919.1551 Vac Ion Plus 150 Pumps User Manual
User Manual: 919.1551
Open the PDF directly: View PDF .
- Procedura per l’installazione
- Anleitung zur Installation
- Procédure pour l’installation
- Installation Procedure
- Technical Information
- Description of the VacIon Pump
- Technical Specification
- Vaclon Plus 150 Starcell
- Vaclon Plus 150 Diode
- Vaclon Plus 150 Noble Diode
- Outline Drawing
- Stray Magnetic Field
- VacIon Plus Pump Installation
- Typical Installation
- Inlet Flange Connection
- Control Unit Connection
- Safety Interlock
- Heater Installation
- Heater Replacement
- Bakeout Operation
- Operating Procedure
- Maintenance
- Pump Troubleshooting
- VacIon Plus Pump Replacement Parts and Accessories
Notices
© Agilent Technologies, Inc. 2011
No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Agilent Technologies, Inc. as governed
by United States and international
copyright laws.
Manual Part Number
Publication Number: 87-900-102-01 (E)
Edition
Edition 05/2011
Printed in ITALY
Agilent Technologies Italia S.p.A.
Vacuum Products Division
Via F.lli Varian, 54
10040 Leinì (TO)
ITALY
Warranty
The material contained in this
document is provided “as is,” and is
subject to being changed, without
notice, in future editions. Further, to
the maximum extent permitted by
applicable law, Agilent disclaims all
warranties, either express or implied,
with regard to this manual and any
information contained herein,
including but not limited to the
implied warranties of merchantability
and fitness for a particular purpose.
Agilent shall not be liable for errors
or for incidental or consequential
damages in connection with the
furnishing, use, or performance of
this document or of any information
contained herein. Should Agilent and
the user have a separate written
agreement with warranty terms
covering the material in this
document that conflict with these
terms, the warranty terms in the
separate agreement shall control.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
Restricted Rights Legend
If software is for use in the
performance of a U.S. Government
prime contract or subcontract,
Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June
1995), or as a “commercial item” as
defined in FAR 2.101(a) or as
“Restricted computer software” as
defined in FAR 52.227-19 (June 1987)
or any equivalent agency regulation or
contract clause. Use, duplication or
disclosure of Software is subject to
Agilent Technologies’ standard
commercial license terms, and non-
DOD Departments and Agencies of the
U.S. Government will receive no
greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will
receive no greater than Limited Rights
as defined in FAR 52.227-14 (June
1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Trademarks
Windows and MS Windows are U.S.
registered trademarks of Microsoft
Corporation.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION notice
until the indicated conditions are fully
understood and met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or the
like that, if not correctly performed
or adhered to, could result in
personal injury or death. Do not
proceed beyond a WARNING notice
until the indicated conditions are
fully understood and met.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
VacIon Plus 150
VacIon Plus 150
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 3/88
VacIon Plus 150
4/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Contents
Contents
1 Procedura per l’installazione 9
Informazioni Generali 10
Preparazione per l’installazione 11
Installazione 12
Uso 14
Procedure di uso 15
Manutenzione 17
Smaltimento 17
2 Anleitung zur Installation 19
Allgemeine Hinweise 20
Vor der Installation 21
Installation 22
Gebrauch 24
Bedienungsschritte 25
Wartung 27
Entsorgung 27
3 Procédure pour l’installation 29
Indications Generales 30
Préparation de l'installation 31
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 5/88
Contents
Installation 32
Utilisation 34
Procédure d’utilisation 35
Maintenance 37
Mise au rebut 37
4 Installation Procedure 39
General Information 40
Preparation for Installation 41
Installation 42
Use 44
Operating Procedure 45
Maintenance 47
Disposal 47
5 Technical Information 49
VacIon Plus Pump Controllers 50
Description of the VacIon Pump 51
Technical Specification 53
Vaclon Plus 150 Starcell 55
Vaclon Plus 150 Diode 56
Vaclon Plus 150 Noble Diode 57
Outline Drawing 58
Stray Magnetic Field 59
6/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Contents
VacIon Plus Pump Installation 62
Typical Installation 65
Inlet Flange Connection 66
Control Unit Connection 69
Safety Interlock 70
Heater Installation 70
Heater Replacement 72
Bakeout Operation 72
Operating Procedure 74
Maintenance 75
Pump Troubleshooting 79
VacIon Plus Pump Replacement Parts and Accessories 80
VacIon Plus Pump Controllers 81
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 7/88
Contents
8/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
1 Procedura per l’installazione
Informazioni Generali
Informazioni Generali
Questa apparecchiatura è destinata ad uso professionale.
L'utilizzatore deve leggere attentamente il presente manuale di
istruzioni ed ogni altra informazione addizionale fornita dalla Agilent
prima dell'utilizzo dell'apparecchiatura. La Agilent si ritiene sollevata
da eventuali responsabilità dovute all'inosservanza totale o parziale
delle istruzioni, ad uso improprio da parte di personale non
addestrato, ad interventi non autorizzati o ad uso contrario alle
normative nazionali specifiche.
Le pompe della serie VacIon Plus sono pompe ioniche utilizzate
comunemente per applicazioni di ultra alto vuoto, grazie alla loro
pulizia, capacità di pompare qualsiasi tipo di gas, e del loro
funzionamento senza vibrazioni e necessità di manutenzione.
Nei paragrafi seguenti sono riportate tutte le informazioni necessarie
a garantire la sicurezza dell'operatore durante l'utilizzo
dell'apparecchiatura. Informazioni dettagliate sono fornite
nell'appendice “Technical information”.
Questo manuale utilizza le seguenti convenzioni:
AVVERTENZA!
I messaggi di avvertenza attirano l'attenzione dell'operatore su una procedura
o una pratica specifica che, se non eseguita in modo corretto, potrebbe
provocare gravi lesioni personali.
ATTENZIONE! I messaggi di attenzione sono visualizzati prima di procedure che, se non
osservate, potrebbero causare danni all'apparecchiatura.
NOTA Le note contengono informazioni importanti estrapolate dal testo.
10/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Procedura per l’installazione
Preparazione per l’installazione
1
Preparazione per l’installazione
La pompa viene fornita in un imballo protettivo speciale; se si
presentano segni di danni, che potrebbero essersi verificati durante il
trasporto, contattare l'ufficio vendite locale. Durante l'operazione di
disimballaggio, prestare particolare attenzione a non lasciar cadere la
pompa e a non sottoporla ad urti o vibrazioni. Non disperdere
l'imballo nell'ambiente. Il materiale è completamente riciclabile e
risponde alla direttiva CEE 85/399 per la tutela dell'ambiente.
ATTENZIONE!
Onde evitare problemi di degassamento, non toccare con le mani nude i
componenti destinati ad essere esposti al vuoto. Utilizzare sempre i guanti o
altra protezione adeguata.
NOTA La pompa non può essere danneggiata rimanendo semplicemente esposta
all'atmosfera. Si consiglia comunque di mantenerla chiusa fino al momento
dell'installazione sul sistema onde evitare eventuale inquinamento da polvere.
Figura 1
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 11/88
1 Procedura per l’installazione
Installazione
Installazione
A
VVERTENZA!
La pompa, a causa del suo peso, deve essere maneggiata tramite appositi
attrezzi di sollevamento e spostamento.
Non installare e/o utilizzare la pompa in ambienti esposti ad agenti
atmosferici (pioggia, gelo, neve), polveri, gas aggressivi, in ambienti
esplosivi o con elevato rischio di incendio. Durante il funzionamento,
per ottenere le specifiche tecniche dichiarate, la temperatura
ambiente deve essere compresa tra 0 °C e +85 °C.
ATTENZIONE!
La pompa deve essere tenuta sigillata con il suo tubo di ingresso schiacciato
finchè non è pronta per essere collegata al sistema.
A
VVERTENZA!
Per evitare lesioni alla persona, non collegare l’alta tensione alla pompa
prima che sia installata nel sistema e che tutte le flange di ingresso siano
adeguatamente collegate o chiuse.
NO!
12/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Procedura per l’installazione
Installazione
1
Il funzionamento delle pompe è ottimizzato solo con l’uso delle
apposite unità di controllo Agilent (Dual, MidiVac o MiniVac).
ATTENZIONE!
Il rispetto delle normative di sicurezza nell’uso delle pompe è garantito solo con
l’uso delle unità di controllo Agilent.
La pompa VacIon Plus può essere installata in qualsiasi posizione.
Per convenienza normalmente viene montata in posizione verticale
con la flangia di ingresso in alto. Per installazioni con posizione
diversa dalla verticale, la pompa deve essere sostenuta tramite i
piedini di montaggio posti sul suo fondo.
Per informazioni dettagliate sull’installazione della pompa, vedere
l’appendice "Technical Information".
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 13/88
1 Procedura per l’installazione
Uso
Uso
Tutte le istruzioni per il corretto funzionamento delle pompe VacIon
Plus sono contenute nel manuale dell'unità di controllo. Leggere
attentamente tale manuale prima dell'utilizzo. Si raccomanda di
portare la pompa ad una pressione di 1x10-4 Torr (mbar) in modo da
ottenere un avvio più rapido. A questo scopo è meglio non utilizzare
una pompa meccanica sigillata con olio, ma se deve essere utilizzata
una tale pompa, si raccomanda di inserire una trappola lungo la linea
di vuoto per ridurre la pressione causata dai vapori di acqua e di olio
provenienti dalla pompa meccanica. Cercare di ridurre al minimo il
tempo in cui la pompa meccanica è aperta verso il sistema e verso la
pompa ionica, poiché i suoi vapori si diffondono nel sistema a
pressioni inferiori a 1x10-1 Torr (mbar) causando una
contaminazione dello stesso. Nei sistemi in cui l’olio deve essere
completamente assente, è meglio utilizzare pompe Turbo per la
suddetta operazione. Depositi igroscopici e l’assorbimento
dell’idrogeno all’interno del composto di titanio possono provocare
l’allungamento del tempo di avvio con l’invecchiamento della pompa.
Durante l’esposizione all’aria, il deposito del composto di titanio
assorbe vapore acqueo; in conseguenza di ciò, al successivo avvio, il
riscaldamento della pompa provoca il rilascio del vapore e di parte
dell’idrogeno pompato precedentemente, con l’effetto
dell’allungamento del tempo di avvio.
14/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Procedura per l’installazione
Procedure di uso
1
Procedure di uso
Controllare che la polarità dell’unità di controllo sia corretta per la
pompa: polarità positiva per le pompe Diode e negativa per le
StarCell e le Triode. Fare riferimento al relativo manuale ed
osservare la seguente procedura per l’uso della pompa:
1 Tramite una pompa di pre-vuoto portare il sistema da vuoto ad
una pressione minima di avvio in accordo alla tabella di pagina 9
ed a seconda del tipo di pompa ionica.
2 Collegare l’unità di controllo ad una apposita fonte di
alimentazione ed accenderla.
3 Osservare la tensione, la corrente e la pressione. Se l’avvio è
avvenuto ad una pressione di 5x10-2 Torr (mbar), è tipica una
tensione di circa 300 - 400 V. Un valore di corrente prossimo alla
corrente di corto circuito dell’unità di controllo è indice
dell’esistenza di una perdita nella pompa e nel sistema. Un
temporaneo incremento della pressione di pre-vuoto è normale
durante la fase di avvio.
4 Lasciare la valvola di pre-vuoto aperta durante l’avvio della
pompa ionica finchè non si raggiunge una adeguata pressione di
avvio. Se la tensione della pompa ionica scende dopo la chiusura
della valvola, riaprirla per un pre-pompaggio aggiuntivo. Appena
la pressione diminuisce, la tensione cresce nuovamente e la
valvola di pre-vuoto può essere chiusa.
5 Quando la tensione ha raggiunto i 2 - 3kV, mettere l’unità di
controllo nello stato PROTECT. Il sistema sarà così autoprotetto
dall’incremento della pressione oltre 1x10-4 Torr (mbar) quando
la pompa non è controllata. Nel caso in cui ci fosse un tale
incremento, l’unità di controllo viene spenta automaticamente.
6 La pressione nella pompa può anche essere calcolata leggendo la
corrente e convertendo la lettura in pressione tramite l’apposito
diagramma pressione-corrente illustrato nell’appendice
“Technical Information” di questo manuale.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 15/88
1 Procedura per l’installazione
Procedure di uso
I gradini nei diagrammi pressione - corrente sono una caratteristica del modo di
funzionamento del Dual. Quando la corrente assorbita dalla pompa raggiunge
determinati valori, l'unità di controllo cambia il valore dell'alta tensione in uscita
ad un valore più basso.
NOTA
7 Quando si porta la pompa alla pressione atmosferica, usare azoto
secco in modo da evitare l’assorbimento di vapore acqueo da
parte delle pareti della pompa.
A
VVERTENZA!
Quando la pompa viene utilizzata per il pompaggio di gas tossici,
infiammabili o radioattivi, seguire le appropriate procedure tipiche di
ciascun gas. Non usare la pompa in presenza di gas esplosivi.
A
VVERTENZA!
Quando è installato il riscaldatore, non toccare la pompa durante le
operazioni di riscaldamento e di raffreddamento. L'elevata temperatura può
causare lesioni alle persone.
ATTENZIONE!
Non avvicinare dispositivi elettronici alla pompa. Il campo magnetico attorno ad
essa può provocare dei malfunzionamenti dei dispositivi stessi.
16/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Procedura per l’installazione
Manutenzione
1
Manutenzione
Le pompe della serie VacIon Plus non richiedono alcuna
manutenzione. Qualsiasi intervento deve essere eseguito da
personale autorizzato.
A
VVERTENZA!
Prima di effettuare qualsiasi intervento sulla pompa scollegarla dall’alta
tensione.
Qualora una pompa dovesse essere rottamata, procedere alla sua
eliminazione nel rispetto delle normative nazionali specifiche.
Smaltimento
Significato del logo "WEEE" presente sulle etichette.
Il simbolo qui sotto riportato applicato in ottemperanza alla direttiva
CE denominata "WEEE".
Questo simbolo (valido solo per i paesi della Comunità Europea)
indica che il prodotto sul quale è applicato, NON deve essere smaltito
insieme ai comuni rifiuti domestici o industriali, ma deve essere
avviato ad un sistema di raccolta differenziata.
Si invita pertanto l'utente finale a contattare il fornitore del
dispositivo, sia esso la casa madre o un rivenditore, per avviare il
processo di raccolta e smaltimento, dopo opportuna verifica dei
termini e condizioni contrattuali di vendita.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 17/88
1 Procedura per l’installazione
Smaltimento
18/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
2 Anleitung zur Installation
Allgemeine Hinweise
Allgemeine Hinweise
Dieses Gerät ist für den professionellen Gebrauch bestimmt. Vor dem
Gebrauch soll der Benutzer dieses Handbuch sowie alle weiteren von
Agilent mitgelieferten Zusatzdokumentationen genau lesen. Bei
vollständiger bzw. teilweiser Nichtbeachtung der enthaltenen
Hinweise, unsachgemäßem Gebrauch durch ungeschultes Personal,
nicht autorisierten Eingriffen und Mißachtung der nationalen
Bestimmungen übernimmt Firma Agilent keinerlei Haftung.
Die Pumpen der Serie VacIon Plus sind Ionenpumpen, die aufgrund
ihrer Reinheit, ihrer Fähigkeit, alle Arten von Gas zu pumpen, und
ihres vibrations- und wartungsfreien Betriebes, allgemein für
Ultrahochvakuumanwendungen zum Einsatz kommen.
In den folgenden Abschnitten sind alle erforderlichen Informationen
für die Sicherheit des Bedieners bei der Anwendung des Geräts
aufgeführt. Detaillierte technische Informationen sind im Anhang
"Technical Information" enthalten.
In dieser Gebrauchsanleitung werden Sicherheitshinweise
folgendermaßen hervorgehoben:
WARNUNG! Die Warnhinweise lenken die Aufmerksamkeit des Bedieners auf eine
spezielle Prozedur oder Praktik, die bei unkorrekter Ausführung schwere
Verletzungen hervorrufen könnte.
Die Vorsichtshinweise vor bestimmten Prozeduren machen den Bediener darauf
aufmerksam, daß bei Nichteinhaltung Schäden am Gerät entstehen können.
VORSICHT!
HINWEIS Die Hinweise enthalten wichtige Informationen, die aus dem Text hervorgehoben
werden.
20/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Anleitung zur Installation
Vor der Installation
2
Vor der Installation
Die Pumpe wird in einer speziellen Schutzverpackung geliefert.
Eventuelle Transportschäden sind der zuständigen örtlichen
Verkaufsstelle zu melden. Beim Auspacken vorsichtig vorgehen,
damit die Pumpe nicht fällt oder Stößen oder Vibrationen ausgesetzt
wird. Das Verpackungsmaterial ist ordnungsgemäß zu entsorgen. Es
ist vollständig recyclebar und entspricht der EG-Richtlinie 85/399 für
den Umweltschutz.
Um Entgasungsprobleme zu vermeiden, dürfen die Komponenten, die mit dem
Vakuum in Kontakt kommen, nicht mit bloßen Händen berührt werden. Stets
Handschuhe oder einen anderen geeigneten Schutz tragen.
VORSICHT!
Die Pumpe kann, wenn sie einfach der Atmosphäre ausgesetzt ist, nicht
beschädigt werden. Sie sollte jedoch bis zur Installation an der Anlage
geschlossen bleiben, um Verunreinigungen durch Staub zu vermeiden.
HINWEIS
Abbildung 1
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 21/88
2 Anleitung zur Installation
Installation
Installation
WARNUNG! Die Pumpe kann aufgrund ihres Gewichtes mittels geeigneter Hebe- und
Förderzeuge gehandhabt werden.
Die Pumpe darf nicht in Umgebungen installiert und/oder benutzt
werden, die ungeschützt vor Witterungsbedingungen (Regen, Frost,
Schnee), Staub und aggressiven Gasen sind, und in denen
Explosions- und erhöhte Brandgefahr besteht. Während des
Betriebes soll die Umgebungstemperatur zwischen 0 °C und +85 °C
betragen, um die angegebenen technischen Merkmale zu
gewährleisten.
Die Pumpe soll versiegelt und mit flachgedrücktem Eintrittsschlauch gehalten
werden, bis sie für den Anschluß an das System bereit ist.
VORSICHT!
WARNUNG!
Um Personenschäden zu vermeiden, darf die Hochspannungsleitung der
Pumpe erst angeschlossen werden, wenn die Pumpe im System installiert ist
und alle Eintrittsflansche entsprechend angeschlossen oder geschlossen
sind.
FALSCH!
22/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Anleitung zur Installation
Installation
2
Der Pumpenbetrieb wird nur durch den Einsatz speziell dafür
vorgesehener Agilent Steuereinheiten (Dual, MidiVac oder MiniVac)
optimiert.
Die Einhaltung der Sicherheitsvorschriften beim Gebrauch der Pumpen ist nur
bei Verwendung von Agilent Steuereinheiten gewährleistet.
VORSICHT!
Die Pumpe Vaclon Plus kann in einer beliebigen Position installiert
werden. Aus praktischen Gründen wird sie in der Regel senkrecht
mit nach oben gerichtetem Eintrittsflansch montiert. Bei nicht
senkrechten Einbaulagen soll die Pumpe auf Montagefüßen aufliegen,
die an ihrer Unterseite angebracht werden.
Detaillierte Informationen zur Installation der Pumpe sind dem
Anhang “Technical Information” zu entnehmen.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 23/88
2 Anleitung zur Installation
Gebrauch
Gebrauch
Sämtliche Hinweise für den korrekten Betrieb der Pumpen VacIon
Plus sind im Handbuch der Steuereinheit enthalten. Dieses
Handbuch ist vor der Inbetriebnahme genau durchzulesen. Es wird
empfohlen, die Pumpe auf einen Druck von 1x10-4 Torr (mbar) zu
bringen, um einen rascheren Anlauf zu gewährleisten. Zu diesem
Zweck empfiehlt es sich, keine ölversiegelte mechanische Pumpe zu
verwenden. Falls hingegen eine derartige Pumpe benutzt werden
muß, soll an der Vakuumleitung eine Falle eingesetzt werden, um den
durch die Wasser- und Öldämpfe aus der mechanischen Pumpe
erzeugten Druck zu verringern. Es ist zu versuchen, die Zeit auf ein
Mindestmaß zu reduzieren, während der die mechanische Pumpe
zum System und zur Ionenpumpe geöffnet ist, da sich ihre Dämpfe
bei Drücken unter 1x10-1 Torr (mbar) im System verbreiten und
dessen Verunreinigung verursachen. Bei Systemen, die vollkommen
ölfrei sein müssen, erweisen sich für den obengenannten Vorgang
Turbopumpen als geeigneter.
Hygroskopische Ablagerungen und die Absorption von Wasserstoff in
der Titanverbindung können die Anlaufzeit verlängern und eine
kürzere Standzeit der Pumpe verursachen. Die Ablagerungen der
Titanverbindungen absorbieren Wasserdampf, wenn sie der Luft
ausgesetzt werden. Dadurch bewirkt beim anschließenden
Anlaufvorgang die Aufheizung der Pumpe die Abgabe des Dampfes
und eines Teils des zuvor gepumpten Wasserstoffs, so daß sich die
Anlaufzeit verlängert.
24/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Anleitung zur Installation
Bedienungsschritte
2
Bedienungsschritte
Es ist zu kontrollieren, daß die Steuereinheit in bezug auf die Pumpe
richtig gepolt ist: positive Polarität für Dioden Pumpen und negative
für StarCell oder Triode Pumpen.
Es ist nach dem diesbezüglichen Handbuch vorzugehen, für den
Gebrauch der Pumpe sind die folgenden Bedienungsschritte zu
beachten:
1 Mit einer Vorvakuumpumpe das System auf einen
Mindestanlaufdruck gemäß Tabelle 9 in Abhängigkeit vom
Ionenpumpentyp bringen.
2 Die Steuereinheit an eine entsprechende Versorgungsquelle
anschließen und einschalten.
3 Die Spannung, die Stromstärke und den Druck beobachten. Wenn
der Anlauf bei einem Druck von 5x10-2 Torr (mbar) erfolgt ist, ist
eine Spannung von ca. 300 – 400 V typisch. Ein Stromwert, der
sich dem Kurzschlußstromwert der Steuereinheit annähert, weist
auf eine Leckstelle an der Pumpe und am System hin. Ein
zeitweiliger Anstieg des Vorvakuumdruckes ist während der
Anlaufphase normal.
4 Während des Anlaufes der Ionenpumpe soll das Vorvakuumventil
geöffnet bleiben, bis ein angemessener Anlaufdruck erreicht ist.
Wenn die Spannung der Ionenpumpe nach Schließung des
Ventils abfällt, ist das Ventil für eine zusätzliche
Vorvakuumpumpung zu öffnen. Sobald der Druck sinkt, steigt die
Spannung erneut an und kann das Vorvakuumventil geschlossen
werden.
5 Wenn die Spannung 2 – 3 kV erreicht hat, ist die Steuereinheit
auf den Status PROTECT zu schalten. Das System verfügt auf
diese Weise über einen Selbstschutz vor einem Druckanstieg über
1x10-4 Torr (mbar), wenn die Pumpe nicht kontrolliert ist. Falls
ein solcher Anstieg stattfindet, wird die Steuereinheit
automatisch ausgeschaltet.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 25/88
2 Anleitung zur Installation
Bedienungsschritte
6 Der Pumpendruck kann auch berechnet werden, indem die
Stromstärke abgelesen wird und der abgelesene Wert mittels des
Druck-Stromstärke-Diagramms im Anhang “Technical
Information” des vorliegenden Handbuches in einen Druckwert
umgerechnet wird.
Die Stufen im Druck-Stromstärke-Diagramm sind charakteris-tisch für die
Funktionsweise des Dual. Wenn die Stromauf-nahme der Pumpe ein bestimmtes
Ausmaß erreicht, ändert die Steuereinheit die ausgangsseitige Hochspannung
auf einen niedrigeren Wert ab.
HINWEIS
7 Wenn die Pumpe auf den atmosphärischen Druck gebracht wird,
ist trockener Stickstoff zu verwenden, um die Aufnahme von
Wasserdampf durch die Pumpenwände zu verhindern.
WARNUNG! Wenn die Pumpe zur Förderung von giftigen, leicht entflammbaren oder
radioaktiven Gasen benutzt wird, sind die für das jeweilige Gas
vorgeschriebenen Vorgänge zu befolgen. Die Pumpe nie bei Vorhandensein
von explosivem Gas benutzen.
WARNUNG! Wenn die Heizung installiert ist, darf die Pumpe während der Aufheizung und
Abkühlung nicht berührt werden. Die hohe Temperatur kann zu
Personenschäden führen.
VORSICHT! Keine elektronischen Geräte in die Nähe der Pumpe bringen. Das darum
befindliche Magnetfeld kann zu Funktionsstörungen der Geräte führen.
26/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Anleitung zur Installation
Wartung
2
Wartung
Die Pumpen der Serie VacIon Plus erfordern keine Wartung.
Sämtliche Eingriffe dürfen nur von autorisiertem Personal
vorgenommen werden.
WARNUNG! Vor Eingriffen an der Pumpe ist diese von der Hochspannungsquelle zu
trennen.
Bei eventueller Verschrottung einer Pumpe ist diese entsprechend
der einschlägigen nationalen Vorschriften zu entsorgen.
Entsorgung
Bedeutung des "WEEE" Logos auf den Etiketten.
Das folgende Symbol ist in Übereinstimmung mit der EU-Richtlinie
WEEE (Waste Electrical and Electronic Equipment) angebracht.
Dieses Symbol (nur in den EU-Ländern gültig) zeigt an, dass das
betreffende Produkt nicht zusammen mit Haushaltsmüll entsorgt
werden darf sondern einem speziellen Sammelsystem zugeführt
werden muss.
Der Endabnehmer sollte daher den Lieferanten des Geräts - d.h. die
Muttergesellschaft oder den Wiederverkäufer - kontaktieren, um den
Entsorgungsprozess zu starten, nachdem er die Verkaufsbedingungen
geprüft hat.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 27/88
2 Anleitung zur Installation
Entsorgung
28/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
3 Procédure pour l’installation
Indications Generales
Indications Generales
Cet appareillage a été conçu en vue d’une utilisation professionnelle.
Il est conseillé à l’utilisateur de lire attentivement cette notice
d’instructions ainsi que toute autre indication supplémentaire
fournie par Agilent avant d’utiliser l’appareil. Agilent décline toute
responsabilité en cas de non respect total ou partiel des instructions
fournies, d’opérations non autorisées, d’utilisation impropre par du
personnel non formé ou contraires aux réglementations nationales
spécifiques.
Grâce à leur propreté, à leur capacité de pomper tous les types de
gaz, à leur fonctionnement sans vibrations et à l’absence d’entretien,
les pompes de la série VacIon Plus sont des pompes ioniques
généralement utilisées pour des applications de vide ultra poussé.
Les paragraphes suivants fournissent toutes les indications
nécessaires à garantir la sécurité de l'opérateur pendant l'utilisation
de l'appareillage. Des renseignements plus détaillés se trouvent dans
l'appendice "Technical Information".
Cette notice utilise les signes conventionnels suivants:
A
VERTISSEMENT!
Les messages d’avertissement attirent l'attention de l'opérateur sur une
procédure ou une manœuvre spéciale dont la mauvaise exécution risque de
provoquer de graves lésions.
Les messages d'attention apparaissent avant certaines procédures dont le non-
respect peut endommager sérieusement l'appareillage.
ATTENTION!
NOTE Les notes contiennent des renseignements importants, extrapolés du texte.
30/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Procédure pour l’installation
Préparation de l'installation
3
Préparation de l'installation
La pompe est fournie dans un emballage de protection spécial; si l'on
constate des signes d’endommagement imputables au transport,
contacter aussitôt le revendeur local. Pendant l'opération de
déballage, veiller tout particulièrement à ne pas laisser tomber la
pompe et à ne lui faire subir aucun choc ni aucune vibration. Ne pas
abandonner l'emballage dans la nature. Le matériel est entièrement
recyclable et il est conforme à la directive CEE 85/399 en matière de
protection de l'environnement.
En vue d'éviter tout problème de dégazage, ne pas toucher à mains nues les
éléments devant être exposés au vide. Mettre toujours des gants ou toute autre
protection appropriée.
ATTENTION!
La pompe ne peut être endommagée si elle reste simplement exposée à
l'atmosphère. Il est quoi qu’il en soit conseillé de ne pas la retirer de son
emballage avant le moment de l'installation, afin d'éviter toute pollution due à la
poussière.
NOTE
Figure 1
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 31/88
3 Procédure pour l’installation
Installation
Installation
A
VERTISSEMENT!
À cause de son poids, il est indispensable de se servir des équipements de
levage et de manutention pour déplacer la pompe.
Ne pas installer et/ou utiliser la pompe dans des milieux exposés à
des agents atmosphériques (pluie, gel, neige), à des poussières, à des
gaz de combat ainsi que dans des milieux explosifs ou à fort risque
d'incendie. Pendant le fonctionnement, pour respecter les
spécifications techniques déclarées la température ambiante doit être
comprise entre 0 °C et +85 °C;
La pompe doit être conservée scellée avec son tuyan d’entrée aplati jusqu’à ce
qu’elle soit prête à être branchée au système.
ATTENTION!
A
VERTISSEMENT!
Pour éviter toute lésion aux personnes, ne pas brancher la haute tension à la
pompe avant que celle-ci soit installée dans le système et avant que toutes
les brides d’entrées soient correctement assemblées ou fermées.
NON!
Le fonctionnement de la pompe n'est optimisé que si celle-ci est
utilisée avec l'une des unités de contrôle Agilent spécifiques (Dual,
MidiVac ou MiniVac).
32/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Procédure pour l’installation
Installation
3
Lors de l'utilisation des pompes, le respect des normes de sécurité est
impérativement subordonné à l'emploi des unités de contrôle Agilent.
ATTENTION!
La pompe VacIon Plus peut être installée dans toutes les positions.
Par facilité elle est généralement montée en position verticale avec la
bride d'entrée en partie haute. Pour toute installation différente de la
position verticale, la pompe doit reposer sur ses pieds de montage
placé sur le fond.
Pour plus de détails sur l’installation de la pompe, consulter
l’appendice “Technical Information”.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 33/88
3 Procédure pour l’installation
Utilisation
Utilisation
Toutes les instructions pour le fonctionnement correct de la pompe
VacIon Plus sont fournies dans la notice de l'unité de contrôle. Il est
conseillé de lire attentivement cette notice avant d'utiliser la pompe.
Il est recommandé de porter la pompe à une pression de 1x10-4 Torrs
(mbars) de façon à obtenir un démarrage plus rapide. Pour ce faire il
est préférable de ne pas utiliser une pompe mécanique scellée à
l’huile mais au cas où il serait nécessaire d’utiliser une telle pompe, il
est recommandé d’introduire un dispositif de retenue le long de la
ligne de vide afin de réduire la pression due aux vapeurs d’eau et
d’huile engendrées par la pompe mécanique. Essayer de réduire au
minimum le temps où la pompe mécanique est ouverte vers le
système et vers la pompe ionique car ses vapeurs se répandent dans
le système à des pressions inférieures à 1x10-1 Torrs (mbars) causant
une contamination de celui-ci. Dans les systèmes où l’huile doit être
complètement absente, il est préférable d’effectuer ladite opération à
l’aide d’une pompe Turbo. Des dépôts hygroscopiques et l’absorption
d’hydrogène dans le composé de titane peuvent provoquer, par effet
du vieillissement de la pompe, l’allongement du temps de démarrage.
Pendant l’exposition à l’air, le dépôt du composé de titane absorbe de
la vapeur d’eau et cette action a pour conséquence qu’au démarrage
suivant le chauffage de la pompe entraînera la dispersion de la
vapeur et d’une partie de l’hydrogène pompée précédemment ce qui
aura pour effet d’allonger le temps de démarrage.
34/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Procédure pour l’installation
Procédure d’utilisation
3
P
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 35/88
rocédure d’utilisation
Contrôler que la polarité de l’unité de contrôle soit correcte pour la
pompe: polarité positive pour les pompes Diode et négative pour les
pompes StarCell et Triode. Se reporter au manuel correspondant et
observer la procédure suivante pour l’utilisation de la pompe:
1 A l'aide d'une pompe de pré-vide, porter le système à une
pression minimale de démarrage en suivant le tableau de la page
9 et en fonction du type de pompe ionique.
2 Brancher l’unité de contrôle à une source d’alimentation
appropriée et l’allumer.
3 Contrôler la tension, le courant et la pression. Lorsque la mise en
route est effectuée à une pression de 5x10-2 Torrs (mbars) la
tension est généralement d’environ 300 - 400 V. Une valeur de
courant proche du courant de court-circuit de l’unité de contrôle
révèle l’existence d’une fuite dans la pompe et dans le système.
Au cours de la phase de démarrage, un accroissement
momentané de la pression de pré-vide est normal.
4 Laisser la soupape de pré-vide ouverte pendant le démarrage de
la pompe ionique tant qu’une pression de démarrage appropriée
n’a pas été atteinte. Si la tension de la pompe ionique descend
après la fermeture de la soupape, la rouvrir pour un pré-pompage
supplémentaire. Dès que la pression commence à diminuer, la
tension augmente à nouveau et la valve de pré-vide peut être
fermée.
5 Lorsque la tension a atteint 2 - 3 kV, mettre l’unité de contrôle en
état de PROTECT. Le système sera ainsi protégé contre toute
augmentation de pression au-delà de 1x10-4 Torrs (mbars)
lorsque la pompe n’est pas contrôlée. Si une telle augmentation
devait être enregistrée, l’unité de contrôle s’éteindrait
automatiquement.
6 La pression de la pompe peut également être calculée en relevant
le courant et en convertissant la lecture en pression à l’aide du
diagramme pression-courant illustré dans l’appendice “Technical
Information” de ce manuel.
3 Procédure pour l’installation
Procédure d’utilisation
Les paliers du diagramme de pression – courant sont une caractéristique du
mode de fonctionnement du Dual. Lorsque le courant absorbé par la pompe
atteint des valeurs déterminées, l'unité de contrôle porte la valeur de haute
tension en sortie à une valeur inférieure.
NOTE
7 Lorsque l’on porte la pompe à la pression atmosphérique, utiliser
de l’azote sec de façon à éviter que les parois de la pompe
n’absorbent de la vapeur aqueuse.
A
VERTISSEMENT!
Lorsque la pompe est utilisée pour le pompage de gaz toxiques,
inflammables ou radioactifs, suivre les procédures appropriées à chaque
gaz. Ne pas utiliser la pompe en présence de gaz explosifs.
A
VERTISSEMENT!
Lorsque le réchauffeur est installé, éviter de toucher la pompe pendant les
opérations de chauffage et de refroidissement. La température élevée peut
provoquer des brûlures.
ATTENTION! Ne pas approcher de dispositifs électroniques de la pompe. Le champ
magnétique environnant cette dernière peut entraîner des dysfonctionnements
desdits dispositifs.
36/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Procédure pour l’installation
Maintenance
3
Maintenance
Les pompes de la série VacIon Plus ne demandent aucun entretien.
Toute intervention doit être exécutée par un personnel agréé.
A
VERTISSEMENT!
Avant toute intervention sur la pompe, la débrancher de la haute tension.
En cas de mise au rebut d’une pompe, procéder à son élimination
dans le respect des normes nationales en vigueur.
Mise au rebut
Signification du logo "WEEE" imprimé sur les étiquettes.
Le symbole ci-dessous est appliqué conformément à la directive CE
nommée "WEEE".
Ce symbole (uniquement valide pour les pays de la Communauté
européenne) indique que le produit sur lequel il est appliqué NE doit
PAS être mis au rebut avec les ordures ménagères ou les déchets
industriels ordinaires, mais passer par un système de collecte
sélective.
Après avoir vérifié les termes et conditions du contrat de vente,
l’utilisateur final est donc prié de contacter le fournisseur du
dispositif, maison mère ou revendeur, pour mettre en œuvre le
processus de collecte et mise au rebut.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 37/88
3 Procédure pour l’installation
Mise au rebut
38/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
4 Installation Procedure
General Information
General Information
This equipment is destined for use by professionals. The user should
read this instruction manual and any other additional information
supplied by Agilent before operating the equipment. Agilent will not
be held responsible for any events occurring due to non-compliance,
even partial, with these instructions, improper use by untrained
persons, non-authorized interference with the equipment or any
action contrary to that provided for by specific national standards.
The VacIon Plus series pumps are ion pumps commonly used to
create ultra-high vacuum, due to their cleanliness, ability to pump
different gases, and maintenance- and vibration-free operation.
The following paragraphs contain all the information necessary to
guarantee the safety of the operator when using the equipment.
Detailed information is supplied in the appendix "Technical
Information".
This manual uses the following standard protocol:
The warning messages are for attracting the attention of the operator to a
particular procedure or practice which, if not followed correctly, could lead
to serious injury.
WARNING!
The caution messages are displayed before procedures which, if not followed,
could cause damage to the equipment.
CAUTION!
The notes contain important information taken from the text.
NOTE
40/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Installation Procedure
Preparation for Installation
4
Preparation for Installation
The pump is supplied in a special protective packing. If this shows
signs of damage which may have occurred during transport, contact
your local sales office. When unpacking the pump, be sure not to
drop it and avoid any kind of sudden impact or shock vibration to it.
Do not dispose of the packing materials in an unauthorized manner.
The material is 100 % recyclable and complies with EEC Directive
85/399.
In order to prevent outgassing problems, do not use bare hands to handle
components which will be exposed to vacuum. Always use gloves or other
appropriate protection.
CAUTION!
Normal exposure to the environment cannot damage the pump. Nevertheless, it
is advisable to keep it closed until it is installed in the system, thus preventing
any form of pollution by dust.
NOTE
Figure 1
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 41/88
4 Installation Procedure
Installation
Installation
The pump is heavy and must be handled by means of suitable moving and
handling tools.
WARNING!
Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, to obtain the declared functioning specification, the
ambient temperature must be between 0 °C and +85 °C.
CAUTION! The pump should be kept sealed with its pinch-off tube until it is ready for
attachment to the vacuum system.
WARNING!
To avoid injury, never connect the high voltage to the pump before it is
installed into the system and all the inlet flanges are properly connected or
blanked off.
NO!
The pump operation is optimized using one of the special Agilent
controllers (Dual, MidiVac or MiniVac) only.
42/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Installation Procedure
Installation
4
The safety specifications agreement using the pump is guaranteed using the
Agilent controller only.
CAUTION!
The VacIon Plus pump can be installed in any position. For
convenience, a pump is usually mounted vertically with the inlet up.
For mounting other than vertical, the pump should be supported
using the mounting feet at the bottom of the pump.
For detailed information about the pump installation, see the
appendix “Technical Information”.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 43/88
4 Installation Procedure
Use
Use
All the instructions for the correct use of the VacIon Plus pumps are
contained in the control unit manual. Read the manual carefully
before using the pump. Rough pumping down to 1x10-4 Torr (mbar) is
recommended for the most rapid starting. Roughing with an oil-
sealed mechanical pump is not desirable, but when used, a trap in
the roughing line is recommended to reduce pressure due to water
vapor and oils from the mechanical pump. Be careful to minimize the
time that the mechanical pump is open to the system and ion pump,
since mechanical pump vapors will start diffusing into the system at
pressures below 1x10-1 Torr (mbar) and cause contamination. In
systems where oils must be completely eliminated, turbopump
roughing pumps should be used. Hygroscopic deposits and hydrogen
absorption into titanium may cause starting times to increase with
age. During exposure to air, the deposits of titanium compound
absorb water vapor. In subsequent start ups, pump heating causes
release of the water vapor and some previously pumped hydrogen;
thus, the starting time may be lengthened.
44/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Installation Procedure
Operating Procedure
4
O
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 45/88
perating Procedure
Check that the controller HV polarity is correct for the pump:
positive polarity for Diode pumps and negative for StarCell or Triode
pumps. Refer to the relevant pump control unit instruction manual
and follow the procedure below when operating the pump:
1 With a clean roughing pump, establish a minimum starting
pressure in the vacuum system per the table on page 9 for the
type of ion pump.
2 Plug the control unit into a suitable power source and switch the
power ON.
3 Observe the voltage, current, and roughing pressure. If started at
5x10-2 Torr (mbar), a controller voltage of approximately 300 to
400 volts is typical. A current value near the short-circuit current
of the control unit could indicate that an unconfined flow
discharge exists in the pump and system. A temporary rise in
roughing pressure will usually be noticed during any starting
procedure.
4 Allow the roughing valve to remain open after turning on the ion
pump until an adequate starting pressure is reached. If the ion
pump voltage drops after closing the roughing valves, reopen the
valve for additional rough pumping. As the pressure decreases,
the voltage again will rise, and the roughing valve may be closed.
5 When the voltage has increased to 2-3 kV, place the control unit
in the PROTECT condition. The system is now automatically
protected against pressure increases to 10-4 Torr (mbar) when the
pump is left unattended. If such an increase should occur, the
control unit will be turned off automatically.
6 The pressure in the pump can also be determined by reading the
current and converting this reading to pressure with the
appropriate pressure versus current graph shown in the
appendix “Technical Information” of this manual.
4 Installation Procedure
Operating Procedure
NOTE The steps on the pressure – current charts are a characteristic of the Dual step
voltage operation. When the current drawn by the VacIon pump reaches the
specified values, the controller will change the high voltage output to a lower
applied voltage.
7 When venting the pump, use dry nitrogen. This will avoid water
vapor absorption on the pump walls.
WARNING!
When employing the pump for pumping toxic, flammable, or radioactive
gases, please follow the required procedures for each gas disposal. Do not
use the pump in the presence of explosive gases.
When the heating element is installed, do not touch the pump during the
heating and cooling phases. The high temperature may cause a serious
damage.
WARNING!
CAUTION! Do not put any electronic device near the pump since the magnetic field may
cause a device malfunction.
46/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Installation Procedure
Maintenance
4
Maintenance
The VacIon Plus series pump does not require any maintenance. Any
work performed on the pump must be carried out by authorized
personnel.
Before carrying out any work on the pump, disconnect it from the High
Voltage supply.
WARNING!
If a pump is to be scrapped, it must be disposed of in accordance
with the specific national standards.
Disposal
Meaning of the "WEEE" logo found in labels.
The following symbol is applied in accordance with the EC WEEE
(Waste Electrical and Electronic Equipment)
Directive. This symbol (valid only in countries of the European
Community) indicates that the product it applies to must NOT be
disposed of together with ordinary domestic or industrial waste but
must be sent to a differentiated waste collection system.
The end user is therefore invited to contact the supplier of the device,
whether the Parent Company or a retailer, to initiate the collection
and disposal process after checking the contractual terms and
conditions of sale.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 47/88
4 Installation Procedure
Disposal
48/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
VacIon Plus 150 pumps Instruction Manual
5
Technical Information
Description of the VacIon Pump 51
Technical Specification 53
Vaclon Plus 150 Starcell 55
Vaclon Plus 150 Diode 56
Vaclon Plus 150 Noble Diode57
Outline Drawing 58
Stray Magnetic Field 59
VacIon Plus Pump Installation 62
Inspection Procedure 62
Visual Inspection 62
Vacuum Evaluation 63
Short Circuits 64
Typical Installation 65
Inlet Flange Connection 66
Control Unit Connection 69
Safety Interlock 70
Heater Installation 70
Heater Replacement 72
Bakeout Operation 72
Bakeout of VacIon Pump with the Integral Heaters 73
Operating Procedure 74
Original Instructions
49/88
5 Technical Information
Disposal
Maintenance 75
Exchange of Pumping Element 75
Exchange of the High Voltage Feedthrough 76
Leakage Current Check 77
High-potting 78
Pump Troubleshooting 79
VacIon Plus Pump Replacement Parts and Accessories 80
VacIon Plus Pump Controllers 81
50/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Description of the VacIon Pump
5
Description of the VacIon Pump
The Agilent VacIon Plus 150 pumps are ion pumps and are available
in three types:
StarCell
Noble Diode
Diode
They operate in the pressure range from 10-2 to below 10-11 mbar.
Virtually all gases and vapors can be pumped successfully with the
ion pump. The pumping speed will vary depending on the system
pressure, the gas type, and the applied operating voltage. The latter
can be optimized using Agilent controllers to achieve the lowest
possible operating pressure.
The VacIon Plus StarCell pump allows starting from as high as 5x10-2
mbar, because ions are prevented from bombarding the system and
pump walls at starting pressures, and have high speed for inert gases.
The VacIon Plus Diode and Noble Diode pumps require a lower
starting pressure (i.e. 10-3 mbar).
The tantalum in the Noble Diode pump enhances the noble gas
pumping capacity, although it somewhat reduces the high getterable
gas pumping speed that the standard diode pump provides.
The VacIon Plus StarCell is the latest variation of the Triode
configuration. Its patented design makes this ion pump the only one
that can handle a high amount of Noble Gases (better than Noble
Diode and Triode) and Hydrogen (comparable to the Diode).
A positive polarity, high voltage supply is required to operate Diode
and Noble Diode pumps. A negative polarity high voltage supply is
required to operate StarCell pumps since the anode is grounded and
the cathode is held at negative potential.
The inlet ConFlat flange is a 6” (NW 100).
Fig. 3 shows the VacIon Plus 150 pump, Fig. 4 shows the main
assemblies of the pump.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 51/88
5 Technical Information
Description of the VacIon Pump
Figure 2 VacIon Plus 150 pump
Inlet flange
Cover High voltage
connector
Magnet
Element
Figure 3 Pump main assemblies
52/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Technical Specification
5
Technical Specification
The following table details the main technical specifications of the
VacIon Plus 150 pumps.
Tab. 1
Specification Model
StarCell Noble Diode Diode
Nominal pumping speed for Nitrogen
(*) (l/s)
125 135 150
Operating life at 1 x 10-6 mbar (hours) 80,000 50,000 50,000
Max starting current 300 mA
Max baking current 25 mA
Protect current 50 mA
Operating voltage (max) -7000 Vdc
+/- 10 %
+7000 Vdc
+/- 10 %
Maximum starting pressure (mbar) ≤1 x 10-2 ≤1 x 10-3
Ultimate pressure Below 10-11 mbar
Inlet flange 6” CFF (NW 100) AISI 304 ESR
Internal volume (litres) 12.1
Maximum baking temperature (°C) 350
Temperature limits (°C):
Pump
Magnet
Flange
400
350
500
Material: Body
Cathode
Anode
Magnet
AISI 304 SST
Titanium Titanium/
Tantalum
Titanium
AISI 304 SST
Ferrite
Weight, lbs (kg) 94 (43)
(*) Tested according to ISO/DIS 3556-1-1992
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 53/88
5 Technical Information
Technical Specification
Figures 5 to 12 show the pumping speed vs. pressure diagrams for
saturated and unsaturated pumps and the pressure vs. current
diagrams for the same pump. The diagrams are for pumps controlled
by means of a Dual controller.
The pumping speed of a newly regenerated (i.e. baked) sputter ion
pump decreases during operation until it reaches a stabilized level
known as "saturation" (nominal pumping speed). To saturate the
VacIon Plus 150 pumps, it normally requires an amount of gas equal
to 5 mbar-liters. Consequently, pumps can operate for extended
periods of time at low pressures in the non-saturated state, if they
are properly conditioned.
54/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Vaclon Plus 150 Starcell
5
Vaclon Plus 150 Starcell
Figure 4 Pumping speed vs pressure for Nitrogen
Figure 5 Pumping speed vs pressure for Argon
Figure 6 Pressure vs current diagram
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 55/88
5 Technical Information
Vaclon Plus 150 Diode
Vaclon Plus 150 Diode
Figure 7 Pumping speed vs pressure for Nitrogen
Figure 8 Pressure vs current diagram
56/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Vaclon Plus 150 Noble Diode
5
Vaclon Plus 150 Noble Diode
Figure 9 Pumping speed vs pressure for Nitrogen
Figure 10 Pumping speed vs pressure for Argon
Figure 11 Pressure vs current diagram
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 57/88
5 Technical Information
Outline Drawing
Outline Drawing
The following figure shows the outline drawing for the VacIon Plus
pump.
Figure 12 VacIon Plus pump outline drawing
58/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Stray Magnetic Field
5
Stray Magnetic Field
Curves of stray magnetic field strength along the centre line of the
pump and in the plane of the flange as a function of distance from
the pump are shown in Figs. 15, 16, 17, 18 and 19, where Bx is the
magnetic field along X axis, By is along Y axis and Bz is along Z axis.
Figure 13 VacIon Plus pump axis identification
Figure 14 Stray magnetic field from flange along X axis (z = 0 mm)
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 59/88
5 Technical Information
Stray Magnetic Field
Figure 15 Stray magnetic field from flange along Y axis (z = 0 mm)
Figure 16 Stray magnetic field from flange along X axis (z = 100 mm)
60/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Stray Magnetic Field
5
Figure 17 Stray magnetic field from flange along Y axis (z = 100 mm)
Figure 18 Stray magnetic field from flange along Z axis
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 61/88
5 Technical Information
VacIon Plus Pump Installation
VacIon Plus Pump Installation
Inspection Procedure
VacIon Plus pumps are evacuated, baked out, sealed and leak-
checked at below 1x10-10 mbar prior to shipping. The following
information and procedures can be used to evaluate the vacuum
integrity of a VacIon Plus pump before installation.
Visual Inspection
Inspect the pump and magnet for physical damage which may have
occurred during shipment. Inspect the pinch-off seal. If it is open, the
pump is at atmospheric pressure.
The pinch-off seal is extremely sharp. Be careful.
WARNING!
A VacIon Plus pump that has been exposed to atmosphere during
shipment, or while in storage, will operate properly if it has not been
damaged.
The pump is not harmed by such exposure, although it is good
practice to keep it under vacuum when not in use to exclude dust
and the accumulation of water vapor from the environment.
62/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
VacIon Plus Pump Installation
5
Vacuum Evaluation
The ion pump is shipped in an evacuated condition. Before removing
the shipping flange for installation on a vacuum system, it is
recommended that the pump be started briefly to verify vacuum
integrity and proper operation.
To verify the vacuum integrity of the new pump before venting:
1 Connect the pump to the control unit as directed in the
instruction manual of the control unit.
The high voltage which is present in the ion pump from the control unit can
cause severe injury or death. Be sure that the high voltage cable is proper
connected to the ion pump and its power unit. Make sure that the pump is
well grounded.
WARNING!
2 With the main power switch in the OFF position, plug the control
unit into a suitable power source.
3 Turn the power to ON.
4 Observe the reading for an indication of one of the following
conditions:
a If the pump is free of leaks and is at a low pressure, the
pressure indication shall quickly fall to or below the
10-8 mbar range as the volume of gas is pumped.
b If the pressure inside the pump is at or near atmospheric
level, an arc may strike inside the high voltage feedthrough
giving a popping sound and the pump current will fluctuate.
If this occurs, turn the power OFF immediately.
5 If the vacuum integrity has been lost, the pump should be leak-
checked with a mass spectrometer leak detector before
installation on the system.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 63/88
5 Technical Information
VacIon Plus Pump Installation
Short Circuits
If there is a short circuit between the anode and cathodes in the
pump (or cathode to pump body), the short-circuit current of the
control unit will be drawn and low voltage will be indicated. If a
short circuit exists in the control unit or high voltage cable and
connector, low voltage will also be observed when the high voltage
connector is disconnected from the pump (refer to the control unit
manuals).
An ohm meter reading on the pump feedthrough may not be effective
in finding a short.
Short circuits may be caused by mechanical shock to the pump. If the
pump is shorted, contact Agilent.
64/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Typical Installation
5
Typical Installation
A typical installation is shown in the following figure and consists of:
1. VacIon Plus pump.
2. A Valve to seal off the pump from the rest of the system (if
required).
3. The control unit.
4. A clean roughing pump (i.e. turbo or sorption).
5. A thermocouple gauge capable of indicating pressure from
atmosphere to 10-3 mbar range.
6. A valve to seal off the roughing pump from the vacuum chamber.
Roughing lines are usually made of stainless steel or copper
tubing, or other low vapor pressure material.
7. High voltage cable.
8. Backing pump.
Figure 19 Typical installation
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 65/88
5 Technical Information
Inlet Flange Connection
Inlet Flange Connection
The pump should be mounted allowing a sufficient clearance for
installation and removal of the high voltage connector.
The pump should be kept sealed with its pinch-off tubulation until it
is ready for attachment to the vacuum system. This avoids
accumulation of dust and other undesirable materials. Before
breaking the pinch-off seal, consult the inspection procedure (see
preceding paragraph).
Vent the ion pump by opening the pinch-off tubulation in a clean area
free from smog, dust, pollen, etc. Venting with dry nitrogen gas is
further recommended. This can be done by placing a clean
polyethylene bag over the ion pump flange. Small pumps can be
placed inside of the bag. Purge the bag with clean, dry nitrogen for
several minutes, then reach into the bag and release the internal
vacuum using pliers to open the copper tube pinch-off.
CAUTION! Do not open the pinch off-seal with a saw or grinder. These methods will cause
metal particles to be drawn into the pump by the inrushing air as the pump is
opened.
The pinch-off seal is extremely sharp. Be careful when opening. Watch your
fingers.
WARNING!
Use appropriate procedures to maintain the clean condition of the
pump and vacuum system.
Unscrew the main flange bolts and lift the blank flange with the help
of a bolt screwed into the available threaded hole. Remove the
ConFlat flange and the copper gasket plate. Some particles of copper
oxide may adhere to the outer edge of the flange gasket. Be careful
not to allow them or any other foreign materials to fall into the pump.
66/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Inlet Flange Connection
5
Connect the ion pump to the vacuum chamber with a short length
and large diameter tubulation in order to retain as much pumping
speed as possible. Proceed as follows:
1 Inspect the mating flanges for cleanliness and absence of
scratches on the knife edge.
2 Place a new copper gasket between pump flange and vacuum
chamber flange.
3 Bolt mating flanges of the pump to the chamber with the screws
provided with the ion pump. For flanges over NW 35 (2.75" o.d.)
also mount washers below the nuts and screw heads.
4 Use silver-plated screws or apply high temperature lubricant to
the screw threads. Lubrication simplifies sealing and
disassembly. A recommended lubricant is Fel-Pro C-100.
Lubrication is essential to prevent galling of the nut and screw after bakeout.
NOTE
5 Attach the nuts and tighten each one to 6 – 11 Nm (4.5 - 8 ft.-lbs)
of torque. After tightening a nut, always tighten the opposite nut
with respect to the center of the flange. This will partially close
the gap between the flange faces.
6 Repeat the sequential tightening for two more cycles.
7 Continue tightening the bolts until the flange faces meet and a
pronounced increase in torque is felt.
Note that it’s not possible to install the screws from the lower side,
but only from the upper side of the flange. To allow complete
flexibility in the installation, a stud mounting kit is available (see Fig.
22). For ordering information refer to the “VacIon Plus pump
replacement parts and accessories” paragraph.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 67/88
5 Technical Information
Inlet Flange Connection
Figure 20 Stud mounting kit
68/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Control Unit Connection
5
Control Unit Connection
WARNING!
The high voltage present in the high voltage cable which connects the control
unit to the ion pump, can cause severe injury or death. Before mounting the
high voltage connector of the cable on the pump high voltage feedthrough, or
before removing it, be sure the main power is removed from the control unit.
WARNING!
To avoid injury, never connect the high voltage to the pump before it is
installed into the system and all the inlet flanges are properly connected or
blanked off. Make sure that the pump is well connected to the grounded
vacuum system.
Before removing the high voltage connector of the cable from the control unit,
be sure the main power is removed from the control unit. Wait at least 10
seconds after removing the main power from the control unit, to allow
capacitors to discharge completely.
WARNING!
To disconnect the coaxial high voltage cable from the controller, slide
the safety locking sleeve (very little sleeve travel is required) from the
control unit and at the same time pull on the male end of the cable
connector to remove it from the socket on the control unit.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 69/88
5 Technical Information
Safety Interlock
Safety Interlock
The VacIon Plus pump feedthrough in conjunction with the cable P/N
929-0705, -0707, 0708 and -0709, when used with the Agilent
MidiVac, DUAL and 4UHV control units, allows the operation of the
“High Voltage Cable Safety Interlock” feature.
When the high voltage cable connector is disconnected from the
VacIon Plus pump feedthrough, the high voltage is automatically
switched off by the control unit.
Heater Installation
The heating element (Fig. 22) is composed of two plates to be fixed to
the pump case.
It is advisable to make a 10 – 15 minutes initial operation of the heater in a
ventilated room to allow the evaporation of residual internal alloying elements.
NOTE
To install the element proceed as follows (Fig. 23):
1 Before removing the pole piece, mark the position and
orientation for an easier assembly later on.
2 Position the smaller heating plate (1) into the guides on the pump
side. Be sure that the written heating plate side is towards you;
3 Position the other heating plate (2) inserting the threaded pins
positioned on the pump jacket into the holes of the plate. Be sure
that the written heating plate side is towards you;
4 Reinstall the pole piece;
5 Position the connection box (3) on the upper side of the pump in
an oppose position of the HV feedthrough with respect to the
flange, and fix it to the pump case.
70/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Heater Installation
5
Figure 21 Heating element
Figure 22 Heating element installation
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 71/88
5 Technical Information
Heater Replacement
Heater Replacement
To replace the heater element proceed as follows:
1 Disconnect the pump from the high voltage;
2 Disconnect the heater element from the main supply;
3 Disconnect the heater plates from the pump case;
4 Disconnect the heater connection box from the upper side of the
pump;
5 Install the new heater elements as described in the paragraph
“HEATER INSTALLATION”.
Bakeout Operation
When a VacIon pump does not reach the desired base pressure, and
there are no leaks, it is necessary to bake the system. This is done by
heating the pump and all the components in the system, and is
generally required to achieve base pressure less than 10-8 mbar.
1 Heat the pump body and the system with a bakeout oven unit or
heating strips to temperatures between 150 °C and 220 °C (220
°C is the maximum allowable for most bakeable high voltage
cables). This temperature is high enough to degas the pump
surfaces of water vapor without damaging the magnet and high
voltage connector. Note that the system components must be
compatible with the bakeout temperature. The heating must be
approximately even on all surfaces or evaporated water will
condence on the colder surfaces resulting in an incomplete bake
and preventing achievement of UHV vacuum pressure.
2 Leave the pump control unit on and monitor the pressure.
Agilent recommends that current during bake not exceed 25 mA;
if this value is exceeded, turn the bakeout off and then on again
when low pressure is restored. To control the heaters and to
monitor the high pressure limit during bakeout in automatic
mode, an automatically controlled relay may be used.
72/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Bakeout Operation
5
3 Bake the VacIon Plus pump for at least 24 hours. Longer bakeout
periods are recommended when the pump has been used with
heavy gas loads or when UHV pressure, 10-9 mbar or less is
desired.
4 As the pump and system cool down to room temperature, a drop
in pressure should be observed.
Note that the ion pump can be also baked when switched off, into an
external turbo pump through a bakeable isolation valve.
Bakeout of VacIon Pump with the Integral Heaters
1 The integral heaters are to be powered with the appropriate
voltage. (Please refer to the inscription on the heaters to apply
correct voltage).
2 The integral heaters are designed to provide a temperature of
250 °C to 300 °C when the pump is wrapped in a 3-fold
aluminium foil.
A three-layer foil wrapping is advisable and sufficient to achieve full
bakeout/regeneration if the standard European heaters (230 Vac) are operated
at 240 Vac, thus preventing overheating.
NOTE
3 Bakeout the VacIon pump for 24 hours. If the pump is used in
heavy gas load applications, it is recommended to bakeout the
pump for a longer period.
4 Wait until the pump cools down to room temperature and
recovers its base pressure before using it in the application.
Do not touch the pump during the heating and cooling phases. The high
temperature may cause serious damage.
WARNING!
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 73/88
5 Technical Information
Operating Procedure
Operating Procedure
1 Using a clean roughing pump, evacuate the system to a minimum
starting pressure as stated by the technical specification. 10-6
mbar or less is recommended. To reach this pressure the use of
an oil free turbo-molecular roughing pump is recommended.
2 When starting an ion pump, a slight increase in vacuum pressure
is normal as the internal components are heated and outgassed.
If possible leave the roughing pump connected to the system
while starting the ion pump. This will make the startup faster
and easier.
3 Connect the control unit to a suitable power source and switch
the power on.
4 Switch on high voltage to the pump and observe the current and
voltage. Fastest starting is obtained using a high applied voltage,
7 kilovolts for example. The applied voltage may be reduced later
to optimize pumping and achieve the lowest vacuum pressure.
5 If started at 10-4 mbar, the voltage will start at approximately
800 volts and increase to full voltage as the pump starts
operating. The current will start at several milliamps and slowly
decrease to microamps or nanoamps as low vacuum pressure is
achieved.
6 When first starting the pump, if the voltage decreases instead of
increasing, reduce the vacuum pressure then start the ion pump
again.
7 When the pump reaches its full operating voltage, you may close
the roughing valve.
8 If the pump does not start after 30 minutes of pumping, see the
section PUMP TROUBLESHOOTING.
9 Once the pump reaches its base pressure with stable voltage and
current, the ion pump may be baked at high temperature if
required per the section titled BAKEOUT OPERATION.
10 To stop the ion pump, simply switch off the high voltage. The
pump surfaces will continue to pump for a few minutes
depending on the system pressure.
11 When venting the pump use clean, dry nitrogen. This will avoid
water absorption on the pump surfaces and make subsequent
pump downs easier.
74/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Maintenance
5
Maintenance
The high voltage present in the high voltage cable which connects the control
unit to the ion pump, can cause severe injury or death. Before mounting the
high voltage connector of the cable on the pump high voltage feedthrough, or
before removing it, be sure the main power is removed from the control unit.
Before removing the high voltage connector of the cable from the control unit,
be sure the main power is removed from the control unit. Wait at least 10
seconds after removing the main power from the control unit, to allow
capacitors to discharge completely.
WARNING!
VacIon Plus pumps are maintenance free. In case of life time expiry
or premature failure of the pump, please contact your nearest Agilent
sales/service office for repair.
The VacIon Plus 150 pumps are designed with ex-changeable high-
voltage feedthrough and pump elements.
Exchange of Pumping Element
The high voltage feedthrough contains a ceramic insulator that can be damaged
if excessive force is applied in torque, in bending, or in tension. When installing
or removing the feedthrough, the applied torque should not exceed 2 Nm. Rotate
the feedthrough gently when making the threaded connection to avoid excessive
force.
CAUTION!
1 Remove the internal high-voltage connections.
2 Remove the element holding brackets.
3 Pull the element out of the pocket and remove it through the inlet
flange.
4 Clean the internal part of the pump with the appropriate solvent.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 75/88
5 Technical Information
Maintenance
5 Put in the new element through the inlet flange. Make sure that
the element is fixed at its rear by the two press-formed buttons of
the element pocket, and at its front by the element holding clamp.
6 Reattach the high-voltage connections and test for continuity
from the feedthrough.
7 Place a blank flange over the pump inlet and leak-check it with a
leak detector.
8 Bake out the blanked off pump while it is operating, and leave it
to cool down; then verify that the base pressure is below
10-9 mbar.
Exchange of the High Voltage Feedthrough
1 Remove the 6 bolts of the Mini-ConFlat flange connection (see
Fig. 24).
2 Remove the cable connection disc.
3 Gently relieve the feedthrough from the metal gasket connection
and turn the feedthrough counterclockwise until it is completely
detached from the internal high voltage threaded connection.
4 Replace the feedthrough and the copper gasket, making sure that
the feedthrough is connected to the pumps internal connector.
Check that the feedthrough is not shorted.
5 Bake out the pump while it is operating, and leave it to cool
down; then verify that the base pressure is below 10-9 mbar.
Figure 23 High voltage feedthrough connection
76/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Maintenance
5
L
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 77/88
eakage Current Check
If the pump current reading is to be used as pressure measurement,
check the pump leakage currents as follows:
1 Turn off the pump control unit.
2 Remove the pump magnet.
3 Turn on the pump control unit and wait for current stabilization.
The current reading should not be higher than 150 nA. Make sure
that the control unit and the high voltage cable leakage current is
negligible.
4 If leakage current comes from the pump, perform "high-potting"
per the following procedure; then recheck the pump and install
the magnet.
5 If it is not possible to "high-pot" the pump, the pressure reading is
biased by the leakage current value.
5 Technical Information
Maintenance
High-potting
A constant pump current when no vacuum leak exists is often caused
by field emission currents which prevent the use of the pump current
as a UHV pressure indicator. To reduce this field emission current,
"high-potting" should be performed. "High-potting" is the term used to
describe the application of higher than normal operating voltage
(8 kV AC, max. 50 mA) for the purpose of burning off "whiskers"
(sharp edges) on the pump cathode.
High-potting should be done under vacuum and preferably without
pump magnets installed (to reduce the drawn current).
High-potting must be done carefully and in voltage steps since
uncontrolled arcing inside the pump can cause permanent damage.
Therefore slowly increase the applied voltage and watch the current
meter for indication of arcing inside the pump as whiskers are
burned away. If arcing occurs wait at this voltage until the current is
stable again. Then slowly increase voltage again in steps up to a
maximum voltage. The maximum voltage may be applied to the pump
for a period of about 15 seconds. This should effectively remove any
"whiskers" or sharp edges on the pump cathode.
Voltages developed in the High Potter power sup-ply are potentially lethal.
Use caution during operation and ensure correct grounding connection.
WARNING!
78/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
Pump Troubleshooting
5
Pump Troubleshooting
Tab. 1
Symptom Possible cause Correction procedure
1) – Slow starting Starting vacuum pressure too high. Reduce pressure to 10-5 mbar
recommended 10-4 mbar
minimum.
2) – Slow starting
(more than 30 minutes).
Air leaks which limit pressure to above 10-6
mbar and cause longer starting time.
Leak check the vacuum system
with a helium leak detector.
3) – Slow pump-down
due to long exposure of
viton parts to air.
Viton releases considerable gas after long
exposure to air. (A bell-jar system which
reached 2 x 10-8 mbar in 24 hours after 30
minutes air exposure, will only reach 1 x 10-7
mbar in 24 hours after 20 hours air exposure)
With the system under vacuum,
pump for several days, or heat
to 100-150 °C for up to 15
hours.
4) – Slow pump-down
due to absorption of
vapours on pump and
system walls.
Vapours and gases admitted to a system are
absorbed on the walls of the system and
pump. Subsequent reduction in pressure
depends on the rate of depletion of this
vapour. Heavy hydrocarbons are most
troublesome because of their relative low
vapour pressure and they are very difficult to
remove, even by baking.
Bake the system walls, thereby
accelerating the desorption
process. Baking mobilizes the
vapours so they can be cracked
and pumped by discharge (see
section “BAKEOUT
OPERATION”).
5) – Slow starting or
slow pump-down.
High voltage feedthrough is leaking. Replace the feedthrough.
6) – Current higher than
expected at any given
pressure.
Ion pump leakage current causing higher
pressure reading.
Highpot the pump.
7) – System fails to
achieve desired UHV
vacuum pressure.
System not fully baked. Water vapor limits
base pressure.
Bake the system walls, thereby
accelerating the desorption
process. Baking mobilizes the
vapours so they can be cracked
and pumped by discharge (see
section “BAKEOUT
OPERATION”).
8) – System fails to
achieve desired UHV
vacuum pressure.
System not appropriately cleaned for UHV.
Excessive outgassing from walls limits base
pressure.
Clean all components for UHV
and bake the system again.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 79/88
5 Technical Information
VacIon Plus Pump Replacement Parts and Accessories
VacIon Plus Pump Replacement Parts and Accessories
Tab. 2
Part number
Diode Noble Diode StarCell
Basic pump 919-1510 919-1520 919-1540
Pump with options
919-1511
919-1512
919-1511
919-1512
919-1511
919-1512
919-1511
919-1512
919-1550 919-1550 919-1550 919-1550
919-1551
919-1552
919-1551
919-1552
919-1551
919-1552
919-1551
919-1552
Cables and accessories
HV rad. Resist. cable, with
interlock
929-0705
Replacement parts
HV Feedthrough 959-5125
Pumping elements (q.ty 2 needed) 919-9040 919-9045 919-9030
Heaters (power: 480 W)
(120V)
(220V)
919-0073,
919-0072
Copper gasket for 6” inlet flange
(10-pack, individually sealed)
FG-0600-CI
Nut and bolt set for 6” inlet flange FB-0450-C12
Copper gasket for 1 1/3" CF
feedthrough flange
(10-pack, individually sealed)
FG-0133-CI
Nut and bolt set for 1 1/3" CF
feedthrough flange, 8-32 x 3/4
(25-pack)
FB-0133-C
For a complete overview of Agilent's extensive vacuum product line,
please refer to the Agilent Vacuum Catalogue.
80/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Technical Information
VacIon Plus Pump Controllers
5
VacIon Plus Pump Controllers
The following controller series are available to supply the VacIon
Plus pumps:
MiniVac
Dual
4UHV
Please refer to the Agilent Vacuum Catalogue to choose the correct
controller for each pump.
VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E) 81/88
5 Technical Information
VacIon Plus Pump Controllers
82/88 VacIon Plus 150 pumps Instruction Manual / 87-900-102-01 (E)
Request f or Return Form
Sales and Service Offices
United States
Agilent Technologies
Vacuum Products Division
121 Hartwell Avenue
Lexington, MA 02421 - USA
Tel.: +1 781 861 7200
Fax: +1 781 860 5437
Toll-Free: +1 800 882 7426
India
Agilent Technologies India Pvt. Ltd.
Vacuum Product Division
G01. Prime corporate Park,
230/231, Sahar Road, Opp. Blue Dart Centre,
Andheri (East), Mumbai – 400 099.India
Tel: +91 22 30648287/8200
Fax: +91 22 30648250
Toll Free: 1800 113037
Southeast Asia
Agilent Technologies Sales Sdn Bhd
Vacuum Products Division
Unit 201, Level 2 uptown 2,
2 Jalan SS21/37, Damansara Uptown
47400 Petaling Jaya,
Selangor, Malaysia
Tel : +603 7712 6106
Fax: +603 6733 8121
Benelux
Agilent Technologies Netherlands B.V.
Vacuum Products Division
Herculesweg 8
4338 PL Middelburg
The Netherlands
Tel.: +31 118 671570
Fax: +31 118 671569
Toll-Free: 00 800 234 234 00
Italy
Agilent Technologies Italia S.p.A.
Vacuum Products Division
Via F.lli Varian, 54
10040 Leini, (Torino) - Italy
Tel.: +39 011 997 9111 Fax: +39 011 997 9350
Toll-Free: 00 800 234 234 00
Taiwan
Agilent Technologies Taiwan Limited
Vacuum Products Division (3F)
20 Kao-Shuang Rd.,
Pin-Chen City, 324
Taoyuan Hsien , Taiwan, R.O.C.
Tel. +886 34959281
Toll Free: 0800 051 342
Canada
Central coordination through: Agilent Technologies
Vacuum Products Division
121 Hartwell Avenue
Lexington, MA 02421 - USA
Tel.: +1 781 861 7200
Fax: +1 781 860 5437
Toll-Free: +1 800 882 7426
Japan
Agilent Technologies Japan, Ltd.
Vacuum Products Division
8th Floor Sumitomo Shibaura Building
4-16-36 Shibaura Minato-ku Tokyo 108-0023 - Japan
Tel.: +81 3 5232 1253
Fax: +81 3 5232 1710
Toll-Free: 0120 655 040
UK and Ireland
Agilent Technologies UK, Ltd.
Vacuum Products Division
6 Mead Road Oxford Industrial Park
Yarnton, Oxford OX5 1QU – UK
Tel.: +44 (0) 1865 291570
Fax: +44 (0) 1865 291571
Toll free: 00 800 234 234 00
China
Agilent Technologies (China) Co. Ltd
Vacuum Products Division
No.3, Wang Jing Bei Lu,
Chao Yang District,
Beijing, 100102
China
Tel.: +86 (10) 6439 7718
Toll-Free: 800 820 6556
Korea
Agilent Technologies Korea, Ltd.
Vacuum Products Division
Shinsa 2nd Bldg. 2F 966-5 Daechi-dong
Kangnam-gu, Seoul
Korea 135-280
Tel.: +82 2 3452 2452
Fax: +82 2 3452 2451
Toll-Free: 080 222 2452
Other Countries
Agilent Technologies Italia S.p.A.
Vacuum Products Division
Via F.lli Varian, 54 10040 Leini, (Torino) -
Italy
Tel.: +39 011 997 9111
Fax: +39 011 997 9350
Toll-Free: 00 800 234 234 00
France
Agilent Technologies France
Vacuum Products Division
7 Avenue des Tropiques
Z.A. de Courtaboeuf - B.P. 12
91941 Les Ulis cedex - France
Tel.: +33 (0) 1 69 86 38 84
Fax: +33 (0) 1 69 86 29 88
Toll free: 00 800 234 234 00
Mexico
Agilent Technologies
Vacuum Products Division
Concepcion Beistegui No 109 Col Del Valle
C.P. 03100 – Mexico, D.F.
Tel.: +52 5 523 9465
Fax: +52 5 523 9472
Customer Support & Service
NORTH AMERICA:
Toll Free: 800 882 7426, Option 3
vpl-ra@agilent.com
EUROPE:
Toll Free: 00 800 234 234 00
vpt-customercare@agilent.com
PACIFIC RIM:
please visit our website for individual office
information http://www.agilent.com
Worldwide Web Site, Catalog and Order
On-line:
www.agilent.com
Representative in most countries
12/10
Germany and Austria
Agilent Technologies
Vacuum Products Division Alsfelder Strasse 6 Postfach 11
14 35
64289 Darmstadt – Germany
Tel.: +49 (0) 6151 703 353
Fax: +49 (0) 6151 703 302
Toll free: 00 800 234 234 00
Singapore
Agilent Technologies Singapore Pte. Ltd,
Vacuum Products Division
Agilent Technologies Building,
1 Yishun Avenue 7,
Singapore 768923
Tel : (65) 6215 8045
Fax : (65) 6754 0574
© Agilent Technologies, Inc. 2011
Printed in ITALY
05/2011
Publication Number: 87-900-102-01 (E)