Fluke 114, 115, 116 and 117
Digital Multimeters
Extended specifications

Technical Data
General specifications (all models)
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity of 0 % to 90 %.
Maximum voltage between any
terminal and earth ground

600 V

Surge protection

6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2

W Fuse for A input

11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293)

Display

Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec

Temperature

Operating: -10 °C to +50 °C; Storage: -40 °C to +60 °C

Humidity

0 % to 90 % to 35 °C; 75 % to 40 °C; 45 % to 50 °C

Temperature coefficient

0.1 x (specified accuracy/°C) (< 18 °C or > 28 °C)

Operating altitude

2,000 meters

Battery

9 Volt Alkaline, NEDA 1604A/IEC 6LR61

Battery life

Alkaline: 400 hours typical, without backlight

Safety compliances

ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN
61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1

Certifications

UL, CSA, TUV, N10140    , VDE

IP rating (dust and water protection)

IP42

N10140

Accuracy specifications (all models)
Accuracy ± ([% of Reading]
+ [Counts])

Model

0.1 mV

0.5 % + 2

114, 115, 116, 117

0.001 V
0.01 V
0.1 V

0.5 % + 2

114, 115, 116, 117

Function

Range

Resolution

DC millivolts

600.0 mV

DC volts

6.000 V
60.00 V
600.0 V

DC, 45 Hz to
500 Hz
Auto-V LoZ true-rms
1

600.0 V

0.1 V

500 Hz to 1 kHz

2.0 % + 3

4.0 % + 3

45 Hz to 500 Hz

500 Hz to 1 kHz

114, 116, 117

AC millivolts1 true-rms

600.0 mV

0.1 mV

1.0 % + 3

2.0 % + 3

114, 115, 116, 117

AC volts1 true-rms

6.000 V
60.00 V
600.0 V

0.001 V
0.01 V
0.1 V

1.0 % + 3

2.0 % + 3

114, 115, 116, 117

Continuity

600 Ω

1Ω

Ohms

600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
40.00 MΩ

Diode test
Capacitance

LoZ capacitance (power-up option)

1 nF to 500 µF

Beeper on < 20 Ω, off > 250 Ω;
detects opens or shorts of
500 µs or longer

114, 115, 116, 117

0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ

0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 2

114, 115, 116, 117

2.000 V

0.001 V

0.9 % + 2

115, 116, 117

1000 nF
10.00 µF
100.0 µF
9999 µF

1 nF
0.01 µF
0.1 µF
1 µF

1.9 % + 2
1.9 % + 2
1.9 % + 2
100 µF to 1000 µF: 1.9 % + 2
> 1000 µF: 5 % + 20

115, 116, 117

10 % + 2 typical

115, 116, 117

	 All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of
range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from
a circuit or are shorted together. For volts, crest factor of ≤ 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of ≤ 3.
AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.

1

Accuracy specifications (all models) cont.
Accuracy ± ([% of Reading]
+ [Counts])

Model

0.1 °C
0.2 °F

1 % + 102
1 % + 182

116

6.000 A
10.00 A
20 A overload for
30 seconds maximum

0.001 A
0.01 A

1.5 % + 3

115, 117

AC µAmps true-rms1 (45 Hz to 1 kHz)

600.0 µA

0.1 µA

1.5 % + 3 (2.5 % + 3 > 500 Hz)

116

DC amps

6.000 A
10.00 A
20 A overload for
30 seconds maximum

0.001 A
0.01 A

1.0 % + 3

115, 117

DC µAmps true-rms

600.0 µA

0.1 µA

1.0 % + 2

116

Hz (V or A input)2

99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz

0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz

0.1 % + 2

115, 117

99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz

0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz

0.1 % + 2

116

Function

Range

Resolution

Temperature (K-Type thermocouple)

-40 °C to 400 °C
-40 °F to 752 °F

AC amps true-rms1 (45 Hz to 500 Hz)

Hz (V input)3

	 All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range
are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit
or are shorted together. For volts, crest factor of ≤ 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of ≤ 3. AC volts is
ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.
2
	 AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC Amps Hz is dc-coupled and specified from 45 Hz to 5 kHz. Amps input burden
voltage (typical): 6 A input 2 mV/A, 10 A input 37 mV/A.
3
	 Frequency is ac-coupled, 45 Hz to 50 kHz.
1

Frequency counter sensitivity (models 115, 116, 117)
Typical sensitivity (rms sine wave)
Input range
Volts AC

AC Amps
(115, 117 only)

5 Hz to 45 Hz

45 Hz to 5 kHz

5 kHz to 10 kHz

10 kHz to 50 kHz

6V
60 V
600 V

0.2 V
2V
20 V

0.2 V to 0.3 V
2 V to 3 V
20 V to 30 V

0.3 V to 0.4 V
3 V to 4 V
30 V to 40 V

0.4 V to 1.0 V
4 V to 10 V
40 V to 100 V

6A
10 A

N/A
N/A

0.4 A
0.5 A

N/A
N/A

N/A
N/A

Input characteristics (all models)
Function

Input impedance
(nominal)

Common mode rejection ratio(1 kΩ unbalanced)

Volts AC

> 5 MΩ < 100 pF

> 60 dB at dc, 50 or 60 Hz

Volts DC

> 10 MΩ < 100 pF

> 100 dB at dc, 50 or 60 Hz

~3 kΩ < 500 pF

> 60 dB at dc, 50 or 60 Hz

Auto-V LoZ

Open circuit test voltage
Ohms

< 2.7 V dc

Diode Test

< 2.7 V dc

Full scale voltage

Normal mode rejection

> 60 dB at 50 or 60 Hz

Short circuit current

To 6.0 MΩ

40 MΩ

< 0.7 V dc

< 0.9 V dc
2.000 V dc

Fluke.	Keeping your world
	
up and running.™
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands

< 350 μA
< 1.2 mA

For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006 Fluke Corporation. All rights reserved.
Printed in U.S.A. 11/2006 2793260 D-EN-N Rev A

 Fluke Corporation

Fluke 11X Series Digital Multimeters Extended Specifications



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
XMP Toolkit                     : 3.1.1-111
About                           : d2fb1c22-880f-11da-9b03-9ca4bb25846e
Create Date                     : 2006:12:01 09:22:23-08:00
Metadata Date                   : 2006:12:01 09:22:25-08:00
Modify Date                     : 2006:12:01 09:22:25-08:00
Creator Tool                    : Adobe InDesign CS2 (4.0.4)
Thumbnail Format                : JPEG
Thumbnail Width                 : 256
Thumbnail Height                : 256
Thumbnail Image                 : (Binary data 10965 bytes, use -b option to extract)
Instance ID                     : uuid:82f2012a-8160-11db-bef4-000a95b26bba
Document ID                     : adobe:docid:indd:1b431827-6585-11db-9e2f-f0739d794614
Rendition Class                 : proof:pdf
Derived From Instance ID        : f7f27a5e-6538-11db-b823-9e3c16ba1a48
Derived From Document ID        : adobe:docid:indd:f7f27a5d-6538-11db-b823-9e3c16ba1a48
Format                          : application/pdf
Producer                        : Adobe PDF Library 7.0
Trapped                         : False
Page Count                      : 2
Creator                         : Adobe InDesign CS2 (4.0.4)
EXIF Metadata provided by EXIF.tools

Navigation menu