Fluke 28 Ii Users Manual EMCReport

2015-09-09

: Fluke Fluke-28-Ii-Users-Manual-805568 fluke-28-ii-users-manual-805568 fluke pdf

Open the PDF directly: View PDF PDF.
Page Count: 2

DownloadFluke Fluke-28-Ii-Users-Manual- EMCReport  Fluke-28-ii-users-manual
Open PDF In BrowserView PDF
TABLE OF CONTENTS
Report # EMC2012Feb09-0733

ELECTROMAGNETIC COMPATIBILITY
TEST CERTIFICATE
for the
Fluke 28IIEx
Digital Multimeter

The Fluke 28IIEx was tested to the following standards
at the EMC laboratories of Fluke Corporation.
6920 Seaway Blvd Everett WA 98203

EN 61326-1:2006, EN61326-2-2:2006, CISPR 11:2004
Class A Emissions and Immunity
The Fluke 28IIEx passes test requirements for equipment used for:
□ Industrial Locations
□ Controlled EM Environments
▣ Portable Equipment
▣ Non-Domestic Use (Class A)
□ Domestic Use (Class B)

Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low voltage power
supply network which supplies buildings used for domestic purposes.

Prepared by:
John Morton _________________________________________
Sr. Specialist Technician

Date: 02/09/12

Approved by:
Michael Meisner______________________________________
Test Engineer .

Date: 2/9/12

I.

Test Results

The Fluke 28IIEx Digital Multimeter was tested to the following Electromagnetic Compatibility [EMC]
requirements:

Port
Enclosure
AC mains

Frequency
MHz
30 to 230
230 to 1000
0.15 to 0.5
0.5 to 5.0
5 to 30

Adapted from CISPR 11 Table 2a
Emissions Limits for Class A Equipment
Limits

Standard

Pass/Fail

CISPR 11

Pass
Pass
N/A
N/A
N/A

40dB (uV/m) quasi peak, measured at 10 meters.
47dB (uV/m) quasi peak, measured at 10 meters.
79dB (uV/m) quasi peak, 66dB (uV/m) average.
73dB (uV/m) quasi peak, 60dB (uV/m) average.
73dB (uV/m) quasi peak, 60dB (uV/m) average.

Adapted from EN 61326-1:2006 Table A.1
Immunity test requirements for portable test and measurement equipment
Port
Phenomenon
Basic standard
Test value
Criteria
ESD
EN 61000-4-2
4 kV/8 kV contact/air
B
3 V/m
EM Field
EN 61000-4-3
A1
(80 MHz to 1 GHz)
Enclosure
3 V/m
EM Field
EN 61000-4-3
A1
(1,4 GHz to 2 GHz)
1 V/m
EM Field
EN 61000-4-3
A1
(2,0 GHz to 2,7 GHz)
1. See performance criteria defiinitions on following page.

Pass/Fail
Pass
Pass
Pass
Pass



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 4.0-c321 44.398116, Tue Aug 04 2009 14:24:39
Producer                        : Acrobat Distiller 8.3.1 (Windows)
Keywords                        : EMC template report
Company                         : Fluke Corporation
Category                        : EMC
Source Modified                 : D:20121023163058
Creator Tool                    : Acrobat PDFMaker 8.1 for Word
Modify Date                     : 2012:10:23 09:32:41-07:00
Create Date                     : 2012:10:23 09:31:59-07:00
Metadata Date                   : 2012:10:23 09:32:41-07:00
Document ID                     : uuid:30e5e94e-0112-495a-abaa-febc59761ee8
Instance ID                     : uuid:6162f266-8ded-4c30-bda3-e5fd7e4113dc
Subject                         : 3
Format                          : application/pdf
Creator                         : mmeisner
Title                           : EMCTestReport
Description                     : Fluke EMC Test Report
Headline                        : Fluke EMC Test Report
Page Count                      : 2
Page Layout                     : OneColumn
Author                          : mmeisner
EXIF Metadata provided by EXIF.tools