Manual Supplement
Manual Title:
87 V Ex Users
Part Number: 2518115
Print Date:
December 2005
Revision/Date:

Supplement Issue:
Issue Date:
Page Count:

6
6/11
3

This supplement contains information necessary to ensure the accuracy of the
above manual.

© 2006-2011 Fluke Corporation. All rights reserved. Printed in the U.S.A.

x

Manual Supplement

87V Ex Users

Change#1
On page 55, under, Maximum output values at the voltage input change:
From: Co = 2.52 F
To:
Co = 2.52 μF

Under, Maximum output values at the current-measuring inputs change:
From:
To:

Co = 1000 F
Co = 1000 μF

Change #2, 39785
On page 52, under Electrical Characteristics of the Terminals, replace the Diode row with the
following:
G

1000 V rms

< 7.9 V dc

3.000 V dc

1

1.0 mA typical

1/07

87V Ex Users

Manual Supplement

Change #3, 44008
On page 45, under General Specifications,
Change: Relative Humidity: 0 % to 80 % (0 °C to 35 °C); 0 % to 70 % (35 °C to 50 °C)
To:
Relative Humidity: 0 % to 80 % (0 °C to 35 °C)
On page 47, under DC Voltage, Resistance, and Conductance Function, add the following
note:
[3] Maximum of 60 % relative humidity or a maximum of 28 °C.

Under Range, change the following:
From:

6.000 MΩ
50.00 MΩ

To:

6.000 MΩ[3]
50.00 MΩ[3]

In the 60.00 nS range change the Accuracy,

1/07

From:

± (1.0 % + 10)[1,2]

To:

unspecified

2

Manual Supplement

87V Ex Users

Change #4, W2041055
On page 6, under EX-Certification Data,
Ex-designation: ( II 2 G EEx ia IIC T4
Change:
To :
Ex-designation: ( II 2 G Ex ia IIC T4

Change #5, 55377
On page 7, in Table 1 remove the following:

®

Inspected and licensed by TÜV Product Services.

On page 45, under General Specifications, replace Safety with:
Safety: Meets 1000V CAT III/600V CAT IV per ANSI/ISA-61010-1 (82.02.01):2004, UL 61010-1:2004, CAN/CSA
C22.2 No. 61010-1-04, and EN/IEC 61010-1:2001. The product is designed for industrial applications in areas with
an increased risk of explosion in zones 2 and 1, in accordance with directive 1999/92/EC (ATEX 137).

Change #6, 58805
On page 45, under the General Specifications, replace Electromagnetic Compatibility with:
Electromagnetic Compatibility: In an RF field of 3 V/m total accuracy = specified accuracy + 20 counts
Except: 600 μA dc range total accuracy = specified accuracy + 60 counts.
Temperature not specified.
Some mobile devices that transmit RF energy may transmit levels that far exceed 3 V/m and may damage sensitive electronic
circuits. To insure the best performance, do not allow a device which is transmitting RF energy in excess of 3 V/m to be within
30 cm of the meter while in use.

3

6/11



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
XMP Toolkit                     : 3.1-702
Producer                        : Acrobat Distiller 7.0.5 (Windows)
Creator Tool                    : PScript5.dll Version 5.2.2
Modify Date                     : 2011:06:23 08:54:40-07:00
Create Date                     : 2011:06:22 13:59:55-07:00
Metadata Date                   : 2011:06:23 08:54:40-07:00
Format                          : application/pdf
Title                           : Microsoft Word - MS6.doc
Creator                         : otto
Document ID                     : uuid:b1c2137d-10c3-4c7b-a59d-54616d829c58
Instance ID                     : uuid:3a8bbb58-27a2-49d5-a2ae-03f5e73e9da6
Page Count                      : 4
Author                          : otto
EXIF Metadata provided by EXIF.tools

Navigation menu