Fluke Electronics VERSIV2 CableAnalyzer User Manual Man Users DSX 600 us

Fluke Electronics CableAnalyzer Man Users DSX 600 us

User Manual DSX 302

May 2017©2017 Fluke CorporationAll product names are trademarks of their respective companies.DSX-602CableAnalyzer™Users ManualSoftware Version 5.1
LIMITED WARRANTY AND LIMITATION OF LIABILITYEach Fluke Networks product is warranted to be free from defects in material and workmanship under normal use and service unless stated otherwise herein. The warranty period for the mainframe is one year and begins on the date of purchase. Parts, accessories, product repairs and services are warranted for 90 days, unless otherwise stated. Ni-Cad, Ni-MH and Li-Ion batteries, cables or other peripherals are all considered parts or accessories. The warranty extends only to the original buyer or end user customer of a Fluke Networks authorized reseller, and does not apply to any product which, in Fluke Networks’ opinion, has been misused, abused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke Networks warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke Networks does not warrant that software will be error free or operate without interruption.Fluke Networks authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke Networks. Warranty support is available only if product is purchased through a Fluke Networks authorized sales outlet or Buyer has paid the applicable international price. To the extent permitted by law, Fluke Networks reserves the right to invoice Buyer for repair/replacement when a product purchased in one country is submitted for repair in another country.For a list of authorized resellers, visit www.flukenetworks.com/wheretobuy.Fluke Networks warranty obligation is limited, at Fluke Networks option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke Networks authorized service center within the warranty period.To obtain warranty service, contact your nearest Fluke Networks authorized service center to obtain return authorization information, then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB destination). Fluke Networks assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB destination). If Fluke Networks determines that failure was caused by neglect, misuse, contamination, alteration, accident or abnormal condition of operation or handling, or normal wear and tear of mechanical components, Fluke Networks will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping point).THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE NETWORKS SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.4/15Fluke NetworksPO Box 777Everett, WA 98206-0777USA
iContentsChapter 1 Get AcquaintedOverview of Features ......................................................1Contact Fluke Networks  .................................................2Register Your Product  ....................................................2Additional Resources  ......................................................2Supplements and Updated Manuals  .............................3Kit Contents  ....................................................................3Symbols ............................................................................3WSafety Information ....................................................4Connectors, Keys, and LEDs ............................................8About Link Interface Adapters  ......................................12AC Adapter and Battery  .................................................14Charge the Battery ...................................................14Check the Battery Status ..........................................15Verify Operation  .............................................................16How to Use the Touchscreen  .........................................18Change the Language  ....................................................20Buttons to Do Tests and Save Results ............................20Overview of Memory Functions .....................................22Options for Cable IDs ......................................................22About LinkWare Applications ........................................23LinkWare PC Cable Test Management Software ....23The LinkWare Live Web Application .......................23LinkWare Stats ..........................................................24
iiDSX-602 CableAnalyzerUsers ManualChapter 2 Certify Twisted Pair CablingThe DSX-602 CableAnalyzer Home Screen  ................... 25Make Sure Your Tester is Ready to Certify Cabling ............................................................................ 28Set the Reference  ........................................................... 29Settings for Twisted Pair Tests ....................................... 31How to Do an Autotest .................................................. 36“Bad Patch Cord” Message ............................................ 39Twisted Pair Autotest Results  ........................................ 40PASS*/FAIL* Results  ................................................. 41WIRE MAP Tab  ......................................................... 42PERFORMANCE Tab  ................................................. 44Frequency-Domain Results  ...................................... 45How to Save Frequency-Domain Results as a Plot or a Table ............................................ 45DIAGNOSTIC Tab ...................................................... 50Continuous Tests ...................................................... 51Chapter 3 Certify Coaxial CablingSet the Reference for Coaxial Tests ............................... 53Settings for Coaxial Tests ............................................... 55How to Do an Autotest .................................................. 57Coaxial Autotest Results  ................................................ 60About Splitters  ............................................................... 61Tests Without a Remote  ................................................ 62Continuous Tests  ............................................................ 65Chapter 4 Manage Test ResultsView Saved Results ......................................................... 67How to Replace a Saved Result that Failed  .................. 70
ContentsiiiDelete, Rename, and Move Results ............................... 70Manage Results on a Flash Drive ................................... 72Upload Results to a PC  ................................................... 73View the Memory Status ................................................ 74Chapter 5 Use ProjectsWhy Use Projects?  .......................................................... 75Set Up a Project  .............................................................. 76The PROJECT Screen  ....................................................... 76The CABLE ID SETUP Screen ........................................... 78About Next ID Sets  ......................................................... 80Manage Projects on a Flash Drive  ................................. 81Copy Project Settings to Other Testers  ......................... 82Chapter 6 Sync Projects with LinkWare™ LiveSign Up for a LinkWare Live Account  ........................... 83How to See the Tester’s MAC Address  .......................... 84Use LinkWare Live Through a Wired Ethernet Network .......................................................................... 84Use LinkWare Live Through a Wi-Fi Ethernet Network .......................................................................... 85Change the Network Settings  ....................................... 88Settings for the Wired Port  ..................................... 88Settings for the Wi-Fi Port ....................................... 88Delete Wi-Fi Settings and Passwords ...................... 90About the Asset Management Service .......................... 90Sign Your Tester Out of LinkWare Live ......................... 91Sign In to LinkWare Live from a Desktop or Mobile Device  .......................................................... 91Import Projects from LinkWare Live into LinkWare PC ..................................................................................... 91
ivDSX-602 CableAnalyzerUsers ManualLearn More About LinkWare Live ................................. 91Chapter 7 MaintenanceVerify Operation  ............................................................ 94Clean the Tester  ............................................................. 94See Information About the Tester  ................................ 95Update the Software  ..................................................... 95Use a PC to Update the Software  ........................... 96Use an Updated Main Tester to Update Other Testers ....................................................................... 97Use LinkWare Live to Update the Software ........... 99Extend the Life of the Battery ....................................... 100Store the Tester  .............................................................. 100Remove the Battery  ....................................................... 100Traceable Calibration Period  ......................................... 101Options and Accessories  ................................................ 101If the Tester Does Not Operate as Usual ....................... 102
1List of FiguresFigure Page1. Main Tester Connectors, Keys, and LEDs.....................................82. Remote Tester Connectors, Keys, and LEDs ................................103. How to Attach and Remove Link Interface Adapters.................124. How to Prevent Damage to the Permanent Link Adapter Cables......................................................................135. LEDs Show the Remote’s Battery Status......................................156. Connections to See the Status of a Remote’s Battery ................177. How to Zoom the Screen..............................................................198. FIX LATER, TEST AGAIN, and TEST Buttons and the TEST Key..................................................................................209. The Home Screen for the DSX-602 CableAnalyzer .....................2610. Reference Connections for Twisted Pair Cable ...........................3011. Outlet Configurations ..................................................................3512. Equipment for Autotests on Twisted Pair Cable.........................3613. Permanent Link Connections .......................................................3814. Channel Connections....................................................................3915. PASS* and FAIL* Results...............................................................4116. WIRE MAP Tab ..............................................................................4217. PERFORMANCE Tab ......................................................................4418. Tabular Results Screen for a Frequency-Domain Test ................4619. Plot Screen for a Frequency-Domain Test ...................................4820. Examples of Diagnostic Screens...................................................5021. Reference Connections for Tests on Coaxial Cabling .................5422. Equipment for Tests on Coaxial Cabling .....................................5723. Examples of Connections for Tests on Coaxial Cabling..............59
2DSX-602 CableAnalyzerUsers Manual24. Autotest Results for Coaxial Cabling .......................................... 6025. Connections for Coaxial Tests Without a Remote...................... 6426. RESULTS Screen ............................................................................ 6827. How to Connect the Tester to a PC............................................. 7428. PROJECT Screen ............................................................................ 7729. CABLE ID SETUP Screen(after you enter the first and last IDs) ........................................ 7930. SYNC PROJECTS Screen ................................................................ 8631. How to Connect the Tester to a PC............................................. 9732. How to Connect Units Together to Update the Software......... 9833. How to Remove the Battery ........................................................ 101
1Chapter 1: Get AcquaintedOverview of FeaturesThe DSX-602 CableAnalyzer main and remote units are rugged, hand-held testers that let you certify, troubleshoot, and document copper network cabling. The DSX-602 includes these features:The testers certify twisted pair cabling to Cat 6A/Class EA limits (500 MHz) in less than 10 seconds.Gives a PASS or FAIL result based on a test limit that you specify.You can save approximately 12,700 Cat 6A Autotest results, with plot data, in the tester’s internal memory. You can save more results on a removable flash drive.Taptive™ user interface lets you quickly navigate through different views of the results and see more information about cables.ProjX™ management system lets you set up projects to specify the types of tests and the cable IDs necessary for a job and monitor the progress and status of the job.You can connect the tester to a wired or Wi-Fi network and use the LinkWare™ Live web application to manage your projects from a desktop or mobile device.LinkWare PC software lets you upload test results to a PC and make professional-quality test reports. LinkWare Stats software makes browsable, graphical reports of cable test statistics.
DSX-602 CableAnalyzerUsers Manual2Contact Fluke Networkswww.flukenetworks.com/supportinfo@flukenetworks.com1-800-283-5853, +1-425-446-5500Fluke Networks6920 Seaway Boulevard, MS 143FEverett WA 98203 USAFluke Networks operates in more than 50 countries worldwide. For more contact information, go to our website.Register Your ProductRegistering your product with Fluke Networks gives you access to valuable information on product updates, troubleshooting tips, and other support services. If you purchased a Gold Support plan, registration also activates your plan.To register your product, use LinkWare PC software.Additional ResourcesThe Fluke Networks Knowledge Base answers common questions about Fluke Networks products and provides articles on cable testing techniques and technology. To access the Knowledge Base, log on to www.flukenetworks.com, then click SUPPORT > Knowledge Base.Supplements and Updated ManualsIf necessary, Fluke Networks will put a supplement for this manual, or an updated manual, on the Fluke Networks website. To see if a supplement or updated manual is available, log on to www.flukenetworks.com, click SUPPORT > Manuals, then select a product.
Chapter 1: Get AcquaintedKit Contents3Kit ContentsFor a list of the contents of your DSX-602 kit, see the list that came in the product’s box or see the lists of models and accessories on the Fluke Networks website. If something is damaged or missing, contact the place of purchase immediately.SymbolsTable 1 shows the symbols used on the tester or in this manual. Table 1. SymbolsXWarning: Risk of fire, electric shock, or personal injury.WWarning or Caution: Risk of damage or destruction to equipment or software. See explanations in the manuals.Consult the user documentation.jDo not connect this equipment to public communications networks, such as telephone systems.40 year Environment Friendly Use Period (EFUP) under China Regulation - Administrative Measure on the Control of Pollution Caused by Electronic Information Products. This is the period of time before any of the identified hazardous substances are likely to leak out, causing possible harm to health and the environment.-continued-~This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste.To return unwanted products, contact the manufacturer’s web site shown on the product or your local sales office or distributor.
DSX-602 CableAnalyzerUsers Manual4WSafety InformationWWarningXTo prevent possible fire, electric shock, or personal injury:Read all safety information before you use the Product.Carefully read all instructions.Do not open the case. You cannot repair or replace parts in the case.Do not modify the Product.Use only replacement parts that are approved by Fluke Networks.PConformite Europeene. Conforms to requirements of European Union directives. Safety requirements for electrical equipment for measurement, control, and laboratory use.)Conforms to relevant North American standards.Conforms to relevant Australian standards.«Conforms to relevant Russian standards.ÃKCC-REM-FKN-012001001: EMC approval for KoreaClass A Equipment (Industrial Broadcasting & Communication Equipment) This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.This key turns the tester on and off.Table 1. Symbols
Chapter 1: Get AcquaintedWSafety Information5Do not touch voltages > 30 V AC rms, 42 V AC peak, or 60 V DC.Do not use the Product around explosive gas, vapor, or in damp or wet environments.Use this Product indoors only.Use the Product only as specified, or the protection supplied by the Product can be compromised.Do not use and disable the Product if it is damaged.Do not use the Product if it operates incorrectly.Do not connect the tester to telephony inputs, systems, or equipment, including ISDN inputs. Doing so is a misapplication of this product, which can cause damage to the tester and make a possible shock hazard for the user.Always turn on the tester before you connect it to a link. Doing so activates the tester’s input protection circuitry.Do not operate the Product with covers removed or the case open. Hazardous voltage exposure is possible.Remove the input signals before you clean the Product.Do not put metal objects into connectors.Batteries contain hazardous chemicals that can cause burns or explode. If exposure to chemicals occurs, clean with water and get medical aid.Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.Replace the rechargeable battery after 5 years of moderate use or 2 years of heavy use. Moderate use is defined as recharged twice a week. Heavy use is defined as discharged to cutoff and recharged daily.
DSX-602 CableAnalyzerUsers Manual6Disconnect the battery charger and move the Product or battery to a cool, non-flammable location if the rechargeable battery becomes hot (>50 °C, >122 °F) during the charge period.The battery door must be closed and locked before you operate the Product.The battery door must be closed and locked before you operate the Product.Repair the Product before use if the battery leaks.Recharge the batteries when the low battery indicator shows to prevent incorrect measurements.Turn off the Product and disconnect all test leads, patch cords, and cables before you replace the battery.Do not disassemble or crush battery cells and battery packs.Do not put battery cells and battery packs near heat or fire. Do not put in direct sunlight.Have an approved technician repair the Product.Use only AC adapters approved by Fluke Networks for use with the Product to supply power to the Product and charge the battery.
Chapter 1: Get AcquaintedWSafety Information7WCautionTo prevent damage to the tester or cables under test, to prevent data loss, and to make sure your test results are as accurate as possible:Do not connect the tester to an active network. Doing so causes unreliable test results, can disrupt network operations, and can cause damage to the tester.Connect only RJ45 plugs to the adapters. Other types of plugs, such as RJ11 (telephone) plugs, can cause permanent damage to the jacks.To make sure your test results are as accurate as possible, do the reference procedure every 30 days. See “Set the Reference” on page 29.Do not operate portable transmitting devices, such as walkie-talkies and cell phones, during a cable test. Doing so can cause errors in test results.For permanent link adapters, do not twist, pull on, pinch, crush, or make kinks in the cables. See Figure 4 on page 13.Do not remove the USB flash drive while the LED on the drive flashes. Doing so can corrupt the data on the drive.You can lose a USB flash drive, cause damage to it, or accidentally erase the contents of the drive. Thus, Fluke Networks recommends that you save no more than one day of test results on a flash drive, or that you upload results to LinkWare Live. See Chapter 4.
DSX-602 CableAnalyzerUsers Manual8Connectors, Keys, and LEDsBK88.EPSFigure 1. Main Tester Connectors, Keys, and LEDsConnector for a link interface adapterLCD display with touchscreen: Starts a test. Turns on the tone generator if a remote tester is not connected to the main tester. To start a test, you can also tap TEST on the display. : Power key
Chapter 1: Get AcquaintedConnectors, Keys, and LEDs9: Press  to go to the home screen.Connector for the AC adapter. The LED is red when the battery charges, and green when the battery is fully charged. The LED is yellow if the battery will not charge. See “Charge the Battery” on page 14.RJ45 connector: Lets you connect to a network for access to Fluke Networks cloud services.Micro-AB USB port: This USB port lets you connect the tester to a PC so you can upload test results to the PC and install software updates in the tester.Type A USB port: This USB host port lets you save test results on a USB flash drive.Headset jackNoteIf you have two main testers, you can use one as a remote. To select the remote function, tap TOOLS > Main as Remote.
AFBGCDEHDSX-602 CableAnalyzerUsers Manual10BK42.EPSFigure 2. Remote Tester Connectors, Keys, and LEDsConnector for a link interface adapterPASS LED comes on when a test passes.TEST LED comes on during a test.FAIL LED comes on when a test fails.TALK LED comes on when the talk function is on (). The LED flashes until the main tester accepts the request to talk.
Chapter 1: Get AcquaintedConnectors, Keys, and LEDs11TONE LED flashes and the tone generator comes on if you press  when a main tester is not connected to the remote.LOW BATTERY LED comes on when the battery is low. The LEDs also have these functions: Battery gauge (see Figure 5 on page 15)Volume indicator for the TALK functionProgress indicator for software updates: Starts a test. Turns on the tone generator if a main tester is not connected to the remote.: Power key: Press  to use the headset to speak to the person at the other end of the link. Press again to adjust the volume. To turn off the talk function, hold down .Connector for the AC adapter. The LED is red when the battery charges, and green when the battery is fully charged. The LED is yellow if the battery will not charge. See “Charge the Battery” on page 14.Micro-AB USB port: This USB port lets you connect the tester to a PC so you can install software updates in the tester.Headset jack
DSX-602 CableAnalyzerUsers Manual12About Link Interface AdaptersLink interface adapters let you connect the DSX-602 CableAnalyzer to different types of twisted pair links. Figure 3 shows how to attach and remove adapters.WCautionTo prevent damage to the cables on the permanent link adapters and to make sure your test results are as accurate as possible, do not twist, pull on, pinch, crush, or make kinks in the cables. See Figure 4 on page 13.BK109.EPSFigure 3. How to Attach and Remove Link Interface Adapters
Chapter 1: Get AcquaintedAbout Link Interface Adapters13GPU108.EPSFigure 4. How to Prevent Damage to the Permanent Link Adapter Cables5 in (13 cm)minimum
DSX-602 CableAnalyzerUsers Manual14AC Adapter and BatteryYou can use the AC adapter (model PWR-SPLY-30W) or the lithium ion battery (model VERSIV-BATTERY) to supply power to the tester.To remove the battery, see “Remove the Battery” on page 100.Charge the BatteryBefore you use the battery for the first time, charge the battery for about 2 hours with the tester turned off. To charge the batteryConnect the AC adapter to the 15V jack on the left side of the tester. The LED near the AC adapter connector is red when the battery charges, and green when the battery is fully charged.A fully-charged battery operates for approximately 8 hours of typical use. The battery takes approximately 4 hours to fully charge when the tester is turned off.NotesYou do not need to fully discharge the battery before you recharge it.The battery will not charge if its temperature is outside the range of 32 °F to 104 °F (0 °C to 40 °C). The LED near the connection for the AC adapter is yellow if the battery will not charge.
Chapter 1: Get AcquaintedAC Adapter and Battery15Check the Battery StatusOn a main testerThe battery status icon is in the upper-left corner of the screen:  Battery is full.  Battery is approximately half full.If the AC adapter is not connected, the red bar shows that the battery is very low. Connect the AC adapter to charge the battery and make sure the tester continues to operate.The red bar also shows if the AC adapter is connected, but the battery is not installed.On a remoteThe LEDs show the battery status at the end of the power-up sequence, as shown in Figure 5.BK102.EPSFigure 5. LEDs Show the Remote’s Battery Status84 % - 100 %67 % - 83 %51 % - 66 %34 % - 50 %18 % - 33 %0 % - 17 %
DSX-602 CableAnalyzerUsers Manual16To see more information about a remote’s battery status1Make the connections shown in Figure 6 and turn on both testers.2Make sure the connection icon shows at the top of the screen (). 3Tap TOOLS, then tap Battery Status.When the AC adapter is not connected, the screen shows the Time Remaining, which is the approximate battery life at the present rate of use.Verify OperationThe tester does a self test when you turn it on. If the tester shows an error or does not turn on, refer to “If the Tester Does Not Operate as Usual” on page 102.
Chapter 1: Get AcquaintedVerify Operation17BK148.EPSFigure 6. Connections to See the Status of a Remote’s BatteryDSX-602 with two channel adapters and a patch cordDSX-602 with permanent link and channel adapters
DSX-602 CableAnalyzerUsers Manual18How to Use the TouchscreenThe DSX-602 main unit’s Taptive™ user interface lets you use a touchscreen to control the tester. You can operate the touchscreen with your fingertip or with a stylus that is made for projected capacitance touchscreens.WCautionFor correct operation and to prevent damage to the touchscreen:Touch the screen only with your fingers or with a stylus that is made for projected capacitance touchscreens. Do not use too much force.Do not touch the screen with sharp objects.NoteThe touchscreen will not respond if you tap it with your fingernail or an incorrect type of stylus or if you wear non-conductive gloves.To use the touchscreenTo select an item on the screen, tap the item lightly with your fingertip.To scroll a screen, lightly touch the screen then move your fingertip in the direction you want the screen to move.On screens that show a plot, you can drag the measurement cursor. These screens also have a zoom function, as shown in Figure 7.To clean the touchscreenTurn off the tester, then use a soft, lint-free cloth that is moist with a mild detergent.WCautionWhen you clean the touchscreen, do not let liquid get under the plastic around the touchscreen.
Chapter 1: Get AcquaintedHow to Use the Touchscreen19BA45.EPSFigure 7. How to Zoom the ScreenTo quickly go back to 1:1 magnification, double-tap the screen.To zoom in, use the reverse-pinch gestureTo zoom out, use the pinch gestureTo move the image, drag it in any direction.
DSX-602 CableAnalyzerUsers Manual20Change the LanguageOn the home screen, tap the TOOLS icon, tap Language, then tap a language.Buttons to Do Tests and Save ResultsWhen a test is completed and more than one button shows at the bottom of the screen, the tester highlights one in yellow to recommend which one to tap. Figure 8 shows the buttons you will see.NoteTo change the Auto Save setting, tap the Next ID panel on the home screen.BK40.EPSFigure 8. FIX LATER, TEST AGAIN, and TEST Buttons and the TEST KeyABCDFEG
Chapter 1: Get AcquaintedButtons to Do Tests and Save Results21SAVE (yellow),  TEST (gray): These buttons show if the test passed and Auto Save is off. When you tap SAVE, you can save the results with an ID that you make or select. When you tap TEST, you can select to save the results or do the test again and not save the results.UNSAVED RESULT: This button shows if Auto Save is off and you go to the home screen when a test is completed. Tap this button to see the result.FIX LATER: This button shows if the test failed or had a PASS* result and the result has not been saved. TEST AGAIN: This button shows if the test failed or had a PASS* result. Tap this button to do the test again. If Auto Save is on, the tester saves subsequent results with the same ID. If the test fails again, you can tap FIX LATER to save the result if necessary. When you look at a saved result that failed, tap TEST AGAIN to do the test again for the same ID and with the same test settings as the saved result.TEST (yellow): This button shows if the test passed and Auto Save is on. When Auto Save is on, the tester saves results with the next available ID when the test is completed. When you tap TEST, the tester does a test for the next available ID.: The  key does the same function as the TEST button. When TEST AGAIN shows, you can press  to do a test on the next ID.
DSX-602 CableAnalyzerUsers Manual22Overview of Memory FunctionsYou can save approximately 12,700 Cat 6A Autotest results, with plot data included, in the DSX-602 main tester.The capacity available for test results depends on the space used by the software and custom test limits in the tester. To see the memory statusOn the home screen, tap the TOOLS icon, then tap Memory Status.To make more memory available, you can export results to a USB flash drive, then delete the results in the tester. See “Manage Results on a Flash Drive” on page 72.Options for Cable IDsWhen you save the test results for a cable, you usually give the results the name that is the ID for the cable. There are several methods you can use to make IDs for test results:You can use the CABLE ID SETUP screen to make a set of sequential IDs. The tester uses the IDs in sequence as the names for the results you save. When Auto Save is on, the tester automatically saves each result with the next available ID in the set.A cable ID set also lets you use IDs again so you can add different results to tests you saved before.You can enter an ID each time you do a test. To do this, turn off the Auto Save function (see page 23). Each time a test is completed, tap SAVE (if the test passed) or FIX LATER (if the test failed), then enter an ID manually.You can use LinkWare PC software to make a set of IDs, download the set to the tester, then import it into a project. After you do a test, you can enter the ID for a test you saved before. This lets you replace results.
Chapter 1: Get AcquaintedAbout LinkWare Applications23If the test failed before, and you saved the results, you can select it on the RESULTS screen, then press TEST AGAIN to replace the results for that ID.NotesCable IDs are case-sensitive. For example, the tester saves result with the names “A0” and “a0” in two different records.A cable ID can have a maximum of 60 characters.If you delete all the ID sets in a project, the tester makes a default set that starts with 001.To turn the Auto Save function on or off1On the home screen, tap the Next ID panel.2On the CHANGE ID screen, tap the On/Off control next to Auto Save.3Tap DONE.About LinkWare ApplicationsLinkWare PC Cable Test Management SoftwareThe LinkWare PC Cable Test Management software lets you upload test records to a PC, organize and examine test results, print professional-quality test reports, and do software updates and other maintenance procedures on your tester.You can download LinkWare PC from the Fluke Networks website.The LinkWare Live Web ApplicationThe LinkWare Live web application lets you manage your projects from a desktop or mobile device. To get started with LinkWare Live, see Chapter 6.
DSX-602 CableAnalyzerUsers Manual24LinkWare StatsThe LinkWare Stats Statistical Report software that is included with LinkWare PC software provides statistical analysis of cable test reports and generates browsable, graphical reports. For instructions about LinkWare PC and LinkWare Stats software, see the guides for getting started and the online help available under Help on the LinkWare PC and LinkWare Stats menus.
25Chapter 2: Certify Twisted Pair CablingWWarningXBefore you use the DSX-602 CableAnalyzer, read the safety information that starts on page 3.The DSX-602 CableAnalyzer Home ScreenThe home screen (Figure 9) shows important test settings. Before you do a test, make sure these settings are correct.
DSX-602 CableAnalyzerUsers Manual26BK110.EPSFigure 9. The Home Screen for the DSX-602 CableAnalyzerPROJECT: The project contains the settings for a job and helps you monitor the status of a job. When you save test results, the tester puts them in the project. Tap the PROJECT panel to edit the project settings, select a different project, or make a new project. Shows a summary of the test results in the project: : The number of tests that passed.AEBFKJCGHDLMNI
Chapter 2: Certify Twisted Pair CablingThe DSX-602 CableAnalyzer Home Screen27: The number of tests that failed.: The number of tests with an overall marginal result.The test setup panel shows the settings the tester will use when you tap TEST or press . To change these settings, tap the panel.Icons show the status of the Store Plot Data and AC Wire Map settings. See Table 2 on page 33.Next ID: The Next ID panel shows the ID that the tester gives to the next test results you save. Tap Next ID to do these tasks:Enter an ID, select a different ID in the ID set, select a different set of IDs, or make a new set. The tester adds the IDs and ID sets you make to the project that shows on the home screen.Turn Auto Save on or off.Operator: The name of the person who does the job. You can enter a maximum of 20 operator names. For each operator you can also enter the email address that the operator will use as an ID to sign in to LinkWare Live.TOOLS: The TOOLS menu lets you set the reference, see the status of the tester, and set user preferences such as the language and the display brightness. RESULTS: Tap RESULTS to see and manage the results that are saved in the tester.SYNC: Tap SYNC to sync projects with LinkWare Live.TEST: Tap TEST to do the test shown in the test setup panel.The percentage of the project that is completed. The percentage is the number of IDs used for saved results divided by the total number of used and available IDs in the project.  % Tested does not show if your project contains only a Next ID list. See “Each project can have up to 5000 IDs. If an ID set does
DSX-602 CableAnalyzerUsers Manual28not have a Last ID, the tester counts the set as one ID. An ID can have a maximum of 60 characters. Symbols, such as the asterisk, and accented characters do not increment.” on page 78 for more information about the Next ID list. This icon shows when the tester’s link interface adapter is connected to the adapter on a DSX-602 remote and the remote is turned on.  The asset management icon shows when the owner of a LinkWare Live account has enabled the asset management service on the tester. See “About the Asset Management Service” on page 90. This icon shows when the talk function is on. To use the talk function: 1Connect the main and remote testers together through a link that has one or more good wire pairs.2Connect headsets to the headset jacks on the testers.3Press the button on one of the headset microphones or press  on the remote, then speak into the microphone. Make Sure Your Tester is Ready to Certify CablingTo make sure your tester meets its accuracy specifications, follow these guidelines:Keep the tester’s software current. The latest software is available on the Fluke Networks website. See “Update the Software” on page 95.Set the reference for the twisted pair adapters every 30 days. See page 29.Make sure that you select the correct cable type for the job, and that the NVP for the cable is correct. See Table 2 on
Chapter 2: Certify Twisted Pair CablingSet the Reference29page 32.Make sure you select the correct test limit for the job. See Table 2 on page 32.Make sure the cords and connectors for all test equipment and patch cords are in good condition.Make sure the battery is fully charged.Send the testers to a Fluke Networks service center every 12 months for factory calibration.Set the ReferenceThe reference procedure for twisted pair cable sets the baseline for insertion loss, ACR-F, and DC resistance measurements.Set the reference at these times:Every 30 days, at minimum.To ensure maximum accuracy of test results, set the reference daily.It is not necessary to set the reference when you change the link interface adapters.To set the reference1Turn on the tester and the remote a minimum of 5 minutes before you set the reference. NoteSet the reference only after the testers are at an ambient temperature between 50 °F and 104 °F (10 °C and 40 °C).2Use the reference patch cord and two channel adapters or permanent link and channel adapters to connect the main and remote testers together as shown in Figure 10.3On the home screen, tap TOOLS, then tap Set Reference.4On the SET REFERENCE screen tap TEST.
DSX-602 CableAnalyzerUsers Manual30BK89.EPSFigure 10. Reference Connections for Twisted Pair CablePermanent link adapterChannel adapter6 inch (15 cm) reference patch cordChannel adapters
Chapter 2: Certify Twisted Pair CablingSettings for Twisted Pair Tests31Settings for Twisted Pair TestsTable 2 gives descriptions of the settings for twisted pair tests. To set up a project, which includes the settings in Table 2, cable IDs, and operator names, see Chapter 5.To set up a twisted pair test1On the home screen, tap the test setup panel.2On the CHANGE TEST screen, select a twisted pair test to change, then tap EDIT. Or to set up a new twisted pair test, tap NEW TEST.3On the TEST SETUP screen, tap the panels to change settings for the test. See Table 2.4On the TEST SETUP screen, tap SAVE when your test setup is completed.5On the CHANGE TEST screen, make sure the button next to the test is selected, then tap USE SELECTED.
DSX-602 CableAnalyzerUsers Manual32Table 2. Settings for Twisted Pair Tests Setting DescriptionCable Type Select a cable type that is correct for the type you will test. To see a different group of cable types, tap MORE, then tap a group. To make a custom cable type, tap Custom in the Cable Groups list.NVPNominal velocity of propagation. The tester uses the NVP and the propagation delay to calculate the length of the cable.The default value is defined by the selected cable type and is the typical NVP for that cable type. To enter a different value, tap the NVP panel, then tap  or   on the NVP screen to increase or decrease the value. To find the actual value for a cable, connect a known length of the cable to the tester, tap MEASURE on the NVP screen, then change the NVP until the measured length matches the known length. Use a cable at least 30 m (100 ft) long.When you increase the NVP value, the calculated length increases.Shield Test This setting shows only when you select a shielded cable type. On: The wire map test includes a DC test for shield continuity and AC tests for shield quality. The wire map test fails if the shield is open or the AC test results are unsatisfactory. Off: The wire map shows the shield if the shield has continuity. The tester does not do AC tests for shield quality. The wire map test does not fail or show the shield if the shield is open.Test Limit Select the correct test limit for the job. To see a different group of limits, tap MORE, then tap the name of a group.
Chapter 2: Certify Twisted Pair CablingSettings for Twisted Pair Tests33Setting DescriptionStore Plot Data Off : The tester does not save plot data for frequency-domain tests or for the HDTDR/HDTDX analyzers. You can see the plots before you save the test and exit the results screen. The saved results show frequency-domain measurements in a table and do not include the HDTDR/HDTDX plots.   On : The tester saves plot data for all frequency-domain tests required by the selected test limit and for the HDTDR/HDTDX analyzers. HDTDR/HDTDX Fail/Pass* only: The tester shows HDTDR and HDTDX analyzer results only for Autotests with PASS*, FAIL*, or FAIL results.All Autotests: The tester shows HDTDR and HDTDX analyzer results for all Autotests.To get HDTDR/HDTDX analyzer results you can also tap TOOLS > Diagnostics.For more information about the HDTDR and HDTDX analyzers, see the Technical Reference Handbook.Outlet ConfigurationThe Outlet Configuration specifies which wire pairs are tested and which wire numbers the wire map shows for the pairs. See Figure 11. To see the wire map for a configuration, tap Outlet Configuration, tap the configuration name on the OUTLET CONFIG screen, then tap SAMPLE. To select a configuration, tap a name on the OUTLET CONFIG screen, then tap USE SELECTED. NoteThe OUTLET CONFIG screen shows only the configurations that are applicable to the selected Test Limit.To make a custom outlet configuration, tap CUSTOM on the OUTLET CONFIG screen, then tap MANAGE, then tap Create.Table 2. Settings for Twisted Pair Tests (continued)
DSX-602 CableAnalyzerUsers Manual34Setting DescriptionAC Wire Map The AC Wire Map test lets you do tests on links connected through midspan PoE (Power over Ethernet) devices. See the Technical Reference Handbook. When the AC Wire Map test is on, this icon shows on the home screen: NotesAlways turn off the AC wire map test when you will not do tests through a PoE device. The AC wire map test increases the time for an Autotest. It also disables the resistance and shield continuity tests.The DSX-8000 modules do not support the AC wire map test.Table 2. Settings for Twisted Pair Tests (continued)
Chapter 2: Certify Twisted Pair CablingSettings for Twisted Pair Tests35GPU85.EPSFigure 11. Outlet ConfigurationsT568ARolloverCSU/DSUATM/TP-PMDStraightATM/TP-PMDCrossedEthernet Two-Pair CrossedEthernet Two-Pair Token RingT568BUSOC Single-PairUSOC Two-PairCrossover1000BASE-T Crossover 2 x Two-Pair CrossedOne Pair (1,2)M12 Two-Pair CrossedM12 Two-Pair
DSX-602 CableAnalyzerUsers Manual36How to Do an AutotestWhen you tap TEST on the main tester or press  on the main or remote tester, the testers do an Autotest. The Autotest includes all the tests necessary to certify that the cabling meets or exceeds the performance requirements specified in the selected test limit.Figure 12 shows the equipment for Autotests on twisted pair cable.Figure 12. Equipment for Autotests on Twisted Pair CableBK111.EPSMain and remote testersFor tests on permanent links: two permanent link adapters (PRO model or optional)For tests on channels: two channel adaptersAC adapters (optional)ACDB
Chapter 2: Certify Twisted Pair CablingHow to Do an Autotest37To do an Autotest on twisted pair cable1Attach permanent link or channel adapters to the main and remote testers.2Make sure that the home screen shows the correct settings for the job. To make sure that other settings are correct, tap the test setup panel, make sure the correct test is selected on the CHANGE TEST SCREEN, then tap EDIT to see more settings. Table 2 on page 32 describes the settings.3Connect the testers to the link as shown in Figure 13 or 14.4Tap TEST on the main tester or press  on the main or remote tester.If the tester at the other end of the cable is in sleep mode or is off, your tester’s tone generator turns on the other tester. If the two testers are not connected: Your tester’s tone generator stays on. Then, you can use a tone probe if necessary to find the cable to connect to the other tester.Or, tap MEASURE to do the tests that do not require a remote tester. Because the tester cannot complete all tests and some tests always fail with no remote connected, the result for an Autotest without a remote is always FAIL.
DSX-602 CableAnalyzerUsers Manual38BK97.EPSFigure 13. Permanent Link ConnectionsEnd permanent linkRemote with permanent link adapterOptional consolidation pointWall outletTester with permanent link adapterStart permanent linkPatch panelHorizontal cabling
Chapter 2: Certify Twisted Pair Cabling“Bad Patch Cord” Message39BK96.EPSFigure 14. Channel Connections“Bad Patch Cord” MessageTo comply with standards for tests on channels, the tester removes the effects of the channel adapters and their connections from the test results. Before it removes these effects, the tester makes sure that the plugs on the patch cord do not have too much near-end crosstalk (NEXT). Too much NEXT is frequently caused by too much untwisted wire in the plug. If a plug is bad, the tester shows the message Bad patch cord at main or Bad patch cord at remote, and does not remove the effects of the channel adapters and their connections. The tester saves the message with the results.If you see one of these messages, replace the patch cord or install a new plug at the bad end. End channelRemote with channel adapterOptional consolidation pointWall outletTester with channel adapterStart channelHub or switchHorizontal cablingPatch cord from hub or switchPatch cord from PCPatch panels
DSX-602 CableAnalyzerUsers Manual40Twisted Pair Autotest ResultsThe tests listed below apply to twisted pair cabling. NoteSome tests are not included in some test limits.Wire mapResistanceLengthPropagation delayDelay skewInsertion loss (attenuation)ImpedanceNEXT (near-end crosstalk)PS NEXT (power-sum near-end crosstalk)ACR-N (attenuation to crosstalk ratio at the near end)PS ACR-N (power-sum attenuation to crosstalk ratio, near end)ACR-F (attenuation to crosstalk ratio at the far end)PS ACR-F (power-sum attenuation to crosstalk ratio, far end)Return lossHDTDR and HDTDX analyzers (optional tests, not required by any test limit)
Chapter 2: Certify Twisted Pair CablingTwisted Pair Autotest Results41PASS*/FAIL* ResultsA result shows an asterisk when measurements are in the tester’s accuracy uncertainty range (Figure 15) and the asterisk is required by the selected test limit. These results are marginal.  A PASS* shows that the cable’s performance is satisfactory. If a cable must get a PASS result to agree with your requirements for quality, identify and correct the problems with the cable and do the Autotest again. Usually, a FAIL* is not a satisfactory result. The tester shows a FAIL for the overall result. Identify and correct the problems with the cable and do the Autotest again.GPU87.EPSFigure 15. PASS* and FAIL* ResultsTester’s accuracy uncertainty rangePASSPASS*FAILLimit FAIL*PASS*
DSX-602 CableAnalyzerUsers Manual42WIRE MAP TabThe WIRE MAP tab shows the connections between the ends of the cable under test. The tester compares the connections to the selected Outlet Configuration to get a PASS or FAIL result. If the wire map test fails, you can continue or stop the Autotest. Or, you can tap SCAN ON to do the wire map test continuously while you look for the fault. To continue the Autotest after you correct the fault, tap SCAN OFF, then tap CONTINUE.If you leave then return to the WIRE MAP FAIL screen, the SCAN ON button goes away. To see the button again, tap TEST AGAIN. You can also select the continuous wire map test as a single test from the TOOLS menu. See “Continuous Tests” on page 51.Figure 16 shows an example of a wire map screen. For information on AC wire map screens, see the Technical Reference Handbook.BA59.EPSFigure 16. WIRE MAP TabADEFBC
Chapter 2: Certify Twisted Pair CablingTwisted Pair Autotest Results43The name of the outlet configuration used for the test. The outlet configuration is a setting on the TEST SETUP screen. The wire map of the cabling. The main tester is at the left side of the wire map.Tap   to see information about wire map faults. If   shows, tap it to see a message about the results, such as Bad patch cord at remote.The overall result for the Autotest. If the result shows an asterisk, See “PASS*/FAIL* Results” on page 41.The result for the wire map test: The wire map does not agree with the outlet configuration selected for the test. The wire map agrees with the outlet configuration selected for the test.When more than one button shows at the bottom of the screen, the tester highlights one in yellow to recommend which one to tap. See “Buttons to Do Tests and Save Results” on page 20.
DSX-602 CableAnalyzerUsers Manual44PERFORMANCE TabThe PERFORMANCE tab (Figure 17) shows the overall result for each test that is required by the selected test limit.BA86.EPSFigure 17. PERFORMANCE TabThe test limit and cable type used for the test. To see all the settings used for the test, tap the panel.To see detailed results for a test, tap the panel.The overall result for the Autotest. If the result shows an asterisk, See “PASS*/FAIL* Results” on page 41.The overall result for the test: The results exceed the limit.BCDEA
Chapter 2: Certify Twisted Pair CablingTwisted Pair Autotest Results45 The results are within the limit.   The selected test limit does not have a limit for the test, or a dB rule applies. See the Technical Reference Handbook.   The results are within the range of accuracy uncertainty for the tester. See “PASS*/FAIL* Results” on page 41.The measurement shown for frequency-domain results is the worst margin. (The insertion loss plot is different. See the Technical Reference Handbook.)When more than one button shows at the bottom of the screen, the tester highlights one in yellow to recommend which one to tap. See “Buttons to Do Tests and Save Results” on page 20.Frequency-Domain ResultsFrequency-domain results are the measurements that change with frequency, such as insertion loss and crosstalk.How to Save Frequency-Domain Results as a Plot or a TableIf Store Plot Data is on when you do a test, the saved results show as plots. If Store Plot Data is Off, the tester does not save plot data for frequency-domain tests or for the HDTDR/HDTDX analyzers. You can see the plots before you save the test and exit the results screen. Figures 18 and 19 show examples of the two types of screens. Also see “Store Plot Data” on page 33.
DSX-602 CableAnalyzerUsers Manual46BA104.EPSFigure 18. Tabular Results Screen for a Frequency-Domain TestThe location where the tester made the measurements. To switch between results for the main and remote, tap REMOTE or MAIN ().The results are for the wire pair or pairs shown. To see the results for a different pair or pairs, tap a tab on the right side of the screen ().WORST MARGIN is the measurement that is nearest to the limit line or exceeds the limit by the largest amount. WORST VALUE is the worst measurement.DBHGICEFA
Chapter 2: Certify Twisted Pair CablingTwisted Pair Autotest Results47The measured value.The limit specified by the selected test limit.MARGIN is the difference between the measured value and the limit. The value is in a red box if the measurement exceeds the limit.To switch between results for the main unit and the remote, tap REMOTE or MAIN.To see the results for a different pair or pairs, tap a tab.The result for the pair. If the result shows an asterisk, see “PASS*/FAIL* Results” on page 41.
DSX-602 CableAnalyzerUsers Manual48BA71.EPSFigure 19. Plot Screen for a Frequency-Domain TestThe location of the measurements. To switch between results for the main and remote, tap REMOTE or MAIN ().Measured values for the wire pairs. The limit line (in red) for the measurement. NoteIf the limit line is black, the tester does not evaluate the measurement at those frequencies because a dB rule applies. See the Technical Reference Handbook. BFJGKMACDEH IL
Chapter 2: Certify Twisted Pair CablingTwisted Pair Autotest Results49The vertical scale is the measured value in decibels. The horizontal scale is the frequency range in megahertz. To see help for the screen, tap  .To switch between results for the main unit and the remote, tap REMOTE or MAIN.The margin at the cursor’s location. The margin is the difference between the measured value and the limit. The margin is negative if the pair failed. The measured value at the cursor’s location. When you first look at the plot, the cursor is at the frequency of the worst margin. To move the cursor to the worst value, tap WORST VALUE. (The insertion loss plot is different. See the Technical Reference Handbook.)To see the plots for pairs, tap   or  . To select pairs to show on the plot, touch   or   for one second to see the SELECT PAIRS window. Select the pairs you want to see, then tap OK.When you first look at the plot, the cursor is at the frequency of the worst margin. To move the cursor to the worst value, tap WORST VALUE (). The box at the bottom of the cursor shows the frequency at the cursor’s position. To move the cursor to other points, touch and drag the yellow circle at the top of the cursor. To move the cursor in small increments, tap the yellow circle, then tap the arrow buttons that show on the plot (  or  ).The overall result for the test. If you look at pairs, the result is for those pairs. If the result shows an asterisk, see “PASS*/FAIL* Results” on page 41.To zoom in and out, use the pinch, reverse-pinch, and double-tap gestures on the touchscreen. You can also use the zoom controls to change the magnification on the frequency and decibels scales independently.
DSX-602 CableAnalyzerUsers Manual50DIAGNOSTIC TabIf an Autotest on twisted pair cabling fails or has marginal results, the DSX-602 CableAnalyzer automatically gives you HDTDR and HDTDX plots to help you find faults. To see the plots, tap the DIAGNOSTIC tab, then tap the HDTDR or HDTDX panel (Figure 20).To get only diagnostics results, select Diagnostics from the TOOLS menu. These results do not include a PASS/FAIL status.For more information on the HDTDR and HDTDX plots, see the Technical Reference Handbook.BA192.EPSFigure 20. Examples of Diagnostic Screens
Chapter 2: Certify Twisted Pair CablingTwisted Pair Autotest Results51Continuous TestsTo do the wire map, length, or resistance test continuously, go to the home screen, tap TOOLS > Single Tests, then tap a test. The wire map test compares the results to the outlet configuration specified by the selected test limit and shows   if the connections agree or   if they do not.The length and resistance tests do not compare the results to a test limit.To save the result, tap SCAN OFF > SAVE. The saved test has an   for the overall result.
DSX-602 CableAnalyzerUsers Manual52
53Chapter 3: Certify Coaxial CablingThe optional DSX-CHA003 coaxial adapters let you use the DSX-600 CableAnalyzer to certify coaxial cabling for network and video applications.Set the Reference for Coaxial TestsTo use DSX-CHA003 adapters, you must set the reference for coaxial tests. The reference procedure sets a baseline for insertion loss and resistance measurements. Set the reference at these times:Every 30 days, at minimum. To ensure maximum accuracy of test results, set the reference daily.NoteIt is not necessary to set the reference again if you use different coaxial adapters.To set the reference1Make the connections shown in Figure 21. 2Turn on the tester and the remote a minimum of 5 minutes before you set the reference. -continued-
DSX-602 CableAnalyzerUsers Manual54NotesSet the reference only after the testers are at an ambient temperature between 10 °C and 40 °C (50 °F and 104 °F).The tester will not let you set the reference if the patch cord is longer than 30 cm (12 in).You can also set the reference with a 50 patch cord3On the home screen, select a coaxial cable test.4On the home screen, tap TOOLS, then tap Set Reference.5On the SET REFERENCE screen tap TEST.BK179.EPSFigure 21. Reference Connections for Tests on Coaxial CablingDSX-CHA003COAX ADAPTERDSX-CHA003COAX ADAPTERCoaxial adapter with F connector adapter30 cm (12 in) 75 coaxial patch cordCoaxial adapter with F connector adapter
Chapter 3: Certify Coaxial CablingSettings for Coaxial Tests55Settings for Coaxial TestsTable 3 gives descriptions of the settings for coaxial tests. To set up a project, which includes the settings in Table 3, cable IDs, and operator names, see Chapter 5.To set up a coaxial test1On the home screen, tap the test setup panel.2On the CHANGE TEST screen, select a coaxial test to change, then tap EDIT. Or to set up a new coaxial test, tap NEW TEST. 3On the TEST SETUP screen, tap the panels to change settings for the test. See Table 3.4On the TEST SETUP screen, tap SAVE when your test setup is completed.5On the CHANGE TEST screen, make sure the button next to the test is selected, then tap USE SELECTED.
DSX-602 CableAnalyzerUsers Manual56Table 3. Settings for Coaxial Tests Setting DescriptionCable Type Select a cable type that is correct for the type you will test. To see a different group of cable types, tap MORE, then tap a group. To make a custom cable type, tap Custom in the Cable Groups list.NVPNominal velocity of propagation. The tester uses the NVP and the propagation delay to calculate the length of the cable.The default value is defined by the selected cable type and is the typical NVP for that cable type. To enter a different value, tap the NVP panel, then tap  or   on the NVP screen to increase or decrease the value. To find the actual value for a cable, connect a known length of the cable to the tester, tap MEASURE on the NVP screen, then change the NVP until the measured length matches the known length. Use a cable at least 30 m (100 ft) long.When you increase the NVP value, the calculated length increases.Test Limit Select the correct test limit for the job. To see a different group of limits, tap MORE, then tap the name of a group. Store Plot Data Off : The tester does not save plot data for insertion loss or for the HDTDR analyzer. You can see the plots before you save the test and exit the results screen. The saved results show insertion loss measurements in a table and do not include the HDTDR plot.On : The tester saves plot data for the insertion loss test and for the HDTDR analyzer.
Chapter 3: Certify Coaxial CablingHow to Do an Autotest57How to Do an AutotestFigure 22 shows the equipment for tests on coaxial cabling.NotesYou can do the HDTDR, length, and resistance tests without a remote tester. See “Tests Without a Remote” on page 62.If you have two main testers, you can use one as a remote. To select the remote function, tap TOOLS > Main as Remote.Figure 22. Equipment for Tests on Coaxial CablingBK181.EPSMain and remote testersTwo DSX-CHA003 adapters For tests on network cabling: F-connector to BNC adapters and patch cords (typically 50 )For tests on video cabling: F-connector adapters and patch cords (typically 75 )AC adapters (optional)ACBDE
DSX-602 CableAnalyzerUsers Manual58To do an Autotest1Attach coaxial adapters to the main and remote testers.2Make sure that the home screen shows the correct settings for the job. To make sure that other settings are correct, tap the test setup panel, make sure the correct test is selected on the CHANGE TEST screen, then tap EDIT to see more settings. Table 3 on page 56 describes the settings.3Connect the testers to the link as shown in Figure 23.WCautionTo make sure your results are reliable:Disconnect all taps and devices from the cable.Do not do tests through splitters (See “About Splitters” on page 61).4Tap TEST on the main tester or press  on the main or remote tester.If the tester at the other end of the cable is in sleep mode or is off, your tester’s tone generator turns on the other tester. If the two testers are not connected:Your tester’s tone generator stays on so you can use a tone probe to find the cable to connect to the other tester.Or, tap MEASURE to do the length and resistance tests, which do not require a remote tester. Because the tester cannot complete all tests, and the reflection at the end of the cable exceeds the 15% limit for the HDTDR test, the result for an Autotest without a remote is always FAIL.
Chapter 3: Certify Coaxial CablingHow to Do an Autotest59BK184.EPSFigure 23. Examples of Connections for Tests on Coaxial CablingDSX-CHA003COAX ADAPTERDSX-CHA003COAX ADAPTERDSX-CHA003COAX ADAPTERDSX-CHA003COAX ADAPTERDSX-CHA003COAX ADAPTERDSX-CHA003COAX ADAPTERDSX-CHA003COAX ADAPTERDisconnect all drop cablesMoCA adapter MoCA adapterRouterIP camerasData networkSurveillance networkCable TVCable spoolsCoaxial segmentNote: Do not do tests through splitters.
DSX-602 CableAnalyzerUsers Manual60Coaxial Autotest ResultsNoteNot all test limits include all the tests shown in Figure 24.GPU182.EPSFigure 24. Autotest Results for Coaxial CablingThe test limit and cable type used for the test.To see detailed results for a test, tap the panel.The overall result for the Autotest. If the result shows an asterisk, See “PASS*/FAIL* Results” on page 41.ACEFBD
Chapter 3: Certify Coaxial CablingAbout Splitters61The DIAGNOSTIC tab shows the HDTDR analyzer button, which you can tap to see the HDTDR plot. The plot helps you find faults on the cable. The HDTDR plot for coaxial cable includes limit lines and a PASS/FAIL result.The overall result for the test: The results exceed the limit. The results are within the limit.   The selected test limit does not have a limit for the test.   The results are within the range of accuracy uncertainty for the tester. See “PASS*/FAIL* Results” on page 41.The measurement shown for the insertion loss plot is the worst value if the test passed, or the worst margin if the test failed.When more than one button shows at the bottom of the screen, the tester highlights one in yellow to recommend which one to tap. See “Buttons to Do Tests and Save Results” on page 20.To do the resistance test continuously, select the test from the TOOLS menu. This function helps you locate intermittent faults.About SplittersIf you get these results, there might be a splitter on the cable:The tester cannot find the remote.The tester looses communication with the remote. The test might continue, then loose communication again as the splitter interferes with the communication signal.The length test shows End not found.The resistance test shows an open.The HDTDR plot shows a reflection that has an unusual shape.Because splitters can cause unreliable test results, you should not do tests through them.
DSX-602 CableAnalyzerUsers Manual62Tests Without a RemoteYou can do the length, resistance, and HDTDR tests without a remote tester. Table 4 describes the effects of a remote on tests.1Attach a coaxial adapter to the main tester.2Make sure that the home screen shows the correct settings for the job. To make sure that other settings are correct, tap the test setup panel, make sure the correct test is selected on the CHANGE TEST screen, then tap EDIT to see more settings. Table 3 on page 56 describes the settings.3Connect the tester as shown in Figure 25.4To do an Autotest: Tap TEST on the main tester or press  on the main or remote tester. When the MEASURE button shows, tap it to do the length and resistance tests, which do not require a remote tester. To do only the length or resistance test: On the home screen tap TOOLS > Single Tests, then tap a test. To do only the HDTDR test: On the home screen tap TOOLS > Diagnostics, then tap HDTDR. NoteBecause the tester cannot complete all tests, and the reflection at the end of the cable exceeds the 15% limit for the HDTDR test, the result for an Autotest without a remote is always FAIL.
Chapter 3: Certify Coaxial CablingTests Without a Remote63Table 4. Remote Requirements for Coaxial TestsTest Remote Requirements*HDTDR analyzerOptional. Without a remote, the plot shows large reflections at the end of the cabling.Resistance A remote or a terminator is required for a loop resistance measurement.Length Not required. Because a coaxial cable terminator eliminates signal reflections, the tester cannot measure the length of terminated cabling. The tester shows End not found in this case.Impedance Optional.Without a remote or terminator, the tester cannot measure the impedance of cables longer than 984 ft (300 m). The tester shows Unknown in this case.Propagation delayNot required.Because a coaxial cable terminator eliminates signal reflections, the tester cannot measure the propagation delay of terminated cabling. In this case, the tester shows End not found.Insertion Loss Required.* If a remote is not required for a test, the test runs without activating the toner when no remote is detected.
DSX-602 CableAnalyzerUsers Manual64BK183.EPSFigure 25. Connections for Coaxial Tests Without a RemoteDSX-CHA003COAX ADAPTERDSX-CHA003COAX ADAPTERFor length measurements, disconnect terminators. See Table 4 on page 63.
Chapter 3: Certify Coaxial CablingContinuous Tests65Continuous TestsTo do the length or resistance test continuously, go to the home screen, tap TOOLS > Single Tests, then tap a test.The length and resistance tests do not compare the results to a test limit.To save the result, tap SCAN OFF > SAVE. The saved test has an   for the overall result.
DSX-602 CableAnalyzerUsers Manual66
67Chapter 4: Manage Test ResultsView Saved ResultsOn the home screen, tap the RESULTS icon. The RESULTS screen shows the results in the active project. See Figure 26.To view results saved on a USB flash drive, connect the drive, then tap RESULTS, TRANSFER, USB Flash Drive, Import. See “Manage Results on a Flash Drive” on page 72.To organize results and make reports you can give to customers, use LinkWare PC software.
DSX-602 CableAnalyzerUsers Manual68BA24.EPSFigure 26. RESULTS ScreenThe name of the active project.: The number of results that passed. This includes individual results for each ID and tests that have an   result. : The number of results that failed. This includes individual results for each ID. KCDABE F GIHJ
Chapter 4: Manage Test ResultsView Saved Results69: The number of DSX CableAnalyzer results that have PASS* results. PASS* results have measurements within the range of accuracy uncertainty for the tester. See “PASS*/FAIL* Results” on page 41.The cable IDs that have FAIL results and must be tested again. The cable IDs that have an overall PASS or   result. Tap VIEW ALL to see a summary of the results in all the projects in the tester.TRANSFER lets you export or import results to or from a flash drive and delete results on the flash drive.MANAGE lets you move results to a different project, rename results, or delete results that are in the tester.The scroll bar shows when the list of results is long. To use the scroll bar, tap on the bar or slide your fingertip on the bar. For example, to see the 12th result in the list, tap on “12” in the scroll bar. When you slide your fingertip on the bar, the number of the result you can see is next to your fingertip.Tap the ID/Date control to sort the results by cable ID or by date. When you sort by ID, the results show in ascending order. When you sort by date, the latest result is at the top of the list.These icons show when you connect the tester to a network to use Fluke Networks cloud services (see Chapter 6): The tester is connected to a wireless network.  The tester is connected to a wired network.  The asset management icon shows when the owner of a LinkWare Live account has enabled the asset management service on the tester. See “About the Asset Management Service” on page 90.
DSX-602 CableAnalyzerUsers Manual70How to Replace a Saved Result that FailedTo use the same test settings that were used for the saved result1On the home screen, tap the RESULTS icon.2On the RESULTS screen, tap a result that failed.3Tap TEST AGAIN.4When the test is completed, and if Auto Save is on, the tester asks you if you want to overwrite the results. Tap Yes. If Auto Save is off, tap FIX LATER (if the test failed) or SAVE (if the test passed) to save the result.To replace a result with a result that uses different test settings1Turn off Auto Save.2Make sure that the home screen shows the project that contains the result you want to replace.3Select the necessary test settings.4Do the test, tap FIX LATER (if the test failed) or SAVE (if the test passed), then enter the ID of the saved result.5The tester asks you if you want to overwrite the results. Tap Yes. Delete, Rename, and Move ResultsBefore you delete, rename, or move results, select the project that contains the results and go to the MANAGE RESULTS screen:1On the home screen, tap the RESULTS icon. The RESULTS screen shows the results in the active project. 2To see the results in another project, tap VIEW ALL, then tap a project.3Tap MANAGE to see the MANAGE RESULTS screen.
Chapter 4: Manage Test ResultsDelete, Rename, and Move Results71To delete results1On the MANAGE RESULTS screen, select the results you want to delete.To select all the tests that failed or all the tests that passed, tap Select All Retests or Select All Passes.2Tap DELETE, then tap DELETE in the confirmation dialog.To rename results1On the MANAGE RESULTS screen, select one result to rename.2Tap RENAME.3Enter a new name, then tap DONE.To move results to a different project1On the MANAGE RESULTS screen, select the results you want to move.2Tap MOVE.To move the results to a project shown in the list, tap the project name, then tap MOVE in the confirmation dialog.To make a new project and move the results to the new project, tap NEW PROJECT, enter a project name, tap DONE, then tap MOVE in the confirmation dialog.NoteWhen you move results to a different project, that project becomes the active project.
DSX-602 CableAnalyzerUsers Manual72Manage Results on a Flash DriveYou can export or import results to or from a flash drive, and delete results on the flash drive.To export or import project settings along with results, see “Manage Projects on a Flash Drive” on page 81.WCautionDo not remove the USB flash drive while the LED on the drive flashes. Doing so can corrupt the data on the drive.You can lose a USB flash drive, cause damage to it, or accidentally erase the contents of the drive. Thus, Fluke Networks recommends that you save no more than one day of test results on a flash drive. NoteThe tester reads only USB drives that use the FAT format.1Connect a USB flash drive to the type A USB port. The tester makes a bell sound when it detects the drive.2On the home screen, tap the RESULTS icon, then tap TRANSFER.3On the TRANSFER RESULTS screen, tap USB Flash Drive, then tap a function:Export: On the EXPORT RESULTS screen, select New or All, select the project that contains the results you want to export to the flash drive, then tap EXPORT.New: Export only results that do not have the same IDs as results that are already on the flash drive. All: Export all the results from all projects in the tester.NoteCable IDs are case-sensitive. For example, the tester saves result with the names “A0” and “a0” in two different records.
Chapter 4: Manage Test ResultsUpload Results to a PC73Import: On the IMPORT RESULTS screen select the project that contains the results you want to import from the flash drive, then tap IMPORT.Delete: On the DELETE RESULTS screen select the project that contains the results you want to delete on the flash drive, then tap DELETE.Upload Results to a PCNoteTo upload results to the cloud for transfer to a PC, use the LinkWare Live web application. See Chapter 6.To upload results to a PC from the tester or a flash drive, use LinkWare PC software.1Install the latest version of LinkWare PC software on the PC.2Turn on the tester and start LinkWare PC on the PC.3Use the USB cable supplied to connect the Micro-AB USB port on the tester to a type A USB port on the PC. See Figure 27.Or connect a USB flash drive to the PC.4On the LinkWare PC toolbar, click  , then select a product, or select Test Files (.tst) to upload from a flash drive.5In the Import dialog box in LinkWare PC, select options for the location and the number of results to import.
DSX-602 CableAnalyzerUsers Manual74BK46.EPSFigure 27. How to Connect the Tester to a PCView the Memory StatusTo see the memory statusOn the home screen, tap the TOOLS icon, then tap Memory Status.The MEMORY STATUS screen shows these values:The percentage of memory availableThe number of test records that are savedThe number of .id files that have been downloaded to the tester from LinkWare PC softwareThe memory space taken by other files, such as the databases for projects and test limitsMicro-AB USB portType A USB portAC adapter
75Chapter 5: Use ProjectsWhy Use Projects?The tester’s ProjX™ management system lets you set up projects that help you monitor the status of a job and make sure that your work agrees with the requirements of the job.You can use a project to do these tasks: Specify the tests that are necessary for a job. Specify settings for tests.Specify an operator for the job.Make sets of sequential IDs to use as names for test results. Automatically save test results with IDs from a set.Add the results from other necessary tests to each saved result in the project. See which IDs do not have results for a specified test.See what percentage of a job is completed. See how many links passed and how many failed.Keep the test results from a job in one place for easy access. When you use a project, you can do tests and use IDs that are not specified in the project if necessary. You can also easily change the settings in a project if necessary.NoteTo manage projects in the cloud, use the LinkWare Live web application. See Chapter 6.
DSX-602 CableAnalyzerUsers Manual76Set Up a ProjectRefer to the PROJECT screen in Figure 28 on page 77.1On the home screen, tap the PROJECT panel, tap CHANGE PROJECT, then tap NEW PROJECT.2On the NEW PROJECT screen, enter a name for the project, then tap DONE.3On the PROJECT screen, tap the Operator panel to enter an operator name for the project.4On the PROJECT screen, tap the NEW TEST button to enter the tests and test settings necessary for the project.5On the PROJECT screen, tap the NEW ID SET button to make one or more sets of cable IDs for the project. See the CABLE ID SETUP screen in Figure 29 on page 79.6On the PROJECT screen, tap DONE. The PROJECT ScreenTo start a new project, tap the PROJECT panel on the screen. Figure 28 shows the PROJECT screen and describes the items you enter to make a project.The name of the project. See also item .Operator: The name of the person who will do the tests for the project. For each operator you can also enter the email address that the operator will use as an ID to sign in to LinkWare Live.The date range for the results in the project.Results: A summary of the test results in the project:: The number of tests that failed.: The number of tests that passed.
Chapter 5: Use ProjectsThe PROJECT Screen77GPU08.EPSFigure 28. PROJECT Screen: The number of DSX CableAnalyzer results that have PASS* results. PASS* results have measurements within the range of accuracy uncertainty for the tester. See “PASS*/FAIL* Results” on page 41.Test Setup: The tests that are available in the project.To add a test to the project, tap NEW TEST.Cable ID Sets: The sets of IDs the tester can use for the names of test results.To add a set of IDs to the project, tap NEW ID SET. See Figure 29.JABGEFHCDJI
DSX-602 CableAnalyzerUsers Manual78To use a different project, tap CHANGE PROJECT, then tap a project.To make a new project, tap CHANGE PROJECT, then tap NEW PROJECT.TRANSFER lets you export or import projects to or from a flash drive and delete projects on the flash drive. The project data includes all project settings and test results.MANAGE lets you rename, copy, or delete a project that is in the tester.To delete the test setup or ID set, tap  . To copy the test setup or ID set so you can edit it to make a new one, tap  .NotesIf you delete an imported ID set from a project, the ID set is still available in the tester. To delete imported ID sets from the tester, use LinkWare PC software. A project must have at least one Test Setup and one Cable ID set. If you delete them all, the tester makes a default Test Setup and Cable ID set.The CABLE ID SETUP ScreenTo see the CABLE ID SETUP screen, tap the PROJECT panel on the home screen, then tap NEW ID SET on the PROJECT screen. See Figure 29.Each project can have up to 5000 IDs. If an ID set does not have a Last ID, the tester counts the set as one ID. An ID can have a maximum of 60 characters. Symbols, such as the asterisk, and accented characters do not increment.
Chapter 5: Use ProjectsThe CABLE ID SETUP Screen79BA09.EPSFigure 29. CABLE ID SETUP Screen(after you enter the first and last IDs)First ID and Last ID: The first and last IDs in a set of sequential IDs. If you do not enter a Last ID when you make an ID set, the tester will increment the First ID to make subsequent IDs. NoteThe tester does not increment symbols or accented characters.When you use an ID set that does not have a Last ID, the set under IDs Untested on the CHANGE ID screen shows only the next ID. ABD EC
DSX-602 CableAnalyzerUsers Manual80Total IDs: The number of IDs in the set. This section does not show for ID sets that do not have a Last ID.Tap IMPORT to use an ID set that you downloaded to the tester from LinkWare PC software. Tap REVIEW to see the CABLE ID REVIEW screen, which shows the ID set an the total number of IDs. NoteThe REVIEW button does not show if you do not enter a Last ID.SAVE: To save the ID set, tap SAVE.About Next ID SetsIf you do not enter a Last ID when you make an ID set, the tester uses the First ID as the Next ID. The tester increments the Next ID each time you save a result. Numbers increment sequentially:1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, ... 99, 100, 101...Letters increment through the English alphabet:A, B, C, D, ... Z, AA, AB, AC, AD, ... AZ, BA, BB, BC... Numbers and letters do not cause each other to increment:1Y, 1Z, 1AA, 1AB, ... 1ZZ, 1AAA, 1AAB...The tester does not increment symbols or accented characters.When you use a Next ID set, the set under IDs Untested on the CHANGE ID screen shows only the next ID. To save the next test with a different ID, tap the Next ID: panel, then enter a different ID.Each project can have one Next ID set.If your project has only a Next ID set, the tester cannot calculate the percentage of the project that is completed, so the % Tested value does not show on the home screen.
Chapter 5: Use ProjectsManage Projects on a Flash Drive81If your project has both a Next ID set and sets with first and last IDs, the % Tested value includes tests you saved with Next ID. For example, if you have one Next ID set and one set with 10 IDs, and you save 10 results with next IDs, the % Tested shows 50% (10 saved results divided by 20 IDs). Manage Projects on a Flash DriveYou can export or import projects to or from a flash drive, and delete projects on the flash drive. The project data includes all project settings, test results, and ID sets.WCautionDo not remove the USB flash drive while the LED on the drive flashes. Doing so can corrupt the data on the drive.You can lose a USB flash drive, cause damage to it, or accidentally erase the contents of the drive. Thus, Fluke Networks recommends that you save no more than one day of test results on a flash drive. NoteThe tester reads only USB drives that use the FAT format.1Connect a USB flash drive to the type A USB port. The tester makes a bell sound when it detects the drive.2On the home screen, tap the PROJECT panel.3On the PROJECT screen, tap TRANSFER. 4On the TRANSFER PROJECTS screen, select a function:Export: On the EXPORT PROJECTS screen, select the projects you want to export to the flash drive, then tap EXPORT.-continued-
DSX-602 CableAnalyzerUsers Manual82Import: On the IMPORT PROJECTS screen select the projects you want to import from the flash drive, then tap IMPORT.Delete: On the DELETE PROJECTS screen select the projects you want to delete on the flash drive, then tap DELETE.Copy Project Settings to Other TestersTo copy the settings in a project to other DSX-602 testers, use the Read Project Setups and Write Project Setups utilities in LinkWare PC software. You can use LinkWare PC to read project settings from a tester or from a project you exported to a flash drive.
83Chapter 6: Sync Projects withLinkWare™ LiveThe LinkWare Live web application lets you manage your projects from a desktop or mobile device. Sign Up for a LinkWare Live Account1Go to www.linkwarelive.com/signin.2If you already have a LinkWare Live account, enter your email address and password on the LinkWare Live Sign In page.3If you do not have a LinkWare Live account, click New user? Sign up now!. Enter the information for your account, then click CREATE ACCOUNT.Fluke Networks sends you an email with a LinkWare Live activation code.4Open the email, copy the activation code, click the LinkWare Live activation link in the email, paste the activation code into the box in the activation window, then click ACTIVATE.5The LinkWare Live Sign In page shows again. Enter your email address and password, then click SIGN IN.
DSX-602 CableAnalyzerUsers Manual84How to See the Tester’s MAC AddressSome networks require users to register their device’s MAC address before they can connect to the network.There are two MAC addresses: one for the wired port and one for the Wi-Fi adapter.To see the tester’s or Wi-Fi adapter’s MAC address1On the home screen tap TOOLS, then tap Network.2On the NETWORK screen, tap the Wired or Wi-Fi panel. The MAC addresses shows at the top of the WIRED and WI-FI PORT screens.Use LinkWare Live Through a Wired Ethernet Network1Use an appropriate cable to connect the tester’s RJ45 Ethernet port to an active network port. If the cable is good and the port is active, the LEDs on the tester’s port will come on.2On the home screen, make sure the Operator name is correct. The tester uses the email address associated with the operator name as the ID when you sign in to LinkWare Live. If no address is associated with the name, or if you enter a different address on the LWL SIGN IN screen, the tester associates the name with the address you enter.3On the home screen, tap the SYNC icon.4When the tester connects to the network, the wired network connection icon shows at the top of the screen: 5On the LWL SIGN IN screen, if necessary, enter the ID and password for your LinkWare Live account, then tap SIGN IN.6If you use other people’s LinkWare Live accounts, the ORGANIZATION screen shows. Tap the organization you want to use.
Chapter 6: Sync Projects with LinkWare™ LiveUse LinkWare Live Through a Wi-Fi Ethernet Network857On the SYNC PROJECTS screen (Figure 30 on page 86), select the projects you want to sync, then tap SYNC.Use LinkWare Live Through a Wi-Fi Ethernet NetworkThe tester can use wireless network channels 1 through 11.To connect the tester to a Wi-Fi network1On the home screen, make sure the Operator name is correct. The tester uses the email address associated with the operator name as the ID when you sign in to LinkWare Live. If no address is associated with the name, or if you enter a different address on the LWL SIGN IN screen, the tester associates the name with the address you enter.2On the home screen, tap the SYNC icon.3Select a wireless network if necessary.If the network setting are not correct, the tester shows the NETWORK screen and a message about the necessary settings. a. To change the settings, tap the Wi-Fi panel.b. To test the settings, go back to the WI-FI PORT screen, then tap CONNECT.When the tester connects to the network, the Wi-Fi network connection icon shows at the top of the screen:  c. Go back to the TRANSFER RESULTS screen, then tap the LinkWare Live panel.4On the LWL SIGN IN screen, if necessary, enter the ID and password for your LinkWare Live account, then tap SIGN IN.5Enter a user name and password for the network, if necessary.-continued-
DSX-602 CableAnalyzerUsers Manual866If you use other people’s LinkWare Live accounts, the ORGANIZATION screen shows. Tap the organization you want to use.7On the SYNC PROJECTS screen (Figure 30), select the projects you want to sync, then tap SYNC.See Figure 30:Projects in the On this tester section are possibly also in LinkWare Live. You can upload results from these projects to LinkWare Live. If the project in LinkWare Live has test settings or cable IDs that are different from those in the tester, you can download the new settings to the tester.GPU190.EPSFigure 30. SYNC PROJECTS ScreenABC
Chapter 6: Sync Projects with LinkWare™ LiveUse LinkWare Live Through a Wi-Fi Ethernet Network87By default, the active project is selected to sync.Projects in the Available only on LinkWare Live section have project settings you can download to the tester. If you download these projects, only test settings and cable IDs are downloaded. Any test results saved in the project in LinkWare Live are not downloaded.If you select Assigned to me only, you will see only projects that are assigned to you in the project settings in LinkWare Live.
DSX-602 CableAnalyzerUsers Manual88Change the Network SettingsIt is not usually necessary to change the wired or Wi-Fi network settings before you try to make a connection. But for example if you must use static addressing, you can get to the settings on the NETWORK screen.To get to the network settingsOn the home screen, tap TOOLS, then tap Network, then tap the Wired or Wi-Fi panel.To test the settingsTap CONNECT on the WI-FI PORT or WIRED PORT screen.Settings for the Wired PortThe tester can use DHCP (dynamic host configuration protocol) or Static for the address method. Most networks use DHCP.For Static, enter an IP address for the tester and the Subnet Mask, Gateway address, and DNS1 and DNS2 addresses for the network. If you are not sure what to enter, speak to the network administrator. Settings for the Wi-Fi PortTable 5 shows the Wi-Fi settings.
Chapter 6: Sync Projects with LinkWare™ LiveChange the Network Settings89Table 5. Settings for the Wi-Fi Connection Setting DescriptionAddress Most networks use DHCP.DHCP address settingsSSID: The tester does a scan for wireless networks and shows a list of available networks. Select the correct SSID. To connect to a hidden network, tap ADD SSID.Security: Authentication: Select the authentication type that the network uses:Open: Security credentials are not necessary.WEP: Select an encryption method and enter the necessary keys.WPA/WPA2 Personal: Enter the password for the network.WPA/WPA2 Enterprise:EAP: Select an EAP type that is appropriate for the authentication server. User and Password: Enter a user name (login name) and password for the network.Alternate ID: The tester can use the Alternate ID with some EAP methods to send an empty or anonymous identity while the tester makes a private connection. The tester then uses the private connection to send the User name and Password you entered.The tester can also use the Alternate ID to send the User name and Password to an authentication server in a different realm. In this situation, the Alternate ID can have a format such as anonymous@MyCompany.com or /MyCompany/anonymous.
DSX-602 CableAnalyzerUsers Manual90Delete Wi-Fi Settings and PasswordsThe tester saves the security settings and passwords for the Wi-Fi connections you use.To delete all Wi-Fi settings and passwordsOn the home screen, tap TOOLS, then tap Network, tap the Wi-Fi panel, then tap FORGET ALL.About the Asset Management ServiceLinkWare Live’s asset management service lets you see the locations of your DSX-602 testers in the field. The owner of the LinkWare Live account can enable or disable the service remotely for each DSX-602 tester.When this service is enabled on a DSX-602 tester, the asset management icon ( ) shows on the tester’s home screen. When a technician uses the tester to sign in to LinkWare Live, the tester’s location shows on a map on LinkWare Live’s ASSETS page. NoteYou can enable or disable the asset management service only with LinkWare Live. There is no setting on the DSX-602 tester that enables or disables this service.Static address settingsEnter an IP address for the tester and the Subnet Mask, Gateway address, and DNS1 and DNS2 addresses for the network. If you are not sure what to enter, speak to the network administrator.The Security settings are the same as for the DHCP address settings.Table 5. Settings for the Wi-Fi Connection (continued)Setting Description
Chapter 6: Sync Projects with LinkWare™ LiveSign Your Tester Out of LinkWare Live91Sign Your Tester Out of LinkWare Live1On the home screen, tap the TOOLS icon, then tap Sign In.2On the LWL SIGN IN screen, tap SIGN OUT.Or, turn off the tester.Sign In to LinkWare Live from a Desktop or Mobile Device1Go to https://www.linkwarelive.com/signin.2Enter your LinkWare Live user name and password, then click SIGN IN.For more information about how to use LinkWare Live, click HELP on the LinkWare Live web page.Import Projects from LinkWare Live into LinkWare PC1Install the latest version of LinkWare PC software on the PC.2Turn on the tester and start LinkWare PC on the PC.3Use the USB cable supplied to connect the Micro-AB USB port on the tester to a type A USB port on the PC. 4On the LinkWare PC toolbar, click  .5Sign in to your LinkWare Live account, then use the LinkWare PC dialog boxes to select and import projects.Learn More About LinkWare LiveGo to http://www.flukenetworks.com/linkwarelive.
DSX-602 CableAnalyzerUsers Manual92
93Chapter 7: MaintenanceWWarningXTo prevent possible fire, electric shock, personal injury, or damage to the tester:Do not open the case. You cannot repair or replace parts in the case.Use only replacement parts that are approved by Fluke Networks.If you replace parts that are not specified as replacement parts, the warranty will not apply to the product and you can make the product dangerous to use. Use only service centers that are approved by Fluke Networks.WCautionIf you replace electrical parts yourself, the tester will possibly not have the correct calibration and can give incorrect test results. If the calibration is not correct, cable manufacturers can remove their warranty from the cabling you install.
DSX-602 CableAnalyzerUsers Manual94Verify OperationThe tester does a self test when you turn it on. If the tester shows an error or does not turn on, refer to “If the Tester Does Not Operate as Usual” on page 102.Clean the TesterTo clean the touchscreen, turn off the tester, then use a soft, lint-free cloth that is moist with water or water and a mild detergent. To clean the case, use a soft cloth that is moist with water or water and a mild detergent.WWarningXDo not put the tester or the battery pack in water.WCautionTo prevent damage to the touchscreen or the case, do not use solvents or abrasive materials.When you clean the touchscreen or the case, do not let liquid get under the plastic around the touchscreen.
Chapter 7: MaintenanceSee Information About the Tester95See Information About the TesterTo see information about your tester On the home screen, tap the TOOLS icon, then tap Version Information. To see information about your remote testerConnect the main and remote testers together (see Figure 6 on page 17), then tap REMOTE on the Version Information screen.Update the SoftwareNew software gives you access to new features and the latest test limits and cable types. Software updates are available on the Fluke Networks website. You can use a PC to install a software update, or connect an updated main unit to a remote or to another main unit to update those units.You can also use LinkWare Live to download a software update from the cloud to your main tester, then use the main tester and a USB cable to install the update in the remote.
DSX-602 CableAnalyzerUsers Manual96Use a PC to Update the SoftwareWCautionTo prevent unexpected loss of power, connect the ac adapter to the tester when you update the software.NoteThe software update procedure does not delete the test records, project settings, or user preferences in the tester, but can possibly change the factory-installed cable types or test limits.1Install the latest version of LinkWare PC software on your PC. LinkWare PC is available on the Fluke Networks website.2Connect the AC adapter to the tester and connect the Micro-AB USB port on the tester to a type A USB port on the PC. See Figure 31.3LinkWare PC automatically tells you if new software for the tester is available on the Fluke Networks website, and lets you install the software. NoteOlder versions of LinkWare PC do not start the update procedure automatically. For older versions, you must have the software update file on a disk or USB drive and click   to start the update procedure.4The tester reboots when the update is completed. To make sure the update was installed correctly, tap the TOOLS icon on the home screen, tap Version Information, then make sure the correct version shows.5Do steps 2 through 4 again for the remote. On a remote tester, the LEDs show the progress of the installation.
Chapter 7: MaintenanceUpdate the Software97BK46.EPSFigure 31. How to Connect the Tester to a PCUse an Updated Main Tester to Update Other Testers1Turn on both testers and connect the AC adapters to both testers.2Use the USB cable provided to connect the updated main unit to the remote or to another main. See Figure 32.3Follow the instructions shown on the display of the updated main unit.NoteIf a remote has newer software than the main unit, the main unit can install the older software in the remote so that you can use the two units together. The remote cannot install the newer software in the main unit.Micro-AB USB portType A USB portAC adapter
DSX-602 CableAnalyzerUsers Manual98BK116.EPSFigure 32. How to Connect Units Together to Update the SoftwareType A USB port Micro-AB USB portType A USB port Micro-AB USB portUpdated main unitUpdated main unitAC adaptersAC adaptersRemoteMain unit
Chapter 7: MaintenanceUpdate the Software99Use LinkWare Live to Update the Software1Use an appropriate cable to connect the tester’s RJ45 Ethernet port to a network port, or connect a Wi-Fi adapter to the tester’s type A USB port.2On the home screen, tap the SYNC icon.3For a wireless network, if the network selection screen shows, connect to a network:a. Select a network.b. Enter security information.c. Go back to the previous screen, then tap CONNECT. d. Go back to the home screen, then tap SYNC.4On the LWL SIGN IN screen, enter the ID and password for your LinkWare Live account, then tap SIGN IN.5If a software update is available, the tester shows a message. Tap YES to install new software in your tester.6The tester reboots when the update is completed. 7To install the update in a remote or another main tester, use the USB cable provided to connect the updated main tester to the other tester. See Figure 32.8Follow the instructions shown on the display of the updated main tester.
DSX-602 CableAnalyzerUsers Manual100Extend the Life of the BatteryWWarningXTo prevent possible fire, electric shock, or personal injury, read the warnings about the rechargeable battery under “WarningX” on page 4.Do not frequently let the battery discharge completely. Do not keep the battery at temperatures below -20 oC (-4 oF) or above +50 oC (+122 oF) for periods longer than one week.Before you put a battery into storage, charge it to approximately 50 % of full charge.Store the TesterBefore you store a tester or an extra battery for a long period, charge the battery to approximately 50 % of full charge. The discharge rate of the battery is 5 % to 10 % each month. Check the battery every 4 months and charge it if necessary.Keep a battery attached to the tester during storage. If you remove the battery for more than approximately 24 hours, the tester will not keep the correct time and date. Storage temperature: -22 °F to +140 °F (-30 °C to +60 °C)Remove the BatteryFigure 33 shows how to remove the battery.NotesIf you remove the battery and do not connect the AC adapter, the clock keeps the current date and time for a minimum of 24 hours.The screw does not come out of the battery door.
Chapter 7: MaintenanceTraceable Calibration Period101BA21.EPSFigure 33. How to Remove the BatteryTraceable Calibration PeriodTo make sure that the testers operate within the published specifications for accuracy, have them calibrated at a Fluke Networks authorized service center every 12 months. To get information on factory calibration, contact an authorized Fluke Networks Service Center.To see when the tester last received a factory calibration, tap the TOOLS icon on the home screen, then tap Version Information.Options and AccessoriesFor a complete list of options and accessories go to the Fluke Networks website at www.flukenetworks.com.To order options and accessories, contact an authorized Fluke Networks distributor.ABCModel VERSIV-BATTERYOFF
DSX-602 CableAnalyzerUsers Manual102If the Tester Does Not Operate as UsualIf the tester does not operate as usual or if it shows an unusual message, see Table 6 for possible explanations and solutions to some conditions that can occur.If the condition continues, contact Fluke Networks for assistance, or search the Fluke Networks Knowledge Base for a solution. If you contact Fluke Networks, have available the serial number, software and hardware versions, and calibration date for the tester, if possible. To see this information, tap the TOOLS icon on the home screen, then tap Version Information.You can also use LinkWare PC software to upload the system log from the tester. This file contains information that can possibly help Fluke Networks find a solution to an unusual problem.The serial numbers of the main and remote units are on a sticker on the back of the units.Table 6. Possible Solutions for Unusual Behavior  The touchscreen or the keys do not respond.Press and hold  until the tester turns off. Then turn on the tester. If the problem continues, install the latest version of software in the tester.The tester will not turn on even though the battery is charged.The safety switch or the fuse in the battery is possibly open. Wait a few seconds, then try again. If the tester does not turn on, the battery is possibly bad and must be replaced. The tester can operate on AC power if the battery does not operate. Test results appear to be incorrect.The tester possibly has incorrect settings. Make sure you selected the correct cable type and test limit.