Fluke 789 Users Manual EMCReport

2015-09-09

: Fluke Fluke-789-Users-Manual-806298 fluke-789-users-manual-806298 fluke pdf

Open the PDF directly: View PDF PDF.
Page Count: 2

DownloadFluke Fluke-789-Users-Manual- EMCReport  Fluke-789-users-manual
Open PDF In BrowserView PDF
Report # EMC2008Nov17-1420

ELECTROMAGNETIC COMPATIBILITY
TEST CERTIFICATE
for the
Fluke 789
Processmeter

The Fluke 789 was tested to the following standard
at the EMC laboratories of Fluke Corporation.

EN 61326-1 :2006
Class A Emissions and Immunity
The Fluke 789 passes test requirements for equipment used for:

□ Industrial Locations

□ Controlled EM Environments

▣ Non-Domestic Use (Class A)

▣ Portable Equipment

□ Domestic Use (Class B)

Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low voltage power
supply network which supplies buildings used for domestic purposes.

Prepared by:
Michael Meisner ______________________________________
EMC Technician Specialist

17 Nov 2008
Date: _______________________

Reviewed by:
Thomas Smith _______________________________________
Test Engineer Sr.

17 Nov 2008
Date: _______________________

Approved by:
Product Evaluation Manager
Division Engineering Manager

„
†

TABLE

I.

Test Results

The Fluke 789 Processmeter was tested to the following Electromagnetic Compatibility [EMC] requirements:

Port
Enclosure
AC mains

Frequency
MHz
30 to 230
230 to 1000
0.15 to 0.5
0.5 to 5.0
5 to 30

Adapted from CISPR 11 Table 2a
Emissions Limits for Class A Equipment
Limits

Standard

Pass/Fail

CISPR 11

Pass1
Pass1
N/A
N/A
N/A

40dB (uV/m) quasi peak, measured at 10 meters.
47dB (uV/m) quasi peak, measured at 10 meters.
79dB (uV/m) quasi peak, 66dB (uV/m) average.
73dB (uV/m) quasi peak, 60dB (uV/m) average.

73dB (uV/m) quasi peak, 60dB (uV/m) average.
1. Initial Report: 789 EMC Test Report 9 Jul 2002.pdf

Portable test & measurement equipment that is not capable of operating while being charged
is tested to the following immunity requirments:
Adapted from IEC 61326-1:2005 Table A.1
Immunity test requirements for portable test and measurement equipment
Port
Phenomenon
Basic standard
Test value
Criteria
ESD
IEC 61000-4-2
4 kV/8 kV contact/air
B
3 V/m
EM Field
IEC 61000-4-3
A1
(80 MHz to 1 GHz)
Enclosure
3 V/m
EM Field
IEC 61000-4-3
A1
(1,4 GHz to 2 GHz)
1 V/m
EM Field
IEC 61000-4-3
A1
(2,0 GHz to 2,7 GHz)
1. See performance criteria defiinitions on following page.
2. Initial Report: 789 EMC Test Report 9 Jul 2002.pdf

Pass/Fail
Pass2
Pass2
Pass
Pass



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 4.0-c321 44.398116, Tue Aug 04 2009 14:24:39
Producer                        : Acrobat Distiller 8.3.1 (Windows)
Keywords                        : EMC template report
Company                         : Fluke Corporation
Category                        : EMC
Source Modified                 : D:20130107211305
Creator Tool                    : Acrobat PDFMaker 8.1 for Word
Modify Date                     : 2013:01:07 13:14:01-08:00
Create Date                     : 2013:01:07 13:13:35-08:00
Metadata Date                   : 2013:01:07 13:14:01-08:00
Document ID                     : uuid:c67d458b-2404-4be1-9871-c8575e764374
Instance ID                     : uuid:cd84a419-688f-49f8-a8e0-4980f84e3211
Subject                         : 2
Format                          : application/pdf
Creator                         : mmeisner
Title                           : EMCTestReport
Description                     : Fluke EMC Test Report
Headline                        : Fluke EMC Test Report
Page Count                      : 2
Page Layout                     : OneColumn
Author                          : mmeisner
EXIF Metadata provided by EXIF.tools

Navigation menu