1523, 1524
Thermometer Readout
User’s Guide

Rev. 891001_EN

Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA
Phone: +1.801.763.1600 • Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
www.hartscientific.com
Specifications subject to change without notice. • Copyright © 2008 • Printed in USA

Limited Warranty & Limitation of Liability
Each product from Fluke Corporation, Hart Scientific Division (“Hart”) is warranted to be free from defects in material
and workmanship under normal use and service. The warranty period is one year(s) for the Reference Thermometer. The
warranty period begins on the date of the shipment. Parts, product repairs, and services are warranted for 90 days. The
warranty extends only to the original buyer or end-user customer of a Hart authorized reseller, and does not apply to fuses,
disposable batteries or to any other product, which in Hart’s opinion, has been misused, altered, neglected, or damaged
by accident or abnormal conditions of operation or handling. Hart warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Hart
does not warrant that software will be error free or operate without interruption. Hart does not warrant calibrations on the
Reference Thermometer.
Hart authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no
authority to extend a greater or different warranty on behalf of Hart. Warranty support is available if product is purchased
through a Hart authorized sales outlet or Buyer has paid the applicable international price. Hart reserves the right to
invoice Buyer for importation costs of repairs/replacement parts when product purchased in one country is submitted for
repair in another country.
Hart’s warranty obligation is limited, at Hart’s option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Hart authorized service center within the warranty period.
To obtain warranty service, contact your nearest Hart authorized service center or send the product, with a description
of the difficulty, postage, and insurance prepaid (FOB Destination), to the nearest Hart authorized service center. Hart
assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation
prepaid (FOB Destination). If Hart determines that the failure was caused by misuse, alteration, accident or abnormal condition or operation or handling, Hart will provide an estimate or repair costs and obtain authorization before commencing
the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed
for the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. HART SHALL NOT BE LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL. OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF

DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE
OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of
incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any
provision of this Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.

Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA
Phone: +1.801.763.1600 • Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
www.hartscientific.com
Specifications subject to change without notice. •Copyright © 2008 • Printed in USA




Table of Contents
1	

Before You Start........................................................................1
1.1	 Introduction................................................................................................ 1
1.2	 Standard Equipment.................................................................................. 1
1.3	 Safety Information...................................................................................... 2
1.3.1	
Warning......................................................................................................2
1.3.2	 Cautions...........................................................................................................3

1.4	 CE Comments............................................................................................ 5
1.4.1	 EMC Directive..................................................................................................5
1.4.2	 Immunity Testing..............................................................................................5

1.5	
1.6	
1.7	
1.8	

2	

Using Clamp-On Ferrites........................................................................... 5
Emissions Testing...................................................................................... 6
Low Voltage Directive (Safety)................................................................... 6
Authorized Service Centers....................................................................... 7

Quick Start.................................................................................9
2.1	 Setup.......................................................................................................... 9
2.2	 Specifications.......................................................................................... 24

i

1523, 1524 Thermometer Readout


Figures
Figure 1 Clamp-On Ferrite.................................................................................. 6
Figure 2 Input/Output Connections - 1523......................................................... 9
Figure 3 Input/Output Connections - 1524....................................................... 10
Figure 4 Keys.................................................................................................... 11
Figure 5 1523 Menu.......................................................................................... 14
Figure 6 1523 Menu (cont).............................................................................. 15
Figure 7 1523 Menu (cont).............................................................................. 16
Figure 8 1524 Menu......................................................................................... 19
Figure 9 1524 Menu (cont).............................................................................. 20
Figure 10 1524 Menu (cont)............................................................................ 21
Figure 11 1524 Menu (cont)............................................................................ 22
Figure 12 1524 Menu (cont)............................................................................ 23
Figure 13 1524 Menu (cont)............................................................................ 24

ii




Tables
Table 1 International Symbols............................................................................. 4
Table 2 1523 Input/Output Connections............................................................. 9
Table 3 1524 Input/Output Connections........................................................... 10
Table 4 1523 Key Functions.............................................................................. 12
Table 5 1524 Key Functions.............................................................................. 17
Table 6 General Specifications......................................................................... 24
Table 7 Millivolt Measurement........................................................................... 25
Table 8 Ohms Measurement, RTDs.................................................................. 25
Table 9 Ohms Measurement, Thermistor.......................................................... 26
Table 10 Temperature, Thermocouples............................................................ 26
Table 11 Temperature, RTD Ranges, and Accuracies (ITS-90)........................ 28
Table 12 Temperature, Thermistor.................................................................... 28

iii

Before You Start
Standard Equipment

1	 Before You Start
1.1	 Introduction
The Reference Thermometer Readouts (1523, 1524) are designed to be reliable, stable, temperature measuring instruments that can be used in the field or laboratory. They offer accuracy, portability, and speed for nearly every field calibration application. The instruments have been designed with the field user in mind and are
easy to use while maintaining stability, uniformity, and accuracy comparable to some laboratory instruments.
Your Fluke 1523 and 1524 thermometer readout is a handheld, battery operated instrument that measures temperature using Platinum resistance Thermometers (PRT), Thermistors, and Thermocouples (TC).

1.2	 Standard Equipment
Unpack the instrument carefully and inspect it for any damage that may have occurred during shipment. If
there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
●●
●●
●●
●●
●●
●●
●●

1523/1524 Reference Thermometer Readout with 3 AA batteries
AC Adapter, with power cord
RS-232 Cable
User’s Guide
Documentation CD
Report of Calibration and calibration label
Clamp-on ferrite(s)

If all items are not present, contact an Authorized Service Center. (See Section 1.8, Authorized Service Centers on page 7.)
1

1523, 1524 Thermometer Readout
Safety Information

1.3	 Safety Information
The Reference Thermometer is designed in accordance with EN 61010-1 {2nd Edition}, and CAN/CSA 22.2
No 61010.1-04. Use this instrument only as specified in this manual, otherwise the protection provided by the
instrument may be impaired.
A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions
and actions that may damage the instrument being used.
International symbols used on the reference thermometer and in this manual are explained in Table 1 on page
4.
1.3.1	

Warning
To avoid possible electric shock or personal injury:
●●
●●

●●

●●

●●
●●
●●
●●

2

Do not use the reference thermometer in environments other than those listed in the user’s guide.
Do not use the reference thermometer for any application other than that which is specified. The
instrument was designed for temperature measurement and calibration. Any other use of the instrument
may cause unknown hazards to the user.
If the reference thermometer is used in a manner not in accordance with the equipment design, the
operation and the protection provided by the instrument may be impaired. In addition, safety hazards
may arise.
Do not apply more than the rated voltage, as marked on the reference thermometer, between the inputs,
or between any input and earth ground (30 V, 24 mA max all terminals).
Follow all equipment safety procedures.
Calibration equipment should only be used by trained personnel.
The reference thermometer is intended for indoor use only.
Before you use the instrument, inspect the case. Look for cracks or missing plastic. Pay particular
attention to the insulation surrounding the connectors. Do not use the reference thermometer if it

Before You Start
Safety Information

●●
●●
●●
●●
●●

appears damaged or operates abnormally. Protection may be impaired. When in doubt, have the
instrument serviced.
Always use an isolated RTD or PRT (metal sheath isolated from lead wires).
Make sure the battery door is closed and latched before you operate the reference thermometer.
Do not operate the reference thermometer around explosive gas, vapor, or dust.
For battery operation use only 3 AA batteries, properly installed in the reference thermometer case.
1524 model, thermocouples can only be used on channel 1.

1.3.2	 Cautions
To avoid possible damage to the reference thermometer or to equipment under test:
●●

●●

●●

●●
●●
●●

Do not apply more than the rated voltage, as marked on the reference thermometer, between the inputs,
or between any input and earth ground (30 V 24 mA max all terminals).
Unless recalibrating the instrument DO NOT change the values of the calibration constants from
the factory set values. The correct setting of these parameters is important to the safety and proper
operation of the instrument.
The instrument and any thermometer probes used with it are sensitive instruments that can be easily
damaged. Always handle these devices with care. DO NOT allow them to be dropped, struck, stressed,
or overheated.
DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty environment.
Use the proper probes, function and range for your measurement.
Ensure probe coefficients are downloaded.

3

1523, 1524 Thermometer Readout
Safety Information
Table 1 International Symbols
Symbol

4

Description

Symbol

Description

AC (Alternating Current)

PE Ground

AC-DC

Hot Surface (Burn Hazard)

Battery

Read the User’s Guide (Important Information)

Complies with European Union directives

Off

DC

On

Double Insulated

Canadian Standards Association

Electric Shock

C-TICK Australian EMC mark

Fuse

The European Waste Electrical and Electronic
Equipment (WEEE) Directive (2002/96/EC) mark.

Before You Start
Using Clamp-On Ferrites

1.4	 CE Comments
1.4.1	 EMC Directive
Hart Scientific’s equipment has been tested to meet the European Electromagnetic Compatibility Directive
(EMC Directive, 2004/108/EC ). The Declaration of Conformity for your instrument lists the specific standards to which the unit was tested.
The instrument was designed specifically as a test and measuring device. Compliance to the EMC directive is through EN 61326-1:2006 Electrical equipment for measurement, control and laboratory use – EMC
requirements
As noted in the EN 61326-1, the instrument can have varying configurations. The instrument was tested in a
typical configuration with shielded RS-232 cables.
1.4.2	 Immunity Testing
The instrument was tested to the requirements for laboratory locations.

1.5	 Using Clamp-On Ferrites
Clamp-on ferrites are provided for use in improving the instrument’s electromagnetic (EM) immunity in environments of excessive EM interference, like areas of heavy industrial equipment. We recommend placing the
ferrites on the cables of probes attached to the instrument.
To attach a ferrite to a probe cable, make a loop in the cable near the connector and clamp the ferrite around
half of the loop as shown in the diagram. The ferrite can be easily detached and moved to a new probe when
needed. (See Figure 1 on next page.)

5

1523, 1524 Thermometer Readout
Emissions Testing

probe cable
clamp-on ferrite
Figure 1 Clamp-On Ferrite

1.6	 Emissions Testing
The instrument fulfills the limit requirements for Class B.

1.7	 Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (2006/95/EC), Fluke equipment has been designed to meet the EN 61010-1.

6

Before You Start
Authorized Service Centers

1.8	 Authorized Service Centers
Please contact one of the following authorized Service Center to coordinate service on your Fluke product:
Fluke Corporation
Hart Scientific Division
Phone: +1.801.763.1600
Fluke Nederland B.V.
Phone: +31-402-675300
Fluke Int'l Corporation - CHINA
Phone: +86-10-6-512-3436
Fluke South East Asia Pte Ltd. - SINGAPORE
Phone: +65-6799-5588
When contacting a Service Center for support, please have the following information available:
●●
●●
●●

Model Number
Serial Number
Complete description of the problem

7

Quick Start
Setup

2	 Quick Start
2.1	 Setup

Table 2 1523 Input/Output Connections

RS 232

1

2

30 V MAX

12 V DC

No.

Name

Description

1

Serial

Serial interface connector

2

Connector, T1

Sensor Connector, Channel 1

4

Power

External Power adapter connection

4

Figure 2 Input/Output Connections - 1523

1523 CALIBRATION
THERMOMETER
READOUT

9

1523, 1524 Thermometer Readout
Setup

Table 3 1524 Input/Output Connections

1
RS 232

2 30 V MAX 3
T1
T2

4
12 V DC

Figure 3 Input/Output Connections - 1524

1524 CALIBRATION
THERMOMETER
READOUT

10

No

Name

Description

1

Serial

Serial interface connector

2

Connector, T1

Sensor Connector, Channel 1

3

Connector, T2

Sensor Connector, Channel 2

4

Power

External Power adapter connection

RESET

RECALL

SAVE

LOG

STATS

11

8

4

1

RS 232

30 V MAX

mV

SETUP

NEXT

ENTER

HOME

10

6

12

HOLD

3

12 V DC

TREND

READOUT

°C °F

2

9

5
7

T2

1524 CALIBRATION
THERMOMETER

T1

Quick Start
Setup

Figure 4 Keys

11

1523, 1524 Thermometer Readout
Setup
Table 4 1523 Key Functions
No

Key

Description

1

Power on or off

2

Yellow Second or Special Function Key

3

Turns the backlight on or off

4

STATS

5

°C °F

Units, °C/°F

6

HOLD

1st press - Holds value on screen "-- HOLD --" across bottom of screen. 2nd press - Releases Screen hold.

7

SETUP

Enters setup menu, see menu structure

8

SAVE

12

Saves measurement as a logged data point

Arrows increment or decrement selections in an active field. In Graph Mode, Arrows change the scale of the graph.

9

10

1st Press: MAX, 2nd Press: MIn, 3rd Press: AVE, 4th Press: STD DEV

ENTER

Selects highlighted selection, Saves a new selection.

Quick Start
Setup
No

Key

11

RECALL

12

NEXT

13

+

14

15

16

Description

Moves down to next option on screen.

STATS

+

+

+

1st Press - Enters Recall menu, 2nd Press - Exits Recall Menu

“RESET” - Resets Stats Data

°C °F

“Ω mV” - Toggles from °C to Ω or Ω to °C (PRT, thermistor), °C to mV or mV to °C (TC)

HOLD

“TREND” - Starts Graphing data

ENTER

“HOME” Returns user to main screen

13

1523, 1524 Thermometer Readout
Setup

Figure 5 1523 Menu

14

Quick Start
Setup

Figure 6 1523 Menu (cont)

15

1523, 1524 Thermometer Readout
Setup

Figure 7 1523 Menu (cont)

16

Quick Start
Setup
Table 5 1524 Key Functions
No

Key

Description

1

Power on or off

2

Yellow Second or Special Function Key

3

Turns the backlight on or off

4

STATS

5

°C °F

Units, °C/°F

6

HOLD

1st press - Holds value on screen "-- HOLD --" across bottom of screen. 2nd press - Releases Screen hold.

7

SETUP

Enters setup menu, see menu structure

8

SAVE

9

1st Press: Max, 2nd Press: Min, 3rd Press: Ave, 4th Press: STD DEV

Saves measurement as a logged data point
Arrows increment or decrement selections in an active field. In Graph Mode, Arrows change the scale of the
graph.

17

1523, 1524 Thermometer Readout
Setup
No

Key

10

ENTER

Selects highlighted selection, Saves a new selection.

11

RECALL

1st press - Enters Recall Menu, 2nd press - Exits Recall Menu

12

NEXT

13

+

14

15

Description

Moves down to next option on screen.

STATS

+

+

16

“RESET” - Resets Stats Data

°C °F

“Ω mV” - Toggles from °C to Ω or Ω to °C (PRT, thermistor), °C to mV or mV to °C (TC)

HOLD

“TREND” - Starts Graphing data

SAVE

“LOG” - Log a series of measurements, see Auto Log in menu structure

+
17

18

+

ENTER

“HOME” Returns user to main screen

Quick Start
Setup

Figure 8 1524 Menu

19

1523, 1524 Thermometer Readout
Setup

Figure 9 1524 Menu (cont)

20

Quick Start
Setup

Figure 10 1524 Menu (cont)

21

1523, 1524 Thermometer Readout
Setup

Figure 11 1524 Menu (cont)

22

Quick Start
Setup

Figure 12 1524 Menu (cont)

23

1523, 1524 Thermometer Readout
Specifications

Figure 13 1524 Menu (cont)

2.2	 Specifications
Specifications are based on a one year calibration cycle and apply from 13 °C to 33 °C unless stated
otherwise. All specifications assume a five minute warm up period.
Table 6 General Specifications

24

Operating
Temperature†

–10 °C to 60 °C

Storage Temperature

–20 °C to 70 °C

Operating altitude

10,000 meters above mean sea level

Relative Humidity (%
RH operating without
condensation)

0 % to 90 % (non condensing)

Vibration

Random, 2g, 5–500 Hz

Quick Start
Specifications
Power requirements

3 AA alkaline batteries
12 V dc universal power supply

Size

96 x 200 x 47 mm
(3.75 x 7.9 x 1.86 inches)

Weight

0.65 kg (1.4 lb)

Safety

EN 61010-1:2001, CAN/CSA C22.2 No.
61010.1-04

Environmental conditions for all specifications: 13 °C to 33 °C

†

Table 7 Millivolt Measurement
Range

Resolution

Accuracy

–10 mV to 75 mV

0.001 mV

± (0.005 % + 5 µV)

Temperature Coefficient ( –10 °C to 13 °C, +33 °C to 60 °C):
± (0.001 %/°C + 1 µV/°C)
Table 8 Ohms Measurement, RTDs
Ohms Range

Accuracy ± Ω
4 Wire

0 W to 400 W

± (0.004 % + 0.002 Ω)

Temperature Coefficient ( –10 °C to 13 °C, +33 °C to 60 °C):
0.0008 %/°C + 0.0004 Ω
Excitation Current: 1 mA

25

1523, 1524 Thermometer Readout
Specifications
Table 9 Ohms Measurement, Thermistor
Ohms Range

Accuracy ± Ω, 4 Wire

200 W to 50 kW

± (0.01 % + 0.5 Ω)

50 kW to 500 kW

± (0.03 %)

Temperature Coefficient ( –10 °C to 13 °C , +33 °C to 60 °C):
0.002 %/°C + 0.1 Ω (0 W to 50 kW)
0.06 %/°C + 0.1 Ω (50 kW to 500 kW)
Excitation Current:

10 µA (0 W to 50 kW)
2 µA (50 kW to 500 kW)

Equivalent temperature accuracies derived from primary specifications (Ω, mV)
Table 10 Temperature, Thermocouples

26

Type

Range

Measure
Accuracies (ITS-90)

B

600 °C to 800 °C
800 °C to 1000 °C
1000 °C to 1800 °C

0.85 °C
0.68 °C
0.57 °C

C

100 °C to 550 °C
550 °C to 2300 °C

0.32 °C
0.71 °C

E

–200 °C to 0 °C
0 °C to 950 °C

0.52 °C
0.22 °C

J

–200 °C to 0 °C
0 °C to 1200 °C

0.52 °C
0.23 °C

K

–200 °C to 0 °C
0 °C to 1370 °C

0.61 °C
0.24 °C

Quick Start
Specifications

Type

Range

Measure
Accuracies (ITS-90)

L

–200 °C to 0 °C
0 °C to 900 °C

0.36 °C
0.23 °C

M

–20 °C to 0 °C
0 °C to 400 °C
400 °C to 1400 °C

0.26 °C
0.25 °C
0.22 °C

N

–200 °C to 0 °C
0 °C to 1300 °C

0.72 °C
0.28 °C

R

–20 °C to 0 °C
0 °C to 500 °C
500 °C to 1750 °C

1.09 °C
0.97 °C
0.49 °C

S

–20 °C to 0 °C
0 °C to 500 °C
500 °C to 1750 °C

1.05 °C
0.95 °C
0.56 °C

T

–200 °C to 0 °C
0 °C to 400 °C

0.60 °C
0.25 °C

U

–200 °C to 0 °C
0 °C to 400 °C

0.54 °C
0.24 °C

Resolution: 0.01 °
Note 1: Accuracies are based on internal Reference
Junction Compensation. Refer to Technical manual for
equivalent external reference accuracies.

27

1523, 1524 Thermometer Readout
Specifications
Table 11 Temperature, RTD Ranges,
and Accuracies (ITS-90)
Accuracy ± °C 4 Wire Probe
± 0.011 at –100 °C
± 0.015 at 0 °C
± 0.019 at 100 °C
± 0.023 at 200 °C
± 0.031 at 400 °C
± 0.039 at 600 °C
Resolution: 0.001°C (0.001°F)
Table 12 Temperature, Thermistor
Accuracy ± °C
± 0.002 at 0 °C
± 0.003 at 25 °C
± 0.006 at 50 °C
± 0.014 at 75 °C
± 0.030 at 100 °C
Resolution: 0.001 °C (0.001 °F)
Based on a 10 kW (at 25 °C) thermister
with a beta value of 4000 W. See
technical manual for details.

28



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Page Mode                       : UseOutlines
XMP Toolkit                     : Adobe XMP Core 4.0-c316 44.253921, Sun Oct 01 2006 17:14:39
Instance ID                     : uuid:d8098649-e8f1-417c-99f1-3a8e774140ef
Document ID                     : adobe:docid:indd:11f57bba-5453-11dd-99c8-ad4a72abbf98
Rendition Class                 : proof:pdf
Manifest Link Form              : ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream
Manifest Placed X Resolution    : 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00
Manifest Placed Y Resolution    : 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00
Manifest Placed Resolution Unit : Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches
Manifest Reference Instance ID  : uuid:ffe18eca-b74e-4b68-8372-8a1a412deab2, uuid:1bcb8852-ffbf-4c87-aef9-9dbbe6ae62f1, uuid:b0166900-01d5-4718-8d18-6deebf7d8613, uuid:d20c7022-2c00-47f2-879b-bac6878caabc, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:3778969b-27fb-467c-bd0e-cfc480131ebd, uuid:2113d185-d973-4d9d-bed4-27bdeb887bd0, uuid:0cc43329-3430-4eb8-9906-36482bfc873a, uuid:1a46a56c-2b31-4507-8bf1-a082cb2ae8f6, uuid:d7b970ac-d7fd-4773-8beb-4a5711afa7fb, uuid:1f2c990c-dada-4cc7-8def-d874e90417cf, uuid:a64c74e5-0498-4585-b7c3-528097596def, uuid:1e55cd0c-a1df-4165-93fb-afda9a64ee5a, uuid:840f32b2-7b2b-4187-b373-79de3d540153, uuid:9d2202b7-9c31-4ea7-9646-41e0c990b359, uuid:700d4ebf-31ee-4a9d-ab16-be61e175be6c, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:2113d185-d973-4d9d-bed4-27bdeb887bd0, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:0cc43329-3430-4eb8-9906-36482bfc873a, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:1a46a56c-2b31-4507-8bf1-a082cb2ae8f6, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:840f32b2-7b2b-4187-b373-79de3d540153, uuid:d20c7022-2c00-47f2-879b-bac6878caabc, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:3778969b-27fb-467c-bd0e-cfc480131ebd, uuid:2113d185-d973-4d9d-bed4-27bdeb887bd0, uuid:0cc43329-3430-4eb8-9906-36482bfc873a, uuid:1a46a56c-2b31-4507-8bf1-a082cb2ae8f6, uuid:d7b970ac-d7fd-4773-8beb-4a5711afa7fb, uuid:1f2c990c-dada-4cc7-8def-d874e90417cf, uuid:a64c74e5-0498-4585-b7c3-528097596def, uuid:1e55cd0c-a1df-4165-93fb-afda9a64ee5a, uuid:840f32b2-7b2b-4187-b373-79de3d540153, uuid:9d2202b7-9c31-4ea7-9646-41e0c990b359, uuid:700d4ebf-31ee-4a9d-ab16-be61e175be6c, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:2113d185-d973-4d9d-bed4-27bdeb887bd0, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:0cc43329-3430-4eb8-9906-36482bfc873a, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:1a46a56c-2b31-4507-8bf1-a082cb2ae8f6, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:1f2c990c-dada-4cc7-8def-d874e90417cf, uuid:edc275e9-52d6-45cd-b720-7a283b261ead, uuid:840f32b2-7b2b-4187-b373-79de3d540153
Manifest Reference Document ID  : uuid:457893501F54DD11B74F864454EC5B92, uuid:477893501F54DD11B74F864454EC5B92, uuid:4BDD89DE4754DD11B264E32A0C98CABD, uuid:153106BE3354DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:D99796903454DD11AD20F15EDE266BEE, uuid:DB9796903454DD11AD20F15EDE266BEE, uuid:DD9796903454DD11AD20F15EDE266BEE, uuid:DF9796903454DD11AD20F15EDE266BEE, uuid:E19796903454DD11AD20F15EDE266BEE, uuid:13EF100E3554DD11AD20F15EDE266BEE, uuid:1BEF100E3554DD11AD20F15EDE266BEE, uuid:1DEF100E3554DD11AD20F15EDE266BEE, uuid:17EF100E3554DD11AD20F15EDE266BEE, uuid:15EF100E3554DD11AD20F15EDE266BEE, uuid:19EF100E3554DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:DB9796903454DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:DD9796903454DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:DF9796903454DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:17EF100E3554DD11AD20F15EDE266BEE, uuid:153106BE3354DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:D99796903454DD11AD20F15EDE266BEE, uuid:DB9796903454DD11AD20F15EDE266BEE, uuid:DD9796903454DD11AD20F15EDE266BEE, uuid:DF9796903454DD11AD20F15EDE266BEE, uuid:E19796903454DD11AD20F15EDE266BEE, uuid:13EF100E3554DD11AD20F15EDE266BEE, uuid:1BEF100E3554DD11AD20F15EDE266BEE, uuid:1DEF100E3554DD11AD20F15EDE266BEE, uuid:17EF100E3554DD11AD20F15EDE266BEE, uuid:15EF100E3554DD11AD20F15EDE266BEE, uuid:19EF100E3554DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:DB9796903454DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:DD9796903454DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:DF9796903454DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:13EF100E3554DD11AD20F15EDE266BEE, uuid:173106BE3354DD11AD20F15EDE266BEE, uuid:17EF100E3554DD11AD20F15EDE266BEE
Create Date                     : 2008:09:10 15:21:13-06:00
Modify Date                     : 2008:09:10 15:22:32-06:00
Metadata Date                   : 2008:09:10 15:22:32-06:00
Creator Tool                    : Adobe InDesign CS3 (5.0.3)
Thumbnail Format                : JPEG
Thumbnail Width                 : 256
Thumbnail Height                : 256
Thumbnail Image                 : (Binary data 5515 bytes, use -b option to extract)
Format                          : application/pdf
Producer                        : Adobe PDF Library 8.0
Trapped                         : False
Page Count                      : 36
Page Layout                     : SinglePage
Creator                         : Adobe InDesign CS3 (5.0.3)
EXIF Metadata provided by EXIF.tools

Navigation menu