Flowline LRXX Echo Pulse Radar Level Transmitter User Manual
Flowline, Inc. Echo Pulse Radar Level Transmitter
Flowline >
User Manual
Rev 9_4 EchoPulse™ Radar Level Transmitter EL EA SE LR10, LR15, LR20, LR25 & LR30 Series Manual Flowline Inc. 10500 Humbolt Street Los Alamitos, CA 90720 Tel: (562) 598‐3015 Fax: (562) 431‐8507 www.flowline.com INTRODUCTION Section One EL EA SE Table of Contents Page Section One | Introduction: .................................................................................................................................................. 2 Table of Contents: ..................................................................................................................................................... 2 Sensor Models: .......................................................................................................................................................... 4 Operating Principle: .................................................................................................................................................. 4 Features: ................................................................................................................................................................... 4 Benefits: .................................................................................................................................................................... 4 Limitations: ................................................................................................................................................................ 4 Specifications: ........................................................................................................................................................... 5 Dimensions: ............................................................................................................................................................... 8 Safety Precautions: ................................................................................................................................................. 11 Section Two | Getting Started: ........................................................................................................................................... 12 Setup Overview: ...................................................................................................................................................... 12 Part Number: ........................................................................................................................................................... 14 Section Three | Install Sensor: ............................................................................................................................................ 16 Installation Requirements: ...................................................................................................................................... 16 LR15 Antenna Preparation: ..................................................................................................................................... 17 Mounting Position: .................................................................................................................................................. 18 Flange Riser Installation: ......................................................................................................................................... 20 Stand Pipe Installation: ........................................................................................................................................... 22 Sight Glass Installation: ........................................................................................................................................... 23 LR30 Sensor Installation: ......................................................................................................................................... 24 LR98 Display Installation: ........................................................................................................................................ 25 Section Four | Wire Sensor: ................................................................................................................................................ 26 Terminal Wiring: ..................................................................................................................................................... 26 LR30 Sensor to LR98 Display: .................................................................................................................................. 27 Wiring to Displays, Controllers & PLCs: ................................................................................................................... 28 Section Five | Configuration: .............................................................................................................................................. 30 Basic Configuration Overview: ................................................................................................................................ 30 Using the Display: .................................................................................................................................................... 31 Changing Display Values: ........................................................................................................................................ 32 Step 1 ‐ Measure the Tank: ..................................................................................................................................... 33 Step 2 ‐ Set the Units of Measurement: ................................................................................................................. 34 Step 3 ‐ Set the Empty Configuration (4mA): .......................................................................................................... 35 Step 4 ‐ Set the Full Configuration (20mA): ............................................................................................................ 36 Step 5 ‐ Set the Range (Maximum Range): ............................................................................................................. 37 Step 6 ‐ Set the Dead Band: .................................................................................................................................... 38 Step 7 ‐ Check the Echo Curve: ............................................................................................................................... 39 Section Six | Process Adjustments: ..................................................................................................................................... 40 Process Adjustments Overview: ............................................................................................................................. 40 Fast Filling or Emptying of Liquid: ........................................................................................................................... 41 Liquid Surface is Turbulent or Agitated: ................................................................................................................. 42 Foam on the Surface of the Liquid: ......................................................................................................................... 43 Sensor Installed in a Stand Pipe or Sight Glass: ...................................................................................................... 44 Page 2 of 72 MN301700 Rev 9_4 INTRODUCTION Section One EL EA SE Section Seven | Advanced Adjustments: ........................................................................................................................... 45 Advanced Adjustments Overview: .......................................................................................................................... 45 4‐20mA Reverse Output: ........................................................................................................................................ 46 Fail‐Safe Output: ..................................................................................................................................................... 47 Minimum Current Output: ...................................................................................................................................... 48 Create a New False Echo Curve: ............................................................................................................................. 49 Update an Existing False Echo Curve: ..................................................................................................................... 50 Section Eight | Troubleshooting: ......................................................................................................................................... 51 Troubleshooting Overview: ..................................................................................................................................... 51 Measurement Status: .............................................................................................................................................. 52 Peak Values: ............................................................................................................................................................ 53 Simulation: .............................................................................................................................................................. 54 First Echo Adjustment: ............................................................................................................................................ 55 Echo Curve Zoom: ................................................................................................................................................... 56 False Echo Curve Delete: ......................................................................................................................................... 57 Reset: ...................................................................................................................................................................... 58 Section Nine | Appendix: ..................................................................................................................................................... 59 Configuration Menu: ............................................................................................................................................... 59 Empty Configuration: .............................................................................................................................................. 60 Full Configuration: ................................................................................................................................................... 60 Medium: .................................................................................................................................................................. 61 Low Dielectric Medium: .......................................................................................................................................... 62 Dampen: .................................................................................................................................................................. 63 Scaled Units: ............................................................................................................................................................ 63 Range: ..................................................................................................................................................................... 63 Dead Band: .............................................................................................................................................................. 63 Display Menu: ......................................................................................................................................................... 64 Display Value: .......................................................................................................................................................... 64 LCD Contrast: .......................................................................................................................................................... 64 Diagnostics Menu: ................................................................................................................................................... 65 Peak Values: ............................................................................................................................................................ 65 Measurement Status: .............................................................................................................................................. 65 Echo Curve: ............................................................................................................................................................. 66 Simulation: .............................................................................................................................................................. 66 Service Menu: ......................................................................................................................................................... 67 False Echo: ............................................................................................................................................................... 68 Output Settings: ...................................................................................................................................................... 68 Reset: ...................................................................................................................................................................... 69 Units of Measurement: ........................................................................................................................................... 69 Language: ................................................................................................................................................................ 69 Information: ............................................................................................................................................................ 70 Factory Settings: ...................................................................................................................................................... 70 User Configuration: ................................................................................................................................................. 70 Notes: ...................................................................................................................................................................... 71 Warranty: ................................................................................................................................................................ 72 Rev 9_4 MN301700 Page 3 of 72 INTRODUCTION Section One Sensor Models Offered in five different models, EchoPulse™ is a general‐purpose, two‐wire, pulse radar level sensor that provides a continuous 4‐20 mA current output that’s proportional to the liquid level in a tank or sump. Make sure that the model purchased is appropriate for your application. Series Max Range Material Mounting Application LR10 32.81’ (10m) PTFE 1‐1/2” NPT Corrosive liquids under simple process conditions Storage tanks & process tanks under difficult process conditions 98.42’ (30m) 316L SS LR20 65.61’ (20m) 316L SS with PTFE cover 3” ANSI flange 4” ANSI flange LR25 114.83’ (35m) 316L SS with PTFE cover 4” ANSI flange 6” ANSI flange LR30 98.42’ (30m) PA66 Bracket or top mounted (1”) conduit Aggressive liquids under extremely difficult process conditions Storage tank & process tanks under extremely difficult process conditions Water processing, lift stations, storm water and sump process conditions EA SE LR15 1‐1/2” NPT 3” ANSI flange 4” ANSI flange 6” ANSI flange EL Operating Principle The sensor emits a microwave pulse from its antenna, which travels at the speed of light to the surface of the medium below. A portion of that energy reflects off the medium and returns to the antenna. The time gap between energy emission and receipt is called the “time of flight”, and is proportional to the distance between the medium surface and the sensors measurement location, which is typically located at the bottom of the antenna. The sensor measures the time of flight and translates this value into a continuous 4‐20mA signal output that’s proportionate to level within a defined measurement span. Features Easy configuration with LCD push button display module Adjustable loop fail‐safe, no change, 20.5 mA, 22 mA Small 12” (30.48cm) dead band enables full tank measurement Recognition, storage and deletion of false echo signal returns Benefits Unaffected by physical process and environmental conditions Ideal for applications with higher temp, pressure, foam and vapor Strong signal penetrability with minimal attenuation over distance Limitations (factors that could influence performance) Air particulates with a high dielectric constant value such as lead or ferroalloy Highly dense air particulates that attenuate microwave emission and receipt Material build‐up on the antenna that degrades microwave emission and receipt Mediums with an extremely low dielectric constant value with little reflectivity Page 4 of 72 MN301700 Rev 9_4 INTRODUCTION Section One Specifications Measurement Range: Series Range (Max) Dead Band: LR10 32.81 ft (10m) LR15 98.42 feet (30m) LR20 65.61 feet (20m) LR30 98.42 feet (30m) 12” (30.48cm) / Factory Set Note: Can be lowered to 2” from the bottom of the antenna (see charts on Page 7) Measurement Accuracy: Series Accuracy LR10 ±5 mm Display Resolution: Frequency Range: LR15 ±3mm LR20 ±3mm LR25 ±10mm LR30 ±3mm 1 mm LR10, LR15, LR20, LR30: ...... 26 GHz LR25: ................................... 6.3 GHz About 1 sec (dependent on configuration settings) About 1 sec (dependent on configuration settings) See part number description LR15 LR20 LR25 316L SS Aluminum 316L SS 316L SS with PTFE Cover N/A Viton Silicone (between housing and cap) Polycarbonate Stainless Steel N/A EA LR10 N/A EL PTFE PBT‐FR Series Flange Enclosure Antenna Extension Seal Seal Ring Window Ground Terminal Bracket Process Connection: Material: SE Measurement Interval: Adjustment Time: LR30 N/A PA66 PA66 N/A 304 SS Depends on process connection size and housing Weight: LR25 114.83 feet (35m) Series Weight LR10 2.20 lbs (1kg) LR15 4.41 lbs (2kg) LR20 6.61 lbs (3kg) LR25 6.61 lbs (3kg) LR30 2.2 lbs (1kg) Temperature: Series Process Temp Storage Temp Relative Humidity: Rev 9_4 LR10 F: ‐40° to 266° C: ‐40° to 130° LR15 F: ‐76° to 302° C: ‐60° to 150° LR20 F: ‐40° to 302° C: ‐40° to 150° F: ‐40° to 176° C: ‐40° to 80° LR25 F: ‐40° to 266° C: ‐40° to 130° LR30 F: ‐40° to 212° C: ‐40° to 100° <95% MN301700 Page 5 of 72 INTRODUCTION Section One Process Pressure: Series Pressure LR10 ‐14.5 to 43.5 psi (‐1 to 3 bar) LR15 ‐14.5 to 150 psi (‐1 to 10.3 bar) LR20 ‐14.5 to 72.5 psi (‐1 to 5 bar) LR25 ‐14.5 to 580 psi (‐1 to 40 bar) LR30 Atmospheric Mechanical vibration 10m/s, 10m2/s, 10 ‐150 Hz Vibration Proof: Output: Signal Output: Resolution: Fail‐Safe Setting: Load Resistance: EA SE 4‐20mA 1.6µA 20.5mA, 22mA or no change See chart below 0‐40 sec, adjustable Cable Connection: 24 VDC (16 to 26 VDC) The same two‐wire connection cable carries power supply and current signal. 22.5mA maximum Cable Entry/Plug: Spring Connection Terminal: Certification: Compliance: Classification: Page 6 of 72 <1V <10mV LR10, LR15, LR20, LR25: ………… IP67 (NEMA 6) LR30: ………………………………………IP68 (NEMA 6P) Standard 2‐wire shielded cable with earth ground wire and outside diameter of 5‐9mm is recommended. LR10, LR15, LR20, LR25: ............. Dual cable entry (½” NPT with adapter, M20x1.5) LR30: ........................................ One cable entry (1” NPT) Applicable for cables with cross section if 2.5mm2 General Purpose: cTUVus (UL 61010‐1:2012 & CAN/CSA‐C22.2 No. 61010‐1‐12) Communication: FCC (US) CE [EN61326 (1997 w/A1:98, A2:01 & A3:03) Class A] RoHS General Purpose Power Consumption: Ripple Allowed: <100Hz: 100 to 100 KHz: Enclosure Rating: EL Integration Time: Power: Power Supply: MN301700 Rev 9_4 INTRODUCTION Section One EL EA SE Accuracy Charts Rev 9_4 MN301700 Page 7 of 72 INTRODUCTION Section One LR15 Series (Threaded) EA SE LR10 Series LR15 Series (Threaded) Antenna Dimensions Diameter (X) Length (Y) 2” (48mm) 5.51” (140mm) 3” (78mm) 8.94” (227mm) 4” (98mm) 11.34” (288mm) EL LR15 Series (Flange) LR15 Series (Flange) Dimensions Flange (A) Diameter (B) Thickness (C) 3” ANSI 7.5” (190.5mm) 0.88” (22.3mm) 4” ANSI 9.0” (228.6mm) 0.88” (22.3mm) 6” ANSI 11.0” (279.4mm) 0.94” (23.9mm) LR15 Series (Flange) Antenna Dimensions Diameter (D) Length (E) Page 8 of 72 MN301700 2” (48mm) 6.36” (161.5mm) 3” (78mm) 9.78” (248.5mm) 4” (98mm) 12.18 (309.5mm) Rev 9_4 INTRODUCTION Section One LR25 Series EL EA SE LR20 Series LR20 Series Flange / Antenna Dimensions Flange Diameter Thickness Length (A) (B) (C) (D) 3” 7.5” 0.88” 6.50” ANSI (190.5mm) (22.3mm) (165mm) 4” 9.0” 0.88” 8.00” ANSI (228.6mm) (22.3mm) (203mm) Rev 9_4 LR25 Series Flange / Antenna Dimensions Flange Diameter Thickness Length (A) (B) (C) (D) 4” 9.0” 0.57” 5.71” ANSI (228.6mm) (14.5mm) (145mm) 9.75” 6” 11.0” 0.63” (247.7mm) ANSI (279.4mm) (16.1mm) MN301700 Page 9 of 72 INTRODUCTION Section One LR30 Series Bracket (Top View) LR30 Series (Side View A) EL EA SE LR30 Series (Side View B) LR98 Series Display (Front View) Page 10 of 72 LR98 Series Display (Bottom View) MN301700 LR98 Series Display (Side View) Rev 9_4 INTRODUCTION Section One Safety Precautions About this Manual: PLEASE READ THE ENTIRE MANUAL PRIOR TO INSTALLING OR USING THIS PRODUCT. This manual includes information on the EchoPulse™ Radar Level Transmitter from FLOWLINE. Please refer to the part number located on the sensor label to verify the exact model, which you have purchased. User’s Responsibility for Safety: FLOWLINE manufactures a broad range of level sensing technologies. While each of these sensors is designed to operate in a wide variety of applications, it is the user’s responsibility to select a sensor model that is appropriate for the application, install it properly, perform tests of the installed system, and maintain all components. The failure to do so could result in property damage or serious injury. Proper Installation and Handling: Only professional staff should install and/or repair this product. Never over tighten the sensor within the fitting. Always check for leaks prior to system start‐up. Wiring and Electrical: A supply voltage of 16 to 26 VDC is used to power the EchoPulse™. Electrical wiring of the sensor should be performed in accordance with all applicable national, state, and local codes. EA SE Material Compatibility: The enclosure is made of either Aluminum or 316 Stainless Steel (refer to sensor part number). The antenna is made of Stainless Steel (SS), Polytetrafluoroethylene (PTFE) or Nylon (PA66) with a Viton seal (refer to sensor part number). Make sure that the model, which you have selected, is chemically compatible with the application media. EL Enclosure: The sensor housing is liquid‐resistant, but is not designed to be operational when immersed (LR30 series is designed for submersion but will not provide a true level reading while submersed). Mount the sensor in such a way that the enclosure and antenna do not come into contact with the application media under normal operational conditions. The enclosure has a cover that provides access to the push button display module and terminal strip for wiring. To open the enclosure, you will need to twist the cover counter‐clockwise. Before closing the enclosure, make sure that the enclosure gasket is properly seated, and that any conduit fittings, cable connectors or plugs are installed correctly and sealed. Note: If using the Flowline LM90‐1001 (liquid tight fitting) on the ½” conduit, the cable minimum is 0.170” (4.3mm) and the maximum is 0.450” (11.4mm). Make a Fail‐Safe System: Design a fail‐safe system that accommodates the possibility of sensor and/or power failure. FLOWLINE recommends the use of redundant back‐up systems and alarms in addition to the primary system. Flammable, Explosive or Hazardous Applications: EchoPulse™ is approved for use within general purpose applications ONLY and should NOT be used within classified hazardous environments. Handling Static‐Sensitive Circuits and Devices: When handling the instrument (part), the technician should follow the below guidelines to reduce the possibility of an electrostatic charge build‐up on the technician’s body from being transferred to the electronic part. Always touch a known good ground source before handling a part. This should be repeated while handling the part and more frequently after sitting down from a standing position, sliding across the seat or walking a distance. Avoid touching electrical terminals of the part unless making connections. DO NOT open the unit cover until it is time to work on the part. Rev 9_4 MN301700 Page 11 of 72 GETTING STARTED Section Two Setup Overview The below highlights the initial steps in setting up your sensor for operation. EL EA SE 1. Part Number (Section Two) 1. Prior to purchasing the sensor, you should have submitted a Level Application Questionnaire (www.flowline.com/LAQ), which based upon the information provided, may have resulted in a suggested part number. Where so, confirm that the suggested part number matches the part number of the purchased sensor. If any of the above does not match and/or meet your application requirements, please contact your distributor. 2. Install Sensor (Section Three) 1. Information on the location and mechanical installation of the sensor. 3. Wire Sensor (Section Four) 1. Information on the electrical wiring and power requirements of the sensor. 4. Basic Configuration (Section Five) 1. Begin by measuring the tank for all key dimensions. a. Accuracy in measurement will result in accuracy of sensor performance. 2. Set the Units of Measurement for the sensor. a. Units can be configured in basic engineering units of length: Feet, Meters 3. Set the Empty Configuration for the sensor in the tank. a. This is the 4mA setting for the output. 4. Set the Full Configuration for the sensor in the tank. a. This is the 20mA setting for the output. 5. Set the Range (Maximum Range or MaxR) for the sensor in the tank. a. The sensor will ignore any echo signal returns beyond this setting. 6. Set the Dead Band (Minimum Range or MinR) for the sensor in the tank. a. The sensor will ignore any echo signal returns closer than this setting. 7. Check the Echo Curve a. This is a quick check to determine if the sensor is reading the correct level. 5. Process Adjustments (Section Six) 1. Information on OPTIONAL adjustments for specific process conditions that may exist in your application. a. Fast filling or emptying of liquid. b. Liquid surface is turbulent or agitated. c. Foam on the surface of the liquid. d. Sensor installed in a still well or sight glass. Page 12 of 72 MN301700 Rev 9_4 GETTING STARTED Section Two EL EA SE 6. Advanced Adjustments (Section Seven) 1. Reverse 4‐20 mA Output a. Reverses the current out from 4mA @ bottom and 20mA @ top of tank to 20mA @ bottom and 4mA @ top of the tank. 2. Fail‐Safe Setting a. Allows for the presetting of the current output when a sensor failure occurs. 3. Minimum Current Setting a. Sets the minimum current output for the sensor. 4. Create a New False Echo Curve a. A method to map out false echo signal returns within the tank. 5. Update an Existing False Echo Curve a. A method to update false echo signal returns for a section of the tank that was not exposed during the creation of the original False Echo Curve. 7. Troubleshooting (Section Eight) 1. Measurement Status a. Determines the measurement reliability and general status of the sensor. 2. Peak Values a. Displays the lowest and highest level height that the sensor has measured in distance (d). 3. Simulation a. Simulates and helps to determine the accuracy and linearity of the sensor. 4. First Echo Adjustment a. Increases or decreases the strength of the first echo signal return. 5. Echo Curve Zoom In a. A method to zoom in and view the Echo Curve over a specific range. 6. False Echo Curve Delete a. A method to delete a previously saved False Echo Curve from memory. 7. Reset a. A method to reset the sensor’s configuration to the original factory setting. Rev 9_4 MN301700 Page 13 of 72 GETTING STARTED Section Two Part Number Prior to purchasing the sensor, you should have submitted a Level Application Questionnaire (www.flowline.com/LAQ), which based upon the information provided, may have resulted in a suggested part number. Where so, confirm that the suggested part number matches the part number of the sensor. The part number can be found on the outside label on the sensor as shown below: LR15 Series LR20 Series LR25 Series LR30 Series LR10 Series LR10 ‐ 0 EA SE The part number will indicate the size and type of mounting fitting required for installing the sensor. Refer to the below part number description for specific information. If any of the above does not match and/or meet your application requirements, please contact your distributor. EL Housing Material 0 ‐ Aluminum Process Connection 0 ‐ Thread 1 ½" NPT Output 1 ‐ 4‐20 mA Approval 0 ‐ General Purpose (cTUVus) LR15 ‐ 0 Page 14 of 72 ‐ Housing Material 0 ‐ Aluminum Process Connection 0 ‐ Thread 1 ½" NPT 3 ‐ ANSI Flange Output 1 ‐ 4‐20 mA Approval 0 ‐ General Purpose (cTUVus) Antenna Size 2 ‐ 2" (48mm) Horn 3 ‐ 3" (78mm) Horn 4 ‐ 4" (98mm) Horn Flange Size 0 ‐ Threaded Conn. MN301700 Rev 9_4 GETTING STARTED Section Two LR20 ‐ 0 1 0 ‐ Housing Material 0 ‐ Aluminum Process Connection 3 ‐ ANSI Flange Output 1 ‐ 4‐20 mA Approval 0 ‐ General Purpose (cTUVus) Flange Size 3 ‐ 3" Flange (ANSI) 4 ‐ 4" Flange (ANSI) LR25 ‐ 0 1 0 ‐ EL EA SE Housing Material 0 ‐ Aluminum Process Connection 3 ‐ ANSI Flange Output 1 ‐ 4‐20 mA Approval 0 ‐ General Purpose (cTUVus) Flange Size 4 ‐ 4" Flange (ANSI) 6 ‐ 6" Flange (ANSI) LR30 ‐ 0 ‐ Housing Material 0 ‐ Nylon (PA66) Mount 0 ‐ Bracket Output 1 ‐ 4‐20 mA Approval 0 ‐ General Purpose (cTUVus) Display (LI98‐1001) 1 ‐ With Display Rev 9_4 MN301700 Page 15 of 72 INSTALL SENSOR Section Three Installation Requirements EchoPulse™ measures the distance between the sensor and the liquid surface below. Typically, all measurements from the sensor originate from the bottom of the antenna. Refer to the Measurement Reference Chart to determine the location where measurement originates on your sensor. To ensure reliable measurement, adhere to the following minimum installation requirements: EA SE 1. There are no obstructions between the bottom edge of the installed antenna and the surface of the liquid below including ladders, walls, tank seams, liquid inflows, rails, other sensors, mixer blades, heating coils, pumps, struts or apparatus. Note: Additionally, when the sensor transmits a microwave pulse, the RF signal spreads in a conical shape (determined by its beam angle) over distance. Refer to the Beam Angle Chart to determine, what if any, additional measurement space is required to be free of such obstacles. If such items are present, then a False Echo Curve configuration must be done (Section Seven). 2. The sensor must be installed with the antenna perpendicular to the surface of the liquid. 3. The sensor must be installed with a distance ≥ 19.7” (500mm) from the side wall of the tank. 4. The liquid level must not be allowed to enter into the dead band (blanking zone) of the sensor. 5. The sensor installation must be done in accordance with relevant local or federal safety regulations. 6. The sensor must be connected to electrical ground. 7. Do not use the housing to screw the sensor into the installation fitting (LR10 & LR15 Series). 1. Applying a tightening force against the housing may damage the sensor. 8. Make sure that all parts of the sensor exposed to the application, specifically any portion installed within the tank, are suitable for the process. 1. Consider any effects from the application temperature, pressure or media. EL FCC Conformity This instrument complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this instrument may not cause harmful interference, and; (2) this instrument must accept any interference received, including interference that may cause undesired operation. Changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment. Warning: User must keep a safety distance of at least 20cm from the antenna. NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Page 16 of 72 This equipment is not allowed to be connected to public utility power lines. MN301700 Rev 9_4 INSTALL SENSOR Section Three LR15 Antenna Preparation The LR15 Series antenna (only) may be removed from the sensor to allow a flange or reducer bushing accessory to be attached to the 1 ½” NPT mounting threads and/or, the antenna may be inserted from within the inside of the tank through the bottom of an existing fitting (where the base of the antenna is too wide to pass through the fitting from the top). Referencing the illustrations, follow the below steps to disconnect, mount and reattach the antenna. 1) Loosen and remove the four (4) socket screws using a 3mm Allen wrench. 2) Carefully remove the antenna. Note: Do not remove or damage the plastic cone within the antenna socket. 3) Insert the antenna through the bottom of the fitting. Note: If doing so from the inside of the tank, make sure to secure it, so as to prevent the antenna from falling into the tank. 4) Connect the sensor to the antenna socket and reattach the four (4) screws using a 3mm Allen wrench. 5) Attach the sensor to the fitting as necessary. Connect Antenna to Sensor, Attach Screws Remove Antenna Thread Sensor to Flange Rev 9_4 Remove Antenna Remove Screws EL EA SE Remove Screws Add a Flange Insert Antenna Add a Reducer Bushing Insert Antenna MN301700 Connect Antenna to Sensor, Attach Screws Thread Sensor to Reducer Bushing Page 17 of 72 INSTALL SENSOR Section Three Mounting Position The minimum distance (independent of beam angle) that the sensor can be mounted next to the straight side wall of the tank is 19.7” (500mm) as measured from the sensor centerline to the side wall. If you are not able to install the sensor more than 19.7” (500mm) away from the side wall, or if there is material build up on the side wall (within the beam angle), perform a False Echo Curve during initial configuration. Avoid mounting the sensor in the center of a dome top tank. The center of such a tank will multiply the echoes, making sensor operation difficult. Correct Mounting Sensor mounted off center in a dome top tank. EL EA SE Incorrect Mounting Sensor mounted in the center of a domed top tank resulting in multiple echoes. In cone bottom tanks, it can be advantageous to mount the sensor in the center of the tank, making it possible for the sensor to measure closer to the bottom of the tank. If the sensor is mounted over an angled bottom, and the level drops below the angle, the echo will be deflected away from the sensor, resulting in poor operation. The sensor can be mounted over an angled bottom as long as the level is maintained within the straight side wall so the sensor will receive echo returns. Center Mount ‐ Correct Page 18 of 72 Level Below Side Wall ‐ Incorrect MN301700 Level Above Side Wall – Correct Rev 9_4 INSTALL SENSOR Section Three Mount the Sensor Perpendicular to the Liquid Level Always mount the sensor perpendicular to the surface of the liquid. This will enable the return echoes to reach the sensor. Mounting the sensor off‐axis will result in weaker return echoes or no return echoes, depending on the degree of angle. EL EA SE Consider the Dead Band The sensor has a dead band (blanking distance) of 12” (30.48mm) as its default. The dead band can be lowered to within 2” of the bottom of the antenna (consult with factory). This is an up close distance where the sensor is not able to measure the level within this range. Typically, the measurement location for the sensor is at the bottom of the antenna. When identifying a location for sensor installation, take into account the length of the antenna combined with the dead band of the sensor. Avoid Condensation in the Conduit You can give your instrument additional protection against moisture penetration by leading the conduit connection or cable downward in front of the cable entry. Condensation in the conduit will thus not enter the sensor enclosure. Rev 9_4 MN301700 Page 19 of 72 INSTALL SENSOR Section Three Avoid Obstructions in the Beam Path Do not mount the sensor in or above the fill stream, other equipment (ladders, pumps, mixers) or structures within the beam path of the sensor. Such items can create false echo returns and prevent the actual level from being seen by the sensor. Find a location where the sensor has a clear view of the liquid surface. If your tank has other equipment near or within the beam path of the sensor, a False Echo Curve should be performed during initial configuration. Fill Stream Mounting Mounting on left incorrectly positions sensor above the tank fill stream inlet. Mounting on right is correct as the sensor has an unobstructed view to the liquid level below. SE Reflector Installation Mounting on left incorrectly allows the sensor to receive false echo returns from the step. EL EA Mounting on right has an angled baffle‐board mounted over the step, which prevents the false echo from returning to the sensor. As such, the sensor only receives correct echo returns from the liquid level. Flange Riser Installation When installing the sensor on a flange with a riser (or any fitting that is tall and narrow), the antenna must protrude at least 0.4” (10mm) from the bottom of the riser. The sensor can be installed within the riser as long as the liquid has a strong reflective property (dielectric constant) providing a strong echo return. The below information describes the maximum distance that the antenna can be recessed within a riser based on the diameter and height of the fitting. Antenna Extension LR15 Series LR10 & LR15 Series Height (H) Diameter (D) 1‐½” 7.9” (200mm) 2” (50mm) 9.8” (250mm) 3” (80mm) 11.8” (300mm) 4” (100mm) 19.7” (500mm) 6” (150mm) 31.5” (800mm) LR20 Series LR20 & LR25 Series Diameter (D) Height (H) 3” (80mm) 11.8” (300mm) 4” (100mm) 19.7” (500mm) 6” (150mm) 31.5” (800mm) Page 20 of 72 MN301700 Rev 9_4 INSTALL SENSOR Section Three EL EA SE Agitator or Mixer If there are agitators or mixers in the tank, the sensor should be mounted as far away from the blades as possible. Once the installation is complete, a False Echo Curve should be preformed while the agitator or mixer is in motion to map out and eliminate false echo returns from the blades. If significant foam and/or agitation exists within the application, a stand‐pipe installation should be considered. Rev 9_4 MN301700 Page 21 of 72 INSTALL SENSOR Section Three Stand Pipe Installation To avoid issues from turbulence, substantial foam or other equipment in the sensors beam path, install the sensor within a stand pipe (still well). A stand pipe installation can be used with liquids with a dielectric constant as low as 1.9. Note: The use of a stand pipe is not recommended with liquids that significantly coat or scale. As a rule, if the inside wall of the tank has material build‐up, then the inside of the stand pipe will also have build‐up that will affect the sensor’s operation. When installing a sensor in a stand pipe, follow the below guidelines: EL EA SE 1. The Full Configuration setting (20mA) must be below the upper vent hole and the bottom of the antenna. 2. The Empty Configuration setting (4mA) is typically placed at or near the bottom of the stand pipe. 3. The Stand Pipe function must be activated. It can be found under Medium in the Configuration Menu. The inner diameter of the stand pipe must be entered within this function. 4. It is recommended to perform a False Echo Curve when the sensor is installed in a stand pipe. Stand Pipe Construction 1. The stand pipe material must be metal with a smooth inner pipe wall. The minimum pipe size is dependent upon the Series and antenna length. The LR10 and LR15 sensor can be applied in pipe sizes ≤ 3” (76.2mm). 2. Any welded joint must be straight with a gap size ≤ 1/254” (0.1 mm). 3. Flanges should be welded to the stand pipe tube. 4. In the case of a pipe extension with a welded neck flange or pipe collar, make sure the inner surfaces are aligned and accurately joined together. 5. When securing the pipe to the tank, do not weld through the pipe wall. 1. Roughness on the inside caused by unintentional pipe penetration should be removed. 2. Not doing so will cause strong false echo returns and encourage buildup within the pipe. 6. The diameter of any holes along the pipe must be ≤ 1/5” (5 mm). 1. The top ventilation hole must be above the Full Configuration setting (20mA). 2. The holes must be vertically aligned on one side of the pipe with all burrs removed. 3. The number of holes does not matter. 7. The inner diameter of the pipe cannot change over the entire pipe length. Page 22 of 72 MN301700 Rev 9_4 INSTALL SENSOR Step Three Sight Glass Installation An alternative to a stand pipe is installing the sensor within a sight glass mounted outside of the tank. Sight glass installations can avoid issues from turbulence, substantial foam or other equipment in the sensors beam path. Note: The use of a sight glass is not recommended with liquids that significantly coat or scale. As a rule, if the inside wall of the tank has material build‐up, then the inside of the sight glass will also have build‐up that will affect the sensor’s operation. When installing a sensor in a sight glass, follow the below guidelines: SE 1. The Full Configuration setting (20mA) must be placed at or below the upper tank connection pipe. 2. The Empty Configuration setting (4mA) must be placed at or above the bottom tank connection pipe. 3. The Stand Pipe feature must be activated. It can be found under Medium in the Configuration Menu. The inner diameter of the sight glass must be entered within this function. 4. It is recommended to perform a False Echo Curve when the sensor is installed in a sight glass. EL EA Sight Glass Construction 1. The sight glass material must be metal with a smooth inner pipe wall. The minimum pipe size is dependent upon the Series and antenna length. The LR10 and LR15 sensor can be applied in pipe sizes ≤ 3” (76.2mm). 2. There is a minimum distance >11.8” (>300mm) between the bottom of the antenna and the top edge of the upper tank connection pipe. 3. Any welded joints must be straight with a gap size ≤ 1/254” (0.1 mm). 4. Flanges should be welded to the sight glass tube. 5. The inner diameter of the sight glass cannot change over the entire pipe length. Stand Pipe / Sight Glass Pipe Size vs. Series Series LR10: All LR15: 2” Antenna 3” Antenna 4” Antenna LR20: 3” Antenna 4” Antenna LR25: 4” Antenna 6” Antenna LR30: All Rev 9_4 Minimum Pipe 2” 2” 3” 4” 3” 4” 4” 6” N/A MN301700 Maximum Pipe 3” 8” 8” 8” N/A Page 23 of 72 INSTALL SENSOR Section Three LR30 Sensor Installation The LR30 is unique within the EchoPulse™ sensor family. It is designed for use in water processing, lift stations, storm water and sump process conditions, which require the sensor to be installed in locations that are often below grade where flooding is a possibility. The design of the sensor allows for the unit to be submersed. While the sensor will not be damaged while submersed, the sensor will not provide correct level readings when submersed. Note: Do not attempt to open the sensor housing. Doing so will damage the seal, allow moisture into the sensor, and cause a sensor failure. The sensor also features a submersion resistant 10m cable and IP67 remote push button display module (LR98 described on the following page) through which the sensor can be configured and the level will be displayed. EL EA SE Mount the Sensor Perpendicular to the Liquid Level Bracket or Conduit Mount Mounting from the bracket or from the 1” conduit connector are both acceptable mounting methods. The measurement location for all readings is located at the bottom of the sensor. Remember to take into account the installed insertion length of the sensor when calculating level height within the sump. Note: Never mount the sensor hanging from the cable. This type of installation will not secure the sensor, may damage the cable connection and will result in inconsistent level readings as the sensor sways. Page 24 of 72 MN301700 Rev 9_4 INSTALL SENSOR Section Three LR98 Display Installation The LR98 is a wall mount IP67 remote push button display module that’s used to configure and display level readings from the LR30 sensor. The LR98 should be mounted in a location where the display can be easily read. Note: The LR30 cable length is 32.8’ (10m) which is also the maximum signal distance between the LR30 and LR98. Take that into account when selecting the LR98 mounting location and use the below drill‐hole template for installing the display. EL EA SE LR98 Rear View Note: Make sure that the LR98 display is mounted in an above grade location where it will not become submersed. Rev 9_4 MN301700 Page 25 of 72 WIRE SENSOR Section Four Remove the Display To access the terminal strip and conduit ports, you first need to remove the display. Gently twist the display counter‐clockwise until you feel the display unlock from the housing. Next, lift the display from the housing to view the terminal strip and wire access ports. Note: This procedure applies to all sensors including the LR30 with its LR98 remote display. Supply Voltage The sensor power supply and current signal share the same two‐wire shielded cable. The sensor supply voltage should never exceed 26 VDC. Always provide complete electrical and physical separation between the sensor supply circuit and the main circuit. Note: Remember that the output voltage of the power supply can be lower under nominal load (with a sensor current of 20.5 mA or 22 mA) and/or with the addition of other instruments placed within the circuit. If voltage spikes or surges are expected, adequate isolation protection must also be provided. EA SE Terminal Wiring EL The positive (+) and Negative (‐) terminals are for connection to a 24 VDC power supply or to a 4‐20 mA loop power source. The wire to the terminals can be extended up to 1,000 feet using 22 gauge or larger wire. The sensor should be wired with shielded 2‐conductor cable (16 to 22 AWG) to protect from electromagnetic interference. If using a liquid tight connector, select a cable with an outer diameter that is designed to ensure an effective seal with the connector [typically between 0.20” to 0.35” (5 to 9 mm)]. Electrical, Usage and Safety 1. Wiring should always be done by a licensed electrician in accordance with national, state and local codes. 2. Never use a general purpose (cTUVus) sensor (LR10, LR15, LR20, LR25, LR30 Series) in environments classified as hazardous. 3. Where personal safety or significant property damage can occur due to a spill, the installation must have a redundant fail‐safe backup system installed which accounts for sensor and/or power failure. Page 26 of 72 MN301700 Rev 9_4 WIRE SENSOR Section Four LR30 Sensor to LR98 Display The LR98 display is used with the LR30 sensor. The attached 8‐conductor sensor cable will wire directly into the display terminals. A shielded two‐wire cable (user supplied) is required to provide power to and the current output signal from the display. LR98 Display Terminals EL EA SE LR98 Display Removal Remove the display (as described on the previous page) to access the input and output terminals within the LR98 display. 2. Referencing the below diagram, connect the appropriately colored 6‐conductors (of 8 total) from the LR30 sensor cable to Out [(+) & (‐)] & terminals 1‐4 on the LR98 display. 3. The remaining 2‐conductors (Green and Yellow w/ Stripe) will not be used. 4. Finally, connect the 2‐conductors (from the user supplied Cable) for loop power input and current output to the (+) and (‐) terminals on the LR98 display. 1. Note: The IN [Positive (+) and Negative (‐)] terminals are for connection to a 24 VDC power supply or 4‐20 mA loop power source. The 4‐20mA wires to the LR98 terminals can be extended up to 1,000 feet using 22‐gauge or larger wires. These terminals are equivalent to the (+) and (‐) terminals described on the previous page. Rev 9_4 MN301700 Page 27 of 72 WIRE SENSOR Section Four Wiring to Displays, Controllers & PLC’s Below are examples of how to wire EchoPulse™ to common displays, controllers and PLC’s. DataView™ LI55 Series Level Controller DataLoop™ LI25 Series Level Indicator (With Backlight) EL DataLoop™ LI25 Series Level Indicator (Without Backlight) EA SE Commander™ LI90 Series Multi‐Tank Level Controller Page 28 of 72 MN301700 Rev 9_4 WIRE SENSOR Section Four Wiring to Displays, Controllers & PLC’s DataPoint™ LC52 Series Level Controller (*JWA Mode ‐ Factory Setting) Generic PLC EL Generic Loop Powered Display EA SE DataPoint™ LC52 Series Level Controller (*JWB Mode) * Refer to the DataPoint™ LC52 Series Level Controller manual for information on JWA mode and JWB mode settings in the controller. Rev 9_4 MN301700 Page 29 of 72 CONFIGURATION Section Five Basic Configuration Overview Below are the 7 basic steps to configure the sensor for operation. Each step is described in detail on the following pages EL EA SE 1. Measure the Tank 1. Begin by measuring the key tank and fitting dimensions. Correct tank dimensions will result in accurate sensor measurement. 2. Set the Units of Measurement 1. Units can be configured in basic engineering units of length including Feet or Meters. 3. Set the Empty Configuration 1. This is the empty setting (4mA) for the tank. 4. Set the Full Configuration 1. This is the full setting (20mA) for the tank. 5. Set the Range (Maximum Range or MaxR) 1. This is the maximum measurement range for the sensor. The sensor will ignore all echo returns beyond this setting. 6. Set the Dead Band (Minimum Range or MinR) 1. This is the minimum measurement range for the sensor. The sensor will ignore all echo returns closer than this setting. 7. Check the Echo Curve 1. This is a quick diagnostic tool to determine if the sensor is reading the correct level. Page 30 of 72 MN301700 Rev 9_4 CONFIGURATION Section Five Using the Display The display module features a dot matrix LCD display with 4 push buttons on a removable puck. Out of the box, the display indicates level in feet and depicts the level within the 4‐20mA span on a bar graph at the right side of the display. The four buttons perform the following functions: Up Arrow o Modify parameter values o Choose display mode Right Arrow o Choose configuration options o Choose parameter digits to edit o Display contents of parameters ENTER o Enter Menu and Options o Confirm configuration options o Confirm changes to parameters ESCAPE o Exit configuration mode o Return to a higher menu level o Display Echo Curve EL EA SE Menu Introduction 1. To enter the Main Menu (from the Main Screen), press the ENTER button. 2. Use the Right Arrow button to scroll through the Main Menu options. 1. Configuration ‐ Below are the configuration menu functions: a. Empty Configuration b. Full Configuration c. Medium d. Dampening e. Output Mapping f. Scaled Units g. Scaling h. Range i. Dead Band j. Sensor ID 2. Display ‐ This menu function sets the display mode and contrast. 3. Diagnostics ‐ Below are the diagnostic menu functions: a. Measurement of Peak Values b. Measurement Status c. Echo Curve d. Simulation 4. Service ‐ Within the service menu functions, you can store a False Echo Curve, set units of measurement, change output settings, reset configuration settings, set language or set a PIN for the sensor. 5. Info ‐ This item provides information on the sensor’s type, serial number, date of manufacture and software version. 3. To select one of the functions, press ENTER. 4. To exit the programming mode, press ESC. Rev 9_4 MN301700 Page 31 of 72 CONFIGURATION Section Five Changing Display Values The numeric values are set using the Right Arrow and Up Arrow buttons. Press the Right Arrow button to select the next digit and the Up Arrow button to increment the digit value. The digit being changed is highlighted. Press the Enter button to accept a setting or the Esc button to exit without saving changes. The below exercise illustrates how to change the value of an Empty configuration. Follow the steps to change the setting from 10.00 ft to 12.00 ft. This example applies to all functional settings starting from the Main Menu. From the Main Menu, press ENTER to advance into the Configuration menu. 1. The “+” sign will be highlighted on the screen. 2. This is the adjustment for the percentage setting. Press ENTER to move down to the distance setting. 1. 4. 1. SE 3. From Empty configuration, press ENTER. The first digit, “1”, will be highlighted. EA 2. Empty configuration will appear on the top line of the screen. Press Right Arrow to move one digit to the right. 1. Use the Right Arrow button to move the digit one space to the right. 2. Pressing Right Arrow on the last digit will jump back to the first digit. EL 1. Press UP ARROW to increase the digit from “0” to “1”. 6. Press UP ARROW to increase the digit from “1” to “2”. 5. Page 32 of 72 1. Use the UP ARROW button to increase the digit by one unit. 2. After “9”, the display will jump back to “0”. 7. Press ENTER to accept the setting as 12.00. 8. Press ESCAPE to move back to the Main Menu. MN301700 Rev 9_4 CONFIGURATION Section Five Step 1 ‐ Measure the Tank Measuring the tank is one of the most important aspects in configuring the sensor. When measuring the tank, take into account the location of the sensor with respect to fittings, risers, dome tops and bottoms, and identify where the measurements are taken from the sensor. Note: The location for measurement may be different among different sensor Series, based upon the type of antenna. Refer to the Measurement Reference Chart for the measurement location of your sensor. The basic measurements for configuration are described below: Measurement Reference Chart LR15 Series LR10 Series (Threaded Version) LR20 Series EL EA SE LR15 Series (Flange Version) 1. Distance from the sensor’s measurement location to the bottom of the tank is the Range value. The Range value is typically set at the bottom of the tank. LR25 Series LR30 Series 2. Distance from the sensor’s measurement location to the empty or lowest liquid level in the tank is the Empty Configuration. 1. Empty Configuration = 4mA setting. 2. With flat bottom tanks, the Range and Empty Configuration values can be the same. 3. Distance from the sensor’s measurement location to the full or highest liquid level in the tank is the Full Configuration. 1. Full Configuration = 20mA setting. Rev 9_4 MN301700 Page 33 of 72 CONFIGURATION Section Five Step 2 ‐ Set the Units of Measurement This function sets the units for all measurement values to be entered into the sensor. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Service. 3. Press Enter to advance into the Service menu (and Echo curve will appear). 4. Press Right Arrow repeatedly until the menu shows Units of Measurement. 5. Press Enter to advance into Units of Measurement. 6. Press Right Arrow to change the setting between feet [ft (d)] and meters [m (d)]. SE 7. When the units are correct, press Enter to save the setting. EL EA 8. When done, press ESC to return to the Main Menu, and press ESC a second time to return to the Main Screen. Page 34 of 72 MN301700 Rev 9_4 CONFIGURATION Section Five Step 3 ‐ Set the Empty Configuration (4mA) This function sets the Empty Configuration point for the sensor corresponding 4mA to empty. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Enter to advance into the Configuration Menu. 3. Press Enter to advance into Empty Configuration. The first percentage segment will be highlighted. 4. Press Enter again to switch to the distance (d) setting. SE 5. Press Right Arrow to move one segment to the right. Right Arrow will scroll left to right and then back to the first segment. 7. EA 6. Press Up Arrow to increase the value of the number highlighted. Up Arrow will scroll from 0 to 9 and back again. When the value is correct, press Enter to save the setting. EL 8. When done, press ESC to return to the Main Menu, and press ESC a second time to return to the Main Screen or; If you want to advance directly into Full Configuration, press Right Arrow. Rev 9_4 MN301700 Page 35 of 72 CONFIGURATION Section Five 2. Press Enter to advance into the Configuration Menu. 3. Press Right Arrow to advance into Full Configuration. 4. Press Enter to advance into Full Configuration. The first percentage segment will be highlighted. 5. Press Enter again to switch to the distance (d) setting. 6. Press Right Arrow to move one segment to the right. Right Arrow will scroll left to right and then back to the first segment. 7. Press Up Arrow to increase the value of the number highlighted. Up Arrow will scroll from 0 to 9 and back again. 8. When the value is correct, press Enter to save the setting. 9. When done, press ESC to return to the Main Menu, and press ESC a second time to return to the Main Screen or; If you want to advance directly into Range, press Right Arrow repeatedly until Range appears. SE From the Main Screen, press Enter to advance into the Main Menu. EL EA 1. Step 4 ‐ Set the Full Configuration (20mA) This function sets the Full Configuration point for the sensor corresponding 20mA to full. Page 36 of 72 MN301700 Rev 9_4 CONFIGURATION Section Five Step 5 ‐ Set the Range (Maximum Range) This function sets the maximum operational range (MaxR) for the sensor. This setting defines the maximum distance that the sensor will detect valid echo returns. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Enter to advance into the Configuration Menu. 3. Press Right Arrow repeatedly until the menu shows Range. 4. Press Enter to edit Range value. The first segment will be highlighted. SE 5. Press Right Arrow to move one segment to the right. Right Arrow will scroll left to right and then back to the first segment. 7. When the value is correct, press Enter to save the setting. EL 8. When done, press ESC to return to the Main Menu, and press ESC a second time to return to the Main Screen or; if you want to advance directly into Dead Band, press Right Arrow repeatedly until Dead Band appears. EA 6. Press Up Arrow to increase the value of the number highlighted. Up Arrow will scroll from 0 to 9 and back again. Rev 9_4 MN301700 Page 37 of 72 CONFIGURATION Section Five Step 6 ‐ Set the Dead Band This function sets the Dead Band for the sensor. This setting defines the minimum distance that the sensor will detect valid echo returns. While the Dead Band setting is typically configured to be equal with or slightly above (higher in the tank) the Full Configuration setting (20 mA), its functions independently of Full Configuration. Note: If the Dead Band setting is placed below the Full Configuration setting, then the sensor will not measure above the Dead Band. Dead Band Below Full Config. Dead Band Above Full Config. SE Dead Band Equals Full Config. 2. Press Enter to advance into the Configuration Menu. EL 3. Press Right Arrow repeatedly until menu shows Dead Band. EA 1. From the Main Screen, press Enter to advance into the Main Menu. 4. Press Enter to edit Dead Band value. The first segment will be highlighted. 5. Press Right Arrow to move one segment to the right. Right Arrow will scroll left to right and then back to the first segment. 6. Press Up Arrow to increase the value of the number highlighted. Up Arrow will scroll from 0 to 9 and back again. 7. When the value is correct, press Enter to save the setting. 8. When done, press ESC to return to the Main Menu, and press ESC a second time to return to the Main Screen. Page 38 of 72 MN301700 Rev 9_4 CONFIGURATION Section Five Step 7 ‐ Check the Echo Curve This function displays the primary echo return(s) that the sensor is seeing graphically, the location and amplitude of the return(s), and the numeric air gap distance from the sensor’s measurement location to the liquid level below. Note: This step should only be performed after having completed the prior six configuration steps with the sensor installed on the tank. Additionally, if the sensor was installed in a stand pipe or sight glass, now go forward to Section Six and turn on the still well function (Sensor Installed in a Stand Pipe or Sight Glass) before continuing with this step. EA SE 1. From the Main Screen, press ESC and the Echo Curve Screen will appear. The curve graphically represents the primary echo return(s) amplitude (Y‐axis) over distance (X‐axis). Above the echo return peak is a floating arrow and triangle symbol (which under normal conditions are often merged together or seen as a single triangle because it’s the larger of the two symbols). The arrow represents the measured liquid level and the triangle represents the peak amplitude location of the echo return. Under normal conditions, expect to see a stable triangle (or overlapping arrow and triangle) floating above a pronounced peak at the expected air gap distance between the measurement location and liquid level. EL 2. In the upper right hand corner of the screen are two lines of numbers that represent the air gap distance from the measurement location to the liquid level (arrow) on the top, and peak amplitude location (triangle) of the echo return on the bottom. Under normal conditions, these values should be relatively close to one another and consistent with the expected air gap distance between the measurement location and liquid level. 3. Assuming that the sensor is properly installed, if the measured liquid level and peak amplitude location data (symbols and values) are unstable, substantially different from one another and/or inconsistent with the actual air gap distance, then this likely indicates that the sensor requires additional process adjustment(s) described in the following Section Six. 4. When done, press ESC to return to the Main Menu. Rev 9_4 MN301700 Page 39 of 72 PROCESS ADJUSTMENTS Section Six Process Adjustments Overview These optional functions are intended to improve sensor performance in applications with the below process and/or installation characteristics. Note: These adjustments should only be performed when (after having completed the seven configuration steps described in Section Five with the sensor installed on the tank) the sensor is not performing to your satisfaction. Where so, perform the following applicable Process Adjustments. Fast Filling or Emptying of the Liquid Liquid Surface is Turbulent or Agitated Surface Foam on the Surface of the Liquid Sensor Installed in a Stand Pipe or Sight Glass EL EA SE 1. 2. 3. 4. Page 40 of 72 MN301700 Rev 9_4 PROCESS ADJUSTMENTS Section Six Fast Filling or Emptying of Liquid If the speed of liquid level rise or fall within the tank is greater than a rate of 1” per second (25.4mm/sec), set Fast Level Change to Yes. Note: Fast filling or emptying can occur when multiple pumps are operating or when a weather event increases the amount of liquid entering the tank. 1. From the Main Screen, press Enter to advance into the Main Menu. Press Right Arrow to advance from Empty Configuration to Full Configuration. 4. Press Right Arrow to advance from Full Configuration to Medium. 5. Press Enter to advance into Medium. Liquid, Solid, Low Dielectric will appear. 6. Press Enter to advance into Liquid. Fast Level Change will appear first. 7. Press Enter to advance into Fast Level Change. 8. Press Right Arrow to change the Fast Level Change setting. 9. When the setting is correct, press Enter to save. 3. SE Press Enter to advance into the Configuration Menu. Fast Level Change EL EA 2. Normal Level Change 10. When done, press ESC to return to Medium, press ESC again to return to the Configuration Menu, and press ESC a third time to return to the Main Screen or; If you want to advance directly into Turbulent Surface, press Right Arrow repeatedly until Turbulent Surface appears. Rev 9_4 MN301700 Page 41 of 72 PROCESS ADJUSTMENTS Section Six Liquid Surface is Turbulent or Agitated If the liquid surface is turbulent or agitated, set Turbulent Surface to Yes. Note: Turbulent or agitated surfaces can occur when tanks are filled from the top without a down pipe, or when a mixer or air agitation is used within the tank. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Enter to advance Configuration Menu. 3. Press Right Arrow to advance from Empty Configuration to Full Configuration. 4. Press Right Arrow to advance from Full Configuration to Medium. 5. Press Enter to advance into Medium. Liquid, Solid, Low Dielectric will appear. 6. Press Enter to advance into Liquid. Fast Level Change will appear first. 7. Press Right Arrow to advance from Fast Level Change to First Echo. 8. Press Right Arrow to advance from First Echo to Turbulent Surface. 9. Press Enter to advance into Turbulent Surface. SE the Agitation from Mixer EL EA into Turbulence from Tank Fill 10. Press Right Arrow to change the Turbulent Surface setting. to save. 11. When the setting is correct, press Enter 12. When done, press ESC to return to Medium, press ESC again to return to the Configuration Menu, and press ESC a third time to return to the Main Screen or; If you want to advance directly into Foam, press Right Arrow repeatedly until Foam appears. Page 42 of 72 MN301700 Rev 9_4 PROCESS ADJUSTMENTS Section Six Foam on the Surface of the Liquid If the entire liquid surface is covered with foam, set Foam to Yes. This is not necessary if the liquid surface is partially covered with foam. Note: Foam can occur when tanks are filled from the top without a down‐fill pipe, or when a mixer or air agitation is used within the tank. Press Enter to advance into the Configuration Menu. 3. Press Right Arrow to advance from Empty Configuration to Full Configuration. 4. Press Right Arrow to advance from Full Configuration to Medium. 5. Press Enter to advance into Medium. Liquid, Solid, Low Dielectric will appear. 6. Press Enter to advance into Liquid. Fast Level Change will appear first. 7. Press Right Arrow repeatedly until Foam 1.3.4 appears. 8. Press Enter to advance into Foam. 9. Press Right Arrow to change the Foam setting. 2. Heavy Foam – If foam covers the entire surface of the liquid, set Foam to Yes. SE From the Main Screen, press Enter to advance into the Main Menu. EL EA 1. Light Foam – If foam partially covers the surface of the liquid, set Foam to No. 10. When the setting is correct, press Enter to save. 11. When done, press ESC to return to Medium, press ESC again to return to the Configuration Menu, and press ESC a third time to return to the Main Screen or; If you want to advance directly into Still Well, press Right Arrow repeatedly until Still Well appears. Rev 9_4 MN301700 Page 43 of 72 PROCESS ADJUSTMENTS Section Six Sensor Installed in a Stand Pipe or Sight Glass If the sensor is installed in a metal stand pipe (still well) or metal sight glass, set Still Well to yes and enter the inner Pipe Diameter dimension. Note: The Pipe Diameter will be entered in millimeters. For example, a 3” pipe can have an inner diameter of 3.042”. To convert inches to mm, multiple inches by 25.4mm. Thus, a 3.042” pipe inner diameter equals 77.26mm. You would then enter the value of 77mm. 1. From the Main Screen, press Enter to advance into the Main Menu. Press Right Arrow repeatedly until menu shows Medium. 4. Press Enter to advance into Medium. Liquid, Solid & Low Dielectric will appear. 5. Press Enter to advance into Liquid. Fast Level Change will appear first. 6. Press Right Arrow repeatedly until Still Well appears. 7. Press Enter to advance into Still Well. 3. SE Press Enter to advance into the Configuration Menu. EA 2. Still Well or Stand Pipe 8. Press Right Arrow to change the setting from No to Yes. EL 9. Press Enter to enter the Pipe Diameter. Sight Glass 10. Use the Right Arrow to move one segment to the right. Right Arrow will scroll left to right and then back to the first segment. 11. Use the Up Arrow to increase the value of the number highlighted. Up Arrow will scroll from 0 to 9 and back again. 12. When the value is correct, press Enter to save. 13. When done, press ESC to return to Medium, press ESC again to return to the Configuration Menu, and press ESC a third time to return to the Main Screen. Page 44 of 72 MN301700 Rev 9_4 ADVANCED ADJUSTMENTS Section Seven Advanced Adjustments Overview These optional functions are used to change the sensor output characteristics, or Create a False Echo Curve to filter out false echo returns within the tank (improving sensor performance), or Update an existing False Echo Curve filter if the original filter was not created during an empty tank condition. 1. 4‐20 mA Rev Output ‐ Reverses the current output from 4mA @ bottom and 20mA @ top of tank to 20mA @ bottom and 4mA @ top of the tank. 2. Fail‐Safe ‐ Allows for the presetting of the current output when a sensor failure occurs. Options are no change to current, 20.5mA or 22mA. SE EL 5. Update an existing False Echo Curve ‐ A method to update an existing False Echo Curve to include a lower section of the tank that was not exposed during the creation of the original Echo Curve. Note: If you don't know the location (level position) or validity of the original False Echo Curve, it is recommended to delete the original Echo Curve, and then create a new False Echo Curve (versus updating an existing False Echo Curve). EA 4. Create a new False Echo Curve ‐ A method to filter out false echo returns within the tank. This should be performed when the tank is at its lowest level (empty). 3. Minimum Current ‐ Sets the minimum current output for the sensor. Options are 4.0mA or 3.9mA. Rev 9_4 MN301700 Page 45 of 72 ADVANCED ADJUSTMENTS Section Seven 4‐20 mA Reverse Output This function sets the current output at either 4‐20 mA or 20‐4 mA. Selecting 4‐20 mA sets the output with 4mA @ bottom and 20mA @ top of the tank. This is the standard output used in the majority of applications. Selecting 20‐4 mA sets the output with 20mA @ bottom and 4mA @ top of the tank. This is an optional output sometimes used in applications where the level is maintained at a high level. 1. From the Main Screen, press Enter to advance into the Main Menu. 4 to 20 mA Output 2. Press Right Arrow repeatedly until the arrow is next to Service. 3. Press Enter to advance into Service Menu. 4. Press Right Arrow to move from Echo Curve to Output Current. 6. Press Enter to advance into 4‐20mA Rev. 7. Press Right Arrow to change the setting between 4‐20mA and 20‐4mA. EA 8. When the setting is correct, press Enter to save. SE 5. Press Enter to advance into Output Current. 20 to 4mA Output EL 9. When done, press ESC to return to the Service Menu and press ESC a second time to return to the Main Screen or; if you want to advance directly into Fail‐Safe, press Enter and then Right Arrow until Fail‐Safe appears. Page 46 of 72 MN301700 Rev 9_4 ADVANCED ADJUSTMENTS Section Seven Fail‐Safe Output This function is used to set the current output to a designated state if the sensor loses measurement confidence. Selecting No Change will hold the current at its last valid current output. Selecting 20.5mA will force the current to jump to 20.5mA. Selecting 22.5 mA will force the current to jump to 22.5 mA. Note: The latter two high current output states are above the standard 4‐20 mA operational range, and can be used to indicate that a failure has occurred. 1. From the Main Screen, press Enter to advance into the Main Menu 2. Press Right Arrow repeatedly until the arrow is next to Service. 3. Press Enter to advance into Service Menu and Echo Curve will appear. SE 5. Press Enter to advance into Output Current menu. 4. Press Right Arrow to move from Echo Curve to Output Current. 6. Press Right Arrow to move from Reverse 4‐20mA to Fail‐safe. EA 7. Press Enter to enter Fail‐safe. 8. Press Right Arrow to change the setting between No change, 20.5mA and 22.0mA. EL 9. When the setting is correct, press Enter to save. 10. When done, press ESC to return to the Service Menu and press ESC a second time to return to the Main Screen or; if you want to advance directly into Min Current, press Enter and then Right Arrow until Min Current appears. Rev 9_4 MN301700 Page 47 of 72 ADVANCED ADJUSTMENTS Section Seven Minimum Current Output This function sets the minimum current output for the sensor at either 4.0mA or 3.9mA. Note: 4.0mA is the default minimum current output and is used in the majority of applications. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Service. 3. Press Enter to advance into Service Menu and Echo Curve will appear. 4. Press Right Arrow to move from Echo Curve to Output Current. 5. Press Enter to advance into Output Current menu. 7. Press Enter to enter Min Current. SE 6. Press Right Arrow repeatedly to move from Reverse 4‐20mA to Min Current. EA 8. Press Right Arrow to change the setting between 3.9mA and 4.0mA. 9. When the setting is correct, press Enter to save. EL 10. When done, press ESC to return to the Service Menu and press ESC a second time to return to the Main Screen. Page 48 of 72 MN301700 Rev 9_4 ADVANCED ADJUSTMENTS Section Seven Create a New False Echo Curve Obstructions in the tank (mixer blades, side wall weld joints or material build‐up, submersible pumps, piping, other apparatus) or tall tank risers or installation fittings can create false echo returns that impair the sensor’s measurement. This function maps all echo returns within the tank, differentiating between good and false echoes, and stores those identified as false into the False Echo Curve, so they will not be considered in the level measurement. Note: A False Echo Curve should only be performed when the tank is empty so that all false reflections will be detected. Before starting, measure and note the exact distance from the sensor’s measurement location to the liquid surface. Setting the distance value too large or too short can force the sensor into ignoring the true level. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Service. 3. Press Enter to advance into Service Menu and Echo Curve will appear. SE 4. Press Enter to make a change to the Echo Curve settings. EA 5. Press Right Arrow repeatedly until the arrow is next to Create New. 6. Press Enter to advance into Create New. Enter the distance from the sensor’s measurement location to the liquid surface. EL 7. Use the Right Arrow to move one segment to the right. The Right Arrow will scroll left to right and then back to the first segment. 8. Use the Up Arrow to increase the value of the number highlighted. The Up Arrow will scroll from 0 to 9 and back again. 9. When the value is correct, press Enter to save the setting and begin the False Echo Curve mapping. The process may take a few minutes. When complete, the display will return to the Echo Curve screen. 10. When done, press ESC to return to the Service Menu and press ESC a second time to return to the Main Screen. 11. From the Main Screen, press ESC and the Echo Curve Screen will appear. Follow the Check Echo Curve procedure described at the end of Section Five to confirm that the sensor is performing correctly. Rev 9_4 MN301700 Page 49 of 72 ADVANCED ADJUSTMENTS Section Seven Update an Existing False Echo Curve This function enables an Existing False Echo Curve to be updated under the circumstances that the curve was created when the level was higher than an empty tank condition. Note: This function should only be performed to update an Existing False Echo Curve when the level is BELOW the original False Echo Curve. Do not use this function to update an Existing False Echo Curve when the level is above the original False Echo Curve. Before starting, measure and note the exact distance from the sensor’s measurement location to the liquid surface. Setting the distance value too large or too short can force the sensor into ignoring the true level. Original False Echo Curve 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Service. 3. Press Enter to advance into Service Menu and Echo Curve will appear. SE 4. Press Enter to make a change to the Echo Curve settings. 5. Press Right Arrow repeatedly until the arrow is next to Update. 6. Press Enter to advance into Update. Enter the actual distance from the sensor’s measurement location to the liquid surface. EA Good Update EL 7. Use the Right Arrow to move one segment to the right. The Right Arrow will scroll left to right and then back to the first segment. 9. When the value is correct, press Enter to save the setting and begin the False Echo Curve mapping. The process may take a few minutes. When complete, the display will return to the Echo Curve screen. Bad Update 8. Use the Up Arrow to increase the value of the number highlighted. The Up Arrow will scroll from 0 to 9 and back again. 10. When done, press ESC to return to the Service Menu and press ESC a second time to return to the Main Screen. 11. From the Main Screen, press ESC and the Echo Curve Screen will appear. Follow the Check Echo Curve procedure described at the end of Section Five to confirm that the sensor is performing correctly. Page 50 of 72 MN301700 Rev 9_4 TROUBLESHOOTING Section Eight EL EA SE Troubleshooting Overview These functions provide troubleshooting information; enable Echo Curve adjustments and deletion of settings. 1. Measurement Status 1. Displays the signal strength (dB) of the echo returns and the functional status of the sensor including diagnostic error codes. 2. Peak Values 1. Displays the lowest and highest level height that the sensor has measured in distance (d). 3. Simulation 1. Simulates the 4‐20mA current output from percent of span, current or distance inputs. 4. First Echo Adjustment 1. Provides the ability to increase or decrease the peak signal strength of the First Echo return. 5. Echo Curve Zoom 1. Provides the ability to zoom in and magnify the Echo Curve over a specified range. 6. False Echo Curve Delete 1. Provides the ability to delete a saved False Echo Curve. 7. Reset 1. Provides the ability to reset configuration settings and memory. Rev 9_4 MN301700 Page 51 of 72 TROUBLESHOOTING Section Eight Measurement Status This function displays the signal strength (dB) of the echo returns and the operational status of the sensor including diagnostic error codes. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Diagnostics. 3. Press Enter to advance into Diagnostics. 4. Press Right Arrow to switch to Measurement Status. 5. Measurement reliability indicates the decibel (dB) strength of RF energy that’s reflecting back to the sensor less any noise. For reliable function, the dB value should be ≥ 10 dB. SE 6. Sensor status indicates the functional status of the sensor (either OK or error code) EL Solution Check power supply Check wiring for open circuit Check for obstacles under the sensor, either clear the obstacle, move the sensor or perform a False Echo Curve Contact your distributor Sensor Status Error Codes Error Code Problem E11 Insufficient power E12 Open circuit EA 7. When done, press ESC to return to the Service Menu and press ESC a second time to return to the Main Screen. Weak echo return (< 10dB) E15 or E17 ROM error E14 Current Output Conditions Problem Sensor output < 3.9mA Sensor output > 22mA Sensor output reaches 4mA before the tank is empty Sensor output will not reach 4mA Page 52 of 72 Solution Check wiring for open circuit Check wiring for short circuit Check the Empty Configuration setting, and if incorrect, extend the setting to the empty tank position Check the Range setting, and if incorrect, extend the setting to or slightly below the Empty Configuration setting MN301700 Rev 9_4 TROUBLESHOOTING Section Eight Peak Values This function displays the lowest and highest level height that the sensor has measured in distance (d). 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until arrow is next to Diagnostics. 3. Press Enter to advance into Diagnostics and view Peak Values. 4. Distance‐min is the lowest measured level and Distance‐max is the highest measured level. Confirm that these values are within the sensor’s operational range. 5. If the values appear too high or too low, check the tank for obstructions that could cause that problem. EL EA SE 6. When done, press ESC to return to the Service Menu and press ESC a second time to return to the Main Screen. Rev 9_4 MN301700 Page 53 of 72 TROUBLESHOOTING Section Eight Simulation This function simulates the 4‐20mA current output, when the sensor is configured and installed on the tank, but the level cannot be changed easily for testing. Percent (of span), Current Output or Distance can be used as the input method to set the current output. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Diagnostics. 3. Press Enter to advance into Diagnostics. 4. Press Right Arrow repeatedly until Simulation appears. 5. Press Enter to advance into Simulation. SE 7. Use Enter to select the method of simulation. 6. Use the Right Arrow to scroll among the three simulation methods (Percent, Current Out or Distance). EA 8. Upon selection of a method, the simulation will start and the current output will proportionately reflect the value shown. While the simulation is running, the value can be changed. EL 9. Use the Right Arrow to move one segment to the right. The Right Arrow will scroll left to right and then back to the first segment. 10. Use the Up Arrow to increase the value of the number highlighted. The Up Arrow will scroll from 0 to 9 and back again. 11. When the value is correct, press Enter and the current output will reflect the value shown. 12. When done, press ESC to return to Simulation, press ESC again to return to the Configuration Menu, and press ESC a third time to return to the Main Screen. Percent Current Output This method uses percentage of span This method directly sets the current (0‐100%) to set the current output (4‐ output (4‐20mA) to the desired value. 20mA) between the Empty and Full Configuration values with 100% equal to 20 mA and 0% equal to 4mA. Page 54 of 72 MN301700 Distance This method uses distance from the sensor’s measurement location to set the current output (4‐20mA) between the Empty and Full Configuration values. Rev 9_4 TROUBLESHOOTING Section Eight First Echo Adjustment This function increases or decreases the peak strength of the sensor’s First Echo return, and should only be performed if: 1) The liquid has a very high dielectric constant value and primarily stays in the near full range of the tank, resulting in a very high First Echo peak strength or; 2) Process conditions, such as when the liquid has a very low dielectric constant value, or when obstructions, heavy foam or turbulence exist in the tank, resulting in little or no First Echo peak strength. Note: Under condition one (Example 1), it can be beneficial to decrease the First Echo peak strength. Under condition two (Example 3), it can be beneficial to increase the First Echo peak strength. Example 1 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Enter to advance into the Configuration Menu. Example 2 3. Press Right Arrow to move from Empty Configuration to Full Configuration. 4. Press Right Arrow to move from Full Configuration to Medium. SE First Echo peak is very strong and can be reduced. 6. Press Enter to advance into Liquid and Fast Level Change will appear. EL First Echo peak is normal and no adjustment is required. EA 5. Press Enter to advance into Medium and Liquid, Solid & Low Dielectric will appear. 7. Press Right Arrow to move from Fast Level Change to First Echo. 8. Press Enter to advance into First Echo. Example 3 9. Press Right Arrow to change the First Echo setting. First Echo peak is weak and can be increased. First Echo Adjustments Normal ‐ No adjustment Small ‐ Decrease by 10 dB Big ‐ Increase by 10 db Bigger ‐ Increase by 20 db Biggest ‐ Increase by 40 db Rev 9_4 10. When setting is correct, press Enter to save. 11. When done, press ESC to return to Medium, press ESC again to return to the Configuration Menu, and press ESC a third time to return to the Main Screen. MN301700 Page 55 of 72 TROUBLESHOOTING Section Eight Echo Curve Zoom This function zooms in and magnifies an Echo Curve over a specified range. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Diagnostics. 3. Press Enter to advance into Diagnostics. 4. Press Right Arrow repeatedly until Echo Curve appears. 5. Press Enter to advance into Echo Curve. 6. Use the Right Arrow to scroll among the three settings (X‐ zoom, Y‐zoom, Unzoom). 7. Press Enter to select X‐zoom. SE 8. Press Enter to advance into the first X‐zoom boundary (Example 1). Press Right Arrow to move the boundary to its desired location. Press Enter to set the boundary and a second X‐zoom boundary will appear. EA 9. Press Right Arrow to move the boundary to its desired location (Example 2). Press Enter to set the boundary and the screen will show the expanded X‐axis (Example 3). Press ESC to exit. EL 10. Press Enter to advance into Y‐zoom (Example 4). Press Right Arrow to move to the desired zoom magnification (Example 5). Press Enter to set the zoom and the screen will show the expanded Y‐axis (Example 6). Note: If you wish to start over, press ESC, return to Echo Curve, select Unzoom, and begin the procedure again. 11. When done, press ESC to return to Echo Curve, press ESC again to return to Diagnostics, and ESC a third time to return to the Main Screen. X‐zoom (Example 1) X‐zoom (Example 2) X‐zoom (Example 3) Y‐zoom (Example 4) Y‐zoom (Example 5) Y‐zoom (Example 6) Page 56 of 72 MN301700 Rev 9_4 TROUBLESHOOTING Section Eight False Echo Curve Delete This function deletes a saved False Echo Curve. Note: If you are dissatisfied with the sensor’s performance (operating with a False Echo Curve), you may delete it, and consider creating a new False Echo Curve. 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Service. 3. Press Enter to advance into Service and Echo Curve will appear. 4. Press Enter to make a change to Echo Curve. 5. Press Enter to select Delete. 6. Press Enter to Delete All. EL EA SE 7. Press ESC to return to Service and press ESC a second time to return to the Main Screen. Rev 9_4 MN301700 Page 57 of 72 TROUBLESHOOTING Section Eight Reset This function resets the sensor’s configuration (basic or factory) settings and memory (peak values measured). 1. From the Main Screen, press Enter to advance into the Main Menu. 2. Press Right Arrow repeatedly until the arrow is next to Service. 3. Press Enter to advance into Service. 4. Press Right Arrow repeatedly until Reset appears. 5. Press Enter to advance into Reset. 6. Use the Right Arrow to scroll among the three reset types (Basic Reset, Factory Settings, Peak Value Meas). SE 1. Basic Reset a. Resets basic configuration settings, process adjustment settings, and peak level values in memory (retaining any advanced adjustment settings). EA 2. Factory Settings a. Resets ALL settings and memory to factory default. EL 3. Peak Values Measured a. Resets the minimum and maximum peak level values in memory (retaining any basic configuration, process adjustment and advanced adjustment settings). 7. Use Enter to select the desired type of reset. 8. When done, press ESC to return to Service, and press ESC a second time to the Main Screen. Page 58 of 72 MN301700 Rev 9_4 APPENDIX Section Nine EL EA SE Configuration Menu The basic configuration functions are found under Configuration. The below tree shows the 10 function settings and how to navigate between them. Note: Press ESCAPE to back‐up to the previous level. Rev 9_4 MN301700 Page 59 of 72 APPENDIX Section Nine Empty Configuration This function adjusts the empty linear scaled current output (4mA) and provides two different adjustment methods: The primary method involves setting the value based upon the distance from the bottom of the sensor. This is a measured value using the units of operation for the sensor. For example, if the units of operation are in meters, then the setting must also be in meters. A second method involves setting a value based upon the percentage of the Range value. For example, if the Range is set to 10ft, then a 10% setting is equivalent to 1ft of liquid height or 9ft of air gap away from the sensor. Note: Set the Range value before setting the Empty Configuration value. Shows the percentage of empty based upon the Range setting. Shows the distance from the bottom of the sensor to the empty tank. Shows the air gap distance from the liquid to the bottom of the sensor. EL EA SE Full Configuration This function adjusts the full linear scaled current output (20mA) and provides two different adjustment methods: The primary method involves setting the value based upon the distance from the bottom of the sensor. This is a measured value using the units of operation for the sensor. For example, if the units of operation are in meters, then the setting must also be in meters. The second method involves setting a value based upon the percentage of the Range value. For example, if the Range is set to 10ft, then a 95% setting is equivalent to 9.5ft of liquid height or 0.5ft or air gap away from the sensor. Note: Set the Range value before setting the Full configuration value. Shows the percentage of full based upon the Range setting. Shows the distance from the bottom of the sensor to the full tank. Shows the air gap distance from the liquid to the bottom of the sensor. Page 60 of 72 MN301700 Rev 9_4 APPENDIX Section Nine Medium This function identifies the type of media that the sensor is measuring including Liquid, Solid or Low Dielectric. Each has various settings to address the different reflective properties associated with each medium. Default is Liquid. The Liquid Medium function has the following settings: Fast Level Change ‐ Used when the media’s level rise or fall within the tank is greater than a rate of 1” per second (25.4mm/sec). Selections are YES or NO. Default is NO. Turbulent Surface ‐ Used when the surface of the liquid is turbulent or agitated. Selections are YES or NO. Default is NO. Foam ‐ Used when the entire liquid surface is covered with foam. Selections are YES or NO. Default is NO. Low Dielectric ‐ Used when the liquid has a low dielectric constant (and should not be performed without first consulting with the factory). Selections are YES or NO. Default is NO. Still Well ‐ Used when the sensor is installed in a still well, stand pipe or sight glass. Selections are YES or NO. Default is NO. Note: If YES is selected, the pipe (inner) diameter must be entered. SE First Echo ‐ Used when the sensor has difficulty seeing the first echo return. This setting adjusts the peak strength (dB) of the first echo. Default is Normal. Below are the settings: o Normal ‐ No adjustment o Small ‐ Decreases by 10dB o Big ‐ Increases by 10dB o Bigger ‐ Increases by 20dB o Biggest ‐ Increases by 40dB EL EA Rev 9_4 MN301700 Page 61 of 72 APPENDIX Section Nine Large Repose Angle ‐ Used when the repose angle of the material is steep. Repose angle is the steepest angle of assent or descent relative to the horizontal plane that a material can be piled without sliding. The repose angle varies between different materials. Selections are Yes or No. Default is No. Powder / Dust ‐ Used when powder or dust are present in the atmosphere above the material. Selections are Yes or No. Default is No. Low Dielectric ‐ Used when the material has a low dielectric constant (and should not be performed without first consulting with the factory). Selections are YES or NO. Default is NO. EL SE First Echo ‐ Used when the sensor has difficulty seeing the first echo return. This setting adjusts the peak strength (dB) of the first echo. Default is Normal. Below are the settings: o Normal ‐ No adjustment o Small ‐Decreases by 10dB o Big ‐ Increases by 10dB o Bigger ‐ Increases by 20dB o Biggest ‐ Increases by 40dB EA The Solid Medium function has the following settings (and should not be performed without first consulting with the factory): Fast Level Change ‐ Used when the media’s level rise or fall within the tank is greater than a rate of 1” per second (25.4mm/sec). Selections are YES or NO. Default is NO. The Low Dielectric Medium function has the following settings (and should not be performed without first consulting with the factory): When the dielectric constant of the media is ≤ 4, the amplitude of the direct echo from the media may be low and difficult to detect. However, by measuring the echo reflected from the bottom of the tank, the liquid level can be measured. Two parameters must be input to complete the setup: 1) Distance from the bottom of the sensor to the tank bottom (Empty Span) and; 2) Distance from the tank bottom to the liquid level (True Level). Empty Span ‐ Distance from the bottom of the sensor to the tank bottom. True Level ‐ Distance from the tank bottom to the liquid level.. DK ‐ Media dielectric constant (determined by sensor). Note: Once this feature has been activated, only a factory setting can undo its activation. A factory reset will erase all active settings including Echo Curves. Page 62 of 72 MN301700 Rev 9_4 APPENDIX Section Nine Dampen This function sets the sampling rate for which the sensor updates the current output. The value is entered in seconds. Default is 6 seconds. Decreasing the value will make the output more responsive to level changes. Increasing the value will make the output less responsive to level changes. Output Mapping This function sets the correlation between the measured value and the current output. Selections are Linear or Non‐Linear. Default is Linear. Keep the setting on Linear. Note: This setting should not be changed without first consulting with the factory. EA SE Scaled Units This function sets the type of measurement (Height, Volume, Mass, Flow or No‐ Units) and units of measurement (which vary based upon the type of measurement) displayed and input during configuration. Default is Height in Units of feet (ft). EL Scaling This function sets the span values for 0% and 100% as an alternative method of adjusting the Empty Configuration and Full Configuration settings. Defaults are 0% and 100%. Range This function sets the maximum range that the sensor can measure which is typically the distance from the bottom of the sensor to the empty tank bottom. Default is the maximum range of the sensor. Dead Band This function sets the dead band or the minimum distance that the sensor will measure. Default is 12” (30cm). Minimum setting is 2” from the bottom of the antenna. Consult factory when setting the dead band less than the factory setting. Sensor ID This function sets an 11‐digit identification code. The setting can use alphanumeric values from A to Z and 0 to 9. Default is the sensor’s 4‐digit Series number. Rev 9_4 MN301700 Page 63 of 72 APPENDIX Section Nine EL EA SE Display Menu This menu sets the Display Value and LCD Contrast. The below tree shows the 2 function settings and how to navigate between them. Display Value This function sets the Display Value. Default is (air gap) Distance. Below are the settings: OFF ‐ Turns the display OFF. Distance ‐ Displays the distance from the bottom of the sensor to the liquid surface. Height ‐ Displays the height of media in the tank from the Maximum Range Setting. Percent Span ‐ Displays the level based on its percentage of operational span. Map Percentage ‐ Displays the level based on its percentage of operational range. Scaled ‐ Displays the level based on a scaled value between 0 (empty) and 10 (full). Current Output ‐ Displays the 4‐20 mA current output. LCD Contrast This function sets the B/W display contrast. Press the Up Arrow to increase the contrast, and press the Right Arrow to decrease the contrast. Page 64 of 72 MN301700 Rev 9_4 APPENDIX Section Nine EL EA SE Diagnostics Menu This menu provides information about the operational status of the sensor and diagnostic tools. The below tree shows the 5 function settings and how to navigate between them. Peak Values This function displays the lowest and highest level height distances (d) that the sensor has measured during operation. To reset these values, use the Reset >> Peak Measured Values option found under the Service Menu. Measurement Status This function displays the signal strength (dB) of the echo returns (Measurement Reliability) and the operational status of the sensor (Sensor Status). A signal strength of 10dB or higher is acceptable. Anything under 10dB is considered a weak signal. The cause of a weak signal is often obstructions within the beam path or a poor installation. Sensor status confirms whether the sensor is operating as expected. If normal, the descriptor will be OK. If abnormal, the descriptor will be one of the following error codes: E11 – Insufficient power Rev 9_4 E12 – Sensor open circuit E14 – Weak return echo E15 – ROM error MN301700 E17 – ROM error Page 65 of 72 APPENDIX Section Nine EL EA Simulation This function sets a fixed 4‐20mA current output from the sensor for purposes of testing devices receiving information from the sensor. The simulated current output may be set with the below three input methods: SE Echo Curve This function shows the Echo Curve and allows the user to zoom in and magnify information along the X‐axis and Y‐axis. X‐zoom ‐ Expands echo information along the X‐axis. Used to determine the location of an echo. Y‐zoom ‐ Expands echo information along the y‐axis. Used to determine the energy of an echo. Unzoom ‐ Returns the display to the original setting. Choose Curve This function sets which information is displayed. Default is Echo Curve. Echo Curve ‐ Shows the echo return energy received by the sensor over distance. False Echo Curve ‐ Shows the false echo returns which are mapped out (not considered) by the sensor. Output Trend ‐ Shows recent history of where the past level readings have occurred. Percent ‐ This method uses percentage of span (0‐100%) to set the current output (4‐20mA) between the Empty and Full Configuration values with 100% equal to 20 mA and 0% equal to 4mA. Current Out ‐ This method directly sets the current output (4‐20mA) to the desired value. Distance ‐ This method uses distance from the sensor’s measurement location to set the current output (4‐20mA) between the Empty and Full Configuration values. Page 66 of 72 MN301700 Rev 9_4 APPENDIX Section Nine EL EA SE Service Menu This menu is used to make more advanced adjustments to the sensor. Note: Some of these functions should only be performed by a trained technician. Rev 9_4 MN301700 Page 67 of 72 APPENDIX Section Nine False Echo Obstructions in the tank (mixer blades, side wall weld joints or material build‐up, submersible pumps, piping, other apparatus) or tall tank risers or installation fittings can create false echo returns that impair the sensor’s measurement. This function maps all echo returns within the tank, differentiating between good and false echoes, and stores those identified as false into the False Echo Curve, so they will not be considered in the level measurement. Note: A False Echo Curve should only be performed when the tank is empty so that all false reflections will be detected. This function allows the selection of the below False Echo Curve settings: Delete ‐ Used to remove a stored False Echo Curve. Edit ‐ Used to edit the false EA echoes stored in a False Echo Curve. Note: This function should only be performed by a trained technician. SE Create New ‐ Used to create a new False Echo Curve. Update ‐ Used to update a stored False Echo Curve. Note: Do not use this function to update an existing False Echo Curve when the level is above the original False Echo Curve. EL Output Settings This function allows the selection of the below of 4‐20mA Reverse, Fail‐Safe and Minimum Current output settings. The initial screen indicates the status of the three functions. 4‐20 mA Reverse ‐ Used to set the current output at 4‐20 mA or 20‐4 mA. Selecting 4‐20 mA sets the output with 4mA @ bottom and 20mA @ top of the tank. Selecting 20‐ 4 mA sets the output with 20mA @ bottom and 4mA @ top of the tank. Default is 4‐20mA. Fail‐Safe – Used to set the current output to a designated if the sensor loses measurement confidence. Selecting No Change will hold the current at its last valid current output. Selecting 20.5mA will force the current to jump to 20.5mA. Selecting 22.5 mA will force the current to jump to 22.5 mA. Default is 22.0mA. Minimum Current ‐ Used to set the minimum current output at either 4.0mA or 3.9mA Default is 4.0mA. Page 68 of 72 MN301700 Rev 9_4 APPENDIX Section Nine Reset This function allows the configuration (basic or factory) settings and memory (peak values measured) to be reset. Below are the three reset types: Basic Reset – Used to reset the basic configuration settings, process adjustment settings, and peak level values in memory (retaining any advanced adjustment settings). Factory Settings – Used to reset ALL settings and memory to factory default. Peak Values Measured – Used to reset the minimum and SE maximum peak level values in memory (retaining any basic configuration, process adjustment and advanced adjustment settings). EA Units of Measurement This function allows the units of measurement to be changed between Metric and English system units. EL Language This function sets the display language. Default is English. Operational Mode This function allows the sensor digital address to be changed from the Standard mode with an address of 0 to a Multidrop mode where the address can be set from 1 to 15. Default is Address 0. Note: This function should only be performed by a trained technician. Copy Sensor Data This function allows configuration and memory data to be uploaded to or downloaded from the sensor. Note: This function should only be performed by a trained technician. PIN This function allows the sensor configuration to be locked via a preset PIN. Note: This function should only be performed by a trained technician. Rev 9_4 MN301700 Page 69 of 72 APPENDIX Section Nine Distance Adjustment This function allows the factory set distance of a measured value to be adjusted. Note: This function should only be performed by a trained technician. Threshold This function allows the factory set echo threshold to be adjusted. Note: This function should only be performed by a trained technician. SE Information This function displays basic information about your sensor including sensor type, serial number, date of manufacture and software version. EA Factory Settings Below are the Empty Configuration and Full Configuration factory settings for each sensor. Empty Configuration (4mA) LR10 LR25 LR15 LR20 Full Configuration (20mA) EL Series 1.64’ (0.5m) 98.42’ (30m) 1.64’ (0.5m) 65.61’ (20m) 1.64’ (0.5m) 114.83’ (35m) 1.64’ (0.5m) 94.82’ (30m) 1.64’ (0.5m) LR30 32.81’ (10m) User Configuration Fill out the below chart and keep a record of your sensor configuration. Configuration Units of Measurement: Feet Empty Configuration: Full Configuration: Range (Maximum): Dead Band: Meters Process Adjustments Fast Level Change: No Yes Agitated Surface: No Yes Foam: No Yes Low Dielectric: No Yes Still Well: No Yes Page 70 of 72 Diameter: MN301700 Rev 9_4 APPENDIX Section Nine Notes Comment EL EA SE Subject Rev 9_4 MN301700 Page 71 of 72 APPENDIX Section Nine Warranty Flowline warrants to the original purchaser of its products that such products will be free from defects in material and workmanship under normal use and service in accordance with instructions furnished by Flowline for a period of two years from the date of manufacture of such products. Flowline's obligation under this warranty is solely and exclusively limited to the repair or replacement, at Flowline's option, of the products or components, which Flowline's examination determines to its satisfaction to be defective in material or workmanship within the warranty period. Flowline must be notified pursuant to the instructions below of any claim under this warranty within thirty (30) days of any claimed lack of conformity of the product. Any product repaired under this warranty will be warranted only for the remainder of the original warranty period. Any product provided as a replacement under this warranty will be warranted for the full two years from the date of manufacture. Returns SE Products cannot be returned to Flowline without Flowline's prior authorization. To return a product that is thought to be defective, go to www.flowline.com, and submit a customer return (MRA) request form and follow the instructions therein. All warranty and non‐warranty product returns to Flowline must be shipped prepaid and insured. Flowline will not be responsible for any products lost or damaged in shipment. EA Limitations EL This warranty does not apply to products which: 1) are beyond the warranty period or are products for which the original purchaser does not follow the warranty procedures outlined above; 2) have been subjected to electrical, mechanical or chemical damage due to improper, accidental or negligent use; 3) have been modified or altered; 4) anyone other than service personnel authorized by Flowline have attempted to repair; 5) have been involved in accidents or natural disasters; or 6) are damaged during return shipment to Flowline. Flowline reserves the right to unilaterally waive this warranty and dispose of any product returned to Flowline where: 1) there is evidence of a potentially hazardous material present with the product; or 2) the product has remained unclaimed at Flowline for more than 30 days after Flowline has dutifully requested disposition. This warranty contains the sole express warranty made by Flowline in connection with its products. ALL IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY DISCLAIMED. The remedies of repair or replacement as stated above are the exclusive remedies for the breach of this warranty. IN NO EVENT SHALL FLOWLINE BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND INCLUDING PERSONAL OR REAL PROPERTY OR FOR INJURY TO ANY PERSON. THIS WARRANTY CONSTITUTES THE FINAL, COMPLETE AND EXCLUSIVE STATEMENT OF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES OR REPRESENTATIONS ON BEHALF OF FLOWLINE. This warranty will be interpreted pursuant to the laws of the State of California. If any portion of this warranty is held to be invalid or unenforceable for any reason, such finding will not invalidate any other provision of this warranty. For complete product documentation, video training, and technical support, go to www.flowline.com. For phone support, call 562‐598‐3015 from 8am to 5pm PST, Mon ‐ Fri. (Please make sure you have the Part and Serial number available.) Page 72 of 72 MN301700 Rev 9_4
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.5 Linearized : No Author : Phil Create Date : 2014:11:24 09:39:00-08:00 Modify Date : 2014:11:24 09:48:39-08:00 Has XFA : No XMP Toolkit : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:08:04 Producer : Acrobat Distiller 9.5.5 (Windows) Creator Tool : PScript5.dll Version 5.2.2 Metadata Date : 2014:11:24 09:48:39-08:00 Format : application/pdf Title : Microsoft Word - MN301700r9_4 EchoPulse 24Nov14 FCC_209 Creator : Phil Document ID : uuid:872aa881-9854-4405-a88b-f811fb2427c8 Instance ID : uuid:6aca0ab9-dd14-4568-9022-177b410de90b Page Count : 72EXIF Metadata provided by EXIF.tools