Fluke 115 Data Sheet

2015-09-09

: Fluke Fluke-115-Data-Sheet-805268 fluke-115-data-sheet-805268 fluke pdf

Open the PDF directly: View PDF PDF.
Page Count: 2

Fluke 114, 115, 116 and 117
Digital Multimeters
Extended specications
Technical Data
General specications (all models)
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity of 0 % to 90 %.
Maximum voltage between any
terminal and earth ground
600 V
Surge protection 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2
W Fuse for A input 11 A, 1000 V FAST 17 kA Fuse (Fluke PN 803293)
Display Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec
Temperature Operating: -10 °C to +50 °C; Storage: -40 °C to +60 °C
Humidity 0 % to 90 % to 35 °C; 75 % to 40 °C; 45 % to 50 °C
Temperature coefficient 0.1 x (specified accuracy/°C) (< 18 °C or > 28 °C)
Operating altitude 2,000 meters
Battery 9 Volt Alkaline, NEDA 1604A/IEC 6LR61
Battery life Alkaline: 400 hours typical, without backlight
Safety compliances ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN
61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1
Certifications UL, CSA, TUV, N10140 , VDE
IP rating (dust and water protection) IP42
N10140
Accuracy specications (all models)
Function Range Resolution
Accuracy ± ([% of Reading]
+ [Counts]) Model
DC millivolts 600.0 mV 0.1 mV 0.5 % + 2 114, 115, 116, 117
DC volts 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V 0.5 % + 2 114, 115, 116, 117
DC, 45 Hz to
500 Hz 500 Hz to 1 kHz
Auto-V LoZ1 true-rms 600.0 V 0.1 V 2.0 % + 3 4.0 % + 3 114, 116, 117
45 Hz to 500 Hz 500 Hz to 1 kHz
AC millivolts1 true-rms 600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3 114, 115, 116, 117
AC volts1 true-rms 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V 1.0 % + 3 2.0 % + 3 114, 115, 116, 117
Continuity 600 W1 WBeeper on < 20 W, off > 250 W;
detects opens or shorts of
500 µs or longer
114, 115, 116, 117
Ohms 600.0 W
6.000 kW
60.00 kW
600.0 kW
6.000 MW
40.00 MW
0.1 W
0.001 kW
0.01 kW
0.1 kW
0.001 MW
0.01 MW
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 2
114, 115, 116, 117
Diode test 2.000 V 0.001 V 0.9 % + 2 115, 116, 117
Capacitance 1000 nF
10.00 µF
100.0 µF
9999 µF
1 nF
0.01 µF
0.1 µF
1 µF
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 µF to 1000 µF: 1.9 % + 2
> 1000 µF: 5 % + 20
115, 116, 117
LoZ capacitance (power-up option) 1 nF to 500 µF 10 % + 2 typical 115, 116, 117
1 All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of
range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from
a circuit or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3.
AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.
2 Fluke Corporation Fluke 11X Series Digital Multimeters Extended Specifications
Function Range Resolution
Accuracy ± ([% of Reading]
+ [Counts]) Model
Temperature (K-Type thermocouple) -40 °C to 400 °C
-40 °F to 752 °F 0.1 °C
0.2 °F 1 % + 102
1 % + 182116
AC amps true-rms1 (45 Hz to 500 Hz) 6.000 A
10.00 A
20 A overload for
30 seconds maximum
0.001 A
0.01 A 1.5 % + 3 115, 117
AC µAmps true-rms1 (45 Hz to 1 kHz) 600.0 µA 0.1 µA 1.5 % + 3 (2.5 % + 3 > 500 Hz) 116
DC amps 6.000 A
10.00 A
20 A overload for
30 seconds maximum
0.001 A
0.01 A 1.0 % + 3 115, 117
DC µAmps true-rms 600.0 µA 0.1 µA 1.0 % + 2 116
Hz (V or A input)299.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 % + 2 115, 117
Hz (V input)399.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 % + 2 116
1 All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range
are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit
or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3. AC volts is
ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.
2 AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC Amps Hz is dc-coupled and specified from 45 Hz to 5 kHz. Amps input burden
voltage (typical): 6 A input 2 mV/A, 10 A input 37 mV/A.
3 Frequency is ac-coupled, 45 Hz to 50 kHz.
Accuracy specications (all models) cont.
Frequency counter sensitivity (models 115, 116, 117)
Typical sensitivity (rms sine wave)
Input range 5 Hz to 45 Hz 45 Hz to 5 kHz 5 kHz to 10 kHz 10 kHz to 50 kHz
Volts AC 6 V
60 V
600 V
0.2 V
2 V
20 V
0.2 V to 0.3 V
2 V to 3 V
20 V to 30 V
0.3 V to 0.4 V
3 V to 4 V
30 V to 40 V
0.4 V to 1.0 V
4 V to 10 V
40 V to 100 V
AC Amps
(115, 117 only) 6 A
10 A N/A
N/A 0.4 A
0.5 A N/A
N/A N/A
N/A
Input characteristics (all models)
Function
Input impedance
(nominal) Common mode rejection ratio(1 kW unbalanced) Normal mode rejection
Volts AC > 5 MW < 100 pF > 60 dB at dc, 50 or 60 Hz
Volts DC > 10 MW < 100 pF > 100 dB at dc, 50 or 60 Hz > 60 dB at 50 or 60 Hz
Auto-V LoZ ~3 kW < 500 pF > 60 dB at dc, 50 or 60 Hz
Open circuit test voltage Full scale voltage Short circuit current
Ohms < 2.7 V dc To 6.0 MW40 MW
< 350 µA
< 0.7 V dc < 0.9 V dc
Diode Test < 2.7 V dc 2.000 V dc < 1.2 mA
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
Fluke. Keeping your world
up and running.
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006 Fluke Corporation. All rights reserved.
Printed in U.S.A. 11/2006 2793260 D-EN-N Rev A