs (mbars) causant une contamination de celui-ci. Dans les systèmes où l’huile doit être complètement absente, il est préférable d’effectuer ladite opération à l’aide d’une pompe Turbo. Des dépôts hygroscopiques et l’absorption d’hydrogène dans le composé de titane peuvent provoquer, par effet du vieillissement de la pompe, l’allongement du temps de démarrage. Pendant l’exposition à l’air, le dépôt du composé de titane absorbe de la vapeur d’eau et cette action a pour conséquence qu’au démarrage suivant le chauffage de la pompe entraînera la dispersion de la vapeur et d’une partie de l’hydrogène pompée précédemment ce qui aura pour effet d’allonger le temps de démarrage. 34/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 3 Procédure pour l’installation Procédure d’utilisation Procédure d’utilisation Contrôler que la polarité de l’unité de contrôle soit correcte pour la pompe: polarité positive pour les pompes Diode et négative pour les pompes StarCell et Triode. Se reporter au manuel correspondant et observer la procédure suivante pour l’utilisation de la pompe: 1 A l'aide d'une pompe de pré-vide, porter le système à une pression minimale de démarrage en suivant le tableau de la page 9 et en fonction du type de pompe ionique. 2 Brancher l’unité de contrôle à une source d’alimentation appropriée et l’allumer. 3 Contrôler la tension, le courant et la pression. Lorsque la mise en route est effectuée à une pression de 5x10-2 Torrs (mbars) la tension est généralement d’environ 300 - 400 V. Une valeur de courant proche du courant de court-circuit de l’unité de contrôle révèle l’existence d’une fuite dans la pompe et dans le système. Au cours de la phase de démarrage, un accroissement momentané de la pression de pré-vide est normal. 4 Laisser la soupape de pré-vide ouverte pendant le démarrage de la pompe ionique tant qu’une pression de démarrage appropriée n’a pas été atteinte. Si la tension de la pompe ionique descend après la fermeture de la soupape, la rouvrir pour un pré-pompage supplémentaire. Dès que la pression commence à diminuer, la tension augmente à nouveau et la valve de pré-vide peut être fermée. 5 Lorsque la tension a atteint 2 - 3 kV, mettre l’unité de contrôle en état de PROTECT. Le système sera ainsi protégé contre toute augmentation de pression au-delà de 1x10-4 Torrs (mbars) lorsque la pompe n’est pas contrôlée. Si une telle augmentation devait être enregistrée, l’unité de contrôle s’éteindrait automatiquement. 6 La pression de la pompe peut également être calculée en relevant le courant et en convertissant la lecture en pression à l’aide du diagramme pression-courant illustré dans l’appendice “Technical Information” de ce manuel. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 35/88 3 Procédure pour l’installation Procédure d’utilisation NOTE Les paliers du diagramme de pression – courant sont une caractéristique du mode de fonctionnement du Dual. Lorsque le courant absorbé par la pompe atteint des valeurs déterminées, l'unité de contrôle porte la valeur de haute tension en sortie à une valeur inférieure. 7 Lorsque l’on porte la pompe à la pression atmosphérique, utiliser de l’azote sec de façon à éviter que les parois de la pompe n’absorbent de la vapeur aqueuse. AVERTISSEMENT! Lorsque la pompe est utilisée pour le pompage de gaz toxiques, inflammables ou radioactifs, suivre les procédures appropriées à chaque gaz. Ne pas utiliser la pompe en présence de gaz explosifs. AVERTISSEMENT! Lorsque le réchauffeur est installé, éviter de toucher la pompe pendant les opérations de chauffage et de refroidissement. La température élevée peut provoquer des brûlures. ATTENTION! 36/88 Ne pas approcher de dispositifs électroniques de la pompe. Le champ magnétique environnant cette dernière peut entraîner des dysfonctionnements desdits dispositifs. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 3 Procédure pour l’installation Maintenance Maintenance Les pompes de la série VacIon Plus ne demandent aucun entretien. Toute intervention doit être exécutée par un personnel agréé. AVERTISSEMENT! Avant toute intervention sur la pompe, la débrancher de la haute tension. En cas de mise au rebut d’une pompe, procéder à son élimination dans le respect des normes nationales en vigueur. Mise au rebut Signification du logo "WEEE" imprimé sur les étiquettes. Le symbole ci-dessous est appliqué conformément à la directive CE nommée "WEEE". Ce symbole (uniquement valide pour les pays de la Communauté européenne) indique que le produit sur lequel il est appliqué NE doit PAS être mis au rebut avec les ordures ménagères ou les déchets industriels ordinaires, mais passer par un système de collecte sélective. Après avoir vérifié les termes et conditions du contrat de vente, l’utilisateur final est donc prié de contacter le fournisseur du dispositif, maison mère ou revendeur, pour mettre en œuvre le processus de collecte et mise au rebut. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 37/88 3 Procédure pour l’installation Mise au rebut 38/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) VacIon Plus 150 pumps Instruction Manual 4 Installation Procedure General Information 40 Preparation for Installation 41 Installation 42 Use 44 Operating Procedure 45 Maintenance 47 Disposal 47 Original Instructions 39/88 4 Installation Procedure General Information General Information This equipment is destined for use by professionals. The user should read this instruction manual and any other additional information supplied by Agilent before operating the equipment. Agilent will not be held responsible for any events occurring due to non-compliance, even partial, with these instructions, improper use by untrained persons, non-authorized interference with the equipment or any action contrary to that provided for by specific national standards. The VacIon Plus series pumps are ion pumps commonly used to create ultra-high vacuum, due to their cleanliness, ability to pump different gases, and maintenance- and vibration-free operation. The following paragraphs contain all the information necessary to guarantee the safety of the operator when using the equipment. Detailed information is supplied in the appendix "Technical Information". This manual uses the following standard protocol: WARNING! CAUTION! NOTE 40/88 The warning messages are for attracting the attention of the operator to a particular procedure or practice which, if not followed correctly, could lead to serious injury. The caution messages are displayed before procedures which, if not followed, could cause damage to the equipment. The notes contain important information taken from the text. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 4 Installation Procedure Preparation for Installation Preparation for Installation The pump is supplied in a special protective packing. If this shows signs of damage which may have occurred during transport, contact your local sales office. When unpacking the pump, be sure not to drop it and avoid any kind of sudden impact or shock vibration to it. Do not dispose of the packing materials in an unauthorized manner. The material is 100 % recyclable and complies with EEC Directive 85/399. CAUTION! NOTE In order to prevent outgassing problems, do not use bare hands to handle components which will be exposed to vacuum. Always use gloves or other appropriate protection. Normal exposure to the environment cannot damage the pump. Nevertheless, it is advisable to keep it closed until it is installed in the system, thus preventing any form of pollution by dust. Figure 1 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 41/88 4 Installation Procedure Installation Installation WARNING! The pump is heavy and must be handled by means of suitable moving and handling tools. Do not install or use the pump in an environment exposed to atmospheric agents (rain, snow, ice), dust, aggressive gases, or in explosive environments or those with a high fire risk. During operation, to obtain the declared functioning specification, the ambient temperature must be between 0 °C and +85 °C. CAUTION! WARNING! The pump should be kept sealed with its pinch-off tube until it is ready for attachment to the vacuum system. NO! To avoid injury, never connect the high voltage to the pump before it is installed into the system and all the inlet flanges are properly connected or blanked off. The pump operation is optimized using one of the special Agilent controllers (Dual, MidiVac or MiniVac) only. 42/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 4 Installation Procedure Installation CAUTION! The safety specifications agreement using the pump is guaranteed using the Agilent controller only. The VacIon Plus pump can be installed in any position. For convenience, a pump is usually mounted vertically with the inlet up. For mounting other than vertical, the pump should be supported using the mounting feet at the bottom of the pump. For detailed information about the pump installation, see the appendix “Technical Information”. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 43/88 4 Installation Procedure Use Use All the instructions for the correct use of the VacIon Plus pumps are contained in the control unit manual. Read the manual carefully before using the pump. Rough pumping down to 1x10-4 Torr (mbar) is recommended for the most rapid starting. Roughing with an oilsealed mechanical pump is not desirable, but when used, a trap in the roughing line is recommended to reduce pressure due to water vapor and oils from the mechanical pump. Be careful to minimize the time that the mechanical pump is open to the system and ion pump, since mechanical pump vapors will start diffusing into the system at pressures below 1x10-1 Torr (mbar) and cause contamination. In systems where oils must be completely eliminated, turbopump roughing pumps should be used. Hygroscopic deposits and hydrogen absorption into titanium may cause starting times to increase with age. During exposure to air, the deposits of titanium compound absorb water vapor. In subsequent start ups, pump heating causes release of the water vapor and some previously pumped hydrogen; thus, the starting time may be lengthened. 44/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 4 Installation Procedure Operating Procedure Operating Procedure Check that the controller HV polarity is correct for the pump: positive polarity for Diode pumps and negative for StarCell or Triode pumps. Refer to the relevant pump control unit instruction manual and follow the procedure below when operating the pump: 1 With a clean roughing pump, establish a minimum starting pressure in the vacuum system per the table on page 9 for the type of ion pump. 2 Plug the control unit into a suitable power source and switch the power ON. 3 Observe the voltage, current, and roughing pressure. If started at 5x10-2 Torr (mbar), a controller voltage of approximately 300 to 400 volts is typical. A current value near the short-circuit current of the control unit could indicate that an unconfined flow discharge exists in the pump and system. A temporary rise in roughing pressure will usually be noticed during any starting procedure. 4 Allow the roughing valve to remain open after turning on the ion pump until an adequate starting pressure is reached. If the ion pump voltage drops after closing the roughing valves, reopen the valve for additional rough pumping. As the pressure decreases, the voltage again will rise, and the roughing valve may be closed. 5 When the voltage has increased to 2-3 kV, place the control unit in the PROTECT condition. The system is now automatically protected against pressure increases to 10-4 Torr (mbar) when the pump is left unattended. If such an increase should occur, the control unit will be turned off automatically. 6 The pressure in the pump can also be determined by reading the current and converting this reading to pressure with the appropriate pressure versus current graph shown in the appendix “Technical Information” of this manual. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 45/88 4 Installation Procedure Operating Procedure NOTE The steps on the pressure – current charts are a characteristic of the Dual step voltage operation. When the current drawn by the VacIon pump reaches the specified values, the controller will change the high voltage output to a lower applied voltage. 7 When venting the pump, use dry nitrogen. This will avoid water vapor absorption on the pump walls. WARNING! When employing the pump for pumping toxic, flammable, or radioactive gases, please follow the required procedures for each gas disposal. Do not use the pump in the presence of explosive gases. WARNING! When the heating element is installed, do not touch the pump during the heating and cooling phases. The high temperature may cause a serious damage. CAUTION! Do not put any electronic device near the pump since the magnetic field may cause a device malfunction. 46/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 4 Installation Procedure Maintenance Maintenance The VacIon Plus series pump does not require any maintenance. Any work performed on the pump must be carried out by authorized personnel. WARNING! Before carrying out any work on the pump, disconnect it from the High Voltage supply. If a pump is to be scrapped, it must be disposed of in accordance with the specific national standards. Disposal Meaning of the "WEEE" logo found in labels. The following symbol is applied in accordance with the EC WEEE (Waste Electrical and Electronic Equipment) Directive. This symbol (valid only in countries of the European Community) indicates that the product it applies to must NOT be disposed of together with ordinary domestic or industrial waste but must be sent to a differentiated waste collection system. The end user is therefore invited to contact the supplier of the device, whether the Parent Company or a retailer, to initiate the collection and disposal process after checking the contractual terms and conditions of sale. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 47/88 4 Installation Procedure Disposal 48/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) VacIon Plus 150 pumps Instruction Manual 5 Technical Information Description of the VacIon Pump 51 Technical Specification 53 Vaclon Plus 150 Starcell 55 Vaclon Plus 150 Diode 56 Vaclon Plus 150 Noble Diode57 Outline Drawing 58 Stray Magnetic Field 59 VacIon Plus Pump Installation 62 Inspection Procedure 62 Visual Inspection 62 Vacuum Evaluation 63 Short Circuits 64 Typical Installation 65 Inlet Flange Connection 66 Control Unit Connection 69 Safety Interlock 70 Heater Installation 70 Heater Replacement 72 Bakeout Operation 72 Bakeout of VacIon Pump with the Integral Heaters 73 Operating Procedure 74 Original Instructions 49/88 5 Technical Information Disposal Maintenance 75 Exchange of Pumping Element 75 Exchange of the High Voltage Feedthrough 76 Leakage Current Check 77 High-potting 78 Pump Troubleshooting 79 VacIon Plus Pump Replacement Parts and Accessories 80 VacIon Plus Pump Controllers 81 50/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information Description of the VacIon Pump Description of the VacIon Pump The Agilent VacIon Plus 150 pumps are ion pumps and are available in three types: StarCell Noble Diode Diode They operate in the pressure range from 10-2 to below 10-11 mbar. Virtually all gases and vapors can be pumped successfully with the ion pump. The pumping speed will vary depending on the system pressure, the gas type, and the applied operating voltage. The latter can be optimized using Agilent controllers to achieve the lowest possible operating pressure. The VacIon Plus StarCell pump allows starting from as high as 5x10-2 mbar, because ions are prevented from bombarding the system and pump walls at starting pressures, and have high speed for inert gases. The VacIon Plus Diode and Noble Diode pumps require a lower starting pressure (i.e. 10-3 mbar). The tantalum in the Noble Diode pump enhances the noble gas pumping capacity, although it somewhat reduces the high getterable gas pumping speed that the standard diode pump provides. The VacIon Plus StarCell is the latest variation of the Triode configuration. Its patented design makes this ion pump the only one that can handle a high amount of Noble Gases (better than Noble Diode and Triode) and Hydrogen (comparable to the Diode). A positive polarity, high voltage supply is required to operate Diode and Noble Diode pumps. A negative polarity high voltage supply is required to operate StarCell pumps since the anode is grounded and the cathode is held at negative potential. The inlet ConFlat flange is a 6” (NW 100). Fig. 3 shows the VacIon Plus 150 pump, Fig. 4 shows the main assemblies of the pump. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 51/88 5 Technical Information Description of the VacIon Pump Figure 2 VacIon Plus 150 pump Inlet flange Element Cover Figure 3 52/88 Magnet High voltage connector Pump main assemblies VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information Technical Specification Technical Specification The following table details the main technical specifications of the VacIon Plus 150 pumps. Tab. 1 Specification Model StarCell Noble Diode Diode Nominal pumping speed for Nitrogen (*) (l/s) 125 135 150 Operating life at 1 x 10-6 mbar (hours) 80,000 50,000 50,000 Max starting current 300 mA Max baking current 25 mA Protect current 50 mA Operating voltage (max) -7000 Vdc +/- 10 % +7000 Vdc +/- 10 % Maximum starting pressure (mbar) ≤1 x 10-2 ≤1 x 10-3 Ultimate pressure Inlet flange Below 10-11 mbar 6” CFF (NW 100) AISI 304 ESR Internal volume (litres) 12.1 Maximum baking temperature (°C) 350 Temperature limits (°C): Pump Magnet Flange 400 350 500 Material: Body Cathode AISI 304 SST Titanium Titanium/ Tantalum Anode AISI 304 SST Magnet Ferrite Weight, lbs (kg) Titanium 94 (43) (*) Tested according to ISO/DIS 3556-1-1992 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 53/88 5 Technical Information Technical Specification Figures 5 to 12 show the pumping speed vs. pressure diagrams for saturated and unsaturated pumps and the pressure vs. current diagrams for the same pump. The diagrams are for pumps controlled by means of a Dual controller. The pumping speed of a newly regenerated (i.e. baked) sputter ion pump decreases during operation until it reaches a stabilized level known as "saturation" (nominal pumping speed). To saturate the VacIon Plus 150 pumps, it normally requires an amount of gas equal to 5 mbar-liters. Consequently, pumps can operate for extended periods of time at low pressures in the non-saturated state, if they are properly conditioned. 54/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Technical Information Vaclon Plus 150 Starcell 5 Vaclon Plus 150 Starcell Figure 4 Pumping speed vs pressure for Nitrogen Figure 5 Pumping speed vs pressure for Argon Figure 6 Pressure vs current diagram VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 55/88 5 Technical Information Vaclon Plus 150 Diode Vaclon Plus 150 Diode 56/88 Figure 7 Pumping speed vs pressure for Nitrogen Figure 8 Pressure vs current diagram VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Technical Information Vaclon Plus 150 Noble Diode 5 Vaclon Plus 150 Noble Diode Figure 9 Pumping speed vs pressure for Nitrogen Figure 10 Pumping speed vs pressure for Argon Figure 11 Pressure vs current diagram VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 57/88 5 Technical Information Outline Drawing Outline Drawing The following figure shows the outline drawing for the VacIon Plus pump. Figure 12 58/88 VacIon Plus pump outline drawing VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Technical Information Stray Magnetic Field 5 Stray Magnetic Field Curves of stray magnetic field strength along the centre line of the pump and in the plane of the flange as a function of distance from the pump are shown in Figs. 15, 16, 17, 18 and 19, where Bx is the magnetic field along X axis, By is along Y axis and Bz is along Z axis. Figure 13 VacIon Plus pump axis identification Figure 14 Stray magnetic field from flange along X axis (z = 0 mm) VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 59/88 5 Technical Information Stray Magnetic Field 60/88 Figure 15 Stray magnetic field from flange along Y axis (z = 0 mm) Figure 16 Stray magnetic field from flange along X axis (z = 100 mm) VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Technical Information Stray Magnetic Field Figure 17 Stray magnetic field from flange along Y axis (z = 100 mm) Figure 18 Stray magnetic field from flange along Z axis VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 61/88 5 Technical Information VacIon Plus Pump Installation VacIon Plus Pump Installation Inspection Procedure VacIon Plus pumps are evacuated, baked out, sealed and leakchecked at below 1x10-10 mbar prior to shipping. The following information and procedures can be used to evaluate the vacuum integrity of a VacIon Plus pump before installation. Visual Inspection Inspect the pump and magnet for physical damage which may have occurred during shipment. Inspect the pinch-off seal. If it is open, the pump is at atmospheric pressure. WARNING! The pinch-off seal is extremely sharp. Be careful. A VacIon Plus pump that has been exposed to atmosphere during shipment, or while in storage, will operate properly if it has not been damaged. The pump is not harmed by such exposure, although it is good practice to keep it under vacuum when not in use to exclude dust and the accumulation of water vapor from the environment. 62/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information VacIon Plus Pump Installation Vacuum Evaluation The ion pump is shipped in an evacuated condition. Before removing the shipping flange for installation on a vacuum system, it is recommended that the pump be started briefly to verify vacuum integrity and proper operation. To verify the vacuum integrity of the new pump before venting: 1 WARNING! Connect the pump to the control unit as directed in the instruction manual of the control unit. The high voltage which is present in the ion pump from the control unit can cause severe injury or death. Be sure that the high voltage cable is proper connected to the ion pump and its power unit. Make sure that the pump is well grounded. 2 With the main power switch in the OFF position, plug the control unit into a suitable power source. 3 Turn the power to ON. 4 Observe the reading for an indication of one of the following conditions: 5 a If the pump is free of leaks and is at a low pressure, the pressure indication shall quickly fall to or below the 10-8 mbar range as the volume of gas is pumped. b If the pressure inside the pump is at or near atmospheric level, an arc may strike inside the high voltage feedthrough giving a popping sound and the pump current will fluctuate. If this occurs, turn the power OFF immediately. If the vacuum integrity has been lost, the pump should be leakchecked with a mass spectrometer leak detector before installation on the system. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 63/88 5 Technical Information VacIon Plus Pump Installation Short Circuits If there is a short circuit between the anode and cathodes in the pump (or cathode to pump body), the short-circuit current of the control unit will be drawn and low voltage will be indicated. If a short circuit exists in the control unit or high voltage cable and connector, low voltage will also be observed when the high voltage connector is disconnected from the pump (refer to the control unit manuals). An ohm meter reading on the pump feedthrough may not be effective in finding a short. Short circuits may be caused by mechanical shock to the pump. If the pump is shorted, contact Agilent. 64/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Technical Information Typical Installation 5 Typical Installation A typical installation is shown in the following figure and consists of: 1. VacIon Plus pump. 2. A Valve to seal off the pump from the rest of the system (if required). 3. The control unit. 4. A clean roughing pump (i.e. turbo or sorption). 5. A thermocouple gauge capable of indicating pressure from atmosphere to 10-3 mbar range. 6. A valve to seal off the roughing pump from the vacuum chamber. Roughing lines are usually made of stainless steel or copper tubing, or other low vapor pressure material. 7. High voltage cable. 8. Backing pump. Figure 19 Typical installation VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 65/88 5 Technical Information Inlet Flange Connection Inlet Flange Connection The pump should be mounted allowing a sufficient clearance for installation and removal of the high voltage connector. The pump should be kept sealed with its pinch-off tubulation until it is ready for attachment to the vacuum system. This avoids accumulation of dust and other undesirable materials. Before breaking the pinch-off seal, consult the inspection procedure (see preceding paragraph). Vent the ion pump by opening the pinch-off tubulation in a clean area free from smog, dust, pollen, etc. Venting with dry nitrogen gas is further recommended. This can be done by placing a clean polyethylene bag over the ion pump flange. Small pumps can be placed inside of the bag. Purge the bag with clean, dry nitrogen for several minutes, then reach into the bag and release the internal vacuum using pliers to open the copper tube pinch-off. CAUTION! Do not open the pinch off-seal with a saw or grinder. These methods will cause metal particles to be drawn into the pump by the inrushing air as the pump is opened. WARNING! The pinch-off seal is extremely sharp. Be careful when opening. Watch your fingers. Use appropriate procedures to maintain the clean condition of the pump and vacuum system. Unscrew the main flange bolts and lift the blank flange with the help of a bolt screwed into the available threaded hole. Remove the ConFlat flange and the copper gasket plate. Some particles of copper oxide may adhere to the outer edge of the flange gasket. Be careful not to allow them or any other foreign materials to fall into the pump. 66/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information Inlet Flange Connection Connect the ion pump to the vacuum chamber with a short length and large diameter tubulation in order to retain as much pumping speed as possible. Proceed as follows: NOTE 1 Inspect the mating flanges for cleanliness and absence of scratches on the knife edge. 2 Place a new copper gasket between pump flange and vacuum chamber flange. 3 Bolt mating flanges of the pump to the chamber with the screws provided with the ion pump. For flanges over NW 35 (2.75" o.d.) also mount washers below the nuts and screw heads. 4 Use silver-plated screws or apply high temperature lubricant to the screw threads. Lubrication simplifies sealing and disassembly. A recommended lubricant is Fel-Pro C-100. Lubrication is essential to prevent galling of the nut and screw after bakeout. 5 Attach the nuts and tighten each one to 6 – 11 Nm (4.5 - 8 ft.-lbs) of torque. After tightening a nut, always tighten the opposite nut with respect to the center of the flange. This will partially close the gap between the flange faces. 6 Repeat the sequential tightening for two more cycles. 7 Continue tightening the bolts until the flange faces meet and a pronounced increase in torque is felt. Note that it’s not possible to install the screws from the lower side, but only from the upper side of the flange. To allow complete flexibility in the installation, a stud mounting kit is available (see Fig. 22). For ordering information refer to the “VacIon Plus pump replacement parts and accessories” paragraph. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 67/88 5 Technical Information Inlet Flange Connection Figure 20 68/88 Stud mounting kit VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information Control Unit Connection Control Unit Connection WARNING! The high voltage present in the high voltage cable which connects the control unit to the ion pump, can cause severe injury or death. Before mounting the high voltage connector of the cable on the pump high voltage feedthrough, or before removing it, be sure the main power is removed from the control unit. WARNING! To avoid injury, never connect the high voltage to the pump before it is installed into the system and all the inlet flanges are properly connected or blanked off. Make sure that the pump is well connected to the grounded vacuum system. WARNING! Before removing the high voltage connector of the cable from the control unit, be sure the main power is removed from the control unit. Wait at least 10 seconds after removing the main power from the control unit, to allow capacitors to discharge completely. To disconnect the coaxial high voltage cable from the controller, slide the safety locking sleeve (very little sleeve travel is required) from the control unit and at the same time pull on the male end of the cable connector to remove it from the socket on the control unit. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 69/88 5 Technical Information Safety Interlock Safety Interlock The VacIon Plus pump feedthrough in conjunction with the cable P/N 929-0705, -0707, 0708 and -0709, when used with the Agilent MidiVac, DUAL and 4UHV control units, allows the operation of the “High Voltage Cable Safety Interlock” feature. When the high voltage cable connector is disconnected from the VacIon Plus pump feedthrough, the high voltage is automatically switched off by the control unit. Heater Installation The heating element (Fig. 22) is composed of two plates to be fixed to the pump case. NOTE It is advisable to make a 10 – 15 minutes initial operation of the heater in a ventilated room to allow the evaporation of residual internal alloying elements. To install the element proceed as follows (Fig. 23): 70/88 1 Before removing the pole piece, mark the position and orientation for an easier assembly later on. 2 Position the smaller heating plate (1) into the guides on the pump side. Be sure that the written heating plate side is towards you; 3 Position the other heating plate (2) inserting the threaded pins positioned on the pump jacket into the holes of the plate. Be sure that the written heating plate side is towards you; 4 Reinstall the pole piece; 5 Position the connection box (3) on the upper side of the pump in an oppose position of the HV feedthrough with respect to the flange, and fix it to the pump case. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Technical Information Heater Installation Figure 21 Heating element Figure 22 Heating element installation VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 71/88 5 Technical Information Heater Replacement Heater Replacement To replace the heater element proceed as follows: 1 Disconnect the pump from the high voltage; 2 Disconnect the heater element from the main supply; 3 Disconnect the heater plates from the pump case; 4 Disconnect the heater connection box from the upper side of the pump; 5 Install the new heater elements as described in the paragraph “HEATER INSTALLATION”. Bakeout Operation When a VacIon pump does not reach the desired base pressure, and there are no leaks, it is necessary to bake the system. This is done by heating the pump and all the components in the system, and is generally required to achieve base pressure less than 10-8 mbar. 72/88 1 Heat the pump body and the system with a bakeout oven unit or heating strips to temperatures between 150 °C and 220 °C (220 °C is the maximum allowable for most bakeable high voltage cables). This temperature is high enough to degas the pump surfaces of water vapor without damaging the magnet and high voltage connector. Note that the system components must be compatible with the bakeout temperature. The heating must be approximately even on all surfaces or evaporated water will condence on the colder surfaces resulting in an incomplete bake and preventing achievement of UHV vacuum pressure. 2 Leave the pump control unit on and monitor the pressure. Agilent recommends that current during bake not exceed 25 mA; if this value is exceeded, turn the bakeout off and then on again when low pressure is restored. To control the heaters and to monitor the high pressure limit during bakeout in automatic mode, an automatically controlled relay may be used. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information Bakeout Operation 3 Bake the VacIon Plus pump for at least 24 hours. Longer bakeout periods are recommended when the pump has been used with heavy gas loads or when UHV pressure, 10-9 mbar or less is desired. 4 As the pump and system cool down to room temperature, a drop in pressure should be observed. Note that the ion pump can be also baked when switched off, into an external turbo pump through a bakeable isolation valve. Bakeout of VacIon Pump with the Integral Heaters NOTE WARNING! 1 The integral heaters are to be powered with the appropriate voltage. (Please refer to the inscription on the heaters to apply correct voltage). 2 The integral heaters are designed to provide a temperature of 250 °C to 300 °C when the pump is wrapped in a 3-fold aluminium foil. A three-layer foil wrapping is advisable and sufficient to achieve full bakeout/regeneration if the standard European heaters (230 Vac) are operated at 240 Vac, thus preventing overheating. 3 Bakeout the VacIon pump for 24 hours. If the pump is used in heavy gas load applications, it is recommended to bakeout the pump for a longer period. 4 Wait until the pump cools down to room temperature and recovers its base pressure before using it in the application. Do not touch the pump during the heating and cooling phases. The high temperature may cause serious damage. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 73/88 5 Technical Information Operating Procedure Operating Procedure 1 Using a clean roughing pump, evacuate the system to a minimum starting pressure as stated by the technical specification. 10-6 mbar or less is recommended. To reach this pressure the use of an oil free turbo-molecular roughing pump is recommended. 2 When starting an ion pump, a slight increase in vacuum pressure is normal as the internal components are heated and outgassed. If possible leave the roughing pump connected to the system while starting the ion pump. This will make the startup faster and easier. 3 Connect the control unit to a suitable power source and switch the power on. 4 Switch on high voltage to the pump and observe the current and voltage. Fastest starting is obtained using a high applied voltage, 7 kilovolts for example. The applied voltage may be reduced later to optimize pumping and achieve the lowest vacuum pressure. 5 If started at 10-4 mbar, the voltage will start at approximately 800 volts and increase to full voltage as the pump starts operating. The current will start at several milliamps and slowly decrease to microamps or nanoamps as low vacuum pressure is achieved. 6 When first starting the pump, if the voltage decreases instead of increasing, reduce the vacuum pressure then start the ion pump again. 7 When the pump reaches its full operating voltage, you may close the roughing valve. 8 If the pump does not start after 30 minutes of pumping, see the section PUMP TROUBLESHOOTING. 9 Once the pump reaches its base pressure with stable voltage and current, the ion pump may be baked at high temperature if required per the section titled BAKEOUT OPERATION. 10 To stop the ion pump, simply switch off the high voltage. The pump surfaces will continue to pump for a few minutes depending on the system pressure. 11 When venting the pump use clean, dry nitrogen. This will avoid water absorption on the pump surfaces and make subsequent pump downs easier. 74/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information Maintenance Maintenance WARNING! The high voltage present in the high voltage cable which connects the control unit to the ion pump, can cause severe injury or death. Before mounting the high voltage connector of the cable on the pump high voltage feedthrough, or before removing it, be sure the main power is removed from the control unit. Before removing the high voltage connector of the cable from the control unit, be sure the main power is removed from the control unit. Wait at least 10 seconds after removing the main power from the control unit, to allow capacitors to discharge completely. VacIon Plus pumps are maintenance free. In case of life time expiry or premature failure of the pump, please contact your nearest Agilent sales/service office for repair. The VacIon Plus 150 pumps are designed with ex-changeable highvoltage feedthrough and pump elements. Exchange of Pumping Element CAUTION! The high voltage feedthrough contains a ceramic insulator that can be damaged if excessive force is applied in torque, in bending, or in tension. When installing or removing the feedthrough, the applied torque should not exceed 2 Nm. Rotate the feedthrough gently when making the threaded connection to avoid excessive force. 1 Remove the internal high-voltage connections. 2 Remove the element holding brackets. 3 Pull the element out of the pocket and remove it through the inlet flange. 4 Clean the internal part of the pump with the appropriate solvent. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 75/88 5 Technical Information Maintenance 5 Put in the new element through the inlet flange. Make sure that the element is fixed at its rear by the two press-formed buttons of the element pocket, and at its front by the element holding clamp. 6 Reattach the high-voltage connections and test for continuity from the feedthrough. 7 Place a blank flange over the pump inlet and leak-check it with a leak detector. 8 Bake out the blanked off pump while it is operating, and leave it to cool down; then verify that the base pressure is below 10-9 mbar. Exchange of the High Voltage Feedthrough 1 Remove the 6 bolts of the Mini-ConFlat flange connection (see Fig. 24). 2 Remove the cable connection disc. 3 Gently relieve the feedthrough from the metal gasket connection and turn the feedthrough counterclockwise until it is completely detached from the internal high voltage threaded connection. 4 Replace the feedthrough and the copper gasket, making sure that the feedthrough is connected to the pumps internal connector. Check that the feedthrough is not shorted. 5 Bake out the pump while it is operating, and leave it to cool down; then verify that the base pressure is below 10-9 mbar. Figure 23 76/88 High voltage feedthrough connection VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information Maintenance Leakage Current Check If the pump current reading is to be used as pressure measurement, check the pump leakage currents as follows: 1 Turn off the pump control unit. 2 Remove the pump magnet. 3 Turn on the pump control unit and wait for current stabilization. The current reading should not be higher than 150 nA. Make sure that the control unit and the high voltage cable leakage current is negligible. 4 If leakage current comes from the pump, perform "high-potting" per the following procedure; then recheck the pump and install the magnet. 5 If it is not possible to "high-pot" the pump, the pressure reading is biased by the leakage current value. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 77/88 5 Technical Information Maintenance High-potting A constant pump current when no vacuum leak exists is often caused by field emission currents which prevent the use of the pump current as a UHV pressure indicator. To reduce this field emission current, "high-potting" should be performed. "High-potting" is the term used to describe the application of higher than normal operating voltage (8 kV AC, max. 50 mA) for the purpose of burning off "whiskers" (sharp edges) on the pump cathode. High-potting should be done under vacuum and preferably without pump magnets installed (to reduce the drawn current). High-potting must be done carefully and in voltage steps since uncontrolled arcing inside the pump can cause permanent damage. Therefore slowly increase the applied voltage and watch the current meter for indication of arcing inside the pump as whiskers are burned away. If arcing occurs wait at this voltage until the current is stable again. Then slowly increase voltage again in steps up to a maximum voltage. The maximum voltage may be applied to the pump for a period of about 15 seconds. This should effectively remove any "whiskers" or sharp edges on the pump cathode. WARNING! 78/88 Voltages developed in the High Potter power sup-ply are potentially lethal. Use caution during operation and ensure correct grounding connection. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information Pump Troubleshooting Pump Troubleshooting Tab. 1 Symptom Possible cause Correction procedure 1) – Slow starting Starting vacuum pressure too high. Reduce pressure to 10-5 mbar recommended 10-4 mbar minimum. 2) – Slow starting (more than 30 minutes). Air leaks which limit pressure to above 10-6 mbar and cause longer starting time. Leak check the vacuum system with a helium leak detector. 3) – Slow pump-down due to long exposure of viton parts to air. Viton releases considerable gas after long exposure to air. (A bell-jar system which reached 2 x 10-8 mbar in 24 hours after 30 minutes air exposure, will only reach 1 x 10-7 mbar in 24 hours after 20 hours air exposure) With the system under vacuum, pump for several days, or heat to 100-150 °C for up to 15 hours. 4) – Slow pump-down due to absorption of vapours on pump and system walls. Vapours and gases admitted to a system are absorbed on the walls of the system and pump. Subsequent reduction in pressure depends on the rate of depletion of this vapour. Heavy hydrocarbons are most troublesome because of their relative low vapour pressure and they are very difficult to remove, even by baking. Bake the system walls, thereby accelerating the desorption process. Baking mobilizes the vapours so they can be cracked and pumped by discharge (see section “BAKEOUT OPERATION”). 5) – Slow starting or slow pump-down. High voltage feedthrough is leaking. Replace the feedthrough. 6) – Current higher than expected at any given pressure. Ion pump leakage current causing higher pressure reading. Highpot the pump. 7) – System fails to achieve desired UHV vacuum pressure. System not fully baked. Water vapor limits base pressure. Bake the system walls, thereby accelerating the desorption process. Baking mobilizes the vapours so they can be cracked and pumped by discharge (see section “BAKEOUT OPERATION”). 8) – System fails to achieve desired UHV vacuum pressure. System not appropriately cleaned for UHV. Excessive outgassing from walls limits base pressure. Clean all components for UHV and bake the system again. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 79/88 5 Technical Information VacIon Plus Pump Replacement Parts and Accessories VacIon Plus Pump Replacement Parts and Accessories Tab. 2 Part number Basic pump Diode Noble Diode StarCell 919-1510 919-1520 919-1540 Pump with options 919-1511 919-1512 919-1511 919-1512 919-1511 919-1512 919-1511 919-1512 919-1550 919-1550 919-1550 919-1550 919-1551 919-1552 919-1551 919-1552 919-1551 919-1552 919-1551 919-1552 Cables and accessories HV rad. Resist. cable, with interlock 929-0705 Replacement parts HV Feedthrough 959-5125 Pumping elements (q.ty 2 needed) 919-9040 Heaters (power: 480 W) (120V) (220V) Copper gasket for 6” inlet flange (10-pack, individually sealed) Nut and bolt set for 6” inlet flange 919-9045 919-9030 919-0073, 919-0072 FG-0600-CI FB-0450-C12 Copper gasket for 1 1/3" CF feedthrough flange (10-pack, individually sealed) FG-0133-CI Nut and bolt set for 1 1/3" CF feedthrough flange, 8-32 x 3/4 (25-pack) FB-0133-C For a complete overview of Agilent's extensive vacuum product line, please refer to the Agilent Vacuum Catalogue. 80/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5 Technical Information VacIon Plus Pump Controllers VacIon Plus Pump Controllers The following controller series are available to supply the VacIon Plus pumps: MiniVac Dual 4UHV Please refer to the Agilent Vacuum Catalogue to choose the correct controller for each pump. VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 81/88 5 Technical Information VacIon Plus Pump Controllers 82/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) Request for Return Form United States Agilent Technologies Vacuum Products Division 121 Hartwell Avenue Lexington, MA 02421 - USA Tel.: +1 781 861 7200 Fax: +1 781 860 5437 Toll-Free: +1 800 882 7426 Sales and Service Offices India Agilent Technologies India Pvt. Ltd. Vacuum Product Division G01. Prime corporate Park, 230/231, Sahar Road, Opp. Blue Dart Centre, Andheri (East), Mumbai – 400 099.India Tel: +91 22 30648287/8200 Fax: +91 22 30648250 Toll Free: 1800 113037 Italy Agilent Technologies Italia S.p.A. Vacuum Products Division Via F.lli Varian, 54 10040 Leini, (Torino) - Italy Tel.: +39 011 997 9111 Fax: +39 011 997 9350 Toll-Free: 00 800 234 234 00 Southeast Asia Agilent Technologies Sales Sdn Bhd Vacuum Products Division Unit 201, Level 2 uptown 2, 2 Jalan SS21/37, Damansara Uptown 47400 Petaling Jaya, Selangor, Malaysia Tel : +603 7712 6106 Fax: +603 6733 8121 Taiwan Agilent Technologies Taiwan Limited Vacuum Products Division (3F) 20 Kao-Shuang Rd., Pin-Chen City, 324 Taoyuan Hsien , Taiwan, R.O.C. Tel. +886 34959281 Toll Free: 0800 051 342 Canada Central coordination through: Agilent Technologies Vacuum Products Division 121 Hartwell Avenue Lexington, MA 02421 - USA Tel.: +1 781 861 7200 Fax: +1 781 860 5437 Toll-Free: +1 800 882 7426 Japan Agilent Technologies Japan, Ltd. Vacuum Products Division 8th Floor Sumitomo Shibaura Building 4-16-36 Shibaura Minato-ku Tokyo 108-0023 - Japan Tel.: +81 3 5232 1253 Fax: +81 3 5232 1710 Toll-Free: 0120 655 040 UK and Ireland Agilent Technologies UK, Ltd. Vacuum Products Division 6 Mead Road Oxford Industrial Park Yarnton, Oxford OX5 1QU – UK Tel.: +44 (0) 1865 291570 Fax: +44 (0) 1865 291571 Toll free: 00 800 234 234 00 China Agilent Technologies (China) Co. Ltd Vacuum Products Division No.3, Wang Jing Bei Lu, Chao Yang District, Beijing, 100102 China Tel.: +86 (10) 6439 7718 Toll-Free: 800 820 6556 France Agilent Technologies France Vacuum Products Division 7 Avenue des Tropiques Z.A. de Courtaboeuf - B.P. 12 91941 Les Ulis cedex - France Tel.: +33 (0) 1 69 86 38 84 Fax: +33 (0) 1 69 86 29 88 Toll free: 00 800 234 234 00 Germany and Austria Agilent Technologies Vacuum Products Division Alsfelder Strasse 6 Postfach 11 14 35 64289 Darmstadt – Germany Tel.: +49 (0) 6151 703 353 Fax: +49 (0) 6151 703 302 Toll free: 00 800 234 234 00 Korea Agilent Technologies Korea, Ltd. Vacuum Products Division Shinsa 2nd Bldg. 2F 966-5 Daechi-dong Kangnam-gu, Seoul Korea 135-280 Tel.: +82 2 3452 2452 Fax: +82 2 3452 2451 Toll-Free: 080 222 2452 Mexico Agilent Technologies Vacuum Products Division Concepcion Beistegui No 109 Col Del Valle C.P. 03100 – Mexico, D.F. Tel.: +52 5 523 9465 Fax: +52 5 523 9472 Other Countries Agilent Technologies Italia S.p.A. Vacuum Products Division Via F.lli Varian, 54 10040 Leini, (Torino) Italy Tel.: +39 011 997 9111 Fax: +39 011 997 9350 Toll-Free: 00 800 234 234 00 Benelux Agilent Technologies Netherlands B.V. Vacuum Products Division Herculesweg 8 4338 PL Middelburg The Netherlands Tel.: +31 118 671570 Fax: +31 118 671569 Toll-Free: 00 800 234 234 00 Singapore Agilent Technologies Singapore Pte. Ltd, Vacuum Products Division Agilent Technologies Building, 1 Yishun Avenue 7, Singapore 768923 Tel : (65) 6215 8045 Fax : (65) 6754 0574 © Agilent Technologies, Inc. 2011 Printed in ITALY 05/2011 Publication Number: 87-900-102-01 (E) Customer Support & Service NORTH AMERICA: Toll Free: 800 882 7426, Option 3 vpl-ra@agilent.com EUROPE: Toll Free: 00 800 234 234 00 vpt-customercare@agilent.com PACIFIC RIM: please visit our website for individual office information http://www.agilent.com Worldwide Web Site, Catalog and Order On-line: www.agilent.com Representative in most countries 12/10
Source Exif Data:
ExifTool Version Number : 10.40 File Name : download Directory : /home/ubuntu/manual/scrapers/tmp/349/73b4e9cec0748b2040e9037474a6a50f21ae2c548ba5f1c757ac9e2b7b5e0cdf File Size : 2.5 MB File Modification Date/Time : 2018:03:04 13:23:18+00:00 File Access Date/Time : 2018:03:04 13:23:18+00:00 File Inode Change Date/Time : 2018:03:04 13:23:18+00:00 File Permissions : rw-rw-r-- Error : Unknown file typeEXIF Metadata provided by EXIF.tools