Combination pH/ORP Sensor
Instruction Manual
PN 51-396P/rev.H
March 2006
MODEL 396P
MODEL 396PVP
with mating cable
Models 396P and 396PVP
CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY
The wetted sensor materials may not be
compatible with process composition and
operating conditions. Application
compatibility is entirely the responsibility
of the user.
DANGER
HAZARDOUS AREA
INSTALLATION
This sensor is not Intrinsically Safe. or Explosion
Proof. Installations near flammable liquids or in
hazardous area locations must be carefully eval-
uated by qualified on site safety personnel.
To secure and maintain an intrinsically safe
installation, an appropriate transmitter/ safety
barrier/sensor combination must be used. The
installation system must be in accordance with
the governing approval agency (FM, CSA or
BASEEFA/CENELEC) hazardous area classifi-
cation requirements. Consult your analyzer/
transmitter instruction manual for details.
Proper installation, operation and servicing of
this sensor in a Hazardous Area Installation is
entirely the responsibility of the user.
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests its products to meet
many national and international standards. Because these instruments are
sophisticated technical products, you must properly install, use, and maintain
them to ensure they continue to operate within their normal specifications. The
following instructions must be adhered to and integrated into your safety pro-
gram when installing, using, and maintaining Rosemount Analytical products.
Failure to follow the proper instructions may cause any one of the following sit-
uations to occur: Loss of life; personal injury; property damage; damage to this
instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If this Instruction Manual is not the correct manual, telephone 1-800-654-
7768 and the requested manual will be provided. Save this Instruction
Manual for future reference.
If you do not understand any of the instructions, contact your Rosemount
representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with
the product.
Inform and educate your personnel in the proper installation, operation, and
maintenance of the product.
Install your equipment as specified in the Installation Instructions of the
appropriate Instruction Manual and per applicable local and national codes.
Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate,
update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use
replacement parts specified by Rosemount. Unauthorized parts and proce-
dures can affect the product’s performance and place the safe operation of
your process at risk. Look alike substitutions may result in fire, electrical
hazards, or improper operation.
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2006
About This Document
This manual contains instructions for installation and operation of the Models 396P and 396PVP
TUpH Combination pH/ORP Sensors. The following list provides notes concerning all revisions of
this document.
Rev. Level Date Notes
0 8/95-2/01 This is the initial release of the product manual. The manual has been reformatted to reflect the
Emerson documentation style and updated to reflect any changes in the product offering.
A 12/01 Added wiring diagram to page 14.
B 4/02 Revised drawings on pages 6, 7, and 14.
C 7/02 Added 1055 wiring diagrams.
D 8/03 Added Silcore information.
E 5/04 Added 5081 and Xmt wiring info. Deleted obsolete model information.
G 11/05 Changed Figure 4-6 drawing on page 17. Resized Figure 1-1 drawing on page 2.
H 3/06 Changed PN on Fig. 2-5, page 8, and in Table 7-3, page 29. Added a note on page 11.
i
MODEL 396P and 396PVP TABLE OF CONTENTS
MODEL 396P and 396PVP
COMBINATION pH/ORP SENSOR
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS................................................................. 1
1.1 Features and Applications........................................................................................ 1
1.2 Physical Specifications............................................................................................. 2
1.3 Ordering Information ................................................................................................ 3
2.0 INSTALLATION ....................................................................................................... 5
2.1 Unpacking and Inspection........................................................................................ 5
2.2 Mounting .................................................................................................................. 5
3.0 WIRING MODEL 396P-01 ....................................................................................... 10
4.0 WIRING MODELS 396P-02 AND 396PVP.............................................................. 16
5.0 START UP AND CALIBRATION ............................................................................. 24
5.1 Model 396P pH ........................................................................................................ 24
5.2 Model 396P ORP ..................................................................................................... 24
6.0 MAINTENANCE ...................................................................................................... 25
6.1 Automatic Temperature Compensator ..................................................................... 25
6.2 Model 396P pH ........................................................................................................ 25
6.3 Model 396P ORP ..................................................................................................... 26
7.0 DIAGNOSTICS AND TROUBLESHOOTING.......................................................... 27
7.1 Diagnostics & Troubleshooting With the Models 54e, 1055, 3081, 81, 4081, 5081.
and Xmt.................................................................................................................... 27
7.2 Troubleshooting Without Advanced Diagnostics...................................................... 28
8.0 RETURN OF MATERIAL......................................................................................... 30
LIST OF TABLES
Number Title Page
1-1 Recommended Accessories for First Time Installations........................................... 4
3-1 Wiring Matrix Guide - Model 396P ........................................................................... 10
4-1 Wiring Matrix Guide - Model 396PVP....................................................................... 16
5-1 ORP of Saturated Quinhydrone Solution (In Millivolts)............................................. 24
6-1 Ro and R1 Values for Temperature Compensation Elements .................................. 25
6-2 Temperature vs Resistance of Auto T.C. Elements .................................................. 25
7-1 Troubleshooting With Advanced Diagnostics ........................................................... 27
7-2 Troubleshooting Without Advanced Diagnostics ...................................................... 28
7-3 Model 396P Replacement Parts and Accessories.................................................... 29
MODEL 396P and 396PVP TABLE OF CONTENTS
LIST OF FIGURES
Number Title Page
1-1 Cross Section Diagram of the TUpH Patented Reference Technology ............................................................................... 2
2-1 Dimensional Drawing .......................................................................................................................................................... 6
2-2 Flow-Through Tee with Adapter (PN 915240-xx)................................................................................................................. 6
2-3 Flow-Through and Insertion Installations ........................................................................................................................... 7
2-4 Model 396P/396PVP with Insertion Mounting Adapter (PN 23242-02) ............................................................................... 7
2-5 Submersion Installations .................................................................................................................................................... 8
2-6 Low Flow Cell PN 23728-00................................................................................................................................................ 9
2-7 Jet Spray Cleaner PN 12707-00.......................................................................................................................................... 10
3-1 Wire Functions for Model 396P with & without Preamplifier................................................................................................ 10
3-2 Wiring Details Model 396P-01-50 for use with & without Junction Box (PN 22719-02) for Models 1181, 1050, & 1060 .... 12
3-3 Wiring Details Model 396P-01-54 for use with & sithout Junction Box (PN 22719-02) for Models 1054/A/B, 2054, 2081
and Extension Cable (PN 9200254).................................................................................................................................... 13
3-4 Wiring Details Model 396P-01-55 for use with Junction Box (PN 23550-00) for Models 54, 3081, 4081, 81, and 5081 ... 14
3-5 Wiring Details Model 396P-01-55 for use with Models 54, 81, 3081, 4081, and 5081........................................................ 15
3-6 Wiring Model 396P-01)-( )-55 Sensor to Model 1055-10-22-32 Analyzer ........................................................................... 15
4-1 Wire Functions for Mating Variopol Cable used with Model 396PVP .................................................................................. 17
4-2 Connector Pins and their Functions .................................................................................................................................... 17
4-3 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Model 81 ....... 17
4-4 Wiring Details for Models 396PVP or 396P-02-50 with Mating Variopol Cable (PN 23645-07) for use with Model 1181 ... 17
4-5 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-07) for use with Model 2081 ... 17
4-6 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Models 3081,
4081, and 5081 ................................................................................................................................................................... 17
4-7 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote ........
Junction Box (PN 23555-00) to Model 81 ........................................................................................................................... 18
4-8 Wiring Details for Models 396PVP or 396P-02-50 with Mating Variopol Cable (PN 23645-07) for use with Remote ........
Junction Box (PN 23309-03) to Model 1181........................................................................................................................ 18
4-9 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Model 54 ...... 18
4-10 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote ........
Junction Box (PN 23555-00) to Model 54 ........................................................................................................................... 18
4-11 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-06) for use with Remote ........
Junction Box (PN 23309-03) to Model 2081 ....................................................................................................................... 19
4-12 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote ........
Junction Box (PN 23555-00) to Models 3081 and 4081...................................................................................................... 19
4-13 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-07) for use with Model 1054 ... 19
4-14 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-07) for use with ......................
Model SCL-(P/Q)................................................................................................................................................................. 19
4-15 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Model 1055 ... 20
4-16 Wiring Model 396P-(02)-( )-54/55-(61) Sensor to Model 1055-01-10-22-32 Analyzer ........................................................ 20
4-17 Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Remote ........
Junction Box (PN 23557-00) to Model 1055 ....................................................................................................................... 20
4-18 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-06) for use with Models ..........
1054A/B and 2054............................................................................................................................................................... 21
4-19 Wiring Details for Models 396PVP or 396P-02-54 with Mating Variopol Cable (PN 23645-06) for use with Remote ........
Junction Box (PN 23309-04) to Models 1054A/B and 2054................................................................................................ 21
4-20 Wiring to Model 54e pH/ORP with Variopol........................................................................................................................ 21
4-21 Wiring Details Model 396P-02-50 for use with Junction Box (PN 23309-03), Remote Preamplifier, and Extension ..........
Cable (PN 9200254)............................................................................................................................................................ 22
4-22 Wiring Details Model 396P-02-54 for use with Junction Box (PN 23309-04), Remote Preamplifier (PN 22698-03), and ..
Extension Cable (PN 9200254)........................................................................................................................................... 22
4-23 Wiring Model 396P-01 to Model Xmt-P-HT-10 ................................................................................................................... 23
4-24 Wiring Models 396P-02 and 396PVP to Model Xmt-P-HT-10 ............................................................................................ 23
ii
Model 396PVP is offered with a watertight sensor-to-
cable connector which eliminates re-wiring and cable
twisting when replacing sensors. The Variopol VP
multiple pin connector is an integral part of each sen-
sor model and uses a mating VP cable. Once the
cable is installed and wired to the analyzer, sensors
are easily replaced without replacing the cable, and, if
the replacement sensor is the same as its predeces-
sor, without rewiring the analyzer. Also the cable can
be disconnected from the sensor before removal from
the process which eliminates cable twisting.
The entire line of TUpH model sensors now incor-
porate the new SILCORE technology contaminant
barrier. This triple-seal barrier prevents moisture and
material impurities from migrating to the pH sensor’s
reference electrode’s metal lead wire. By preventing
these contaminants from compromising the integrity
of the pH measurement, sensor life is increased,
especially at higher temperatures where increased
migrations occur. In addition, the SILCORE technolo-
gy provides added protection against sensor failure
due to vibrations and shock by transferring damaging
energy away from the glass-to-metal seal.
The AccuGLASS pH glass formulations exceed
industry standards. The AccuGlass pH glass is a
result of many years of glass research resulting in a
formulation which has been found to increase the life
of the sensor. Unlike other pH glasses presently on
the market, this glass resists cracking especially at
higher temperatures and reduces sodium ion error
commonly found in high pH applications. Overall, the
AccuGlass formulation enhances the sensor perform-
ance to measure pH more accurately and have a
longer sensor life than ever before.
AccuGlass and Silcore are trademarks of Rosemount Analytical.
1
MODEL 396P and 396PVP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.1 FEATURES AND APPLICATIONS
The Rosemount Analytical Model 396P and 396PVP
Sensors measure the pH or the ORP of aqueous solu-
tions in pipelines, open tanks, or ponds. It is designed
for harsh, dirty applications such as sourwater waste
treatment and scrubbers, where a high performance,
low maintenance, disposable sensor is required.
The combination electrode features a patented, large
area polypropylene reference junction for resistance to
process coatings, generally found in dirty, high solid
applications. The patented helical reference pathway
provides added resistance for poisoning applications.
In addition, it is available in two configurations:
glass/platinum electrode completely recessed within
the reference junction for added life in abrasive or
rough applications and the glass/platinum electrode
partially exposed for added life and accuracy in vis-
cous and low flow applications.
Both Models feature a titanium solution ground con-
structed in an annular design around the pH/ORP elec-
trode. The solution ground provides glass and refer-
ence sensor diagnostics for preventative maintenance
when used with the Model 54e, 1055, 5081, 3081,
4081, 81 pH/ORP, and Xmt-P transmitters. In addition
Models 396P and 396PVP can be used with most non-
diagnostic Rosemount Analytical and other manufac-
turer’s instruments.
Each sensor is housed in a molded polypropylene
body with EPDM seals, making it virtually indestructi-
ble and chemically resistant. Complete encapsulation
eliminates leakage or high humidity problems tradi-
tionally found in other pH/ORP designs. The simplified
construction, designed with user convenience in mind,
does not require electrolyte (KCl) replenishment or any
high maintenance troubleshooting procedures.
The Model 396P is offered with or without an optional
integral preamplifier to convert the high impedance pH
or ORP mV signal into a stable, noise-free signal with
transmission capability of up to three miles. A remote
preamplifier must be used with Model 396PVP.
Installation is easily achieved through the wide variety
of mounting configurations. The Models 396P and
396PVP feature 1 in. (MNPT) front and rear facing
connections for insertion, submersion or flow through
pH and ORP applications.
2
MODEL 396P and 396PVP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
Measurements and Ranges *pH: 0-14
ORP: -1500 to 1500 mv
Available pH ACCUGLASS Types GPLR hemi or flat glass
Wetted Materials Titanium, Polypropylene, EPDM, glass; platinum (ORP only)
Process Connection 1 in. MNPT front and rear facing threads
Temperature Range 0-100°C (32-212°F)
Pressure Range-Hemi bulb 100-1135 kPa [abs] (0-150 psig)
Pressure Range-Flat bulb 100-790 kPa [abs] (0-100 psig)
Minimum Conductivity 75 μS/cm, nominal; 100 μS/cm
Integral Cable 396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial
396PVP: none - must use mating VP cable
Preamplifier Options 396P: Remote or Integral
396PVP: Remote only
Weight/Shipping Weight 0.45 kg/0.9 kg (1 lb/2 lb)
1.2 SPECIFICATIONS
FIGURE 1-1. Cross Section Diagram of the TUpH Patented Reference Technology
All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an AccuGlass pH
glass bulb. This patented sensor technology ensures superior performance while only requiring minimal maintenance.
PERCENT LINEARITY*
pH range GPLR hemi bulb GPLR flat bulb
0 - 2 pH 94%
2 - 12 pH 99% 98%
12 - 13 pH 97% 95%
13 - 14 pH 92%
3
MODEL 396P and 396PVP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
1.3 ORDERING INFORMATION
The Model 396P Sensor is housed in a molded reinforced polypropylene
body with 1 in. MNPT threads suitable for insertion, submersion or flow
through installation. The sensor includes a general purpose pH electrode
or a platinum ORP electrode, a patented reference junction and a solution
ground. The Model 396P comes standard with a recessed electrode; an
optional slotted tip is also available. In addition, the 396P features an
optional integral hermetically sealed preamplifier and 15 ft or 25 ft cable
lengths. Automatic temperature compensation, Pt 100 or 3K Balco, is
standard with the Model 396P.
CODE PREAMPLIFIER/CABLE (Required Selection)
01 With integral preamplifier, 25 ft cable
02 Without integral preamplifier, 15 ft cable
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 GPLR hemi bulb, General Purpose Low Resistivity (0-14 pH)
12 ORP
13 GPLR flat bulb, General Purpose Low Resistivity (2 - 13 pH)
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC)
54 For Models 1054A/B, 2054, 2081 (Pt 100 RTD)
55 For Models 54e, 3081, 81, 4081, 1055, 5081, Xmt (PT-100 RTD)
396P - 01 - 10 - 55 EXAMPLE
NOTE: The Model 396P is also compatible with Model SCL-P/Q (option 02-54 only).
CODE OPTIONAL SELECTION
41 Slotted Tip (not available on flat bulb sensors)
1. Mounting Accessories (optional)
Choose one: PN 915240-03, PVC flow through tee, 3/4 in. NPT process connection
PN 915240-04, PVC flow through tee, 1 in. NPT process connection
PN 915240-05, PVC flow through tee, 1-1/2 in. NPT process connection
PN 23728-00, acrylic low flow cell
PN 2002011, 1-1/2 in. CPVC tee with 1-in. FNPT connection
PN 11275-01, sensor handrail assembly
2. Junction Boxes (optional)
Remote Junction Boxes (used with option -02 sensors, for sensor to analyzer distances of more than 15 ft)
Choose one: PN 23555-00 includes preamplifier for Models 54, 81, 3081, 4081, 5081, Xmt
PN 23309-03 and PN 22698-02 plug-in preamplifier for Model 1181Analyzer
PN 23309-04 and PN 22698-03 plug-in preamplifier for Models 1054A/B series, 2054, 2081 Analyzers
Choose one: PN 9120516 BNC Adapter for use with remote junction boxes PNs 23309-03, 23309-04
Remote Junction Box (used with option -01 sensors)
Choose one: PN 23550-00 cable extension board
3. Extension cables (used with remote junction boxes)
Choose one: PN 23646-01, 11 conductor, shielded, prepped
PN 9200273, 11 conductor, shielded, unprepped
TABLE 1-1. RECOMMENDED ACCESSORIES FOR FIRST TIME INSTALLATIONS
The Model 396P insertion/submersion
sensor with integral cable is offered with
or without a built-in preamplifier
MODEL
396P TUpH INSERTION/SUBMERSION POLYPROPYLENE pH/ORP SENSOR
4
MODEL 396P and 396PVP SECTION 1.0
DESCRIPTIONS AND SPECIFICATIONS
Optional Slotted Tip
is available on all
hemi bulb sensors,
ordered as option -41
Flat Tip is avail-
able with flat
glass bulb sen-
sors
Examples of all sensing tip offerings
Shrouded Tip is
standard on all
hemi bulb sensors
Variopol connector
shown with mating variopol
cable receptacle
MODEL
396PVP TUpH INSERTION/SUBMERSION POLYPROPYLENE pH/ORP SENSOR
CODE MEASURING ELECTRODE TYPE (Required Selection)
10 GPLR hemi bulb, General Purpose Low Resistivity (0-14 pH)
12 ORP
13 GPLR flat bulb, General Purpose Low Resistivity (2 - 13 pH)
CODE ANALYZER/TC COMPATIBILITY (Required Selection)
50 For Models 1181 (3K TC)
54 For Models 1054A/B, 2054, 2081 (Pt 100 RTD)
55 For Models 54e, 3081, 81, 4081, 1055, 5081, Xmt (PT-100 RTD)
CODE OPTIONAL SELECTION
41 Slotted Tip (not available on flat bulb sensors)
PN MATING VP CONNECTOR CABLE (Required for all new installations)
23645-06 15 ft. cable with mating VP connector, prepped with BNC on analyzer end
23645-07 15 ft. cable with mating VP connector, prepped without BNC on analyzer end*
The Model 396PVP Sensor has similar features to the Model 396P.
However, the Model 396PVP is offered with the new Variopol (VP)
connector and uses a mating VP cable (purchased separately).
A remote preamplifier must be used with this sensor.
A Variopol cable is required for all new installations. See below
for cable selection.
The Model 396PVP insertion/ submersion sen-
sor with the VP (Variopol) connector
OTHER ACCESSORIES FOR MODELS 396P AND 396PVP
PART DESCRIPTION
22698-00 Preamplifier plug-in for junction box, for Model 1003,
22698-02 Preamplifier plug-in for junction box, for Models 1181, 1050
22698-03 Preamplifier plug-in for junction box, for Models 1054A/B, 2054, 2081
22743-01 Pt100 preamplifier for Model 1181
22744-01 3K Preamplifier for Model 1181
23557-00 Preamplifier for junction box for Models 54, 3081, 81, 4081, 5081
9210012 Buffer solution, 4.01 pH, 16 oz
9210013 Buffer solution, 6.86 pH, 16 oz
9210014 Buffer solution, 9.18 pH, 16oz
R508-160Z ORP solution, 460 mv ± 10 at 20°C
12707-00 Jet Spray Cleaner
*For use with connections to Models 1181, 1054A/B, 2081, 54e, 81, 3081, 4081, 5081, Xmt, and remote junction box PN 23555-00.
5
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
SECTION 2.0
INSTALLATION
2. Do not install the sensor on the horizontal. The
sensor must be 10° off the horizontal to ensure
accuracy.
3. Do not install the sensor upside down.
4. Air bubbles may become trapped in the sensor end
between the glass bulb and the sensor body. This
problem is most commonly encountered in areas
of low flow or during calibration. Shake the probe
while immersed in solution to remove bubbles.
This problem can be avoided by ordering the sen-
sor with the slotted tip (option -41).
In most cases, the pH sensor can simply be installed
as shipped and readings with an accuracy of ± 0.6 pH
may be obtained. To obtain greater accuracy or to ver-
ify proper operation, the sensor must be calibrated as
a loop with its compatible analyzer or transmitter.
2.2.1 Flow Through and Insertion Mounting.
Model 396P and 396PVP Sensors have a 1-inch
MNPT process connection at the front of the sensor for
mounting into a 1-1/2 inch tee or the process pipes.
See Figure 2-1 through Figure 2-7 for installation con-
figurations.
NOTE
LARGE PIPE WRENCHES MUST NOT BE
USED TO TIGHTEN THE SENSOR INTO A
FLANGE OR OTHER TYPE OF MOUNTING.
2.2.2 Submersion Mounting. Model 396P and
396PVP Sensors also have a 1 inch MNPT process
connection at the back of the sensor. Utilizing a stan-
dard 1 inch union, the sensor may be mounted to a 1
inch SCH 80 CPVC or PVDF standpipe (see Figure 2-
x). Tapered pipe threads in plastic tend to loosen after
installation. It is therefore recommended that Teflon1
tape be used on the threads and that the tightness of
the connection be checked frequently to assure that no
loosening has occurred. To prevent rain water or con-
densation from running into the sensor, a weatherproof
junction box is recommended (see Figure 2-x). The
sensor cable must be run through a protective conduit
for isolation from electrical interference or physical
abuse from the process. The sensor should be installed
within 80° of vertical, with the electrode facing down.
The sensor’s cable should not be run with power or
control wiring.
1Teflon is a registered trademark of E.I. du Pont de Nemours & Co.
2.1 UNPACKING AND INSPECTION.
Inspect the outside of the carton for any damage. If dam-
age is detected, contact the carrier immediately. Inspect
the hardware. Make sure all the items in the packing list
are present and in good condition. Notify the factory if any
part is missing. If the sensor appears to be in satisfactory
condition, proceed to Section 2.2, Mounting.
NOTE
Save the original packing cartons and materi-
als as most carriers require proof of damage
due to mishandling, etc. Also, if it is necessary
to return the sensor to the factory, you must
pack the sensor in the same manner as it was
received. Refer to Section 6.0 for return
instructions. If the sensor is to be stored, the
vinyl boot should be filled with pH buffer solu-
tion and replaced on sensor tip until ready to
use.
CAUTION
Buffer solution, in the vinyl boot, may cause
skin or eye irritation.
WARNING
Glass electrode must be wetted at all times (in
storage and in line) to maximize sensor life.
2.2 MOUNTING.
The sensor has been designed to be located in indus-
trial process environments. Temperature and pressure
limitations must not be exceeded at any time. A caution
label regarding this matter is attached to the sensor.
Please do not remove the label. See Figure 2-1.
CAUTION
Internal electrolyte fill solution may cause skin
or eye irritation.
Mounting Guidelines:
1. Shake the sensor in a downward motion to
remove any air bubbles that may be present inside
the tip of the pH glass.
6
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
FIGURE 2-1. Dimensional Drawing
DWG. NO. REV.
40396P01 D
MILLIMETER
INCH
FIGURE 2-2. Flow-Through Tee with Adapter (PN 915240-xx*)
MILLIMETER
INCH
xx* Process Connection
Threads
03 3/4 inch
04 1 inch
05 1-1/2 inch
SENSOR CABLE (OR VP CONNECTOR - NOT SHOWN)
SENSOR CABLE
(OR VP CONNECTOR
- NOT SHOWN)
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
FIGURE 2-4. Model 396P with Insertion Mounting Adapter (PN 23242-02).
Not for use with Model 396PVP.
Mounting adapter allows for sensor removal without twisting or
disconnecting interconnecting cable for ease of maintenance.
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396P02 A
FIGURE 2-3. Flow-Through and Insertion Installations
1-1/2” PIPE “Y”
1-1/2” X 1”
Reducing
Bushing
1-1/2” X 1”
Reducing
Bushing
1-1/2” X 1”
Reducing
Bushing
1-1/2” Pipe Tee
PN 2002011
1-1/2” Pipe Tee
PN 2002011
PIPE “Y”
INSTALLATION
SHOWN
STRAIGHT
FLOW
SHOWN
FLOW
ANGLE
FLOW
SHOWN
7
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
FIGURE 2-5. Submersion Installations
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396P04 A
DWG. NO. REV.
40396P03 A
Handrail Mounting
Assembly PN 11275-01
8
9
MODEL 396P and 396PVP SECTION 2.0
INSTALLATION
FIGURE 2-6. Low Flow Cell PN 23728-00
MILLIMETER
INCH
FIGURE 2-7. Jet Spray Cleaner PN 12707-00
SENSOR CABLE
(OR VP CONNECTOR
- NOT SHOWN)
10
Figures in this section provide the guidelines for wiring
the 396P-01 sensor to various Analyzer/Transmitter
instruments.
To determine which wiring guideline to use, locate the
model number of the sensor to be installed.
1. If the cable needs to be extended, use a high qual-
ity eleven conductor double shielded instrument
cable available from Rosemount Analytical. Refer
to Table 3-1 for the appropriate junction box to use
and the corresponding wiring details.
NOTE
If the cable is too long, loop up the excess
cable. If the cable has to be shortened, cut
and terminate each conductor neatly and
make sure that the overall (outermost)
drain wire is not shorted out with either
of the two inner drain wires (shields).
2. Signal cable should be run in a dedicated conduit
(preferably an earth grounded metallic conduit)
and should be kept away from AC power lines. For
your convenience, a wire nut kit is furnished (in a
plastic bag wrapped around the cable).
NOTE
For maximum EMI/RFI protection when
wiring from the sensor to the junction box,
the outer braid of the sensor should be con-
nected to the outer braided shield of the
extension cable. The outer braid of the
extension cable to the instrument must be
terminated at earth ground or by using an
appropriate metal cable gland fitting that
provides a secure connection to the instru-
ment cable.
Preamplifier Location Analyzer/Transmitter
Sensor Integral to Integral to Remote Interconnect 1054A, 54e, 81, 1055 1181 1050 1003 SCL or Refer to
Model Sensor Analyzer Junction (unprepped 1054B, 2054 3081, 1060 1023 SoluCube Figure
Options Box (PN) PN) 2081 4081, 5081 #
396P-01-()-50  3-2
396P-01-()-50 Ext. box 9200254  3-2
(22719-02)
396P-01-()-54  3-3
396P-01-()-54 Ext. box 9200254 3-3
(22719-02)
396P-01-()-55 3-5
396P-01-()-55 Ext. box 9200273 3-4
(23550-00)
396P-01-()-55 9200273 3-6
or 23646-01
TABLE 3-1. Wiring Matrix Guide
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
SECTION 3.0
WIRING — MODEL 396P-01
11
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
Wiring. The Model 396P has an optional built-in pre-
amplifier and is offered with a shielded cable. The
cable should be handled carefully and kept dry and
free of corrosive chemicals at all times. Extreme care
should be used to prevent it from being twisted, dam-
aged or scraped by rough, sharp edges or surfaces.
Please refer to Figures 3-1 through 3-6 for wiring
Model 396P-01.
FIGURE 3-1. Wire Functions for Model 396P with and without Preamplifier
DANGER
DO NOT CONNECT SENSOR CABLE TO
POWER LINES. SERIOUS INJURY MAY
RESULT.
Model 396P
Option 01 (with
preamplifier) cable
Model 396P
Option 02 (without
preamplifier) cable
NOTE
Remove electrical tape or shrink sleeve from gray
reference wire before connecting wire to terminal.
12
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
FIGURE 3-2. Wiring Details Model 396P-01-50 for use with and without Junction Box
(PN 22719-02) for Model 1181.
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396P15 B
13
FIGURE 3-3. Wiring Details Model 396P-01-54 for use with and without Junction Box
(PN 22719-02) for Models 1054A, 1054B, 2054, 2081, and Extension Cable (9200254).
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
DWG. NO. REV.
40396P13 D
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
14
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
FIGURE 3-4. Wiring Details Model 396P-01-55 for use with Junction Box
(PN 23550-00) for Models 54e, 3081, 4081, 81, and 5081.
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
DWG. NO. REV.
40396P08 B
15
MODEL 396P and 396PVP SECTION 3.0
WIRING - MODEL 396P
FIGURE 3-5. Wiring Details Model 396P-01-55 for use with Models 54e, 81, 3081, 4081, and 5081
FIGURE 3-6. Wiring Model 396P-01-( )--55 Sensor to Model 1055-10-22-32 Analyzer
16
Figures 4-1 thru 4-22 provide the guidelines for wiring
the sensor to various Analyzer/Transmitter instruments.
To determine which wiring guideline to use, locate the
model number of the sensor to be installed.
1. If the cable needs to be extended, use a high qual-
ity eleven conductor double shielded instrument
cable available from Rosemount Analytical. Refer
to Table 4-1 to refer to the appropriate junction box
to use and the corresponding wiring details.
NOTE
If the cable is too long, loop up the excess
cable. If the cable has to be shortened, cut and
terminate each conductor neatly and make
sure that the overall (outermost) drain wire
is not shorted out with either of the two
inner drain wires (shields).
2. Signal cable should be run in a dedicated conduit
(preferably an earth grounded metallic conduit)
and should be kept away from AC power lines. For
your convenience, a wire nut kit is furnished (in a
plastic bag wrapped around the cable).
NOTE
For maximum EMI/RFI protection when wiring
from the sensor to the junction box, the outer
braid of the sensor should be connected to the
outer braided shield of the extension cable.
The outer braid of the extension cable to the
instrument must be terminated at earth ground
or by using an appropriate metal cable gland
fitting that provides a secure connection to the
instrument cable.
TABLE 4-1. Wiring Matrix Guide
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
SECTION 4.0
WIRING — MODELS 396P-02 AND 396PVP
Preamplifier Location Analyzer/Transmitter
Sensor Integral to Integral to Remote Interconnect 1054A, 54, 81, 1055 1181 1050 1003 SCL or Refer to
Model Sensor Analyzer Junction (unprepped 1054B, 2054 3081, 1060 1023 SoluCube Figure
Options Box (PN) PN) 2081 4081, 5081 #
396P-02-()-55 4-3, 4-6,
396PVP-55 4-9
396P-02-()-50 4-4
396PVP-50
396P-02-()-50 23309-03 9200254 4-22
396PVP-50
396P-02-()-50 23309-03 9200273 or 4-8
396PVP-50 23646-01
396P-02-()-54 23309-04 9200254 4-23
396P-02-()-54 23309-04 9200273 or 4-20
396PVP-54 23646-01
396P-02-()-54  4-5, 4-18
396PVP-54 4-19
396P-02-()-55 23555-00 9200273 or 4-7, 4-10
396PVP-55 23646-01 4-12
396P-02-()-54 661-898695 4-14
396PVP-54
396P-02-54 23309-04 9200273 or 4-11
396PVP-54 23646-01 4-21
396P-02-55 4-15
396PVP-55 4-16
396P-02-55 9200273 or 4-17
396PVP-55 23646-01
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-1. Wire Functions
for Mating Variopol Cable used
with Model 396PVP
FIGURE 4-2. Connector Pins
and Their Functions
FIGURE 4-3. Wiring Details for Models 396PVP
or 396P-02-55 with Mating Variopol Cable
(PN 23645-07) for use with Model 81
FIGURE 4-4. Wiring Details for Models 396PVP
or 396P-02-50 with Mating Variopol Cable
(PN 23645-07) for use with Model 1181
FIGURE 4-5. Wiring Details for Models 396PVP
or 396P-02-54 with Mating Variopol Cable
(PN 23645-07) for use with Model 2081
FIGURE 4-6. Wiring Details for Models 396PVP
or 396P-02-55 with Mating Variopol Cable
(PN 23645-07) for use with Models 3081, 4081, 5081
17
18
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-8. Wiring Details for Models 396PVP
or 396P-02-50 with Mating Variopol Cable
(PN 23645-06) for use with Remote Junction Box
(PN 23309-03) to Model 1181
FIGURE 4-7. Wiring Details for Models 396PVP
or 396P-02-55 with Mating Variopol Cable
(PN 23645-07) for use with Remote Junction Box
(PN 23555-00) to Model 81
FIGURE 4-9. Wiring Details for
Models 396PVP or 396P-02-55 with
Mating Variopol Cable (PN 23645-07)
for use with Model 54/54e
FIGURE 4-10. Wiring Details for Models 396PVP or 396P-02-55
with Mating Variopol Cable (PN 23645-07) for use with Remote
Junction Box (PN 23555-00) to Model 54/54e
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-12. Wiring Details for Models 396PVP or 396P-02-55
with Mating Variopol Cable (PN 23645-07) for use with Remote
Junction Box (PN 23555-00) to Models 3081, 4081, & 5081
FIGURE 4-11. Wiring Details for Models 396PVP or 396P-
02-54 with Mating Variopol Cable (PN 23645-06) for use
with Remote Junction Box (PN 23309-04) to Model 2081
FIGURE 4-14. Wiring Details for Models 396PVP or
396P-02-54 with Mating Variopol Cable (PN 23645-07)
for use with Model SCL-(P/Q)
19
FIGURE 4-13. Wiring Details for Models 396PVP
or 396P-02-54 with Mating Variopol Cable
(PN 23645-07) for use with Model 1054
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-15. Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07) for use with Model 1055
FIGURE 4-16. Wiring Model 396P-(02)-( )--54/55-(61) Sensor to Model 1055-01-10-22-32 Analyzer
FIGURE 4-17. Wiring Details for Models 396PVP or 396P-02-55 with Mating Variopol Cable (PN 23645-07)
for use with Remote Junction Box (PN 23557-00) to Model 1055
NOTE: This wiring diagram can also be used for wiring a Model 396P-01. The wire colors and functions for
the extension cable (PN 9200273) are the same as the 396P-01 wire colors and functions.
20
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-18. Wiring Details for Models 396PVP
or 396P-02-54 with Mating Variopol Cable
(PN 23645-06) for use with Models 1054A/B & 2054
FIGURE 4-19. Wiring Details for Models 396PVP or
396P-02-54 with Mating Variopol Cable (PN 23645-06)
for use with Remote Junction Box (PN 23309-04) to
Models 1054A/B and 2054
21
FIGURE 4-20. Wiring to Model 54e pH/ORP
with Variopol
22
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-21. Wiring Details Model 396P-02-50 for use with Junction Box
(PN 23309-03) and
Remote Preamplifier, Extension Cable
(PN 9200254).
DWG. NO. REV.
40396P25 A
FIGURE 4-22. Wiring Details Model 396P-02-54 for use with Junction Box (PN 23309-04) and
Remote Preamplifier (PN 22698-03), Extension Cable (9200254).
DWG. NO. REV.
40396P24 A
MODEL 396P and 396PVP SECTION 4.0
WIRING - MODEL 396PVP
FIGURE 4-23. Wiring Model 396P-01 to Model Xmt-P-HT-10
FIGURE 4-24. Wiring Models 396P-02 and 396PVP to Model Xmt-P-HT-10
23
1. While obtaining a process solution sample (it is recommended that
the sample is taken close to the sensor), record the pH value that
is shown on the analyzer/transmitter display.
2. Measure and record the pH of the process solution sample with
another temperature compensated, calibrated pH instrument. For
best results, standardization should be performed at the process
temperature.
3. Adjust the analyzer/transmitter value to the standardized value.
5.2 MODEL 396P and 396PVP ORP SENSORS
5.2.1 SENSOR PREPARATION. Most industrial applications have
a number of ORP reactions occurring in sequence or simultaneously.
There can be several components that are oxidized or reduced by
the reagents that are used. Theoretically, the ORP potential is
absolute because it is the result of the oxidation-reduction equilibri-
um. However, the actual measured potential is dependent on many
factors, including the condition of the surface of the ORP platinum
electrode. Therefore, the sensor should be allowed 1-2 hours to
become “conditioned” to the stream when first set-up or after
being cleaned.
5.2.2 ORP CALIBRATION
1. Make a temporary electrical connection between the sensor and
the instrument.
2. Obtain an ORP standard solution, or a standard solution can also be
made quite simply by adding a few crystals of quinhydrone to either
pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble therefore a
few crystals will be required. (Refer to Section 4.3. for an alternate
ORP standard solution).
3. Immerse the sensor in the standard solution. Allow 1-2 minutes for
the ORP sensor to stabilize.
4. Adjust the standardize control of the instrument to the solution value
shown in Table 5-1 (below) or on the label of the standard solution.
The resulting potentials, measured with a clean platinum electrode
and saturated KCl/AgCl reference electrode, should be within ±20
millivolts of the value. Solution temperature must be noted to ensure
accurate interpretation of results. The ORP value of saturated quin-
hydrone solution is not stable over long periods of time. Therefore,
these standards should be made up fresh each time they are used.
5. Remove the sensor from the buffer, rinse and install in the process.
MODEL 396P and 396PVP SECTION 5.0
START UP AND CALIBRATION
SECTION 5.0
START UP AND CALIBRATION
5.1 MODELS 396P and 396PVP pH SENSORS
5.1.1 SENSOR PREPARATION.Shake down the sensor
to remove any air bubbles that may be present at the tip of
the pH glass bulb. In most cases, the pH sensor can simply
be installed as shipped and readings with an accuracy of ±
0.6 pH may be obtained. To obtain greater accuracy or to
verify proper operation, the sensor must be calibrated as a
loop with its compatible analyzer or transmitter.
5.1.2 pH CALIBRATION. After a temporary connection is
established between the sensor and the instrument, a buffer
calibration may be performed. Consult appropriate pH/ORP
analyzer or transmitter instruction manual for specific calibra-
tion and standardization procedures, or see below for recom-
mended two-point buffer calibration procedure.
Recommended two-point buffer calibration procedure:
Select two stable buffer solutions, preferably pH 4.0 and 10.0
(pH buffers other than pH 4.0 and pH 10.0 can be used as long
as the pH values are at least two pH units apart).
NOTE
A pH 7.0 buffer solution reads a mV value of approxi-
mately zero, and pH buffers read approximately 59.1
mV for each pH unit above or below pH 7.0. Check the
pH buffer manufacturer specifications for millivolt values
at various temperatures since it may affect the actual
value of the buffer solution mV/pH value.
1. Immerse sensor in the first buffer solution. Allow sensor
to adjust to the buffer temperature (to avoid errors due to
temperature differences between the buffer solution and
sensor temperature) and wait for reading to stabilize.
Value of buffer can now be acknowledged by
analyzer/transmitter.
2. Once the first buffer has been acknowledged by the ana-
lyzer/transmitter, rinse the buffer solution off of the sensor
with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solution.
4. Once the analyzer/transmitter has acknowledged both
buffer solutions, a sensor slope (mV/pH) is established (the
slope value can be found within the analyzer/ transmitter).
5. The slope value should read about 59.1 mV/pH for a new
sensor and will decrease over time to approximately 47-
49 mV/pH. Once the slope reads below the 47-49 mV/pH
range, a new sensor should be installed to maintain accu-
rate readings.
Recommended pH Sensor Standardization:
For maximum accuracy, the sensor can be standardized online
or with a process grab sample after a buffer calibration has
been performed and the sensor has been conditioned to the
process. Standardization accounts for the sensor junction
potential and other interferences. Standardization will not
change the sensor’s slope but will simply adjust the analyzer’s
reading to match that of the known process pH.
pH 4 Solution pH 7 Solution
Temp °°C20 25 30 20 25 30
Millivolt Potential 268 264 260 94 87 80
TABLE 5-1. ORP of Saturated Quinhydrone
Solution (In Millivolts)
24
25
MODEL 396P and 396PVP SECTION 6.0
MAINTENANCE
The Model 396P and 396PVP Sensors require mini-
mum maintenance. The sensor should be kept clean
and free of debris and sediment at all times. The fre-
quency of cleaning by wiping or brushing with a soft
cloth or brush is determined by the nature of the solu-
tion being measured. The sensor should be removed
from the process periodically and checked in buffer
solutions.
WARNING
BEFORE REMOVING THE SENSOR, be
absolutely certain that the process pres-
sure is reduced to 0 psig and the
process temperature is lowered to a
safe level!
If the sensor will not calibrate, refer to your analyzer/
transmitter instruction manual for proper test proce-
dures. If it is determined that the sensor has failed, it
should be discarded and replaced.
6.1 AUTOMATIC TEMPERATURE
COMPENSATOR.
The temperature compensator element is temp-
erature sensitive and can be checked with an ohm-
meter. Resistance increases with temperature.
The 3K element will read 3000 ohms ± 1% at 25°C
(77°F) and a Pt100 will read 110 ohms. Resistance
varies with temperature for a 3K and Pt100 element
and can be determined according to Table 6-2 or the
following formula:
RT=Ro[l+R1(T-20)]
Where RT= Resistance
T = Temperature in °C
Refer to Table 6-1 for Roand R1values
6.2 MODEL 396P & 396PVP pH SENSORS
4.2.1 ELECTRODE CLEANING. If the electrode is
coated or dirty, clean as follows:
1. Remove the sensor from process.
2. Wipe the glass bulb with a soft, clean, lint free
cloth or tissue. If this does not remove the dirt or
coating, go to Step 3. (Detergents clean oil and
grease; acids remove scale.)
3. Wash the glass bulb in a mild detergent solution
and rinse it in clean water. If this does not clean
the glass bulb, go to Step 4.
SECTION 6.0
MAINTENANCE
CAUTION
The solution used during the following
check is an acid and should be handled
with care. Follow the directions of the acid
manufacturer. Wear the proper protective
equipment. Do not let the solution come in
contact with skin or clothing. If contact with
skin is made, immediately rinse with clean
water.
4. Wash the glass bulb in a dilute 5% hydrochloric
acid solution and rinse with clean water. Soaking
the sensor overnight in the acid solution can
improve cleaning action.
NOTE
Erroneous pH results may result immedi-
ately after acid soak, due to reference
junction potential build-up.
Replace the sensor if cleaning does not
restore sensor operation.
TABLE 6-1
Roand R1VALUES FOR TEMPERATURE
COMPENSATION ELEMENTS
Resistance
Temperature °C (Ohms) ±1%
3K Pt100
0 2670 100.0
10 2802 103.8
20 2934 107.7
25 3000 109.6
30 3066 111.5
40 3198 115.4
50 3330 119.2
60 3462 123.1
70 3594 126.9
80 3726 130.8
90 3858 134.6
100 3990 138.5
Temperature
Compensation Element
RoR1
3K 2934 .0045
Pt100 107.7 .00385
TABLE 6-2
TEMPERATURE vs RESISTANCE OF AUTO
T.C. ELEMENTS
26
MODEL 396P and 396PVP SECTION 6.0
MAINTENANCE
6.3 MODEL 396P and 396PVP ORP
4.3.1 Platinum Electrode Check. The platinum elec-
trode may be checked as follows: There are two types
of standard solutions which may be used to check the
ORP electrode/transmitter system.
Type 1: One type of commonly used ORP standard
solution is the saturated quinhydrone solution. Refer
to Section 5.2.
CAUTION
The solution used during the following
check is an acid and should be handled
with care. Follow the directions of the acid
manufacturer. Wear the proper protective
equipment. If contact with skin of clothing is
made, immediately rinse with plenty of
clean water.
Type 2: A second ORP standard solution is the Ferric-
Ferrous Ammonium Sulfate Solution (PN R508-16OZ),
and it can be ordered as a spare part; otherwise, it can
be prepared from the following recipe: Dissolve 39.2
grams of reagent grade ferrous ammonium sulfate,
Fe(NH4)2(SO4)2• 6H2O and 48.2 grams of reagent
grade ferric ammonium sulfate, FeNH4(SO4)2
12H2O, in approximately 700 milliliters of water (dis-
tilled water is preferred, but tap water is acceptable).
Slowly and carefully add 56.2 milliliters of concentrated
sulfuric acid. Add sufficient water to bring the total solu-
tion volume up to 1000 ml. This standard ORP solution,
although not as simple to prepare as the quinhydrone
recipe, is much more stable, and will maintain its milli-
volt value for approximately one year when stored in
glass containers. This solution (ferric/ferrous ammoni-
um sulfate) will produce a nominal ORP of 476 +20 mV
at 25°C when used with a saturated KCl/AgCl refer-
ence electrode and platinum measuring electrode.
Some tolerance in mV values is to be expected due to
the rather large liquid reference junction potentials
which can arise when measuring this strongly acidic
and concentrated solution. However, if the measuring
electrodes are kept clean and in good operating condi-
tion, consistently repeatable calibrations can be carried
out using this standard solution.
6.3.2 Cleaning Platinum Electrode. The electrode can
be restored to normal operation by simply cleaning the
platinum electrode with baking soda. Polish it by rubbing
it with a damp paper towel and baking soda until a bright,
shiny appearance is attained.
27
MODEL 396P and 396PVP SECTION 7.0
DIAGNOSTICS AND TROUBLESHOOTING
SECTION 7.0
DIAGNOSTICS AND TROUBLESHOOTING
7.1 MODEL 54e/81/3081/4081/5081/Xmt DIAGNOSTICS AND TROUBLESHOOTING
The Model 54 Analyzer and Models 81 and 3081 Transmitters automatically search for fault conditions that would
cause an error in the measured pH value, as does the Model 1054B pH/ORP Analyzer to a lesser degree. Refer to
the applicable Instruction Manual for a complete description of the analyzer’s fault conditions.
Table 7-1, below, lists the Model 54’s, 4081’s, 3081’s, and 81’s diagnostic messages that indicate a possible sen-
sor problem. A more complete description of the problem and a suggested remedy corresponding to each mes-
sage is also listed.
DIAGNOSTIC MESSAGE
54
3081/81/4081 DESCRIPTION OF PROBLEM REMEDY
“Calibration Warning” 1. Aged glass. 1. Perform buffer calibration.
2. Sensor not immersed. 2. Be sure electrode measuring tip is in
/-[5ES-U1 process.
“Cracked glass failure” Broken or cracked glass. Replace sensor.
([-<<!I-5[
“High reference impede” 1. Liquid junction coated. 1. Clean sensor; replace if necessary.
2. Reference Cell gel depleted. 2. Replace sensor.
S12!I-5[ or S12!@M-SP 3. Sensor not immersed. 3. Be sure electrode measuring tip is in
process.
“Input voltage high” pH input shorted or sensor Check wiring. Replace sensor if
“Input voltage low” miswired. necessary.
“Old glass warning” 1. Glass electrode worn out. 1. Replace sensor.
2. Sensor not immersed. 2. Be sure electrode measuring tip is in
([D<<!@M-SP process.
“Reference offset err” Reference electrode poisoned. Replace sensor.
(offline only)
<UG!1SS
“Ref voltage high” 1. Reference shorted or sensor Check wiring and installation. Replace
“Ref voltage low” miswired. sensor if necessary.
2. Sensor not immersed
“Sensor line open” 1.
Open wire between sensor and analyzer.
1. Check sensor wiring.
2. Interconnecting cable greater than 2. Relocate analyzer.
[5P1!2-5[ 1000 ft.
“Sensor miswired” 1.
Open wire between sensor and analyzer.
1. Check wiring.
2. Bad preamplifier. 2. Replace preamplifier. (Code 02 only)
“Temp error high” 1. Open or shorted RTD. 1. Replace sensor.
“Temp error low” 2. Temperature out of range. 2. Check process temperature.
U17:!45
U17:![9
TABLE 7-1. Troubleshooting with Advanced Diagnostics
28
7.2 TROUBLESHOOTING WITHOUT ADVANCED DIAGNOSTICS.
Table 7-2, below, lists common problems, causes and remedies typically encountered in process measurement.
Problem Probable Cause Remedy
Meter reads off scale. (Display Defective preamplifier Replace preamplifier (for code 02
reads overrange).
sensors). For code 01, replace sensor.
T.C. element shorted Check T.C. element as instructed
in Section 6.1 and
replace sensor if defective.
Sensor not in process. Sample Make sure sensor is in process with
stream is low or air bubbles are sufficient sample stream (refer to
present. Section 2.0 for installation details).
Open glass electrode Replace sensor.
Reference element open - no contact
Replace sensor.
Display reads between 3 and 6 pH Electrode cracked Replace sensor.
regardless of actual pH of solution
or sample.
Meter or display indication swings T.C. element shorted Check T.C. element as instructed
or jumps widely in AUTO T.C. Mode. in Section 6.1 and replace
sensor if defective.
Span between buffers extremely T.C. element open Check T.C. element as instructed
short in AUTO T.C. Mode. in Section 6.1 and replace sensor
if defective.
Sluggish or slow meter indication Electrode coated Clean sensor as instructed in
for real changes in pH level. Sections 6.2 or Section 6.3.2.
Replace sensor if cracked.
Electrode defective Replace sensor.
Transmitter cannot be standardized. Electrode coated or cracked Clean Sensor as instructed in
Sections 6.2 or Section 6.3.2
Replace sensor if cracked.
Defective preamplifier Replace preamplifier.
Transmitter short spans between Aged glass electrode or high Replace sensor.
two different buffer values. temperature exposure
Electrode coated Clean Sensor as instructed in
Section 6.2 or Section 6.3.2.
Replace sensor if cracked.
Air bubbles trapped in sensor end Shake the sensor in solution. See
between glass bulb and sensor body Section 2.0 for mounting guidelines.
TABLE 7-2. Troubleshooting without Advanced Diagnostics
MODEL 396P and 396PVP SECTION 7.0
TROUBLESHOOTING
29
MODEL 396P and 396PVP SECTION 7.0
TROUBLESHOOTING
TABLE 7-3. Model 396P and 396PVP pH/ORP Replacement Parts and Accessories
PN DESCRIPTION QUANTITY
11275-01 Sensor Handrail Mounting Assembly
2002011 Flow Cell, CPVC, 1 inch FNPT
23242-02 Mounting Adapter, Insertion, 1¼-inch MNPT (304 S.S.) X 1” FNPT (PEEK)
23309-03 Junction Box, for remote preamplifier Code-50
23309-04 Junction Box, for remote preamplifier Code-54
23646-01 Cable, Extension (Prepped) for Models 54, 81, 3081, 4081, and 5081
23555-00 Junction Box with preamplifier, Models 54, 81, 3081, 4081, and 5081 compatible
23557-00 Preamplifier, remote for Junction Box, Models 54, 81, 3081, 4081, and 5081 compatible
22698-00 Preamplifier, Plug-in, Model 1003 compatible (for Code 02-50) 1
22698-02 Preamplifier, Plug-in, Models 1181 and 1050 compatible (for Code 02-50) 1
22698-03 Preamplifier, Plug-in, Models 1054, 1054A, 1054B, 2054, and 2081 compatible (for Code 02-54) 1
22719-02 Junction Box, w/o Preamplifier
33081-00 Adapter Insert, PEEK, 1 X 3/4-inch, for 23242-02
7901631 Shroud, PVC
9200254 Cable, 4 conductor, 22 AWG, shielded pair, for 1054/A/B, 2054, and 1181
9200273 Cable, Extension (Unprepped) for Models 54, 81, 3081, 4081, and 5081
23645-06 15 ft (4.6 m) cable with mating VP connector with BNC on transmitter end
23645-07 15 ft (4.6 m) cable with mating VP connector with bare wires on transmitter end
9210012 Buffer Solution, 4.01pH, 16 oz 4
9210013 Buffer Solution, 6.86pH, 16 oz 4
9210014 Buffer Solution, 9.18pH, 16 oz 4
9322014 Union, KYNAR1
9320057 Union, PVC
9120516 BNC Adapter
915240-04 Tee, Flow-through, 2” PVC, 1” NPT
9550175 O-ring for Mounting Adapter (23242-02)
R508-160Z ORP Standard Solution, 460mV ±10 at 20°C
23550-00 Junction Box with Extension Board, Models 54, 81, 3081, 4081, and 5081 compatible
661-898695 Cable 5 Conductor (for Model 2700 only)
30
MODEL 396P and 396PVP SECTION 8.0
RETURN OF MATERIAL
SECTION 8.0
RETURN OF MATERIAL
8.1 GENERAL.
To expedite the repair and return of instruments, prop-
er communication between the customer and the fac-
tory is important. Before returning a product for
repair, call 1-949-757-8500 for a Return Materials
Authorization (RMA) number.
8.2 WARRANTY REPAIR.
The following is the procedure for returning instru-
ments still under warranty:
1. Call Rosemount Analytical for authorization.
2. To verify warranty, supply the factory sales order
number or the original purchase order number.
In the case of individual parts or sub-assemblies,
the serial number on the unit must be supplied.
3. Carefully package the materials and enclose
your “Letter of Transmittal” (see Warranty). If
possible, pack the materials in the same manner
as they were received.
4. Send the package prepaid to:
Rosemount Analytical Inc., Uniloc Division
Uniloc Division
2400 Barranca Parkway
Irvine, CA 92606
Attn: Factory Repair
RMA No. ____________
Mark the package: Returned for Repair
Model No. ____
8.3 NON-WARRANTY REPAIR.
The following is the procedure for returning for repair
instruments that are no longer under warranty:
1. Call Rosemount Analytical for authorization.
2. Supply the purchase order number, and make
sure to provide the name and telephone number
of the individual to be contacted should addition-
al information be needed.
3. Do Steps 3 and 4 of Section 8.2.
NOTE
Consult the factory for additional informa-
tion regarding service or repair.
WARRANTY
Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured
or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira-
tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen
(18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,
membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90
days from the date of shipment or provision.
Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by
the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial
effort to arrange for procurement and shipping of the Resale Products.
If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at
its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man-
ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por-
tion of the Goods/Services.
All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit-
able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other
cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli-
gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by
an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per-
sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri-
od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only
in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE
ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT-
NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair
Model
_______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information (make a
copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or non-warranty service is requested.
5. Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges
for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Credit Cards for U.S. Purchases Only.
The right people,
the right answers,
right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
Emerson Process Management
Liquid Division
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com