Making sense of the
product rating systems
Each approved intrinsically safe device is rated to ATEX,
NEC, FM or other country standards. The corresponding
rating system allows you to understand which zones, type
of protection, gas groups and temperature classes the
instrument is approved for.

ATEX Example
Fluke 707Ex is

ATEX Compliant II 2 G Ex ia IIC T4

The ATEX examination mark is required on all devices for use in
European hazardous areas.

ATEX Markings
II 2 G

The classification of zones. “II” designates the tool is approved for
all non-mining areas. “2” represents the category of the device, in
this case the device is rated for the second most hazardous areas.
“G” designates atmosphere, in this case gas, vapors and mist.

Ex

Explosion protection based on European Ex-regulations.

ia

The type of protection from explosion, in this case the energy in a
device or connector has been reduced to a safe value.

IIC

Gas Group. “IIC” rating indicates compatibility with the most
dangerous gas groups.

T4

Temperature class is the maximum temperature of a surface that
may be.

FM Example
Fluke 707Ex is
Groups A-D, T4

FM-classified N.I. Class I, Div 2,

The Factory Mutual Approved mark.

Factory Mutual Markings
N.I.

Non-incendive apparatus, internal energy is limited so a specified
atmosphere cannot be ignited by its use.

Class I

For use with gases, vapors and liquids (not dust, fibers or filings).

Div 2

Certified for use in Zone 2, explosive atmospheres not normally
present, may rarely exist for short duration.

Groups
A-D

Rated for use with explosive gasses as defined by groups A-D,
including acetylene, hydrogen, acetylene and propane.

Fluke.	 Keeping your world
	
up and running.®
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2005-2014 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 4/2014 2432675C_EN
Modification of this document is not permitted
without written permission from Fluke Corporation.

Fluke tools are designed
to keep you safe in
dangerous environments
Fluke’s intrinsically safe tools can be used to help you perform
maintenance and calibration tasks in potentially explosive and
hazardous areas. Learn about the importance of intrinsic safety
on the job and the industries that use intrinsically safe tools.
Fluke offers a wide range of reliable, accurate and intrinsically
safe tools. Including the NEW 28II Ex True-rms Multimeter
and the NEW Fluke 568 Ex Infrared Thermometer, which
compliments the line of Fluke intrinsically safe field calibrators.

Fluke 725Ex
Multifunction
Process Calibrator

Fluke 700PEx
Pressure Modules

Fluke 28II Ex
True-rms Industrial
Multimeter

Fluke 568 Ex
Infrared
Thermometer

Fluke 718Ex
Pressure
Calibrator

700G Series
Precision Pressure
Test Gauges

Fluke 707Ex
Loop Calibrator

1551/1552 “Stik”
Thermometers

Fluke Intrinsically Safe products
help to keep you safe on the job
Fluke 700PEx
Pressure Modules
“Intrinsically Safe” or I.S. is a
protection method employed
in potentially explosive atmospheres. Certified I.S. tools are
designed to prevent the release
of sufficient energy to cause
ignition of flammable material. I.S. standards apply to all
equipment that can create one
or more of a range of defined
potential explosion sources:
•	 Electrical sparks
•	 Electrical arcs
•	Flames
•	 Hot surfaces
•	 Static electricity
•	 Electromagnetic radiation
•	 Chemical reactions
•	 Mechanical impact
•	 Mechanical friction
•	 Compression ignition
•	 Acoustic energy
•	 Ionizing radiation
Industries that use or should
use I.S. tools include:
•	Petro-chemical
•	 Oil platforms and refineries
•	 Pharmaceutical
•	 Pipelines
•	 Any environment where
explosive gases or vapors
could be present

Best suited for:
•	 Use with the Fluke 725Ex Multifunctional Process Calibrator and

Fluke 718Ex Pressure Calibrator to cover the most commonly used
pressure calibration ranges from 0-25 mbar up to 0-200 bar

Features:

•	 I.S. Class I Div 1 Groups A-D T4, Ta = 0 °C to +50 °C
•	 ATEX II 1G Ex ia IIC T4 compliant
•	 Very high accuracy up to 0.025 %

1551/1552 “Stik”
Thermometers
Best suited for:
•	 Daily checks of working thermometers
•	 Custody transfer temperature determination, PET

calibration, LIG calibration, temperature transmitter
calibration and verification

Features:

•	 ATEX II 2 G Ex ib IIB T4 Gb, Ta -10 °C to 50 °C

1551A range: -50 °C to 160 °C (-58 °F to 320 °F)
1552A range: -80 °C to 300 °C (-112 °F to 572 °F)
•	 Accuracy (1 year): ± 0.05 °C (± 0.09 °F)
•	 Easy data logging available with 1552A

Legend
Motor Control Center 	
	
Flow Computer 	

Process Measurements

Tank Farm	

Pumping Station	

Process Controller

For more information on these ATEX-compliant tools designed

Fluke 725Ex Multifunction
Process Calibrator

Fluke 568 Ex Intrinsically Safe
Infrared Thermometer

Best suited for:

Best suited for:

•	 Calibrating virtually any process parameter

•	 Taking spot, differential or scanning temperature

Features:
•	 Class I Div 1 Groups B-D 171 °C
•	 ATEX II IG Ex ia IIB 171 °C Compliant
•	 Measure, source or simulate volts dc, mA, RTDs,
thermocouples, frequency and ohms
•	 2-channel simultaneous source and measure capability
for calibration of transmitters
•	 Internal loop supply to power transmitters
•	 Store frequently-used test setups for later use
•	 Pressure measurement to 200 bar and pressure switch
test using any of the 8 Fluke 700PEx pressure modules

measurements in any environment where flammable gases or vapors may be present

Features:

•	 Class I Div 1 and Div 2 or Zone 1 and 2
hazardous environments

•	 ATEX/IECEx, NEC-500/NEC-505, PCEC,
INMETRO certified

•	 Enhanced 50:1 distance to spot ratio allows user to measure
•	
•	
•	
•	
•	
•	

smaller objects from farther away
Comfortable pistol grip makes target pointing an easy job
Backlit display for poorly lit areas
Displays Min/Max/DIF/AVG measurements and saves in memory
Adjustable emissivity setting for more accurate measurements
Log up to 99 measurements
Compatible with mini-connector K-type thermocouple (KTC) probe

Fluke 718Ex Pressure Calibrator
Best suited for:

700G Series Precision
Pressure Test Gauges
Best suited for:
•	 Precision pressure measurement from

0 inH2O/20 mbar to 10,000 psi/690 bar and
absolute pressure measurement

Features:
•	 CSA; Class 1, Div 2, Groups A-D rating
•	 ATEX rating: II 3 G Ex nA IIB T6
•	 Accuracy to 0.05 % of full scale
•	 Reference class gauge accuracies to 0.04 % of reading
•	 Log up to 8,493 pressure measurements to memory

Fluke 707Ex Loop Calibrator
Best suited for:
•	 Calibration and maintenance of 4 to 20 mA control loop
Features:
•	 F.M. Class I Div 2 Groups A-D T4
•	 ATEX II 2G Ex ia IIC T4 compliant
•	 Large display and simple, quick-click push/rotary
button for easy one-handed operation
•	 Simultaneous mA and % readout for quick,
easy interpretation of readings
•	 mA accuracy of 0.015 %, superior to other loop calibrators
•	 Pushbutton with 25 % steps for fast, easy linearity checks
•	 0 and 100 % ‘span check’ for fast confirmation of zero and span

d for tough process maintenance and calibration tasks, visit www.fluke.com/ex

•	 Pressure measurements and calibration with internal sensor
at pressure ranges of 2, 7 and 20 bar

Features:
•	 Class I Div 1 Groups A-D T4 Compliant
•	 ATEX II IG Ex ia IIC T4 compliant
•	 Built-in pressure/vacuum hand pump, with vernier and
bleed valve
•	 Pressure measurement to 200 bar using any of the 8 Fluke
700PEx Pressure Modules
•	 Pressure measurement to 0.05 % of full span using internal
pressure sensor
•	 Pressure switch test function

Fluke 28II Ex True-rms
Industrial Multimeter
Best suited for:
•	 Solving problems on motor drives, in plant automation,
power distribution and electromechanical equipment

Features:
•	 Class I Div 1 Groups A-D
•	 ATEX II 2 G Ex ia IIC T4 Gb
•	 ATEX II 2 D Ex ia IIIC T130C Db
•	 ATEX I M1 Ex ia I Ma
•	 Low voltage troubleshooting in hazardous areas
•	 Built-in thermometer allows you to take temperature
readings
•	 Large digit display with bright, two-level
•	 IP67 rated, waterproof and dustproof



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.5-c021 79.154911, 2013/10/29-11:47:16
Instance ID                     : uuid:0c513ced-bfec-0b4c-8f93-312b4a9226df
Original Document ID            : adobe:docid:indd:27c71146-3b65-11db-98cb-b78f614f3e2d
Document ID                     : xmp.id:dcc3c3cc-46a8-4831-a051-355747357c7f
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:56c0b195-f96f-4e3d-bac3-943594766c73
Derived From Document ID        : adobe:docid:indd:27c71146-3b65-11db-98cb-b78f614f3e2d
Derived From Original Document ID: adobe:docid:indd:27c71146-3b65-11db-98cb-b78f614f3e2d
Derived From Rendition Class    : default
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CC (Macintosh)
History Changed                 : /
History When                    : 2014:05:03 09:27:11-07:00
Create Date                     : 2014:05:03 09:27:11-07:00
Modify Date                     : 2014:05:03 09:27:28-07:00
Metadata Date                   : 2014:05:03 09:27:28-07:00
Creator Tool                    : Adobe InDesign CC (Macintosh)
Format                          : application/pdf
Producer                        : Adobe PDF Library 11.0
Trapped                         : False
Page Count                      : 4
Creator                         : Adobe InDesign CC (Macintosh)
EXIF Metadata provided by EXIF.tools

Navigation menu