Manual Supplement
© 2006-2011 Fluke Corporation. All rights reserved. Printed in the U.S.A. x
Manual Title: 87 V Ex Users Supplement Issue: 6
Part Number: 2518115 Issue Date: 6/11
Print Date: December 2005 Page Count: 3
Revision/Date:
This supplement contains information necessary to ensure the accuracy of the
above manual.
Manual Supplement 87V Ex Users
1 1/07
Change#1
On page 55, under, Maximum output values at the voltage input change:
From: Co = 2.52 F
To: Co = 2.52 μF
Under, Maximum output values at the current-measuring inputs change:
From: Co = 1000 F
To: Co = 1000 μF
Change #2, 39785
On page 52, under Electrical Characteristics of the Terminals, replace the Diode row with the
following:
G 1000 V rms < 7.9 V dc 3.000 V dc 1.0 mA typical
87V Ex Users Manual Supplement
1/07 2
Change #3, 44008
On page 45, under General Specifications,
Change: Relative Humidity: 0 % to 80 % (0 °C to 35 °C); 0 % to 70 % (35 °C to 50 °C)
To: Relative Humidity: 0 % to 80 % (0 °C to 35 °C)
On page 47, under DC Voltage, Resistance, and Conductance Function, add the following
note:
[3] Maximum of 60 % relative humidity or a maximum of 28 °C.
Under Range, change the following:
From:
6.000 MΩ
50.00 MΩ
To: 6.000 MΩ[3]
50.00 MΩ[3]
In the 60.00 nS range change the Accuracy,
From: ± (1.0 % + 10)[1,2]
To: unspecified
Manual Supplement 87V Ex Users
3 6/11
Change #4, W2041055
On page 6, under EX-Certification Data,
Change: Ex-designation: ( II 2 G EEx ia IIC T4
To : Ex-designation: ( II 2 G Ex ia IIC T4
Change #5, 55377
On page 7, in Table 1 remove the following:
® Inspected and licensed by TÜV Product Services.
On page 45, under General Specifications, replace Safety with:
Safety: Meets 1000V CAT III/600V CAT IV per ANSI/ISA-61010-1 (82.02.01):2004, UL 61010-1:2004, CAN/CSA
C22.2 No. 61010-1-04, and EN/IEC 61010-1:2001. The product is designed for industrial applications in areas with
an increased risk of explosion in zones 2 and 1, in accordance with directive 1999/92/EC (ATEX 137).
Change #6, 58805
On page 45, under the General Specifications, replace Electromagnetic Compatibility with:
Electromagnetic Compatibility: In an RF field of 3 V/m total accuracy = specified accuracy + 20 counts
Except: 600 μA dc range total accuracy = specified accuracy + 60 counts.
Temperature not specified.
Some mobile devices that transmit RF energy may transmit levels that far exceed 3 V/m and may damage sensitive electronic
circuits. To insure the best performance, do not allow a device which is transmitting RF energy in excess of 3 V/m to be within
30 cm of the meter while in use.