Fluke 28 Ii Users Manual EMCReport

2015-09-09

: Fluke Fluke-28-Ii-Users-Manual-805568 fluke-28-ii-users-manual-805568 fluke pdf

Open the PDF directly: View PDF PDF.
Page Count: 2

TABLE OF CONTENTS
Report # EMC2012Feb09-0733
ELECTROMAGNETIC COMPATIBILITY
TEST CERTIFICATE
for the
Fluke 28IIEx
Digital Multimeter
The Fluke 28IIEx was tested to the following standards
at the EMC laboratories of Fluke Corporation.
6920 Seaway Blvd Everett WA 98203
EN 61326-1:2006, EN61326-2-2:2006, CISPR 11:2004
Class A Emissions and Immunity
The Fluke 28IIEx passes test requirements for equipment used for:
Industrial Locations Controlled EM Environments Portable Equipment
Non-Domestic Use (Class A) Domestic Use (Class B)
Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low voltage power
supply network which supplies buildings used for domestic purposes.
Prepared by:
John Morton _________________________________________ Date: 02/09/12
Sr. Specialist Technician
Approved by:
Michael Meisner ______________________________________ Date: 2/9/12
Test Engineer .
I. Test Results
The Fluke 28IIEx Digital Multimeter was tested to the following Electromagnetic Compatibility [EMC]
requirements:
Adapted from CISPR 11 Table 2a
Emissions Limits for Class A Equipment
Port Frequency
MHz Limits Standard Pass/Fail
Enclosure 30 to 230 40dB (uV/m) quasi peak, measured at 10 meters. Pass
230 to 1000 47dB (uV/m) quasi peak, measured at 10 meters. Pass
AC mains 0.15 to 0.5 79dB (uV/m) quasi peak, 66dB (uV/m) average. CISPR 11 N/A
0.5 to 5.0 73dB (uV/m) quasi peak, 60dB (uV/m) average. N/A
5 to 30 73dB (uV/m) quasi peak, 60dB (uV/m) average. N/A
Adapted from EN 61326-1:2006 Table A.1
Immunity test requirements for portable test and measurement equipment
Port Phenomenon Basic standard Test value Criteria Pass/Fail
Enclosure
ESD EN 61000-4-2 4 kV/8 kV contact/air B Pass
EM Field EN 61000-4-3 3 V/m
(80 MHz to 1 GHz) A1 Pass
EM Field EN 61000-4-3 3 V/m
(1,4 GHz to 2 GHz) A1 Pass
EM Field EN 61000-4-3 1 V/m
(2,0 GHz to 2,7 GHz) A1 Pass
1. See performance criteria defiinitions on following page.