LC-4000 Series

Integrated HPLC and UHPLC systems

A quick brief - LC-4000 Series
The LC-4000 Series is the latest in a long history of innovative
HPLC systems developed by JASCO reaching all the way back to
the start of commercial HPLC in the early 1970s.
The concept of the integrated LC-4000 series HPLC provides
key separation platforms at 50MPa, 70MPa and 130MPa which
correspond to conventional HPLC, the increasingly popular
Rapid Analysis (RHPLC) and sub 2um UHPLC, respectively.
Each platform is supplied with a dedicated pump and
autosampler matched to the operating pressure and all three
platforms share common detectors optimized for high-speed
100Hz acquisition and the narrow peak shapes common to
both RHPLC and UHPLC.

Advances in Solvent Delivery
For over two decades JA SCO
analy tical HPLC pumps have
employed an asymmetric twinpiston solvent deliver y system
SSQD (Slow Suction, Quick Delivery)
which provides significantly better
flow and pressure profiles than the

2

commonly used conventional twinpiston reciprocating design. In the
LC-4000 series, SSQD has been
re-developed, with a completely
new solvent delivery mechanism
ExReFT (Extremely Reliable Flow
Technology) offering the highest

stability in solvent delivery across
the entire analytical flow rate range
used in the PU-4180/85 RHPLC and
PU-4280/85 UHPLC pump models.

Pioneering Optical Design

New Technologies for Method Development

As a pioneer in optical spectroscopy, JASCO has access
to optical designs that few can match. Adopting designs
from some of our more powerful spectrometers we
have developed a range of HPLC optical detectors with
unrivalled performance, like the new class leading FP4020 fluorescence detector with S/N of over 2300:1, the
latest high-speed refractive index detector with micro cell
optimized for RHPLC and UHPLC and a completely revised
circular dichroismdetector for chiral chromatography.

With the range of column technologies becoming ever more
diverse - the traditional 5um fully porous particle is finally
losing ground to some of the newer technologies such as the
new superficially porous solid core (Coreshell) and the wide
variety of sub-2um particles, the requirements for method
development are becoming more demanding. The LC-4000
series has a number of newly developed accessories to help
you make better separations faster, with accessories such
as, flexible and stackable column ovens in a range of sizes to
accommodate analytical to preparative columns and multiple
columns at every scale and with options for integrated 6 or
10 solvent and column selection valves.

C h r o m N A V, t h e c r o s s - p l a t f o r m
chromatography data system, is evolving
too. Version 2.0 now offers many new
features and a great new look to make
working with the LC-4000 HPLC system
an even better user experience.

Compact and Easy to Use
Despite the extra power delivered by the LC-4000 series
HPLC, the standard footprint is only 300 mm wide, a standard
system requires only a very small amount of bench space.
For easy user-maintenance, all LC-4000 instruments feature
front-access for replacing consumables such as pump
components and light sources, even the autosampler
consumables such as the syringe parts are easy to access.
For those users who require front-panel control, the LC4000 series retains the popular keypad and display (with a
backlit LED display as standard). Keypad control can still be
configured for convenient override when under PC control.

Each LC-4000 module
is designed for easy
maintenance.
Front-panel access to light
source and consumables

Front panel control is included as standard

The LC-4000 Series pump modules requires smaller bench
space than ordinary HPLC pumps. An integrated solvent
delivery system can be incorporated into a single pump
module.

3

System Configurations

The LC-4000 series HPLC analytical and preparative HPLC systems
have a number of common features
300mm wide footprint
Each system component is completely stackable into a modest height
The system is fully integrated with complete system monitoring
All user replaceable components are accessible through the front panel
Vertical leak/drain system to channel away solvent
Integrated earthquake securing option
All detectors include 100Hz acquistion with interchangeable flowcells

Conventional HPLC and RHPLC Analytical Systems up to 70MPa
The LC-4000 Series 70MPa system has been designed to future-proof your HPLC
requirements. The LC-4000 70MPa system can be used with conventional 3 and 5µm
particle size analytical columns at typical lower pressures around 10 to 20MPa (1500
to 3000psi) and can also be used with smaller particles such as shorter length UHPLC
columns and superficially porous (SPP) or Coreshell that require slightly higher
pressure solvent delivery to provide optimum linear velocity through the column.

UHPLC Analytical Systems at 130MPa
Ten years after the new segment of the HPLC market, UHPLC has found a niche for
those that need to run a large number of samplers or get results quickly. Pioneering
column technology is pushing the particle size even smaller. The near 20,000 psi
operation of the LC-4000 130MPa is designed to take advantage of the separation
efficiency of the very small particles couple with longer and narrower columns. The
LC-4000 130MPa system incorporates many of the same features as the 70MPa
system, but with materials designed to withstand higher pressures.

Preparative HPLC Systems
The LC-4000 preparative HPLC systems includes three platform options
Up to 20mL/min for columns up to 21.2mm ID
Up to 50mL/min for columns up to 3cm ID
Up to 120mL/min for columns up to 5cm ID
Large column oven for multiple preparative columns and optional automatic column
switching
Preparative detector options including high flow short path length flow-cells
Open-bed fraction collector with sample trays for tubes, plates or large containers
ChromNAV-FC for fraction control using time, threshold or slope from up to 4 different
signals

Semi-prep HPLC LC-4000 system

4

LC-4000 Series Extensive Range of Modules
The LC-4000 Series, enables multiple system design based on the following modules. This concept can be used
to create many different system configurations to match your application.
The extensive range of detectors based on JASCO's spectroscopy technology is a key factor in providing a
solution to virtually any application.

PUMPS

HPLC, RHPLC
PU-4185/80
PU-4185B

UHPLC
PU-4285
PU-4285B

PREP
PU-4086/87
PU-4086B

DETECTORS

UV/Vis visible
UV-4070/75

PDA
MD-4010/15/17

Refractive Index
RI-4030/35

AUTOSAMPLERS

Fluorescence
FP-4020/25

Circular Dichroism
CD-4095

Optical rotation
OR-4090

COLUMN OVENS
CO - 4 0 6 0 high
capacity

CO-4065 Prep

RO-4068
Reaction oven
CO- 4061/62
compact
HPLC
AS-4050/50i

RHPLC
AS-4150
UHPLC
AS-4250
Preparative
AS-4058
5

The LC-4000 Series next generation HPLC pumps with flexible expandability
The LC-4000 series HPLC pumps include expansion bays that allow a simple pump module to be
transformed into a powerful solvent delivery system. Accessory modules can be installed in the LC-4000
series pump for use with virtually any solvent delivery requirement.

Upgrade and Expansion
The unique expansion bays in the pump platform can be fitted with plug-in
modules to quickly and easily transform a simple isocratic pumping system
into a sophisticated multi-solvent delivery system by installing a second pump,
degasser, quaternary gradient unit, solvent selection valve or dynamic mixer
without increasing the footprint.
Adding a second pump module enables an isocratic pump to be upgraded to a
dual pump system for either binary gradient or two isocratic flows.
PU-4180
Analytical pump platform

A powerful configuration of the single-unit design is the quaternary gradient
solvent delivery system including degasser, with a post-column reagent pump
configured in a compact chassis only 300mm wide.

Excellent flow rate repeatability

PU-4180 Flow rate accuracy
(1.0mL/min, n=10)
Mobile phase:
Flow Rate:		
Column: 		
Column temerature:
Detection wave length:
Sample:		
		
Injection:		

Acetonitrile/Water (80/20)
1.0mL/min
CrestPak C18S (4.6mmI.D. x 150mm, 5µm)
40 deg C
260nm, Response; 1sec
Polycyclic aromatic hydrocarbon (10µg/mL each)
(Naphthalene, Fluorene, Anthracene, Pyrene, Chrysene, Benzo[a]pyrene)
10µL

Retention time reproducibility (n=10)

PU-4185
Semi micro pump platform

#
1
2
3
4
5
6
7
8
9
10
Average
Standard diviation(n-1)
CV%(n-1)

Naphthalene
3.1400
3.1417
3.1417
3.1400
3.1400
3.1400
3.1416
3.1400
3.1400
3.1400
3.1405
0.001
0.025

Fluorene
3.9367
3.9683
3.9367
3.9367
3.9367
3.9367
3.9383
3.9367
3.9367
3.9367
3.9370
0.001
0.018

Anthracene
4.5000
4.5017
4.5000
4.5002
4.5000
4.5000
4.5016
4.5000
4.5000
4.5017
4.5005
0.001
0.017

Pyrene
5.6283
5.6317
5.6300
5.6283
5.6283
5.6300
5.6317
5.6300
5.6300
5.6301
5.6299
0.001
0.022

Chrysene
6.4934
6.4950
6.4933
6.4950
6.4950
6.4933
6.4967
6.4950
6.4949
6.4950
6.4947
0.001
0.016

Benzo[a]pyrene
9.4267
9.4300
9.4285
9.4283
9.4283
9.4283
9.4317
9.4317
9.4317
9.4317
9.4297
0.002
0.020

Eliminating pulsation

6

The problems caused by pressure
variations in solvent delivery when
switching between the two pump
heads has been improved with a
new design that precisely controls
the plunger drive. The LC- 4 0 0 0
series analytical pumps offer flow
rate precision with RSDs of better

than 0.05% (difficult to achieve with
commercially available pumps). The
PU-4180 and 4185 pumps operate
at pressures up to 70 MPa, designed
mainly for use with RHPLC for coreshell columns or similar micro
packing materials as well as working
at pressures commonly used for

conventional HPLC.
The PU- 4280 and PU- 4285 UHPLC
pumps provide precise flow at
pressures up to 130 MPa; for columns
with particle sizes of less than 2 µm.
The LC-4000 series UHPLC system can
also be used for rapid analysis HPLC
(RHPLC) and conventional HPLC.

Pump Configurations
The PU-4180/85 and PU-4280/85 analytical pumps can
be configured for both single pump gradient and multiple
pump gradient operation. The single pump quaternary
gradient module offers excellent gradient formation for
up to four solvents. The newly designed 4-way solvent
valve for switching the four solvents is synchronized
to the operation of the pump motor control - variable
cycle adaption optimizes delivery according to gradient
resolution and flow rate. This highly efficient quaternary
gradient mode can be applied to a wide flow-rate range
from low to high flow rates.
Two and three pump binary or ternary gradient modules
with optional built-in degasser are available for fast cycle
high-resolution gradient formation; the response time is
superior to the single pump quaternary gradient system.

PU-4185-Binary pump
Two pump binary gradient

Solvent Mixing
Multi-pump gradient mixing is done using a variable speed
dynamic mixer with a range of a small volume mixing
chambers incorporating a micro-stirrer to provide the
smallest possible dead volume, offering higher mixing
efficiencies for all HPLC solvents.
The optional in-line TERA Mixer is a turbulent-flow solvent
mixer uniquely designed to offer high efficiency mixing
with minimal dead volume.

PU-4180 Quaternary gradient mixing performance
(1.0mL/min, n=10)

Solvent Degassing
The in-line vacuum degasser module uses a new Teflon
AF membrane with a hold-up volume of only 400µL (per
solvent line); this offers very quick solvent replacement
with minimal waste. Despite its small size, the new
degasser offers more effective and stable degassing
than larger previous generation models. A degasser for
preparative-HPLC is also available.

Mobile phase:
A; Water, B; Acetonitrile
Gradient:		
(A/B) (65/35) → (5/95) Linear Gradient
Flow rate:		
1.0mL/min
Column:		
CrestPak C18S (4.6mmI.D. x 150mm, 5µm)
Column temerature:
40 deg C
Detection wave length:
254nm, Response; 1sec
Sample:		
Alkylphenon samles 10µg/mL (Benzohenon 5µg/mL)
		
(Acetanilide, Acetophenone, Propiophenone,
Butyrophenone, Benzophenone, Valerophenone, Hexanophenone, 		
						
Heptanophenone,Octanophenone)
Injection:		
10µL

Retention time reproducibility (n=10)
#
1
2
3
4
5
6
7
8
9
10
Average
Standard diviation (n-1)
CV% (n-1)

Acetanilide
3.117
3.117
3.113
3.117
3.117
3.117
3.117
3.117
3.117
3.117
3.116
0.001
0.034

Acetophenone
5.617
5.617
5.617
5.617
5.617
5.620
5.620
5.617
5.617
5.620
5.618
0.002
0.029

Propiophenone
7.590
7.590
7.590
7.590
7.590
7.593
7.590
7.590
7.593
7.593
7.591
0.002
0.021

Butyrophenone
9.303
9.307
9.307
9.307
9.307
9.310
9.307
9.310
9.310
9.313
9.308
0.003
0.030

Benzophenone
9.950
9.950
9.950
9.953
9.953
9.953
9.953
9.953
9.957
9.957
9.953
0.002
0.025

Valerophenone
10.980
10.983
10.983
10.983
10.983
10.987
10.983
10.987
10.987
10.990
10.985
0.003
0.026

Hexanophemone
12.583
12.587
12.587
12.587
12.587
12.587
12.587
12.590
12.593
12.593
12.588
0.003
0.026

Heptanophenone
14.067
14.067
14.067
14.067
14.067
14.067
14.067
14.067
14.073
14.073
14.070
0.002
0.016

Octanophenone
15.430
15.433
15.433
15.433
15.433
15.433
15.433
15.433
15.437
15.437
15.434
0.002
0.012

PU-4180
(compact quaternary gradient pump with
micro-volume degasser ) *Option

Preparative and Semi-preparative Pumps
The LC-4000 Preparative-HPLC
System includes pumps at three
scales. The PU-4086 pump offers
flow rates up to 20 mL/min, the
PU-4087 pump up to 50 mL/min
and the PU-4088 up to 120mL/
min. Both single and multi-pump
gradient modes can be used with
the PU-4086 pump platform whilst
the PU-4087 and PU-4088 are
used in multi-pump mode only.
The new MK2 technology used in
the PU-4086 and PU-4087 motor
driver control has been increased

so that the maximum pressure
can be used at higher flow-rates
for all pump models. Solvents can
be delivered at higher flow rates
and higher pressures for use with
longer preparative HPLC columns
that generate more back pressure
and for faster and more productive
preparative separations.
Dynamic mixers, solvent selection
valves and the prep degasser can
be installed in the expansion bays
in both the PU-4086 and PU-4087
modules.

PU-4086 semi prep pump
Recycling inlet switching unit, Recycling valve
unit (Option)

7

A comprehensive range of high performance optical detectors
With JASCO’s long history of spectrometer development, you will find that each detector is well
designed; these next generation detectors employ the latest optical and electronic technologies for high
sensitivity, excellent stability and easy maintenance.

MD-4010, MD-4015
and MD-4017 PDA Detectors
The LC-4000 Series includes three
PDA detec tor s. T he MD - 4 017 is
a useful replacement for a single
wavelength detector, the MD-4015
is the workhor se for UV- visible
detection in the range 200 to 600nm
and the MD-4010 is designed for high
sensitivity and high spectral and time
resolution.
The MD-4010 PDA detector offers
the highest detection sensitivity,
equivalent to a single wavelength
detec tor. Making use of a 1024
element photometric diode array,
the measurement wavelength range
is from 190nm to 90 0nm with a

MD-4010 PDA

resolution of 1nm and fast spectrum
acquisition rate of up to 100 spectra/
sec. When used with the new
temperature controlled SP t ype
semi-micro flow cell the MD-4010
is well suited to the narrower peaks
typically found with UHPLC or RHPLC
separations. The MD-4010 is fitted
with an internal mercury lamp for
automatic wavelength calibration.
The MD-4015 PDA detector is the
routine detector for conventional
HPLC, RHPLC and UHPLC. The
sensitivity is almost equivalent to
that of the MD-4010 but only covers
the wavelength range from 200nm to
600nm using a 512 element PDA for
1nm resolution and with high speed
100Hz spectrum acquisition.

The MD-4017 detector has a price-point comparable
to a typical single wavelength detector and has been
designed for the user only interested in the UV region
from 200-400nm, the resolution is excellent at 1nm and
the maximum spectrum acquisition is 20 spectra per
second (20Hz), useful for peak widths as small as 1 to 2
seconds.

ChromNAV PDA data processing
included as standard

Front panel access for easy lamp
replacement

When used with ChromNAV 2.0 CDS the PDA detector
offers multi-functional analysis with a variety of
functions such as spectral analysis, library search for
peak identification, peak purity and multi-wavelength
quantitation.

UV- 4070/4075UV-Visible
Detectors
Using the advanced CzernyTu r n e r m o n o c h r o m a t o r d e s i g n
with optimization of the optical
sys tem, baseline change due to
refrac tive index is minimized.
Baseline fluctuation from changes
in environmental temperature
is reduced with a temperature controlled cell, baseline drift caused
by temperature fluctuation in the
light source is also reduced with a
temperature controlled light source.
The result is excellent stability with

UV-4070 UV Visible

UV spectra auto-scan

Automatic spectral scannning can be
controlled using ChromNAV, when a peak
exceeds a set threshold a scan is executed
automatically.

0.24

220000
200000

Intensity[µV]

the noise level +/- 0.2x10 -5 AU and
drift 1x10 -4 AU/h. Both detectors offer
simultaneous dual-wavelength data
acquisition and spectral scanning.
Us er maintenan c e c an b e d o n e
without removing the module from
the system. Front-panel access is
used for light source replacement and
automatic wavelength calibration can
be carried out using a mercury lamp.
Both detectors include high speed
data acquisition at up to 100Hz and
can be used for conventional HPLC
with the standard flow cell and for
UHPLC/RHPLC when used with the SP
type semi-micro cell.

0.2

Abs
0.1
100000

Threshold 50000

0
-0.01

Threshold 10000

200

400

600

800

900

Wavelength[nm]

8

Spectral measurement view

0
0.0

0.2

0.4

0.6

0.8

Retention Time[min]

10.0

12.0

14.0

UV detector Auto scan parameters

Fluorescence detector FP-4020/4025
The FP-4020 fluorescence detector has been designed for the pursuit of extreme
sensitivity. Re-designing the optics and electronics from the ground up, as well
as removing the effects of ambient temperature with a new temperatureregulated cell body has led to the highest sensitivity with stable detection. The
new axially irradiated cell design enables the very highest sensitivity detection.
These changes have resulted in a giant leap in sensitivity for the FP-4020 to
achieve a signal noise ratio of 2300:1 or higher (based on the water Raman
peak). The FP-4025 is an excellent routine fluorescence detector that has a
slightly lower specification with a water Raman S/N of 1400:1 .
Both detectors can be used for simultaneous detection at two wavelength pairs;
excitation and emission wavelengths can be changed during the acquisition using
a time program and automatic spectra scanning can be triggered manually, by
time or automatically by threshold. A built-in mercury lamp enables automatic
wavelength correction and the front-access design allows easy replacement of
the flow cell assembly and light source.

FP-4020 Fluorescence

26000

20000
Intensity[µV]

Anthracene 1.4 fg (S/N=3)
(Anthracene 100 fg Injection)

10000

0.0

1.0

2.0

0

3.0

4.0

5.0

Retention Time[min]

High sensitivity measurement of Anthracene

Dual wavelength simultaneous measurement

Refractive Index detector RI-4030/4035

RI-4030 Refractive Index

The Refractive Index detectors include the optical system mounted in a precisely
temperature-regulated housing to eliminate the influence of changes in ambient
temperature. The RI-4030 and 4035 detectors are fully stable just one hour
after powering on. The RI-4030 can be used for both analytical and preparative
scales. In preparative mode the maximum flow rate is 120 ml/min. The RI-4035
is a dedicated semi-micro RI detector with a micro-volume flow cell designed for
use with semi-micro GPC analysis using GPC columns with a fine pore size. The
100Hz data acquisition rate makes the RI-4035 suitable for RHPLC and UHPLC
separations. The LED light source in both detectors has a long life time (over
10,000 hours) for low maintenance operation. Both the RI-4030 and RI-4035
detectors can be fitted with an optional pressure release safety-valve to prevent
damage to the flow cell in the event blockage.

Circular Dichroism detector CD-4095

CD-4095 Circular Dichroism

The CD-4095 is a unique HPLC detector for measuring circular dichroism
developed from the market leading J-Series circular dichroism
spectropolarimeter. Optically active compounds with a chromophore close to a
chiral center may absorb circularly polarized light, which can be detected with
excellent sensitivity and selectivity. When chiral compounds are measured using
a UV-Visible detector, the d- and l- enantiomers cannot be distinguished even
when separated by a chiral column. The CD-4095 circular dichroism detector
makes it possible to positively distinguish between the d- and l- enantiomers
by measuring the positive and negative peaks. It is also possible to quantify the
d- and l-forms even when they are not separated, a calibration curve based on
the g-factor can be used to determine the enantiomeric ratio. The ChromNAV
CDS can be used to simultaneously measure the CD, g-factor data and UV-visible
signal. The CD spectra of eluted compounds can be measured by triggering
scanning manually, by timed program or automatically by threshold.

Chiral detector OR-4090

OR-4090 Optical Rotation

The OR-4090 Chiral detector measures the angle of rotation of plane polarized
light caused by optical active isomers and is useful for chiral compounds with no
absorption. The OR-4090 uses a high intensity Hg-Xe arc lamp (150W) providing
a strong output in the UV/Visible region where the largest optical rotations offer
the highest sensitivity. The wide wavelength range from 350-900nm also allows
measurement of a wide variety of samples. One of the key benefits of optical
rotation detection is that it is unaffected by changes in solvent and can be used
with gradient elution, unlike an RI detector.

9

Autosamplers - AS-4050/4050i/4150/4250/4058
The LC-4000 Series autosamplers provide a range of options to suit any sample
injection requirements, from the AS-4050 entry-level model, to the AS-4250 UHPLC
model or the AS-4058 preparative scale autosampler.
The AS-4050 is a 60 sample autosampler for use with conventional HPLC with options
for analytical and semi-preparative injection volumes. The AS-4150 and AS-4250 are
high-capacity, high performance autosamplers designed for RHPLC (up to 70MPa)and
UHPLC (up to 130MPa), respectively, with a minimum injection cycle-time of only 30
seconds. Peltier cooling is available for all autosamplers and light protection for photolabile samples is available on the higher level models.

AS-4050 60 sample HPLC
autosampler

Pre-injection sample preparation is available on the high performance autosamplers
with programmable options for sample dilution and pre-column derivitization.
The injection mode can be switched between full loop injection and partial loop
injections.
Up to 5 different washing solvents can be used to achieve excellent carry-over at less
than 0.005%.
An optional degasser can be installed for degassing up to 5 flushing solvents.
The AS-4058 is the LC-4000 Series largest volume autosampler with injection volumes
up to 5mL as standard (up to 10mL as an option). The AS-4058 is typically used in
combination with either a PU-4087 or PU-4088 pump for preparative separations.
All LC-4000 series autosamplers share the same simple maintenance features, with
consumable parts such as injection ports, rotor seals, and loops accessible from the
front panel for easy replacement.

AS-4250 High capacity UHPLC
autosampler

Simple replacement
of syringe in AS-4050

Designed for easy maintenance
Metering Tube

Air-drain
Valve

Waste
Bottle

Needle

Flushing
Port

Solvent-stop
Valve
NC
Air Pump

IN

3way
Valve

5

Sample
Vial

4

Injection
Port
Sample
loop

NO

Column

1

2
Injection
Valve

Flushing
Solvent

Metering
Pump

AS-4050/4150/4250/4058 flow diagram (loop injection)
10

To Detector

3

6

HPLC Pump

Column Ovens
The LC-4000 Series column ovens have
been designed to provide excellent
utility for a wide range of different user
requirements. For simple requirements
the stackable CO-4061 can be used with
column lengths up to 200mm (or up to
300mm with an optional extension).
CO-4061

For RHPLC and UHPLC the CO-4062 has
capacity for up to six 150 mm length
columns, with optional one or t wo
internally mounted automatic switching
valves for 2 or 6 column selection.
For GPC and larger preparative columns,
select either of the two high-capacity CO4060 or CO-4065 ovens. The CO-4060
can accommodate columns with lengths
up to 300mm or larger ID preparative

columns. The CO-4065 can accommodate
up to 10 columns with lengths up to
300mm and a pair of column selection
valves is included as standard.
All column ovens have built- in
electronic cooling for controlling
column temperature at sub-ambient
temperatures.
The CO-4062 and CO-4061 block-heater
type ovens can be fitted with an optional
micro-volume pre-heating unit for preheating the mobile phase before the
column to improve resolution and peak
shape.
The RO-4068 reaction oven is used for
high temperature post-column sample
processing with capacity for columns and
reaction coils.

CO-4060
600000

WIth Pre coulmn stabilizer —
WIthout Pre coulmn stabilizer —

Temp : 25 degree C

Toluene

X-Presspak VC18-W
(3 mmI.D. x 50 mmL, 2 µm)
Water / Acetonitrile (Gradient)
1.0 mL/min
Gasoline Oxidization Products

Eluent
Flow Rate
Sample

400000

Intensity

Column

200000

Analysis conditions

0
0.0

0.4

0.8

1.2

1.6

2.0

600000

Temp : 40 degree C

Toluene

Intensity

400000

2.4

2.8

WIth Pre coulmn stabilizer —
WIthout Pre coulmn stabilizer —

Room Temp.

Oven Temp.

25˚C

200000

21˚C
0
0.0

0.4

0.8

1.2

1.6

2.0

2.4

2.8

Efficiency comparison of stabilizer

40˚C

Pre-Column
Stabilizer

Theoretical

Peak Width

tR(min)

With
Without
With
Without

56095
54747
46021
36402

0.019
0.019
0.020
0.023

1.893
1.887
1.799
1.851

Comparison of Toluene peaks

UniFinePak HPLC Columns
UniFinePak is a new type of column that uses a hybrid packing material. Compared
with traditional packing materials that use silica as the base material, UniFinePak
columns can be used with a wider pH range from alkali to acid. UniFinePak columns
offer an extended analysis range due to its chemical strength and resilience to high
pressure operation.
The entire range of UniFinePak columns uses a common packing material in a range
of particle sizes from semi-micro to analytical to semi preparative scales, which
simplifies method development when scaling-up separations from analytical to semipreparative .

Method Conversion Program

Optional Preparative HPLC fraction collection kit
The LC-4000 series preparative system fraction collection kit includes
Fraction collector
Flow divertor valve
ChromNav FC fraction control program for ChromNAV

11

ChromNAV 2.0 The 'PowerTool' for HPLC Analysis
ChromNAV 2.0 (and ChromNAV-CFR 2.0) are JASCO’s next generation CDS developed from the powerful
and easy-to-use ChromNAV 1.0 with a host of exciting new features.
With a customizable graphical-user-interface (GUI), the user can set-up the system to display only the
functions necessary for their application. This latest intuitive GUI allows the user to quickly learn the
operation and explore the extensive functionality for data processing.
ChromNAV 2.0 is a universal CDS which can be used with any type of separation - HPLC, UHPLC, RHPLC,
prep-HPLC, analytical SFC and prep-SFC. ChromNav can also satisfiy the demands of dedicated analysis
and multi-purpose systems.

Control method
The pump flow rate and gradient profile display is
flexible and can be overlaid with a chromatogram
for adjusting gradient conditions.

PDA analysis
PDA data processing is included as standard. Data is displayed
in 2D contour plot and 3D with simultaneous overlay of
spectra and chromatograms.Extract chromatograms at single
or multiple wavelengths for quantitation

ChromNAV 2.0 offers powerful system
control and data acquisition with
options for 2D (or up to 4D) gradient
control and multi-zone oven control
with control of up to 3 column ovens
for method development. Solvent and
column selection control is increased
to include up to 10 positions for both.
During acquisition, the run-time for
a chromatogram can be increased to
capture later eluting peaks which may
exceed the set run-time.
12

Previously acquired chromatograms
can be overlaid for visual comparison
with the data currently being
acquired, as a quick indicator to
t h e u s e r o f c u r r e nt s e p a r a t i o n
performance.

manually, by time or by threshold.
A newly added feature of ChromNAV
2.0 is the automatic message that can
be emailed to your PC or smartphone
to give an update on the progress of
the sequence.

Data acquisition and analysis for a
PDA detector are included as standard
in ChromNAV.

With the increasing requirements
for Green Energy, the LC-4000 series
has the facility not only to power off
and save lamp life at the end of the
run, but also a ‘soft-off’ function to
automatically power off a module at
the end of the sequence.

The fluorescence and UV-visible
detectors include spectral scanning
as standard, which can be triggered

ChromNav 2.0 standard features
•
•
•
•
•
•
•

Peak integration and peak identification
Peak grouping
Linear and non-linear quantitation
3D chromatogram analysis,
Spectral analysis for UV-visible, Fluorescence and PDA detectors
Style report generator
User calculation

All raw data is protected and saved; which can then be analyzed and re-analyzed, reported and saved with both the raw
data and with any data processing from the user’s analysis.
A comprehensive audit trail records the acquisition method along with a history of instrument performance indicators in
each sample data file, which provides the user a snap-shot of the condition of the system during the run and can warn about
requirements for impending maintenance.

ChromNAV 2.0 Optional Applications
•
•
•
•

ChromNAV CFR for 21 CFR Part 11 compliance and electronic registration of data
GPC molecular weight dispersion
Molecular weight dispersion of low molecular weight heparin
FUMI theory (Function of Mutual Information) for theoretical precision analysis

ChromNAV-GPC
Molecular weight distribution program ( Option )
ChromNAV GPC/SEC calculation program for GPC analysis
ChromNAV- FC Collection view window ( Option)
ChromNAV-FC fraction collector control for the CHF122SC fraction collector

ChromNAV-FC for preparative HPLC
The optional ChromNAV-FC fraction
collector control is available as
part of the Prep-HPLC kit and
provides ChromNav with the
capability to control the LC-4000
Series preparative HPLC system for
fraction collection.

or circular dichroism. Fraction
simulation is applied to an acquired
chromatogram to optimize the
fraction collection parameters and
the graphical display can be used
to locate peaks of interest in the
fraction collector tray.

ChromNAV-FC offers control for
two different capacity fraction
collectors, with sophisticated
fractionation control with several
methods for peak collection, such
as time, threshold and slope, and
can even be programmed to use
data collected simultaneously
from multiple detectors such
as UV-visible, refractive index

ChromNAV-FC also supports
additional functions for prep-HPLC
applications, including automatic
multi-channel fractions, fraction
repeat and fraction re-cycle.
*1: ChromNAV CFR is an optional
software package for ChromNAV.

13

Specifications
Pump
Model

PU-4185

Flow rate setting range 0.001 ~ 4.0 mL/min
Applicable flow rate
range
Maximum pressure
Flow rate accuracy

0.05 ~ 3.0 mL/min (~ 70 MPa)
~ 4.0 mL/min (~ 40 MPa)

PU-4180

PU-4285

0.001 ~ 10.0 mL/min

0.001 ~ 2.0 mL/min

0.5 ~ 6.0 mL/min (~ 70 MPa)

0.050 ~ 1.5 mL/min (~ 130 MPa)

~ 10.0 mL/min (~ 35 MPa)

1.501 ~ 2.0mL/min (~ 100 MPa)

70 MPa (~ 3.0 mL/min)

70 MPa ( ~ 6.0 mL/min)

130.0 MPa (~ 1.5 mL/min)

40 MPa (~ 4.0 mL/min)

35 MPa (~ 10.0 mL/min)

100.0 MPa (~ 2.0 mL/min)

+/- 1% or +/- 2 µL/min, whichever is larger.

+/- 1% or +/- 2 µL/min, whichever is larger.

+/- 1% or +/- 2 µL/min, whichever is larger.

(0.05 ~ 4.0 mL/min)

(0.5 ~ 10.0 mL/min)

(0.05 ~ 2.0 mL/min)

0.05% RSD or +/- 0.04min SD, whichever is larger. 0.05% RSD or +/- 0.04min SD, whichever is larger. 0.05 % RSD or +/- 0.04 min SD, whichever is
Flow rate precision

(0.05 ~ 4.0 mL/min)

(0.5 ~ 5.0 mL/min)

(measurement by chromatogram)

(measurement by chromatogram)

(measurement by chromatogram)

Dimensions, weight

300(W) x 470(D) x 150(H) mm, 13 kg

300(W) x 470(D) x 150(H) mm, 13 kg

300(W) x 470(D) x 150(H) mm, 14.5 kg

Power requirement

AC 100 ~ 240 V, 50/60 Hz, 80 VA

AC 100 ~ 240V, 50/60 Hz, 80 VA

AC 100 ~ 240V, 50/60 Hz, 60 VA

larger. (0.1 ~ 2.0 mL/min)

(option) temp. range: 4 ~ 40ºC
Gradient mode with low-pressure gradient unit
Mixing accuracy

Mixing precision

Binary pump

+/-0.6% (5 ~ 95%, 0.2 ~ 1.0 mL/min)
+/-1.2% (5 ~ 95%, ~ 4.0 mL/min)

+/-0.8% (5 ~ 95%, 0.5 ~ 5.0 mL/min)

+/-0.6% (5 ~ 95%, 0.2 ~ 1.0 mL/min)
+/-1.2% (5 ~ 95%, ~ 2.0 mL/min)

0.15% RSD or +/- 0.01min, whichever is larger.

0.25% RSD or +/-0.02min, whichever is larger.

0.15% RSD or +/-0.01min, whichever is larger.

(0.2 ~ 4.0 mL/min)

(0.5 ~ 5.0 mL/min)

(0.2 ~ 2.0 mL/min)

(measurement by chromatogram)

(measurement by chromatogram)

(measurement by chromatogram)

PU-4185-Binary

PU-4285-Binary

Gradient mode
Mixing accuracy

+/-0.4%

-

(5 ~ 95%, 0.2 ~ 4.0 mL/min)
0.15% RSD or +/-0.01 min, whichever is larger.

Mixing precision

(5 ~ 95%, 0.2 ~ 2.0 mL/min)
0.15% RSD or +/-0.01 min, whichever is larger.

(0.2 ~ 4.0 mL/min)

-

(0.2 ~ 2.0 mL/min)

(measurement by chromatogram)

(measurement by chromatogram)

Dimensions, weight

300(W) x 470(D) x 150(H) mm, 19.5 kg

-

300(W) x 470(D) x 150(H) mm, 22.5 kg

Power requirement

AC 100 ~ 240 V, 50/60 Hz, 130 VA

-

AC 100 ~ 240 V, 50/60 Hz, 110 VA

Model

PU-4086

Flow rate setting range 0.001 ~ 20.0 mL/min
Applicable flow rate

0.5 ~ 20.0 mL/min

range

(0 ~ 50 MPa)

PU-4087
0.01 ~ 50.0 mL/min
5.0 ~ 30.0 mL/min (~ 50 MPa)
30.01 ~ 40.0 mL/min (~ 40 MPa)
40.01 ~ 50.0 mL/min (~ 30 MPa)
50 MPa (~ 30.0 mL/min)

Maximum pressure

50 MPa

40 MPa (~ 40.0 mL/min)
30 MPa (~ 50.0 mL/min)

Flow rate accuracy

+/- 1% or +/- 10 µL/min, whichever is larger.

+/- 1% or +/- 50 µL/min, whichever is larger.

(0.5 ~ 20.0 mL/min)

(5.0 ~ 50.0 mL/min)

0.05% RSD or +/- 0.04min SD, whichever is larger. 0.1% RSD or +/- 0.04min SD, whichever is larger.
Flow rate precision

(0.5 ~ 20.0 mL/min)

(5 ~ 40 mL/min)

(measurement by chromatogram)

(measurement by chromatogram)

Dimensions, weight

300(W) x 470(D) x 150(H) mm, 14.5 kg

300(W) x 470(D) x 150(H) mm, 18 kg

Power requirement

AC 100 ~ 240 V, 50/60 Hz, 90 VA

AC 100 ~ 240 V, 50/60 Hz, 120 VA

Binary pump

PU-4086-Binary

Gradient
Mixing accuracy
Mixing precision

14

+/-0.4%

+/-0.75% (5 ~ 95%, 0.5 ~ 20.0 mL/min)
0.2% RSD or +/-0.03min, whichever is larger.
(0.5 ~ 20.0mL/min)

-

Dimensions, weight

300(W) x 470(D) x 150(H) mm, 21.5 kg

-

Power requirement

AC 100 ~ 240 V, 50/60 Hz, 100 VA

-

Model

MD-4010

MD-4015

MD-4017

Light source

D2 & WI lamp

D2 lamp

D2 lamp

Measurement range

190 ~ 900 nm

200 ~ 600 nm

200 ~ 400 nm

PDA element number

1024 ch

512 ch

512 ch

Slit width

1, 4, 8 nm

4 nm

4 nm

Data acquisition rate

Maximum 100 spectra/sec (full range)

Maximum 100 spectra/sec (full range)

Maximum 20 spectra/sec (full range)

Noise level

+/- 3.0 x 10-6 AU (specified conditions)

+/- 3.0 x 10-6 AU (specified conditions)

+/- 7.0 x 10-6 AU (specified conditions)

Drift

< 0.5 x 10-6 AU/h (specified conditions)

< 0.5 x 10-6 AU/h (specified conditions)

< 1.0 x 10-6 AU/h (specified conditions)

Linearity

2.0 AU or more (specified conditions)

2.0 AU or more (specified conditions)

2.0 AU or more (specified conditions)

Flow cell

Path length: 10 mm, front loading cassette type

Path length: 10 mm, front loading cassette type

Path length: 10 mm, front loading cassette type

PC communication

USB2.0

USB2.0

USB2.0

Dimensions, weight

300(W) x 470(D) x 150(H) mm, 14.5 kg

300(W) x 470(D) x 150(H) mm, 13.5 kg

300(W) x 470(D) x 150(H) mm, 13.5 kg

Power requirement

AC 100 ~ 240 V, 50/60 Hz, 180 VA

AC 100 ~ 240 V, 50/60 Hz, 150 VA

AC 100 ~ 240 V, 50/60 Hz, 120 VA

Model

UV-4070 / UV-4075

Light source

D2 lamp + WI lamp (UV-4070)

Wavelength range

190 ~ 900 nm (UV-4070)

Monochromator

Czerny-T urner mount

D2 lamp (UV-4075)
190 ~ 600 nm (UV-4075)

Spectral bandwidth

8 nm

Noise level

+/-0.2 x 10-5 AU (230 nm, 1.5 sec)

Drift

1 x 10-4 AU/h (230 nm, constant room temperature)

Data output

Maximum 100 Hz
Front loading cassette cell (temperature controlled), tapered.

Flow cell

Path length: 10 mm

Spectrum measurement
(On-flow measurement)

200 ~ 900 nm (D2 & WI), 200 & 370 nm (D2), 371 ~ 900 nm (WI) (UV-4070)
200 ~ 600 nm (D2) (UV-4075), applicable measurement range; fixed

T wo-wavelengths monitoring

Any two wavelengths in 190 ~ 370 nm, 371 ~ 700 nm, 701 ~ 900 nm (UV-4070)
Any two wavelengths in 190 ~ 370 nm, 371 ~ 600 nm (UV-4075)

Dimensions, weight

300(W) x 470(D) x 150(H) mm, 10 kg

Power requirement

AC100 ~ 240 V, 50/60 Hz, 175VA (UV-4070), 110VA (UV-4075)

Model

FP-4020 (high performance model) / FP-4025 (standard model)

Light source

Xenon short arc lamp

Wavelength setting range

200 ~ 900 nm (both EX and EM)

(measurement range)

220 ~ 700 nm (both EX and EM), 200 ~ 900 nm (both EX and EM, option)

Spectral bandwidth

EX: 20 nm, EM: 20, 40 nm (switchable)
Raman peak of water, S/N > 2300 (FP-4020)、 S/N > 1400 (FP-4025)

Sensitivity

(EX wavelength; 350 nm)

Data output

Maximum 100 Hz

Flow cell

Front loading cassette cell

T emperature control

Available setting: OFF, 4 ~ 40ºC, applicable range: [ambient -10ºC] ~ 40ºC (only for FP-4020 )

Spectrum measurement

Ex and Em spectrum measurement, spectrum output, output of subtracted spectrum

T wo-wavelength monitoring

T wo sets of Ex/Em wavelength setting, maximum wavelength difference; 200 nm or shorter.

Dimensions, weight

300(W) x 470(D) x 225(H) mm, 20 kg

Power requirement

AC 100 ~ 240 V, 50/60 Hz, 270VA (FP-4020), 230VA (FP-4025)

Model

RI-4030

RI-4035

Measurement system

Deflection type

Refractive index range

1.00 ~ 1.75
5.0 x 10

Linearity

-5

RIU (HIGH)

5.0 x 10

5.0 x 10-4 RIU (ST D)

-5

RIU (HIGH)

5.0 x 10-4 RIU (ST D) (H2O)

5.0 x 10-3 RIU (LOW) (H2O)
Noise level

0.20 x 10-8 RIU or less (HIGH, ST D)

0.50 x 10-8 RIU or less (HIGH, ST D)

Cell capacity

10 µL

2.7 µL

Maximum usable flow- 10 mL/min (Low Flow Rate tubing)
rate
Maximum pressure

120 mL/min (High Flow Rate
0.1 MPa (Low Flow Rate tubing)
0.3 MPa (High Flow Rate tubing)

Model

CD-4095

Light source

150W Hg-Xe lamp

Polarizer

Glan-T aylor prism

Phase modulation

PEM (photo-elastic modulator)

Wavelength accuracy

+/- 5 nm

Wavelength reproducibility

+/- 0.5 nm

Wavelength setting range

200 ~ 460 nm

Measurement range

220 ~ 460 nm

Noise level

1.2 mL/min (H2 O)

Drift

0.1 MPa

Data output

T emperature control

Ambient +10ºC ~ ambient +25ºC (setting range: 0 ~ 60ºC)

Dimensions, weight

300(W) x 470(D) x 150(H) mm, 14 kg

Power requirement

AC 100 ~ 240 V, 50/60 Hz, 80 VA

Flow cell
Spectrum measurement

0.04 mdeg (at specified conditions)
0.1 mdeg/h (room temperature variation is
within +/-1ºC)
Maximum 100 Hz
Front loading cassette cell, tapered, 44 µ
L, 25 mmL
200 ~ 460 nm

Dimensions, weight

300 (W) x 470 (D) x 230 (H) mm, 21 kg

Power requirement

AC 100 ~ 240V, 50/60 Hz, 210 VA

15

Model

AS-4050/AS-4050i

AS-4150

AS-4250

AS-4058

Full or partially fill loop injection (zero Full or partially fill loop injection (zero
Sample injection

Full or partially fill loop injection (zero sample loss available), direct line

method

sample loss available)

injection (option, change of flow-line
is required.)

is required.)

Number of samples

60 (2 mL vial)

180 (2 mL vial)

180 (2 mL vial)

0.1 ~ 100 µL

0.1 ~ 100 µL

0.1 ~ 100 µL

Injection volume

Injection precision

Injection accuracy

Carry over

sample loss available), direct line

Full or partially fill loop injection (zero

injection (option, change of flow-line

sample loss available)

(large volume injection kit: 1 ~ 1000 µ (large volume injection kit: 1 ~ 1000 µ (large volume injection kit: 1 ~ 1000 µ 1 ~ 10000 µL
L, option)

L, option)

L, option)

RSD 0.3 % or less (specified

RSD 0.25% or less (specified

RSD 0.25% or less (specified

conditions)

conditions)

conditions)

+/- 0.1% or less (with correction

+/- 0.1% or less (with correction

+/- 0.1% or less (with correction

+/- 0.1% or less (with correction

function)

function)

function)

function)

0.01% or less

0.01% or less

(2 µL) (specified conditions)

(2 µL) (specified conditions)

0.005% or less (with multiple solvent

0.005% or less (with multiple solvent

0.01% or less
(10 µL, specified conditions)

pressure
Sample
cooling/heating
Pre-column
derivatization, etc

RSD 1.0 % or less (2500 µL, zero
sample loss)
(specified conditions)

flushing valve) (specified conditions) flushing valve) (specified conditions)
Maximum usable

40 (20 mL vial, standard)

0.05% or less
(2000 µL, zero sample loss)
(specified conditions)
0.02% or less (with multiple solvent
flushing valve) (specified conditions)

30 MPa (AS-4050), 34 MPa (AS-4050i) 70 MPa

130 MPa

41 MPa

(option) temp. range: 4 ~ 40ºC

(option) temp. range: 4 ~ 40ºC

-

(option) temp. range: 4 ~ 40ºC

(option) temp. range: 4 ~ 40ºC

Pre-column derivatization (up to two

Pre-column derivatization (up to two

reagents), dilution, other user

reagents), dilution, other user

programs

programs

-

Dimensions, weight

300 (W) x 470 (D) x 300 (H) mm, 21 kg 300 (W) x 470 (D) x 385.5 (H) mm, 25 kg300 (W) x 470 (D) x 385.5 (H) mm, 25 kg300 (W) x 470 (D) x 385.5 (H) mm, 25 kg

Power requirement

AC 100 ~ 240V, 50/60 Hz, 60 VA

AC 100 ~ 240V, 50/60 Hz, 75 VA

AC 100 ~ 240V, 50/60 Hz, 75 VA

AC 100 ~ 240V, 50/60 Hz, 75 VA

Model

CO-4061

CO-4062

CO-4060

CO-4065

RO-4068

T emperature range

Ambient -15ºC ~ 100ºC

Ambient -15ºC ~ 100ºC

Ambient -15ºC ~ 80ºC

Ambient -15ºC ~ 90ºC

Ambient +10ºC ~ 200ºC

280(W) x 25(D) x 60(H) mm

280(W) x 25(D) x 100(H) mm

280(W) x 25(D) x 100(H) mm

110(W) x 83(D) x 415(H) mm

270(W) x 120(D) x 415(H) mm

Column
compartment
dimensions
Heating/cooling power shut off when unusually high temperatures or solvent leaks are detected.

Safety features
Dimensions, weight
Power consumption

300(W) x 470(D) x 150(H) mm, 300(W) x 470(D) x 150(H) mm, 150(W) x 470(D) x 465(H) mm, 300(W) x 470(D) x 465(H) mm, 150(W) x 470(D) x 150(H) mm,
10 kg

10 kg

16 kg

25 kg

AC 100 ~ 240 V, 50/60 Hz,

AC 100 ~ 240 V, 50/60 Hz,

AC 100 ~ 240 V, 50/60 Hz,

AC 100 ~ 240 V, 50/60 Hz,

AC 100 ~ 240 V, 50/60 Hz,

160 VA

200 VA

350 VA

450 VA

325 VA

Language

English/Japanese, selectable

Windows OS

Windows 7 Professional 32/64 bit
Windows 8.1 Professional 32/64 bit

Controllable hardware

LC-4000 Series, X-LC 3000 Series, LC-2000 Series,

System control

Direct control of instrument by Direct Mode,

Maximum 4 systems control, Maximum 20 systems registration
T ime programming by Control Method,
Spectrum measurement,
Prevention from unauthorized use by security key,
Analog signal input,
Up to 4 channel simultaneous input per one LC-Net II/ADC

JASCO INTERNATIONAL CO., LTD.

11-10, Myojin-cho 1-chome, Hachioji, Tokyo 192-0046, Japan
Tel: +81-42-649-3247 Fax: +81-42-649-3518 http://www.jascoint.co.jp/english/
Australia, China, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand,
Pakistan, Philippines, Russia, Singapore, Taiwan, Thailand

JASCO INCORPORATED

28600 Mary's Court, Easton, MD 21601, U.S.A
Tel: +1-800-333-5272 +1-410-822-1220 Fax: +1-410-822-7526 http://www.jascoinc.com
U.S.A., Canada, Costa Rica, Mexico, Puerto Rico, Argentina, Brazil, Chile, Colombia, Paraguay,
Peru, Uruguay, Guatemala, , Ecuador, Bolivia

JASCO EUROPE s.r.l.

Via Luigi Cadorna 1, 23894 Cremella (Lc), Italy
Tel: +39-039-9215811 Fax: +39-039-9215835 http://www.jasco-europe.com
JASCO Deutschland www.jasco.de, JASCO UK www.jasco.co.uk,
JASCO France www.jascofrance.fr, JASCO Benelux www.jasco.nl,
JASCO Spain www.jasco-spain.com
Italy, Germany, U.K., France, Netherlands, Belgium, Luxembourg, Spain, Sweden, Norway,
Denmark,
Austria, Finland, Greece, Hungary, Poland, Portugal, Romania, Switzerland, Algeria, Cyprus,
Egypt, Israel, Jordan, Kuwait, Lebanon, Morocco, Saudi Arabia, South Africa, Tunisia,
Turkey, U.A.E., Yemen

12 kg



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.6-c011 79.156289, 2014/03/31-23:39:12
Instance ID                     : uuid:ad0d5631-2d84-4467-bf63-ff0e901126dd
Original Document ID            : adobe:docid:indd:f2355600-2bd1-11e4-9d06-bbfd56ae698a
Document ID                     : xmp.id:806b0494-b5ee-c44f-86c1-e3701d95a2da
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:a6eff5ac-d704-664d-8723-af6ce4ac0bd8
Derived From Document ID        : xmp.did:a6eff5ac-d704-664d-8723-af6ce4ac0bd8
Derived From Original Document ID: adobe:docid:indd:f2355600-2bd1-11e4-9d06-bbfd56ae698a
Derived From Rendition Class    : default
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CC 2014 (Windows)
History Changed                 : /
History When                    : 2014:09:29 10:14:19-04:00
Create Date                     : 2014:09:29 10:14:19-04:00
Modify Date                     : 2014:09:29 10:15:28-04:00
Metadata Date                   : 2014:09:29 10:15:28-04:00
Creator Tool                    : Adobe InDesign CC 2014 (Windows)
Format                          : application/pdf
Producer                        : Adobe PDF Library 11.0
Trapped                         : False
Page Count                      : 16
Creator                         : Adobe InDesign CC 2014 (Windows)
EXIF Metadata provided by EXIF.tools

Navigation menu